Extech 380340 Instruction Manual

User's Manual
Model 380340 Heavy Duty Datalogging Module
Stores data for later recall and analysis
Can be used with any Extech Heavy Duty meter
Selectable recording interval
Battery powered and portable
Heavy Duty meters are available for the
measurement of Temperature, Air Velocity and Flow, Relative Humidity, Light Level, Wood Moisture, pH, Conductivity, Dissolved Oxygen, and Pressure
Introduction
Congratulations on your purchase of Extech’s 380340 Datalogging Module. This professional device, with proper care, will provide years of safe reliable service. The Datalogger connects to and records data from an Extech Heavy Duty Meter. The sample rate at which data is recorded is user selectable. Readings stored in the Datalogger can later be transferred to a PC using the supplied software and communication cables.
1 380340 Ver. 1.6 11/00
Specifications
13
2 3 4 5 6 7 8
10
14
Input type RS-232 data stream Data storage 8000 readings (16 bit each) Display 0.5" red 2 digit LED, 100 counts (0-99) Adjustable Sample Rate Seconds: 1 to 99, Minutes: 1 to 99, Hours: 1 to 99 Internal Clock Crystal oscillator Memory Volatility Data is retained at power down Batteries Four 1.5V AA (alkaline or heavy duty type) Battery life 36 hours continuous (typical) Power consumption Approx. 20 to 54 mA DC Operating temperature 32 to 122oF (0 to 50oC) Operating humidity < 80 % RH Weight 0.45 lbs. (205g) including battery Dimensions 5.2x2.8x1.0" (131x70x26mm) (HxWxD)
Datalogger Description
1 Display 2 Memory ‘Full’ indicator 3 ‘Data’ indicator 4 ‘Pause’ indicator 5 ‘Power/Low Power’ indicator 6 ‘Auto/Manual/Out’ switch 7 Sample rate switch 8 Power ON/Off switch 9 Reset> / INC 10 button 10 Pause / INC 1 button 11 ‘Set’ button 12 ‘Data in’ port (meter to Datalogger) 13 ‘Data out’ port (Datalogger to PC) 14 Battery compartment (rear)
12
1
9
11
Datalogging Operation
Note: The Heavy Duty Meters have an Auto-Shutoff feature that will automatically power off the meter after a fixed time period. This feature should be disabled during a datalogging session. Press the "RECORD" button on the meter to disable the Auto-Shutoff feature.
Automated Datalogging Preparation
This function permits the Datalogger to automatically collect readings at the sample rate set by the user.
1. Connect the Datalogger's uP IN port to the Heavy Duty Meter as shown in the illustration below. (The supplied dual 3.5mm cable links the Heavy Duty meter to the Datalogger).
2. Turn the Datalogger OFF ("0" switch position).
3. Slide the Auto/Man/Out switch to the Auto position.
4. Slide the Time switch to the "S" (seconds), "M" (minutes) or "H" (hours) position depending on the data sampling (recording) rate desired (1-99 seconds, 1-99 minutes, or 1-99 hours).
5. If this is the first time the datalogger is being used, reset (erase) the datalogger's memory per the Reset Function section below .
6. Power the datalogger ("1" position).
2 380340 Ver. 1.6 11/00