Operating Instructions
Flowfit W CCA250
Flow assembly for chlorine sensors
BA062C/07/en/07.09
71099066
Brief overview
Fig. 1: CCA250 with sensors
1 pH sensor (accessories, not in scope of delivery)
2PML connection
3Vent screw
4 Chlorine sensor (accessories, not in scope of delivery)
5 Outlet (adapter depending on version)
6 Needle valve for flow <120 l/h (32 gal/h)
7 Locking cap with calibration vessel
8 Inlet (adapter depending on version)
9 Inductive limit switch for flow measurement (optional)
10 Installation position for a second pH/ORP sensor
a0007322
Table of contents
1 Safety instructions . . . . . . . . . . . . . . . . 4
1.1 Designated use . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.2 Installation, commissioning and operation . . . . . . . . 4
1.3 Operational safety . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.4 Return . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.5 Notes on safety icons and symbols . . . . . . . . . . . . . . 5
2 Identification . . . . . . . . . . . . . . . . . . . . 6
2.1 Nameplate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
2.2 Product structure . . . . . . . . . . . . . . . . . . . . . . . . . . 6
2.3 Scope of delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
3 Installation . . . . . . . . . . . . . . . . . . . . . . 7
3.1 Incoming acceptance, transport, storage . . . . . . . . . . 7
3.2 Installation conditions . . . . . . . . . . . . . . . . . . . . . . . 7
3.3 Installation instructions . . . . . . . . . . . . . . . . . . . . . . 8
3.4 Post-installation check . . . . . . . . . . . . . . . . . . . . . . 10
4 Commissioning. . . . . . . . . . . . . . . . . . 10
5 Maintenance. . . . . . . . . . . . . . . . . . . . 11
5.1 Cleaning the assembly . . . . . . . . . . . . . . . . . . . . . . 11
5.2 Cleaning the sensor . . . . . . . . . . . . . . . . . . . . . . . . 11
5.3 Cleaning agents . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
6 Accessories. . . . . . . . . . . . . . . . . . . . . 13
6.1 Accessories kits . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
6.2 Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
7 Trouble-shooting . . . . . . . . . . . . . . . . 15
7.1 Replacing damaged parts . . . . . . . . . . . . . . . . . . . . 15
7.2 Return . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
7.3 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
8 Technical data . . . . . . . . . . . . . . . . . . 16
8.1 Environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
8.2 Process . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
8.3 Mechanical construction . . . . . . . . . . . . . . . . . . . . 17
Index . . . . . . . . . . . . . . . . . . . . . . . . . 18
3
Safety instructions
1 Safety instructions
1.1 Designated use
The flow assembly has been designed for the installation of the membrane covered chlorine sensors
CCS120, CCS140, CCS141, CCS240, CCS241 and CCS142D. Additionally, there are two more
installation positions for the installation of sensors with Pg 13.5 and installation length 120 mm
(4.72 inch), e.g. pH resp. ORP sensors.
Its mechanical design permits its use in pressurised systems (see "Technical data").
Any other use than the one described here compromises the safety of persons and the entire
measuring system and is not permitted.
The manufacturer is not liable for damage caused by improper or non-designated use.
1.2 Installation, commissioning and operation
Please note the following items:
• Installation, commissioning, operation and maintenance of the measuring system must only be
carried out by trained technical personnel.
Trained personnel must be authorized for the specified activities by the system operator.
• Electrical connection must only be carried out by a certified electrician.
• Technical personnel must have read and understood these Operating Instructions and must
adhere to them.
• Before commissioning the entire measuring point, check all the connections. Ensure that
electrical cables and hose connections are not damaged.
• Do not operate damaged products and secure them against unintentional commissioning.
Mark the damaged product as being defective.
• Measuring point faults may only be rectified by authorized and specially trained personnel.
• If faults can not be rectified, the products must be taken out of service and secured against
unintentional commissioning.
• Repairs not described in these Operating Instructions may only be carried out at the
manufacturer’s or by the service organization.
1.3 Operational safety
The assembly has been designed and tested in accordance with the latest industry standards and left
the factory in perfect functioning order.
Relevant regulations and standards have been met.
As the user, you are responsible for complying with the following safety conditions:
• Installation instructions
• Local prevailing standards and regulations.
1.4 Return
If the assembly has to be repaired, please return it cleaned to the appropriate sales center.
Please use the original packaging, if possible.
Please enclose the completed "Declaration of contamination" (copy the second to last page of these
Operating Instructions) with the packaging and the transportation documents.
No repair without completed "Declaration of contamination"!
4
#
"
!
Safety instructions
1.5 Notes on safety icons and symbols
Warning!
This symbol alerts you to hazards that can cause serious damage to the instrument or to persons if
ignored.
Caution!
This symbol alerts you to possible faults which could arise from incorrect operation. They could
cause damage to the instrument if ignored.
Note!
This symbol indicates important items of information.
5
Identification
2 Identification
2.1 Nameplate
You can read the following information from the nameplate on the assembly:
•Product name
•Order code
• Serial number or week code
• Permissible process pressure and maximum process temperature
You can find possible assembly versions and the resulting order codes in the product structure.
2.2 Product structure
Process connection, adapter
A G1/2, none
B G1/2, 2x NV 1/2"
C G1/2, 2x D 6/12
D G1/2, D 6/12 + D 16
M NPT 1/2", none
Q NPT 1/4", none
Proximity switch
0none
1 with inductive proximity switch
CCA250- complete order code
2.3 Scope of delivery
The scope of delivery comprises:
• Flowfit assembly (ordered version)
• a PML adapter
• Operating Instructions (English)
If you have any questions, please contact your supplier or your local sales center.
6