Emerson Process Management 8732 User Manual

Page 1
Reference Manual
00809-0100-4663, Rev BA January 2010
Rosemount 8732
Integral Mount or Remote Mount Magnetic Flowmeter System with F
OUNDATION
fieldbus
www.rosemount.com
Page 2
Page 3
Reference Manual
NOTICE
00809-0100-4663, Rev BA January 2010
Rosemount 8732
Integral Mount or Remote Mount Magnetic Flowmeter System with
OUNDATION
F
Read this manual before working with the product. For personal and system safety, and for optimum product performance, make sure you thoroughly understand the contents before installing, using, or maintaining this product.
Rosemount Inc. has two toll-free assistance numbers:
Customer Central
Technical support, quoting, and order-related questions. United States - 1-800-999-9307 (7:00 am to 7:00 pm CST) Asia Pacific- 65 777 8211 Europe/ Middle East/ Africa - 49 (8153) 9390
North American Response Center
Equipment service needs. 1-800-654-7768 (24 hours—includes Canada) Outside of these areas, contact your local Rosemount representative.
fieldbus
The products described in this document are NOT designed for nuclear-qualified applications. Using non-nuclear qualified products in applications that require nuclear-qualified hardware or products may cause inaccurate readings.
For information on Rosemount nuclear-qualified products, contact your local Rosemount Sales Representative.
www.rosemount.com
Page 4
Page 5
Reference Manual
00809-0100-4663, Rev BA January 2010
Rosemount 8732

Table of Contents

SECTION 1 Introduction

SECTION 2 Installation

System Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-1
Safety Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2
Service Support . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2
Safety Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-1
Transmitter Symbols. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-2
Pre-Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-2
Mechanical Considerations. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-2
Environmental Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-3
Installation Procedures. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-3
Mount the Transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-3
Identify Options and Configurations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-4
Hardware Switches . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-4
Conduit Ports and Connections. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-5
Conduit Cables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-6
Electrical Considerations. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-6
Installation Category . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-7
Overcurrent Protection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-7
Connect Transmitter Power. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-7
Connect F
Transmitter Communication Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-8
Power Conditioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-8
Field Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-8
Transmitter Wiring Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-9
Sensor Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-11
Rosemount Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-11
Transmitter to Sensor Wiring. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-11
Conduit Cables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-12
Sensor to Remote Mount Transmitter Connections . . . . . . . . . . . . . . . . . .2-13
OUNDATION fieldbus Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-8

SECTION 3 Configuration

Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-1
Local Operator Interface. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-1
Basic Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-1
Data Entry . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-2
LOI Examples. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-2
Table Value Example . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-3
Select Value Example . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-3
Display Lock . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-3
Start Totalizer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-3
Stop Totalizer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-3
Reset Totalizer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-3
Diagnostic Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-5
Review. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-5
Process Variables. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-5
PV - Primary Variable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-6
PV -% Range. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-6
PV - Analog Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-6
Totalizer Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-6
Pulse Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-7
TOC-1
Page 6
Rosemount 8732
Reference Manual
00809-0100-4663, Rev BA
January 2010
Basic Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
Tag . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
Flow Units. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
Line Size. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-9
PV URV (Upper Range Value). . . . . . . . . . . . . . . . . . . . . . . . . . . 3-10
PV LRV (Lower Range Value) . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-10
Calibration Number. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-11
PV Damping . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-11

SECTION 4 Operation

SECTION 5 Sensor Installation

Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
Diagnostics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
Diagnostic Controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
Basic Diagnostics. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
Advanced Diagnostics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
Diagnostic Variables. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-8
Trims. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-9
Status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-11
Advanced Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-12
Detailed Setup. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-12
Additional Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-12
Display Language. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-13
Signal Processing. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-13
Device Info . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-15
Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-17
Block Mode: Target . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-18
Block Mode: Actual. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-18
Block Mode: Permitted . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-18
Block Mode: Normal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-18
Safety Messages. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
Sensor Handling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
Sensor Mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
Upstream/Downstream
Piping . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
Sensor Orientation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
Flow Direction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-6
Installation (Flanged Sensor). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-7
Gaskets . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-7
Flange Bolts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-7
Installation
(Wafer Sensor) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-10
Gaskets . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-10
Flange Bolts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-11
Installation
(Sanitary Sensor). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-12
Gaskets . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-12
Alignment and Bolting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-12
Grounding . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-12
Process Leak Protection (Optional). . . . . . . . . . . . . . . . . . . . . . . . . . 5-16
Standard Housing Configuration . . . . . . . . . . . . . . . . . . . . . . . . . 5-16
Relief Valves. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-17
Process Leak Containment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-17
TOC-2
Page 7
Reference Manual
00809-0100-4663, Rev BA January 2010
Rosemount 8732

SECTION 6 Maintenance and Troubleshooting

APPENDIX A Reference Data

APPENDIX B Approval Information

Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
Installation Check and Guide. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2
Diagnostic Messages. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-3
Transmitter Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-5
Quick Troubleshooting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-7
Step 1: Wiring Errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-7
Step 2: Process Noise . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-7
Step 3: Installed Sensor Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-7
Step 4: Uninstalled Sensor Tests. . . . . . . . . . . . . . . . . . . . . . . . . . 6-9
Functional Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-1
Foundation
Performance Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-5
Physical Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-7
Product Certifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .B-1
Approved Manufacturing Locations. . . . . . . . . . . . . . . . . . . . . . . . . . .B-1
European Directive Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .B-1
ATEX Directive. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .B-1
European Pressure Equipment Directive (PED) (97/23/EC) . . . . .B-1
Electro Magnetic Compatibility (EMC) (2004/108/EC) . . . . . . . . . .B-2
Low Voltage Directive (93/68/EEC) . . . . . . . . . . . . . . . . . . . . . . . .B-2
Low Voltage Directive (2006/95/EC) . . . . . . . . . . . . . . . . . . . . . . .B-2
Other important guidelines . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .B-2
IECEx Scheme. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .B-2
Hazardous Locations Product Approvals Offering. . . . . . . . . . . . . . . .B-3
Hazardous Location Certifications. . . . . . . . . . . . . . . . . . . . . . . . . . . .B-5
Transmitter Approval Information. . . . . . . . . . . . . . . . . . . . . . . . . .B-5
fieldbus Specifications . . . . . . . . . . . . . . . . . . . . . . . . . .A-4

APPENDIX C Diagnostics

Diagnostic Availability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C-1
Licensing and Enabling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C-2
Licensing the 8732 Diagnostics . . . . . . . . . . . . . . . . . . . . . . . . . . .C-2
Tunable Empty Pipe Detection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C-2
Tunable Empty Pipe Parameters . . . . . . . . . . . . . . . . . . . . . . . . . .C-2
Optimizing Tunable Empty Pipe. . . . . . . . . . . . . . . . . . . . . . . . . . .C-3
Troubleshooting Empty Pipe . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C-4
Ground/Wiring Fault Detection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C-4
Ground/Wiring Fault Parameters . . . . . . . . . . . . . . . . . . . . . . . . . .C-4
Troubleshooting Ground/Wiring Fault. . . . . . . . . . . . . . . . . . . . . . .C-5
Ground/Wiring Fault Functionality . . . . . . . . . . . . . . . . . . . . . . . . .C-5
High Process Noise Detection. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C-5
High Process Noise Parameters . . . . . . . . . . . . . . . . . . . . . . . . . .C-6
Troubleshooting High Process Noise . . . . . . . . . . . . . . . . . . . . . . .C-6
High Process Noise Functionality. . . . . . . . . . . . . . . . . . . . . . . . . .C-7
8714i Meter Verification. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C-8
Sensor Signature Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . .C-8
8714i Meter Verification Test Parameters . . . . . . . . . . . . . . . . . . .C-9
8714i Meter Verification Test Results Parameters . . . . . . . . . . . .C-10
Optimizing the 8714i Meter Verification . . . . . . . . . . . . . . . . . . . .C-13
Troubleshooting the 8714i Meter Verification Test. . . . . . . . . . . .C-14
8714i Meter Verification Functionality. . . . . . . . . . . . . . . . . . . . . .C-14
Rosemount Magnetic Flowmeter Calibration Verification Report . . .C-16
TOC-3
Page 8
Rosemount 8732
Reference Manual
00809-0100-4663, Rev BA
January 2010

APPENDIX D Digital Signal Processing

APPENDIX E Universal Sensor Wiring Diagrams

Safety Messages. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-1
Warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-1
Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-2
Auto Zero . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-2
Signal Processing. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-2
Rosemount Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-3
Rosemount 8705/8707/8711/8721 Sensors to
Rosemount 8732 Transmitter. . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-3
Rosemount 8701 Sensor to Rosemount 8732 Transmitter . . . . . . E-4
Connecting Sensors of Other Manufacturers. . . . . . . . . . . . . . . . . E-5
Brooks Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-6
Model 5000 Sensor to Rosemount 8732 Transmitter. . . . . . . . . . . E-6
Model 7400 Sensor to Rosemount 8732 Transmitter. . . . . . . . . . . E-7
Endress And Hauser Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-8
Endress and Hauser Sensor to Rosemount 8732 Transmitter. . . . E-8
Fischer And Porter Sensors. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-9
Model 10D1418 Sensor to Rosemount 8732 Transmitter . . . . . . . E-9
Model 10D1419 Sensor to Rosemount 8732 Transmitter . . . . . . E-10
Model 10D1430 Sensor (Remote) to
Rosemount 8732 Transmitter. . . . . . . . . . . . . . . . . . . . . . . . . . . . E-11
Model 10D1430 Sensor (Integral) to
Rosemount 8732 Transmitter. . . . . . . . . . . . . . . . . . . . . . . . . . . . E-12
Model 10D1465 and Model 10D1475 Sensors (Integral) to
8732 Transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-13
Fischer and Porter Sensor to Rosemount 8732 Transmitter . . . .E-14
Foxboro Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .E-15
Series 1800 Sensor to Rosemount 8732 Transmitter . . . . . . . . .E-15
Series 1800 (Version 2) Sensor to
Rosemount 8732 Transmitter. . . . . . . . . . . . . . . . . . . . . . . . . . . .E-16
Series 2800 Sensor to 8732 Transmitter . . . . . . . . . . . . . . . . . . .E-17
Foxboro Sensor to 8732 Transmitter. . . . . . . . . . . . . . . . . . . . . .E-18
Kent Veriflux VTC Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .E-19
Veriflux VTC Sensor to 8732 Transmitter. . . . . . . . . . . . . . . . . . .E-19
Kent Sensors. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .E-20
Kent Sensor to Rosemount 8732 Transmitter . . . . . . . . . . . . . . .E-20
Krohne Sensors. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .E-21
Krohne Sensor to Rosemount 8732 Transmitter . . . . . . . . . . . . .E-21
Taylor Sensors. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .E-22
Series 1100 Sensor to Rosemount 8732 Transmitter . . . . . . . . .E-22
Taylor Sensor to Rosemount 8732 Transmitter . . . . . . . . . . . . . .E-23
Yamatake Honeywell Sensors. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .E-24
Yamatake Honeywell Sensor to Rosemount 8732 Transmitter . .E-24
Yokogawa Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .E-25
Yokogawa Sensor to Rosemount 8732 Transmitter. . . . . . . . . . .E-25
Generic Manufacturer Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . .E-26
Generic Manufacturer Sensor to Rosemount 8732 Transmitter. .E-26
Identify the Terminals. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .E-26
Wiring Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .E-26
TOC-4
Page 9
Reference Manual
00809-0100-4663, Rev BA January 2010
Rosemount 8732

APPENDIX F Resource Block

APPENDIX G Transducer Block

APPENDIX H 375 Field Communicator Operation

Definition. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .F-1
Parameters and Descriptions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . F-1
Resource Block Errors. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . F-5
Modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . F-5
Alarm Detection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .F-6
Status Handling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .F-6
VCR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . F-6
Troubleshooting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .F-6
Definition. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .G-1
Parameters and Descriptions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .G-2
Flow-Specific Block Configuration Values . . . . . . . . . . . . . . . . . . . . . .G-3
Transducer Block Errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .G-4
Transducer Block Diagnostics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .G-5
Modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .G-5
Alarm Detection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .G-5
Status Handling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .G-5
Troubleshooting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .G-6
HandHeld Communicator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .H-1
Connections and Hardware. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .H-2
Basic Features. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .H-3
Action Keys. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .H-3
Alphanumeric and Shift Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . .H-4
Menus and Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .H-4
Main Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .H-5
Online Menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .H-5
Diagnostic Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .H-6
TOC-5
Page 10
Rosemount 8732
Reference Manual
00809-0100-4663, Rev BA
January 2010
TOC-6
Page 11
Reference Manual
00809-0100-4663, Rev BA January 2010

Section 1 Introduction

System Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 1-1
Safety Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .page 1-2
Service Support . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 1-2
Rosemount 8732

SYSTEM DESCRIPTION The Rosemount

sensor and transmitter, and measures volumetric flow rate by detecting the velocity of a conductive liquid that passes through a magnetic field.
There are four Rosemount magnetic flowmeter sensors:
Flanged Rosemount 8705
Flanged High-Signal Rosemount 8707
Wafer-Style Rosemount 8711
Sanitary Rosemount 8721
There are two Rosemount magnetic flowmeter transmitters:
Rosemount 8712
Rosemount 8732
The sensor is installed in-line with process piping — either vertically or horizontally. Coils located on opposite sides of the sensor create a magnetic field. Electrodes located perpendicular to the coils make contact with the process fluid. A conductive liquid moving through the magnetic field generates a voltage at the two electrodes that is proportional to the flow velocity.
The transmitter drives the coils to generate a magnetic field, and electronical ly conditions the voltage detected by the electrodes to provide a flow signal. T he transmitter can be integrally or remotely mounted from the sensor.
This manual is designed to assist in the installation and ope ration of the Rosemount 8732 Magnetic Flowmeter Transmitter and the Rosemount 8700 Series Magnetic Flowmeter Sensors.
®
8700 Series Magnetic Flowmeter System consists of a
www.rosemount.com
Page 12
Reference Manual
See “Safety Messages” on page D-1 for complete warning information.
00809-0100-4663, Rev BA
Rosemount 8732
January 2010

SAFETY MESSAGES Procedures and instructions in this manual may require special preca utions to

ensure the safety of the personnel performing the operations. Refer to the safety messages listed at the beginning of each section before performing any operations.
Attempting to install and operate the Rosemount 8705, 8707 High-Signal, 8711 or 8721 Magnetic Sensors with the Rosemount 8712 or 8732 Magnetic Flowmeter Transmitter without reviewing the instructions contained in this manual could result in personal injury or equipment damage.

SERVICE SUPPORT To expedite the return process outside the United States, contac t the nearest

Rosemount representative. Within the United States and Canada, call the North American Response
Center using the 800-654-RSMT (7768) toll-free number. The Response Center, available 24 hours a day, will assist you with any needed information or materials.
The center will ask for product model and serial numbers, and will provide a Return Material Authorization (RMA) number. The center will also ask for the name of the process material to which the product was last exposed.
Mishandling products exposed to a hazardous subst ance may result in death or serious injury. If the product being returned was exposed to a hazardous substance as defined by OSHA, a copy of the required Material Safety Data Sheet (MSDS) for each hazardous substance identified must be included with the returned goods.
The North American Response Center will detail the additional information and procedures necessary to return goods exposed to hazardous substances.
1-2
Page 13
Reference Manual
00809-0100-4663, Rev BA January 2010

Section 2 Installation

Safety Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .page 2-1
Transmitter Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 2-2
Pre-Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .page 2-2
Mechanical Considerations . . . . . . . . . . . . . . . . . . . . . . . .page 2-2
Environmental Considerations . . . . . . . . . . . . . . . . . . . . . page 2-3
Installation Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . page 2-3
Sensor Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 2-11
This section covers the steps required to physically install the magnetic flowmeter. Instructions and procedures in this section may require special precautions to ensure the safety of the personnel performing the operations. Please refer to the following safety messages before performing any operation in this section.
Rosemount 8732

SAFETY MESSAGES This symbol is used throughout this manual to indicate that special attention

to warning information is required.
Failure to follow these installation guidelines could result in death or serious injury: Installation and servicing instructions are for use by qualified personnel only. Do not perform
any servicing other than that contained in the operating instructions, unless qualified. Verify that the operating environment of the sensor and transmitter is consistent with the appropriate hazardous area approval.
Do not connect a Rosemount 8732 to a non-Rosemount sensor that is located in an explosive atmosphere.
Explosions could result in death or serious injury: Installation of this transmitter in an explosive environment must be in accordance with the
appropriate local, national, and international standards, codes, and practices. Please review the approvals section of the 8732 reference manual for any restrictions associated with a safe installation.
Before connecting a handheld communicator in an explosive atmosphere, make sure the instruments in the loop are installed in accordance with intrinsically safe or non-incendive field wiring practices.
Electrical shock can result in death or serious injury Avoid contact with the leads and terminals. High voltage that may be present on leads can
cause electrical shock.
www.rosemount.com
Page 14
Rosemount 8732
Reference Manual
00809-0100-4663, Rev BA
January 2010
The sensor liner is vulnerable to handling damage. Never place anything through the sensor for the purpose of lifting or gaining leverage. Liner damage can render the sensor useless.
To avoid possible damage to the sensor liner ends, do not use metallic or spiral-wound gaskets. If frequent removal is anticipated, take precautions to protect the liner ends. Short spool pieces attached to the sensor ends are often used for protection.
Correct flange bolt tightening is crucial for proper sensor operation and life. All bolts must be tightened in the proper sequence to the specified torque limits. Failure to observe these instructions could result in severe damage to the sensor lining and possible sensor replacement.
Emerson Process Management can supply lining protectors to prevent liner damage during removal, installation, and excessive bolt torquing.

TRANSMITTER SYMBOLS

Caution symbol — check product documentation for details Protective conductor (grounding) terminal

PRE-INSTALLATION Before installing the Rosemount 8732 Magnetic Flowmeter Transmitter, there

are several pre-installation steps that should be completed to make the installation process easier:
Identify the options and configurations that apply to your application
Set the hardware switches if necessary
Consider mechanical, electrical, an d en vir onm en tal req uir eme nts

MECHANICAL CONSIDERATIONS

The mounting site for the 8732 transmitter should provide enough room for secure mounting, easy access to conduit ports, full openin g of the transmitter covers, and easy readability of the LOI screen (see Figure 2-1). The transmitter should be mounted in a manner that prevent s moisture in conduit from collecting in the transmitter.
If the 8732 is mounted remotely from the sensor, it is not subject to limitations that might apply to the sensor.
2-2
Page 15
Reference Manual
5.82
(148)
6.48 (165)
7.49 (190) LOI Cover
4.97
(126)
8.81
(224)
3.00 (76)
3.07 (78)
4.97
(126)
1
/2”-14 NPT Electrical
Conduit Connections
(2 places with a 3rd
optional)
1
/2”-14 NPT Remote Junction
Box Conduit Connections (2
places)
00809-0100-4663, Rev BA January 2010
Figure 2-1. Rosemount 8732 Dimensional Drawing
Rosemount 8732

ENVIRONMENTAL CONSIDERATIONS

To ensure maximum transmitter life, avoid temperature extremes and vibration. Typical problem areas include:
high-vibration lines with integrally mounted transmitters
warm-climate installations in direct sunlight
outdoor installations in cold climates.
Remote-mounted transmitters may be installed in the control room to protect the electronics from a harsh environment and provides easy access for

INSTALLATION PROCEDURES

configuration or service. Rosemount 8732 transmitters require external power so there must be access
to a suitable power source.
Rosemount 8732 installation includes both detailed mecha nical and electrical installation procedures.

Mount the Transmitter Remote-mounted transmitters may be mounted on a pipe up to two inches in

diameter or against a flat surface.
Pipe Mounting
To mount the transmitter on a pipe:
1. Attach the mounting bracket to the pip e using the m ounting har dware.
2. Attach the 8732 to the mounting bracket using the mounting screws.
2-3
Page 16
Rosemount 8732
Reference Manual
00809-0100-4663, Rev BA
January 2010
Surface Mounting
To surface mount the transmitter:
1. Attach the 8732 to the mounting location using the mounting screws.

Identify Options and Configurations

The standard application of the Rosemount 8732 includes a FOUNDATION fieldbus output. Be sure to identify options and configurations that apply to your situation, and keep a list of them nearby for consideration during the installation and configuration procedures.

Hardware Switches The 8732 electronics board is equipped with two user-selectable hardware

switches. These switches set the Transmitter Security and Simulate Mode. The standard configuration for these switches when shipped from the factory are as follows:
Transmitter Security: OFF Simulate Mode OFF
Definitions of these switches and their functions are provided below. If you determine that the settings must be changed, see below .
Transmitter Security
The security switch on the 8732 allows the user to lock out any configuration changes attempted on the transmitter. No changes to the configuration are allowed when the switch is in the ON position. The flow rate indication function remains active at all times.
With the switch in the ON position, you may still access and review any of the operating parameters and scroll through the available choices, but no actual data changes are allowed. Transmitter security is set in the OFF position when shipped from the factory.
Simulate Mode
The Simulate Mode switch is used in conjunction with the Analog Input (AI) function block. The switch is used to enable flow measurement simulation. To enable the simulate enable feature, the switch must transition from OFF to ON after power is applied to the transmitter, preventing the transmitter from being accidentally left in simulate mode. Simulate Mode is set in the OFF position when shipped from the factory.
Changing Hardware Switch Settings
In most cases, it is not necessary to change the setting of the hardware switches. If you need to change the switch settings, complete the steps below:
NOTE
The hardware switches are located on the top side of the electronics board and changing their settings requires opening the electronics housing. If possible, carry out these procedures away from the plant environment in order to protect the electronics.
2-4
Page 17
Reference Manual
00809-0100-4663, Rev BA January 2010
Figure 2-2. Rosemount 8732 Electronics Board and Hardware Switches
Rosemount 8732
1. Disconnect power to the transmitter.
2. Remove electronics cover.
3. Remove display if applicable.
4. Identify the location of each switch (see Figure 2-2).
5. Change the setting of the desired switche s with a sm all scr ewd r iver.
6. Replace the electronics cover.

Conduit Ports and Connections

Both the sensor and transmitter junction boxes have ports for 1/2-inch NPT conduit connections, with optional CM20 and PG 13.5 connections available. These connections should be made in accordance with national, local or plant electrical codes. Unused ports should be sealed with metal plugs and PTFE tape or other thread sealant. Connections should also be made in accordance with area approval requirements, see examples below for details. Proper electrical installation is necessary to prevent errors due to electrical noise and interference. Separate conduit s are not necessary for the coil drive and signa l cables connecting the transmitter to the sensor, but a dedicated conduit line between each transmitter and sensor is required. A shielded cable must be used.
Example 1: Installing flanged sensors into an IP68 area. Sensors must be installed with IP68 cable glands and cable to maintain IP68 rating. Unused conduit connections must be properly sealed to prevent water ingress. For added protection, dielectric gel can be used to pot the sensor terminal block.
Example 2: Installing flowmeters into explosion proof/flameproof areas. Conduit connections and conduit must be rated for use in the hazardous area to maintain flowmeter approval rating.
2-5
Page 18
Reference Manual
NOTE Dimensions are in inches (millimeters).
1.00 (26)
Cable Shield
00809-0100-4663, Rev BA
Rosemount 8732
January 2010

Conduit Cables Run the appropriate size cable through the conduit connections in your

magnetic flowmeter system. Run the power cable from the power source to the transmitter . Do n ot run power cab les and outpu t signal cables in the same conduit. For remote mount installations, run the coil drive and electrode cables between the flowmeter and transmitter. Refer to Electrical Considerations for wire type. Prepare the ends of the coil drive and electrode cables as shown in Figure 2-3. Limit the unshielded wire length to 1-in. on both the electrode and coil drive cables. Excessive lead length or failure to connect cable shields can create electrical noise resulting in unstable meter readings.
Figure 2-3. Cable Preparation Detail

Electrical Considerations Before making any electrical connections to the Rosemount 8732, consider

the following standards and be sure to have th e proper power supply, conduit, and other accessories. When prepar ing all wir e conne ctions, remove only the insulation required to fit the wire completely under the terminal connection. Removal of excessive insulation may result in an unwanted electrical short to the transmitter housing or other wire connections.
Transmitter Input Power
The 8732 transmitter is designed to be powered b y 90-250 V AC, 50–60 Hz or 12–42 V DC. The eighth digit in the transmitter model number designates the appropriate power supply requirement.
Model Number Power Supply Requirement
1 90-250 V AC 2 12-42 V DC
Supply Wire Temperature Rating
Use 12 to 18 AWG wire. For connections in ambient temperatures exceeding 140 °F (60 °C), use wire rated to at least 194 °F (90 °C).
Disconnects
Connect the device through an external disconnect or circuit breaker. Clearly label the disconnect or circuit breaker and locate it near the transmitter.
Requirements for 90-250 V AC Power Supply
2-6
Wire the transmitter according to national, local, and plant electrical requirements for the supply voltage. In addition, follow the supply wire and disconnect requirements on page2-7.
Page 19
Reference Manual
MaximumResis cetan
SupplyVoltage 12VDC
1amp
--------------------------------------------------------------------=
Power Supply (Volts)
I = 10/V I = Supply current requirement (Amps) V = Power supply voltage (Volts)
Supply Current (Amps)
12 18
24
30
36
42
0.2
0.3
0.4
0.5
0.6
0.7
0.8
0.9
00809-0100-4663, Rev BA January 2010
Figure 2-4. Supply Current versus Input Voltage
Rosemount 8732
Requirements for 12-42 V DC Power Supply
Units powered with 12-42 V DC may draw up to 1 amp of cu rren t. As a result, the input power wire must meet certain gauge requirements.
Figure 2-4 shows the supply current for each corresponding supply voltage . For combinations not shown, you can calculate the maximum distance given the supply current, the voltage of the source, and the minimum start-up voltage of the transmitter, 12 V DC, using the following equation:

Installation Category The installation category for the Rosemount 87 32 is (overvoltage) Category II. Overcurrent Protection The Rosemount 8732 Flowmeter Transmitter requires overcurrent protection

of the supply lines. Maximum ratings of overcurrent devices are as follows:
Power System Fuse Rating Manufacturer
110 V AC 250 V; 1 Amp, Quick Acting Bussman AGCI or Equivalent 220 V AC 250 V; 2 Amp, Quick Acting Bussman AGCI or Equivalent
42 V DC 50 V, 3 Amp, Quick Acting Bussman AGCI or Equivalent

Connect Transmitter Power

To connect power to the transmitter, complete the following steps.
1. Ensure that the power source and connecting cable meet the requirements outlined on page 2-8.
2. Turn of f the power source.
3. Open the power terminal cover.
4. Run the power cable through the conduit to the transmitter.
5. Connect the power cable leads as shown in Figure 2-5. a. Connect AC Neutral or DC- to terminal 9. b. Connect AC Line or DC+ to terminal 10. c. Connect AC Ground or DC Ground to the ground screw mounted
inside the transmitter enclosure.
2-7
Page 20
Rosemount 8732
AC Line or DC +
Transmitter Power Cable
AC Neutral or DC –
AC or DC Ground
See “Safety Messages” on page 2-1 for complete warning information.
Figure 2-5. AC Transmitter Power Connections
Reference Manual
00809-0100-4663, Rev BA
January 2010

Connect FOUNDATION fieldbus Wiring

Transmitter Communication Input

Power Conditioning Each fieldbus power supply requires a power conditioner to decouple the

Field Wiring Power independent of the coil power supply must be supplied for FOUNDATION

Table 2-1. Ideal Cable Specifications for Fieldbus Wiring
The FOUNDATION fieldbus signal provides the output information from the transmitter.
The FOUNDATION fieldbus communication requires a minimum of 9 V dc and a maximum of 32 V dc at the transmitter communication terminals.
NOTES
Do not exceed 32 V dc at the transmitter communication terminals.
Do not apply ac line voltage to the transmitter
communication terminals.
Improper supply voltage can damage the transmitter.
power supply output from the fieldbus wiring segment.
fieldbus communications. Use shielded, twisted pair for best results. For new installations or to get maximum performance, twisted pair cable designed especially for fieldbus should be used. Table 2-1 details cable characteristics and ideal specifications.
Characteristic Ideal Specification
Impedance 100 Ohms ± 20% at 31.25 kHz
Wire Size 18 AWG (0,8 mm2)
Shield Coverage 90%
Attenuation 3 db/km
Capacitive Unbalance 2 nF/km
2-8
Page 21
Reference Manual
–FF signal
+FF signal
00809-0100-4663, Rev BA January 2010
Rosemount 8732
NOTE
The number of devices on a fieldbus segment is limited by the power supply voltage, the resistance of the cable, and the amount of current drawn by each device.

Transmitter Wiring Connection

Figure 2-6. F Signal Connections
OUNDATION fieldbus
To connect the 8732 to the FOUNDATION fieldbus (FF) segment, complete the following steps.
1. Ensure that the power source and connecting cable meet the requirements outlined above and in “Field Wiring” on page 2-8.
2. Turn of f the transmitter and power sources.
3. Run the F
4. Connect -FF to Terminal 1.
5. Connect +FF to Terminal 2.
NOTE
Foundation fieldbus signal wiring for the 8732 is not polarity sensitive.
Refer to Figure 2-6 on page 2-9.
OUNDATION fieldbus cable into the transmitter.
2-9
Page 22
Rosemount 8732
Integrated
Power
Conditioner
and Filter
Terminators
6234 ft (1900 m) max
(depending upon cable
characteristics)
Fieldbus
Segment
(Spur)
(Trunk)
(The power supply, filter, first terminator, and configuration tool are typically located in the control room.)
*Intrinsically safe installations may allow fewer devices per I.S. barrier.
Power
Supply
FOUNDATION
Fieldbus
Configuration
Tool
(Spur)
Devices 1 through 11*
Figure 2-7. Rosemount 8732 Transmitter Field Wiring
Reference Manual
00809-0100-4663, Rev BA
January 2010
2-10
Page 23
Reference Manual
Coil Drive and Electrode Cables
Power
Power
Outputs
Outputs
Coil Drive and Electrode Cables
Power
Outputs
Power
Outputs
00809-0100-4663, Rev BA January 2010
Rosemount 8732

SENSOR CONNECTIONS This section covers the steps required to physically install the transmitter

including wiring and calibration.

Rosemount Sensors To connect the transmitter to a non-Rosemount sensor, refer to the

appropriate wiring diagram in “Universal Sensor Wiring Diagrams” on page E-1. The calibration procedure listed is not required for use with Rosemount sensors.

Transmitter to Sensor Wiring

Figure 2-8. Conduit Preparation
Correct Incorrect
Flanged and wafer sensors have two conduit ports as shown in Figur e 2-8. Either one may be used for both the coil drive and electrode cables. Use the stainless steel plug that is provided to seal the unused conduit port. Use Teflon tape or thread sealant appropriate for the installation when sealing the conduit.
A single dedicated conduit run for the coil drive and electrode cables is needed between a sensor and a remote transmitter. Bundled cables in a single conduit are likely to create interference and noise problems in your system. Use one set of cables per conduit run. See Figure 2-8 for proper conduit installation diagram and Table 2-2 for recommended cable. For integral and remote wiring diagrams refer to Figure 2-10.
Table 2-2. Cable Requirements
Description Units Part Number
Signal Cable (20 AWG) Belden 8762, Alpha 2411 equivalent ft
Coil Drive Cable (14 AWG) Belden 8720, Alpha 2442 equivalent ft
Combination Signal and Coil Drive Cable (18 AWG)
(1) Combination signal and coil drive cable is not recommended for high-signal magmeter system. For remote mount installa tions, combination signal and coil
(1)
drive cable should be limited to less than 330 ft. (100 m).
08712-0061-0001
m
m ft m
08712-0061-0003 08712-0060-0001 08712-0060-0003 08712-0752-0001 08712-0752-0003
2-11
Page 24
Reference Manual
1.00 (26)
NOTE Dimensions are in inches (millimeters).
Cable Shield
00809-0100-4663, Rev BA
Rosemount 8732
January 2010
Rosemount recommends using the combination signal and coil drive for N5, E5 approved sensors for optimum performance.
Remote transmitter installations require equal lengths of signal and coil drive cables. Integrally mounted transmitters are factory wired and do not require interconnecting cables.
Lengths from 5 to 1,000 feet (1.5 to 300 meters) may be specified, and will be shipped with the sensor.

Conduit Cables Run the appropriate size cable through the conduit connections in your

magnetic flowmeter system. Run the power cable from the power source to the transmitter. Run the coil drive and electrode cables between the sensor and transmitter .
Prepare the ends of the coil drive and electrode cables as shown in Figure 2-9. Limit the unshielded wire length to 1-inch on both the electrode and coil drive cables.
NOTE
Excessive lead length or failure to connect cable shields ca n create electrical noise resulting in unstable meter readings.
Figure 2-9. Cable Preparation Detail
2-12
Page 25
Reference Manual
00809-0100-4663, Rev BA January 2010
Rosemount 8732

Sensor to Remote Mount Transmitter Connections

Figure 2-10. Wiring Diagram
Connect coil drive and electrode cables as shown in Figure 2-10.
Do not connect AC power to the sensor or to terminals 1 an d 2 of the transmitter, or replacement of the electronics board will be necessary.
Rosemount 8732 Transmitter
11 2 2
17 17 18 18 19 19
Rosemount 8705/8707/8711/8721
Sensors
2-13
Page 26
Rosemount 8732
Reference Manual
00809-0100-4663, Rev BA
January 2010
2-14
Page 27
Reference Manual
00809-0100-4663, Rev BA January 2010
Rosemount 8732

Section 3 Configuration

Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 3-1
Local Operator Interface . . . . . . . . . . . . . . . . . . . . . . . . . . page 3-1
Basic Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 3-1
LOI Examples . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .page 3-2
Diagnostic Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 3-5
Process Variables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .page 3-5
Basic Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .page 3-7

INTRODUCTION This section covers basic operation, software functionality, and configuration

procedures for the Rosemount 8732 Magnetic Flowmeter Transmitter. For information on connecting anothe r m anufacturer’s flowtube sensor, refer to “Universal Sensor Wiring Diagrams” on page E-1.
The Rosemount 8732 features a full range of software functions for configuration of output from the transmitter. Software functions are accessed through the LOI, AMS, a Handheld Communicator, or a control system. Configuration variables may be changed at any time and specific instructions are provided through on-screen instructions.
Table 3-1. Parameters
Basic Set-up Parameters Page
Review page 3-5
Process Variables page 3-5
Basic Setup page 3-7
Flow Units page 3-7
Range Values page 3-10
PV Sensor/Flowtube Sensor Calibration Number page 3-11
Totalizer Setup page 3-6

LOCAL OPERATOR INTERFACE

The optional Local Operator Interface (LOI) provides an operator communications center for the 8732. By using th e LO I, th e op er at or can access any transmitter function for changing configuration parameter settings, checking totalized values, or other functions. The LOI is integral to the transmitter electronics.

BASIC FEATURES The basic features of the LOI include 4 navigational arrow keys that are used

to access the menu structure. See Figure 3-1.
www.rosemount.com
Page 28
Rosemount 8732
Figure 3-1. Local Operator Interface Keypad
Reference Manual
00809-0100-4663, Rev BA
January 2010

Data Entry The LOI keypad does not have numerical keys. Numerical data is entered by

the following procedure.
1. Access the appropriate function.
2. Use the RIGHT ARROW key to move to the value to change.
3. Use the UP and DOWN ARROWS to change the highlighted value. For numerical data, toggle through the digits 0–9, decimal point, and
dash. For alphabetical data, toggle through the letters of the alphabet A–Z, digits 0–9, and the symbols ,&, +, -, *, /, $, @,%, and the blank space.
4. Use the RIGHT ARRO WS to highlight other digit s you want to change and change them.
5. Press “E” (the lef t arrow key) when all cha nges are compl ete to save the entered values.

LOI EXAMPLES Use the DOWN ARROW to access the menu structure in Table 3-2. Use the

ARROW KEYS to select the desired parameters to review/change.
Parameters are set in one of two ways, Table Values or Select Values.
Table Values:
Parameters such as units, that are available from a predefined list
Select Values:
Parameters that consist of a user-created number or character string, such as calibration number; values are entered one character at a time using the ARROW KEYS.
3-2
Page 29
Reference Manual
00809-0100-4663, Rev BA January 2010

Table Value Example Setting the TUBE SIZE:

1. Press the DOWN arrow to access the menu.
2. Select line size from the Basic set-up menu.
3. Press the UP or DOWN arrow to increase/decrease (incrementally) the tube size to the next value.
4. When you reach the desired size, press “E” (the left arrow).
5. Set the loop to manual if necessary, and press “E” again.
After a moment, the LCD will display the new tube size and the maximum flow rate.

Select Value Example Changing the ANALOG OUTPUT RANGE:

1. Press the DOWN arrow to access the menu.
2. Using the arrow keys, select PV URV from the Basic Setup menu.
3. Press RIGHT arrow key to posit ion the cursor.
4. Press UP or DOWN to set the number.
5. Repeat steps 2 and 3 until desired number is displayed.
6. Press “E”.
Rosemount 8732
After a moment, the LCD will display the new analog output range.

Display Lock The display can be locked to prevent unintentional configuration changes.

The display lock can be activated through a HART communication device, or by holding the UP arrow for 10 seconds. When the display lock is activated, DL will appear in the lower left hand corner of the display. To deactivate the display lock (DL), hold the UP arrow for 10 seconds. Once deactivated, the DL will no longer appear in the lower left hand corner of the display.

Start Totalizer To start the totalizer, press the DOWN arrow to display the totalizer screen

and press “E” to begin totalization. A symbol will flash in the lower right hand corner indicating that the meter is totalizing.

Stop Totalizer To stop the totalizer, press the DOWN arrow to display the totalizer screen

and press “E” to end totalization. The flashing symbol will no longer display in the lower right hand corner indicating that the meter has stopped totalizing.

Reset Totalizer To reset the totalizer, press the DOWN arrow to display the totalizer screen

and follow the procedure above to stop totalization. Once totalization has stopped, press the RIGHT arrow key to reset the NET total value to zero.
To reset the gross total value, you must change the line size. See “Line Size” on page 3-9 for details on how to change the line size.
3-3
Page 30
Rosemount 8732
Diag Controls Basic Diag A dvanced D iag Variables Trims Status
E mpty P ipe Process Noise G round/W i ri ng Elec Temp
Self Test A O L oop Test Pulse Out Test E mpty P ipe Elec Temp
G round/W i ri ng Process Noise 8714i 4-20 mA V erif y Licensing
R un 8714i V iew Results Tube Signature Test Criteria M easur ements
Values R e-Signature Recall V alues
Coil R esist C oil S ignature E lectr ode R es
No Flow Flowing, Full E mpty P ipe
Coil R esist C oil S ignature E lectr ode R es
E mpty P ipe Elec Temp L ine N oise 5Hz SN R 37Hz SN R Signal Power 8714i R esults
D/A Trim Digital Trim A uto Z ero Universal Trim
4-20 mA V erif y View Results
Tag Flow Units L ine S ize PV UR V PV L R V Cal Number PV Damping
C oi l F requency Proc Density PV LSL PV USL PV Min Span
Analog Pulse DI/DO Config Totalizer Reverse Flow HART
PV UR V PV L R V Alarm Type Test
Pulse Scaling Pulse W idth Pulse M ode Test
DI 1 DO 2
Totalize Units Total Dis
play
Burst Mode B urst C ommand
F lange T ype F lange M atl E lectr ode T ype E lectr ode M atl L iner Material
Software Rev Final A smbl #
Tag Description Message Device ID PV Sensor S/N Flowtube Tag Revision Num M ateri al s
Operating Mode SP Config C oi l F requenc y PV Damping Lo-Flow Cuto
Flow Display Total Display L anguage
More Params Output C onfi g LOI Config Si g Processing Device Info
PV Units Special Units Totalize Units
Diagnostics
Basic Setup
Detailed S etup
Ta ble 3-2. LOI Menu Tree
Reference Manual
00809-0100-4663, Rev BA
January 2010
3-4
Page 31
Reference Manual
00809-0100-4663, Rev BA January 2010
Rosemount 8732

DIAGNOSTIC MESSAGES

The following error messages may appear on the LOI screen. See Table 6-4 on page 6-5 for potential causes and corrective actions for these errors:
Electronics Failure
Coil open circuit
Digital trim failure
Auto zero failure
Auto trim failure
Flowrate > sensor limit
Analog out of range
PZR activated
Electronics Temp Fail
Pulse out of range
Empty pipe
Reverse flow
Electronics temp out of range
The following error messages may appear on the LOI screen. See Table 6-4 on page 6-5 for potential causes and corrective actions for these errors:
High Process Noise
Grounding/Wiring Fault
4-20 mA Loop Verification Failed
8714i Failed

Review The 8732 includes a capability that enables you to review the configuration

Fast Keys 1, 5
variable settings. The flowmeter configuration parameters set at the factory should be reviewed
to ensure accuracy and compatibility with your particular application of the flowmeter.
NOTE
If you are using the LOI to review variables, each variable must be accessed as if you were going to change its setting. The value displayed on the LOI screen is the configured value of the variable.

PROCESS VARIABLES The process variables measure flow in several ways that reflect your needs

Fast Keys 1, 1
and the configuration of your flowmeter. When commissioning a flowmeter, review each process variable, its function and output, and take co rrective action if necessary before using the flowmete r in a proc es s application
Process Variable (PV) – The actual measured flow rate in the line. Use the Process Variable Units function to select the units for yourapplication.
Percent of Range – The process variable as a percentage of the Analog Output range, provides an indication where the current flow of the meter is within the configured range of the flowmeter. For example, the Analog Output range may be defined as 0 gal/min to 20 gal/min. If the measured flow is 10 gal/min, the percent of range is 50 percent.
3-5
Page 32
Reference Manual
00809-0100-4663, Rev BA
Rosemount 8732
Analog Output – The analog output variable provides the analog value fo r the flow rate. The analog output refers to the industry stand ard output in th e 4–20 mA range. The analog output and 4-20 mA loop can be verified using the Analog Feedback diagnostic capability internal to the transmitter (See “8714i Meter Verification” on page C-8).

Totalizer Setup – Provides a reading of the total flow of the flowmeter since the totalizer was last reset. The totalizer value should be zero during commissioning on the bench, and the units should reflect the volume units of the flow rate. If the totalizer value is not zero, it may need to be reset. This function also allows for configuration of the totalizer parameters.

Pulse Output – The pulse output variable provides the pulse value for the flow rate.
January 2010

PV - Primary Variable The Primary Variable shows the current measured flow rate. This value

Fast Keys 1, 1, 1
determines the analog output from the transmitter.

PV -% Range The PV% Range shows where in the flow range the current flow value is as a

Fast Keys 1, 1, 2
percentage of the configured span.

PV - Analog Output The PV Analog Output displays the mA output of the transmitter

Fast Keys 1, 1, 3
corresponding to the measured flow rate.
Totalizer Setup The Totalizer Setup menu allows for the viewing and configuration of the
Fast Keys 1, 1, 4
totalizer parameters.
Totalizer Units
Fast Keys 1, 1, 4, 1
Totalizer units allow for the configuration of the units that the totalized value will be displayed as. These units are independent of the flow units.
Measured Gross Total
Fast Keys 1, 1, 4, 2
Measured gross total provides the output reading of the tot alizer. This value is the amount of process fluid that has passed through the flowmeter since the totalizer was last reset.
NOTE
To reset the measured gross total value, the line size must be changed.
Measured Net Total
Fast Keys 1, 1, 4, 3
Measured net total provides the output reading of the totalizer. This value is the amount of process fluid that has passed through the flowmeter since the totalizer was last reset. When re verse flow is enabled, the net tot al represent s the difference between the total flow in the forward dir ection less the tot al flow in the reverse direction.
3-6
Page 33
Reference Manual
00809-0100-4663, Rev BA January 2010
Rosemount 8732
Measured Reverse Total
Fast Keys 1, 1, 4, 4
Measured reverse total provides the output reading of th e tota lizer. This value is the amount of process fluid that has passed through the flowmeter in the reverse direction since the totalizer was last reset. This value is only totalized when reverse flow is enabled.
Start Totalizer
Fast Keys 1, 1, 4, 5
Start totalizer starts the totalizer counting from its current value.
Stop Totalizer
Fast Keys 1, 1, 4, 6
Stop totalizer in terrupts the totalizer count until it is restarted again. This feature is often used during pipe cleaning or other mainte na nce operations.
Reset Totalizer
Fast Keys 1, 1, 4, 7
Reset totalizer resets the net totalizer value to zero. The totalizer must be stopped before resetting.
NOTE
The totalizer value is saved in the Non-Volatile memory of the electronics every three seconds. Should power to the transmitter be interrupted, the totalizer value will start at the last saved value when power is re-applied.

Pulse Output The Pulse Output displays the current value of the pulse signal.

Fast Keys 1, 1, 5

BASIC SETUP The basic configuration functions of the Rosemount 8732 must be set for all

Fast Keys 1, 3
applications of the transmitter in a magnetic flowmeter system. If your application requires the advanced functionality features of the Rosemount 8732, see Section 4 "Operation" of this manual.
Tag Tag is the quickest and shortest way of identifying and distinguishing between
Fast Keys 1, 3, 1
transmitters. Transmitters can be tagged according to the requirements of your application. The tag may be up to eight characters long.

Flow Units Flow Units set the output units for the Primary Variable which controls the

Fast Keys 1, 3, 2
analog output of the transmitter.
Primary Variable Units
Fast Keys 1, 3, 2, 1
The Primary Variable Units specifies the format in which the flow rate will be displayed. Units should be selected to meet your particular metering needs.
3-7
Page 34
Rosemount 8732
Reference Manual
00809-0100-4663, Rev BA
January 2010
Options for Flow Rate Units
• ft/sec • B31/sec (1 Barrel = 31.5 gallons)
•m/sec • B31/min (1 Barrel = 31.5 gallons)
• gal/sec • B31/hr (1 Barrel = 31.5 gallons)
• gal/min • B31/day (1 Barrel = 31.5 gallons)
• gal/hr • lbs/sec
• gal/day •lbs/min
•l/sec •lbs/hr
•l/min • lbs/day
• l/hr • kg/sec
• l/day •kg/min
3
•ft
/sec • kg/hr
•ft3/min • kg/day
3
/hr • (s)tons/min
•ft
•ft3/day • (s)tons/hr
3
/sec • (s)tons/day
•m
•m3/min • (m)tons/min
3
/hr • (m)tons/hr
•m
•m3/day • (m)tons/day
• Impgal/sec • Special (User Defined, see
• Impgal/min
• Impgal/hr
• Impgal/day
• B42/sec (1 Barrel = 42 gallons)
• B42/min (1 Barrel = 42 gallons)
• B42/hr (1 Barrel = 42 gallons)
• B42/day (1 Barrel = 42 gallons)
“Special Units” on page 3-8)
3-8
Special Units
Fast Keys 1, 3, 2, 2
The Rosemount 8732 provides a selection of standard u nit configurations that meet the needs of most applications (see “Flow Units” on page 3-7). If your application has special needs and the standard configurations do not apply, the Rosemount 8732 provides the flexibility to configure the transmitter in a custom-designed units format using the special units variable.
Special Volume Unit
Fast Keys 1, 3, 2, 2, 1
Special volume unit enables you to display the volume unit format to which you have converted the base volume units. For example, if the special units are abc/min, the special volume variable is abc. The volume units variable is also used in totalizing the special units flow.
Page 35
Reference Manual
00809-0100-4663, Rev BA January 2010
Rosemount 8732
Base Volume Unit
Fast Keys 1, 3, 2, 2, 2
Base volume unit is the unit from which the conversion is being made. Set this variable to the appropriate option.
Conversion Number
Fast Keys 1, 3, 2, 2, 3
The special units conversion number is used to convert base units to special units. For a straight conversion of volume units from one to another, the conversion number is the number of base units in the new unit. For example, if you are converting from gallons to barrels and there are 31 gallons in a barrel, the conversion factor is 31.
Base Time Unit
Fast Keys 1, 3, 2, 2, 4
Base time unit provides the time unit from which to calculate the special units. For example, if your special units is a volume per minute, select minutes.
Special Flow Rate Unit
Fast Keys 1, 3, 2, 2, 5
Special flow rate unit is a format variable that provides a record of the units to which you are converting. The Handheld Communicator will display a special units designator as the units format for your primary variable. The actual special units setting you define will not appear. Four characters are available to store the new units designation. The 8732 LOI will display the four character designation as configured.
Example To display flow in barrels per hour, and one barrel is equal to 31.0 gallons, the
procedure would be: Set the Volume Unit to BARL.
Set the Base Volume Unit to gallons. Set the Input Conversion Number to 31. Set the Time Base to Hour. Set the Rate Unit to BR/H.

Line Size The line size (flowtube sensor size) must be set to match the actual flowtube

Fast Keys 1, 3, 3
sensor connected to the transmitter. The size must be specified in inches according to the available sizes listed below. If a value is entered from a control system or Handheld Communicator that does not match one of these figures, the value will go to the next highest option.
The line size (inches) options are as follows:
0.1, 0.15, 0.25, 0.30, 0.50, 0.75, 1, 1.5, 2, 2.5, 3, 4, 6, 8, 10, 12, 14, 16, 18, 20, 24, 28, 30, 32, 36, 40, 42, 44, 48, 54, 56, 60, 64, 72, 80
3-9
Page 36
Rosemount 8732
Reference Manual
00809-0100-4663, Rev BA
January 2010

PV URV (Upper Range Value)

Fast Keys 1, 3, 4

PV LRV (Lower Range Value)

Fast Keys 1, 3, 5
The upper range value (URV), or analog output range, is preset to 30 ft/s at the factory . The units that appear will be the same as those selected under the units parameter.
The URV (20 mA point) can be set for both forward or reverse flow rate. Flow in the forward direction is represented by positive values and flow in the reverse direction is represented by negative values. The URV can be any value from –39.3 ft/s to +39.3 ft/s (-12 m/ s to +12 m/s) , as long as it is at least 1 ft/s (0.3 m/s) from the lower range value (4 mA point). The URV can be set to a value less than the lower range value. This will cause the transmitter analog output to operate in reverse, with the current increasing for lower (or more negative) flow rates.
NOTE
Line size, special units, and density must be selected prior to configuration of URV and LRV.
Set the lower range value (LRV), or analog output zer o, to change the size of the range (or span) between the UR V an d LR V. Under normal circumstances, the LRV should be set to a value near the minimum expected flow rate to maximize resolution. The LRV must be between –39.3 ft/s to +39.3 ft/s (-12 m/s to +12 m/s).
NOTE
Line size, special units, and density must be selected prior to configuration of URV and LRV.
Example
If the URV is greater than the LRV, the analog output will saturate at 3.9 mA when the flow rate falls below the selected 4 mA point.
The minimum allowable span between the URV and LRV is 1 ft/s (0.3 m/s). Do not set the LRV within 1 ft/s (0.3 m/s) of the 20 mA point. For example, if the URV is set to 15.67 ft/s (4.8 m/s) and if the desired URV is greater than the LRV, then the highest allowable analog zero setting would be 14.67 ft/s (4.5 m/s). If the desired URV is less than the LRV, then the lowest allowable LRV would be 16.67 ft/s (5.1 m/s).
3-10
Page 37
Reference Manual
00809-0100-4663, Rev BA January 2010
Rosemount 8732

Calibration Number The tube calibration number is a 16-digit number used to identify flowtube

Fast Keys 1, 3, 6
sensors calibrated at the Rosemount factory. The calibration number is also printed inside the flowtube sensor terminal block or on the flowtube sensor name plate. The number provides detailed calibration information to the Rosemount 8732. To function properly within accuracy specifications, the number stored in the transmitter must match the calibration number on the flowtube sensor exactly.
NOTE
Flowtube Sensors from manufacturers other than Rosemount Inc. can also be calibrated at the Rosemount factory. Check the tube for Rosemount calibration tags to determine if a 16-digit tube calibration number exists for your flowtube sensor.
NOTE
Be sure the calibration number reflects a calibration to a Rosemount reference transmitter. If the calibration number was generated by a means other than a certified Rosemount flow lab, accuracy of the system may be compromised.
If your flowtube sensor is not a Rosemount flowtube sensor and was not calibrated at the Rosemount factory, contact your Rosemount representative for assistance.
If your flowtube sensor is imprinted with an eight-digit number or a k-factor, check in the flowtube sensor wiring compartment for the sixteen-digit calibration number. If there is no serial number, contact the factory for a proper conversion.
PV Damping Adjustable between 0.0 and 256 seconds
Fast Keys 1, 3, 7

PV Damping allows selection of a response time, in seconds, to a step change in flow rate. It is most often used to smooth fluctuations in output.

3-11
Page 38
Rosemount 8732
Reference Manual
00809-0100-4663, Rev BA
January 2010
3-12
Page 39
Reference Manual
00809-0100-4663, Rev BA January 2010
Rosemount 8732

Section 4 Operation

Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 4-1
Diagnostics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .page 4-1
Advanced Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . page 4-12
Detailed Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 4-12
Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 4-17

INTRODUCTION This section contains information for advanced configuration parameters and

diagnostics. The software configuration settings for the Rosem ount 8732 can be accessed
through a 375 Field Communicator or through a contro l system. The sof tware functions for the 375 Field Communicator are described in detail in this section of the manual. It provides an overview and summary of communicator functions. For more complete instructions, see the communicator manual. Before operating the Rosemount 8732 in an actual installation, you should review all of the factory set configuration data to ensure that they reflect the current application.

DIAGNOSTICS Diagnostics are used to verify that the transmitter is functioning properly, to

375 Transducer Block
assist in troubleshooting, to identify potential causes of error messages, and to verify the health of the transmitter and sensor. Diagnostic tests can be initiated through the use of a 375 Field Communicator or through the control system.
Rosemount offers several different diagnostic suites providing various functionality.
Standard diagnostics included with every Rosemount 8732 transmitter are Empty Pipe detection, Electronics Temperature monitoring, Coil Fault detection, and various loop and transmitter tests.
Advanced diagnostics suite option one (D01 option) contains advanced diagnostics for High Process Noise detection and Grounding and Wiring fault detection.
Advanced diagnostics suite option two (D02 option) contains advanced diagnostics for the 8714i Meter Verification. This diagnostic is used to verify the accuracy and performance of the magnetic flow meter installation.

Diagnostic Controls The diagnostic controls menu provides a centralized location for enabling or

375 Transducer Block, Diagnostics
disabling each of the diagnostics that are available. Note that for some diagnostics to be available, a diagnostics suite package is required.
www.rosemount.com
Page 40
Rosemount 8732
Reference Manual
00809-0100-4663, Rev BA
January 2010
Empty Pipe Detection
Turn the empty pipe diagnostic on or off as required by the application. For more details on the empty pipe diagnostic, see Appendix C: Diagnostics.
Electronics Temperature Out of Range
Turn the electronics temperature diagnostic on or off as required by the application. For more details on the electronics temperature diagnostic, see Appendix C: Diagnostics.
High Process Noise Detection
Turn the high process noise diagnostic on or off as required by the application. For more details on the high process noise diagn os tic , see Appendix C: Diagnostics.
Grounding / Wiring Fault Detection
Turn the grounding / wiring diagnostic on or of f as required by the application . For more details on the grounding / wiring diagnostic, see Appendix C: Diagnostics.

Basic Diagnostics The basic diagnostics menu contains all of the st anda rd diagno stics and test s

375 Transducer Block, Diagnostics
that are available in the 8732E transmitter.
Empty Pipe Limits
375 Transducer Block, Diagnostics, Basic Diagnostics
Empty Pipe allows you to view the current value and configure the diagnostic parameters. For more detail on this parameter see Appendix C: Diagnostics.
EP Value
375 Transducer Block, Diagnostics, Basic Diagnostics, Empty Pipe Limits
Read the current Empty Pipe Value. This number is a unitless number and is calculated based on multiple installation and process variables. For more detail on this parameter see Appendix C: Diagnostics.
4-2
Page 41
Reference Manual
00809-0100-4663, Rev BA January 2010
Rosemount 8732
EP Trigger Level
375 Transducer Block, Diagnostics, Basic Diagnostics, Empty Pipe Limits
Limits: 3 to 2000 Configure the threshold limit that the empty pipe value must exceed before
the diagnostic alert activates. Default from the factory is set to 100. For more detail on this parameter see Appendix C: Diagnostics.
EP Counts
375 Transducer Block, Diagnostics, Basic Diagnostics, Empty Pipe Limits
Limits: 5 to 50 Configure the number of consecutive times that the empty pipe value must
exceed the empty pipe trigger level before the diagnostic alert activates. Counts are taken at 1.5 second intervals. Default from the factory is set to 5. For more detail on this parameter see Appendix C: Diagnostics.
Electronics Temp Value
375 Transducer Block, Diagnostics, Basic Diagnostics
Electronics Temperature allows you to view the current value for the electronics temperature.

Advanced Diagnostics The advanced diagnostics menu contains information on all of the additional

375 Transducer Block, Diagnostics
diagnostics and tests that are available in the 8732 transmitter if one of the diagnostics suite packages was ordered.
Rosemount offers two advanced diagnostic suites. Functionality under this menu will depend on which of these suites are ordered.
Advanced diagnostics suite option one (D01 option) contains advanced diagnostics for High Process Noise detection and Grounding and Wiring fault detection.
Advanced diagnostics suite option two (D02 option) contains advanced diagnostics for the 8714i Meter Verification. This diagnostic is used to verify the accuracy and performance of the magnetic flow meter installation.
8714i Meter Verification
375 Transducer Block, Diagnostics, Advanced Diagnostics
This diagnostic allows you to test and verify that the sensor, transmitter, or both are working within specifications. For more details on this diagnostic, see Appendix C: Diagnostics.
Run 8714i
375 Transducer Block, Diagnostics, Advanced Diagnostics, 8714i Meter Verification
Run the meter verification test to check the transmitter, sensor, or entire installation.
Full Meter Verification
Run the internal meter verification to check the entire installation, sensor and transmitter at the same time.
Transmitter Only
Run the internal meter verification to check the transmitter only.
Sensor Only
Run the internal meter verification to check the sensor only.
4-3
Page 42
Rosemount 8732
Reference Manual
00809-0100-4663, Rev BA
January 2010
8714i Results
375 Transducer Block, Diagnostics, Advanced Diagnostics, 8714i Meter Verification
Review the results of the most recently performed 8714i Meter Verification test. Information in this section details the measurements taken and if the meter passed the verification test. For more details on these results and what they mean, see Appendix C: Diagnostics.
Test Condition
375 Transducer Block, Diagnostics, Advanced Diagnostics, 8714i Meter Verification,
8714i Results
Displays the conditions that the 8714i Meter Verification test was performed under. For more details on this parameter see Appendix C: Diagnostics.
Test Criteria
375 Transducer Block, Diagnostics, Advanced Diagnostics, 8714i Meter Verification,
8714i Results
Displays the criteria that the 8714i Meter Verification test was performed against. For more details on this parameter see Appendix C: Diagnostics.
8714i Result
375 Transducer Block, Diagnostics, Advanced Diagnostics, 8714i Meter Verification,
8714i Results
Displays the results of the 8714i Meter Verification test as pass or fail. For more details on this parameter see Appendix C: Diagnostics.
Simulated Velocity
375 Transducer Block, Diagnostics, Advanced Diagnostics, 8714i Meter Verification,
8714i Results
Displays the test velocity used to verify transmitter calibration. For more details on this parameter see Appendix C: Diagnostics.
Actual Velocity
375 Transducer Block, Diagnostics, Advanced Diagnostics, 8714i Meter Verification,
8714i Results
Displays the velocity measured by the transmitter during the transmitter calibration verification test. For more details on this parameter see Appendix C: Diagnostics.
Velocity Deviation
375 Transducer Block, Diagnostics, Advanced Diagnostics, 8714i Meter Verification,
8714i Results
Displays the deviation of the transmitter calibration verification test. For more details on this parameter see Appendix C: Diagnostics.
Transmitter Calibration Result
375 Transducer Block, Diagnostics, Advanced Diagnostics, 8714i Meter Verification,
8714i Results
Displays the result of the transmitter calibration verification test as pass or fail. For more details on this parameter see Appendix C: Diagnostics.
4-4
Sensor Calibration Deviation
375 Transducer Block, Diagnostics, Advanced Diagnostics, 8714i Meter Verification,
8714i Results
Displays the deviation of the sensor calibration verification test. For more details on this parameter see Appendix C: Diagnostics.
Page 43
Reference Manual
00809-0100-4663, Rev BA January 2010
Rosemount 8732
Sensor Calibration Result
375 Transducer Block, Diagnostics, Advanced Diagnostics, 8714i Meter Verification,
8714i Results
Displays the result of the sensor calibration verification test as pass or fail. For more details on this parameter see Appendix C: Diagnostics.
Coil Circuit Result
375 Transducer Block, Diagnostics, Advanced Diagnostics, 8714i Meter Verification,
8714i Results
Displays the result of the coil circuit test as pass or fail. For more details on this parameter see Appendix C: Diagnostics.
Electrode Circuit Result
375 Transducer Block, Diagnostics, Advanced Diagnostics, 8714i Meter Verification,
8714i Results
Displays the result of the electrode circuit test as pass or fail. For more details on this parameter see Appendix C: Diagnostics.
Sensor Signature
375 Transducer Block, Diagnostics, Advanced Diagnostics, 8714i Meter Verification
The sensor signature describes the sensor characteristics to the transmitter and is an integral part of the sensor meter verification test. From this menu you can view the current stored signature, have the transmitter t ake and store the sensor signature, and re-call the last saved go o d valu e s for the sens or signature. For more details on this parameter see Appendix C: Diagnostics.
Signature Values
375 Transducer Block, Diagnostics, Advanced Diagnostics, 8714i Meter Verification,
Sensor Signature
Review the current values stored for the sensor signature. For more details on this parameter see Appendix C: Diagnostics.
Coil Resistance
375 Transducer Block, Diagnostics, Advanced Diagnostics, 8714i Meter Verification,
Sensor Signature, Signature Values
View the reference value for the coil resist ance taken during the sensor signature process.
Coil Signature
375 Transducer Block, Diagnostics, Advanced Diagnostics, 8714i Meter Verification,
Sensor Signature, Signature Values
View the reference value for the coil signature taken during the sensor signature process.
Electrode Resistance
375 Transducer Block, Diagnostics, Advanced Diagnostics, 8714i Meter Verification,
Sensor Signature, Signature Values
View the reference value for the electrode resistance taken during the sensor signature process.
4-5
Page 44
Rosemount 8732
Reference Manual
00809-0100-4663, Rev BA
January 2010
Re-Signature Meter
375 Transducer Block, Diagnostics, Advanced Diagnostics, 8714i Meter Verification,
Sensor Signature
Have the transmitter measure and store the sensor signature values. These values will then be used as the baseline for the meter verification test. Use this when connecting to older Rosemount or competitors’ sensors or installing the magnetic flowmeter system for the first time. For more details on this parameter see Appendix C: Diagnostics.
Recall Last Saved Values
375 Transducer Block, Diagnostics, Advanced Diagnostics, 8714i Meter Verification,
Sensor Signature
Recalls the last saved “good” values for the sensor signature.
Set Pass/Fail Criteria
375 Transducer Block, Diagnostics, Advanced Diagnostics, 8714i Meter Verification
Set the maximum allowable deviation percentage test criteria for the 8714i Meter Verification test. There are three tests that this criteria can be set for:
Full Pipe; No Flow (Best test condition) – Default is 2%
Full Pipe; Flowing – Default is 3%
Empty Pipe – Default is 5%
NOTE
If the 8714i Meter Verification test is done with an empty pipe, the electrode circuit will NOT be tested.
No Flow Limit
375 Transducer Block, Diagnostics, Advanced Diagnostics, 8714i Meter Verification,
Set Pass/Fail Criteria
Limits: 1 to 10 percent Set the pass/fail test criteria for the 8714i Meter Verification test at Full Pipe,
No Flow conditions.
Flowing Limit
375 Transducer Block, Diagnostics, Advanced Diagnostics, 8714i Meter Verification,
Set Pass/Fail Criteria
Limits: 1 to 10 percent
4-6
Set the pass/fail test criteria for the 8714i Meter Verification test at Full Pipe, Flowing conditions.
Empty Pipe Limit
375 Transducer Block, Diagnostics, Advanced Diagnostics, 8714i Meter Verification,
Set Pass/Fail Criteria
Limits: 1 to 10 percent Set the pass/fail test criteria for the 8714i Meter Verification test at Empty Pipe
conditions.
Page 45
Reference Manual
00809-0100-4663, Rev BA January 2010
Rosemount 8732
Measurements
375 Transducer Block, Diagnostics, Advanced Diagnostics, 8714i Meter Verification
View the measured values taken during th e meter verification process. These values are compared to the signature values to de termine if the test p asses or fails. V alues are shown fo r the Coil Resist ance, Coil Signatur e, and Electrod e Resistance.
Coil Resistance
375 Transducer Block, Diagnostics, Advanced Diagnostics, 8714i Meter Verification,
measurements
View the measured value for the coil resist ance taken during the meter verification test.
Coil Signature
375 Transducer Block, Diagnostics, Advanced Diagnostics, 8714i Meter Verification,
measurements
View the measured value for the coil signature taken during the meter verification test.
Electrode Resistance
375 Transducer Block, Diagnostics, Advanced Diagnostics, 8714i Meter Verification,
measurements
View the measured value for the electrode resistance taken during the meter verification test.
Licensing
375 Transducer Block, Diagnostics, Advanced Diagnostics
If a diagnostic suite was not ordered initially, advanced diagnostics can be licensed in the field. Access the licensing information from this menu. For more details on licensing, see Appendix C: Diagnostics.
License Status
375 Transducer Block, Diagnostics, Advanced Diagnostics, Licensing
Determine if a diagnostics suite has been licensed, and if so, which diagnostics are available for activation.
License Key
375 Transducer Block, Diagnostics, Advanced Diagnostics, Licensing
A license key is required to activate diagnostics in the field if the diagnostic suite was not initially ordered. This menu allows for gathering of necessary data to generate a license key and also the ability to enter the license key once it has been received.
Device ID
375 Transducer Block, Diagnostics, Advanced Diagnostics, Licensing, License Key
This function displays the Device ID and Software Revision for the transmitter. Both of these pieces of information are required to generate a license key.
License Key
375 Transducer Block, Diagnostics, Advanced Diagnostics, Licensing, License Key
Allows you to enter a license key to activate a diagnostic suite.
4-7
Page 46
Reference Manual
00809-0100-4663, Rev BA
Rosemount 8732
January 2010

Diagnostic Variables From this menu, all of the diagnostic variable values can be reviewed. This

375 Transducer Block, Diagnostics
information can be used to get more information about the transmitter, sensor, and process, or to get more detail about an alert that may have activated.
Empty Pipe Value
375 Transducer Block, Diagnostics, Diagnostic Variables
Read the current value of the Empty Pipe parameter . This value will read zero if Empty Pipe is turned off.
Electronics Temperature
375 Transducer Block, Diagnostics, Diagnostic Variables
Read the current value of the Electronics Temperature.
Line Noise
375 Transducer Block, Diagnostics, Diagnostic Variables
Read the current value of the amplitude of AC line noise measured on the transmitter’s electrode inputs. This value is used in the grounding / wiring diagnostic.
5Hz SNR
375 Transducer Block, Diagnostics, Diagnostic Variables
Read the current value of the signal to noise ratio at 5 Hz. For optimum performance, a value greater than 100 is preferred. Values less than 25 will cause the High Process Noise alert to activate.
37Hz SNR
375 Transducer Block, Diagnostics, Diagnostic Variables
Read the current value of the signal to noise ratio at 37.5 Hz. For optimum performance, a value greater than 100 is preferred. Values less than 25 will cause the High Process Noise alert to activate.
Signal Power
375 Transducer Block, Diagnostics, Diagnostic Variables
Read the current value of the velocity of the fluid through the sensor. Higher velocities result in greater signal power.
8714i Results
375 Transducer Block, Diagnostics, Diagnostic Variables
Review the results of the 8714i Meter Verification tests. For more details on these results and what they mean, see Appendix C: Diagnostics.
Test Condition
375 Transducer Block, Diagnostics, Diagnostic Variables, 8714i Results
Displays the conditions that the 8714i Meter Verification test was performed under. For more details on this parameter see Appendix C: Diagnostics.
4-8
Test Criteria
375 Transducer Block, Diagnostics, Diagnostic Variables, 8714i Results
Displays the criteria that the 8714i Meter Verification test was performed against. For more details on this parameter see Appendix C: Diagnostics.
Page 47
Reference Manual
00809-0100-4663, Rev BA January 2010
Rosemount 8732
8714i Result
375 Transducer Block, Diagnostics, Diagnostic Variables, 8714i Results
Displays the results of the 8714i Meter Verification test as pass or fail. For more details on this parameter see Appendix C: Diagnostics.
Simulated Velocity
375 Transducer Block, Diagnostics, Diagnostic Variables, 8714i Results
Displays the test velocity used to verify transmitter calibration. For more details on this parameter see Appendix C: Diagnostics.
Actual Velocity
375 Transducer Block, Diagnostics, Diagnostic Variables, 8714i Results
Displays the velocity measured by the transmitter during the transmitter calibration verification test. For more details on this parameter see Appendix C: Diagnostics.
Velocity Deviation
375 Transducer Block, Diagnostics, Diagnostic Variables, 8714i Results
Displays the deviation of the transmitter calibration verification test. For more details on this parameter see Appendix C: Diagnostics.
Transmitter Calibration Result
375 Transducer Block, Diagnostics, Diagnostic Variables, 8714i Results
Displays the result of the transmitter calibration verification test as pass or fail. For more details on this parameter see Appendix C: Diagnostics.
Sensor Calibration Deviation
375 Transducer Block, Diagnostics, Diagnostic Variables, 8714i Results
Displays the deviation of the sensor calibration verification test. For more details on this parameter see Appendix C: Diagnostics.
Sensor Calibration Result
375 Transducer Block, Diagnostics, Diagnostic Variables, 8714i Results
Displays the result of the sensor calibration verification test as pass or fail. For more details on this parameter see Appendix C: Diagnostics.
Coil Circuit Result
375 Transducer Block, Diagnostics, Diagnostic Variables, 8714i Results
Displays the result of the coil circuit test as pass or fail. For more details on this parameter see Appendix C: Diagnostics.
Electrode Circuit Result
375 Transducer Block, Diagnostics, Diagnostic Variables, 8714i Results
Displays the result of the electrode circuit test as pass or fail. For more details on this parameter see Appendix C: Diagnostics.

Trims Trims are used to calibrate the analog loop, calibrate the transmitter, re-zero

375 Transducer Block, Diagnostics
the transmitter, and calibrate the transmitter with another manufacturer’s sensor. Proce ed with caution whenever performing a trim function.
4-9
Page 48
Rosemount 8732
Reference Manual
00809-0100-4663, Rev BA
January 2010
Electronics Trim
375 Transducer Block, Diagnostics, Trims
Electronics trim is the function by which the factory calibrates the transmitter. This procedure is rarely needed by customers. It is only necessary if you suspect the Rosemount 8732E is no longer accurate. A Rosemount 8714 Calibration Standard is required to complete a digital trim. Attempting an Electronics trim without a Rosemount 8714 Calibration S t andard may result in an inaccurate transmitter or an error message. Electronics trim must be performed only with the coil drive mode set to 5 Hz and with a nominal sensor calibration number stored in the memory.
NOTE
Attempting an Electronics trim without a Rosemount 8714 may result in an inaccurate transmitter, or a “DIGITAL TRIM FAILURE” message may appear. If this message occurs, no values were changed in the transmitter. Simply power down the Rosemount 8732E to clear the message.
To simulate a nominal sensor with the Rosemount 8714, you must chang e the following five parameters in the Rosemount 8732E:
1. Sensor Calibration Number—1000015010000000
2. Units—ft/s
3. PV URV—AI EU at 100 = 30.00 ft/s
4. PV LRV—AI EU at 0 = 0 ft/s
5. Coil Drive Frequency—5 Hz
The instructions for changing the Sensor Calibration Number, Units, PV URV, and PV LRV are located in “Basic Setup” on pag e3-14. Instructions for changing the Coil Drive Frequency can be found on page 4-12 in this section.
Set the loop to manual, if necessary , before you begin. Complete the following steps:
1. Power down the transmitter.
2. Connect the transmitter to a Rosemount 8714 sensor simulator.
3. Power up the transmitter with the Rosemount 8714 connected and read the flow rate. The electronics need about a 5-minute warm-up time to stabilize.
4. Set the 8714 calibrator to the 30 ft/s setting.
5. The flow rate reading after warm-up should be between 29.97 and
30.03 ft/s.
6. If the reading is within the range, return the transmitter to the original configuration parameters.
7. If the reading is not within this range, initiate a digital trim with the Handheld Communicator. The digital trim takes about 90 seconds to complete. No transmitter adjustments are required.
4-10
Page 49
Reference Manual
00809-0100-4663, Rev BA January 2010
Rosemount 8732
Auto Zero
375 Transducer Block, Diagnostics, Trims
The auto zero function initializes the transmitter for use with the 37 Hz coil drive mode only. Run this function only with the transmitter and sensor installed in the process. The sensor must be filled with process fluid at zero flow. Before running the auto zero function, be sure the coil drive mode is set to 37 Hz (Auto Zero will not run with the coil drive frequency set at 5 Hz).
Set the loop to manual if necessary and begin the auto zero procedure. The transmitter completes the procedure automatically in about 90 seconds. A symbol appears in the lower right-hand corner of the display to indicate that the procedure is running.
Universal Trim
375 Transducer Block, Diagnostics, Trims
The universal auto trim function enables the Rosemount 8732E to calibrate sensors that were not calibrated at the Rosemount factory. The function is activated as one step in a procedure known as in-process calibration. If your Rosemount sensor has a 16-digit calibration number , in -pro cess ca libr ation is not required. If it does not, or if your sensor is made by another manufactu rer, complete the following steps for in-process calibration.
1. Determine the flow rate of the process fluid in the sensor.
NOTE
The flow rate in the line can be determined by using another sensor in the line, by counting the revolutions of a centrifug al pump , or by pe rfo rm in g a bucket test to determine how fast a given volume is filled by the process fluid.
2. Complete the universal auto trim function.
3. When the routine is completed, the sensor is ready for use.

Status Review status information regarding the operation of the transducer block.

375 Transducer Block, Diagnostics
This is where additional information can be reviewed regard ing transmitter health and diagnostic messages.
4-11
Page 50
Rosemount 8732
Reference Manual
00809-0100-4663, Rev BA
January 2010

ADVANCED CONFIGURATION

In addition to the basic configuration options and the diagnostic information and controls, the 8732 has many advanced functions that can also be configured as required by the application .

DETAILED SETUP The detailed setup function provides access to other parameters within the

375 Transducer Block
transmitter that can be configured such as coil drive frequency, output parameters, local display configuration, and other general information about the device.

Additional Parameters The additional parameters menu provides a means to configure optional

375 Transducer Block, Detailed
Setup
parameters within the 8732E transmitter.
Coil Drive Frequency
375 Transducer Block, Detailed Setup, Additional Params
Coil drive frequency allows pulse-rate selection of the sensor coils.
5 Hz
The standard coil drive frequency is 5 Hz, which is sufficient for nearly all applications.
37 Hz
If the process fluid causes a noisy or unstable output, increase the coil drive frequency to 37 Hz. If the 37 Hz mode is selected, perform the auto zero function.
Density Value
375 Transducer Block, Detailed Setup, Additional Params
The density value is used to convert from a volumetric flow rate to a mass flow rate using the following equation:
= Qv
Q
m
Where: Q
is the mass flow rate
m
Q
is the volumetric flow rate, and
v
is the fluid density
NOTE
A density value is required to configure the flow units for mass flow rate measurement.
Sensor Range: EU at 100%
375 Transducer Block, Detailed Setup, Additional Params
This parameter is the maximum value that the PV Range value can be set to. This is the upper measuring limit of the transmitter and sensor.
Sensor Range: EU at 0%
375 Transducer Block, Detailed Setup, Additional Params
This parameter is the minimum value that the PV Range value can be set to. This is the lower measuring limit of the transmitter and sensor.
4-12
Page 51
Reference Manual
00809-0100-4663, Rev BA January 2010
Rosemount 8732
Cal Min Span
375 Transducer Block, Detailed Setup, Additional Params
The PV minimum span is the minimum flow range that must separate th e minimum and maximum configured PV Range values.
Reverse Flow
375 Transducer Block, Detailed Setup, Additional Params
Enable or disable the transmitter’s ability to read reverse flow. Reverse Flow allows the transmitter to read negative flow. This may occur
when flow in the pipe is going the negative direction, or when either electrode wires or coil wires are reversed. This also enables the totalizer to count in the reverse direction.

Display Language This allows you to configure the language shown on the display. There are

375 Transducer Block, Detailed
Setup
five options available:
English
Spanish
Portuguese
•German
•French

Signal Processing The 8732E contains several advanced functions that can be used to stabilize

375 Transducer Block, Detailed
Setup
erratic outputs caused by process noise. The signal processing menu contains this functionality.
Operating Mode
375 Transducer Block, Detailed Setup, Signal Processing
The Operating Mode should be used only when the signal is noisy and gives an unstable output. Filter mode automatically uses 37 Hz coil drive mode an d activates signal processing at the factory set default values. When using filter mode, perform an auto zero with no flow and a full sensor. Either of the parameters, coil drive mode or signal processing, may still be changed individually. T urning Signal Processing off or changing the coil drive frequency to 5 Hz will automatically change the Operating Mode from filter mode to normal mode.
Man Config DSP
375 Transducer Block, Detailed Setup, Signal Processing
Manually configure the digital signal processing parameters. The 8732E transmitter includes digital signal processing capabilities that can
be used to condition the output from the transmitter by enabling noise rejection. See Appendix D: Digital Signal Processing for more information on the DSP functionality.
4-13
Page 52
Rosemount 8732
Reference Manual
00809-0100-4663, Rev BA
January 2010
Control
375 Transducer Block, Detailed Setup, Signal Processing, Man Config DSP
When ON is selected, the Rosemount 8732E output is derived using a running average of the individual flow inputs. Signal processing is a software algorithm that examines the quality of the electrode signal against user-specified tolerances. This average is updated at the rate of 10 samples per second with a coil drive frequency of 5 Hz, and 75 samples per second with a coil drive frequency of 37 Hz. The three parameters that make up signal processing (number of samples, maximum percent limit, and time limit) are described below.
Samples
375 Transducer Block, Detailed Setup, Signal Processing, Man Config DSP
0 to 125 Samples The number of samples function sets the amount of time that inputs are
collected and used to calculate the average value. Each second is divided into tenths (1/10) with the number of samples equaling the number of 1/10 second increments used to calculate the average.
For example, a value of: 1 averages the inputs over the past 1/10 second 10 averages the inputs over the past 1 second 100 averages the inputs over the past 10 second s 125 averages the inputs over the past 12.5 seconds
% Limit
375 Transducer Block, Detailed Setup, Signal Processing, Man Config DSP
0 to 100 Percent The maximum percent limit is a tolerance band set up on either side of the
running average. The percentage value refers to deviation from the running average. For example, if the running average is 100 gal/min, and a 2 percent maximum limit is selected, then the acceptable range is from 98 to 102 gal/min.
V alues within the limit are accepted while values out side the limit are analyzed to determine if they are a noise spike or an actual flow change.
Time Limit
375 Transducer Block, Detailed Setup, Signal Processing, Man Config DSP
0 to 256 Seconds The time limit parameter forces the output and running average values to the
new value of an actual flow rate change that is outside the percent limit boundaries. It thereby limits response time to flow changes to the time limit value rather than the length of the running average.
4-14
Page 53
Reference Manual
00809-0100-4663, Rev BA January 2010
Rosemount 8732
For example, if the number of samples selected is 100, then the response time of the system is 10 seconds. In some cases this may be unacceptable. By setting the time limit, you can force the 8732E to clear the value of the running average and re-establish the output and average at the new flow rate once the time limit has elapsed. This parameter limits the response time added to the loop. A suggested time limit value of two seconds is a good starting point for most applicable process fluids. The selected signal processing configuration may be turned ON or OFF to suit your needs.
Coil Drive Frequency
375 Transducer Block, Detailed Setup, Signal Processing
Coil drive frequency allows pulse-rate selection of the sensor coils.
5 Hz
The standard coil drive frequency is 5 Hz, which is sufficient for nearly all applications.
37 Hz
If the process fluid causes a noisy or unstable output, increase the coil drive frequency to 37 Hz. If the 37 Hz mode is selected, perform the auto zero function with no flow and a full sensor.
Low Flow Cutoff
375 Transducer Block, Detailed Setup, Signal Processing
Low flow cutoff allows you to specify the flow rate, between 0.01 and 38.37 feet per second, below which the outputs are driven to zero flow. The units format for low flow cutoff cannot be changed. It is always displ ayed as feet per second regardless of the format selected. The low flow cutof f value applies to both forward and reverse flows.
Primary Variable Damping
375 Transducer Block, Detailed Setup, Signal Processing
0 to 256 Seconds Primary Variable Damping allows selection of a response time, in seconds, to
a step change in flow rate. It is most often used to smooth fluctuations in output.

Device Info Information variables are used for identification of flowmeters in the field and

375 Transducer Block, Detailed
Setup
to store information that may be useful in service situations. Information variables have no effect on flowmeter output or process variables.
Device ID
375 Transducer Block, Detailed Setup, Device Info
This function displays the Device ID of the transmitter. This is one piece of information required to generate a license code to enable diagnostics in the field.
PV Sensor S/N
375 Transducer Block, Detailed Setup, Device Info
The PV sensor serial number is the serial number of the sensor connected to the transmitter and can be stored in the transmitter configuration for future reference. The number provides easy identification if the sensor needs servicing or for other purposes.
4-15
Page 54
Rosemount 8732
Reference Manual
00809-0100-4663, Rev BA
January 2010
Sensor Tag
375 Transducer Block, Detailed Setup, Device Info
Sensor tag is the quickest and shortest way of identifyin g and distinguishing between sensors. Sensors can be tagged according to th e requirements of your application. The tag may be up to eight characters long.
DSP Software Rev
375 Transducer Block, Detailed Setup, Device Info
This function displays the software revision number of the transmitter.
Construction Materials
375 Transducer Block, Detailed Setup, Device Info
Construction materials contain information about the sensor that is con nected to the transmitter. This information is configured into the transmitter for later reference. This information can be helpful when calling the factory for support.
Flange Type
375 Transducer Block, Detailed Setup, Device Info, Construction Materials
Flange type enables you to select the flange type for your magnetic transmitter system. This variable only needs to be changed if you have changed your sensor. Options for this value are:
• ANSI 150 • PN 10
• ANSI 300 •PN 16
• ANSI 600 • PN 25
• ANSI 900 •PN 40
• ANSI 1500 • PN 64
• ANSI 2500 • Other
•Wafer
Flange Material
375 Transducer Block, Detailed Setup, Device Info, Construction Materials
Flange material enables you to select the flange material for your magnetic transmitter system. This variable only needs to be changed if you have changed your sensor. Options for this value are:
Carbon Ste el
304L Stainless Steel
316L Stainless Steel
Wafer
•Other
4-16
Page 55
Reference Manual
00809-0100-4663, Rev BA January 2010
Rosemount 8732
Electrode Type
375 Transducer Block, Detailed Setup, Device Info, Construction Materials
Electrode type enables you to select the electrode type for your magnetic transmitter system. This variable only needs to be changed if you have replaced electrodes or if you have replaced your sensor. Options for this value are:
Standard
St d & Ground
Bullet
•Other
Electrode Material
375 Transducer Block, Detailed Setup, Device Info, Construction Materials
Electrode Material enables you to select the electrode material for your magnetic transmitter system. This variable only needs to be changed if you have replaced electrodes or if you have replaced your sensor. Options for this value are:
316L SST
Nickel Alloy 276 (UNS N10276)
•Tantalum
Titanium
80% Platinum – 20% Iridium
Alloy 20
•Other
Liner Material
375 Transducer Block, Detailed Setup, Device Info, Construction Materials
Liner material enables you to select the liner material for the attached sensor. This variable only needs to be changed if you have replaced your sensor. Options for this value are:
PTFE
ETFE
•PFA
Polyurethane
Linatex
Natural Rubber
Neoprene
•Other

MODE Set and review the mode configuration for the transducer function block.

375 Transducer Block
4-17
Page 56
Reference Manual
00809-0100-4663, Rev BA
Rosemount 8732
January 2010

Block Mode: Target Operator requested mode for the function block. Only one selection may be

375 Transducer Block, Mode
made. Options include:
Auto
Use this mode when all configuration changes to the block are complete and the transmitter is ready to be returned to service.
OOS
Out of service mode. Use this mode when making configuration changes to parameters found in the function block. This removes the transmitter from operation until the mode is set back to Auto.

Block Mode: Actual This is the current mode of the function block. This mode may differ from the

375 Transducer Block, Mode
Target mode based on operating conditions.

Block Mode: Permitted This parameter defines which modes are available for a given function block.

375 Transducer Block, Mode

Block Mode: Normal Displays the mode that the function block should be set to for normal

375 Transducer Block, Mode
operation.
4-18
Page 57
Reference Manual
00809-0100-4663, Rev BA January 2010

Section 5 Sensor Installation

Safety Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .page 5-1
Sensor Handling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 5-3
Sensor Mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 5-4
Installation (Flanged Sensor) . . . . . . . . . . . . . . . . . . . . . . page 5-7
Installation (W afer Sensor) . . . . . . . . . . . . . . . . . . . . . . . .page 5-10
Installation (Sanitary Sensor) . . . . . . . . . . . . . . . . . . . . . . page 5-12
Grounding . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 5-12
Process Leak Protection (Optional) . . . . . . . . . . . . . . . . . page 5-16
This section covers the steps required to physically install the magnetic sensor. For electrical connections and cabling see Section 2: "Installation". Instructions and procedures in this section may require special precautions to ensure the safety of the personnel performing the op erations. Please refer to the following safety messages before performin g any op eration in thi s section.
Rosemount 8732

SAFETY MESSAGES This symbol is used throughout this manual to indicate that special attention

to warning information is required.
Failure to follow these installation guidelines could result in death or serious injury: Installation and servicing instructions are for use by qualified personnel only. Do not perform
any servicing other than that contained in the operating instructions, unless qualified. Verify that the operating environment of the sensor and transmitter is consistent with the appropriate hazardous area approval.
Do not connect a Rosemount 8732 to a non-Rosemount sensor that is located in an explosive atmosphere.
www.rosemount.com
Page 58
Rosemount 8732
Reference Manual
00809-0100-4663, Rev BA
January 2010
Explosions could result in death or serious injury: Installation of this transmitter in an explosive environment must be in accordance with the
appropriate local, national, and international standards, codes, and practices. Please review the approvals section of the 8732 reference manual for any restrictions associated with a safe installation.
Before connecting a Field Communicator in an explosive atmosphere, make sure the instruments in the loop are installed in accordance with intrinsically safe or non-incendive field wiring practices.
Electrical shock can result in death or serious injury Avoid contact with the leads and terminals. High voltage that may be present on leads can
cause electrical shock.
The sensor liner is vulnerable to handling damage. Never place anything through the sensor for the purpose of lifting or gaining leverage. Liner damage can render the sensor useless.
To avoid possible damage to the sensor liner ends, do not use metallic or spiral-wound gaskets. If frequent removal is anticipated, take precautions to protect the liner ends. Short spool pieces attached to the sensor ends are often used for protection.
Correct flange bolt tightening is crucial for proper sensor operation and life. All bolts must be tightened in the proper sequence to the specified torque limits. Failure to observe these instructions could result in severe damage to the sensor lining and possible sensor replacement.
Emerson Process Management can supply lining protectors to prevent liner damage during removal, installation, and excessive bolt torquing.
5-2
Page 59
Reference Manual
See ”Safety Messages” on pages 5-1 and 5-2 for complete warning information.
½- through 4-Inch
Sensors
6-Inch and Larger
Sensors
00809-0100-4663, Rev BA January 2010
Rosemount 8732

SENSOR HANDLING Handle all parts carefully to prevent damage. Whenever possible, transport

the system to the installation site in the original shipping containers. PTFE-lined sensors are shipped with end covers that protect it from both mechanical damage and normal unrestrained distortion. Remove the end covers just before installation.
Flanged 6- through 36-inch sensors come with a lifting lug on each flange. The lifting lugs make the sensor easier to handle when it is transported and lowered into place at the installation site.
Flanged ½- to 4-inch sensors do not have lugs. They must be supported with a lifting sling on each side of the housing.
Figure 5-1 shows sensors correctly supported for handling and installation. Notice the plywood end pieces are still in place to protect the sensor liner during transportation.
Figure 5-1. Rosemount 8705 Sensor Support for Handling
5-3
Page 60
Reference Manual
FLOW
5 Pipe Diameters
2 Pipe Diameters
FLOW
FLOW
00809-0100-4663, Rev BA
Rosemount 8732
January 2010

SENSOR MOUNTING Physical mounting of a sensor is similar to installing a typical section of pipe.

Conventional tools, equipment, and accessories (bolts, gaskets, and grounding hardware) are required.

Upstream/Downstream Piping

Figure 5-2. Upstream and Downstream Straight Pipe Diameters
To ensure specification accuracy over widely varying process conditions, install the sensor a minimum of five straight pipe diameters upstream and two pipe diameters downstream from the electrode plane (see Figure 5-2).

Sensor Orientation The sensor should be installed in a position that ensures the sensor remains

full during operation. Figures 5-3, 5-4, and 5-5 show the proper sensor orientation for the most common installations. The following orientations ensure that the electrodes are in the optimum plane to minimize the ef fects of entrapped gas.
Vertical installation allows upward process fluid flow and is generally preferred. Upward flow keeps the cross-sectional area full, regardless of flow rate. Orientation of the electrode plane is unimportant in vertical installations. As illustrated in Figures 5-3 and 5-4, avoid downward flows where back pressure does not ensure that the se nsor remains full at all times.
Figure 5-3. Vertical Sensor Orientation
Installations with reduced straight runs from 0 to five pipe diameters are possible. In reduced straight pipe run installations, performance will shift to as much as 0.5% of rate. Reported flow rates will still be highly repeatable.
5-4
Page 61
Reference Manual
FLOW
FLOW
FLOW
00809-0100-4663, Rev BA January 2010
Figure 5-4. Incline or Decline Orientation
Rosemount 8732
Horizontal installation should be restricted to low piping sections that are normally full. Orient the electrode plane to within 45 degrees of horizontal in horizontal installations. A deviation of more than 45 degrees of horizontal would place an electrode at or near the top of the sensor thereby making it more susceptible to insulation by air or entrapped gas at the top of the sensor.
Figure 5-5. Horizontal Sensor Orientation
5-5
Page 62
Rosemount 8732
45° Electrode Plane
45° Electrode Plane
FLOW
Figure 5-6. Rosemount 8711 Mounting Position
Reference Manual
00809-0100-4663, Rev BA
January 2010
The electrodes in the Rosemount 8711 are properly oriented when the top of the sensor is either vertical or horizontal, as shown in Figure 5-6. Avoid any mounting orientation that positions the top of the sensor at 45 degrees from the vertical or horizontal position.

Flow Direction The sensor should be mounted so that the FORWARD end of the flow arrow,

shown on the sensor identification tag, points in the direction of flow through the sensor (see Figure 5-7).
Figure 5-7. Flow Direction
5-6
Page 63
Reference Manual
See ”Safety Messages” on pages 5-1 and 5-2 for complete warning information.
Gasket (Supplied by user)
Gasket (Supplied by user)
Grounding Ring
Gasket (Supplied by user)
00809-0100-4663, Rev BA January 2010
Rosemount 8732

INSTALLATION (FLANGED SENSOR)

The following section should be used as a guide in the installation of the flange-type Rosemount 8705 and Rosemount 8707 High-Signal Sensors. Refer to page 5-10 for installation of the wafer-type Rosemount 8711 Sensor.

Gaskets The sensor requires a gasket at each of its connections to adjacent devices or

piping. The gasket material selected must be compatible with the process fluid and operating conditions. Metallic or spiral-wound gaskets can damage the liner. If the gaskets will be removed frequently, protect the liner ends. All other applications (including sensors with lining protectors or a grounding electrode) require only one gasket on each end connection, as shown in Figure 5-8. If grounding rings are used, gaskets are required on each side of the groundin g ring, as shown in Figure 5-9.
Figure 5-8. Gasket Placement
Figure 5-9. Gasket Placement with Non-attached Grounding Rings

Flange Bolts Suggested torque values by sensor line size and liner type are listed in Table

5-1 on page 5-8 for ASME B16.5 (ANSI) flanges and Table 5-2 and Table 5-3 for DIN flanges. Consult the factory for other flange ratings. Tighten flange bolts in the incremental sequence as shown in Figure 5- 10. See Table 5-1 and Table 5-2 for bolt sizes and hole diameters.
5-7
Page 64
Rosemount 8732
See ”Safety Messages” on pages 5-1 and 5-2 for complete warning information.
Reference Manual
00809-0100-4663, Rev BA
January 2010
NOTE
Do not bolt one side at a time. Tighten each side simultaneously. Example:
1. Snug left
2. Snug right
3. Tighten left
4. Tighten right Do not snug and tighten the upstream side and then snug and tighten the downstream side. Failure to alternate between the upstream a nd downstream flanges when tightening bolts may result in liner damage.
Always check for leaks at the flanges after tightening the flange bolts. Failure to use the correct flange bolt tightening methods can result in severe dam age. All sensors require a second torquing 24 hours after initial flange bolt tightening.
Table 5-1. Flange Bolt Torque Specifications for Rosemount 8705 and 8707 High-Signal Sensors
PTFE/ETFE liner Polyurethane liner
Class 150
Size Code Line Size
005 010 1 inch (25 mm) 8 12 — 015 1 020 2 inch (50 mm) 19 17 14 11 030 3 inch (80 mm) 34 35 23 23 040 4 inch (100 mm) 26 50 17 32 060 6 inch (150mm) 45 50 30 37 080 8 inch (200 mm) 60 82 42 55 100 10 inch (250 mm) 55 80 40 70 120 12 inch (300 mm) 65 125 55 105 140 14 inch (350 mm) 85 110 70 95 160 16 inch (400 mm) 85 160 65 140 180 18 inch (450 mm) 120 170 95 150 200 20 inch (500 mm) 110 175 90 150 240 24 inch (600 mm) 165 280 140 250 300 30 inch (750 mm) 195 415 165 375 360 36 inch (900 mm) 280 575 245 525
1
/2-inch (15 mm) 8 8
1
/2 inch (40 mm) 13 25 7 18
(pound-feet)
Class 300
(pound-feet)
Class 150
(pound-feet)
Class 300
(pound-feet)
5-8
Page 65
Reference Manual
4-Bolt
8-Bolt
12-Bolt
14-Bolt
20-Bolt
Torque the flange bolts
in increments according to
the above numerical sequence.
00809-0100-4663, Rev BA January 2010
Table 5-2. Flange Bolt Torque and Bolt Load Specifications for Rosemount 8705
PTFE/ETFE liner
Size
Code
005 010 1 inch (25 mm) 13 6983 13 6983 13 6983 13 8816 015 1 020 2 inch (50 mm) 25 10420 25 10420 25 10420 25 14457 030 3 inch (80 mm) 14 5935 14 5935 18 7612 18 12264 040 4 inch (100 mm) 17 7038 17 7038 30 9944 30 16021 060 6 inch (150mm) 23 7522 32 10587 60 16571 60 26698 080 8 inch (200 mm) 35 11516 35 11694 66 18304 66 36263 100 10 inch (250 mm) 31 10406 59 16506 105 25835 105 48041 120 12 inch (300 mm) 43 14439 82 22903 109 26886 109 51614 140 14 inch (350 mm) 42 13927 80 22091 156 34578 156 73825 160 16 inch (400 mm) 65 18189 117 28851 224 45158 224 99501 180 18 inch (450 mm) 56 15431 99 24477 67953 200 20 inch (500 mm) 66 18342 131 29094 225 45538 225 73367 240 24 inch (600 mm) 104 25754 202 40850 345 63940 345 103014
Line Size (Newton-meter) (Newton) (Newton-meter) (Newton) (Newton-meter) (Newton) (Newton-meter) (Newton)
1
/2-inch (15 mm) 7 3209 7 3809 7 3809 7 4173
1
/2 inch (40 mm) 24 9983 24 9983 24 9983 24 13010
PN10 PN 16 PN 25 PN 40
Rosemount 8732
Figure 5-10. Flange Bolt To rquing Sequence
5-9
Page 66
Reference Manual
00809-0100-4663, Rev BA
Rosemount 8732
Table 5-3. Flange Bolt Torque and Bolt Load Specifications for Rosemount 8705
Polyurethane Liner
Size
Code
005 010 1 inch (25 mm) 2 1191 3 1890 5 2958 10 5555 015 1 020 2 inch (50 mm) 6 2535 10 4021 15 6294 26 10831 030 3 inch (80 mm) 5 2246 9 3563 13 5577 24 19998 040 4 inch (100 mm) 7 3033 12 4812 23 7531 35 11665 060 6 inch (150mm) 16 5311 25 8425 47 13186 75 20829 080 8 inch (200 mm) 27 8971 28 9487 53 14849 100 24687 100 10 inch (250 mm) 26 8637 49 13700 87 21443 155 34547 120 12 inch (300 mm) 36 12117 69 19220 91 22563 165 36660 140 14 inch (350 mm) 35 11693 67 18547 131 29030 235 47466 160 16 inch (400 mm) 55 15393 99 24417 189 38218 335 62026 200 20 inch (500 mm) 58 15989 114 25361 197 39696 375 64091 240 24 inch (600 mm) 92 22699 178 36006 304 56357 615 91094
Line Size
1
/2-inch (15 mm) 1 521 1 826 2 1293 6 3333
1
/2 inch (40 mm) 5 1960 7 3109 12 4867 20 8332
(Newton-meter) (Newton) (Newton-meter) (Newton) (Newton-meter) (Newton) (Newton-meter) (Newton)
PN 10 PN 16 PN 25 PN 40
January 2010

INSTALLATION (WAFER SENSOR)

The following section should be used as a guide in the installation of the Rosemount 8711 Sensor. Refer to page 5-7 for installation of the flange-type Rosemount 8705 and 8707 High-Signal sensor.

Gaskets The sensor requires a gasket at each of its connections to adjacent devices or

piping. The gasket material selected must be compatible with the process fluid and operating conditions. Metallic or spiral-wound gaskets can damage the liner. If the gaskets will be removed frequently, protect the liner ends. If grounding rings are used, a gasket is required on each side of the grounding ring.
Alignment and Bolting
1. On 11/2 - through 8-inch (40 through 200 mm) line sizes, place centering rings over each end of the sensor. The smaller line sizes,
0.15- through 1-inch (4 through 25 mm), do not require centering rings.
2. Insert studs for the bottom side of the sensor between the pipe flanges. Stud specifications are listed in Table 5-4. Using carbon
steel bolts on smaller line sizes, 0.15- through 1-inch (4 through 25 mm), rather than the required stainless steel bolts, will degrade performance.
5-10
Page 67
Reference Manual
Customer-supplied
Gasket
FLOW
Installation, Studs Nuts and Washers
Centering Rings
See ”Safety Messages” on pages 5-1 and 5-2 for complete warning information.
00809-0100-4663, Rev BA January 2010
Rosemount 8732
Table 5-4. Stud Specifications
Nominal Sensor Size Stud Specifications
0.15 – 1 inch (4 – 25 mm) 316 SST ASTM A193, Grade B8M Class 1 threaded mounted studs
11/2 – 8 inch (40 – 200 mm) CS, ASTM A193, Grade B7, threaded mounting studs
3. Place the sensor between the flanges. Make sure that the centering rings are properly placed in the studs. The studs should be aligned with the markings on the rings that correspond to the flange you are using.
4. Insert the remaining studs, washers, and nuts.
5. Tighten to the torque specifications shown in Table 5-5. Do not overtighten the bolts or the liner may be damaged .
NOTE
On the 4- and 6- inch PN 10-16, insert the sensor with rings first and then insert the studs. The slots on this ring scen ario are located on the insid e of the ring.
Figure 5-11. Gasket Placement with Centering Rings

Flange Bolts Sensor sizes and torque values for both Class 1 50 and Class 300 flan ges are

listed in Table 5-5. Tighten flange bolts in the incremental sequence, shown in Figure 5-10.
NOTE
Do not bolt one side at a time. Tighten each side simultaneously. Example:
1. Snug left
2. Snug right
3. Tighten left
4. Tighten right Do not snug and tighten the upstream side and then snug and tighten the downstream side. Failure to alternate between the upstream a nd downstream flanges when tightening bolts may result in liner damage.
Always check for leaks at the flanges after tightening the flange bolts. All sensors require a second torquing 24 hours after initial flange bolt tightening.
5-11
Page 68
Reference Manual
User supplied clamp
User supplied gasket
00809-0100-4663, Rev BA
Rosemount 8732
January 2010
Table 5-5. Flange bolt Torque Specifications of Rosemount 8711 Sensors
Size Code Line Size Pound-feet Newton-meter
15F 0.15 inch (4 mm) 5 6.8 30F 0.30 inch (8 mm) 5 6.8 005 010 1 inch (25 mm) 10 13.6 015 1 020 2 inch (50 mm) 25 34.1 030 3 inch (80 mm) 40 54.6 040 4 inch (100 mm) 30 40.1 060 6 inch (150 mm) 50 68.2 080 8 inch (200 mm) 70 81.9
1
/2-inch (15 mm) 5 6.8
1
/2 inch (40 mm) 15 20.5

INSTALLATION (SANITARY SENSOR)

Gaskets The sensor requires a gasket at each of its connections to adjacent devices or

piping. The gasket material selected must be compatible with the process fluid and operating conditions. Gaskets ar e supplied with all Rosemo unt 8721 Sanitary sensors except when the process connection is an IDF sanitary screw type.

Alignment and Bolting Standard plant practices should be followed when installing a magmeter with

Figure 5-12. Rosemount 8721 Sanitary Installation
sanitary fittings. Unique torque values and bolting techniq ue s ar e no t required.
5-12
Process grounding the sensor is one of the most important details of sensor installation. Proper process grounding ensures that th e transmitter amplifier is referenced to the process. This creates the lowest noise environment for the transmitter to make a stable reading. Use Table 5-6 to determine which grounding option to follow for proper installation.
Page 69
Reference Manual
00809-0100-4663, Rev BA January 2010
NOTE
Consult factory for installations requiring cathodic protection or situations where there are high currents or high potential in the process.
The sensor case should always be earth grounded in accordance with national and local electrical codes. Failure to do so may impair the protection provided by the equipment. The most effective grounding method is direct connection from the sensor to earth ground with minimal impedance.
The Internal Ground Connection (Protective Ground Connection) located in side the junction box is the Internal Ground Connection screw. This screw is identified by the ground symbol:
Table 5-6. Grounding Installation
Grounding Options
Type of Pipe No Grounding Options Grounding Rings Grounding Electrodes Lining Protectors
Conductive Unlined Pipe See Figure 5-13 Not Required Not Required See Figure 5-14
Conductive Lined Pipe Insufficient Grounding See Figure 5-14 See Figure 5-13 See Figure 5-14
Non-Conductive Pipe Insufficient Grounding See Figure 5-15 See Figure 5-16 See Figure 5-15
Rosemount 8732
Figure 5-13. No Grounding Options or Grounding Electrode in Lined Pipe
5-13
Page 70
Rosemount 8732
Grounding Rings or
Lining Protectors
Grounding Rings or
Lining Protectors
Figure 5-14. Grounding with Grounding Rings or Lining Protectors
Reference Manual
00809-0100-4663, Rev BA
January 2010
Figure 5-15. Grounding with Grounding Rings or Lining Protectors
5-14
Page 71
Reference Manual
00809-0100-4663, Rev BA January 2010
Figure 5-16. Grounding with Grounding Electrodes
Rosemount 8732
5-15
Page 72
Rosemount 8732
1
/2–14 NPT Conduit
Connection
(no relief valve)
Reference Manual
00809-0100-4663, Rev BA
January 2010

PROCESS LEAK PROTECTION (OPTIONAL)

Standard Housing Configuration

The Rosemount 8705 and 8707 High-Signal Sensor housing is fabricated from carbon steel to perform two separate functions. First, it provides shielding for the sensor magnetics so that external disturbances cannot interfere with the magnetic field and thus affect the flow measu re m en t. Second, it provides the physical protection to the coils and other internal components from contamination and physical d amage that might occur in an industrial environment. The housing is completely welded and gasket-free.
The three housing configurations are identified by the W0, W1, or W3 in the model number option code when ordering. Below are brief description s of each housing configuration, which are followed by a more detailed overview.
Code W0 — sealed, welded coil housing (standard configuration)
Code W1 — sealed, welded coil housing with a relief valve capable of
venting fugitive emissions to a safe location (additional plumbing from the sensor to a safe area, installed by the user, is required to vent properly)
Code W3 — sealed, welded coil housing with separate electrode
compartments capable of venting fugitive emissions (additional plumbing from the sensor to a safe area, installed by the user, is required to vent properly)
The standard housing configuration is identified by a code W0 in the model number. This configuration does not provide separate electrode compartments with external electrode access. In the even t of a pr oc es s leak, these models will not protect the coils or other sensitive areas around the sensor from exposure to the pressure fluid (Figure 5-17).
Figure 5-17. Standard Housing Configuration — Sealed Welded Housing (Option Code W0)
5-16
Page 73
Reference Manual
Optional:
Use drain port to
plumb to a safe area
(Supplied by user)
1
/2 – 14 NPT Conduit
Connection
¼'' NPT – 5 psi
Pressure Relief Valve
00809-0100-4663, Rev BA January 2010
Rosemount 8732

Relief Valves The first optional configuration, identified by the W1 in the model number

option code, uses a completely welded coil housing. This configuration does not provide separate electrode compartments with external electrode access. This optional housing configuration provides a relief valve in the housing to prevent possible overpressuring caused by damage to the lining or other situations that might allow process pressure to enter the housing. The relief valve will vent when the pressure inside the sensor housing exceeds 5 psi. Additional piping (provided by the user) may be connecte d to this relief valve to drain any process leakage to safe containment (see Figure 5-18).
Figure 5-18. Coil-Housing Configuration — Standard Welded Housing With Relief Valve (Option Code W1)

Process Leak Containment

The second optional configuration, identified as option code W3 in the model number, divides the coil housing into three compartments: one for each electrode and one for the coils. Should a damaged liner or electrode fault allow process fluid to migrate behind the electrode seal s, the fluid is contained in the electrode compartment. The sealed electrode compartment prevents the process fluid from entering the coil compartm ent wh er e it w oul d da m age the coils and other internal components.
The electrode compartments are designed to contain the process fluid at full line pressure. An o-ring sealed cover provides access to each of the electrode compartments from outside the se nsor; drain port s are provided in ea ch cover for the removal of fluid.
NOTE
The electrode compartment could contain full line pressure and it must be depressurized before the cover is removed.
5-17
Page 74
Rosemount 8732
Fused Glass Seal
Sealed Electrode Compartment
1
/2 - 27 NPT
O-Ring Seal
Optional:
Use drain port to
plumb to a safe area
(Supplied by user)
Grounding Electrode Port
Figure 5-19. Housing Configuration — Sealed Electrode Compartment (Option Code W3)
Reference Manual
00809-0100-4663, Rev BA
January 2010
If necessary, capture any process fluid leakage, connect the appropriate piping to the drainports, and provide for proper disposal (see Figure 5-19).
5-18
Page 75
Reference Manual
00809-0100-4663, Rev BA January 2010
Section 6 Maintenance and
Troubleshooting
Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 6-1
Installation Check and Guide . . . . . . . . . . . . . . . . . . . . . .page 6-2
Diagnostic Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 6-3
Transmitter Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . page 6-5
Quick Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . .page 6-7
This section covers basic transmitter and sensor troubleshooting. Problems in the magnetic flowmeter system are usually indicated by incorrect output readings from the system, error messages, or failed tests. Consider all sources when identifying a problem in your system. If the problem persists, consult your local Rosemount representative to determine if the material should be returned to the factory. Emerson Process Management offers several diagnostics that aid in the troubleshooting process.
Instructions and procedures in this section may require special precautions to ensure the safety of the personnel performing the operations. Please read the following safety messages before performing any operation described in this section. Refer to these warnings when appropriate throughout this section.
Rosemount 8732

SAFETY INFORMATION

Failure to follow these installation guidelines could result in death or serious injury: Installation and servicing instructions are for use by qualified personnel only. Do not perform
any servicing other than that contained in the operating instructions, unless qualified. Verify that the operating environment of the sensor and transmitter is consistent with the appropriate FM or CSA approval.
Do not connect a Rosemount 8732 to a non-Rosemount sensor that is located in an explosive atmosphere.
Mishandling products exposed to a hazardous substance may result in death or serious injury. If the product being returned was exposed to a hazardous substance as defined by OSHA, a copy of the required Material Safety Data Sheet (MSDS) for each hazardous substance identified must be included with the returned goods.
The 8732 performs self diagnostics on the entire magnetic flowmeter system: the transmitter, the sensor, and the interconnecting wiring. By sequentially troubleshooting each individual piece of the magmeter system, it becomes easier to pin point the problem and make the appropriate adjustments.
If there are problems with a new magmeter installation, se e “In stallation Check and Guide” on page 6-2 for a quick guide to solve the most common installation problems. For existing magmeter installations, Table 6-4 lists the most common magmeter problems and corrective actions.
www.rosemount.com
Page 76
Rosemount 8732
Reference Manual
00809-0100-4663, Rev BA
January 2010

INSTALLATION CHECK AND GUIDE

Use this guide to check new installations of Rosemount magnetic flowmeter systems that appear to malfunction.
Before You Begin Transmitter
Apply power to your system before making the following transmitter checks.
1. Verify that the correct sensor calibration number is entered in the transmitter. The calibration number is listed on the sensor nameplate.
2. Verify that the correct sensor line size is entered in the transmitter. The line size value is listed on the sensor nameplate.
3. Verify that the function blocks are not in Out of Service mode.
4. Verify that the transmitter is functioning correctly by using the 8714i Meter Verification diagnostic or the 8714D Calibration Reference Standard.
Sensor
Be sure that power to your system is removed before beginning sensor checks.
1. For horizontal flow installations, ensure that the electrodes remain covered by process fluid.
For vertical or inclined installations, ensure that the process fluid is flowing up into the sensor to keep the electrodes covered by process fluid.
2. Ensure that the grounding straps on the sensor are connected to grounding rings, lining protectors, or the adjacent pipe flanges. Improper grounding will cause erratic operation of the system.
Wiring for Remote Configurations
1. The signal wire and coil drive wire must be twisted shielded cable. Emerson Process Management, Rosemount division. recommends 20 AWG twisted shielded cable for the elec trodes and 14 AWG twisted shielded cable for the coils.
2. The cable shield must be connected at both ends of the electrode and coil drive cables. Connection of the signal wire shield at both ends is necessary for proper operation. It is recommended that the coil drive wire shield also be connected at both ends for maximum flowmeter performance
3. The signal and coil drive wires must be separate cables, unless Emerson Process Management specified combo cable is used. See Table 2-2 on page 2-11.
4. The single conduit that houses both the signal and coil drive cables should not contain any other wires.
Process Fluid
1. The process fluid conductivity should be 5 microsiemens (5 micro mhos) per centimeter minimum.
2. The process fluid must be free of air and gasses.
3. The sensor should be full of process fluid.
6-2
Page 77
Reference Manual
00809-0100-4663, Rev BA January 2010
Rosemount 8732

DIAGNOSTIC MESSAGES

Problems in the magnetic flowmeter system are usually indicated by incorrect output readings from the system, error messages, or failed te sts. Consider all sources in identifying a problem in your system.
Table 6-1. Rosemount 8732 Basic Diagnostic Messages
Message Potential Cause Corrective Action
“Fieldbus Not
Communicating”
“Sensor Processor Not
Communicating”
“Empty Pipe” Empty Pipe None - message will clear when pipe is full
“Coil Open Circuit” Improper wiring Check coil drive wiring and sensor coils
“Auto Zero Failure”
(Cycle power to clear
messages, no changes
were made)
“Universal Trim Failure” No flow in pipe while performing
“Electronics Failure” Electronics self check failure Replace Electronics
“Electronics Temp Fail” Ambient temperature exceeded the
“Reverse Flow” Electrode or coil wires reverse Verify wiring between sensor and transmitter
“Flow Rate > Sensor Limit” Flow rate is greater than 43 ft/sec Lower flow velocity, increase pipe diameter
“Digital Trim Failure”
(Cycle power to clear
messages, no changes
were made)
Fieldbus segment is disconnected Connect the fieldbus segment Fieldbus segment power missing Verify the segment fieldbus voltage Electronics failure Replace electronics Transmitter input power (AC/DC) is
not connected Electronics failure Replace electronics
Wiring Error Check that wiring matches appropriate wiring diagrams - see Appendix E:
Electrode Error Perform sensor tests C and D (see Table 6-5 on page 6-8) Conductivity less than 5
microsiemens per cm Intermittent Diagnostic Adjust tuning of Empty Pipe parameters
Other manufacturer’s sensor Change coil current to 75 mA
Circuit Board Failure Replace Rosemount 8732 Electronics Coil Circuit OPEN Fuse Return to factory for fuse replacement Flow is not set to zero Force flow to zero, perform autozero Unshielded cable in use Change wire to shielded cable Moisture problems See moisture problems in “Accuracy Section” Empty pipe is present Fill sensor with process fluid
Universal Auto Trim Wiring error Check that wiring matches appropriate wiring diagrams - see “Universal
Flow rate is changing in pipe while performing Universal Auto-Trim routine
Flow rate through sensor is significantly different than value entered during Universal Auto-Trim routine
Incorrect calibration number entered into transmitter for Universal Auto-Trim routine
Wrong sensor size selected Correct sensor size setting - See “Line Size” on page 3-9 Sensor failure Perform sensor tests C and D (see Table 6-5 on page 6-8)
electronics temperature limits
Flow is reverse Turn ON Reverse Flow Enable to read flow Sensor installed backwards Re-install sensor correctly, or switch either the electrode wires (18 and 19)
Improper wiring Check coil drive wiring and sensor coils
The calibrator (8714B/C/D) is not connected properly
Incorrect calibration number entered into transmitter
Calibrator is not set to 30 FPS Change calibrator setting to 30 FPS Bad calibrator Replace calibrator
Connect the input power. If the LCD displays a message, the input power is applied
Universal Sensor Wiring Diagrams
Increase Conductivity to greater than or equal to 5 microsiemens per cm
Perform sensor test A - Sensor Coil
Perform a Universal Auto Trim to select the proper coil current
Establish a known flow in sensor, and perform Universal Auto-Trim calibration
Sensor Wiring Diagrams” on page E-1 Establish a constant flow in sensor, and perform Universal Auto-Trim
calibration
Verify flow in sensor and perform Universal Auto-Trim calibration
Replace sensor calibration number with 1000005010000001
Move transmitter to a location with an ambient temperature range of -40 to 165 °F (-40 to 74 °C)
or the coil wires (1 and 2)
Perform sensor test A - Sensor Coil (see Table 6-5 on page 6-8) Review calibrator connections
Replace sensor calibration number with 1000005010000001
6-3
Page 78
00809-0100-4663, Rev BA
Rosemount 8732
Table 6-2. Rosemount 8732 Advanced Diagnostic Messages (Suite 1 - Option Code D01)
Message Potential Cause Corrective Action
Grounding/Wiring Fault Improper installation of wiring See “Sensor Connections” on page 2-11
Coil/Electrode shield not connected
Improper process grounding See “Grounding” on page 5-12 Faulty ground connection Check wiring for corrosion, moisture in the terminal block, and refer to
Sensor not full Verify sensor is full and empty pipe diagnostic is on
High Process Noise Slurry flows - mining/pulp stock Decrease the flow rate below 10 ft/s (3 m/s)
Chemical additives upstream of the sensor
Electrode not compatible with the process fluid
Air in line Move the sensor to another location in the process line to ensure that it is
Electrode coating Use bulletnose electrodes
Styrofoam or other insulating particles
Low conductivity fluids (below 10 microsiemens/cm)
See “Sensor Connections” on page 2-11
“Grounding” on page 5-12
Complete the possible solutions listed under “Step 2: Process Noise” on page 6-7
Move injection point downstream of the sensor, or move the sensor Complete the possible solutions listed under “Step 2: Process Noise” on page 6-7
Refer to the Rosemount Magnetic Flowmeter Material Selection Guide (00816-0100-3033)
full under all conditions
Downsize sensor to increases flowrate above 3 ft/s (1 m/s) Periodically clean sensor
Complete the possible solutions listed under “Step 2: Process Noise” on page 6-7 Consult factory
Trim electrode and coil wires - refer to “Installation” on page 2-1
Reference Manual
January 2010
Table 6-3. Rosemount 8732 Advanced Diagnostic Messages (Suite 2 - Option Code D02)
Message Potential Cause Corrective Action
8714i Failed
Transmitter Calibration Verification test failed
Sensor Calibration test failed Verify pass/fail criteria
Sensor Coil Circuit test failed Verify pass/fail criteria
Sensor Electrode Circuit test failed Verify pass/fail criteria
Verify pass/fail criteria Rerun 8714i Meter Verification under no flow conditions Verify calibration using 8714D Calibration Standard Perform digital trim Replace electronics board
Perform sensor test - see Table 6-5 on page 6-8
Perform sensor test - see Table 6-5 on page 6-8
Perform sensor test - see Table 6-5 on page 6-8
6-4
Page 79
Reference Manual
00809-0100-4663, Rev BA January 2010

TRANSMITTER TROUBLESHOOTING

Table 6-4. Advanced Troubleshooting–Rosemount 8732
Symptom Potential Cause Corrective Action
Does not appear to be within rated accuracy
Noisy Process Chemical additives upstream of
Transmitter, control system, or other receiving device not configured properly
Electrode Coating Use bulletnose electrodes;
Air in line Move the sensor to another location in the process line to
Moisture problem Perform the sensor Tests A, B, C, and D
Improper wiring If electrode shield and signal wires are switched, flow indication
Flow rate is below 1 ft/s (specification issue)
Auto zero was not performed when the coil drive frequency was changed from 5 Hz to 37 Hz
Sensor failure–Shorted electrode Perform the sensor Tests C and D
Sensor failure–Shorted or open coil Perform the sensor Tests A and B
Transmitter failure Verify transmitter operation with an 8714 Calibration S tandard or
magnetic flowmeter
Sludge flows–Mining/Coal/ Sand/Slurries (other slurries with hard particles)
Styrofoam or other insulating particles in process
Electrode coating Use replaceable electrodes in Rosemount 8705.
Air in line Move the sensor to another location in the process line to
Low conductivity fluids (below 10 microsiemens/cm)
Advanced Troubleshooting continued on next page
Rosemount 8732
Check all configuration variables for the transmitter, sensor, communicator, and/or control system
Check these other transmitter settings:
•Sensor calibration number
•Units
•Line size
Downsize sensor to increase flow rate above 3 ft/s; Periodically clean sensor
ensure that it is full under all conditions.
(see Table 6-5 on page 6-8)
will be about half of what is expected. Check wiring diagrams for your application.
See accuracy specification for specific transmitter and sensor
Set the coil drive frequency to 37 Hz, verify the sensor is full, verify there is no flow, and perform the auto zero function.
(see Table 6-5 on page 6-8)
(see Table 6-5 on page 6-8)
replace the electronic board Complete the Noisy Process Basic procedure. Move injection
point downstream of magnetic flowmeter, or move magnetic flowmeter.
Decrease flow rate below 10 ft/s
Complete the Noisy Process Basic procedure; Consult factory
Use a smaller sensor to increase flow rate above 3 ft/s. Periodically clean sensor.
ensure that it is full under all conditions.
• Trim electrode and coil wires – see “Conduit Cables” on page 2-6
• Keep flow rate below 3 FPS
• Integral mount transmitter
• Use 8712-0752-1,3 cable
• Use N0 approval sensor
6-5
Page 80
Rosemount 8732
Table 6-4. Advanced Troubleshooting–Rosemount 8732
Symptom Potential Cause Corrective Action
Meter output is unstable Medium to low conductivity fluids (10–
25 microsiemens/cm) combined with cable vibration or 60 Hz interference
Electrode incompatibility Check the Technical Data Sheet, Magnetic Flowmeter Material
Improper grounding Check ground wiring – see “Mount the Transmitter” on page 2-3
High local magnetic or electric fields Move magnetic flowmeter (20–25 ft away is usually acceptable) Control loop improperly tuned Check control loop tuning Sticky valve (look for periodic
oscillation of meter output) Sensor failure Perform the sensor Tests A, B, C, and D
Reading does not appear to be within rated accuracy
Transmitter, control system, or other receiving device not configured properly
Electrode coating Use bulletnose electrodes in the Rosemount 8705 Sensor.
Air in line Move the sensor to another location in the process line to
Flow rate is below 1 ft/s (specification issue)
Insufficient upstream/downstream pipe diameter
Cables for multiple magmeters run through same conduit
Auto zero was not performed when the coil drive frequency was changed from 5 Hz to 37.5 Hz
Sensor failure—shorted electrode See Table 6-5 on page 6-8 Sensor failure—shorted or open coil See Table 6-5 on page 6-8 Transmitter failure Replace the electronics board Transmitter wired to correct sensor Check wiring
Reference Manual
00809-0100-4663, Rev BA
January 2010
Eliminate cable vibration:
• Integral mount
• Move cable to lower vibration run
• Tie down cable mechanically
• Trim electrode and coil wires
• See “Conduit Cables” on page 2-6
• Route cable line away from other equipment powered by 60 Hz
• Use 8712-0752-1,3 cable
Selection Guide (document number 00816-0100-3033), for chemical compatibility with electrode material.
for wiring and grounding procedures
Service valve
(See Table 6-5 on page 6-8) Check all configuration variables for the transmitter, sensor,
communicator, and/or control system
Check these other transmitter settings: Sensor calibration number Units Line size
Downsize the sensor to increase the flow rate above 3 ft/s. Periodically clean the sensor
ensure that it is full under all conditions See the accuracy specification for specific transmitter and
sensor Move sensor to location where 5 pipe diameters upstream and 2
pipe diameters downstream is possible Run only one conduit cable between each sensor and
transmitter Perform the auto zero function with full pipe and no flow
6-6
Page 81
Reference Manual
00809-0100-4663, Rev BA January 2010
Rosemount 8732

QUICK TROUBLESHOOTING

Step 1: Wiring Errors The most common magmeter problem is wiring between the sensor and the

transmitter in remote mount installations. The signal wire and coil drive wire must be twisted shielded cable: 20 AWG twisted shielded cable for the electrodes and 14 AWG twisted shielded cable for the coils. Ensure that the cable shield is connected at both ends of the electrode and coil drive cables. Signal and coil drive wires must have their own cable s. The single conduit that houses both the signal and coil drive cables should not contain any other wires. For more information on proper wir ing practices, refe r to “T ran smitter to Sensor Wiring” on page 2-11.

Step 2: Process Noise In some circumstances, process conditions rather than the magmeter can

cause the meter output to be unstable. Possible solutions for addressing a noisy process situation are given below. When the output attains the desired stability, no further steps are required.
Use the Auto Zero function to initialize the transmitter for use with the 37 .5 Hz coil drive mode only. Run this function only with the transmitter and sensor installed in the process. The sensor must be filled with process fluid with zero flow rate. Before running the auto zero function, be sure the coil drive mode is set to 37.5 Hz.

Step 3: Installed Sensor Tests

Set the loop to manual if necessary and begin the auto zero procedure. The transmitter completes the procedure automatically in about 90 seconds. A symbol appears in the lower right-hand corner of the display to indicate that the procedure is running.
1. Change the coil drive to 37.5 Hz. Complete the Auto Zero function, if possible (see “Coil Drive Frequency” on page 4-13).
2. Turn on Digital Signal Processing (see “Signal Processing” on page 4-25)
3. Increase the damping (see “Damping” on page 3-17).
If the preceding steps fail to resolve the process noise symptoms, consult your Rosemount sales representative about using a high-sig nal magnetic flowmeter system.
If a problem with an installed sensor is identified, Table 6-5 can assist in troubleshooting the sensor. Before performing any of the sensor tests, disconnect or turn off power to the transmitter. To interpret the results, the hazardous location certification for the sensor must be known. Applicable codes for the Rosemount 8705 are N0, N5, and KD. Applicable codes for the Rosemount 8707 are N0 and N5. Applicable codes for the Rosemount 8711 are N0, N5, E5, and KD. Always check the operation of test equipment before each test.
If possible, take all readings from inside the sensor junction box. If the sensor junction box is inaccessible, take measurements as close as possible. Readings taken at the terminals of remote-mount transmitters that are more than 100 feet away from the sensor may provide incorrect or inconclusive information and should be avoided. A sensor circuit diagram is provided in Figure 6-1 on page 6-9.
6-7
Page 82
Rosemount 8732
2R18
0.2


R1R–2300
R1R21500
1nanosiemens
1
1gigaohm
----------------------------=
1nanosiemens
1
1109ohm
------------------------------ -=
Table 6-5. Sensor Test
Test
A. Sensor Coil
B. Shields to Case
C. Coil Shield to Coil
D. Electrode Shield to Electrode
Sensor Location
Installed or Uninstalled
Installed or Uninstalled
Installed or Uninstalled
Installed LCR (Set to
Required Equipment
Multimeter 1 and 2 = R
Multimeter 17 and
Multimeter 1 and
Resistance and 120 Hz)
Measuring at Connections
and case
ground
17 and case ground
2 and
18 and 17 = R 19 and 17 = R
Reference Manual
00809-0100-4663, Rev BA
January 2010
Expected Value Potential Cause Corrective Action
(< 1nS) (< 1nS)
R1 and R2 should be stable
1
NO:
2
N5, E5, CD, ED:
• Open or Shorted Coil
• Moisture in terminal block
• Leaky electrode
• Process behind liner
• Process behind liner
• Leaky electrode
• Moisture in terminal block
• Unstable R1 or
values
R
2
confirm coated electrode
• Shorted electrode
• Electrode not in contact with process
• Empty Pipe
• Low conductivity
• Leaky electrode
• Remove and replace sensor
• Clean terminal block
• Remove sensor
• Remove sensor and dry
• Clean terminal block
• Confirm with sensor coil test
• Remove coating from sensor wall
• Use bulletnose electrodes
• Repeat measurement
• Pull sensor, complete test in Table 6-6 and Table 6-7 on page 6-10 out of line.
To test the sensor, a multimeter capable of measuring conductance in nanosiemens is preferred. Nanosiemens is the reciprocal of resistance.
or
6-8
Page 83
Reference Manual
68.1k(not applicable for sensors with N0 hazardous certification approval option code)
Sensor Housing
68.1k
See “Safety Information” on page 6-1 for complete warning information.
00809-0100-4663, Rev BA January 2010
Figure 6-1. Sensor Circuit Diagram
Rosemount 8732

Step 4: Uninstalled Sensor Tests

An uninstalled sensor can also be used for sensor troubleshooting. To interpret the results, the hazardous location certification for the sensor must be known. Applicable codes for the Rosemount 8705 are N0, N5, and KD. Applicable codes for the Rosemount 8707 are N0 and N5. Applicable codes for the Rosemount 8711 are N0, N5, E5, and KD.
A sensor circuit diagram is provided in Figure 6-1. Take measurements from the terminal block and on the electrode head inside th e se nso r. The measurement electrodes, 18 and 19, are on opposite sides in the inside diameter. If applicable, the third grounding electrode is in between the other two electrodes. On Rosemount 8711 sensors, electrode 18 is ne ar the sensor junction box and electrode 19 is near the bottom of the sensor (Figure 6-2). The different sensor models will have slightly different resistance readings. Flanged sensor resistance readings are in Table 6-6 while wafer sensor resistance readings are in Table 6-7.
6-9
Page 84
Rosemount 8732
275
61kR75k 
275
61kR75k 
0.3
0.3
0.3
0.3
0.3
61kR75k 
275
61kR75k 
0.3
0.3
0.3
0.3
Figure 6-2. 45° Electrode Plane
Reference Manual
00809-0100-4663, Rev BA
January 2010
To insure accuracy of resistance readings, zero out multimeter by shorting and touching the leads together.
Table 6-6. Uninstalled Rosemount 8705 / 8707 Flanged Sensor Tests
Hazardous Location Certifications
Measuring at Connections
18 and Electrode 19 and Electrode 17 and Grounding Electrode
17 and Ground Symbol 17 and 18 Open Open
17 and 19 Open Open 17 and 1 Open Open
(1) It is difficult to tell from visual inspection alone which electrode is wired to which number terminal in
the terminal block. Measure both electrodes. One electrode shoul d result in an open reading, while the other electrode should be less than 275 .
(1)
(1)
N0 N5, KD
Table 6-7. Uninstalled Rosemount 8711 Wafer Sensor Tests
Hazardous Location Certification
Measuring at Connections
18 and Electrode 19 and Electrode 17 and Grounding Electrode
17 and Grounding Symbol 17 and 18 Open Open
17 and 19 Open Open 17 and 1 Open Open
(1) Measure the electrode closest to the junction box (2) Measure the electrode farthest away from the junction box.
(1)
(2)
N0 N5, E5, CD
6-10
Page 85
Reference Manual
Supply Current (Amps)
0.100
0.120
0.140
0.180
0.200
0.220
0.240
0.260
0.280
0.300
0.320
0.160
80 100 120 140 160 180 200 220 240
Power Supply Voltage (AC RMS)
00809-0100-4663, Rev BA January 2010

Appendix A Reference Data

Functional Specifications . . . . . . . . . . . . . . . . . . . . . . . . . page A-1
Foundation™ fieldbus Specifications . . . . . . . . . . . . . . .page A-4
Performance Specifications . . . . . . . . . . . . . . . . . . . . . . . page A-5
Physical Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . .page A-7
Rosemount 8732

FUNCTIONAL SPECIFICATIONS

Sensor Compatibility
Compatible with Rosemount 8705, 8711, 8721, and 570TM sensors. Compatible with Rosemount 8707 sensor with D2 Dual calibration option. Compatible with AC and DC powered sensors of other manufacturers.
Sensor Coil Resistance
350 maximum
Flow Rate Range
Capable of processing signals from fluids that are travelin g between 0.04 an d 39 ft/s (0.01 to 12 m/s) for both forward and reverse flow in all sensor sizes. Full scale continuously adjustable between –39 and 39 ft/s (–12 to 12 m/s).
Conductivity Limits
Process liquid must have a conductivity of 5 microsiemens/cm (5 micromhos/cm) or greater for 8732E. Excludes the effect of interconnecting cable length in remote mount transmitter installations.
Power Supply
90 -250 V AC ±10%, 50–60 Hz or 12-42 V DC
AC Power Supply Requirements
Units powered by 90-250 V AC have the following power requirements.
Figure A-1. AC Current Requirements
www.rosemount.com
Page 86
Rosemount 8732
80 100 120 140 160 180 200 220 240
Power Supply Voltage (AC RMS)
20
22
24
26
28
30
32
34
36
38
Apparent Power (VA)
250
0
0.25
0.5
0.75
1
12 18 24 30 36 42
Power Supply (Volts)
Supply Current (Amps)
Figure A-2. Apparent Power
Figure A-3. DC Current Requirements
Reference Manual
00809-0100-4663, Rev BA
January 2010
DC Supply Current Requirements
Units powered by 12-42 V DC power supply may draw up to 1 amp of curr ent steady state.
Installation Coordination
Installation (overvoltage) Catego ry II
Power Consumption
10 watts maximum
Switch-on current
AC: Maximum 26 A (< 5 ms) at 250 V AC DC: Maximum 30 A (< 5 ms) at 42 V DC
Ambient Temperature Limits
Operating
–58 to 165 °F (–50 to 74 °C) without local operator interface 13 to 149 °F (–25 to 65 °C) with local operator interface
A-2
Storage
–40 to 185 °F (–40 to 85 °C) –22 to 176 °F (–30 to 80 °C) with local operator interface
Humidity Limits
0–100% RH to 150 °F (65 °C)
Page 87
Reference Manual
00809-0100-4663, Rev BA January 2010
Rosemount 8732
Enclosure Rating
NEMA 4X CSA Type 4X, IEC 60529, IP66 (transmitter), Pollution Degree 2
Output Signal
Manchester-encoded digital signal that conforms to IEC 1158-2 and ISA 50.02
A-3
Page 88
Rosemount 8732
Reference Manual
00809-0100-4663, Rev BA
January 2010

FOUNDATION™ FIELDBUS SPECIFICATIONS

Table A-1. Block Information
Schedule Entries
Seven (7)
Links
Twenty (20)
Virtual Communications Relationships (VCRs)
One (1) predefined (F6, F7) Nineteen (19) configurable (see Table 1)
Block Execution Time (Milliseconds)
Resource (RB)
Transducer (TB)
Analog Input (AI) 10
Proportional/Integral/
Derivative (PID)
Integrator (INT) 10 Arithmetic (AR) 10
10
Reverse Flow
Detects and reports reverse flow
Software Lockout
A write-lock switch and software lockout are provided in the resource function block.
Turn-on Time
5 minutes to rated accuracy from power up; 10 seco n ds fr om pow er interruption.
Start-up Time
50 ms from zero flow.
Low Flow Cutoff
Adjustable between 0.01 and 38.37 ft/s (0.003 and 11.7 m/s). Below selected value, output is driven to the zero flow rate signal level.
Overrange Capability
Signal output will remain linear until 1 10% of upper range value or 44 ft/s (13 m/s). The signal output will remain constant above these values. Out of range message displayed on local display and field communicator.
Damping
Adjustable between 0 and 256 seconds.
Sensor Compensation
Rosemount sensors are flow-calibrated and assigned a calibration factor at the factory. The calibration factor is entered into the transmitter, enabling interchangeability of sensors without calculations or a compromise in standard accuracy.
A-4
Page 89
Reference Manual
0
0.50.5
1.01.0
1.51.5
2.02.0
2.52.5
0
3 3 (1)(1)
6 6 (2)(2)
13 13 (4)(4)
20 20 (6)(6)
27 27 (8)(8)
33 33 (10)(10)
40 40 (12)(12)
Velocity in ft/s (m/s)Velocity in ft/s (m/s)
% of Rate
0.25%0.25%
0.15%0.15%
00809-0100-4663, Rev BA January 2010
Rosemount 8732
8732E transmitters and other manufacturer’s sensors can be calibrated at known process conditions or at the Rosemount NIST-Traceable Flow Facility. Transmitters calibrated on site require a two-step procedure to match a known flow rate. This procedure can be found in “Universal Trim” on page 4-11.
Diagnostics
Basic
Self test Transmitter faults Tunable empty pipe Reverse flow Coil circuit fault Electronics temperature
Advanced (D01 Suite)
Ground/wiring fault High process noise
Advanced (D02 Suite)
8714i Meter Verification

PERFORMANCE SPECIFICATIONS

(System specifications are given using the frequency output and with the unit at reference conditions.)
Accuracy
Includes the combined effects of linearity, hysteresis, repeatability, and calibration uncertainty.
Rosemount 8732E with 8705/8707 Sensor:
Standard system accuracy is ±0.25% of rate ±1.0 mm/sec from 0.04 to 6 ft/s (0.01 to 2 m/s); above 6 ft/s (2 m/s), the system has an accuracy of ±0.25% of rate ±1.5 mm/sec.
Optional high accuracy is ±0.15% of rate ±1.0 mm/sec from 0.04 to 13 ft/s (0.01 to 4 m/s); above 13 ft/s (4 m/s), the system has an accuracy of ±0.18% of rate.
(1)
Rosemount 8732E with 8711 Sensor:
Standard system accuracy is ±0.25% of rate ±2.0 mm/sec from 0.04 to 39 ft/s (0.01 to 12 m/s).
(1) For Sensor sizes greater than 12 in. (300 mm) the high accuracy is ±0.25% of rate from 3 to
39 ft/sec (1 to 12 m/sec).
A-5
Page 90
Rosemount 8732
0
0.50.5
1.01.0
1.51.5
2.02.0
2.52.5
0
3 3 (1)(1)
6 6 (2)(2)
13 13 (4)(4)
20 20 (6)(6)
27 27 (8)(8)
33 33 (10)(10)
40 40 (12)(12)
Velocity in ft/s (m/s)Velocity in ft/s (m/s)
% of Rate% of Rate
0.25%0.25%
0.15%0.15%
0
0.50.5
1.01.0
1.51.5
2.02.0
2.52.5
0
3 3 (1)(1)
6 6 (2)(2)
13 13 (4)(4)
20 20 (6)(6)
27 27 (8)(8)
33 33 (10)(10)
40 40 (12)(12)
Velocity in ft/s (m/s)Velocity in ft/s (m/s)
% of Rate% of Rate
0.5%0.5%
0.25%0.25%
Reference Manual
00809-0100-4663, Rev BA
January 2010
Optional high accuracy is ±0.15% of rate ±1.0 mm/sec from 0.04 to 13 ft/s (0.01 to 4 m/s); above 13 ft/s (4 m/s), the system has an accuracy of ±0.18% of rate.
Rosemount 8732E with 8721 Sensor:
Stan dard system accuracy is ±0.5% of rate from 1 to 39 ft/s (0.3 to 12 m/s); between 0.04 and 1.0 ft/s (0.01 and 0.3 m/s), the system has an accuracy of ±0.005 ft/s (0.0015 m/s).
Optional high accuracy is ±0.25% of rate from 3 to 39 ft/s (1 to 12 m/s).
Rosemount 8732E with Legacy 8705 Sensors:
Stan dard system accuracy is ±0.5% of rate from 1 to 39 ft/s (0.3 to 12 m/s); between 0.04 and 1.0 ft/s (0.01 and 0.3 m/s), the system has an accuracy of ±0.005 ft/s (0.0015 m/s).
Rosemount 8732E with Legacy 8711 Sensors:
Standard system accuracy is ±0.5% of rate from 3 to 39 ft/s (1 to 12 m/s); between 0.04 and 3.0 ft/s (0.01 and 1 m/s), the system has an accuracy of ±0.015 ft/s (0.005 m/s).
Rosemount 8732E with Other Manufacturers’ Sensors:
When calibrated in the Rosemount Flow Facility, system accuracies as good as 0.5% of rate can be attained.
There is no accuracy specification for other manufacturers’ sensors calibrated in the process line.
Vibration Effect
IEC 60770-1
Repeatability
±0.1% of reading
A-6
Page 91
Reference Manual
00809-0100-4663, Rev BA January 2010
Rosemount 8732
Stability
±0.1% of rate over six months
Ambient Temperature Effect
±0.25% change over operating temperature range
EMC Compliance
EN61326-1 1997 + A1/A2/A3 (Industrial) electromagnetic compatibility (EMC) for process and laboratory apparatus.

PHYSICAL SPECIFICATIONS

Materials of Construction
Housing
Low copper aluminum, NEMA 4X and IEC 60529 IP66 Pollution Degree 2
Paint
Polyurethane
Cover Gasket
Rubber
Electrical Connections
1
Two
/2–14 NPT connections provided on the transmitter housing (optional third connection available). PG13.5 and CM20 adapters are available. Screw terminals provided for all connections. Powe r wiring co nn e cte d to tran sm itt er only. Integrally mounted transmitters are factory wired to the sensor.
Transmitter Weight
Approximately 7 pounds (3.2 kg). Add 1 pound (0.5 kg) for Option Code M5.
A-7
Page 92
Rosemount 8732
Reference Manual
00809-0100-4663, Rev BA
January 2010
A-8
Page 93
Reference Manual
00809-0100-4663, Rev BA January 2010
Rosemount 8732

Appendix B Approval Information

Product Certifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . page B-1
Approved Manufacturing Locations . . . . . . . . . . . . . . . . .page B-1
European Directive Information . . . . . . . . . . . . . . . . . . . .page B-1
Hazardous Locations Product Approvals Offering . . . . .page B-3
Hazardous Location Certifications . . . . . . . . . . . . . . . . . .page B-5

PRODUCT CERTIFICATIONS

Approved Manufacturing Locations

Rosemount Inc. — Eden Prairie, Minnesota, USA Fisher-Rosemount Technologias de Flujo, S.A. de C.V. —
Chihuahua Mexico Emerson Process Management Flow — Ede, The Netherlands Emerson Process Management Flow Technologies Co., Ltd. — Nanjing, China

European Directive Information

The EC declaration of conformity for all applicable European directives for this product can be found on our website at www.rosemount.com. A hard copy may be obtained by contacting our local sales office.

ATEX Directive

Rosemount Inc. complies with the ATEX Directive.
Type n protection type in accordance with EN50 021
• Closing of entries in the device must be carried out using the appropriate EExe or EExn metal cable gland and metal blanking plug or any appropriate ATEX approved cable gland and blanking plug with IP66 rating certified by an EU approved certification body.
For Rosemount 8732E transmitters:
Complies with Essential Health and Safety Requirements: EN 60079-0: 2006 IEC 60079-1: 2007 EN 60079-7: 2007 EN 60079-11: 2007 EN 60079-26: 2004 EN 50281-1-1: 1998 + A1

European Pressure Equipment Directive (PED) (97/23/EC)

www.rosemount.com
Rosemount 8705 and 8707 Magnetic Flowmeter sensors in line size and flange
combinations:
1
Line Size: 1
Line Size: 30 in. - 36 in. with AWWA 125 flanges
Rosemount 8711 Magnetic Flowmeter Sensors
/2 in. - 24 in. with all DIN flanges and ANSI 150 and
ANSI 300 flanges. Also available with ANSI 600 flanges in limited line sizes.
QS Certificate of Assessment - EC No. PED-H-20 Module H Conformity Assessment
Line Sizes: 1.5, 2, 3, 4, 6, and 8 in.
QS Certificate of Assessment - EC No. PED-H-20 Module H Conformity Assessment
Page 94
Rosemount 8732
Reference Manual
00809-0100-4663, Rev BA
January 2010
Rosemount 8721 Sanitary Magmeter Sensors
in line sizes of 1
Module A Conformity Assessment
All other Rosemount 8705/8707/8711/8721
Sensors — in line sizes of 1 in. and less: Sound Engineering Practice
Sensors that are SEP are outside the scope of PED and cannot be marked for compliance with PED.
Mandatory CE-marking for sensors in accordance with Article 15 of the PED can be found on the sensor body (CE 0575).
Sensor category I is assessed for conformity per module A procedures. Sensor categories II – IV, use module H for conformity assessment procedures.
1
/2 in. and larger:

Electro Magnetic Compatibility (EMC) (2004/108/EC)

Models 8712D - EN 50081-1: 1992, EN 50082-2: 1995, Model 8732E - EN 61326: 1997: A1 + A2 + A3
Installed signal wiring should not be run together and should not be in the same cable tray as AC power wiring.
Device must be properly grounded or earthed according to local electric codes. To improve protection against signal interference, shielded cable is recommended.

Low Voltage Directive (93/68/EEC)

Model 8712D - EN 61010 -1: 1995

Low Voltage Directive (2006/95/EC)

Model 8732E - EN 61010 -1: 2001

Other important guidelines

Only use new, original parts. To prevent the process medium escaping, do not unscrew or remove process flange bolts,
adapter bolts or bleed screws during operation. Maintenance shall only be done by qualified personnel.
CE CE Marking
Compliance with all applicable European Union Directives. (Note: CE Marking is not available on Rosemount 8712H).

IECEx Scheme

For Rosemount 8732E transmitters:
Rosemount complies with all of the stated standards below: IEC 60079-0 : 2004 IEC 60079-1 : 2007-04 IEC 60079-11 : 2006 IEC 60079-26 : 2006 IEC 60079-7 : 2006-07 IEC 61010-1 : 2001 IEC 61241-0 : 2004 IEC 61241-1 : 2004
C-Tic Marking
Complies with IEC 61326-1 : 1997 + A1, A2, A3.
B-2
Page 95
Reference Manual
00809-0100-4663, Rev BA January 2010
Rosemount 8732

HAZARDOUS LOCATIONS PRODUCT APPROVALS OFFERING

Table B-1. Factory Mutual (FM) Approvals Offering
The Rosemount 8700 Series magnetic flowmeters offer many different hazardous locations certifications. The table below provides an overview of the available hazardous area approval options. Equivalent hazardous locations certifications for sensor and transmitter must match in integrally mounted magnetic flowmeter systems. Remote mounted magnetic flowmeter systems do not require matched hazardous location certifications. For complete information about the hazardous area approval codes listed, see Hazardous Location Certifications starting on page B-5.
Transmitter 8732E 8712D
Sensor 8705 8707 8711 8705 8707 8711 8707
FM Category Hazardous Area Approval Code
Non-Classified Locations
Suitable for Class I, Division 1 Explosion-Proof
Explosion-Proof with Intrinsically Safe Output
Suitable for Class I, Division 2 Non-Flammable Fluids
Flammable Fluids
Non-Flammable Fluids with Intrinsically Safe Output
Other Certifications Product Certification Code
European Pressure Equipment Directive (PED) PD - PD PD - PD ­NSF 61 Drinking Water
(1) Remote Transmitter Only (2) Available in line sizes 0.5 in. to 8 in. (15 mm to 200 mm) only (3) For I.S. Output, Output Code B must be ordered (4) 8707 Sensor has Temp Code - T3C (5) Product Certification Codes are added to the Sensor model number only (6) Only available with PTFE (all line sizes) or Polyurethane (4 in. or larger) Lining Materials and
316L SST Electrodes
Transmitter NANANANANANA N0
Sensor NA N0 NA NA N0 NA N0
(2)
Trans: Groups C, D T6 E5
Sensor: Groups C, D T6 E5
Trans: Groups C, D T6 E5
Sensor: Groups C, D T6 E5
Trans: Groups A,B,C,D T4 N0 N0 N0 N0 N0 N0 N0
Sensor: Groups A,B,C,D T5 N0 N0
Trans: Groups A,B,C,D T4 N5 N5 N5 N5 N5 N5 N5
Sensor: Groups A,B,C,D T5 N5 N5
Trans: Groups A,B,C,D T4 N0
Sensor: Groups A,B,C,D T5 N0 N0
(6)
-E5- - - -
(2)
- E5 - - - -
(2)(3)
(2)
(3)N0(3)N0(3)
DW - DW DW - DW -
(3)
- E5
-E5- - - -
(4)
(4)
(4)
- - - -
N0 N0 N0
N5 N5 N5
- - - -
N0 - - - -
(1)
(4)
(4)
N0 N0
N5 N5
(5)
8712H
(1)
(4)
(4)
Table B-2. Canadian Standards Association (CSA) Approvals Offering
Transmitter 8732E 8712D
Sensor 8705 8707 8711 8705 8707 8711 8707
CSA Category Hazardous Area Approval Code
Non-Classified Locations
Suitable for Class I, Division 2 Non-Flammable Fluids
Other Certifications Product Certification Code
European Pressure Equipment Directive (PED) PD - PD PD - PD -
(1) Remote Transmitter Only (2) 8707 Sensor has Temp Code - T3C (3) Product Certification Codes are added to the Sensor model number only (4) Only available with PTFE (all line sizes) or Polyurethane (4 in. or larger) Lining Materials and
316L SST Electrodes
Transmitter NA - NA NA - NA -
Sensor NA - NA NA - NA -
Trans: Groups A,B,C,D T4 N0 N0 N0 N0 N0 N0 N0
Sensor: Groups A,B,C,D T5 N0 N0
NSF 61 Drinking Water
(4)
(2)
N0 N0 N0
DW - DW DW - DW -
(1)
(2)
N0 N0
(3)
8712H
(1)
(2)
B-3
Page 96
Rosemount 8732
Reference Manual
00809-0100-4663, Rev BA
January 2010
Table B-3. ATEX Approvals Offering
Transmitter 8732E 8712D
Sensor 8705 8707 8711 8705 8707 8711 8707
ATEX Category Hazardous Area Approval Code
(1)
8712H
Non-Hazardous
Trans: LVD and EMC NA - NA NA - NA -
Sensor: LVD and EMC NA - NA NA - NA -
Equipment Category 2
Gas Group IIB
Trans: Ex d IIB T6 ED - ED - - - -
Sensor: Ex e ia IIC T3...T6 KD
(2)
- KD
(2)
- - - -
Gas Group IIC
Trans: Ex d IIC T6 E1 - E1 - - - -
Sensor: Ex e ia IIC T3...T6 E1 - E1 - - - -
Gas Group IIB with Intrinsically Safe Output
Trans: Ex de [ia] IIB T6 ED
Sensor: Ex e ia IIC T3...T6 KD
Gas Group IIC with Intrinsically Safe Output
Trans: Ex de [ia] IIC T6 E1
Sensor: Ex e ia IIC T3...T6 E1 - E1 - - - -
(3) (2)
(3)
(3)
-ED
(2)
- KD
(3)
- E1
--- -
- - - -
- - - -
Equipment Category 3
Gas Group IIC
Trans: Ex nA nL IIC T4 N1 - N1 N1 - N1 -
Sensor: Ex nA [L] IIC T3...T6 N1 - N1 N1 - N1 -
Equipment Category 1 - Dust Environment
Dust Environment Only
Trans: Dust Ignition Proof ND - ND - - - -
Other Certifications Product Certification Code
European Pressure Equipment Directive (PED) PD - PD PD - PD ­NSF 61 Drinking Water
(1) Remote Transmitter Only (2) With integral mount transmitter, approval is valid for Gas Group IIB (3) For I.S. Output, Output Code B must be ordered (4) Product Certification Codes are added to the Sensor model number only (5) Only available with PTFE (all line sizes) or Polyurethane (4 in. or larger) Lining Materials and
Sensor: Dust Ignition Proof ND - ND - - - -
(5)
316L SST Electrodes
DW - DW DW - DW -
(4)
(1)
Table B-4. IECEx Approvals Offering
B-4
Transmitter 8732E
(1)
Sensor 8705 8707 8711 IECEx Category Hazardous Area Approval Code
Non-Hazardous
Trans: LVD and EMC NA - NA
Sensor: LVD and EMC NA - NA
Equipment Category 2
Gas Group IIB
Trans: Ex d IIB T6 EF - EF
Gas Group IIC
Trans: Ex d IIC T6 E7 - E7
Gas Group IIB with Intrinsically Safe Output
Trans: Ex de [ia] IIB T6 EF
(2)
-EF
(3)
Gas Group IIC with Intrinsically Safe Output
Trans: Ex de [ia] IIC T6 E1
(3)
-E1
(3)
Equipment Category 3
Gas Group IIC
Trans: Ex nA nL IIC T4 N7 - N7
Page 97
Reference Manual
00809-0100-4663, Rev BA January 2010
Rosemount 8732
Equipment Category 1 - Dust Environment
Dust Environment Only
Trans: Dust Ignition Proof NF - NF
Other Certifications Product Certification Code
European Pressure Equipment Directive (PED)
NSF 61 Drinking Water
(1) Available in remote mount configuration only. Requires equivalent ATEX approval on the sensor (2) For I.S. Output, Output Code B must be ordered (3) Product Certification Codes are added to the Sensor model number only (4) Only available with PTFE (all line sizes) or Polyurethane (4 in. or larger) Lining Materials and
316L SST Electrodes
(4)
PD - PD
DW - DW
(3)

HAZARDOUS LOCATION CERTIFICATIONS

Transmitter Approval Information

Table B-5. Transmitter Option Codes
Equivalent Hazardous Location Certifications for sensor and transmitter must match in integrally-mounted magnetic flowmeter systems. Remote-mounted systems do not require matched hazardous location certifica tion option codes.
Rosemount
8732E
FOUNDATION
Approval Codes HART
NA • N0 • N1 • N5
N7 • ND • NF
E1
E5
E7 • ED
fieldbus
Rosemount
8712D
Rosemount
8712H
B-5
Page 98
Rosemount 8732
Reference Manual
00809-0100-4663, Rev BA
January 2010
North American Certifications
Factory Mutual (FM)
NOTE
For intrinsically safe (IS) outputs on the 8732E output option code B must be selected. IS outputs for Class I, Division 1, Groups A, B, C, D. Temp Code – T4 at 60°C
NOTE
For the 8732E transmitters with a local operator interface (LOI) or display, the lower ambient temperature limit is -20 °C.
N0 Division 2 Approval (All transmitters)
Reference Rosemount Control Drawing 08732-1052 (8732E). Class I, Division 2, Groups A, B, C, D
Temp Codes – T4 (8712 at 40°C) T4 (8732 at 60°C: -50 °C Ta 60 °C)
Dust-ignition proof Class II/III, Division 1, Groups E, F, G Temp Codes – T4 (8712 at 40°C), T5 (8732 at 60°C)
Enclosure Type 4X
N5 Division 2 Approval (All Transmitters)
For sensors with IS electrodes only
Reference Rosemount Control Drawing 08732-1052 (8732E). Class I, Division 2, Groups A, B, C, D
Temp Codes – T4 (8712 at 40°C), T4 (8732 at 60°C: -50 °C Ta 60 °C)
Dust-ignition proof Class II/III, Division 1, Groups E, F, G Temp Codes – T4 (8712 at 40°C), T5 (8732 at 60°C)
Enclosure Type 4X
E5 Explosion-Proof Approval (8732E)
Reference Rosemount Control Drawing 08732-1052 Explosion-Proof for Class I, Division 1, Groups C, D
Temp Code – T6 at 60°C Dust-ignition proof Class II/III, Division 1, Groups E, F, G
Temp Code – T5 at 60°C Class I, Division 2, Groups A, B, C, D
Temp Codes – T4 (8732 at 60°C) Enclosure Type 4X
B-6
Canadian Standards Association (CSA)
N0 Division 2 Approval
Reference Rosemount Control Drawing 08732-1051 (8732E) Class I, Division 2, Groups A, B, C, D
Temp Codes – T4 (8732 at 60°C: -50 °C Ta 60 °C) Dust-ignition proof Class II/III, Division 1, Groups E, F, G
Temp Codes – T4 (8712 at 40°C), T5 (8732 at 60°C) Enclosure Type 4X
Page 99
Reference Manual
00809-0100-4663, Rev BA January 2010
European Certifications
E1 ATEX Flameproof
Hydrogen gas group 8732 - Certificate No.: KEMA 07ATEX0073 X II 2G Ex de [ia] IIC T6 (-50 °C Ta +60 °C) with LOI T6 (-20 °C Ta +60 °C)
= 250 V AC or 42 V DC
V
max
0575
ED ATEX Flameproof
8732 - Certificate No.: KEMA 07ATEX0073 X II 2G
Ex de [ia] IIB T6 (-50 °C Ta +60 °C) with LOI T6 (-20 °C Ta +60 °C)
= 250 V AC or 42 V DC
V
max
0575
ND ATEX Dust
8732 - Certificate No.: KEMA 06ATEX0006 II 1D
max T = 40 °K Amb. Temp. Limits: (-20 °C Ta  + 65 °C)
= 250 V AC or 42 V DC
V
max
IP 66
0575
(1)
Rosemount 8732
SPECIAL CONDITIONS FOR SAFE USE (KEMA 07ATEX0073 X):
If the Rosemount 8732 Flow Transmitter is used integrally with the Rosemount 8705 or 8711 Sensors, it shall be assured that the mechanical contact areas of the Sensor and Flow Transmitter comply with the requirements for flat joints according to standard EN/IEC 60079-1 clause 5.2.
The relation between ambient temperature, process temperature, and temperature class is to be taken from Table B-8 on page B-13
The electrical data is to be taken from Table B-7 on page B-12
If the Rosemount 8732 Flow Transmitter is used integrally with the Junction Box, it shall be assured that the mechanical contact areas of the Junction Box and Flow Transmitter comply with the requirements for flanged joints according to standard EN/IEC 60079-1 clause 5.2.
Per EN60079-1: 2004 the gap of the joint between transmitter and remote junction box/sensor is less than required per table 1 clause 5.2.2, and is only approved for use with an approved Rosemount transmitter and approved junction box/sensor.
INSTALLATION INSTRUCTIONS:
The cable and conduit entry devices and blanking elements shall be of a certified flameproof type, suitable for the conditions of use and correctly installed. With the use of conduit, a certified stopping box shall be provided immediately to the entrance of the enclosure.
SPECIAL CONDITIONS FOR SAFE USE (X) (03ATEX2159X):
The relation between ambient temperature, process temperature and temperature class is to be taken from Table B-8 on page B-13.
(1) Max surface temperature is 40 °C above the ambient temperature conditions. Tmax = 100
°C
B-7
Page 100
Rosemount 8732
Reference Manual
00809-0100-4663, Rev BA
January 2010
INSTALLATION INSTRUCTIONS:
The cable and conduit entry devices and the closing elements shall be of a certified increased safety type, suitable for the conditions of use and correctly installed.
At ambient temperatures above 50 °C, the flow meter shall be used with heat resistant cables with a temperature rating of at least 90 °C.
A Junction Box in type of explosion protection increased safety “e” may be attached to the base of the Rosemount 8732E Flow Transmitter , permitting remote mounting of the Rosemount 8705 and 8711 Sensors.
Ambient temperature range of the Junction Box: -50 °C to +60 °C.
The Junction Box is classified as: II 2 G Ex e IIB T6 and certified under KEMA 07ATEX0073 X.
N1 ATEX Type n
8712D - ATEX Certificate No: BASEEFA 05ATEX0170X
EEx nA nL IIC T4 (Ta = -50 °C to + 60 °C)
= 42 V DC
V
max
0575 8732 - ATEX Certificate No: BASEEFA 07ATEX0203X Ex nA nL IIC T4 (Ta = -50 °C to + 60 °C)
= 42 V DC
V
max
0575
Remote Junction Box
8732 - Certificate No.: KEMA 07ATEX0073 X II 2G
ATEX Ex e When installed per drawing 08732-1060 After de-energizing, wait 10 minutes before opening cover
0575
(1)
IIC for E1
IIB for ED
(1)
T6 (Ta = -50 °C to +60 °C)
International Certifications
E7 IECEx Flameproof
8732 - Certificate No.: KEM 07.0038X
Ex de [ia] IIC T6 (-50 °C Ta +60 °C)
= 250 V AC or 42 V DC
V
max
EF IECEx Flameproof
8732 - Certificate No.: KEM 07.0038X
Ex de [ia] IIB T6 (-50 °C Ta +60 °C)
= 250 V AC or 42 V DC
V
max
NF IECEx Dust
8732 - Certificate No.: KEM 07.0038X
Ex tD A20 IP66 T 100 °C T6 (-20 °C Ta +60 °C)
= 250 V AC or 42 V DC
V
max
B-8
Loading...