EUTECH INSTRUMENTS 2-WIRE DISSOLVED OXYGEN TRANSMITTER, EC-PH-XTR, EC-ORP-XTR, EC-CON-XTR, EC-DO-XTR Operating Instructions Manual

...
EUTECH INSTRUMENTS
2-Wire Dissolved Oxygen
Transmitter
Operating Instructions
1 Preface
This manual serves to explain the correct installation and operations of the XTR Transmitters. Other Transmitters available individually are: pH, ORP (REDOX), Conductivity, Dissolved Oxygen and Temperature.
EC-PH-XTR;
EC-ORP-XTR;
EC-CON-XTR;
EC-DO-XTR; and
EC-TEMP-XTR.
The information presented in this manual is subject to change without prior notice, as improvements are made to the product. While Eutech will endeavor to try its best to inform the market, this does not represent a commitment on the part of Eutech Instruments Pte Ltd.
Eutech Instruments does not accept any responsibility for damage or malfunction of the unit due to improper installation or operation of the Transmitter.
Copyright 1999 Eutech Instruments Pte Ltd. Version 1.0. All rights reserved.
Eutech Instruments Pte Ltd. Blk 55 Ayer Rajah Crescent #04-14/24, Singapore 139 949. Tel: (65) 778 6876; Fax: (65) 773 0836; e-mail:
marketing@eutechinst.com; Home page: http://www.eutechinst.com
2 Introduction
The Transmitter is used for the continuous measurement of pH, ORP (REDOX), Conductivity, Dissolved Oxygen or Temperature, in conjunction with an appropriate sensor. Data output is via 4-20 mA current output.
The Transmitter is housed in a IP65 enclosure, with openings for input and 4-20mA output.
3 Diagram and Dimensions (in mm)
4 Installation
Operating with one or several Transmitters:
Figure 1 - Operating with 1 Transmitter
Figure 2 - Operating with several transmitters
5 Electrical Connections
6 Before Calibration
Remove cover of Transmitter and ensure that electrode is correctly connected as above. Connect Amp meter to the 4/20mA output of Transmitter (Appendix 1).
7 Dissolved Oxygen Transmitter Calibration
1) Short Dissolved Oxygen input and observe Amp meter display;
2) Amp meter should read 4mA; if not, calibration is necessary;
3) Locate “Offset adjustment” trimmer; Using a fine screwdriver, slowly turn trimmer till Amp meter reads 4mA;
4) For greater accuracy, adjust the slope using Air as a reference;
5) Hold the dry probe in Air;
6) In 100% Saturation (in Air), output is 12 mA;
7) In 200% Saturation, output is 20 mA;
8) Using a fine screwdriver, slowly turn trimmer, marked ‘Slope’, till Amp meter reads correct mA value with respective Oxygen Saturation levels.
9) Remember to disconnect Amp meter and close cover.
NB: For other % Saturation values, expected current output is based on:
mA= {(16/200) x % Saturation of Dissolved Oxygen} + 4mA
8 Specifications
Model ECDOXTR
Range 0 to 200% Accuracy
± 1.5% Full Scale Calibration 1 point (100% air saturation) Input Only with Galvanic DO probe – 2 wires Output 4-20 mA Operating
Voltage
DC 12 V to 24 V
Load 100 Ohms max. for 12 V;
600 Ohms max. for 24V
Housing 77 x 77 x 28 mm, field mountable
9 Appendix 1
Connecting Amp Meter for Calibration
Figure 3 - Method One
Figure 4 - Method Two
10 Ordering Information
pH
Item Cat Code
pH Transmitter ECPHXTR Combination pH Electrode, General, integral 5m cable ECARTSO05B Combination pH Electrode, w/PMP, integral 5m cable ECARGTSO05B
Dissolved Oxygen
Item Cat Code
Dissolved Oxygen Transmitter ECDOXTR General Purpose D.O. probe, 0.5 – 40 ppm, Industrial,
integral 5m cable
ECDOGEN
General Purpose D.O. probe, 0 – 10 ppm, Industrial, integral 5m cable
ECDOTPll
ORP
Item Cat Code
ORP Transmitter ECORPXTR Combination Platinum ORP Electrode, w/PMP,
integral 5m cable
ECHTPTTSO05B
Combination Gold ORP Electrode, w/PMP, integral 5m cable
ECHTAUTTSO05B
Conductivity
Item Cat Code
Conductivity Transmitter ECCONXTR 2-Pin SS,k=1.0, integral 3m cable ECCONSEN46
Temperature
Item Cat Code
Temperature Transmitter ECTEMPXTR Pt 100 probe, SS316 ECPT10001M or
ECPT10005M
Power Supply
Item Cat Code
120V Power Supply EC-120-XTR 220V Power Supply EC-220-XTR