Extech EC600 Operating Manual

User's Guide
Digital Conductivity / pH Meter
Model EC600
Introduction
Congratulations on your purchase of the Extech Conductivity / pH meter. For best
results, please read the entire manual before use.
The EC600 meter measures Conductivity, pH, and temperature parameters. Conductivity measurements also yield TDS (total dissolved solids), Salinity, and Resistivity readings.
The built-in microprocessor provides automatic calibration, automatic temperature compensation, data storage, and self-diagnostics. The meter can recognize up to 13 types of pH standard buffer solutions and 8 types of Conductivity standards.
The meter’s digital filter improves measurement speed and accuracy.
Two special modes (for distilled water and distilled water mixed with ammonia) are provided.
The meter offers a backlit LCD display and is dust-proof and water-proof, meeting the IP57 rating
This meter is shipped fully tested and calibrated and, with proper use, will provide years of reliable service.
Supplied Equipment List
EC6000 Meter
Conductivity and pH Electrodes
Standard pH buffer solutions (4.00pH, 7.00pH, and 10.01pH) / 50ml
Standard Conductivity solution (1413µS/cm) / 50ml
Screwdriver (for removing battery compartment)
Batteries (2 x ’AA’ 1.5V)
User’s Guide
Carrying Case
EC600 V1.0 7/09
Meter Description
1. LCD Display (detailed in section below)
2. Rubber protective jacket (remove to access the rear battery compartment)
3. Keypad (detailed in section below)
4. Electrode connection jack
5. Electrode plug
6. Electrode cable
Display Description
1. Measurement type icon
2. Measured reading
3. 888 (Stored data serial number), M+ (measurement to be stored icon), RM (reading to be recalled icon)
4. Units of measure
5. Temperature reading
6. Stable measurement icon
7. Calibration icons
8. Low battery icon
EC600 V1.0 7/09
Keypad Description
The meter has five (5) front panel function keys:
ON/OFF Power Key
Calibration Key:
a) When taking a measurement, press this key to enter the calibration mode b) When in the programming mode, press this key to change the displayed
number or the ON/OFF state
Function Key
a) Press momentarily (< 1.5 seconds) to select the desired measurement mode b) Press and hold for at least 2 seconds to enter the P1 programming parameter.
Subsequent presses will scroll through the remaining parameters
Backlight and ENTER Key
a) In the measurement mode, momentary presses toggle the backlight on and off b) When in the programming or calibration mode, press to ENTER data c) When in the pH mode, press and hold to change the resolution in the range of
0.01 to 0.1pH
d) When in the Conductivity mode, press and hold to scroll through the TDS,
Salinity, Resistivity, and Conductivity modes
MEMORY STORE and RECALL Key
a) In the measurement mode, momentary presses store the displayed readings.
Press and hold for at least 2 seconds to recall a saved reading
b) In programming mode, press to change the displayed number or ON/OFF state
Store, Recall, and Clear Datalogger Memory
STORE Readings
The meter can store up to 100 pH, 100 mV, and 100 Conductivity readings for a total of 300 data points. To store a reading, wait until the reading stabilizes (the smiling face icon
appears when the reading stabilizes). Press the M+/RM key momentarily to store a
reading. The M+ icon appears and the data point serial number increments.
RECALL Readings
In the measurement mode, press the M+/RM key to recall the most recently stored reading. RM and the data point serial number for the displayed reading will appear on the LCD. The measurement information will appear on the lower right hand side of the LCD. Use the CAL or M+/RM buttons to scroll the remaining stored readings. Press the ENTER key to return to the normal measurement mode.
CLEAR Readings
From the RECALL mode, press and hold the ENTER key for at least five (5) seconds. The LCD display will show ‘CLR’ indicating that all of the stored readings have been erased. The unit will automatically return to the normal measurement mode after approximately 2 seconds.
EC600 V1.0 7/09
pH Measurement Mode
Preparation for Measurement
1. Turn the meter ON using the power key
2. Use the MODE key momentarily to switch to the pH mode
3. Unscrew the protective cap on the probe jack located on the bottom of the meter (store the protective cap in the carrying case for later use)
4. Carefully connect the pH probe to the meter’s probe jack. The probe can only be inserted in one orientation. Once it is firmly connected, screw the probe collar onto the meter to secure the probe
3-Point Calibration (7.00pH, 4.00pH and 10.01pH)
1. Press the CAL key to enter the Calibration mode. The meter’s display will show a blinking ‘C1’
2. Rinse the probe in distilled water, allow it to air dry, and submerge it into a pH 7.00 buffer solution. Stir the solution briefly and allow it to stay in the buffer solution until a stable reading is reached
3. Press CAL again and the display will show a blinking ‘7.00’
4. This portion of the calibration procedure is complete when the display stops blinking and shows the ‘C2’ icon. The unit will automatically switch to the second point of the calibration
5. Rinse the probe in distilled water again, allow it to dry, and submerge it into a pH 4.00 buffer solution. Stir the solution briefly and allow it to stay in the buffer solution until a stable reading is reached
6. Press CAL again and the display will show a blinking ‘4.00’
7. This portion of the calibration procedure is complete when the display stops blinking and shows the ‘C3’ icon. The unit will automatically switch to the third point of the calibration
8. Rinse the probe in distilled water again, allow it to dry, and submerge it into a pH 10.01 buffer solution. Stir the solution briefly and allow it to stay in the buffer solution until a stable reading is reached
9. Press CAL again and the display will show a blinking ‘10.01’
10. After the display stabilizes the 3-point calibration icon will appear
1-Point and 2-Point Calibration
Note that the user can calibrate one or two points only, if the expected measurement is known.
For example, if the expeced pH is 4pH, it is acceptable to perform only a 1-Point Calibration (4pH). If the expected measurement is between 4.00pH and 7.00pH, the user can perform a 2-Point Calibration (4.00 and 7.00pH), and so on.
For a 4pH calibration, only the circled L will appear on the LCD. For a 7.00pH, only the circled M will appear on the LCD. For a 10.01pH Calibration, only the circled H will appear (Low, Medium, and High).
Note: Use the ENTER key to exit the Calibration mode and return to the normal measurement mode when performing 1 or 2 point calibrations.
For all other applications, a 3-Point Calibration is recommended. Always perform a 3-Point Calibration on new probes and probes that have been in use for long periods. This maximizes measurement slope linearity.
EC600 V1.0 7/09
Testing the pH of a Sample
1. Perform the pH Calibratoin as described above
2. Rinse and dry the pH Probe and submerge it in a sample liquid
3. Stir the solution briefly with the probe and allow it to stand until the display stabilizes
4. Note that the closer the temperature of the sample solution to the calibration solution, the more accurate readings
Programming pH Parameters
The Table below shows the available programming menu items P1 ~ P7. Each parameter is explained in detail in the subsequent sections.
Parameters Code Selections
P1 pH buffer solution series selection
Distilled water pH temperature
P2
compensation setting (see notes
below this Table)
Distilled water with Ammonia pH
P3
temperature compensation setting
(see Notes below this Table)
P4 Temperature unit setting /
P5 Back light display time setting 0-1-3-6min
P6 Auto power off setting 0-10-20min
P7 Restore to default factory settings OFF / ON
Notes on P2 and P3 Parameters: Measurements of distilled water and distilled water
mixed with ammonia affect the temperature compensation and the slope linearity of the pH probe. Such measurements are sometimes used in the electrical power and petrochemical industries. Set these parameters to ON only if necessary, otherwise leave these parameters in the OFF state.
OFF / ON
OFF / ON
USA (Europe & U.S.A)
NIS (NIST)
CH (China)
EC600 V1.0 7/09