EUTECH INSTRUMENTS 60010-00, 60010-10, 60010-20 Instruction Manual

Technology Made Eas y . ..
Thermocouple
Thermometer Models
60010-00 (Type J) 60010-10 (Type K) 60010-20 (Type T)
68X309905 Rev 1 04/04
TABLE OF CONTENTS
1.
INTRODUCTION 4
2.
SAFETY PRECAUTIONS 6
3.
SPECIFICATIONS 7
4.
BATTERY INSTALLATION AND REPLACEMENT 10
5.
SELF TEST 12
6.
CONNECTING A THERMOCOUPLE 12
7.
SELECTING TEMPERATURE SCALE 16
8.
HOLD FUNCTON 17
9.
CALIBRATION 17
10.
CANCELING A FIELD CALIBRATION 18
11.
FIELD CALIBRATION LOCKOUT AND RE-ENABLE 19
12.
LOCKOUT PROCEDURE 19
13.
RE-ENABLE PROCEDURE 20
14.
MAINTENANCE 20
15.
CLEANING 21
16.
BATTERIES 21
17.
ACCESSORIES 22
18.
SERVICE 24
19.
WARRANTY 25
20.
PRODUCT RETURN 25
4
1. INTRODUCTION
This versatile hand-held instrument provides highly accurate temperature measurements. The instrument is designed for easy operation and includes the following features:
• Operator selection of Celsius or Fahrenheit scale
• Resolution of 0.1° from –99.9 to
299.9°
• Four-digit LCD
• Hold feature for temporarily retaining a reading
• Field calibration capability
5
• Low battery warning
• Built-in tilt stand for easy hands-free operation
• Two blade female ANSI mini­connector input
• Operates with a wide selection of probes 2
6
2. SAFETY PRECAUTIONS
WARNING: THIS INSTRUMENT IS
DESIGNED TO ACCEPT LOW LEVEL SIGNALS SUPPLIED BY STANDARD THERMOCOUPLES. UNDER NO CIRCUMSTANCES SHOULD THE INPUT VOLTAGE EXCEED THE SPECIFIED 50V RMS.
CAUTION: DO NOT USE OR STORE THIS INSTRUMENT IN MICROWAVE OVENS OR ANY ABNORMALLY HOT OR COLD AREAS.
CAUTION: WEAK BATTERIES SHOULD NOT BE LEFT IN THE INSTRUMENT. DEAD BATTERIES CAN LEAK AND CAUSE DAMAGE TO UNIT.
DANGER: VOLTAGES PRESENT AT THE THERMOCOUPLE MAY ALSO BE PRESENT AT THE BATTERY TERMINALS. ALWAYS DISCONNECT THE THERMOCOUPLE WHEN CHANGING BATTERIES.
WARNING: TO PREVENT IGNITION OF A HAZARDOUS ATMOSPHERE, BATTERIES MUST ONLY BE CHANGED IN AN AREA KNOWN TO BE NON-HAZARDOUS.
7
3. SPECIFICATIONS
THERMOCOUPLE PROBES
Type Temperature range
Type J –200°C to 1000°C
(–328°F to 1832°F )
Type K –250°C to 1372°C
(–418°F to 2501°F )
Type T –250°C to 400°C
(–418°F to 752°F )
Out of range display: - - - -
Resolution
0.1°/1° autoranging, 0.1° from –99.9° to
299.9°, 1° outside this range.
Accuracy
> –99.9°: +(0.2% of reading +0.5°C) +(0.2% of reading +0.9°F) < –99.9°: +(0.25% of reading, +1°C) +(0.25% of reading +2°F)
Display
4-digit LCD with 0.5 in high numerals.
Display update rate
0.6 sec per update.
Input
One thermocouple with ANSI connector.
8
Input Protection
50V rms
Battery
Two AA, 1.5V alkaline ANSI-L40, IEC­LR6. Battery Life: 750 hours continuous, typical.
Low battery indication
Battery Symbol on when 8 to 10 hours of battery life remains.
Battery symbol blinking: replace battery
Operating Conditions
Stated accuracy: 18°C to 28°C (64°F to 82°F)
Useful range: 0°C to 40°C (32°F to 104°F)
Storage: –40°C to 65°C (–40°F to 149°F)
Humidity (non-condensing): 10% to 90% 4
Dimensions
3 cm D x 8.4 cm W x 15.8 cm H (1.2 in x 3.3 in x 6.2 in)
Weight with batteries
227 grams (8 ounces)
9
Ingress protection
Meets IEC-529 IP-54 for dust and
water resistant enclosures.
Intrinsic safety
This product is energy limited for intrinsically safe operation in hydrogen atmospheres per Class I, Division 1, Groups A, B, C and D hazardous(classified) locations for UL per UL913 and CSA per C22.2 No. 0­M1982 and No. 157-M1987. Maximum surface temperature: 135°C (T4); UL file No. E182612 (1997).
Electromagnetic Compatibility
(For CE Mark) EN61326-1/1A: 1998 (EU EMC Directive)