Operating Instructions
Liquistation CSF33
Automatic sampler for liquid media
Calibration

About this manual
This manual describes how to calibrate the distribution arm and the sample volume.
This manual does not include the following:
• Setup/General settings
--> Operating Instructions BA00479C "Commissioning"
• Display/Operation
--> Operating Instructions BA00479C "Commissioning"
•Inputs
--> Operating Instructions BA00487C "Operation & settings"
•Outputs
--> Operating Instructions BA00487C "Operation & settings"
• Sampling programs
--> Operating Instructions BA00487C "Operation & settings"
• Additional functions
--> Operating Instructions BA00487C "Operation & settings"
• Data management
--> Operating Instructions BA00487C "Operation & settings"
•Expert
--> Internal Service Manual
•Diagnostics
--> Operating Instructions BA00488C "Maintenance & diagnostics"

Endress+Hauser
Liquistation CSF33
Table of contents
1 Calibrating the distribution
arm . . . . . . . . . . . . . . . . . . . . . . . . 4
2 Calibrating the sample volume 5
2.1 Vacuum pump . . . . . . . . . . . . . . . . . . . . . . 5
2.2 Peristaltic pump . . . . . . . . . . . . . . . . . . . . . 7
Index. . . . . . . . . . . . . . . . . . . . . . . 8

Calibrating the distribution arm Liquistation CSF33
4 Endress+Hauser
1 Calibrating the distribution arm
It is only possible to calibrate the distribution arm in the version with multiple bottles.
The distribution arm must be calibrated if:
• The distribution arm motor has been replaced
• Error message "F236 Distribution arm" appears on the display
1. Select the number of bottles in the "Setup/Basic setup" menu.
2. Proceed as follows to calibrate the distribution arm:
3. Afterwards perform a distribution arm test in the "Diagnostics/Systemtest/Reset/Distribution
arm" menu.
Path: Menu/Calibration active
Function Options Info
Distribution arm
Go to ref. point Action The reference run is started. The reference point is
in the middle at the front.
For versions with distributor plate, the reference
point is at the arrow in the middle of the plate. For
versions with distribution assembly the reference
point is between bottle number 1 and the last one.
With Adjust you can correct the distribution arm if the unit does not move to the reference point correctly. Use
the two arrow keys to correct the position.