This manual serves to explain the use of the Waterproof CON 400/410 hand-held meters. It
functions in two ways, firstly as a step by step guide to help you to operate the meter.
Secondly, it serves as a handy reference guide. It is written to cover as many anticipated
applications of the Waterproof CON 400/410 meters as possible. If there are doubts in the
use of the meter, please do not hesitate to contact the nearest Authorised Distributor.
Eutech Instruments/ Oakton Instruments cannot accept any responsibility for damage or
malfunction to the meter caused by improper use of the instrument.
The information presented in this manual is subject to change without notice as improvements
are made, and does not represent a commitment on the part of Eutech Instruments Pte Lt d/
Oakton Instruments.
18 RETURN OF ITEMS...................................................................................50
Instruction Manual CON 400/410
1 INTRODUCTION
Thank you for selecting the CON 400/410 waterproof portable meter. This meter is a
microprocessor-based instrument that is designed to be user-friendly and allow one-hand
operation. It has a built-in real time clock, expanded memory, and many other user-friendl y
features, all of which are accessible through the membrane keypad. It is completely
WATERPROOF --- and it FLOATS! You have one of the following models:
• CON 400 meter: reads conductivity and temperature.
• CON 410 meter: reads conductivity, TDS and temperature.
Your meter includes 4 ‘AAA’ batteries and a conductivity electrode (cell constant K = 1.0) with
built-in temperature sensor. Please read this manual thoroughly before operating your meter.
- 1 -
Instruction Manual CON 400/410
2 DISPLAY AND KEYPAD FUNCTIONS
2.1 Display
The LCD has a primary and secondary display.
• The primary display shows the measured conductivity or TDS reading.
• The secondary display shows the measured temperature.
The display also shows error messages, keypad functions and program functions.
See Figure 1.
Primary Display
12
3
4
SETUPMEASCALMEM
5
6
7
9
19
20
18
17
READY
HOLD
ON
OFF
K =
-8.8.8.8
MEM
ERR
mS
µ
S
ppt
ppm
°C °F
8
-1.8.8.8
16
1. SETup mode indicator 8. parts per million indicator (CON
7. parts per thousand indicator
(CON 410 meter only)
AM PM
12
131415
Figure 1: Full LCD Screen
410 meter only)
Compensation indicator
14. Low battery indicator
Secondary Display
ATC
1011
15. Probe indicator
17. Cell constant indicator
- 2 -
Instruction Manual CON 400/410
2.2 Keypad
The large membrane keypad makes the instrument easy to use. Each button, when pressed,
has a corresponding graphic indicator on the LCD. See Figure 2. Some buttons have several
functions depending on its mode of operation.
Key Function
ON/OFF Powers on and shuts off the meter. When you switch on the meter, the meter starts up
HOLD Freezes th e m easured reading. To activate, press HOLD while in measurement mode.
MODE Selects the measurement parameter.
CAL/MEAS Toggles between Calibration and Measurement mode.
ENTER /
RANGE
MI/S and
MR/T
SETUP Takes you into the SETUP mode. This mode lets you customise meter preference and
in the mode that you last switched off from. For example, if you shut the meter off in
TDS measurement mode (only in CON 410 meter), the meter will be in TDS
measurement mode when you switch the meter on.
To release, press HOLD again.
~ CON 400 meter: Toggles between conductivity and time.
~ CON 410 meter: Toggles between conductivity, TDS and time.
NOTE: Temperature calibration is available from conductivity calibration mode; see
page 9 for directions.
ENTER function: Press to confirm values in Calibration mode and to confirm selections
in SETUP mode.
RANGE function: Press to enter manual ranging function.
The MEAS indicator blinks while in manual ranging function.
In Measurement mode:
Press MI/S (memory input) to store values with its corresponding temperature values in
the memory. Press MR/T (memory recall) to retrieve data from memory.
In Calibration mode:
Press to scroll through calibration values.
In SETUP mode:
Press to scroll through the setup subgroup programmes.
defaults, and view calibration, electrode offset data and select cell constant.
Figure 2 - Keypad
- 3 -
Instruction Manual CON 400/410
3 PREPARATION
3.1 Inse rting the Batteries
Four AAA batteries are included with your meter.
1. Use a Philips screwdriver to remove the two screws holding the battery cover. See Figure
3 below.
2. Remove battery cover to expose batteries.
3. Insert batteries. Follow the diagram inside the cover for correct polarity.
4. Replace the battery cover into its original position using the two screws removed earlier.
Figure 3 - Back panel of meter showing meter compartment
- 4 -
Instruction Manual CON 400/410
3.2 Conducti vity Electrode Information
The CON400/410 meter uses a conductivity / TDS cell with a
sturdy 6-pin connector.
Your meter includes a conductivity probe (Part No: ECCONSEN91W / 35608-50) Ultem / Stainless Steel cells with a cell
constant of K = 1.0. This conductivity / TDS cell features a built-in
temperature sensor for Automatic Temperature Compensation
(ATC). It has a specially designed housing that provides fast
temperature response and reduces air bubble entrapment, which
makes it easy to obtain accurate, stable readings.
Wetted parts include:
1. Polyetherimide (Ultem™)
2. Polybutylterphalate (Valox™)
3. Stainless Steel (SS 304)
NOTE: We recommend that you do not submerge the probe above
the protective yellow cap for prolonged periods. The protective
yellow cap can be removed for cleaning, but must be attached at
all times during calibration or measurment.
See Section 8 on page 40 for “Probe Care and Maintenance”
information.
1
Figure 4 - Conductivity probe
(EC-CONSEN91W/ 35608-50)
1
Valox and Ultem are trademark names of General Electric Co.
- 5 -
Instruction Manual CON 400/410
3.3 Connecting the p robe to the meter
1. Line up the notch and 6 pins on the meter with the holes in the 6-pin connector. Push
down and turn the locking ring clockwise to lock into place. See figure below.
2. To remove probe, turn the locking ring counterclockwise on the probe connector. Pull
probe away from the meter.
CAUTION: Do not pull on the probe cord or the probe wires might disconnect.
NOTE: Keep connectors clean. Do not touch connector with soiled hands.
Figure 5 - Connection for Conductivity probe (6-pin connector)
- 6 -
Instruction Manual CON 400/410
4 CALIBRATION
4.1 Imp ortant Information on Meter Calibrat ion
Your meter has five measuring ranges. You can calibrate one point in each of the measuring
ranges (up to five points). If you are measuring values in more than one range, make sure to
calibrate each of the ranges you are measuring.
To view current calibration points, see SETUP section Program 2.0 on page 28.
The following table lists the corresponding conductivity and TDS ranges. You should calibrate
each range using a solution that falls between the values in the “recommended calibration
solution range” column.
Conductivity
Measuring
Range
0.00 Æ 19.99 µS 6.00 to 17.00 µS 0.00 Æ 9.99 ppm 3.00 to 8.50 ppm
0.0 Æ 199.9 µS 60.0 to 170.0 µS 10.0 Æ 99.9 ppm 30.0 to 85.0 ppm
0 Æ 1999 µS 600 to 1700 µS 100 Æ 999 ppm 300 to 850 ppm
0.00 Æ 19.99 mS 6.00 to 17.00 mS 1.00 Æ 9.99 ppt 3.00 to 8.50 ppt
0.0 Æ 199.9 mS 60.0 to 170.0 mS 10.0 Æ 200 ppt 30.0 to 170 ppt
When you recalibrate your meter, old calibrations are replaced on a range by range basis. For
example, if you previously calibrated your conductivity meter at 1413 µS in the 0 to 1999 µS
range and you recalibrate at 1500 µS (also in the 0 to 1999 µS range), the meter will replace
the old calibration data (1413 µS) in that range. The meter will retain all calibration data in
other ranges.
To completely recalibrate your meter, or when you use a replacement probe, it is best to clear
all calibration data in memory. To erase all the old conductivity and TDS calibration data
completely from memory, see SETUP section Program 8.0 on page 39.
For information on how to calibrate your meter:
• See section 4.3 for Temperature Calibration.
• See section 4.4 for Conductivity
• See section 4.5 for TDS Calibration (CON 410 meter only).
• See Addendum 1 for more calibration tips.
Recommended
Calibration Solution
Range
TDS Measuring
Range
Recommended
Calibration Solution
Range
- 7 -
Instruction Manual CON 400/410
4.2 Prep aring the Meter for Calib ration
Before starting calibration, make sure you are in the correct measurement mode. When you
switch on the meter, the meter begins with the units you shut it off in.
For best results, select a standard value close to the sample value you are measuring.
Alternatively use a calibration solution value that is approximately 2/3 the full-scale value of
the measurement range you plan to use. For example, in the 0 to 1999 µS conductivity range,
use a 1413 µS solution for calibration.
Do not reuse calibration solutions after calibration. Contaminants in the solution can affect the
calibration, and eventually the accuracy of the measurements. Use fresh calibration solution
each time you calibrate your meter.
NOTE: Your meter is factory set to a temperature coefficient of 2.1% per °C. For most
applications this will provide good results. See Program P5.1 on page 34 to set the
temperature coefficient to different value. See Addendum 3, “Calculating Temperature
Coefficients” to determine the appropriate temperature coefficient for your solution.
NOTE: The factory default value for normalization temperature is 25 °C. If you need to
normalize to a value other than 25 °C, see Program P5.2 on page 35.
- 8 -
Instruction Manual CON 400/410
4.3 Te mperature Calibration
Your probe features a built-in temperature sensor. The
temperature sensor is factory calibrated. Calibrate your
sensor only if you suspect temperature errors that may
have occurred over a long period of time or if you have a
replacement probe.
Temperature Calibration
1. Make sure the cell is attached to the 6-pin connector.
The ATC annunciator will appear at the right-hand
side of the LCD*.
2. Switch the meter on. Press the MODE key to select
conductivity or TDS mode.
3. Press the CAL/MEAS key to enter conductivity or
TDS calibration mode. The CAL indicator appears
above the primary display.
4. While in conductivity or TDS calibration mode, press
the MODE key to enter temperature calibration
mode. The primary display shows the current
temperature reading and the secondary display
shows the factory default temperature value.
5. Dip the cell into a solution of known temperature (i.e.
A temperature bath). Allow time for the built-in
temperature sensor to stabilise.
6. Scroll with the MI/S or MR/Tkeys to set the correct
temperature value (i.e. the temperature of the
temperature bath). You can adjust the reading in
increments of 0.1 °C.
7. Once you have selected the correct temperature,
press the ENTER key.
* If the ATC indicator does not light up, see Program P4.3
to switch it on.
NOTES:
•You can offset the temperature reading up to ±5°C from
the original reading (default reading).
•To exit this program without confirming the temperature
calibration value, DO NOT press ENTER. Press
CAL/MEAS instead.
•If the ATC indicator does not light, see Prog ram P4.3,
on page 32 to switch it on.
•Since temperature readings affect the accuracy of the
conductivity/TDS measurements, it is strongly
recommended to carry out a conductivity/TDS
calibration after a temperature calibrati on is done.
- 9 -
MEAS
1413
L
A
C
S
A
E
M
CAL
1413
E
D
O
M
22.3
I
M
22.0
R
E
T
N
E
E
G
N
A
R
MEAS
READY
1425
Figure 10 - Temperature
Calibration
22.3
22.3
CAL
22.3
CAL
22.3
22.0
µ
S
°C
ATC
µ
S
°C
ATC
°C
ATC
R
M
°C
ATC
µ
S
°C
ATC
Instruction Manual CON 400/410
4.4 Conductivity Calibration
The CON 400 and CON 410 meters are capable of up
to 5-point conductivity calibration at one point per
conductivity range (0.00 - 19.99 µS; 0.0 - 199.9 µS; 0 1999 µS; 0.00 - 19.99 mS; 0.0 - 199.9 mS).
All new calibration data will over-ride existing stored
calibration data for each measuring range calibrated.
Calibrating for Conductivity:
1. If necessary, press the MODE key to select
conductivity mode.
2. Rinse the probe thoroughly with de-ionized (DI)
water or a rinse solution, then rinse with a small
amount of calibration standard.
3. With the yellow probe guard attached, dip the
probe into the calibration standard. Immerse the
probe tip beyond the upper steel band. Stir the
probe gently to create a homogeneous sample.
Allow time for the reading to stabilizse.
See Figure 7.
4. Press CAL/MEAS to enter conductivity
calibration mode. The CAL indicator will appear
in the upper right corner of the display.
5. Press the MI/S or MR/T key to change the
value on the primary display to match the value
of the calibration standard referenced to your
normalisation temperature (usually 25°C).
6. Press ENTER to confirm calibration value. The
meter returns to the MEAS (measurement) mode.
7. Repeat steps 1 to 6 for other measuring ranges.
See figure 6.
Figure 6 - Conductivity Calibration
- 10 -
Instruction Manual CON 400/410
NOTES:
When entering calibration mode, the meter
displays the factory default value. If the meter
was previously calibrated, the display may
seem to “jump” to the factory default value
when switching from measurement to
calibration mode. This is expected.
To exit from Conductivity calibration
mode without confirming calibration, DO
NOT press the ENTER key in step 6. Press
CAL/MEAS instead. This will retain the
meter’s old calibration data in the measuring
range of the calibration.
You can offset the conductivity reading up to
±40% from factory default value. If your
measured value differs by more than ±40%,
clean or replace probe if needed.
Figure 7 - Proper Immersion of
the conductivity probe
The minimum offset reading allowable is
limited to 10% of full scale reading of the
range you are working in. The maximum offset reading allowed is the range full scale reading
Eutech Instruments/ Oakton Instruments offers a wide selection of high-quality calibration
standards. See page 44 for more information.
- 11 -
Instruction Manual CON 400/410
4.5 TDS Calibration
4.5.1 Calibrating for TDS directly
For CON 410 meter only
The factory default setting for the TDS conversion factor is
0.50. If your solution has a different TDS factor, you can
improve calibration accuracy by setting the TDS factor prior to
calibration. See page 33 for directions.
1. If necessary, press the MODE key to select TDS mode.
2. Rinse the probe thoroughly with DI water or a rinse
solution, then rinse with a small amount of calibration
standard.
3. With the yellow probe guard attached, dip the probe into
the calibration standard. Immerse the probe tip beyond
the upper steel band. Stir the probe gently to create a
homogeneous sample. Allow time for the reading to
stabilise.
4. Press CAL/MEAS to enter TDS calibration mode. The
CAL indicator will appear in the upper right corner of the
display.
5. Press the MI/S or MR/T to change the value on the
primary display to match the value of the calibration
standard referenced to your normalisation temperature
(usually 25°C).
6. Press ENTER to confirm the calibration value. The meter
returns to the MEAS (measurement) mode.
7. Repeat steps 1 to 6 for other measuring ranges.
NOTES
To exit from TDS Calibration mode without confirming
calibration, DO NOT press the ENTER key in step 6. Press
CAL/MEAS instead. This will retain the meter’s old calibration
data in the measuring range of the calibration.
You can offset the TDS reading up to ±40% from the default
setting. If your measured value differs by more than ±40%,
clean or replace probe as needed.
The minimum offset reading allowable is limited to 10% of full scale reading of the range you
are working in. The maximum offset reading allowed is the range full scale reading
Eutech Instruments/ Oakton Instruments offer a wide selection of high-quality calibration
standards. See page 44 for more information.
Figure 8 - TDS Calibration
- 12 -
Instruction Manual CON 400/410
4.5.2 Calibration with Conductivity Standard and TDS factor
For CON 410 meter only
The concentration of salts dissolved in solution increases the conductivity of that solution. This
relationship varies from salt to salt and is roughly linear over a given range for a given salt.
The TDS conversion factor is the number used by the meter to convert f rom conductivity to
TDS.
Instead of calibrating for TDS directly (described in Section 4.5.1), you can calibrate the CON
410 meter by:
1. calibrating to conductivity standards (as described on pages 10 to 11) and then
2. entering the appropriate TDS conversion factor into the meter.
To determine the conductivity to TDS conversion factor for your solution:
• Addendum 2 explains the formula to calculate the conductivity-to-TDS
conversion factor.
Enter the TDS conversion factor into your meter as described under Section 7.5.4, in P4.4 Setting the TDS Factor on page 33.
- 13 -
Instruction Manual CON 400/410
From measurement mode
1. Press Setup key to enter Set Up mode.
2. Press the S and T keys to scroll through subgroups
until you view parameter P4.0.
SETUP
cof
p 4.0
See Figure 9.
I
M
R
M
3. Press the ENTER key again. The upper display shows
a value and the lower display shows “tdS”.
4. Calculate the TDS factor of your solution. See
SETUP
ppt
ppm
Addendum 2 on page 46 for information on how to
calculate the TDS factor.
p 4.4
5. Press the S and T keys to select your calculated TDS
M
R
M
I
conversion factor.
6. Press the ENTER key to confirm selection and to
return to the subgroup menu. Press the CAL/MEAS
SETUP
0.50
ppt
ppm
key to return to measurement mode.
R
E
T
N
E
E
G
N
A
R
Figure 9 - TDS Conversion Factor
- 14 -
Instruction Manual CON 400/410
5 MEASUREMENT
This meter is capable of taking measurements with automatic or manual temperature
compensation. Factory default is set with automatic temperature compensation (ATC) on.
5.1 Automatic Temperature Compensation
For automatic temperature compensation (ATC) simply
plug the conductivity/TDS probe into the meter (see page
6 for directions). The ATC indicator will light on the LCD.
NOTE: If the ATC indicator does not light, manual
temperature compensation may be selected in
the meter’s SETUP mode. See Program P4.3 on
page 32 for directions on selecting Automatic
Temperature Compensation.
Figure 11 - ATC annunciator will
light up when connected to
temperature probe
- 15 -
Instruction Manual CON 400/410
5.2 Manual Temperature Compensation
IMPORTANT: For manual compensation, you must deactivate the automatic temperature
compensation feature.
5.2.1 Selecting Manual Temperature
Compensation
To select between Automatic Temperature Compensation
(ATC) and Manual Temperature Compensation in the
SETUP program P4.3. Meter default is ATC on.
P4.0 Configuration Setup
From measurement mode
1) Press SETUP key to enter Set Up mode.
2) Press the MI/S or MR/T keys to scroll through
subgroups until you view parameter P4.0.
3) Press the ENTER key three times to select parameter
4.3. The upper display shows “ATC” and the lower
display shows “P4.3”.
4) Press the ENTER key again. The upper display shows
“ATC” and the lower display shows “YES” or “NO”.
5) Press the MI/S or MR/T to select between Automatic
or Manual Temperature Compensation feature.
• YES = Automatic Temperature Compensation
activated.
• NO = Manual Temperature Compensation
activated.
6) Press the ENTER key to confirm selection and to
return to the subgroup menu. Press the CAL/MEAS
key to return to measurement mode.
NOTE:
Manual Temperature Compensation should only be
selected if the ATC fails. For non-compensated
measurements, change the temperature coefficient to
0.0% (see section 7.6.1)
Selecting Automatic or Manual
Temperature Compensation
- 16 -
Instruction Manual CON 400/410
5.2.2 Setting a manual temperature compensation value
To use manual temperature compensation, you need to
manually enter the temperature value of your process into
the meter. You can select any temperature between 0 and
100 °C (32 to 212 °F). Default value is 25 °C.
To select a manual temperature compensation value
1. Switch the meter on. Press the MODE key to select
measurement mode.
2. If necessary, select ATC off as described in section
5.2.1 on page 16. The ATC indicator will not appear
on the display.
3. Press the CAL/MEAS key to enter conductivity or TDS
calibration mode. The CAL indicator will appear above
the primary display.
4. While in conductivity or TDS calibration mode, press
the MODE key to enter temperature calibration mode.
The primary display shows the current temperature
setting and the secondary display shows the default
value 25 °C (77 °F) or its last set value.
5. Check the temperature of your sample using an
accurate thermometer.
6. Press the MI/S or MR/T keys to offset the
temperature to the measured value from step 5.
7. Press ENTER to confirm the selected temperature and
to return to the conductivity or TDS measurement
mode.
The meter will now compensate conductivity or TDS readings for manually set temperature.
NOTES:
To exit this program without confirming the manual temperature compensation value, DO NOT
press ENTER in step 7. Press CAL/MEAS instead.
Figure 9 - Manual Temperature
Compensation
- 17 -
Instruction Manual CON 400/410
5.3 Taking Measurements
To take readings:
1. Rinse the probe with DI or distilled water before use
to remove any impurities adhering to the probe
body. Shake or air dry. To avoid contamination or
dilution of your sample, rinse probe with a small
volume of your sample liquid.
2. Press ON to switch on meter and the MEAS
annunciator appears on the top center of the LCD.
3. Dip the probe into the sample.
When dipping the probe into the sample, take care to ensure that the liquid
level is above its upper steel band. Stir the probe gently in the sa mple to
create a homogenous sample.
4. Allow time for the reading to stabilise. Note the reading on the display.
5. CON 410 meter only
: Press the MODE key to toggle between conductivity and TDS
readings.
Taking measurements with READY indicator selected on
If the READY indicator has been activated, the READY annunciator lights when the reading is
stable. Switch the READY indicator on or off in SETUP program P 4.1. See page 30 for
directions.
Taking measurements with the auto endpoint feature selected on
When a reading is stable for more than 5 seconds, the auto endpoint feature will automatically
“hold” the reading. The “hold” indicator appears on the left side of the display. Press the HOLD
key to release the reading. Switch the Auto endpoint feature on or off in SETUP program P 4.1,
see page 30 for instructions.
Figure 10 - During measurement
- 18 -
Instruction Manual CON 400/410
5.4 Using Manu al Ranging Function
Your meter automatically selects the range in which your readings appear. The manual
ranging function lets you select the specific range (and corresponding resolution) that you
want to work in:
CON 400 and CON 410 meters: CON 410 meter:
1. 0.00 --- 19.99 µS
2. 0.0 --- 199.9 µS
3. 0 --- 1999 µS
4. 0.00 --- 19.99 mS 4. 0.00 --- 9.99 ppt
5. 0.0 --- 199.9 mS 5. 0 --- 100 ppt
1. To select the desired measuring range, press the
RANGE key while in Measurement mode. The first
range will appear on the display and the “MEAS”
indicator blinks.
2. Press the RANGE key again (if needed) until desired
range is selected.
3. To re-select the Auto-ranging function, repeatedly
press the RANGE key until the “MEAS” indicator
appears without blinking.
NOTES
If the value of the solution you are measuring is higher than
the range selected “Or” will appear on the primary display.
Press RANGE until the correct range is selected.
1. 0.00 --- 9.99 ppm
2. 0.0 --- 99.9 ppm
3. 0 --- 999 ppm
(up to 199.9 ppt depending on TDS factor
setting)
Figure 11 - Manual ranging
Figure 12 - Out-of-range
The meter resets to the Auto-ranging function once it is
turned off. You will have reset the manual ranging function
each time you turn the meter off.
- 19 -
Instruction Manual CON 400/410
5.5 HOLD Function
This feature lets you freeze the display for a
delayed observation. HOLD can be used any time
in MEAS mode.
1. To hold a measurement, press the HOLD
key while in measurement mode. “HOLD” will
appear on the display.
2. To release the held value, press the HOLD
again. Continue to take measurements.
NOTE:
• This meter shuts off automatically after 20 minutes of nonuse (without key press).
• If the meter is shut off either automatically or manually, the HOLD value will be lost.
For longer storage, use the memory functions (see pages 21-22).
•Your meter has an auto endpoin t feature. When this feature is switched on, and
when a reading is stable for more than 5 seconds, the display will automatically
“HOLD” the reading. The “HOLD” indicator appears. Press the HOLD key to release
the reading. To switch on or off the auto endpoint feature, see SETUP program P4.1
on page 30.
Figure 13 - HOLD function
- 20 -
Instruction Manual CON 400/410
6 MEMORY AND DATA INPUT FUNCTIONS
6.1 Memory Input
Your meter stores up to 50 sets of data. Data sets
include conductivity, temperature, date and time.
To store a reading:
1. During any measurement function (MEAS), press
MI/S key to input any data into the memory.
2. MEM, “Sto” and memory number will flash. The
meter then returns to measurement mode.
3. If necessary, measure the next sample solution and
press MI/Skey to input the next data into the
memory.
See figure 14.
NOTE: If the memory is full, the first value stored will be
erased to create space for the new value.
- 21 -
Figure 14 - Memory input
Instruction Manual CON 400/410
6.2 Memory Recall
This function recalls the previous readings stored in the
memory. You can access MR from the measurement
mode only. Memory recall is in "Last In First Out" order.
To recall readings:
1. Press the MR/T key once to retrieve the last
reading stored. The memory location screen – MEM, “Loc” and the memory number – will flash on the
display.
2. Press the ENTER key to recall the reading stored
under that memory number.
3. Press ENTER key again to view the date and time
the reading was taken.
4. Press ENTER key again to return to the "memory
location" screen. The display automatically moves
to the next memory location screen.
5. If necessary, press the MI/S or MR/Tkeys to
select the next "memory location" screen; press the
ENTER/RANGE key to select the previous "memory
location" screen.
6. Repeat steps 2 - 5 to review additional stored data
sets.
7. To exit Memory Recall, press MEAS key to return to
Measurement mode.
NOTES:
Readings stored in memory are retained even if the units
are turned off. To erase all readings stored in memory,
use the SETUP mode P1.0 on page 27.
Figure 20 - Memory recall
- 22 -
Instruction Manual CON 400/410
7 ADVANCED SETUP FUNCTIONS
The advanced setup mode lets you customised your meter’s preferences and defaults. The
Waterproof CON400 / CON410 meters features different sub-groups that organise setup
parameters.
The sub-groups are:
1. P1.0: Memory clear (CLr)
2. P2.0: Viewing calibration data (CAL)
3. P3.0: Viewing electrode data (ELE)
4. P4.0: Unit configuration (COF)
5. P5.0: Temperature (tPr)
6. P6.0: Selecting cell constant (CEL)
7. P7.0: Setting clock (CLO)
8. P8.0: Reset to factory defaults (rSt)
See Figure 22 on next page.
- 23 -
Instruction Manual CON 400/410
Figure 22 - Overall view of SETUP programs
- 24 -
Instruction Manual CON 400/410
7.1 Ad vanced SETUP Mode Overview
Press the SETUP key to enter Set up mode.
Press the MI/S or MR/T keys to scroll through sub groups.
Press ENTER key to enter a particular parameter.
See Addendum 4 on page 48 for a table of meter factory default settings.
P1.0: Memory clear
P1.0 Clear all stored readings
P2.0: Viewing previous calibration data
P2.1 First range calibration point
P2.2 Second rang e cal ibration point
P2.3 Third range cal ib ration point
P2.4 Fourth range calibration point
P2.5 Fifth range calibration point
P3.0: Viewing electrode data
P3.1 Effective cell constant for first range
P3.2 Effective cell constant for second range
P3.3 Effective cell constant for third range
P3.4 Effective cell constant for fourth range
P3.5 Effective cell constant for fifth range
P4.0: Unit configuration
P4.1 READY indicator and auto endpoint function –
select on or off
P4.2 Select °C or °F
P4.3 Select Automatic or Manual Temperature
P5.1 Adjusting temperature coefficient
P5.2 Adjusting normalization temperature
P6.0: Selecting cell constant
P6.1 Selecting cell constant:
K = 1.0, 10, or 0.1
P7.0: Setting Clock
• Setting Year
• Setting Date (month/day)
• Setting Time (hour / minute / second)
P8.0: Reset to factory defaults
P8.0 Reset meter to factory defaults
- 26 -
Instruction Manual CON 400/410
7.2 P 1.0: Memory Clear (CLr)
Use this parameter to clear all stored memory values when you need to store a new series of
values. This lets you avoid confusing the old values with the new ones.
NO is the default setting.
Selecting YES will wipe out all memory.
NOTE:
From measurement mode:
1. Press the SETUP key to enter Set Up mode.
2. Press the MI/S or MR/T keys to scroll through
subgroups until you view the parameter P1.0.
3. Press the ENTER key to enter parameter P1.0.
4. Press the MI/S or MR/T keys to toggle between
NO and YES. NO retains current memory; YES
clears all memory. See Figure 23.
5. Press the ENTER key to confirm selection and
return to the subgroup menu. Press CAL/MEAS key
to return to measurement mode.
- 27 -
Figure 23 - P1.0: Memory Clear
Instruction Manual CON 400/410
7.3 P2.0: View ing calibration data
This mode lets you recall previous calibration data, which
helps you know when to re-calibrate your meter. This is
a “view only” mode. See Figure 24.
From measurement mode:
1. Press the SETUP key to enter Set up mode.
2. Press the MI/S or MR/T keys to scroll through
subgroups until you view parameter P2.0.
3. Press ENTER key repeatedly to view previous
calibration data. The meter will first display the
calibration point, and then display the date and time
of calibration to help you comply with the Good
Laboratory Practice (GLP) standards.
4. When you have scrolled through all calibration data,
you will automatically return to the subgroup menu.
Press CAL/MEAS key to return to measurement
mode.
NOTES:
If there is no previous calibration data at a particular point,
the primary display will show “ ----“.
CON 410 meter only: If you entered Set Up mode from
Conductivity measurement mode, calibration data will be
in µS or mS. If you entered Set Up mode from TDS
measurement mode, calibration data will be in ppm or ppt.
Figure 24 - View calibration data
and its relevant time and date
- 28 -
Instruction Manual CON 400/410
7.4 P3.0: View ing probe data
Program 3 has five “view only” options that let you check
the probe’s parameters for diagnostic purposes. These
options show you the effective cell constant for each range.
The cell constant is adjusted according to your calibration.
From measurement mode
1. Press the SETUP key to enter Set Up mode.
2. Press the MI/S or MR/T keys to scroll through
subgroups until you view parameter P3.0.
3. Press the ENTER key repeatedly to view the effective
cell constant for each range.
4. When you have scrolled through all probe data, you
will automatically return to the subgroup menu. Press
the CAL/MEAS key to return to measurement mode.
See Figure 25.
NOTES:
Cell constants may degrade with time and usage. You
can use this feature to alert you to the need for a new
probe prior to total failure.
- 29 -
Figure 25 - View probe data for
each measurement range
Instruction Manual CON 400/410
7.5 P4.0: Unit configuration
7.5.1 P4.1: READY indicator and auto
endpoint function
Program P4.1 lets you select “READY indicator on” to
indicate when your measurement is stable, or select
“READY indicator off” for faster meter response.
Program P4.1 also lets you switch the Auto endpoint
function on or off. Select auto endpoint on to “hold” your
measurement when it is stable for more than 5 seconds.
The display automatically freezes, and HOLD indicator
appears on the left side of the display. Press the HOLD
key to release the display and access other functions.
Select auto endpoint off to deactivate this feature.
From measurement mode
1. Press SETUP key to enter Set Up mode.
2. Press the MI/S or MR/T keys to scroll through
subgroups until you view parameter P4.0.
3. Press ENTER key to select parameter 4.1.
4. Press the MI/S or MR/T keys to select the
configuration you require.
• OFF switches the READY indicator off; ON
switches the READY indicator on. ON and HOLD
together switches the auto endpoint feature on.
5. Press ENTER key to confirm selection and to
proceed to step 4 of P4.2. You can also press the
CAL/MEAS key to return back to measurement
mode.
NOTE: Meter default is set for Ready Indicator on, and auto
endpoint function off.
Figure 26 - Configuring the
READY feature
- 30 -
Instruction Manual CON 400/410
7.5.2 P4.2 Selecting °C or °F
You can select between °C and °F units for temperature readings. Meter default is °C.
From measurement mode
1. Press SETUP key to enter Set Up mode.
2. Press the MI/S or MR/T keys to scroll through
subgroups until you view parameter P4.0.
3. Press the ENTER key two times to select parameter
4.2.
4. Press the MI/S or MR/T keys to toggle between °C and °F.
5. Press the ENTER key to confirm selection and to
proceed to step 3 of P4.3. Press the CAL/MEAS
key to return to measurement mode.
See Figure 27.
Figure 27 - Change Temperature
measurement unit
- 31 -
Instruction Manual CON 400/410
7.5.3 P4.3 Selecting Automatic or Manual Temperature Compensation
This feature lets you select between Automatic Temperature Compensation (ATC) and
Manual Temperature Compensation. Meter default is ATC activated.
From measurement mode
1. Press SETUP key to enter Set Up mode.
2. Press the MI/S or MR/T keys to scroll through
subgroups until you view parameter P4.0.
3. Press the ENTER key three times to select parameter
4.3. The upper display shows “ATC” and the lower
display shows “P4.3”.
4. Press the ENTER key again. The upper display
shows “ATC” and the lower display shows “YES” or
“NO”.
5. Press MI/S or MR/T keys to select Automatic or
Manual Temperature Compensation.
• YES = Automatic Temperature Compensation
activated.
• NO = Manual Temperature Compensation
activated.
6. Press the ENTER key to confirm selection and to
return to the subgroup menu. Press the CAL/MEAS
key to return to measurement mode.
- 32 -
Figure 28 - Selecting Automatic or
Manual Temperature Compensation
Instruction Manual CON 400/410
7.5.4 P4.4 Setting the TDS factor
For CON 410 meter only
The concentration of salts dissolved in solution increases the conductivity of that solution. This
relationship varies from salt to salt and is roughly linear over a given range for a given salt.
The TDS conversion factor is the number used by the meter to convert f rom conductivity to
TDS.
To determine the conductivity to TDS conversion
factor for your solution:
Addendum 2 on page 46 explains the calculation in
determining the conductivity to TDS conversion factor.
You can set the TDS conversion factor between 0.40 and
1.00; meter default is 0.50.
From measurement mode
1. Press SETUP key to enter Set Up mode.
2. Press the MI/S or MR/T keys to scroll through
subgroups until you view parameter P4.0.
3. Press the ENTER key five times to select parameter
4.4. The upper display shows “tdS” and the lower
display shows “P4.4”.
4. Press the ENTER key again. The upper display
shows a value and the lower display shows “tdS”.
5. Calculate the TDS factor of your solution. See
Addendum 2 on page 46 for information on how to
calculate the TDS factor.
6. Press the MI/S or MR/T keys to select your
calculated TDS conversion factor.
7. Press the ENTER key to confirm selection and to
return to the subgroup menu. Press the CAL/MEAS
key to return to measurement mode.
Figure 29 - Change of TDS factor
- 33 -
Instruction Manual CON 400/410
7.6 P5.0 Temperature
7.6.1 P5.1 Adjusting the temperature coefficient
The temperature coefficient is the amount of change in conductivity per degree of temperature;
it is expressed in percent per °C. Entering the exact temperature coefficient of your solution
lets you accurately compensate temperature for almost any solution. You can adjust 0.0 to
10.0 % per °C. Meter default is 2.1% per °C. Adjustment to 0.0% will not correct for
temperature effects so the displayed value will be the actual value at the displayed
temperature.
From measurement mode
1. Press SETUP key to enter Set Up mode.
2. Press the MI/S or MR/T keys to scroll through
subgroups until you view parameter P5.0.
3. Press the ENTER key to select parameter 5.1. The
display shows “t.CO” on the upper display.
4. Press the ENTER key again. The upper display
shows the temperature coefficient and the lower
display shows “t.CO”.
5. Press the MI/S or MR/T keys to select the
temperature coefficient of your solution.
6. Press the ENTER key to confirm selection and to
proceed to step 3 of P5.2. Press the CAL/MEAS key
twice to return to measurement mode.
NOTES:
If you do not know the temperature coefficient of your
solution you can determine the correct value using the
formula in Addendum 3 “Calculating Temperature
Coefficients” on page 47.
•
- 34 -
Figure 30 - Changing the
temperature coefficient
Instruction Manual CON 400/410
7.6.2 P5.2 Adjusting the normalization temperature
Your meter will normalize its conductivity measurements to a standard temperature that you
can select. You can adjust the normalization temperature from 15 to 30 °C (59 to 86 °F). Meter
default is 25 °C (77 °F).
From measurement mode
1. Press SETUP key to enter Set Up mode.
2. Press the MI/S or MR/T keys to scroll through
subgroups until you view parameter P5.0.
3. Press the ENTER key three times to select
parameter 5.2. The display shows “t.nr” on the
upper display.
4. Press the ENTER key again. The upper display
shows the normalisation temperature and the lower
display shows “t.nr”.
5. Press the MI/S or MR/T keys to select the
normalisation temperature.
6. Press the ENTER key to confirm selection and to
return to the subgroup menu. Press CAL/MEAS key
to return to measurement mode.
NOTE:
Most calibration standards are referenced to 25°C. If a
normalization temperature other than 25°C is entered, the
calibration standard value at this temperature must be
known.
- 35 -
Figure 31 - Adjusting the
normalization temperature
Instruction Manual CON 400/410
7.7 P6.0 Sele cting the cell constant
Your meter lets you select a cell constant of K = 1.0, 10, or 0.1.
Use a cell of K = 1.0 for midrange measurements
Use a cell of K = 10 for high range measurements (above 20 mS or 10 ppt).
Use a cell of K = 0.1 for low range measurements (below 20 µS or 10 ppm).
The cell included with your meter has a cell constant of K = 1.0.
From measurement mode
1. Press SETUP key to enter Set Up mode.
2. Press the MI/S or MR/T keys to scroll through
subgroups until you view parameter P6.0.
3. Press the ENTER key to select parameter 6.1.
4. Press the MI/S or MR/T keys to select the cell
constant between K = 1.0, 0.1, or 10.
5. Press the ENTER key to confirm selection and to
return to the subgroup menu. Press the CAL/MEAS
key to return to measurement mode.
NOTES
When using a cell of K = 0.1, the lowest measuring range
will be: 0 to 1.999 µS (0 to 0.999 ppm). The 0 to 199.9 ms
(10 to 200 ppt) range will not be accessible.
When using a cell of K = 10, the highest measuring range
will be: 0 to 1999 mS (100 to 2000 ppt). The 0 to 19.99 µS
(0.00 to 9.99 ppm) range will not be accessible.
- 36 -
Figure 32 - Selecting cell constant
Instruction Manual CON 400/410
7.8 P7.0: Settin g the real-time clock
The meter features a real-time calendar and clock. This
helps you to meet the Good Laboratory Practice (GLP)
standards.
From measurement mode
1. Press SETUP key to enter Set Up mode.
2. Press the MI/S or MR/T keys to scroll through
subgroups until you view parameter P7.0.
3. Press the ENTER key to enter parameter P7.0. The
meter lets you select the century: “19—“ or “20—“. The
century digits will flash.
4. Press the MI/S or MR/T keys to toggle to the correct
century.
5. Press ENTER key to confirm the century and move to
“year” selection. The “year” digits will flash.
6. Press the MI/S or MR/T keys to toggle to the correct
year. See Figure 33.
7. Press ENTER key to confirm the year and move to
“month” selection. The “month” digits will flash.
8. Press the MI/S or MR/T keys to toggle to the correct
month.
9. Press the ENTER key to confirm the month and move
to “date” selection. The “date” digits will flash. See
Figure 33.
10. Press the MI/S or MR/T keys to toggle to the correct
date.
11. Press the ENTER key to confirm the date and move to
“hour” selection. The “hour” digits will flash. See Figure
34 (next page).
- 37 -
Figure 33 - P5.0: Setting century,
year, month and date
Instruction Manual CON 400/410
12. Press the MI/S or MR/T keys to toggle to the correct
hour. Note the “AM” and “PM” indicator on the lower
portion of the display.
13. Press the ENTER key to confirm the hour and move
to “minute” selection. The “minute” digits will flash.
14. Press the MI/S or MR/T keys to toggle to the correct
minutes.
15. Press the ENTER key to confirm the minutes and
move to “second” digits will flash. See Figure 34.
16. Press the MI/S or MR/T keys to toggle to the correct
seconds.
17. Press the ENTER key to confirm the seconds and
return to “century” selection.
18. Press the CAL/MEAS key to return to the subgroup
menu. Press the CAL/MEAS key again to return to
measurement mode.
NOTES
Press the CAL/MEAS key at any point while setting the
time to return to the subgroup menu.
Figure 34 - Selecting hour,
minutes and seconds
- 38 -
Instruction Manual CON 400/410
7.9 P8.0: Re setting to factory default settings
Program 8.0 lets you reset all parameters to factory default settings. This clears all calibration
data, memory, and any other setup functions you might have changed. It does not clear clock
settings.
From measurement mode
1. Press SETUP key to enter Set Up mode.
2. Press the MI/S or MR/T keys to scroll through
subgroups until you view parameter P8.0.
3. Press the ENTER key to enter parameter P8.0.
4. Press the MI/S or MR/T keys to toggle between NO
and YES. NO retains current settings; YES resets to
factory default settings.
5. Press the ENTER key to confirm selection and to
return to the measurement mode. Otherwise press
CAL/MEAS key to return to measurement mode
without resetting to factory default.
See Figure 35.
NOTES:
See Addendum 4 on page 48 for a table of factory default
settings.
Figure 35 - P6.0: Reset to
factory default values
- 39 -
Instruction Manual CON 400/410
8 PROBE CARE AND MAINTENANCE
Keep the conductivity probe clean. Rinse the probe twice, and
gently swirl it while you take readings. Rinse the probe with DI
or clean water before storing. Never scratch the bands with a
hard substance. Do not strike the probe against any hard
surface.
Do not immerse the probe in oily solutions. Clean the
electrode thoroughly by stirring it in a mild detergent bath or
isopropyl alcohol if needed. Wipe the probe with a soft tissue
paper. Rinse thoroughly in tap water and then in DI water.
Recalibrate the meter after cleaning the probe.
The conductivity probe (Order Part No. EC-CONSEN91W/
35608-50) which is included with your meter features a
removable probe guard to make cleaning easy.
NOTE: The yellow probe guard must be attached during
measurement and calibration. Readings are unstable and not
accurate without this guard in place.
To remove probe guard for cleaning:
1. Grip yellow probe guard and twist clockwise. The locking
notch will release.
2. Slide probe guard off end of probe.
Figure 36 - Conductivity probe
(EC-CONSEN91W/ 35608-50)
- 40 -
Instruction Manual CON 400/410
9 TROUBLE SHOOTING GUIDE
Problem Cause Solution
No display when
turned ON
Unstable readings a) Air bubbles in probe.
Slow response a) Dirty / Oily probe. a) Clean probe. See “Probe
a) Batteries not in place or
battery housing is not
attached to meter correctly
b) Batteries not in correct
polarity (+ and – position).
c) Weak batteries
b) Dirty probe.
c) Probe not deep enough in
sample.
d) External noise pickup or
induction caused by nearby
electric motor.
e) Broken probe.
f) probe guard not in place
CAL & Err
annunciators on /
Buffer and
electrode
indicators blink.
Battery indicator
blinks
Err 1 (in primary
display)
Err. 2 (in primary
display)
Err. 3 (in primary
display)
Err. 4 (in primary
display)
from keypad
Calibration error. Wrong value input at
Low battery level. Need new batteries or
Memory write error. Hardware failure. Turn meter on and off
Memory checksum
error.
A/D converter error. Hardware error. Turn meter on and off
Keypad error. One or more keys on
* See “Warranty” and “Return of Items” on pages 49 and 50.
If an error message appears in the primary display (the
upper row of larger digits), switching off the meter and
switching it on again may eliminate the error message.
Refer to diagram on right.
If error persists, or the meter shows incorrect values, return the meter.
For a complete diagram of the display see page 2.
Wrong input in selected
mode.
calibration.
Dirty probe.
battery connection is
bad.
Hardware failure. Turn meter on and off
the keypad are stuck or
fault in keypad.
Release key. Select valid
operations depending on
mode.
Check your input value,
clean probe.
See Calibration sections or
Probe Maintenance
section.
Clean battery contacts.
Replace batteries with fresh
ones, noting polarity.
again. If message persists,
return unit*.
again. If message persists,
return unit*.
again. If message persists,
return unit*.
Turn meter on and off
again. If message persists,
return unit*.
- 42 -
Instruction Manual CON 400/410
11 SPECIFICATIONS
SPECIFICATIONS DESCRIPTIONS CON 400 CON 410
Conductivity Range
TDS Range
Resolution 0.05 % Full Scale • •
Accuracy ±1% F.S. ±1 digit • •
Temperature Range
Resolution
Accuracy
Cell Constant 0.1, 1.0, 10.0 (selectable) • •
Temperature Compensation Automatic / Manual (from 0 to 80 ° C) • •
Temperature Coefficient 0.0 to 10.0% / ° C • •
Normalization Temperature 15.0 to 30.0 ° C (adjustable) • •
Conductivity to TDS
Conversion factor
Number of calibration points 5: Maximum 1 per range • •
Auto-ranging Yes (manual select available) • •
HOLD Function Yes • •
Memory Yes 50 50
Auto Power off 20 minutes after last key operation • •
Averaging/Stability
Waterproof CON 400 Hand-held conductivity meter
complete with conductivity probe of k=1.0 (ECCONSEN91W).
Waterproof CON 400 Hand-held conductivity meter
complete with conductivity probe of k=1.0 (ECCONSEN91W) packed in hard plastic carrying case
Waterproof CON 410 Hand-held conductivity/TDS
meter complete with conductivity/temp probe with
k=1.0 (EC-CONSEN91W).
Waterproof CON 410 Hand-held conductivity/TDS
meter complete with conductivity/temp probe with
k=1.0 (EC-CONSEN91W). packed in hard plastic
carry case
Ultem body conductivity/temp probe for CON 400
series meters, k= 1.0, 16 X 110 mm, 1-m cable
Ultem body conductivity/temp probe for CON 400
series meters, k= 1.0, 16 X 110 mm, 5-m cable
Epoxy body conductivity/temp probe for CON 400
series meters, k= 0.1, 12 X 110 mm, 1-m cable
Epoxy body conductivity/temp probe for CON 400
series meters, k= 10.0, 12 X 110 mm, 1-m cable
Hard carrying case WPDRYKIT 35632-98
Eutech Instruments
Ordering Code
EC-CONWP400/03 35608-00
EC-CONWP410/03K -
EC-CONWP410/03 35608-10
EC-CONWP410/03K -
EC-CONSEN91W 35608-50
- 35608-57
EC-CONSEN72W 35608-55
EC-CONSEN73W 35608-51
Oakton Instruments
Ordering Code
Calibration Solutions
Calibration standard solutions referenced to 25°C
10 µS single use pouches 20 mL x 20 pcs
447 µS single use pouches 20 mL x 20 pcs
1413 µS 480-mL bottle (1 pint) EC-CON-1413BT 00653-18
1413 µS single use pouches 20 mL x 20 pcs
2764 µS 480-mL bottle (1 pint) EC-CON-2764BT 00653-20
12.88 mS 480-mL bottle (1 pint) EC-CON-1288BT 00606-10
15.00 mS 480-mL bottle (1 pint) - 35653-13
Eutech Instruments
Ordering Code
EC-CON-10BS
EC-CON-447BS 35653-10
EC-CON-1413BS 35653-11
Oakton Instruments
Ordering Code
35653-09
- 44 -
Instruction Manual CON 400/410
13 ADDENDUM 1: CALIBRATION TIPS
You only need one calibration for measurement throughout the entire range of the meter. If a
range was not calibrated, the meter automatically detects the closest range calibrated and
uses that calibration information. However, only the ranges that were calibrated have
maximum accuracy.
• If you are measuring in ranges near to or greater than 20 mS (10 ppt), or
near to or lower than 100 µS (50 ppm), calibrate the meter at least once a
week to get specified ±1% F.S. accuracy.
• If you are measuring in the mid-ranges and you washed the probe in
deionized water and stored it dry, calibrate the meter at least once a month.
•
Wet the probe for 10 minutes before calibrating or taking readings to saturate the probe
surface and minimize drift. If you make measurements at extreme temperatures, calibrate the
meter at least once a week.
You should only use the conductivity / TDS / temperature probe specified for these meters.
These probes have a built-in temperature sensor. If you use a different probe without a
temperature sensor, you must measure the solution temperature separately and manually
enter the solution temperature (see manual temperature compensation section on pages 1617 ).
- 45 -
Instruction Manual CON 400/410
14 ADDENDUM 2: CALCULATING TDS CONVERSION
FACTORS
You can calibrate your meter using TDS calibration standard solutions. The calibration
standard only needs to give the TDS value at a standard temperature such as 25 °C. To
determine the conductivity-to-TDS conversion factor use the following formula:
Factor = Actual TDS ÷ Actual Conductivity @ 25 °C
Definitions:
Both the Actual TDS and the Actual Conductivity values must be in the same magnitude of
units. For example, if the TDS value is in ppm the conductivity value must be in µS; if the TDS
value is in ppt the conductivity value must be in mS.
Check your factor by multiplying the conductivity reading by the factor in the above formula.
The result should be in TDS value.
• Actual TDS: Value from the solution bottle label or as a standard you make
using high purity water and precisely weighed salts.
• Actual Conductivity: Value measured using a properly calibrated
Conductivity/Temperature meter.
14.1 “Preparing your ow n TDS calibration standards”.
If you are trying to measure a concentration of a fertilizer, sea salt, or other sample that uses
multiple salts that may not be relevant to commercially available KCl based standards, it may
be desirable to prepare your own TDS standards.
Since ppm is expressed as milligrams per Liter, the preparation is relatively simple with an
analytical balance. One gram of water = 1 mL = 1 cubic centimeter.
To prepare a TDS calibration standard of 750 ppm, weigh out 750 milligrams (0.75 grams) of
your salt or fertilizer and dissolve into 1000 grams (1 liter) of purified water
- 46 -
Instruction Manual CON 400/410
15 ADDENDUM 3: CALCULATING TEMPERATURE
COEFFICIENTS
To determine the temperature coefficient of your sample solution use this formula:
Where:
tc = Temperature coefficient 25 = 25 °C
= Conductivity at Temp 1 CT2 = Conductivity at Temp 2
C
T1
= Temp 1 T2 = Temp 2
T
1
NOTE: A controlled temperature water bath is ideal for this procedure.
1. Immerse the probe into a sample of your solution and adjust the temperature coefficient
to 0% (that is, no compensation) by performing the following:
A. From measurement mode, press the SETUP key to enter Setup mode.
B. Press the MI/S or MR/T key until the lower display reads P5.0.
C. Press the ENTER key twice. The lower display reads tCO and the upper display
shows the temperature coefficient value.
D. Press the MR/T key until the upper display shows 0.0.
E. Press ENTER key to confirm the value.
F. Press CA/MEAS key twice to return to measurement mode.
2. Wait for 5 minutes. Note T
3. Condition the sample solution and probe to a temperature (T
different from T
, and note the conductivity reading CT2.
1
and CT1 (conductivity at T1).
1
) that is about 5 °C to 10 °C
2
NOTE: Record your results for future reference. Ideally T
and T2 should bracket your
1
measurement temperature, and should not different by more than 5 °C.
4. Calculate the temperature coefficient of your solution according to the formula shown
above.
5. Enter the temperature coefficient you calculated into the meter.
A. From measurement mode, press the SETUP key to enter Setup mode.
B. Press the MR/T key until the lower display reads P5.0.
C. Press the ENTER key twice. The lower display reads tCO and the upper display
shows the temperature coefficient value (should be 0, as per step 1 above).
D. Press the MI/S key until the upper display shows your calculated temperature
coefficient.
E. Press ENTER key to confirm the value.
F. Press CAL/MEAS key twice to return to measurement mode.
The calculated temperature coefficient will not be applied to all the meter readings.
- 47 -
Instruction Manual CON 400/410
16 ADDENDUM 4: METER FACTORY DEFAULT
SETTINGS
Type Parameter Default Remarks
P1.0 Memory clear No Retains current memory
P2.1 Viewing previous –– No calibration data for 1st range
P2.2 calibration data –– No calibration data for 2nd range
P2.3 –– No calibration data for 3rd range
P2.4 –– No calibration data for 4th range
P2.5 –– No calibration data for 5th range
P3.1 Viewing probe data 1.0 No offset for effective cell constant (1st range)
P3.2 1.0 No offset for effective cell constant (2nd range)
P3.3 1.0 No offset for effective cell constant (3rd range)
P3.4 1.0 No offset for effective cell constant (4th range)
P3.5 1.0 No offset for effective cell constant (5th range)
P4.1 Ready indicator / auto
endpoint
P4.2 Select °C/°F °C Degrees Celsius
P4.3 ATC on or off ATC on ––
P4.4 TDS factor 0.50 Adjustable from 0.40 to 1.00 (available in TDS
P5.1 Temperature coefficient 2.1 % per ° C Adjustable from 0.0 to 10%
P5.2 Normalisation temperature 25° C Adjustable from 15 to 30° C
P6.1 Cell constant 1.0 Select from k=1.0, 0.1 or 10
P8.0 Factory default No Retains your current settings
NOTES:
The P7.0 clock function retains the year and time that you set, even after reverting to factor y
default settings.
READY/ON Ready indicator on; auto endpoint off
400 and CON 410 meters only)
- 48 -
Instruction Manual CON 400/410
17 WARRANTY
This meter is supplied with a three -year warranty, six-month warranty for probe against
significant deviations in material and workmanship.
If repair or adjustment is necessary and has not been the result of abuse or misuse within the
designated period, please return – freight pre-paid – and correction will be made without
charge. Eutech Instruments/ Oakton Instruments will determine if the product problem is due
to deviations or customer misuse.
Out of warranty products will be repaired on a charged basis.
Exclusions
The warranty on your instrument shall not apply to defects resulting from:
Improper or inadequate maintena nce by customer
Unauthorised modification or misuse
Operation outside of the environment specifications of the products
- 49 -
Instruction Manual CON 400/410
18 RETURN OF ITEMS
Authorisation must be obtained from our Customer Service Department or authorised
distributor before returning items for any reason. A “Return Goods Authorisation” (RGA) form
is available through our authorised distributor. Please include data regarding the reason the
items are to be returned. For your protection, items must be carefully packed to prevent
damage in shipment and insured against possible damage or loss. Eutech Instruments/
Oakton Instruments will not be responsible for damage resulting from careless or insufficient
packing. A restocking charge will be made on all unauthorised returns.
NOTE: Eutech Instruments Pte Ltd/ Oakton Instruments reserves the right to make
improvements in design, construction, and appearance of products without notice.
- 50 -
NOTES
For more information on Eutech Instruments/ Oakton Instruments’ products, contact
your nearest distributor or visit our website listed below:
Oakton Instruments
P.O Box 5136,
Vernon Hills, IL 60061, USA
Tel: (1) 888-462-5866
Fax: (1) 847-247-2984
E-mail: info@4oakton.com
Web-sites: