Ergotron StyleView SV32-S, StyleView SV32-T, StyleView SV32 PHD, StyleView SV32, StyleView SV42 Installation Manual

1 of 10
888-83-065-G-01 rev. E • 03/19
10mm (3/8") or 8mm (5/16")
SV32-S SV32-T
SV32
3 6 8
www.ergotron.com |
USA: 1-800-888-8458
|
Europe: +31 (0)33-45 45 600
|
China: 400-120-3051
|
Japan: japansupport@ergotron.com
English, Español, Français, Deutsch, Nederlands, Italiano, Svenska, 日本語, 汉语
10 °C 50 °F
29 °C
86 °F
Relative
Humidity
Range
5-95% rH
Operational
-20 °C
- 4 °F
50 °C 122 °F
Relative
Humidity
Range
5-95% rH
Storage
BATTERY
LEAD
Please dispose of all batteries in accordance with local law
Two Sealed Lead Acid, Absorbed Glass Mat, 12VDC batteries.
Depends on battery
StyleView® SLA
Change Power System Batteries
· SV32-S
· SV32-T
· SV32
· SV32 PHD
· SV42
English
For the latest User Installation Guide please visit: www.ergotron.com
User's Guide
2 of 10
888-83-065-G-01 rev. E • 03/19
Maintenance & Safety
Color Symbol
Signal
Word
Level of Hazard
Red
DANGER
Indicates an imminently hazardous situation which, if not avoided, will result in death or serious injury.
Orange
WARNING
Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
Yellow
CAUTION
Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury.
None
CAUTION
Used without the safety alert symbol indicates a poten­tially hazardous situation which, if not avoided, may result in property damage.
Red,
Orange
or
Yellow
Electrical
Indicates an impending electrical hazard which, if not avoided, may result in personal injury,  re and/or death.
Hazard Symbols Review
The Meaning of Symbols appearing in this Guide, on the Cart or on the Power System These symbols alert you to a safety condition that demands your attention. You should be able to recognize and understand the signi cance of the following Safety Hazards if you encounter them on the Cart or within Cart documentation such as this Set-up Guide.
33 Amp-Hour Batteries
BATTERY
LEAD
WARNING: RISK OF ELECTRICAL DISCHARGE
• Do NOT swap battery cables, doing so will cause arcing and trip the circuit breaker.
• Do not remove or replace the batteries while cart is located in an oxygen rich or hazardous environment, arcing may occur and cause combustion.
• Replace  rst battery before removing second battery to reduce the risk of cables touching terminals and causing arcing.
• Always replace both batteries!
• Replace  rst battery before removing second battery.
• Replace batteries with same Amp/Hour rating batteries only.
• Only Ergotron-speci ed batteries may be used in the StyleView Power System. Please call customer care for more details.
• Recycle battery or contact Ergotron for proper battery disposal guidelines.
20 Amp-hour Batteries
BATTERY REPLACEMENT
The power module is congured for two, 33 -Ah batteries. Only the following batteries are compatible with this system:
• B&B EP33-12 • SBS S-12330
• Longway 6FM33G • Vision 6FM33D
• Vision 6FM33HD-X • PBQ PBQ33-12L
• Werker WKA12-33C • Werker WKA12-35C
• Duracell Dura 12-35C
Installing batteries other than the 33 -Ah listed above will void the product warranty. Failure to heed this warning may result in severe damage to batteries, power module and possible re hazard.
REMPLACEMENT DE LA BATTERIE
Le module d’alimentation est conguré pour deux batteries 33 A-h. Seules les batteries suivantes sont compatibles avec ce système :
• B&B EP33-12 • SBS S-12330
• Longway 6FM33G • Vision 6FM33D
• Vision 6FM33HD-X • PBQ PBQ33-12L
• Werker WKA12-33C • Werker WKA12-35C
• Duracell Dura 12-35C
L’installation de batteries autres que les batteries 33 A-h citées ci-dessus annulera la garantie du produit. Si vous ne vous conformez pas à cette précaution d’utilisation, les batteries et le module d’alimentation risquent d’être sérieusement endommagés, ce qui pourrait provoquer un incendie.
WARNING
AVERTISSEMENT
822-386
www.ergotron.com
3 of 10
888-83-065-G-01 rev. E • 03/19
bacd
12
3
4
5
SV32-S SV32-T SV32
Turn o all mounted equipment. Disconnect Power System from
power source.
Turn power system o by holding down the AC Outlet Power button for 1 - 3 seconds. Power light will shut o .
If your cart has circuit breakers, switch them both to "OFF".
Loading...