Congratulations on your purchase of the Extech 380395 (120V) or 380396 (220V) High
Voltage Insulation Tester. This meter provides four Insulation Resistance test ranges
plus Continuity, AC/DC Voltage, Polarization Index and Dielectric Absorption Ratio
measurements. This meter is shipped fully tested and calibrated and, with proper care
and use, will provide years of reliable service.
Features
• Auto ranging 6000 count LCD display with bargraph
• Selectable LCD backlight level function facilitates working in dimly illuminated areas.
• MAX/MIN, PEAK, Relative, and Data hold functions for DC/AC voltage measurements.
• Designed to the following safety standards:
IEC 61010-1 (CAT IV 600V Pollution degree 2)
IEC 61010-031 (Requirements for hand-held probes)
• Insulation test range: 0.1M to 60G.
• Insulation test voltages: 500V, 1000V, 1500V, and 5000V.
• AC/DC voltage: 0.5 V to 600 V.
• 200mA Continuity.
• Resistance: 0.1 to 6k.
• Auto-discharge function & voltage output warning function.
• Electric charges stored in capacitive circuits are automatically discharged after
measurements. Discharge status can be checked using the real time voltage bargraph.
• LIVE circuit warning symbols plus audible warning.
• Live circuit detection prevents insulation test if voltage > 30 V is detected.
• Auto-power off function and battery check.
• Programmable Test Timer measurement function.
• Polarization Index measurement (PI).
• Dielectric Absorption Ratio measurement (DAR).
• Power source: Eight (8) 1.5V ‘C’ Cell Batteries.
2
380395/380396 V1.1 05/10
Safety
International Safety Symbols
This symbol, adjacent to another symbol or terminal, indicates the user must refer
to the manual for further information.
This symbol, adjacent to a terminal, indicates that, under normal use, hazardous
voltages may be present.
Danger: identifies conditions and actions that pose hazard(s) to the user.
Warning: alerts the user to avoid electric shock.
Caution: identifies conditions and actions that may damage the Insulation
Resistance tester.
Operating Caution: identifies conditions where the user needs to take extra care
during the operation of the Insulation resistance tester
Danger: Use of this instrument in a manner not specified by the manufacturer may
impair safety features/protection provided by the equipment. Read all of the safety
information carefully before using or servicing the instrument.
Safety Notes
This symbol indicates that the meter uses double or reinforced insulation.
• Do not exceed the maximum allowable input range for any of the meter’s functions.
• Set the function switch OFF when the meter is not in use.
• Remove the batteries if the meter is to be stored for longer than 60 days.
Warnings
• Set the function switch to the appropriate position before measuring.
• Do not measure current on a circuit whose voltage exceeds 600V.
• When changing ranges always disconnect the test leads from the circuit under test.
• Do not use the Insulation resistance tester if it is damaged or if metal parts are
exposed. Also, check the meter for cracks or missing plastic.
• Be careful when working above 30V AC rms or 30V DC. Such voltages pose a shock
hazard. Discharge all circuit loads under test after measuring high voltage.
• Do not replace the batteries with the tester in a wet environment.
• Place the test leads in the proper input terminals. Ensure that the test leads are firmly
connected to the Insulation Resistance tester’s input terminals.
• Ensure that the Insulation Resistance tester is switched off when opening the battery
compartment.
3
380395/380396 V1.1 05/10
Cautions
• Improper use of this meter can cause damage, shock, injury or death. Read and
understand this user manual before operating the meter.
• Always remove the test leads before replacing the batteries.
• Inspect the condition of the test leads and the meter itself for any damage before
operating the meter. Repair or replace any damage before use.
• Voltage checks on electrical outlets can be difficult and misleading because of the
uncertainty of a definitive connection to the recessed electrical contacts. Other means
should be used to ensure that the outlet terminals are not "live".
• If the equipment is used in a manner not specified by the manufacturer, the protection
provided by the equipment may be impaired.
• Do not use the Insulation resistance tester near explosive gas, vapor or dust.
• When using the test leads, keep fingers away from the lead contacts. Keep fingers
behind the test lead finger guards.
• Do not use the tester if its cover or any of its parts are removed.
• When performing resistance tests, remove all power from the circuit under test.
• When servicing the Insulation Resistance tester, use only the supplied test leads and
power adaptor.
• Do not use the Insulation resistance tester if the battery indicator shows a low battery.
• Do not use or store the meter in areas of high humidity, high temperature, potentially
explosive or flammable environments, or in strong magnetic fields.
• Soft cloth and mild detergent should be used to clean the surface of the Insulation
resistance tester when servicing. Do not use abrasives or solvents.
• Dry the Insulation resistance tester before storing if it is wet.
• Remove the test leads and set the Range selector switch to the OFF position after
use.
Safety Category Information
This Insulation resistance tester complies with IEC61010 standard safety measurement
requirements: pollution degree 2, over-voltage category (CAT. IV) 600V, with double
insulation.
4
380395/380396 V1.1 05/10
Meter Description
1. Insulation resistance
Test/Lock switch
2. REL /down button
3. HOLD button
4. BATT /up button
5. MAX /MIN /STOP button
6. LCD Display
7. DC power adaptor Input
8. EARTH Input
9. V input and IR GUARD
terminal
10. V COM input and High
voltage shielding terminal
11. LINE: 500V to 5000V High
Voltage Output
12. PI /DAR /PEAK button
13. Backlight /AUTO OFF button
14. Function Selector Switch
Display Description
7
891011
6
5
4
3
MAX
MIN
12
13
2
1
14
1. Timed Test Indicators
2. AC/DC Indicators
3. Polarization Index and
Dielectric Absorption
Ratio Indicators
4. Analog Bargraph
5. Auto Range Indicator
6. Low Battery Indicator
7. High Voltage Indicator
8. Auto discharge Indicator
9. Continuity Buzzer Indicator
10. PASS /FAIL Indicator
11. Live Circuit Voltage Indicator
12. Units Indicator
13. Function Indicators
7
6
5
4
3
2
1
5
8
9
10
11
12
13
380395/380396 V1.1 05/10
Insulation Resistance Testing
CAUTIONS:
• Ensure that no electrical charge exists on the circuit under test.
• Insulated gloves should be worn while testing.
• Use extreme caution to avoid touching the tips of the test leads or the circuit under
test when the PRESS TO TEST button is pressed; High Voltage is present.
• Do not make measurements when the battery cover is removed.
• Always connect the Earth Cord (black) to the Earth terminal of the circuit under
test.
PRESS TO TEST Knob
Emergency Stop
CAUTION: When the live circuit warning indicates “>30V” or the warning buzzer
sounds and the
made even if the “Press to Test” button is pressed.
DANGER
•Do not touch the circuit under test immediately after testing. Voltage stored in the
circuit may cause electric shock.
•Leave the test leads connected to the circuit under test and never touch the circuit
until the discharge is complete.
All insulation resistance tests are initiated with the PRESS TO TEST knob.
1. Press and HOLD the knob to perform a test. Release the
knob to end the test.
2. Press and TURN the knob 90 degrees clockwise to the
LOCK position to lock the meter in the continuous test
mode. To end a test, either rotate the knob 90 degrees
counterclockwise or press the STOP button.
3. Always return the knob to the vertical position before
beginning a test.
To stop a test at any time, press the STOP button or rotate the PRESS TO TEST knob in a
counter-clockwise direction.
The STOP button also clears the display and beeper after a discharge is complete.
display flashes, the measurement cannot be
6
380395/380396 V1.1 05/10