Extech Extech 407026 Manual

User's Guide

Heavy Duty Meter Model 407026
99 Washington Street Melrose, MA 02176
Phone 781-665-1400 Toll Free 1-800-517-8431
Visit us at www.TestEquipmentDepot.com
Back to the Extech 407026 Product Page
Introduction
Congratulations on your purchase of the Extech Heavy Duty Light Meter. This light meter offers selectable lighting types, data record/recall, relative display mode, and PC interface. This professional meter, with proper care, will provide years of safe reliable service.
Specifications
General Specifications
Circuit Custom one-chip LSI microprocessor circuit Display 3-1/2 digit (2000 count) LCD display with contrast adjustment Measurement ranges LUX: 0 to 50,000 LUX (3 range); Fc: 0 to 5000 Fc (3 range);
Relative mode: 0 to 1999% Data Hold Freezes displayed reading Lighting Types Sodium, Daylight/Tungsten, Fluorescent and Mercury Sensor Structure Cosine/color corrected photo-diode meets C.I.E. Memory Store/Recall Records/Recalls Max/Min/Avg readings Sample rate 0.4 seconds (approx.) per reading Zero Adjust Push-button procedure Auto Power Off After approx. 10 minutes Data Output RS-232 PC serial interface (optional software/cable PN 407001) Operating conditions 32 °F to 122 °F (0 °C to 50 °C); <80% RH Power Supply 9V battery Power consumption Approx. 5 mA DC. (approx. 200 hr battery life) Weight 0.71 lbs. (320 g) Dimensions Instrument: 7.1 x 2.8 x1.3" (180 x 72 x 32 mm)
Sensor: 3.3 x 2.2 x 0.7" (85 x 55 x 17.5 mm)
Range Specifications
Measurement
LUX 2,000 LUX 0-1,999 LUX 1 LUX
Foot Candles 200 Fc 0-186.0 Fc 0.1 Fc 2,000 Fc 167-1,860 Fc 1 Fc 5,000 Fc 1,670-5,000 Fc 10 Fc Relative mode 0-1999% 1%
Range Display Resolution Accuracy
± (4% + 2 digits)
20,000 LUX 1,800-19,990 LUX 10 LUX 50,000 LUX 18,000-50,000 LUX 100 LUX
of full scale
Note: The accuracy specification above applies to calibration performed using a precision standard incandescent tungsten light source of 2856
Test Equipment Depot, 99 Washington Street, Melrose, MA 02176
781.665.1400 | 800.517.8431 | Fax: 781.665.0780
3
O
K with meter on the tungsten setting.
Model 407026 Version 2.3 January 2004
Meter Description
1 LCD Display
2 Keypad
3 Light sensor
4 Battery compartment (rear)
5 Sensor input
6 RS-232 PC Interface jack
7 Protective Holster
8 LCD Contrast Adjust
6 7
4
5
1
8
2
3
4
Model 407026 Version 2.3 January 2004
Operation
Meter Power and Automatic Power Off
1. Press the POWER button to turn the meter on. If the display does not switch on, check the 9V battery.
2. Press the POWER button again to turn the meter off.
3. The meter has an automatic power off feature that conserves battery energy. After 10 minutes the meter automatically shuts off. To defeat this feature, press the RECORD button to put the meter in the recording mode.
Display ZERO Calibration
Please perform a zero calibration before each use. This will ensure the highest accuracy.
1. Place the sensor cover over the light sensor, effectively blocking all light to the sensor.
2. Select the 2,000 LUX range using the RANGE switch.
3. Press the ZERO button. The reading should be zero.
4. Remove the sensor cover from the light sensor and proceed.
Selecting the unit of measure
Press the LUX/Fc button to select the desired unit of measure. The display icons will reflect the current setting.
Selecting a light source
Select the type of lighting to be measured by pressing the LIGHT SOURCE button. The display will indicate the lighting type icon (see icon list below).
L = Tungsten/Daylight (use this setting when calibrating with a Tungsten lamp) F = Fluorescent S = Sodium C = Mercury
Note: For Halogen and Metal Halide light use the Tungsten setting
Taking Measurements
Hold the light sensor in the area where light is to be collected. The light being measured must encompass the entire surface of the light sensor dome (pin point lighting such as with LED light cannot be measured).
The display will indicate the light intensity value in LUX or Foot candles (Fc). Note that since the main display area is limited to a reading of ‘1999’, the right-most digit in the 20,000 LUX and 5,000 Fc ranges appears on the lower LCD line. In the 50,000 LUX range, the last
two digits appear on the lower LCD line
It may be necessary to adjust the display contrast due to a change in viewing angle or voltage drift. Use the LCD Contrast adjustment located on the right side of the meter to set the preferred contrast.
Display Range Selection
Take a measurement starting at the highest range and then work down using the RANGE switch. If the display shows dashes along the top of the reading area ( ), the input exceeds the maximum value for the selected range; select a higher range. If the display shows dashes along the bottom of the reading area ( ), the input is too low; select a lower range.
5
Model 407026 Version 2.3 January 2004
Data Hold
While measuring, press the HOLD button to freeze the reading. The LCD will display D.H. indicating that the Data Hold function is engaged. Press HOLD again to return to normal operation.
Relative % Mode
In the RELATIVE mode of operation, the meter displays light level as a comparison to a reference value. The reference value is stored by pressing the ‘%’ button while the desired light level reference is on the display. When the button is pressed, ‘100%’ will be displayed and the ‘%’ icon will appear.
After the ‘%’ button is pressed, light measurements will be displayed as a percentage of the reference value. For example, if the light level reference is 1000 LUX, a display of 500 LUX will be represented as 50%. Further, a light level of 250 LUX will display as 25%. Use the equation below:
Display % = --------------------------------------------------------------------------- X 100
Press the ‘%’ button again to return to normal operation.
Maximum (MAX), Minimum (MIN), and Average (AVG) readings
The meter can keep track of the highest, lowest, and average readings for later recall.
1. Press the RECORD button to start tracking the MAX, MIN, and AVG readings. The
REC display will switch on.
2. When desired, press the RECALL button.
3. The MAX display icon appears next to the REC display icon. The reading shown is the
highest reading encountered since the RECORD button was pressed.
4. Press RECALL twice more to view the MIN and AVG readings.
5. Press the RECORD button to return to normal operation.
RS-232 PC Interface
This meter is equipped with a RS-232 serial data port. This interface is for use with the Extech Data Acquisition Software (Part number 407001) and enables the user to capture, store and display readings on a PC. Contact Extech or refer to the 407001 user manual for details on the PC interface.
Battery Replacement
When the low battery indicator appears (LBT), replace the battery as soon as possible. Reliable readings can be obtained for several hours after the first appearance of the low battery indication. To replace the battery:
1. Remover the meter's rubber protective holster.
2. The battery compartment is located on the lower back of the meter. Pry the battery compartment cover off using a small coin or screwdriver and remove the battery.
3. Replace the 9V battery and reinstall the cover.
4. Ensure that the battery cover is secured after replacing the battery.
Stored reference light measurement
Light measurement
99 Washington Street Melrose, MA 02176
Phone 781-665-1400 Toll Free 1-800-517-8431
Visit us at www.TestEquipmentDepot.com
Back to the Extech 407026 Product Page
Typical Light Levels
Lux Foot
Factories Home 20-75 2-7 Emergency Stairs,
75-150 7-15 Exit/Entrance Passages 150-200 15-20 Recreational Activities 150-300 15-30 Packing Work 200-300 20-30 Drawing Room, Table 300-750 30-75 Visual Work: Production Line 300-500 30-50 Makeup 750-1,500 75-150 Typesetting: Inspection Work 500-1,500 50-150 Reading, Study 1,500-
3,000 Office Restaurant 75-100 7-10 Indoor Emergency Stairs 75-150 7-15 Corridor Stairs 100-200 10-20 Corridor Stairs 150-300 15-30 Entrance, Wash Room 200-750 20-75 Conference, Reception Room 300-750 30-75 Cooking\Dinning Room 750-1,500 75-150 Clerical Work 750-1,500 75-150 Show Window 1,500-
2,000 Store Hospital 75-150 7-15 Indoors 30-75 3-7 Emergency Stairs 150-200 15-20 Corridor/Stairs 75-100 7-10 Stairs 200-300 20-30 Reception 100-150 10-15 Sick Room,
300-500 30-50 Display Stand 150-200 15-20 Waiting Room 500-750 50-75 Elevator 200-750 20-75 Medical Exam Room 750-1,500 75-150 Show Window, Packing Table 750-1,500 75-150 Operating Room 1,500-
3,000
Candles
150-300 Electronic Assembly, Drafting 1,000-2,000 100-200 Sewing
150-2000 Typing, Drafting
150-300 Storefront, Show Window 5,000-
Lux Foot
Candles
100-150 10-15 Washing
Warehouse
500-1000 Eye Inspection
10,000
Warehouse
Common Conversion Factors
Illuminance (Visible Flux Density)
Luminance (Visible Flux Density per Solid Angle)
1 lm/m2 =
1 lm/m2/sr = 1 candela/m2 Luminous Intensity (Visible Flux per Solid Angle) 1 lm/sr = 1 candella Luminous Flux ( Visible Flux) 1 lumen (lm) =
1 lux (lx) 10-4 lm/cm2 10-4 phot (ph)
9.290 x 10-2 lm/ft2
9.290 x 10-2 foot-candles
1.464 x 10-3 watts @ 555 nm