This guide provides basic guidelines for Rosemount Smart Pressure Gauges. It does not provide instructions for configuration, diagnostics, maintenance, service, troubleshooting, or intrinsically safe (I.S.) installations. Refer to the Rosemount Smart Pressure Gauge Reference Manual for more instruction. The manual and this guide are also available electronically.
NOTICE
Shipping considerations
The device is shipped with the battery installed. Each device contains one “D” size primary lithium-thionyl chloride battery. Primary lithium
batteries are regulated in transportation by the U.S. Department of Transportation, and are also covered by IATA (International Air Transport Association), ICAO (International Civil Aviation Organization), and ARD (European Ground Transportation of Dangerous Goods). It is the responsibility of the shipper to ensure compliance with these or any other local requirements. Consult current regulations and requirements before shipping.
WARNING
Explosions could result in death or serious injury.
Installation of device in an explosive environment must be in accordance with appropriate local, national, and international standards, codes, and practices.
Ensure device is installed in accordance with intrinsically safe or non-incendive field practices.
Electrical shock could cause death or serious injury.
Device must be installed to ensure a minimum antenna separation distance of 8 in. (20 cm) from all persons.
Care must be taken during transportation of device to prevent electrostatic charge build-up.
Process leaks could result in death or serious injury.
Handle the device carefully.
Failure to follow these installation guidelines could result in death or serious injury.
Ensure only qualified personnel perform the installation.
Contents
Required equipment...................................... 3
What’s in the box........................................... 4
Optional: Power/device check........................6
Optional: Normal range indication option...... 7
2 Emerson.com/Rosemount
Installation Procedure.................................... 8
Troubleshooting.......................................... 12
Product Certifications.................................. 13
February 2019 Quick Start Guide

1 Required equipment

Figure 1-1: Anti-seize paste or PTFE tape (for NPT threaded connection)
Figure 1-2: Standard tools (e.g., screwdriver, wrench, pliers)
Quick Start Guide 3
Quick Start Guide February 2019

2 What’s in the box

Smart Pressure Gauge Quick Start Guide
The following options are also available and will ship with the Rosemount Smart Pressure Gauge if ordered.
Rosemount 306 Integral Manifold (model code S5)
4 Emerson.com/Rosemount
B4 bracket (model code B4)
February 2019 Quick Start Guide
Rosemount 1199 Seal(model code S1) Product Certification
Q4: Calibration Cert QG: Calibration Cert and GOST
Verification Cert QP: Calibration Cert and Tamper Evident
Seal Q8: Material Traceability cert per EN
102043.1 Q15: Cert of Compliance to NACE
MR0175/ISO 15156 for wetted materials Q25: Cert of Compliance to NACE
MR00103 for wetted materials
Normal range Indication (model code LK) Product Certification model codes
Quick Start Guide 5
Quick Start Guide February 2019

3 Optional: Power/device check

The device is designed to be installation-ready.
Procedure
1. To check device battery prior to installation, perform Turn on device
2. Slide the ON/OFF switch to the OFF position until ready for use.
6 Emerson.com/Rosemount
February 2019 Quick Start Guide

4 Optional: Normal range indication option

Important
The stickers are intended to be installed on the dial only and should not be
applied on the inside or outside of the housing cover.
Stickers should be applied in an environment where the ambient
temperature is above 50°F (10°C)
Procedure
1. Modify each of the stickers to desired size.
2. Remove housing cover.
3. Slide ON/OFF switch to the OFF position and wait for the LED to stop flashing.
4. Gently move the needle in the clockwise direction until it is pointing at the red X.
NOTICE
Equipment Damage
Use caution as the electronics assembly is connected to the needle.
5. Remove any debris on the dial, so it does not become trapped under the sticker.
6. Peel back the white paper backing of the sticker.
7. Slowly lower the sticker onto the surface of the dial in the desired location and rub it in place firmly. Repeat steps Step 6 and Step 7 until desired indication locations are set.
Moving the sticker after initial contact is not recommended as this decreases the amount of adhesive on the back of the sticker.
8. Slide ON/OFF switch to the ON position.
9. Replace housing cover.
Quick Start Guide 7
OR
Quick Start Guide February 2019

5 Installation Procedure

Procedure
1. Seal and protect threads
2. Mount device
Note
Use wrench on flats, not on housing.
5.1

Mounting orientation

The low side pressure port (atmospheric reference) is located in the neck of the device behind the housing.
The vent path is between the housing and sensor.
CAUTION
Keep the vent path free of any obstruction, including but not limited to paint, dust, and lubrication by mounting the device so the process can drain away.
8 Emerson.com/Rosemount
A
February 2019 Quick Start Guide
Figure 5-1: Low Side Pressure Port
A. Low side pressure port (atmospheric reference)

5.2 Turn on device

Check to ensure the device and battery are working properly.
Procedure
1. Twist the cover counterclockwise to remove it.
2. Slide the OFF/ON switch to the ON position to initiate the power sequence.
Note
During the power sequence, the dial tests full range of motion and LED flashes amber.
3. Once the power sequence ends, verify that the LED flashes green. The LED may display several colors; see Device status indicator .
Quick Start Guide 9
1 2 3
64 5
8 709
A
B
C
Quick Start Guide February 2019

5.3 Connect to the device

Image displaying the device connection diagram
A. Field Communicator B. HART® Modem C. AMS Configurator

5.3.1 AMS Configurator

Procedure to connect the AMS Configurator to the smart pressure gauge.
Procedure
1. Start AMS Configurator.
2. From the View menu, select Device Connection View.
3. Double-click the device under the HART modem.

5.3.2 Field communicator

Procedure to connect the smart pressure gauge to the field communicator.
Procedure
1. Turn on the Field Communicator.
2. From the Main menu, select the HART symbol.
5.4

Eliminate mounting effects

Devices are factory-calibrated. Once installed, it is recommended to perform this step to eliminate potential error caused by mounting position or static pressure.
10 Emerson.com/Rosemount
February 2019 Quick Start Guide
Note
See the Rosemount Smart Pressure Gauge Reference Manual for the following:
Using AMS Configurator
Sensor trim function on absolute gauge
Procedure
1. Vent the device.
2. Connect the Field Communicator.
3. From the HOME screen, enter the HART Fast Key sequence.
Device dashboard Fast Keys 2, 1, 1
4. Follow the commands to perform the procedure.
Quick Start Guide 11
Quick Start Guide February 2019

6 Troubleshooting

6.1 Device status indicator

Table 6-1: Status Descriptions
LED color Device status
Green Functioning properly
Amber Battery is low, battery replacement
Red Battery replacement required OR
No color No power, verify ON/OFF switch is in

6.2 Pressure measurement

If the mounting effects have not been eliminated after completing, perform this alternative procedure for verifying the pressure value.
Procedure
1. From the HOME screen, enter the HART Fast Key sequence.
recommended
Device is malfunctioning
the on position
Device dashboard Fast Keys
2, 2, 1, 1, 1
2. Follow the commands to perform the procedure.
12 Emerson.com/Rosemount
February 2019 Quick Start Guide

7 Product Certifications

Rev: 2.0
European Directive Information
A copy of the EU Declaration of Conformity can be found at the end of the Quick Start Guide. The most recent revision of the EU Declaration of Conformity can be found at Emerson.com/Rosemount.
Ordinary Location Certification from CSA
The product has been examined and tested to determine that the design meets the basic electrical, mechanical, and fire protection requirements by CSA, a nationally recognized test laboratory (NRTL) as accredited by the Federal Occupational Safety and Health Administration (OSHA).
Installing in North America
The US National Electrical Code (NEC) and the Canadian Electrical Code (CEC) permit the use of Division marked equipment in Zones and Zone marked equipment in Divisions.The markings must be suitable for the area classification, gas, and temperature class. This information is clearly defined in the respective codes.
USA
I5 U.S.A. Intrinsically Safe (IS)
Certificate:
Standards:
Markings:
Special Conditions for Safe Use(X):
1. Do not replace battery when explosive atmosphere is present.
2. Use only 00G45-9000-0001 batteries.
3.
4. Substitution of components may impair intrinsic safety.
Quick Start Guide 13
[CSA] 70047656 FM 3600 – 2011, FM 3610 – 2010, UL Standard 50 – Eleventh
Edition, UL 61010-1 – 3rd Edition, ANSI/ISA-60079-0 (12.00.01) – 2013, ANSI/ISA-60079-11 (12.02.01) – 2013, ANSI/IEC 60529 – 2004
IS CL I, DIV 1, GP A, B, C, D T4; Class 1, Zone 0, AEx ia IIC T4 Ga; T4 (–40 °C ≤ Ta ≤ +70 °C) when installed per Rosemount drawing 00G45-1020; Type 4X; IP66/67;
The surface resistivity of the housing is greater than 1G . To avoid electrostatic charge build-up, it must not be rubbed or cleaned with solvents or a dry cloth.
Quick Start Guide February 2019
Canada
I6 Canada Intrinsically Safe (IS)
Certificate:
Standards:
[CSA]70047656 CAN/CSA C22.2 No. 0-10, CAN/CSA C22.2 No. 94-M1991
(R2011), CAN/CSA-60079-0-11, CAN/CSA-60079-11-14, CSA Std C22.2 No. 60529-05, CAN/CSA-C22.2 No. 61010-1-12
Markings:
Intrinsically Safe for Class I, Division 1, Groups A, B, C, D T4; Ex ia IIC T4 Ga T4 (–50 °C ≤ Ta ≤ +70 °C) when installed per Rosemount drawing 00G45-1020; Type 4X; IP66/67;
Special Conditions for Safe Use (X):
1. Do not replace battery when explosive atmosphere is present. Ne pas remplacer les accumulateurs si une atmosphère explosive peut être présente.
2. Use only 00G45-9000-0001 batteries. Utiliser uniquement des accumulateurs 00G45-9000-0001.
3.
The surface resistivity of the housing is greater than 1G
. To avoid electrostatic charge build-up, it must not be rubbed or cleaned with solvents or a dry cloth. La résistivité de surface du boÎtier est supérieure à un gigaohm. Pour éviter l’accumulation de charge électrostatique, ne pas frotter ou nettoyer avec des produits solvants ou un chiffon sec.
4. Substitution of components may impair intrinsic safety. La substitution de composants peut compromettre la sécurité intrinsèque.
Europe
I1
Certificate:
Standards:
Markings:
ATEX Intrinsic Safety
Baseefa16ATEX0005X EN 60079-0: 2012 + A11: 2013, EN 60079-11: 2012
II 1 G Ex ia IIC T4 Ga, T4 (–40 °C ≤ Ta ≤ +70 °C) IP66/67;
Special Conditions for Safe Use (X):
1. The plastic enclosure may constitute a potential electrostatic ignition risk and must not be rubbed or cleaned with a dry cloth.
2. The measured capacitance between the equipment enclosure and metallic inline sensor module is 4.7pF. This must be considered only when the SPG is integrated into a system where the process connection is not grounded.
14 Emerson.com/Rosemount
February 2019 Quick Start Guide
3. Do not change the battery when an explosive atmosphere is present.
4. Only replace battery with Rosemount Part No. 00G45-9000-0001.
International
I7 IECEx Intrinsic Safety
Certificate:
Standards:
Markings:
IECEx BAS 16.0012X IEC 60079-0: 2011, IEC 60079-11: 2011 Ex ia IIC T4 Ga, T4 (–40 °C ≤ Ta ≤ +70 °C) IP66/67;
Special Conditions for Safe Use (X):
1. The plastic enclosure may constitute a potential electrostatic ignition risk and must not be rubbed or cleaned with a dry cloth.
2. The measured capacitance between the equipment enclosure and metallic inline sensor module is 4.7pF. This must be considered only when the SPG is integrated into a system where the process connection is not grounded.
3. Do not change the battery when an explosive atmosphere is present.
4. Only replace battery with Rosemount Part No. 00G45-9000-0001.
Quick Start Guide 15
EU Declaration of Conformity
No: RMD 1108 Rev. E
Page 1 of 3
We,
Rosemount Inc. 8200 Market Boulevard Chanhassen, MN 55317-9685 USA
declare under our sole responsibility that the product,
Models WPG & SPG:
Wireless Pressure Gauge & Smart Pressure Gauge
manufactured by,
Rosemount Inc. 8200 Market Boulevard Chanhassen, MN 55317-9685 USA
to which this declaration relates, is in conformity with the provisions of the European Community Directives, including the latest amendments, as shown in the attached schedule.
Assumption of conformity is based on the application of the harmonized standards and, when applicable or required, a European Community notified body certification, as shown in the attached schedule.
(signature)
Vice President of Global Quality
(function name - printed)
Chris LaPoint
(name - printed)
1-Feb-19
(date of issue)
Quick Start Guide February 2019
Figure 7-1: Declaration of Conformity
16 Emerson.com/Rosemount
EU Declaration of Conformity
No: RMD 1108 Rev. E
Page 2 of 3
EMC Directive (2014/30/EU)
Models WPG & SPG
Harmonized Standards:
EN 61326-1: 2013
Radio Equipment Directive (RED) (2014/53/EU)
Model WPG (Wireless Pressure Gauge only)
Harmonized Standards:
EN 300 328 V2.1.1 EN 301 489-1 V2.2.0 EN 301 489-17: V3.2.0 EN 61010-1: 2010 EN 62479: 2010
ATEX Directive (2014/34/EU)
Models WPG & SPG
Baseefa16ATEX0005X - Intrinsic Safety Certificate
Equipment Group II Category 1 G
Ex ia IIC T4 Ga, T4(-40°C ≤ Ta ≤ +70°C)
Harmonized Standards:
EN 60079-0: 2012 + A11: 2013 EN 60079-11: 2012
February 2019 Quick Start Guide
Quick Start Guide 17
EU Declaration of Conformity
No: RMD 1108 Rev. E
Page 3 of 3
ATEX Notified Bodies
SGS FIMCO OY [Notified Body Number: 0598]
P.O. Box 30 (Särkiniementie 3) 00211 HELSINKI Finland
ATEX Notified Body for Quality Assurance
SGS FIMCO OY [Notified Body Number: 0598]
P.O. Box 30 (Särkiniementie 3) 00211 HELSINKI Finland
Quick Start Guide February 2019
18 Emerson.com/Rosemount
ᴹ
China RoHS
㇑᧗⢙䍘䎵䗷ᴰབྷ⎃ᓖ䲀٬Ⲵ䜘Ԧරࡇ㺘
Rosemount SPG
List of Rosemount SPG Parts with China RoHS Concentration above MCVs
䜘Ԧ〠
Part Name
ᴹᇣ⢙䍘
/ Hazardous Substances
䫵
Lead
(Pb)
⊎
Mercury
(Hg)
䭹
Cadmium
(Cd)
ޝԧ䬜
Hexavalent
Chromium
(Cr +6)
ཊⓤ㚄㤟
Polybrominated
biphenyls
(PBB)
ཊⓤ㚄㤟䟊
Polybrominated diphenyl ethers
(PBDE)
⭥ᆀ㓴Ԧ
Electronics
Assembly
X O O O O O
༣փ㓴Ԧ
Housing
Assembly
O O O O O O
Րᝏಘ㓴Ԧ
Sensor
Assembly
X O O O O O
⭥⊐㓴Ԧ
Battery
Assembly
X O O O O O
ᵜ㺘Ṭ㌫׍ᦞ
SJ/T11364
Ⲵ㿴ᇊ㘼ࡦ֌
This table is proposed in accordance with the provision of SJ/T11364.
O:
᜿Ѫ䈕䜘ԦⲴᡰᴹ൷䍘ᶀᯉѝ䈕ᴹᇣ⢙䍘Ⲵ䟿൷վҾ
GB/T 26572
ᡰ㿴ᇊⲴ䲀䟿㾱≲
O: Indicate that said hazardous substance in all of the homogeneous materials for this part is below the limit requirement of GB/T 26572.
X:
᜿Ѫ൘䈕䜘Ԧᡰ֯⭘Ⲵᡰᴹ൷䍘ᶀᯉ䟼ˈ㠣ቁᴹа㊫൷䍘ᶀᯉѝ䈕ᴹᇣ⢙䍘Ⲵ䟿儈Ҿ
GB/T 26572
ᡰ㿴ᇊⲴ䲀䟿㾱≲
X: Indicate that said hazardous substance contained in at least one of the homogeneous materials used for this part is above the limit requirement of GB/T 26572.
February 2019 Quick Start Guide
Quick Start Guide 19
*00825-0100-4145*
00825-0100-4145, Rev. BB
Quick Start Guide
February 2019
Global Headquarters
Emerson Automation Solutions 6021 Innovation Blvd. Shakopee, MN 55379, USA
+1 800 999 9307 or +1 952 906 8888 +1 952 949 7001 RFQ.RMD-RCC@Emerson.com
Latin America Regional Office
Emerson Automation Solutions 1300 Concord Terrace, Suite 400 Sunrise, FL 33323, USA
+1 954 846 5030 +1 954 846 5121 RFQ.RMD-RCC@Emerson.com
Asia Pacific Regional Office
Emerson Automation Solutions 1 Pandan Crescent Singapore 128461
+65 6777 8211 +65 6777 0947 Enquiries@AP.Emerson.com
North America Regional Office
Emerson Automation Solutions 8200 Market Blvd. Chanhassen, MN 55317, USA
+1 800 999 9307 or +1 952 906 8888 +1 952 949 7001 RMT-NA.RCCRF@Emerson.com
Europe Regional Office
Emerson Automation Solutions Europe GmbH Neuhofstrasse 19a P.O. Box 1046 CH 6340 Baar Switzerland
+41 (0) 41 768 6111 +41 (0) 41 768 6300 RFQ.RMD-RCC@Emerson.com
Middle East and Africa Regional Office
Emerson Automation Solutions Emerson FZE P.O. Box 17033 Jebel Ali Free Zone - South 2 Dubai, United Arab Emirates
+971 4 8118100 +971 4 8865465 RFQ.RMTMEA@Emerson.com
Linkedin.com/company/Emerson-
Automation-Solutions
Twitter.com/Rosemount_News Facebook.com/Rosemount Youtube.com/user/
RosemountMeasurement
Google.com/+RosemountMeasurement
©
2019 Emerson. All rights reserved.
Emerson Terms and Conditions of Sale are available upon request. The Emerson logo is a trademark and service mark of Emerson Electric Co. Rosemount is mark of one of the Emerson family of companies. All other marks are the property of their respective owners.
Loading...