Technical
Information
TI213T/02/en
RTD Thermometer
omnigrad TST602
Surface thermometer suitable for application
measurement on flat surfaces or pipes
Connection cable, radius or flat mounting block
Description
TST602 RTD thermometer consists of a
radius or flat mounting block in
aluminium including a built-in RTD,
matched to a connection cable.
The sensing element includes a single
Pt100, in fiberglass insulated cable,
having accuracy class A or B, with 3 or
4 wires. The connection cable, with free
end wires and PVC or PTFE insulated,
can be selected from a wide choice of
standard lengths for any application up
to 70°C or 180°C. It is supplied with free
end wires.
Application
TST602 RTD thermometer is a general
purpose thermometer for surface
measurements. It is suitable for
application on flat surfaces or pipes and
is specially designed for climatisation
and general application in building
automation and heat exchangers.
Technical data
Probe
Sensing element: Platinum resistance, 1 x Pt100 Ω at 0°C, standard type
Tolerances: class A or B to IEC 751
Insulation: potting compound
Wiring: 3 or 4 wire connections
Insulation resistance: ≥ 100 MΩ, test voltage 250 V at ambient temperature
Mounting
Versions: radius block for pipe diameters up to 4"
flat block for pipes from 5" and flat surface
Standard material: Aluminium
Dimensions: 50 x 30 x 15 mm
Connection cable
Versions: PVC insulated / PVC sheathed, 4 wires
PTFE insulated / Silicone rubber sheathed, 4 wires
PTFE insulated / PTFE sheathed, 4 wires
Standard lengths: 1000, 2000, 2500, 3500 mm; others on request
Operating temperature: PVC / PVC : -20°C to +70°C
PTFE / Silicone : -20°C to +180°C
PTFE / PTFE : -20°C to +180°C
Electrical connection: free end wires
Endress+Hauser
The Power of Know How
Product structure
TI213T/02/en/12.99
EHT/CV5
71104991
TST602- RTD Thermometer
Surface thermometer suitable for application measurement on flat
surfaces or pipes, with connection cable and radius or flat mounting
block in aluminium
Mounting version, material
A - 50 x 30 x 15 radius block mounting
B - 50 x 30 x 15 flat block mounting
RTD class and type of construction
B - 1 Pt100, class B, 3 wires
H - 1 Pt100, class A, 3 wires
M - 1 Pt100, class A, 4 wires
Y - RTD class and type to specification
Connection cable length L (500-30000mm)
1-1000mm
2-2000mm
3-2500mm
4-3500mm
8 - .... mm length to specification
9 - .... mm special length
Connection cable type
A - PVC insulated - PVC sheathed - 4 wires
B - PTFE insulated - Silicone sheathed - 4 wires
C - PTFE insulated - PTFE sheathed - 4 wires
Y - Connection cable type to specification
Connection cable, terminal type
11 - Free end wires
99 - Connection to specification
Built-in transmitter
0 - Without built-in transmitter
C99MT021.CDR
TST602
(all dimensionsin millimeters)
Accessories
TAF10
United Kingdom
UK Head Office and
Production Centre:
Floats Road
Manchester
M23 9NF
Tel. no. (0161) 2 86 50 00
Fax no. (0161) 9 98 18 41
TST602- Complete Order Code
Field housing TAF10
For field mounting a TAF10 housing for
transmitter installation is available.
For further information please refer to:
Field housing TAF10
Operating Instructions KA 093R/09/a2
Regional Offices:
London
Tel. no. (0181) 9 03 95 69
Fax no. (0181) 9 03 93 12
Glasgow
Tel. no. (0141) 2 48 62 34
Installation
Pipe mounting example using a hose
clamp (not included):
Hose clamp
TST602
mounting block
Pipe section
Export Division
+
Endress
GmbH+Co.
Instruments International
P.O. Box 2222
D-79574 Weil am Rhein
Germany
Tel. (07621) 975-02
Fax (07621) 975-345
http://www.endress.com
info@ii.endress.com
Hauser
12.99/MTM
TI213T/02/en/12.99
EHT/CV5
Endress+Hauser
The Power of Know How