eta K PLUS, K PLUS 1504, KA PLUS, K PLUS 1005, K PLUS 0510 Operating Manual

...
K PLUS - K CL PLUS - K CO PLUS
1
2
3
4
5
1
2
3
4
5
1
2
3
4
5
Self-venting version: KA PLUS
K CL PLUS
EN
K CO PLUS
K PLUS
SOLENOID DRIVEN METERING PUMPS
WITH DIAPHRAGM
OPERATING MANUAL
This operating instructions contains safety information that if ignored can endanger life or result in serious injury.
Read these instructions carefully before use and keep them for future reference. The original instruction is in Italian. All non-Italian instructions are translations of the original instruction.
Information and specifications on this manual could be uncorrect or could have printing errors. Specifications are subject to change without notice.
Version: R1-11-13
NORME CE EC RULES (STANDARD EC) NORMAS DE LA CE
Direttiva Basso Voltaggio Low Voltage Directive Directiva de baja tensión
Direttiva EMC Compatibilità Elettromagnetica EMC electromagnetic compatibility directive EMC directiva de compatibilidad electromagnética
Norme armonizzate europee nell’ambito della direttiva European harmonized standards underdirective Las normas europeas armonizadas conforme a la directiva
K METERING PUMP IS TESTED AND CERTIFIED BY WQA TO NSF/ANSI 50 AND 61 FOR MATERIALS SAFETY.
2006/95/CE
2004/108/CE
2006/42/CE
ICON
2

GENERAL SAFETY GUIDELINES

Operating, installing, or maintaining the unit in any way that is not covered in this manual could cause death, serious personal injury, or damage to the equipment.
This manual use the following safety message icon:
Danger! Indicates a hazardous situation which, if not avoided, will result in death or serious injury.
Warning!
Indicates a hazardous situation which, if not avoided, could result in death or serious injury.
Important - A practice not related to personal injury or additional information.
Cross reference - An instance which refers to related information elsewhere in the same document

PURPOSE OF USE AND SAFETY

METERING PUMP IS INTENDED FOR CHEMICAL DOSING.
Do not use in explosive area (EX). Do not use with flammable chemicals. Do not use with radioactive chemicals.
Use after a proper installation.
Use the pump in accordance with the data and specifications printed on the label.
Do not modify or use in a manner inconsistent with the provisions of the operating manual.
Keep the pump protected from sun and water. Avoid water splashes.
In emergencies the pump should be switched off immediately. Disconnect the power cable from the power supply.
When using pump with aggressive chemicals observe the regulations concerning the transport and storage of aggressive fluids.
When installing always observe national regulations.
Manufacturer is not liable for any unauthorized use or misuse of this product that may cause injury, damage to persons or materials.
Pump must be accessible at all times for both operating and servicing. Access must not be obstructed in any way.
Feeder should be interlocked with a no-flow protection device.
Pump and accessories must be serviced and repaired by qualified and authorized personnel only.
Before any operation:
always read chemical Material Safety Data Sheet (MSDS);
always wear protective clothing;
always discharge the liquid end before servicing the pump.
empty and rinse the liquid end before work on a pump which has been used with hazardous or unknown chemicals.
This equipment requires regular maintenance to ensure potability requirements of the water and maintenance of improvements as declared by the manufaturer.
Feeder should be interlocked with a no-flow protection device to automatically shut-off the pumps when there is no flow!
Adequate measures shall be taken to prevent cross connection of chemicals!
Chemical feeding must be stopped during backwash cycles and periods of noflow as these conditions may introduce the potential for chemical overdosing. Not doing so may result in elevated chemical concentrations and hazerdous gas introduction into the pool or spa.
3

ENVIRONMENTAL SAFETY

Work area
Always keep the pump area clean to avoid and/or discover emissions.
Recycling guidelines
EWC code: 16 02 14
Always recycle according to these guidelines:
1. If the unit or parts are accepted by an authorized recycling company, then follow local recycling
laws and regulations.
2. If the unit or parts are not accepted by an authorized recycling company, then return them to
the nearest representative.
Waste and emissions regulations
Observe these safety regulations regarding waste and emissions:
• Dispose appropriately of all waste.
• Handle and dispose of the dosed chemical in compliance with applicable environmental
regulations.
• Clean up all spills in accordance with safety and environmental procedures.
• Report all environmental emissions to the appropriate authorities.

LABEL

Distributor
CODE: pump code MODEL: pump model
PUMP’S DATA
S/N: serial number

SPARE PARTS

Fig. 1. Product label.

QR code
For spare parts orders or any other communication, refer to the pump’s label. Code (CODE) and serial number (S / N) uniquely identify the pump.

Fig. 2. WQA label.

THIS METERING PUMP IS TESTED AND CERTIFIED BY WQA TO NSF/ANSI 50 AND 61 FOR MATERIALS SAFETY.
Data matrix
4
Transportation and storage
A not suitable transportation or storage can cause damages.
Use origianal box to pack the pump.
Observe storage conditions also for transportation.
Although packed, always protect the unit against humidity and the action of chemicals.
Before return the dosing pump to the manufacturer Repair service, drain the chemical from pump head and rinse it. Refer to Shutdown procedure.
Fill the PRODUCT SERVICE REPAIR FORM and send it with the dosing pump. Repair service is not accepted if PRODUCT SERVICE REPAIR FORM is missing.
DO NOT TRASH PACKAGING. USE IT TO RETURN THE PUMP.
Transportation and storage temperature ..... 10 ÷ 50°C (32 ÷ 122°F)
Umidity ...................................................... 95% relative humidity (not condensed)

Included into package

5
QUANTITY CONTENT K PLUS K CL PLUS K CO PLUS
n. 4 ø6 dibbles
n. 4
4,5 x 40 self tapping screws
n. 1 5 X 20 delayed fuse
n. 1
n. 1
level probe with axial foot filter (PVDF)
0,3 bar injection valve (PVDF)
m 2 delivery hose (PVDF)
m 2 suction hose (PE)
m 2
venting hose (PVC 4x6 transparent)
m 2,5 input signal cable
n.1 operating manual

DESCRIPTION

K PLUS

NSF/ANSI 61

K CO PLUS

NSF/ANSI 61

K CL PLUS

K PLUS is a constant or proportional dosing pump with level control for chemical feeding into water.
In Constant dosing mode pump doses a constant quantity regularly as configured by the user. In Proportional dosing mode pump doses a quantity proportionally to an input signal, digital (voltage free contact) or current (mA).
Working modes available:
constant
constant with 1-10 pulses divider
multiplier with 1-10 pulses divider
divider with 1-10 pulses divider
divider with 1-100 pulses divider
divider with 1-1000 pulses divider
mA current signal (0/4 mA = 0 pulses: 20mA = max pulses)
Flow rate is determined by the stroke length and by the stroke speed. The stroke length is adjustable from 0 to 100% using the stroke length adjustment knob. However dosing accuracy is guarantee within an adjustment range from 30% to 100%.
K CO PLUS works in constant mode.
The pump can be set to work in
constant
constant with 1-10 speed reducer.
K CL PLUS works in constant mode and has got level control.
NSF/ANSI 61
Self venting: KA PLUS
NSF/ANSI 61
6
The pump can be set to work in
constant
constant with 1-10 speed reducer.
KA is the K version with self-venting pump head. Self-venting pump head must be used when using chemicals that produce gas (i.e. hydrogen peroxide, ammonium, sodium hypoclorite at particular conditions). For connections Self-Venting pump head installation”.

Fig. 3. K PLUS - KA PLUS

1
2
3
4
5
K PLUS- detail on pump head with manual venting
Venting knob
Delivery valve
To manual venting hose
Suction valve
KA PLUS - detail on self venting pump head mod. LA, MA, NA
To self venting hose
Delivery valve
Suction valve
Level probe input
External signal input
Power supply
7

Features

Power Supply Fuse 230 VAC (180-270 VAC) 115 VAC (90-135 VAC) 24 VAC (20-32 VAC)
- 50/60 Hz 1 A
- 50/60 Hz 500 mA
- 50/60 Hz 2 A
12 VDC (10-16 VDC) 2 A
Environment temperature .........................10 ÷ 45°C (55 ÷ 113°F) - 50/60 Hz
Chemical temperature ..............................0 ÷ 50°C (32 ÷ 122°F) - 50/60 Hz
Transportation and storage temperature ...10 ÷ 50°C (55 ÷ 122°F) - 50/60 Hz
Installation class ......................................II
Pollution level ..........................................2
Audible noise ...........................................73,4 dbA (silenced: 70,4 dbA; ultrasilenced: 69,4 dbA)
Protection degree .....................................IP 65
Working UR % .......................................... 85% at t 40 °C; 70% at 50 °C (no condensing water)
Max installation height .............................1,5 m
Capacity .................................................. Table 1 and 2

Tab. 1. Capacity (manual and self venting models)

cc per
stroke *
min max
cc per
stroke *
min max
max
l/h
max
l/h
Flow
Flow
Min GPH
Min GPH
Max GPH
Max GPH
Mod.
K PLUS
1802 0,06 2 0,000016 0,53 0,06 0,19
1504 0,11 4 0,000029 1,06 0,11 0,37
1005 0,14 5 0,000037 1,32 0,14 0,46
0808 0,22 8 0,000058 2,11 0,22 0,74
0510 0,28 10 0,000074 2,64 0,28 0,93
0501 0,28 1 0,000008 0,3 0,03 0,09
0218 0,50 18 0,00013 4,76 0,50 1,67
Mod.
KA PLUS
1801 0,03 1 0,000008 0,26 0,03 0,09
1503 0,08 3 0,000021 0,79 0,08 0,28
103.5 0,10 3,5 0,000026 0,92 0,10 0,32
085.5 0,15 5,5 0,000040 1,45 0,15 0,51
057.5 0,21 7,5 0,000055 1,98 0,21 0,69
0213 0,37 13 0,000098 3,43 0,37 1,20
* cc per stroke: referred to cc/stroke with stroke length knob on 100%.
cc/h
cc/h
min
min
CAPACITY
pulse/
min
180
180
180
180
180
180
180
pulse/
min
180
180
180
180
180
180
Max
pressure
bar PSI
18 261
15 217
10 145
8 116
5 72
5 72
2 29
Max
pressure
bar PSI
18 261
15 217
10 145
8 116
5 72
2 29
peak Amps
(A)
230
115
VAC
VAC
2.7 1.45
2.7 1.45
2.7 1.45
2.7 1.45
2.7 1.45
2.7 1.45
2.7 1.45
peak Amps
(A)
230
115
VAC
VAC
2.7 1.45
2.7 1.45
2.7 1.45
2.7 1.45
2.7 1.45
2.7 1.45
Delivery
(PVDF)
Delivery
(PVDF)
Suction
hose
4 x 6 4 x 6 L
4 x 6 4 x 6 L
4 x 6 4 x 6 L
4 x 6 4 x 6 L
4 x 6 4 x 6 L
4 x 6 4 x 6 L
6 x 8 6 x 8 M
hose
4 x 6 4 x 6 LA
4 x 6 4 x 6 LA
4 x 6 4 x 6 LA
4 x 6 4 x 6 LA
4 x 6 4 x 6 LA
6 x 8 6 x 8 MA
hose
Suction
hose
Pump
head
Pump
head
8

Tab. 2. Capacity (compressed air model)

CAPACITY
Flow cc per
Modello
K AC PLUS
* cc per stroke: referred to cc/stroke with stroke length knob on 100%.
min
max
cc/h
1018 0.6 18 0.16 4.7 0.6 2 10 145 6 x 8 6 x 8 M
Min
l/h
GPH
Max GPH
stroke *
min max
Max
pressure
bar PSI
Delivery
hose
(PVDF)
Suction
hose
Pump
head

Manual stroke length adjustment

Materials

Max cc/stroke ( Construction Materials and Technical info) are referred to cc/stroke with stroke length knob on 100%. The stroke length knob adjusts the pump capacity. Press and rotate the knob when the pump is powered.
Dosing accuracy is guarantee within an adjustment range from 30% to 100%.
Note:if knob isn’t on 100% position then the pump will dose at pressure greater than the one declared on label.
✓ : standard ✗: options available
PVDF PP PPV0 PMMA PVC PE CE VETRO PTFE SS FKM B EPDM WAX SI
BOX
PUMP HEAD
DIAPHRAGM
BALLS
SUCTION HOSE
DELIVERY HOSE
VENTING HOSE
O RING
LEVEL PROBE/ FOOT FILTER
LEVEL PROBE CABLE
9

INSTALLATION

How to install metering pump

5 steps to install and start-up the pump:
1. Pump location
2. Piping connections (hoses, level probe, injection valve)
3. Wirings
4. Pump priming
5. Programming and start-up
The operator must be aware of safety precautions to prevent physical injury.

User health and safety

The work area

Pump location

POWER SUPPLY DISCONNECTION Disconnect power supply before you perform any installation or maintenance tasks. Failure to disconnect power will result in serious physical injury.
SAFETY EQUIPMENT Use safety equipment according to the company regulations. Use this safety equipment within the work area:
• Helmet
• Safety goggles (with side shields)
• Protective shoes
• Protective gloves
• Gas mask
THE WORK AREA Observe these regulations and warnings in the work area:
• Always keep the work area clean.
• Pay attention to the risks presented by gas and vapors in the work area.
• Avoid all electrical dangers. Pay attention to the risks of electric shock or arc flash hazards.
• Avoid water splashs and direct sun!
Pump must be installed on a stable support at a max 1,5 mt height from tank’s bottom.
Injection point must be higher than tank to avoid accidental chemical injection.
Otherwise, connect a multifunction valve on delivery pipeline.
INSTALLATION PUMP GUIDELINES
Install the pump
- in a safety place and fixed to the table / wall to avoid vibration problems;
- in an easy accessible place;
- in horizontal position.

Requirements for product positioning

10
Use only hoses compatibles with product to dose. See “Chemical compatibility table” page 31. If dosing product is not listed please consult full compatibility table or contact chemical’s manufacturer.
REQUIREMENTS FOR PRODUCT POSITIONING Only use fasteners of the proper size and material. Replace all corroded fasteners. Make sure that all fasteners are properly tightened and that there are no missing fasteners.

Fig. 4. Installation

1 - Dosing Pump 2 - Suction Hose 3 - Delivery Hose 4 - Injection Valve 5 - Air discharge 6 - Level Probe 7 - Foot Filter 8 - Power Cable 9 - Stand-by/alarm (if any)
4
3
8
9
1
Max 1,5 mt
2
5
6
7
11

PIPING CONNECTIONS

Foot filter / Level probe (included only in some models)
Level probe is assembled with a foot filter that avoid sediments priming probles. Install level probe on the bottom of the tank. Connect BNC level probe to the pump BNC input.
Warning: If there is a mixer installed into tank, install a suction lance instead of level probe / foot filter.
In case of replacement of level probe parts, follow the diagram below.

Fig. 5. Level probe assembling diagram.

STEP 5
STEP 4
INSERT RING AS SHOWN
STEP 3
A CLICK
STEP 2
INSERT FLOATER
INSERT PROBE
WITH N.O. CONTACT
UNTIL TO HEAR
STEP 1
INSERT RING AS SHOWN
12

Suction hose connection

Suction piping should be as short as possible and installed in vertical position to avoid air bubbles suction.
Completely unscrew tightening nut from pump’s head and remove assembling components: tightening nut, holding ring and pipe holder. Assembly as shown in fig. 5. Insert hose into pipe holder until it reaches the bottom. Lock hose on pump’s head by screwing down the tightening nut.
Hand-tighten the nuts firmly.
Do not use tongs or any other tool.

Fig. 6. Suction hose assembling

Suction hose
Threading nut
Holding ring
Pipe holder

Pump head / delivery hose assembling procedure

13
Suction and delivery valves must be in vertical position.
Delivery hose must be firmly fixed to avoid suddenly movements that could damage near objects
Completely unscrew tightening nut from pump’s head and remove assembling components: tightening nut, holding ring and pipe holder. Assembly as shown in fig.6. Insert hose into pipe holder until it reaches the bottom. Lock hose on pump’s head by screwing down the tightening nut.
Hand-tighten the nuts firmly.
Do not use tongs or any other tool.
Connect the other end of the hose to the injection valve using the same procedure.

Fig. 7. Delivery hose / pump head assembling

Delivery hose
Threading nut
Holding ring
Pipe holder
3
4
5
6

Injection valve

Injection valve must be installed on plant from water’s input. Injection valve will open at pressure greater than 0,3 bar. On request 1, 2, 3, 4 or 5 bar injection valve are available.

Venting hose

Insert one side of venting hose into discharge connector as shown in fig 8.
Insert other side of venting hose into product’s tank. During priming procedure product exceeding will flow into tank.

Fig. 8. Manual venting pump head model (K PLUS).

to delivery hose
Venting knob
To manual venting hose
To suction hose
For priming procedure see PRIMING.
it’s allowed to lightly bend venting hose.
During calibration procedure (“TEST”) insert venting hose into BECKER test-tube.
14
2
3
4
5
6

KA PLUS self venting pump head connection

Refer to fig. 9 for delivery and venting hose. Assembling procedures are the same described before.

Fig. 9. Self-venting models pump head: IA, LA, MA, (KA PLUS).

To self venting hose
To delivery hose
To suction hose
Suction, delivery and discharge valve are different.
15

WIRING

Preliminary checks

THE ELECTRICAL WIRINGS SHOULD BE CARRIED OUT BY AUTHORIZED AND QUALIFIED PERSONNEL ONLY IN ACCORDANCE WITH LOCAL REGULATIONS.
Before to proceed, verify the following steps:
1. Verify the data on nameplate.
Make sure that the electrical data on the nameplate of the motor corresponds to the electrical supply.
2. Verify the grounded power outlet.
The pump must be plugged to a grounded power outlet. Pump must be connected to a motor protection switch (Residual Current Circuit Breaker - MCCB).
3. Install a relay switch. Do not install it in parallel with heavy inductance load (for example: engines). See fig. 10.

Fig. 10. Electrical installation.

P - Dosing pump R - Relay I - Switch or safety device E - Electrovalve or inductance load A - Power supply
4. Verify peak Amps. 115 or 230 VAC pumps do not use motor overload protection.
Power supply
12 VDC connect the pump to a 55 Ah-12VDC battery
24 VDC connect the pump to a 200W stabilized power supply (verify
peak Amps)
5. Verify level probe “BNC” is connected as described in Foot filter / Level probe”..
16

Pump’s wiring

1
2
3
4
5
Connect external signal “BNC”to pump “INPUT”. This signal can be:
- water meter input
- mA signal input.

Fig. 11. Wirings

Power supply
Level probe input (not on KCO PLUS model)
INPUT for water meter or mA signal (only K PLUS)

Level alarm output (option)

If present, connect level alarm (blue and brown wires).
Level alarm is free contact and not fuse protected output. Max load relay output: 2A 250VAC.
17

PRIMING

Warnings

Manual priming

Feeder should be interlocked with a no-flow protection device to automatically shut-off the pumps when there is no flow!
Adequate measures shall be taken to prevent cross connection of chemicals!
Chemical feeding must be stopped during backwash cycles and periods of noflow as these conditions may introduce the potential for chemical overdosing. Not doing so may result in elevated chemical concentrations and hazerdous gas intro­duction into the pool or spa.
Never operate any pumping system with a blocked suction and discharge. You must take all necessary measures to avoid this condition.
SAFETY EQUIPMENT Use safety equipment according to the company regulations. Use this safety equipment within the work area:
• Helmet
• Safety goggles (with side shields)
• Protective shoes
• Protective gloves
• Gas mask
To prime the pump (only in CONSTANT working mode):
1. perform al pipings (delivery, suction and venting hose);
2. turn completely the venting knob to open discharge valve;
3. set STROKE LENGTH KNOB on 100% (for viscous liquids set between 50 and 70%);
4. supply the pump.
5. When the product will start to flow into venting hose, close the discharge valve turning the knob (not for self-venting model).

Automatic priming

18
For viscous liquids, to facilitate priming: insert a 20 cc syringe on venting pipe and suck;
When syringe is almost full close the discharge valve turning the knob..
1. Turn OFF the pump.
2. Keep pressed OFF key for 4 seconds.
3. Pump primes for 30 seconds.
4. Turn ON the pump.
The pump returns to the last working mode.

CONTROL PANEL

K PLUS K CL PLUS K CO PLUS

Keyboard function

PROGRAMMING MODE ENTER/EXIT
ON/OFF - SCROLL PROGRAMS
STROKE LENGHT ADJUSTMENT KNOB (0-100%)

K PLUS

- STROKE FREQUENCY ADJUSTMENT (yellow labelled scale 0-100%) or
- DIVIDER MULTIPLIER FACTOR ADJUSTMENT N (grey labelled scale N: 1-10)
K CO PLUS / K CL PLUS
- STROKE FREQUENCY ADJUSTMENT (yellow labelled scale 0-100%) WITH CONSTANT 0-100% PROGRAM
- DIVIDER FACTOR ADJUSTMENT N (grey labelled scale N: 1-10) WITH CONSTANT 0-10% PROGRAM

Tab. 3. Keys functions

OPERATION KEY
ON / OFF / AUTOMATIC PRIMING ON/OFF - SCROLL
ENTER / EXIT from PROGRAMS MENU P
CONFIRM PROGRAM P
SCROLL PROGRAMS ON/OFF - SCROLL
PROGRAMMS LED
Select a program to turn on the corresponding LED
Set the PROGRAM.
LED LEVEL (NOT ON K CO PLUS) Functions described in Led LEVEL.
19

LEVEL led

Red level led blinks in different ways described in the table

Tab. 4. Led LEVEL

LED STATE SOLUTION
Permanent red
3 blinks per second
2 blinks per second
1 blink per second
Product end (if present a level probe) / tank empty
Power supply is over the range (refer to pump label)
Power supply is under the range (refer to pump label)
Pump is waiting program setting
Fill the tank
Check power supply correspond to pump label. Shutdown and restart.
Check power supply correspond to pump label. Shutdown and restart.
Press scroll key and choose a program. Confirm with P key

PROGRAMS led

PROGRAMS led shows the current working program. Press repeatedly SCROLL to select hte working program

Tab. 5. Led PROGRAMS

LED STATE Permanent on Pump ON. Current pump working mode. 1 blink every 2
seconds on last
Pump OFF..
working program. All leds blinking
together
Pump is waiting for programming. Press P and SCROLL to select the program or wait 30 seconds to exit without changing.
20

PROGRAMMING THE PUMP

Start/shutdown

Set the PROGRAM

PROGRAMS

Connet power supply cable and start the pump with ON/OFF key. Led will be on the last program set (default setting: ). In OFF mode led will blink once every 2 seconds on the last program set (default setting: ).
• Keep pressed P for 4 seconds.
• Leds blink together.
• Press P.
• Press SCROLL and choose a program.
• Press P to confirm. Led will be on the program set.
If you do not press any key, after 30 seconds pump will esc from programming mode.
Each program has its own led.

Tab. 6. Programs menu

PROGRAMS WORKING MODE
1
mA
proportional dosing mode based on mA current signal
CONSTANT constant dosing mode
CONSTANT / DIVIDE
MULT 1÷10
1
costant dosing mode with pulses divider (to reduce up to 10 times pump capacity)
External pulses from a water meter are multiplied by a factor “N” from 1 to 10. Set “N” value turning DIVIDER MULTIPLIER FACTOR ADJUSTMENT KNOB (grey labelled scale N: 1-10).
DIV 1÷10
1
External pulses from a water meter are divided by a factor “N” from 1 to
10. Set “N” value turning DIVIDER MULTIPLIER FACTOR ADJUSTMENT KNOB (grey labelled scale N: 1-10).
External pulses from a water meter are divided by a factor “N” from 10 to
100. Set “N” value turning DIVIDER MULTIPLIER FACTOR ADJUSTMENT
DIV 10÷100
1
KNOB (grey labelled scale N: 1-10). Grey labelled scela 1-10 is proportional to the range 10-100. Adjust the knob on maximum value (10) is equivalent to setting the scale on 100.
External pulses from a water meter are divided by a factor “N” from 100 to 1000. Set “N” value turning DIVIDER MULTIPLIER FACTOR ADJUSTMENT KNOB (grey labelled scale N: 1-10).
DIV 100÷1000
1
Grey labelled scale 1-10 is proportional to the range 10-100. Adjust the knob on maximum value (10) is equivalent to setting the scale on 1000.
1
Not available on K CO PLUS and K CL PLUS models.
21

mA mode

Current from an external device (BNC input) drives the pump that doses proportionally according to the minimum and maximum set (0-20 mA or 4-20 mA).
To set press SCROLL until mA led turn on (red for 0-20 mA; green for 4-20 mA) and confirm with P key.

CONSTANT mode

CONSTANT with divider mode

To choose if...
... there is a mA current signal (controllers provided with proportional output in current), and you have to dose a certain amount of product.
STROKE LENGHT ADJUSTMENT KNOB (0-100%) acts percentually on pump capacity.
STROKE FREQUENCY ADJUSTMENT (yellow labelled scale 0-100%) acts on injection per minutes.
Pump doses at a constant rate set with stroke length adjustment knob.
To set press SCROLL until CONSTANT led turn on and confirm with P key.
To choose if...
... there is not an external signal, and you have to dose a certain amount of product regularly.
STROKE LENGHT ADJUSTMENT KNOB (0-100%) acts percentually on pump capacity.
STROKE FREQUENCY ADJUSTMENT (yellow labelled scale 0-100%) acts on injection per minutes.
Pump doses at a constant rate set with stroke length adjustment knob but this rate is divided by a factor up to 10.
To set press SCROLL until CONSTANT and DIV 1÷10 led turn on together, then confirm with P key.
To choose if...
... there is not an external signal, and you have to dose a certain amount of product regularly but pump capacity is too high
STROKE LENGHT ADJUSTMENT KNOB (0-100%) acts percentually on pump capacity.
DIVIDER FACTOR ADJUSTMENT KNOB (grey labelled scale 1-10%) set the divider factor 1-10 to reduce pump capacity.
22

MULT 1÷10 mode

External pulses are multiplied by a value set by MULTIPLIER FACTOR ADJUSTMENT KNOB.
To set press SCROLL until MULT 1÷10 led turn on, then confirm with P key.
DIV 1÷10 DIV 10÷100 DIV 100÷1000 mode

Calculate the N factor

To choose if...
STROKE LENGHT ADJUSTMENT KNOB (0-100%) acts percentually on pump capacity.
MULTIPLIER FACTOR ADJUSTMENT KNOB (grey labelled scale 1-10) set the multiplier factor 1-10 to increase pump capacity.
External pulses are multiplied by a value set by DIVIDER FACTOR ADJUSTMENT KNOB.
To set press SCROLL until DIV 1÷10 or DIV 10÷100 or DIV 100÷1000 led turn on, then confirm with P key.
To choose if...
STROKE LENGHT ADJUSTMENT KNOB (0-100%) acts percentually on pump capacity.
DIVIDER FACTOR ADJUSTMENT KNOB (grey labelled scale 1-10) set the divider factor to reduce pump capacity:
from 1 to 10 if in DIV 1÷10 mode
from 10 to 100 if in DIV 10÷100 mode
from 100 to 1000 if in DIV 100÷1000 mode
Use the formula:
[imp/l] x [cc]
[ppm] x [K]
... an external signal produces low pulses number. This working mode multiplies pulses from 1 to 10 to dose the correct product amount.
... an external signal produces high pulses number. This working mode divides pulses to dose the correct product amount.
——————— x 1000 = N
N value to set with FACTOR ADJUSTMENT KNOB [imp/l] pulses/litre from pulse emitter water meter [cc] single injection product amount of dosing pump (cubic centimetres) [ppm] part per million product amount (gr/m3) [K] product diluition coefficient.
Depending on N set working mode:
Result Working mode
N>1 DIV 1÷10 or DIV 10÷100 or DIV 100÷1000
N<1 Calculate 1/N then set the resul in MULT 1÷10
N=1 DIV 1÷10 or DIV 10÷100 or DIV 100÷1000 or MULT 1÷10
23

TROUBLESHOOTING

Tab. 7. Guide to troubleshooting

PROBLEM CAUSE REMEDY
• Collegare la pompa alla rete elettrica
• Replace fuse Fuse replacement procedure.
• Replace main board Main board replacement
procedure
.
• Clean the foot filter
• Prime the pump PRIMING
• Check valves, pipes and fittings
• Open venting knob and let air flow out. Use a self-venting pump head.
• Clean valves and ball valve. Feed 2-3 litres of water to wash valves and pump head
• Change valves
Pump does not start
Pump does not feed but solenoid runs
Pump does not feed, solenoid does not run or slightly run
• Pump not powered
• Protection fuse
• Main board
• Foot filter obstruction
• Pump head empty (suction pipe empty)
• Air bubbles into pump head or into suction pipe
• Product generates gas
• Valves and/or ball valves blocked
• Injection valve obstruction
If the problem can not be solved, please contac after-sales service or return the dosing pump to the manufacturer.

Repair service

Before return the dosing pump to the manufacturer Repair service, drain the chemical from pump head and rinse it. Refer to
Shutdown procedure.
If there is the possibility that residual corrosive liquid into pump head could cause damages, declare it on REPAIR FORM.
Fill the PRODUCT SERVICE REPAIR FORM and send it with the dosing pump. Repair service is not accepted if PRODUCT SERVICE REPAIR FORM is missing.
24

Fuse replacement procedure

Make sure that the product is isolated from the power supply and cannot be powered by mistake.
This procedure SHOULD BE CARRIED OUT BY AUTHORIZED AND QUALIFIED PERSONNEL
In order to replace fuse, you need these tools:i:
• a 3x16 screwdriver
• a 3x15 screwdriver
• fuse (see Features)
• Unplug power supply and pipings.
• Turn STROKE LENGHT ADJUSTMENT KNOB on 0%.
• Remove screws on the back of the pump.
• Pull back cover until it’s completed separated from pump’s front. Be careful of the knob’s spring.
• Locate the fuse and replace with a new one.
• Reassemble the pump. Be careful to put back the knob’s spring.
• Reinsert screws

Main board replacement procedure

Make sure that the product is isolated from the power supply and cannot be powered by mistake.
This procedure SHOULD BE CARRIED OUT BY AUTHORIZED AND QUALIFIED PERSONNEL
In order to replace main board, you need these tools:i:
• a 3x16 screwdriver
• a 3x15 screwdriver
• new main board
• Unplug power supply and pipings.
• Turn STROKE LENGHT ADJUSTMENT KNOB on 0%.
• Remove screws on the back of the pump.
• Pull back cover until it’s completed separated from pump’s front. Be careful of the knob’s spring.
• Remove boards screws..
• Completely disconnect wires from main board and replace it. Reinsert screws.
• Reconnect wires to the main board ( Main board scheme).
• Reassemble the pump. Be careful to put back the knob’s spring.
• Reinsert screws.
25

Main board

Fig. 12. Main board scheme

L N
COIL
­N.O.
OPTION
C
N.C.
+ level
input
-
+
26

MAINTENANCE

Maintenance schedule

Maintenance inspection

In order to ensure the requirements of potable drinking water treated and the maintenance of the improvements as declared by the manufacturer, this equipment must be checked at least once a month.
OPERATOR PROTECTION Use safety equipment according to the company regulations.
Use this safety equipment within the work area during installation, service and when handling chemicals:
• protective mask
• protective gloves
• safety goggles
• ear plugs or hear muffs
• further security device, if necessary.
POWER SUPPLY DISCONNECTION Always disconnect power to the motor before you perform any installation or maintenance tasks. Failure to disconnect power will result in serious physical injury.
Installation
and maintenance tasks should be carried out by AUTHORIZED AND
QUALIFIED PERSONNEL only in accordance with local regulations.
Use original spare parts.
Shutdown the dosing pump before any maintenance operation Shutdown
procedure.
A maintenance schedule includes these types of inspections:
• Routine maintenance and inspoections
• Three-month inspections
• Annual inspections
Shorten the inspection intervals appropriately if the pumped chemical is abrasive or corrosive.
Routine maitenance and inspections
Perform these tasks whenever you perform routine maintenance:
• Inspect the seal. Ensure that there are no leaks from the mechanical seal.
• Check electrical wiring
• Check for unusual noise and vibration (noise allowed 74 dbA; ± 5 dB).
• Check the pump and piping for leaks.
• Check for corrosion on parts of the pump and / or on hoses.
Three-month inspections
Perform these tasks every three months:
• Check that the tightenings.
• Check the mechanical seal if the pump has been left idle.
Annual inspections
Perform these inspections one time each year:
• Check the pump capacity (as per nameplate).
• Check the pump pressure (as per nameplate).
• Check the pump power (as per nameplate).
27
f the pump performance does not satisfy your process requirements, and the process requirements have not changed, then perform these steps:
1. Disassemble the pump.
2. Inspect it.
3. Replace worn parts.

Shutdown procedure

This procedure SHOULD BE CARRIED OUT BY AUTHORIZED AND QUALIFIED PERSONNEL
OPERATOR PROTECTION Use safety equipment according to the company regulations.
Use this safety equipment within the work area during installation, service and when handling chemicals:
• protective mask
• protective gloves
• safety goggles
• ear plugs or hear muffs
• further security device, if necessary.
Shutdown the dosing pump before any maintenance operation or before long downtimes. Disconnect power and ensure it cannot be restarted.
Depressurize the system. The liquid may leak splashing.
Drain the chemical from pump head. Release the pressure and disconnect the disharge pipe from the discharge valve. Rinse the pump head and clean all valves.
28

Delivery curves

Flow rate indicated is for H Dosing accuracy ± 2% at constant pressure ± 0,5 bar.
O at 20°C at the rated pressure.
2

Fig. 13. K PLUS delivery curves

1802: l/h 2 bar 18
Pump head mod. L
l/h
1005: l/h 5 bar 10
Pump head mod. L
l/h
1504: l/h 4 bar 15
Pump head mod. L
bar
bar
l/h
bar
0808: l/h 8 bar 8
Pump head mod. L
l/h
bar
0510: l/h 10 bar 5
Pump head mod. L
l/h
bar
29
0501: l/h 1 bar 5 Pump head mod. I
l/h
0218: l/h 18 bar 2
Pump head mod. M
l/h
bar
bar
30

Fig. 14. K A PLUS delivery curves

1801: l/h 1 bar 18
Pump head mod. LA
l/h
103,5: l/h 3,5 bar 10
Pump head mod. LA
l/h
057,5,5: l/h 7,5 bar 5
Pump head mod. LA
l/h
1503: l/h 3 bar 15
Pump head mod. LA
l/h
bar
bar
085,5: l/h 5,5 bar 8
Pump head mod. LA
l/h
bar
bar
0213: l/h 13 bar 2
Pump head mod. MA
l/h
bar
bar
31
32

Dimensions






 

Fig. 15. Dimensions
 



33

COMPATIBILITY TABLE

Chemical compatibility table

Solenoid driven metering pumps are widely used to dose chemical fluids and it is important that the most suitable material in contact with fluid is selected for each application. This compatibility table serves as a useful help in this respect. All the informations in this list are verified periodically and believed to be correct on the date of issuance. All the informations in this list are based on manufacturer’s data and its own experience but since the resistance of any material depends by several factors this list is supplied only as an initial guide, in no way manufacturer makes warranties of any matter respect to the informations provided in this list.

Tab. 8. Chemical compatibility table.

Product Formula Ceram. PVDF PP PVC SS 316 PMMA Hastel. PTFE FPM EPDM NBR PE
Acetic Acid, Max 75% CH3COOH 2 1 1 1 1 3 1 1 3 1 3 1
Hydrochloric Acid, Concentrate HCl 1 1 1 1 3 1 1 1 1 3 3 1
Hydrofluoric Acid 40% H2F2 3 1 3 2 3 3 2 1 1 3 3 1
Phosphoric Acid, 50% H3PO4 1 1 1 1 2 1 1 1 1 1 3 1
Nitric Acid, 65% HNO3 1 1 2 3 2 3 1 1 1 3 3 2
Sulphuric Acid, 85% H2SO4 1 1 1 1 2 3 1 1 1 3 3 1
Sulphuric Acid, 98.5% H2SO4 1 1 3 3 3 3 1 1 1 3 3 3
Amines R-NH2 1 2 1 3 1 - 1 1 3 3 1 1
Sodium Bisulphite NaHSO3 1 1 1 1 2 1
Sodium Carbonate (Soda) Na2CO3 2 3 1 1 1 1 1 1 2 1 1 1
Ferric Chloride FeCl3 1 1 1 1 3 1 1 1 1 1 1 1
Calcium Hydroxide (Slaked Lime) Ca(OH)2 1 1 1 1 1 1 1 1 1 1 1 1
Sodium Hydroxide (Caustic Soda) NaOH 2 3 1 1 1 1 1 1 2 1 2 1
Calcium Hypochlor.(Chlor.ted Lime)
Sodium Hypochlorite, 12.5% NaOCl + NaCl 1 1 2 1 3 1 1 1 1 1 2 3
Potassium Permanganate, 10% KMnO4 1 1 1 1 1 1 1 1 1 1 3 1
Hydrogen Peroxide, 30% (Perydrol) H2O2 1 1 1 1 1 3 1 1 1 3 3 1
Aluminium Sulphate Al2(SO4)3 1 1 1 1 1 1 1 1 1 1 1 1
Copper-II-Sulphate (Roman Vitriol)
1
Calcium Hypochlor.(Chlor.ted Lime): WQA test was based on 1% Calcium Hypochlorite solution.
1
Ca(OCl)2 1 1 1 1 3 1 1 1 1 1 3 1
CuSO4 1 1 1 1 1 1 1 1 1 1 1 1
1 1 1 1 1 1

Materials

34
Polyvinyldene fluoride (PVDF) .................Pump heads, Valves, Fittings
Polypropylene (PP) ..................................Pump heads, Valves, Fittings
PVC .......................................................Pump heads
Stainless steel (SS 316) ...........................Pump heads, Valves
Polymethyl Metacrilate Acrylic (PMMA) ...Pump heads
Polytetrafluoroethylene (PTFE) ................Diaphragm
Fluorocarbon (FPM) ................................O-ring
Ethylene propylene (EPDM) .....................O-ring
Nitrile (NBR) ...........................................O-ring

Hose resistance table

Hose features are very important for a reliable dosage. Every pump’s model is made to work in the best way using selected hoses according to pump’s capacity / model. Information reported here are intended for standard use only. For extended information ask to hose’s manufacturer.

Tab. 9. Hoses features

35
36
MOD 7.5 B1 Q
PRODUCT SERVICE REPAIR FORM
Ed. 1 - rev. 0 21/02/2012
ENCLOSE THE PRESENT FORM TO THE DELIVERY NOTE
DATE ............................................
SENDER
Company name ............................................................................................................................................
Address ................................................................................................................................................
Phone no. ................................................................................................................................................
Contact person.............................................................................................................................................
PRODUCT TYPE (see product label)
DEVICE CODE ..............................................................................................................................................
S/N (serial number).......................................................................................................................................
OPERATING CONDITIONS
Location/installation description
..................................................................................................................
...................................................................................................................................................................
Chemical ................................................................................................................................................
Start-up (date) ............................................ Running time (approx. hours) ....................................................
REMOVE ALL THE LIQUID INTO THE PUMP HEAD AND DRY IT BEFORE PACKAGING IN ITS ORIGINAL BOX.
DESCRIPTION OF PROBLEM
MECHANICAL
Wear parts
.................................................................................................................................
Brekage/other damages .............................................................................................................
Corrosion ...................................................................................................................................
Other .........................................................................................................................................
ELECTRICAL
Connections, connector, cables ...................................................................................................
Operating controls (keyboard, display, etc.) .................................................................................
Elettronics ..................................................................................................................................
Other .........................................................................................................................................
LEAKS
Connections ...............................................................................................................................
Pump head ................................................................................................................................
NOT OR INADEQUATE FUNCTION/OTHER
.................................................................................................................................................
.................................................................................................................................................
.................................................................................................................................................
I declare that the dosing pump is free of any hazardous chemical.
37
Signature of the compiler Company stamp
TABLE OF CONTENTS
GENERAL SAFETY GUIDELINES
PURPOSE OF USE AND SAFETY
ENVIRONMENTAL SAFETY
LABEL
..........................................................4
SPARE PARTS
Included into package
DESCRIPTION
K PLUS ........................................................... 6
K CO PLUS ..................................................... 6
K CL PLUS ...................................................... 6
Self venting: KA PLUS .................................... 6
Features ......................................................... 8
Manual stroke length adjustment.................... 9
Materials ........................................................ 9
INSTALLATION
How to install metering pump......................... 10
User health and safety .................................... 10
The work area ................................................ 10
Pump location ................................................ 10
Requirements for product positioning ............. 10
PIPING CONNECTIONS
Foot filter / Level probe .................................. 12
(included only in some models) ....................... 12
Suction hose connection ................................. 13
Pump head / delivery hose assembling procedure
...................................................................... 13
Injection valve ................................................ 14
Venting hose .................................................. 14
KA PLUS self venting pump head connection . 15
WIRING
Preliminary checks .......................................... 16
Pump’s wiring ................................................ 17
Level alarm output (option) ............................ 17
PRIMING
Warnings ....................................................... 18
Manual priming .............................................. 18
Automatic priming ......................................... 18
CONTROL PANEL
Keyboard function .......................................... 19
K PLUS
K CL PLUS....................................................19
K CO PLUS
LEVEL led
PROGRAMS led .............................................. 20
PROGRAMMING THE PUMP
Start/shutdown .............................................. 21
Set the PROGRAM .......................................... 21
PROGRAMS ................................................... 21
mA mode ....................................................... 22
CONSTANT mode ........................................... 22
38
...............................................4
.................................... 5
..............................................6
.............................................10
.......................................................16
.....................................................18
.........................................19
........................................................19
...................................................19
....................................................... 20
...................2
...................3
...........................4
................................12
........................21
CONSTANT with divider mode MULT 1÷10 mode DIV 1÷10 DIV 10÷100 DIV 100÷1000 mode
Calculate the N factor..................................... 23
TROUBLESHOOTING
Repair service ................................................. 24
Fuse replacement procedure ........................... 25
Main board replacement procedure ................ 25
Main board .................................................... 26
MAINTENANCE
Maintenance schedule .................................... 27
Maintenance inspection ................................ 27
Shutdown procedure
Delivery curves ............................................... 29
Dimensions
COMPATIBILITY TABLE
Chemical compatibility table ........................... 34
Materials ........................................................ 34
Hose resistance table ...................................... 35
.......................................... 23
...........................................27
.................................................... 33
....................... 22
....................................24
...................................... 28
................................34
. 23
FIGURES
Fig. 1. Product label. ................................................................ 4
Fig. 2. WQA label. .................................................................... 4
Fig. 3. K PLUS - KA PLUS ......................................................... 7
Fig. 4. Installation .................................................................... 11
Fig. 5. Level probe assembling diagram. ................................... 12
Fig. 6. Suction hose assembling ................................................ 13
Fig. 7. Delivery hose / pump head assembling ........................... 13
Fig. 8. Manual venting pump head model (K PLUS). .................. 14
Fig. 9. Self-venting models pump head: IA, LA, MA, (KA PLUS). 15
Fig. 10. Electrical installation. ..................................................... 16
Fig. 11. Wirings ......................................................................... 17
Fig. 12. Main board scheme ....................................................... 26
Fig. 13. K PLUS delivery curves ................................................... 29
Fig. 14. K A PLUS delivery curves ................................................ 31
Fig. 16. Dimensions ................................................................... 33
TABLES
Tab. 1. Capacity (manual and self venting models) ................... 8
Tab. 2. Capacity (compressed air model) ................................... 9
Tab. 3. Keys functions .............................................................. 19
Tab. 4. Led LEVEL .................................................................... 20
Tab. 5. Led PROGRAMS ........................................................... 20
Tab. 6. Programs menu ............................................................ 21
Tab. 7. Guide to troubleshooting .............................................. 24
Tab. 8. Chemical compatibility table. ........................................ 34
Tab. 9. Hoses features .............................................................. 35
39
ETA Ekipman Teknoloji Ar. Sis. San. Tic. Ltd. Şti.
www.etaekipman.com info@etaekipman.com Tel:0 216 540 99 64 Fax:0 216 540 99 68
Loading...