Emerson Rosemount FCL Quick Start Guide

Page 1
Quick Start Guide
00825-0100-3412, Rev AA
Rosemount™ FCL
Free Chlorine System with Rosemount 1056 Transmitter
June 2019
Page 2
Quick Start Guide June 2019
Read this page before proceeding!
Emerson designs, manufactures, and tests its products to meet many national and international standards. Because these instruments are sophisticated technical products, you must properly install, use, and maintain them to ensure they continue to operate within their normal specifications. The following instructions must be adhered to and integrated into your safety program when installing, using, and maintaining Emerson products. Failure to follow the proper instructions may cause any one of the following situations to occur: loss of life, personal injury, property damage, damage to this instrument, and warranty invalidation.
Read all instructions prior to installing, operating, and servicing the product.
If this Reference Manual is not the correct one, call 1-800-999-9307 to request the correct
Reference Manual. Save this Reference Manual for future reference.
If you do not understand any of the instructions, contact your Emerson representative for
clarification.
Follow all warnings, cautions, and instructions marked on and supplied with the product.
Inform and educate your personnel in the proper installation, operation, and maintenance of the
product.
Install equipment as specified in the installation instructions of the appropriate Reference Manual
and per applicable local and national codes. Connect all products to the proper electrical and pressure sources.
To ensure proper performance, use qualified personnel to install, operate, update, program, and
maintain the product.
When replacement parts are required, ensure that qualified people use replacement parts
specified by Rosemount. Unauthorized parts and procedures can affect the product's performance, place the safe operation of your process at risk, and may result in fire, electrical hazards, or improper operation.
Ensure that all equipment doors are closed and protective covers are in place, except when
maintenance is being performed by qualified people, to prevent electrical shock and personal injury.
WARNING
Hazardous area installation
Installations near flammable liquids or in hazardous area locations must be carefully evaluated by qualified on site safety personnel. This device is not Intrinisically Safe or Explosion Proof.
To secure and maintain intrinsically safe installation, use an appropriate transmitter/safety barrier/sensor combination. The installation system must be in accordance with the governing approval agency (FM, CSA, or BASEEFA/CENELEC) hazardous are classification requirements. Consult your transmitter Reference Manual for details.
Proper installation, operation, and servicing of this sensor in a hazardous area installation are entirely the operator's responsibility.
2 Emerson.com/Rosemount
Page 3
June 2019 Quick Start Guide
WARNING
Electrical shock
Making cable connections to and servicing this instrument require access to shock hazard level voltages, which can cause death or serious injury.
Equipment protected throughout by double insulation. Be sure to disconnect all hazardous voltages before opening the enclosure. Disconnect relay contacts made to separate power sources before servicing. Electrical installation must be in accordance with the National Electrical Code (ANSI/NFPA-70)
and/or any other national or local codes. Unused cable conduit entries must be securely sealed by non-flammable closures to provide
exposure integrity in compliance with personal safety and environmental protection requirements. Unused conduit openings must be sealed with NEMA 4X or IP65 conduit plugs to maintain the ingress protection rating (IP65).
Safety and performance require that this instrument be connected and properly grounded through a three-wire power source.
Proper use and configuration is the operator's responsibility. No external power to the instrument of more than 69 Vdc or 43 V peak is allowed, with the
exception of power and relay terminals. Any violation will impair the safety protection provided.
Do not operate this instrument without the front cover secured. Refer installation, operation, and servicing to qualified personnel.
WARNING
This product is not intended for use in the light industrial, residential, or commercial environments per the instrument's certification to EN50081-2.
CAUTION
Sensor/process application compatibility
The wetted sensor materials may not be compatible with process composition and operating conditions.
Application compatibility is entirely the operator's responsibility.
WARNING
Physical access
Unauthorized personnel may potentially cause significant damage to and/or misconfiguration of end users’ equipment. This could be intentional or unintentional and needs to be protected against.
Physical security is an important part of any security program and fundamental to protecting your system. Restrict physical access by unauthorized personnel to protect end users’ assets. This is true for all systems used within the facility.
Contents
Specifications...............................................................................................................................5
Quick Start Guide 3
Page 4
Quick Start Guide June 2019
Install........................................................................................................................................... 7
Wire........................................................................................................................................... 12
EU Declaration of Conformity.....................................................................................................24
China RoHS Table....................................................................................................................... 26
4 Emerson.com/Rosemount
Page 5
June 2019 Quick Start Guide

1 Specifications

Rosemount 1056 Transmitter
For Rosemount 1056 Transmitter specifications, see the Rosemount 1056 Transmitter Reference Manual on Emerson.com/Rosemount: Manual: Rosemount 1056 Dual-Input Transmitter.
Table 1-1: General Specifications
Characteristic Specification
Sample requirements Pressure: 3 to 65 psig (122 to 549
kPa abs). A check valve in the inlet prevents the sensor flow cells from going dry if sample flow is lost. The check valve opens at 3 psig (122 kPa abs). If the check valve is removed, minimum pressure is 1 psig (108 kPa abs).
Temperature: 32 to 122 °F (0 to 50
°C)
Minimum flow: 3 gal/hr (11 L/hr)
Maximum flow: 80 gal/hr (303 L/hr);
high flow causes the overflow tube to back up.
Sample conductivity >50 µS/cm at 77 °F (25 °C)
Process connection ¼-in. OD tubing compression fitting (can
Drain connection ¾-in. barbed fitting. Sample must drain
Wetted parts Overflow sampler and flow cell: acrylic,
Response time to step change in chlorine concentration
Quick Start Guide 5
be removed and replaced with barbed fitting for soft tubing)
to open atmosphere.
polycarbonate, Kynar®, nylon, and silicone
Chlorine sensor: Noryl®, Viton®, wood, silicone, polyethersulfone, polyester, and platinum
pH sensor (Rosemount™ 3900VP): Stainless steel, glass, Teflon®, polyphenylene sulfide, EPDM, and silicone
< 80 sec to 95% of final reading for inlet sample flow of 3 gph (11 L/hr)
Page 6
Quick Start Guide June 2019
Table 1-1: General Specifications (continued)
Characteristic Specification
Weight/shipping weight (rounded up to nearest 1 lb. or 0.5 kg)
Rosemount FCL-01: 10 lb./13 lb. (4.5 kg/6.0 kg)
Rosemount FCL-02: 11 lb./14 lb. (5.0 kg/6.5 kg)
Table 1-2: Sensor Specifications
Characteristic Specification
Free chlorine range 0 to 10 ppm as Cl2. For higher ranges,
pH correction range 6.0 to 9.5. For samples having pH
Accuracy Accuracy depends on the accuracy of the
Interferences Monochloramine, permangante, and
Electrolyte volume 25 mL (approx.)
Electrolyte life 3 months (approx.); for best results,
consult the factory.
between 9.5 and 10.0, consult the factory. If pH <6.0, correction is not necessary. For manual pH correction, choose option -01. For continuous pH correction, choose option -02.
chemical test used to calibrate the sensor.
peroxides
replace electrolyte monthly.
6 Emerson.com/Rosemount
Page 7
June 2019 Quick Start Guide

2 Install

2.1 Unpack and inspect

Procedure
1. Inspect the shipping container(s). If there is damage, contact the shipper immediately for instructions.
2. If there is no apparent damage, unpack the container(s).
3. Ensure that all items shown on the packing list are present. If items are missing, notify Emerson immediately.

2.1.1 Rosemount™ FCL-01 (free chlorine without continuous pH correction)

The Rosemount FCL-01 consists of the following items mounted on a back plate.
1. Rosemount 1056-03-24-38-AN transmitter with sensor cable attached.
2. Constant head overflow sampler with flow cell for chlorine sensor.
The free chlorine sensor (Rosemount 499ACL-01-54-VP), three membrane assemblies, and a bottle of electrolyte solution are in a separate package.

2.1.2 Rosemount™ FCL-02 (free chlorine with continuous pH correction)

The Rosemount FCL-02 consists of the following items mounted on a back plate:
1. Rosemount 1056-03-24-32-AN transmitter with sensor cables attached.
2. Constant head overflow sampler with flow cells for pH and chlorine sensors.
3. Stand to hold pH buffer solution during calibration.
The free chlorine sensor (Rosemount 499ACL-01-54-VP), shipped with three membrane assemblies and a bottle of electolyte solution, and the Rosemount 3900VP-02-10 pH sensor are in separate packages.
2.2
Quick Start Guide 7

General installation information

1. Although the system is suitable for outdoor use, do not install it in direct sunlight or in areas of extreme temperature.
Page 8
Quick Start Guide June 2019
CAUTION
Hazardous areas
The system is not suitable for use in hazardous areas.
2. To keep the transmitter enclosure watertight, install plugs (provided) in the unused conduit openings.
3. Install the system in an area where vibrations and electromagnetic and radio frequency interference are minimized or absent.
4. Be sure there is easy access to the transmitter and sensor(s).

2.3 Sample requirements

Be sure the sample meets the following requirements:
1. Temperature: 32 to 122 °F (0 to 50 °C )
2. Pressure: 3 to 65 psig (122 to 549 kPa abs)
3. Minimum flow: 3 gal/hr (11 L/hr)

2.4 Mounting, inlet, and drain connections

The Rosemount™ FCL is intended for wall mounting only.
Refer to Figure 2-1 or Figure 2-2 for details. The sensor(s) screw into the flow cell adapters as shown in the figures. For Rosemount FCL-02 (free chlorine with continuous pH adjustment), you must also install the pH sensor.
8 Emerson.com/Rosemount
Page 9
June 2019 Quick Start Guide
Figure 2-1: Rosemount FCL-01
A. Chlorine sensor B. Inlet C. Drain
Quick Start Guide 9
Page 10
Quick Start Guide June 2019
Figure 2-2: Rosemount FCL-02
A. pH sensor B. Chlorine sensor C. Inlet
D. Drain
A ¼-in. OD tubing compression fitting is provided for the sample inlet. If desired, you can remove the compression fitting and replace it with a barbed fitting. The fitting screws into a ¼-in. FNPT check valve. The check valve prevents the sensor flow cell from going dry if sample flow is lost.
The sample drains through a ¾-in. barbed fitting.
1. Attach a piece of soft tubing to the fitting and allow the waste to drain to open atmosphere.
Important
Do not restrict the drain line.
10 Emerson.com/Rosemount
Page 11
June 2019 Quick Start Guide
2. Adjust the sample flow until the water level is even with the central overflow tube and excess water is flowing down the tube.
3. Confirm that sample is flowing through the flow cells.

2.5 Install the sensor(s)

Emerson provides the Rosemount™ FCL with the sensor cable pre-wired to the transmitter.
Procedure
1. Connect the chlorine sensor (Rosemount 499ACL-01-54-VP) to the cable labeled CL.
2. Connect the pH sensor (Rosemount 3900-VP-02-10) to the cable labeled pH.
The terminal end of the sensor is keyed to ensure proper mating with the cable receptacle.
3. Once the key has slid into the mating slot, tighten the connection by turning the knurled ring clockwise.
4. Screw the sensor(s) into the plastic fitting(s), which are held in the flow cell(s) by the union nut.
Do not remove the protective cap on the sensor(s) until ready to put the sensor(s) in service.
Quick Start Guide 11
Page 12
Quick Start Guide June 2019

3 Wire

3.1 Wire power

Wire AC mains power supply to the power supply board, which is mounted vertically on the left hand side of the transmitter enclosure.
WARNING
Electrical shock
Electrical installation must be in accordance with the National Electrical Code (ANSI/NFPA-70) and/or any other applicable national or local codes.
The power connector is at the top of the board.
Procedure
1. Unplug the connector from the board and wire the power cable to it. Lead connections are marked on the connector. (L is live or hot; N is
neutral; the ground connection has the standard symbol.)
2. Run the power wiring through the conduit opening nearest the power terminal.
AC power wiring should be 14 gauge or greater.
3. Provide a switch or breaker to disconnect the transmitter from the main power supply.
4. Install the switch or breaker near the transmitter and label it as the disconnecting device for the transmitter.
3.2
12 Emerson.com/Rosemount

Wire analog outputs

Two analog output currents are located on the main circuit board, which is attached to the inside of the enclosure door.
Figure 3-1 shows the locations of the terminals. The connectors can be
detached for wiring. TB-1 is output 1. TB-2 is output 2. Polarity is marked on the circuit board.
Page 13
June 2019 Quick Start Guide
Figure 3-1: Analog output connections
The analog outputs are on the main board near the hinged end of the enclosure door.
For best EMI/RFI protection, use shielded output signal cable enclosed in earth-grounded metal conduit.
Keep output signal wiring separate from power wiring. Do not run signal and power or relay wiring in the same conduit or close together in a cable tray.
3.3
Quick Start Guide 13

Alarm wiring

The alarm relay terminal strip is located just below the power connector on the power supply board.
See Figure 3-2.
Page 14
Quick Start Guide June 2019
Figure 3-2: Alarm relay connections
A. Alarm relay 1 B. Alarm relay 2 C. Alarm relay 3
D. Alarm relay 4
1. To remove the cover, grab it by the upper edges and pull straight out. The relay terminal strip is at the top of the board.
2. Bring the relay wires through the rear conduit opening on the left hand side of the enclosure and make connections to the terminals strip.
3. Replace the cover. The two tabs on the back edge of the cover fit into slots at the rear of the enclosure, and the three small slots in the front of the cover snap into the three tabs next to the relay terminal strip. See Figure 3-2. Once the tabs are lined up, push the cover to snap it in place.
14 Emerson.com/Rosemount
Page 15
June 2019 Quick Start Guide
Keep alarm relay wiring separate from signal wiring. Do not run signal and power or relay wiring in the same conduit or close together in a cable tray.

3.4 Wire sensor

The Rosemount™ FCL is provided with sensor cables pre-wired to the transmitter. If it is necessary to replace the sensor cable, refer to the instructions below.
Procedure
1. Shut off power to the transmitter.
2. Loosen the four screws holding the front panel in place and let it drop down.
3. Locate the appropriate signal board.
Slot 1 (left) Slot 2 (center) Slot 3 (right)
communication input 1 (chlorine) input 2 (optional)
4. Loosen the gland fitting and carefully push the sensor cable up through the fitting as you pull the board forward to gain access to the wires and terminal screws.
5. Wire the sensor to the signal board. Refer to the wiring diagrams in Figure 3-3 and Figure 3-4.
Quick Start Guide 15
Page 16
Quick Start Guide June 2019
Figure 3-3: Wiring Diagram for Free Chlorine Sensor
A. White B. Resistance temperature device return C. White/red
D. Resistance temperature device sense
E. Red
F. Resistance temperature device in G. Clear H. Resistance temperature device shield
I. +5 V out J. -4.5 V out
K. Anode shield
L. Orange
M. Anode
N. Cathode shield O. Gray
P. Cathode
Connect green wire to metal conduit ground plate in bottom of enclosure.
16 Emerson.com/Rosemount
Page 17
June 2019 Quick Start Guide
Figure 3-4: Wiring Diagram for 3900VP-10 pH Sensor (Blue Cable)
A. White B. White/red C. Red
D. Blue
E. Clear (not used)
F. Clear G. Orange H. White/gray
I. Gray J. Resistance temperature device return
K. Resistance temperature device sense
L. Resistance temperature device in
M. Ground solution
N. pH shield O. In pH/ORP
P. Reference shield
Q. In reference
Green (connect to green grounding screw at bottom of enclosure).
6. Once the cable has been connected to the board, slide the board fully into the enclosure while taking up the excess cable through the cable gland.
7. Tighten the gland nut to secure the cable and ensure a sealed enclosure.
3.5
Quick Start Guide 17

Quick Start

Procedure
1. Once connections are secured and verified, apply power to the transmitter. When the transmitter is powered up for the first time, Quick Start screens appear. Using Quick Start is easy.
Page 18
Quick Start Guide June 2019
a. A backlit field shows the position of the cursor.
b. To move the cursor left or right, use the keys to the left or
right of the ENTER key. To scroll up or down or to increase or decrease the value of a digit, use the keys above and below the ENTER key. Use the left and right keys to move the decimal point.
c. Press ENTER to store a setting. Press EXIT to leave without
storing changes. Pressing EXIT also returns the display to the initial Quick Start screen.
d. A vertical black bar with a downward pointing arrow on the
right side of the screen means there are more items to display. Continue scrolling down to display all the items. When you reach the bottom of the list, the arrow points up.
2. Choose the desired language. Scroll down to display more choices.
3. Choose Free Chlorine for sensor 1 (S1).
4. Choose the desired units for chlorine.
18 Emerson.com/Rosemount
Page 19
June 2019 Quick Start Guide
The screens in Step 5 and Step 6 only appear if you have a Rosemount™ FCL-02.
5. If you have a Rosemount FCL-01, go to Step 8. Otherwise, choose pH for Sensor 2 (S2).
6. Choose Analyzer.
7. Choose Live/Continuous. Go to Step 9.
8. The screen below appears only if you have an FCL-01. Enter the pH of the process liquid.
Quick Start Guide 19
Page 20
Quick Start Guide June 2019
9. Choose the desired temperature units.
The main display appears. The outputs and alarms (if an alarm board is present) are assigned to default values.
10. To change outputs, alarms, and other settings, go to the Main Menu and choose Program. Follow the prompts.
20 Emerson.com/Rosemount
Page 21
June 2019 Quick Start Guide
Quick Start Guide 21
Page 22
Quick Start Guide June 2019
22 Emerson.com/Rosemount
Page 23
June 2019 Quick Start Guide
Quick Start Guide 23
Page 24
Quick Start Guide June 2019

A EU Declaration of Conformity

24 Emerson.com/Rosemount
Page 25
June 2019 Quick Start Guide
Quick Start Guide 25
Page 26
Quick Start Guide June 2019

B China RoHS Table

26 Emerson.com/Rosemount
Page 27
June 2019 Quick Start Guide
Quick Start Guide 27
Page 28
GLOBAL HEADQUARTERS
6021 Innovation Blvd. Shakopee, MN 55379
+1 866 347 3427 +1 952 949 7001 liquid.csc@emerson.com
*00825-0100-3412*
Quick Start Guide
00825-0100-3412, Rev. AA
June 2019
EUROPE
Emerson Automation Solutions Neuhofstrasse 19a PO Box 1046 CH-6340 Baar Switzerland
+41 (0) 41 768 6111 +41 (0) 41 768 6300 liquid.csc@emerson.com
ASIA-PACIFIC
Emerson Automation Solutions 1 Pandan Crescent Singapore 128461 Republic of Singapore
+65 6 777 8211 +65 6 777 0947 liquid.csc@emerson.com
Linkedin.com/company/Emerson-
Automation-Solutions
twitter.com/rosemount_news
Facebook.com/Rosemount
youtube.com/RosemountMeasurement
MIDDLE EAST AND AFRICA
Emerson Automation Solutions Emerson FZE Jebel Ali Free Zone Dubai, United Arab Emirates, P.O. Box 17033
+971 4 811 8100 +971 4 886 5465 liquid.csc@emerson.com
©
2019 Emerson. All rights reserved.
The Emerson logo is a trademark and service mark of Emerson Electric Co. Rosemount is a mark of one of the Emerson family of companies. All other marks are the property of their respective owners.
Loading...