DO NOT APPLY VOLTAGE TO TERMINALS
DISCHARGE CAPACITOR BEFORE T E S TING
+
AUTO-POWER OFF
INTRODUCTION
Congratulations on your purchase of Extech's Model 380193 LCR meter. This meter will
accurately measure capacitors, inductors and resistors using the test frequencies of
120Hz and 1 kHz. The dual display will simultaneously display the associated quality
factor, dissipation or resistance value using a series or parallel equivalent circuit.
The included RS-232c PC interface feature with Data Acquisition permits the user to
capture readings to a PC for data storage, viewing, printing, and exporting to spreadsheet
for graphing and other data manipulation tasks.
This meter is shipped fully tested and calibrated and, with proper use, will provide years of
reliable service.
International Safety Symbols
Safety Precautions
Caution ! Refer to the explanation in this Manual
!
Caution ! Risk of electric shock
Earth (Ground)
1. Make sure any covers or battery doors are properly closed and secured.
2. Always remove the test leads before replacing the battery or fuses.
3. Inspect the condition of the test leads and the meter itself for any damage before
operating the meter. Repair or replace any damage before use.
4. To reduce the risk of fire or electric shock, do not expose this product to rain or
moisture.
5. Do not exceed the maximum rated input limits.
6. Always discharge capacitors and remove power from the device under test before
performing Inductance, Capacitance, or Resistance.
7. Remove the battery from the meter if the meter is to be stored for long periods.
2
380193 V2.1 5/07
SPECIFICATIONS
Capacitance @ 120Hz
Range Cx accuracy DF accuracy Note
9.999mF ±(5.0% rdg + 5d)
(DF<0.1)
1999.9µF ±(1.0% rdg + 5d)
(DF<0.1)
199.99µF ±(0.7% rdg + 3d)
(DF<0.5)
19.999µF ±(0.7% rdg + 3d)
(DF<0.5)
1999.9nF ±(0.7% rdg + 3d)
(DF<0.5)
199.99nF ±(0.7% rdg + 5d)
(DF<0.5)
19.999nF ±(1.0% rdg + 5d)
(DF<0.1)
Capacitance @ 1kHz
Range Cx accuracy DF accuracy Note
999.9µF ±(5.0% rdg + 5d)
(DF<0.1)
199.99µF ±(1.0% rdg + 3d)
(DF<0.5)
19.999µF ±(0.7% rdg + 3d)
(DF<0.5)
1999.9nF ±(0.7% rdg + 3d)
(DF<0.5)
199.99nF ±(0.7% rdg + 5d)
(DF<0.5)
19.999nF ±(0.7% rdg + 5d)
(DF<0.1)
1999.9pF ±(1.0% rdg + 5d)
(DF<0.1)
Inductance @ 120Hz
Range Lx accuracy (DF<0.5) DF accuracy (DF<0.5) Note
10000H Not specified Not specified
1999.9H ±(1.0%rdg + Lx/10000 + 5d) ±(2.0%rdg + 100/Lx + 5d) after open cal
±(2.0%rdg + 8d) after open cal*
±(0.5%rdg + 5d) after open cal*
±(0.5%rdg + 3d)
±(0.5%rdg + 3d)
±(0.5%rdg + 3d)
±(0.8%rdg + 5d) after short cal
±(1.2%rdg + 8d) after short cal
*Note: For resistance readings above 1MΩ, series and parallel impedances may affect
readings (especially at 1kHz). This effect is often noticed on decade resistance boxes
where the AC measured value can vary from the DC calibrated value. Use fixed value
low inductance resistors (film or equivalent) for high resistance calibration or certification.
Note: In the 20Ω range, effective readings must be over 20 counts.
Test Frequency (accuracy) 122.88Hz (±4Hz) and 1kHz (±4Hz)
Display: Dual 4 ½ digit backlit LCD
Overload indication: “OL”
Low battery indication:
Measurement rate: One time per second
Auto-power off: After 10 minutes of inactivity
Operating environment: 32
Storage environment: 14
o
F to 122oF (0oC to 50oC), < 80% RH
o
F to 140oF (-20oC to 60oC), < 80% RH, battery removed
Power: 9V battery or optional external 12V-15V @ 50mA (approx.)
Fuse 0.1A/250V fast blow
Dimensions: 7.56x3.6x2.1” (19.2x9.1x5.25cm)
Weight: 12.9oz (365gm)
METER DESCRIPTION
1. Q/D/R display
2. L/C/R display
3. Keypad
4. Test fixture
5. Input jacks
6. External power input
7. Protective holster
8. Battery compartment (rear)
6
7
1
2
3
8
4
5
4
LCR Meter 380193
APO
AUTO
D
PAL
c
-0 ENTER
L/C/R Q/D/R
123
RANGE L/C/R Q/D/R
456
>2sec
MIN
SET
HOLD
MAX
789
Hi / Lo
REL
TOL
LIMITS
DO NOT APPLY VOLTAGE TO TERMINALS
DISCHARGE CAPACITOR BEFO RE TESTING
+
UTO-POWER OFF
380193 V2.1 5/07
1kHz
F
µ
Display Symbols and Annunciators
APO Auto Power Off 1KHz 1kHz test frequency
R Recording mode active 120Hz 120Hz test frequency
6
MAX Maximum reading M Mega (10
MIN Minimum reading K kilo (10
AVG Average reading p pico (10
)
3
)
-12
)
AUTO AutoRanging active n nano (10-9)
H Data hold active µ micro (10
SET SET mode m milli (10
-6
)
-3
)
TOL Tolerance mode H Henry (inductance units)
PAL Parallel equivalent circuit F Farad (capacitance units
SER Series equivalent circuit ΩOhms (resistance units)
D Dissipation factor Upper limit
Q Quality factor Lower limit
R Resistance Relative mode
L Inductance Low battery
C Capacitance % Tolerance (percentage)
5
380193 V2.1 5/07
OPERATING INSTRUCTIONS
CAUTION: Measuring a DUT (device under test) in a live circuit will produce false readings
and may damage the meter. Always remove power and isolate the component from the
circuit to obtain an accurate reading.
CAUTION: Do not apply voltage to the input terminals. Discharge capacitors before testing
Note: Measurement considerations for resistance <0.5 ohms.
1. Use positive contact alligator clips.
2. Perform a SHORT calibration zero to remove stray impedances.
3. Clean the DUT leads / contacts of oxidation or film to minimize contact resistance.
Power
1. Press the power key to turn the meter on or off
2. Auto-Power OFF (APO)
If the keypad is inactive for 10 minutes, the meter will automatically shut down. If this
occurs, press the key to resume operation.
3. Auto-Power Off disable.
To disable the auto-power off feature, from the off position, press and hold the power
on key until “APO OFF” appears in the display. Auto-Power Off will also be disabled
if the MIN MAX record mode is used or if the meter is powered by an external power
supply.
Frequency Selection
Press the FREQ key to select either 120Hz or 1kHz as the test frequency. The frequency
selected appears in the display.
Generally, 120Hz will be used for large electrolytic capacitors and 1kHz used for most
other tests.
Parallel/Series Selection
Press the PAL SER key to select either a parallel (PAL) or series (SER) equivalent circuit.
The selected mode will appears as “SER” or “PAL” in the display.
This mode defines the R loss of an inductor or capacitor as a series loss or a parallel loss.
Generally, high impedances are measured in the parallel mode and low impedances are
measured in the series mode.
Range Selection
The meter turns on in the autoranging mode with “AUTO” indicated in the display. Press
the RANGE key and the “AUTO” indicator will disappear. Each press of the RANGE key
will now step through and hold the available ranges for the parameter selected. To exit the
manual range mode, press and hold the RANGE key for 2 seconds.
6
380193 V2.1 5/07
Inductance, Capacitance, Resistance Selection
The L/C/R key selects the primary parameter measurement function. Each press of the
key will select either inductance (L), capacitance (C) or resistance (R) along with the
proper units of H (henries), F (farads) or
Quality, Dissipation, Resistance Selection
The Q/D/R key selects the secondary parameter measurement function. Each press of the
key will select either the quality (Q) or dissipation (D) indicators or resistance (Ω) units in
the small secondary display.
Hold and Backlight Selection
The HOLD >2sec key selects the Hold feature and also enables the display backlight.
Press the key and the H indicator will appear in the display and the last reading will
“freeze” in the display. Press the key again and the reading will begin to update again.
Press and Hold the key for 2 seconds and the display backlight will turn-on. To extinguish
the backlight, press and hold the key again for 2 seconds or wait 1 minute for it to
automatically disable.
Minimum, Maximum and Average Selection
The MAX MIN key selects the recording function. Press the key and the “R“ indicator will
appear in the display and the meter will begin recording the minimum, maximum and
average measured values. When this mode is entered, the auto power off and the function
keys are disabled.
Max-Min operation
1. Set all the function parameters for the test.
2. Press the MAX MIN key. The “R“ indicator will appear and a “beep” will sound after
approximately six seconds. Two “beeps” will sound every time the max or min is
updated.
3. Press the MAX MIN key. The “MAX” indicator and the maximum recorded value will
appear in the display
4. Press the MAX MIN key. The “MIN” indicator and the minimum recorded value will
appear in the display
5. Press the MAX MIN key. The “MAX - MIN” indicator and the difference between the
maximum - minimum value will appear in the display
6. Press the MAX MIN key. The “AVG” indicator and the average of the recorded
values will appear in the display.
7. Press and hold the MAX MIN key for 2 seconds to exit the mode.
Notes:
The average value is a true average and averages up to 3000 values. If the 3000 limit is
exceeded, the AVG indicator will flash and no further averaging will take place. The max
and min values will continue to update.
If the HOLD key is pressed during min max recording, the recording is halted until the
HOLD key is pressed again.
Relative Mode
The relative mode displays the difference between the measured value and the value of a
stored reference.
1. Press the REL key to enter the Relative mode.
2. The value in the display when the REL key is pressed will become the stored
reference value and the display will indicate zero or a value close to zero (since the
measured value and the reference value are the same at this point).
3. All subsequent measurements will be displayed as a value relative to the stored
value.
4. The reference value may also be a value that was stored into memory using the
SET Relative procedure (see the Setting a Relative Reference paragraph).
5. To use the SET relative value, press the SET key while in the Relative mode.
6. To exit the Relative mode, press and hold the REL key for 2 seconds.
Ω (ohms) in the main large display.
7
380193 V2.1 5/07
Hi / Lo Limits Mode
The Hi / Lo limits mode compares the measured value to the stored high and low limit
values and gives an audible and visible indication if the measured value is outside the
limits. See the setting Hi/Lo limits paragraph below to store the limits in memory.
1. Press the Hi/Lo LIMITS key to enter the mode. The display will briefly show the
stored upper limit with the “ “ indicator and then the stored lower limit with the “ “
indicator before displaying the measured value.
2. The meter will sound an audible tone and blink the upper or lower limit indicator if
the measured value is outside the limits.
3. The meter will ignore an “OL” overload reading.
4. Press the Hi/Lo LIMITS key to exit the mode.
% Tolerance Mode
The % Tolerance Limits mode compares the measured value to a % high and low limit
based on a stored reference value and gives an audible and visible indication if the
measured value is outside the limits. Any % limit can be entered in the SET % Limit mode
(see paragraph below) or standard 1%, 5%, 10% and 20% symmetrical limits can be
selected directly in the % Tolerance mode.
1. Press the TOL key to enter the mode. The display will briefly show the stored
reference value in the main display and the small display will indicate the %
difference between the measured value and the reference value. See the SET %
Limit paragraph to change the reference value.
2. Press the TOL key to step through and select the 1, 5, 10 or 20% settings. The
selected % will briefly appear in the small display.
3. Previously stored user defined % limits are accessed by pressing the SET key.
4. The meter will sound an audible tone and blink the upper or lower limit indicator if
the measured value is outside the limits.
5. Press and hold the TOL key for 2 seconds to exit the mode.
Set Limits and Open/Short Calibration Selection
The SET key is used to; 1. Set Hi/Lo limits, 2. Set % limits, 3. Set Tolerance reference
value and 4. Perform the Open / Short calibration. The SET mode can only be activated if
no other function is active.
Entering the SET mode
1. Power ON and press the SET key.
2. The display will clear, “SEt“ will appear in the small display and a flashing TOL
and flashing indicators will appear in the display.
3. The 5 keys that are now active are; Power, SET, REL, Hi/Lo, and TOL
Open and Short Calibration
The Open and Short function removes stray parallel and series fixture impedances from
the measured value. This feature improves accuracy for very high or low impedances.
(Note: Remove any leads from the meter during this procedure. Leaving them connected
will add impedance to the circuit causing the calibration to fail indicated by OUT UAL
appearing on the display.)
1. Press the SET key 2 times and the display will indicate “CAL OPEn”.
2. Remove any devices or test leads and press “ENTER” (PAL SER). After several
seconds the calibration will complete and display “CAL SHrt”.
3. Short circuit the test fixture and press “ENTER” (PAL SER). After several seconds
the calibration will complete and the meter will return to normal operation.
4. Press the “SET” to bypass either the open or short calibration.
8
380193 V2.1 5/07
Setting Absolute Hi/Lo Limits
The Hi/Lo limits set allows the user to enter an upper and lower limit value into memory
for comparison to the measured value.
1. Press the SET key and then the Hi / Lo LIMITS key. The upper limit “ “ indicator
will flash and the previously stored upper limit will appear with the first digit flashing.
2. Set the value of the flashing digit by pressing the appropriate numeric key. The
adjustment selection will then proceed through each digit from left to right.
3. Press the - 0 key after the last digit is set to change the value of the sign to negative
or positive.
4. Press the “ENTER” key to store the value and continue on to the lower limit
adjustment.
5. The lower limit “ “ indicator will flash and the previously stored lower limit will
appear.
6. Adjust the limits as described for the upper limit and press the “ENTER” key when
complete.
Setting % Tolerance Limits
The % tolerance set allows the user to enter an upper and lower percentage limit into
memory for comparison of the measured value to a reference value.
1. Press the SET key and then the TOL key. The “TOL” indicator will flash and the
previously stored reference will appear with the first digit flashing.
2. To adjust the reference, set the value of the flashing digit by pressing the
appropriate numeric key. The adjustment selection will then proceed through each
digit from left to right.
3. Press the “ENTER” key to store the value and continue on to the % upper limit
adjustment. The upper limit “ “ indicator will flash and the previously stored upper
% limit will appear.
4. Adjust the % limit as described for the reference value and press the “ENTER” key
when complete. The lower limit “ “ indicator will flash and the previously stored
lower % limit will appear.
5. Adjust the lower % limit and press “ENTER” when complete.
Setting a Relative Reference
The relative set allows the user to store a relative reference value into memory for later
use in the REL mode.
1. Press the SET key and then the REL key. The “ ” indicator will flash and the
previously stored reference will appear with the first digit flashing.
2. To adjust the reference, set the value of the flashing digit by pressing the
appropriate numeric key. The adjustment selection will then proceed through each
digit from left to right.
3. Press the - 0 key after the last digit is set to change the value of the sign to negative
or positive.
4. Press the “ENTER” key to store the reference value.
9
380193 V2.1 5/07
PC INTERFACE
The Model 380193 LCR Meter includes a PC interface feature for use with the supplied
Windows
Instructions for use of the PC interface are included on the supplied Program Disk and are
beyond the scope of this operation manual. For complete details and instructions refer to
the HELP file on the supplied Program Disk.
TM
software. The interface allows the user to:
View measurement data in real time on the PC
Save, Print, and Export measurement data.
Set standard and high / low limits for data analysis
Generate calibration reports in spreadsheet format
Plot SPC (statistical process control) analyses
Database compatibility (supports ODBC) for use with: SQL server, Access
and other database utilities
TM
Warranty
EXTECH INSTRUMENTS CORPORATION warrants this instrument to be free of defects
in parts and workmanship for one year from date of shipment (a six month limited warranty
applies to sensors and cables). If it should become necessary to return the instrument for
service during or beyond the warranty period, contact the Customer Service Department at
(781) 890-7440 ext. 210 for authorization or visit our website www.extech.com for contact
information. A Return Authorization (RA) number must be issued before any product is
returned to Extech. The sender is responsible for shipping charges, freight, insurance and
proper packaging to prevent damage in transit. This warranty does not apply to defects
resulting from action of the user such as misuse, improper wiring, operation outside of
specification, improper maintenance or repair, or unauthorized modification. Extech
specifically disclaims any implied warranties or merchantability or fitness for a specific
purpose and will not be liable for any direct, indirect, incidental or consequential damages.
Extech's total liability is limited to repair or replacement of the product. The warranty set
forth above is inclusive and no other warranty, whether written or oral, is expressed or
implied.
Calibration and Repair Services
Extech offers repair and calibration services for the products we sell. Extech also
provides NIST certification for most products. Call the Customer Care Department for
information on calibration services available for this product. Extech recommends that
annual calibrations be performed to verify meter performance and accuracy.