A grounded terminal which, as far as the operator is concerned, is grounded via a
grounding system.
Protective Earth (PE)
A terminal which must be connected to ground prior to establishing any other
connections.
The ground terminals are situated inside and outside the device:
• Inner ground terminal: Connects the protectiv earth to the mains supply.
• Outer ground terminal: Connects the device to the plant grounding system.
Communication symbols
SymbolMeaning
Wireless Local Area Network (WLAN)
Communication via a wireless, local network.
LED
Light emitting diode is off.
LED
Light emitting diode is on.
LED
Light emitting diode is flashing.
Symbols for certain types of information
SymbolMeaning
Permitted
Procedures, processes or actions that are permitted.
Preferred
Procedures, processes or actions that are preferred.
Forbidden
Procedures, processes or actions that are forbidden.
Tip
Indicates additional information.
Reference to documentation.
Reference to page.
Reference to graphic.
Visual inspection.
4Endress+Hauser
Page 5
Proline Promass A 300
1.
-
.
2
1
3
Symbols in graphics
SymbolMeaning
1, 2, 3, ...Item numbers
, 2., 3., …Series of steps
A, B, C, ...Views
A-A, B-B, C-C, ...Sections
Hazardous area
Safe area (non-hazardous area)
Flow direction
Function and system design
Measuring principle
The measuring principle is based on the controlled generation of Coriolis forces. These forces are
always present in a system when both translational and rotational movements are superimposed.
Fc = 2 · ∆m (ν · ω)
Fc = Coriolis force
∆m = moving mass
ω = rotational velocity
ν = radial velocity in rotating or oscillating system
The amplitude of the Coriolis force depends on the moving mass ∆m, its velocity ν in the system and
thus on the mass flow. Instead of a constant rotational velocity ω, the sensor uses oscillation.
In the sensor, an oscillation is produced in the measuring tube. The Coriolis forces produced at the
measuring tube cause a phase shift in the tube oscillations (see illustration):
• If there is zero flow (i.e. when the fluid stands still), the oscillation measured at points A and B has
the same phase (no phase difference) (1).
• Mass flow causes deceleration of the oscillation at the inlet of the tubes (2) and acceleration at the
outlet (3).
A0029932
The phase difference (A-B) increases with increasing mass flow. Electrodynamic sensors register the
tube oscillations at the inlet and outlet. The measuring principle operates independently of
temperature, pressure, viscosity, conductivity and flow profile.
Endress+Hauser5
Page 6
Proline Promass A 300
Density measurement
The measuring tube is continuously excited at its resonance frequency. A change in the mass and
thus the density of the oscillating system (comprising measuring tube and fluid) results in a
corresponding, automatic adjustment in the oscillation frequency. Resonance frequency is thus a
function of medium density. The microprocessor utilizes this relationship to obtain a density signal.
Volume measurement
Together with the measured mass flow, this is used to calculate the volume flow.
Temperature measurement
The temperature of the measuring tube is determined in order to calculate the compensation factor
due to temperature effects. This signal corresponds to the process temperature and is also available
as an output signal.
Measuring system
The device consists of a transmitter and a sensor.
The device is available as a compact version:
The transmitter and sensor form a mechanical unit.
Transmitter
Promass 300
Device versions and materials:
• Transmitter housing
• Aluminum, coated: aluminum, AlSi10Mg, coated
• Stainless, hygienic: stainless steel, 1.4404
• Material of window in transmitter housing:
• Aluminum, coated: glass
• Stainless, hygienic: polycarbonate
Configuration:
A0026708
• External operation via 4-line, backlit, graphic local display with touch
control and guided menus ("Make-it-run" wizards) for applicationspecific commissioning.
• Via service interface or WLAN interface:
• Operating tools (e.g. FieldCare, DeviceCare)
• Web server (access via Web browser, e.g. Microsoft Internet
Explorer, Microsoft Edge)
Sensor
Promass A• Bent single-tube system for high-precision measurement of minimum
flow rates
• Simultaneous measurement of flow, volume flow, density and
temperature (multivariable)
• Immune to process influences
• Nominal diameter range: DN 1 to 4 (¹⁄₂₄ to ¹⁄₈ ")
• Process connections: stainless steel, 1.4404 (316/316L); 1.4435
(316L); Alloy C22, 2.4602 (UNS N06022)
6Endress+Hauser
Page 7
Proline Promass A 300
2
1
6
5
7
4
3
3
Equipment architecture
A0027512
1Possibilities for integrating measuring devices into a system
1Control system (e.g. PLC)
2Connecting cable (0/4 to 20 mA HART etc.)
3Fieldbus
4Coupler
5Non-hazardous area
6Hazardous area: Zone 2; Class I, Division 2
7Hazardous area: Zone 1; Class I, Division 1
SafetyIT security
Our warranty is valid only if the device is installed and used as described in the Operating
Instructions. The device is equipped with security mechanisms to protect it against any inadvertent
changes to the settings.
IT security measures, which provide additional protection for the device and associated data transfer,
must be implemented by the operators themselves in line with their security standards.
Device-specific IT security
The device offers a range of specific functions to support protective measures on the operator's side.
These functions can be configured by the user and guarantee greater in-operation safety if used
correctly. An overview of the most important functions is provided in the following section.
Function/interfaceFactory settingRecommendation
Write protection via hardware write
protection switch → 8
Access code
(also applies for Web server login or
FieldCare connection) → 8
WLAN
(order option in display module)
WLAN security modeEnabled (WPA2-
WLAN passphrase
(password) → 8
WLAN modeAccess PointOn an individual basis following risk
Web server→ 8Enabled.On an individual basis following risk
CDI-RJ45 service interface → 9–On an individual basis following risk
Not enabled.On an individual basis following risk
assessment.
Not enabled
(0000).
Enabled.On an individual basis following risk
PSK)
Serial numberAssign an individual WLAN passphrase during
Assign a customized access code during
commissioning.
assessment.
Do not change.
commissioning.
assessment.
assessment.
assessment.
Endress+Hauser7
Page 8
Proline Promass A 300
Protecting access via hardware write protection
Write access to the device parameters via the local display, Web browser or operating tool (e.g.
FieldCare, DeviceCare) can be disabled via a write protection switch (DIP switch on the
motherboard). When hardware write protection is enabled, only read access to the parameters is
possible.
Hardware write protection is disabled when the device is delivered.
Protecting access via a password
Different passwords are available to protect write access to the device parameters or access to the
device via the WLAN interface.
• User-specific access code
Protect write access to the device parameters via the local display, Web browser or operating tool
(e.g. FieldCare, DeviceCare). Access authorization is clearly regulated through the use of a userspecific access code.
• WLAN passphrase
The network key protects a connection between an operating unit (e.g. notebook or tablet) and the
device via the WLAN interface which can be ordered as an option.
• Infrastructure mode
When the device is operated in infrastructure mode, the WLAN passphrase corresponds to the
WLAN passphrase configured on the operator side.
User-specific access code
Write access to the device parameters via the local display, Web browser or operating tool (e.g.
FieldCare, DeviceCare) can be protected by the modifiable, user-specific access code.
WLAN passphrase: Operation as WLAN access point
A connection between an operating unit (e.g. notebook or tablet) and the device via the WLAN
interface, which can be ordered as an optional extra, is protected by the network key. The WLAN
authentication of the network key complies with the IEEE 802.11 standard.
When the device is delivered, the network key is pre-defined depending on the device. It can be
changed via the WLAN settings submenu in the WLAN passphrase parameter.
Infrastructure mode
A connection between the device and WLAN access point is protected by means of an SSID and
passphrase on the system side. Please contact the relevant system administrator for access.
General notes on the use of passwords
• The access code and network key supplied with the device should be changed during
commissioning.
• Follow the general rules for generating a secure password when defining and managing the access
code or network key.
• The user is responsible for the management and careful handling of the access code and network
key.
Access via Web server
The device can be operated and configured via a Web browser with the integrated Web server. The
connection is via the service interface (CDI-RJ45) or the WLAN interface. For device versions with
the EtherNet/IP and PROFINET communication protocols, the connection can also be established via
the terminal connection for signal transmission with EtherNet/IP or PROFINET (RJ45 connector).
The Web server is enabled when the device is delivered. The Web server can be disabled if necessary
(e.g. after commissioning) via the Web server functionality parameter.
The device and status information can be hidden on the login page. This prevents unauthorized
access to the information.
For detailed information on device parameters, see:
The "Description of Device Parameters" document → 109
Access via OPC-UA
The "OPC UA Server" application package is available in the device version with the HART
communication protocol → 105.
8Endress+Hauser
Page 9
Proline Promass A 300
The device can communicate with OPC UA clients using the "OPC UA Server" application package.
The OPC UA server integrated in the device can be accessed via the WLAN access point using the
WLAN interface - which can be ordered as an optional extra - or the service interface (CDI- RJ45) via
Ethernet network. Access rights and authorization as per separate configuration.
The following Security Modes are supported as per the OPC UA Specification (IEC 62541):
• None
• Basic128Rsa15 – signed
• Basic128Rsa15 – signed and encrypted
Access via service interface (CDI-RJ45)
The device can be connected to a network via the service interface (CDI-RJ45). Device-specific
functions guarantee the secure operation of the device in a network.
The use of relevant industrial standards and guidelines that have been defined by national and
international safety committees, such as IEC/ISA62443 or the IEEE, is recommended. This includes
organizational security measures such as the assignment of access authorization as well as technical
measures such as network segmentation.
Transmitters with an Ex de approval may not be connected via the service interface (CDI-RJ45)!
Order code for "Approval transmitter + sensor", options (Ex de): BA, BB, C1, C2, GA, GB, MA,
MB, NA, NB
The device can be integrated in a ring topology. The device is integrated via the terminal
connection for signal transmission (output 1) and the connection to the service interface (CDIRJ45) .
Endress+Hauser9
Page 10
Input
Measured variableDirect measured variables
• Mass flow
• Density
• Temperature
Calculated measured variables
• Volume flow
• Corrected volume flow
• Reference density
Measuring rangeMeasuring range for liquids
Proline Promass A 300
DNMeasuring range full scale values
[mm][in][kg/h][lb/min]
1¹⁄₂₄0 to 200 to 0.735
2¹⁄₁₂0 to 1000 to 3.675
4¹⁄₈0 to 4500 to 16.54
min(F)
to
Measuring range for gases
The full scale value depends on the density and the sound velocity of the gas used and can be
calculated with the formula below:
= minimum (
max(G)
max(G)
max(F)
<
max(G)
ρ
G
xConstant dependent on nominal diameter
c
G
d
i
max(F)
· ρG : x ; ρG · cG · π/2 · (di)2 · 3600)
max(F)
Maximum full scale value for gas [kg/h]
Maximum full scale value for liquid [kg/h]
can never be greater than
max(G)
Gas density in [kg/m³] at operating conditions
Sound velocity (gas) [m/s]
Measuring tube internal diameter [m]
max(F)
max(F)
DNx
[mm][in][kg/m3]
1¹⁄₂₄32
2¹⁄₁₂32
4¹⁄₈32
To calculate the measuring range, use the Applicator sizing tool → 107
Calculation example for gas
• Sensor: Promass A, DN 2
• Gas: Air with a density of 11.9 kg/m³ (at 20 °C and 10 bar)
Flow rates above the preset full scale value do not override the electronics unit, with the result that
the totalizer values are registered correctly.
Input signalInput and output versions
→ 13
External measured values
To increase the accuracy of certain measured variables or to calculate the corrected volume flow for
gases, the automation system can continuously write different measured values to the measuring
device:
• Operating pressure to increase accuracy (Endress+Hauser recommends the use of a pressure
measuring device for absolute pressure, e.g. Cerabar M or Cerabar S)
• Medium temperature to increase accuracy (e.g. iTEMP)
• Reference density for calculating the corrected volume flow for gases
Various pressure transmitters and temperature measuring devices can be ordered from Endress
+Hauser: see "Accessories" section → 108
It is recommended to read in external measured values to calculate the corrected volume flow.
HART protocol
The measured values are written from the automation system to the measuring device via the HART
protocol. The pressure transmitter must support the following protocol-specific functions:
• HART protocol
• Burst mode
Current input
The measured values are written from the automation system to the measuring device via the
current input → 11.
Digital communication
The measured values can be written from the automation system to the measuring via:
• FOUNDATION Fieldbus
• PROFIBUS DP
• PROFIBUS PA
• Modbus RS485
• EtherNet/IP
• PROFINET
Current input 0/4 to 20 mA
Current input0/4 to 20 mA (active/passive)
Current span• 4 to 20 mA (active)
• 0/4 to 20 mA (passive)
Resolution1 µA
Voltage dropTypically: 0.6 to 2 V for 3.6 to 22 mA (passive)
Maximum input voltage≤ 30 V (passive)
Open-circuit voltage≤ 28.8 V (active)
Possible input variables• Pressure
• Temperature
• Density
•
Endress+Hauser11
Page 12
Status input
Maximum input values• DC –3 to 30 V
• If status input is active (ON): Ri >3 kΩ
Response timeConfigurable: 5 to 200 ms
Input signal level• Low signal: DC –3 to +5 V
• High signal: DC 12 to 30 V
Assignable functions• Off
• Reset the individual totalizers separately
• Reset all totalizers
• Flow override
Proline Promass A 300
12Endress+Hauser
Page 13
Proline Promass A 300
Output
Output and input variants
Depending on the option selected for output/input 1, different options are available for the other
outputs and inputs. Only one option can be selected for each output/input 1 to 3. The following
tables must be read vertically (↓).
Example: If the option BA "4–20 mA HART" was selected for output/input 1, one of the options A, B,
D, E, F, H, I or J is available for output 2 and one of the options A, B, D, E, F, H, I or J is available for
output 3.
Output/input 1 and options for output/input 2
Options for output/input 3
Order code for "Output; input 1" (020) →Possible options
Current output 4 to 20 mA HARTBA
Current output 4 to 20 mA HART Ex i passive↓ CA
Current output 4 to 20 mA HART Ex i active↓ CC
FOUNDATION Fieldbus↓ SA
FOUNDATION Fieldbus Ex i↓ TA
PROFIBUS DP↓ LA
PROFIBUS PA↓ GA
PROFIBUS PA Ex i↓ HA
Modbus RS485↓ MA
EtherNet/IP 2-port switch integrated↓ NA
PROFINET 2-port switch integrated↓ RA
Order code for "Output; input 2" (021) →↓↓↓↓↓↓↓↓↓↓↓
Not assignedAAAAAAAAAAA
Current output 4 to 20 mABBBBBBB
Current output 4 to 20 mA Ex i passiveCCCC
User-configurable input/output
Pulse/frequency/switch outputEEEEEEE
Double pulse output
Pulse/frequency/switch output Ex i passiveGGGG
Relay outputHHHHHHH
Current input 0/4 to 20 mAIIIIIII
Status inputJJJJJJJ
2)
1)
DDDDDDD
FF
1)A specific input or output can be assigned to a user-configurable input/output → 20.
2)If double pulse output (F) is selected for output/input 2 (021), only the double pulse output (F) option is available for selection for output/input 3
(022).
Endress+Hauser13
Page 14
Proline Promass A 300
Output/input 1 and options for output/input 3
Options for output/input 2 → 13
Order code for "Output; input 1" (020) →Possible options
Current output 4 to 20 mA HARTBA
Current output 4 to 20 mA HART Ex i passive↓ CA
Current output 4 to 20 mA HART Ex i active↓ CC
FOUNDATION Fieldbus↓ SA
FOUNDATION Fieldbus Ex i↓ TA
PROFIBUS DP↓ LA
PROFIBUS PA↓ GA
PROFIBUS PA Ex i↓ HA
Modbus RS485↓ MA
EtherNet/IP 2-port switch integrated↓ NA
PROFINET 2-port switch integrated↓ RA
Order code for "Output; input 3" (022) →↓↓↓↓↓↓↓↓↓↓↓
Not assignedAAAAAAAAAAA
Current output 4 to 20 mABBBBB
Current output 4 to 20 mA Ex i passiveCC
User-configurable input/outputDDDDD
Pulse/frequency/switch outputEEEEE
Double pulse output (slave)
Pulse/frequency/switch output Ex i passiveGG
Relay outputHHHHH
Current input 0/4 to 20 mAIIIII
Status inputJJJJJ
1)
FF
1)If double pulse output (F) is selected for output/input 2 (021), only the double pulse output (F) option is available for output/input 3 (022).
14Endress+Hauser
Page 15
Proline Promass A 300
Output signalCurrent output 4 to 20 mA HART
Order code"Output; Input 1" (20):
Option BA: current output 4 to 20 mA HART
Signal modeCan be set to:
• Active
• Passive
Current rangeCan be set to:
• 4 to 20 mA NAMUR
• 4 to 20 mA US
• 4 to 20 mA
• 0 to 20 mA (only with signal mode active)
• Fixed current value
Open-circuit voltageDC 28.8 V (active)
Maximum input voltageDC 30 V (passive)
Load250 to 700 Ω
Resolution0.38 µA
DampingConfigurable: 0 to 999 s
Assignable measured
variables
• Mass flow
• Volume flow
• Corrected volume flow
• Density
• Reference density
• Temperature
• Electronics temperature
• Oscillation frequency 0
• Oscillation damping 0
• Signal asymmetry
• Exciter current 0
The range of options increases if the measuring device has one or more
application packages.
Current output 4 to 20 mA HART Ex i
Order code"Output; Input 1" (20) can be set to:
• Option CA: current output 4 to 20 mA HART Ex i passive
• Option CC: current output 4 to 20 mA HART Ex i active
Signal modeDepending on the ordered variant.
Current rangeCan be set to:
• 4 to 20 mA NAMUR
• 4 to 20 mA US
• 4 to 20 mA
• 0 to 20 mA (only with signal mode active)
• Fixed current value
Open-circuit voltageDC 21.8 V (active)
Maximum input voltageDC 30 V (passive)
Load• 250 to 400 Ω (active)
• 250 to 700 Ω (passive)
Resolution0.38 µA
Endress+Hauser15
Page 16
DampingConfigurable: 0 to 999 s
Assignable measured
variables
• Mass flow
• Volume flow
• Corrected volume flow
• Density
• Reference density
• Temperature
• Electronics temperature
• Oscillation frequency 0
• Oscillation damping 0
• Signal asymmetry
• Exciter current 0
The range of options increases if the measuring device has one or more
application packages.
FOUNDATION Fieldbus
FOUNDATION FieldbusH1, IEC 61158-2, galvanically isolated
Data transfer31.25 kbit/s
Current consumption10 mA
Permitted supply voltage9 to 32 V
Bus connectionWith integrated reverse polarity protection
Proline Promass A 300
PROFIBUS DP
Signal encodingNRZ code
Data transfer9.6 kBaud…12 MBaud
PROFIBUS PA
PROFIBUS PAIn accordance with EN 50170 Volume 2, IEC 61158-2 (MBP), galvanically
isolated
Data transmission31.25 kbit/s
Current consumption10 mA
Permitted supply voltage9 to 32 V
Bus connectionWith integrated reverse polarity protection
Modbus RS485
Physical interfaceRS485 in accordance with EIA/TIA-485 standard
Terminating resistorIntegrated, can be activated via DIP switches
EtherNet/IP
StandardsIn accordance with IEEE 802.3
PROFINET
StandardsIn accordance with IEEE 802.3
16Endress+Hauser
Page 17
Proline Promass A 300
Current output 4 to 20 mA
Order code"Output; Input 2" (21), "Output; Input 3" (022):
Option B: current output 4 to 20 mA
Signal modeCan be set to:
• Active
• Passive
Current rangeCan be set to:
• 4 to 20 mA NAMUR
• 4 to 20 mA US
• 4 to 20 mA
• 0 to 20 mA (only with signal mode active)
• Fixed current value
Maximum output values22.5 mA
Open-circuit voltageDC 28.8 V (active)
Maximum input voltageDC 30 V (passive)
Load0 to 700 Ω
Resolution0.38 µA
DampingConfigurable: 0 to 999 s
Assignable measured
variables
• Mass flow
• Volume flow
• Corrected volume flow
• Density
• Reference density
• Temperature
• Electronics temperature
• Oscillation frequency 0
• Oscillation damping 0
• Signal asymmetry
• Exciter current 0
The range of options increases if the measuring device has one or more
application packages.
Current output 4 to 20 mA Ex i passive
Order code"Output; Input 2" (21), "Output; Input 3" (022):
Option C: current output 4 to 20 mA Ex i passive
Signal modePassive
Current rangeCan be set to:
• 4 to 20 mA NAMUR
• 4 to 20 mA US
• 4 to 20 mA
• Fixed current value
Maximum output values22.5 mA
Maximum input voltageDC 30 V
Load0 to 700 Ω
Resolution0.38 µA
Endress+Hauser17
Page 18
DampingConfigurable: 0 to 999 s
Assignable measured
variables
• Mass flow
• Volume flow
• Corrected volume flow
• Density
• Reference density
• Temperature
• Electronics temperature
• Oscillation frequency 0
• Oscillation damping 0
• Signal asymmetry
• Exciter current 0
The range of options increases if the measuring device has one or more
application packages.
Pulse/frequency/switch output
FunctionCan be set to pulse, frequency or switch output
VersionOpen collector
Can be set to:
• Active
• Passive
• Passive NAMUR
Ex-i, passive
Proline Promass A 300
Maximum input valuesDC 30 V, 250 mA (passive)
Open-circuit voltageDC 28.8 V (active)
Voltage dropFor 22.5 mA: ≤ DC 2 V
Pulse output
Maximum input valuesDC 30 V, 250 mA (passive)
Maximum output current22.5 mA (active)
Open-circuit voltageDC 28.8 V (active)
Pulse widthConfigurable: 0.05 to 2000 ms
Maximum pulse rate10000 Impulse/s
Pulse valueAdjustable
Assignable measured
variables
Frequency output
Maximum input valuesDC 30 V, 250 mA (passive)
Maximum output current22.5 mA (active)
Open-circuit voltageDC 28.8 V (active)
Output frequencyAdjustable: end value frequency 2 to 10000 Hz (f
DampingConfigurable: 0 to 999 s
Pulse/pause ratio1:1
• Mass flow
• Volume flow
• Corrected volume flow
= 12500 Hz)
max
18Endress+Hauser
Page 19
Proline Promass A 300
Assignable measured
variables
Switch output
Maximum input valuesDC 30 V, 250 mA (passive)
Open-circuit voltageDC 28.8 V (active)
Switching behaviorBinary, conductive or non-conductive
Switching delayConfigurable: 0 to 100 s
Number of switching
cycles
Assignable functions• Off
• Mass flow
• Volume flow
• Corrected volume flow
• Density
• Reference density
• Temperature
• Electronics temperature
• Oscillation frequency 0
• Oscillation damping 0
• Signal asymmetry
• Exciter current 0
The range of options increases if the measuring device has one or more
application packages.
Unlimited
• On
• Diagnostic behavior
• Limit value
• Mass flow
• Volume flow
• Corrected volume flow
• Density
• Reference density
• Temperature
• Totalizer 1-3
• Flow direction monitoring
• Status
• Partially filled pipe detection
• Low flow cut off
The range of options increases if the measuring device has one or more
application packages.
Double pulse output
FunctionDouble pulse
VersionOpen collector
Can be set to:
• Active
• Passive
• Passive NAMUR
Maximum input valuesDC 30 V, 250 mA (passive)
Open-circuit voltageDC 28.8 V (active)
Voltage dropFor 22.5 mA: ≤ DC 2 V
Output frequencyConfigurable: 0 to 1000 Hz
DampingConfigurable: 0 to 999 s
Endress+Hauser19
Page 20
Pulse/pause ratio1:1
Assignable measured
variables
• Mass flow
• Volume flow
• Corrected volume flow
• Density
• Reference density
• Temperature
The range of options increases if the measuring device has one or more
application packages.
Relay output
FunctionSwitch output
VersionRelay output, galvanically isolated
Switching behaviorCan be set to:
• NO (normally open), factory setting
• NC (normally closed)
Maximum switching
capacity (passive)
Assignable functions• Off
• DC 30 V, 0.1 A
• AC 30 V, 0.5 A
• On
• Diagnostic behavior
• Limit value
• Mass flow
• Volume flow
• Corrected volume flow
• Density
• Reference density
• Temperature
• Totalizer 1-3
• Flow direction monitoring
• Status
• Partially filled pipe detection
• Low flow cut off
The range of options increases if the measuring device has one or more
application packages.
Proline Promass A 300
User-configurable input/output
One specific input or output is assigned to a user-configurable input/output (configurable I/O)
during device commissioning.
The following inputs and outputs are available for assignment:
• Choice of current output: 4 to 20 mA (active), 0/4 to 20 mA (passive)
• Pulse/frequency/switch output
• Choice of current input: 4 to 20 mA (active), 0/4 to 20 mA (passive)
• Status input
The technical values correspond to those of the inputs and outputs described in this section.
Signal on alarm
Depending on the interface, failure information is displayed as follows:
HART current output
Device diagnosticsDevice condition can be read out via HART Command 48
20Endress+Hauser
Page 21
Proline Promass A 300
PROFIBUS PA
Status and alarm
messages
Failure current FDE (Fault
Disconnection Electronic)
Diagnostics in accordance with PROFIBUS PA Profile 3.02
0 mA
PROFIBUS DP
Status and alarm
messages
Diagnostics in accordance with PROFIBUS PA Profile 3.02
EtherNet/IP
Device diagnosticsDevice condition can be read out in Input Assembly
PROFINET
Device diagnosticsAccording to "Application Layer protocol for decentralized periphery", Version 2.3
FOUNDATION Fieldbus
Status and alarm
messages
Failure current FDE (Fault
Disconnection Electronic)
Diagnostics in accordance with FF-891
0 mA
Modbus RS485
Failure modeChoose from:
• NaN value instead of current value
• Last valid value
Current output 0/4 to 20 mA
4 to 20 mA
Failure modeChoose from:
• 4 to 20 mA in accordance with NAMUR recommendation NE 43
• 4 to 20 mA in accordance with US
• Min. value: 3.59 mA
• Max. value: 22.5 mA
• Freely definable value between: 3.59 to 22.5 mA
• Actual value
• Last valid value
0 to 20 mA
Failure modeChoose from:
• Maximum alarm: 22 mA
• Freely definable value between: 0 to 20.5 mA
Endress+Hauser21
Page 22
Pulse/frequency/switch output
Pulse output
Failure modeChoose from:
• Actual value
• No pulses
Frequency output
Failure modeChoose from:
• Actual value
• 0 Hz
• Defined value (f
Switch output
Failure modeChoose from:
• Current status
• Open
• Closed
Relay output
Failure modeChoose from:
• Current status
• Open
• Closed
2 to 12500 Hz)
max
Proline Promass A 300
Local display
Plain text displayWith information on cause and remedial measures
BacklightRed backlighting indicates a device error.
Status signal as per NAMUR recommendation NE 107
Interface/protocol
• Via digital communication:
• HART protocol
• FOUNDATION Fieldbus
• PROFIBUS PA
• PROFIBUS DP
• Modbus RS485
• EtherNet/IP
• PROFINET
• Via service interface
• CDI-RJ45 service interface
• WLAN interface
Plain text displayWith information on cause and remedial measures
Additional information on remote operation → 89
Web browser
Plain text displayWith information on cause and remedial measures
22Endress+Hauser
Page 23
Proline Promass A 300
Light emitting diodes (LED)
Status informationStatus indicated by various light emitting diodes
Ex connection dataSafety-related values
The following information is displayed depending on the device version:
• Supply voltage active
• Data transmission active
• Device alarm/error has occurred
• EtherNet/IP network available
• EtherNet/IP connection established
• PROFINET network available
• PROFINET connection established
• PROFINET blinking feature
Order code for
Output typeSafety-related values
"Output; input 1"
Option BACurrent output
4 to 20 mA HART
UN = 30 V
UM = 250 V
Option GAPROFIBUS PAUN = 30 V
UM = 250 V
Option LAPROFIBUS DPUN = 30 V
UM = 250 V
Option MAModbus RS485UN = 30 V
UM = 250 V
Option SAFOUNDATION FieldbusUN = 30 V
UM = 250 V
Option NAEtherNet/IPUN = 30 V
UM = 250 V
Option RAPROFINETUN = 30 V
UM = 250 V
Order code for
"Output; input 2";
"Output; input 3"
Output typeSafety-related values
Output; input 2Output; input 3
24 (+)25 (–)22 (+)23 (–)
Option BCurrent output
4 to 20 mA
Option DUser-configurable input/
output
Option EPulse/frequency/switch
output
UN = 30 V
UM = 250 V
UN = 30 V
UM = 250 V
UN = 30 V
UM = 250 V
Option FDouble pulse outputUN = 30 V
UM = 250 V
Option HRelay outputUN = 30 V
IN =100 mADC/500 mA
UM = 250 V
Option ICurrent input 4 to 20 mA UN = 30 V
UM = 250 V
Option JStatus inputUN = 30 V
UM = 250 V
"Output; input 1"
26 (+)27 (–)
DC
AC
DC
AC
DC
AC
DC
AC
DC
AC
DC
AC
DC
AC
DC
AC
DC
AC
DC
AC
DC
AC
DC
AC
AC
DC
AC
DC
AC
Endress+Hauser23
Page 24
Intrinsically safe values
Proline Promass A 300
Order code for
Output typeIntrinsically safe values
"Output; input 1"
Option CACurrent output
4 to 20 mA HART Ex i
passive
Option CCCurrent output
4 to 20 mA HART Ex i
active
Option HAPROFIBUS PA Ex i
(FISCO Field Device)
Option TAFOUNDATION Fieldbus
Ex i
"Output; input 1"
26 (+)27 (–)
Ui = 30 V
li = 100 mA
Pi = 1.25 W
Li = 0 µH
Ci = 6 nF
Ui = 32 V
li = 570 mA
Pi = 8.5 W
Li = 10 µH
Ci = 5 nF
4)
Ex ic
Ui = 32 V
li = 570 mA
Pi = 8.5 W
Li = 10 µH
Ci = 5 nF
Low flow cut off
Galvanic isolation
1)Only available for the Zone 1; Class I, Division 1 version
2)Only available for the Zone 2; Class I, Division 2 version transmitter
3)Only available for the Zone 1; Class I, Division 1 version
4)Only available for the Zone 2; Class I, Division 2 version transmitter
Order code for
"Output; input 2";
"Output; input 3"
Output typeIntrinsically safe values or NIFW values
Output; input 2Output; input 3
24 (+)25 (–)22 (+)23 (–)
Option CCurrent output
4 to 20 mA Ex i
Ui = 30 V
li = 100 mA
Pi = 1.25 W
Li = 0
Ci = 0
Option GPulse/frequency/switch
output Ex i
Ui = 30 V
li = 100 mA
Pi = 1.25 W
Li = 0
Ci = 0
The switch points for low flow cut off are user-selectable.
The outputs are galvanically isolated from one another and from earth (PE).
24Endress+Hauser
Page 25
Proline Promass A 300
Protocol-specific dataHART
Manufacturer ID0x11
Device type ID0x3B
HART protocol revision7
Device description files
(DTM, DD)
HART loadMin. 250 Ω
System integrationInformation on system integration: Operating Instructions → 109.
FOUNDATION Fieldbus
Manufacturer ID0x452B48 (hex)
Ident number0x103B (hex)
Device revision1
DD revisionInformation and files under:
CFF revision
Interoperability Test Kit (ITK) Version 6.2.0
ITK Test Campaign NumberInformation:
Link Master capability (LAS)Yes
Choice of "Link Master" and
"Basic Device"
Node addressFactory setting: 247 (0xF7)
Supported functionsThe following methods are supported:
Virtual Communication Relationships (VCRs)
Number of VCRs44
Number of link objects in VFD 50
Permanent entries1
Client VCRs0
Server VCRs10
Source VCRs43
Sink VCRs0
Subscriber VCRs43
Publisher VCRs43
Device Link Capabilities
Slot time4
Min. delay between PDU8
Information and files under:
www.endress.com
• Measured variables via HART protocol
• Burst Mode functionality
• www.endress.com
• www.fieldbus.org
• www.endress.com
• www.fieldbus.org
Yes
Factory setting: Basic Device
• Restart
• ENP Restart
• Diagnostic
• Set to OOS
• Set to AUTO
• Read trend data
• Read event logbook
Endress+Hauser25
Page 26
Proline Promass A 300
Max. response delay16
System integrationInformation regarding system integration: Operating Instructions → 109.
• Cyclic data transmission
• Description of the modules
• Execution times
• Methods
PROFIBUS DP
Manufacturer ID0x11
Ident number0x156F
Profile version3.02
Device description files (GSD,
DTM, DD)
Supported functions• Identification & Maintenance
Configuration of the device
address
Compatibility with
earlier model
System integrationInformation regarding system integration: Operating Instructions → 109.
Information and files under:
• www.endress.com
On the product page for the device: Documents/Software → Device drivers
• www.profibus.org
Simplest device identification on the part of the control system and
nameplate
• PROFIBUS upload/download
Reading and writing parameters is up to ten times faster with PROFIBUS
upload/download
• Condensed status
Simplest and self-explanatory diagnostic information by categorizing
diagnostic messages that occur
• DIP switches on the I/O electronics module
• Via operating tools (e.g. FieldCare)
If the device is replaced, the measuring device Promass 300 supports the
compatibility of the cyclic data with previous models. It is not necessary to
adjust the engineering parameters of the PROFIBUS network with the
Promass 300 GSD file.
Previous model:
Promass 83 PROFIBUS DP
• ID No.: 1529 (hex)
• Extended GSD file: EH3x1529.gsd
• Standard GSD file: EH3_1529.gsd
Description of the function scope of compatibility:
Operating Instructions → 109.
• Cyclic data transmission
• Block model
• Description of the modules
PROFIBUS PA
Manufacturer ID0x11
Ident number0x156D
Profile version3.02
Device description files (GSD,
DTM, DD)
Information and files under:
• www.endress.com
• www.profibus.org
26Endress+Hauser
Page 27
Proline Promass A 300
Supported functions• Identification & Maintenance
Simplest device identification on the part of the control system and
nameplate
• PROFIBUS upload/download
Reading and writing parameters is up to ten times faster with PROFIBUS
upload/download
• Condensed status
Simplest and self-explanatory diagnostic information by categorizing
diagnostic messages that occur
Configuration of the device
address
Compatibility with
earlier model
System integrationInformation regarding system integration: Operating Instructions → 109.
• DIP switches on the I/O electronics module
• Local display
• Via operating tools (e.g. FieldCare)
If the device is replaced, the measuring device Promass 300 supports the
compatibility of the cyclic data with previous models. It is not necessary to
adjust the engineering parameters of the PROFIBUS network with the
Promass 300 GSD file.
Earlier models:
• Promass 80 PROFIBUS PA
• ID No.: 1528 (hex)
• Extended GSD file: EH3x1528.gsd
• Standard GSD file: EH3_1528.gsd
• Promass 83 PROFIBUS PA
• ID No.: 152A (hex)
• Extended GSD file: EH3x152A.gsd
• Standard GSD file: EH3_152A.gsd
Description of the function scope of compatibility:
Response times• Direct data access: typically 25 to 50 ms
• Auto-scan buffer (data range): typically 3 to 5 ms
Device typeSlave
Slave address range1 to 247
Broadcast address range0
Function codes• 03: Read holding register
• 04: Read input register
• 06: Write single registers
• 08: Diagnostics
• 16: Write multiple registers
• 23: Read/write multiple registers
Broadcast messagesSupported by the following function codes:
• 06: Write single registers
• 16: Write multiple registers
• 23: Read/write multiple registers
Supported baud rate• 1200 BAUD
• 2400 BAUD
• 4800 BAUD
• 9600 BAUD
• 19200 BAUD
• 38400 BAUD
• 57600 BAUD
• 115200 BAUD
Data transfer mode• ASCII
• RTU
Endress+Hauser27
Page 28
Proline Promass A 300
Data accessEach device parameter can be accessed via Modbus RS485.
For Modbus register information
Compatibility with
earlier model
System integrationInformation on system integration: Operating Instructions → 109.
If the device is replaced, the measuring device Promass 300 supports the
compatibility of the Modbus registers for the process variables and the
diagnostic information with the previous model Promass 83. It is not
necessary to change the engineering parameters in the automation system.
Description of the function scope of compatibility:
Operating Instructions → 109.
• Modbus RS485 information
• Function codes
• Register information
• Response time
• Modbus data map
EtherNet/IP
Protocol• The CIP Networks Library Volume 1: Common Industrial Protocol
• The CIP Networks Library Volume 2: EtherNet/IP Adaptation of CIP
Communication type• 10Base-T
• 100Base-TX
Device profileGeneric device (product type: 0x2B)
Manufacturer ID0x11
Device type ID0x103B
Baud ratesAutomatic ¹⁰⁄₁₀₀ Mbit with half-duplex and full-duplex detection
PolarityAuto-polarity for automatic correction of crossed TxD and RxD pairs
Supported CIP connectionsMax. 3 connections
Explicit connectionsMax. 6 connections
I/O connectionsMax. 6 connections (scanner)
Configuration options for
measuring device
Configuration of the EtherNet
interface
Configuration of the device
address
Device Level Ring (DLR)Yes
System integrationInformation regarding system integration: Operating Instructions
• DIP switches on the electronics module for IP addressing
• Manufacturer-specific software (FieldCare)
• Add-on Profile Level 3 for Rockwell Automation control systems
• Web browser
• Electronic Data Sheet (EDS) integrated in the measuring device
• Speed: 10 MBit, 100 MBit, auto (factory setting)
• Duplex: half-duplex, full-duplex, auto (factory setting)
• DIP switches on the electronics module for IP addressing (last octet)
• DHCP
• Manufacturer-specific software (FieldCare)
• Add-on Profile Level 3 for Rockwell Automation control systems
• Web browser
• EtherNet/IP tools, e.g. RSLinx (Rockwell Automation)
→ 109.
• Cyclic data transmission
• Block model
• Input and output groups
PROFINET
ProtocolApplication layer protocol for decentral device periphery and distributed
automation, Version 2.3
Communication type100 MBit/s
28Endress+Hauser
Page 29
Proline Promass A 300
Conformity classConformance Class B
Netload ClassNetload Class II
Baud ratesAutomatic 100 Mbit/s with full-duplex detection
Cycle timesFrom 8 ms
PolarityAuto-polarity for automatic correction of crossed TxD and RxD pairs
Media Redundancy Protocol
(MRP)
System redundancy supportSystem redundancy S2 (2 AR with 1 NAP)
Terminal assignment of the remote display and operating module → 33.
Device plugs available
Device plugs may not be used in hazardous areas!
30Endress+Hauser
Page 31
Proline Promass A 300
Device plugs for fieldbus systems:
Order code for "Input; output 1"
• Option SA "FOUNDATION Fieldbus" → 31
• Option GA "PROFIBUS PA" → 31
• Option NA "EtherNet/IP" → 31
• Option RA "PROFINET" → 31
Device plug for connecting to the service interface:
Order code for "Accessory mounted"
option NB, adapter RJ45 M12 (service interface) → 43
Order code for "Input; output 1", option SA "FOUNDATION Fieldbus"
Order code forCable entry/connection → 32
"Electrical connection"23
M, 3, 4, 57/8" connector–
Order code for "Input; output 1", option GA "PROFIBUS PA"
Order code forCable entry/connection → 32
"Electrical connection"23
L, N, P, UConnector M12 × 1–
Order code for "Input; output 1", option NA "EtherNet/IP"
Order code forCable entry/connection → 32
"Electrical connection"23
L, N, P, UConnector M12 × 1–
1) 2)
1) 2)
1) 2)
R
, S
, T
, V
1) 2)
Connector M12 × 1Connector M12 × 1
1)Cannot be combined with an external WLAN antenna (order code for "Enclosed accessories", option P8) of
an RJ45 M12 adapter for the service interface (order code for "Accessories mounted", option NB) or of the
remote display and operating module DKX001
2)Suitable for integrating the device in a ring topology.
Order code for "Input; output 1", option RA "PROFINET"
Order code forCable entry/connection → 32
"Electrical connection"23
L, N, P, UConnector M12 × 1–
1) 2)
1) 2)
1) 2)
R
, S
, T
1)Cannot be combined with an external WLAN antenna (order code for "Enclosed accessories", option P8) of
an RJ45 M12 adapter for the service interface (order code for "Accessories mounted", option NB) or of the
remote display and operating module DKX001.
2)Suitable for integrating the device in a ring topology.
, V
1) 2)
Connector M12 × 1Connector M12 × 1
Order code for "Accessory mounted", option NB "Adapter RJ45 M12 (service interface)"
Order codeCable entry/coupling → 32
"Accessory mounted"Cable entry
2
Cable entry
3
NBPlug M12 × 1–
Endress+Hauser31
Page 32
Proline Promass A 300
1
2
3
4
Supply voltage
Order code for
"Power supply"
Option DDC24 V±20%–
Option EAC100 to 240 V –15…+10%50/60 Hz
Option I
Power consumptionTransmitter
Max. 10 W (active power)
switch-on currentMax. 36 A (<5 ms) as per NAMUR Recommendation NE 21
Current consumptionTransmitter
• Max. 400 mA (24 V)
• Max. 200 mA (110 V, 50/60 Hz; 230 V, 50/60 Hz)
Power supply failure
• Totalizers stop at the last value measured.
• Depending on the device version, the configuration is retained in the device memoryor in the
pluggable data memory (HistoROM DAT).
• Error messages (incl. total operated hours) are stored.
Electrical connectionConnecting the transmitter
• Terminal assignment→ 30
• Device plugs available→ 30
Terminal voltageFrequency range
DC24 V±20%–
AC100 to 240 V –15…+10%50/60 Hz
1Terminal connection for supply voltage
2Terminal connection for signal transmission, input/output
3Terminal connection for signal transmission, input/output or terminal for network connection via service
interface (CDI-RJ45); Optional: terminal connection for external WLAN antenna or connection for remote
display and operating module DKX001
4Protective ground (PE)
An adapter for RJ45 and the M12 connector is optionally available:
Order code for "Accessories", option NB: "Adapter RJ45 M12 (service interface)"
The adapter connects the service interface (CDI-RJ45) to an M12 connector mounted in the
cable entry. Therefore the connection to the service interface can be established via an M12
connector without opening the device.
Network connection via service interface (CDI-RJ45) → 95
Connecting in a ring topology
Device versions with EtherNet/IP and PROFINET communication protocols can be integrated into a
ring topology. The device is integrated via the terminal connection for signal transmission (output 1)
and the connection to the service interface (CDI-RJ45).
Integrate the transmitter into a ring topology:
• EtherNet/IP
• PROFINET
32Endress+Hauser
A0026781
Page 33
Proline Promass A 300
1
2
3
4
1
3
81
Vcc
82
Gnd
83A84
B
81
Vcc
82
Gnd
83A84
B
2
4
5
A0026781
1Terminal connection for supply voltage
2Terminal connection for signal transmission: PROFINET or EtherNet/IP (RJ45 connector)
3Connection to service interface (CDI-RJ45)
4Protective ground (PE)
If the device has additional input/outputs, these are routed via the cable entry for the
connection to the service interface (CDI-RJ45).
Connecting the remote display and operating module DKX001
The remote display and operating module DKX001 is available as an optional extra → 105.
• The remote display and operating module DKX001 is only available for the following housing
version: order code for "Housing": option A "Aluminum, coated"
• The measuring device is always supplied with a dummy cover when the remote display and
operating module DKX001 is ordered directly with the measuring device. Display or operation
at the transmitter is not possible in this case.
• If ordered subsequently, the remote display and operating module DKX001 may not be
connected at the same time as the existing measuring device display module. Only one
display or operation unit may be connected to the transmitter at any one time.
2Connection example for 4 to 20 mA HART current output (active)
1Automation system with current input (e.g. PLC)
2Cable shield provided at one end. The cable shield must be grounded at both ends to comply with EMC
requirements; observe cable specifications → 43
3Connection for HART operating devices → 89
4Resistor for HART communication (≥ 250 Ω): observe maximum load → 15
5Analog display unit: observe maximum load → 15
6Transmitter
A0029055
3Connection example for 4 to 20 mA HART current output (passive)
1Automation system with current input (e.g. PLC)
2Power supply
3Cable shield provided at one end. The cable shield must be grounded at both ends to comply with EMC
4Connection example for HART input with a common negative (passive)
1Automation system with HART output (e.g. PLC)
2Active barrier for power supply (e.g. RN221N)
3Cable shield provided at one end. The cable shield must be grounded at both ends to comply with EMC
requirements; observe cable specifications
4Analog display unit: observe maximum load → 15
5Pressure measuring device (e.g. Cerabar M, Cerabar S): see requirements
6Transmitter
PROFIBUS PA
5Connection example for PROFIBUS PA
1Control system (e.g. PLC)
2PROFIBUS PA segment coupler
3Cable shield provided at one end. The cable shield must be grounded at both ends to comply with EMC
4T-box
5Measuring device
6Local grounding
7Bus terminator
8Potential matching line
requirements; observe cable specifications
A0028768
Endress+Hauser35
Page 36
Proline Promass A 300
2
1
A
B
3
4
4
A
B
A
B
1
2
4
3
5
5
PROFIBUS DP
6Connection example for PROFIBUS DP, non-hazardous area and Zone 2/Div. 2
1Control system (e.g. PLC)
2Cable shield provided at one end. The cable shield must be grounded at both ends to comply with EMC
1Control system (e.g. PLC)
2Ethernet switch
3Observe cable specifications → 44
4Transmitter
5Connecting cable between the two transmitters
A0028767
A0027544
Endress+Hauser37
Page 38
PROFINET: system redundancy S2
1
2
555
4
3
2134
78
66
6
6
5
6
6
5
9Connection example for system redundancy S2
1Control system 1 (e.g. PLC)
2Synchronization of control systems
3Control system 2 (e.g. PLC)
4Industrial Ethernet Managed Switch
5Transmitter
Proline Promass A 300
A0039553
FOUNDATION Fieldbus
A0028768
10Connection example for FOUNDATION Fieldbus
1Control system (e.g. PLC)
2Power Conditioner (FOUNDATION Fieldbus)
3Cable shield provided at one end. The cable shield must be grounded at both ends to comply with EMC
12Connection example for 4-20 mA current output (active)
1Automation system with current input (e.g. PLC)
2Analog display unit: observe maximum load → 15
3Transmitter
A0028759
13Connection example for 4-20 mA current output (passive)
1Automation system with current input (e.g. PLC)
2Active barrier for power supply (e.g. RN221N)
3Analog display unit: observe maximum load → 15
4Transmitter
Endress+Hauser39
Page 40
Pulse/frequency output
1
2
3
12345
1
2
3
1
2
3
4
14Connection example for pulse/frequency output (passive)
1Automation system with pulse/frequency input (e.g. PLC)
2Power supply
3Transmitter: Observe input values → 18
Switch output
Proline Promass A 300
A0028761
15Connection example for switch output (passive)
1Automation system with switch input (e.g. PLC)
2Power supply
3Transmitter: Observe input values → 18
Double pulse output
16Connection example for double pulse output (active)
1Automation system with relay input (e.g. PLC)
2Power supply
3Transmitter: Observe input values → 20
Current input
A0028760
A0028915
19Connection example for 4 to 20 mA current input
1Power supply
2Terminal box
3External measuring device (to read in pressure or temperature, for instance)
4Transmitter
Endress+Hauser41
Page 42
Status input
1
2
3
1
2
4
3
1
2
4
3
20Connection example for status input
1Automation system with status output (e.g. PLC)
2Power supply
3Transmitter
Potential equalizationRequirements
No special measures for potential equalization are required.
Please consider the following to ensure correct measurement:
• Same electrical potential for the fluid and sensor
• Company-internal grounding concepts
Proline Promass A 300
A0028764
terminals
Spring-loaded terminals: Suitable for strands and strands with ferrules.
Conductor cross-section 0.2 to 2.5 mm2 (24 to 12 AWG).
Cable entries
• Cable gland: M20 × 1.5 with cable ⌀ 6 to 12 mm (0.24 to 0.47 in)
• Thread for cable entry:
• NPT ½"
• G ½"
• M20
• Device plug for digital communication: M12
Only available for certain device versions → 30.
Pin assignment, device plugFOUNDATION Fieldbus
PROFIBUS PA
PinAssignmentCodingPlug/socket
1+Signal +APlug
2-Signal –
3Grounding
4Not assigned
PinAssignmentCodingPlug/socket
1+PROFIBUS PA +APlug
2Grounding
3-PROFIBUS PA –
4Not assigned
42Endress+Hauser
Page 43
Proline Promass A 300
3
2
4
3
2
4
3
2
4
PROFINET
PinAssignment
1+TD +
2+RD +
3-TD –
4-RD –
A0032047
Recommended plug:
• Binder, series 763, part no. 99 3729 810 04
• Phoenix, part no. 1543223 SACC-M12MSD-4Q
CodingPlug/socket
DSocket
EtherNet/IP
PinAssignment
1+Tx
2+Rx
3-Tx
4-Rx
A0032047
CodingPlug/socket
DSocket
Recommended plug:
• Binder, series 763, part no. 99 3729 810 04
• Phoenix, part no. 1543223 SACC-M12MSD-4Q
Service interface
Order code for "Accessories mounted", option NB: Adapter RJ45 M12 (service interface)
PinAssignment
1+Tx
2+Rx
3-Tx
4-Rx
A0032047
Recommended plug:
• Binder, series 763, part no. 99 3729 810 04
• Phoenix, part no. 1543223 SACC-M12MSD-4Q
CodingPlug/socket
Cable specificationPermitted temperature range
• The installation guidelines that apply in the country of installation must be observed.
• The cables must be suitable for the minimum and maximum temperatures to be expected.
DSocket
Power supply cable
Standard installation cable is sufficient.
Endress+Hauser43
Page 44
Proline Promass A 300
Signal cable
Current output 4 to 20 mA HART
A shielded cable is recommended. Observe grounding concept of the plant.
PROFIBUS PA
Twisted, shielded two-wire cable. Cable type A is recommended .
For further information on planning and installing PROFIBUS networks see:
• Operating Instructions "PROFIBUS DP/PA: Guidelines for planning and commissioning"
(BA00034S)
• PNO Directive 2.092 "PROFIBUS PA User and Installation Guideline"
• IEC 61158-2 (MBP)
PROFIBUS DP
The IEC 61158 standard specifies two types of cable (A and B) for the bus line which can be used for
every transmission rate. Cable type A is recommended.
Cable typeA
Characteristic impedance135 to 165 Ω at a measuring frequency of 3 to 20 MHz
Cable capacitance< 30 pF/m
Wire cross-section
Cable typeTwisted pairs
Loop resistance≤110 Ω/km
Signal dampingMax. 9 dB over the entire length of the cable cross-section
ShieldCopper braided shielding or braided shielding with foil shield. When grounding
> 0.34 mm2 (22 AWG)
the cable shield, observe the grounding concept of the plant.
For further information on planning and installing PROFIBUS networks see:
• Operating Instructions "PROFIBUS DP/PA: Guidelines for planning and commissioning"
(BA00034S)
• PNO Directive 2.092 "PROFIBUS PA User and Installation Guideline"
• IEC 61158-2 (MBP)
EtherNet/IP
The standard ANSI/TIA/EIA-568-B.2 Annex specifies CAT 5 as the minimum category for a cable
used for EtherNet/IP. CAT 5e and CAT 6 are recommended.
For more information on planning and installing EtherNet/IP networks, please refer to the
"Media Planning and Installation Manual. EtherNet/IP" of ODVA Organization
PROFINET
Standard IEC 61156-6 specifies CAT 5 as the minimum category for a cable used for PROFINET. CAT
5e and CAT 6 are recommended.
For more information on planning and installing PROFINET networks, see: "PROFINET Cabling
and Interconnection Technology", Guideline for PROFINET
FOUNDATION Fieldbus
Twisted, shielded two-wire cable.
For further information on planning and installing FOUNDATION Fieldbus networks see:
• Operating Instructions for "FOUNDATION Fieldbus Overview" (BA00013S)
• FOUNDATION Fieldbus Guideline
• IEC 61158-2 (MBP)
44Endress+Hauser
Page 45
Proline Promass A 300
Modbus RS485
The EIA/TIA-485 standard specifies two types of cable (A and B) for the bus line which can be used
for every transmission rate. Cable type A is recommended.
Cable typeA
Characteristic impedance135 to 165 Ω at a measuring frequency of 3 to 20 MHz
Cable capacitance< 30 pF/m
Wire cross-section
Cable typeTwisted pairs
Loop resistance≤110 Ω/km
Signal dampingMax. 9 dB over the entire length of the cable cross-section
ShieldCopper braided shielding or braided shielding with foil shield. When grounding
> 0.34 mm2 (22 AWG)
the cable shield, observe the grounding concept of the plant.
Current output 0/4 to 20 mA
Standard installation cable is sufficient.
Pulse/frequency/switch output
Standard installation cable is sufficient.
Double pulse output
Standard installation cable is sufficient.
Relay output
Standard installation cable is sufficient.
Current input 0/4 to 20 mA
Standard installation cable is sufficient.
Status input
Standard installation cable is sufficient.
Connecting cable for transmitter - remote display and operating module DKX001
Standard cable
A standard cable can be used as the connecting cable.
Standard cable4 cores (2 pairs); pair-stranded with common shield
Operating temperatureWhen mounted in a fixed position: –50 to +105 °C (–58 to +221 °F); when cable
1)UV radiation can impair the cable outer sheath. Protect the cable from direct sunshine where possible.
2 × 2 × 0.34 mm2 (22 AWG) PVC cable
stranded)
can move freely: –25 to +105 °C (–13 to +221 °F)
1)
with common shield (2 pairs, pair-
Performance characteristics
• Error limits based on ISO 11631
• Water with +15 to +45 °C (+59 to +113 °F) at2 to 6 bar (29 to 87 psi)
• Specifications as per calibration protocol
• Accuracy based on accredited calibration rigs that are traced to ISO 17025.
To obtain measured errors, use the Applicator sizing tool → 107
Maximum measured error
o.r. = of reading; 1 g/cm³ = 1 kg/l; T = medium temperature
Base accuracy
Design fundamentals → 49
Mass flow and volume flow (liquids)
±0.10 % o.r.
Mass flow (gases)
±0.35 % o.r.
46Endress+Hauser
Page 47
Proline Promass A 300
Density (liquids)
Under
reference operating conditions
[g/cm³][g/cm³][g/cm³]
±0.0005±0.02±0.002
1)Valid over the entire temperature and density range
2)Valid range for special density calibration: 0 to 2 g/cm³, +5 to +80 °C (+41 to +176 °F)
3)Order code for "Application package", option EE "Special density" only in combination with the order code
for "Measuring tube mat., wetted surface", option BB, BF, HA, SA
Standard density
calibration
1)
Wide-range
Density specification
2) 3)
Temperature
±0.5 °C ± 0.005 · T °C (±0.9 °F ± 0.003 · (T – 32) °F)
Zero point stability
Standard version: order code for "Measuring tube mat., wetted surface", option BB, BF, HA, SA
DNZero point stability
[mm][in][kg/h][lb/min]
1¹⁄₂₄0.00050.000018
2¹⁄₁₂0.00250.00009
4¹⁄₈0.01000.00036
High-pressure version: order code for "Measuring tube mat., wetted surface", option HB
DNZero point stability
[mm][in][kg/h][lb/min]
1¹⁄₂₄0.00080.0000288
2¹⁄₁₂0.00400.000144
4¹⁄₈0.01600.000576
Flow values
Flow values as turndown parameter depending on nominal diameter.
The outputs have the following base accuracy specifications.
Current output
Accuracy±5 µA
Pulse/frequency output
o.r. = of reading
Repeatability
Response time
Influence of ambient
temperature
AccuracyMax. ±50 ppm o.r. (over the entire ambient temperature range)
o.r. = of reading; 1 g/cm3 = 1 kg/l; T = medium temperature
Base repeatability
Design fundamentals → 49
Mass flow and volume flow (liquids)
±0.05 % o.r.
Mass flow (gases)
±0.15 % o.r.
Density (liquids)
±0.00025 g/cm
3
Temperature
±0.25 °C ± 0.0025 · T °C (±0.45 °F ± 0.0015 · (T–32) °F)
The response time depends on the configuration (damping).
Current output
Temperature coefficientMax. 1 μA/°C
Pulse/frequency output
Temperature coefficientNo additional effect. Included in accuracy.
Influence of medium
temperature
Mass flow and volume flow
o.f.s. = of full scale value
When there is a difference between the temperature for zero point adjustment and the process
temperature, the additional measured error of the sensor is typically ±0.0002 % o.f.s./°C (±0.0001 %
o. f.s./°F).
48Endress+Hauser
Page 49
Proline Promass A 300
2
1
3
[°C]
-40 0-8040 80 120 160 200 240 280 320 360 400
[°F]
-50050100150200
0
2
4
6
8
10
ZeroPoint
BaseAccu
⋅ 100
³
± BaseAccu
ZeroPoint
BaseAccu
⋅ 100
<
ZeroPoint
MeasValue
⋅ 100
±
½ ⋅ ZeroPoint
BaseRepeat
⋅ 100
³
½ ⋅ ZeroPoint
BaseRepeat
⋅ 100
<
ZeroPoint
MeasValue
⋅ 100
½ ⋅±
The effect is reduced if zero point adjustment is performed at process temperature.
Density
When there is a difference between the density calibration temperature and the process
temperature, the typical measured error of the sensor is
±0.00005 g/cm3 /°C (±0.000025 g/cm3 /°F). Field density calibration is possible.
Wide-range density specification (special density calibration)
If the process temperature is outside the valid range (→ 46) the measured error is
±0.00005 g/cm3 /°C (±0.000025 g/cm3 /°F)
Influence of medium
pressure
Design fundamentals
1Field density calibration, for example at +20 °C (+68 °F)
2Special density calibration
A difference between the calibration pressure and process pressure does not affect accuracy.
o.r. = of reading, o.f.s. = of full scale value
BaseAccu = base accuracy in % o.r., BaseRepeat = base repeatability in % o.r.
MeasValue = measured value; ZeroPoint = zero point stability
Calculation of the maximum measured error as a function of the flow rate
Flow rateMaximum measured error in % o.r.
A0021332
A0021333
Calculation of the maximum repeatability as a function of the flow rate
Flow rateMaximum repeatability in % o.r.
A0016616
A0021339
A0021334
A0021335
A0021336
Endress+Hauser49
A0021340
A0021337
Page 50
Proline Promass A 300
0102030405060
0
0.5
1.0
1.5
2.0
2.5
708090100
Q [%]
E [%]
Example for maximum measured error
EMaximum measured error in % o.r. (example)
QFlow rate in % of maximum full scale value
Installation
No special measures such as supports etc. are necessary. External forces are absorbed by the
construction of the device.
Use a sensor holder for additional stability in the process pipe and for non-flanged process
connections. → 52
A0030378
Mounting location
A0028772
To prevent measuring errors arising from accumulation of gas bubbles in the measuring tube, avoid
the following mounting locations in the pipe:
• Highest point of a pipeline.
• Directly upstream of a free pipe outlet in a down pipe.
Installation in down pipes
However, the following installation suggestion allows for installation in an open vertical pipeline.
Pipe restrictions or the use of an orifice with a smaller cross-section than the nominal diameter
prevent the sensor running empty while measurement is in progress.
50Endress+Hauser
Page 51
Proline Promass A 300
1
2
3
4
5
A0028773
21Installation in a down pipe (e.g. for batching applications)
1Supply tank
2Sensor
3Orifice plate, pipe restriction
4Valve
5Batching tank
Orientation
DNØ orifice plate, pipe restriction
[mm][in][mm][in]
1¹⁄₂₄0.80.03
2¹⁄₁₂1.50.06
4¹⁄₈3.00.12
The direction of the arrow on the sensor nameplate helps you to install the sensor according to the
flow direction (direction of medium flow through the piping).
OrientationRecommendation
AVertical orientation
BHorizontal orientation, transmitter at
top
CHorizontal orientation, transmitter at
bottom
DHorizontal orientation, transmitter at
side
A0015591
A0015589
A0015590
A0015592
1)
2)
3)
1)This orientation is recommended to ensure self-draining.
2)Applications with low process temperatures may decrease the ambient temperature. To maintain the
minimum ambient temperature for the transmitter, this orientation is recommended.
3)Applications with high process temperatures may increase the ambient temperature. To maintain the
maximum ambient temperature for the transmitter, this orientation is recommended.
If a sensor is installed horizontally with a curved measuring tube, match the position of the sensor to
the fluid properties.
Endress+Hauser51
Page 52
Proline Promass A 300
RUPTURE DISK
i
1
Inlet and outlet runs
Special mounting
instructions
No special precautions need to be taken for fittings which create turbulence, such as valves, elbows
or T-pieces, as long as no cavitation occurs → 61.
Drainability
When the device is installed in a vertical position, the measuring tube can be drained completely and
protected against deposit buildup if the properties of the measured liquid allow this. Furthermore, as
only one measuring tube is used the flow is not impeded and the risk of product being retained in the
measuring device is reduced to a minimum. The larger internal diameter of the measuring tube
1)
also reduces the risk of particles getting trapped in the measuring system. Due to the larger crosssection of the individual measuring tube, the tube is also generally less susceptible to clogging.
Sanitary compatibility
When installing in hygienic applications, please refer to the information in the "Certificates and
approvals/hygienic compatibility" section → 101.
Rupture disk
Information that is relevant to the process: → 61.
WARNING
L
Danger from medium escaping!
Medium escaping under pressure can cause injury or material damage.
Take precautions to prevent danger to persons and damage if the rupture disk is actuated.
‣
Observe information on the rupture disk sticker.
‣
Make sure that the function and operation of the rupture disk is not impeded through the
‣
installation of the device.
Do not remove or damage the rupture disk.
‣
The position of the rupture disk is indicated on a sticker applied over it. If the rupture disk is
triggered, the sticker is destroyed. The disk can therefore be visually monitored.
If the sensor holder is used with a measuring device fitted with a rupture disk, it is important to
ensure that the opening in the neck is not covered over and that the cover of the rupture disk is not
damaged.
A0032051
1Rupture disk label
For information on the dimensions: see the "Mechanical construction" section (accessories)
Sensor holder
The sensor holder is used to secure the device to a wall, tabletop or pipe (order code for "Enclosed
accessories", option PR).
1)Compared with the double-tube design with a similar flow capacity with measuring tubes and a smaller internal diameter
52Endress+Hauser
Page 53
Proline Promass A 300
A
A
!20...70
(!0.79...2.75)
4 x
2 x
M8
1
2
4
5
3
M8
AA
1
A0036471
12 x Allen screw M8 x 50, washer and spring washer A4
21 x clamp (measuring device neck)
34 x securing screw for wall, tabletop or pipe mounting (not supplied)
41 x base profile
52 x clamp (pipe mounting)
AMeasuring device central line
If the holder is used with a measuring device fitted with a rupture disk, it is important to ensure that
the rupture disk in the neck is not covered over and that the cover of the rupture disk is not
damaged.
Lubricate all threaded joints prior to mounting. The screws for wall, tabletop or pipe mounting
are not supplied with the device and must be chosen to suit the individual installation position.
WARNING
L
Strain on pipes!
Excessive strain on unsupported pipes can cause the pipe to break.
Install the sensor in a pipe that is adequately supported.
‣
The following mounting versions are recommended for the installation:
Using the sensor holder.
1Sensor holder (order code for "Enclosed accessories", option PR)
A0036492
Mounting on a wall
Screw the sensor holder to the wall with four screws. Two of the four holes to secure the holder are
designed to hook into the screws.
Endress+Hauser53
Page 54
Proline Promass A 300
146 (5.75)
48 (1.9)
12 (0.47)
280 (11.0)255 (10.0)
134 (5.3)
30 (1.18)
Mounting on a table
Screw the sensor holder onto the tabletop with four screws.
Mounting on a pipe
Secure the sensor holder to the pipe with two clamps.
Zero point adjustment
All measuring devices are calibrated in accordance with state-of-the-art technology. Calibration
takes place under reference conditions→ 46. Therefore, a zero point adjustment in the field is
generally not required.
Experience shows that zero point adjustment is advisable only in special cases:
• To achieve maximum measuring accuracy even with low flow rates.
• Under extreme process or operating conditions (e.g. very high process temperatures or very highviscosity fluids).
Protective cover
Ambient temperature range
Storage temperature
Environment
Measuring device• –40 to +60 °C (–40 to +140 °F)
• Order code for "Test, certificate", option JP:
–50 to +60 °C (–58 to +140 °F)
Readability of the local
display
Dependency of ambient temperature on medium temperature→ 55
If operating outdoors:
‣
Avoid direct sunlight, particularly in warm climatic regions.
You can order a weather protection cover from Endress+Hauser. → 105.
–50 to +80 °C (–58 to +176 °F)
–20 to +60 °C (–4 to +140 °F)
The readability of the display may be impaired at temperatures outside the
temperature range.
A0029553
Climate class
DIN EN 60068-2-38 (test Z/AD)
54Endress+Hauser
Page 55
Proline Promass A 300
Degree of protection
Vibration- and shockresistance
Interior cleaning
Measuring device
• As standard: IP66/67, type 4X enclosure
• When housing is open: IP20, type 1 enclosure
• Display module: IP20, type 1 enclosure
• With the order code for "Sensor options", option CM: IP69 can also be ordered
External WLAN antenna
IP67
Vibration broad-band random, according to IEC 60068-2-6
• 2 to 8.4 Hz, 3.5 mm peak
• 8.4 to 2000 Hz, 1 g peak
Vibration broad-band random, according to IEC 60068-2-64
• 10 to 200 Hz, 0.003 g2/Hz
• 200 to 2000 Hz, 0.001 g2/Hz
• Total: 1.54 g rms
Shock half-sine, according to IEC 60068-2-27
6 ms 30 g
Rough handling shocks, according to IEC 60068-2-31
• Cleaning in place (CIP)
• Sterilization in place (SIP)
Options
Oil- and grease-free version for wetted parts, without declaration
Order code for "Service", option HA
Electromagnetic
compatibility (EMC)
Medium temperature range
• As per IEC/EN 61326 and NAMUR Recommendation 21 (NE 21)
• Device version with PROFIBUS DP: Complies with emission limits for industry as per EN 50170
Volume 2, IEC 61784
The following applies for PROFIBUS DP: If baud rates > 1.5 MBaud, an EMC cable entry must be
used and the cable shield must continue as far as the terminal wherever possible.
Details are provided in the Declaration of Conformity.
Process
–50 to +205 °C (–58 to +401 °F)
Endress+Hauser55
Page 56
Proline Promass A 300
T
a
T
m
B
A
Dependency of ambient temperature on medium temperature
22Exemplary representation, values in the table below.
TaAmbient temperature range
TmMedium temperature
AMaximum permitted medium temperature Tm at T
require a reduced ambient temperature T
a
BMaximum permitted ambient temperature Ta for the maximum specified medium temperature Tm of the
sensor
= 60 °C (140 °F); higher medium temperatures T
a max
A0031121
m
Values for devices used in the hazardous area:
Separate Ex documentation (XA) for the device → 109.
Not insulatedInsulated
ABAB
T
a
T
m
TaT
m
T
a
T
m
T
a
60 °C (140 °F)205 °C (401 °F)––60 °C (140 °F)120 °C (248 °F)55 °C (131 °F)205 °C (401 °F)
Density
Pressure-temperature
ratings
0 to 5000 kg/m3 (0 to 312 lb/cf)
The following pressure/temperature diagrams apply to all pressure-bearing parts of the device and
not just the process connection. The diagrams show the maximum permissible medium pressure
depending on the specific medium temperature.
T
m
56Endress+Hauser
Page 57
Proline Promass A 300
PN40
-80
080160
240
320400
0
50
100
150
200
250
300
350
400
450
500
0
1000
2000
3000
4000
5000
6000
7000
[°C]
[°F]
[bar][psi]
20015010050
0
-50
PN010
PN040
PN40
-80
080160
240
320400
0
50
100
150
200
250
300
350
400
450
500
0
1000
2000
3000
4000
5000
6000
7000
[°C]
[°F]
[bar][psi]
20015010050
0
-50
PN010
PN040
-80
080160
240
320400
0
10
20
30
40
0
[°C]
[°F]
[bar][psi]
20015010050
0
-50
PN40
Flange connection according to EN 1092-1 (DIN 2501)
30Lap joint flange with flange material: 1.4301 (F304), wetted parts Alloy C22: 2.4602 (UNS N06022)
A0036206-EN
Endress+Hauser59
Tri-Clamp process connection
The clamp connections are suitable up to a maximum pressure of 40 bar (580 psi). Please observe
the operating limits of the clamp and seal used as they could be under 40 bar (580 psi). The clamp
and seal are not included in the scope of supply.
The sensor housing is filled with dry nitrogen gas and protects the electronics and mechanics inside.
If a measuring tube fails (e.g. due to process characteristics like corrosive or abrasive fluids), the
fluid will initially be contained by the sensor housing.
In the event of a tube failure, the pressure level inside the sensor housing will rise according to the
operating process pressure. If the user judges that the sensor housing pressure rating/burst pressure
does not provide an adequate safety margin, the device can be fitted with a rupture disk. This
prevents excessively high pressure from forming inside the sensor housing. Therefore, the use of a
60Endress+Hauser
rupture disk is strongly recommended in applications involving high gas pressures, and particularly
in applications in which the process pressure is greater than 2/3 of the sensor housing burst
pressure.
High-pressure devices are always fitted with a rupture disk: order code for "Measuring tube
mat., wetted surface", option HB
A0036210-EN
A0036211-EN
Page 61
Proline Promass A 300
Sensor housing nominal pressure rating and burst pressure
If the device is fitted with a rupture disk (order code for "Sensor option", option CA "Rupture disk"), the
rupture disk trigger pressure is decisive for the maximum nominal pressure .
The sensor housing burst pressure refers to a typical internal pressure which is reached prior to
mechanical failure of the sensor housing and which was determined during type testing. The
corresponding type test declaration can be ordered with the device (order code for "Additional
approval", option LN "Sensor housing burst pressure, type test").
Rupture disk
Flow limit
DNSensor housing nominal
pressure
(designed with a safety factor
≥ 4)
[mm][in][bar][psi][bar][psi]
1¹⁄₂₄253621001450
2¹⁄₁₂253621001450
4¹⁄₈253621001450
Sensor housing burst pressure
For information on the dimensions: see the "Mechanical construction" section
To increase the level of safety, a device version with a rupture disk with a trigger pressure of
10 to 15 bar (145 to 217.5 psi)can be used (order code for "Sensor option", option CA "rupture disk").
High-pressure devices are always fitted with a rupture disk: order code for "Measuring tube
mat., wetted surface", option HB
Select the nominal diameter by optimizing between the required flow range and permissible pressure
loss.
For an overview of the full scale values for the measuring range, see the "Measuring range"
section → 10
• The minimum recommended full scale value is approx. 1/20 of the maximum full scale value
• In most applications, 20 to 50 % of the maximum full scale value can be considered ideal
• A low full scale value must be selected for abrasive media (such as liquids with entrained solids):
flow velocity < 1 m/s (< 3 ft/s).
• For gas measurement the following rules apply:
• The flow velocity in the measuring tubes should not exceed half the sound velocity (0.5 Mach).
• The maximum mass flow depends on the density of the gas: formula → 10
To calculate the flow limit, use the Applicator sizing tool → 107
Pressure loss
System pressure
To calculate the pressure loss, use the Applicator sizing tool → 107
It is important that cavitation does not occur, or that gases entrained in the liquids do not outgas.
This is prevented by means of a sufficiently high system pressure.
For this reason, the following mounting locations are recommended:
• At the lowest point in a vertical pipe
• Downstream from pumps (no danger of vacuum)
A0028777
Thermal insulation
In the case of some fluids, it is important to keep the heat radiated from the sensor to the
transmitter to a low level. A wide range of materials can be used for the required insulation.
Endress+Hauser61
Page 62
Proline Promass A 300
NOTICE
Electronics overheating on account of thermal insulation!
Maximum permissible temperature at the lower end of the transmitter housing: 80 °C (176 °F)
‣
Thermal insulation with extended neck free: We recommend that you do not insulate the
‣
extended neck in order to ensure optimum dissipation of heat.
34Thermal insulation with extended neck free
A0034391
Heating
Vibrations
Some fluids require suitable measures to avoid loss of heat at the sensor.
Heating options
• Electrical heating, e.g. with electric band heaters
• Via pipes carrying hot water or steam
• Via heating jackets
Heating jackets for the sensors can be ordered as accessories from Endress+Hauser. → 106
NOTICE
Danger of overheating when heating
Ensure that the temperature at the lower end of the transmitter housing does not exceed
‣
80 °C (176 °F).
Ensure that sufficient convection takes place at the transmitter neck.
‣
Ensure that a sufficiently large area of the transmitter neck remains exposed. The uncovered part
‣
serves as a radiator and protects the electronics from overheating and excessive cooling.
When using in potentially explosive atmospheres, observe the information in the device-specific
‣
Ex documentation. For detailed information on the temperature tables, see the separate
document entitled "Safety Instructions" (XA) for the device.
The high oscillation frequency of the measuring tubes ensures that the correct operation of the
measuring system is not influenced by plant vibrations.
62Endress+Hauser
Page 63
Proline Promass A 300
D
L
E
F
A
CB
K
G
H
I
M
Mechanical construction
Dimensions in SI unitsCompact version
A0033783
Order code for "Housing", option A "Aluminum, coated"
1)Depending on the cable gland used: values up to + 30 mm
2)For version without local display: values - 13 mm
3)Dependent on the respective process connection
2)
HI
2)
KLM
3)
3)
3)
34
48
51
Endress+Hauser63
Page 64
Proline Promass A 300
L
A
B
L
A
B
C
Threaded glands
VCO coupling
Length tolerance for dimension L in mm:
+1.5 / –2.0
4-VCO-4
Order code for "Process connection", option HAW
1.4435 (316/316L): order code for "Measuring tube mat., wetted surface", option BB, BF, SA
Alloy C22: order code for "Measuring tube mat., wetted surface", option HA
Alloy C22, high pressure: order code for "Measuring tube mat., wetted surface", option HB
DN
[mm]
1AF ¹¹⁄₁₆1.11187
2AF ¹¹⁄₁₆2.52.1264
4AF ¹¹⁄₁₆3.93.2310
A
[in]
Option BB, BF, HA,
SA
B
[mm]
Option HB
A0015624
L
[mm]
G ¹⁄₄ "
Order code for "Process connection", option G06
1.4404 (316L): order code for "Measuring tube mat., wetted surface", option SA
Alloy C22: order code for "Measuring tube mat., wetted surface", option HA
Alloy C22, high pressure: order code for "Measuring tube mat., wetted surface", option HB
DN
[mm]
Option HA, SAOption HB
122.525G ¹⁄₄ "AF 21257
222.525G ¹⁄₄ "AF 21334
422.525G ¹⁄₄ "AF 21380
A
[mm]
B
[in]
C
[mm]
L
[mm]
64Endress+Hauser
A0036429
Page 65
Proline Promass A 300
NPT ¹⁄₄ "
Order code for "Process connection", option P06
1.4404 (316L): order code for "Measuring tube mat., wetted surface", option SA
Alloy C22: order code for "Measuring tube mat., wetted surface", option HA
Alloy C22, high pressure: order code for "Measuring tube mat., wetted surface", option HB
DN
[mm]
Option HA, SAOption HB
122.525NPT ¹⁄₄ "AF 19257
222.525NPT ¹⁄₄ "AF 19334
422.525NPT ¹⁄₄ "AF 19380
A
[mm]
B
[in]
C
[mm]
L
[mm]
Endress+Hauser65
Page 66
Proline Promass A 300
L
A
B
Clamp connections
Tri-Clamp
Length tolerance for dimension L in mm:
+1.5 / –2.0
½" Tri-Clamp
Order code for "Process connection", option FBW
1.4435 (316L): order code for "Measuring tube mat., wetted surface", option BB, BF, SA
DN
[mm]
1259.5193
2259.5270
4259.5316
3-A version available (Ra ≤ 0.76 µm/30 µin, Ra ≤ 0.38 µm/15 µin):
Order code for "Measuring tube mat., wetted surface", option BB, BF in conjunction with order code for
"Additional approval" , option LP
A
[mm]
B
[mm]
[mm]
A0015625
L
66Endress+Hauser
Page 67
Proline Promass A 300
C
D
A
B
E
L
Flange connections
Fixed flange EN 1092-1, ASME B16.5, JIS B2220
A0015621
Length tolerance for dimension L in mm:
+1.5 / –2.0
Flange according to EN 1092-1 (DIN 2501 / DIN 2512N), PN 40
1.4404 (F316/F316L): order code for "Process connection", option D2S
Alloy C22: order code for "Process connection", option D2C
Flange with groove according to EN 1092-1 Form D (DIN 2512N), PN 40
1.4404 (F316/F316L): order code for "Process connection", option D6S
Alloy C22: order code for "Process connection", option D6C
DN
[mm]
195654 × Ø141617.3262
295654 × Ø141617.3339
495654 × Ø141617.3385
Surface roughness (flange): EN 1092-1 Form B1 (DIN 2526 Form C), Ra 3.2 to 12.5 µm
Flange according to EN 1092-1 (DIN 2501 / DIN 2512N), PN 100
1.4404 (F316/F316L): order code for "Process connection", option D4S
Alloy C22: order code for "Process connection", option D4C
Flange with groove according to EN 1092-1 Form D (DIN 2512N), PN 100
1.4404 (F316/F316L): order code for "Process connection", option D8S
Alloy C22: order code for "Process connection", option D8C
DN
[mm]
1105754 × Ø142017.3292
2105754 × Ø142017.3369
4105754 × Ø142017.3415
Surface roughness (flange): EN 1092-1 Form B1 (DIN 2526 Form C), Ra 3.2 to 12.5 µm
A
[mm]
A
[mm]
B
[mm]
B
[mm]
C
[mm]
C
[mm]
D
[mm]
D
[mm]
E
[mm]
E
[mm]
L
[mm]
L
[mm]
Endress+Hauser67
Page 68
Proline Promass A 300
Flange according to EN 1092-1 (DIN 2501 / DIN 2512N), PN 400
1.4404 (F316/F316L): order code for "Process connection", option DNS
Alloy C22: order code for "Process connection", option DNC
Flange with groove according to EN 1092-1 Form D (DIN 2512N), PN 400
1.4404 (F316/F316L): order code for "Process connection", option DPS
Alloy C22: order code for "Process connection", option DPC
DN
[mm]
11451004 × Ø223017.3336
21451004 × Ø223017.3413
41451004 × Ø223017.3459
Surface roughness (flange): EN 1092-1 Form B1 (DIN 2526 Form C), Ra 3.2 to 12.5 µm
Flange according to ASME B16.5, Class 150 RF, Schedule 40
1.4404 (F316/F316L): order code for "Process connection", option AAS
Alloy C22: order code for "Process connection", option AAC
DN
[mm]
19060.34 × Ø15.911.615.7262
29060.34 × Ø15.911.615.7339
49060.34 × Ø15.911.615.7385
Surface roughness (flange): Ra 3.2 to 6.3 µm
A
[mm]
A
[mm]
B
[mm]
B
[mm]
C
[mm]
C
[mm]
D
[mm]
D
[mm]
E
[mm]
E
[mm]
L
[mm]
L
[mm]
Flange according to ASME B16.5, Class 300 RF, Schedule 40
1.4404 (F316/F316L): order code for "Process connection", option ABS
Alloy C22: order code for "Process connection", option ABC
DN
[mm]
19566.74 × Ø15.914.715.7262
29566.74 × Ø15.914.715.7339
49566.74 × Ø15.914.715.7385
Surface roughness (flange): Ra 3.2 to 6.3 µm
Flange according to ASME B16.5, Class 600 RF, Schedule 80
1.4404 (F316/F316L): order code for "Process connection", option ACS
Alloy C22: order code for "Process connection", option ACC
DN
[mm]
19566.74 × Ø15.921.313.9292
29566.74 × Ø15.921.313.9369
49566.74 × Ø15.921.313.9415
Surface roughness (flange): Ra 3.2 to 6.3 µm
A
[mm]
A
[mm]
B
[mm]
B
[mm]
C
[mm]
C
[mm]
D
[mm]
D
[mm]
E
[mm]
E
[mm]
L
[mm]
L
[mm]
68Endress+Hauser
Page 69
Proline Promass A 300
Flange according to ASME B16.5, Class 900/1500 RF, Schedule 80
1.4404 (F316/F316L): order code for "Process connection", option ARS
Alloy C22: order code for "Process connection", option ARC
Flange according to ASME B16.5, Class 900/1500 RTJ, Schedule 80
1.4404 (F316/F316L): order code for "Process connection", option ASS
Alloy C22: order code for "Process connection", option ASC
DN
[mm]
112082.64 × Ø22
212082.64 × Ø22
412082.64 × Ø22
A
[mm]
B
[mm]
C
[mm]
D
[mm]
1)
1)
1)
29.314324
29.314401
29.314447
E
[mm]
Surface roughness (flange): Ra 3.2 to 6.3 µm
1)option ARC/ARS: 4 × Ø22.2
Flange according to ASME B16.5, Class 2500 RF, Schedule 80
1.4404 (F316/F316L): order code for "Process connection", option ATS
Alloy C22: order code for "Process connection", option ATC
Flange according to ASME B16.5, Class 2500 RTJ, Schedule 80
1.4404 (F316/F316L): order code for "Process connection", option AUS
Alloy C22: order code for "Process connection", option AUC
DN
[mm]
A
[mm]
B
[mm]
C
[mm]
D
[mm]
E
[mm]
113588.94 × Ø22.237.214351
213588.94 × Ø22.237.214428
413588.94 × Ø22.237.214474
Surface roughness (flange): Ra 3.2 to 6.3 µm
L
[mm]
L
[mm]
Flange JIS B2220, 20K
1.4404 (F316/F316L): order code for "Process connection", option NES
Alloy C22: order code for "Process connection", option NEC
DN
[mm]
A
[mm]
B
[mm]
C
[mm]
D
[mm]
E
[mm]
195704 × Ø151415262
295704 × Ø151415339
495704 × Ø151415385
Surface roughness (flange): Ra 3.2 to 6.3 µm
Flange JIS B2220, 40K
1.4404 (F316/F316L): order code for "Process connection", option NGS
Alloy C22: order code for "Process connection", option NGC
DN
[mm]
A
[mm]
B
[mm]
C
[mm]
D
[mm]
E
[mm]
1115804 × Ø192015292
2115804 × Ø192015369
4115804 × Ø192015415
Surface roughness (flange): Ra 3.2 to 6.3 µm
L
[mm]
L
[mm]
Endress+Hauser69
Page 70
Proline Promass A 300
Flange JIS B2220, 63K
1.4404 (F316/F316L): order code for "Process connection", option NHS
Alloy C22: order code for "Process connection", option NHC
DN
[mm]
1120854 × Ø192312312
2120854 × Ø192312389
4120854 × Ø192312435
Surface roughness (flange): Ra 3.2 to 6.3 µm
A
[mm]
B
[mm]
C
[mm]
D
[mm]
E
[mm]
L
[mm]
70Endress+Hauser
Page 71
Proline Promass A 300
L
C
D
A
B
F
E
Lap joint flange EN 1092-1, ASME B16.5, JIS B2220
Length tolerance for dimension L in mm:
+1.5 / –2.0
Lap joint flange according to EN 1092-1 Form D: PN 40
1.4301 (F304), wetted parts Alloy C22: order code for "Process connection", option DAC
DN
[mm]
195654 × Ø1414.54517.3262
295654 × Ø1414.54517.3339
495654 × Ø1414.54517.3385
Surface roughness (flange): Ra 3.2 to 12.5 µm
A
[mm]
B
[mm]
C
[mm]
D
[mm]
E
[mm]
F
[mm]
A0022221
L
[mm]
Lap joint flange according to ASME B16.5: Class 150, Schedule 40
1.4301 (F304), wetted parts Alloy C22: order code for "Process connection", option ADC
DN
[mm]
19060.34 × Ø15.91535.115.7262
29060.34 × Ø15.91535.115.7339
49060.34 × Ø15.91535.115.7385
Surface roughness (flange): Ra 3.2 to 12.5 µm
Lap joint flange according to ASME B16.5: Class 300, Schedule 40
1.4301 (F304), wetted parts Alloy C22: order code for "Process connection", option AEC
DN
[mm]A[mm]
19566.74 × Ø15.916.535.115.7268+6
29566.74 × Ø15.916.535.115.7245+6
49566.74 × Ø15.916.535.115.7391+6
Surface roughness (flange): Ra 3.2 to 12.5 µm
1)Difference to installation length of the welding neck flange (order code for "Process connection", option
AAC)
A
[mm]
B
[mm]
B
[mm]
C
[mm]
C
[mm]
D
[mm]
D
[mm]E[mm]F[mm]L[mm]
E
[mm]
F
[mm]
L
[mm]
L
diff
[mm]
1)
Endress+Hauser71
Page 72
Proline Promass A 300
F
A
B
D
C
K
GHGI
L
G
M
ED
N
O
C
E
D
Lap joint flange according to ASME B16.5: Class 600, Schedule 80
1.4301 (F304), wetted parts Alloy C22: order code for "Process connection", option AFC
DN
[mm]
19566.74 × Ø15.91735.113.9292
29566.74 × Ø15.91735.113.9369
49566.74 × Ø15.91735.113.9415
Surface roughness (flange): Ra 3.2 to 12.5 µm
Lap joint flange JIS B2220: 20K
1.4301 (F304), wetted parts Alloy C22: order code for "Process connection", option NIC
DN
[mm]
195704 × Ø15145115262
295704 × Ø15145115339
495704 × Ø15145115385
Surface roughness (flange): Ra 3.2 to 12.5 µm
A
[mm]
A
[mm]
B
[mm]
B
[mm]
C
[mm]
C
[mm]
[mm]
D
[mm]
D
E
[mm]
E
[mm]
F
[mm]
F
[mm]
L
[mm]
L
[mm]
Accessories
Sensor holder
A
[mm]
106117187023.52123
B
[mm]
C
[mm]
D
[mm]
E
[mm]
F
[mm]
A0036633
G
[mm]
H
[mm]
62129012092915
I
[mm]
K
[mm]
L
[mm]
M
[mm]
N
[mm]
72Endress+Hauser
O
[mm]
Page 73
Proline Promass A 300
146 (5.75)
48 (1.9)
12 (0.47)
280 (11.0)255 (10.0)
134 (5.3)
30 (1.18)
136 (5.35)
114 (4.49)
86 (3.39)
78 (3.07)
!103 (4.06)
!12 (0.47)
7.0 (0.27)
68 (2.7)105 (4.1)
173 (6.8)
Protective cover
A0029553
Remote display and operating module DKX001
35Engineering unit mm (in)
External WLAN antenna
External WLAN antenna mounted on device
36Engineering unit mm (in)
A0028921
The external WLAN antenna is not suitable for use in hygienic applications.
A0028923
Endress+Hauser73
Page 74
Proline Promass A 300
70 (2.8)
1500 (59.1)
72 (2.8)
D
L
E
F
A
CB
K
G
H
I
M
External WLAN antenna mounted with cable
The external WLAN antenna can be mounted separately from the transmitter if the transmission/
reception conditions at the transmitter mounting location are poor.
A0033597
37Engineering unit mm (in)
Dimensions in US unitsCompact version
Order code for "Housing", option A "Aluminum, coated"
1)Depending on the cable gland used: values up to + 1.18 in
2)For version without local display: values - 0.51 in
3)Dependent on the respective process connection
Endress+Hauser75
Page 76
Proline Promass A 300
L
A
B
L
A
B
C
Threaded glands
VCO coupling
Length tolerance for dimension L in inch:
+0.06 / –0.08
4-VCO-4
Order code for "Process connection", option HAW
1.4435 (316/316L): order code for "Measuring tube mat., wetted surface", option BB, BF, SA
Alloy C22: order code for "Measuring tube mat., wetted surface", option HA
Alloy C22, high pressure: order code for "Measuring tube mat., wetted surface", option HB
DN
[in]
¹⁄₂₄AF ¹¹⁄₁₆0.040.047.36
¹⁄₁₂AF ¹¹⁄₁₆0.10.0810.39
¹⁄₈AF ¹¹⁄₁₆0.150.1312.2
A
[in]
Option BB, BF, HA,
SA
B
[in]
Option HB
A0015624
L
[in]
G ¹⁄₄ "
Order code for "Process connection", option G06
1.4404 (316L): order code for "Measuring tube mat., wetted surface", option SA
Alloy C22: order code for "Measuring tube mat., wetted surface", option HA
Alloy C22, high pressure: order code for "Measuring tube mat., wetted surface", option HB
DN
[in]
Option HA, SAOption HB
¹⁄₂₄0.890.98G ¹⁄₄ "AF ¹³⁄₁₆ "10.12
¹⁄₁₂0.890.98G ¹⁄₄ "AF ¹³⁄₁₆ "13.15
¹⁄₈0.890.98G ¹⁄₄ "AF ¹³⁄₁₆ "14.96
A
[in]
B
[in]
C
[in]
L
[in]
76Endress+Hauser
A0036429
Page 77
Proline Promass A 300
NPT ¹⁄₄ "
Order code for "Process connection", option P06
1.4404 (316L): order code for "Measuring tube mat., wetted surface", option SA
Alloy C22: order code for "Measuring tube mat., wetted surface", option HA
Alloy C22, high pressure: order code for "Measuring tube mat., wetted surface", option HB
DN
[in]
Option HA, SAOption HB
¹⁄₂₄0.890.98NPT ¹⁄₄ "AF ³⁄₄ "10.12
¹⁄₁₂0.890.98NPT ¹⁄₄ "AF ³⁄₄ "13.15
¹⁄₈0.890.98NPT ¹⁄₄ "AF ³⁄₄ "14.96
A
[in]
B
[in]
C
[in]
L
[in]
Endress+Hauser77
Page 78
Proline Promass A 300
L
A
B
Clamp connections
Tri-Clamp
Length tolerance for dimension L in inch:
+0.06 / –0.08
½" Tri-Clamp
Order code for "Process connection", option FBW
1.4435 (316L): order code for "Measuring tube mat., wetted surface", option BB, BF, SA
DN
[in]
¹⁄₂₄0.980.377.6
¹⁄₁₂0.980.3710.63
¹⁄₈0.980.3712.44
3-A version available (Ra ≤ 0.76 µm/30 µin, Ra ≤ 0.38 µm/15 µin):
Order code for "Measuring tube mat., wetted surface", option BB, BF in conjunction with order code for
"Additional approval" , option LP
A
[in]
B
[in]
[in]
A0015625
L
78Endress+Hauser
Page 79
Proline Promass A 300
C
D
A
B
E
L
Flange connections
Fixed flange ASME B16.5
A0015621
Length tolerance for dimension L in inch:
+0.06 / –0.08
Flange according to ASME B16.5, Class 150 RF, Schedule 40
1.4404 (F316/F316L): order code for "Process connection", option AAS
Alloy C22: order code for "Process connection", option AAC
DN
[in]
¹⁄₂₄3.542.374 × Ø0.630.460.6210.31
¹⁄₁₂3.542.374 × Ø0.630.460.6213.35
¹⁄₈3.542.374 × Ø0.630.460.6215.16
Surface roughness (flange): Ra 3.2 to 6.3 µm
Flange according to ASME B16.5, Class 300 RF, Schedule 40
1.4404 (F316/F316L): order code for "Process connection", option ABS
Alloy C22: order code for "Process connection", option ABC
DN
[in]
¹⁄₂₄3.742.634 × Ø0.630.580.6210.31
¹⁄₁₂3.742.634 × Ø0.630.580.6213.35
¹⁄₈3.742.634 × Ø0.630.580.6215.16
Surface roughness (flange): Ra 3.2 to 6.3 µm
A
[in]
A
[in]
B
[in]
B
[in]
C
[in]
C
[in]
D
[in]
D
[in]
E
[in]
E
[in]
L
[in]
L
[in]
Flange according to ASME B16.5, Class 600 RF, Schedule 80
1.4404 (F316/F316L): order code for "Process connection", option ACS
Alloy C22: order code for "Process connection", option ACC
DN
[in]
¹⁄₂₄3.742.634 × Ø0.630.840.5511.5
¹⁄₁₂3.742.634 × Ø0.630.840.5514.53
¹⁄₈3.742.634 × Ø0.630.840.5516.34
Endress+Hauser79
Surface roughness (flange): Ra 3.2 to 6.3 µm
A
[in]
B
[in]
C
[in]
D
[in]
E
[in]
L
[in]
Page 80
Proline Promass A 300
Flange according to ASME B16.5, Class 900/1500 RF, Schedule 80
1.4404 (F316/F316L): order code for "Process connection", option ARS
Alloy C22: order code for "Process connection", option ARC
Flange according to ASME B16.5, Class 900/1500 RTJ, Schedule 80
1.4404 (F316/F316L): order code for "Process connection", option ASS
Alloy C22: order code for "Process connection", option ASC
DN
[in]
¹⁄₂₄4.723.254 × Ø0.871.150.5512.76
¹⁄₁₂4.723.254 × Ø0.871.150.5515.79
¹⁄₈4.723.254 × Ø0.871.150.5517.6
Surface roughness (flange): Ra 3.2 to 6.3 µm
Flange according to ASME B16.5, Class 2500 RF, Schedule 80
1.4404 (F316/F316L): order code for "Process connection", option ATS
Alloy C22: order code for "Process connection", option ATC
Flange according to ASME B16.5, Class 2500 RTJ, Schedule 80
1.4404 (F316/F316L): order code for "Process connection", option AUS
Alloy C22: order code for "Process connection", option AUC
DN
[in]
¹⁄₂₄5.313.54 × Ø0.871.460.5513.82
¹⁄₁₂5.313.54 × Ø0.871.460.5516.85
¹⁄₈5.313.54 × Ø0.871.460.5518.66
Surface roughness (flange): Ra 3.2 to 6.3 µm
A
[in]
A
[in]
B
[in]
B
[in]
[in]
C
[in]
C
D
[in]
D
[in]
E
[in]
E
[in]
L
[in]
L
[in]
80Endress+Hauser
Page 81
Proline Promass A 300
L
C
D
A
B
F
E
Lap joint flange ASME B16.5
Length tolerance for dimension L in mm:
+1.5 / –2.0
Lap joint flange according to ASME B16.5: Class 150, Schedule 40
1.4301 (F304), wetted parts Alloy C22: order code for "Process connection", option ADC
DN
[in]
¹⁄₂₄3.542.374 × Ø0.630.591.650.6210.31
¹⁄₁₂3.542.374 × Ø0.630.591.650.6213.35
¹⁄₈3.542.374 × Ø0.630.591.650.6215.16
Surface roughness (flange): Ra 3.2 to 12.5 µm
A
[in]
B
[in]
C
[in]
D
[in]
E
[in]
F
[in]
A0022221
L
[in]
Lap joint flange according to ASME B16.5: Class 300, Schedule 40
1.4301 (F304), wetted parts Alloy C22: order code for "Process connection", option AEC
DN
[in]
¹⁄₂₄3.742.634 × Ø0.630.651.770.6210.550.24
¹⁄₁₂3.742.634 × Ø0.630.651.770.629.650.24
¹⁄₈3.742.634 × Ø0.630.651.770.6215.390.24
Surface roughness (flange): Ra 3.2 to 12.5 µm
1)Difference to installation length of the welding neck flange (order code for "Process connection", option
AAC)
Lap joint flange according to ASME B16.5: Class 600, Schedule 80
1.4301 (F304), wetted parts Alloy C22: order code for "Process connection", option AFC
DN
[in]
¹⁄₂₄3.742.634 × Ø15.90.671.890.5511.5
¹⁄₁₂3.742.634 × Ø15.90.671.890.5514.53
¹⁄₈3.742.634 × Ø15.90.671.890.5516.34
Surface roughness (flange): Ra 3.2 to 12.5 µm
A
[in]
A
[in]
B
[in]
B
[in]
C
[in]
C
[in]
D
[in]
D
[in]
E
[in]
E
[in]
F
[in]
L
[in]
F
[in]
L
diff
[in]
L
[in]
1)
Endress+Hauser81
Page 82
Accessories
F
A
B
D
C
K
GHGI
L
G
M
ED
N
O
C
E
D
146 (5.75)
48 (1.9)
12 (0.47)
280 (11.0)255 (10.0)
134 (5.3)
30 (1.18)
Sensor holder
Proline Promass A 300
A0036633
A
[in]
4.174.60.72.40.90.830.9
H
[in]
2.40.53.544.73.60.350.6
[in]
B
[in]
I
[in]
K
[in]
C
D
[in]
L
[in]
E
[in]
M
[in]
F
[in]
N
[in]
Protective cover
G
[in]
O
[in]
A0029553
82Endress+Hauser
Page 83
Proline Promass A 300
136 (5.35)
114 (4.49)
86 (3.39)
78 (3.07)
!103 (4.06)
!12 (0.47)
7.0 (0.27)
68 (2.7)105 (4.1)
173 (6.8)
Remote display and operating module DKX001
A0028921
38Engineering unit mm (in)
External WLAN antenna
The external WLAN antenna is not suitable for use in hygienic applications.
External WLAN antenna mounted on device
A0028923
39Engineering unit mm (in)
External WLAN antenna mounted with cable
The external WLAN antenna can be mounted separately from the transmitter if the transmission/
reception conditions at the transmitter mounting location are poor.
Endress+Hauser83
Page 84
70 (2.8)
1500 (59.1)
72 (2.8)
40Engineering unit mm (in)
Proline Promass A 300
A0033597
Weight
All values (weight exclusive of packaging material) refer to devices with VCO couplings. Weight
specifications including transmitter as per order code for "Housing", option A "Aluminum, coated".
Different values due to different transmitter versions:
• Transmitter version for the hazardous area
(Order code for "Housing", option A "Aluminum, coated"; Ex d): +2 kg (+4.4 lbs)
• Transmitter version for hygienic area
(Order code for "Housing", option B "Stainless, hygienic"): +0.2 kg (+0.44 lbs)
Weight in SI units
DN [mm]Weight [kg]
15.35
26.9
48.75
Weight in US units
DN [in]Weight [lbs]
1/2412
1/1215
1/819
MaterialsTransmitter housing
Order code for "Housing":
• Option A "Aluminum, coated": aluminum, AlSi10Mg, coated
• Option B "Stainless, hygienic": stainless steel, 1.4404 (316L)
Window material
Order code for "Housing":
• Option A "Aluminum, coated": glass
• Option B "Stainless, hygienic": polycarbonate
Seals
Order code for "Housing":
Option B "Stainless, hygienic": EPDM and silicone
84Endress+Hauser
Page 85
Proline Promass A 300
1
2
3
4
Cable entries/cable glands
A0028352
41Possible cable entries/cable glands
1Female thread M20 × 1.5
2Cable gland M20 × 1.5
3Adapter for cable entry with internal thread G ½" or NPT ½"
4Device plugs
Order code for "Housing", option A "Aluminum, coated"
The various cable entries are suitable for hazardous and non-hazardous areas.
Cable entry/cable glandMaterial
Coupling M20 × 1.5
Adapter for cable entry with internal thread G ½"Nickel-plated brass
Adapter for cable entry with internal thread NPT ½"
Non-Ex: plastic
Z2, D2, Ex d/de: brass with plastic
Order code for "Housing", option B "Stainless, hygienic"
The various cable entries are suitable for hazardous and non-hazardous areas.
Cable entry/cable glandMaterial
Cable gland M20 × 1.5Plastic
Adapter for cable entry with internal thread G ½"Nickel-plated brass
Adapter for cable entry with internal thread NPT ½"
• Antenna: ASA plastic (acrylic ester-styrene-acrylonitrile) and nickel-plated brass
• Adapter: Stainless steel and nickel-plated brass
• Cable: Polyethylene
• Plug: Nickel-plated brass
• Angle bracket: Stainless steel
Process connections
Surface roughness
• Fixed flange connections:
• EN 1092-1 (DIN 2501) flange
• EN 1092-1 (DIN 2512N) flange
• ASME B16.5 flange
• JIS B2220 flange
• Clamp connections:
Tri-Clamp (OD tubes), DIN 11866 series C
• VCO connections:
4-VCO-4
• Internal thread:
• Cylindrical internal thread BSPP (G) in accordance with ISO 228-1 with sealing surfaces in
accordance with DIN 3852-2/ISO 1179-1
• NPT
Process connection materials → 86
All data relate to parts in contact with fluid. The following surface roughness quality can be ordered.
• Not polished
• Ra
• Ra
= 0.76 µm (30 µin) mechanically polished
max
= 0.38 µm (15 µin) mechanically polished
max
Human interface
Operating concept
Operator-oriented menu structure for user-specific tasks
• Commissioning
• Operation
• Diagnostics
• Expert level
Fast and safe commissioning
• Guided menus ("Make-it-run" wizards) for applications
• Menu guidance with brief descriptions of the individual parameter functions
• Access to the device via Web server → 107
• WLAN access to the device via mobile handheld terminal, tablet or smart phone
Reliable operation
• Operation in local language → 88
• Uniform operating philosophy applied to device and operating tools
• If replacing electronic modules, transfer the device configuration via the integrated memory
(HistoROM backup) which contains the process and measuring device data and the event logbook.
No need to reconfigure.
• Troubleshooting measures can be called up via the device and in the operating tools
• Diverse simulation options, logbook for events that occur and optional line recorder functions
Endress+Hauser87
Page 88
Proline Promass A 300
Languages
Can be operated in the following languages:
• Via local operation
English, German, French, Spanish, Italian, Dutch, Portuguese, Polish, Russian, Turkish, Chinese,
Japanese, Korean, Bahasa (Indonesian), Vietnamese, Czech, Swedish
• Via Web browser
English, German, French, Spanish, Italian, Dutch, Portuguese, Polish, Russian, Turkish, Chinese,
Japanese, Korean, Bahasa (Indonesian), Vietnamese, Czech, Swedish
• Via "FieldCare", "DeviceCare" operating tool: English, German, French, Spanish, Italian, Chinese,
Japanese
Local operationVia display module
Equipment:
• Order code for "Display; operation", option F "4-line, illuminated, graphic display; touch control"
• Order code for "Display; operation", option G "4-line, illuminated, graphic display; touch control +
WLAN"
Information about WLAN interface → 95
42Operation with touch control
Display elements
• 4-line, illuminated, graphic display
• White background lighting; switches to red in event of device errors
• Format for displaying measured variables and status variables can be individually configured
• Permitted ambient temperature for the display: –20 to +60 °C (–4 to +140 °F)
The readability of the display may be impaired at temperatures outside the temperature range.
Operating elements
• External operation via touch control (3 optical keys) without opening the housing: , ,
• Operating elements also accessible in the various zones of the hazardous area
Via remote display and operating module DKX001
The remote display and operating module DKX001 is available as an optional extra → 105.
• The remote display and operating module DKX001 is only available for the following housing
version: order code for "Housing": option A "Aluminum, coated"
• The measuring device is always supplied with a dummy cover when the remote display and
operating module DKX001 is ordered directly with the measuring device. Display or operation
at the transmitter is not possible in this case.
• If ordered subsequently, the remote display and operating module DKX001 may not be
connected at the same time as the existing measuring device display module. Only one
display or operation unit may be connected to the transmitter at any one time.
A0026785
88Endress+Hauser
Page 89
Proline Promass A 300
A0026786
43Operation via remote display and operating module DKX001
Display and operating elements
The display and operating elements correspond to those of the display module → 88.
Material
The housing material of the display and operating module DKX001 depends on the choice of
transmitter housing material.
Transmitter housingRemote display and operating module
Order code for "Housing"MaterialMaterial
Option A "Aluminum, coated"AlSi10Mg, coatedAlSi10Mg, coated
Cable entry
Corresponds to the choice of transmitter housing, order code for "Electrical connection".
Connecting cable
→ 45
Dimensions
→ 73
Remote operationVia HART protocol
This communication interface is available in device versions with a HART output.
Endress+Hauser89
Page 90
Proline Promass A 300
1235
8
47
6
1
10
2
3
457
69
8
44Options for remote operation via HART protocol (active)
1Control system (e.g. PLC)
2Field Communicator 475
3Computer with Web browser (e.g. Internet Explorer) for access to the integrated device Web server or
computer with an operating tool (e.g. FieldCare, DeviceCare, AMS Device Manager, SIMATIC PDM) with COM
DTM "CDI Communication TCP/IP"
4Commubox FXA195 (USB)
5Field Xpert SFX350 or SFX370
6Field Xpert SMT70
7VIATOR Bluetooth modem with connecting cable
8Transmitter
A0028747
45Options for remote operation via HART protocol (passive)
1Control system (e.g. PLC)
2Transmitter power supply unit, e.g. RN221N (with communication resistor)
3Connection for Commubox FXA195 and Field Communicator 475
4Field Communicator 475
5Computer with Web browser (e.g. Internet Explorer) for access to the integrated device Web server or
computer with an operating tool (e.g. FieldCare, DeviceCare, AMS Device Manager, SIMATIC PDM) with COM
DTM "CDI Communication TCP/IP"
6Commubox FXA195 (USB)
7Field Xpert SFX350 or SFX370
8Field Xpert SMT70
9VIATOR Bluetooth modem with connecting cable
10 Transmitter
Via FOUNDATION Fieldbus network
This communication interface is available in device versions with FOUNDATION Fieldbus.
90Endress+Hauser
A0028746
Page 91
Proline Promass A 300
12
3
1
9898
4
66
5
12
3
444
A0028837
46Options for remote operation via FOUNDATION Fieldbus network
1Automation system
2Computer with FOUNDATION Fieldbus network card
3Industry network
4High Speed Ethernet FF-HSE network
5Segment coupler FF-HSE/FF-H1
6FOUNDATION Fieldbus FF-H1 network
7Power supply FF-H1 network
8T-box
9Measuring device
Via PROFIBUS DP network
This communication interface is available in device versions with PROFIBUS DP.
47Options for remote operation via PROFIBUS DP network
1Automation system
2Computer with PROFIBUS network card
3PROFIBUS DP network
4Measuring device
Via PROFIBUS PA network
This communication interface is available in device versions with PROFIBUS PA.
A0020903
Endress+Hauser91
Page 92
2
3
1
76
5
4
48Options for remote operation via PROFIBUS PA network
12
3
1Automation system
2Computer with PROFIBUS network card
3PROFIBUS DP network
4Segment coupler PROFIBUS DP/PA
5PROFIBUS PA network
6T-box
7Measuring device
Proline Promass A 300
A0028838
Via Modbus RS485 protocol
This communication interface is available in device versions with a Modbus-RS485 output.
A0029437
49Options for remote operation via Modbus-RS485 protocol (active)
1Control system (e.g. PLC)
2Computer with Web browser (e.g. Internet Explorer) for accessing the integrated device Web server or with
operating tool (e.g. FieldCare, DeviceCare) with COM DTM "CDI Communication TCP/IP" or Modbus DTM
3Transmitter
Via EtherNet/IP network
This communication interface is available in device versions with EtherNet/IP.
92Endress+Hauser
Page 93
Proline Promass A 300
4
213
555
4
213
5
55
Star topology
A0032078
50Options for remote operation via EtherNet/IP network: star topology
1Automation system, e.g. "RSLogix" (Rockwell Automation)
2Workstation for measuring device operation: with Custom Add-On Profile for "RSLogix 5000" (Rockwell
Automation) or with Electronic Data Sheet (EDS)
3Computer with Web browser (e.g. Internet Explorer) for accessing the integrated device Web server or
computer with operating tool (e.g. FieldCare, DeviceCare) with COM DTM "CDI Communication TCP/IP"
4Ethernet switch
5Measuring device
Ring topology
The device is integrated via the terminal connection for signal transmission (output 1) and the
service interface (CDI-RJ45).
51Options for remote operation via EtherNet/IP network: ring topology
1Automation system, e.g. "RSLogix" (Rockwell Automation)
2Workstation for measuring device operation: with Custom Add-On Profile for "RSLogix 5000" (Rockwell
Automation) or with Electronic Data Sheet (EDS)
3Computer with Web browser (e.g. Internet Explorer) for accessing the integrated device Web server or
computer with operating tool (e.g. FieldCare, DeviceCare) with COM DTM "CDI Communication TCP/IP"
4Ethernet switch
5Measuring device
A0033725
Via PROFINET network
This communication interface is available in device versions with PROFINET.
Endress+Hauser93
Page 94
Proline Promass A 300
312
444
3
12
444
Star topology
52Options for remote operation via PROFINET network: star topology
1Automation system, e.g. Simatic S7 (Siemens)
2Computer with Web browser (e.g. Internet Explorer) for accessing the integrated device Web server or
computer with operating tool (e.g. FieldCare, DeviceCare, SIMATIC PDM) with COM DTM "CDI Communication
TCP/IP"
3Switch, e.g. Scalance X204 (Siemens)
4Measuring device
A0026545
Ring topology
The device is integrated via the terminal connection for signal transmission (output 1) and the
service interface (CDI-RJ45).
53Options for remote operation via PROFINET network: ring topology
1Automation system, e.g. Simatic S7 (Siemens)
2Computer with Web browser (e.g. Internet Explorer) for accessing the integrated device Web server or
computer with operating tool (e.g. FieldCare, DeviceCare, SIMATIC PDM) with COM DTM "CDI Communication
TCP/IP"
3Switch, e.g. Scalance X204 (Siemens)
4Measuring device
A0033719
94Endress+Hauser
Page 95
Proline Promass A 300
2
3
213
4
3
4
567
Service interfaceVia service interface (CDI-RJ45)
A point-to-point connection can be established to configure the device onsite. With the housing
open, the connection is established directly via the service interface (CDI-RJ45) of the device.
An adapter for RJ45 and the M12 connector is optionally available:
Order code for "Accessories", option NB: "Adapter RJ45 M12 (service interface)"
The adapter connects the service interface (CDI-RJ45) to an M12 connector mounted in the
cable entry. Therefore the connection to the service interface can be established via an M12
connector without opening the device.
A0027563
54Connection via service interface (CDI-RJ45)
1Computer with Web browser (e.g. Microsoft Internet Explorer, Microsoft Edge) for accessing the integrated
device Web server or with "FieldCare", "DeviceCare" operating tool with COM DTM "CDI Communication TCP/IP"
or Modbus DTM
2Standard Ethernet connecting cable with RJ45 connector
3Service interface (CDI-RJ45) of the measuring device with access to the integrated Web server
Via WLAN interface
The optional WLAN interface is available on the following device version:
Order code for "Display; operation", option G "4-line, illuminated; touch control + WLAN"
1Transmitter with integrated WLAN antenna
2Transmitter with external WLAN antenna
3LED lit constantly: WLAN reception is enabled on measuring device
4LED flashing: WLAN connection established between operating unit and measuring device
5Computer with WLAN interface and Web browser (e.g. Microsoft Internet Explorer, Microsoft Edge) for
6Mobile handheld terminal with WLAN interface and Web browser (e.g. Microsoft Internet Explorer, Microsoft
7Smart phone or tablet (e.g. Field Xpert SMT70)
accessing the integrated device Web server or with operating tool (e.g. FieldCare, DeviceCare)
Edge) for accessing the integrated device Web server or operating tool (e.g. FieldCare, DeviceCare)
Endress+Hauser95
A0034570
Page 96
Proline Promass A 300
FunctionWLAN: IEEE 802.11 b/g (2.4 GHz)
• Access point with DHCP server (default setting)
• Network
EncryptionWPA2-PSK AES-128 (in accordance with IEEE 802.11i)
Configurable WLAN channels1 to 11
Degree of protectionIP67
Available antennas• Internal antenna
• External antenna (optional)
In the event of poor transmission/reception conditions at the place of
installation.
Available as an accessory → 105.
Only one antenna active in each case!
Range• Internal antenna: typically 10 m (32 ft)
• External antenna: typically 50 m (164 ft)
Materials (external antenna)• Antenna: ASA plastic (acrylic ester-styrene-acrylonitrile) and nickel-
plated brass
• Adapter: Stainless steel and nickel-plated brass
• Cable: Polyethylene
• Connector: Nickel-plated brass
• Angle bracket: Stainless steel
Network integration
With the optional "OPC-UA Server" application package, the device can be integrated into an Ethernet
network via the service interface (CDI-RJ45 and WLAN) and communicate with OPC-UA clients. If
the device is used in this way, IT security must be considered.
Transmitters with an Ex de approval may not be connected via the service interface (CDI-RJ45)!
Order code for "Approval transmitter + sensor", options (Ex de):
BB, C2, GB, MB, NB
For permanent access to device data and for device configuration via the Web server, the device is
incorporated directly in a network via the service interface (CDI-RJ45). In this way, the device can be
accessed any time from the control station. The measured values are processed separately via the
inputs and outputs through the automation system.
96Endress+Hauser
Page 97
Proline Promass A 300
213
555
4
6
7
8
A0033618
1Automation system, e.g. Simatic S7 (Siemens)
2Ethernet switch
3Edge Gateway
4Cloud
5Measuring device
6Ethernet network
7Measured values via inputs and outputs
8Optional WLAN interface
Supported operating tools
The optional WLAN interface is available on the following device version:
Order code for "Display; operation", option G "4-line, illuminated, graphic display; touch control +
WLAN"
Special Documentation for the OPC-UA Server application package → 109.
Different operating tools can be used for local or remote access to the measuring device. Depending
on the operating tool used, access is possible with different operating units and via a variety of
interfaces.
Supported operating
tools
Web browserNotebook, PC or tablet
DeviceCare SFE100Notebook, PC or tablet
Operating unitInterfaceAdditional information
with Web browser
with Microsoft Windows
system
• CDI-RJ45 service
interface
• WLAN interface
• Ethernet-based
fieldbus (EtherNet/IP,
PROFINET)
• CDI-RJ45 service
interface
• WLAN interface
• Fieldbus protocol
Special Documentation for
device → 109
→ 107
Endress+Hauser97
Page 98
Proline Promass A 300
Supported operating
tools
FieldCare SFE500Notebook, PC or tablet
Device XpertField Xpert SFX
Operating unitInterfaceAdditional information
with Microsoft Windows
system
100/350/370
• CDI-RJ45 service
interface
• WLAN interface
• Fieldbus protocol
HART and
FOUNDATION Fieldbus
fieldbus protocol
→ 107
Operating Instructions
BA01202S
Device description files:
Use update function of
handheld terminal
Other operating tools based on FDT technology with a device driver such as DTM/iDTM or
DD/EDD can be used for device operation. These operating tools are available from the
individual manufacturers. Integration into the following operating tools, among others, is
supported:
• FactoryTalk AssetCentre (FTAC) by Rockwell Automation → www.rockwellautomation.com
• Process Device Manager (PDM) by Siemens → www.siemens.com
• Asset Management Solutions (AMS) by Emerson → www.emersonprocess.com
• FieldCommunicator 375/475 by Emerson → www.emersonprocess.com
• Field Device Manager (FDM) by Honeywell → www.honeywellprocess.com
• FieldMate by Yokogawa → www.yokogawa.com
• PACTWare → www.pactware.com
The associated device description files are available at: www.endress.com → Downloads
Web server
Thanks to the integrated Web server, the device can be operated and configured via a Web browser
and via a service interface (CDI-RJ45) or via a WLAN interface. The structure of the operating menu
is the same as for the local display. In addition to the measured values, status information on the
device is also displayed and allows the user to monitor the status of the device. Furthermore the
device data can be managed and the network parameters can be configured.
A device that has a WLAN interface (can be ordered as an option) is required for the WLAN
connection: order code for "Display; operation", option G "4-line, illuminated; touch control + WLAN".
The device acts as an Access Point and enables communication by computer or a mobile handheld
terminal.
Supported functions
Data exchange between the operating unit (such as a notebook for example) and the measuring
device:
• Upload the configuration from the measuring device (XML format, configuration backup)
• Save the configuration to the measuring device (XML format, restore configuration)
• Export event list (.csv file)
• Export parameter settings (.csv file or PDF file, document the measuring point configuration)
• Export the Heartbeat verification log (PDF file, only available with the "Heartbeat Verification"
application package)
• Flash firmware version for device firmware upgrade, for instance
• Download driver for system integration
• Visualize up to 1000 saved measured values (only available with the Extended HistoROM
application package → 104)
Web server special documentation → 109
HistoROM data management
The measuring device features HistoROM data management. HistoROM data management
comprises both the storage and import/export of key device and process data, making operation and
servicing far more reliable, secure and efficient.
When the device is delivered, the factory settings of the configuration data are stored as a
backup in the device memory. This memory can be overwritten with an updated data record, for
example after commissioning.
98Endress+Hauser
Page 99
Proline Promass A 300
Additional information on the data storage concept
There are different types of data storage units in which device data are stored and used by the device:
Device memoryT-DATS-DAT
Available data• Event logbook such as diagnostic events for example
• Parameter data record backup
• Device firmware package
• Driver for system integration for exporting via Web
server, e.g:
• GSD for PROFIBUS DP
• GSD for PROFIBUS PA
• GSDML for PROFINET
• EDS for EtherNet/IP
• DD for FOUNDATION Fieldbus
Storage location Fixed on the user interface board in the connection
compartment
Data backup
Automatic
• The most important device data (sensor and transmitter) are automatically saved in the DAT
modules
• If the transmitter or measuring device is replaced: once the T-DAT containing the previous device
data has been exchanged, the new measuring device is ready for operation again immediately
without any errors
• If the sensor is replaced: once the sensor has been replaced, new sensor data are transferred from
the S-DAT in the measuring device and the measuring device is ready for operation again
immediately without any errors
• If exchanging the electronics module (e.g. I/O electronics module): Once the electronics module
has been replaced, the software of the module is compared against the current device firmware.
The module software is upgraded or downgraded where necessary. The electronics module is
available for use immediately afterwards and no compatibility problems occur.
Manual
Additional parameter data record (complete parameter settings) in the integrated device memory
HistoROM backup for:
• Data backup function
Backup and subsequent restoration of a device configuration in the device memory HistoROM
backup
• Data comparison function
Comparison of the current device configuration with the device configuration saved in the device
memory HistoROM backup
• Measured value logging
("Extended HistoROM" order
option)
• Current parameter data record
(used by firmware at run time)
• Peakhold indicator (min/max
values)
• Totalizer values
Attachable to the user interface
board in the connection
compartment
• Sensor data: nominal diameter
etc.
• Serial number
• Calibration data
• Device configuration (e.g. SW
options, fixed I/O or multi I/O)
In the sensor plug in the transmitter
neck part
Data transfer
Manual
• Transfer of a device configuration to another device using the export function of the specific
operating tool, e.g. with FieldCare, DeviceCare or Web server: to duplicate the configuration or to
store in an archive (e.g. for backup purposes)
• Transmission of the drivers for system integration via Web server, e.g.:
• GSD for PROFIBUS DP
• GSD for PROFIBUS PA
• GSDML for PROFINET
• EDS for EtherNet/IP
• DD for FOUNDATION Fieldbus
Event list
Automatic
• Chronological display of up to 20 event messages in the events list
• If the Extended HistoROM application package (order option) is enabled: up to 100 event
messages are displayed in the events list along with a time stamp, plain text description and
remedial measures
• The events list can be exported and displayed via a variety of interfaces and operating tools e.g.
DeviceCare, FieldCare or Web server
Endress+Hauser99
Page 100
Proline Promass A 300
Data logging
Manual
If the Extended HistoROM application package (order option) is enabled:
• Record up to 1000 measured values via 1 to 4 channels
• User configurable recording interval
• Record up to 250 measured values via each of the 4 memory channels
• Export the measured value log via a variety of interfaces and operating tools e.g. FieldCare,
DeviceCare or web server
Certificates and approvals
Currently available certificates and approvals can be called up via the product configurator.
CE mark
RCM-tick symbol
Ex approval
The device meets the legal requirements of the applicable EU Directives. These are listed in the
corresponding EU Declaration of Conformity along with the standards applied.
Endress+Hauser confirms successful testing of the device by affixing to it the CE mark.
The measuring system meets the EMC requirements of the "Australian Communications and Media
Authority (ACMA)".
The measuring device is certified for use in hazardous areas and the relevant safety instructions are
provided in the separate "Safety Instructions" (XA) document. Reference is made to this document on
the nameplate.
The separate Ex documentation (XA) containing all the relevant explosion protection data is
available from your Endress+Hauser sales center.
ATEX/IECEx
Currently, the following versions for use in hazardous areas are available:
Ex db eb
CategoryType of protection
II1/2GEx db eb ia IIC T6…T1 Ga/Gb
II2GEx db eb ia IIC T6…T1 Gb
1)The following applies for sensors with nominal diameter DN 01: Ex db eb ia IIC T6…T1 Gb
1)
Ex db
CategoryType of protection
II1/2GEx db ia IIC T6…T1 Ga/Gb
II2GEx db ia IIC T6…T1 Gb
1)The following applies for sensors with nominal diameter DN 01: Ex db eb ia IIC T6…T1 Gb
1)
Ex ec
CategoryType of protection
II3GEx ec IIC T5…T1 Gc
Ex tb
CategoryType of protection
II2DEx tb IIIC T** °C Db
100Endress+Hauser
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.