EWS ES2030 CV Instruction Manual

Control for ion exchangers and filter systems
Instruction manual
Software version 2.03
ES2030 CV
Functional description..................................................................................................................…. 1
Illustration .....................................................................................................................................…. 2
Service and regeneration displays …………...............................................................................…. 3
LED indicator lights…...………………………………………………………………....……………. 3
LCD display………....….....………………………………………………………………....………… 3
First line…….…...……………………………………………………………………...………. 3 Second line during service…….………………………………………………………...……. 3 Second line during regeneration………………………………………………………...……. 3
Displaying and altering program values.......................................................……………………..…. 4
Hardness of supply water / filter capacity..………………………………………………………..... 4
Current time………………………………...………………………………………………………..... 4
Info key ....................................…...........................................................……………………….………. 5
Flush……….……………….………………………………………………………………........…….. 5
Regeneration time..……….………………………………………………………………........…….. 5
Regeneration restrictions…………………………………………………………………........…….. 5
Additional program run……………………………………………………………………........…….. 5
Filter capacity….………………………………………………………………………………………. 5 Water delivery…….………….………………………………………………………………………... 6
Input states……………..………………………………………………………………………..…..... 6
Output states……………..………………………………………………………………………........ 6
Service number….……….……………………………………………………………………………. 6 Software version….…………………………………………………………………………………... 6
Programmed inputs………………………………………………………………………........…….. 6
Programmed outputs……………………………………………………………………………........ 6
Last regeneration.……………………….……………………………………………………………. 6 Regeneration ratio……………………….……………………………………………………………. 6
Messages……………………......................................................................……………………………. 7
Capacity exceeded…. .……………...……………………………………………………………….. 7 Power failure…………..……………...……………………………………………………………….. 7 Refill regeneration medium.…………..……………………………………………………………... 7
Delayed regeneration…..….…......…………………………………………………………………... 7
Stop regeneration……..…....……….………………………………………………………………... 8
Stop service…….…………..……………………………....……………………………………….…. 8
Minimum regeneration interval.…………………………………………………………………….… 8
Cancel buzzer
………………......................................................................……………………………. 8
Switching the OUT1 and OUT2 relays on and off .........................…..........................................… 8
Additional output function.....….……………………………………………………………………... 8
Regeneration cycle function….….…..…………………………………………………………….…. 8 Supply pulse function…………………..……………………………………………………………... 8 Warning function..………..……………….…………………………..…………………………….… 8
Desalinate function….……………………...........…………………………………………………... 8
Initiating regeneration manually...............................................................……………………………. 9
Special functions.....................................................................…...............…………………………….. 9
Filter change over without program initiation.……………..………………………………………… 9 Immediate stop…………………...…………………………..………………………………………... 9 Regeneration of standby filter……………………………..…………………………………..……… 9 Switching from parallel to alternate.……………………..…………………………………………... 10
Fast cycle..………...........……….……..………………....…………………………………………... 10
Regeneration without initialisation..…..……………..…..…………………………………………... 10
Regeneration of Filter 1 ONLY…....…..…………..……..…………………………………………... 10
Regeneration of Filter 2 ONLY…....…..………..………..…………………………………………... 10
Displaying and modification of the basic settings...............................……………………………… 11
General information on programming and language settings…….……………………………….. 11
1 Electrical control.…..………………………......………………………………………………….. 12
2 Number of valve switch phases ………….....…………………………………………………... 15
3 Pulse length……………………………...…………..……………………………………..……… 15 4 Regeneration times….…………………..…………..………………………………………..…... 15 5 Filter switching…………..…………………………………………………………………………. 15 6 Delayed regeneration.……………….……………………………………………………………. 17 Starting on real time clock…………..……………………………………………………………. 18
7 Interval start of regeneration…………...…………....................…..…………………………… 18
8 Minimum regeneration distance….....................………………………………………….……. 19
9 Definition of input functions…………………….…….…………..………………………….…… 20 10 “Water meter” input…..……...……………………..…………………..…………………….…… 21 Filter capacity…..…..……...……………………..…………………..…………………..……..… 22 11 “Stop service” input…….…...……………………..…………………..…………………..…...… 22
12 “Start regeneration” input…….…………………..…………………..…………………..…....… 23
13 “Chemical shortage” input…….…………………..…………………..…………………..…....… 23
14 “Stop regeneration” input…….…………………..…………………..…………………..…....… 23
15 Definition of output functions…………………….…….…………..…………………………..… 24 16 “Additional program” output...…………………..……………………..………………………… 25 17 “Regeneration” output.……...…………………..……………………..……………………….… 25 18 “Flow pulse” output..…………………….…………..…………………………..……………...… 26 19 “Warning” output..…………………….…………..…………………………..………………...… 26 20 “Desalinate” output………………….…………..…………………………..…………………..… 27 21 Buzzer…….…………………………………………..…………………..………………………… 27 22 Programming mode.……...…………...…..……………………………..……………………….. 27
Examples of systems……........................................................................……………….…………… 28
Typical electrical wiring diagrams............................................................……………….…………… 29
Wiring diagram ES2030 CV.......................................................................……………….…………… 31
Notes on installation and initial use........................................................……………….…………… 32
Technical data….........................................................................……………………………............… 33
Declaration of conformity...................................................…………….………………….……...…… 34
ES2030 CV Functional description 1
Functional description
The controller ES2030 CV (wall mounted) is used for the automatic control and monitoring of single and double filter systems.
If additional control functions are required, they can be obtained with the IF2030 card, which can also be installed subsequently.
The software’s flexible programming capacity and the individually adaptable hardware make a wide range of uses in water treatment systems possible. In combination remote control valves or pilot distributors, these control units can operate water-softening systems, partial flushing systems and filter systems.
NOTE: For the sake of simplicity, in these instructions the treatment process carried out by a filter system (e.g. deferrization) is also referred to as “REGENERATION”, as is usual in the case of ion exchangers.
A regeneration can be initiated :
1. by manual switch
2. by remote switch (water hardness monitor, conductivity meter, manual switch, etc..)
3. by pre selected amount (pulse water meter required)
4. after set time intervals (e.g. every 72 hours)
5. starting on real time clock A time window can be set to determine times
when regeneration must not take place (delayed regeneration).
A minimum regeneration distance between regenerations prevents regenerations being initiated constantly if the water meter or the remote switch is faulty.
Terminal diagram of ES 2030 CV
2 relays for remote control valves or pilot distributors (control valve 1 and control valve 2) 2 relays for service valves (service valve 1 and service valve 2) 1 relay for automatic synchronization of the remote control valves or pilot distributors, with a connection for the message ‘program running’ (program)
1 signal input, programmable for the following functions : water meter, stop, start or chemical shortage
1 12V= output for an external electronic application, e.g. turbines with Hall effect switch (auxiliary power)
1 power output for “control on” message and power supply to the potential free contacts “OUT1” and “OUT2” (power out)
Separately available card IF2030 : 1 signal input, programmable for the following functions : water meter, stop, start or chemical
shortage 2 output relays programmable for the following functions : additional program, regeneration,
flow pulse, warning or desalination. (OUT1 and OUT2).
Control valve 1
Service valve 1
Control valve 2
Service valve 2
Program IN1auxiliary
power
IN2power
out
power in
OUT1 OUT2
ES2030 CV Illustration 2
Illustration
Wall-mounted
1 Service filter 2 6 Main switch 11 Output 2 16 Input 2 LED 2 Service filter 1 7 Start regeneration 12 Information 17 Output 2 LED 3 Regeneration filter 1 8 Supply water 13 Programming 18 Output 1 LED 4 Regeneration filter 2 9 Time 14 LCD display 5 Main fuse 10 Output 1 15 Input 1 LED
Service unit 1
100.00m3 Mo17:10
MT 4A
1
2 3 4
5
6
8 9 10 11 12 13
14
12
IN1 IN2 OUT1 OUT2
#
Enter
info
OUT1OUT
2
7
15 16
17
18
ES2030 CV Service and regeneration display 3
Service and regeneration display
LED control lamps
Coloured control lamps show the unit’s major states : Filter 1 service (green)
Filter 2 service (green) Filter 1 regeneration (orange) Filter 2 regeneration (orange)
IN1 = Input 1 active (orange) IN2 = Input 2 active (orange) OUT1 = Output 1 active (orange) OUT2 = Output 2 active (orange)
LCD display
First LCD line
The first line of the LCD display shows the present state of the system, e.g. “Filter1 in service”, “Filter 2 in regeneration” or “not in service”.
Second LCD line during service
The second line of the LCD display shows the following information during service:
1. The amount of water remaining until the next regeneration
or: The time of the next regeneration if a ‘delayed regeneration’ has been initiated (see program step 6).
or: The number of hours until the next regeneration (see program step 7).
or: Alternating with the ‘amount of water remaining’ of the current through flow (see program step 10.1 : pulse count)
or: The flushing time remaining (see program step 19)
or: “No Autom. Reg” if no automatic initiation of regeneration was selected (water meter, time interval).
2. The current time
Second LCD line during regeneration
During regeneration, the second line of the LCD display shows the remaining time for the current phase and after the oblique the remaining time for the whole regeneration.
Or: Alternating with the regeneration times, the remaining time for the additional program (see program step 15).
Service unit 1
100.00m3 Mo12:00
Regener.unit 1 Phase:2 80/100m
12
IN1 IN2 OUT1 OUT
2
Service unit 1
100.00m3 Mo12:00
ES2030 CV Displaying and altering program values 4
Displaying and altering program values
The main program values can be displayed and altered if required by pressing a key.
Hardness of supply water / filter capacity
Press the key with the symbol . In the case of an ion exchanger the bottom line will show the present supply water hardness, for a filter system it will show the capacity entered. (see program step 10.3).
If you wish to change the value shown, use the ‘ ‘ key to move the cursor under the figure to be changed, and change the value with the number key ( ‘#’).
The following values may be entered, according to the units entered at phase 10.3 of the basic programming:
Unit of the supply water hardness : entry value : °D 2 - 99 °F 4 - 199 °E 2 - 99 mg/l CaCO3 40 -1999 gpg 2 - 99 no unit 0,01 - 9999,99 m3
(0,02 mmol/l = 0,10°D = 0,13 °E =0,18 °F = 1,79ppm = 0,11gpg(USA) = 0,13gpg(UK))
The calculation of the soft water quantity with altered values is carried out at the start of each regeneration using the formula:
NOTE: In the case of remote control valves with water supply by-pass ports, the water meter records the dilution amount. The hardness of the diluted water must therefore be subtracted from the value to be entered.
Example : Supply water hardness = 15 °D Diluted water = 6 °D This gives an input value of : 15 °D – 6 °D = 9 °D
Current time
Press the time key with the symbol . The current time appears in the lower line.
If you wish to change the value shown, use the ‘ ‘ key to move the cursor under the digit to be changed, and change the value with the number key ( ‘#’).
Water hardness: 1
8°D
Current time M
o 15:50
#
Exhanger capacity (°D m3) . Supply water hardness (°D)
= soft water quanitity (m3)
#
ES2030 CV Info 5
Info key
The information key is used to display various information and values. Only the service telephone number can be changed using the info key.
If the info key is pressed during programming, the full texts of some abbreviated texts are shown in the LCD display.
Flushing
The following values are shown : Top right : Flush time in seconds
Bottom left : quantity of water remaining
before flushing Bottom right : quantity of water between two flushes. NOTE: This display shown only if the flushing function is selected (see program step 15).
Regeneration time
The total time for a regeneration cycle is shown.
Regeneration restrictions
NoReg 16:00 – 18:00
If “delayed regeneration” was selected during programming at step 6, the period in which no regeneration is to occur is shown. Otherwise “NoReg ----“ is displayed.
IntRg 72
If “interval start” was selected during programming at step 7, the time interval in hours is shown. Otherwise “IntRg --“ is displayed.
MinRg 4
If a “minimum regeneration distance” was selected during programming at step 8, the time interval in hours is shown. Otherwise “MinRg --“ is displayed.
Additional program
Bottom left : the starting point of the additional program is shown. Bottom right : the start time entered is shown or the time remaining if the additional program is currently running.
If phase ‘0’ is displayed, the additional program will first run for its full time, followed by the regeneration program.
If phase ‘E’ is displayed the regeneration program will first run completely, then the additional program.
NOTE: If the additional program function was not selected during programming at step 14, “no additional program” will be displayed. If the IF card is not fitted this display will be omitted.
Filter capacity
The water quantity delivered by a filter between two regenerations is shown. In the case of ion exchangers, the calculation is always carried out using the currently entered values for filter capacity and the supply water hardness.
NoReg16:00-18:00 IntRg72 MinRg4
Additional prog. Phase: 2 20
Unit capacity 100m3
Flushing 500l 500l 20s
info
Regen.time Σ95m rest.0m
ES2030 CV Info 6
Water delivery
The total quantity of water delivered by the system is shown.
Input states
Indicates the current state of each input. (- = Input not activated, | = Input activated)
WM = Water meter SP = Stop service ST = Start regeneration CH= Chemicals shortage HO = Stop regeneration
Output states
The current switch states of the output relays are shown. Each figure has a relay allocated to it (see switching diagram on p.31). A horizontal stroke ‘-‘ under a figure means ‘relay switched off’. A vertical stroke ‘ | ‘ under a figure means ‘relay switched on’.
NOTE: Relays 7 and 8 are only shown if an IF card is fitted.
Service number
A service phone number is displayed. You can also change the number here.
Change the phone number : Select number : Lower number : Higher number :
Software version
The software is continuously updated by the factory. Where necessary changes are made
to reflect new technology and customer requirements. The number of the version currently installed is displayed.
Programming the inputs
The programmed functions of input IN1 (and IN2 if the IF2030 card is fitted) are displayed.
Programming the outputs
The programmed functions of output OUT1 and OUT2 are displayed. NOTE: This display is only shown if the IF2030 card is fitted.
Last regeneration
The display shows how much time has passed since the last regeneration. For example : 3d 12h 15min It has been 13 days, 12 hours and 15 minutes since the last regeneration.
Regeneration ratio
The regeneration ratio entered is displayed, and after the oblique the current state if filter 2’s regeneration counter is shown.
Example 1: 1:3/2 Regeneration ratio of filter 1 : filter 2 = 1:3 Filter 1 has already been regenerated once.
Example 2: 1:3/1 Regeneration ratio of filter 1 : filter 2 = 1:3 Filter 1 has already been regenerated twice. Both filters will be regenerated at the next regeneration. NOTE: This display only appears where a connection in series with two filters and a regeneration ratio greater than 1:1 has been selected.
Service 0031 73 443755
Softwareversion ES2030cv2.03.00g
IN1=Water meter IN2=prog.initi
OUT1=Add. Progra OUT2=Status
last regenera. 3d 12h 15Min.
Treated water
0.1m3
Output 12345 78
-|--- --
Relation 1:3/3
Input WM-ST-
ES2030 CV Messages 7
Messages
During service and during the regeneration of the system, various signals are given depending on the type of controller and its programming. These signals can be signalled with the built-in buzzer and displayed in the LCD display. If the extension card IF2030 is installed an additional relay can be selected as warning relay (step
15).
Press the key OUT1 or OUT2 to clear the buzzer and any activated warning relay. The LCD display is only cleared when the warning signal is no longer active.
Capacity exceeded
This display can only appear with a double filter system. While one of the filters is regeneration, the other was also called on to regenerate. The warning in the LCD display is cleared when this filter starts regeneration.
Possible causes where activated by the water meter : Incorrect setting of capacity, supply water hardness or of the water meter itself. Overloading of the system e.g. by filling a large container.
Possible causes where activated externally by a water analysis device : Saturation of a system newly put into use, caused by the negative ion effect. Solution : fit a flushing valve or circulation pump. Reduce the sensitivity of the analysis Device.
Other possible causes : Poor regeneration of the filter due e.g. to regeneration medium not being present or being incorrectly primed.
NOTE: In two-filter systems, after a flow dependent regeneration activation the regeneration of the second filter will follow immediately the current regeneration ends. However in the case of external activation of the regeneration, e.g. by a water analysis device, the regeneration does not follow, as it can be assumed that the hardness warning
occurred as a result of a standstill hardening of the standby filter. The second filter is only regenerated if the relevant start signal is still present at the end of the current regeneration or if it reset. In the case of ion exchangers with a salt release valve, if no brine has yet formed for the second filter, stop the regeneration by switching off the unit.
Power failure
No data are lost if there is a power failure. When power is restored the control panel will return to the same setting with the same values. Only the current time must be reset after a longer power loss. NOTE: If the system is at the regeneration setting when power is lost, the filter may become over saturated again if the water pressure continues during this time and it is washed by supply water over a period of hours. If so, stop the regeneration and then re-start.
Refill regeneration medium
Refill regeneration medium. NOTE: An imminent regeneration will not be carried out unless either regeneration medium is again available or the “start regeneration” key with the symbol is pressed. In the case of alternately operating two filter systems, the unit switches over to the standby filter.
Delayed regeneration
The required regeneration will not start until the time shown in the LCD display. However the regeneration can be started immediately by pressing the “start regeneration” key ( ). This display is only shown if activation of “delayed regeneration” was selected in step 19 or 21.
S T A T U S Instal.exceeded
S T A T U S Supply failure
OUT1OUT
2
S T A T U S Refill RegMedium
S T A T U S Prohibited Reg.
ES2030 CV Messages / Cancel Buzze r / Switching O UT1 an d OUT2 8
Stop regeneration
This warning may have various causes depending on the function of the switch contact connected, for instance two controls may be blocking each other, or the control pressure for a pneumatic valve may have been cut off. Find the cause. In alternately operating two filter systems, if the “Stop” warning is already displayed at the start
of a regeneration cycle the unit switches over to the standby filter. NOTE: The stop signal can be cancelled for the duration of the regeneration cycle by pressing the ‘start’ key with the symbol. Regeneration then continues.
Stop Service
This display only appears if activation of “stop service” was selected at program step 19 or
20. The LCD display is cleared automatically as soon as the input signal is no longer present.
Minimum regeneration distance
Possible causes if activated by the water meter:
Incorrect setting of capacity, supply water hardness or of the meter itself. Overloading of the system e.g. by filling a large container. Possible causes where activated externally by a water analysis device.
Saturation of a system newly put into use, caused by the negative ion effect. Solution: fit a flushing valve or circulation pump. Reduce the sensitivity of the analysis device. NOTE: You determine at program step 8.3 whether regeneration is to follow automatically at the end of set “minimum regeneration distance” or whether the next regeneration has to be started manually. NOTE: The message in the LCD display is not cleared until regeneration is started.
Cancel buzzer
If the built-in buzzer sounds, it can be cancelled immediately by pressing the ‘OUT1’ or ‘OUT2’ key.
Switching the OUT1 and OUT2 relays on and off
If the controller has been fitted with the IF expansion card, the two additional relays with the functions selected at program step 14 may be switched on and off manually by pressing the relevant key for approx. 5 seconds. The ‘OUT1’ key is assigned to relay 7 and the LED display ‘OUT1’. The same applies to ‘OUT2’ and relay 8.
‘Additional program’ function
The relay can be switched on and off during the ‘service’ or ‘regeneration’ phases. The test function is automatically deactivated at the beginning and end of a regeneration.
‘Regeneration’ function
The relay can be switched on and off (e.g. for control purposes) during the ‘service’ phase. It
is switched off automatically at the end of a regeneration.
‘Flow pulse’ function
The relay is switched on for the length of time set at program step 17.
‘Warning’ function
The relay is switched on (e.g. for control purposes) for as long as the key is pressed. If the relay has been switched on by a warning the relay is cleared.
‘Flush’ function
The relay is switched on for the length of time set at program step 19. If a flush cycle is already running, it can be stopped prematurely.
S T A T U S StopRegeneration
S T A T U S Stop service
S T A T U S Min.regen.period
OUT1OUT
2
ES2030 CV Initiating regeneration manually / Special functions 9
Initiating regeneration manually
A regeneration cycle can be initiated manually at any time pressing the ‘Start’ key with the symbol . Regeneration of the filter in service commences after six seconds.
- In the case of systems operating alternately, the standby filter is put into service.
- If ‘delayed regeneration’ was selected at step 6.1 of the programming, the time function is activated and the time at which the delayed regeneration will be initiated automatically is displayed at the bottom left of the LCD display.
- No regeneration is yet initiated.
- If the time function for ‘delayed regeneration’ has already been activated (and the time regeneration will be initiated is already displayed at the bottom left of the LCD display), regeneration will be initiated after four seconds regardless of the time shown.
- The filter’s flow counter is reset to full capacity after regeneration.
- If initiation at intervals was selected at step
7.1 during the input of basic values, the hour interval meter is set to its preset interval.
- If a minimum regeneration interval was selected at step 8.1 during the input of basic values, the timer for the regeneration interval is reset.
Special functions
These functions should only be used by a professional water treatment specialist, as their impr oper use may lead to malfunctions.
Filter change over without program initiation
Simultaneously press the function keys with the symbols and . Change over of the filters will take place after 4 seconds on two filter systems.
NOTE: Separate flow counters are used for each filter. Where an almost saturated filter is switched into the standby position, it may happen that regeneration becomes necessary shortly after it has been brought into service, and this may be at a time when the other filter is still regenerating. In this event, the fault warning ‘capacity exceeded’ appears.
Immediate Stop
Simultaneously press the function keys with the symbols and . Any regeneration program running will stop after 4 seconds and the system will be switched into service position.
NOTE: Multi stage valves which have no connection to return them automatically to the service setting will remain at a regeneration setting and are no longer synchronized with the control panel.
Regeneration of standby filter
Simultaneously press the function keys with the symbols and . The regeneration of the standby filter will begin after 4 seconds on two filter systems
NOTE: this only applies for alternate filter service (program step 5.3 = YES).
OUT
1
OUT
1
OUT
1
OUT
1
OUT
1
OUT
1
Loading...
+ 27 hidden pages