Endress+Hauser Cerabar M PMC51? Cerabar M PMP51, Cerabar M PMP55 Operating Instructions Manual

Page 1
BA00385P/00/EN/19.14 71269382
Valid from software version:
01.00.zz
Operating Instructions
Cerabar M PMC51, PMP51, PMP55
Process pressure measurement
Page 2
Cerabar M Analog
2 Endress+Hauser
A0023555
TAG No.: XXX000
Ser. No.: X000X000000
Order code 00X00-XXXX0XX0XXX
www.endress.com/deviceviewer Endress+Hauser Operations App
Serial number
Page 3
Cerabar M Analog Table of contents
Endress+Hauser 3
Table of contents
1 Safety instructions . . . . . . . . . . . . . . . . 4
1.1 Designated use . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.2 Installation, commissioning and operation . . . . . . . . 4
1.3 Operational and process safety . . . . . . . . . . . . . . . . . 4
1.4 Notes on safety conventions and icons . . . . . . . . . . . 5
2 Identification . . . . . . . . . . . . . . . . . . . . 6
2.1 Product identification . . . . . . . . . . . . . . . . . . . . . . . 6
2.2 Device designation . . . . . . . . . . . . . . . . . . . . . . . . . 6
2.3 Scope of delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2.4 CE mark, Declaration of Conformity . . . . . . . . . . . . 9
2.5 Registered trademarks . . . . . . . . . . . . . . . . . . . . . . . 9
3 Installation . . . . . . . . . . . . . . . . . . . . . 10
3.1 Incoming acceptance, transport, storage . . . . . . . . . 10
3.2 Installation conditions . . . . . . . . . . . . . . . . . . . . . . 10
3.3 Installation instructions . . . . . . . . . . . . . . . . . . . . . 11
3.4 Closing the housing cover . . . . . . . . . . . . . . . . . . . 20
3.5 Mounting of the profile seal for universal process
mounting adapter . . . . . . . . . . . . . . . . . . . . . . . . . 20
3.6 Post-installation check . . . . . . . . . . . . . . . . . . . . . . 20
4 Wiring . . . . . . . . . . . . . . . . . . . . . . . . 21
4.1 Connecting the device . . . . . . . . . . . . . . . . . . . . . . 21
4.2 Connecting the measuring unit . . . . . . . . . . . . . . . 23
4.3 Potential equalization . . . . . . . . . . . . . . . . . . . . . . 23
4.4 Overvoltage protection (optional) . . . . . . . . . . . . . . 24
4.5 Post-connection check . . . . . . . . . . . . . . . . . . . . . . 25
5 Operation . . . . . . . . . . . . . . . . . . . . . . 26
5.1 Position of operating elements . . . . . . . . . . . . . . . . 26
5.2 Using the device display (optional) . . . . . . . . . . . . . 27
6 Commissioning. . . . . . . . . . . . . . . . . . 29
6.1 Function check . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
6.2 Commissioning . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
7 Maintenance. . . . . . . . . . . . . . . . . . . . 30
7.1 Cleaning instructions . . . . . . . . . . . . . . . . . . . . . . . 30
7.2 Exterior cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . 30
8 Troubleshooting . . . . . . . . . . . . . . . . . 31
8.1 Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
8.2 Measures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
8.3 Response of output to errors . . . . . . . . . . . . . . . . . 31
8.4 Repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
8.5 Spare Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
8.6 Return . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
8.7 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
8.8 Software history . . . . . . . . . . . . . . . . . . . . . . . . . . 32
9 Technical data . . . . . . . . . . . . . . . . . . . 33
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Page 4
Safety instructions Cerabar M Analog
4 Endress+Hauser
1 Safety instructions
1.1 Designated use
The Cerabar M is a pressure transmitter for measuring pressure and level.
The manufacturer accepts no liability for damages resulting from incorrect use or use other than that designated.
1.2 Installation, commissioning and operation
The device is designed to meet state-of-the-art safety requirements and complies with applicable standards and EU regulations. If used incorrectly or for applications for which it is not intended, however, it can be a source of application-related danger, e.g. product overflow due to incorrect installation or configuration. For this reason, installation, connection to the electricity supply, commissioning, operation and maintenance of the measuring system must only be carried out by trained, qualified specialists authorized to perform such work by the facility's owner-operator. The specialist staff must have read and understood these Operating Instructions and must follow the instructions they contain. Modifications and repairs to the devices are permissible only if they are expressly approved in the Operating Instructions. Pay particular attention to the technical data and information on the nameplate.
1.3 Operational and process safety
Alternative monitoring measures have to be taken while configuring, testing or servicing the device to ensure the operational and process safety.
#
Warning! Dismantle device only when depressurized!
Page 5
Cerabar M Analog Safety instructions
Endress+Hauser 5
1.4 Notes on safety conventions and icons
In order to highlight safety-relevant or alternative operating procedures in the manual, the following conventions have been used, each indicated by a corresponding icon in the margin.
Symbol Meaning
#
Warning!
A warning highlights actions or procedures which, if not performed correctly, will lead to serious personal injury, a safety hazard or the destruction of the device.
"
Caution!
Caution highlights actions or procedures which, if not performed correctly, can lead to personal injury or the incorrect operation of the device.
!
Note!
A note highlights actions or procedures which, if not performed correctly, can have an indirect effect on operation or trigger an unexpected response on the part of the device.
%
Direct current
A terminal to which DC voltage is applied or through which direct current flows.
&
Alternating current
A terminal to which alternating voltage (sine-wave) is applied or through which alternating current flows.
)
Ground connection
A grounded terminal, which as far as the operator is concerned, is already grounded by means of a grounding system.
*
Protective ground connection
A terminal which must be connected to ground prior to establishing any other connections.
+
Equipotential connection
A connection that has to be connected to the plant grounding system: This may be a potential equalization line or a star grounding system depending on national or company codes of practice.
Connecting cable immunity to temperature change
Indicates that the connecting cables have to withstand a temperature of 85°C at least.
Safety instructions
Observe the safety instructions in the associated Operating Instructions.
>
=
t 85°C
Page 6
Identification Cerabar M Analog
6 Endress+Hauser
2 Identification
2.1 Product identification
The following options are available for identification of the measuring device:
• Nameplate specifications
• Order code with breakdown of the device features on the delivery note
• Enter serial numbers from nameplates in W@M Device Viewer (www.endress.com/deviceviewer): All information about the measuring device is displayed.
For an overview of the technical documentation provided, enter the serial number from the nameplates in the W@M Device Viewer (www.endress.com/deviceviewer).
2.2 Device designation
2.2.1 Nameplate
!
Note!
• The MWP (maximum working pressure) is specified on the nameplate. This value refers to a reference temperature of 20°C (68°F) or 100°F (38 °C) for ANSI flanges.
• The pressure values permitted at higher temperatures can be found in the following standards: – EN 1092-1: 2001 Tab. 18
1)
– ASME B 16.5a – 1998 Tab. 2-2.2 F316 – ASME B 16.5a – 1998 Tab. 2.3.8 N10276 – JIS B 2220
• The test pressure corresponds to the over pressure limit (OPL) of the device = MWP x 1.5
2)
.
• The Pressure Equipment Directive (EC Directive 97/23/EC) uses the abbreviation "PS". The abbreviation "PS" corresponds to the MWP (maximum working pressure) of the measuring device.
1) With regard to their stability-temperature property, the materials 1.4435 and 1.4404 are grouped together under 13EO
in EN 1092-1 Tab. 18. The chemical composition of the two materials can be identical.
2) The equation does not apply for PMP51 and PMP55 with a 40 bar (600 psi) or a 100 bar (1500 psi) measuring cell.
Page 7
Cerabar M Analog Identification
Endress+Hauser 7
Aluminum housing
P01-xMx5xxxx-18-xx-xx-xx-000
Fig. 1: Nameplate
1 Device name 2 Order code (for re-orders) 3 Serial number (for identification) 4 Extended order code (complete) 5 MWP (maximum working pressure) 6 Electronic version (output signal) 7 Min./max. span 8 Nominal measuring range 9 Supply voltage 10 Unit of length 11 No entry 12 ID number of notified body with regard to Pressure Equipment Directive (optional) 13 Approvals 14 Device version 15 Software version 16 Degree of protection 17 Wetted materials 18 Approval-specific information
Devices suitable for oxygen applications are fitted with an additional nameplate.
P01-xxxxxxxx-18-xx-xx-xx-000
Fig. 2: Additional nameplate for devices suitable for oxygen applications
1 Maximum pressure for oxygen applications 2 Maximum temperature for oxygen applications 3 Layout identification of the nameplate
ex works FW:
Dev.Rev. :
250002755-C
Ext. order code:
Ser. no.:
U=
Span
P
MWP
Order code:
Made in Germany, D-79689 Maulburg
Mat.:
250002756-E
L=
15
1
2
3
4
6
7
8
9
10
11
12
13
14
5
16
17
18
18
Pmax Tmax
for oxygen service
Bei Sauerstoffeinsatz/
3
1
2
Page 8
Identification Cerabar M Analog
8 Endress+Hauser
Stainless steel housing, hygienic
P01-xMx5xxxx-18-xx-xx-xx-001
Fig. 3: Nameplate
1 Device name 2 Order code (for re-orders) 3 Serial number (for identification) 4 Extended order code (complete) 5 Nominal measuring range 6 MWP (maximum working pressure) 7 Length data 8 Electronic version (output signal) 9 Supply voltage 10 Min./max. span 11 Wetted materials 12 Approval-specific information 13 No entry 14 ID number of notified body with regard to Pressure Equipment Directive (optional) 15 No entry 16 Software version 17 Device version 18 Degree of protection
Devices with certificates are fitted with an additional plate.
P01-xMx5xxxx-18-xx-xx-xx-002
Fig. 4: Additional nameplate for devices with certificates
1 Approval-specific information
2.3 Scope of delivery
The scope of delivery comprises:
• Device
• Optional accessories
Documentation supplied:
• The Operating Instructions BA00385P is available on the Internet. See: www.endress.com Download
• Brief Operating Instructions: KA01036P
• Final inspection report
• Optional: factory calibration form, test certificates
002757-A
p
Order code:
Ext. order code:
Ser. no.:
MWP
D-79689 Maulburg
Made in Germany,
ex works FW
U=
Dev.Rev.:
L=
Span Mat.:
002758-C
16
17
15
1
2
3
4
4
4
6
7
8
9
10
11
12
12
13
14
5
18
002759-B
1
1
Page 9
Cerabar M Analog Identification
Endress+Hauser 9
2.4 CE mark, Declaration of Conformity
The devices are designed to meet state-of-the-art safety requirements, have been tested and left the factory in a condition in which they are safe to operate. The devices comply with the applicable standards and regulations as listed in the EC Declaration of Conformity and thus comply with the statutory requirements of the EC Directives. Endress+Hauser confirms the conformity of the device by affixing to it the CE mark.
2.5 Registered trademarks
KALREZ, VITON, TEFLON Registered trademark of E.I. Du Pont de Nemours & Co., Wilmington, USA
TRI-CLAMP Registered trademark of Ladish & Co., Inc., Kenosha, USA
GORE-TEX
®
Registered trademark of W.L. Gore & Associates, Inc., USA
Page 10
Installation Cerabar M Analog
10 Endress+Hauser
3 Installation
3.1 Incoming acceptance, transport, storage
3.1.1 Incoming acceptance
• Check the packaging and the contents for damage.
• Check the shipment, make sure nothing is missing and that the scope of supply matches your order.
3.1.2 Transport
"
Caution! Follow the safety instructions and transport conditions for devices of more than 18 kg (39.69 lbs). Transport the measuring device to the measuring point in its original packaging or at the process connection.
3.1.3 Storage
The device must be stored in a dry, clean area and protected against damage from impact (EN 837-2). Storage temperature range: See Technical Information for Cerabar M TI00436P.
3.2 Installation conditions
3.2.1 Dimensions
For dimensions, please refer to the Technical Information for Cerabar M TI00436P, "Mechanical construction" section.
Page 11
Cerabar M Analog Installation
Endress+Hauser 11
3.3 Installation instructions
!
Note!
• Due to the orientation of the Cerabar M, there may be a shift in the zero point, i.e. when the container is empty or partially full, the measured value does not display zero. You can correct this zero point shift ä 27, Section 5.1.2 "Function of the operating elements".
• For PMP55, please refer to Section 3.3.2 "Installation instructions for devices with diaphragm seals – PMP55", ä 14.
• Endress+Hauser offers a mounting bracket for installing on pipes or walls. ä 17, Section 3.3.5 "Wall and pipe mounting (optional)".
3.3.1 Installation instructions for devices without diaphragm seals –
PMP51, PMC51
!
Note!
• If a heated Cerabar M is cooled during the cleaning process (e.g. by cold water), a vacuum develops for a short time, whereby moisture can penetrate the sensor through the pressure compensation (1). If this is the case, mount the Cerabar M with the pressure compensation (1) pointing downwards.
• Keep the pressure compensation and GORE-TEX
®
filter (1) free from contamination.
• Cerabar M transmitters without diaphragm seals are mounted as per the norms for a manometer (DIN EN 837-2). We recommend the use of shutoff devices and siphons. The orientation depends on the measuring application.
• Do not clean or touch process isolating diaphragms with hard or pointed objects.
• The device must be installed as follows in order to comply with the cleanability requirements of the ASME-BPE (Part SD Cleanibility).:
1
1
1
Page 12
Installation Cerabar M Analog
12 Endress+Hauser
Pressure measurement in gases
P01-PMx5xxxx-11-xx-xx-xx-003
Fig. 5: Measuring arrangement for pressure measurement in gases
1 Cerabar M 2 Shutoff device
• Mount the Cerabar M with the shutoff device above the tapping point so that any condensate can flow into the process.
Pressure measurement in steams
P01-PMx5xxxx-11-xx-xx-xx-004
Fig. 6: Measuring arrangement for pressure measurement in steams
1 Cerabar M 2 Shutoff device 3 U-shaped siphon 4 Circular siphon
• Mount Cerabar M with siphon above the tapping point.
• Fill the siphon with liquid before commissioning. The siphon reduces the temperature to almost the ambient temperature.
Page 13
Cerabar M Analog Installation
Endress+Hauser 13
Pressure measurement in liquids
P01-PMx5xxxx-11-xx-xx-xx-005
Fig. 7: Measuring arrangement for pressure measurement in liquids
1 Cerabar M 2 Shutoff device
• Mount Cerabar M with shutoff device below or at the same level as the tapping point.
Level measurement
P01-xMx5xxxx-11-xx-xx-xx-000
Fig. 8: Measuring arrangement for level
• Always install the Cerabar M below the lowest measuring point.
• Do not mount the device in the filling curtain or at a point in the tank which could be affected by pressure pulses from an agitator.
• Do not mount the device in the suction area of a pump.
• The calibration and functional test can be carried out more easily if you mount the device downstream of a shutoff device.
PVDF interchangeable threaded boss
!
Note! A maximum torque of 7 Nm (5.16 lbs ft) is permitted for devices with a PVDF interchangeable threaded boss. The thread connection may become loose at high temperatures and pressures. This means that the integrity of the thread must be checked regularly and may need to be tightened using the torque given above. Teflon tape is recommended for sealing the 1/2 NPT thread.
Page 14
Installation Cerabar M Analog
14 Endress+Hauser
3.3.2 Installation instructions for devices with diaphragm seals – PMP55
!
Note!
• Cerabar M devices with diaphragm seals are screwed in, flanged or clamped, depending on the
type of diaphragm seal.
• A diaphragm seal and the pressure transmitter together form a closed, oil-filled calibrated system.
The fill fluid hole is sealed and may not be opened.
• Do not clean or touch the process isolating diaphragm of the diaphragm seal with hard or pointed
objects.
• Do not remove process isolating diaphragm protection until shortly before installation.
• When using a mounting bracket, sufficient strain relief must be ensured for the capillaries in order
to prevent the capillary bending down (bending radius 100 mm (3.94 in)).
• Please note that the hydrostatic pressure of the liquid columns in the capillaries can cause zero
point shift. The zero point shift can be corrected.
• Please observe the application limits of the diaphragm seal filling oil as detailed in the Technical
Information for Cerabar M TI00436P, "Planning instructions for diaphragm seal systems" section.
In order to obtain more precise measurement results and to avoid a defect in the device, mount the capillaries as follows:
• Vibration-free (in order to avoid additional pressure fluctuations)
• Not in the vicinity of heating or cooling lines
• Insulate if the ambient temperature is below or above the reference temperature
• With a bending radius of 100 mm (3.94 in).
Page 15
Cerabar M Analog Installation
Endress+Hauser 15
Vacuum application
For applications under vacuum, Endress+Hauser recommends mounting the pressure transmitter below the diaphragm seal. This prevents vacuum loading of the diaphragm seal caused by the presence of filling oil in the capillaries.
When the pressure transmitter is mounted above the diaphragm seal, the maximum height difference H1 in accordance with the illustration below left must not be exceeded. The maximum height difference depends on the density of the filling oil and the smallest ever pressure that is permitted to occur at the diaphragm seal (empty container), see illustration below right.
Mounting with temperature isolator
P01-PMx5xxxx-11-xx-xx-xx-008
Endress+Hauser recommends the use of temperature isolators in the event of constant extreme medium temperatures which lead to the maximum permissible electronics temperature of +85°C (+185°F) being exceeded. To minimize the influence of rising heat, Endress+Hauser recommends the device be mounted horizontally or with the housing pointing downwards. The additional installation height also brings about a zero point shift of approx. 21 mbar (0.315 psi) due to the hydrostatic column in the temperature isolator. You can correct this zero point shift. ä 27 "Function of the operating elements".
P01-PMx5xxxx-11-xx-xx-xx-006
Fig. 9: Installation above the diaphragm
seal
P01-PMP75xxx-05-xx-xx-en-011
Fig. 10: Diagram of maximum installation height above the
diaphragm seal for vacuum applications, depending on the pressure at the diaphragm seal
H1
0.0
2.0
4.0
6.0
8.0
10.0
12.0
50 100 300 400 500 600 700 800 900 1000
200
Inert oil
High temperature oil
Vegetableoil
Silicone oil
Pressure at diaphragm seal[mbar
abs
]
Heightdifference H1 [m]
Low temperature oil
max. 115
Page 16
Installation Cerabar M Analog
16 Endress+Hauser
3.3.3 Seal for flange mounting
P01-FMD7xxxx-11-xx-xx-xx-002
Fig. 11: Mounting the versions with a flange
1 Process isolating diaphragm 2Seal
#
Warning! The seal is not allowed to press against the process isolating diaphragm as this could affect the measurement result.
3.3.4 Thermal insulation – PMP55
The PMP55 may only be insulated up to a certain height. The maximum permitted insulation height is indicated on the devices and applies to an insulation material with a heat conductivity 0.04 W/(m x K) and to the maximum permitted ambient and process temperature ( see table below). The data were determined under the most critical application "quiescent air".
P01-PMx5xxxx-11-xx-xx-xx-010
Fig. 12: Maximum permitted insulation height, here indicated on a PMP55 with a flange
l 0.04£
W
mK
T
P
T
A
PMP55
Ambient temperature (T
A
) 70°C (158°F)
Process temperature (T
P
) Max. 400°C (752°F), depending on the diaphragm seal filling oil used (see
TI00436PEN)
Page 17
Cerabar M Analog Installation
Endress+Hauser 17
3.3.5 Wall and pipe mounting (optional)
Endress+Hauser offers a mounting bracket for installing on pipes or walls (for pipes from 1
1/4
" up
to 2" diameter).
P01-xMx5xxxx-06-xx-xx-xx-001
Please note the following when mounting:
• Devices with capillary tubes: mount capillaries with a bending radius 100 mm (3.94 in).
• When mounting on a pipe, tighten the nuts on the bracket uniformly with a torque of at least 5 Nm (3.69 lbs ft).
52 (2.05)
140 (5.51)
122 (4.8)
70 (2.76)
6 (0.24)
mm (in)
Page 18
Installation Cerabar M Analog
18 Endress+Hauser
3.3.6 Assembling and mounting the "separate housing" version
P01-XMx5xxxx-11-xx-xx-xx-009
Fig. 13: "Separate housing" version
1 In the case of the "separate housing" version, the sensor is delivered with the process connection and cable ready
mounted. 2 Cable with connection jack 3 Pressure compensation 4 Connector 5Locking screw 6 Housing mounted with housing adapter, included 7 Mounting bracket provided, suitable for pipe and wall mounting (for pipes from 1
1/4
" up to 2" diameter)
Assembly and mounting
1. Insert the connector (item 4) into the corresponding connection jack of the cable (item 2).
2. Plug the cable into the housing adapter (item 6).
3. Tighten the locking screw (item 5).
4. Mount the housing on a wall or pipe using the mounting bracket (item 7).
When mounting on a pipe, tighten the nuts on the bracket uniformly with a torque of at least 5 Nm (3.69 lbs ft). Mount the cable with a bending radius (r) 120 mm (4.72 in).
r ³ 120 mm
1
2
3
4
5
6
7
7
Page 19
Cerabar M Analog Installation
Endress+Hauser 19
3.3.7 PMP51, version prepared for diaphragm seal mount – welding recommendation
P01-PMP71xxx-11-xx-xx-xx-000
Fig. 14: Version U1: prepared for diaphragm seal mount
1 Hole for fill fluid 2Bearing 3Setscrew A1 See the "Welding recommendation" table below
Endress+Hauser recommends welding on the diaphragm seal as follows for the "XSJ - Prepared for diaphragm seal mount" version in feature 110 "Process connections" in the order code up to, and including, 40 bar (600 psi) sensors: the total welding depth of the fillet weld is 1 mm (0.04 in) with an outer diameter of 16 mm (0.63 in). Welding is performed according to the WIG method.
Information on filling
The diaphragm seal must be filled as soon as it has been welded on.
• After welding into the process connection, the sensor assembly must be properly filled with a
filling oil and sealed gas-tight with a sealing ball and lock screw. Once the diaphragm seal has been filled, the device display should not exceed 10% of the full scale value of the cell measuring range at the zero point. The internal pressure of the diaphragm seal must be corrected accordingly.
• Adjustment / calibration:
– The device is operational once it has been fully assembled. – Perform a reset. The device then has to be calibrated to the process measuring range as
explained in the Operating Instructions.
ø2.5 (0.1)
5 (0.2)
➀➂
ø7.95 (0.31)
A1
mm (in)
Consecutive seam no.
Sketch/welding groove shape, dimension as per DIN 8551
Base material matching
Welding process DIN EN ISO 24063
Welding position
Inert gas, additives
A1 for sensors 40 bar (600 psi)
P01-PMP71xxx-11-xx-xx-xx-001
Adapter made of AISI 316L (1.4435) to be welded to diaphragm seal made of AISI 316L (1.4435 or 1.4404)
141 PB Inert gas
Ar/H 95/5
Additive: ER 316L Si (1.4430)
s1 a0.8
Page 20
Installation Cerabar M Analog
20 Endress+Hauser
3.4 Closing the housing cover
!
Note! When closing the housing cover, please ensure that the thread of the cover and housing are free from dirt, e.g. sand.If you feel any resistance when closing the cover, check the thread on both again to ensure that they are free from dirt.
3.4.1 Closing the cover on the stainless steel housing
P01-XMx5xxxx-17-xx-xx-xx-001
Fig. 15: Closing the cover
The cover for the electronics compartment is tightened by hand at the housing until the stop. The screw serves as DustEx protection (only available for devices with DustEx approval).
3.5 Mounting of the profile seal for universal process mounting adapter
For details on mounting, see KA00096F/00/A3.
3.6 Post-installation check
After installing the device, carry out the following checks:
• Are all screws firmly tightened?
• Is the housing cover screwed down tight?
Page 21
Cerabar M Analog Wiring
Endress+Hauser 21
4 Wiring
4.1 Connecting the device
#
Warning! Risk of electric shock and/or explosion in hazardous areas! In a wet environment, do not open the cover if voltage is present.
!
Note!
• A suitable circuit breaker must be provided for the device in accordance with IEC/EN61010.
• Devices with integrated overvoltage protection must be grounded.
• Protective circuits against reverse polarity, HF influences and overvoltage peaks are integrated.
The procedure
1. Check that the supply voltage corresponds to the supply voltage indicated on the nameplate.
2. Switch off the supply voltage before connecting the device.
3. Remove housing cover.
4. Guide the cable through the gland. Preferably use a twisted, shielded two-wire cable.
5. Connect the device in accordance with the following diagram.
6. Screw down the housing cover.
7. Switch on the supply voltage.
P01-PMD55xxx-04-xx-xx-xx-010
Electrical connection 4 to 20 mA
1 Terminals for supply and signal 2 Test terminals 3 Ground terminal 4 Supply voltage: 11.5 to 45 VDC (versions with plug-in connectors: 35 V DC) 5 External ground terminal
4…20 mA
11.5 V DC
- +
1
2
3
4
5
Page 22
Wiring Cerabar M Analog
22 Endress+Hauser
4.1.1 Devices with Harting connector Han7D
P01-xMD7xxxx-04-xx-xx-xx-000
Fig. 16: Left: electrical connection for devices with a Harting connector Han7D
Right: view of the connection at the device
4.1.2 Connecting devices with an M12 connector
PIN assignment for M12 connector
4.1.3 Devices with 7/8" plug
PIN assignment for 7/8" connector
4.1.4 Devices with valve connector
P01-xMx5xxxx-04-xx-xx-xx-005
Fig. 17: BN = brown, BU = blue, GNYE = green/yellow; Left: electrical connection for devices with a valve connector
Right: view of the connector at the device
Han7D
+
+–
+
1
5
4
6
7
8 2
3
A0011175
PIN Meaning
1 Signal +
2Not assigned
3 Signal –
4Earth
A0011176
PIN Meaning
1 Signal –
2 Signal +
3Shield
4Not assigned
21
3
4
+
nc
2
1
4
3
+
BNBU GNYE
1
2
3
+
123
Page 23
Cerabar M Analog Wiring
Endress+Hauser 23
4.2 Connecting the measuring unit
4.2.1 Supply voltage
Taking 4 to 20 mA test signal
A 4 to 20 mA test signal may be measured via the test terminals without interrupting the measurement. To keep the corresponding measured error below 0.1 %, the current measuring device should exhibit an internal resistance of < 0.7 .
4.2.2 Cable specification
• Endress+Hauser recommends using twisted, shielded two-wire cables.
• Terminals for wire cross-sections 0.5 to 2.5 mm
2
(20 to 14 AWG)
• Cable outer diameter: 5 to 9 mm (0.2 to 0.35 in) depends on the used cable gland (see technical information)
4.2.3 Load
P01-xxxxxxxx-05-xx-xx-xx-003
Load diagram
Power supply 11.5 to 45 V DC (versions with plug connectors: 35 V DC) for other types of protection and for uncertified device versions
R
Lmax
Maximum load resistance
U Supply voltage
4.2.4 Shielding/potential equalization
You achieve optimum shielding against disturbances if the shielding is connected on both sides (in the cabinet and on the device). If potential equalization currents are expected in the plant, only ground shielding on one side, preferably at the transmitter.
4.3 Potential equalization
Observe the applicable regulations.
Electronic version
4 to 20 mA 11.5 to 45 V DC
(Versions with plug-in connectors: 35 V DC)
30
20
11.5
U
[V]
40 45
1239
1456
804
369
[]W
R
L
max
U – 11.5 V
R
L
max
0.023A
£
Page 24
Wiring Cerabar M Analog
24 Endress+Hauser
4.4 Overvoltage protection (optional)
Devices showing version "NA" in feature 610 "Accessory mounted" in the order code are equipped with overvoltage protection (see also Technical Information TI00436P "Ordering information"). The overvoltage protection is mounted at the factory on the housing thread for the cable gland and is approx. 70 mm (2.76 in) in length (take additional length into account when installing). The device is connected as specified in the following graphic. For details, see TI001013KEN, XA01003KA3 and BA00304KA2.
4.4.1 Wiring
P01-xMx5xxxx-04-xx-xx-en-006
4.4.2 Installation
P01-xMx5xxxx-04-xx-xx-en-007
Unit to be protected
Connection cables
Incoming connection cables
HAW569-DA2B
+
-
+
-
Red
Black
+
-
Shield grounding
without
direct
5
red
red
+
-
+
-
black
black
5
7
10
10
15
min. 1.5 mm²
max. 3.75 Nm
max. 1.5 mm²
0.4 Nm
max. 2.5 mm²
Shield grounding
➁ ➂
without direct
Screw coupling glued at the factory: hold in place with a wrench when loosening/tightening the coupling nut
Page 25
Cerabar M Analog Wiring
Endress+Hauser 25
4.5 Post-connection check
Perform the following checks after completing electrical installation of the device:
• Does the supply voltage match the specifications on the nameplate?
• Is the device connected as per Section 4.1?
• Are all screws firmly tightened?
• Is the housing cover screwed down tight?
As soon as voltage is applied to the device, the green LED on the electronic insert lights up for a few seconds or the connected local display lights up.
Page 26
Operation Cerabar M Analog
26 Endress+Hauser
5Operation
5.1 Position of operating elements
The operating keys and DIP switch are located on the electronic insert in the device.
5.1.1 Function of the DIP switch
P01-Mxxxxxxx-19-xx-xx-xx-010
Fig. 18: Electronic insert
1 Operating keys for lower range value (zero) and upper range value (span) 2 Green LED to indicate successful operation 3 Slot for optional local display 4 DIP switch for switching damping on/off
Switch position
"off" "on"
Damping is switched off. The output signal follows measured value changes without any delay.
Damping is switched on. The output signal follows measured value changes with the delay time Factory setting: = 2 s or as per order specifications).
on
off
damping
1
on
off
Display
Zero
Span
damping
Page 27
Cerabar M Analog Operation
Endress+Hauser 27
5.1.2 Function of the operating elements
5.2 Using the device display (optional)
A 4-line liquid crystal display (LCD) is used. The local display shows measured values, fault messages and notice messages. The display can be removed for easy operation (see diagram, steps 1 - 3). It is connected to the device via a 90 mm (3.54 in) long cable. The device display can be rotated in 90 ° stages (see diagram, steps 4 - 6). Depending on the orientation of the device, this makes it easy to read the measured values.
P01-Mxxxxxxx-19-xx-xx-xx-011
Functions:
• 8-digit measured value display including sign and decimal point, bar graph for 4 to 20 mA as current display.
• Diagnostic functions (fault and warning message etc.)
Operating key(s) Meaning
"Zero"
pressed briefly
Display lower range value
"Zero"
pressed for at least 3 seconds
Get lower range value
The pressure present is accepted as the lower range value (LRV).
"Span"
pressed briefly
Display upper range value
"Span"
pressed for at least 3 seconds
Get upper range value
The pressure present is accepted as the upper range value (LRV).
"Zero" and "Span" pressed together briefly Display position adjustment
"Zero" and "Span" pressed simultaneously
for at least 3 seconds
Position adjustment
The sensor characteristic curve is shifted parallel to itself, so that the pressure present becomes the zero value.
"Zero" and "Span" pressed simultaneously for at least 12 seconds
Reset
All parameters are reset to the order configuration.
1.
2. 3.
4. 5. 6.
Page 28
Operation Cerabar M Analog
28 Endress+Hauser
P01-Mxxxxxxx-07-xx-xx-xx-002
The following table illustrates the symbols that can appear on the local display. Four symbols can occur at one time.
Symbol Meaning
Error message "Out of specification"
The device is being operated outside its technical specifications (e.g. during warmup or cleaning processes).
Error message "Service mode"
The device is in the service mode (during a simulation, for example).
Error message "Maintenance required"
Maintenance is required. The measured value remains valid.
Error message "Failure detected"
An operating error has occurred. The measured value is no longer valid.
Symbol
F
F
Value
Measured value display
Unit
Bargraph
Information line
Main line
S
C
M
F
Page 29
Cerabar M Analog Commissioning
Endress+Hauser 29
6 Commissioning
#
Warning!
• If the pressure present at the device is less that the permitted minimum pressure or greater than the permitted maximum pressure, the message "S" and "Warning" are output alternately.
"
Caution! The measuring range and the unit in which the measured value is displayed correspond to the specifications on the nameplate.
6.1 Function check
Carry out a post-installation and a post-connection check as per the checklist before commissioning the device.
• "Post-installation check" ä 20 checklist
• "Post-connection check" ä 25 checklist
6.2 Commissioning
The following functions are possible via the keys on the electronic insert:
• Position adjustment (zero point correction) A pressure shift resulting from the orientation of the measuring device can be corrected by performing the position adjustment.
• Setting lower range value and upper range value
• Device reset
!
Note!
• The pressure applied must be within the nominal pressure limits of the sensor. See information on the nameplate.
1.) Carrying out position adjustment
2.) Setting lower range value 3.) Setting upper range value 4.) Check configuration
Pressure is present at device. Desired pressure for lower range
value is present at device.
Desired pressure for upper range value is present at device.
Press "Zero" key briefly to display the lower range value.

Press the "Zero" and "Span" keys simultaneously for at least 3 s.
Press the "Zero" key for at least 3 s. Press the "Span" key for at least 3 s. Press "Span" key briefly to display the
upper range value.

Does the LED on the electronic insert light up briefly?
Does the LED on the electronic insert light up briefly?
Does the LED on the electronic insert light up briefly?
Press "Zero" and "Span" keys together briefly to display the calibration offset.
Yes No Yes No Yes No
  
Applied pressure for position adjustment has been accepted.
Applied pressure for position adjustment has not been accepted. Observe the input limits.
Applied pressure for lower range value has been accepted.
Applied pressure for lower range value has not been accepted. Observe the input limits.
Applied pressure for upper range value has been accepted.
Applied pressure for upper range value has not been accepted. Observe the input limits.
Page 30
Maintenance Cerabar M Analog
30 Endress+Hauser
7 Maintenance
Keep the pressure compensation and GORE-TEX® filter (1) free from contamination.
P01-xMx5xxxx-17-xx-xx-xx-000.
7.1 Cleaning instructions
Endress+Hauser offer flushing rings as accessories to clean process isolating diaphragms without taking the transmitters out of the process. For further information please contact your local Endress+Hauser Sales Center.
7.1.1 Cerabar M PMP55
We recommend you perform CIP (cleaning in place (hot water)) before SIP (sterilization in place (steam)) for pipe diaphragm seals. A frequent use of sterilization in place (SIP) will increase the stress on the process isolating diaphragm. Under unfavorable circumstances in the long term view we cannot exclude that a frequent temperature change could lead to a material fatigue of the process isolating diaphragm and possibly to a leakage.
7.2 Exterior cleaning
Please note the following points when cleaning the device:
• The cleaning agents used should not corrode the surface and the seals.
• Mechanical damage to the diaphragm, e.g. due to sharp objects, must be avoided.
• Observe the degree of protection of the device. See the nameplate if necessary (? 6 ff).
1
1
Page 31
Cerabar M Analog Troubleshooting
Endress+Hauser 31
8 Troubleshooting
8.1 Messages
The following is a list of the messages that can occur. The device has four different status information codes in accordance with NE107:
• F = failure
• M (warning) = maintenance required
• C (warning) = function check
• S (warning) = out of specification (deviations from the permitted ambient or process conditions determined by the device with the self-monitoring function, or errors in the device itself indicate that the measuring uncertainty is greater than what would be expected under normal operating conditions).
8.2 Measures
When a message is a displayed, the following steps can be taken:
• Check cable/pressure value
•Restart device
• Perform a reset
If these steps do not correct the error, please contact your Endress+Hauser subsidiary.
8.3 Response of output to errors
In the event of an error, the current output assumes a value of 3.6mA.
8.4 Repair
The Endress+Hauser repair concept provides for measuring devices to have a modular design and that the customer can also carry out repairs (see ä 31, on Section 8.5 "Spare Parts").
!
Note!
• For more information on service and spare parts, contact Endress+Hauser Service. See www.endress.com/worldwide.
8.5 Spare Parts
• Some replaceable measuring device components are identified by means of a spare part nameplate. This contains information about the spare part.
• All the spare parts for the measuring device along with the order code are listed In the W@M Device Viewer (www.endress.com/deviceviewer) and can be ordered. If available, users can also download the associated Installation Instructions.
!
Note! Measuring device serial number:
• Located on the device and spare part nameplate.
Page 32
Troubleshooting Cerabar M Analog
32 Endress+Hauser
8.6 Return
The measuring device must be returned if repairs or a factory calibration are required, or if the wrong measuring device has been ordered or delivered. According to legal regulations, Endress+Hauser, as a ISO-certified company, is required to follow certain procedures when handling returned products that are in contact with process fluids. To ensure swift, safe and professional device returns, please read the return procedures and conditions on the Endress+Hauser website at www.services.endress.com/return-material.
8.7 Disposal
When disposing, separate and recycle the device components based on the materials.
8.8 Software history
Date Software version Software modifications Documentation
Operating Instructions
10.2009 01.00.zz Original software. BA385P/00/EN/10.09 71102503
BA00385P/00/EN/13.10 71125888
BA00385P/00/EN/15.11 71134887
BA00385P/00/EN/16.12 71157152
BA00385P/00/EN/17.12 71191314
BA00385P/00/EN/18.14 71241498
Page 33
Cerabar M Analog Technical data
Endress+Hauser 33
9 Technical data
For the technical data, please refer to the Technical Information TI00436P.
Page 34
Cerabar M Analog Index
34 Endress+Hauser
Index
Numerics
4 to 20 mA test signal. . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
C
Cable specification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
D
Diaphragm seals, installation instructions . . . . . . . . . . . . . 14
Diaphragm seals, vacuum application . . . . . . . . . . . . . . . . 15
Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
E
Electrical connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
I
Incoming acceptance . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Installation instructions for devices with diaphragm seals. . 14 Installation instructions for devices without diaphragm seals . . 11
K
Keys, local, function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
L
Load . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Local display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
M
Measuring arrangement for level measurement . . . . . . . . . 13
Measuring arrangement for pressure measurement . . . 12–13
N
Nameplate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
O
Operating elements, function . . . . . . . . . . . . . . . . . . . . . . 27
Operating elements, position . . . . . . . . . . . . . . . . . . . . . . 26
Operating keys, position . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Overvoltage protection . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
P
Pipe mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Potential equalization . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
R
Repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Returning devices. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
S
Scope of delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Separate housing, assembly and mounting . . . . . . . . . . . . 18
Shielding . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Software history . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Spare Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Storage. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Supply voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
T
Temperature isolator, installation instructions . . . . . . . . . . 15
W
Wall mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Welding recommendation. . . . . . . . . . . . . . . . . . . . . . . . . 19
Page 35
Cerabar M Analog Index
Endress+Hauser 35
Page 36
www.endress.com/worldwide
BA00385P/00/EN/19.14 71269382 CCS/FM+SGML 9
71269382
Page 37
Configurator 2016 SC United States v2
Endress+Hauser
2350 Endress Place
Greenwood, IN 46143
Pressure, capacitive. compact. Application: pressure/level. Membrane: Ceraphire, dry, vacuum proof. :: Easiest on-site operation. :: Modular.
SPK: PAB
71086208
Cerabar M PMC51
PMC51-AA21JD1HGFRKJA
Basic model
Cerabar M PMC51
Approval:
Non-hazardous area
AA
Output:
4-20mA HART
2
Display, Operation:
LCD, push button on display/electronics
1
Housing:
F31 Alu, glass window
J
Electrical Connection:
Thread NPT1/2, IP66/68 NEMA4X/6P
D
Sensor Range:
1bar/100kPa/15psi gauge, 10mH2O/33ftH2O/400inH2O overload: 10bar/1MPa/150psi
1H
Reference Accuracy:
Standard
G
Calibration; Unit:
Sensor range; psi
F
Process Connection:
Thread ANSI MNPT1/2 hole 11.4mm, 316L
RKJ
Seal:
FKM Viton
A
>Additional Operation Language:
>Calibration:
>>Service:
>>Test, Certificate:
>>Additional Approval:
>Separate Housing:
>>Accessory Mounted:
>>Accessory Enclosed:
>Firmware Version:
>>Marking: