Extech EA15 Operating Manual

User Guide
Thermocouple Datalogging Thermometer
Seven (7) Thermocouple input types K, J, T, E, R, S, N
Dual thermocouple Input with PC Interface
Model EA15
Introduction
Congratulations on your purchase of the Extech Datalogging Thermometer. This device offers dual thermocouple inputs with manual and automatic datalogging capabilities. Careful use of this meter will provide years of reliable service.
Specifications
Measurement Ranges K - Type : -200.0oF to +1999.9oF (-150.0oC to +1370.0oC)
J - Type : -200.0
T - Type : -200.0
E - Type : -200.0
R - Type : +32
S - Type : +32
N - Type : -200.0
Resolution J, K, T, E, N Type: 0.1oF/oC/K
R, S Type: 1.0
Accuracy
J, K, T, E, N Type: ±(0.05% reading + 1.5oF (0.75°C)
below -148°F (-100°C): add 0.15% of reading for J, K, E and N;
0.45% of reading for T; ±0.05 of reading +2°F(1°C) for R, S
Temperature Coefficient 0.01% of reading +0.06°F per °F (+0.03 °C) outside the specified
+64°F to 82°F (+18°C to 28°C) range
Below -148°F (-100°C): add 0.04% of reading for J, K, E and N
type; and 0.08% of reading for T type
Note: Temperature accuracy does not include the accuracy of the probe.
Note: The temperature scale is based on the international temperature scale of 1990 (ITS90).
Display Dual Display Multi-function LCD with Backlight
Input Protection 60VDC; 24VAC rms
Measurement Rate 1 time per 1.5 second
Over range indication "OL" appears on the LCD
Open input indication "OL" appears on the LCD
Low battery indication “BT” appears on the LCD
Manual Memory Capacity 98 sets
Datalogging Memory Capacity 8800 sets
Power supply 6 ‘AAA’ Batteries
Battery life Approximately 200 hours with alkaline batteries
Operating Temperature 32 to 122
Operating Humidity < 80% RH
Storage Temperature 14 to 140
Storage Humidity 10 to 80% RH
Dimensions 5.91 x 2.8 x 1.4" (150 x 72 x 35mm)
Weight Approx. 8.29 oz. (235g) with batteries
o
F to +1994.0oF (-150.0oC to +1090.0oC)
o
F to +752.0oF (-150.0oC to +400.0oC)
o
F to +1598.0oF (-150.0oC to +870.0oC)
o
F to +1999.9oF (0.0oC to +1767.0oC)
o
F to +1999.9oF (0.0oC to +1767.0oC)
o
F to +1999.9oF (-150.0oC to +1300.0oC)
oF/o
C/K (0.1° for reference only)
o
F (0 to 50oC)
o
F (-10 to 60oC)
2
Model EA15 Version 1.5 9/03
Meter Description
1. RS-232 and thermocouple input jacks
2. LCD display
3. Power, Data Hold, and Backlight buttons
4. Max/Min, display select, and units buttons
5. MEM, input TYPE, and READ buttons
6. SET button
7. Battery compartment (rear)
8. Up/Down scroll button
9. Enter Button
Operation
Connecting thermocouples
1. This meter accepts two thermocouples with spade plugs (sub-miniature type with one spade wider than the other).
2. Plug the thermocouple(s) into the meter's T1 and/or T2 thermocouple input jack(s).
3. Always check LCD display for the proper thermocouple type selection..
Turn Power ON
Press the button to turn power on. The meter will perform a short self-test. If a thermocouple is not plugged into the selected input or if the thermocouple is "open", the display will show "- - - - -". The default thermocouple type, units and display is determined from the parameters selected when the meter was last used.
Selecting thermocouple type
Press the TYPE button to select the type of thermocouple being used.
Selecting °C, °F or °K units of measure
Press the C/F/K button to select the unit of measure.
Selecting T1, T2, and T1-T2 Displays
Press the T1 T2 T1-T2 button to step through and select from the following displays: a. T1 (thermocouple 1) primary display and T2 (thermocouple 2) lower display. b. T2 primary display and T1 lower display c. (T1-T2) (difference between T1 and T2) primary display and T1 lower display d. (T1-T2)primary display and T2 lower display
Notes:
To avoid electrical shock or personal injury, do not apply more than 20Vrms, between
the thermocouples, or between any thermocouple and earth ground.
If voltage on the measurement surface results in potentials of more than 1V,
measurement errors may occur.
If a potential exists between the thermocouples, use electrically insulated
thermocouples.
1
2
3
4
5
6
T1
T2
INTERVAL
T1
min: se c hour:min
T2
TYPE
KJTERSN
MX AVG-MN
7 8 9
C F K
3
Model EA15 Version 1.5 9/03
Loading...
+ 5 hidden pages