FasTest MediMate Connector Body User Manual

REPLACEMENT INSTRUCTIONS
FasTest, Inc. Product Warranty
FasTest, Inc. warrants its products against defects the date of the sale by FasTest, Inc. This warranty is void if the product is misused, tampered with or used in a manner that is not in accordance with FasTest, Inc. recommendations and/or instructions. FasTest, Inc. is not liable for consequential or other damages including, but not limited to, loss, damage, personal injury, or any other expense directly or indirectly arising from the use of or inability to use its products either separately or in combination with other products. ALL OTHER WARRANTIES EXPRESSED OR IMPLIED, WHETHER ORAL OR WRITTEN, INCLUDING BUT NOT LIMITED TO WARRANTIES OR MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE ARE EXPRESSLY EXCLUDED. Remedy under this warranty is limited to replacement of the product or an account credit in the amount of the original selling price, at the option on FasTest, Inc. All allegedly defective products must be returned prepaid transportation to FasTest, Inc. along with information describing the products performance, unless disposition in the field is authorized in writing by FasTest, Inc.
of workmanship and/or material for 12 months from
G870041 MediMate® Oxygen Connector Body Replacement (G870041-1) & (G870041A-1)
Description:
Please thoroughly read the instructions prior to rebuilding the MediMate® Connector. If you do not understand instructions, or if components are missing, call FasTast before rebuilding connector.
Remove connector from service. Select a clean
nvironment to transfer the components. A laboratory or
e clean office is acceptable. Avoid any location where the connector could come in contact with external contaminants. Wear rubber gloves or wash hands thoroughly before handling.
WARNING: Failure to take proper precautions to assure
the cleanliness of the connector and components can be hazardous.
Replacement of MediMate® Connector Body
Roseville, MN
®
Ph: 1-800-444-2373 Fax: 651-645-7390
www.fastestinc.com
REPLACEMENT INSTRUCTIONS
Core Sub-Assembly Removal and Replacement:
CLEVIS PIN
HANDLE
RETAINING RING
1. To remove the core sub-assembly from the old body, place the body of the MediMate connector in a clamp or vise.
2. Utilizing flats on shaft, unscrew shaft from the body. Use care so not to scratch or damage the connector components. The core sub-assembly should easily slide out of the body. If the components do not remove easily, gently remove parts by hand. Do not use any sharp instruments.
3. Make sure the core sub-assembly is dry and clean. If necessary, blow-off the assembly with clean dry air.
4. Apply a drop of LocTite 242 to the shaft threads. Use care to assure that the LocTite does not contaminate any other components.
ACTIVATION PISTON
AND SPRING
BODY
CORE SUB
ASSEMBLY
SHAFT THREADS
SHAFT
SHAFT FLATS
SHAFT THREADS
5. Place the core sub-assembly into the new body Do not over tighten.
and tighten down with wrench until snug.
Handle / Piston Removal and Replacement:
1. Remove the retaining ring from the clevis pin.
2. Slide the clevis pin out through the handle and body.
HANDLE
3. The activation piston will now be free and easily removed from the old body. Remove the piston and the piston spring and transfer to the new body.
4. Orientate the handle into the new body and slide the clevis pin through the body and handle.
5. Replace the retaining ring into the clevis pin.
RETAINING RING
CLEVIS PIN
ACTIVATION PISTON AND SPRING
Roseville, MN
®
Ph: 1-800-444-2373 Fax: 651-645-7390
www.fastestinc.com