Extech 380365 Instruction Manual

User's Manual
Autoranging Insulation Tester
with PC Interface
Model 380365
WARRANTY
EXTECH INSTRUMENTS CORPORATION warrants the basic instrument to be free of defects in parts and workmanship for one year from date of shipment (a six month limited warranty applies on sensors and cables). If it should become necessary to return the instrument for service during or beyond the warranty period, contact the Customer Service Department at (781) 890-7440 for authorization. A Return Authorization (RA) number must be issued before any product is returned to Extech. The sender is responsible for shipping charges, freight, insurance and proper packaging to prevent damage in transit. This warranty does not apply to defects resulting from action of the user such as misuse, improper wiring, operation outside of specification, improper maintenance or repair, or unauthorized modification. Extech specifically disclaims any implied warranties or merchantability or fitness for a specific purpose and will not be liable for any direct, indirect, incidental or consequential damages. Extech's total liability is limited to repair or replacement of the product. The warranty set forth above is inclusive and no other warranty, whether written or oral, is expressed or implied.
Introduction
Congratulations on your purchase of Extech’s Insulation Tester. Special features include Autoranging measurement display, RS-232 PC Interface for data acquisition, and backlit LCD display with analog indicator. This professional meter, with proper care, will provide years of safe reliable service.
Safety
1. Circuits under test must be de-energized and isolated before connections are made (except for voltage measurements).
2. Circuit connections must not be touched during a test. Use extreme caution when working near bare conductors and bus bars. Accidental contact with conductors could
result in electrical shock.
3. Use caution when working near voltages above 60VDC or 30VACrms.
4. After insulation tests, capacitors must be discharged.
5. Test leads (including alligator clips) must be in good working order, clean and without broken or cracked insulation.
6. When servicing, use only specified replacement parts.
International Safety Symbols
Caution, refer to this manual before using this meter
Dangerous Voltages
Meter is protected throughout by double or reinforced insulation
Specifications
General specifications
Display Backlit 3-3/4 digit (4000 count) LCD with analog indicator Sampling rate 2.5 samples/sec. (digital); 10 samples/sec. (analog) Audible Continuity Over range indicator "OL” displayed when measurement exceeds range limits Low battery indicator " - +" symbol displayed when battery voltage is low Power source Eight 1.5V AA cells Pow er consumption 20 to 90mA depending upon function Operating conditions 32 to 104oF (0 to 40oC); < 80% RH Storage conditions 14 to 140oF (-10 to 60oC); < 80% RH Dimensions 7.5 x 5.5 x 3” (190 x 140 x 77mm) Weight 2 lbs (900g)
Range Specifications
Resistance
Range Res. Accuracy Max. open
400 0.1
Short circuit current: 280mA (min. 200mA)
Continuity
Range Res. Audible tone Max. open circuit Volt Overload Protect.
200 0.1 < 40
AC Voltage (40 to 500Hz)
Range Res. Accuracy Input impedance Overload Protect.
600VAC 1V ±1.5% + 3d
Megohmmeter (Auto Range)
Range Res. Accuracy Terminal voltage
4M /40M /400M / 4000MΩ /
250V
4M /40M /400M / 4000MΩ /
500V
4M /40M /400M / 4000MΩ /
1000V
Test Current / Load Short circuit I
4000MΩ / 250V 250K 4000MΩ / 500V 500K
4000MΩ / 1000V
Note: Accuracies are specified as % reading + digits at 23oC < 80% RH
200 range with audible tone when resistance < 40
±1% + 5d 12.8V 220Vrms
1k
1mA
circuit Volt
12.8V (200mA) 220Vrms
4.5M
±3% + 5d (<2000MΩ) ±5% + 5d (>2000MΩ)
1M
250V + 20% ~ -0%
500V + 20% ~ -0%
1000V + 10% ~ -0%
< 1.5mA
Overload
Protect.
220Vrms