Emerson Process Management 755R User Manual

Instruction Manual
748213-S April 2002
Model 755R
Oxygen Analyzer
http://www.processanalytic.com

ESSENTIAL INSTRUCTIONS

READ THIS PAGE BEFORE PROCEEDING!
Rosemount Analytical designs, manufactures and tests its products to meet many national and international standards. Because these instruments are sophisticated technical products, you MUST properly install, use, and maintain them to ensure they continue to operate within their normal specifications. The following instructions MUST be adhered to and integrated into your safety program when installing, using, and maintaining Rosemount Analytical products. Failure to follow the proper instructions may cause any one of the following situations to occur: Loss of life; personal injury; property damage; damage to this instrument; and warranty invalidation.
Read all instructions prior to installing, operating, and servicing the product.
If you do not understand any of the instructions, contact your Rosemount Analytical representative
for clarification.
Follow all warnings, cautions, and instructions marked on and supplied with the product.
Inform and educate your personnel in the proper installation, operation, and maintenance of the product.
Install your equipment as specified in the Installation Instructions of the appropriate Instruction Manual and per applicable local and national codes. Connect all products to the
proper electrical and pressure sources.
To ensure proper performance, use qualified personnel to install, operate, update, program, and
maintain the product.
When replacement parts are required, ensure that qualified people use replacement parts specified by
Rosemount. Unauthorized parts and procedures can affect the product’s performance, place the safe operation of your process at risk, and VOID YOUR WARRANTY. Look-alike substitutions may result in fire, electrical hazards, or improper operation.
Ensure that all equipment doors are closed and protective covers are in place, except when maintenance is being performed by qualified persons, to prevent electrical shock and personal injury.
The information contained in this document is subject to change without notice.
Teflon and Viton are registered trademarks of E.I. duPont de Nemours and Co., Inc. Paliney No.7 is a trademark of J.M. Ney Co., Hartford, CT SNOOP is a registered trademark of NUPRO Co.
Emerson Process Management
Rosemount Analytical Inc. Process Analytic Division
1201 N. Main St. Orrville, OH 44667-0901 T (330) 682-9010 F (330) 684-4434 e-mail: gas.csc@EmersonProcess.com
http://www.processanalytic.com
Model 755R
PREFACE...........................................................................................................................................P-1
Definitions ...........................................................................................................................................P-1
Intended Use Statement.....................................................................................................................P-2
Safety Summary .................................................................................................................................P-2
General Precautions For Handling And Storing High Pressure Gas Cylinders .................................P-4
Documentation....................................................................................................................................P-5
Compliances .......................................................................................................................................P-5
1-0 DESCRIPTION AND SPECIFICATIONS..............................................................................1-1
1-1 Description.............................................................................................................................1-1
1-2 Recorder Output Ranges.......................................................................................................1-1
1-3 Mounting................................................................................................................................1-1
1-4 Isolated Current Output Option .............................................................................................1-1
1-5 Alarm Option..........................................................................................................................1-2
1-6 Electrical Options...................................................................................................................1-2
1-7 Remote Range Change Option .............................................................................................1-2
1-8 Specifications ........................................................................................................................1-3
a. Performance....................................................................................................................1-3
b. Sample ............................................................................................................................1-3
c. Electrical..........................................................................................................................1-4
d. Physical...........................................................................................................................1-4
Instruction Manual
748213-S
April 2002

TABLE OF CONTENTS

2-0 INSTALLATION ....................................................................................................................2-1
2-1 Facility Preparation................................................................................................................2-1
a. Installation Drawings.......................................................................................................2-1
b. Electrical Interconnection Diagram ...............................................................................2-1
c. Flow Diagram ..................................................................................................................2-1
d. Location and Mounting....................................................................................................2-1
2-2 Calibration Gas Requirements ..............................................................................................2-2
a. Zero Standard Gas..........................................................................................................2-2
b. Span Standard Gas ........................................................................................................2-2
2-3 Sample...................................................................................................................................2-2
a. Temperature Requirements ............................................................................................2-2
b. Pressure Requirements - General ..................................................................................2-3
c. Normal Operation at Positive Gauge Pressures.............................................................2-3
d. Operation at Negative Gauge Pressures........................................................................2-4
e. Flow Rate ........................................................................................................................2-4
f. Materials in Contact with Sample...................................................................................2-4
g. Corrosive Gases .............................................................................................................2-4
2-4 Leak Test ...............................................................................................................................2-5
2-5 Electrical Connections ...........................................................................................................2-6
a. Line Power Connection...................................................................................................2-6
b. Recorder Output Selection and Cable Connections .......................................................2-6
c. Potentiometric Output .....................................................................................................2-7
d. Isolated Current Output (Optional)..................................................................................2-7
e. Output Connections and Initial Setup for Dual Alarm Option .........................................2-8
2-6 Remote Range Change Option .............................................................................................2-12
Rosemount Analytical Inc. A Division of Emerson Process Management Contents i
Instruction Manual
748213-S April 2002
3-0 OPERATION .........................................................................................................................3-1
3-1 Overview................................................................................................................................3-1
3-2 Operating Range Selection ...................................................................................................3-1
3-3 Startup Procedure .................................................................................................................3-1
3-4 Calibration..............................................................................................................................3-1
a. Calibration with Zero and Span Standard Gases ...........................................................3-1
3-5 Compensation For Composition Of Background Gas ...........................................................3-2
a. Oxygen Equivalent Value of Gases ................................................................................3-4
b. Computing Adjusted Settings for Zero and Span Controls.............................................3-4
3-6 Selection Of Setpoints And Deadband On Alarm Option......................................................3-7
3-7 Current Output Board (Option) ..............................................................................................3-7
3-8 Routine Operation .................................................................................................................3-8
3-9 Effect of Barometric Pressure Changes on Instrument Readout ..........................................3-8
3-10 Calibration Frequency ...........................................................................................................3-8
4-0 THEORY................................................................................................................................4-1
4-1 Principles of Operation ..........................................................................................................4-1
4-2 Variables Influencing Paramagnetic Oxygen Measurements ...............................................4-2
a. Pressure Effects..............................................................................................................4-2
4-3 Electronic Circuitry.................................................................................................................4-4
a. Detector/Magnet Assembly.............................................................................................4-4
b. Control Board and Associated Circuitry..........................................................................4-4
c. Power Supply Board Assembly.......................................................................................4-5
d. Isolated Current Output Board (Optional) .......................................................................4-6
Model 755R
5-0 CIRCUIT ANALYSIS.............................................................................................................5-1
5-1 Circuit Operation....................................................................................................................5-1
5-2 ±15 VDC Power Supply.........................................................................................................5-1
5-3 Case Heater Control Circuit...................................................................................................5-1
5-4 Detector Heater Control Circuit .............................................................................................5-6
5-5 Detector Light Source Control Circuit....................................................................................5-7
5-6 Detector with First Stage Amplifier ........................................................................................5-8
5-7 Buffer Amplifiers U8 and U10 with Associated Anticipation Function ...................................5-10
5-8 Digital Output Circuit..............................................................................................................5-10
5-9 Analog Output Circuits for Recorder and Alarms ..................................................................5-11
a. First Stage Amplifier........................................................................................................5-11
b. Second Stage Amplifier ..................................................................................................5-11
6-0 MAINTENANCE AND SERVICE ..........................................................................................6-1
6-1 Initial Checkout With Standard Gases...................................................................................6-1
a. Control Board Checkout..................................................................................................6-1
6-2 Heating Circuits .....................................................................................................................6-2
a. Case Heater Control Circuit ............................................................................................6-2
6-3 Detector/Magnet Heating Circuit ...........................................................................................6-2
6-4 Detector Check......................................................................................................................6-4
a. Source Lamp...................................................................................................................6-5
b. Photocell .........................................................................................................................6-5
c. Suspension .....................................................................................................................6-5
ii Contents Rosemount Analytical Inc. A Division of Emerson Process Management
Model 755R
6-5 Replacement Of Detector/Magnet Components ...................................................................6-5
a. Source Lamp...................................................................................................................6-5
b. Photocell .........................................................................................................................6-5
c. Detector...........................................................................................................................6-7
6-6 Control Board Setup ..............................................................................................................6-7
a. Power Supply Test..........................................................................................................6-7
b. Detector zero...................................................................................................................6-7
c. U4 Zero ...........................................................................................................................6-8
d. U8 Zero ...........................................................................................................................6-8
e. U10 Zero .........................................................................................................................6-8
f. Fullscale ..........................................................................................................................6-8
g. Recorder Fullscale ..........................................................................................................6-8
7-0 REPLACEMENT PARTS ......................................................................................................7-1
7-1 Circuit Board Replacement Policy .........................................................................................7-1
7-2 Matrix – Model 755R Oxygen Analyzer.................................................................................7-2
7-3 Selected Replacement Parts.................................................................................................7-3
Instruction Manual
748213-S
April 2002
8-0 RETURN OF MATERIAL ......................................................................................................8-1
8-1 Return Of Material .................................................................................................................8-1
8-2 Customer Service ..................................................................................................................8-1
8-3 Training..................................................................................................................................8-1

LIST OF ILLUSTRATIONS

Figure 1-1. Model 755R Oxygen Analyzer – Front Panel ........................................................ 1-1
Figure 2-1. Interconnect of Typical Gas Manifold to Model 755R............................................ 2-3
Figure 2-2. Model 755R Rear Panel ........................................................................................ 2-5
Figure 2-3. Connections for Potentiometric Recorder with Non-Standard Span ..................... 2-6
Figure 2-4. Model 755R Connected to Drive Several Current-Actuated Output Devices........ 2-7
Figure 2-5. Relay Terminal Connections for Typical Fail-Safe Applications............................ 2-8
Figure 2-6. Typical Alarm Settings ......................................................................................... 2-10
Figure 2-7. Alarm Relay Assembly Schematic Diagram ........................................................ 2-11
Figure 3-1. Control Board - Adjustment Locations................................................................... 3-3
Figure 3-2. Dial Settings for Alarm Setpoint Adjustments........................................................ 3-7
Figure 4-1. Functional Diagram of Paramagnetic Oxygen Measurement System................... 4-3
Figure 4-2. Spherical Body in Non-Uniform Magnetic Field..................................................... 4-4
Figure 4-3. Detector/Magnet Assembly.................................................................................... 4-7
Figure 5-1. Two-Comparator OR Circuit .................................................................................. 5-2
Figure 5-2. Case Heater Control Circuit................................................................................... 5-3
Figure 5-3. Ramp Generator Circuit......................................................................................... 5-3
Figure 5-4. Detector Heater Control Circuit.............................................................................. 5-6
Figure 5-5. Detector Light Source Control Circuit .................................................................... 5-7
Figure 5-6. Detector with First Stage Amplifier ........................................................................ 5-9
Figure 5-7. Buffer, Anticipation, and Digital Output Circuits................................................... 5-10
Figure 5-8. Simplified Analog Recorder Output Circuit .......................................................... 5-12
Figure 6-1. Detector/Magnet Assembly.................................................................................... 6-3
Figure 6-2. Pin/Lead Removal ................................................................................................. 6-4
Figure 6-3. Detector Optical Bench.......................................................................................... 6-4
Figure 6-4. Lamp Replacement................................................................................................ 6-6
Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii
Instruction Manual
748213-S April 2002
Table 2-1. Remote Range Switching Truth Table................................................................. 2-12
Table 3-1. Calibration Range for Various Zero-Based Operating Ranges ............................. 3-4
Table 3-2. Oxygen Equivalent of Common Gases ................................................................. 3-6
Model 755R

LIST OF TABLES

DRAWINGS
617186 Schematic Diagram, Case Board 620434 Schematic Diagram, Isolated Current Output Board 646090 Schematic Diagram, Remote Range Board 652826 Schematic Diagram, Control Board 654014 Pictorial Wiring Diagram, Model 755R 654015 Installation Drawing, Model 755R 656081 Instructions, Remote Range Selection
(LOCATED IN REAR OF MANUAL)
iv Contents Rosemount Analytical Inc. A Division of Emerson Process Management
Instruction Manual
Model 755R

PREFACE

The purpose of this manual is to provide information concerning the components, functions, installation and maintenance of the 755R.
Some sections may describe equipment not used in your configuration. The user should become thoroughly familiar with the operation of this module before operating it. Read this instruction manual completely.

DEFINITIONS

The following definitions apply to DANGERS, WARNINGS, CAUTIONS and NOTES found throughout this publication.
DANGER .
748213-S
April 2002
Highlights the presence of a hazard which will cause severe personal injury, death, or substantial property damage if the warning is ignored.
WARNING .
Highlights an operation or maintenance procedure, practice, condition, statement, etc. If not
strictly observed, could result in injury, death, or long-term health hazards of personnel.
CAUTION.
Highlights an operation or maintenance procedure, practice, condition, statement, etc. If not strictly observed, could result in damage to or destruction of equipment, or loss of effectiveness.
NOTE
Highlights an essential operating procedure, condition or statement.
Rosemount Analytical Inc. A Division of Emerson Process Management Preface P-1
Instruction Manual
748213-S April 2002
Model 755R

INTENDED USE STATEMENT

The Model 755R is intended for use as an industrial process measurement device only. It is not intended for use in medical, diagnostic, or life support applications, and no independent agency certifications or approvals are to be implied as covering such application.

SAFETY SUMMARY

If this equipment is used in a manner not specified in these instructions, protective systems may be impaired.

AUTHORIZED PERSONNEL

To avoid explosion, loss of life, personal injury and damage to this equipment and on-site property, all personnel authorized to install, operate and service the this equipment should be thoroughly familiar with and strictly follow the instructions in this manual. SAVE THESE INSTRUCTIONS.
DANGER.
ELECTRICAL SHOCK HAZARD
Do not operate without doors and covers secure. Servicing requires access to live parts which can cause death or serious injury. Refer servicing to qualified personnel.
For safety and proper performance this instrument must be connected to a properly grounded three-wire source of power.
Optional alarm switching relay contacts wired to separate power sources must be disconnected before servicing.
WARNING
POSSIBLE EXPLOSION HAZARD
This analyzer is of a type capable of analysis of sample gases which may be flammable. If used for analysis of such gases, internal leakage of sample could result in an explosion causing death, per­sonal injury, or property damage. Do not use this analyzer on flammable samples. Use explosion­proof version instruments for analysis of flammable samples.
P-2 Preface Rosemount Analytical Inc. A Division of Emerson Process Management
Instruction Manual
Model 755R
WARNING.
PARTS INTEGRITY
Tampering or unauthorized substitution of components may adversely affect safety of this product. Use only factory documented components for repair.
CAUTION
PRESSURIZED GAS
This module requires periodic use of pressurized gas. See General Precautions for Handling and Storing High Pressure Gas Cylinders, page P-4
CAUTION
TOPPLING HAZARD
This instrument’s internal pullout chassis is equipped with a safety stop latch located on the left side of the chassis.
748213-S
April 2002
When extracting the chassis, verify that the safety latch is in its proper (counter-clockwise) orienta­tion.
If access to the rear of the chassis is required, the safety stop may be overridden by lifting the latch; however, further extraction must be done very carefully to insure the chassis does not fall out of its enclosure.
If the instrument is located on top of a table or bench near the edge, and the chassis is extracted, it must be supported to prevent toppling.
Rosemount Analytical Inc. A Division of Emerson Process Management Preface P-3
Instruction Manual
748213-S April 2002
Model 755R
GENERAL PRECAUTIONS FOR HANDLING AND STORING HIGH
PRESSURE GAS CYLINDERS
Edited from selected paragraphs of the Compressed Gas Association's "Handbook of Compressed Gases" published in 1981
Compressed Gas Association 1235 Jefferson Davis Highway Arlington, Virginia 22202
Used by Permission
1. Never drop cylinders or permit them to strike each other violently.
2. Cylinders may be stored in the open, but in such cases, should be protected against extremes of weather and, to prevent rusting, from the dampness of the ground. Cylinders should be stored in the shade when located in areas where extreme temperatures are prevalent.
3. The valve protection cap should be left on each cylinder until it has been secured against a wall or bench, or placed in a cylinder stand, and is ready to be used.
4. Avoid dragging, rolling, or sliding cylinders, even for a short distance; they should be moved by using a suitable hand-truck.
5. Never tamper with safety devices in valves or cylinders.
6. Do not store full and empty cylinders together. Serious suckback can occur when an empty cylinder is attached to a pressurized system.
7. No part of cylinder should be subjected to a temperature higher than 125°F (52°C). A flame should never be permitted to come in contact with any part of a compressed gas cylinder.
8. Do not place cylinders where they may become part of an electric circuit. When electric arc welding, precautions must be taken to prevent striking an arc against the cylinder.
P-4 Preface Rosemount Analytical Inc. A Division of Emerson Process Management
Instruction Manual
Model 755R

DOCUMENTATION

The following Model 755R instruction materials are available. Contact Customer Service Center or the local representative to order.
748213 Instruction Manual (this document)

COMPLIANCES

This product satisfies all obligations of all relevant standards of the EMC framework in Australia and New Zealand.
748213-S
April 2002
N
9
6
Rosemount Analytical Inc. A Division of Emerson Process Management Preface P-5
Instruction Manual
748213-S April 2002
Model 755R
P-6 Preface Rosemount Analytical Inc. A Division of Emerson Process Management
Model 755R
y
Instruction Manual
748213-S
April 2002
SECTION 1

DESCRIPTION AND SPECIFICATIONS

1-1 DESCRIPTION

The Model 755R Oxygen Analyzer provides continuous readout of the oxygen content of a flowing gas sample. The determination is based on measurement of the magnetic sus­ceptibility of the sample gas. Oxygen is strongly paramagnetic while most other com­mon gases are weakly diamagnetic.
The instrument provides direct readout of 0 to 100% oxygen concentration on a front panel digital display. In addition, a field-selectable voltage output is provided as standard. An isolated current output of 0 to 20 mA or 4 to 20 mA is obtainable through plug-in of an op­tional circuit board. Current and voltage out­puts may be utilized simultaneously if desired. An alarm option is also available by way of a relay assembly that mounts at the rear of the case with a cable that plugs into the Control Board. Customer connections are available on this assembly.
The basic electronic circuitry is incorporated into two master boards designated the Control
Board assembly and the Power Supply Board assembly. The Control Board has receptacles that accept optional plug-in current output board and alarm features.

1-2 RECORDER OUTPUT RANGES

Seven zero-based ranges are available with the Model 755R: 0 to 1%, 0 to 2.5%, 0 to 5%, 0 to 10%, 0 to 25%, 0 to 50%, and 0 to 100%. Each range is jumper selectable.

1-3 MOUNTING

The Model 755R is a rack-mounted instru­ment, standard for a 19-inch relay rack (Refer to IEC Standard, Publication 297-1, 1986).

1-4 ISOLATED CURRENT OUTPUT OPTION

An isolated current output is obtainable by using an optional current output board, either during factory assembly or subsequently in the field. The board provides ranges of 0 to 20 or 4 to 20 mA into a maximum resistive load of 1000 ohms.
Digital Displa
O
%
2
SPANZERO
Rosemount Analytical
Zero Control Span Control
Model 755R
Figure 1-1. Model 755R Oxygen Analyzer – Front Panel
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-1
Instruction Manual
748213-S April 2002
Model 755R

1-5 ALARM OPTION

The alarm option contains:
An alarm circuit incorporating two com-
parator amplifiers, one each for the ALARM 1 and ALARM 2 functions. Each amplifier has associated setpoint and deadband adjustments. Setpoint is ad­justable from 1% to 100% of fullscale. Deadband is adjustable from 1% to 20% of fullscale.
An alarm relay assembly, containing two
single-pole, double-throw relays (one each for the ALARM 1 and ALARM 2
contacts). These relays may be used to drive external, customer-supplied alarm and/or control devices.

1-6 ELECTRICAL OPTIONS

The analyzer is supplied, as ordered, for op­eration on either 115 VAC, 50/60 Hz or 230 VAC, 50/60 Hz.

1-7 REMOTE RANGE CHANGE OPTION

This option allows the customer to remotely control the recorder scaling. It disables the internal recorder fullscale range select without affecting the front panel display.
1-2 Description and Specifications Rosemount Analytical Inc. A Division of Emerson Process Management
Model 755R
Instruction Manual
748213-S
April 2002

1-8 SPECIFICATIONS

a. Performance

Operating Range (Standard)......... 0 to 5, 0 to 10, 0 to 25, 0 to 50, and 0 to 100% oxygen
Operating Range (Optional) .......... 0 to 1, 0 to 2.5, 0 to 5, 0 to 10, 0 to 25, 0 to 50, and 0 to 100%
Response Time ............................. 90% of fullscale, 20 seconds
Reproducibility............................... 0.01% oxygen or ±1% of fullscale, whichever is greater
Ambient Temperature Limits ......... 32°F (0°C) to 113°F (45°C)
Zero Drift........................................ ±1% fullscale per 24 hours, provided that ambient temperature
Span Drift....................................... ±1% fullscale per 24 hours, provided that ambient temperature

b. Sample

Dryness ......................................... Sample dewpoint below 110°F (43°C), sample free of entrained
Temperature Limits ....................... 50°F (10°C) to 150°F (65°C)
Operating Pressure ....................... Maximum: 10 psig (68.9 kPa)
....................................................... Minimum: 5 psig vacuum (34.5 kPa vacuum)
Flow Rate ...................................... 50 cc/min. to 500 cc/min.
....................................................... Recommended 250 ±20 cc/min.
Materials in Contact with Sample.. Glass, 316 stainless steel, titanium, Paliney No. 7, epoxy resin,
1
oxygen
does not change by more than 20°F (11.1°C)
±2.5% of fullscale per 24 hours with ambient temperature change
over entire range
does not change by more than 20°F (11.1°C)
±2.5% of fullscale per 24 hours with ambient temperature change
over entire range
liquids.
Viton-A, platinum, nickel, and MgF2
1
Performance specifications are measured at recorder output and are based on constant sample pressure and deviation from
set flow held to within 10% or 20 cc/min., whichever is smaller.
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-3
Instruction Manual
748213-S April 2002

c. Electrical

Supply Voltage and Frequency
(selectable when ordered)............ Standard: 115 VAC ±10 VAC, 50/60 Hz
Power Consumption ...................... Maximum: 300 watts
Outputs .......................................... Standard: Field selectable voltage output of 0 to 10mV, 0 to
Alarm Option.................................. High-Low Alarm
Contact Ratings ..................... 5 amperes, 240V AC, resistive 3 amperes, 120 VAC inductive
Setpoint ......................................... Adjustable from 1% to 100% of fullscale
Deadband ...................................... Adjustable from 1% to 20% of fullscale (Factory set at 10% of
Model 755R
Optional: 230 VAC ±10 VAC, 50/60 Hz
100mV, 0 to 1V, or 0 to 5VDC
Optional: Isolated current output of 0 to 20mA or 4 to 20mA (with Current Output Board)
1 amperes, 24V DC, resistive 5 amperes, 30 VDC resistive 5 amperes, 120V AC, resistive 3 amperes, 30 VDC inductive
fullscale)

d. Physical

Mounting........................................ 19 inch rack (IEC 297-1, 1986)
Case Classification........................ General Purpose
Weight ........................................... 46 lbs. (21 kg)
Dimensions.................................... 19.0 x 8.7 x 19.2 inches (482.2 x 221 x 487 mm) W x H x D
1-4 Description and Specifications Rosemount Analytical Inc. A Division of Emerson Process Management
Model 755R
Instruction Manual
748213-S
April 2002
SECTION 2

INSTALLATION

2-1 FACILITY PREPARATION

Observe all precautions given in this section when installing the instrument.

a. Installation Drawings

For outline and mounting dimensions, gas connections, and other installation infor­mation, refer to Installation Drawing 654015 at the back of this manual.

b. Electrical Interconnection Diagram

Electrical interconnection is also shown in drawing 654015. Refer also to Section 2­5, page 2-6.

c. Flow Diagram

The flow diagram of Figure 2-1 (page 2-3) shows connection of a typical gas selector manifold to the Model 755R.

d. Location and Mounting

Install the Model 755R only in a non-hazardous, weather-protected area. Permissible ambient temperature range is 32°F to 113°F (0°C to 45°C). Avoid mounting where ambient temperature may exceed the allowable maximum.
Magnetic susceptibilities and partial pres­sures of gases vary with temperature. In the Model 755R, temperature-induced readout error is avoided by control of temperatures in the following areas:
tector, the sample is preheated by passage through a coil maintained at approximately the same temperature as the detector (See Figure 4-3A, page 4-7).
3. The detector is maintained at a con­trolled temperature of 150°F (66°C).
Also, avoid excessive vibration. To mini­mize vibration effects, the detec­tor/magnet assembly is contained in a shock-mounted compartment.
WARNING
POSSIBLE EXPLOSION HAZARD
This analyzer is of a type capable of analy­sis of sample gases which may be flam­mable. If used for analysis of such gases, internal leakage of sample could result in an explosion causing death, personal in­jury, or property damage. Do not use this analyzer on flammable samples. Use ex­plosion-proof version instruments for analysis of flammable samples.
Use reasonable precautions to avoid ex­cessive vibration. In making electrical connections, do not allow any cable to touch the shock-mounted detector as­sembly or the associated internal sample inlet and outlet tubing. This precaution ensures against possible transmission of mechanical vibration through the cable to the detector, which could cause noisy readout.
1. Interior of the analyzer is maintained at 140°F (60°C) by an electrically controlled heater and associated fan.
2. Immediately downstream from the inlet port, prior to entry into the de-
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-1
Instruction Manual
748213-S April 2002
Model 755R

2-2 CALIBRATION GAS REQUIREMENTS

WARNING
HIGH PRESSURE GAS CYLINDERS
Calibration gas cylinders are under pres­sure. Mishandling of gas cylinders could result in death, injury, or property damage. Handle and store cylinders with extreme caution and in accordance with the manu­facturer’s instructions. Refer to GENERAL PRECAUTIONS FOR HANDLING & STOR­ING HIGH PRESSURE CYLINDERS, page P-4.
Analyzer calibration consists of establishing a zero calibration point and a span calibration point.
Zero calibration is performed on the range that will be used during sample analysis. In some applications, however, it may be desir­able to perform span calibration on a range of higher sensitivity (i.e., more narrow span) and then jumper to the desired operating range. For example, if the operating range is to be 0 to 50% oxygen, span calibration may be per­formed on the 0 to 25% range to permit use of air as the span standard gas.
Recommendations on calibration gases for various operating ranges are tabulated in Ta­ble 3-1 (page 3-4) and are explained in Sec­tions 2-2a (page 2-2) and 2-2b (page 2-2).

a. Zero Standard Gas

In the preferred calibration method, de­scribed in Section 3-4a (page 3-1), a suit­able zero standard gas is used to establish a calibration point at or near the lower range limit. Composition of the zero standard normally requires an oxy­gen-free zero gas, typically nitrogen.

b. Span Standard Gas

A suitable span standard gas is required to establish a calibration point at or near the upper range limit. If this range limit is 21% or 25% oxygen, the usual span standard gas is air (20.93% oxygen).

2-3 SAMPLE

Basic requirements for sample are:
1. A 2-micron particulate filter, inserted into the sample line immediately upstream from the analyzer inlet.
2. Provision for pressurizing the sample gas to provide flow through the analyzer.
3. Provision for selecting sample, zero stan­dard, or span standard gas for admission to the analyzer, and for measuring the flow of the selected gas.

a. Temperature Requirements

Each standard gas should be supplied from a cylinder equipped with dual-stage, metal dia­phragm type pressure regulator, with output pressure adjustable from 0 to 50 psig (0 to 345 kPa).
Instrument response to most non-oxygen sample components is comparatively slight, but is not in all cases negligible. During initial installation of an instrument in a given appli­cation, effects of the background gas should be calculated to determine if any correction is required (See Section 3-4, page 3-1).
2-2 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
Sample temperature at the analyzer inlet should be in the range of 50°F to 150°F (10°C to 66°C).
Normally, however, a maximum entry temperature of 110°F (43°C) is recom­mended so that the sample temperature will rise during passage of the sample through the analyzer. This precaution prevents cooling of the sample and possi­ble analyzer-damaging condensation. With a thoroughly dry sample, entry tem­perature can be as high as 150°F (66°C) without affecting readout accuracy.
Model 755R
Instruction Manual
748213-S
April 2002
Needle
Valves
Sample In
Zero Standard Gas
Span Standard Gas
Figure 2-1. Interconnect of Typical Gas Manifold to Model 755R

b. Pressure Requirements - General

Operating pressure limits are as follows: maximum, 10 psig (68.9 kPa); minimum, 5 psig vacuum (34.5 kPa vacuum).
CAUTION
Model 755R
Oxygen Analyzer
Two Micron
Flowmeter
Filter
To Vent
c. Normal Operation at Positive Gauge
Pressures
Normally, the sample is supplied to the analyzer inlet at a positive gauge pres­sure in the range of 0 to 10 psig (0 to 68.9 kPa).
RANGE LIMITATIONS
Operation outside the specified pressure limits may damage the detector, and will void the warranty.
HIGH PRESSURE GAS CYLINDERS
Pressure surges in excess of 10 psig dur-
CAUTION
ing admission of sample or standard
The basic rule for pressure of sample and
gases can damage the detector.
standard gases supplied to the inlet is to calibrate the analyzer at the same pres­sure that will be used during subsequent operation, and to maintain this pressure during operation. The arrangement re­quired to obtain appropriate pressure control will depend on the application. When inputting sample or calibration gases, use the same pressure that will be used during subsequent operation. Refer to Section 2-3c (page 2-3), Normal Op-
Maximum permissible operating pressure is 10 psig (68.9 kPa). To ensure against over-pressurization, insert a pressure re­lief valve into the sample inlet line. In ad­dition, a check valve should be placed in the vent line if the analyzer is connected to a manifold associated with a flare or other outlet that is not at atmospheric pressure. If the detector is over­pressurized, damage will result.
eration at Positive Gauge Pressures, or Section 2-3d (page 2-4) Operation at Negative Gauge Pressures.
The analyzer exhaust port is commonly vented directly to the atmosphere. Any change in barometric pressure results in a
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-3
Instruction Manual
748213-S April 2002
Model 755R
directly proportional change in the indi­cated percentage of oxygen.
Example:
Range, 0% to 5% O
2.
Barometric pressure change after calibration, 1%.
Instrument reading, 5% O2. Readout error = 0.01 x 5% O2 =
0.05% O Fullscale span is 5% O
2
.
2.
Therefore, the 0.05% O2 error is equal to 1% of fullscale.
Thus, if the exhaust is vented to the at­mosphere, the pressure effect must be taken into consideration. This may be ac­complished in various ways, including manual computation and computer cor­rection of data.
d. Operation at Negative Gauge Pres-
sures
Operation at negative gauge pressures is not normally recommended, but may be used in certain special applications. A suction pump is connected to the analyzer exhaust port to draw sample into the inlet and through the analyzer. Such operation necessitates special precautions to en­sure accurate readout. First is the basic consideration of supplying the standard gases to the analyzer at the same pres­sure that will be used for the sample dur­ing subsequent operation. In addition, any leakage in the sample handling system will result in decreased readout accuracy as compared with operation at atmos­pheric pressure.
mum, 500 cc/min. A flow rate of less than 50 cc/min is too weak to sweep out the detector and associated flow system effi­ciently. Incoming sample may mix with earlier sample, causing an averaging or damping effect. Too rapid a flow will cause back pressure that will affect the readout accuracy. The optimum flow rate is between 200 and 300 cc/min.
Deviation from the set flow should be held to within 10% or 20 cc/min, whichever is smaller. If deviation is held to within these parameters and operating pressure re­mains constant, zero and span drift will remain within specification limits.
The analyzer should be installed near the sample source to minimize transport time. Otherwise, time lag may be appreciable. For example, assume that sample is sup­plied to the analyzer via a 100-foot (30.5 m) length of 1/4-inch (6.35 mm) tubing. With a flow rate of 100 cc/min, sample transport time is approximately 6 minutes.
Sample transport time may be reduced by piping a greater flow than is required to the analyzer, and then routing only the appropriate portion of the total flow through the analyzer. The unused portion of the sample may be returned to the stream or discarded.

f. Materials in Contact with Sample

Within the Model 755R, the following materials are exposed to the sample: 316 stainless steel, glass, titanium, Paliney No.7, epoxy resin, Viton-A, platinum, nickel and MgF
2
coating on mirror.
The minimum permissible operating pres-

g. Corrosive Gases

sure is 5 psig vacuum (34.5 kPa vacuum). Operation of the analyzer below this limit may damage the detector, and will void the warranty.
In applications where the sample stream contains corrosive gases, a complete drying of the sample is desirable, as most of these gases are practically inert when

e. Flow Rate

totally dry. For corrosive applications consult the factory.
Operating limits for sample flow rate are as follows: minimum, 50 cc/min; maxi-
2-4 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
Model 755R
Instruction Manual
748213-S
April 2002

2-4 LEAK TEST

WARNING
TOXIC OR CORROSIVE HAZARD
The sample containment system must be carefully leak checked upon installation and before initial start-up, during routine maintenance and any time the integrity of the sample containment system is broken, to ensure the system is in leak proof con­dition.
Internal leaks resulting from failure to ob­serve these precautions could result in personal injury or property damage.
A B I C D E
For proper operation and safety, system leak­age must be corrected, particularly before in­troduction of toxic or corrosive samples and/or application of electrical power.
To check system for leaks, liberally cover all fittings, seals, and other possible sources of leakage with suitable leak test liquid such as SNOOP (P/N 837801). Check for leak indica­tive bubbling or foaming. Leaks that are inac­cessible to SNOOP application could evade detection by this method.
L1/HOT
L2/NEUT
GND
CUR VOLT OUTPUT OUTPUT
+ - G + -
(Rear terminal cover removed for clarity)
A. Sample outlet. 1/4” O.D. tube fitting. B. Sample Inlet. 1/4” O.D. tube fitting. C. 5/8” diameter hole for optional Dual Alarm Cable. Cable supplied by customer, minimum 24 AWG. D. 5/8” diameter hole fitted with liquid-tight gland for Recorder Output Cable. Cable supplied by customer, conductor, minimum 24 AWG. E. 13/16” diameter hole for Power Cable. Cable supplied by customer, 3 conductor, minimum 18 AWG. F. TB1: Customer hook-up for Power. G. TB2: Customer hook-up for Recorder Output. H. Optional Dual Alarm connections. I. Connections for Optional Remote Range Change.
H G H
Figure 2-2. Model 755R Rear Panel
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-5
Instruction Manual
I
t
(Verif
748213-S April 2002
Model 755R

2-5 ELECTRICAL CONNECTIONS

WARNING
ELECTRICAL SHOCK HAZARD
For safety and proper performance, this instrument must be connected to a prop­erly grounded three-wire source of supply.
Cable connections for AC power, recorder output, and alarm output are shown in In­stallation Drawing, 654015, and are ex­plained in the following sections.

a. Line Power Connection

The analyzer is supplied, as ordered, for operation on 115 VAC or 230 VAC, 50/60 Hz. Ensure that the power source con­forms to the requirements of the individual instrument, as noted on the name-rating plate.
Electrical power is supplied to the ana­lyzer via a customer-supplied three-conductor cable, type SJT, mini­mum wire size 18 AWG. Route power ca­ble through conduit and into appropriate opening in the instrument case. Connect power leads to HOT, NEUT, and GND terminals on the I/O board. Connect ana­lyzer to power source via an external fuse or breaker, in accordance with local codes. Do not draw power for associated
equipment from the analyzer power cable (Refer to Figure 2-3 below).
If the analyzer is mounted in a protected rack or cabinet or on a bench, an acces­sory kit (P/N 654008) is available which provides a 10-foot North American power cord set and a liquid-tight feed through gland for the power cable hole. The kit also contains four enclosure support feet for bench top use.
b. Recorder Output Selection and Cable
Connections
If a recorder, controller, or other output device is used, connect it to the analyzer via a number 22 or number 24 AWG two-conductor shielded cable. Route the cable into the case through the liquid-tight feed through gland in the Recorder Out­put opening (See Installation Drawing,
654015). Connect the shield only at the recorder end or the analyzer end, not to both at the same time because a ground loop may occur.
NOTE:
Route recorder cable through a separate cable gland (P/N 899329) or conduit not with power cable or alarm output cable. Cable connections and output selection for potentiometric and current-actuated devices are explained below.
755R
Analyzer
(Customer Supplied)
Position of Recorder Output
Selector Plug
10 mV 1K 100 mV 10K 1 V 100K 5 V 2K
Minimum Permissible
Resistance for R1 + R2
Potentiometric
Recorder
npu
Terminals
y polarity
is correct)Voltage Divider
(ohms)
Figure 2-3. Connections for Potentiometric Recorder with Non-Standard Span
2-6 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
Model 755R
Not
1000 oh
R
Controll
R
Instruction Manual
748213-S
April 2002

c. Potentiometric Output

1. Insert RECORDER OUTPUT Selector Plug (See Figure 3-1) in position ap­propriate to the desired output: 10 mV, 100 mV, 1V or 5V.
2. Connect leads of shielded recorder cable to “REC OUT +” and “-” termi­nals on the I/O board.
3. Connect the output cable to the appro­priate terminals of the recorder or other potentiometric device:
a. For device with span of 0 to 10
mV, 0 to 100 mV, 0 to 1V, or 0 to 5V, connect cable directly to input terminals of the device, ensuring correct polarity and range selec­tion.
b. For a device with intermediate
span (i.e., between the specified values), connect the cable to the device via a suitable external voltage divider (See Figure 2-3, page 2-6).

d. Isolated Current Output (Optional)

1. Verify that the optional current output board appropriate to desired output is properly in place in its connector. See Figure 3-1, page 3-3. If originally or­dered with the analyzer, the board is factory installed.
2. On I/O board, connect leads of shielded recorder cable to “CUR­RENT OUT+” and “-” terminals.
3. Connect free end of output cable to input terminals of recorder or other current-actuated device, making sure that polarity is correct. If two or more current-actuated devices are to be used, they must be connected in se­ries (See Figure 2-4 below). Do not exceed the maximum load resistance of 1000 ohms.
Current and voltage outputs may be util­ized simultaneously if desired.
+
mA
-
755R
Analyzer
e: Total series resistance of all devices is not to exceed
ms.
+
-
+
-
+
Indicator
-
ecorder
er
emote
Figure 2-4. Model 755R Connected to Drive Several Current-Actuated Output Devices
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-7
Instruction Manual
748213-S April 2002
Model 755R
e. Output Connections and Initial Setup
for Dual Alarm Option
If so ordered, the analyzer is factory equipped with alarm output. Alternatively, the alarm feature is obtainable by subse­quent installation of the 654019 Alarm Kit.
No. 1
NO
Low Alarm, Fail-Safe
High Alarm, Fail-Safe
Low Control Limit, Fail-Safe
RESET
RESET
No. 2
No. 1
RESET
RESET
No. 2
No. 1
RESET
RESET
No. 2
COM
NC
NO
COM
NC
NO
COM
NC
NO
COM
NC
NO
COM
NC
NO
COM
NC
Alarm Bell or Lamp
Alarm Bell or Lamp
Solenoid
Valve
115 VAC
115 VAC
115 VAC
The alarm output provides two sets of re­lay contacts for actuation of alarm and/or process-control functions. Leads from the customer-supplied external alarm system connect to terminals on the 654019 Alarm Assembly (See Figure 2-5 below and In­terconnect Drawing 654014).
REQUIREMENT TYPICAL CONNECTIONSREQUIREMENT TYPICAL CONNECTIONS
Solenoid
No. 1
RESET
RESET
No. 2
No. 1
RESET
RESET
No. 2
No. 1
RESET
RESET
No. 2
NO
COM
NC
NO
COM
NC
NO
COM
NC
NO
COM
NC
NO
COM
NC
NO
COM
NC
N
H
Low Control Limit, Fail-Safe
Lower Low Alarm Indicator, Fail-Safe
N
H
H
N
Low Control, Fail-Safe
High Control, Fail-Safe
Higher High Alarm Indicator, Fail-Safe
Valve
Alarm Bell or Lamp
Solenoid
Valve
Solenoid
Valve
Alarm Bell or Lamp
115 VAC
115 VAC
115 VAC
115 VAC
115 VAC
H
N
N
H
H
N
H
N
N
H
Figure 2-5. Relay Terminal Connections for Typical Fail-Safe Applications
Note the following recommendations:
quency interference (RFI), it should be arc suppressed. The 858728 Arc
A fuse should be inserted into the line
Suppressor is recommended.
between the customer-supplied power supply and the alarm relay terminals on the Alarm Relay Assembly.
If at all possible, the analyzer should operate on a different AC power source, to avoid RFI.
If the alarm contacts are connected to any device that produces radio fre-
2-8 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
Model 755R
Instruction Manual
748213-S
April 2002
Alarm 1 and Alarm 2 output through the 654019 Alarm Relay Assembly is pro­vided by two identical single-pole, dou­ble-throw relays. These relay contacts are rated at the following values:
5 amperes 240 VAC resistive 1 ampere 240 VAC inductive 5 amperes 120 VAC resistive 3 amperes 120 VAC inductive 5 amperes 30 VDC resistive 3 amperes 30 VDC inductive
Removal of AC power from the analyzer (such as power failure) de-energizes both relays, placing them in alarm condition. Switching characteristics of the Alarm 1 and Alarm 2 relays are as follows:
The Alarm 1 relay coil is de-energized when the display moves downscale through the value that corresponds to setpoint minus deadband. This relay coil is energized when the display moves up­scale through the value that corresponds to setpoint plus deadband.
The Alarm 2 relay coil is de-energized when the display moves upscale through the value that corresponds to the setpoint plus deadband. This relay coil is ener­gized when the display moves downscale through the value that corresponds to setpoint minus deadband.
Both the ALARM 1 and ALARM 2 func­tions generally incorporate automatic rest. When the display goes beyond the pre­selected limits, the corresponding relay is de-energized. When the display returns within the acceptable range, the relay is turned on.
The ALARM 1 and/or ALARM 2 alarm functions may be converted to manual re­set. The conversion requires the substitu­tion of an external pushbutton or other momentary contact switch for the jumper that connects the RESET terminals on the Alarm Relay Assembly. If the corre­sponding relay is now de-energized (i.e., in alarm condition), the relay remains
de-energized until the operator momen­tarily closes the switch.
By appropriate connection to the dou­ble-throw relay contacts, it is possible to obtain either a contact closure or a con­tact opening for an energized relay. Also, either a contact closure or a contact opening may be obtained for a de-energized relay. It is important, for fail-safe applications, that the user under­stands what circuit conditions are desired in event of power failure and the resultant relay de-energization. Relay contacts should then be connected accordingly (See Figure 2-5, page 2-8).
The ALARM 1 and ALARM 2 circuits have independent setpoint and deadband ad­justments (See Figure 3-1, page 3-3). Ini­tially, the ALARM 1 and ALARM 2 Setpoint Adjustments must be calibrated by means of the ALARM 1 and ALARM 2 Calibration Adjustments by the following procedure:
1. Set RANGE Select in a position ap-
propriate to the span standard gas.
2. Inject span standard gas through ana-
lyzer at 50 to 500 cc/min.
3. Verify that ALARM 1 and ALARM 2
Deadband Adjustments (See Figure 3-1, page 3-3) are set for minimum value (turned fully counterclockwise). These potentiometers should be fac­tory-set for minimum deadband. Both potentiometers MUST REMAIN at this setting throughout calibration of the alarm setpoint adjustments.
4. Adjust ALARM 1 control function as
follows:
a. With ALARM 1 Setpoint Adjust-
ment at 100% (i.e., position 10 on dial), adjust front panel SPAN Control so that the display or re­corder reads exactly fullscale.
b. Set ALARM 1 Calibrate Adjust-
ment (R63) to its clockwise limit.
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-9
Instruction Manual
748213-S April 2002
Model 755R
Carefully rotate R63 counter­clockwise the minimum amount required to obtain energization of ALARM 1 Relay K1 (See Figure 2-6 below and Figure 3-1, page 3-
3). Energization may be verified by connecting an ohmmeter to relay terminals on 654019 Alarm Relay Assembly.
c. To verify correct adjustment of
R63, adjust front panel SPAN Control so that the display or re­corder reads 99% of fullscale. Relay K1 should now be DE-ENERGIZED.
5. Adjust ALARM 2 control function as follows:
a. With ALARM 2 Setpoint Adjust-
ment at 100% (i.e., Position 10 on the dial), adjust front panel SPAN Control so that
b. the display or recorder reads ex-
actly fullscale.
c. Set ALARM 2 Calibrate Adjust-
ment (R67) to its clockwise limit. Carefully rotate R67 counter­clockwise the minimum amount required to obtain energization of ALARM 2 Relay K2 (See Figure 2-5, page 2-8).
d. To verify correct adjustment of
R67, adjust front panel SPAN Control so that the display or re­corder reads 99% of fullscale. Relay K2 should now be DE-ENERGIZED.
The ALARM 1 and ALARM 2 Setpoint Adjustments are now properly calibrated and may be used to select the desired alarm setpoints, as described in Section 3-6 (page 3-7).
A. Typical ALARM 1 Setting
DEADBAND SET FOR
20% OF FULLSCALE
B. Typical ALARM 2 Setting
DEADBAND SET FOR
10% OF FULLSCALE
40
INPUT SIGNAL
Percent of Fullscale
30
20
55
INPUT SIGNAL
Percent of Fullscale
50
45
Figure 2-6. Typical Alarm Settings
When input signal moves upscale through this point, the coil of ALARM 1 relay (K1) is energized, providing continuity between the common and normally-closed contacts of the relay.
ALARM 1 Setpoint
When input signal moves downscale through this point, the coil of ALARM 1 relay (K1) is de-energized, providing continuity between the common and normally-open contacts of the relay.
When input signal moves upscale through this point, the coil of ALARM 2 relay (K2) is de-energized, providing continuity between the common and normally-open contacts of the relay.
ALARM 2 Setpoint
When input signal moves upscale through this point, the coil of ALARM 2 relay (K2) is energized, providing continuity between the common and normally-closed contacts of the relay.
2-10 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
Model 755R
2. CR! AND CR2 ARE ANY 600V, 1 AMP DIODE.
1. RELAYS SHOWN IN ENERGIZED POSITION. NOTES:
+15V
-15V
Alarm
No. 1
Command
Alarm
No. 2
Command
Instruction Manual
748213-S
April 2002
TB4
1
J5
1
2
4
6
CR1
14
K1
CR2
14 13
K2
5
13
12
1
5
12
9
8
9
8
NO
COM
NC
Reset
NO
COM
NC
Reset
Alarm
No. 1
Alarm
No. 2
Figure 2-7. Alarm Relay Assembly Schematic Diagram
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-11
Instruction Manual
748213-S April 2002
Model 755R

2-6 REMOTE RANGE CHANGE OPTION

The power supply circuitry on the Remote Range Board 646004 must be jumpered for the correct line voltage, either 115 VAC or 230 VAC. See Drawings # 656081 (Table 2) and 646090 for correct jumper locations.
On the Remote Range Board, an additional option exists: for using either the on-board 12 V to drive the range select relays or an exter­nal 12 V supply.
To use an external supply:
1. Remove the E to F jumper (DWG 646090).
2. Apply the external 12 V to J3-5.
3. Program the remote controller to pull the range bits, J3-1 through J3-4, low. (See Table 2-1 below.)
To use the internal 12 V supply:
1. Verify the E to F jumper is in place.
2. Connect the controller's common to J3-6 to reference the instrument's common to the controller's common.
NOTE
DO NOT connect anything to J3-5.
3. Connect J3-1 to J3-4, as shown in the truth table below, to switch ranges.
Remember that you are dealing with inverse logic and not normal binary addresses. Also, this process switches the recorder output only, and does not affect the front panel dis­play.
J3-4 J3-3 J3-2 J3-1 Hex
Range 1
Range 2
Range 3
Range 4
Note: 1 = 12 V, 0 << 1 V.

Table 2-1. Remote Range Switching Truth Table

1 1 1 0 E
1 1 0 1 D
1 0 1 1 B
0 1 1 1 7
2-12 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
Model 755R
Instruction Manual
748213-S
April 2002
SECTION 3

OPERATION

3-1 OVERVIEW

Preparatory to operation, a familiarization with Figure 3-1 (page 3-3) is recom­mended. This figure gives locations and summarized descriptions of operating ad­justments of the Model 755R Oxygen Ana­lyzer.

3-2 OPERATING RANGE SELECTION

The Model 755R is designed to operate on a single, field-selectable range. A new range may be selected any time the ana­lyzer application changes or any time cali­bration may require a range change.
To select the operating range, reposition the jumper shown in Figure 3-1 (page 3-3) to the desired location. Each position is la­beled as to its fullscale range. Only the analog output (voltage and optional current) is affected by range selection. The digital display always reads 100% oxygen.

3-3 STARTUP PROCEDURE

Inject a suitable on-scale gas (not actual sample) through the analyzer. Turn power ON. If digital display gives overrange indi­cation, the probable cause is the suspen­sion in the detector is hung up. To correct this condition, turn power OFF, tap detector compartment with fingers, wait 30 seconds, turn power ON.
proceed to Section 3-4 below. Otherwise, refer to Section 6, Maintenance and Serv­ice.

3-4 CALIBRATION

Calibration consists of establishing a zero calibration point and a span calibration point (see Table 3-1, page 3-4). Zero and span calibration should be performed on the range that will be used during sample analysis. In some applications, however, it may be desirable to perform span calibra­tion on a range of higher sensitivity (i.e., more narrow span) and then move the jumper to the desired operating range. For example, if the operating range is to be 0 to 50% oxygen, span calibration may be per­formed on the 0 to 25% range to permit use of air as the span standard gas.
a. Calibration with Zero and Span
Standard Gases
NOTE:
The same flow rate must be main­tained for zero, span, and sample to avoid measure error. The exhaust is vented to the atmosphere to avoid back pressure. The following pro­cedure is based on the standards in Table 3-2 (page 3-6). Performance specifications are based on recorder output.
When on-scale reading is obtained, allow analyzer to warm-up for a minimum of one hour with gas flowing. This warm-up is necessary because a reliable calibration is obtainable only after the analyzer reaches temperature stability. Moreover, the resul­tant elevated temperature will ensure against condensation within, and possible damage to the detector assembly. After warm-up, the digital display or recorder should give stable, drift-free readout. If so,
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-1
Set Zero Calibration Point
Inject nitrogen zero standard gas through analyzer at suitable flow rate, preferably 250 cc/min. Allow gas to purge analyzer for a minimum of three minutes.
Adjust ZERO control so that the read­ing on the digital display or recorder is zero
Instruction Manual
748213-S April 2002
Set Span Calibration Point
Inject span standard gas (see Table 3-1, page 3-4) through the analyzer at the same flow rate as was used for zero standard gas. Allow gas to purge analyzer for a minimum of three min­utes.
Adjust SPAN control so that reading on display or recorder is appropriate to the span standard gas.
3-5 COMPENSATION FOR COMPOSITION OF
BACKGROUND GAS
Any gas having a composition other than 100% oxygen contains background gas. The background gas comprises all non­oxygen constituents. Although instrument response to most gases other than oxygen is comparatively slight, it is not in all cases negligible. The contribution of these com-
Model 755R
ponents to instrument response is a func­tion of the span and range used, and can be computed for each individual case.
If the zero and span standard gases con­tain the same background gas as the sam­ple, the routine standardization procedure automatically compensates for the back­ground components. Therefore, the zero and span standard gases would introduce no error.
If the background gas in the sample is dif­ferent from that in the zero and/or span standard gas(es), background effects must be taken into consideration to ensure cor­rect readout. During adjustment of the front panel ZERO and SPAN controls (see Fig­ure 1-1, page 1-1), the instrument is not set to indicate the true oxygen content of the zero and span standard gases. It is set to indicate a slightly different value, relative to background gas, calculated to provide cor­rect readout during subsequent analysis of the sample gas.
3-2 Operation Rosemount Analytical Inc. A Division of Emerson Process Management
Model 755R
4
Instruction Manual
748213-S
April 2002
16
15
SPAN
E10
100% 50%
R8 R9
CR2
U6
I G O
1 2 3 4
I G O
U5
C5
I G O
U3
C4
U4
1
C59
E21 E1 E3 E5 E7
C60
C56
U18
C58
U16
E22 E2 E4 E6 E8
C49
C57
8
E11 E13 E15
E23E9
E12 E14 E16 E20
E24
10% 2.5%
25%
R3 R4
R2 R1 R5 R6
U2
C2
C3 CR1 C1
U1
J1
5V 1V .1V .01V
U13
R57
R58
R59
R60
7
E17
E19
TP5 TP6
TP7 TP 8 T P9 TP10 TP11 TP16 TP17 TP18
E18
1%
5%
C14
C61
R52
R87
R53
R82
U17
R47
R84
R85
R86
R102
C48
R43 CR5
R49
T1
R40
U12
C55
U15
C13
C12
U4
R21
R22R23R2
J2
R50
SPAN
SPAN
C65
C53
R20
CW
CW S
S CCW
CCW
U14
R56
R55
R42 CR4
C45
TP19
TP20
R89
R89
C68
+
T1
U11
U11
C51
C44
U21
C31
C31
5 4
3
2
R90
R90
R88 R30 R29 R100
R88 R30 R29 R100
C64
C64
R68
C17
C17
C39
C39
R25
R25
C29
C29 C26
C26
R13
R13
CW
CW S
S CCW
CCW
U20
U20
C38 C41
C38 C41
C16
C16 C18
C36
C36
R27
R27
C28 C27
C28 C27
R68
C8
C8
C1
C1
CR1
CR2R1R2R3R4R5R6
CR1
CR2R1R2R3R4R5R6
C30
C30
U8
U8
C63
C63
U19
U19
R11 R66 R77 R80 R82 R72 R70
R11 R66 R77 R80 R82 R72 R70
BL
CR3 R12 R76 R69 R81 R75 R71 R85
CR3 R12 R76 R69 R81 R75 R71 R85
R101
C50
C50
C37
C37
U10
U10
R54
R37
R37
R36
R31
R61
R39
R38
R28
R36
R31
R61
R39
R38
R28
ZERO
ZERO
6 10
R9
R9
R64
R64
Q1
Q1
E
E
R10 C6 6
R10 C6 6
R8
R8
R7
R7
C2
C2
C5
C5
J5
J5
13
R79 R74
R79 R74
Q2
Q2
U1
C4
C4
12
R63
R63
9
R67
R67
E
E
R73
R73
14
11
R78
R78
U2
U2
C9
C9
C6
C6
C10
C10
J4
J4
1
1
2
2
3
3
4
4
C67
C67
5
5
6
6
7
7
8
8
9
9
C7C3
C7C3
10
10
11
11
J6
J6
12
12
13
13
14
14
15
15
16
16
652830 SIGNAL CONTROL BOARD
652830 SIGNAL CONTROL BOARD
1. RECORDER OUTPUT selector plug Provides selectable output of 10 mV, 100 mV, 1 V or 5 V for a voltage recorder.
2. DIGITAL READOUT (R100) Calibration of digital readout.
3. AMPLIFIER U8 ZERO (R29) Initial factory zeroing of amplifier U8.
4. RESPONSE TIME (R30) Adjustment of electronic response time.
5. FULLSCALE OUTPUT (R88) Setting fullscale for 1 V, 0.1 V and 10 mV outputs.
6. DETECTOR COARSE ZERO (R9) Coarse adjustment of detector zero by shifting the position of the detector within the mag­netic field. It is adjusted during factory checkout, and does not require readjustment except if detector is replaced.
7. CURRENT OUTPUT ZERO (R1) Located on Current Output Board, adjustment for zero-level current output, i.e., 4mA or 0mA
8. CURRENT OUTPUT SPAN (R2) Located on Current Output Board, adjustment for fullscale current output: 20mA
9. ALARM 2 CALIBRATION (R67) Initial calibration of ALARM 2 circuit.
10. ALARM 2 SETPOINT (R68) Continuously variable adjustment of setpoint for ALARM 2 circuit, for actuation of external, customer supplied control device(s). Adjustment range is 0 to 100% of fullscale span.
11. ALARM 2 DEADBAND (R78) Adjustment of ALARM 2 deadband circuit from 1% to 20% of fullscale. Deadband is essen­tially symmetrical with respect to setpoint.
12. ALARM 1 CALIBRATION (R63) Initial calibration of ALARM 1 circuit.
13. ALARM 1 SETPOINT (R64) Continuously variable adjustment of setpoint for ALARM 1 circuit, for actuation of external, customer supplied control device(s). Adjustment range is 0 to 100% of fullscale span.
14. ALARM 1 DEADBAND (R73) Adjustment of ALARM 1 deadband circuit from 1% to 20% of fullscale. Deadband is essen­tially symmetrical with respect to setpoint.
15. OUTPUT RANGE selector plug Selectable fullscale output range.
16. DETECTOR ISOLATION plug For servicing and testing of the Control Board.
DIGITAL DISPLAY Display (viewed on front panel) indicates oxygen content of sample.
ZERO control (R13) Accessible on front panel, use to establish zero-calibration point.
SPAN control (R20) Accessible on front panel, use to establish span calibration point.
Figure 3-1. Control Board - Adjustment Locations
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-3
Instruction Manual
748213-S April 2002
Model 755R
RANGE % OXYGEN
0 to 1 Nitrogen 0.9% O2, balance N2 0 to 2.5 Nitrogen 2.3% O2, balance N2 0 to 5 Nitrogen 4.5% O2, balance N2 0 to 10 Nitrogen 9% O2, balance N2 0 to 25 Nitrogen Air (20.93% O2) 0 to 50 Nitrogen 45% O2, balance N2 0 to 100 Nitrogen 100% O2

Table 3-1. Calibration Range for Various Zero-Based Operating Ranges

a. Oxygen Equivalent Value of Gases

For computation of background correc­tions, the analyzer response to each component of the sample must be shown. Table 3-2 (page 3-6) lists the percentage oxygen equivalent values for many com­mon gases.
The percentage oxygen equivalent of a gas is the instrument response to the given gas compared to the response to oxygen, assuming that both gases are supplied at the same pressure .
RECOMMENDED ZERO
STANDARD GAS
RECOMMENDED SPAN
STANDARD GAS
In equation form:
%O2 Equivalent of Gas =
Analyzer Response to Gas
Analyzer Response to O
To select a random example from Table 3-2, if analyzer response to oxygen is +100%, the response to xenon would be
-1.34%.
The oxygen equivalent of a gas mixture is the sum of the contribution of the individ­ual gas components.
2
X 100%
Example: Zero Based Range
At lower range limit, i.e., 0% oxygen, composition of sample is 80% CO2, 20% N2.
From Table 3-1 (page 3-4), the % oxygen equivalents are CO2. -0.623 and N2, -0.358%.
% oxygen equivalent of mixture =
0.8 x (-0.623) + 0.2 x (-0.358) = (-0.4984) + (-0.0716) = -0.570% Oxygen

b. Computing Adjusted Settings for Zero and Span Controls

During instrument calibration, adjusted values may be required in setting the ZERO and SPAN controls to correct for the magnetic susceptibility of the background gas.
The quantities are defined as follows:
BGGst = Oxygen equivalent of background gas in standard gas (see Table 3-2, page 3-6)
BGGs = Oxygen equivalent of background gas sample (see Table 3-2, page 3-6)
OP = operating pressure. Unless special pressure corrections are to be made, the zero stan­dard, span standard and sample gases must all be admitted at the same pressure.
3-4 Operation Rosemount Analytical Inc. A Division of Emerson Process Management
Model 755R
Use the following equation to compute the adjusted settings for the ZERO and SPAN controls:
Adjusted % O2 for standard gas =
(A)[100 + (B-C)] - 100 [B-C]
Where:
A = true % O2 of standard gas
B = BGGs
C = BGGst
Example:
Instruction Manual
748213-S
April 2002
100
Background gas in sample is CO2, oxygen equivalent = -0.623% Zero gas is 100% N2 Span standard gas is air: 21% O2, 79% N2 Background gas in zero and span standard gases is N2, oxygen equivalent = -0.358% With N2 zero standard gas flowing, ZERO control is adjusted so digital display reads:
0[100+(-0.623-(-0.358))] - 100[-0.623-(-0.358)]
100
With air flowing, SPAN control is adjusted so the digital display reads:
21[100 - 0.265) - 100 (-0.265)
= 21.209% O
100
In two limiting cases, the general equation is reduced to simpler forms.
1. If the span standard gas is 100% oxygen, the adjusted oxygen value for setting the SPAN control is the same as the true value (i.e., 100% oxygen).
0.265% O
=
2
2
21.21
2. If the zero standard is an oxygen-free zero gas, the adjusted value for setting the ZERO control = BGGst - BGGs. (If the oxygen-free zero gas is more diamagnetic than the background gas in the sample, this difference is negative. The negative value may be set on the digital display or the re­corder if provided with below-zero capability.)
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-5
Instruction Manual
748213-S April 2002
Model 755R
GAS
Acetylene, C2H Allene, C3H
4
Ammonia, NH
2
3
EQUIV. % AS
-0.612
-0.744
-0.479 Argon, A -0.569 Bromine, Br 1,2-Butadiene C4H 1,3-Butadiene C4H n-Butane, C4H iso-Butane, C4H Butene-1, C4H cis Butene-2, C4H iso-Butene, C4H trans butene-2, C4H Carbon Dioxide CO
2
6
6
10
10
8
8
8
8
2
-1.83
-1.047
-1.944
-1.481
-1.485
-1.205
-1.252
-1.201
-1.274
-0.623 Carbon Monoxide, CO -0.354 Ethane, C2H Ethylene, C2H Helium, H n-Heptane, C7H n-Hexane, C6H cyclo-Hexane, C6H Hydrogen, H
6
4
e
16
12
12
2
-0.789
-0.553
-0.059
-2.508
-2.175
-1.915
-0.117 Hydrogen Bromide, Hbr -0.968 Hydrogen Chloride, HC1 -0.651 Hydrogen Fluoride, HF -0.253 Hydrogen Iodide, HI -1.403 Hydrogen Sulphide, C2S -0.751 Kryton, Kr -0.853 Methane, CH
4
-0.512 Neon, Ne -0.205 Nitric Oxide, NO +44.2 Nitrogen, N Nitrogen Dioxide, NO n-Octane, C8H Oxygen, O n-Pentane, C5H iso-Pentane, C5H neo-Pentane, C5H Propane, C3H Propylene, C3H
2
2
18
2
12
12
12
8
6
-0.358
+28.7
-2.840
+100.0
-1.810
-1.853
-1.853
-1.135
-0.903 Water, H2O -0.381 Xenon, Xe -1.340
Table 3-2. Oxygen Equivalent of Common Gases
3-6 Operation Rosemount Analytical Inc. A Division of Emerson Process Management
Model 755R
3-6 SELECTION OF SETPOINTS AND DEAD-
BAND ON ALARM OPTION
The ALARM 1 and ALARM 2 setpoint adjust­ments (see Figure 3-1, page 3-3) are adjust­able for any desired value from 1% to 100% of the fullscale analyzer span. The adjustment screws are graduated from 0 to 10.
Required dial settings for both setpoint ad­justments may be determined from either Fig­ure 3-2 or the appropriate equation that follows:
Zero-based operating range
Required control setting =
(desired alarm setpoint)(10)
fullscale span
Figure 3-2 example:
Operating range, 0 to 5% oxygen
Desired ALARM 1 setpoint = 4% oxygen
Turn potentiometer R64 to 8
Desired ALARM 2 setpoint = 3% oxygen
Turn potentiometer R68 to 6
The desired deadband may be selected via the appropriate trimming potentiometer, R73, for ALARM 1 deadband adjustment and R78 for ALARM 2 deadband adjustment. For any setpoint, deadband is adjustable from 1% of fullscale (counterclockwise limit) to 20% of fullscale (clockwise limit). Deadband is es­sentially symmetrical with respect to setpoint.

3-7 CURRENT OUTPUT BOARD (OPTION)

RANGE
%
OXYGEN
0 to 1
0 to 1
0 to 2.5
0 to 5
0 to 10
0 to 25
0 to 50
0 to 100
Instruction Manual
748213-S
April 2002
PERCENTAGE OXYGEN READOUT
versus
ALARM SETPOINT DIAL READING
Percentage Oxygen Readout
0.0 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9 1.0
0 1 2 3 4 5 6 7 8 9 10
Setpoint Dial Reading
Percentage Oxygen Readout
0.0 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9 1.0
0 1 2 3 4 5 6 7 8 9 10
Setpoint Dial Reading
Percentage Oxygen Readout
0.0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0
0 1 2 3 4 5 6 7 8 9 10
Setpoint Dial Reading
Percentage Oxygen Readout
0 1 2 3 4 5 6 7 8 9 10
0 1 2 3 4 5 6 7 8 9 10
Setpoint Dial Reading
Percentage Oxygen Readout
0 2.5 5 7.5 10 12.5 15 17.5 20 22.5 25
0 1 2 3 4 5 6 7 8 9 10
Setpoint Dial Reading
Percentage Oxygen Readout
0 5 10 15 20 25 30 35 40 45 50
0 1 2 3 4 5 6 7 8 9 10
Setpoint Dial Reading
Percentage Oxygen Readout
0 10 20 30 40 50 60 70 80 90 100
0 1 2 3 4 5 6 7 8 9 10
Setpoint Dial Reading
The Current Output is set at the factor for 4 to 20 mA. If a 0 to 20 mA output is required, re­adjust both the zero and span potentiometers
Figure 3-2. Dial Settings for Alarm Setpoint
Adjustments
(R1 and R2) on the Current Output Board.
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-7
Instruction Manual
748213-S April 2002

3-8 ROUTINE OPERATION

After the calibration procedure of Section 3-4 (page 3-1), admit sample gas to the analyzer at the same pressure and the same flow rates used for the zero and span gases. The in­strument will now continuously indicate the oxygen content of the sample gas.
3-9 EFFECT OF BAROMETRIC PRESSURE
CHANGES ON INSTRUMENT READOUT
If the analyzer exhaust port is vented through a suitable absolute backpressure regulator, barometric pressure changes do not affect the percent oxygen readout. However, if the ana­lyzer exhaust port is vented directly to the at­mosphere, any change in barometric pressure after instrument standardization will result in a directly proportional change in the indicated percentage of oxygen. This effect may be compensated in various ways. If desired, cor­rection may be made by the following equa­tion:
Model 755R

3-10 CALIBRATION FREQUENCY

The appropriate calibration interval will de­pend on the accuracy required in the particu­lar application, and is best determined by keeping a calibration log. If the analyzer ex­haust port is vented directly to the atmos­phere, the greatest source of error is normally the variation in barometric pressure. If de­sired, effects of barometric pressure variation can be minimized by calibrating immediately before taking readings, for example, at the beginning of each shift.
True % Oxygen = (Pst/Pan)(Indicated % Oxy­gen)
Where:
Pst = Operating pressure during stan­dardization
Pan = Operating pressure sample analy­sis
Example: U.S. Units
Pst = 760 mm Hg Pan = 740 mm Hg Indicated % O2 True % O2 = (760/740)(40%) = 41.1% O2
Example: S.I. Units
Pst = 101 kPa Pan = 98.2 kPa Indicated % O2 = 40% True % O2 = (101/98.2)(40%) = 41.1%
O2
3-8 Operation Rosemount Analytical Inc. A Division of Emerson Process Management
Model 755R
Instruction Manual
748213-S
April 2002
SECTION 4

THEORY

4-1 PRINCIPLES OF OPERATION

Oxygen is strongly paramagnetic while most other common gases are weakly diamagnetic. The paramagnetism of oxygen may be re­garded as the capability of an oxygen mole­cule to become a temporary magnet when placed in a magnetic field. This is analogous to the magnetization of a piece of soft iron. Diamagnetic gases are analogous to non-magnetic substances.
With the Model 755R, the volume magnetic susceptibility of the flowing gas sample is sensed in the detector/magnet assembly. As shown in the functional diagram of Figure 5-1, a dumbbell-shaped, nitrogen-filled, hollow glass test body is suspended on a plati­num/nickel alloy ribbon in a non-uniform mag­netic field.
Because of the “magnetic buoyancy” effect, the spheres of the test body are subjected to displacement forces, resulting in a displace­ment torque that is proportional to the volume magnetic susceptibility of the gas surrounding the test body.
Measurement is accomplished by a null-balance system, where the displacement torque is opposed by an equal, but opposite, restorative torque. The restorative torque is due to electromagnetic forces on the spheres, resulting from a feedback current routed through a titanium wire conductor wound lengthwise around the dumbbell.
In effect, each sphere is wound with a one-turn circular loop. The current required to restore the test body to null position is directly proportional to the original displacement torque, and is a linear function of the volume magnetic susceptibility of the sample gas.
The restoring current is automatically main­tained at the correct level by an electro-optical feedback system. A beam of light from the source lamp is reflected off the square mirror attached to the test body, and onto the dual photocell.
The output current from the dual photocell is equal to the difference between the signals developed by the two halves of the photocell. This difference, which constitutes the error signal, is applied to the input of an amplifier circuit that provides the restoring current.
When the test body is in null position, both halves of the photocell are equally illuminated, the error signal is zero, and the amplifier is unequal. This condition results in application of an error signal to the input of the amplifier circuit. The resultant amplifier output signal is routed through the current loop, thus creating the electromagnetic forces required to restore the test body to null position.
Additionally, the output from the amplifier is conditioned as required to drive the digital display, and recorder if used. The electronic circuitry involved is described briefly in Sec­tion 4-3 (page 4-4) and in greater detail in Section 5.
Rosemount Analytical Inc. A Division of Emerson Process Management Theory 4-1
Instruction Manual
748213-S April 2002
4-2 VARIABLES INFLUENCING PARAMAG-
NETIC OXYGEN MEASUREMENTS
Variables that influence paramagnetic oxygen measurements include: operating pressure (See Section 4-3a, page 4-4), sample tem­perature, interfering sample components, and vibration (See Section 2-1d, page 2-1).

a. Pressure Effects

Although normally calibrated for readout in percent oxygen, the Model 755R actu­ally responds to oxygen partial pressure. The partial pressure of the oxygen com­ponent in a gas mixture is proportional to the total pressure of the mixture. Thus readout is affected by pressure variations.
For instance, assume that an instrument is calibrated for correct readout with a standard gas containing 5% oxygen, ad­mitted at the normal sea level atmos­pheric pressure of 14.7 psia (101.3 kPa). If the operating pressure now drops to one-half the original value (i.e., to 7.35 psia {50.65 kPa}) and the calibration con­trols are left at the previously established settings, the display reading for the stan­dard gas will drop to 2.5%.
Model 755R
It is therefore necessary to calibrate the instrument at the same pressure that will be used during subsequent operation, and to maintain this pressure during op­eration.
Typically, the sample gas is supplied to the analyzer inlet at slightly above ambi­ent pressure, and is discharged to ambi­ent pressure from the analyzer outlet. However, in some applications, it is nec­essary to insert an absolute back pres­sure regulator into the exhaust line to prevent the readout error that would oth­erwise result from fluctuations in exhaust pressure. The regulator must be mounted in a temperature-controlled housing (See Section 2-3c, page 2-3).
Operation at negative gauge pressure is not normally recommended, but is used in certain special applications (See Section 2-3d, page 2-4).
CAUTION
PRESSURE MINIMUM
Never subject the sensing unit to an ab­solute pressure of less than 500 mm Hg (66.7 kPa).
4-2 Theory Rosemount Analytical Inc. A Division of Emerson Process Management
Model 755R
A
Displacement
Torque
Instruction Manual
748213-S
April 2002
Balancing Weight
Electromagnetic
xis
Platinum/Nickel Alloy Suspension Ribbon
TEST BODY DETAIL
Displacement
Torque
Restoring Torque
Restoring
Current
Mirror
Restoring Torque
Titanium Wire Conductor
Restoring Current
Electromagnetic Axis
Balancing Weight
Nitrogen-Filled Hollow Glass Test Body
Restoring
Magnet
Shaded Pole Pieces (4)
Dual Photocell BT1, BT2
Test Body
Source Lamp
DS1
DETECTOR/MAGNET
Current
ASSEMBLY
Zero
CONTROL
ASSEMBLY
Span
% Oxygen
Readout
Figure 4-1. Functional Diagram of Paramagnetic Oxygen Measurement System
Rosemount Analytical Inc. A Division of Emerson Process Management Theory 4-3
Instruction Manual
748213-S April 2002
Shaded Pole Piece
Model 755R
Figure 4-2. Spherical Body in Non-Uniform Magnetic Field

4-3 ELECTRONIC CIRCUITRY

Electronic circuitry is shown in the Control Board schematic diagram, Drawing 652826, and is described briefly in the following sec­tions. For detailed circuit analysis, refer to Section 5 Circuit Analysis.

a. Detector/Magnet Assembly

A cross-sectional view of the optical bench and detector assemblies is shown in Figure 4-3B, page 4-7. Source lamp DS1, powered by a supply section within the Power Supply Board assembly (See Section 4-3c, page 4-5) directs a light beam onto the mirror attached to the test body. The mirror reflects the beam onto dual photocell BT1, BT2.
Sphere (Magnetic Susceptibility = k
F
k
Sample Gas (Magnetic Susceptibility = k )
As percentage of oxygen in sample gas increases, displacement force (F
Note:
) increases.
k
in turn, supplies the restoring current to the titanium wire loop on the test body (See Section 4-1, page 4-1).
Detector temperature is sensed by ther­mistor RT1, an integral part of the detec­tor assembly (See Figure 4-3B, page 4-7). The thermistor provides the input signal to the detector temperature control section of the Power Supply Board assembly: HR1, mounted on the top of the magnet, and HR2, mounted permanently on the rear of the detector assembly.
b. Control Board and Associated Cir-
cuitry
The Control Board consists of signal con­ditioning and control circuitry.
)
o
The difference between the signals de-
This circuitry includes the following: veloped by the two halves of the photocell constitutes the error signal supplied to the input of amplifier U1 on the Control Board assembly. Amplifier U1 drives U2 which,
4-4 Theory Rosemount Analytical Inc. A Division of Emerson Process Management
Model 755R
Instruction Manual
748213-S
April 2002
Input Amplifier U1
This amplifier receives the error signal from the dual photocell of the detector as­sembly and drives amplifier U2.
Amplifier U2 and Associated Zero Ad­justment
Amplifier U2 supplies the restoring current to the titanium wire loop of the test body within the detector assembly. Front panel ZERO Control R13 applies an adjustable zero biasing signal to the input of U2 to permit establishing a zero calibration point on the display or recorder. With zero standard gas flowing through the ana­lyzer, the ZERO control is adjusted for the appropriate reading.
Amplifier U4 and Associated Span Adjustment
Amplifier U4 and associated feed back resistors provide a signal amplification of X4. Front panel SPAN adjustment R20 modifies the value of the input resistance and hence the signal amplification factor. Adjustment range is approximately ±30%.
Amplifier U8
This unity gain amplifier provides zeroing capability and a buffered output for the anticipation circuit feeding U10.
Amplifier U10
U10 is an inverting buffer amplifier that in­corporates an anticipation arrangement in its input network, thus providing slightly faster response on the readout device(s).
Potentiometer R30 provides a continu­ously variable adjustment of 5 to 25 sec­onds for the electronic anticipation time and is factory-set for 20 seconds.
Since the anticipation network attenuates the signal, a gain of 10 is provided in U10 to restore the signal to the desired full­scale range of 0 to 10 VDC.
The output signal from U10 is routed to
two output circuits: a digital and an ana-
log.
In the Digital Output Circuit, the signal
from U10 passes to an integrating ana-
log-to-digital converter. The resulting
digital signal drives the liquid crystal dis-
play.
In the Analog Output Circuit, the output
from U10 is provided as an input to the
recorder output amplifier. This circuitry
provides scale expansion, and amplifica-
tion preparatory to use for potentiometric
recorder, voltage-to-current conversion for
current recorder, and/or alarm functions.
Potentiometric output is strap-selectable
for 0 to 10 mV, 0 to 100 mV, 0 to 1 V, or 0
to 5VDC. Potentiometer R88 permits ad-
justment of recorder span on 0 to 1 V, 100
mV and 10 mV outputs.

c. Power Supply Board Assembly

The Power Supply Board assembly con-
tains power supply and temperature con-
trol circuitry. The assembly is mounted
within the analyzer case.
As shown in DWG 617186, the various
circuits operate on main power trans-
former T1. During instrument assembly,
the two primary windings of T1 are fac-
tory-connected for operation on either 115
VAC or 230 VAC, as noted on the name
rating plate.
The same circuit board contains the fol-
lowing:
Source Lamp Power Supply Section
This circuit provides a regulated output of
2.20 VDC to operate incandescent source
lamp DS1 within the optical bench as-
sembly. One secondary of main power
transformer T1 drives a fullwave rectifier
consisting of CR7 and CR8. The output of
DS1 is held constant by a voltage regu-
lator circuit utilizing U7, Q4 and Q5.
Rosemount Analytical Inc. A Division of Emerson Process Management Theory 4-5
Instruction Manual
748213-S April 2002
Model 755R
±15 V Power Supply Section
This section provides DC voltage required for various amplifiers and other circuits.
Fullwave rectifier bridge CR5 provides both positive and negative outputs. Each is routed through an associated series type integrated circuit, voltage regulator, providing regulated outputs of +15 V and
-15 V.
Detector Temperature Control Section
This section maintains the detector at a controlled temperature of 150°F (66°C). Temperature is sensed by RT1, a resis­tance element permanently attached to the detector assembly. The signal from the sensor is applied to amplifier AR6, which drives transistors Q2 and Q3, thus controlling application of DC power from full wave rectifier bridge CR6 to two heat­ers within the detector/magnet assembly: HR1, mounted on the top of the magnet and HR2, permanently mounted on the rear of the detector assembly.
Detector Compartment Temperature
Control Section
This section maintains the interior of the
detector compartment at a controlled
temperature of 140°F (60°C). Tempera-
ture is sensed by a thermistor located in
the detector compartment and plugged
into the Control Board assembly.
The circuit provides an on-off control of
heater element HR3 via TRIAC element
Q7. Heater HR3 is a part of the heater/fan
assembly.
a. Isolated Current Output Board (Op-
tional)
An isolated current output is obtainable by
insertion of an optional plug-in circuit
board into receptacle J1 on the Control
Board (see Figure 3-1, page 3-3). The
current outputs available by this board are
0 to 20 mA or 4 to 20 mA.
4-6 Theory Rosemount Analytical Inc. A Division of Emerson Process Management
Model 755R
t
Instruction Manual
748213-S
April 2002
Sample Pre-Heating Coil
Sample Inlet Tube
Sample Outlet Tube
Magnet Assembly
Detector Assembly
Optical Bench Assembly
Mounting Screws (2)
A. Exploded View of Detector/Magnet Assembly
Integral Heater (HR2)
Sample Ou
Sample In
Dual Photocell
Optical Bench Assembly Detector Assembly
Sensor (RT1)
Integral 5-Micron Diffusion Screen
Test Body
Mirror
Source Lamp
B. Sectional Top View of Optical Bench and Detector Assemblies
Connector J12
Connector J12
Photocell Lock S crews (2 )
Lamp Retaining Set Screw
Lamp Viewing Hole
Source Lamp Assembly
C. Exploded View of Optical Bench Assembly
Dual Photocell
Figure 4-3. Detector/Magnet Assembly
Rosemount Analytical Inc. A Division of Emerson Process Management Theory 4-7
Instruction Manual
748213-S April 2002
Model 755R
4-8 Theory Rosemount Analytical Inc. A Division of Emerson Process Management
Model 755R
Instruction Manual
748213-S
April 2002
SECTION 5

CIRCUIT ANALYSIS

5-1 CIRCUIT OPERATION

The electronic circuitry of the Model 755R Oxygen Analyzer consists of the following:
A detector compartment heater circuit.
A detector heater circuit.
A ±15 VDC power supply.
A voltage regulating circuit for a stable
light source.
A detector circuit with a first-stage am-
plifier to provide a feedback current for mechanical feedback to the detector and a scaling amplifier circuit to give an output change of 0 to +2.5 V for a 0 to 100% change of the operating span.
A digital output circuit for the digital
read-out.
An analog output circuit for recorder,
optional alarms and current output.
±15 VDC POWER SUPPLY
5-2
Refer to Drawing 617186. The components of the ±VDC power supply circuit are located in the lower left-hand corner of the Power Supply Board. 19 VAC should be measured with respect to ground at CR5 (WO4). +15 VDC should be measured at the C27 (+) lead and -15 VDC at the C28 (-) lead. If the specified voltage measurements are ob­tained, the power supply is working cor­rectly.

5-3 CASE HEATER CONTROL CIRCUIT

The case heater control circuit utilizes four voltage-comparators (LM339 quad com­parator). An understanding of how one of these comparators functions is necessary before any circuit analysis can be attempted.
In Figure 5-1 (page 5-2), comparators 1 and 2 are depicted having a comparator within an overall comparator symbol. Also within this symbol, the base of the NPN transistor is connected to the output of the compara­tor. A -15 VDC is supplied to the emitter. The collector is illustrated as the overall out­put for the comparator package.
When the non-inverting terminal of com­parator 2 is more positive than the inverting terminal, the transistor does not conduct and the collector of the transistor or comparator output is at whatever potential is then pres­ent on the collector.
When the non-inverting terminal of com­parator 2 is less positive (more negative) than the inverting terminal, the transistor conducts and the output of the comparator is
-15 V. This value is the output of the OR cir­cuit.
Comparator 2 is biased at 0 volts on the in­verting terminal. Comparator 1 is biased at about 159 mV on the non-inverting terminal. Positive feedback or hysteresis is built into each comparator circuit for stability or posi­tive action. This is achieved by the 20 M re­sistances, R70 and R73.
An approximate 8 V peak-to-peak AC signal is applied to comparators 1 and 2. As the signal starts going positive, comparator 2 transistor ceases conducting and compara­tor 1 transistor is off.
When the signal exceeds the +159 mV on the non-inverting terminal, it turns on com­parator 1 and the output is -15 V. Com­parator 1 stays on until the signal drops below +159 mV, at which time the output will be the value of the OR bus.
As the AC signal goes negative with respect to ground, the transistor of comparator 2 conducts and the output is again -15 V. The
Rosemount Analytical Inc. A Division of Emerson Process Management Circuit Analysis 5-1
Instruction Manual
748213-S April 2002
Model 755R
output remains at -15 VDC until the incom­ing signal crosses zero value and the posi­tive signal causes the comparator 2 transistor to cease to conduct.
Summing the effects of the two comparators in the OR circuit results in no output from the comparators for about 4° of the sine wave, 2° after the signal goes positive (0 to 2°) and 2° before the positive signal reaches 180° (178° to 180°).
During the period that neither comparator is conducting, the value on the OR bus is the potential from the temperature-sensing bridge plus the effect of the ramp generator, probably -1.88 ±0.03 V.
The on-off effect of the comparators to the OR circuit results in application of a posi­tive-going pulse (from -15 V to -1.89 V) to
-1.7V
the temperature bridge at the rate of 120 pulses per second.
Capacitor C36 is added to the input circuit to delay the incoming AC signal so that the pulses will occur at or just after the line fre­quency crossover point.
Circuits for a ramp generator and a tem­perature-sensing bridge are part of the case heater control circuit (See Figure 5-2, page 5-3 and Figure 5-3, page 5-3).
On initial application of power to comparator of Figure 5-2 (page 5-3), no potential exists on the inverting terminal because no charge exists on capacitor, C37. If the transistor of comparator 3 does not conduct, +15V is at the output terminal. With +15V at the output, the potential on the non-inverting terminals will be about ±2.3 V because of the resis­tance divider, R75, R76.
100µ
-15V
159mV
360
°
°
0
°
-
+
INPUT
R69
R71
21.5K
4.75K
2M
R72
COMP 1
COMP 2
+15V
R68
3.3K
0
°
ON ONOFF
OFF OFFON
-
+
C38
0.18uF
1
180
+15V
-15V R70
20M
Figure 5-1. Two-Comparator OR Circuit
2
+15V
-15V R73
20M
180
-1.88 VDC
°
OUTPUT
Source
5-2 Circuit Analysis Rosemount Analytical Inc. A Division of Emerson Process Management
Model 755R
Instruction Manual
748213-S
April 2002
120 V
RMS
T1
R72
4.75K
19 VAC
12
TO POWER SUPPLY
19 VAC
R67 10K
C36
.18uF
+15V
C39 .01uF
-
1
+
R70
20M
-
2
+
R73
20M
9.07K
RT1
R82
R74
590K
1.0uF
R84 169K
C37
R83
63.4K
R85
11.0K
-15V
CR9
-15V
CR10
R69 2 M
R71
21.5K
R68
3.3K
Figure 5-2. Case Heater Control Circuit
-
+
R76
37.4K
-
+
R78
249K
3
R75
210K
R86
20M
4
C40 2200uF
R77
10K
R79 10K
R80 10K
CR11
Q6
R81
56.2
.18uF
R87 10K
T2
C38
-15V
INPUT FROM MULTIVIBRATOR
OFF OFF
R82
9.09K
RT1
OFF
+2.3V
-2.3V
R78
249K
+15V
-15V to 1.88V ±0.3V
R83
63.4K
-15V
R84 169K
R74
590K
C37
1.0uF
TO COMPARATOR
R76
37.4K
-
3
+
-15V R75
210K
C40 2200uF
Figure 5-3. Ramp Generator Circuit
R77 10K
R79 10K
R80 10K
6 Hz
+15V
Q6
T2
R81
56.2 C38
.18uF
R87
10K
-15V
Rosemount Analytical Inc. A Division of Emerson Process Management Circuit Analysis 5-3
Instruction Manual
748213-S April 2002
Model 755R
Capacitor C37 will now start to charge posi­tively through R78. When the positive poten­tial across C37 and at the inverting terminal of comparator 3 exceeds the potential on the non-inverting terminals, the transistor con­ducts. The output is -15 V. A full 30 V drop appears across R77.
The potential on the non-inverting terminal will now be about -2.3 V. C37 will not discharge through R78 until its potential exceeds that on the non-inverting terminal. At that time, com­parator 3 will switch polarity and start charging C37 again. The result is that the potential across C37 will vary almost linearly with time and form a ramp signal of about 6 Hz.
As the potential across C37 increases and decreases linearly, it affects the potential at the top of the bridge circuit between R82 and R83 through R74. Because of the ramp action charging and discharging C37, the potential between R82 and R83 varies approximately from -1.85 V to -1.92 VDC.
The temperature sensing device, RT1, in the bridge circuit is a thermistor. The bridge is de­signed to control the temperature in the case at 135°F (57°C). When the temperature is 135°F (57°C), the resistance of the thermistor RT1 will be at its lowest and the potential at the junction of RT1 and R84 should be the same as the junction of R82 and R83. Com­parator 4 (See Figure 5-4, page 5-6) does not allow pulses from the OR circuit (comparators 1 and 2) to operate Q6 or Triac Q7 in the case heater (See Figure 5-5, page 5-7).
vary from 0mV to some absolute value. The polarity of the error signal will depend on the deviation from the desired temperature and the ramp value at the function of R82 and R83.
The input from the OR circuit comparator (See Figure 5-1, page 5-2) is either -15 VDC or the ramp effect on the bridge. When -15V, the junction of R82 and R83 is also this value. The error signal into comparator 4 is nega­tively large to the inverting terminal. Com­parator 4 output transistor does not conduct. The base of Q6 is positive; therefore, Q6 does not conduct and a charge builds up on ca­pacitor C38.
The input from the OR comparators 1 and 2 form multivibrator circuit, pulses 120 times a second. For about 100 microseconds the junction of R82 and R83 is some value be­tween -1.85 V and -1.92 V, depending on the ramp generator. For this brief period of time (one pulse), comparator 4 compares the po­tential of junction R82, R83 with junction RT1, R84 of the bridge circuit. If the temperature at RT1 is low, the potential at the non-inverting terminal of comparator 4 is more negative and the output is -15 V.
The base of Q6 is zero, because of the volt­age drops across R79 and R80. Therefore, Q6 conducts. Energy, stored in C38, flows through Q6 as current and capacitor C38 dis­charges to zero potential. No current flows through the primary winding of transformer T2.
Theoretically, at 135°F (57°C) the potential at the junction of RTR1 and R84 is -1.85VDC. This is equivalent to a resistance of 21.2 K. By substituting a decade box for the thermistor and placing 20.2 K into the bridge, the heater should be off. With 22.7 K, the heater should be full on.
Since the potential at the junction of R82 and R83 can vary between 1.85V and 1.92V ac­cording to the 6 Hz ramp, and the potential at the junction of RT1 and R84 may vary around or within these limits, depending on tempera­ture, the error signal to comparator 4 may
5-4 Circuit Analysis Rosemount Analytical Inc. A Division of Emerson Process Management
At the end of the 100 microsecond pulse, the NPN transistor in the output of comparator 4 ceases to conduct, so the signal on the base of Q6 is +15V. Q6 ceases to conduct. C38 starts to charge, driving electrons (current) through the primary of T2. This induces a pulse into the secondary of T2 and to the gate of Triac Q7 turning it on.
At the beginning of the next 100 microsecond pulse, comparator 4 output is again -15V, with zero volts on the base of Q6. Q6 again con­ducts, discharging C38. At the end of the 100 microsecond pulse, Q6 ceases to conduct.
Model 755R
Instruction Manual
748213-S
April 2002
C38 charges and a pulse appears at the gate of Triac Q7, turning it on again.
The charging time for C38 is about one-half a time constant (C38, R87) and ten time con­stants (R81, C38) are available for discharg­ing C38.
The above action is repeated as long as the temperature is low, causing an error between R82, R83 junction and RT1, R84 junction. As the temperature approaches the desired case temperature of 135°F (57°C), differences be­tween these two junctions will exist for only part of each ramp and the number of pulses operating Q7 will be proportional to the amount of error sensed by the 6 Hz ramp.
The pulses arrive at Q7 just as the supply AC line voltage is passing the zero-volt crossover
point. The purpose of C36 is to delay the tim­ing pulse, relative to line frequency, so that a pulse arrives at the gate of Triac Q7 as the line potential just passes the zero-volt cross­over point (0° and 180° of line phase).
Varistor, RV1 is a temperature sensitive re­sistance device. When case temperature is low, such as ambient, the value of RV1 is low. Applying power at that temperature might cause a current surge to damage Triac Q7. RV1 with its low initial value of resistance acts as a bypass and most of the current is shunted through it. As the temperature in­creases and approaches the desired case temperature, the resistance of RV1 increases to a large value. This limits the current through it and gives fine control of the heater to Triac Q7 and the temperature-sensing cir­cuit.
Rosemount Analytical Inc. A Division of Emerson Process Management Circuit Analysis 5-5
Instruction Manual
748213-S April 2002
Model 755R

5-4 DETECTOR HEATER CONTROL CIRCUIT

Figure 5-4 below is a simplified heater control circuit drawing for the detector. Heaters 1 and 2 are actually connected in parallel and have a combined resistance of about 17 ohms.
The thermistor resistance (RT1) in the resis­tance bridge varies inversely with tempera­ture. The bridge is designed to maintain the temperature of the detector at 150°F (65.5°C).
The junction point between R55 and R56 is maintained at a specific voltage since these resistances maintain a definite ratio. The thermistor resistance is 149 K at 150F (65.5°C) and increases rapidly as the tem­perature decreases. R59 in this bridge circuit represents the setpoint value for temperature. Suppose that, at temperature, resistance of the bridge (R55, R56, R59 and RT1) equals 149 K.
If the temperature goes down, RT1 increases in resistance and causes the junction of RT1 and R59 to go positive in voltage value. Since R55 and R56 are of equal resistance, their junction is at zero volts. Therefore, terminal 3 of AR6 is more positive than terminal 2 and the base of Q2 is positive. Q2 conducts, al­lowing alternating current to flow through heaters 1 and 2. The voltage drop across the heaters, when completely cold, would be about 20 VAC and, when controlling, would be AC of very low amplitude.
As the temperature increases, the resistance of RT1 decreases and the junction point be­tween RT1 and R59 becomes less positive. Terminal 3 of AR6 becomes less positive with respect to terminal 2. The output of AR causes Q2 and Q3 to conduct less. When terminal 3 equals terminal 2, or is less than terminal 2, the output of AR6 is zero or less. Q2 and Q3 do not conduct and the heater would not be supplying heat energy to the detector.
120 V
RMS
R59 700K
RT1
HR1 +2
C31
.01uF
2
3
R88 5M
R58
-
+
F1
25 VAC
+15V
R55
700K
R56
149K
-15V
Figure 5-4. Detector Heater Control Circuit
CR6 WO4
5M
U6
R60 100
R62
1K
6
CR12
Q3
Q2
R61
2.0
5-6 Circuit Analysis Rosemount Analytical Inc. A Division of Emerson Process Management
Model 755R
Instruction Manual
748213-S
April 2002
5-5 DETECTOR LIGHT SOURCE CONTROL
CIRCUIT
Refer to Figure 5-5 below. The detector light source control circuit maintains the light out­put from the bulb (DS1) as uniform as possi­ble, regardless of voltage fluctuations or aging of the bulb.
The power source for the light bulb is a cen­ter-tapped secondary of transformer T1. This AC voltage is rectified by CR7 and CR8 and filtered (C32), presenting an approximate +8.5 V bus to the current-limiting Darlington con­figuration of Q4.
Q4 controls the basic amount of current through DS1.
Amplifier AR7 has a fixed value, approxi­mately +2.2 VDC on terminal 3. The output of AR7 is positive, causing Q4 to conduct. As Q4 conducts, electrons flow from the center-tap of T1 to ground and from ground through DS1 for an input voltage to terminal 2 of AR7, through R66 to develop a bias on the base of Q5, through Q4 to the +8.5 V bus, and back to the secondary. As Q5 conducts, some of the
current going through DS1 is shunted from the main current path, and goes through Q5, which acts as a variable feedback resistance, goes to the positive output potential of AR7.
As DS1 ages, its light emission decreases and its resistance increases. The current through DS1 tends to decrease, causing a decrease in the voltage drop across DS1 and the input potential to terminal 2 of AR7. Now the output AR7 will increase, causing Q4 to conduct more current through R66. As the potential across R66 increases, Q5 will con­duct more current, causing a further increase in current flow through DS1. The net result is that the voltage across DS1 will remain uni­form and the operation of Q4 and Q5 will ad­just the gain of AR7 to maintain the light emission from DS1 uniform for a long period of time.
Voltage fluctuations in the 115 VAC supply could cause some variation in the amount of current flowing through the bulb DS1. How­ever, the voltage drop across DS1 would cause AR7 to adjust Q4 and the voltage drop across R66 to adjust Q5. The net result would still be uniform current flow through DS1 and uniform light emission.
T1
6.1 VAC
120 V
RMS
6.1 VAC
CR7
CR8
2000uF
C31
VR3
9.0V
+
+15V
R63
7.5K
R64 14K
R65 4530
α
α
+8.5V BUS
2
-
+
3
2.2V
AR7
C34 .01uF
C35 .01uF
Q5
Q4
R66
1.0
DS1
Figure 5-5. Detector Light Source Control Circuit
Rosemount Analytical Inc. A Division of Emerson Process Management Circuit Analysis 5-7
Instruction Manual
748213-S April 2002
Model 755R

5-6 DETECTOR WITH FIRST STAGE AMPLIFIER

Refer to Figure 5-6, page 5-9. The detector assembly consists of a test body suspended on a platinum wire and located in a non-uniform magnetic field.
The test body is constructed of two hollow glass spheres forming a dumbbell shape. They are filled and sealed with pure, dry nitro­gen. Around the test body, a titanium wire is chemically etched in order to form a feedback loop that can create a counteracting magnetic force to the test body displacement caused by oxygen concentration in the test assembly magnetic field.
Attached to the center arm of the test body dumbbell is a diamond-shaped mirror. At­tached to the mirror are two separate platinum wires in tension with the supports for the test body. The supports are isolated from ground and are electrically connected to the feedback loop and the electronics for that loop. The platinum wires form a fulcrum around which the test body pivots.
The detector operates in the following fashion. If the sample gas contains oxygen, it collects in the non-uniform magnetic field around the test body. Oxygen, because of its paramag­netic qualities, gathers along the magnetic lines of flux and forces the dumbbell of the test body out of the magnetic field.
A light source is focused on the test body mir­ror. As the test body moves out of the mag­netic field, the mirror distributes light unevenly on two photocells (BT1 and BT2). The photo­cells create a current proportional to light. This current is converted to a ± voltage by U1 and U2 located on the connector board in the de­tector housing. This voltage is then presented to comparator U1 on the controller board. The output of U1 goes to U2. The output of U2 causes current to flow through the feedback loop attached to the dumbbell.
This feedback current creates an elec­tro-magnetic field that attracts the dumbbell and mirror into the test assembly magnetic field until the mirror reflects light almost uni-
formly on each photocell. A current propor­tional to the oxygen concentration in the mag­netic field of the test assembly has to be flowing through the feedback loop in order to maintain balance and provide a reading of the oxygen content of a sample.
Resistances R7, R8 and the resistance of the wire in the feedback loop determine the gain of amplifier U2. The mirror on the dumbbell is positioned by the amount of current in the feedback loop. The mirror reflects light from the source (DS1) to the photocells (BT1, BT2). This repositioning of the mirror is a form of mechanical feedback to the input of the amplifier U1.
The net result is that the output of U1 could vary from 0 to -70 mV, or 0 to -7.0 V, de­pending on the range of the instrument. R4, C3 and R5, C7 form damping circuits for the input amplifier U1 and to smooth out noise that might be introduced by the measurement source.
Diode CR2 is a low-leakage device. Its pur­pose in the circuit is to ensure that the dumb­bell and mirror are positioned correctly with respect to the photocells on initial application of power.
If the dumbbell was out of position on start-up, the mirror might reflect light from the source onto one of the photocells. If the photocell output was positive, the current in the feed­back loop would be in the wrong direction and its electromagnetic field would cause the dumbbell to be further repelled from the per­manent magnetic field. The result would be error, not balance.
On application of AC power, capacitor C1 has no charge. The current will have to flow through R2. Initially the full 30 V drop (the difference between the +15 VDC and -15 VDC power) will appear cross R2. The cath­ode of CR2 will be initially at -15 VDC. The anode of CR2 will be some value more posi­tive than -15 VDC. CR2 will conduct. The in­put terminal of U1 will be negative and the current through the feedback loop around U2
5-8 Circuit Analysis Rosemount Analytical Inc. A Division of Emerson Process Management
Model 755R
p
Instruction Manual
748213-S
April 2002
will cause the dumbbell and mirror to be posi­tioned correctly in the test body.
As the charge on C1 increases, the cathode of CR2 becomes more positive. When it ex­ceeds that on the anode, CR2 ceases to con­duct and isolates the +15 VDC and -15 VDC power supply from the input circuit.
The front panel zero potentiometer R13 and detector coarse zero potentiometer add or subtract current to the input of U2 to offset any currents that may occur because of any
+15V
-15V
CONTROL
BOARD
C1
3.3uF
TP6
R3
110
TP7
1000
R1
10
R2
249K
C2
BT1
BT2
CURRENT FEEDBACK LOOP
DS1
DETECTOR
HOUSING
R1 1K
-
U2
+
R3 1K
-
U1
+
R2 1K
R2 1K
imbalance in the detector and the photocells BT1 and BT2.
The output current that U2 must provide to re­store the dumbbell is a measure of the dis­placing force and thus is a function of both (a) the % oxygen concentration of the sample and (b) the sample pressure.
The output from the U1 and U2 loop is further amplified by U4 to provide a 0 to 10 VDC out­put that constitutes signal V.
R23
TP8
E21 E22
C4
.01uF
CR2
-
U1
+
C3
.47uF
R6
118K
R4
1.13K
f
R5 2M
C7 .47uF
R7
1.77K
R20 20K
TP20
R8
1.77K
R21
49.9K
R22
49.9K
R12
200K
R10
3.01K
C8
.0022uF
-
U2
+
150K
-
+
+15V
-15V
+15V
-15V
U4
R13 20K
R9 20K
TP10
SIG. Vx 0 -+ 10V
FRONT PANEL ZERO
DETECTOR COARSE ZERO
Figure 5-6. Detector with First Stage Amplifier
Rosemount Analytical Inc. A Division of Emerson Process Management Circuit Analysis 5-9
Instruction Manual
748213-S April 2002
Model 755R
5-7 BUFFER AMPLIFIERS U8 AND U10 WITH
ASSOCIATED ANTICIPATION FUNCTION
Refer to Figure 5-8, page 5-12. U8 is a unity gain amplifier that provides zeroing capability and a buffered output for the anticipation cir­cuit feeding U10.
U10 is an inverting buffer amplifier that incor­porates an anticipation arrangement in its in­put network, thus providing slightly faster response on the readout device(s).
Potentiometer R30 provides a continuously variable adjustment of 5 to 25 seconds for the electronic response time (90% of fullscale) and is factory-set for 20 seconds.
Since the anticipation network attenuates the signal, a gain of 10 is provided by the feed­back network associated with U10 to restore the signal to the desired fullscale range of 0 to 10 VDC.
The output signal from U10 is routed to two output circuits:
Analog output circuits for recorder, V/I and alarms (See Section 5-9, page 5-11).

5-8 DIGITAL OUTPUT CIRCUIT

Refer to Figure 5-7 below. The output signal from buffer amplifier U10 is routed through an attenuator and filter network to an integrating analog-to-digital converter. It converts the sig­nal into an equivalent digital value in the range of 0.00% to 99.99%. Any value above
99.99% will be preceded by an over-range bit, for example, 1.1123.
The output of the ADC consists of bi­nary-coded decimal characters that are input to the liquid crystal controller and display chip characters sequentially in time. The BCD characters are converted into seven-line codes to drive the bar segments of the liquid crystal display.
A separate regulator circuit, which operates from the +15 VDC supply, provides a regu­lated 5 VDC for the digital functions associ­ated with the display.
Digital output circuit (See Section 5-8, page 5-10).
R37
2M
C38
.22uF
-
U10
+
R30 20K
R31
2K
TP11
C31
1.0uF
R36
1M
R61
1M
.47uF
C36
Figure 5-7. Buffer, Anticipation, and Digital Output Circuits
R38
100K
R39 11K
To Analog Output Circuit (Figure 6-8)
TP16
R49 20K
REF
C40
1.0uF
8052A
ADC
71C03
DRIVER AND
TP16
DISPLAY
CONTROL
5V
DIGITAL
DISPLAY
U18
5V
REGULATOR
+15V
5-10 Circuit Analysis Rosemount Analytical Inc. A Division of Emerson Process Management
Model 755R
Instruction Manual
748213-S
April 2002
5-9 ANALOG OUTPUT CIRCUITS FOR RE-
CORDER AND ALARMS
Refer to Figure 5-8, page 5-12. The analog output circuits utilize two amplifiers, first-stage amplifier and second-stage amplifier.

a. First Stage Amplifier

Permits selection of the desired fullscale oxygen range for the recorder via jumper-selectable signal amplification for scale expansion. This amplifier permits selecting the desired fullscale oxygen range for the recorder by an appropriate jumper selection of one of seven recorder spans. The following recorder spans are available: 1, 2.5, 5, 10, 25, 50, and 100%.

b. Second Stage Amplifier

Provides (a) a jumper-selectable output for a potentiometric recorder and (b) an output to drive the voltage-to-current and/or alarm option(s), if used. This am­plifier is an inverting configuration that provides a signal attenuation of 2X, thus reducing the 10-volt fullscale input signal to obtain a 5-volt fullscale output. This output is routed to:
1. Recorder Output Resistor Network. It provides a jumper-selectable output of 0 to 10 mV, 0 to 100 mV, 0 to 1 V, or 0 to 5 VDC for a potentiometric re­corder.
2. Current Output Receptacle J1. This connector accepts the optional plug-in current-output board.
3. Dual Alarm Amplifier Circuit. This cir­cuit drives the optional 654019 Alarm Relay Assembly.
Oxygen is strongly paramagnetic while most other common gases are weakly diamagnetic. The paramagnetism of oxy­gen may be regarded as the capability of an oxygen molecule to become a tempo­rary magnet when placed in a magnetic field. This is analogous to the magnetiza­tion of a piece of soft iron. Diamagnetic gases are analogous to non-magnetic substances.
With the Model 755R, the volume mag­netic susceptibility of the flowing gas sample is sensed in the detector/magnet assembly. As shown in the functional dia­gram of Figure 5-1 (page 5-2), a dumb­bell-shaped, nitrogen-filled, hollow glass test body is suspended on a plati­num/nickel alloy ribbon in a non-uniform magnetic field.
Because of the “magnetic buoyancy” ef­fect, the spheres of the test body are subjected to displacement forces, result­ing in a displacement torque that is pro­portional to the volume magnetic susceptibility of the gas surrounding the test body.
Rosemount Analytical Inc. A Division of Emerson Process Management Circuit Analysis 5-11
Instruction Manual
+
+
748213-S April 2002
Model 755R
FROM U10
R84
20K
R85
2M
R102
40K
R86
80K
R52
200K
R87
400K
R53
800K
R82
2M
100
50
25
10
5
2.5
1
C46
.1uF
-
U13
Recorder Span (Jumper Selectable)
TP18
R50A
20K
-
R50B
20K
R50C
20K
C55
.1uF
U16
R88
500
R57
3.83K
5V
E1 E2
1V
To Alarm and V/I
To Recorder
R58
909
R59
90.9
R60
E3 E4
100mV
E5 E6
10mV
E7 E8
10
Recorder Output (Jumper Selectable)
Figure 5-8. Simplified Analog Recorder Output Circuit
5-12 Circuit Analysis Rosemount Analytical Inc. A Division of Emerson Process Management
Model 755R
Instruction Manual
748213-S
April 2002
SECTION 6

MAINTENANCE AND SERVICE

The information provided in this section will aid in isolation of a malfunction to a particular assembly or circuit board. A few detailed checks are included, to aide in locating the defective assembly.
It is recommended that those familiar with cir­cuit analysis, refer to Section 6 Circuit Analy­sis of this manual.
WARNING
ELECTRICAL SHOCK HAZARD
Do not operate without doors and covers secure. Servicing requires access to live parts which can cause death or serious injury. Refer servicing to qualified per­sonnel.
For safety and proper performance this in­strument must be connected to a properly grounded three-wire source of power.
Optional alarm switching relay contacts wired to separate power sources must be disconnected before servicing.
WARNING
PARTS INTEGRITY
Tampering or unauthorized substitution of components may adversely affect safety of this product. Use only factory docu­mented components for repair
6-1 INITIAL CHECKOUT WITH STANDARD
GASES
If instrument readings do not meet specifica­tions, the first step in troubleshooting is to isolate the analyzer from the sample stream and the sample handling system.
Admit zero and span standard gases to the analyzer. Observe readout on digital display, and on recorder, if used.

Digital display gives correct reading with standard gases, but not with sample gas

The sample and the sample handling system are suspect. Check these areas.

Digital display gives correct readings with standard gases, but the alarm or output devices do not

Check these devices individually.

Digital display gives overrange readings with standard gases, as well as sample gas

The problem is likely with detector or the electronic circuitry. Turn power OFF. Tap detector compartment with fingers, wait 30 seconds, reapply power. If the suspension within the detector assembly is hung up, this may correct the problem. If not, proceed with checks of the detector and electronic circuitry.

Digital display gives erratic readings with standard gases, as well as sample gas

If zero and span standard gases give noisy or drifting readings, the problem is probably in the detector or the temperature control cir­cuits. Proceed with checks of the detector and electronics. In general, before concluding that the detector is defective and must be re­placed, verify correct operation of all circuits that could cause erratic readings.

a. Control Board Checkout

The Detector Isolation Plug located on the Control Board (Figure 3-1, page 3-3), re­moves the detector signal, allowing the input voltage to go to zero. The display should register near zero or on scale, and TP20 should read zero voltage. To test the remainder of the measuring circuit, do the following:
Voltage
1. Set RANGE Select to lowest range.
Rosemount Analytical Inc. A Division of Emerson Process Management Maintenance and Service 6-1
Instruction Manual
748213-S April 2002
Model 755R
2. Adjust R29 clockwise and counter­clockwise. The display should follow accordingly and remain steady within the adjustment limits of R29. If this condition is met, refer to Section 6-6a (page 6-7) for Control Board setup. Before replacing the Control Board, test for -15V at the junction of C1/J4-
7. Use the junction of CR1/R2 for +15V, or any source of ±15V on the board for the respective voltages.
3. If adjustment of R29 is not possible, replace the Control Board.
Alarms
Set RANGE Select to lowest range or use zero and span gases.
Current Output
Set RANGE Select to lowest range or use zero and span gases.
When checkout complete, re-install De­tector Isolation Plug. Configure Control Board to original setup.
If the Control Board functions correctly, the problem is either located in the De­tector/Magnet Assembly or related to temperature control.

6-2 HEATING CIRCUITS

To ensure against damage from overheating in the event of malfunction, the heating cir­cuits receive power via thermal fuses F2 and F3. If temperature of a heated area exceeds the permissible maximum, the associated fuse melts, opening the circuit.
NOTE:
The thermal fuses should be plugged in, NOT SOLDERED, as the fuse element might melt and open the circuit.

a. Case Heater Control Circuit

The case heater control circuit receives power via thermal fuse F2 (setpoint
75°C). This fuse, accessible on the Power Supply Board, may be checked for continuity.
Detector compartment heater element HR3, mounted on the heater/fan assem­bly, has a normal resistance of 20 ohms.
To verify heater operation, carefully place a hand on top of detector compartment. Heat should be felt. If not, check the case heating circuit.
Temperature sensor RT1 has a cold re­sistance of 22.7K ohms and a normal op­erating resistance of 20.2K ohms, indicating normal operating temperature.
As a further check, disconnect plug P6 on the Control Board, thus disconnecting temperature sensor RT1. Substitute a decade resistor box to simulate the re­sistance of RT1. Also, connect an AC voltmeter from the hot side of the line to the neutral side of F2, located inside the detector compartment.
Set the decade box for 20.2K ohms to simulate RT1 at controlling temperature. The voltmeter should show pulses of 1 VAC.
CAUTION
OVERHEATING
Avoid prolonged operation with the dec­ade box set at 22.2K ohms, overheating may result.
Set the decade box for 22.2K ohms to simulate RT1 resistance at ambient tem­perature. The voltmeter should show pulses of 120 VAC.

6-3 DETECTOR/MAGNET HEATING CIRCUIT

Heater HR1 is attached to the magnet. Heater HR2 is attached to the rear of the de­tector. Combined resistance of these two parallel-connected heaters, as measured at
6-2 Maintenance and Service Rosemount Analytical Inc. A Division of Emerson Process Management
Model 755R
Instruction Manual
748213-S
April 2002
pins 15 and 16 of the detector connector J12, should be approximately 89 ohms.
VDC when cold and will drop to approximately
0.4 VDC at control temperature. Temperature
sensor RT1 is mounted in the detector, with If resistance is correct, and the combined re­sistance is incorrect, heater HR1 may be open.
leads accessible at pins 10 and 11 of detector
connector J12. The sensor resistance should
be 1M ohms at 25°C and approximately 149K
ohms at operating temperature of 65°C. To reach the leads of HR1, remove the circuit board on the heater assembly. Resistance of HR1 should be approximately 21 ohms.
To check operation of the heater circuit, con­nect a voltmeter across R61 on the Power Supply Board. Normally, the voltage will be 4
A. Detector/Magnet Assembly - Exploded View B. Optical Bench - Exploded View
Sample Pre-Heating Coil
Sample Inlet Tube
Sample Out­let Tube
Mounting Screws (2)
Connector Board
Detector Assembly
Optical Bench Assembly
Photocell Lock Screws (2)
Lamp Retaining Set Screw
Magnet Assembly
Lamp Viewing Hole
Figure 6-1. Detector/Magnet Assembly
Connector J12
Dual Photocell
Source Lamp Assembly
Rosemount Analytical Inc. A Division of Emerson Process Management Maintenance and Service 6-3
Instruction Manual
748213-S April 2002
Model 755R

6-4 DETECTOR CHECK

To isolate the detector as the problem, it is necessary to check the source lamp, photo­cells, and suspension (see Figure Figure 6-1B, page 6-3). These components are con­nected via J12 on the optical bench assembly.
Pin/leads may be removed from connector J12 by use of an improvised pin removal tool, such as a paper clip (see Figure 6-2 below).
Upper Slot
Side View of Connector
Lower Slot
Connect J12 has slots at top and bottom. To
remove a connector pin/lead, insert the tool
into the upper or lower slot and push down on
the end to release the keeper on the pin.
When inserting a pin/lead, its keeper must
face toward the slot opening in the connector
in order to lock in. If inserted otherwise, the
pin/lead will be forced out when the two con-
nectors are joined.
Keeper
Connector Pin Removed
Connector Pin/ Leads in Place
Improvised Pin Removal Tool, Such as a Paper Clip
Figure 6-2. Pin/Lead Removal
When dual photocell is installed, the gap between the two photocells should be in position indicated by this line.
Figure 6-3. Detector Optical Bench
10
J12
Optical Bench
18
WHT WHT
BLK BLK
PUR GRN
1
Hole for Source Lamp
RT1
HR2
Suspension
Heater
Suspension Terminals
6-4 Maintenance and Service Rosemount Analytical Inc. A Division of Emerson Process Management
Model 755R
Instruction Manual
748213-S
April 2002

a. Source Lamp

The simplest check of the source lamp is to verify that it is lit. Another check is done by removing the housing cover and viewing the lamp through the photocell alignment hole (see Figure 6-3, page 6-4). If the photocell is not illuminated, test the voltage across TP2 to TP5 (ground). This voltage should be 2.2 V ±0.2 VDC. If reading is correct, the lamp may be burned out; also inspect the cable for continuity. If voltage reading is not 2.2 V ±0.2 VDC, the Power Supply Board must be replaced.

b. Photocell

If the photocells are on, observe through the photocell alignment hole. The image should be steady. Disconnect the line power and observe the image when you reconnect the power. It should come up from the side and seek a position that equally illuminates the photocells.

c. Suspension

Turn electrical power to instrument OFF. Remove optical bench assembly (see Figure 6-1A, page 6-3). With 100% nitro­gen flowing through the analyzer, note position of the suspension. Then admit air and note response of the suspension. It should rotate clockwise as viewed from the top, and to the right as viewed though the window. Failure to rotate indicates that the suspension has been damaged and detector assembly must be replaced. See Section 6-5c, page 6-7.
If the suspension has been changed, the cause may be improper operating condi­tions.
6-5 REPLACEMENT OF DETECTOR/MAGNET
COMPONENTS

a. Source Lamp

Removal/Installation
The source lamp is held in the optical bench assembly by a set screw (see Fig­ure 6-1B, page 6-3). The two lamp leads are connected to J12.
The red line on the lamp base must align with the set screw (see Figure 6-4A). The base of the lamp should extend from the hole approximately 1/4 inch. Tighten set screw when lamp is aligned.
Realign the photocell per Section 6-5b, page 6-5.

b. Photocell

Removal/Installation
Refer to Figure 6-1B, page 6-3. Note lo­cation of photocell leads in connector J12. Remove leads. Remove photocell lock screws (2), slide photocell out.
Reverse the removal procedure for in­stallation. Align photocell (see below).
Alignment
The adjustments in this procedure are made on the Control Board. With zero gas flowing:
1. Place a digital voltmeter between the wiper of zero potentiometer (R13) and TP7 (ground). Adjust for 0 VDC.
2. Remove the voltmeter from R13 and place on R10 (see Figure 6-4B, page 6-6). Adjust R9 for 0 VDC.
3. Remove the voltmeter from R10 and place on TP8. Move the photocell to obtain a DC voltage as close to 0 mV as possible, but no more than ±750 mV.
Rosemount Analytical Inc. A Division of Emerson Process Management Maintenance and Service 6-5
Instruction Manual
Alig
748213-S April 2002
Model 755R
4. Apply power to instrument and allow to warm-up approximately one hour.
5. Set front panel ZERO at mid-range (i.e., five turns from either end).
6. Connect digital voltmeter from slider of R9 to chassis ground. With a steady flow of 50 to 500 cc/min of ni­trogen zero gas going through instru­ment, adjust R9 for 0 V.
DETAIL A
Set Screw
7. Connect the voltmeter between TP10 and circuit ground (TP7). Adjust front panel ZERO for reading of exactly zero on voltmeter.
All internal adjustments are now properly set. The instrument may be calibrated per Section 3-4, page 3-1.
1/4"
Red Mark for
nment
DETAIL B
C8
Voltmeter Lead
C1
R10
R8
R7
Figure 6-4. Lamp Replacement
6-6 Maintenance and Service Rosemount Analytical Inc. A Division of Emerson Process Management
Model 755R
Instruction Manual
748213-S
April 2002

c. Detector

Removal
Prior to removal of the detector, remove power from instrument and stop flow of sample gas.
1. Remove the four screws securing the detector cover plate.
2. Disconnect cable from J12 on the detector assembly.
NOTE:
Note how the rubber sample lines are looped into a "long coil". When rein­stalling the sample lines they must be configured in the same way. This pre­caution isolates the detector from the effects of mechanical vibration. Oth­erwise vibration waves could travel upward along the tubing walls, result­ing in noisy readout.
3. Refer to Figure 6-1, page 6-3. Using needle-nose pliers, squeeze the hose clamps to disconnect the rubber sam­ple lines from the metal inlet and out­let tubes of the detector assembly.
4. Remove the two screws at the bottom of the detector assembly, slide de­tector out.
Installation
1. Install replacement detector assembly and connect cable to J12.
2. Seat the detector assembly firmly against the magnet pole pieces and tighten attaching screws.
Calibration
1. On the Control Board, set the front panel ZERO control (R13) at mid­range (i.e., five turns from either end).
2. Connect a digital voltmeter from the slider of R9 to chassis ground. With a steady flow of 50 to 500 cc/min. of ni­trogen gas passing through the in­strument, adjust R9 for zero volts.
3. Connect the voltmeter between TP10 and circuit ground (TP7). Adjust front panel ZERO control (R13) for reading of exactly zero on voltmeter.
4. With all internal adjustments now properly set, the instrument may be calibrated per Section 3-4, page 3-1.

6-6 CONTROL BOARD SETUP

a. Power Supply Test

1. TP7 (circuit ground) is ground point for all voltage tests.
2. Counterclockwise end of front panel ZERO potentiometer (R13 on Control Board): -15 VDC ±0.5 VDC.
3. Clockwise end of ZERO potentiome­ter: +15 VDC ±5 VDC.
4. Set ZERO potentiometer to obtain a reading of .0 VDC ±10 mV at slider.
5. Measure TP19: +5 VDC ±0.25 VDC.
3. Reconnect rubber sample lines to metal inlet and outlet tubes on detec­tor assembly.
4. Apply power to instrument and allow to warm up approximately one hour.
Rosemount Analytical Inc. A Division of Emerson Process Management Maintenance and Service 6-7

b. Detector zero

1. Flow 250 cc/min nitrogen.
2. Monitor TP8, adjust R9 for 0 VDC ±2mV.
Instruction Manual
748213-S April 2002
Model 755R

c. U4 Zero

1. Monitor TP5, adjust R100 for 1 VDC ±2mV.
2. Monitor TP10, adjust R13 (ZERO) for
0.0 VDC ±5mV.

d. U8 Zero

Monitor TP11, adjust R29 for 0.0 VDC ±5mV.

e. U10 Zero

1. Monitor TP16, adjust R29 for 0.0 VDC ±5mV.
NOTE:
This adjustment requires a "long time" constant. Allow adequate time.
2. Adjust R29 to obtain a reading of
00.00 ±5 counts on the display.

f. Fullscale

1. Monitor TP11. Flow 100% oxygen, adjust SPAN potentiometer R20 for
10.00 V ±5mV.
2. Monitor TP16, adjust R20 for 10.00 V
±5mV. The display must read 100.00 ±5 counts.

g. Recorder Fullscale

1. Flow nitrogen at 250 cc/min, monitor TP16, and adjust front panel ZERO potentiometer for .000 VDC.
2. Flow 100% oxygen for span gas. Re­corder output for 1 V, 100 mV, or 10 mV should read 100% of span gas. Adjust R88 if necessary.
6-8 Maintenance and Service Rosemount Analytical Inc. A Division of Emerson Process Management
Model 755R
Instruction Manual
748213-S
April 2002
SECTION 7

REPLACEMENT PARTS

The following parts are recommended for rou­tine maintenance and troubleshooting of the Model 755R Oxygen Analyzer. If the trouble­shooting procedures do not resolve the prob­lem, contact Rosemount Analytical Customer Service Center.
WARNING
PARTS INTEGRITY
Tampering or unauthorized substitution of components may adversely affect safety of this product. Use only factory-documented components for repair.
7-1 CIRCUIT BOARD REPLACEMENT POLICY
In most situations involving a malfunction of a circuit board, it is more practical to replace the board than to attempt isolation and replace­ment of the individual component. The cost of test and replacement will exceed the cost of a rebuilt assembly. As standard policy, rebuilt boards are available on an exchange basis.
Because of the exchange policy covering cir­cuit boards the following list does not include individual electronic components. If circum­stances necessitate replacement of an indi­vidual component, which can be identified by inspection or from the schematic diagrams, obtain the replacement component from a lo­cal source of supply.
Rosemount Analytical Inc. A Division of Emerson Process Management Replacement Parts 7-1
Instruction Manual
748213-S April 2002
7-2 MATRIX – MODEL 755R OXYGEN ANALYZER
755R Model 755R Oxygen Analyzer
Code Ranges
01 0-5, 10, 25, 50, and 100% O2 (Standard) 02 0-1, 2.5, 5, 10, 25, 50, and 100% O2 (Extended) 99 Special
Code
Output
01 0-10 mV, 0-100 mV, 0-1 V or 0-5 VDC (Standard) 02 0, 4-20 mA (Current) 99 Special
Code Alarm Relays
00 No Alarm 01 Dual Alarm 99 Special
Model 755R
Code Case
01 Standard 02 Standard with Tropicalization 03 EMC Kit 04 EMC Kit with Tropicalization 99 Special
Code Operation
115 VAC, 50/60 Hz (Standard)
01 02 230 VAC, 50/60 Hz 99 Special
Code Remote Range
00 None
01 Standard ( 0-5, 10, 25, 50, 100%)
02 Extended ( 0-1, 2.5, 5, 10, 25%)
03 0-1, 2.5, 5, 10, 50%
04 0-1, 2.5, 5, 25, 50%
0-1, 2.5, 5, 25, 100%
05
0-1, 2.5, 10, 25, 100%
06
Special
99
Code Feature
Features as selected
00
Special
99
(1)
(1)
(1)
(1)
(1)
755R 01 01 01 01 01 00 00 Example
7-2 Replacement Parts Rosemount Analytical Inc. A Division of Emerson Process Management
Model 755R

7-3 SELECTED REPLACEMENT PARTS

092114 Fuse, 1/2A (240VAC) (Package of 5)
777362 Fuse, Heater 3A (120VAC) (Package of 15)
777361 Fuse, Heater 1.5A (240VAC) (Package of 15)
861649 Thermal Fuse (F2,F3)
656189 Detector/Optical Bench Assembly (0 to 1%)
616418 Source Lamp Kit
622356 Photocell
621023 Current Output Board (0 to 20mA, 4 to 20mA)
622351 Connector Board
631773 Power Supply Board
652830 Control Board Kit
654004 Thermistor - Case Heater
654022 Display Assembly
654078 Viton Tubing (Sample In)
654079 Viton Tubing (Sample Out)
654080 Fan Assembly
654081 Case Heater
809374 Fuse, 3/4A (Power Transformer, 115VAC)
860371 Alarm Relay
645407 Shock Mount (Package of 4)
Instruction Manual
748213-S
April 2002
Rosemount Analytical Inc. A Division of Emerson Process Management Replacement Parts 7-3
Instruction Manual
748213-S April 2002
Model 755R
7-4 Replacement Parts Rosemount Analytical Inc. A Division of Emerson Process Management
Model 755R
Instruction Manual
748213-S
April 2002
SECTION 8

RETURN OF MATERIAL

8-1 RETURN OF MATERIAL

If factory repair of defective equipment is required, proceed as follows:
1. Secure a return authorization from a Rosemount Analytical Inc. Sales Office or Representative before returning the equipment. Equipment must be returned with complete identification in accordance with Rosemount instructions or it will not be accepted.
Rosemount CSC will provide the shipping address for your instrument.
In no event will Rosemount be responsible for equipment returned without proper authorization and identification.
2. Carefully pack the defective unit in a sturdy box with sufficient shock absorbing material to ensure no additional damage occurs during shipping.
3. In a cover letter, describe completely:
The symptoms that determined the
equipment is faulty. The environment in which the
equipment was operating (housing, weather, vibration, dust, etc.). Site from where the equipment was
removed. Whether warranty or non-warranty
service is expected. Complete shipping instructions for the
return of the equipment.
4. Enclose a cover letter and purchase order and ship the defective equipment according to instructions provided in the Rosemount Return Authorization, prepaid, to the address provided by Rosemount CSC.
Rosemount Analytical Inc.
Process Analytical Division
Customer Service Center
1-800-433-6076
If warranty service is expected, the defective unit will be carefully inspected and tested at the factory. If the failure was due to the conditions listed in the standard Rosemount warranty, the defective unit will be repaired or replaced at Rosemount’s option, and an operating unit will be returned to the customer in accordance with the shipping instructions furnished in the cover letter.
For equipment no longer under warranty, the equipment will be repaired at the factory and returned as directed by the purchase order and shipping instructions.

8-2 CUSTOMER SERVICE

For order administration, replacement Parts, application assistance, on-site or factory repair, service or maintenance contract information, contact:
Rosemount Analytical Inc.
Process Analytical Division
Customer Service Center
1-800-433-6076

8-3 TRAINING

A comprehensive Factory Training Program of operator and service classes is available. For a copy of the
Training Schedule
Services Department at:
Rosemount Analytical Inc.
Customer Service Center
Current Operator and Service
contact the Technical
1-800-433-6076
Rosemount Analytical Inc. A Division of Emerson Process Management Return of Material 8-1
Instruction Manual
748213-S April 2002
Model 755R
8-2 Return of Material Rosemount Analytical Inc. A Division of Emerson Process Management

WARRANTY

Goods and part(s) (excluding consumables) manufactured by Seller are warranted to be free from defects in workmanship and material under normal use and service for a period of twelve (12) months from the date of shipment by Seller. Consumables, glass electrodes, membranes, liquid junctions, electrolyte, o-rings, etc., are warranted to be free from defects in workmanship and material under normal use and service for a period of ninety (90) days from date of shipment by Seller. Goods, part(s) and consumables proven by Seller to be defective in workmanship and/or material shall be replaced or repaired, free of charge, F.O.B. Seller's factory provided that the goods, part(s) or consumables are returned to Seller's designated factory, transportation charges prepaid, within the twelve (12) month period of warranty in the case of goods and part(s), and in the case of consumables, within the ninety (90) day period of warranty. This warranty shall be in effect for replacement or repaired goods, part(s) and the remaining portion of the ninety (90) day warranty in the case of consumables. A defect in goods, part(s) and consumables of the commercial unit shall not operate to condemn such commercial unit when such goods, part(s) and consumables are capable of being renewed, repaired or replaced.
The Seller shall not be liable to the Buyer, or to any other person, for the loss or damage directly or indirectly, arising from the use of the equipment or goods, from breach of any warranty, or from any other cause. All other warranties, expressed or implied are hereby excluded.
IN CONSIDERATION OF THE HEREIN STATED PURCHASE PRICE OF THE GOODS, SELLER GRANTS ONLY THE ABOVE STATED EXPRESS WARRANTY. NO OTHER WARRANTIES ARE GRANTED INCLUDING, BUT NOT LIMITED TO, EXPRESS AND IMPLIED WARRANTIES OR MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.
Limitations of Remedy. SELLER SHALL NOT BE LIABLE FOR DAMAGES CAUSED BY DELAY IN PERFORMANCE. THE SOLE AND EXCLUSIVE REMEDY FOR BREACH OF WARRANTY SHALL BE LIMITED TO REPAIR OR REPLACEMENT UNDER THE STANDARD WARRANTY CLAUSE. IN NO CASE, REGARDLESS OF THE FORM OF THE CAUSE OF ACTION, SHALL SELLER'S LIABILITY EXCEED THE PRICE TO BUYER OF THE SPECIFIC GOODS MANUFACTURED BY SELLER GIVING RISE TO THE CAUSE OF ACTION. BUYER AGREES THAT IN NO EVENT SHALL SELLER'S LIABILITY EXTEND TO INCLUDE INCIDENTAL OR CONSEQUENTIAL DAMAGES. CONSEQUENTIAL DAMAGES SHALL INCLUDE, BUT ARE NOT LIMITED TO, LOSS OF ANTICIPATED PROFITS, LOSS OF USE, LOSS OF REVENUE, COST OF CAPITAL AND DAMAGE OR LOSS OF OTHER PROPERTY OR EQUIPMENT. IN NO EVENT SHALL SELLER BE OBLIGATED TO INDEMNIFY BUYER IN ANY MANNER NOR SHALL SELLER BE LIABLE FOR PROPERTY DAMAGE AND/OR THIRD PARTY CLAIMS COVERED BY UMBRELLA INSURANCE AND/OR INDEMNITY COVERAGE PROVIDED TO BUYER, ITS ASSIGNS, AND EACH SUCCESSOR INTEREST TO THE GOODS PROVIDED HEREUNDER.
Force Majeure. Seller shall not be liable for failure to perform due to labor strikes or acts beyond Seller's direct control.
Instruction Manual
748213-S April 2002
Model 755R
Emerson Process Management
Rosemount Analytical Inc. Process Analytic Division
1201 N. Main St. Orrville, OH 44667-0901 T (330) 682-9010 F (330) 684-4434 E gas.csc@emersonprocess.com
ASIA - PACIFIC Fisher-Rosemount Singapore Private Ltd.
1 Pandan Crescent Singapore 128461 Republic of Singapore T 65-777-8211 F 65-777-0947
http://www.processanalytic.com
© Rosemount Analytical Inc. 2001
Fisher-Rosemount GmbH & Co.
Industriestrasse 1 63594 Hasselroth Germany T 49-6055-884 0 F 49-6055-884209
EUROPE, MIDDLE EAST, AFRICA Fisher-Rosemount Ltd.
Heath Place Bognor Regis West Sussex PO22 9SH England T 44-1243-863121 F 44-1243-845354
LATIN AMERICA Fisher - Rosemount
Av. das Americas 3333 sala 1004 Rio de Janeiro, RJ Brazil 22631-003 T 55-21-2431-1882
Loading...