201000433000 • Version 2017-01-12 / Rev 01 • English
ergoselect 4 / 5
Ergometer
Operator‘s Manual
201000433000 • Version 2017-01-12 / Rev 01 • English
This manual was written with the utmost care. Should you still nd details that do not correspond
with the system, please letus know and we will correct the issue as soon as possible.
We reserve the right to modify the design and technical features of the device and are not
bound by the information and illustrations provided in this manual.
All trademarks appearing in this document are trademarks of their respective owners. Their
protection is acknowledged.
No part of this manual may be reprinted, translated or reproduced without the manufacturer‘s
written permission.
This manual will not be automatically updated. Please contact the manufacturer for the
latest document revision.
• The product ergoselect bears the CE marking CE-0123
(Notied Body: TÜV), indicating its compliance with the
provisions of the Council Directive 93/42/EEC about
medical devices and fullls the essential requirements
of Annex I of this directive.
The CE marking covers only the accessories listed in the
Order Information chapter. The device is an MDD class
IIa product.
• The device fullls the requirements of standard
EN 60601-1 “Medical electrical equipment, Part 1:
General Requirements for Safety” as well as the
in terference protection requirements of standard
EN 60601-1-2 “Electromagnetic Compatibility –
Medical Electrical Devices”.
The radio-interference emitted by this product is within
the limits specied in EN 55011, class B.
• The symbol means: protection class II.
• This manual is an integral part of the device. It should
be available to the device operator at all times. Close
observance of the information given in the manual
isaprerequisite for proper device performance and
correct operation and ensures patient and operator
safety. Please note that information pertinent to several chapters is given only once. Therefore carefully read
the manual once in its entirety.
• The symbols mean:
Consult accompanying documents.
They indicate points that are of particular importance
in the operation of the device.
• Observance of the safety information protects from
injuries and prevents inappropriate use of the device.
All device users and persons responsible for assembly,
maintenance, inspection and repair of the device must
read and understand the content of this manual, before
using or working with it. Paragraphs with special symbols are of particular importance.
• The ergoline quality management system complies
with the standards ISO 9001: 2008 and EN ISO 13485:
2003-AC2007.
• The safety information given in this manual is classied
as follows:
Danger
indicates an imminent hazard. If not avoided, the hazard will
result in death or serious injury.
Warning
indicates a hazard. If not avoided, the hazard may result in
minor injury and/or product/property damage.
Caution
indicates a potential hazard. If not avoided, the hazard may
result in minor injury and/or product/property damage.
• To ensure patient safety, the specied measuring
accuracy, and interference-free operation, we recommend using only original ergoline accessories. The user
is responsible if accessories from other manufacturers
are used.
• ergoline is responsible for the safety, reliability, and
performance of the device, only if
— modications and repair are carried out by ergoline
GmbH or by an organization expressly authorized
by ergoline GmbH.
— the device is used in accordance with the instruc-
tions given in this operator’s manual.
• If unauthorized individuals open the control terminal,
damaging the calibration sticker, any warranty claim
shall become void.
• This manual reflects the device specifications and
applicable safety standards valid at the time of printing.
All rights are reserved for devices, circuits, techniques,
software programs, and names appearing in this manual.
• On request ergoline will provide a Field Service Manual.
The device is not designed for use in areas where an explosion
hazard may occur.
Explosion hazards may result from the use of ammable anesthetics, skin cleansing agents, or disinfectants.
Warning
• Patient Hazard, Equipment Damage •
Do not expose the ergoselect to direct sunlight to prevent
system components from reaching inadmissible high temperatures.
Do NOT use the ergoselect outdoors (medical device). Furthermore the device has no additional protection against the
ingress of humidity. Humidity inside the device may cause
equipment malfunctions and increases the risk of an electric
shock.
Additionally, the device should not be operated in the vicinity
of power systems, because they may impair equipment functions.
The ergoselect may only be used in combination with accessories approved by ergoline GmbH.
Warning
• Shock Hazard •
When the ergometer is connected to other equipment or if a
medical system is created, it must be ensured that the added
leakage currents do not present a hazard. In case of questions,
please contact your ergoline dealer or the ergoline GmbH
Service Department.
For use, the ergometer must always be connected to electric
installations that fulll the local requirements.
• Patient Hazard •
The German Medical Device Operator Ordinance (MPBetreibV, § 5)
demands that users
• must be trained in the use of the ergometer
• must be familiar with the routines for handling and
assembly of the ergometer
• must be familiar with and observe the safety rules and
regulations for operation of this type of equipment
• must be informed about any other pertinent rules and
regulations (e.g. safety instructions)
• must be informed about the potential hazards arising from
the use of this type of equipment.
• make sure that no unauthorised changes are carried out.
• Risk to Persons •
Before using the ergometer, the operator must ascertain that it
is in correct working order and operating condition. The cables
and connectors, in particular, must be checked for signs of
damage. Damaged parts must be replaced immediately.
• Equipment Malfunction •
Only the special shielded cables supplied by ergoline may be
used to connect the device to other pieces of equipment.
• Equipment Malfunction •
Cellular telephones may not be used in the immediate vicinity
of the ergometer, because they might interfere with the proper
functioning of the ergometer.
Electromagnetic interference most probably exists when the
watt reading is unstable. If the displayed value changes frequently even though the speed is above 30 RPM, this may be
due to electro magnetic interference.
Note
Only the removal of the power cord will result in an all pole
disconnection of the device from mains.
Caution
Additional equipment connected to medical electrical equipment must comply with the respective IEC or ISO standards
(e.g., IEC 60950 for data processing equipment).
Furthermore all congurations shall comply with the requirements for medical electrical systems (see IEC 60601-1-1 or
clause 16 of the 3Ed. of IEC 60601-1, respectively).
Anybody connecting additional equipment to medical electrical equipment congures a medical system and is therefore
responsible that the system complies with the requirements for
medical electrical systems. Attention is drawn to the fact that
local laws take priority over the above mentioned requirements .
If in doubt, please consult your local dealer or ergoline GmbH.
Safety Information for
Non‑Invasive Blood Pressure Measurement
Warning
• Patient Hazard •
Do not take blood pressure measurements with a cuff on
patients suering from sickle cell anemia or if skin lesions are
likely to occur.
The cu may cause hematomas in patients with severe blood
coagulation disease. In these instances, the user must take a
decision for or against automatic blood pressure measurements .
Caution
• Compromised Measuring Accuracy •
Arrhythmias occurring frequently during a measurement may
compromise the accuracy of the measurement.
Valid measurements may not be possible under certain circumstances.
Electromagnetic elds are also capable of impairing the measuring accuracy.
Note
• If the cuff pressure exceeds the maximum value of
300 mmHg during ination, the ination procedure will
be aborted and the cu deated. As a redundant safety
precaution, the cu is immediately deated when the cu
pressure exceeds 320 mmHg.
You can check the proper functioning of this safety precaution by abruptly bending your arm while the cu is being
inated, causing a brief overpressure in the cu. The cu
must deate immediately.
• Measurements that do not yield a valid measurement will
not be repeated during the exercise test.
• If the ination phase takes longer than 40 seconds or if an
adequate pressure does not build up in the cu within a
reasonable period of time, the measurement will be aborted
and the cu deated.
• If a valid measurement cannot be completed within
120 seconds, the measurement will be aborted and the cu
deated.
• If the cu pressure remains constant for some time, the
measurement will also be aborted and the cu deated.
Intended Use
The ergoselect is a computer-controlled medical ergometer,
which operates at pedal speeds between 30 and 130 RPM
and loads between 6 and 999 W.
The speed-independent range is shown in section11.5 on
page53.
The ergoselect ergometer may only be used in exercise
testing and for rehabilitation of cardiac and cardiovascular
patients according to the instructions given in this manual.
If the ergometer is used for other purposes, the manufacturer cannot be held liable for personal injuries or property
damage resulting from the unintended use of the equipment.
Note ‑ Applied Parts
• Applied parts are components that are directly in contact
with the human body (e.g., blood pressure measuring
devices).
Note ‑ Stability
• Ensure the stability of the ergometer. If the maximum permitted patient weight is exceeded by 10%, the stability of
the ergometer can no longer be guaranteed. It may become
unstable as a result.
Biocompatibility
The parts of the product described in this manual, including
all accessories that come in contact with the patient during
the intended use, fulll the biocompatibility requirements
of the applicable standards if applied as intended.
If you have questions in this matter, please contact ergoline
or a representative.
• EN 1060-1 Non-invasive sphygmomanometers, Part1:
General requirements
• EN 1060-3 Non-invasive sphygmomanometers, Part3:
Supplementary requirements for electro-mechanical
blood pressure measuring systems Symbole
ergoselect 4/5 7
3 Symbols
3 Symbols
Symbol ’type B applied part’.
Type B applied parts have no direct contact with
patients and oer the lowest protection against
electric shock.
Symbol ’type BF applied part’.
Type BF applied parts are connected to the body of
the patient and provide a higher degree of protection against electric shock. The applied parts are
isolated.
Caution: Consult accompanying documents.
Protection class II equipment.
This symbol indicates that the waste of electrical
and electronic equipment must not be disposed of
as unsorted municipal waste and must be collected
separately.
Consult Operating Manual!
Order number.
Date of manufacture.
The number found under this symbol is the year of
manufacture in the YYYY format.
PVC-free.
Latex-free.
Suitable for indicated arm circumference.
Small size.
Standard size.
Large size.
18
17
01
gemäss MPBetreibV
02
03
19
nächster
Prüftermin
04
05
Serial number.
20
21
Scheduled date of the next inspection
22
12
(e.g., March 2017).
11
10
09
08
07
06
Toggle switch ON (voltage).
Toggle switch OFF (voltage).
CE mark per the Medical Device
Directive 93/42/EEC of the European Union.
Notiedbody: TÜV SÜD Product Service GmbH,
Ridlerstr. 65, 80339 München, Germany.
Nationally Recognized Testing Laboratory
NRTL label for the USA and Canada.
Manufacturer’s identication.
Transport and storage label:
top.
Transport and storage label:
keep dry.
Transport and storage label:
fragile.
Transport and storage label:
approved temperature range.
Transport and storage label:
approved humidity, non-condensing.
Transport and storage label:
approved pressure range.
Transport and storage label:
do not stack.
8 ergoselect 4/5
4 Setup and Mains Connection
4.1 Controls and Indicators
4 Setup and Mains Connection
1
Control terminal M
2
Speed readout for patient
3
Connectors (e.g., for blood pressure cu, SpO2)
4
Adjustment of handlebar angle
5
Castors
6
Baseplate (small)
7
Leveling feet to adjust the ergometer to uneven oors
8
Sockets for power cord and connection cables
(underside of ergometer)
9
Power switch (toggle switch [I / 0])
10
Saddle adjustment with clamping lever
1
3
2
4
6
5
Figure 4 – 1: ergoselect 4 M
10
9
8
7
1
Control terminal T
2
Speed readout for patient
3
Connectors (e.g., for blood pressure cu, SpO2)
4
Adjustment of handlebar angle
5
Height adjustment for handlebar
(by means of toggle switch, option)
6
Castors
7
Baseplate (large)
8
Leveling feet to adjust the ergometer to uneven oors
9
Sockets for power cord and connection cables
(underside of ergometer)
10
Power switch (toggle switch [I / 0])
11
Saddle adjustment
(optionally with gas pressure spring or motor)
1
3
5
2
4
7
6
Figure 4 – 2: ergoselect 5 T
11
10
9
8
ergoselect 4/5 9
4 Setup and Mains Connection
4.2 Transport
For short distances, the ergometer can be lifted at the
saddle and rolled away on its castors.
To cover greater distances, however, we recommend the
following method:
• Disconnect the power cord and the connection cables.
• Rotate the handlebar towards the front and tighten the
clamping lever.
• Stand in front of the ergometer, grasp the handlebar
and tilt the ergometer towards you until it is standing
on the castors only and is balanced.
• It is now possible to transport the ergometer.
• When you have reached the new location, lower the
ergometer very carefully to avoid damage.
Caution
• Equipment Damage •
Avoid strong vibrations of the ergometer during transport.
Figure 4 – 3: Transporting the ergoselect
4.3 Setup
Place the ergometer on a horizontal level oor.
The ergometer must be set up in a secure and stable
position; the two leveling feet at the back make for easy
adjustment to uneven oors. An optional stabilizer plate is
available to enhance the stability.
Extend the foot concerned until the ergometer no longer
wobbles.
In case of delicate ooring, it is recommended to place
a mat under the ergometer to protect the ooring from
damage by the feet.
Figure 4 – 4: Levelling feet of the ergoselect ergometer
10 ergoselect 4/5
4.4 Mounting the Control Terminal
The control terminal can be installed with the display either
facing the patient or the operator.
It is recommended to install the terminal with the display
and control keys towards the operator and the speed readout towards the patient.
4.5 Connecting the Power Cord
Stand in front of the ergometer (looking at the control
panel) and rmly grasp the handlebar with both hands.
4 Setup and Mains Connection
Figure 4 – 5: Different orientations of the control terminal
Then tilt the ergometer carefully to one side (it is recommended to do this with the help of a second person) and
place it on the oor so that it rests on the handlebar.
Figure 4 – 6: Assembly position
Caution
• Equipment Damage •
Before connecting the ergometer to the power line, check that
the line voltage corresponds to the ratings on the type plate.
The type plate is located on the back of the ergometer, at the
bottom.
ergoselect 4/5 11
4 Setup and Mains Connection
The connection panel is located on the underside of the
ergometer.
• Connect the power cord to socket (a).
• Using the supplied strain relief, attach the cable to the
metal frame (b).
Return the ergometer carefully to its upright position and
make sure that it is not standing on the power cord.
b
a
Figure 4 – 7: Underside of the Ergometer
b
a
Figure 4 – 8: Connecting the power cord
a Power input
b Strain relief
Note
• Disconnection from Power Supply •
Pressing the power switch or removing the power cord disconnects the device from the power supply.
Removing the power cord results in a complete disconnection
of the device from the power supply (all poles).
Ensure that the power plug is readily accessible at all times.
12 ergoselect 4/5
4.6 Connecting the ECG Cable
The ergometers can be connected to electrocardiographs
and PC-based ECG systems of most manufacturers.
Dierent connection cables are available to support dierent
communication modes (digital, analog, remote start, etc.).
All ergometers are equipped with a digital interface. (Special adapters are needed for analog control or the remote
start capability. Please contact ergoline for these adapters.)
The connection cable is plugged into the 9-pole socket of
the connection panel (Port 1) or the USB port and secured
at the metal frame with an additional strain relief.
4 Setup and Mains Connection
USBPort 1USBPort 1
Figure 4 – 9: Connection to electrocardiograph/PC-based ECG system
USB PC connection via USB (virtual COM)
PORT 1 Digital connection (remote control from
PC or ECG recorder), connection for cable
adapter (analog interface + remote start)
Note
• Connection Cables •
Use only connection cables approved by ergoline.
A special PC driver software, which can be obtained from
ergoline, is required for operation via the USB port.
4.7 Connecting ECG Leadwires
Plug the ECG leadwires (L, R, N) into the appropriate
sockets (1) in the control terminal.
N
L
R
1
Figure 4 – 10: ECG leadwire connections
ergoselect 4/5 13
4 Setup and Mains Connection
4.8 Connecting the Blood Pressure Cu
• Connect the microphone cable for blood pressure measurement to the intended port (2) so that it clicks into
place.
• Slip the cu tubing onto the tting (3) and engage.
To disconnect, push back the connector‘s knurled
sleeve.
Artifacts that may be caused by patient movements during
the exercise test, must be avoided, while the blood pressure
is being taken.
Therefore, do not forget to attach the cu tubing to the
handlebar with the supplied Velcro tape:
• Open the large Velcro tape and wrap it around the
handle bar.
• Secure the cu tubing with the small Velcro tape, but
do not exert pressure on the tubing.
4.9 Connecting the SpO2 Sensor
Connect the sensor cable to the corresponding socket (4)
on the underside of the control terminal.
2
3
Figure 4 – 11: Blood pressure cuff connections
2 Microphone connection
3 Fitting for connection of cuff tubing
Figure 4 – 12: Velcro tape to secure the cuff tubing
Insert the index nger into the SoftTip sensor. The cable
should run over the back of the hand.
Within a few seconds the current oxygen saturation in
percent (%) will be indicated on the display.
4
Figure 4 – 13: SpO2 sensor connection
4 SpO2 connection
Figure 4 – 14: SpO2 SoftTip
14ergoselect 4/5
5 Preparing the Patient for Blood Pressure Measurements
5 Preparing the Patient for Blood Pressure Measurements
5.1 Cu Size
Always choose the cu size suitable for the patient’s arm.
The maximum arm circumference is indicated on the cu.
Figure 5 – 1: Correct cuff size
When you close the Velcro strap, check that the metal clasp
(a) is inside the marked index range (b), and not outside.
5.2 Microphone Position
Before applying the cu, check the position of the microphone inside the red pocket (on the inside of the cu):
When the microphone is inside the pocket, its metal side must face the arm.
Figure 5 – 2: Wrong cuff size
Figure 5 – 3: Correct microphone position
ergoselect 4/5 15
5 Preparing the Patient for Blood Pressure Measurements
5.3 Applying the Cu
The center of the microphone must be located exactly
on the brachial artery. Locate the artery by palpation,
ifrequired.
The red tab identies the position of the microphone.
The accurate placement of the microphone is the primary
condition for reliable pressure measurement during exercise tests. The cu must be applied directly on the skin,
itmay not be applied on top of clothing, paper, etc.
Apply the cu approx. 2 cm above the bend of the elbow.
The cu should be tight, but it should not constrict blood
vessels. Make sure that the cu cannot shift when the
patient moves during the exercise test.
The cu tab must be located below the metal clasp (see
illustration at right).
Figure 5 – 4: Microphone placement on the artery
5.4 Checking the Cu Tubing
Check that the cu tubing does not knock against the
patient’s knee, when the patient is pedalling and the hand
is on the handlebar.
Secure the cu tubing with the Velcro tape attached to the
handlebar.
Instruct your patient to move as little as possible during
a blood pressure measurement and, in particular, to avoid
excessive contractions of the muscles in the upper arm.
Caution
• Patient Hazard •
Apply the cuff directly on the skin. Make sure that rolled up
sleeves do not impede blood circulation in the upper arm.
Loose cus will cause erroneous measurements; overtight cus
may constrict blood vessels or cause skin lesions and hematomas.
• Incorrect Measurements •
A loose cu would degrade the accuracy of the measurement.
Therefore, the computer aborts the measurement, if a minimum pressure is not attained within a few seconds.
Figure 5 – 5: Correct cuff position (tab)
Figure 5 – 6: Distance between knee and tubing
Warning
• Patient Hazard •
If, by accident, an excessive pressure builds up inside the cu,
either remove the cu immediately from the arm or disconnect
the cu tubing from the control terminal.
The same measures are recommended, if the cu does not
deate correctly.
16 ergoselect 4/5
6 Operation
The ergometer is available with dierent control terminals
which dier to some extent in terms of their functionality.
The following sections describe the control and conguration of the ergometer.
6.1 Speed Readout
A speed readout as well as three LEDs on the control terminal inform the patient of the speed: too slow, too fast
or correct (this readout is located at the top of the P and T
versions of the control terminal and on the side of control
terminal M).
The ranges for the respective speed ratings depend on the
selected load (see chapter11 “Technical Specications” on
page49).
6 Operation
1
2
3
Figure 6 – 1: Speed Readout
1 speed low (patient should pedal faster)
2 correct speed
3 speed high (= patient should pedal slower)
Note
If, during an exercise test, the speed drops below 30RPM,
theload readout starts blinking on the display and the load is
reduced to zero.
6.2 Handlebar Adjustment
Set the handlebar to a position where it is comfortable to
reach while sitting upright.
Fold up the clamping lever to unlock the handlebar.
When the handlebar is in the correct position, fold down
the clamping lever to lock the handlebar and check that the
lever is tightened.
With the ergometer standing rmly, check that the handlebar is tight. Adjust the clamping force of the clamping lever
if necessary.
Adjust the clamping force by opening the clamping lever
and turning the setting screw clockwise about a quarter
revolution with a at-blade screwdriver. Then check the
clamping force. Repeat these steps if necessary. When the
clamping force is appropriate, lock the clamping lever by
folding it down.
1
Figure 6 – 2: Handlebar adjustment
1 Clamping lever
Note
• Lock the clamping levers only as tight as necessary, NOT
with maximum force.
• Lubricate the thread of the saddle clamping lever periodically with a suitable lubricant (e.g., OKS470).
The handlebar is not designed to support the full body
weight.
ergoselect 4/5 17
6 Operation
Optionally, the height of the handlebar can be adjusted by
means of a motor, that is controlled via the toggle switch
on the handlebar.
6.3 Saddle Adjustment
Figure 6 – 3: Toggle switch for adjusting the height of the
handlebar
On the dierent ergometer models, the saddle height is
adjusted in dierent ways:
— with a clamping lever (mechanical)
— with a gas pressure spring
— with a motor.
When adjusting the height at the display, press the appropriate arrow key on the right to raise or lower the saddle.
The RPM display shows the actual saddle hight.
Note
Press the [Saddle] or the key to enable the adjustment
of the saddle: the adjustment keys will be displayed (or via the
and buttons on the control terminal).
12
Saddle
0
Figure 6 – 4: Saddle adjustment
at the control terminal
1 Saddle up
2 Saddle down
1
Figure 6 – 5: Saddle adjustment at the display
1 Saddle up
2
2 Saddle down
18 ergoselect 4/5
7 Control Terminal M
7.1 Turning the System On
You turn the ergometer on by pressing the power switch.
The ergometer runs a self-test. Subsequently, the start
screen displays.
7 Control Terminal M
Figure 7 – 1: Control Terminal Type „M“
Note
• Instruct the patient not to pedal while the ergometer is
being turned on and during the self-test.
• Apply the blood pressure cu to the patient AFTER the
ergometer has been turned on and the self-test completed.
Control terminals M are entirely operated by remote control
(e.g., from an ECG recorder or a PC).
With this key you initiate a blood pressure measurement. Pressing the key a second time during a measurement will stop the measurement.
Figure 7 – 2: Start screen
7.2 Operating Mode with Control Terminal M
Ergometers with the control terminal M support the following operating mode:
PC Mode
An external device (e.g., an ECG recorder, a PC-based ECG
system) controls the ergometer – no intervention at all is
required at the ergometer.
ergoselect 4/5 19
7 Control Terminal M
The display changes – the ergometer is waiting for commands from the external ECG unit.
As soon as the ergometer receives commands from the
controlling ECG unit or PC, the exercise test will start and
the corresponding values will be displayed.
The exercise test can only be terminated with the corresponding command from the controlling ECG unit.
Figure 7 – 3: Start screen
Note
• During an exercise test, the saddle height can be adjusted
with
and .
Figure 7 – 4: Display during exercise test 1
1 current load in watts
2 most recent BP value (systolic/diastolic values
or cuff pressure during inflation and bar graph
indicating microphone signal strength (see below)
3 duration of exercise test (min)
4 heart rate at the time of the BP measurement (BPM)
5 pedal speed (RPM)
Figure 7 – 5: Display during exercise test 2
20 ergoselect 4/5
8 Control Terminal P
8.1 Turning the System On
You turn on the ergometer by pressing the power switch
(toggle switch [I / 0] ).
The ergometer runs a self-test. Subsequently, the main
menu displays.
8 Control Terminal P
Note
• Instruct the patient not to pedal while the ergometer is
being turned on and during the self-test.
• Apply the blood pressure cu to the patient AFTER the
ergometer has been turned on and the self-test completed.
• The device can be configured to default to one of the
operating modes.
If this option is selected, the start screen of the selected
operating mode (e.g., Ergometry) will be displayed instead
of the main menu.
With the
The ergometer software is controlled with 5 keys:
With this key you display the main menu or return
to the previous menu level.
With this key you initiate a blood pressure measurement. A measurement in progress can be aborted with the
same key.
key, you can display the main menu.
Figure 8 – 1: Control Terminal P
Figure 8 – 2: Main Menu
The functions of these three softkeys chan ge
with the displayed menu – the key label describing the
function is shown on the display.
ergoselect 4/521
Figure 8 – 3: Keypad P
8 Control Terminal P
8.2 Operating Modes with Control Terminal P
Ergometers with control terminal P support the following
operating modes:
PC Mode
An external device (e.g., an ECG recorder, a PC-based ECG
system) controls the ergometer – no intervention at all is
required at the ergometer.
Ergometry
The ergometer runs an automatic exercise test – some of
the corresponding test protocols are user-congurable and
stored in the system.
(see chapter8.2.4 “Settings with Control Terminal P” on
page26)
Manual
The ergometer is controlled manually, i.e., the user performs all load changes via the keypad.
Settings
Used to congure the ergometer.
8. 2.1 PC Mode
Use the softkeys on the right and left (↑↓) to position the
bar cursor on PC Mode and conrm the selection with
Select.
The display changes – the ergometer is waiting for commands from the external ECG unit.
Figure 8 – 4: Main Menu
Figure 8 – 5: Start screen
22ergoselect 4/5
As soon as the ergometer receives commands from the
controlling ECG unit or PC, the exercise test will start and
the corresponding values will be displayed.
The exercise test can only be terminated with the corresponding command from the controlling ECG unit.
8 Control Terminal P
Figure 8 – 6: Display during exercise test 1
1 current load in watts
2 most recent BP value (systolic/diastolic values)
or cuff pressure during inflation and bar graph
indicating microphone signal strength (see below)
3 duration of exercise test (min)
4 heart rate at the time of the BP measurement (BPM)
5 pedal speed (RPM)
Note
• All functions are locked while the ergometer is operating in
PC mode, except for the saddle height adjustment and the
blood pressure key.
• To reactivate the saddle height adjustment function, press
and the arrow keys will again be displayed.
• Additional blood pressure measurements can be initiated
with
.
8.2.2 Ergometry
Use the softkeys on the right and on the left (↑↓) to
position the bar cursor on Ergometry and conrm the
selection with Select.
Figure 8 – 7: Display during exercise test 2
Figure 8 – 8: Main Menu
ergoselect 4/5 23
8 Control Terminal P
The stored test protocols available for selection will be
displayed. There are ve xed protocols (protocols 1 – 5,
see chapter11.3 “Exercise Test Protocols” on page52),
whereas protocols 6 – 15 are user-pro grammable.
The protocol menu provides an overview of the test phases.
Example: 50 W / 2 min / 25 W
means: initial (basic) load 50 watts
stage time 2 minutes
load increment 25 watts
Use the softkeys on the right and left (↑ ↓) to position the
bar cursor on one of the protocols and conrm the selection with Select.
The exercise test is started with the “Start” key, a blood
pressure measurement at rest may precede the test
(depending on selected ergometry protocol).
When the basic load appears on the display (after approx.
15 seconds or upon termination of the blood pressure
measurement) and the patient’s RPM indicator blinks, the
patient should start pedalling.
Figure 8 – 9: Selecting an exercise test protocol
The internal protocol will now control the entire exercise
test – the display always indicates the current values.
With the +5 W and –5 W keys, the current load can be
changed (in increments of +/–1 W up to +/–25 W, as congured).
Note
• The saddle height can be adjusted during an exercise test.
• To activate the saddle height adjustment, press
thearrow keys will be displayed then.
• Additional blood pressure measurements can be initiated
with
.
:
Figure 8 – 10: Initial exercise test screen
Figure 8 – 11: Screen display during the test
24 ergoselect 4/5
Terminating an Exercise Test
The exercise phase can be terminated manually at any time
with the Recovery key.
The load will immediately be reduced to 25 watts, but a
higher or lower value can be selected manually.
It is recommended that the patient continue to pedal in the
recovery phase.
The End key in the middle will terminate the test.Figure 8 – 12: Recovery phase
8.2.3 Manual
Use the softkeys on the right and left (↑↓) to position
the bar cursor on Manual and conrm the selection with
Select.
8 Control Terminal P
In this operating mode the user controls the entire exercise
test by selecting the loads, stage times and by initiating
blood pressure measurements.
The exercise test is started with the “Start” key, afterwards
the load can be set and changed with the +5 W and –5 W
keys (in increments of +/–1 W up to +/–25 W, as congured).
Blood pressure measurements can be initiated with .
Terminating an Exercise Test
The exercise test can be terminated manually at any time
with the End key located in the middle.
The load will immediately drop to 0 watt.
Figure 8 – 13: Main menu
Figure 8 – 14: Initial screen of a manual exercise test
There is no recovery phase in the manual mode.
Figure 8 – 15: Screen display during the test
ergoselect 4/5 25
8 Control Terminal P
8.2.4 Settings with Control Terminal P
Some of the device settings are congurable to meet specic requirements. The settings will be saved and remain
stored even when the ergometer is switched o.
Use the softkeys on the right and left (↑↓) to position the
bar cursor on Settings and confirm the selection with
Select.
The conguration menu displays.
When all changes have been made, you can exit the conguration menu with the key.
Use the softkeys on the right and left (↑↓) to position the
bar cursor on the parameter to change and conrm the
selection with Select.
Default Mode
In this menu you choose the default mode activated when
the ergometer is turned on. When rst turned on after
delivery, the ergometer will display this menu.
Use the softkeys on the right and left (↑↓) to position the
bar cursor on your preferred default mode and save the
selection with Select.
Figure 8 – 16: Main Menu
Figure 8 – 17: Settings Menu
Figure 8 – 18: Selecting the default mode
Protocols
Protocols 6 – 15 are user-programmable (protocols 1 – 5
are xed, see chapter11.3 “Exercise Test Protocols” on
page52 for protocol parameter details). Default values
can be entered for the following parameters:
— protocol type (Step/Ramp)
— basic load
— stage time
— load stage (load increase with each stage)
Figure 8 – 19: Selecting the exercise test protocol to edit
Use the softkeys on the right and left (↑↓) to position
the bar cursor on the protocol to change (No. 6 – 15) and
conrm the selection with Select.
26ergoselect 4/5
Use the right and left softkeys (↑↓) to select the parameter
to edit.
At Select, for example, you can choose the protocol type:
— Step (load increase in steps) or
— Ramp (continuous load increase).
Press Select to save the selected protocol type.
To cancel the selection, press the
All other parameters are edited in the same way.
Using the arrow keys (↑↓), highlight a parameter and
conrm the selection with Select: the corresponding value
appears in reverse video and can be changed with the
arrow keys ↑↓.
Pressing Select will save the new value.
You exit the conguration with .
key.
8 Control Terminal P
Figure 8 – 20: Selecting the parameter to edit
Figure 8 – 21: Editing the parameter value
Contrast
The display contrast is adjustable in the range from
0 to 100%.
Load Change
Here you determine the increments for each load change.
Depending on your choice, each key press will change the
load by +/– 1, 5, 10 und 25 Watts.
Figure 8 – 22: Adjusting the display contrast
Figure 8 – 23: Selecting the increment for manual load changes
ergoselect 4/5 27
8 Control Terminal P
Language
The texts can be displayed in dierent languages.
Beep
The audio signal emitted during blood pressure measurements can be turned on and o.
Figure 8 – 24: Language menu
Software Version
Select this option to view the installed software version.
Date / Time
To begin with, you select Date and conrm the selection.
Then the value displayed in reverse video can be edited
with the ↑ ↓ keys and saved with Select.
The time is adjusted in the same way.
You exit the conguration with .
Figure 8 – 25: Beep during BP measurements
Figure 8 – 26: Setting the date
Figure 8 – 27: Setting the day
28 ergoselect 4/5
EKG Type
The selected EKG Type determines the communication
method with the ECG recorder, PC-based ECG system, etc.
To prevent an accidental change of this setting, the menu is
protected with a password.
Using the arrow keys, enter 003 and conrm the entry with
Select.
All ergometers support the following communication
modes:
• Analog with pulse
Remote start mode; before advancing to the next load
level, the ergometer generates a control pulse and
sends the corresponding data via the interface.
8 Control Terminal P
Figure 8 – 28: Entering the EKG Type password
• Analo g / Digital
An analog voltage controls the load – blood pressure
measurements can be initiated with digital commands.
• Digital (default)
The communication with the ergo meter is entirely
controlled with digital commands.
• Analog IN-OUT
The entire communication (load control and BP measurements) is controlled with analog signals.
No digital data will be sent.
Select the communication mode and conrm with Select.
Note
• The EKG Type needs to be selected only when the ergometer
is connected to an ECG unit. The selection is part of the
installation procedure.
• The “Analog/Digital” and “Digital” communication is only
possible when PC Mode is selected from the main menu or
when this is the default mode.
Figure 8 – 29: Selecting the ergometer communication mode
ergoselect 4/5 29
8 Control Terminal P
RPM
Here you determine the RPM limits. When these limits
are exceeded, the LEDs for high or low speed (RPM) will
illuminate.
Select the value to change (Min. or Max.) and conrm with
Select.
Using the arrow keys, change the value and save the new
value with Select.
Note
The limits selected in this menu only apply to the load range
between 6 and 150 watts. At higher loads the RPM limits automatically adapt to the respective loads:
The pulse readout on the display can be turned o.
Figure 8 – 31: Setting the Pulse Display
30 ergoselect 4/5
9 Control Terminal T
9.1 Turning the System On
You turn on the ergometer by pressing the power switch
(toggle switch [I / 0] ).
Note
• Instruct the patient not to pedal while the ergometer is
being turned on and during the self-test.
• Apply the blood pressure cuff to the patient AFTER the
ergometer has been turned on and the self-test completed.
• The device can be congured to default to one of the operating modes.
If this option is selected, the start screen of the selected
operating mode (e.g., Ergometry) will be displayed instead
of the main menu.
9 Control Terminal T
The ergometer runs a self-test. Subsequently, the main
menu displays.
The ergometer software is controlled from the touch panel.
Figure 9 – 1: Control Terminal T
Figure 9 – 2: Self-test screen
Figure 9 – 3: Main Menu
ergoselect 4/5 31
9 Control Terminal T
9.2 Operating Modes with Control Terminal T
Ergometers with control terminal T support the following
operating modes:
PC Mode
An external device (e.g., an ECG recorder, a PC-based ECG
system) controls the ergometer – no intervention at all is
required at the ergometer.Figure 9 – 4: Operating mode „PC Mode“
Ergometry
The ergometer runs an automatic exercise test – the available
protocols (5 precongured, editable protocols and 5 usercongurable protocols) are saved in the ergometer (see
chapter “Settings” on page34).
Training / Test
Ten user-congurable training/test protocols are available
(see chapter9.2.3 “Training / Test” on page36).
A POLAR receiver is integrated in the ergometer and provides the relevant data for heart-rate controlled training
sessions. The test subject’s performance can be assessed
on the basis of these protocols.
Figure 9 – 5: Operating mode „Ergometry“
Figure 9 – 6: Operating mode „Training / Test“
Manual
The ergometer is controlled manually, i.e., the user performs all load changes via the touch panel.
Setup
Used to congure the ergometer.
Figure 9 – 7: Operating mode „Manual“
Figure 9 – 8: Operating mode „Setup“
32 ergoselect 4/5
9.2.1 PC Mode
When the [PC Mode] key has been pressed, the screen
appears as shown at right. The ergometer is waiting for
commands from the external ECG unit.
As soon as the ergometer receives commands from the
controlling ECG unit or PC, the exercise test will start and
the corresponding values will be displayed.
The display shows heart rate (1/min), blood pressure (mmHG),
oxygen saturation in percent (%), duration of the exercise
test (min:ss), pedal speed (1/min) and current load (W).
9 Control Terminal T
The exercise test can only be terminated with the corresponding command from the controlling ECG unit.
A blood pressure measurement can be initiated with the
[RR] key. Pressing the [RR] key a second time during a
measurement will stop the measurement.
When you press the [ECG] key, the display will change.
Theacquired, electrical signals will be displayed.
The amplitude (gain) can be adjusted with the arrow keys
[ ] and [ ].
The pump for the electrode suction system can be switched
on and o with the [I / 0] key.
You can change the vacuum intensity between low, medium and high by touching the [I] key.
To switch off the pump, press the [0] key for about
3seconds.
Conrm all inputs with the [ ] key.
Figure 9 – 9: PC mode screen
Figure 9 – 10: ECG screen
ergoselect 4/5 33
9 Control Terminal T
9.2.2 Ergometry
Pressing the [Ergometry] key in the main menu activates
the ergometry mode.
The dierent exercise test protocols will be displayed (5precongured, editable and 10user-congurable protocols).
All exercise test protocols (including the 5 precongured
protocols) are editable.
Figure 9 – 11: Ergometry Menu
Settings
When you touch a protocol, the dierent parameters will
be displayed.
With the [Edit] key, you can modify each protocol parameter. The new inputs overwrite the existing values.
All protocols can be edited during operation (except for the
PC mode).
User-programmed protocols must be saved with the
[Setup] key in the main menu.
In the conguration menu, the following parameters can
be edited:
• the basic load (from 6 to 100 W),
• the stage time (form 1 to 30 min),
• the stage rate (increment, from 1 to 400 W).
Figure 9 – 12: Exercise test protocol 1
Figure 9 – 13: Exercise test protocol 2
34ergoselect 4/5
With the [ 2.] key, you proceed to the next menu level
where you can edit these parameters:
• the recovery load (from 6 to 100 W) and
• the recovery time (from 1 to 30 min).
With the [ 1.] key, you return to the previous screen.
Touch the light gray eld, e.g., at Basic Load: a submenu
with input eld and numeric keypad opens. You can enter
numeric values directly via the keypad.
Conrm your inputs with the [ ] key. To cancel the input,
press the [ ] key.
9 Control Terminal T
Figure 9 – 14: Basic load of the exercise test 1
The other parameters can be edited in the same way, they
will overwrite the current values.
Figure 9 – 15: Basic load of the exercise test 2
When you touch the [ ] key again after conrming, the
display will change. Touching the [Start] key on the display
will initiate the training session. The session is entirely
controlled by the protocol. The display indicates the current
values.
When you press the [const.] key, the current load will be
maintained for the rest of the session.
To change the load, press the [+ 5 W] and [– 5 W] keys
(the increments can be congured from +/– 1 W up to
+/– 25 W as needed). See section9.2.5 “Setup with Control
Terminal”, section “Load Change” on page41.Figure 9 – 16: Starting an exercise test
ergoselect 4/5 35
9 Control Terminal T
Terminating the Program
Once the full protocol has been completed, it terminates
automatically.
The program can be terminated manually before the end
of the protocol with [Stop]. First, you enter the recovery
phase.
When you touch [Stop] again, the training will be terminated.
Figure 9 – 17: Terminating an exercise test 1
9.2.3 Train ing / Test
Ten dierent protocols are available in the Training/Test
menu.
To edit the protocol parameters, rst touch the Training/
Test protocol you want to edit.
Figure 9 – 18: Terminating an exercise test 2
Figure 9 – 19: Training/Test setup
36ergoselect 4/5
Then press the [Edit] key.
The individual parameters (light gray elds) can now be
edited by touching the display or by repeatedly tapping [ ].
If you need to input characters (numbers or letters), an
(alpha-) numeric keypad will be displayed.
The options for Type are [Pulse ], [Constant ], [Interval ],
[Ramp Test ], [PWC Test 1 ], [PWC Test 2 ], [PWC Test 3 ]
and [Inactive ]. You scroll through the Type options by
tapping the [ ] key. Congure the parameters as required
by the selected type.
9 Control Terminal T
Figure 9 – 20: Modify Training/Test
With the [ ] key (arrow down, [ 2.] or [ 3.]), you
access the next menu level where more parameters can be
congured. With the [ ] key (arrow up, [ 1.] or [ 2.],
you return to the previous screen.
Inputs are conrmed with the [ ] key.
Figure 9 – 21: Modify parameters 1
Figure 9 – 22: Modify parameters 2
ergoselect 4/5 37
9 Control Terminal T
9.2.4 Manual
In this operating mode the user has complete control over
the ergometer and initiates blood pressure measurements.
Pressing the [Start] key initiates the exercise test, the
[+ 5 W] and [– 5 W] keys are used to control the load.
Theincrements for the load change can be dened in the
setup menu.
A blood pressure measurement is initiated with the [RR] key.
Figure 9 – 23: Starting a manual test
Terminating an Exercise Test
The exercise test can be terminated manually at any time
with the [Stop] key.
The load will immediately drop to 0 watt.
There is no recovery phase in the manual mode.
9.2.5 Setup
The [Setup] key opens the conguration menu where
various program functions can be dened.
To access a menu, touch the corresponding key on the
display.
Figure 9 – 24: Terminating a manual test
Conrm inputs with the [ ] key and exit menus with the
[ ] key.
Figure 9 – 25: Setup menu
38 ergoselect 4/5
Default Mode
Select the operating mode to be activated when the
ergometer is turned on:
• PC Mode
• Ergometry
• Training/Test
• Manual
Conrm the selection with the [ ] key.
Protocols
The 5 ve exercise test protocols (WHO, BAL, Hollm, Std Fr
and Standard) are precongured, but all protocols in the
list are editable.
9 Control Terminal T
Figure 9 – 26: Default mode setup
To reach the level for editing of the individual protocol
parameters, rst touch the protocol that you want to modify
(e.g., [5. Program 6]), and then touch the [Edit] key.
To change the name of a protocol, touch the name and
enter the new name from the keypad. Conrm your inputs
with the [ ] key.
Figure 9 – 27: Protocol configuration 1
Figure 9 – 28: Protocol configuration 2
Figure 9 – 29: Protocol configuration 3
ergoselect 4/5 39
9 Control Terminal T
At Type, you can choose [Step ], [Ramp ] or [Inactive ].
You scroll through the Type options with the [ ] key.
When choosing the Step type (load increase in steps), dene
the basic load (from 6 to 100 W), the stage time (from 1 to
30min) and the stage rate (increment, from 1 to 400 W).
When choosing the Ramp type (continuous load increase),
define the basic load (from 6 to 100 W) and the load
increase (from 1 to 50 W).
To congure the protocol parameters (light gray elds),
touch one of the parameters.
Edit the parameter as appropriate and conrm the modication with the [ ] key.
Touch the Cancel key [ ] to conclude the parameter
change.
Figure 9 – 30: Protocol configuration 4
With the [ 2.] and [ 1.] keys, you toggle between the
dierent screens.
EKG Type
The selected EKG Type determines the communication
method with the ECG recorder, PC-based ECG system, etc.
To prevent an accidental change, this setting is protected
with a password.
A submenu opens when you touch EKG Type on the display.
Enter the code number “3” via the numeric keypad and
conrm with the [ ] key.
Figure 9 – 31: Protocol configuration 5
Figure 9 – 32: Setup menu
40ergoselect 4/5
The following communication modes are supported:
• Analog with pulse
Remote start mode; before advancing to the next load
level, the ergometer generates a control pulse and
sends the corresponding data via the interface.
• Analo g / Digital
An analog voltage controls the load – blood pressure
measurements can be initiated with digital commands.
• Digital (default)
The communication with the ergometer is entirely
controlled with digital commands.
• Analog IN-OUT
The entire communication (load control and BP measurements) is controlled with analog signals. No digital
data will be sent.
9 Control Terminal T
Figure 9 – 33: EKG type setup
Choose the appropriate communication mode and conrm
with the [ ] key.
Load Change
With this function, you select the increments for load
changes.
Figure 9 – 34: Load change setup
ergoselect 4/5 41
9 Control Terminal T
HR Belt Number
If, during the training, the test subject wears a chest belt
to measure the heart rate, the corresponding chest belt
number must be entered. You will nd the number on
the back of the housing. It is the unique identier for this
particular belt.
With the [
screen displays.
2.] and [ 1.] keys, you toggle between the
Date / Time
Touch the respective elds to adjust date and time.
Figure 9 – 35: HR Belt Number setup 1
Figure 9 – 36: HR Belt Number setup 2
Enter day, month, year as well as hours, minutes and
seconds via the numeric keypad.
Inputs are conrmed with the [
] key.
Figure 9 – 37: Date/Time setup 1
Figure 9 – 38: Date/Time setup 2
42 ergoselect 4/5
Regulation
At Regulation, you can specify the load details:
• load control (at, normal, steep)
• duration: load + (0 min … 15 min) and
• duration: load – (0 min … 15 min).
You scroll through the load control options (at, normal,
steep) by tapping the light gray text eld.
When you touch the light gray elds to the right of ‘Duration: load +’ or ‘Duration: load –’, the time can be entered
via the numeric keypad.
9 Control Terminal T
Figure 9 – 39: Control setup 1
Inputs are conrmed with the [ ] key.
RPM
In this menu, you determine the limits for the RPM indication.
The three LEDs on the control panel show the patient
whether the pedal speed is high, low or correct.
Figure 9 – 40: Control setup 2
Figure 9 – 41: RPM setup 1
ergoselect 4/5 43
9 Control Terminal T
Touch the light gray eld to the right of Min. or Max. and
enter the value via the numeric keypad.
Conrm the input with the [ ] key or cancel the input
with the [ ] key.
Beep
A short beep conrms each key press. The audible feedback
can be enabled and disabled.
Figure 9 – 42: RPM setup 2
Software Version
This menu shows the software version and the date of the
technical inspection of the measuring system (MTK).
Figure 9 – 43: Beep setup
Figure 9 – 44: Software Version setup
44 ergoselect 4/5
Language
Here you choose the language for the user interface.
Display
Press the [ 2.] key to display the next screen and touch
Display.
9 Control Terminal T
Figure 9 – 45: Language setup
Switch the pulse readout on or o.
Select the blood pressure unit:
mmHg (millimeter of mercury) or kPa (kilopascal).
Figure 9 – 46: Setup screen 1
Figure 9 – 47: Setup screen 2
ergoselect 4/5 45
10 Cleaning, Maintenance, Disposal
10 Cleaning, Maintenance, Disposal
10.1 General Cleaning
Wipe the device surface down with a cloth moistened with
soap water or a disinfectant.
The cloth should not be dripping wet; do not allow liquids
to enter the device.
10.1.1 Cleaning the Saddle
Clean the saddle with a soft and dry or moist cloth (Disinfectants used should not contain any alcohol).
10.1.2 Disinfection
Only the following disinfectants are approved for disinfection:
Schülke & Mayr GmbH:
• Antifect ® AF, FF, FD 10
• Terralin ® (0,5 %)
• Quartamon Med ®
B. Braun Melsungen AG:
• Hexaquart plus ® (0,5 % / 5,0 %)
• Hexaquart S ® (1,5 % / 5,0 %)
• Meliseptol ®
• Melsept SF ® (0,5 % / 5,0 %)
Warning
• Shock Hazard •
• Disconnect the device from the power line before cleaning.
• Equipment Damage •
• Do not allow liquids to enter the equipment.
Devices into which liquids have entered must be immediately cleaned and checked by a service technician, before
they can be reused.
• Do not use acids, alkaline solutions (household cleaners) or
caustic disinfectants.
Hint
The use of cleaning agents and disinfectants containing alcohol is not permitted!
Hint
Strictly observe the manufacturer’s instructions for use.
ECOLAB:
• Incidin Foam ®
46 ergoselect 4/5
10.2 Cleaning the Blood Pressure Cu
10. 2.1 Removing the Microphone
Pull the end of the cu through the metal clasp and fold
out the cu.
Pull on the short Velcro tab to open the microphone pocket
and carefully remove the microphone.
10.2.2 Cleaning
10 Cleaning, Maintenance, Disposal
Figure 10 – 1: Removing the Microphone
Clean the cu and tubing with a moist cloth.
You can use a dishwashing liquid or mild soap water
(no cleaning agents containing alcohol).
Clean the microphone with a cloth moistened with alcohol
or soap water.
Allow the microphone to dry before reinserting it in its
pocket.
10.2.3 Disinfection
For disinfection, spray a disinfectant sparingly on the cu,
the tubing and the microphone.
After the contact time indicated by the manufacturer, wipe
all components dry.
Only the following disinfectants are approved for disinfection:
Schülke & Mayr GmbH:
• Antifect ® AF, FF, FD 10
• Terralin ® (0,5 %)
• Quartamon Med ®
Warning
• Equipment Damage •
• Cu, microphone and tubing may not under any circumstances:
—be immersed in liquids
—be cleaned in a water bath or in running water.
Hint
Strictly observe the manufacturer’s instructions for use.
B. Braun Melsungen AG:
• Hexaquart plus ® (0,5 % / 5,0 %)
• Hexaquart S ® (1,5 % / 5,0 %)
• Meliseptol ®
• Melsept SF ® (0,5 % / 5,0 %)
ECOLAB:
• Incidin Foam ®
ergoselect 4/5 47
10 Cleaning, Maintenance, Disposal
10.2.4 Inserting the Microphone
Slip the microphone into the pocket, the metal side facing
the arm.
Guide the microphone cable out of the pocket and to the
right of the Velcro tab. Then close the tab.
Fold the end of the cu over and introduce it into the
metal clasp.
10.3 Maintenance
10.3.1 Checks before each use
Before each use, visually inspect the device for signs of
damage. If you detect damages or impaired functions
which may result in a hazard to the patient or the operator,
the device must be repaired before it can be used again.
Figure 10 – 2: Inserting the Microphone
10.3.2 Technical Safety Inspections,
Inspections of the Measuring
System
The technical safety inspections and the inspections of
the measuring system must be completed every two years
according to the rules of the art by a Service Engineer
authorized by ergoline.
Similarly, the automatic sphygmomanometer in the control
terminal must be checked and calibrated by an authorized
specialist every two years to fulll legal requirements.
The date of the next inspection is indicated on the
inspection sticker attached next to the type plate on the
ergometer.
10.4 Disposal
The product described in this operator manual must not
be disposed of as unsorted municipal waste and must be
collected separately.
Please contact your authorized manufacturer ergolineGmbH
for information concerning the disposal of your equipment.
There is no waste approval. Proper disposal is documented
by ergoline GmbH.
Consult Operator’s Manual!
48ergoselect 4/5
11 Technical Specications
11 Technical Specications
11.1 Ergometer
Modelmodular ergometer system ergoselect 4 / 5
models ergoselect 4, M / P / T, ergoselect 5, M / P / T
Operating modecontinuous operation
Power100 – 240 V / 50 – 60 Hz (max. 60 VA or 90 VA)
specications of the US power cord:
SPT 2x18AWG 125 V / 10 A
“hospital” or “hospital grade”
specications of the internal backup battery:
IEC: CR 2032 / 3 V 230 mAh
Braking principlecomputer-controlled eddy current brake
with torque measurement;
speed independent to DIN VDE 0750-0238
Load range6 … 999 Watt, speed (RPM)-independent
(see diagrams on page53)
Speed range30 … 130 n/min
Load accuracyto DIN VDE 0750-0238
Load incrementsuser programmable
Internal protocolscontrol terminal P:
• 5 xed and 10 user-programmable protocols
• manual load control
control terminal T:
• 10 protocols (5precongured, editable and 5 user-programmable protocols)
• 10 additional, user-programmable training/exercise test
protocols
• manual load control
• 3 precongured performance tests
Permitted patient weightversion 1: manual adjustment (standard), up to 160 kg
version 2: gas-spring assisted adjustment of the saddle
height (option), up to 200 kg
version 3: electrical adjustment of the saddle height with
digital indication of the current saddle height
(option), up to 200 kg
Handlebar adjustmentfor patient heights between 120 cm and 210 cm
(*) With control terminal P, the recovery load is xed at 25 W.
Stage Time
[min]
Load Stage
[W]
Recovery Load
[W]
Recovery Time
[min]
11.4 Test Protocols (control terminal T only)
ProtocolBasic Load
[W]
ramping protocol 012025 102599
PWC-130 (*)25 0251202599
PWC-150 (*)50 0251202599
PWC-170 (*)50 0501202599
(*) the protocol advances to the recovery phase as soon as the target heart rate (130/150/170) is reached
Duration
[sec]
Load Increment
[W]
Stage Time
[sec]
Recovery Load
[W]
Recovery Time
[min]
52ergoselect 4/5
11 Technical Specications
11.5 Family of characteristics of the braking torque control range
rpm
1
min
130130
120
110
100
90
80
70
60
50
40
30
20
10
0
25 1002003004005006007008009001000
Load
Figure 11 – 1: black: speed-independent range to DIN VDE 0750-0238
black + grey: speed-independent range of the ergoselect ergometer
Watt
11.6 Family of characteristics of the load periods according to IEC 60601‑1
Figure 11 – 2: Under permanent load, the load periods and pauses (white) shall be observed.
ergoselect 4/5 53
12 Electromagnetic Compatibility EN 60601-1-2
12 Electromagnetic Compatibility EN 60601‑1‑2
Changes or modications to this system not expressly
approved by ergoline may cause EMC issues with this or
other equipment.
Use of portable telephones or other radio frequency (RF) emitting
This system is designed to comply with applicable regulations regarding EMC.
Its compliance with these requirements has been veried.
Itneeds to be installed and put into service according to
the EMC information stated as follows.
Guidance and Manufacturer’s Declaration – Electromagnetic Emissions
The ergoselect ergometer is intended for use in the electromagnetic environment specied below. It is the responsibility
ofthe customer or user to ensure that the ergoselect ergometer is used in such an environment.
equipment near the system may cause unexpected or adverse
operation.
• Equipment Malfunction •
The equipment or system should not be used adjacent to, or
stacked with, other equipment. If adjacent or stacked use is
necessary, the equipment or system should be tested to verify
normal operation in the conguration in which it is being used.
RF emissions to EN 55011Group 1The ergoselect ergometer uses RF energy only for its
internal function. Therefore, its RF emissions are very
low and are not likely to cause any interference in
nearby electronic equipment.
RF emissions to EN 55011Class BThe ergoselect ergometer is suitable for use in all estab-
lishments, including domestic and those directly con-
Harmonic emissions to EN 61000-3-2Class A
Voltage uctuations/icker emissions
to EN 61000-3-3
Complies
nected to the public low-voltage power supply network
that supplies buildings used for domestic purposes.
54ergoselect 4/5
12 Electromagnetic Compatibility EN 60601-1-2
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The ergoselect ergometer is intended for use in the electromagnetic environment specied below. It is the
responsibility of the customer or user to ensure that the ergoselect ergometer is used in such an environment.
Immunity TestIEC 60601 Test LevelCompliance Level Electromagnetic Environment –
Guidance
Electrostatic discharge
(ESD) to EN 61000-4-2
Electrical fast transient/
burst to EN 61000-4-4
Surge to EN 61000-4-5± 1 kV dierential mode
Voltage dips, short interruptions and voltage
variations on power
supply input lines to
EN61000-4-11
± 6 kV contact
± 8 kV air
± 2 kV for power supply lines
± 1 kV for input and output lines
± 2 kV common mode
< 5 % UT
(> 95 % dip in UT) for 0.5 cycles
40 % UT
(60 % dip in UT) for 5 cycles
70 % UT
(30 % dip in UT) for 25 cycles
< 5 % UT
(> 95 % dip in UT) for 5 s
± 6 kV
± 8 kV
± 2 kV passedMains power should be that of
± 1 kV
N/A
< 5 % UTMains power should be that of a
40 % UT
70 % UT
< 5 % UT
Floors should be wood, concrete or
ceramic tile. If oors are covered
with synthetic material, the relative
humidity should be at least 30%.
a typical commercial or hospital
environment.
Mains power should be that of
a typical commercial or hospital
environment.
typical commercial or hospital environment. If the user of the ergoselect ergometer requires continued
operation during power mains
interruptions, it is recommended
that the ergoselect ergometer be
powered from an uninterruptible
power supply or a battery.
Power frequency
(50/60Hz) magnetic
eld to EN61000-4-8
NOTE: UT is the a.c. mains voltage prior to application of the test level.
3 A/mpassedPower frequency magnetic elds
should be at levels characteristic of
atypical location in a typical commercial or hospital environment.
The ergoselect ergometer has no
components susceptible to magnetic
elds.
ergoselect 4/5 55
12 Electromagnetic Compatibility EN 60601-1-2
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The ergoselect ergometer is intended for use in the electromagnetic environment specied below. It is the
responsibility of the customer or user to ensure that the ergoselect ergometer is used in such an environment.
Immunity TestIEC 60601 Test LevelCompliance Level Electromagnetic Environment –
Guidance
Portable and mobile RF communications equipment should be used no
closer to any part of the ergoselect
ergometer, including cables, than
the recommended separation distance calculated from the equation
applicable to the frequency of the
transmitter.
Recommended separation distance:
d = 1.2 √P
d = 1.2 √P for 80 MHz to 800 MHz
d = 2.3 √P for 800 MHz to 2.5 GHz
Conducted RF to
EN61000-4-6
Radiated RF to
EN61000-4-3
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
3 V
3 V/m
where P is the rated output power of
the transmitter in watts (W) according to the transmitter manufacturer
and d is the recommended separation distance in meters (m).
Field strengths from xed RF transmitters, as determined by an electromagnetic site survey (a), should
be less than the compliance level in
each frequency range (b).
Interference may occur in the vicinity
of equipment marked with the
following symbol:
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is aected by absorption and reection from
structures, objects, and people.
(a) Field strengths from xed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radio, AM and FM
radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to xed
RF transmitters, an electromagnetic site survey should be considered. If the measured eld strength in the location in which the ergoselect ergometer is used exceeds the applicable RF compliance level above, the ergoselect ergometer should be observed to verify normal
operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the ergoselect
ergometer.
(b) Over the frequency range from 150 kHz to 80 MHz, eld strengths should be less than 3 V/m.
56ergoselect 4/5
12 Electromagnetic Compatibility EN 60601-1-2
Recommended separation distances between portable and mobile RF communications equipment
and the ergoselect ergometer
The ergoselect ergometer is intended for use in an electromagnetic environment, as specied below, in which radiated RF
disturbances are controlled. The customer or the user of the ergoselect ergometer can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters)
and the ergoselect ergometer as recommended below, according to the maximum output power of the communications
equipment.
Rated Maximum Output
separation distance according to frequency of transmitter [m]
Power of Transmitter [W]
150 kHz to 80 MHz
d = 1.2 √P
80 MHz bis 800 MHz
d = 1,2 √P
800 MHz bis 2,5 GHz
d = 2,3 √P
0.010.120.120.23
0.10.370.370 .74
11.171.172.33
103.73.77.37
10011.711.723.3
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated
using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is aected by absorption and reection from
structures, objects, and people.
ergoselect 4/557
201000433000 • Version 2017-01-12 / Rev 01 • English