If necessary, Consequences of noncompliance (if applicable)
Corrective action
‣
WARNING
L
Causes (/consequences)
If necessary, Consequences of noncompliance (if applicable)
Corrective action
‣
CAUTION
L
Causes (/consequences)
If necessary, Consequences of noncompliance (if applicable)
Corrective action
‣
NOTICE
Cause/situation
If necessary, Consequences of noncompliance (if applicable)
Action/note
‣
This symbol alerts you to a dangerous situation.
Failure to avoid the dangerous situation will result in a fatal or serious injury.
This symbol alerts you to a dangerous situation.
Failure to avoid the dangerous situation can result in a fatal or serious injury.
This symbol alerts you to a dangerous situation.
Failure to avoid this situation can result in minor or more serious injuries.
This symbol alerts you to situations which may result in damage to property.
1.2 Symbols used
SymbolMeaning
Additional information, tips
Permitted or recommended
Not permitted or not recommended
Reference to device documentation
Reference to page
Reference to graphic
Result of a step
4Endress+Hauser
Page 5
Flowfit CCA151Document information
1.2.1 Symbols on the device
SymbolMeaning
Reference to device documentation
Endress+Hauser5
Page 6
Basic safety instructionsFlowfit CCA151
2 Basic safety instructions
2.1 Requirements for the personnel
• Installation, commissioning, operation and maintenance of the measuring system may be
carried out only by specially trained technical personnel.
• The technical personnel must be authorized by the plant operator to carry out the specified
activities.
• The electrical connection may be performed only by an electrical technician.
• The technical personnel must have read and understood these Operating Instructions and
must follow the instructions contained therein.
• Faults at the measuring point may only be rectified by authorized and specially trained
personnel.
Repairs not described in the Operating Instructions provided must be carried out only
directly at the manufacturer's site or by the service organization.
2.2 Designated use
The flow assembly has been specially developed to hold the membrane-covered disinfection
sensors CCS5xD.
Thanks to its design, it can be used in pressurized systems → 28.
Use of the device for any purpose other than that described, poses a threat to the safety of
people and of the entire measuring system and is therefore not permitted.
The manufacturer is not liable for damage caused by improper or non-designated use.
2.3 Occupational safety
As the user, you are responsible for complying with the following safety conditions:
• Installation guidelines
• Local standards and regulations
• Regulations for explosion protection
6Endress+Hauser
Page 7
Flowfit CCA151Basic safety instructions
2.4 Operational safety
Before commissioning the entire measuring point:
1.Verify that all connections are correct.
2.Ensure that electrical cables and hose connections are undamaged.
3.Do not operate damaged products, and protect them against unintentional operation.
4.Label damaged products as defective.
During operation:
If faults cannot be rectified:
‣
products must be taken out of service and protected against unintentional operation.
2.5 Product safety
2.5.1 State of the art
The product is designed to meet state-of-the-art safety requirements, has been tested, and
left the factory in a condition in which it is safe to operate. The relevant regulations and
European standards have been observed.
Endress+Hauser7
Page 8
Product descriptionFlowfit CCA151
4
1
2
5
3
6
3 Product description
3.1 Product design
1Disinfection sensor CCS5xD (membrane-covered, ⌀25 mm), e.g. CCS50D
2Coupling nut (for securing a disinfection sensor)
3Flowfit CCA151 flow assembly
4Outlet from Flowfit CCA151 flow assembly (internal thread G 1/8") and hose connection
5Inlet to Flowfit CCA151 flow assembly (internal thread G 1/8")
6Openings (M5 Allen screw x 60) for securing Flowfit CCA151 flow assembly
8Endress+Hauser
A0034257
Page 9
Flowfit CCA151Incoming acceptance and product identification
4 Incoming acceptance and product identification
4.1 Incoming acceptance
1.Verify that the packaging is undamaged.
Notify the supplier of any damage to the packaging.
Keep the damaged packaging until the issue has been resolved.
2.Verify that the contents are undamaged.
Notify the supplier of any damage to the delivery contents.
Keep the damaged goods until the issue has been resolved.
3.Check that the delivery is complete and nothing is missing.
Compare the shipping documents with your order.
4.Pack the product for storage and transportation in such a way that it is protected
against impact and moisture.
The original packaging offers the best protection.
Make sure to comply with the permitted ambient conditions (→ Technical data).
If you have any questions, please contact your supplier or your local Sales Center.
Endress+Hauser9
Page 10
Incoming acceptance and product identificationFlowfit CCA151
4.2 Product identification
4.2.1 Nameplate
The nameplate provides you with the following information on your device:
• Manufacturer identification
• Order code
• Extended order code
• Serial number
• Ambient and process conditions
• Safety information and warnings
Compare the information on the nameplate with the order.
‣
4.2.2 Product identification
Product page
www.endress.com/cca151
Interpreting the order code
The order code and serial number of your product can be found in the following locations:
• On the nameplate
• In the delivery papers
Obtaining information on the product
1.Open the product website.
2.At the bottom of the page, select the link Online Tools and then select Access devicespecific information .
An additional window opens.
3.Enter the order code from the nameplate into the search field. Then select Showdetails .
Details of each feature (selected option) of the order code are displayed.
Manufacturer address
Endress+Hauser Conducta GmbH+Co. KG
Dieselstraße 24
D-70839 Gerlingen
Scope of delivery
The scope of delivery comprises:
• Assembly in the version ordered
• Operating Instructions
10Endress+Hauser
Page 11
Flowfit CCA151Installation
Certificates and approvals
RL 2014/34/EU (ATEX)
The assembly does not fall within the scope of the directive. However, if conditions for safe
use are adhered to, it may be deployed in the hazardous area.
RL 2014/68/EU PED
The assembly was manufactured in accordance with Article 3, paragraph 3, Pressure
Equipment Directive 97/23/EC in accordance with good engineering practice.
5 Installation
5.1 Installation conditions
5.1.1 Installation position
The assembly is designed for installation on panels, walls or level surfaces. G 1/8" connections
and a hose connection with an external diameter of 6 mm and an internal diameter of 4 mm
are provided for this purpose.
The assembly is designed in such a way that it must be installed vertically.
The sensor that is used can restrict the orientation.
Endress+Hauser11
Page 12
InstallationFlowfit CCA151
110 (4.33)
85 (3.35)
!14.5
(0.57)
7.5 (0.3)
70 (2.76)
70 (2.76)
105 (4.13)
130 (5.12)
150 (5.91)
194.5 (7.66)
M5 60·
M5 60·
150 (5.91)
219.5 (8.64)
! 25
(0.98)
! 14.5
(0.57)
! 39.5
(1.56)
5.1.2 Dimensions
1Front view
To allow the sensor to be removed and for operation with Memosens data cables, the spacing
required for mounting is 150 mm (5.91 in).
12Endress+Hauser
A0034258
Page 13
Flowfit CCA151Installation
150 (5.91)
85 (3.35)
60 (2.36)
2Side view
To allow the sensor to be removed and for operation with Memosens data cables, the spacing
required for mounting is 150 mm (5.91 in).
1Transmitter Liquiline CM44x
2Power cable for transmitter
3Disinfection sensor CCS5xD (membrane-covered, ⌀25 mm), e.g. CCS50D
4Outlet from Flowfit CCA151 flow assembly
5Inlet to Flowfit CCA151 flow assembly
6Flowfit CCA151 flow assembly
7Coupling nut for installing sensor CCS50D in Flowfit CCA151 flow assembly
8Measuring cable CYK10
5.2.2 Mounting assembly on a vertical surface
To secure the assembly to a wall, use the wall mounting kit which can be ordered as an
‣
accessory→ 28.
For detailed information on mounting the assembly on a wall using the wall mounting
kit, see kit instructions for wall mounting kit
Endress+Hauser15
Page 16
InstallationFlowfit CCA151
5.2.3 Mounting assembly in the process
WARNING
L
Risk of injury from high pressure, high temperature or chemical hazards if process
medium escapes.
Wear protective gloves, protective goggles and protective clothing.
‣
Mount the assembly only if vessels or pipes are empty and unpressurized.
‣
Prior to installation, check the flange seal between the flanges.
1.Mount assembly on vertical surface→ 15.
2.Connect to the vessel or pipe via the process connection.
Bypass operation
To achieve flow through the assembly with a bypass, pressure p1 must be higher than
pressure p2.
This requires the installation of an orifice plate or throttle valve in the main pipe.
p1 must not exceed the permitted operating pressure for the assembly of 4 bar (58 psi).
If the sensor is installed, the sensor's pressure specifications must also be observed.
1.Mount the assembly vertically.
2.Connect the medium using the usual commercial connection fittings. Depending on
requirements, use the usual sealing materials, e.g. PMMA-compatible thread adhesive,
Teflon tape or O-ring made of FKM.
3.Installing the assembly in the bypass is preferable to installing it directly in the process
line. The bypass line can be blocked off without interrupting the process (a shut-off
valve is required upstream and downstream). This allows, for example, the sensor to be
cleaned without restricting the process.
4.Install a dirt trap (filter) with a mesh size of 500 μm upstream from the assembly. If a
pressure-reducing valve is used, it usually includes a dirt trap.
5.Set flow value upstream from the assembly, e. g. by installation of a flow setting.
6.Install a DN5-8 tap downstream from the assembly outlet to allow samples to be taken
for reference measurements in accordance with the DPD method.
16Endress+Hauser
Page 17
Flowfit CCA151Installation
123
4
p1
p2
p1 > p2
A0034259
4Connection example with bypass and orifice plate in the main pipe
1Shut-off valve (provided by customer)
2Pressure-reducing valve (at p1 > 4 bar (58 psi)) (provided by customer)
3Shut-off valve (provided by customer)
4Orifice plate in process line (provided by customer)
Open outlet operation
p must not exceed the permitted operating pressure for the assembly of 4 bar (58 psi).
If the sensor is installed, the sensor's pressure specifications must also be observed.
If the medium pressure is above 4 bar (58 psi), a pressure-reducing valve is required.
1.Mount the assembly vertically.
2.Connect the medium using the usual commercial connection fittings. Depending on
requirements, use the usual sealing materials, e.g. PMMA-compatible thread adhesive,
Teflon tape or O-ring made of FKM.
3.Installing the assembly in the bypass is preferable to installing it directly in the process
line. The bypass line can be blocked off without interrupting the process (a shut-off
valve is required upstream and downstream). This allows, for example, the sensor to be
cleaned without restricting the process.
4.Install a dirt trap (filter) with a mesh size of 500 μm upstream from the assembly. If a
pressure-reducing valve is used, it usually includes a dirt trap.
5.Set flow value upstream from the assembly, e. g. by installation of a flow setting.
6.Install a DN5-8 tap downstream from the assembly outlet to allow samples to be taken
for reference measurements in accordance with the DPD method.
Endress+Hauser17
Page 18
InstallationFlowfit CCA151
1
2
3
p
A0034260
5Connection example with open outlet
1Shut-off valve (provided by customer)
2Pressure-reducing valve (at p > 4 bar (58 psi)) (provided by customer)
3Shut-off valve (provided by customer)
18Endress+Hauser
Page 19
Flowfit CCA151Installation
1
5.3 Installing sensor in assembly
The CCS50D disinfection sensor (membrane-covered, ⌀25 mm) is designed for installation in
the Flowfit CCA151 flow assembly.
Please note the following during installation:
The flow rate must be at least 5 l/h (1.32 gal/h).
‣
If the medium is fed back into an overflow basin, pipe or similar, the resulting
‣
counterpressure on the sensor may not exceed 1 bar (14.5 psi) and must remain constant.
Negative pressure at the sensor, e.g. due to medium being returned to the suction side of a
‣
pump, must be avoided.
Preparing assembly
1.The assembly is supplied to the customer with a coupling nut screwed onto the
assembly: unscrew coupling nut from assembly.
A0034262
6Flowfit CCA151 flow assembly
1Thread adapter nut
2.The assembly is supplied to the customer with a dummy plug inserted in the assembly:
remove dummy plug from assembly.
Endress+Hauser19
Page 20
InstallationFlowfit CCA151
Removing protection cap from sensor
NOTICE
Negative pressure causes damage to the sensor's membrane cap.
If the protection cap is attached, carefully remove protection cap from sensor.
‣
1.When supplied to the customer and when in storage, the sensor is fitted with a
protection cap: First release just the top part of the protection cap by turning it.
A0034263
7Releasing top part of protection cap by turning
2.Carefully remove protection cap from sensor.
A0034350
8Carefully remove protection cap.
20Endress+Hauser
Page 21
Flowfit CCA151Installation
Fit the adapter to the sensor.
The required adapter (clamping ring) can be ordered as a sensor accessory or as a separate
accessory.
1.Slide the adapter for CCA151 (clamping ring) from the membrane cap towards the
sensor head and into the lower groove.
A0034247
9Slide the adapter (clamping ring) upwards from the membrane camp to the senor
shaft and into the lower groove.
Installing sensor in assembly
2.Slide the sensor with adapter for CCA151 (clamping ring) into the opening in the
assembly.
Endress+Hauser21
Page 22
CommissioningFlowfit CCA151
1
2
3
3.Screw coupling nut onto assembly on block.
A0034261
1Disinfection sensor CCS50D
2Flowfit CCA151 flow assembly
3Coupling nut for securing a CCS50D disinfection sensor
5.4 Post-installation check
1.After mounting, check all connections to ensure they are fully secured and leak-tight.
2.Check all hoses for damage.
6 Commissioning
Prior to commissioning, ensure that:
• all seals are correctly seated (on the assembly and on the process connection)
• the sensor is correctly installed and connected
WARNING
L
Risk of injury from high pressure, high temperature or chemical hazards if process
medium escapes.
Before subjecting the assembly to the process pressure, verify that all connections are
‣
sealed.
22Endress+Hauser
Page 23
Flowfit CCA151Maintenance
7 Maintenance
WARNING
L
Risk of injury if medium escapes
Before each maintenance task, ensure that the process pipe is empty and rinsed.
‣
7.1 Maintenance schedule
The specified intervals serve as a guide. For harsh process or ambient conditions, it is
recommended that the interval be shortened accordingly. Cleaning intervals for the
sensor and assembly are dependent on the medium.
To ensure stable and reliable measurements, the assembly and the sensor must be cleaned
regularly. The frequency and intensity of the cleaning process depend on the medium.
7.2.1 Cleaning the assembly
A typical example of a cleaning interval would be 6 months in the case of drinking water.
1.All parts in contact with the medium, such as the sensor and the sensor guide, must be
cleaned regularly. To do so, remove the sensor → 26.
2.Remove light dirt and fouling using a cloth moistened with suitable cleaning solutions.
3.Remove heavy soiling using a soft brush and a suitable cleaning agent.
4.For very persistent dirt, soak the parts in a cleaning solution. Then clean the parts with
a brush.
24Endress+Hauser
Page 25
Flowfit CCA151Maintenance
Cleaning agent
The choice of cleaning agent depends on the degree and type of contamination. The most
common types of contamination and the appropriate cleaning agents can be found in the
following table.
Type of soilingCleaning agent
Greases and oilsHot water or water-soluble organic solvents (e.g.
Do not use any halogen-containing, organic solvents or acetone. These solvents may
‣
destroy plastic components of the sensor and are also suspected carcinogens (e.g.
chloroform).
NOTICE
Damage to sensor membrane
The sensor membrane must not come into contact with surfactant-containing agents.
‣
7.2.2 Cleaning the sensor
Cleaning the sensor
1.Prior to calibration if dirt is visible on the surface
2.Regularly during operation
3.Before returning it for repairs
For detailed information on "Cleaning the sensor", see Operating Instructions for sensor.
Endress+Hauser25
Page 26
MaintenanceFlowfit CCA151
1
2
3
7.2.3 Removing sensor from assembly
Remove sensor from assembly CCA151
1.Unscrew coupling nut from assembly.
A0034261
1Disinfection sensor CCS50D
2Flowfit CCA151 flow assembly
3Coupling nut for securing a CCS50D disinfection sensor
2.Pull sensor out through opening in assembly.
26Endress+Hauser
Page 27
Flowfit CCA151Repairs
8 Repairs
WARNING
L
Danger resulting from improper repair
Damage to the assembly, which compromises pressure safety, must be repaired only by
‣
authorized and qualified personnel.
Following each repair and maintenance task, the assembly must be checked for leaks using
‣
appropriate procedures. Following this, the assembly must again comply with the
specifications in the technical data.
Replace all other damaged components immediately.
8.1 Spare parts
For more detailed information on spare parts kits, please refer to the "Spare Part Finding Tool"
on the Internet:
www.endress.com/spareparts_consumables
Description and contentsOrder No.
Kit CCA151, complete with blind plug
• Blind plug with O-ring
• Screw plug
8.2 Return
The product must be returned if repairs or a factory calibration are required, or if the wrong
product was ordered or delivered. As an ISO-certified company and also due to legal
regulations, Endress+Hauser is obliged to follow certain procedures when handling any
returned products that have been in contact with medium.
71372020
To ensure the swift, safe and professional return of the device:
Refer to the website www.endress.com/support/return-material for information on the
‣
procedure and conditions for returning devices.
8.3 Disposal
The device contains electronic components. and must therefore be disposed of in accordance
with regulations on the disposal of electronic waste.
Observe the local regulations.
‣
Endress+Hauser27
Page 28
AccessoriesFlowfit CCA151
9 Accessories
The following are the most important accessories available at the time this documentation
was issued.
For accessories not listed here, please contact your Service or Sales Center.
‣
9.1 Device-specific accessories
9.1.1 Disinfection sensor
CCS50D
• Membrane-covered amperometric sensor for chlorine dioxide
• With Memosens technology
• Product Configurator on the product page: www.endress.com/ccs50d
Technical Information TI01353C
9.1.2 Wall mounting kit
Wall mounting kit for CCA151
Order No. 71372109
10 Technical data
10.1 Environment
10.1.1 Ambient temperature range
-20 to +60 °C (-4 to 140 °F)
10.1.2 Storage temperature
-20 to +60 °C (-4 to 140 °F)
10.2 Process
10.2.1 Process temperature
0 to 60 °C (32 to 140 °F), non-freezing
10.2.2 Process pressure
0 to 4 bar (0 to 58 psi) relative
28Endress+Hauser
Page 29
Flowfit CCA151Technical data
p [bar]
T[°C]
2060
40
T[°F]
p [psi]
0
32
0
0
68104140
4
58
10.2.3 Pressure-temperature ratings
A0034376-EN
10Pressure-temperature ratings
10.2.4 Process connections
G1/8"
Hose specification:External diameter 6 mm, internal diameter 4 mm
10.2.5 Flow
remains stable for a minimum of 5 l/h (1.32 gal/hr) when using 25 mm (0.98 in) disinfection
sensors with approx. 77 mm (3.03 in) immersion depth
Endress+Hauser29
Page 30
Technical dataFlowfit CCA151
10.3 Mechanical construction
→ 12
10.3.1 Weight
1.077 kg (2.37 lbs)
10.3.2 Materials
In contact with medium
Assembly:PMMA
Seals:PVDF
Dummy plug:PVC, FKM
30Endress+Hauser
Page 31
Flowfit CCA151Index
Index
A
Accessories ..................... 28
Approvals ...................... 11
C
Certificates ..................... 11
Cleaning agent ................... 25
Commissioning ...................22
D
Designated use ....................6
Dimensions ..................... 12
Disposal ....................... 27
I
Incoming acceptance ................ 9
Installation ................... 11, 14
Installation conditions ...............11
Installing the sensor ................19
M
Maintenance .................... 23
Maintenance schedule ...............23
Maintenance tasks .................24
Measuring system ................. 14
N
Nameplate ..................... 10
U
Use ...........................6
W
Warnings ....................... 4
P
Post-installation check .............. 22
Product identification ................ 9
R
Repairs ........................27
Return ........................ 27
S
Safety instructions ..................6
Scope of delivery .................. 10
Spare parts ..................... 27
Symbols ........................ 4
T
Technical data ................... 28
Endress+Hauser31
Page 32
*71371624*
71371624
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.