Rosemount Analytical designs, manufactures and tests its products to meet many national and
international standards. Because these instruments are sophisticated technical products, you
MUST properly install, use, and maintain them to ensure they continue to operate within their
normal specifications. The following instructions MUST be adhered to and integrated into your
safety program when installing, using, and maintaining Rosemount Analytical products. Failure to
follow the proper instructions may cause any one of the following situations to occur: Loss of life;
personal injury; property damage; damage to this instrument; and warranty invalidation.
•
Read all instructions prior to installing, operating, and servicing the product.
If you do not understand any of the instructions, contact your Rosemount Analytical representative
•
for clarification.
•
Follow all warnings, cautions, and instructions marked on and supplied with the product.
•
Inform and educate your personnel in the proper installation, operation, and maintenance of
the product.
•
Install your equipment as specified in the Installation Instructions of the appropriate
Instruction Manual and per applicable local and national codes. Connect all products to the
proper electrical and pressure sources.
To ensure proper performance, use qualified personnel to install, operate, update, program, and
•
maintain the product.
When replacement parts are required, ensure that qualified people use replacement parts specified by
•
Rosemount. Unauthorized parts and procedures can affect the product’s performance, place the safe
operation of your process at risk, and VOID YOUR WARRANTY. Look-alike substitutions may result
in fire, electrical hazards, or improper operation.
•
Ensure that all equipment doors are closed and protective covers are in place, except when
maintenance is being performed by qualified persons, to prevent electrical shock and personal
injury.
The information contained in this document is subject to change without notice.
Teflon and Viton are registered trademarks of E.I. duPont de Nemours and Co., Inc.
Paliney No.7 is a trademark of J.M. Ney Co., Hartford, CT
SNOOP is a registered trademark of NUPRO Co.
Emerson Process Management
Rosemount Analytical Inc.
Process Analytic Division
1201 N. Main St.
Orrville, OH 44667-0901
T (330) 682-9010
F (330) 684-4434
e-mail: gas.csc@EmersonProcess.com
Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii
Instruction Manual
748213-S
April 2002
Table 2-1.Remote Range Switching Truth Table................................................................. 2-12
Table 3-1.Calibration Range for Various Zero-Based Operating Ranges ............................. 3-4
Table 3-2.Oxygen Equivalent of Common Gases ................................................................. 3-6
Model 755R
LIST OF TABLES
DRAWINGS
617186Schematic Diagram, Case Board
620434Schematic Diagram, Isolated Current Output Board
646090Schematic Diagram, Remote Range Board
652826Schematic Diagram, Control Board
654014Pictorial Wiring Diagram, Model 755R
654015Installation Drawing, Model 755R
656081Instructions, Remote Range Selection
(LOCATED IN REAR OF MANUAL)
iv Contents Rosemount Analytical Inc. A Division of Emerson Process Management
Instruction Manual
Model 755R
PREFACE
The purpose of this manual is to provide information concerning the components,
functions, installation and maintenance of the 755R.
Some sections may describe equipment not used in your configuration. The user should
become thoroughly familiar with the operation of this module before operating it. Read
this instruction manual completely.
DEFINITIONS
The following definitions apply to DANGERS, WARNINGS, CAUTIONS and NOTES found throughout
this publication.
DANGER .
748213-S
April 2002
Highlights the presence of a hazard which will cause severe personal injury, death, or substantial
property damage if the warning is ignored.
WARNING .
Highlights an operation or maintenance procedure, practice, condition, statement, etc. If not
strictly observed, could result in injury, death, or long-term health hazards of personnel.
CAUTION.
Highlights an operation or maintenance procedure, practice, condition, statement, etc. If not
strictly observed, could result in damage to or destruction of equipment, or loss of effectiveness.
NOTE
Highlights an essential operating procedure,
condition or statement.
Rosemount Analytical Inc. A Division of Emerson Process Management Preface P-1
Instruction Manual
748213-S
April 2002
Model 755R
INTENDED USE STATEMENT
The Model 755R is intended for use as an industrial process measurement device only. It is not intended
for use in medical, diagnostic, or life support applications, and no independent agency certifications or
approvals are to be implied as covering such application.
SAFETY SUMMARY
If this equipment is used in a manner not specified in these instructions, protective systems may be
impaired.
AUTHORIZED PERSONNEL
To avoid explosion, loss of life, personal injury and damage to this equipment and on-site
property, all personnel authorized to install, operate and service the this equipment should be
thoroughly familiar with and strictly follow the instructions in this manual. SAVE THESE
INSTRUCTIONS.
DANGER.
ELECTRICAL SHOCK HAZARD
Do not operate without doors and covers secure. Servicing requires access to live parts which can
cause death or serious injury. Refer servicing to qualified personnel.
For safety and proper performance this instrument must be connected to a properly grounded
three-wire source of power.
Optional alarm switching relay contacts wired to separate power sources must be disconnected
before servicing.
WARNING
POSSIBLE EXPLOSION HAZARD
This analyzer is of a type capable of analysis of sample gases which may be flammable. If used for
analysis of such gases, internal leakage of sample could result in an explosion causing death, personal injury, or property damage. Do not use this analyzer on flammable samples. Use explosionproof version instruments for analysis of flammable samples.
P-2 Preface Rosemount Analytical Inc. A Division of Emerson Process Management
Instruction Manual
Model 755R
WARNING.
PARTS INTEGRITY
Tampering or unauthorized substitution of components may adversely affect safety of this product.
Use only factory documented components for repair.
CAUTION
PRESSURIZED GAS
This module requires periodic use of pressurized gas. See General Precautions for Handling and
Storing High Pressure Gas Cylinders, page P-4
CAUTION
TOPPLING HAZARD
This instrument’s internal pullout chassis is equipped with a safety stop latch located on the left
side of the chassis.
748213-S
April 2002
When extracting the chassis, verify that the safety latch is in its proper (counter-clockwise) orientation.
If access to the rear of the chassis is required, the safety stop may be overridden by lifting the
latch; however, further extraction must be done very carefully to insure the chassis does not fall
out of its enclosure.
If the instrument is located on top of a table or bench near the edge, and the chassis is extracted, it
must be supported to prevent toppling.
Rosemount Analytical Inc. A Division of Emerson Process Management Preface P-3
Instruction Manual
748213-S
April 2002
Model 755R
GENERAL PRECAUTIONS FOR HANDLING AND STORING HIGH
PRESSURE GAS CYLINDERS
Edited from selected paragraphs of the Compressed Gas Association's "Handbook of Compressed
Gases" published in 1981
Compressed Gas Association
1235 Jefferson Davis Highway
Arlington, Virginia 22202
Used by Permission
1. Never drop cylinders or permit them to strike each other violently.
2. Cylinders may be stored in the open, but in such cases, should be protected against extremes of weather
and, to prevent rusting, from the dampness of the ground. Cylinders should be stored in the shade when
located in areas where extreme temperatures are prevalent.
3. The valve protection cap should be left on each cylinder until it has been secured against a wall or bench, or
placed in a cylinder stand, and is ready to be used.
4. Avoid dragging, rolling, or sliding cylinders, even for a short distance; they should be moved by using a
suitable hand-truck.
5. Never tamper with safety devices in valves or cylinders.
6. Do not store full and empty cylinders together. Serious suckback can occur when an empty cylinder is
attached to a pressurized system.
7. No part of cylinder should be subjected to a temperature higher than 125°F (52°C). A flame should never be
permitted to come in contact with any part of a compressed gas cylinder.
8. Do not place cylinders where they may become part of an electric circuit. When electric arc welding,
precautions must be taken to prevent striking an arc against the cylinder.
P-4 Preface Rosemount Analytical Inc. A Division of Emerson Process Management
Instruction Manual
Model 755R
DOCUMENTATION
The following Model 755R instruction materials are available. Contact Customer Service Center or the
local representative to order.
748213 Instruction Manual (this document)
COMPLIANCES
This product satisfies all obligations of all relevant standards of the EMC framework in Australia and New
Zealand.
748213-S
April 2002
N
9
6
Rosemount Analytical Inc. A Division of Emerson Process Management Preface P-5
Instruction Manual
748213-S
April 2002
Model 755R
P-6 Preface Rosemount Analytical Inc. A Division of Emerson Process Management
Model 755R
y
Instruction Manual
748213-S
April 2002
SECTION 1
DESCRIPTION AND SPECIFICATIONS
1-1 DESCRIPTION
The Model 755R Oxygen Analyzer provides
continuous readout of the oxygen content of a
flowing gas sample. The determination is
based on measurement of the magnetic susceptibility of the sample gas. Oxygen is
strongly paramagnetic while most other common gases are weakly diamagnetic.
The instrument provides direct readout of 0 to
100% oxygen concentration on a front panel
digital display. In addition, a field-selectable
voltage output is provided as standard. An
isolated current output of 0 to 20 mA or 4 to
20 mA is obtainable through plug-in of an optional circuit board. Current and voltage outputs may be utilized simultaneously if desired.
An alarm option is also available by way of a
relay assembly that mounts at the rear of the
case with a cable that plugs into the Control
Board. Customer connections are available on
this assembly.
The basic electronic circuitry is incorporated
into two master boards designated the Control
Board assembly and the Power Supply Board
assembly. The Control Board has receptacles
that accept optional plug-in current output
board and alarm features.
1-2 RECORDER OUTPUT RANGES
Seven zero-based ranges are available with
the Model 755R: 0 to 1%, 0 to 2.5%, 0 to 5%, 0
to 10%, 0 to 25%, 0 to 50%, and 0 to 100%.
Each range is jumper selectable.
1-3 MOUNTING
The Model 755R is a rack-mounted instrument, standard for a 19-inch relay rack (Refer
to IEC Standard, Publication 297-1, 1986).
1-4 ISOLATED CURRENT OUTPUT OPTION
An isolated current output is obtainable by
using an optional current output board, either
during factory assembly or subsequently in
the field. The board provides ranges of 0 to 20
or 4 to 20 mA into a maximum resistive load
of 1000 ohms.
Digital Displa
O
%
2
SPANZERO
Rosemount Analytical
Zero ControlSpan Control
Model 755R
Figure 1-1. Model 755R Oxygen Analyzer – Front Panel
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-1
Instruction Manual
748213-S
April 2002
Model 755R
1-5 ALARM OPTION
The alarm option contains:
An alarm circuit incorporating two com-
•
parator amplifiers, one each for the
ALARM 1 and ALARM 2 functions. Each
amplifier has associated setpoint and
deadband adjustments. Setpoint is adjustable from 1% to 100% of fullscale.
Deadband is adjustable from 1% to 20%
of fullscale.
An alarm relay assembly, containing two
•
single-pole, double-throw relays (one
each for the ALARM 1 and ALARM 2
contacts). These relays may be used to
drive external, customer-supplied alarm
and/or control devices.
1-6 ELECTRICAL OPTIONS
The analyzer is supplied, as ordered, for operation on either 115 VAC, 50/60 Hz or 230
VAC, 50/60 Hz.
1-7 REMOTE RANGE CHANGE OPTION
This option allows the customer to remotely
control the recorder scaling. It disables the
internal recorder fullscale range select without
affecting the front panel display.
1-2 Description and Specifications Rosemount Analytical Inc. A Division of Emerson Process Management
Model 755R
Instruction Manual
748213-S
April 2002
1-8 SPECIFICATIONS
a. Performance
Operating Range (Standard)......... 0 to 5, 0 to 10, 0 to 25, 0 to 50, and 0 to 100% oxygen
Operating Range (Optional) .......... 0 to 1, 0 to 2.5, 0 to 5, 0 to 10, 0 to 25, 0 to 50, and 0 to 100%
Response Time ............................. 90% of fullscale, 20 seconds
Reproducibility............................... 0.01% oxygen or ±1% of fullscale, whichever is greater
Ambient Temperature Limits ......... 32°F (0°C) to 113°F (45°C)
Zero Drift........................................ ±1% fullscale per 24 hours, provided that ambient temperature
Span Drift....................................... ±1% fullscale per 24 hours, provided that ambient temperature
Mounting........................................ 19 inch rack (IEC 297-1, 1986)
Case Classification........................ General Purpose
Weight ........................................... 46 lbs. (21 kg)
Dimensions.................................... 19.0 x 8.7 x 19.2 inches (482.2 x 221 x 487 mm) W x H x D
1-4 Description and Specifications Rosemount Analytical Inc. A Division of Emerson Process Management
Model 755R
Instruction Manual
748213-S
April 2002
SECTION 2
INSTALLATION
2-1 FACILITY PREPARATION
Observe all precautions given in this section
when installing the instrument.
a. Installation Drawings
For outline and mounting dimensions, gas
connections, and other installation information, refer to Installation Drawing
654015 at the back of this manual.
b. Electrical Interconnection Diagram
Electrical interconnection is also shown in
drawing 654015. Refer also to Section 25, page 2-6.
c. Flow Diagram
The flow diagram of Figure 2-1 (page 2-3)
shows connection of a typical gas selector
manifold to the Model 755R.
d. Location and Mounting
Install the Model 755R only in a
non-hazardous, weather-protected area.
Permissible ambient temperature range is
32°F to 113°F (0°C to 45°C). Avoid
mounting where ambient temperature
may exceed the allowable maximum.
Magnetic susceptibilities and partial pressures of gases vary with temperature. In
the Model 755R, temperature-induced
readout error is avoided by control of
temperatures in the following areas:
tector, the sample is preheated by
passage through a coil maintained at
approximately the same temperature
as the detector (See Figure 4-3A,
page 4-7).
3. The detector is maintained at a controlled temperature of 150°F (66°C).
Also, avoid excessive vibration. To minimize vibration effects, the detector/magnet assembly is contained in a
shock-mounted compartment.
WARNING
POSSIBLE EXPLOSION HAZARD
This analyzer is of a type capable of analysis of sample gases which may be flammable. If used for analysis of such gases,
internal leakage of sample could result in
an explosion causing death, personal injury, or property damage. Do not use this
analyzer on flammable samples. Use explosion-proof version instruments for
analysis of flammable samples.
Use reasonable precautions to avoid excessive vibration. In making electrical
connections, do not allow any cable to
touch the shock-mounted detector assembly or the associated internal sample
inlet and outlet tubing. This precaution
ensures against possible transmission of
mechanical vibration through the cable to
the detector, which could cause noisy
readout.
1. Interior of the analyzer is maintained
at 140°F (60°C) by an electrically
controlled heater and associated fan.
2. Immediately downstream from the
inlet port, prior to entry into the de-
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-1
Instruction Manual
748213-S
April 2002
Model 755R
2-2 CALIBRATION GAS REQUIREMENTS
WARNING
HIGH PRESSURE GAS CYLINDERS
Calibration gas cylinders are under pressure. Mishandling of gas cylinders could
result in death, injury, or property damage.
Handle and store cylinders with extreme
caution and in accordance with the manufacturer’s instructions. Refer to GENERAL
PRECAUTIONS FOR HANDLING & STORING HIGH PRESSURE CYLINDERS, page
P-4.
Analyzer calibration consists of establishing a
zero calibration point and a span calibration
point.
Zero calibration is performed on the range
that will be used during sample analysis. In
some applications, however, it may be desirable to perform span calibration on a range of
higher sensitivity (i.e., more narrow span) and
then jumper to the desired operating range.
For example, if the operating range is to be 0
to 50% oxygen, span calibration may be performed on the 0 to 25% range to permit use of
air as the span standard gas.
Recommendations on calibration gases for
various operating ranges are tabulated in Table 3-1 (page 3-4) and are explained in Sections 2-2a (page 2-2) and 2-2b (page 2-2).
a. Zero Standard Gas
In the preferred calibration method, described in Section 3-4a (page 3-1), a suitable zero standard gas is used to
establish a calibration point at or near the
lower range limit. Composition of the zero
standard normally requires an oxygen-free zero gas, typically nitrogen.
b. Span Standard Gas
A suitable span standard gas is required
to establish a calibration point at or near
the upper range limit. If this range limit is
21% or 25% oxygen, the usual span
standard gas is air (20.93% oxygen).
2-3 SAMPLE
Basic requirements for sample are:
1. A 2-micron particulate filter, inserted into
the sample line immediately upstream
from the analyzer inlet.
2. Provision for pressurizing the sample gas
to provide flow through the analyzer.
3. Provision for selecting sample, zero standard, or span standard gas for admission
to the analyzer, and for measuring the
flow of the selected gas.
a. Temperature Requirements
Each standard gas should be supplied from a
cylinder equipped with dual-stage, metal diaphragm type pressure regulator, with output
pressure adjustable from 0 to 50 psig (0 to
345 kPa).
Instrument response to most non-oxygen
sample components is comparatively slight,
but is not in all cases negligible. During initial
installation of an instrument in a given application, effects of the background gas should
be calculated to determine if any correction is
required (See Section 3-4, page 3-1).
2-2 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
Sample temperature at the analyzer inlet
should be in the range of 50°F to 150°F
(10°C to 66°C).
Normally, however, a maximum entry
temperature of 110°F (43°C) is recommended so that the sample temperature
will rise during passage of the sample
through the analyzer. This precaution
prevents cooling of the sample and possible analyzer-damaging condensation.
With a thoroughly dry sample, entry temperature can be as high as 150°F (66°C)
without affecting readout accuracy.
Model 755R
Instruction Manual
748213-S
April 2002
Needle
Valves
Sample In
Zero
Standard
Gas
Span
Standard
Gas
Figure 2-1. Interconnect of Typical Gas Manifold to Model 755R
b. Pressure Requirements - General
Operating pressure limits are as follows:
maximum, 10 psig (68.9 kPa); minimum,
5 psig vacuum (34.5 kPa vacuum).
CAUTION
Model 755R
Oxygen Analyzer
Two Micron
Flowmeter
Filter
To Vent
c. Normal Operation at Positive Gauge
Pressures
Normally, the sample is supplied to the
analyzer inlet at a positive gauge pressure in the range of 0 to 10 psig (0 to 68.9
kPa).
RANGE LIMITATIONS
Operation outside the specified pressure
limits may damage the detector, and will
void the warranty.
HIGH PRESSURE GAS CYLINDERS
Pressure surges in excess of 10 psig dur-
CAUTION
ing admission of sample or standard
The basic rule for pressure of sample and
gases can damage the detector.
standard gases supplied to the inlet is to
calibrate the analyzer at the same pressure that will be used during subsequent
operation, and to maintain this pressure
during operation. The arrangement required to obtain appropriate pressure
control will depend on the application.
When inputting sample or calibration
gases, use the same pressure that will be
used during subsequent operation. Refer
to Section 2-3c (page 2-3), Normal Op-
Maximum permissible operating pressure
is 10 psig (68.9 kPa). To ensure against
over-pressurization, insert a pressure relief valve into the sample inlet line. In addition, a check valve should be placed in
the vent line if the analyzer is connected
to a manifold associated with a flare or
other outlet that is not at atmospheric
pressure. If the detector is overpressurized, damage will result.
eration at Positive Gauge Pressures, or
Section 2-3d (page 2-4) Operation at
Negative Gauge Pressures.
The analyzer exhaust port is commonly
vented directly to the atmosphere. Any
change in barometric pressure results in a
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-3
Instruction Manual
748213-S
April 2002
Model 755R
directly proportional change in the indicated percentage of oxygen.
Therefore, the 0.05% O2 error is
equal to 1% of fullscale.
Thus, if the exhaust is vented to the atmosphere, the pressure effect must be
taken into consideration. This may be accomplished in various ways, including
manual computation and computer correction of data.
d. Operation at Negative Gauge Pres-
sures
Operation at negative gauge pressures is
not normally recommended, but may be
used in certain special applications. A
suction pump is connected to the analyzer
exhaust port to draw sample into the inlet
and through the analyzer. Such operation
necessitates special precautions to ensure accurate readout. First is the basic
consideration of supplying the standard
gases to the analyzer at the same pressure that will be used for the sample during subsequent operation. In addition, any
leakage in the sample handling system
will result in decreased readout accuracy
as compared with operation at atmospheric pressure.
mum, 500 cc/min. A flow rate of less than
50 cc/min is too weak to sweep out the
detector and associated flow system efficiently. Incoming sample may mix with
earlier sample, causing an averaging or
damping effect. Too rapid a flow will
cause back pressure that will affect the
readout accuracy. The optimum flow rate
is between 200 and 300 cc/min.
Deviation from the set flow should be held
to within 10% or 20 cc/min, whichever is
smaller. If deviation is held to within these
parameters and operating pressure remains constant, zero and span drift will
remain within specification limits.
The analyzer should be installed near the
sample source to minimize transport time.
Otherwise, time lag may be appreciable.
For example, assume that sample is supplied to the analyzer via a 100-foot
(30.5 m) length of 1/4-inch (6.35 mm)
tubing. With a flow rate of 100 cc/min,
sample transport time is approximately 6
minutes.
Sample transport time may be reduced by
piping a greater flow than is required to
the analyzer, and then routing only the
appropriate portion of the total flow
through the analyzer. The unused portion
of the sample may be returned to the
stream or discarded.
f. Materials in Contact with Sample
Within the Model 755R, the following
materials are exposed to the sample: 316
stainless steel, glass, titanium, Paliney
No.7, epoxy resin, Viton-A, platinum,
nickel and MgF
2
coating on mirror.
The minimum permissible operating pres-
g. Corrosive Gases
sure is 5 psig vacuum (34.5 kPa vacuum).
Operation of the analyzer below this limit
may damage the detector, and will void
the warranty.
In applications where the sample stream
contains corrosive gases, a complete
drying of the sample is desirable, as most
of these gases are practically inert when
e. Flow Rate
totally dry. For corrosive applications
consult the factory.
Operating limits for sample flow rate are
as follows: minimum, 50 cc/min; maxi-
2-4 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
Model 755R
Instruction Manual
748213-S
April 2002
2-4 LEAK TEST
WARNING
TOXIC OR CORROSIVE HAZARD
The sample containment system must be
carefully leak checked upon installation
and before initial start-up, during routine
maintenance and any time the integrity of
the sample containment system is broken,
to ensure the system is in leak proof condition.
Internal leaks resulting from failure to observe these precautions could result in
personal injury or property damage.
A B I C D E
For proper operation and safety, system leakage must be corrected, particularly before introduction of toxic or corrosive samples and/or
application of electrical power.
To check system for leaks, liberally cover all
fittings, seals, and other possible sources of
leakage with suitable leak test liquid such as
SNOOP (P/N 837801). Check for leak indicative bubbling or foaming. Leaks that are inaccessible to SNOOP application could evade
detection by this method.
L1/HOT
L2/NEUT
GND
CUR VOLT
OUTPUT OUTPUT
+ - G + -
(Rear terminal cover removed for clarity)
A. Sample outlet. 1/4” O.D. tube fitting.
B. Sample Inlet. 1/4” O.D. tube fitting.
C. 5/8” diameter hole for optional Dual Alarm Cable. Cable supplied by customer, minimum 24 AWG.
D. 5/8” diameter hole fitted with liquid-tight gland for Recorder Output Cable. Cable supplied by customer,
conductor, minimum 24 AWG.
E. 13/16” diameter hole for Power Cable. Cable supplied by customer, 3 conductor, minimum 18 AWG.
F. TB1: Customer hook-up for Power.
G. TB2: Customer hook-up for Recorder Output.
H. Optional Dual Alarm connections.
I. Connections for Optional Remote Range Change.
H G H
Figure 2-2. Model 755R Rear Panel
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-5
Instruction Manual
I
t
(Verif
748213-S
April 2002
Model 755R
2-5 ELECTRICAL CONNECTIONS
WARNING
ELECTRICAL SHOCK HAZARD
For safety and proper performance, this
instrument must be connected to a properly grounded three-wire source of supply.
Cable connections for AC power, recorder
output, and alarm output are shown in Installation Drawing, 654015, and are explained in the following sections.
a. Line Power Connection
The analyzer is supplied, as ordered, for
operation on 115 VAC or 230 VAC, 50/60
Hz. Ensure that the power source conforms to the requirements of the individual
instrument, as noted on the name-rating
plate.
Electrical power is supplied to the analyzer via a customer-supplied
three-conductor cable, type SJT, minimum wire size 18 AWG. Route power cable through conduit and into appropriate
opening in the instrument case. Connect
power leads to HOT, NEUT, and GND
terminals on the I/O board. Connect analyzer to power source via an external fuse
or breaker, in accordance with local
codes. Do not draw power for associated
equipment from the analyzer power cable
(Refer to Figure 2-3 below).
If the analyzer is mounted in a protected
rack or cabinet or on a bench, an accessory kit (P/N 654008) is available which
provides a 10-foot North American power
cord set and a liquid-tight feed through
gland for the power cable hole. The kit
also contains four enclosure support feet
for bench top use.
b. Recorder Output Selection and Cable
Connections
If a recorder, controller, or other output
device is used, connect it to the analyzer
via a number 22 or number 24 AWG
two-conductor shielded cable. Route the
cable into the case through the liquid-tight
feed through gland in the Recorder Output opening (See Installation Drawing,
654015). Connect the shield only at the
recorder end or the analyzer end, not to
both at the same time because a ground
loop may occur.
NOTE:
Route recorder cable through a separate
cable gland (P/N 899329) or conduit not
with power cable or alarm output cable.
Cable connections and output selection
for potentiometric and current-actuated
devices are explained below.
755R
Analyzer
(Customer Supplied)
Position of Recorder Output
Selector Plug
10 mV 1K
100 mV 10K
1 V 100K
5 V 2K
Minimum Permissible
Resistance for R1 + R2
Potentiometric
Recorder
npu
Terminals
y polarity
is correct)Voltage Divider
(ohms)
Figure 2-3. Connections for Potentiometric Recorder with Non-Standard Span
2-6 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
Model 755R
Not
1000 oh
R
Controll
R
Instruction Manual
748213-S
April 2002
c. Potentiometric Output
1. Insert RECORDER OUTPUT Selector
Plug (See Figure 3-1) in position appropriate to the desired output: 10
mV, 100 mV, 1V or 5V.
2. Connect leads of shielded recorder
cable to “REC OUT +” and “-” terminals on the I/O board.
3. Connect the output cable to the appropriate terminals of the recorder or
other potentiometric device:
a. For device with span of 0 to 10
mV, 0 to 100 mV, 0 to 1V, or 0 to
5V, connect cable directly to input
terminals of the device, ensuring
correct polarity and range selection.
b. For a device with intermediate
span (i.e., between the specified
values), connect the cable to the
device via a suitable external
voltage divider (See Figure 2-3,
page 2-6).
d. Isolated Current Output (Optional)
1. Verify that the optional current output
board appropriate to desired output is
properly in place in its connector. See
Figure 3-1, page 3-3. If originally ordered with the analyzer, the board is
factory installed.
2. On I/O board, connect leads of
shielded recorder cable to “CURRENT OUT+” and “-” terminals.
3. Connect free end of output cable to
input terminals of recorder or other
current-actuated device, making sure
that polarity is correct. If two or more
current-actuated devices are to be
used, they must be connected in series (See Figure 2-4 below). Do not
exceed the maximum load resistance
of 1000 ohms.
Current and voltage outputs may be utilized simultaneously if desired.
+
mA
-
755R
Analyzer
e: Total series resistance of all devices is not to exceed
ms.
+
-
+
-
+
Indicator
-
ecorder
er
emote
Figure 2-4. Model 755R Connected to Drive Several Current-Actuated Output Devices
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-7
Instruction Manual
748213-S
April 2002
Model 755R
e. Output Connections and Initial Setup
for Dual Alarm Option
If so ordered, the analyzer is factory
equipped with alarm output. Alternatively,
the alarm feature is obtainable by subsequent installation of the 654019 Alarm Kit.
No. 1
NO
Low Alarm,
Fail-Safe
High Alarm,
Fail-Safe
Low Control
Limit,
Fail-Safe
RESET
RESET
No. 2
No. 1
RESET
RESET
No. 2
No. 1
RESET
RESET
No. 2
COM
NC
NO
COM
NC
NO
COM
NC
NO
COM
NC
NO
COM
NC
NO
COM
NC
Alarm Bell
or Lamp
Alarm Bell
or Lamp
Solenoid
Valve
115 VAC
115 VAC
115 VAC
The alarm output provides two sets of relay contacts for actuation of alarm and/or
process-control functions. Leads from the
customer-supplied external alarm system
connect to terminals on the 654019 Alarm
Assembly (See Figure 2-5 below and Interconnect Drawing 654014).