Emerson B-106-300NH User Manual

Instruction Manual
IB-106-300NH Rev. 4.2 July 2002
World Class 3000
Oxygen Analyzer with IFT 3000 Intelligent Field Transmitter
http://www.processanalytic.com

ESSENTIAL INSTRUCTIONS

READ THIS PAGE BEFORE PROCEEDING!
Rosemount Analytical designs, manufactures and tests its products to meet many national and international standards. Because these instruments are sophisticated technical products, you MUST properly install, use, and maintain them to ensure they continue to operate within their normal specifications. The following instructions MUST be adhered to and integrated into your safety program when installing, using, and maintaining Rosemount Analytical products. Failure to follow the proper instructions may cause any one of the following situations to occur: Loss of life; personal injury; property damage; damage to this instrument; and warranty invalidation.
Read all instructions prior to installing, operating, and servicing the product.
If you do not understand any of the instructions, contact your Rosemount Analytical repre-
sentative for clarification.
Follow all warnings, cautions, and instructions marked on and supplied with the product.
Inform and educate your personnel in the proper installation, operation, and mainte-
nance of the product.
Install your equipment as specified in the Installation Instructions of the appropriate In- struction Manual and per applicable local and national codes. Connect all products to the
proper electrical and pressure sources.
To ensure proper performance, use qualified personnel to install, operate, update, program, and maintain the product.
When replacement parts are required, ensure that qualified people use replacement parts specified by Rosemount. Unauthorized parts and procedures can affect the product’s per­formance, place the safe operation of your process at risk, and VOID YOUR WARRANTY. Look-alike substitutions may result in fire, electrical hazards, or improper operation.
Ensure that all equipment doors are closed and protective covers are in place, except when maintenance is being performed by qualified persons, to prevent electrical shock and personal injury.
If a Model 275 Universal HART® Communicator is used with this unit, the software within the Model 275 may require modification. If a software modification is required, please contact your local Fisher-Rosemount Service Group or National Response Center at 1-800-654-7768.
The information contained in this document is subject to change without notice.
Process Analytic Division
1201 N. Main St. Orrville, OH 44667-0901 T (330) 682-9010 F (330) 684-4434 e-mail: gas.csc@EmersonProcess.c om
http://www.processanalytic.c om

HIGHLIGHTS OF CHANGES

Effective May, 1999 Rev. 4.0
Page Summary
Page P-6 Added new Quick Start Guide. Page 3-1 Added Section 3, Setup. Page 4-1 Removed calibration information from Operation section, and created
Section 4, Calibration. Page 6-2 Expanded explanations of IFT status codes. Section 6 Added new troubleshooting procedures.
Effective November, 2001 Rev. 4.1
Page Summary
Highlights Updated Highlights of Changes Appendix A page. Appendix A Replaced Appendix A, Rev. 3.6 with Rev. 3.7.
Effective July, 2002 Rev. 4.2
Page Summary
Highlights Updated Highlights of Changes Appendix A page. Page 4-1 Added reference to new Calibration Record sheet. Page 4-9 Added new Calibration Record sheet. Appendix A Replaced Appendix A, Rev. 3.7 with Rev. 3.8.
HIGHLIGHTS OF CHANGES
APPENDIX A
Effective May, 1996 Rev. 3
Page Summary
-- General. Updated appendix to reflect probe design changes. Page A-13 Added “Extended temperature by-pass arrangements” to Figure A-13
(Sheet 3 of 3)
Effective June, 1996 Rev. 3.1
Page Summary
Page A-13 Updated part ordering information.
Effective August, 1996 Rev. 3.2
Page Summary
Page A-25 Updated cell replacement kit part numbers for the probe.
Effective October, 1996 Rev. 3.3
Page Summary
Page A-6 Added NOTE to Figure A-7.
Effective January, 1997 Rev. 3.4
Page Summary
Page A-1 Added warning to read new safety instructions. Page A-12 Added protective covers and grounds warning. Page A-16 Added protective covers and grounds warning.
Effective February, 1998 Rev. 3.5
Page Summary
Page A-18 Changed screw torque in paragraph A-11h.
Effective July, 1998 Rev. 3.6
Page Summary
-- Changed test gas to calibration gas and reference gas to reference
air throughout the appendix.
HIGHLIGHTS OF CHANGES (CONTINUED)
Effective November, 2001 Rev. 3.7
Page Summary
A-8 Added new cup type diffusion assembly description, paragraph A-6.e.
and diffusion assembly illustrations, Figure A-13 and A-14. A-26 Added new cup type diffusion assembly part numbers 4851B89G04
and 4851B90G04 to replacement parts list. Deleted stainless steel
diffuser assembly from replacement parts list.
Effective July, 2002 Rev. 3.8
Page Summary
A-13 Added troubleshooting symptoms 5 and 6 to Table A-2.
HIGHLIGHTS OF CHANGES
APPENDIX B
Effective February, 1992 Rev. 2
Page Summary
Page B-1 Figure B-1. New HPS 3000 Optional Class 1, Division 1, Group B
(IP56) Explosion-Proof Enclosure added.
Page B-11 Figure and Index No. column added to Table B-2. Replacement Parts
for Heater Power Supply.
Effective January, 1995 Rev. 2.1
Page Summary
Page B-3 Updated Figure B-3, Heater Power Supply Block Diagram for IB
consistency.
Effective January, 1997 Rev. 2.2
Page Summary
Page B-1 Added warning to read new safety instructions.
Page B-3 Corrected Table B-1 specifications list.
Page B-4 Added protective covers and grounds warning.
Page B-8 Added protective covers and grounds warning.
Page B-11 Added expanded fuse description.
HIGHLIGHTS OF CHANGES
Effective June, 1994 Rev. 2
Page Summary
APPENDIX D
Page D-1 Page D-2
Page D-3
Page D-4
Page D-7
Page D-8
Page D-10
Page D-11
Page Summary
Page D-1 Updated Figure D-1, MPS 3000 to include hinge.
Page Summary
Page D-11 Updated replacement parts list to reflect new part numbers.
MPS outline drawing changed to show new MPS.
MPS interior view replaced with new MPS in Figure D-2.
"Optional" for check valve deleted in Figure D-3.
Drawing showing location of optional Z-Purge added as Figure D-4.
Power supply replacement procedures in paragraph D-7 changed to
reflect new design in the MPS. Solenoid valve replacement proce-
dures in paragraph D-8 changed to reflect new design in the MPS.
Old exploded view of MPS replaced with new MPS.
Paragraph D-11, Adding Probes to the new MPS, added.
Change part numbers for the power supply, solenoid valve, and test
gas flowmeter assembly. Add part numbers for reference gas flow-
meter assembly and all the parts in the probe adder kit.
Effective January, 1995 Rev. 2.1
Effective May, 1996 Rev. 2.2
Effective January, 1997 Rev. 2.3
Page Summary
Page D-1
Page D-2 Page D-5
Page D-7
Page D-11
Page Summary
--- Changed test gas to calibration gas and reference gas to reference
Added warning to read new safety instructions.
Corrected Table D-1 Specifications listing, 1
Added protective covers and grounds warning.
Added protective covers and grounds warning, corrected item num-
ber errors in paragraph D-6.
Added expanded fuse descriptions.
Effective July, 1998 Rev. 2.4
air throughout the appendix.
st
entry.
HIGHLIGHTS OF CHANGES
APPENDIX E
Effective May, 1996 Rev. 4
Page Summary
--- General. Updated text and illustrations to reflect new version of IFT.
Page E-4 Updated IFT display status codes and placed in priority sequence.
Effective June, 1996 Rev. 4.1
Page Summary
Page E-2 Updated specification table.
Effective October, 1996 Rev. 4.2
Page Summary
Page E-4 Added new status displays for password protection features.
Effective January, 1997 Rev. 4.3
Page Summary
Front matter Added "Safety instructions for the wiring and installation of this
apparatus.”
Page E-1 Added warning to read new safety instructions.
Page E-2 Deleted NOTE.
Page E-4 Added protective covers and grounds warning.
Page E-8 Added protective covers and grounds warning.
Page E-15 Added expanded fuse description.
Effective July, 1998 Rev. 4.4
Page Summary
--- Changed test gas to calibration gas throughout the appendix.
Effective June, 1999 Rev. 4.5
Page Summary
Page E-1 Changed “real time clock” to “timer”.
HIGHLIGHTS OF CHANGES
APPENDIX J
Effective April, 1995 Rev. 1
Page Summary
Page J-13 Added statement of reference to the return authorization number.
Effective June, 1995 Rev. 1.1
Page Summary
Figure J-4. Updated figure to include “Status group” and “K3 eff” in
calculations.
World Class 3000
PREFACE........................................................................................................................P-1
Definitions........................................................................................................................P-1
Safety Instructions..........................................................................................................P-2
Glossary of Terms .........................................................................................................P-3
Quick Start Guide ..........................................................................................................P-6
1-0 DESCRIPTION AND SPECIFICATIONS...................................................................... 1-1
1-1 Component Checklist of Typical System (Package Contents).................................. 1-1
1-2 System Overview............................................................................................................ 1-2
2-0 INSTALLATION .............................................................................................................. 2-1
2-1 Oxygen Analyzer (Probe) Installation........................................................................... 2-1
2-2 Intelligent Field Transmitter (IFT) Installation.............................................................. 2-9
2-3 Heater Power Supply Installation ............................................................................... 2-14
2-4 Multiprobe Calibration Gas Sequencer Installation.................................................. 2-22
3-0 SETUP............................................................................................................................. 3-1
3-1 Overview.......................................................................................................................... 3-1
3-2 Configuring the Analog Output..................................................................................... 3-1
3-3 Setting Calibration Parameters....................................................................................... 3-1
3-4 Setting the O
3-5 Configuring Efficiency Calculations............................................................................... 3-2
3-6 Configuring the Relay Outputs ..................................................................................... 3-2
Instruction Manual
IB-106-300NH Rev. 4.2

TABLE OF CONTENTS

Alarm Setpoints.................................................................................... 3-2
2
July 2002
4-0 CALIBRATION................................................................................................................ 4-1
4-1 Analog Output Calibration ............................................................................................. 4-1
4-2 System Calibration ......................................................................................................... 4-1
5-0 GENERAL USER INTERFACE (GUI) OPERATION.................................................. 5-1
5-1 Overview.......................................................................................................................... 5-1
5-2 Deluxe Version IFT Displays and Controls................................................................. 5-2
5-3 Help Key......................................................................................................................... 5-3
5-4 Status Line...................................................................................................................... 5-3
5-5 Quick Reference Chart.................................................................................................. 5-3
5-6 Main Menu......................................................................................................................5-3
5-7 Probe Data Sub-Menu................................................................................................... 5-3
5-8 Calibrate O
5-9 Setup Sub-Menu............................................................................................................. 5-4
6-0 TROUBLESHOOTING.................................................................................................... 6-1
6-1 Overview.......................................................................................................................... 6-1
6-2 Special Troubleshooting Notes...................................................................................... 6-1
6-3 System Troubleshooting................................................................................................. 6-1
6-4 Heater Problem ..............................................................................................................6-3
6-5 Cell Problem................................................................................................................... 6-5
6-6 IFT Problem....................................................................................................................6-7
6-7 MPS Problem .................................................................................................................6-8
6-8 Performance Problem (Process Response is Suspect)............................................. 6-9
Sub-Menu.................................................................................................. 5-4
2
Rosemount Analytical Inc. A Division of Emerson Process Management i
Instruction Manual
IB-106-300NH Rev. 4.2 July 2002
7-0 RETURN OF MATERIAL.............................................................................................. 7-1
8-0 APPENDICES ................................................................................................................. 8-1
Appendix A .....................................................................................................................A-1
Appendix B .....................................................................................................................B-1
Appendix D.....................................................................................................................D-1
Appendix E .....................................................................................................................E-1
Appendix J.......................................................................................................................J-1
9-0 INDEX.............................................................................................................................. 9-1
Figure 1. Complete World Class 3000 System.....................................................................P-5
Figure 2. Wiring Layout for World Class 3000 System without HPS or MPS.......................P-8
Figure 1-1. Typical System Package ....................................................................................... 1-1
Figure 1-2. Typical System Installation .................................................................................... 1-5
Figure 1-3. World Class 3000 Typical Application with Intelligent Field Transmitters ............. 1-6
Figure 2-1. Probe Installation................................................................................................... 2-2
Figure 2-2. Orienting the Optional Vee Deflector..................................................................... 2-7
Figure 2-3. Air Set, Plant Air Connection................................................................................. 2-8
Figure 2-4. Outline of Intelligent Field Transmitter................................................................... 2-9
Figure 2-5. Power Supply Board Jumper Configuration ........................................................ 2-10
Figure 2-6. Signal Wire Routing............................................................................................. 2-11
Figure 2-7. IFT Power Supply Board Jumpers....................................................................... 2-12
Figure 2-8. Wiring Layout for IFT Systems without HPS ....................................................... 2-13
Figure 2-9. Microprocessor Board Jumper Configuration..................................................... 2-14
Figure 2-10. IFT Microprocessor Board ................................................................................... 2-15
Figure 2-11. Interconnect Board Jumper Configuration........................................................... 2-16
Figure 2-12. IFT Interconnect Board Output Connections....................................................... 2-16
Figure 2-13. Outline of Heater Power Supply .......................................................................... 2-17
Figure 2-14. Wiring Layout for Complete IFT 3000 System with HPS (Sheet 1of 2)............... 2-18
Figure 2-15. Heater Power Supply Wiring Connections .......................................................... 2-20
Figure 2-16. Jumper Selection Label ....................................................................................... 2-21
Figure 2-17. Jumpers on HPS Mother Board........................................................................... 2-21
Figure 2-18. MPS Module ........................................................................................................2-22
Figure 2-19. MPS Gas Connections ........................................................................................2-23
Figure 2-20. MPS Probe Wiring ............................................................................................... 2-24
Figure 4-1. Typical Calibration Setup....................................................................................... 4-4
Figure 4-2. Portable Rosemount Oxygen Calibration Gas Kit ................................................. 4-5
Figure 4-3. Typical Portable Calibration Setup ........................................................................ 4-6
Figure 4-4. Typical Automatic Calibration System................................................................... 4-7
Figure 5-1. Deluxe Version IFT Displays and Controls............................................................ 5-2
Figure 5-2. Quick Reference Chart .......................................................................................... 5-5
World Class 3000
LIST OF ILLUSTRATIONS
ii Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
Table 4-1. Automatic Calibration Parameters......................................................................... 4-8
Table 5-1. Sample HELP Messages....................................................................................... 5-3
Table 5-2. MAIN menu............................................................................................................ 5-3
Table 5-3. PROBE DATA Sub-Menu...................................................................................... 5-4
Table 5-4. CALIBRATE O
Table 5-5. SETUP Sub-Menu ............................................................................................... 5-12
Table 5-6. Efficiency Constants ............................................................................................ 5-14
Table 6-1. IFT Status Codes................................................................................................... 6-2
Table 6-2. Heater Troubleshooting ......................................................................................... 6-3
Table 6-3. Cell Troubleshooting.............................................................................................. 6-5
Table 6-4. IFT Troubleshooting............................................................................................... 6-7
Table 6-5. MPS Troubleshooting ............................................................................................ 6-8
Table 6-6. Performance Problem Troubleshooting................................................................. 6-9
Instruction Manual
IB-106-300NH Rev. 4.2

LIST OF TABLES

Sub-Menu.................................................................................. 5-10
2
July 2002
Rosemount Analytical Inc. A Division of Emerson Process Management iii
Instruction Manual
IB-106-300NH Rev. 4.2 July 2002
World Class 3000
iv Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
The purpose of this manual is to provide a comprehensive understanding of the World Class 3000 Oxygen Analyzer components, functions, installation, and maintenance.
This manual is designed to provide information about the World Class 3000 Oxygen Ana­lyzer. We recommend that you familiarize yourself with the Overview and Installation sec­tions before installing your emissions monitor.
The overview presents the basic principles of the oxygen analyzer along with its perform­ance characteristics and components. The remaining sections contain detailed proce­dures and information necessary to install and service the oxygen analyzer.
NOTE
!
Only one probe can be calibrated at a time. Probe calibrations must be scheduled appropriately in multiple probe applications.
PREFACE
DEFINITIONS
Instruction Manual
IB-106-300NH Rev. 4.2
July 2002
The following definitions apply to WARNINGS, CAUTIONS, and NOTES found throughout this publication.
Highlights an operation or maintenance procedure, practice, condition, state­ment, etc. If not strictly observed, could result in injury, death, or long-term health hazards of personnel.
NOTE
Highlights an essential operating procedure, condition, or statement.
: EARTH (GROUND) TERMINAL
: PROTECTIVE CONDUCTOR TERMINAL
: RISK OF ELECTRICAL SHOCK
: WARNING: REFER TO INSTRUCTION BULLETIN
Highlights an operation or maintenance procedure, practice, condition, state­ment, etc. If not strictly observed, could result in damage to or destruction of equipment, or loss of effectiveness.

NOTE TO USERS

The number in the lower right corner of each illustration in this publication is a manual illus­tration number. It is not a part number, and is not related to the illustration in any technical manner.
Rosemount Analytical Inc. A Division of Emerson Process Management P-1
Instruction Manual
IB-106-300NH Rev. 4.2 July 2002
FOR THE WIRING AND INSTALLATION
The following safety instructions apply specifically to all EU member states. They should be strictly adhered to in order to assure compliance with the Low Voltage Directive. Non­EU states should also comply with the following unless superseded by local or National Standards.
1. Adequate earth connections should be made to all earthing points, internal and external, where provided.
2. After installation or troubleshooting, all safety covers and safety grounds must be replaced. The integrity of all earth terminals must be maintained at all times.
3. Mains supply cords should comply with the requirements of IEC227 or IEC245.
World Class 3000
IMPORTANT

SAFETY INSTRUCTIONS

OF THIS APPARATUS
4. All wiring shall be suitable for use in an ambient temperature of greater than 75°C.
5. All cable glands used should be of such internal dimensions as to provide adequate cable anchorage.
6. To ensure safe operation of this equipment, connection to the mains supply should only be made through a circuit breaker which will disconnect all circuits carrying conductors during a fault situation. The circuit breaker may also include a mechanically operated isolating switch. If not, then another means of disconnecting the equipment from the supply must be provided and clearly marked as such. Circuit breakers or switches must comply with a recognized standard such as IEC947. All wiring must conform with any local standards.
7. Where equipment or covers are marked with the symbol to the right, hazard­ous voltages are likely to be present beneath. These covers should only be removed when power is removed from the equipment — and then only by trained service personnel.
8. Where equipment or covers are marked with the symbol to the right, there is a danger from hot surfaces beneath. These covers should only be removed by trained service personnel when power is removed from the equipment. Cer­tain surfaces may remain hot to the touch.
9. Where equipment or covers are marked with the symbol to the right, refer to the Operator Manual for instructions.
10. All graphical symbols used in this product are from one or more of the follow­ing standards: EN61010-1, IEC417, and ISO3864.
P-2 Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
Abrasive Shield
An optional component that shields the probe from high velocity particulate entrained in the flue gas stream.
Automatic Calibration
An automatic calibration can only be performed if the system is equipped with an MPS 3000 Mul­tiprobe Calibration Gas Sequencer. Once a calibration is initiated by the operator or by the IFT on a scheduled interval, all calibration actions are performed by the IFT. The MPS switched calibra­tion gases under direction from the IFT.
Calibration
The process of measuring gases of a known concentration, and comparing that known concen­tration to the actual values sensed by the instrument. After reading the calibration gases, the IFT automatically adjusts the slope and constant values to ensure that the system is correctly reading the process gas O
Cold Junction Compensation
A method for compensating for the small voltage developed at the junction of the thermocouple leads in the probe junction box.
values.
2
Instruction Manual
IB-106-300NH Rev. 4.2
July 2002

GLOSSARY OF TERMS

Dead Band
The range through which a signal can be varied without initiating a response. In the IFT 3000, dead band is used to prevent an oxygen signal near an alarm setpoint from cycling the alarm on and off.
GUI
General User Interface. The GUI is the operator interface for the IFT 3000.
HART
A communications protocol using frequency shift keying (FSK) to transmit data on an analog out­put line without affecting the analog output signal.
HPS
Heater Power Supply. An HPS should be used to provide power for the probe heater if the probe is more than 150 ft (45 m) from the IFT.
IFT
Intelligent Field Transmitter.
In Situ
A method of analyzing process gases without removing them from the process stream.
MPS
Multiprobe Calibration Gas Sequencer. The MPS can provide automatic calibration gas se­quencing for up to four probes.
Rosemount Analytical Inc. A Division of Emerson Process Management P-3
Instruction Manual
IB-106-300NH Rev. 4.2 July 2002
Reference Air
Provides a known oxygen concentration to the reference side of the oxygen sensing cell.
Semiautomatic Calibration
Semiautomatic calibration is performed when the system does not include an MPS 3000 Multi­probe Calibration Gas Sequencer. The IFT 3000 provides prompts to direct the user to switch calibration gases when performing the calibration.
Thermocouple
An electrical device made of two dissimilar metals. A thermocouple develops a millivolt signal proportional to its temperature.
Vee Deflector
Protects the optional ceramic diffusor from the process gases. The vee deflector must be posi­tioned so it points toward the direction of the process gas flow. See Figure 2-2 for an illustration of the vee deflector.
World Class 3000
P-4 Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
BEFORE INSTALLING AND WIRING A ROSEMOUNT IFT 3000
1. What is the line voltage being supplied to the IFT 3000? Write the line voltage here __________
2. Use the following drawing, Figure 1, to identify which parts of the World Class 3000 system are included in your system. Components in the shaded area are optional components.
STACK
THERMOCOUPLE
(OPTIONAL )
Instruction Manual
IB-106-300NH Rev. 4.2
July 2002
WHAT YOU NEED TO KNOW
INTELLIGENT FIELD TRANSMITTER WITH WORLD CLASS 3000 PROBE
LINE
VOLT AGE
LINE VOLT AGE
HPS 3000
(OPTIONAL )
WC PROBE 3000
Figure 1. Complete World Class 3000 System
IFT 3000
LINE VOLT AGE
MPS 3000
(OPTIONAL )
CALIBRA TION
GAS
BOTTLES
29850003
Rosemount Analytical Inc. A Division of Emerson Process Management P-5
Instruction Manual
IB-106-300NH Rev. 4.2 July 2002
Use this Quick Start Guide if ...
1. You are using a World Class 3000 probe.
2. You are NOT using any optional components. Optional components are shown in the shaded area in Figure 1.
3. You are familiar with the installation requirements for the IFT 3000 Intelligent Field Trans­mitter and World Class 3000 probe.
4. You are familiar with the procedures for changing the jumpers located in the IFT 3000, as described in Section 2, Installation.
If you cannot use the Quick Start Guide, turn to Section 2, Installation, in this Instruction Bulletin.
World Class 3000
QUICK START GUIDE
P-6 Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000

QUICK START GUIDE FOR IFT 3000 SYSTEMS

Before using the Quick Start Guide, please read “WHAT YOU NEED TO KNOW BEFORE INSTALLING AND WIRING A ROSEMOUNT IFT 3000 INTELLIGENT FIELD TRANSMIT­TER WITH WORLD CLASS 3000 PROBE” on the preceding page.
1. Install the probe in an appropriate location on the stack or duct. Refer to Section 2, para­graph 2-1a for information on selecting a location for the probe.
2. Connect calibration gas and reference air to the probe.
3. Verify the jumper selection on the IFT 3000 power supply board, microprocessor board, and interconnect board, as shown in Figure 2.
4. Install the IFT 3000 in the desired location. Refer to Section 2, paragraph 2-2a for informa­tion on selecting a location for the IFT 3000.
5. Wire the probe to the IFT as shown in Figure 2.
Instruction Manual
IB-106-300NH Rev. 4.2
July 2002
6. Connect line voltage to the IFT as shown in Figure 2.
7. Apply power to the IFT 3000. Allow sufficient time for the probe to reach normal operating temperature. The time required will vary based on process temperature and other variables.
8. Perform a manual (semiautomatic) calibration. Press the CAL key on the GUI. Select the PERFORM CALIBRATION sub-menu. “Press ENTER to start Manual Calibration” will appear on the LCD display. Press ENTER to start the calibration process. Follow the in­structions on the LCD display. Refer to Section 4, Calibration, for more information on per­forming a calibration.
Rosemount Analytical Inc. A Division of Emerson Process Management P-7
Instruction Manual
IB-106-300NH Rev. 4.2 July 2002
PROBE JUNCTION
BOX WIRING
GN CELL -VE
OR CELL +VE
YE CHROMEL
RD ALUMEL
GN
BK
HEATER
}
World Class 3000
WORLD CLASS
PROBE
BK
123456 78
BL
YE
RD
OR
PROBE MV -
PROBE MV +
PROBE TC +
GNE
PROBE TC -
LINE VOLTAGE SECTION
100 V.A.C. 120 V.A.C. 220 V.A.C. 240 V.A.C.
LINE VOLTAGE
JUMPERS ON IFT
POWER SUPPLY
JUMPER
(INSTALL) JM3, JM7, JM2 JM8, JM7, JM1
JM6, JM5, JM2 JM6, JM5, JM1
BOARD
WH
R
BK
H
J1
3D39122G REV
POWER SUPPLY BOARD
LINE VOLTAGE SECTION
100 V.A.C.
JM3, JM7, JM2
120 V.A.C.
JM8, JM7, JM1
220 V.A.C.
JM6, JM5, JM2 JM6, JM5, JM1
240 V.A.C.
JUMPER
(INSTALL)
J2 J3 J4
3D39120G REV
INTERCONNECT BOARD
J1
J5
J6
J5 J6
J7
SHIELD STACK TC ­STACK TC +
J8 J9
SHIELD
L
LINE
VOLTAGE
BK WH
GN PU
OR
BL YE
RD
H
EN
ER
PROBE TC -
RD YE
PROBE TC +
SHIELD
BL
PROBE MV PROBE
OR
-
MV+
JM1
NOTES:
INSTALL JUMPER ACROSS TERMINALS 13 AND 14.
INSTALL JUMPER ACROSS TERMINALS 7 AND 8.
29850002
Figure 2. Wiring Layout for World Class 3000 System without HPS or MPS
P-8 Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
IFT 3000 INTELLIGENT FIELD TRANSMITTER
Performing a Manual (Semiautomatic) Calibration
1. Connect the high calibration gas to the probe fitting.
2. Press the CAL key.
3. Select the PERFORM CALIBRATION sub-menu.
4. Press the ENTER key.
5. Turn on the high calibration gas.
Instruction Manual
IB-106-300NH Rev. 4.2
July 2002
QUICK REFERENCE GUIDE
6. When the O
7. Turn off the high calibration gas and turn on the low calibration gas.
8. Press Enter.
9. When the O
10. The LCD display will show “Resistance Check”. When the display changes to “Turn off low calibration gas”, turn off the low calibration gas and press ENTER.
11. When the oxygen reading has stabilized at the process value, press ENTER.
Setting up the Analog Output
1. Press the SETUP key.
2. Select the Analog Output sub-menu.
3. Set the SOURCE to O Dual Range O
4. Set the AOUT TYPE to the desired setting. Note that the setting must agree with the position of the analog output selector switch. If you will communicate with the IFT using HART com­munications, the AOUT TYPE must be set to HART 4-20mA.
5. Select Range Setup and press ENTER.
6. Set the Xfer Fnct to Lin or Log, as desired.
7. Select Range Values and press ENTER.
reading is stable, press ENTER.
2
reading is stable, press ENTER.
2
. For information on configuring the analog output for Efficiency or
, refer to Section V, Operation.
2
2
8. Set the High End to the oxygen concentration to be represented by the high analog output value, i.e., 20mA or 10V.
9. Set the Low End to the ox ygen conc entr at ion to be repres ent ed b y the low anal og outp ut value, i.e., 0 or 4mA or 0V.
10. Press the ESC key until you are back at the Main menu.
Rosemount Analytical Inc. A Division of Emerson Process Management P-9
Instruction Manual
IB-106-300NH Rev. 4.2 July 2002

HART COMMUNICATOR FAST KEY SEQUENCES

Toggle Analog Output Tracking View O2 Value
World Class 3000
Perform Calibration Analog Output Upper Range Value
2313 324
Trim Analog Output Analog Output Lower Range Value
24 325
2312 111
View Analog Output
121
Technical Support Hotline:
For assistance with technical problems, please call the Customer Support Center (CSC). The CSC is staffed 24 hours a day, 7 days a week.
Phone: 1-800-433-6076
In addition to the CSC, you may also contact Field Watch. Field Watch coordinates Rosemount’s field service throughout the US and abroad.
Phone: 1-800-654-RSMT (1-800- 6 54-77 68)
Rosemount may also be reached via the Internet through e-mail and the World Wide Web:
E-mail: GAS.CSC@emersonprocess.com
World Wide Web: www.processanalytic.com
P-10 Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
1

DESCRIPTION AND SPECIFICATIONS

Instruction Manual
IB-106-300NH Rev. 4.2
July 2002
SECTION 1
1-1 COMPONENT CHECKLIST OF TYPICAL
SYSTEM (PACKAGE CONTENTS)
A typical Rosemount World Class 3000 Oxygen Analyzer with IFT 3000 Intelligent Field Trans-
1
mitter should contain the items shown in Figure 1-1. Record the part number, serial number, and order number for each component of your sys­tem in the table located on the first page of this manual.
1. Intelligent Field Transmitter
2. Instruction Bulletin
3. Multiprobe Calibration Gas Sequencer (Optional)
4. Heater Power Supply (Optional)
5. Oxygen Analyzer (Probe)
6. System Cable
7. M Adapter Plate with mounting hardware and gasket
2
8. Reference Air Set (If MPS not supplied)
9. HART
®
Communicator Package (Optional)
3
4
5
21190001
HART
MAN4275A00
October1994
Communicator
o
FISHER-ROSEMOUNT
English
TM
8
9
6
7
Figure 1-1. Typical System Package
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-1
Instruction Manual
IB-106-300NH Rev. 4.2 July 2002
World Class 3000

1-2 SYSTEM OVERVIEW

a. Scope
This Instruction Bulletin has been designed to supply details needed to install, startup, operate, and maintain the Rosemount World Class 3000 Oxygen Analyzer with IFT 3000 Intelligent Field Transmitter. The Intel­ligent Field Transmitter (IFT) can be inter­faced with one World Class 3000 probe. The IFT provides all necessary intelligence for controlling the probe and optional MPS 3000 Multiprobe Calibration Gas Se­quencer. Appendices at the back of this manual detail each component and option from the standpoint of troubleshooting, re­pair, and spare parts.
Operator/Technician interface to the IFT can be provided from the displays and key­pads on the front panel, and remotely through HART utilizing the 4-20 mA out-put signal from the IFT interconnect board. HART Communi­cator IFT applications are detailed in Appendix J.
b. System Description
The Rosemount Oxygen Analyzer (Probe) is designed to measure the net concentration of oxygen in an industrial process; i.e., the oxygen remaining after all fuels have been oxidized. The probe is permanently posi­tioned within an exhaust duct or stack and performs its task without the use of a sam­pling system.
The equipment measures oxygen percent­age by reading the voltage developed across a heated electrochemical cell, which consists of a small yttria-stabilized, zirconia disc. Both sides of the disc are coated with porous metal electrodes. When operated at the proper temperature, the millivolt output voltage of the cell is given by the following Nernst equation:
®
communications protocol,
Where:
1. P2 is the partial pressure of the oxygen in the measured gas on one side of the cell,
2. P1 is the partial pressure of the oxygen in the reference air on the other side,
3. T is the absolute temperature,
4. C is the cell constant,
5. K is an arithmetic constant.
NOTE
For best results, use clean, dry, in­strument air (20.95% oxygen) as a ref­erence air.
When the cell is at operating temperature and there are unequal oxygen concentra­tions across the cell, oxygen ions will trave l from the high partial pressure of oxygen side to the low partial pressure side of the cell. The resulting logarithmic output voltage is approximately 50 mV per decade. Be­cause the magnitude of the output is pro­portional to the logarithm of the inverse of the sample of the oxygen partial pressure, the output signal increases as the oxygen concentration of the sample gas decreases. This characteristic enables the oxygen analyzer to provide exceptional sensitivity at low oxygen concentrations.
Oxygen analyzer equipment measures net oxygen concentration in the presenc e of all the products of combustion, including water vapor. Therefore, it may be considered an analysis on a "wet" basis. In comparison with older methods, such as the Orsat ap­paratus, which provides an analysis on a "dry" gas basis, the "wet" analysis will, in general, indicate a lower percentage of oxygen. The difference will be proportional to the water content of the sampled gas stream.
EMF = KT log10(P1/P2) + C
1-2 Description and Specifications Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
1
Instruction Manual
IB-106-300NH Rev. 4.2
July 2002
c. System Configuration
The equipment covered in this manual con­sists of three major components: the oxy­gen analyzer (probe), the intelligent field transmitter (IFT), and an optional heater power supply (HPS). The HPS is required where the cable run between the probe and the electronics is greater than 150 ft (45 m). There is also an optional multiprobe calibra­tion gas sequencer (MPS) to facilitate automatic calibration of the probe.
Probes are available in five length options, giving the user the flexibility to use an in situ penetration appropriate to the size of the stack or duct. The options on length are 18 in. (457 mm), 3 ft (0.91 m), 6 ft (1.83 m), 9 ft (2.7 m), or 12 ft (3.66 m).
The IFT contains electronics that control probe temperature (in conjunction with the optional HPS), supply power, and provide isolated outputs that are proportional to the measured oxygen concentration. The oxy­gen sensing cell is maintained at a constant temperature by modulating the duty cycle of the probe heater. The IFT accepts millivolt signals generated by the sensing cell and produces outputs to be used by remotely connected devices. The IFT output is iso­lated and selectable to provide linearized voltage or current.
The heater power supply (HPS) can provide an interface between the IFT and the probe. The HPS contains a transformer for sup­plying proper voltage to the probe heater. The enclosure has been designed to meet NEMA 4X (IP56) specifications for water tightness; an optional enclosure to meet Class 1, Division 1, Group B (IP56) explo­sion proof is also available.
Systems with multiprobe and multiple IFT applications may employ an optional MPS 3000 Multiprobe Calibration Gas Se-
quencer. The MPS 3000 provides automatic calibration gas sequencing for up to four probes and IFTs to accommodate automatic calibration.
d. System Features
1. Unique and pat ent ed elec tro nic cel l protection action that automatically protects sensor cell when the analyzer detects reducing atmospheres.
2. Output vo lta ge and se ns it iv ity increase as the oxygen concentration decreases.
3. User friendly, menu driven operator interface with context-sensitive on-line help.
4. Field replaceable cell.
5. Analyzer constructed of rugged 316 LSS for all wetted parts.
6. The intelligent field transmitter (IFT) can be located up to 150 ft (45 m) from the probe when used without optional heater power supply (HPS). When the system includes the optional HPS, the HPS can be located up to 150 ft (45 m) from the probe and the IFT may be lo­cated up to 1200 ft (364 m) from the HPS.
7. All electronic modules are adaptable to 100, 120, 220, and 240 line voltages.
8. Five languages may be selected for use with the Intelligent Field Transmitter:
English Italian French Spanish German
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-3
Instruction Manual
IB-106-300NH Rev. 4.2 July 2002
World Class 3000
9. An operator can set up, calibrate, or troubleshoot the IFT in one of two ways:
(a) Optional General User Interface
(GUI). The GUI is housed within the IFT electronics enclosure and makes use of an LCD and keypad.
(b) Optional HART Interface. The IFT's
4-20 mA output line transmits an analog signal proportional to oxy­gen level. The line also carries all information normally accessed by use of the General User Interface LCD and keypad. This information can be accessed through the following:
1 Rosemount Model 275 Hand-
held Communicator - The handheld communicator re­quires Device Descriptor (DD) software specific to the World Class 3000 product. The DD software will be supplied with many model 275 units, but can also be programmed into existing units at most Fisher­Rosemount service offices.
2 Personal Computer (PC) -
The use of a personal com­puter requires Cornerstone software with Module Library (ModLib) specific to the World Class 3000 product.
3 Selected Distributed Control
Systems - The use of distrib­uted control systems requires input/output (I/O) hardware and software which permit HART communications.
e. Handling the Oxygen Analyzer.
It is important that printed circuit boards and integrated circuits are handled only when adequate antistatic precautions have been taken to pre­vent possible equipment damage.
The oxygen analyzer is designed for industrial application. Treat each component of the system with care to avoid physical damage. The probe contains components made from ce­ramics, which are susceptible to shock when mishandled.
NOTE
Retain packaging in which the oxygen analyzer arrived from the factory in case any components are to be shipped to another site. This packag­ing has been designed to protect the product.
f. System Considerations
Prior to installation of your Rosemount World Class 3000 Oxygen Analyzer with Intelligent Field Transmitter make sure that you have all of the components necessary to make the system installation. Ensure that all the components are properly integrated to make the system functional.
Once you have verified that you have all the components, select mounting locations and determine how each component will be placed in terms of available power supply, ambient temperatures, environmental con­siderations, convenience, and serviceability. A typical system installation is illustrated in Figure 1-2. Figure 1-3 shows a typical sys­tem wiring. For details on installing the indi­vidual components of the system, refer to Section 2, Installation.
1-4 Description and Specifications Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
1
Instruction Manual
IB-106-300NH Rev. 4.2
July 2002
CALIBRATION
INSTRUMENT
AIR SUPPLY
(REF. AIR)
GAS
PRESSURE
REGULATOR
GASES
STACK
FLOWMETER
STANDARD
DUCT
OXYGEN ANALYZER (PROBE)
INTELLIGENT FIELD TRANSMITTER
MULTIPROBE
CALIBRATION GAS
SEQUENCER
}
ADAPTER PLATE
LINE VOLTAGE
ADAPTER
PLATE
CALIBRATION
GAS
GASES
STACK
OPTIONS
DUCT
OXYGEN ANALYZER (PROBE)
SUPPLY
INST. AIR
CAL GAS 1
CAL GAS 2
REFERENCE AIR
Figure 1-2. Typical System Installation
HEATER POWER SUPPLY
INTELLIGENT FIELD TRANSMITTER
}
LINE VOLTAGE
27270001
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-5
Instruction Manual
IB-106-300NH Rev. 4.2 July 2002
World Class 3000
Stack Thermocouple
(optional)
World Class 3000
Probe
2-Calibration Gas Lines
World Class 3000
Probe
Stack Thermocouple
(optional)
Wire [150 Ft (45 m) Max]
7-Conductor Cable
[150 Ft (45 m) Max]
by Customer
[300 Ft (90 m) Max]
2-Conductor T/C
(optional)
(HPS not required for lengths of less than 150 feet)
7-Conductor Cable
[150 Feet (45 m) Max]
2-Conductor T/C
Wire [150 Feet (45 m) Max]
(optional)
(OPTIONAL)
HPS 3000
HPS 3000
Explosion Proof Required only for Hazardous Area Applications, otherwise use NEMA 4X. Lengths Exceeding 150 ft (45 m).
(OPTIONAL)
MPS 3000
CALIBRATION GAS
SEQUENCER
Modular Design Up to 4 Probes
Line Voltage
Line Voltage
Calibration Gas
by
Customer
4 Twisted Pair Plus 2 Twisted Pair for Options [1200 Ft (364 m) Max]
Line Voltage
IFT 3000
Intelligent Field Transmitter
NEMA 4X Enclosure
Line Voltage 100 to 120 Volt 220 to 240 Volt
5 Conductor
[1000 Ft (309 m) Max]
Line Voltage
IFT 3000
Intelligent Field Transmitter
NEMA 4X Enclosure
Line Voltage 100 to 120 Volt 220 to 240 Volt
HART Model 275
Hand Held
Interface
Customer's Laptop with
Cornerstone Software
Control System
with HART
27270002
World Class 3000
Probe
7-Conductor Cable
[150 Ft (45 m) Max]
2-Calibration Gas Lines by Customer [300 Ft (90 m) Max]
Line Voltage
HPS 3000
Heater Power Supply
[Optional,
Required for > 150 Ft (45 m)]
4 Twisted Pair, plus 2 Twisted Pair
for Options [1200 Ft (364 m) Max]
Line Voltage
IFT 3000
Intelligent Field Transmitter
NEMA 4X Enclosure
Line Voltage 100 to 120 Volt 220 to 240 Volt
4-20 mA Output (Twisted Pair)
Termination in Control Room
Customer's Distributed
Interface Capability
Figure 1-3. World Class 3000 Typical Application with Intelligent Field Transmitters
1-6 Description and Specifications Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
1
Instruction Manual
IB-106-300NH Rev. 4.2
July 2002
After selecting the probe mounting location, provision should be made for a platform where the probe can be easily serviced. The intelligent field transmitter (IFT) can be located up to 150 ft (45 m) cabling distance from the probe when used without optional heater power supply (HPS). When the sys­tem includes the optional HPS, the HPS can be located up to 150 ft (45 m) cabling dis­tance from the probe and the IFT may be located up to 1200 ft (364 m) cabling dis­tance from the HPS.
A source of instrument air is required at the probe for reference air use. Since the probe is equipped with an in-place calibration feature, provision should be made for con-
necting calibration gas tanks to the oxygen analyzer when the probe is to be calibrated.
If the calibration gas bottles will be perma­nently hooked up, a check valve is required next to the calibration fittings on the probe junction box. This is to prevent breathing of calibration gas line and subsequent flue gas condensation and corrosion. The check valve is in addition to the stop valve in the calibration gas kit or the solenoid valve in the multiprobe calibration gas sequencer units.
An optional Z-purge arrangement is avail­able for applications where hazardous area classification may be required (See Appli­cation Data Bulletin AD 106-300B).
Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-7
Instruction Manual
IB-106-300NH Rev. 4.2 July 2002
World Class 3000
1-8 Description and Specifications Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
2
Instruction Manual
IB-106-300NH Rev. 4.2
July 2002
SECTION 2

INSTALLATION

2-1 OXYGEN ANALYZER (PROBE)
INSTALLATION
Before starting to install this equip­ment, read the "Safety instructions for wiring and installation of this appara­tus" at the front of this Instruction Bulletin. Failure to follow the safety instructions could result in serious injury or death.
a. Selecting Location
1. The location of the probe in the stack or flue is most important for maximum accuracy in the oxygen analyzing pro­cess. The probe must be positioned so that the gas it measures is representa­tive of the process. Best results are normally obtained if the probe is posi­tioned near the center of the duct (40 to 60% insertion). A point too near the edge or wall of the duct may not pro­vide a representative sample because of the possibility of gas stratification. In addition, the sensing point should be selected so that the process gas tem­perature falls within a range of 50° to 1300°F (10° to 704°C). Figure 2-1 pro­vides mechanical installation references.
4. If the probe is to be mounted outside, subject to rain and snow conditions, make sure the back of the probe (out­side of the duct) is insulated to prevent the formation of flue gas condensate in the calibration gas lines.
Do not allow the temperature of the probe junction box to exceed 300°F (149°C) or damage to the unit may re­sult. If the probe junction box tempera­ture exceeds 300°F (149°C), the user must fabricate a heat shield or provide adequate cooling air to the probe junc­tion box.
b. Mechanical Installation
1. Ensure that all components are avail­able for installation of the probe. En­sure that the system cable is the required length. If equipped with the optional ceramic diffusor element, en­sure that it is not damaged.
2. The probe may be installed intact as it is received. It is recommended that you disassemble the adapter plate for each installation.
NOTE
2. Check the flue or stack for holes and air leakage. The presence of this con­dition will substantially affect the accu­racy of the oxygen reading. Therefore, either make necessary repairs or install the probe upstream of any leakage.
3. Ensure that the area is clear of ob­structions internal and externa l that wi ll interfere with installation. All o w ade­quate clearance for removal of probe (Figure 2-1).
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-1
An abrasive shield is recommended for high velocity particulate in the flue stream (such as those in coal fired boilers, kilns, and recovery boilers). Vertical and horizontal brace clamps are provided for 9 ft and 12 ft (2.75 m and 3.66 m) probes to provide me­chanical support of the probe. Refer to Figure 2-1, sheet 5.
3. Weld or bolt adapter plate (Figure 2-1) onto the duct.
Instruction Manual
IB-106-300NH Rev. 4.2 July 2002
World Class 3000
FURNISHED IN - XIT
4512C34
ADAPTER & ACCESSORY
TO AMBIENT
INSULATE IF EXPOSED
WEATHER CONDITIONS
4512C35
4512C36
2.27 (58)
DIA MAX
ROSEMOUNT
5.85 (148.6)
DIM "A"
WITH STANDARD
REF AIR
CAL GAS
7.58 (192)
SNUBBER
DIFFUSER
ELEC
DIM "B" REMOVAL ENVELOPE
1/4 IN. TUBE
6 MM TUBE
6 MM TUBE
ANSI
DIN
JIS
1/2"
CONN
CONDUIT
GAS CAL
JIS
4512C18H01
6.10
(155)
1.88 (48)
0.59
AIR
REF
(15)
AT THE BOTTOM
BOTTOM VIEW
INSTALL WITH CONNECTIONS
5.12
(130)
THESE FLAT FACED FLANGES ARE MANUFACTURED
TO ANSI, DIN, AND JIS BOLT PATTERNS AND ARE NOT
DIMENSIONS ARE IN INCHES WITH MILLIMETERS IN
PARENTHESES.
PRESSURE RATED.
2.
NOTES: 1.
27270009
3.80 (96.5)
DIFFUSER
FOR PROBE
WITH CERAMIC
4.90 (124.5)
ADD TO DIM "A" FOR
ARRESTOR
DIFFUSER AND FLAME
PROBE WITH CERAMIC
DIN
7.28
4512C19H01
ANSI
6.00
4512C17H01
(185)
(153)
0.71
0.75
(18)
(20)
5.71
4.75
(145)
(121)
DIM "B"
DIM "A"
27.3 (694)
16 (406)
45.3 (1151)
34 (864)
81.3 (2065)
117.3 (2980)
70 (1778)
106 (2692)
153.3 (3894)
142 (3607)
3535B18H02
3635B48H01
ANSI
JIS
0.062 THK GASKET
ADD TO DIM "A"
3535B45H01
DIN
TABLE I MOUNTING FLANGE
3FT
6FT
9FT
12 FT
PROCESS FLOW MUST
BE IN THIS DIRECTION
WITH RESPECT TO
DEFLECTOR 3534848G01
FLANGE
DIA.
HOLE
DIA.
(4) HOLES
EQ SP ON BC
TABLE II INSTALLATION/REMOVAL
18 IN.
PROBE
Figure 2-1. Probe Installation (Sheet 1 of 5)
2-2 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
2
7.50
7.48
0.75
9.25 (235)
*
JIS
7.48
0.945
9.25 (235)
*
DIN
* FLANGES ARE MANUFACTURED TO ANSI,
TABLE IV. FLANGE SIZE
BOLT
CIRCLE
0.75
(8) HOLES
DIAMETER
FLANGE
9.00 (153)
DIAMETER
*
ANSI
DIN, AND JIS BOLT PATTERNS AND ARE
FLAT FACED. THESE FLANGES ARE NOT
PRESSURE RATED.
5.7 (145)
14.5
(369)
DIM "D" REMOVAL ENVELOPE
7.00
(178)
Instruction Manual
IB-106-300NH Rev. 4.2
July 2002
27270010
REF AIR AND
CAL GAS
CONNECTOR
ELECTRICAL
CONNECTOR
CAL GAS LINES
CHECK VALVE FOR
31.1
(790)
45.3 (1151)
DIM "D" DIM "E"
27
(686)
DIM "C"
NOMINAL MEASUREMENTS
TABLE III. REMOVAL / INSTALLATION
3FT
67.1
81.3
63
6FT
(1704)
(2065)
(1600)
103.1
(2619)
117.3
(2980)
99
(2515)
9FT
139.1
(3533)
153.3
(3894)
135
(3429)
12 FT
DIM "E" (WITH FLAME ARRESTOR)
IN HARDWARE PACKAGE
DIM "C"
0.06 THK GASKET FURNISHED
(P/N 3535B58G04 - JIS)
(P/N 3535B58G02 - ANSI)
SEE TABLE IV
FOR FLANGE
(P/N 3535B58G06 - DIN)
SIZES
3.6
NOMINAL
(P/N 4843B38G02)
SNUBBER DIFFUSION/
DUST SEAL ASSEMBLY
INSULATE IF
EXPOSED TO
AMBIENT WEATHER
DIMENSIONS ARE IN INCHES WITH
MILLIMETERS IN PARENTHESES.
CONDITIONS
NOTE:
Figure 2-1. Probe Installation (Sheet 2 of 5)
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-3
Instruction Manual
IB-106-300NH Rev. 4.2 July 2002
JIS
(P/N 3535B58G04)
9.25
(235)
4.92
(125)
(M-20 x 2.5)
(200)
7.894
World Class 3000
16860021
8 THREADED HOLES
EQUALLY SPACED ON
D DIA B.C.
ABRASIVE SHIELD
FLANGE O.D.
TABLE VI. ADAPTOR PLATE DIMENSIONS FOR ABRASIVE SHIELD
DIN
ANSI
IN.
DIMENSIONS
JIS
9.25
(P/N 3535B58G06)
9.00
(P/N 3535B58G02)
"A"
(mm)
6.50
(P/N 4512C35G01)
(235)
(229)
(165)
3.94
(100)
4.75
(121)
"B"
DIA
(M-12 x 1.75)
(M-16 x 2)
0.625-11
"C"
THREAD
(130)
5.118
7.48
7.50
"D"
(190)
(191)
DIA
ATTACHING HARDWARE.
NOTE: PART NUMBERS FOR ADAPTOR PLATES INCLUDE
o
22.5
A
OUTSIDE WALL SURFACE.
CROSSHATCHED AREA IN 4
CORNERS MAY BE USED TO
FIELD BOLTING OF PLATE TO
PROVIDE ADDITIONAL HOLES FOR
AND 12 FT ABRASIVE SHIELD
ADAPTOR PLATE FOR 3, 6, 9,
INSTALLATIONS. SEE SHEET 2.
C
A
B
4 STUDS,
LOCKWASHERS AND
NUTS EQUALLY
SPACED ON
C DIA B.C.
B
ADAPTOR PLATE FOR
STD WORLD CLASS 3000
PROBE INSTALLATION.
SEE SHEET 1.
TABLE V. ADAPTOR PLATE DIMENSIONS FOR PROBE
DIN
ANSI
IN.
DIMENSIONS
7.5
(P/N 4512C36G01)
6.00
(P/N 4512C34G01)
"A"
(mm)
(191)
(153)
(M-16 x 2)
0.625-11
"B"
THREAD
(145)
5.708
4.75
(121)
"C"
DIA
A
o
45
A
C
ATTACHING HARDWARE.
2.500 DIA
NOTE: PART NUMBERS FOR ADAPTOR PLATES INCLUDE
Figure 2-1. Probe Installation (Sheet 3 of 5)
2-4 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
2
INSTALLATION FOR METAL
WALL STACK OR DUCT
CONSTRUCTION
Instruction Manual
IB-106-300NH Rev. 4.2
July 2002
INSTALLATION FOR MASONRY
WALL STACK CONSTRUCTION
MTG HOLES SHOWN ROTATED
o
45 OUT OF TRUE POSITION
WELD OR BOLT ADAPTOR PLATE TO METAL WALL OF STACK OR DUCT. JOINT MUST BE AIR TIGHT.
0.50 [13]
3.75 [95]
MIN DIA HOLE IN WALL
STACK OR DUCT METAL WALL
0.50 [13]
BOLT ADAPTOR
PLATE TO OUTSIDE
WALL SURFACE
FIELD WELD
PIPE TO
ADAPTOR PLATE
MTG HOLES
SHOWN ROTATED
o
45 OUT OF
TRUE POSITION
JOINT MUST
BE AIRTIGHT
OUTSIDE WALL
SURFACE
NOTE: ALL MASONRY STACK WORK AND JOINTS EXCEPT
ADAPTOR PLATE NOT FURNISHED BY ROSEMOUNT.
4.50 [114] O.D. REF
PIPE 4.00 SCHED 40 PIPE SLEEVE (NOT BY ROSEMOUNT) LENGTH BY CUSTOMER
MASONRY STACK WALL
WELD OR BOLT ADAPTOR
PLATE TO METAL WALL
OF STACK OR DUCT.
JOINT MUST BE AIR TIGHT.
2.50 [63.5]
MIN DIA HOLE IN WALL
STACK OR DUCT METAL WALL
BOLT ADAPTOR
PLATE TO OUTSIDE
WALL SURFACE
JOINT MUST
BE AIRTIGHT
OUTSIDE WALL
SURFACE
NOTE: DIMENSIONS IN INCHES WITH
MILLIMETERS IN PARENTHESES.
FIELD WELD PIPE TO ADAPTOR PLATE
3.50 [89] O.D. REF
PIPE 3.00 SCHED 40 PIPE SLEEVE (NOT BY ROSEMOUNT) LENGTH BY CUSTOMER
MASONRY STACK WALL
624038
Figure 2-1. Probe Installation (Sheet 4 of 5)
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-5
Instruction Manual
IB-106-300NH Rev. 4.2 July 2002
World Class 3000
o
60 MAX.
o
30 MIN.
4.12
(105)
4.12
(105)
BRACE BARS (NOT BY ROSEMOUNT)
2.00 (51)
1.00 (25)
2 HOLES - 0.625 (16) DIA. FOR
0.50 (12) DIA. BOLT
1.00 (25) MAX.
0.375 (10)
NOTE: DIMENSIONS IN INCHES WITH
5.62
(143)
5.62
(143)
MILLIMETERS IN PARETHESES.
VERTICAL BRACE CLAMP ASSY.
HORIZONTAL BRACE CLAMP ASSY. (BOTH BRACE CLAMP ASSEMBLIES ARE THE SAME. INSTALLATION AND LOCATION OF CLAMP ASSEMBLIES AND BRACE BARS TO BE DONE IN FIELD.)
ABRASIVE SHIELD
BY ROSEMOUNT
}
36.00 (914)
NOTE: BRACING IS FOR VERTICAL AND HORIZONTAL PROBE INSTALLATION.
EXTERNAL BRACING REQUIRED FOR 9 FT AND 12 FT (2.75 M AND 3.66 M) PROBES AS SHOWN ABOVE.
Figure 2-1. Probe Installation (Sheet 5 of 5)
27270008
2-6 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
2
Instruction Manual
IB-106-300NH Rev. 4.2
July 2002
4. If using the optional ceramic diffusor element, the vee deflector must be cor­rectly oriented. Before inserting the probe, check the direction of flow of the gas in the duct. Orient the vee deflector on the probe so that the apex points upstream toward the flow (Figure 2-2). This may be done by loosening the setscrews, and rotating the vee de­flector to the desired position. Retighten the setscrews.
5. In horizontal installations, the probe junction box should be oriented so that the system cable drops vertically from the probe junction box. In a vertical in­stallation, the system cable can be ori­ented in any direction.
6. If the system has an abrasive shield, check the dust seal packings. The joints in the two packings must be staggered 180°. Also, make sure that the packings are in the hub grooves as the probe slides into the 15° forcing cone in the abrasive shield.
NOTE
7. Insert probe through the opening in the mounting flange and bolt the unit to the flange. When probe lengths selected are 9 or 12 ft (2.75 or 3.66 m), special brackets are supplied to provide addi­tional support for the probe inside the flue or stack. See Figure 2-1, sheet 5.
NOTE
Probe Installation To maintain CE compliance, ensure there is a good connection between the chassis of the probe and earth.
GAS FLOW DIRECTION
VEE
DEFLECTOR
APEX
DIFFUSION
ELEMENT
SETSCREW
FILTER
VEE
DEFLECTOR
If process temperatures will exceed 392°F (200°C), use anti-seize com­pound on stud threads to ease future removal of probe.
624017
Figure 2-2. Orienting the Optional Vee Deflector
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-7
Instruction Manual
IB-106-300NH Rev. 4.2 July 2002
World Class 3000
c. Reference Air Package
After the oxygen analyzing (probe) unit is installed, connect the reference air set to the probe junction box. The reference air set should be installed in accord ance with Figure 2-3.
d. Service Required
1. Power in put: 100 , 115 or 220 Vac sin­gle phase, 50 to 60 Hz, 3 amp mini­mum. (See label.)
2. Compressed air: 10 psig (68.95 kPa) minimum, 225 psig (1551.38 kPa)
0.125-27 NPT FEMALE OUTLET CONNECTION
1
4.81 (122.17)
FLOW SET
POINT KNOB
2
OUTLET
1.19 (30.22)
DRAIN VALVE
maximum at 2 scfh (56.6 L/hr) maxi­mum; supplied by one of the following (less than 40 parts-per-million total hy­drocarbons). Regulator outlet pressure should be set at 5 psi (35 kPa).
(a) Instrument air - clean, dry. (b) Bottled standard air with step-down
regulator.
(c) Bottled compressed gas mixture
(20.95% oxygen in nitrogen).
(d) Other equivalent clean, dry, oil-free
air supply.
3.12 (79.25) MAX
3
2.250 (57.15)
2.0
(50.80)
1.50
(38.10)
0.25-18 NPT FEMALE INLET CONNECTION
NOTE: DIMENSIONS ARE IN INCHES WITH
2 MOUNTING HOLES
3.19 (81.03) LG
THROUGH BODY FOR
0.312 (7.92) DIA BOLTS
MILLIMETERS IN PARENTHESES.
8.50 MAX (215.90)
10.0 REF
0.250 OR 6 MM OD
TUBE COMPRESSION
FITTING (SUPPLIED BY WECO)
0.250 OR 6 MM OD TUBING (SUPPLIED BY CUSTOMER)
SCHEMATIC HOOKUP FOR REFERENCE AIR SUPPLY ON OXYGEN ANALYZER PROBE HEAD.
(254)
TO PROBE HEAD
1 FLOWMETER 0.2-2.0 SCFH 771B635H02
2 2" PRESSURE GAGE 0-15 PSIG 275431-006
3 COMBINATION FILTER-REG. 0-30 PSIG 4505C21G01
REF AIR SET
263C152G01
COMPRESSED AIR SUPPLY 10-225 PSIG MAX PRESSURE
27270003
Figure 2-3. Air Set, Plant Air Connection
2-8 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
2
Instruction Manual
IB-106-300NH Rev. 4.2
July 2002
2-2 INTELLIGENT FIELD TRANSMITTER (IFT)
INSTALLATION
a. Mechanical Installation
The outline drawing of the IFT module in Figure 2-4 shows mounting centers and clearances. The NEMA 4X enclosure is de­signed to be mounted on a wall or bulk­head. The IFT should be installed no more than 1200 feet (364 m) from the optional HPS or 150 feet (45 m) from the probe if HPS is not installed in the system.
b. Electrical Connections
To meet the Safety Requirements of IEC 1010 (EC requirement), and ensure safe operation of this equipment, con­nection to the main electrical power supply must be made through a circuit breaker (min 10A) which will discon­nect all current carrying conductors during a fault situation. This circuit breaker should also include a me­chanically operated isolating switch. If not, then another external means of disconnecting the supply from the equipment should be located close by. Circuit breakers or switches must comply with a recognized standard such as IEC 947.
NOTE
Refer to Figure 2-7 for fuse locations and specifications.
1. The IFT can be configured for 100, 120, 220, or 240 line voltages. For 120 Vac usage, install JM8, JM7, and JM1 on the power supply board. For 220 Vac usage, install jumpers JM6, JM5, JM2 (refer to Figure 2-5 and Figure 2-7).
2. For installations where the cable run is less than 150 feet (45 m), the IFT can
3.36
(85.3)
11.24 (285.5)
8.00 (203.2)
5.76 (146.3)
9.00 (228.6)
15.00
(381.0)
13.24
(336.3)
11.5 (292.1) MINIMUM DOOR SWING CLEARANCE
1.25
(31.75)
(152.4)
2.25
(57.15)
0.867
(22.00)
16.00
(406.4)
6.0
21190002
0.31
(7.9)
1.62
(41.1)
2.00
(50.8)
NOTE:
DESIGN DIMENSIONS ARE IN INCHES WITH MILLIMETERS IN PARENTHESES.
Figure 2-4. Outline of Intelligent Field Transmitter
be configured to connect directly to a probe. An optional HPS is available for cable runs over 150 feet (45 m). The electrical connections for a non-HPS equipped system should be made as de­scribed in the electrical installation dia­gram, Figure 2-8. Refer to Figure 2-14 for connections for an HPS equipped system.
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-9
Instruction Manual
IB-106-300NH Rev. 4.2 July 2002
Do not install jumper JM6 on the mi­croprocessor board, or JM1 on the in­terconnect board, if an HPS is installed in the system. This will result in system failure.
World Class 3000
if MPS is installed in the system. Refer to Figure 2-8, note 6.
5. The power cable should comply with the safety regulations in the user's country and should not be smaller than 16 gauge, 3 amp.
3. The IFT must have JM6 on the micro­processor board (Figure 2-9 and Figure 2-10) and JM1 on the interconnect board (Figure 2-11 and Figure 2-12) installed if an HPS is not installed in the system.
4. If an MPS is not used in the system, wire jumper between CAL RET and NO GAS must be installed on the in­terconnect board. Remove wire jumper
CONFIGURATION
LINE VOLTAGE
SELECTION
100 V.A.C.
120 V.A.C.
220 V.A.C.
240 V.A.C.
JUMPER
(INSTALL)
JM3, JM7, JM2
JM8, JM7, JM1
JM6, JM5, JM2
JM6, JM5, JM1
ALWAYS DISCONNECT LINE VOLTAGE
JUMPER
REPLACEMENT" PROBE (115V)
FROM INTELLIGENT FIELD TRANSMITTER BEFORE CHANGING JUMPERS.
PROBE HEATER
VOLTAGE SELECTION
WORLD CLASS PROBE (44V)
218 PROBE (115V)
WORLD CLASS "DIRECT
6. Before supplying power to the IFT, ver­ify that the jumpers have been properly set in the IFT (Figure 2-5, Figure 2-9, and Figure 2-11).
7. Terminal strip J5 on the power supply board is used for supplying the IFT with power. Terminal strip J6 on the power supply board is used to supply the probe heater with power if an HPS is not used (Figure 2-7).
JUMPER
(INSTALL)
JM10
JM9
JM9
21190012
If incorrect heater voltage is selected, damage to the probe may occur. For HPS voltage se­lection jumper, refer to Figure 2-15. Always update the relevant labeling to reflect the set voltage.
Figure 2-5. Power Supply Board Jumper Configuration
2-10 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
Instruction Manual
2
IB-106-300NH Rev. 4.2
World Class 3000
NOTE
General Wiring Recommendations
To maintain CE compliance and ensure proper EMC performance, all signal wires to the Inter­connect Board, with the exception of the probe cable, should be looped through the ferrite beads provided as shown in Figure 2-6 (P/N 1L04253H01). Signal wires may be grouped to­gether and looped through before exiting the enclosure. Ferrite beads should be placed as close as possible to the exit point. Ferrite beads are provided for European Common Market applications only.
July 2002
Figure 2-6. Signal Wire Routing
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-11
Instruction Manual
IB-106-300NH Rev. 4.2 July 2002
FUSES
World Class 3000
FUSES
NOTE: ALL FUSES (F1 THROUGH F5)
ARE 5A @ 250 VAC, ANTI­SURGE, CASE SIZE5X20MM, TYPE T TO IEC127, SCHURTER.
Figure 2-7. IFT Power Supply Board Jumpers
2-12 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
2
PROBE JUNCTION
BOX WIRING
HEATER
GN CELL -VE
OR CELL +VE
YE CHROMEL
RD ALUMEL
123456 78
BL
YE
RD
OR
}
GN
BK
BK
BK
GNE
WH
H
R
J1
3D39122G REV
POWER SUPPLYBOARD
ALWAYS DISCONNECT LINE VOLTAGE FROM INTELLIGENT FIELD TRANSMITTER BEFORE CHANGING JUMPERS.
3D39513G
MICROPROCESSOR
BOARD
Instruction Manual
IB-106-300NH Rev. 4.2
CURRENT/VOLTAGE SELECTOR SWITCH
July 2002
JM7
I
V
JM6
PROBE MV -
PROBE MV +
PROBE TC +
PROBE TC -
WORLD CLASS
PROBE
NOTES:
STACK TC WIRING AS REQUIRED.
SPECIAL PROBE CABLE BETWEEN PROBE AND IFT BY ROSEMOUNT.
INSTALL JM1 ON INTERCONNECT BOARD.
INSTALL JM6 ON MICROPROCESSOR BOARD.
IF STACK TEMPERATURE NOT USED.
IF MPS 3000 NOT USED.
1 RELAY PER PROBE AVAILABLE FOR CALIBRATION STATUS INDICATION. (48 V max, 100 mA max)
CURRENT/VOLTAGE SELECTOR SWITCH MUST BE SELECTED TO CURRENT (I) FOR HART COMMUNICATIONS APPLICATIONS.
JUMPER JM7 INFORMATION APPEARS IN FIGURE 2-9.
LINE VOLTAGE SECTION
100 V.A.C. 120 V.A.C. 220 V.A.C. 240 V.A.C.
JUMPER CONFIGURATION
JUMPER
(INSTALL) JM3, JM7, JM2 JM8, JM7, JM1 JM6, JM5, JM2 JM6, JM5, JM1
PROBE HEATER
VOLTAGE SECTION
WORLD CLASS PROBE
218 PROBE
WORLD CLASS "DIRECT REPLACEMENT" PROBE
JUMPER
(INSTALL)
JM10
JM9 JM9
J2 J3 J4
3D39120G REV
INTERCONNECT BOARD
J1
J5
CAL RET
J5 J6
L
H
EN
LINE
VOLTAGE
BK WH
GN PU
OR
BL YE
RD
INTELLIGENT FIELD TRANSMITTER IFT 3000
ER
PROBE 1
LINE OUT LINE IN
J10
L
N
HI GAS
IN CAL
NO GAS
CAL RET
LOW GAS
NO GAS
LO GAS HI GAS IN GAS
SHIELD
PROBE 2 PROBE 3 PROBE 4
IN CAL
HI GAS
NO GAS
CAL RET
LOW GAS
NC C NO NC C NO NC C NO NC C NO
J11
PROBE 1
PROBE 2 PROBE 3 PROBE 4
MPS 3000 MULTIPROBE CALIBRATION GAS SEQUENCER (OPTIONAL)
MPS TERMINATION BOARD
J6
J7
SHIELD STACK TC ­STACK TC +
J8 J9
SHIELD
PROBE TC -
RD YE
PROBE TC +
SHIELD BL OR
5 CONDUCTOR SHIELDED CABLE PER PROBE #16 AWG BY CUSTOMER
IN CAL
HI GAS
CAL RET
PROBE 1
SOLENOID
J13 J14 J15 J16 J17 J18
PROBE MV
PROBE
NO GAS
LOW GAS
PROBE 2
SOLENOID
-
MV+
HI GAS
PROBE 3
SOLENOID
IN CAL
PROBE 4
CAL RET
SOLENOID
HIGH GAS
NO GAS
SOLENOID
LOW GAS
JM1
LOW GAS
SOLENOID
L
N
SWITCH
PRESSURE
J12
L
LINE
E
VOLTAGE
N
27270011
Figure 2-8. Wiring Layout for IFT Systems without HPS
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-13
Instruction Manual
IB-106-300NH Rev. 4.2 July 2002
World Class 3000
OUTPUT JUMPER
HPS Remove JM6 Probe (No HPS) Install JM6
ANALOG OUTPUT (Condition during microcontroller failure) Jumper
Output = zero Install JM7 Output = maximum Remove JM7
(See Figure 2-10 for jumper locations.)
Figure 2-9. Microprocessor Board
Jumper Configuration
c. Analog Output and Relay Output
Connections
1. The microprocessor board has a se­lector for voltage or current operations. Figure 2-10 shows switch orientation. In voltage mode, output is 0-10 V. In the current mode, the output can be configured from the SETUP menu to be 0-20 mA or 4-20 mA.
2. The analog output and relay outputs are programmed by the user as needed. The analog output is typically sent to recording equipment such as
chart recorders. Relay outputs are typi­cally sent to annunciators.
3. Relays K1 and K2 are user configur­able from the probe SETUP sub-menu (Table 5-5). Typically these are used to indicate O specified tolerances. OK relay is ener­gized when unit is functioning properly.
4. All wiring must conform to local and national codes.
5. Connect the analog output and relay outputs as shown in Figure 2-12.

2-3 HEATER POWER SUPPLY INSTALLATION

a. Mechanical Installation
The outline drawing of the heater power supply enclosure in Figure 2-13 shows mounting centers and clearances. The NEMA 4X enclosure is designed to be mounted on a wall or bulkhead. The heater power supply should be installed no further than 150 feet (45 m) from the probe. The heater power supply must be located in a location free from significant ambient tem­perature changes and electrical noise. Am­bient temperature must be between -20° and 140°F (-30° and 60°C).
values above or below
2
2-14 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
2
Instruction Manual
IB-106-300NH Rev. 4.2
July 2002
JM7
JM7
CURRENT/VOLTAGE SELECTOR SWITCH
TO I/O
BOARD
J4
JM6
SW1
JM6
TO POWER
SUPPLYCARD
TP5 TP6
TP4
+30VISO-C GNDC +5VISO-C
TP3
-15V
TP8
+5V
TP2
TP7
+15V
TP1
-5V
J1
J2
3D39513G REV
TO LDP CARD
J3
TO GUI CARD
29850004
Figure 2-10. IFT Microprocessor Board
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-15
Instruction Manual
IB-106-300NH Rev. 4.2 July 2002
NOTES:
DENOTES SHIELD CONNECTION.
OK RELAY IS ENERGIZED WHEN UNIT IS FUNCTIONING PROPERLY.
World Class 3000
OUTPUT JUMPER
HPS Remove JM1 Probe (No HPS) Install JM1
Figure 2-11. Interconnect Board Jumper Configuration
24
OK-COM OK-NO
23 22
K1-COM K1-NO
21
K2-COM
20 19
K2-NO
18 17 16 15 14 13 12
ANOUT-
11
ANOUT+
10
9
STACK T/C
8
STACK T/C
7 6
PROBE T/C
5 4
PROBE T/C
3 2
PROBE MV
1
PROBE MV+
OK-NC
K1-NC
K2-NC
CALINIT-2
CAL INIT-1
CALRET
NOGAS
LOGAS
HIGAS
INCAL
RELAY
RELAY+
AD590
AD590+
TRIAC
TRIAC+
24 23 22
21 20 19 18 17 16 15 14 13 12
11 10
9
8
-
7
6
5
-
4
3
2
-
1
-
JM1 (UNDER SHIELD)
16860010
Figure 2-12. IFT Interconnect Board Output Connections
2-16 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
2
Instruction Manual
IB-106-300NH Rev. 4.2
July 2002
4.00
(101.6)
3.25
(82.6)
8.50
(215.9)
0.31 (7.9)
#0.31
6.00
(152.4)
4.88
(124)
6.75
(171.5)
NOTE: DIMENSIONS IN INCHES
7.00
(177.8)
#10-32 UNF 2A THREADED INSERT (0.31 x 0.31 FROM CORNER OF PLATE)
WITH MILLIMETERS IN PARENTHESES.
3.63
8.00
(203.2)
(92.2)
11.00
(279.4)
NEMA 4X
(NON-HAZARDOUS)
0.13" (3.3) THK U. L. APPROVED GASKET
1.81 (46)
4.38
(111.3)
10.39 (264)
9.96
(253)
0.38 (9.7)
1.00 (25.4) MINIMUM CLEARANCE
4.72
(120)
CLASS 1, DIVISION 1, GROUP B ENCLOSURE
FOR REMOVING COVER
9.17
(233)
8.50
(215.9)
6.18
(156.9)
0.56 (14) DIA (2) MOUNTING HOLES
686029
Figure 2-13. Outline of Heater Power Supply
b. Electrical Connections
1. Electrical connections should be made as described in the electrical installa­tion diagram, Figure 2-14. The wiring terminals are divided into two layers; the bottom (FROM PROBE) terminals should be connected first, the top (FROM ELECTRONICS) terminals should be connected last (Figure 2-15). Each terminal strip has a protective cover which must be removed when making connections. To remove the terminal covers, remove two slotted screws holding cover in place. Always reinstall terminal covers after making connections. All wiring must conform to local and national codes.
NOTE
Refer to Figure 2-17 for fuse locations and specifications.
2. Power Input: 120, 220 or 240 Vac. For 120 Vac usage, install jumpers JM4 and JM1. For 220 or 240 Vac usage, install jumper JM5 (see label, Figure 2-16).
NOTE
For 100 Vac usage, the heater power supply is factory-supplied with a dif­ferent transformer. When using the HPS with 100 Vac transformer, install jumpers JM1 and JM4.
3. The power cable should comply with safety regulations in the user's country and should not be smaller than 16 gauge, 3 amp.
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-17
Instruction Manual
IB-106-300NH Rev. 4.2 July 2002
PROBE JUNCTION
BOX WIRING
GN CELL -VE
OR CELL +VE
YE CHROMEL
RD ALUMEL
123456 78
BL
YE
RD
OR
PROBE MV -
PROBE MV +
PROBE TC +
PROBE TC -
WORLD CLASS PROBE
HEATER
}
GN
BK
BK
BK
GNE
WH
H
R
LINE VOLTAGE
SELECTION
120 V.A.C./100 V.A.C.
11
220/240 V.A.C.
PROBE HEATER
VOLTAGE SECTION
WORLD CLASS PROBE JM7
TOP
TRIAC RELAY
J9
+
+++
-
CONFIGURATION
JUMPER (INSTALL)
JM4, JM1
JM5
JUMPER (INSTALL)
2
1
STACKTCANALOG
---
World Class 3000
NOTES
RELAY WIRE IS OPTIONAL, RELAY CAN BE BYPASSED
1
WITH JUMPER JM-2 IF NOT WIRED TO THE IFT.
STACK TC WIRING AS REQUIRED.
2
ALL WIRES #16-#22 AWG TWISTED PAIR WITH SHIELD
3
BY CUSTOMER EXCEPT AS NOTED.
STANDARD PROBE CABLE BETWEEN PROBE AND
4
HPS BY ROSEMOUNT.
REMOVE JM1 ON INTERCONNECT BOARD.
5
REMOVE JM6 ON MICROPROCESSOR BOARD.
6
IF RELAY WIRE OF NOTE 1 INSTALLED THEN
7
REMOVE JM2 ON HPS 3000.
8
IF STACK TEMPERATURE NOT USED.
9
IF MPS 3000 NOT USED.
1 RELAY PER PROBE AVAILABLE FOR CALIBRATION
10
STATUS INDICATION (48 V max, 100 mA max).
100 V.A.C. REQUIRES TRANSFORMER PART
11
NUMBER 1M02961G02.
JUMPER JM7 INFORMATION APPEARS IN FIGURE 2-9.
ALWAYS DISCONNECT LINE VOLTAGE
JUMPER
FROM HEATER POWER SUPPLY BEFORE CHANGING JUMPERS.
HEATER POWER
REMOTE
ON
ELECTRONICS
SELECTION
(NEXT GENERATION)
HEATER
BK
WH
ON
REMOVE JM2
INSTALL JM2
JUMPER
REMOVE JM3, JM6DIGITAL
7
1
2 TWISTED PAIR SHIELDED #22 AWG BY CUSTOMER (OPTIONAL)
A
B
22540001
4
OR CELL+ BL CELL -
YE HTR TC +
RD HTR TC
WH BK GN
-
GN/YE
(INTERNAL
AD590
+
-
WIRING)
-
PROBE
PROBE
TC
MV
J8
+
-
BOTTOM
PROBE
J3
+
STACK
2
J2 J1
+
MV
TC
PROBE
TC
+
-
PROBE
HEATER
RH
-
HPS 3000 INTERFACE MODULE
WHBK
4 TWISTED PAIR SHIELDED #22 AWG BY CUSTOMER
LINE
L
N
LINE VOLTAGE
Figure 2-14. Wiring Layout for Complete IFT 3000 System with HPS (Sheet 1of 2)
2-18 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
2
Instruction Manual
IB-106-300NH Rev. 4.2
CURRENT/VOLTAGE SELECTOR SWITCH
July 2002
LINE
VOLTAGE
SECTION 100 V.A.C. 120 V.A.C. 200 V.A.C. 220 V.A.C. 240 V.A.C.
JUMPER
(INSTALL) JM3, JM7, JM2 JM8, JM7, JM1 JM4, JM5, JM2 JM6, JM5, JM2 JM6, JM5, JM1
J5 J6
L
EN
LINE
VOLTAGE
J1
POWER SUPPLYBOARD
ALWAYS DISCONNECT LINE VOLTAGE FROM INTELLIGENT FIELD TRANSMITTER BEFORE CHANGING JUMPERS.
JUMPER CONFIGURATION
PROBE HEATER
VOLTAGE SECTION
NOT USED REMOVE
NOT USED
3D39122G REV
(INSTALL)
JUMPER
JM9, JM10
3D39513G
MICROPROCESSOR
BOARD
CAL RET NO GAS
LO GAS HI GAS IN CAL
SHIELD RELAY – RELAY + SHIELD AD590 –
AD590 + SHIELD TRIAC –
TRIAC +
J2 J3
J4
J5 J6 J7 J8 J9
SHIELD STACK TC – STACK TC +
SHIELD
PROBE TC – PROBE TC +
SHIELD
PROBE MV – PROBE MV +
J1
3D39120G REV
INTERCONNECT BOARD
JM7
I
V
JM6
JM1
A
B
INTELLIGENT FIELD TRANSMITTER IFT 3000
PROBE 1
LINE OUT LINE IN
J10
L
N
HI GAS
IN CAL
NO GAS
CAL RET
NC C NO NC C NO NC C NO NC C NO
J11
PROBE 1
PROBE 2 PROBE 3 PROBE 4
MPS 3000 MULTIPROBE CALIBRATION GAS SEQUENCER (OPTIONAL)
PROBE 2 PROBE 3 PROBE 4
IN CAL
HI GAS
LOW GAS
CAL RET
MPS TERMINATION BOARD
5 CONDUCTOR SHIELDED CABLE PER PROBE #16 AWG BY CUSTOMER
NO GAS
LOW GAS
IN CAL
HI GAS
NO GAS
CAL RET
PROBE 2
PROBE 1
SOLENOID
J13 J14 J15 J16 J17 J18
LOW GAS
SOLENOID
PROBE 3
SOLENOID
HI GAS
IN CAL
PROBE 4
SOLENOID
NO GAS
CAL RET
LOW GAS
LOW GAS
SOLENOID
HIGH GAS
SOLENOID
L
N
SWITCH
PRESSURE
J12
L
LINE
E
VOLTAGE
N
27270012
Figure 2-14. Wiring Layout for Complete IFT 3000 System with HPS (Sheet 2 of 2)
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-19
Instruction Manual
IB-106-300NH Rev. 4.2 July 2002
SCREW
(2 PER COVER)
World Class 3000
TRANSFORMER
TERMINAL
COVERS
FRONT
TERMINAL STRIP
(FROM ELECTRONICS)
TRANSFORMER
(FROM ELECTRONICS)
SIDE
Figure 2-15. Heater Power Supply Wiring Connections
TERMINAL STRIP
29850005
2-20 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
2
1
2
NOTES:
1
100 V.A.C. OPERATION REQUIRES TRANSFORMER PART NUMBER 1M02961G02.
2
REFER TO TABLE 3-5 FOR PROPER SET POINT SELECTION.
Instruction Manual
IB-106-300NH Rev. 4.2
July 2002
0310122
Figure 2-16. Jumper Selection Label
NOTE
Before supplying power to the heater power supply, verify that jumpers JM3, JM6 are removed and JM7 is installed. If relay wire (Figure 2-14, note 1) is in­stalled, JM2 must be removed from HPS Mother Board (Figure 2-17).
4. Before supplying power to the heater power supply, verify that the jumpers on the mother board, Figure 2-17, are properly configured. Jumpers JM3 and JM6 should be removed and JM7 should be installed.
Additionally, make sure that the proper jumper for your line voltage is ins tal led , Figure 2-16. If relay wire (Figure 2-14, note 1) is not installed, JM 2 should be installed on the HPS Mother Board (Figure 2-17).
NOTE
Refer to Figure 2-9 and Figure 2-11 for proper IFT jumper configuration. IFT microprocessor and interconnect board jumper configurations must be set correctly in order for HPS to work properly.
FUSES
JM1
JM2 JM4
JM7
JM5
3D3 080G REV
FUSE
NOTE: ALL FUSES ARE 5A @ 250 VAC,
ANTI-SURGE, CASE SIZE 5 X 20 MM, TYPE T TO IEC127, SCHURTER.
Figure 2-17. Jumpers on HPS Mother Board
29850001
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-21
Instruction Manual
IB-106-300NH Rev. 4.2 July 2002
World Class 3000
2-4 MULTIPROBE CALIBRATION GAS
SEQUENCER INSTALLATION a. Mechanical Installation
The outline drawing of the MPS module in Figure 2-18 shows mounting centers and clearances. The box is designed to be mounted on a wall or bulkhead. The MPS module should be installed no further than 300 feet (91 mz) piping distance from the probe, and no more than 1000 feet (303 m) cabling distance from the IFT. Install the MPS module in a location where the ambi­ent temperature is between -20° and 160°F (-30° and 71°C).
12.00
(304.80)
10.00
(254.00)
b. Gas Connections
Figure 2-19 shows the bottom of the MPS where the gas connections are made. 1/4 inch threaded fittings are used.
1. Connect the reference air supply to INSTR. AIR IN. The air pressure regu­lator valve is set at the factory to 20 psi (138 kPa). If the reference air pressure should need readjustment, turn the knob on the top of the valve until the desired pressure is obtained.
2. Connect the high O
calibration gas to
2
HIGH GAS. The calibration gas pres­sure should be set at 20 psi (138 kPa).
12.00
(304.80)
NOTE: DIMENSIONS ARE IN INCHES
WITH MILLIMETERS IN PARENTHESES.
HIGHCAL
GASIN
PROBE1 PROBE 2 PROBE 3 PROBE4
LOWCAL
CALGAS
GASIN
REFAIR
INSTR
AIR
CALGAS
CALGAS
OUT
REFAIR
OUT
CALGAS
OUT
OUT
OUT
REFAIR
REFAIR
OUT
OUT
OUT
0.84 (21.34)
1.96 (49.78)
3.09 (78.49)
4.21 (106.93)
5.25 (133.35)
5.54 (140.72)
14.00 (355.60) REF
27270013
Figure 2-18. MPS Module
2-22 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
2
Instruction Manual
IB-106-300NH Rev. 4.2
July 2002
3. Connect the low O2 calibration gas to LOW GAS. The calibration gas pres­sure should be set at 20 psi (138 kPa).
4. Connect the REF AIR OUT to the ref­erence air fitting on the probe junction box.
5. Connect the CAL GAS OUT to the calibration gas fitting on the probe junction box.
6. If the MPS is configured for multiple probes (up to four), repeat steps 4 and 5 for each additional probe.
A check valve is required for each probe connected to an MPS to prevent condensation of flue gas in the cali­bration gas lines. The check valve must be located between the calibra­tion fitting and the gas line.
c. Electrical Connections
Electrical connections should be made as described in the electrical installation dia­gram, Figure 2-20. All wiring must conform to local and national codes. The electrical connections will exist only between the electronics package and the MPS to enable automatic and semiautomatic calibration. If
more than one probe system is being used, the additional probes and electric packages would be wired similar to the first probe.
NOTE
Refer to Figure 2-20 for fuse locations and specifications.
1. Run the line voltage through the bulk­head fitting on the bottom of the MPS where marked LINE IN, Figure 2-19. Connect the line voltage as shown in Figure 2-20 to the LINE IN terminal on the MPS termination board located in­side the unit. Tighten the cord grips to provide strain relief.
2. The MPS can accommodate up to four probes. The terminal strips on the MPS termination board are marked PROBE 1, PROBE 2, PROBE 3, and PROBE 4. Select PROBE 1 if this is the first probe and electronic package installed on the MPS.
3. Make the connections from the MPS to the IFT as shown in Figure 2-20. Run wires from the MPS Termination Board inside the unit through the bulkhead fit­ting on the bottom of the unit where marked SIGNAL IN, Figure 2-19. After the connections are made, tighten the cord grips to provide strain relief.
LINE IN
PROBE 1 PROBE 2 PROBE 3 PROBE4
LOW CAL
CAL GAS
CAL GAS
CAL GAS
OUT
REF AIR
OUT
CAL GAS
OUT
REF AIR
OUT
SIGNAL IN
27270014
DRAIN
HIGH CAL
GAS IN
INSTR
GAS IN
OUT
OUT
REF AIR
REF AIR
OUT
OUT
AIR
Figure 2-19. MPS Gas Connections
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-23
Instruction Manual
IB-106-300NH Rev. 4.2 July 2002
NOTE: FUSES FOR 115 VOLT MPS UNIT ARE
FAST ACTING, 1A @ 250 VAC, SIZE: 1/4 IN. DIA X 1-1/4 IN. LG., GLASS BODY, NON-TIME DELAY, BUSSMAN PART NO. BK/AGC-1 (ROSEMOUNT APRT NO. 138799-004).
FUSES FOR 220 VOLT MPS UNIT ARE FAST ACTING, 0.5 A @ 250 VAC, SIZE 1/4 IN. DIA. X 1-1/4 IN. LG., GLASS BODY, NON-TIME DELAY, BUSSMAN PART NO. BK/AGC-1/2 (ROSEMOUNT PART NO. 138799-014).
INTERCONNECT
IFT
BOARD
CAL RET NO GAS LO GAS HI GAS IN CAL
MH1
World Class 3000
J1
J2 J3 J4 J5 J6 J7 J8 J9
MH2
3D39120G REV
MH3
MH4
PROBE 1 PROBE 2 PROBE 3 PROBE 4
LINE OUT LINE IN
L
IN CAL
HI GAS
N
CAL RET
NO GAS
LOW GAS
HI GAS
IN CAL
CAL RET
J10
NC C NO NC C NO NC C NO NC C NO
J11
PROBE 1
PROBE 2 PROBE 3 PROBE 4
NO GAS
LOW GAS
IN CAL
HI GAS
NO GAS
CAL RET
LOW GAS
PROBE 2
SOLENOID
PROBE 1
SOLENOID
J13 J14 J15 J16 J17 J18
PROBE 3
SOLENOID
HI GAS
IN CAL
PROBE 4
CAL RET
SOLENOID
NO GAS
HIGH GAS
SOLENOID
LOW GAS
LOW GAS
SOLENOID
L
N
SWITCH
PRESSURE
J12
MPS TERMINATION BOARD
FUSES LOCATED BEHIND
TERMINATION BOARD
Figure 2-20. MPS Probe Wiring
L
E
N
LINE VOLTAGE
29850006
2-24 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
Instruction Manual
2
IB-106-300NH Rev. 4.2
World Class 3000
NOTE
Upon completing installation, make sure that the probe is turned on and operating prior to firing up the combustion process. Damage can result from having a cold probe exposed to the process gases.
Power down all probes during outages. Sensor chamber is heated to 736°C. Further, if ducts will be washed down during the outage, remove the probes to prevent water damage.
July 2002
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-25
Instruction Manual
IB-106-300NH Rev. 4.2 July 2002
World Class 3000
2-26 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
3
Instruction Manual
IB-106-300NH Rev. 4.2
July 2002
SECTION 3

SETUP

3-1 OVERVIEW
This section provides information on configuring the IFT 3000 Intelligent Field Transmitter. This section assumes that you are familiar with the operation of the IFT and the General User In­terface (GUI). If you need additional information on operating the IFT or using the GUI, refer to Section 5, General User Interface (GUI) Operation.

3-2 CONFIGURING THE ANALOG OUTPUT

Use the following procedure to configure the analog output.
a. Press the SETUP key on the GUI keypad. b. Set the Source to the desired measurement
value to be represented by the analog out­put. The choices are O Rng O
.
2
, Efficiency, or Dual
2
c. Set the Type to the desired output signal
style. The choices are HART 4-20mA, 0­20mA, and 0-10V. The choice selected must agree with the position of the current/voltage selector switch on the IFT microprocessor board. An in va li d cho ice wil l be discarded. Note that if you are using HART to communicate with the IFT, you must set the analog output type to HART 4­20mA.
d. The next choice, Range Setup, will vary
based on the source selected.
3. Source Set to Dual Rng O
. Range
2
setup allows you to set the transfer function (Xfer Fnct) to either linear or log output. You can also specify the O values represented by the high and low analog output values for both the nor­mal and high range.
The Mode Setup sub-menu contains entries for setting the range mode, whether the high range is used during calibration, and the point at which the output switches from normal to high range.
For a complete description of all pa­rameters associated with configuring the analog output, refer to Table 5-5.

3-3 SETTING CALIBRATION PARAMETERS

To successfully calibrate a World Class 3000 system, several calibration paramete rs must be set. These parameters are generally set once and left at those values. These values should only be changed if the system is not calibrating properly, or when changing test gas bottles.
a. Press the SETUP key on the GUI keypad. b. Select the Calibration sub-menu. c. Set the High Gas parameter to the oxygen
concentration of the high calibration gas. For high calibration gas, 8% oxygen with a balance of nitrogen is recommended.
2
1. Source set to Efficiency. No range setup is allowed when the source is set to efficiency. Analog output range is fixed at 0-100% efficiency.
2. Source set to O
. Range setup allows
2
you to set the transfer function (Xfer Fnct) to either linear or log output. You can also specify the O
values repre-
2
sented by the high and low analog out­put values.
Rosemount Analytical Inc. A Division of Emerson Process Management Setup 3-1
d. Set the Low Gas parameter to the oxygen
concentration of the low calibration gas. For low calibration gas, 0.4% oxygen with a balance of nitrogen is recommended.
e. The Auto Cal parameter determines
whether the IFT performs automatic or semiautomatic calibrations. In order to per­form automatic calibration, the system must be equipped with an MPS 3000 Multiprob e Calibration Gas Sequencer. To perform
Instruction Manual
IB-106-300NH Rev. 4.2 July 2002
World Class 3000
automatic calibrations, set the Auto Cal pa­rameter to Yes.
f. The Output Tracks setting determines
whether the analog output tracks the oxy­gen reading during a calibration. Setting Output Tracks to No locks the analog output value to the last measured oxygen reading until the calibration is complete.
g. The Cal Interval parameter sets the time in
hours and days between automatic calibra­tions. When Auto Cal is set to NO, this pa­rameter is set to OFF.
h. The Next Cal parameter displays the time
until the next scheduled automatic calibra­tion. If automatic calibration is not enabled, this parameter displays Disabled.
i. The Gas Time parameter sets the amount
of time that calibration gas flows during an automatic calibration before a reading is taken. This value is not used for semiauto­matic calibrations.
alarms, respectively. The Alarm DB parameter allows the setting of an alarm dead band. When a dead band is set, the O
value must change
2
by the dead band value before the alarm will re­set. For example, if the Hi Alarm is set to 8.00% and the dead band is set to 0.25%, the O centration must drop to below 7.75% before the
alarm will clear. This prevents the alarms
O
2
from continually activating and clearing when the oxygen value is near the alarm setpoint.
3-5 CONFIGURING EFFICIENCY
CALCULATIONS
To enable efficiency calculations and set the ef­ficiency constants, press the SETUP key on GUI keypad, and select the Efficiency Calc sub­menu. The Enable Calc selection turns effi­ciency calculation on and off. Enter the K1, K2, and L3 constant values in the appropria te fie lds. Efficiency constant values are listed in Table 5-6 for oil and gas for the US and Europe.

3-6 CONFIGURING THE RELAY OUTPUTS

con-
2
j. The Purge Time parameter sets the amount
of time after an automatic calibration before the system is returned to normal operation. This allows time for the calibration gases to clear the lines and the system to return to the process gas concentration. This value is not used for semiautomatic calibrations.
k. The Res Alarm parameter displays the set-
point for the Res Hi alarm. Do not change this parameter unless directed by a qualified Rosemount Service Engineer.
l. Press the ESC key twice to return to the
Main menu.
3-4 S ETTING THE O
The IFT has a high and low O
ALARM SETPOINTS
2
alarm. To
2
change the alarm setpoints, press the SETUP key on the GUI keypad and select the O Alarms sub-menu.
The Hi Alarm and Lo Alarm values are the set­tings for the high and low oxygen concentration
The IFT has two relays that can be individually configured. Each relay can be triggered by three separate events selected from a list of eight events. Use the following procedure to configure the relay outputs.
a. Press the SETUP key on the GUI keypad.
Select the Relay Outputs sub-menu.
b. Select K1 Setup or K2 Setup to configure
relay one or relay two, respectively.
c. Set Event 1, Event 2, and Event 3 to the
desired triggering event. The relay will be energized when any of the three events oc­curs. If you do not want a relay to trigger on three events, set the desired trigger or trig­gers and set the remaining events to Off.
Note that the TG Low event will only func­tion if the system includes an MPS 3000 Multiprobe Test Gas Sequencer.
2
d. Press the ESC key and select the other re-
lay. Configure the relay as described above.
e. Press the ESC key three times to return to
the Main menu.
3-2 Setup Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
4
Instruction Manual
IB-106-300NH Rev. 4.2
July 2002
SECTION 4

CALIBRATION

4-1 ANALOG OUTPUT CALIBRATION

For the analog output to perform within the pub­lished specifications, it must be manually cali­brated. The only equipment needed to perform the calibration is a voltage or current meter, de­pending on which mode of operation is to be calibrated. Prior to manual calibration, remove the IFT from any control loops it may be in.
Prior to manual calibration, the IFT should be removed from any auto­matic control loops. Failure to remove the IFT from control loops prior to calibration may result in faulty equip­ment performance.
Once initiated from the Setup - Analog Outputs menu, the calibration procedure is self guiding.

4-2 SYSTEM CALIBRATION

a. Overview
The primary purpose of an oxygen analyzer is to give an accurate representation of the percentage of O system should be calibrated periodically to maintain an accuracy which m ay otherwise be reduced over time due to cell aging. A calibration record sheet is provided at the end of this section to track cell performance.
A requirement for calibration is a set of two accurate calibration gases spanning the oxygen range of most interest. For example,
0.4% and 8% for a 0-10% oxygen range. Under normal conditions the probe should
not need frequent calibration. Because cali­bration is necessary, the system can be equipped with the optional MPS 300 0 Multi­probe Calibration Gas Sequencer for fully automatic calibration at regular intervals. Without an MPS, the probes must be cali­brated manually (semiautomatically).
in the gas stream. The
2
b. Probe Calibration
1. Previous Calibration Constants
Functionality
There are three sets of registers used to store calibration constants. These are: Latest Calibration, Previous Cali­bration, and Calculatio n. Onl y the val­ues in the Calculation register are used to calculate the oxygen value for dis­play and representation on the analog output signal. These values may be changed in two ways.
(a) The operator may change the val-
ues through the SETUP menu. The operator may adjust the slope and constant individually, or reset both to the values calculated during the last good calibration. To reset the values, move the cursor to RESET SLOPE & CONST and push ENTER.
(b) The IFT will automatically change
the values after each calibration as follows:
The values in the Latest Calibra­tion registers are updated after every complete calibration, even if the calibration is not successful. If the calibration is successful, the values in the Latest Calibration registers are copied into the Previ­ous Calibration registers. This is accomplished prior to the update of the Latest Calibration registers. The new slope and constant are copied into the Calculation register.
If the calibration fails, the Previous Calibration registers retain their existing values, while the Latest Calibration registers record the values of the failed calibration. The Calculation register is not updated when the calibration fails.
Rosemount Analytical Inc. A Division of Emerson Process Management Calibration 4-1
Instruction Manual
IB-106-300NH Rev. 4.2 July 2002
World Class 3000
2. Calibration Methods
There are three calibration methods: manual (semiautomatic), manually ini­tiated automatic, and fully automatic. Manual (semiautomatic) calibration is done without an MPS unit. Calibration gases are switched on and off by the operator and the IFT is sequenced through the calibration procedure by the operator with the front panel key­board. The IFT prompts the operator for the correct action. Manually initiated automatic calibration is done with an MPS. The operator manually initiates the calibration at the IFT or through a remote switch, and the IFT controls the operation of the MPS unit and the cali­bration sequencing. Fully automatic calibration requires no action from the operator. The setup is the same as semiautomatic except the IFT auto­matically initiates the calibration at a fixed calibration interval. In this mode the operator can also manually initiate calibrations between the intervals in the same manner as semiautomatic calibrations.
c. Manual (Semiautomatic) Calibration
1. Calibration Gases For Manual (Semiautomatic) Calibration
There are two options for supplying calibration gases to the probe during semiautomatic calibration. The first "A" uses refillable bottles and adjustable 2­stage pressure regulators; the second "B" uses disposable bottles and a fixed single stage regulator to provide a mixed flow. Normally, the first (method "A") will have a higher cost and not be portable. The second ("B") is less costly, portable, and weighs about 10 lbs (4.5 kg).
Test Method "A" Fixed Tanks and Manifolds
(a) Required Equipment
Do not use 100% nitrogen as a zero gas. It is suggested that gas for the zero be between 0.4% and 2.0% O2. Do not use gases with hydrocarbon con­centrations of more than 40 parts per million. Failure to use proper gases will result in erroneous readings.
NOTE
Ambient air is not recommended for use as high calibration gas. An 8% O balance in nitrogen is recommended for high calibration gas.
1 Two tanks of precision cali-
bration gas mixtures. Rec­ommended calibration gases are nominally 0.4% and 8.0% oxygen in nitrogen.
Two sources of calibrated gas mixtures are:
LIQUID CARBONIC GAS CORP. SPECIALTY GAS LABORATORIES
700 South Alameda Street Los Angeles, California 90058 213/585-2154
767 Industrial Road San Carlos, California 94070 415/592-7303
9950 Chemical Road Pasadena, Texas 77507 713/474-4141
12054 S.W. Doty Avenue Chicago, Illinois 60628 312/568-8840
2
603 Bergen Street Harrison, New Jersey 07029 201/485-1995
4-2 Calibration Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
4
Instruction Manual
IB-106-300NH Rev. 4.2
July 2002
255 Brimley Road Scarborough, Ontario, Canada 416/266-3161
SCOTT ENVIRONMENTAL TECHNOLOGY, INC. SCOTT SPECIALTY GASES
2600 Cajon Blvd. San Bernardino, CA 92411 714/887-2571 TWX: 910-390-1159
1290 Combermere Street Troy, MI 48084 314/589-2950
Route 611 Plumsteadville, PA 18949 215/766-8861 TWX: 510-665-9344
2616 South Loop, West Suite 100 Houston, TX 77054 713/669-0469
2 If gas bottles will be perma-
nently hooked up to the probe, a manual block valve is required at the probe (be­tween the calibration fitting and the gas line) to prevent the migration of process gases down the calibration gas line.
If an MPS 3000 Multiprobe Gas Sequencer is used, a check valve is required at the probe.
3 Two, 2-stage pressure regu-
lators with stainless steel dia­phragms for tanks. Maximum output required: 20 psi (138 kPa).
4 One instrument air pressure
regulator: 20 psi (138 kPa)
maximum and a supply of clean, dry instrument air.
5 Two zero-leakage shutoff
valves.
6 Miscellaneous oil-free tubing
and fittings.
(b) Calibration
1 A typical calibration setup is
shown in Figure 4-1. Care must be taken that all fittings are tight and free from oil or other organic contaminants. Small openings can cause back diffusion of oxygen from the atmosphere even though positive pressures are main­tained in the lines.
NOTE
The probe calibration gas fitting has a seal cap which must be in place at all times except during calibration.
In addition to the precision calibration gas mixtures, clean, dry, oil-free instrument air should be used for calibration.
For optimum accuracy, this calibration should be run with the process at normal temperature and operating conditions.
When the calibration gas line exceeds 6 ft (1.8 m) in length from the leak tight valves, check valve, Rosemount P/N 6292A97H02, should be in­stalled next to the calibration gas connection on the probe to prevent breathing of the line with the process gas and subsequent gas condensation and corrosion.
Rosemount Analytical Inc. A Division of Emerson Process Management Calibration 4-3
Instruction Manual
IB-106-300NH Rev. 4.2 July 2002
CALIBRATE
IN-PLACE
FITTING
5 SCFH
CHECK VALVE
PROBE (END VIEW)
REFERENCE AIR
CONNECTION
World Class 3000
2 SCFH
FLOW METER
LEAK TIGHT
VALVES
REG
0.4% O
2
PROBE CALIBRATION GAS FITTING
NOTE:
HAS A SEAL CAP WHICH MUST BE IN PLACE AT ALL TIMES EXCEPT DURING CALIBRATION.
Figure 4-1. Typical Calibration Setup
NOTE
Only set the calibration gas flowmeter upon initial installation and after changing the diffusion element. A slightly lower calibration gas flow rate may indicate a plugged diffusion element.
2 Set the calibration gas pres-
sure regulators and the flow meter for a flow of 5 SCFH at 20 psi (138 kPa) for both gases. The reference air should be flowing as in normal operation.
3 Refer to paragraph 4-2d of
this section for Manual (Semi­automatic) Calibration setup and procedure using the IFT.
4 Calibration gases will be
switched on and off using the shutoff valves.
8.0% O
REFERENCE
AIR
SET
INSTR. AIR IN
2
Test Method "B" Rosemount Oxygen Calibration Gas and Service Kit.
(a) Required Equipment
Do not use 100% nitrogen as a zero gas. It is suggested that gas for the zero be between 0.4% and 2.0% O2. Do not use gases with hydrocarbon con­centrations of more than 40 parts per million. Failure to use proper gases will result in erroneous readings.
NOTE
Ambient air is not recommended for use as high calibration gas. An 8% O balance in nitrogen is recommended for high calibration gas.
1 Portable Rosemount Oxygen
Calibration Gas Kits (Figure 4-2), Rosemount P/N 6296A27G01, containing 8% and 0.4% gases in a portable carrying case with regulator, built-in valve, hose and con­necting adapter to the calibra­tion gas connection.
2
4-4 Calibration Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
4
CALIBRATION
GAS KIT #1
(P/N 6296A27G01)
Instruction Manual
IB-106-300NH Rev. 4.2
July 2002
Rosemount France 165 Boulevard de Vallmy 92706, Colombes, France
Rosemount P/N 3530B07G01 for probe 0.4% oxygen in ni­trogen in disposable bottle.
Rosemount P/N 3530B07G02 for probe 8% oxygen in nitro­gen in disposable bottle.
3 A check valve is required at
the probe (between the cali­bration fitting and the gas line) to prevent the migration of process gases down the cali­bration gas line.
(b) Calibrat ion with a Port able R ose-
mount Oxygen Calibration Gases Kit.
Figure 4-2. Portable Rosemount Oxygen
Calibration Gas Kit
2 Extra gas bottles are available
at: Rosemount Analytical Inc.
Box 901 Orrville, Ohio 44667 U.S.A.
Rosemount Limited Burymead Road Hitchin, Herts. U.K.
Rosemount Italy VIA Guido Cavalcanti 8 20127 Milan, Italy
Rosemount Spain Saturnino Calleja 6 28002 Madrid Spain
27270007
1 A typical portable calibration
setup is shown in Figure 4-3. For manual (semiautomatic) calibration, remove cap plug from the calibrate in place fit­ting. The cap plug must be retained to seal this fitting af­ter calibration is complete; failure to do so may render the probe useless if the sys­tem pressure is slightly nega­tive. The reference air should be flowing as in normal operation.
2 Refer to paragraph 4-2.d of
this section for Manual (Semi­automatic) Calibration setup and procedure using the IFT.
3 Screw the pushbutton regu-
lator with contents gage on to the calibration gas of choice and inject the calibration gas by opening the valve. Gas is on continuously when the valve is opened.
Rosemount Analytical Inc. A Division of Emerson Process Management Calibration 4-5
Instruction Manual
IB-106-300NH Rev. 4.2 July 2002
CALIBRATE
IN PLACE
CONNECTION
GAGE - SET 5 SCFH
NOTE:
PROBE CALIBRATION GAS FITTING HAS A SEAL CAP WHICH MUST BE IN PLACE EXCEPT DURING CALIBRATION.
CHECK
VALV E
PUSHBUTTON
REGULATOR
WITH CONTENTS
Figure 4-3. Typical Portable Calibration Setup
d. Manual (Semiautomatic) Calibration
Procedure
The following procedure relates to an op­erator initiated calibration selected at the IFT by pressing the CAL key. The calibra­tion is manually performed by the operator upon data queues from the IFT. Any system without an MPS 3000 Multiprobe Calibration Gas Sequencer must follow these steps.
1. Press SETUP to display the SETUP menu. Select PROBE CALIBRATION sub-menu. Ensure that Auto Cal is dis­abled. Set the cursor on Auto Cal. Press ENTER. Set Auto Cal to NO if not already done.
REFERENCE AIR CONNECTION
CALIBRATION GAS HOSE CONNECTS TO CHECK VALV E
0.4 %
O
2
8.0 % O
2
27270005
World Class 3000
ENTER to start. Follow the data queues. Refer to Table 5-4, CALI­BRATE O
e. Fully Automatic Calibration
1. Calibration Gases For Fully Automatic Calibration. For fully automatic calibra­tion, an MPS 3000 Multiprobe Calibra­tion Gas Sequencer is required as well as the two types of calibration gas.
Do not use 100% nitrogen as a zero gas. It is suggested that gas for the zero be between 0.4% and 2.0% O2. Do not use gases with hydrocarbon con­centrations of more than 40 parts per million. Failure to use proper gases will result in erroneous readings.
Ambient air is not recommended for use as high calibration gas. An 8% O balance in nitrogen is recommended for high calibration gas.
Two tanks of precision calibration gas mixtures. Recommended calibration gases are nominally 0.4% and 8.0% oxygen in nitrogen set calibration gas pressure at 20 psi (138 kPa).
Sub-menu.
2
NOTE
2
2. Press the CAL key. Select PERFORM CALIBRATION sub-menu. "Press EN-
A typical automatic calibration system
is shown in Figure 4-4. TER to start Manual Calibration" will appear on the LCD display. Press
4-6 Calibration Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
4
OPTIONAL
CHECK
VALV E
CALIBRATION GAS
MPS
PROBE
(END VIEW)
REFERENCE
HPS
MPS-IFT SIGNAL CONNECTIONS
AIR
PROBE SIGNAL CONNECTIONS
Instruction Manual
IB-106-300NH Rev. 4.2
July 2002
IFT
INSTRUMENT
AIR IN
NOTE: THE MPS CAN BE USED WITH UP
NOTE: SHOWN WITH HPS OPTION.
TO FOUR PROBES. ONLY ONE PROBE CAN BE CALIBRATED AT A TIME. PROBE CALIBRATIONS MUST BE SCHEDULED IN MULTIPLE PROBE APPLICATIONS.
CALIBRATION
GAS 1
(HIGH )O
2
CALIBRATION
GAS 2
(LOW O )
2
27270006
Figure 4-4. Typical Automatic Calibration System
Rosemount Analytical Inc. A Division of Emerson Process Management Calibration 4-7
Instruction Manual
IB-106-300NH Rev. 4.2 July 2002
World Class 3000
Table 4-1. Automatic Calibration Parameters
Auto Cal YES/NO Set to YES Output Tracks YES/NO Set as desired to con-
figure analog output tracking.
Cal Intvl XD XH Set the desired time
between calibrations in number of days and hours (1 year max).
Next Cal. XD XH Displays the time left to
the start of the next calibration. Set the de­sired time until the start of the next calibration (1 year max). If nothing is entered here, the unit will automatically enter the Cal Intvl and count down from that.
Gas Time 0:00 Set the amount of time
for calibration gases to be turned on in minutes and seconds; allow enough time for signal value to stabilize.
Gas Time 0:00 Set the amount of time
for calibration gases to be turned on in minutes and seconds; allow enough time for signal value to stabilize.
Purge Time 0:00 Set the amount of time
for the gas lines to clear in number of minutes and seconds.
2. Fully Automatic Calibration Setup. In order for the IFT system to calibrate automatically, the parameters from the CALIBRATE sub-menu (shown in Table 4-1) in the IFT have to be entered.
Once these parameters have been set, the system will initiate calibration with­out operator intervention as set by the CAL INTVL parameter.
3. Manually Initiated Fully Automatic Cali­bration Procedure. The following pro­cedure relates to an operator initiated calibration, either by a remote switch (CAL INIT on interconnect board) or selected at the IFT by pressing the CAL key using an MPS 3000 Multi­probe Gas Sequencer.
(a) Press SETUP to display the
SETUP sub-menu. Select Calibra­tion. Ensure that Auto Cal is en­abled. Set the cursor on Auto Cal. Press ENTER. Set Auto Cal to YES if not already done.
(b) Press the CAL key. Select Perform
Calibration. "Press ENTER to start Automatic Calibration" will appear on the LCD display. Press ENTER to start. Refer to Table 5-5, CALI­BRATE O
Sub-Menu.
2
Abort Time 0:00 Set the amount of time
allowed between key functions before the calibration procedure is aborted in number of minutes and seconds.
Res Alarm ____ Set the desired resis-
tance alarm between 50 to 10,000 ohms.
4-8 Calibration Rosemount Analytical Inc. A Division of Emerson Process Management
Instruction Manual
4
IB-106-300NH Rev. 4.2
World Class 3000
July 2002
Calibration Record
For
Rosemount Analytical In Situ O
Probe Serial Number: Probe Tag Number: Probe Location: Date Placed Into Service:
Probe
2
Date Slope Constant Impedance Response
initial
Response
final
Notes: Response
Response
Rosemount Analytical Inc. A Division of Emerson Process Management Calibration 4-9
When the second calibration gas is turned off, note the number of seconds required for
initial
final
value to begin migrating back to the process value.
the O
2
When the second calibration gas is turned off, note the number of seconds required for
value to settle out at the process value.
the O
2
Instruction Manual
IB-106-300NH Rev. 4.2 July 2002
World Class 3000
4-10 Calibration Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
5

GENERAL USER INTERFACE (GUI) OPERATION

Instruction Manual
IB-106-300NH Rev. 4.2
July 2002
SECTION 5
5-1 OVERVIEW
Ensure that the oxygen analyzer, heater po wer supply, and intelligent field transmitter have been properly connected. It is important to check that grounding and screening of termina­tions are correctly made to prevent the introduc­tion of ground loops. The IFT is equipped with noise suppression circuitry on the power supply and signal input lines. Proper grounding at in­stallation will ensure accuracy of function.
The following five languages are can be se­lected within the IFT:
English Italian French Spanish German
NOTE
Support the keypad with the free hand to prevent bounce back of the IFT door.
a. Intelligent Field Transmitter (IFT)
b. HART Communicator Interface Devices
The HART communications protocol can interface with any of the above IFT ver­sions. To interface a HART communicator with an IFT, one of three interface devices is required. The interface devices are as follows:
1. Rosemount Model 275 Handheld Communicator. The handheld com­municator requires Device Descriptor (DD) software specific to the World Class 3000 product. The DD software will be supplied with many model 275 units, but can also be programmed into existing units at most Fisher­Rosemount service offices.
2. Personal Computer (PC). The use of a personal computer requires Corner­stone software with Module Library (ModLib) specific to the World Class 3000 product.
The Intelligent Field Transmitter may be supplied with either of two configurations. These are the blind version and the deluxe version. The two versions differ as follows:
1. Blind Version. The blind version has no display and no keypad. With this version an external HART communica­tions device is required .
2. Deluxe Version (GUI). The deluxe version is also known as the General User Interface (GUI) version. This IFT contains an LED display, liquid crystal display panel, and an eight-key pad from which the probe and electronics can be configured, calibrated and troubleshooted.
3. Selected Distributed Control Systems. The use of distributed control systems requires input/output (I/O) hardware and software which permit HART communications.
This section of the manual deals with op­erator controls and displays available with the GUI equipped IFT. Operating parame­ters are listed and instructions are included for viewing and changing them.
Any procedures not associated with normal operation are included in Section 2, Instal­lation, or Section 5, Troubleshooting.
Rosemount Analytical Inc. A Division of Emerson Process Management General User Interface (GUI) Operation 5-1
Instruction Manual
IB-106-300NH Rev. 4.2 July 2002

5-2 DELUXE VERSION IFT DISPLAYS AND CONTROLS

World Class 3000
3
1
2
HELP
DATA
CAL
ESC
SETUP
ENTE
R
6
7
8
9
INTERNAL VIEW
4
5
IFT COVER DOOR
NOTE:
SHOWN FOR REFERENCE.
EXTERNAL VIEW
CAL TGH TGL
10
Figure 5-1. Deluxe Version IFT Displays and Controls
Figure 5-1
Index No. Control/LED Description
1 LCD Display Top line displays system status, menu, and probe number.
11
12
13
21190003
2 HELP Context sensitive HELP is displayed when this key is pressed. 3 DATA DATA key is used to access DATA menu. 4 CAL CAL key used to access CALIBRATE menu. 5 SETUP SETUP key used to access SETUP menu. 6 ESC The escape key is used to exit to a high level menu or to abort a
parameter change.
7 The decrease key is used to move the cursor (asterisk) when scrolling
through lists or to decrease a parameter value.
8
The increase key is used to move the cursor (asterisk) when scrolling through lists or to increase a parameter value.
9 ENTER The ENTER key is used to select a lower level menu, initiate calibra-
tion, or select a parameter to change.
10 LED Display Indicates current O
or calibration gas value.
2
11 CAL Calibration in progress indicator light. 12 TGH High calibration gas indicator light. High calibration gas is being used in
calibration process.
13 TGL Low calibration gas indicator light. Low calibration gas is being used in
calibration process.
5-2 General User Interface (GUI) Operation Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
5
Table 5-1. Sample HELP Messages
MENU, SUB-MENU, HELP OR PARAMETER NAME MESSAGE
PROBE DATA Press ENTER key to access DATA menu.
Instruction Manual
IB-106-300NH Rev. 4.2
July 2002
CALIBRATE O
2
The CAL menu is used to start calibration and view calibration.
SETUP The SETUP menu is used to configure the IFT 3000.
5-3 HELP KEY
The HELP key will display explanatory informa­tion about a menu, sub-menu, or parameter that the asterisk is next to when pressed. The HELP key is not available during calibrati on ro uti nes. Refer to Table 4-1 for sample HELP messages.

5-5 QUICK REFERENCE CHART

The quick reference chart (Figure 5-2) is de­signed to help you get where you want to be in the menu system. The chart shows all the avail­able menu and sub-menu options for the IFT. Follow the lines to determine whic h m enu choices to make. Moving down a level on the chart is accomplished by the use of the ENTER

5-4 STATUS LINE

key. To move up a level on the chart, press the
ESCAPE key. The top line of the LCD display (1, Figure 5-1) is a status line that always displays system status, menu name, and O
level. System status dis-
2
plays will be displayed one at a time in priority sequence, as follows:
a. Off - The probe has been turned off be-
cause the IFT cannot control the heater temperature.

5-6 MAIN MENU

When power is first applied to the IFT, the MAIN
menu (Table 5-2) is initially displayed. It is from
the MAIN menu that the PROBE DATA (Table
5-3), CALIBRATE O
(Table 5-4), and SETUP
2
(Table 5-5) menus can be accessed.
Table 5-2. MAIN menu
b. PrbEr - The probe is disconnected, cold, or
leads are reversed.
c. HtrEr - Heater error.
MENU SELECTION DESCRIPTION
PROBE DATA Refer to Table 5-3. CALIBRATE O
2
Refer to Table 5-4.
SETUP Refer to Table 5-5.
d. InCAL - Calibration in progress.

5-7 PROBE DATA SUB-MENU

- O
e. Low O
limit.
value is below the low alarm
2
2
The PROBE DATA sub-menu is a list of all the
parameters of the system as it is currently f. HiO2 - O
value is above the high alarm
2
limit.
g. NoGas - Calibrati on gas press ure is low.
configured. To access the PROBE DATA
sub-menu, press the DATA key at any time. The
increase and decrease keys are used to scr oll
through the list. The PROBE DATA sub-menu
can be viewed but not changed. The operator
must use the SETUP menu to change any of
h. CalEr - Calibration error.
i. ResHi - Resistance is above the high limit.
the parameters.
There are two selections available on the
PROBE DATA sub-menu; Process Data and
Diagnostic Data. Refer to Table 5-3 for contents j. OK - System is functioning correctly.
Rosemount Analytical Inc. A Division of Emerson Process Management General User Interface (GUI) Operation 5-3
of the sub-menu.
Instruction Manual
IB-106-300NH Rev. 4.2 July 2002
World Class 3000

5-8 CALIBRATE O2 SUB-MENU

The CALIBRATE O
sub-menu (Table 5-4) is
2
used to enter the calibration mode. To access the CALIBRATE O
sub-menu, press the CAL
2
key at any time. The increase and decrease keys are used to scroll through the list.
The CALIBRATE O
sub-menu has three se-
2
lections available: Perform Calibration, View Constants, and Calibration Status. Refer to Table 5-4 for contents of the sub-menus.
Perform Calibration has two options depending on how Auto Cal is selected in Probe Setup. Refer to SETUP Setting in Table 5-4.
For information on performing a calibration, re­fer to Section 4, Calibration.
Table 5-3. PROBE DATA Sub-Menu
SUB-MENU
SELECTION PARAMETER DESCRIPTION
Process Data O
Efficiency __% Efficiency display. Stack Temp __DegC Stack temperature.
5-9 SETUP SUB-MENU
The SETUP sub-menu is used to enter all op-
erator set variables into the system. To access
the SETUP sub-menu press the SETUP key at
any time. To select the parameter to be
changed, move the cursor to the desired pa-
rameter using the arrow keys. Press ENTER to
select that parameter. To change the value for
that parameter, use the arrow keys to increase
or decrease the value. Press ENTER to save
changes.
There are six selections available on the
SETUP sub-menu: Calibration, O
Alarms, Efficiency Calc., Relay Outputs, and
O
2
Calculation,
2
Analog Outputs. Refer to Table 5-5 for the con-
tents of the SETUP sub-menu, or ESCAPE to
abort changes.
2
__% O
2
O2 value for the probe.
Diagnostic Data
Temperature Cell __DegC Cell temperature of the probe.
Voltages Cell __mV Cell voltage of the probe.
Output Values Analog __% FS Analog output voltage.
Stack __DegC Stack temperature. Cold Junct __DegC Cold Junction temperature.
Cell T/C __mV Cell thermocouple voltage of the probe. Stk T/C __mV Stack thermocouple voltage. Cold Jnt __mV Cold junction voltage.
K1 State OFF/ON Status of relay 1. K2 State OFF/ON Status of relay 2.
5-4 General User Interface (GUI) Operation Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
5
PROBE DATA DIAGNOSTIC
PROCESS DATA
DATA
O2 Efficiency Stack Temp
TEMPERATURE
VOLTAGES
Instruction Manual
IB-106-300NH Rev. 4.2
July 2002
Cell Stack Cold Junct
Cell Cell T/C Stk T/C Cold Jnt
CALIBRATE O2
(CONTINUED ON
SHEET 2)
PERFORM CALIBRATION
VIEW CONSTANTS
CALIBRATION STATUS
OUTPUT VALUES
LATEST CALIBRATION
PREVIOUS CAL
Next Cal Slope Constant Resist
Analog K1 State K2 State
Slope Constant Resist
Slope Constant Resist
686022
Figure 5-2. Quick Reference Chart (Sheet 1 of 5)
Rosemount Analytical Inc. A Division of Emerson Process Management General User Interface (GUI) Operation 5-5
Instruction Manual
IB-106-300NH Rev. 4.2 July 2002
World Class 3000
(CONTINUED FROM
SHEET 1)
SETUP
CALIBRATION
O2 CALIBRATION
O2 ALARMS
See sheet 4
SLOPE
CONSTANT
SET POINT
RESET SLOPE AND CONST
HI ALARM LO ALARM ALARM DB
ENABLE CALC
34.5 mV/D-
57.5 mV/D
-20.0 mV-
20.0 mV
o
736 C
o
843 C
0.1000% O2-
25.00% O2
0.00% O2-
25.00% O2
Yes No
(CONTINUED ON
SHEET 3)
EFFICIENCY CALC
RELAY OUTPUT
K1 VALUE K2 VALUE
K3 VALUE
K1 SETUP
K2 SETUP
(CONTINUED ON
SHEET 3)
Figure 5-2. Quick Reference Chart (Sheet 2 of 5)
0.0000-
1.000
0.0000-
20.00
EVENT 1 EVENT 2 EVENT 3
EVENT 1 EVENT 2 EVENT 3
Off In Cal Hi O2 Lo O2 Htr Fail Cal Fail TG Low Cell Res High Range
19860023
5-6 General User Interface (GUI) Operation Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
5
(CONTINUED FROM
SHEET 2)
(CONTINUED FROM
SHEET 2)
ANALOG OUTPUTS
SOURCE
AOUT TYPE
Instruction Manual
IB-106-300NH Rev. 4.2
July 2002
O2
Efficiency
Dual Rng O2
HART 4-20 mA
0-20 mA
0-10 V
SETUP
RANGE SETUP
USA
GBR
COUNTRY
FRA
ESP
GER
Figure 5-2. Quick Reference Chart (Sheet 3 of 5)
See sheet 5
27270004
Rosemount Analytical Inc. A Division of Emerson Process Management General User Interface (GUI) Operation 5-7
Instruction Manual
IB-106-300NH Rev. 4.2 July 2002
World Class 3000
CALIBRATION
HIGH GAS
LOW GAS
AUTO CAL
OUTPUT TRACKS
CAL INTRVL
0.1000% O2
25.00% O2
0.1000% O2
25.00% O2
Yes No
Yes No
Off,
1H -
365 D OH
(1 hour to 365 days and no hours)
(CONTINUED FROM
SHEET 2)
Figure 5-2. Quick Reference Chart (Sheet 4 of 5)
NEXT CAL
GAS TIME
PURGE TIME
RES ALARM
Disabled,
1H -
365 D OH
00:30 ­20:00
00:30 ­20:00
50 -
10000
(1 hour to 365 days and no hours)
16860025
5-8 General User Interface (GUI) Operation Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
5
Instruction Manual
IB-106-300NH Rev. 4.2
July 2002
Range Setup
(Source not set to:
Dual Rng O2)
(CONTINUED FROM
SHEET 3)
RANGE SETUP
XFER FNCT
RANGE VALUES
XFER FNCT
Log LIN
HIGH END
LOW END
Log LIN
0.000% O2
25.00% O2
0.000% O2
25.00% O2
Range Setup
(Source not set to:
Dual Rng O2)
HIGH END
NORMAL RANGE VALUES
DUAL RANGE SETUP
LOW END
MODE SETUP
HIGH RANGE VALUES
Figure 5-2. Quick Reference Chart (Sheet 5 of 5)
0.000% O2
25.00% O2
0.000% O2
25.00% O2
RANGE MODE
HIGH IN CAL
SWITCHES AT
LOW END
HIGH END
Normal Auto High
Yes No
0.000% O2
25.00% O2
0.000% O2
25.00% O2
0.000% O2
25.00% O2
16860026
Rosemount Analytical Inc. A Division of Emerson Process Management General User Interface (GUI) Operation 5-9
Instruction Manual
IB-106-300NH Rev. 4.2 July 2002
World Class 3000
Table 5-4. CALIBRATE O2 Sub-Menu
SUB-MENU
SELECTION
Perform Calibration
SETUP SETTING
(SEE TABLE 3-5) DISPLAY DESCRIPTION
Auto Cal in Probe Setup is YES
Press ENTER to start Auto Calibration.
MPS will start calibrating probe.
Starting Automatic Calibrat ion High Gas _____% O
Time Left 0:00
2
Value for high O
calibration gas.
2
Amount of time necessary to com­plete the current testing phase in
min:sec. Cell mV ______mV Cell voltage of the probe. Low Gas _____% O
Time Left 0:00
2
Value for low O
calibration gas.
2
Amount of time necessary to com-
plete the current testing phase in
min:sec. Cell mV ______mV Cell voltage of the probe. Resistance Check
Resistance check in progress. Time Left 0:00
Cell _____mV _____C
Cell voltage and probe temperature. Calibration Complete Purging 0:00
Gas lines are being purged of cali-
bration gas. Cell _____mV _____C
Cell voltage and probe temperature. Calibration Complete
Auto Cal in Probe Setup is NO.
Press ENTER to start Manual Calibration.
Manual calibration sequence will
begin when ENTER is pressed. Switch ON high calibration gas.
Press ENTER when ready. High gas ______% O
2
High O2 calibration gas value. Press ENTER when O2 reading
is stable. Turn OFF high calibration gas
and ON low calibration gas. Press ENTER when ready.
Low gas ______% O
2
Low O2 calibration gas value. Press ENTER when O2 reading
is stable. Resistance Check. Resistance check in progress. Turn off low calibration gas.
Press ENTER when ready. Press ENTER when probe has
returned to process.
5-10 General User Interface (GUI) Operation Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
5
Instruction Manual
IB-106-300NH Rev. 4.2
July 2002
Table 5-4. CALIBRATE O2 Sub-Menu (continued)
SUB-MENU
SELECTION
View Constants
Calibration Status
SETUP SETTING (SEE TABLE 3-5) DISPLAY DESCRIPTION
Latest Calibration
Previous Calibration
N/A Next Cal XD XH Time until next calibration in number of
Slope _____mV/D Constant _____mV Resist _____ohms
Slope _____mV/D Constant _____mV Resist _____ohms
Slope _____ Constant _____ Resist _____
Slope for probe from latest calibration. Latest calibration offset for probe. Latest calibration resistance of probe.
Slope for probe from previous calibration. Previous calibration offset for probe. Previous calibration resistance of probe.
days and number of hours. Status of the slope.
Status of the offset. Status of the resistance.
Rosemount Analytical Inc. A Division of Emerson Process Management General User Interface (GUI) Operation 5-11
Instruction Manual
IB-106-300NH Rev. 4.2 July 2002
Table 5-5. SETUP Sub-Menu
SUB-MENU
SELECTION PARAMETERS DESCRIPTION
Calibration High Gas ____% O
Low Gas ____% O
2
2
Value of high O2 calibration gas (0.1000% - 25.00%
).
O
2
Value of low O2 calibration gas (0.1000% - 25.00%
).
O
2
Auto Cal YES/NO MPS required for Auto Cal. Output Tracks YES/NO NO, locks output during calibration. Cal Intrvl XD XH
Select time between calibrations in number of days and hours (1 year max).
Next Cal XH
Time until next calibration in number of hours (1 year max).
Gas Time 0:30 - 20:00 Amount of time calibration gases will be turned on in
number of minutes and seconds; allow enough time for signal values to stabilize.
Purge Time 0:30 - 20:00 Amount of time for gas lines to clear of calibration
gas.
Res Alarm 50 W – 10 kW Resistance alarm set from 50 to 10,000 ohms.
O2 Calculation Slope ____ mV/D
Constant ____ mV Set Point ____°C
Set value between 34.5 and 57.5. Set value between -20.0 and +20.0 mV. Set either 736 for World Class 3000 probes or 843 for 218 probes.
World Class 3000
Ensure the correct voltage is selected when using HPS 3000 with either World Class 3000 probes or 218 probes. Refer to Figure 2-15, Jumper Selection Label for proper voltage selections. If incorrect SET POINT is selected, damage to the probe may occur.
Reset slope and constants.
O2 Alarms Hi Alarm ____% O
Lo Alarm ____% O Alarm DB ____% O
Efficiency Calc. Enable Calc. YES/NO
K1 Value _______ K2 Value _______ K3 Value _______
Press ENTER to reset slope and constants to values from the latest successful calibratio n.
2 2 2
Set value for high alarm limit (0.1000% - 25.00%). Set value for low alarm limit (0.1000% - 25.00%). Set value for alarm dead band (0.0000% - 25.00%).
Select YES to enable, NO to disable. Set between 0.0000 and 1.000. Refer to Table 5-6. Set between 0.0000 and 1.000. Refer to Table 5-6. Set between 1.000 and 20.00. Refer to Table 5-6.
5-12 General User Interface (GUI) Operation Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
5
Table 5-5. SETUP Sub-Menu (continued)
SUB-MENU
SELECTION PARAMETERS DESCRIPTION
Relay Outputs
NOTE
K1 and K2 relay outputs can be configured for "OFF" or any one of the eight events listed below. Up to three events can control each relay output. Events are selected in the SETUP sub-menu.
K1 Setup
K2 Setup
Analog Output
Event 1 1. In Cal Event 2 2. Hi O Event 3 3. Lo O Event 1 4. Htr Fail Event 2 5. Cal Fail Event 3 6. TG Low
SOURCE O
-Off
2
2
7. Cell Res
8. High Range
2
Efficiency Dual Rng O
2
No effect . Probe goes into calibration status. Output exceeds high end alarm limit. Output goes below low alarm limit. Probe heater fault occurs. Probe failed last calibration. Calibration gas pressure gets too low. Probe resistance exceeds high limit. High analog output range is selected.
Select the measurement value to be represented on the analog output.
Instruction Manual
IB-106-300NH Rev. 4.2
July 2002
AOUT TYPE
HART 4-20mA 0-20mA 0-10V
RANGE SETUP (Source not set to Dual Rng O2) Xfer Fnct Log
Lin
Range Values
High End
0.000% O
- 25.00% O
2
Low End
0.000% O
- 25.00% O
2
Select one of the listed options to define upper and lower limits of probe analog output. Only a selection that matches the position of the analog output selector switch on the microprocessor board (Figure 2-9) will be accepted. The defined limits correspond to the upper-lower %O
values defined in the Range Setup
2
menu.
Select the transfer function used on the analog output. Selecting Log will not effect the output when Efficiency is selected as the Source.
Enter the upper and lower analog output range values. The High End value defines the measured O
2
corresponding to the high analog output value, i.e., 20mA or 10V, and the Low End value corresponds to
2
the low analog output value, i.e., 0mA, 4mA, or 0V.
value
2
Rosemount Analytical Inc. A Division of Emerson Process Management General User Interface (GUI) Operation 5-13
Instruction Manual
IB-106-300NH Rev. 4.2 July 2002
Table 5-5. SETUP Sub-Menu (continued)
SUB-MENU
SELECTION PARAMETERS DESCRIPTION
Analog Output RANGE SETUP
(continued) (Source set to Dual Rng O2)
Xfer Fnct Log
Lin
Select the transfer function used on the analog output. Selecting Log will not effect the output when Efficiency is selected as the Source.
World Class 3000
Normal Range Values
High End
0.000% O
- 25.00% O
2
Low End
0.000% O
- 25.00% O
2
Enter the upper and lower analog output range values for Normal Operating Range. The High End value de­fines the measured O
2
analog output value, i.e, 20mA or 10V, and the Low End value corresponds to the low analog output value,
2
i.e., 0mA, 4mA, or 0V.
value correspondin g to the high
2
Dual Range Setup
Mode Setup
Range Mode Normal Forces the output to the Normal Range.
Auto Allows the IFT to select either the High Range or the
Normal Range based on the present O
2
Mode Setup Values.
High Forces the output to the High Range.
High in Cal Yes/No Selecting Yes will cause the High Range to be used
whenever the probe is being calibrated.
Switches at
0.000% O
- 25.00% O
2
2
Enters the switching point between the High and Nor­mal Ranges. O
values above this point will use the
2
High Range and values below this point wil l use the Normal Range. The O
value must be below the switch
2
point by 10% (of the "Switches at" value) to cause a switch from High to Normal Range.
High Range Values
High End
0.000% O
- 25.00% O
2
Low End
0.000% O
- 25.00% O
2
2
2
Enter the upper and lower analog output range values for High Operating Range. The High End value defines the measured O
value corresponding to the high
2
analog output value, i.e., 20mA or 10V, and the Low End value corresponds to the low analog output value, i.e., 0mA, 4mA, or 0V.
value and the
NOTE: Relay output can be initiated upon range change. (See page 5-12 of Table 5-5.)
Table 5-6. Efficiency Constants
UNITED STATES EUROPE
CONSTANT
GAS OIL GAS OIL
K1 0.407 0.432 0.66 0.69 K2 0.0 0.0 0.0082 0.0051 K3 5.12 5.12 12.28 8.74
5-14 General User Interface (GUI) Operation Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
6
Instruction Manual
IB-106-300NH Rev. 4.2
July 2002
SECTION 6

TROUBLESHOOTING

6-1 OVERVIEW
The system troubleshooting describes how to identify and isolate faults which may develop in the Oxygen Analyzer System. Refer to Probe, IFT, HPS, MPS, and HART Communicator appendices.
Install all protective equipment covers and safety ground leads after trouble­shooting. Failure to replace covers and ground leads could result in seri­ous injury or death.

6-2 SPECIAL TROUBLESHOOTING NOTES

a. Grounding
It is essential that adequate grounding pre­cautions are taken when the system is be­ing installed. A very thorough check must be made at both the probe and electronics to ensure that the grounding quality has not degraded during fault finding. The system provides facilities for 100% effective grounding and the total elimination of ground loops.
b. Electrical Noise
The IFT has been designed to operate in the type of environment normally found in a boiler room or control room. Noise suppres-
sion circuits are employed on all field termi­nations and main inputs. When fault finding, the electrical noise being generated in the immediate circuitry of a faulty system should be evaluated. All cable shields must be connected to earth.
c. Loose Integrated Circuits
The IFT uses a microprocessor and sup­porting integrated circuits. Should the elec­tronics unit receive rough handling during installation in a location where it is sub­jected to severe vibration, an Integrated Circuit (IC) could work loose. The fault find­ing guides in paragraph 6-3 and Table E-2 in Appendix E, show the resulting variety of failure modes. It is recommended that all IC's be confirmed to be fully seated before troubleshooting on the system begins.
d. Electrostatic Discharge
Electrostatic discharge can damage the IC's used in the electronics unit. It is essential that the user ensure he/she is at ground potential before removing or handling the processor board or the IC's used on it.
6-3 SYSTEM TROUBLESHOOTING
The status line of the GUI equipped IFT will dis­play one of ten conditions. The system status displays will be displayed one at a time in prior­ity sequence, as indicated in Table 6-1.
Rosemount Analytical Inc. A Division of Emerson Process Management Troubleshooting 6-1
Instruction Manual
IB-106-300NH Rev. 4.2 July 2002
World Class 3000
Table 6-1. IFT Status Codes
Off Heater power has been turned OFF by the electronics. The display shows 0% O
. Several con-
2
ditions may cause the OFF status:
1. The cell heater temperature is below -50°C. The thermocouple wires may be reversed.
2. The cell temperature is more than 70°C above the set point. The heater is out of control. The triac module may have failed.
3. The cell heater thermocouple voltage has remained within +1.5 mV for more that 4 minutes. The thermocouple may be shorted.
4. The AD590 voltage is below 50.0 mV (50K or -223°C). The AD590 is not connected.
5. The AD590 voltage is above 363 mV (363K or 90°C). If HPS is used with IFT, then IFT in­terconnect board has JM1 in position connecting two AD590s in parallel.
PrbEr The probe is disconnected or cold, or leads are reversed. HtrEr There is a fault within the heater system. The heater temperature is more than +25°C from the
set point. When the unit is first turned ON, HtrEr is normal. The heater may take 0.5 to 1.0 hours to warm up.
InCal The system is currently undergoing calibration. If Output Tracks is set to YES, the output will
show changing O
values. If Output Tracks is set to NO, the output will hold the pre-calibration
2
value.
LowO
2
The measured O2 value is below the low O2 alarm limit. The problem may be in the probe or the process.
HiO
2
The measured O2 value is above the high O2 alarm limit. The problem may be in the probe or the process.
NoGas Test gas pressure is too low. Pressure switches are set to trigger this alarm at 12 to 16 psig (83
to 110 kPa gage). Test gas regulators are usually set at 20 to 25 psig (138 to 172 kPa gage). Possible causes are:
1. At least one test gas pressure switch is open.
2. A test gas cylinder is empty.
3. There is an MPS or piping failure.
4. If MPS is not connected, CALRET and NOGAS signals should be jumpered on the intercon­nect board.
CalEr An error occurred during the last calibration. The error may be one of the following:
1. The new calculated slope value is outside the range 34.5 to 57.6 mV/decade.
2. The new calculated constant value is outside the range +20.0 to -20.0 mV.
3. The test gas pressure switch opened during calibration.
Ensure that the proper test gases are being used, and that the gas flows are set properly. Refer to Appendix D for additional MPS troubleshooting information.
ResHi The resistance calculated during the last good calibration was greater than the High Resistance
Alarm limit set in the calibration setup. The resistance limit may be set wrong, or there is a
problem with the probe. Ok Operation appears to be normal. (blank
screen)
6-2 Troubleshooting Rosemount Analytical Inc. A Division of Emerson Process Management
A possible failure within the IFT. Check the LED on the microprocessor board to help isolate
problems. See IFT Problem in the troubleshooting tables.
World Class 3000
6
Instruction Manual
IB-106-300NH Rev. 4.2
July 2002
6-4 HEATER PROBLEM
For all heater troubleshooting, al lo w at least 30 minutes for the operating temperature to stabi­lize. After the warmup period, observe the sys­tem status and the voltages of the cell TC and the cold junction AD590. For heater related problems:
a. The status line may read: HtrEr or OFF.
Table 6-2. Heater Troubleshooting
Problem
Cause
Corrective Action
Status is HtrEr or OFF. Cell TC < 28.4 mV. Cold Junction 273 to 330 mV (normal).
Display = 0%
O
2
1. Blown fuse or faulty wiring. Check all fuses and wiring for continuity and repair as needed. Verify that input power jumpers are installed correctly. Check jumpers for proper configuration in IFT and HPS if used.
2. Heater failure. In HPS with power OFF, check heater resistance at J2, terminals R/H. For 44 V heater, resis­tance should be 11 to 14 ohms. For 115 V heater, resistance should be 67 to 77 ohms. Check wiring, and replace heater if needed. Heater resistance can also be checked at the probe junc­tion box:
44 V heater: terminals 7 and 8 should measure 11 to 14 ohms.
115 V heater: terminals 5 and 6 should measure 67 to 77 ohms. (Terminals 6 to 7 and 6 to
8 should be open circuits.)
3. Triac open. Check the triac. Repair as needed.
4. Electronics failure. First check and repair all related wiring. Check and repair electronics as needed.
5. Missing insulation around heater. Check that insulation is in place and undamaged. Repair or replace insulation as needed.
b. The displayed O c. Cell TC voltages will vary from normal.
These voltages are found by accessing the proper menu. In the IFT, use the DIAG­NOSTIC DATA sub-menu of the PROBE DATA menu.
Refer to Table 6-2 to troubleshoot heater related problems.
value will read 0%.
2
Rosemount Analytical Inc. A Division of Emerson Process Management Troubleshooting 6-3
Instruction Manual
IB-106-300NH Rev. 4.2 July 2002
Table 6-2. Heater Troubleshooting (continued)
Problem
Cause
Corrective Action
Status is HtrEr or OFF. Cell TC > 28.4 mV. Cold Junction 273 to 330 mV (normal).
Display = 0%
O
2
1. Triac failure. Check the triac. Repair as needed.
2. Wrong TC set point. Check electronics manual and verify the set point; typically 1356°F (736°C).
3. Wrong heater voltage selected. HPS voltage jumpers setup wrong. For 44 V heater, make sure JM7 is installed and JM8 is removed. For 115 V heater, JM7 is removed and JM8 is installed. The 115 V heater has an identifying stainless steel tag attached in the junction box.
Status is HtrEr or OFF. Cell TC < 28.4 mV. Cold Junction < 273 mV.
Display = 0%
O
2
1. Wiring error, thermocouple wires reversed. Verify TC wiring at junction box terminal and electronics. The yellow chromel line connects to terminal 3. The red alumel line connects to terminal 4. Trace line through the HPS (if used) and the electronics. Reverse wires if needed.
2. Faulty thermocouple. At a cold junction reference of 77°F (25°C), the probe TC should read about
29.3 mV. Replace faulty thermocouple.
3. Faulty AD590. At normal ambient temperatures, cold junction sensor should be 273 to 330 mV. Replace faulty sensor.
World Class 3000
Status is HtrEr or OFF. Cell TC = -40 mV. Cold Junction 273 to 330 mV (normal).
Display = 0%
O
2
1. Faulty thermocouple connection or open. Verify TC wiring at junction box terminal and electronics. The yellow chromel line connects to terminal 3. The red alumel line connects to terminal 4. Trace line through the HPS (if used) and the electronics. Repair connection or wiring as needed.
2. Thermocouple fault. At a cold junction reference of 77°F (25°C), the probe TC should read about
29.3 mV. Replace faulty thermocouple.
6-4 Troubleshooting Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
6
Instruction Manual
IB-106-300NH Rev. 4.2
July 2002

6-5 CELL PROBLEM

For cell troubleshooting, as in heater problems, you should allow at least 30 minutes for operat­ing temperature to stabilize. After this warmup period, observe the system status and cell volt­age. If the heater is working, troubleshoot the cell. If the heater is not working, refer to Heater Problem, paragraph 6-4.
The status line may read: Low O CalEr, ResHi.
Access voltage values in the proper menu. Use the DIAGNOSTIC DATA sub-menu of the PROBE DATA menu.
Problem
Cause
Corrective Action
Status is LowO2. Cell mV = -127 mV.
1. Faulty cell connection or open. If the cell circuit is open, the cell output will show about -127 mV. Check cable connection between the probe and the electronics. Check that the probe spring presses the contact pad firmly onto the cell. Repair or replace faulty wires, spring, or connectors.
2. Electronics fault. Cell output is good, and the input to the electronics is good. Check the electronics package. In an IFT, replace the microprocessor or interface board as needed. In a CRE, replace the DPI board if needed.
The displayed O 99%.
It may be helpful to observe the calibration status and parameters from the last calibra­tion: Slope, Constant, and Cell Resistance. In the CALIBRATE menu, VIEW CON­STANTS shows previous calibration values, and CALIBRATION STATUS shows the
, Hi O2,
2
Table 6-3. Cell Troubleshooting
latest values. If these values appear out of range, perform a calibration before trouble­shooting the cell.
Refer to Table 6-3 to troubleshoot cell related problems.
value will read 0% to
2
Status is ResHi or CalEr. Cell mV = -20 to 120 mV (normal).
1. Test gas flow not 5 scfh (2.4 L/min). Check test gas flow and related piping. Rotameter should show 5 scfh. Adjust needle valve for correct flow rate.
2. Incorrect test gas. Confirm labels on test gas bottles are correct. Confirm High Gas and Low Gas values agree with labels on test gas bottles. (Refer to menu map — SETUP-CALIBRATION, High Gas, Low Gas.) Check all ports, cylinders, and gas lines for proper hookup. Change piping if necessary. Label pipes for reference.
Rosemount Analytical Inc. A Division of Emerson Process Management Troubleshooting 6-5
Instruction Manual
IB-106-300NH Rev. 4.2 July 2002
Table 6-3. Cell Troubleshooting (continued)
Problem
Cause
Corrective Action
Status is ResHi or CalEr. Cell mV = -20 to 120 mV (normal) (continued).
1. Reference air contamination (oil/water). Clean or replace lines and valves as needed.
2. Cell leads reversed. Check cell signal wiring from probe junction box to electronics, and correct wiring if needed.
3. Reference/test gas lines reversed. Switch piping as needed.
4. Diffusion element fault. Diffusion element cracked, broken, missing, or plugged. Replace diffusor or snubber as needed. Diffusors are disposable because it is difficult to clean a diffusor and know the tiny pores are open. A flow and pressure test with a manometer is possi­ble but usually not practical. To clean a snubber, blow off surface dirt with pressurized air and clean the unit in an ultrasonic bath.
5. Faulty cell. Low sensor cell output when test gas is applied. If test gas flow is good and there is low cell signal, replace the cell, or call the SCAN line for assistance.
World Class 3000
Typical cell output: Test Gas mV
8.0% 18 to 25
0.4% 76 to 86
6. Cell performance degraded from aging. Replace the sensor cell if its resistance has increased beyond 1 kOhm and the slope calculated during calibration has decreased lower than 40 mV/decade.
7. Electronics fault. Cell output is good, and the input to the electronics is good. Check the electronics package. In an IFT, replace the microprocessor or interface board as needed. In a CRE, replace the DPI board if needed.
Status is Res Hi. Cell mV = -120 to 20 mV.
1. Cell leads reversed. Check cell signal wiring from probe junction box to electronics, and correct wiring if needed.
2. Reference/test gas lines reversed. Switch piping as needed.
3. Reference air (nitrogen). Confirm labels on test gas bottles are correct. 100% nitrogen must NOT be used as a zero gas because cell protection will engage and affect the O instrument air prepared from ambient air with 20.95% O
reading. Reference air should be clean, dry
2
.
2
6-6 Troubleshooting Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
6
Instruction Manual
IB-106-300NH Rev. 4.2
July 2002

6-6 IFT PROBLEM

When an IFT problem is suspected, look at the LED on the microprocessor board. The LED may be OFF, ON, or flashing.
Table 6-4. IFT Troubleshooting
Problem
Cause
Corrective Action
IFT LED is OFF. IFT failure. Fuse fault. Check fuses on power supply board. Replace fuses as needed.
1. Power fault. Check line voltage. Correct or turn main power ON.
2. Power supply fault. Check voltage test points on the microprocessor board. Replace power supply board if needed.
3. Microprocessor board fault. Replace microprocessor board.
IFT LED is steady ON. Heater or cell wiring problem.
1. Faulty wiring. Check thermocouple and heater wires and connections for continuity. Repair as needed.
2. Jumpers set up wrong. JM1 on interconnect board, JM6 on microprocessor board, or JM9 and JM10
on power supply board are configured incorrectly.
Check that jumpers are set up as follows:
Without an HPS, JM1 and JM6 should be installed.
With a 115 V probe heater, JM9 is installed.
With a 44 V probe heater, JM10 is installed.
3. Status line is “OFF”. Turn OFF IFT power and restart. If light stays ON and both wiring and jumpers are OK, then re­place the microprocessor board.
Refer to Table 6-4 to troubleshoot IFT related problems.
Faulty GUI or LDP (IFT LED is Flashing).
1. Microprocessor is normal, but front panel indicators are not working properly. Check connections to GUI or LDP, and repair or replace as needed.
Rosemount Analytical Inc. A Division of Emerson Process Management Troubleshooting 6-7
Instruction Manual
IB-106-300NH Rev. 4.2 July 2002
World Class 3000

6-7 MPS PROBLEM

MPS problems can occur with a status of C Err, R Hi, TGLow. The O
Problem
Cause
Corrective Action
Status is NoGas. Cell mV is between -20 to 120 mV.
1. Regulator or plumbing fault. The test gas pressure is low for the indicated probe [20 to 25 psig (138 to
172 kPa gage)].
Check test gas pressure [should be 20 psig (138 kPa gage)], regulator, and lines. Reset, repair, or replace the regulator as needed. If only one probe has low flow [less than 5 scfh (2.4 L/min)], check lines, needle valve, connectors, and MPS solenoid for that probe.
2. Test gas low.
Replace empty test gas cylinder with full cylinder. Verify O
3. Wiring fault.
Confirm proper wiring and continuity between MPS and electronics. Repair as needed.
4. Pressure switch fault.
Pressure switch is factory set at 16 psig (68.9 kPa gage). Set test gas regulator pressure to 20 psig (138 kPa gage) to avoid nuisance alarms. Replace faulty switch with a new one if test gas supply is good.
reading can be 0% to
2
99%, and probe data will be in the normal ranges. Consider two conditions, A and B.
Refer to Table 6-5 to troubleshoot problems with the MPS.
Table 6-5. MPS Troubleshooting
concentration.
2
Status is ResHi or CalEr. Cell mV is between -20 to 120 mV.
The CalEr occurs when the slope calculated from the last calibration was out of range. CalEr can be
caused by leaks, a faulty diffusor or sensor cell, erroneous test gas values, or not enough test gas time. Each test gas should be supplied for at least three minutes.
1. Flowmeter set incorrectly.
The flowmeter for each probe must be set individually. Flow should be 5 scfh (2.4 L/min).
2. Wiring fault.
Confirm proper wiring and continuity between MPS and electronics.
3. Piping fault. Faulty gas line or regulator. Check gas line, valves, and regulators for blockage or corrosion. Repair or replace as needed.
4. Solenoid fault. Verify nominal 24 VDC at HI GAS, LOW GAS, IN CAL, and CAL RET connections. Voltages should drop to about 4 VDC. If voltage is present but solenoid does not work, replace the solenoid.
5. Termination board fau lt. Verify 24 VDC at J11 on termination board. Repair or replace termination board or connectors as needed.
6. Power supply fault. Verify power supply fuses and output are good and that line voltage is present at J1. Repair or replace the power supply as needed.
7. Power fault. Check fuses, mains, and circuit breakers. Repair or replace as needed.
6-8 Troubleshooting Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
6
Instruction Manual
IB-106-300NH Rev. 4.2
July 2002
6-8 PERFORMANCE PROBLEM (PROCESS
RESPONSE IS SUSPECT)
readings may not always agree with known
O
2
process conditions. Such a discrepancy can be the first sign of a problem either in the process or the World Class 3000. The O
display will
2
Table 6-6. Performance Problem Troubleshooting
Problem
Cause
Corrective Action
Status is OK. Cell mV is -20 to 120 mV (normal).
display is stable but not expected value.
O
2
Such a condition occurs during various kinds of leaks and data output faults.
1. Mounting flange leak.
Reseal the flange, and tighten bolts properly.
2. Test gas line leak.
Since the test gas line is under positive pressure, the line can be tested with a bubbling liquid such as SNOOP
TM
. Repair or replace as needed.
3. Silicon hose break. Leaks may occur in the silicon rubber hose in the probe junction box.
Replace hose.
4. Air ingress from leaky duct.
Check condition of duct, gas lines, and fittings. If duct has air ingress upstream of probe, re-site the probe or fix the leak.
5. Analog output or recorder fault.
Measure analog output in voltage or milliamps as set up on the analog output board and soft­ware. If analog output is not in range, replace the analog output board in a CRE or the micro­processor board in an IFT. Check recorder function, and repair as needed.
6. Random spiking of the analog output to 0 mA dc.
Check the power supply voltage. If suspect, replace the power supply in the CRE or the power supply board in the IFT.
read between 0 to 99%, but the reading may be unstable. The status line may read OK, and PROBE DATA voltages may read normal.
Refer to Table 6-6 to troubleshoot performance problems.
Status is OK. Cell mV is -20 to 120 mV (normal).
display is unstable.
O
2
1. Process variations.
Analyze the process for even flows of gases or materials. Check the operation of dampers and control valves. Repair process devices, procedures, and flows as needed. Depending on the process, some variation may be normal.
2. Pad to cell connection fault.
Check pad and contact for cleanliness, and clean as needed. Check spring tension, and replace as needed.
3. Grounding fault.
Check all wiring for continuity and connections for cleanliness and lack of corrosion. Repair as needed.
4. Improper line voltage.
Check line voltage circuit for proper polarity and/or "hot" and "neutral" circuitry.
Rosemount Analytical Inc. A Division of Emerson Process Management Troubleshooting 6-9
Instruction Manual
IB-106-300NH Rev. 4.2 July 2002
World Class 3000
6-10 Troubleshooting Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
7
Instruction Manual
IB-106-300NH Rev. 4.2
July 2002
SECTION 7

RETURN OF MATERIAL

7-1 If factory repair of defective equipment is re-
quired, proceed as follows: a. Secure a return authorization number from
a Rosemount Analytical Sales Office or rep­resentative before returning the equipment. Equipment must be returned with complete identification in accordanc e wit h Rosemount instructions or it will not be accepted.
In no event will Rosemount be responsible for equipment returned without proper authorization and ident if ic ati on.
b. Carefully pack defective unit in a sturdy box
with sufficient shock absorbing material to insure that no additional damage will occur during shipping.
c. In a cover letter, describe completely:
1. The symptoms from which it was de­termined that the equipment is faulty.
2. The environment in which the equip­ment has been operating (housing, weather, vibration, dust, etc.).
5. Complete shipping instructions for re­turn of equipment.
6. Reference the return authorization number.
d. Enclose a cover letter and purchase order
and ship the defective equipment according to instructions provided in Rosemount Re­turn Authorization, prepaid, to:
Rosemount Analytical Inc. RMR Department 1201 N. Main Street Orrville, Ohio 44667
If warranty service is requested, the defec­tive unit will be carefully insp ec ted and tested at the factory. If failure was due to conditions listed in the standard Rosemount warranty, the defecti ve unit will be repaired or replaced at Rosemount's option, and an operating unit will be returned to the cus­tomer in accordance with shipping instruc­tions furnished in the cover letter.
3. Site from which equipment was removed.
4. Whether warranty or nonwarranty service is requested.
Rosemount Analytical Inc. A Division of Emerson Process Management Return of Material 7-1
For equipment no longer under warranty, the equipment will be repaired at the factory and returned as directed by the purchase order and shipping instructions.
Instruction Manual
IB-106-300NH Rev. 4.2 July 2002
World Class 3000
7-2 Return of Material Rosemount Analytical Inc. A Division of Emerson Process Management
Instruction Manual
8
IB-106-300NH Rev. 4.2
World Class 3000
SECTION 8

APPENDICES

APPENDIX A. WORLD CLASS 3000 OXYGEN ANALYZER (PROBE) APPENDIX B. HPS 3000 HEATER POWER SUPPLY APPENDIX D. MPS 3000 MULTIPROBE CALIBRATION GAS SEQUENCER APPENDIX E. IFT 3000 INTELLIGENT FIELD TRANSMITTER APPENDIX J. HART COMMUNICATOR MODEL 275D9E IFT 3000 APPLICATIONS
July 2002
Rosemount Analytical Inc. A Division of Emerson Process Management Appendices 8-1
Instruction Manual
Appendix A Rev. 3.8 July 2002
NOTE: NOT ALL PARTS SHOWN ARE AVAILABLE FOR
PURCHASE SEPARATELY. FOR LIST OF AVAILABLE PARTS, SEE TABLE A-3.
1
APPENDIX A
26
20
World Class 3000
1. Heater, Strut, and Backplate Assembly
2. Diffusion Assembly
3. Retainer Screw
4. Cell and Flange
5. Corrugated Seal
6. Probe Tube Assembly
7. Screw
8. Washer
9. Cover Chain Screw
10. Cover Chain
11. Probe Junction Box Cover
12. Cover Gasket
13. Wiring Diagram
14. O-Ring
15. Terminal Block Screws
16. Terminal Block
17. Terminal Block Marker
18. Terminal Block Mounting Plate
3
2
4
19. Probe Junction Box Screws
20. Hose Clamp
21. Hose
22. Gas Connection
23. Seal Cap
24. Label
25. Probe Junction Box
26. Ground Wires
27. Insulating Gasket
28. Washer
29. Screw
18
23
21
19
16
14
20
12
29
10
17
11
13
15
8
7
9
21240005
28
27
25
24
11
22
10
5
6
NOTE: ITEM , CALIBRATION GAS TUBE,
FITS INTO HOLES WHEN PROBE IS
ASSEMBLED.
Figure A-1. Oxygen Analyzer (Probe) Exploded View
A-0 Appendices Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
A
WORLD CLASS 3000 OXYGEN ANALYZER (PROBE )
Instruction Manual
Appendix A Rev. 3.8
July 2002
APPENDIX A, REV. 3.8

DESCRIPTION

A-1 OXYGEN ANALYZER (PROBE) - GENERAL
Read the “Safety instructions for the wiring and installation of this appara­tus” at the front of this Instruction Bulletin. Failure to follow the safety instructions could result in serious injury or death.
PROBE EXTERIOR (SENSING CELL INSTALLED)
The Oxygen Analyzer (Probe), Figure A-1, con­sists of three component groups: probe exterior, inner probe, and probe junction box, Figure A-2.
PROBE INTERIOR
PROBE
JUNCTION
BOX
21240006
Figure A-2. Main Probe Components
Rosemount Analytical Inc. A Division of Emerson Process Management Appendices A-1
Instruction Manual
Appendix A Rev. 3.8 July 2002
World Class 3000
Table A-1. Specifications for Oxygen Analyzing Equipment.
1, 2
Probe lengths, nominal ..................................................................18 inches (457 mm), 3 feet (0.91 m), 6 feet
(1.83 m), 9 feet (2.74 m), or 12 feet (3.66 m), depending on duct dimensions
Temperature limits in process
measurement area ............................................................50° to 1300°F (10° to 704°C)
Standard/current output..................................................................4-20 mA dc signal (factory set)
indication (Digital display
O
2
and analog output).............................................................0.1% O
or ±3% of reading, whichever is
2
greater using Rosemount calibration gases
System speed of response.............................................................less than 3 seconds (amplifier output)
Resolution sensitivity......................................................................0.01% O
transmitted signal
2
HPS 3000 housing..........................................................................NEMA 4X (IP56)
Probe reference air flow .................................................................2 scfh (56.6 L/hr) clean, dry, instrument quality
air (20.95% O
), regulated to 5 psi (34 kPa)
2
Calibration gas mixtures.................................................................Rosemount Hagan Calibration Gas Kit Part No.
6296A27G01 contains 0.4% O
8% O
Nominal
2N2
Nominal and
2N2
Calibration gas flow ........................................................................5 scfh (141.6 L/hr)
HPS 3000 Power supply...............................................................100/110/220 ±10% Vac at 50/60 Hz
HPS 3000 Power requirement......................................................200 VA
HPS 3000 Ambient Operating Temperature ................................32° to 120°F (0° to 50°C)
Ambient operating temperature (Probe Junction Box)...................300°F (150°C) max
Approximate shipping weights:
18 inch (457 mm) package................................................55 pounds (24.97 kg)
3 foot (0.91 m) package.....................................................60 pounds (27.24 kg)
6 foot (1.83 m) package.....................................................65 pounds (29.51 kg)
9 foot (2.74 m) package.....................................................72 pounds (32.66 kg)
12 foot (3.66 m) package...................................................78 pounds (35.38 kg)
1
All static performance characteristics are with operating variables constant.
2
Equipment ordered utilizing this document as reference will be supplied to the USA standard design. Custom­ers requiring the EEC standard design sho ul d reques t the EEC doc umentation and utilize its ordering data. Temperatures over 1000°F (537°C) may affect the ease of field cell replaceability.
A-2 Appendices Rosemount Analytical Inc. A Division of Emerson Process Management
Loading...