Emerson pH-ORP Analyzer-Controller 54e, pH-ORP Analyzer-Controller 54e pH, pH-ORP Analyzer-Controller PN 51A-54epH User Manual

pH/ORP HART®Analyzer/Controller
Instruction Sheet
PN 51A-54epH/rev J October 2010
Model 54e pH/ORP
ESSENTIAL INSTRUCTIONS
Rosemount Analytical designs, manufactures, and tests its prod­ucts to meet many national and international standards. Because these instruments are sophisticated technical products, you must properly install, use, and maintain them to ensure they continue to operate within their normal specifications. The following instructions must be adhered to and integrated into your safety program when installing, using, and maintaining Rosemount Analytical products. Failure to follow the proper instructions may cause any one of the following situations to occur: Loss of life; personal injury; property damage; damage to this instrument; and warranty invalidation.
• Read all instructions prior to installing, operating, and servicing the product. If this Instruction Manual is not the correct manu­al, telephone 1-949-757-8500 and the requested manual will be provided. Save this Instruction Manual for future reference.
• If you do not understand any of the instructions, contact your Rosemount representative for clarification.
• Follow all warnings, cautions, and instructions marked on and supplied with the product.
• Inform and educate your personnel in the proper installation, operation, and maintenance of the product.
• Install your equipment as specified in the Installation Instructions of the appropriate Instruction Manual and per applicable local and national codes. Connect all products to the proper electrical and pressure sources.
• To ensure proper performance, use qualified personnel to install, operate, update, program, and maintain the product.
• When replacement parts are required, ensure that qualified people use replacement parts specified by Rosemount. Unauthorized parts and procedures can affect the product’s performance and place the safe operation of your process at risk. Look alike substitutions may result in fire, electrical hazards, or improper operation.
• Ensure that all equipment doors are closed and protective cov­ers are in place, except when maintenance is being performed by qualified persons, to prevent electrical shock and personal injury.
WARNINGS
RISK OF ELECTRICAL SHOCK
Making cable connections to and servicing this instru­ment require access to shock hazard level voltages which can cause death or serious injury, therefore, disconnect all hazardous voltage before accessing the electronics.
Relay contacts made to separate power sources must be disconnected before servicing.
Electrical installation must be in accordance with the National Electrical Code (ANSI/NFPA-70) and/or any other applicable national or local codes.
Unused cable conduit entries must be securely sealed by non-flammable closures to provide enclo­sure integrity in compliance with personal safety and environmental protection requirements. Use NEMA 4X or IP65 conduit plugs supplied with the instrument to maintain the ingress protection rating (IP65).
For safety and proper performance this instrument must be connected to a properly grounded three­wire power source.
Proper relay use and configuration is the responsi­bility of the user. No external connection to the instrument of more than 60VDC or 43V peak allowed with the exception of power and relay termi­nals. Any violation will impair the safety protection provided.
Do not operate this instrument without front cover secured. Refer installation, operation and servicing to qualified personnel.
WARNING
This product is not intended for use in the res­idential, commercial or light industrial environ­ment per certification to EN50081-2.
For additional information, please visit our website
at www.emersonprocess.com/raihome/liquid/.
3
MODEL 54e pH/ORP SPECIFICATIONS
PHYSICAL SPECIFICATIONS – GENERAL
Enclosure: Epoxy-painted aluminum
NEMA 4X (IP65) 144 X 144 X 131 mm, DIN size (5.7 X 5.7 X 5.2 in.)
Front Panel: Membrane keyboard with tactile
feedback and user selectable security. Light gray, blue and white overlay Light gray enclosure, dark gray bezel
Display: Back-lit dot matrix LCD (7.0 x 3.5 cm),
blue on gray-green The display contrast is compensated for ambient temperature.
Process Variable Character Height: 16mm (0.6 in.)
Electrical Classification:
Class I, Division 2, Groups A, B, C, & D. T5 Ta=50°C. Dust ignition proof: Class II, Division 1, Groups E, F, & G; Class III.
CSA-LR34186:
Max. relay contact rating: 28 Vdc; 110 Vac; 230 Vac; 6 amps resistive
FM: Max. relay contact rating: 28 Vdc resistive
150 mA - Groups A & B; 400 mA - Group C; 540 mA - Group D
Power:
Code -01: 100 - 127 VAC, 50/60 Hz ± 6%, 6.0 W
200 - 253 VAC, 50/60 Hz ± 6%, 6.0 W
Code -02: 20 - 30 VDC, 6.0 W
Current Outputs: Output 1: pH, ORP, temperature,
glass impedance, or reference impedance. Output 2: pH, ORP, temperature, glass impedance,
or reference impedance. Each output is galvanically isolated, 0-20 mA or
4-20 mA into 500 ohms maximum load at 24 Vdc or 115/230 Vac or 500 ohms maximum load at 100/200 Vac. Output 1 includes digital signal 4-20 mA superimposed HART (Code -09 only).
RFI/EMI: EN-61326
LVD (Code -01 only): EN-61010-1
Ambient Temperature:
0 to 50°C (32 to 122°F)
NOTE:
The analyzer is operable from -20 to 60°C
(-4 to 140°F)
performance.
Relative Humidity: 95%, non condensing
Alarms:
Relay 1 - Process, Interval, or Time Proportional Control (TPC requires code -20)
Relay 2 - Process, Interval, or Time Proportional Control (TPC requires code -20)
Relay 3 - Process, Interval, or Time Proportional Control (TPC requires code -20)
Relay 4 - Sensor/analyzer and process fault alarm Each relay has a dedicated LED on the front
panel.
Relay Contacts: Relays 1-3: Epoxy sealed form A
contacts, SPST, normally open
Relay 4: Epoxy sealed form C, SPDT
Resistive Inductive 28 Vdc 5.0 Amps 3.0 Amps 115 Vac 5.0 Amps 3.0 Amps 230 Vac 5.0 Amps 1.5 Amps
Weight/Shipping Weight: 1.8 kg/2.3 kg (4 lb/5 lb)
INSTRUMENT SPECIFICATIONS @ 25°C — pH
Measurement Range: 0 to 14 pH
Output Scale Expansion: Zero suppression: up to 13
pH units Span: Any pH from 1 to 14
Accuracy: ± 0.01 pH
Repeatability: ± 0.01 pH
Stability: ± 0.01 pH/month, non-cumulative
Temperature Coefficient: Input: ± 0.003 pH/°C
Output: ± 0.006 pH/°C
Temperature Compensation:
Pt 100 or Pt 1000 RTD,
Automatic or Manual; –15 to 120°C (5 to 248°F)
INSTRUMENT SPECIFICATIONS @ 25°C — ORP
Measurement Range: –1400 to +1400 mV
Output Scale Expansion: Zero suppression: up to
±1300 mV Span: Any ORP range from 100 to 2800 mV
Accuracy: ± 1.0 mV
Repeatability: ± 1.0 mV
Stability: ± 1.0 mV/month, non-cumulative
Temperature Coefficient: Input: ± 0.2 mV/°C
Output: ± 0.4 mV/°C
Temperature Measurement: –15 to 120°C (5 to 248°F)
Pt 100 or Pt 1000 RTD
Sira MC070111/00