Introduction
Congratulations on your purchase of the Extech Milliohm Meter Model 380460 (110V) or Model 380462 (220V). The milliohm meters are enclosed in rugged carrying cases for durability. Measurements are made with Kelvin test leads (supplied) for optimal accuracy in five (5) selectable ranges. Careful use of this meter will provide years of reliable service.
Specifications
General Specifications
Circuit Custom one-chip LSI microprocessor circuit Display 0.7" (18 mm) LCD with 2000 display counts Measurement terminals 4-Terminal Kelvin type Measurement ranges Zero Adjust ±50 count adjustment Sampling Time Approximately 0.4 seconds Over input indication Indication of "1 - - -" Operating Temperature 32 °F to 122 °F (0 °C to 50 °C) Operating Humidity 80% Relative Humidity max. Power Supply 110V (380460) or 220V (380462) ±15%, 50/60Hz Power Consumption Less than 2 VA Weight 1.32 lbs (600g) Dimensions 6.3 x 4.72 x 3.35" (160 x 120 x 85 mm) with cover Accessories Power Cable and 4-wire Kelvin clip leads
Range Specifications
Five (5) ranges (200m, 2, 20, 200, 2000 ohms)
Range Resolution
200 m 0.1mΩ
2 1m
20 10m
200 0.1
2000 1
Note: Specifications tested using RF Field Strength <3V/m and frequency <30MHz
Test
Current
100mA ± 0.75% + 4 digits 3.8V
10mA ± 0.75% + 2 digits 3.4V 10mA ± 0.75% + 2 digits 3.4V
1mA ± 0.75% + 2 digits 3.2V 1mA ± 0.75% + 2 digits 3.2V
2
Accuracy
Model 380460 and 380462 Version 2.1 March 2005
Open Circuit
Voltage
Loading...