EUTECH INSTRUMENTS CON 700 - REV 2, CON 700 Instruction Manual

Instruction Manual
CON 700 Conductivity/ºC/ºF Bench Meter
Technol o
gyM
adeEa
sy...
Part of Thermo Fisher Scientific 68X541702 Rev. 2 Jan 2010
TABLE OF CONTENTS
1. INTRODUCTION.............................................................. 1
2. GETTING STARTED........................................................ 2
2.1 Keypad Functions ...................................................................2
2.2 LCD Annunciators...................................................................3
2.3 Meter Connections..................................................................3
3. ELECTRODE INFORMATION......................................... 4
4. CALIBRATION................................................................. 4
4.1 Automatic or Manual Calibration.............................................4
4.2 Single or Multi-Point Calibration..............................................5
4.3 General Calibration Tips .........................................................6
4.4 Automatic Conductivity Calibration Procedure........................6
4.5 Manual Conductivity & TDS Calibration Procedure.................7
4.6 Temperature Calibration .........................................................7
5. MEASUREMENT.............................................................. 7
5.1 Taking Measurements.............................................................7
5.2 Automatic and Manual Ranging..............................................8
5.3 HOLD Function .......................................................................8
5.4 Storing and Recalling Data .....................................................8
6. SETUP FUNCTIONS........................................................ 9
6.1 P1.0 CAL (Calibration)............................................................9
6.2 P2.0 ELE (Electrode Information)............................................9
6.3 P3.0 ConF (Configuration)......................................................9
6.4 P3.1 rdY (Ready / Stability Indicator)......................................9
6.5 P3.2 ºC ºF (Celcius Or Fahrenheit).........................................10
6.6 P3.3 AtC (Automatic Temperature Compensation).................10
6.7 P3.4 tdS (Total Dissolved Solids factor)..................................10
6.8 P3.5 t.CO (Temperature Coefficient).......................................10
6.9 P3.6 t.nr (Normalization Temperature in ºC)...........................11
6.10 ACAL (Automatic Conductivity Calibration).............................11
6.11 SPC (Single Point Calibration)................................................11
6.12 CELL (Nominal Cell Constant)................................................11
6.13 P4.0 rSt (Reset) .....................................................................12
6.14 P5.0 CLr (Clear Memory).......................................................12
7. CALCULATING TDS CONVERSION FACTOR.............. 13
8. CALCULATING TEMPERATURE COEFFICIENTS........ 14
9. REPLACEMENTS AND ACCESSORIES........................ 15
10. TROUBLESHOOTING GUIDE......................................... 16
11. SPECIFICATIONS............................................................ 17
12. WARRANTY..................................................................... 18
13. RETURN OF ITEMS......................................................... 18
1
1. INTRODUCTION
Thank you for purchasing our Con 700 series benchtop meter. This microprocessor-based meter is economical and simple to use. The design incorporates a large LCD for clear viewing, yet offers a small footprint to conserve space.
The CON 700 measures conductivity or TDS, and temperature (ºC o
r ºF) simultaneously.
Each meter includes a convenient slide-out card for quick referenc
e.
Some configurations include an electrode arm and metal bra
cket which can be easily attached to the left or right side of
the meter according to your preference. The CON 700 series benchtop meter replaces our popular CON
510 serie
s meter that was introduced in 2002.
We take great pride in every instrument we manufacture and hope
this one serves you well.
2. GETTING STARTED
2.1 Keypad Functions
Powers the meter on and off. Upon power on, the meter automatic
ally begins in the mode that was last used. Calibration and memory values are retained even if meter is unplugged.
Toggle between available measurement modes— Con
ductivity with Temperature or TDS with Temperature. Also used to switch to Temperature calibration during Conductivity or TDS calibration mode.
Press and hold for 5 seconds to enter SETUP mode.
Toggles between measurement and calibrati on modes. In SET
UP mode, returns user to the measurement mode.
Confirms calibration values in CAL mode. Confirms selections in SETUP mode. Changes range and resolution in MEAS mode.
View recalled values in memory mode. MI (Memor
y Insert) stores values into memory.
S
Increase value. Scroll up in SETUP & CAL modes.
MR (Memory Recall) recalls values from memory T Decrease valu
e. Scroll down in SETUP & CAL modes.
Freezes measured reading. Press again to resume live reading.
2
2.2 LCD Annunciators
2.3 Meter Connections
CON/
Temp
8-pin DIN connection for 2-cell Con/TDS/Temp electrode
DC
Power supply
3
3. ELECTRODE INFORMATION
The CON700 includes an electrode with a nominal cell constant of k = 1.0, built-in temperature sensor, and 1 meter cable. The Ultem body housing has good chemical resistant properties. The electrode design offers fast temperature response and reduces air entrapment, ensuring accurate, repeatable, and stable readings.
The wetted materials of the probe include:
Polyetherimide (Ultem) – protective probe guard
Polybutylterphalate (Valox) – sensor housing
Stainless Steel (SS 304) – 2 bands
The protective probe guard can be removed tempor arily for cleaning however it must be re-attached during measurement and calibration. Erroneous results will occur while the probe guard
is removed.
Always immerse the probe beyond upper steel band for best results. Use the fill line on the outside of the probe guard for reference.
4. CALIBRATION
4.1 Automatic or Manual Calibration
The CON 700 is capable of automatic or manual calibration for
n the
TA
BLE 1, the CON 700
must be set for manual calibration instead.
conductivity, and manual calibration for TDS. In the automati
c calibration mode, the meter will automatically select one of (4) conductivity calibration standard values depending o range and normalization temperature being used (TA
BLE 1).
If you only use calibration standards that are listed in TABLE 1, automatic calibration is recommended. If you intend to calibrate with one or more standards that are not listed in
4
5
The factory default is automatic conductivity calibration. See Section
6.10 to change this setting. There is no automatic calibration value
available for conductivity range 1.
TABLE 1
Automatic Calibration Values
Normalization Temperature
Range
#
Conductivity
Range
25 ºC 20 ºC
r 1 0.00 – 20.00 μS None None r 2 20.1 – 200.0 μS 84 μS 76 μS r 3 201 – 2000 μS 1413 μS 1278 μS r 4 2.01 – 20.00 mS 12.88 mS 11.67 mS r 5 20.1 – 200.0 mS 111.8 mS 102.1 mS
Range
#
TDS Range
(using 0.5 TDS factor)
Automatic Calibration
Values
r 1 0 – 10.00 ppm none r 2 10.1 – 100.0 ppm
none
r 3 101 – 1000 ppm
none
r 4 1.01 – 10.00 ppt
none
r 5 10.1 – 100 ppt
none
4.2 Single or Multi-Point Calibration
Use Single-Point Calibration to apply a single calibration value across all ranges.
Use Multi-Point Calibration for individual calib r ation in each range. This w
ill restrict an individual calibration so that it is applied to on e rang e only. When using multi-point calibration, perform a calibration in each range that you expect to use for best results.
The factory default is Single-Point Calibration. See Section 6.11 to change this setting.
4.3 General Calibration Tips
For best results, periodic calibration with known accurate standards is recomme
nded. A maximum of one calibration point per range can be performed. If multiple calibration points are used in the same range, the most recent one will replace the previous one.
When the electrode is replaced, it is best to clear the calibration to the factor
y default values (see Section 6.13).
Rinse or immerse the probe before calibration and between samples w
ith clean water (deionized water is ideal).
The CON 700 has non-volatile memory which will retain all calibration valu
es, as well as meter settings and memory values upon meter shut
down or unexpected power loss. To protect from erroneous calibrations, the allowable tolerance is ±40%
of the factory default value. Low conductivity standard solutions (less than 20 µS) are unstable and
are ver
y temperature dependent. As a result, reproducible calibration results are challenging in lowest measurement range #1 (0.00 to 20.0 µS).
4.4 Automatic Conductivity Calibration Procedure
1) Press
as needed to select conductivity (μS or mS).
2) Dip the electrode into the conductivity standard and press .
Provide stirring for best results.
3) The primary display will show the factor
y default value, while the secondary display will lock on the appropriate automatic standard value from TABLE 1.
4) When the RE
ADY indicator appears, press
to accept. The primary reading will flash briefly before returning to measurement mode upon successful calibration.
6
4.5 Manual Conductivity & TDS Calibration Procedure
1) Press
as needed to select conductivity (μS or mS) or TDS
(ppm or ppt) calibration.
2) Dip the electrode into the calibration standard and press
.
Provide stirring for best results.
3) The primary display will show the current readin g, while the secondary display will be the factory default value. Adjust the
temperature reading using
or . Press to accept. The primary reading will flash briefly before returning to measurement mode upon successful calibration.
4.6 Temperature Calibration
The thermistor sensor used for automatic te
mperature compensation and measurement is accurate and stable, so frequent calibration isn’t required. Temperature calibration is recommended upon probe replacement, whenever the temperature reading is suspect, or if matching against a certified thermometer is desired.
1) Place the probe into a solution with a known accurate temperature
such as a constant temperature bath.
2) Press
followed by . The primary display shows the measured temperature while the secondary display shows the factory default temperature.
4) Adjust the tem
perature using
or . Press to accept or
to cancel. The meter allows an adjustable maximum value of ± 5
º
C
(or ± 9
º
F) from the factory default temperature.
5. MEASUREMENT
5.1 Taking Measurements
1) Rinse the electrode with de-ionized or distilled water before use to remove any impurities. Gently shake excess water droplets.
7
2) Dip the probe into the sample beyond the upper steel band (utilize the fill line on the outside of the probe guard for reference).
3) Allow time for the reading to stabilize. Note the reading on the display. The clear yellow protective probe guard must be attached during measurement. Erroneous results will occur while the probe guard is removed.
5.2 Automatic and Manual Ranging
The CON 700 automatically selects the optimum range in which your readings appear. Refer to TABLE 1 for a list of the available ranges.
To turn-off automatic ranging, press
. MEAS will flash, indicating
that manual ranging is active. To manually select the next range, press
again. After range 5, automatic ranging is resumed.
For example, a reading of 465 μS will automatically settle in range 3. Using manual range advancement this will read as 0.47 mS in range 4, and 0.5 mS in range 5. For best resolution, use auto-ranging.
5.3 HOLD Function
For prolonged observation of a reading, press
during measurement mode to freeze the display. The “HOLD” indicator will display when the reading is held. To release the held value and
resume measurement, press
again or insert the held value into
memory by pressing
.
5.4 Storing and Recalling Data
The CON 700 can retain up to 100 data points into memory for later retrieval.
1) In the measurement mode, press
to insert the measured value into memory. The stored memory location value (StO) is briefly displayed.
8
2) To recall data from memory, press . The location of the most recent stored data is displayed first. Press
or to select the
location of the desired data, then press
to accept.
3) Press
to return to the stored data location. Press to return
to measurement mode. To erase stored data, see Section 6.14.
6. SETUP FUNCTIONS
Use the set up mode to customize your instrument operation. During measurement, press and hold
for 5 seconds to enter SETUP
mode. Press or to change programs or change options. Press
to select the program or confirm selection.
Press
to go back one level or return to measurement mode.
6.1 P1.0 CAL (Calibration)
Press
to view stored calibration points in each range.
6.2 P2.0 ELE (Electrode Information)
Press
to view cell efficiency in each range.
6.3 P3.0 ConF (Configuration)
Press
to access set-up programs 3.1 thru 3.9.
9
6.4 P3.1 rdY (Ready / Stability Indicator)
Press
.
Press
or to choose READY “On”, READY “OFF”, or
Auto HOLd.
Press to confirm.
6.5 P3.2 ºC ºF (Celcius
Or Fahrenheit)
Press
.
Press
or to select ºC or ºF.
Press
to confirm.
6.6 P3.3 AtC (Automatic Temperature Compensation)
Press
.
Press
or to select “Yes” or “No”.
Press
to confirm.
6.7 P3.4 tdS (Total Dissolved Solids factor)
Press
.
Press
or to select the desired TDS factor (.40 to
1.00). Press
to confirm.
6.8 P3.5 t.CO (Temperature Coefficient)
The temperature coefficient is the amount of change in conductivity per degr
ee temperature (% per ºC). The CON 700 is factory set to a temperature coefficient of 2.1 % per ºC. For most applications this will provide good results. The meter allows adjustment from 0.0 to 10.0.
TIP: Select 0.0% for uncompensated measurements. The temperature w
ill be measured by the electrode and displayed in measurement
mode—without compensation.
See Section 8
– Calculating Temperature Coefficients.
10
Press . Press
or to select the desired value (0.00 to 10.0).
Press
to confirm.
6.9 P3.6 t.nr (Normalization Temperature in ºC)
When Automatic Temperature Compensation is used, measurements are ad
justed by the temperature coefficient to the normalization
temperature. The default value is 25 ºC.
Press
.
Press
or to select the desired value (15.0 to 30.0).
Press
to confirm.
6.10 P3.7 ACAL (Automatic Conductivity Calibration)
See Section 4.1 for more details on Automatic Conductivity Calibration.
.
Press Press
or to select “Yes” (Automatic) or “No” (Manual).
Press
to confirm.
6.11 P3.8 SPC (Single Point Calibration)
See Section 4.2 for details on Single Poi
nt & Multi Point Calibration.
.
Press Press
or to select “Yes” (Single) or “No” (Multi).
Press
to confirm.
6.12 P3.9 CELL (Nominal Cell C
onstant)
The CON 700 includes a probe with a nominal cell constant (k) of 1.0. Use probes with k = 0.1 and 10 (sold separately) for improved
11
performance in extreme sample ranges. Use this setup function to change the cell constant if necessary. Meter default is 1.0 to match the included probe. k = 0.1 ideal for low measurements <20 µS (<10 ppm). k = 1.0 ideal for mid-range measurements k = 10 ideal for high measurements >20 mS (>10 ppt).
Press
.
Press
or to select 0.1, 1.0, or 10.0
Press
to confirm.
6.13 P4.0 rSt (Reset)
Press
.
Press
or to select “Yes” (Reset) or “No” (Cancel).
If “Yes”, press
or to select “Cal” (calibration reset
only) or “FCt” (complete reset to factory default settings). Press
to confirm.
6.14 P5.0 CLr (Clear Memory)
Press
.
Press
or to select “Yes” (Erase memory) or “No”.
Press
to confirm.
12
13
7. CALCULATING TDS CONVERSION FACTOR
You can calibrate TDS using the value of the calibration standard sol utio n at a standard temperature such as 25 ºC. To determine the conductivity­to-TDS conversion factor use the following formula:
Factor = Actual TDS ÷ Actual Conductivity @ 25 ºC
Actual TDS: Value from the solution bottle label or as a standard made using high purity water and precisely weighed salts. ppm = milligram of salt(s) per liter of water = mg/L ppt = gram of salt(s) per liter of water = g/L
Actual Conductivity: Value
measured using a properly
calibrated Conductivity/Temperature meter.
Both the Actual TDS and the Actual Cond
uctivity values must be in the same magnitude of units. For example, if the TDS value is in ppm the conductivity value must be in µS; if the TDS value is in ppt the conductivity value must be in mS.
Check your factor by multiplying the co
nductivity reading by the factor in
the above formula. The result should be in TDS value. When the TDS factor is set to 1.0, Conductivity = TDS.
8. CALCULATING TEMPERATURE COEFFICIENTS
To determine the temperature coefficient of your sample solution use this formula:
Where: tc = Temperature coefficient 25 = 25 ºC C
T1
= Conductivity at Temp 1 CT2 = Conductivity at Temp 2 T1 = Te mp 1 T2 = Temp 2 NOTE: A controlled temperature water bath is ideal for this procedure.
1. Immerse the probe into a sample of your solution and adjust the
temperatur
e coefficient to 0% (that is, no compensation) by following
instructions as described in Section 6.8.
2. W ait for 5 minu tes. Note T1 and CT1 (conductivity at T1).
3. Condition the sample solution and probe to a temperature ( T2) that is
about 5 ºC to 10 ºC different from T1, and note the conductivity reading C
T2
.
NOTE: Record your results for future referen
ce. Ideally T1 and T2 should bracket your measurement temperature, and should not differ b y more than 5 ºC.
4. Calculate the temper ature coefficient of you r solution acc ording to the formula sh
own above.
5. Enter the calculated temperature coefficient into the meter.
The calculated temperature coefficient will now be applied to all the meter readings.
14
15
9. REPLACEMENTS AND ACCESSORIES
Part number Ordering Code
Item Description
Eutech Instruments Oakton Instruments
CON 700 Benchtop with electrode and integral stand,100/240 VAC
ECCON70043S 35411-00
Replacement electrode, k = 1.0 CONSEN9501D 35608-74 Epoxy/platinum electrode, k = 0.1 93X546101 35608-72 Glass/platinum electrode, k = 1.0 - 35608-76 Epoxy/platinum electrode, k = 10.0 93X546102 35608-78 SMPS, 100/240 VAC, 9V, 6W power adapter 60X030130 35615-07 10 μS, (20) x 20mL Sachets ECCON10BS 35653-09 84 µS, 480 mL bottle* ECCON84BT 00653-16 100 μS, 480 mL bottle ECCON100BT ­447 μS, 480 mL bottle ECCON447BS 00653-47 500 μS, 480 mL bottle ECCON500BT ­1413 μS, 480 mL bottle* ECCON1413BT 00653-18 1413 μS, (20) x 20mL Sachets* ECCON1413BS 35653-11 2764 μS, 480 mL bottle ECCON2764BT 00653-20 2764 μS, (20) x 20mL Sachets ECCON2764BS 35653-12
5.0 mS, 480 mL bottle ECCON5000BT -
12.88 mS, 480 mL bottle* ECCON1288BT 00606-10 15 mS, (20) x 20mL Sachets ECCON15000BS 35653-13
111.8 mS, 480 mL bottle* ECCON1118BT -
* Calibration Standard Value at 25 ºC
10. TROUBLESHOOTING GUIDE
PROBLEM CAUSE SOLUTION
No display Main power not switched
on. AC Adapter socket not inserted properly.
a) Switch on the power supply. b) Re-insert AC Adapter.
“Ur” (Under range)
“Or” (Over range)
in primary or secondar
y
display
Measured value is out of range.
Electrodes not connected. Electrode clogged, dirty
or broken. Meter not calibrated. Wrong temp value.
Check electrode is connected. Clean or replace electrode. Recalibrate the meter. Confirm measurement condition. Treat samples to bring within
meter measuring
range.
Reset meter.
Slow response
Dirty electrode. Temperature of sample is
changing.
Clean the electrode. Allow temperature to stabilize.
Meter not responding to key press
Manual HOLD or Auto HOLD is active.
Worn keypad.
Press
to deactivate manual HOLD. See Section 6.4 to disable Auto Hold feature if enabled.
Contact Technical Service.
Invalid key; Button is not functional in the current operation mod
e.
Press alternate key. Select valid key depending on current mode.
Secondary display continually scrolls
Conductivity calibration standard is not
within 40% of expected value during automatic calibration.
Ensure electrode guard has not been re
moved.
Use fresh or different calibration standard.
16
17
11. SPECIFICATIONS
Conductivity Range
..µS to 200.0 mS
Conductivitity Resolution
0.01 / 0.1 / 1 μS ;
0.01 / 0.1 mS
TDS Range .. to 100.0 ppt @ 0.5 fact (200.0 @ 1 factor) Resolution 0.01 / 0.1 / 1 ppm ; 0.01 / 0.1 ppt Accuracy ±1% Full Scale Cell Constant 0.1, 1.0, 10.0 (selectable) TDS Factor 0.40 to 1.00 (adjustable)
Calibration points
Automatic (4 points); Maximum 1 per range
Manual (5 points
); Maximum 1 per range
Temperature
Range (Meter)
0.0 to 100.0 ºC / 32.0 to 212.0 ºF
Compensation Type & Range
Automatic with supplied cell or Manual.
0.0 to 100 ºC / 32.0 to 212.0 ºF,
0.0 to 80 ºC / 32.0 to 176.0 ºF with supplied cell
Resolution
0.1 ºC / 0.1 ºF
Accuracy
± 0.5ºC / ± 0.9ºF
Calibration
Offset in 0.1 º increments; Offset range: ± 5.0 ºC / ± 9.0 ºF
Coefficient
0.0 to 10.00 per ºC Auto-ranging Yes Hold Function Yes Memory 100 data sets Input 8-pin DIN Power AC/DC 9V, 6W Adapter (100/240 VAC, 50-60Hz) Dimensions (mm) meter only = 175 (L) x 155 (W) x 69 (H)
18
12. WARRANTY
This meter is supplied with a warranty against significant deviations in material and workmanship for a period of THREE years from date of purchase whereas probe with a SIX month warranty.
If repair or adjustment is necessary a
nd has not been the result of abuse or misuse within the designated period, please return – freight prepaid – and correction will be made without charge. Eutech Instruments/Oakton Instruments will determine if the product problem is due to deviations or customer misuse.
Out of warranty products will be repaired on a charged basis. The warranty on your instrument shall not ap
ply to defects resulting from:
Improper or inadequate maintenance by customer
Unauthorized modification or misuse
Operation outside of the environment specifications of
the products
13. RETURN OF ITEMS
Authorization must be obtained from our Customer Service Department or authorized distributor before returning items for any reason. A “Return Material Authorization” (RMA) form is available through our authorized distributor. Please include data regarding the reason the items are to be returned. For your protection, items must be carefully packed to prevent damage in shipment and insured against possible damage or loss. Eutech Instruments will not be responsible for damage resulting from careless or insufficient packing. A restocking charge will be made on all unauthorized returns.
NOTE: Eutech Instruments Pte Ltd /Oakton Instruments res
erve the right to make improvements in design, construction, and appearance of products without notice.
For more information on our products, please contact our channel partner or visit our websites listed below:
Eutech Instruments Pte Ltd
Blk 55, Ayer Rajah Crescent, #04-16/24 Singapore 139949 Tel: (65) 6778 6876 Fax: (65) 6773 0836
eutech@thermofisher.com www.eutechinst.com
Oakton Instruments
625 E Bunker Court Verno
n Hills, IL 60061 USA Tel: (1) 888-462-5866 Fax: (1) 847-247-2984
info@4oakton.com www.4oakton.com
Distributed by:
Part of Thermo Fisher Scientific
Loading...