All users must read this entire manual to fully understand
the safe use of Ettan IPGphor 3.
WARNING!
The following WARNING signs highlight instructions that
must be followed to avoid personal injury. It is important
not to continue until all stated conditions are met and
clearly understood.
WARNING! Lethal high voltage above 1 kV.
WARNING! Read the instructions to avoid
hazardous conditions.
CAUTION!
The Caution sign highlights instructions that
must be followed to avoid damage to the product or other equipment. It is important not
to continue until all stated conditions are met
and clearly understood.
Note
The Note sign is used to indicate information
important for trouble-free and optimal use of
the product.
CE certification
This product meets the requirements of applicable CEdirectives. A copy of the corresponding Declaration of
Conformity is available on request.
The CE symbol and corresponding declaration of
conformity is valid for the instrument when it is:
–used as a stand-alone unit, or
–connected to other CE-marked GE Healthcare
instruments, or
–connected to other products recommended or
described in this manual, and
–used in the same state as it was delivered from GE
Healthcare except for alterations described in this
manual.
Contents
1Ettan IPGphor 3 Isoelectric Focusing System function and
6.1Technical service and repair .............................................................. ......57
6.2Ordering information ........................................... ........................................ 57
viEttan IPGphor 3 Instrument User Manual 11-0034-58 Edition AB
Ettan IPGphor 3 Isoelectric Focusing System function and description 1
1Ettan IPGphor 3 Isoelectric Focusing System function
and description
The Ettan™ IPGphor™ 3 Isoelectric Focusing System performs the first dimension
isoelectric focusing (IEF) step in 2-dimensional (2-D) electrophoresis. The system
comprises the Ettan IPGphor 3 unit, combined with separate strip holders or a
manifold. Two types of manifolds ar e available; the ceramic manifold and the
light manifold. The IEF separation protocols are programmed, run and controlled
from the Ettan IPGphor 3 control panel or from a PC running the Ettan IPGphor
Control Software.
IMPORTANT! The Ettan IPGphor 3 system is intended for research use only
and should not be used in any clinical or in Vitro procedures for diagnostic
purposes.
The Ettan IPGphor 3 Isoelectric Focusing System includes the following features:
•an integrated high-voltage DC power supply delivering up to 10 kVolts
•temperature control by Peltier elements
•light exclusion during the run. Compatible with CyDye™ labeled proteins
and other light sensitive stains
•IEF runs are programmed from the Ettan IPGphor 3 control panel with up to
ten user-defined IEF protocols, each with up to 9 steps per protocol
•programmable functions including: rehydratio n time, platform temperature,
current limit, voltage limit for each step, voltage gradient or step and step
duration
•Ettan IPGphor 3 Control Software. From a PC, protocols are created,
graphically displayed and stored on the Ettan IPGphor 3 unit. The software
records the run parameters over time and presents data as graphs and log
files. Data is saved or can be exported to Microsoft® Excel
•Ettan IPGphor 3 Control Software allows easy programmin g and
recommended protocols are generated by providing instrument
configuration, IPG strip length and pH gradient
•strip holders for rehydration and automated running
•Ettan IPGphor Cup Loading Manifold for easy running of up to 12 IPG strips.
Cup loading improves focusing results particularly when using basic IPG
strips
Ettan IPGphor 3 Instrument User Manual 11-0034-58 Edition AB7
1 Ettan IPGphor 3 Isoelectric Focusing System function and description
•for larger sample volumes the Ettan IPGphor Cup Loading Manifold
facilitates paper bridge loading and in gel rehydration loading
The Ettan IPGphor 3 Control Software has an easy to use interface to the Ettan
IPGphor 3 instrument and provides enhanced p rotocol genera tion, present ation,
and storage.
Recommended standard protocols can be retrieved and used by the user simply
providing the instrument configuration, IPG strip length and pH gradient. Using
the advanced mode setting, earlier made custom protocols or entire new
protocols are made. All protocols are edited and saved for later use. The selected
protocol is loaded to Ettan IPGphor 3 Isoelectric Focusing Instrument.
During the run the control software monitors and records the running
parameters. All data is displayed as graphs and log file ta bles. When the run is
complete, graph and log files are printed or/and saved. If desired, the log files are
exported to programs such as Microsoft Excel.
The Ettan IPGphor 3 Control Software contro ls up to four Ettan IPG phor 3 units at
one time, each running a different set of run parameters.
IPG strip manifolds are available in two models:
•One produced from a ceramic material providing greater temperature
uniformity
•One produced from a light polymeric material for less critical
experimentation
Both are designed for analytical analysis using cup application and preparative
runs applying sample to the gel duri ng the rehydration step or by paper bridge
loading. Using the Manifold Immobiline DryStrips are rehydrated in the
Rehydration Tray.
Fixed length strip holders are available. Once the IPG strip is placed in the fixedlength strip holder, no further handling is required. The holder is placed on the
Ettan IPGphor 3 electrode platform for rehydration and can be immediately
followed by isoelectric focusing. The sample is introduced during the rehydration
step. The Ettan IPGphor 3 accommodates up to 12 strip holders for any one
protocol.
IPG strips are available in five lengths 7, 11, 13, 18, and 24 cm, and a number of
pH ranges, several linear and non-linear (listed in section 6.2 Ordering
information).
8Ettan IPGphor 3 Instrument User Manual 11-0034-58 Edition AB
Ettan IPGphor 3 Isoelectric Focusing System function and description 1
1.1Unpacking
Carefully unpack all items delivered and make sure that all items are presen t by
comparing the contents with the pa c king list. If any part is missing, please
contact your local GE Healthcare sales office. Inspect all components. If any part
appears damaged, please contact the courier immediately. Please keep all
packing material and use if damage d parts need to be returned for repair or
replacement.
1.2System components
Safety lid
Fig 1-1. The Ettan IPGphor 3 instrument.
Positive (anode)
electrode area
Negative (cathode)
electrode area
Control panel
Ettan IPGphor 3 Instrument User Manual 11-0034-58 Edition AB9
1 Ettan IPGphor 3 Isoelectric Focusing System function and description
Fig 1-2. Light protective cover.
Light protective
cover
/
L
Fig 1-3. Location of mains power switch and serial port.
The mains power
switch is located on the
rear panel.
Serial port (RS232) for
connection to a PC
10Ettan IPGphor 3 Instrument User Manual 11-0034-58 Edition AB
A.
Ettan IPGphor 3 Isoelectric Focusing System function and description 1
B.
D.
G
C.
F
I.
E
Fig 1-4. Components of the Ettan IPGphor Cup Loading Manifold system. Only one of each
item shown.
A. Manifold, ceramic or lightF. Pre-cut paper bridges
B. Pre-cut electrode wicksG. Sample cup insertion tool
C. Electrode assemblyH. Cleaning solution
D. Sample cupsI. Forceps
E. Cleaning brushJ. Spirit level
H.
J
Ettan IPGphor 3 Instrument User Manual 11-0034-58 Edition AB11
1 Ettan IPGphor 3 Isoelectric Focusing System function and description
Five strip holder lengths are
available—7, 11, 13, 18, and 24 cm—
one for each IPG strip length.
Fig 1-5. Strip holders.
Note: All strips (and holders) used at any one time must be of the same length.
Fig 1-6. Lid adapter.
12Ettan IPGphor 3 Instrument User Manual 11-0034-58 Edition AB
Ettan IPGphor 3 Isoelectric Focusing System function and description 1
Control panelThe Ettan IPGphor 3 is controlled by a 7-key membrane keypad. Instrument
status is indicated on the LCD screen.
LCD screenThe
always under the protocol number “1.”
Select any protocol by pressing the
number of the desired protocol. Once the selected protocol is started, the LCD
screen indicates the parameters of the step currently being carried out.
In “edit mode” the LCD displays the programmable parameters. Position the
cursor in the field to be edited using the
value by pressing the
COMMAND keys
•The EDIT key switches the display to "edit mode", in which protocol steps
•The left arrow key moves the cursor to the left. Moving the cursor off the
•The right arrow key moves the cursor to the rig ht. Moving the cursor off
•The up arrow key increases the value of the selected field or digit, and the
•The START key begins the selected protocol.
•The STOP key pauses the protocol. Pressing the STOP key a second time
main screen is “Protocol #1, Name.” The default location of the cursor is
up and down arrow keys to display the
right and left arrow keys and adjust the
up and down arrow keys.
can be programmed (see section 3.5). Pressing the EDIT key in edit mode
toggles between the identification screen with general settings for all
steps and the settings in the three first protocol steps. The key also allows
the estimated run time r e ma ini ng to be vi ew ed w h ile the unit i s op era tin g.
screen to the left repositions the cursor at the left side of the next row. On
the bottom row the cursor is repositioned on the left side of the top row.
the screen to the right repositions the cursor at the left of the next row. On
the bottom row the cursor is repositioned on the left side of the top row.
down arrow key decreases the value of the selected field or digit.
Exceptions when the cursor is positioned to the left under the Step
number S1-S9, down arrow change the display to the next group of three
protocol steps. Up arrow change to the previous group of three protocol
steps. During electrophoresis, the up/down arrow keys are used to toggle
between VhS, VhT, and Hrs.
cancels the protocol in progress.
HV ON indicator lampThe HV ON lamp indicates that the high voltage power supply is on and
isoelectric focusing is in progress. This lamp is not illuminated during the
rehydration phase of a protocol.
Serial port RS 232Serial port for connection of a PC.
Ettan IPGphor 3 Instrument User Manual 11-0034-58 Edition AB13
1 Ettan IPGphor 3 Isoelectric Focusing System function and description
PlatformThe platform accommodates one manifold for the first dimension isoelectric
focusing (IEF) run.
Up to 12 fixed-length strip holders can be placed on the platform for rehydration
and IEF.
The embedded electrode areas provide the electrical connection between strip
holder or manifold electrodes and the integral power supply.
Platform temperature is controlled by Peltier thermoelectric modules.
Electrode areasBoth electrode areas are gold-plated copper. Placing a manifold or a strip holder
so that it bridges both electrode areas completes the electrical circuit (once the
safety lid is in position).
Safety lidThe safety lid must be properly closed before power is applied.
Light protective coverWhen analyzing proteins labeled with light sensitive stains like CyDye®, a light
protective cover (Fig 1-2) is used. The cover is applied by magnetic force but is
easy to partially remove for checking the isoelectric focusing procedure,
especially the migration of bromo-phenol-blue at start.
Lid adapter for strip
holders
Ettan IPGphor 3
Control Software
Manifolds
Ettan IPGphor Cup
Loading Manifold
Ceramic
When using the Ettan IPGphor 3 standard strip holders a lid adapter (Fig 1-6) must
be used to apply the correct amount of pressure to the lids of the Ettan IPGphor 3
strip holders.
Ettan IPGphor 3 Control Software allows the user to control the Ettan IPGphor 3
Isoelectric Focusing Instrument. The software provides an enhanced functionality
in the areas of data presentation, data storage, and protocol handling. The PC is
connected to the Ettan IPGphor 3 over a serial cable or USB converter.
Ettan IPGphor 3 Control Software controls up to four Ettan IPGphor 3 units at one
time, each running a different set of run parameters.
The manifold (Fig 1-4) is made of aluminum oxide ceramic for efficient heat
transfer and temperature control during IEF. The manifold accomodates up to 12
IPG strips from 7 cm to 24 cm long. Immobiline DryStrip is placed with the gel side
facing up.
Samples are applied in a cup. Cup loading has been found to improve protein
focusing patterns, particularly on basic IPG strips (pH 6-9 and 6-11). Larger
sample volumes can be applied by gel rehydration loading or anodic paper
bridge loading.
Protrusions along the channel inside the manifold align the rehydrated IPG strip,
keeping it straight and centered when placed inside the manifold. The manifold
has a coating to minimize protein adsorption. The unit may also be used for IPG
strip equilibration prior to second dimension electrophoresis.
The manifold is fragile and should be handled with care.
14Ettan IPGphor 3 Instrument User Manual 11-0034-58 Edition AB
Ettan IPGphor 3 Isoelectric Focusing System function and description 1
Ettan IPGphor Cup
Loading Manifold Light
Electrode AssemblyThese electrodes fit to both the ceramic and light manifold. The movable
Sample CupsSample cups (Fig 1-4) can be placed along the length of the Ettan IPGphor
IEF electrode paper
wicks
Paper bridge loading
pads
The manifold (Fig 1-4) is made of polymers polycarbonate and ABS (acrylonitrilebutadiene-styrene). The light manifold and the ceramic manifold have the same
design/measurements with only the choice of materials the only difference. The
manifold accommodates up to 12 IPG strips from 7 cm to 24 cm long.
The unit may also be used for IPG strip equilibration prior to second dimension
electrophoresis.
electrodes (Fig 1-4) can be placed along the manifold to fit from 7 to 24 cm long
IPG strips.
Electrode pins make contact with the power supply electrode areas on the
platform of the Ettan IPGphor 3. The electrodes have a platinum wire that is strung
across the bottom of the electrode teeth to provide electrical contact to the IPG
strips. The hold down teeth for paper bridge loading are found along the opposite
length of the electrode. Cams on the sides of the electrode secure the position.
Manifold that is not blocked by a centering protrusion.
The sample cups are supplied in strips of 6 for easy handling and placement.
Each sample cup accommodates sample volumes of up to 150 µl.
The small, precut rectangular wicks (Fig 1-4) must be placed at both the anodal
and cathodal ends of the rehydrated IPG strips just under the electrodes.
The wicks absorb excess water, salts, and proteins with pI values that lie outside
the pH range of the IPG strip.
Paper Bridge loading pads (Fig 1-4) are used to load larger volumes of sample
(375-500 µl). The bridges are 0.8 cm
One end of the pad is in contact with the IPG strip and is held in position by the
hold down teeth of the electrode assembly. The other end of the bridge loading
pad makes contact with a paper wick placed under the platinum wire of the
electrode assembly.
× 5.0 cm in size including the pointed ends.
Immobiline DryStrip
Reswelling Tray
Strip holdersThe strip holders are made of aluminum oxide ceramic with electrodes of
Ettan IPGphor 3 Instrument User Manual 11-0034-58 Edition AB15
When using the manifold the Immobiline DryStrip Reswelling Tray is required for
proper strip rehydration. The longer Tray fits 7-24 cm strip lengths.
platinum on titanium. The ceramic is very brittle so the holders are fragile and
must be handled with care.
The ceramic is treated with a special coating to minimize protein binding and
must be cleaned only with the Ettan IPGphor Cleaning Solution supplied to retain
that coating.
Strip holders are available for the following IPG strip lengths: 7, 11, 13, 18, and
24 cm.
Each fixed-length strip holder base holds a single IPG strip throughout
rehydration and isoelectric focusing. A matching length holder is available for
each size strip.
1 Ettan IPGphor 3 Isoelectric Focusing System function and description
Sample application
wells
CoverThe cover is made of acrylic and protects the IPG strip during rehydration. The
Each of the two wells accepts up to 15 µl of sample (7.5 µl on either side of the
strip) in excess of the rehydration volume (i.e. 30 µl maximum for both wells).
Introduce the sample through the wells if it is not included in the rehydration
solution.
shape of the underside of the cover ensures electrical continuity during IEF by
pressing the strip against the electrode while allowing electrolysis gases to
escape.
Rehydration and separation progress can be monitored visually through the clear
cover.
16Ettan IPGphor 3 Instrument User Manual 11-0034-58 Edition AB
Ettan IPGphor 3 Isoelectric Focusing System function and description 1
1.3Specifications
Working surface
Electrode areasGold-plated copper
CapacityUp to 12 fixed-length strip holders or one
manifold
Platform temperature15–30 °C ±2 °C
User interface
Control panel7-key membrane keypad.
Liquid crystal display (LCD)4 lines, 24 characters per line
Programmable parametersRehydration time, platform temperature,
maximum current limit per strip, voltage
limit in each step, voltage step or
gradient, and step duration.
Protocol capacity10 protocols, with up to 9 steps each
Serial portRS232, 1200 or 9600 baud, 8 data bits 1
stop bit, no parity, no flow control.
9600 baud is mandatory for connection
with Ettan IPGphor 3 Control Software.
Operating environment
Temperature15–32 °C
Relative humidity0–70%*
* High relative humidity may result in condensation forming on the platform sur-
face. Runs conducted at a platform temperature of 20 °C and relative humidity
less than 70% will generally not cause condensation. Platform temperatures
greater than 20 °C allow for higher relative humidity without condensation. Environments for which condensation occurs on the platform surface may vary. If
condensation occurs, the unit may require conditioned air to reduce temperature and relative humidity.
Ettan IPGphor 3 Instrument User Manual 11-0034-58 Edition AB17
1 Ettan IPGphor 3 Isoelectric Focusing System function and description
Power requirements
Mains supply voltage100-240 V AC ±10%
Frequency50/60 Hz
Power consumption230 VA
Electrophoresis
High voltageUp to 10 kV
CurrentUp to 1.5 mA
Physical parameters
Dimensions (h
Weight8.3 kg
Safety
Safety featuresAutomatic voltage cutoff when safety lid
× w × d)160 × 278 × 471 mm
is opened
Safety certificationsCE
1.4Associated documentation
•The Ettan IPGphor 3 Software User Manual contains instructions for
controlling Ettan IPGphor 3 Isoelectric Focusing Unit from a PC.
•The Ettan IPGphor Cup Loading Manifold User Manual contains instructions
for first-dimension isoelectric focusing of proteins on IPG strips.
•Instructions attached to each package of Immobiline DryStrip.
18Ettan IPGphor 3 Instrument User Manual 11-0034-58 Edition AB
2Instrument set up
2.1Placing the instrument
WARNING! The vents at the front, back, and bottom of the instrument must
not be obstructed. Air must be free to circulate for the cooling apparatus to
function properly. Minimum free space behind the rear panel should be
10 cm.
IMPORTANT! Select a place for Ettan IPGphor 3 where it is possible to access
the mains power switch on the rear panel of the instrument.
Position the instrument on a flat surface with the spirit level in the ce nter. Turn the
leveling feet as necessary to make the instrument level and stable.
2.2Power connection
Two power cords are supplied: one cord has a “North American” style (UL817)
3-pronged plug for 115 V~ power outlets, and the other has a “Central European”
style (CEE7/VII) plug for 230 V~ power outlets.
Instrument set up 2
1Select the cord with the appropriate plug for the power outlet in your
laboratory. If neither of these cords are suitable for your power outlets,
obtain a detachable cord with an IEC/320/C13 (CEE22/V) outlet.
2Plug the connectors into the power cord receptacle on the rear panel and
into a properly grounded power outlet.
Ettan IPGphor 3 Instrument User Manual 11-0034-58 Edition AB19
2 Instrument set up
2.3Serial port connector for a computer connection
The RS232 serial port for a computer connection is at the rear of the unit. The type
of cable required (shielded straight serial or shielded null modem) depends on the
type of device (DTE or DCE) that it is connected to:
•If the computer is configured to re ce ive d ata o n pin 2 and transmit on pin 3,
a regular straight serial cable is required.
•For any other configuration, a null modem cable is required.
Pin number assignments and device set up details follow:
Ettan IPGphor 3 RS232 signal and pin number assignments:
The Ettan IPGphor 3 requires these settings in the device receiving data:
Baud rate9600
Data bits8
Stop bit 1
Start bit1
ParityNone
Flow controlNone
To set the baud rate:
1Press the Up and Down arrows together to access the Baud menu.
2Use the Up or Down arrows to set the baud rate.
3Use the Right arrow to exit to the main menu.
Serial Port Setup
Baud Rate: 9600
Set Up or Dn Exit>
20Ettan IPGphor 3 Instrument User Manual 11-0034-58 Edition AB
Loading...
+ 44 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.