Elektro-Automatik PS 9000 3U, PS 9040-170 3U, PS 9080-170 3U, PS 9200-70 3U, PS 9360-40 3U User guide

...
Operating Guide
PS 9000 3U
DC High Efficiency Power Supply
Attention! This document is only valid for devices with TFT dis-
play and rmwares “KE: 3.07” (standard version) or “KE: 2.11” (GPIB, 3W) and “HMI: 2.03” or higher. For availability of up-
Elektro-Automatik
dates for your device check our
website or contact us.
Doc ID: PS93UTEN Revision: 09 Date: 02/2020
TABLE OF CONTENTS
GENERAL
1
1.1 About this document ......................................5
1.1.1 Retention and use ..........................................5
1.1.2 Copyright ........................................................5
1.1.3 Validity ............................................................5
1.1.4 Explanation of symbols ..................................5
1.2 Warranty .........................................................5
1.3 Limitation of liability ........................................5
1.4 Disposal of equipment ...................................6
1.5 Product key ....................................................6
1.6 Intended usage ..............................................6
1.7 Safety .............................................................7
1.7.1 Safety notices .................................................7
1.7.2 Responsibility of the user...............................8
1.7.3 Responsibility of the operator .......................8
1.7.4 User requirements .........................................8
1.7.5 Alarm signals ..................................................9
1.8 Technical data ................................................9
1.8.1 Approved operating conditions ......................9
1.8.2 General technical data ...................................9
1.8.3 Specic technical data (400 V AC models) . 10
1.8.4 Specic technical data (208 V AC models) . 18
1.8.5 Views ............................................................26
1.8.6 Control elements ..........................................29
1.9 Construction and function ............................30
1.9.1 General description ......................................30
1.9.2 Block diagram ..............................................30
1.9.3 Scope of delivery .........................................31
1.9.4 Accessories ..................................................31
1.9.5 Options .........................................................31
1.9.6 The control panel (HMI) ...............................32
1.9.7 USB port .......................................................34
1.9.8 Ethernet port ................................................34
1.9.9 Analog interface ...........................................35
1.9.10 Share Bus connection .................................35
1.9.11 Sense connector (remote sensing) .............35
1.9.12 GPIB port (optional) .....................................35
INSTALLATION &
2
COMMISSIONING
2.1 Transport and storage .................................36
2.1.1 Transport ......................................................36
2.1.2 Packaging ....................................................36
2.1.3 Storage .........................................................36
2.2 Unpacking and visual check ........................36
2.3 Installation ....................................................36
2.3.1 Safety procedures before installation and
use ................................................................36
2.3.2 Preparation ...................................................37
2.3.3 Installing the device .....................................37
2.3.4 Connection to AC supply .............................38
2.3.5 Connection to DC loads ...............................41
PS 9000 3U Series
2.3.6 Connection of remote sensing ....................42
2.3.7 Grounding of the DC output ........................43
2.3.8 Connecting the “Share” bus ........................43
2.3.9 Connecting the analog interface .................43
2.3.10 Connecting the USB port .............................43
2.3.11 Initial commission .........................................44
2.3.12 Initial network setup .....................................44
2.3.13 Commission after a rmware update or a
long period of non-use .................................44
OPERATION AND APPLICATION
3
3.1 Important notes ............................................45
3.1.1 Personal safety ............................................45
3.1.2 General .........................................................45
3.2 Operating modes .........................................45
3.2.1 Voltage regulation / Constant voltage .........45
3.2.2 Current regulation / constant current / current
limiting ..........................................................46
3.2.3 Power regulation / constant power / power
limiting ..........................................................46
3.3 Alarm conditions ..........................................47
3.3.1 Power Fail ...................................................47
3.3.2 Overtemperature ..........................................47
3.3.3 Overvoltage ..................................................47
3.3.4 Overcurrent ..................................................47
3.3.5 Overpower ....................................................47
3.4 Manual operation .........................................48
3.4.1 Switching the device on ...............................48
3.4.2 Switching the device o󰀨 ...............................48
3.4.3 Conguration in the setup menu .................48
3.4.4 Adjustment limits ..........................................53
3.4.5 Manual adjustment of set values .................54
3.4.6 Switching the main screen view ..................54
3.4.7 The quick menu ...........................................55
3.4.8 Switching the DC output on or o󰀨 ...............55
3.5 Remote control .............................................56
3.5.1 General .........................................................56
3.5.2 Control locations ..........................................56
3.5.3 Remote control via a digital interface ..........56
3.5.4 Remote control via the analog interface
(AI) ................................................................57
3.6 Alarms and monitoring .................................61
3.6.1 Denition of terms ........................................61
3.6.2 Device alarm handling .................................61
3.7 Control panel (HMI) lock ..............................62
3.8 Loading and saving a user prole ...............63
3.9 Other applications ........................................64
3.9.1 Parallel operation in Share bus mode .........64
3.9.2 Series connection ........................................65
3.9.3 Operation as battery charger .......................65
3.9.4 Two quadrants operation (2QO) ..................66
EA Elektro-Automatik GmbH Helmholtzstr. 31-37 • 41747 Viersen Germany
Fon: +49 2162 / 3785-0 Fax: +49 2162 / 16230
www.elektroautomatik.de
ea1974@elektroautomatik.de
Page 3
SERVICE AND MAINTENANCE
4
4.1 Maintenance / cleaning ................................68
4.2 Fault nding / diagnosis / repair...................68
4.2.1 Firmware updates ........................................68
4.3 Calibration (readjustment) ...........................69
4.3.1 Preface .........................................................69
4.3.2 Preparation ...................................................69
4.3.3 Calibration procedure ..................................69
CONTACT AND SUPPORT
5
5.1 Repairs .........................................................71
5.2 Contact options ............................................71
PS 9000 3U Series
EA Elektro-Automatik GmbH Helmholtzstr. 31-37 • 41747 Viersen Germany
Fon: +49 2162 / 3785-0 Fax: +49 2162 / 16230
www.elektroautomatik.de
ea1974@elektroautomatik.de
Page 4
PS 9000 3U Series

1. General

1.1 About this document

1.1.1 Retention and use

This document is to be kept in the vicinity of the equipment for future reference and explanation of the operation of the device. This document is to be delivered and kept with the equipment in case of change of location and/or user.

1.1.2 Copyright

Reprinting, copying, also partially, usage for other purposes as foreseen of this manual are forbidden and breach may lead to legal process.

1.1.3 Validity

This manual is valid for the following equipment with colour TFT display, including derived variants.
Model Article nr.. Model Article nr.. Model Article nr.. PS 9040-170 3U 06230250 PS 9080-340 3U 06230257 PS 9080-510 3U 06230264 PS 9080-170 3U 06230251 PS 9200-140 3U 06230258 PS 9200-210 3U 06230265 PS 9200-70 3U 06230252 PS 9360-80 3U 06230259 PS 9360-120 3U 06230266 PS 9360-40 3U 06230253 PS 9500-60 3U 06230260 PS 9500-90 3U 06230267 PS 9500-30 3U 06230254 PS 9750-40 3U 06230261 PS 9750-60 3U 06230268 PS 9750-20 3U 06230255 PS 91000-30 3U 06230262 PS 91000-40 3U 06230270 PS 9040-340 3U 06230256 PS 9040-510 3U 06230263 PS 91500-30 3U 06230269

1.1.4 Explanation of symbols

Warning and safety notices as well as general notices in this document are shown in a box with a symbol as follows:
Symbol for a life threatening danger
Symbol for general safety notices (instructions and damage protection bans) or important infor­mation for operation
Symbol for general notices

1.2 Warranty

EA Elektro-Automatik guarantees the functional competence of the applied technology and the stated performance parameters. The warranty period begins with the delivery of free from defects equipment.
Terms of guarantee are included in the general terms and conditions (TOS) of EA Elektro-Automatik.

1.3 Limitation of liability

All statements and instructions in this manual are based on current norms and regulations, up-to-date technology and our long term knowledge and experience. The manufacturer accepts no liability for losses due to:
• Usage for purposes other than designed
• Use by untrained personnel
• Rebuilding by the customer
• Technical changes
• Use of not authorized spare parts
The actual delivered device(s) may di󰀨er from the explanations and diagrams given here due to latest technical changes or due to customized models with the inclusion of additionally ordered options.
EA Elektro-Automatik GmbH Helmholtzstr. 31-37 • 41747 Viersen Germany
Fon: +49 2162 / 3785-0 Fax: +49 2162 / 16230
www.elektroautomatik.de
ea1974@elektroautomatik.de
Page 5
PS 9000 3U Series

1.4 Disposal of equipment

A piece of equipment which is intended for disposal must, according to European laws and regulations (ElektroG, WEEE) be returned to the manufacturer for scrapping, unless the person operating the piece of equipment or an­other, delegated person is conducting the disposal. Our equipment falls under these regulations and is accordingly marked with the following symbol:

1.5 Product key

Decoding of the product description on the label, using an example:
PS 9 080 - 510 3U zzz
Field for identication of installed options and/or special models
S01...S0x = Special models HS = High Speed option installed 3W = Option 3W installed (GPIB port instead of Ethernet port) US208V = 208 V model for US market
Construction (not stated everywhere): 3U = 19" frame with 3U
Maximum current of the device in Ampere
Maximum voltage of the device in Volt
Series : 9 = Series 9000
Type identication:
PS = Power Supply, usually programmable

1.6 Intended usage

The equipment is intended to be used, if a power supply or battery charger, only as a variable voltage and current source, or, if an electronic load, only as a variable current sink.
Typical application for a power supply is DC supply to any relevant user, for a battery charger the charging of var­ious battery types and for electronic loads the replacement of an ohmic resistor by an adjustable DC current sink in order to load relevant voltage and current sources of any type.
• Claims of any sort due to damage caused by non-intended usage will not be accepted.
• All damage caused by non-intended usage is solely the responsibility of the operator.
EA Elektro-Automatik GmbH Helmholtzstr. 31-37 • 41747 Viersen Germany
Fon: +49 2162 / 3785-0 Fax: +49 2162 / 16230
www.elektroautomatik.de
ea1974@elektroautomatik.de
Page 6

1.7 Safety

1.7.1 Safety notices

Mortal danger - Hazardous voltage
• Electrical equipment operation means that some parts can be under dangerous voltage. Therefore all parts under voltage must be covered! This basically applies to all models, though 40 V models according to SELV can not generate hazardous DC voltage!
• All work on connections must be carried out under zero voltage (output not connected to
load) and may only be performed by qualied and informed persons. Improper actions
can cause fatal injury as well as serious material damage!
• Never touch cables or connectors directly after unplugging from mains supply as the danger of electric shock remains!
• Never touch the contacts on DC output terminal directly after switching o󰀨 the DC
output, because there still can dangerous voltage present, sinking more or less slowly depending on the load! There also can be dangerous potential between negative DC output to PE or positive DC output to PE due to charged X capacitors.
• Always follow 5 safety rules when working with electric devices:
• Disconnect completely
• Secure against reconnection
• Verify that the system is dead
• Carry out earthing and short-circuiting
• Provide protection from adjacent live parts
PS 9000 3U Series
• The equipment must only be used as intended
• The equipment is only approved for use within the connection limits stated on the product label.
• Do not insert any object, particularly metallic, through the ventilator slots
• Avoid any use of liquids near the equipment. Protect the device from wet, damp and conden-
sation.
• For power supplies and battery chargers: do not connect users, particularly low resistance, to devices under power; sparking may occur which can cause burns as well as damage to the equipment and to the user.
• For electronic loads: do not connect power sources to equipment under power, sparking may occur which can cause burns as well as damage to the equipment and to the source.
• ESD regulations must be applied when plugging interface cards or modules into the relative slot
• Interface cards or modules may only be attached or removed after the device is switched o󰀨.
It isn’t necessary to open the device.
• Do not connect external power sources with reversed polarity to DC input or outputs! The equipment will be damaged.
• For power supply devices: avoid where possible connecting external power sources to the DC output, and never those that can generate a higher voltage than the nominal voltage of the device.
• For electronic loads: do not connect a power source to the DC input which can generate a voltage more than 120% of the nominal input voltage of the load. The equipment isn’t protected against over voltage and may be irreparably damaged.
• Never insert a network cable which is connected to Ethernet or its components into the mas­ter-slave socket on the back side of the device!
• Always congure the various protecting features against overvoltage overpower etc. for sen­sitive loads to what the currently used application requires
EA Elektro-Automatik GmbH Helmholtzstr. 31-37 • 41747 Viersen Germany
Fon: +49 2162 / 3785-0 Fax: +49 2162 / 16230
www.elektroautomatik.de
ea1974@elektroautomatik.de
Page 7
PS 9000 3U Series

1.7.2 Responsibility of the user

The equipment is in industrial operation. Therefore the operators are governed by the legal safety regulations.
Alongside the warning and safety notices in this manual the relevant safety, accident prevention and environmental
regulations must also be applied. In particular the users of the equipment:
• must be informed of the relevant job safety requirements
• must work to the dened responsibilities for operation, maintenance and cleaning of the equipment
• before starting work must have read and understood the operating manual
• must use the designated and recommended safety equipment.

1.7.3 Responsibility of the operator

Operator is any natural or legal person who uses the equipment or delegates the usage to a third party, and is responsible during its usage for the safety of the user, other personnel or third parties.
The equipment is in industrial operation. Therefore the operators are governed by the legal safety regulations.
Alongside the warning and safety notices in this manual the relevant safety, accident prevention and environmental
regulations must also be applied. In particular the operator has to
• be acquainted with the relevant job safety requirements
• identify other possible dangers arising from the specic usage conditions at the work station via a risk assessment
• introduce the necessary steps in the operating procedures for the local conditions
• regularly control that the operating procedures are current
• update the operating procedures where necessary to reect changes in regulation, standards or operating con-
ditions.
• dene clearly and unambiguously the responsibilities for operation, maintenance and cleaning of the equipment.
• ensure that all employees who use the equipment have read and understood the manual. Furthermore the users
are to be regularly schooled in working with the equipment and the possible dangers.
• provide all personnel who work with the equipment with the designated and recommended safety equipment Furthermore, the operator is responsible for ensuring that the device is at all times technically t for use.

1.7.4 User requirements

Any activity with equipment of this type may only be performed by persons who are able to work correctly and reliably and satisfy the requirements of the job.
• Persons whose reaction capability is negatively inuenced by e.g. drugs, alcohol or medication may not operate the equipment.
• Age or job related regulations valid at the operating site must always be applied.
Danger for unqualied users
Improper operation can cause person or object damage. Only persons who have the necessary training, knowledge and experience may use the equipment.
Delegated persons are those who have been properly and demonstrably instructed in their tasks and the atten­dant dangers.
Qualied persons are those who are able through training, knowledge and experience as well as knowledge of the specic details to carry out all the required tasks, identify dangers and avoid personal and other risks.
EA Elektro-Automatik GmbH Helmholtzstr. 31-37 • 41747 Viersen Germany
Fon: +49 2162 / 3785-0 Fax: +49 2162 / 16230
www.elektroautomatik.de
ea1974@elektroautomatik.de
Page 8
PS 9000 3U Series

1.7.5 Alarm signals

The equipment o󰀨ers various possibilities for signalling alarm conditions, however, not for danger situations. The signals may be optical (on the display as text) acoustic (piezo buzzer) or electronic (pin/status output of an analog interface). All alarms will cause the device to switch o󰀨 the DC output.
The meaning of the signals is as follows:
Signal OT (OverTemperature)
• Overheating of the device
• DC output will be switched o󰀨 temporarily
• Non-critical
Signal OVP (OverVoltage)
• Overvoltage shutdown of the DC output due to high voltage entering the device or gen­erated by the device itself due to a defect
• Critical! The device and/or the load could be damaged
Signal OCP (OverCurrent) Signal OPP (OverPower) Signal PF
(Power Fail)
• Shutdown of the DC output due to excess of the preset limit
• Non-critical, protects the load from excessive current consumption
• Shutdown of the DC output due to excess of the preset limit
• Non-critical, protects the load from excessive power consumption
• DC output shutdown due to AC undervoltage or defect of the AC input circuit
• Critical on overvoltage! AC mains input circuit could take damage

1.8 Technical data

1.8.1 Approved operating conditions

• Use only inside dry buildings
• Ambient temperature 0-50°C (32-122 °F)
• Operational altitude: max. 2000 m (1.242 mi) above sea level
• Maximum 80% humidity, non-condensing

1.8.2 General technical data

Display: Color TFT display, 480pt x 128pt Controls: 2 rotary knobs with button function, 5 pushbuttons The nominal values for the device determine the maximum adjustable ranges.
EA Elektro-Automatik GmbH Helmholtzstr. 31-37 • 41747 Viersen Germany
Fon: +49 2162 / 3785-0 Fax: +49 2162 / 16230
www.elektroautomatik.de
ea1974@elektroautomatik.de
Page 9
PS 9000 3U Series
1.8.3 Specic technical data (400 V AC models)
Model 400 V
3.3 kW / 5 kW
AC Input
Voltage (L-L) 340...460 V AC, 45 - 65 Hz
Input connection 2ph,PE Leak current < 3.5 mA Power factor > 0.99
DC Output
Max. output voltage U Max. output current I Max. output power P Overvoltage protection range 0...44 V 0...88 V 0...220 V 0...396 V 0...550 V Overcurrent protection range 0...187 A 0...187 A 0...77 A 0...44 A 0...33 A Overpower protection range 0…3.63 kW 0…5.5 kW 0…5.5 kW 0…5.5 kW 0…5.5 kW Temperature coe󰀩cient for set
values Δ/K Output capacitance (approx.) 8500 μF 8500 μF 2500 μF 400 μF 250 μF
Voltage regulation
Adjustment range 0...40.8 V 0...81.6 V 0...204 V 0...367.2 V 0...510 V
Accuracy
(1
(at 23±5°C / 73±9°F)) < 0.1% U Line regulation at ±10% ΔU Load regulation at 0...100% load < 0.05% U Rise time 10...90% Max. 30 ms Max. 30 ms Max. 30 ms Max. 30 ms Max. 30 ms Settling time after load step < 1.5 ms < 1.5 ms < 1.5 ms < 1.5 ms < 1.5 ms Display: Resolution See section „1.9.6.4. Resolution of the displayed values“ Display: Accuracy
(2
Ripple
Remote sensing compensation Max. 5% U
Fall time at no load after switching
DC output o󰀨
Current regulation
Adjustment range 0...173.4 A 0...173.4 A 0...71.4 A 0...40.8 A 0...30.6 A
Accuracy
(1
(at 23±5°C / 73±9°F)) < 0.2% I Line regulation at ±10% ΔU Load regulation at 0...100% ΔU
(2
Ripple
Display: Resolution See section „1.9.6.4. Resolution of the displayed values“ Display: Accuracy
Power regulation
Adjustment range 0…3.36 kW 0…5.1 kW 0…5.1 kW 0…5.1 kW 0…5.1 kW
Accuracy
(1
(at 23±5°C / 73±9°F)) < 1% P Line regulation at ±10% ΔU Load reg. at 10-90% ΔU Display: Resolution See section „1.9.6.4. Resolution of the displayed values“ Display: Accuracy E󰀩ciency
(3
Max
Max
Max
AC
(4
AC
(4
AC
* ΔI
OUT
(4
PS 9040-170 PS 9080-170 PS 9200-70 PS 9360-40 PS 9500-30
40 V 80 V 200 V 360 V 500 V 170 A 170 A 70 A 40 A 30 A
3.3 kW 5 kW 5 kW 5 kW 5 kW
Voltage / current: 100 ppm
Max
Max
Max
< 0.1% U
Max
< 0.02% U < 0.05% U
≤ 0.2% U
Max
< 200 mVPP < 16 mV
RMS
Max. 5% U
Max
< 0.02% U
≤ 0.2% U
Max
< 200 mVPP < 16 mV
RMS
- Down from 100% to <60 V: less than 10 s
Max
Max
Max
Max
Max
< 0.2% I
Max
< 0.05% I < 0.15% I < 80 mA
≤ 0.2% I
< 1% P
RMS
Max
Max
< 0.05% P < 0.75% P
≤ 0.8% P
Max
OUT
OUT
Max
< 0.05% I < 0.15% I < 80 mA
≤ 0.2% I
RMS
Max
Max
< 0.05% P < 0.75% P
≤ 0.75% P ≈ 93% ≈ 93% ≈ 95% ≈ 95% ≈ 95,5%
Max
Max
Max
Max
Max
Max
Max
< 0.1% U
Max
< 0.02% U < 0.05% U
≤ 0.2% U
Max
< 300 mVPP < 40 mV
RMS
Max. 5% U
< 0.2% I < 0.05% I < 0.15% I < 22 mA
≤ 0.2% I
< 1% P
Max
Max
Max
RMS
Max
Max
< 0.05% P < 0.75% P
≤ 0.8% P
Max
Max
Max
Max
Max
Max
< 0.1% U
Max
< 0.02% U < 0.05% U
≤ 0.2% U
Max
< 550 mVPP < 65 mV
RMS
Max. 5% U
< 0.2% I < 0.05% I < 0.15% I < 5.2 mA
≤ 0.2% I
< 1% P
Max
Max
Max
RMS
Max
Max
< 0.05% P < 0.75% P
≤ 0.8% P
Max
Max
Max
Max
Max
Max
< 0.1% U < 0.02% U < 0.05% U
≤ 0.2% U < 350 mVPP
< 70 mV Max. 5% U
< 0.2% I < 0.05% I < 0.15% I < 16 mA
≤ 0.2% I
< 1% P < 0.05% P < 0.75% P
≤ 0.8% P
Max
Max
Max
Max
RMS
Max
Max
Max
Max
RMS
Max
Max
Max
Max
Max
(1 Related to the nominal values, the accuracy denes the maximum deviation between an adjusted values and the true (actual) value. Example: a 80 V model has min. 0.1% voltage accuracy, that is 80 mV. When adjusting the voltage to 5 V, the actual value is allowed to di󰀨er max. 80 mV, which means it might be between 4.92 V and 5.08 V. (2 RMS value: LF 0...300 kHz, PP value: HF 0...20MHz (3 Typical value at 100% output voltage and 100% power (4 The display error adds to the error of the related actual value on the DC output
EA Elektro-Automatik GmbH Helmholtzstr. 31-37 • 41747 Viersen Germany
Fon: +49 2162 / 3785-0 Fax: +49 2162 / 16230
www.elektroautomatik.de
ea1974@elektroautomatik.de
Page 10
PS 9000 3U Series
Model 400 V
3.3 kW / 5 kW
Analog interface
Set value inputs U, I, P Actual value outputs U, I Control signals DC on/o󰀨, remote control on/o󰀨 Status signals CV, OVP, OT, OCP, OPP, PF, DC on/o󰀨 Galvanic isolation to the device Max. 725 V DC Sample rate of inputs 500 Hz
Insulation Allowed oat (potential shift) on the DC output:
Negative terminal to PE Max. ±400 V DC ±400 V DC ±400 V DC ±400 V DC ±725 V DC Positive terminal to PE Max. ±400 V DC ±400 V DC ±600 V DC ±600 V DC ±1000 V DC
Miscellaneous
Cooling Temperature controlled fans, front inlet, rear exhaust Ambient temperature 0..50°C (32...122°F) Storage temperature -20...70°C (-4...158°F) Humidity < 80%, not condensing
Standards
Overvoltage category 2
Protection class 1 Pollution degree 2
Operational altitude <2000 m (1,242 mi)
Digital interfaces
Featured 1x USB-B, 1x Ethernet (2, 1x GPIB (optional with option 3W) Galvanic isolation from device Max. 725 V DC
Connectors
Rear side Share Bus, DC output, AC input, remote sensing, analog interface, USB-B, Ethernet
Dimensions
Enclosure (WxHxD) 19“ x 3 U x 609 mm (24”) Total (WxHxD) 483 x 133 x 714 mm (19” x 5,2” x 28,1”)
Weight
Article number
(1
(3
PS 9040-170 PS 9080-170 PS 9200-70 PS 9360-40 PS 9500-30
EN 61010-1:2010 EMC TÜV approved acc. IEC 61000-6-2:2005 and IEC 61000-6-3:2006 Class B
17 kg (37.5 lb) 17 kg (37.5 lb) 17 kg (37.5 lb) 17 kg (37.5 lb) 17 kg (37.5 lb) 06230250 06230251 06230252 06230253 06230254
(1 For technical specications of the analog interface see „3.5.4.4 Analog interface specication“ on page 58 (2 Only in the standard version (3 Article number of the standard EU version, devices with options will have a di󰀨erent number
EA Elektro-Automatik GmbH Helmholtzstr. 31-37 • 41747 Viersen Germany
Fon: +49 2162 / 3785-0 Fax: +49 2162 / 16230
www.elektroautomatik.de
ea1974@elektroautomatik.de
Page 11
PS 9000 3U Series
Model 400 V
5 kW / 6.6 kW / 10 kW
AC Input
Voltage (L-L) 340...460 V AC, 45 - 65 Hz
Input connection 2ph,PE 3ph,PE 3ph,PE 3ph,PE 3ph,PE Leak current < 3.5 mA < 3.5 mA < 3.5 mA < 3.5 mA < 3.5 mA Power factor > 0.99 > 0.99 > 0.99 > 0.99 > 0.99
DC Output
Max. output voltage U Max. output current I Max. output power P Overvoltage protection range 0...825 V 0...44 V 0...44 V 0...88 V 0...220 V Overcurrent protection range 0...22 A 0...374 A 0...561 A 0...374 A 0...154 A Overpower protection range 0…5.5 kW 0…7.26 kW 0…11 kW 0…11 kW 0…11 kW Temperature coe󰀩cient for set
values Δ/K Output capacitance (approx.) 100 μF 16900 μF 25380 μF 16900 μF 5040 μF
Voltage regulation
Adjustment range 0...765 V 0...40.8 V 0...40.8 V 0...81.6 V 0...204 V
Accuracy
(1
(at 23±5°C / 73±9°F)) < 0.1% U Line regulation at ±10% ΔU Load regulation at 0...100% load < 0.05% U Rise time 10...90% Max. 30 ms Max. 30 ms Max. 30 ms Max. 30 ms Max. 30 ms Settling time after load step < 1.5 ms < 1.5 ms < 1.5 ms < 1.5 ms < 1.5 ms Display: Resolution See section „1.9.6.4. Resolution of the displayed values“ Display: Accuracy
(2
Ripple
Remote sensing compensation Max. 5% U
Fall time at no load after switching
DC output o󰀨
Current regulation
Adjustment range 0...20.4 A 0...346.8 A 0...520.2 A 0...346.8 A 0...142.8 A
Accuracy
(1
(at 23±5°C / 73±9°F)) < 0.2% I Line regulation at ±10% ΔU Load regulation at 0...100% ΔU
(2
Ripple
Display: Resolution See section „1.9.6.4. Resolution of the displayed values“ Display: Accuracy
Power regulation
Adjustment range 0…5.1 kW 0…6.72 kW 0…10.2 kW 0…10.2 kW 0…10.2 kW
Accuracy
(1
(at 23±5°C / 73±9°F)) < 1% P Line regulation at ±10% ΔU Load reg. at 10-90% ΔU Display: Resolution See section „1.9.6.4. Resolution of the displayed values“ Display: Accuracy E󰀩ciency
(3
Max
Max
Max
AC
(4
AC
(4
AC
* ΔI
OUT
(4
PS 9750-20 PS 9040-340 PS 9040-510 PS 9080-340 PS 9200-140
750 V 40 V 40 V 80 V 200 V 20 A 340 A 510 A 340 A 140 A 5 kW 6.6 kW 10 kW 10 kW 10 kW
Voltage / current: 100 ppm
Max
Max
RMS
Max
< 0.1% U
Max
< 0.02% U < 0.05% U
≤ 0.2% U
Max
< 320 mVPP < 25 mV
RMS
Max. 5% U
Max
< 0.02% U
≤ 0.2% U
Max
< 800 mVPP < 200 mV
Down from
100% to <60 V:
- -
less than 10 s
Max
Max
Max
Max
< 0.2% I
Max
< 0.05% I < 0.15% I < 160 mA
≤ 0.2% I
< 1% P
Max
Max
< 0.05% P < 0.75% P
≤ 0.7% P
Max
OUT
OUT
Max
< 0.05% I < 0.15% I < 16 mA
≤ 0.2% I
RMS
Max
Max
< 0.05% P < 0.75% P
≤ 0.8% P
Max
≈ 94% ≈ 93% ≈ 93% ≈ 93% ≈ 95%
Max
Max
Max
Max
Max
RMS
Max
Max
< 0.1% U
Max
< 0.02% U < 0.05% U
≤ 0.2% U
Max
< 320 mVPP < 25 mV
RMS
Max. 5% U
< 0.2% I < 0.05% I < 0.15% I < 120 mA
≤ 0.2% I
< 1% P
Max
Max
Max
RMS
Max
Max
< 0.05% P < 0.75% P
≤ 0.7% P
Max
< 0.1% U < 0.02% U
Max
< 0.05% U
Max
≤ 0.2% U < 320 mVPP
< 25 mV Max. 5% U
Max
Down from
100% to <60 V: less than 10 s
< 0.2% I < 0.05% I < 0.15% I < 160 mA
≤ 0.2% I
< 1% P < 0.05% P
Max
< 0.75% P
Max
≤ 0.8% P
Max
RMS
Max
Max
Max
Max
Max
Max
Max
Max
Max
Max
RMS
Max
Max
< 0.1% U < 0.02% U < 0.05% U
≤ 0.2% U < 300 mVPP
< 40 mV Max. 5% U
Down from
100% to <60 V: less than 10 s
< 0.2% I < 0.05% I < 0.15% I < 44 mA
≤ 0.2% I
< 1% P < 0.05% P < 0.75% P
≤ 0.85% P
Max
Max
Max
Max
RMS
Max
Max
Max
Max
RMS
Max
Max
Max
Max
Max
(1 Related to the nominal values, the accuracy denes the maximum deviation between an adjusted values and the true (actual) value. Example: a 80 V model has min. 0.1% voltage accuracy, that is 80 mV. When adjusting the voltage to 5 V, the actual value is allowed to di󰀨er max. 80 mV, which means it might be between 4.92 V and 5.08 V. (2 RMS value: LF 0...300 kHz, PP value: HF 0...20MHz (3 Typical value at 100% output voltage and 100% power (4 The display error adds to the error of the related actual value on the DC output
EA Elektro-Automatik GmbH Helmholtzstr. 31-37 • 41747 Viersen Germany
Fon: +49 2162 / 3785-0 Fax: +49 2162 / 16230
www.elektroautomatik.de
ea1974@elektroautomatik.de
Page 12
PS 9000 3U Series
Model 400 V
5 kW / 6.6 kW / 10 kW
Analog interface
Set value inputs U, I, P Actual value outputs U, I Control signals DC on/o󰀨, remote control on/o󰀨 Status signals CV, OVP, OT, OCP, OPP, PF, DC on/o󰀨 Galvanic isolation to the device Max. 725 V DC Sample rate of inputs 500 Hz
Insulation Allowed oat (potential shift) on the DC output:
Negative terminal to PE Max. ±725 V DC ±400 V DC ±400 V DC ±400 V DC ±400 V DC Positive terminal to PE Max. ±1000 V DC ±400 V DC ±400 V DC ±400 V DC ±600 V DC
Miscellaneous
Cooling Temperature controlled fans, front inlet, rear exhaust Ambient temperature 0..50°C (32...122°F) Storage temperature -20...70°C (-4...158°F) Humidity < 80%, not condensing
Standards
Overvoltage category 2
Protection class 1 Pollution degree 2
Operational altitude <2000 m (1,242 mi)
Digital interfaces
Featured 1x USB-B, 1x Ethernet (2, 1x GPIB (optional with option 3W) Galvanic isolation from device Max. 725 V DC
Connectors
Rear side Share Bus, DC output, AC input, remote sensing, analog interface, USB-B, Ethernet
Dimensions
Enclosure (WxHxD) 19“ x 3 U x 609 mm (24”) Total (WxHxD) 483 x 133 x 714 mm (19” x 5,2” x 28,1”)
Weight
Article number
(1
(3
PS 9750-20 PS 9040-340 PS 9040-510 PS 9080-340 PS 9200-140
EN 61010-1:2010 EMC TÜV approved acc. IEC 61000-6-2:2005 and IEC 61000-6-3:2006 Class B
17 kg (37.5 lb) 24 kg (52.9 lb) 30 kg (66.1 lb) 24 kg (52.9 lb) 24 kg (52.9 lb) 06230255 06230256 06230263 06230257 06230258
(1 For technical specications of the analog interface see „3.5.4.4 Analog interface specication“ on page 58 (2 Only in the standard version (3 Article number of the standard EU version, devices with options will have a di󰀨erent number
EA Elektro-Automatik GmbH Helmholtzstr. 31-37 • 41747 Viersen Germany
Fon: +49 2162 / 3785-0 Fax: +49 2162 / 16230
www.elektroautomatik.de
ea1974@elektroautomatik.de
Page 13
PS 9000 3U Series
Model 400 V
10 kW / 15 kW
AC Input
Voltage (L-L) 340...460 V AC, 45 - 65 Hz
Input connection 3ph,PE Leak current < 3.5 mA Power factor > 0.99
DC Output
Max. output voltage U Max. output current I Max. output power P Overvoltage protection range 0...396 V 0...550 V 0...825 V 0...1100 V 0...88 V Overcurrent protection range 0...88 A 0...66 A 0...44 A 0...33 A 0...561 A Overpower protection range 0…11 kW 0…11 kW 0…11 kW 0…11 kW 0…16.5 kW Temperature coe󰀩cient for set
values Δ/K Output capacitance (approx.) 800 μF 500 μF 210 μF 127 μF 25380 μF
Voltage regulation
Adjustment range 0...367.2 V 0...510 V 0...765 V 0...1020 V 0...81.6 V
Accuracy
(1
(at 23±5°C / 73±9°F)) < 0.1% U Line regulation at ±10% ΔU Load regulation at 0...100% load < 0.05% U Rise time 10...90% Max. 30 ms Max. 30 ms Max. 30 ms Max. 30 ms Max. 30 ms Settling time after load step < 1.5 ms < 1.5 ms < 1.5 ms < 1.5 ms < 2 ms Display: Resolution See section „1.9.6.4. Resolution of the displayed values“ Display: Accuracy
(2
Ripple
Remote sensing compensation Max. 5% U
Fall time at no load after switching
DC output o󰀨
Current regulation
Adjustment range 0...81.6 A 0...61.2 A 0...40.8 A 0...30.6 A 0...520.2 A
Accuracy
(1
(at 23±5°C / 73±9°F)) < 0.2% I Line regulation at ±10% ΔU Load regulation at 0...100% ΔU
(2
Ripple
Display: Resolution See section „1.9.6.4. Resolution of the displayed values“ Display: Accuracy
Power regulation
Adjustment range 0…10.2 kW 0…10.2 kW 0…10.2 kW 0…10.2 kW 0…15.3 kW
Accuracy
(1
(at 23±5°C / 73±9°F)) < 1% P Line regulation at ±10% ΔU Load reg. at 10-90% ΔU Display: Resolution See section „1.9.6.4. Resolution of the displayed values“ Display: Accuracy E󰀩ciency
(3
Max
Max
Max
AC
(4
AC
(4
AC
* ΔI
OUT
(4
PS 9360-80 PS 9500-60 PS 9750-40 PS 91000-30 PS 9080-510
360 V 500 V 750 V 1000 V 80 V 80 A 60 A 40 A 30 A 510 A 10 kW 10 kW 10 kW 10 kW 15 kW
Voltage / current: 100 ppm
Max
Max
Max
< 0.1% U
Max
< 0.02% U < 0.05% U
≤ 0.2% U
Max
< 350 mVPP < 70 mV
RMS
Max. 5% U
Max
< 0.02% U
≤ 0.2% U
Max
< 550 mVPP < 65 mV
RMS
Down from 100% to <60 V: less than 10 s
Max
Max
RMS
Max
Max
< 0.2% I
Max
< 0.05% I < 0.15% I < 32 mA
≤ 0.2% I
< 1% P
RMS
Max
Max
< 0.05% P < 0.75% P
≤ 0.85% P
OUT
OUT
Max
< 0.05% I < 0.15% I < 10.4 mA
≤ 0.2% I
Max
Max
< 0.05% P < 0.75% P
≤ 0.8% P
Max
≈ 93% ≈ 95% ≈ 94% ≈ 95% ≈ 93%
Max
Max
Max
Max
Max
Max
Max
Max
< 0.1% U
Max
< 0.02% U < 0.05% U
≤ 0.2% U
Max
< 800 mVPP < 200 mV
RMS
Max. 5% U
< 0.2% I < 0.05% I < 0.15% I < 32 mA
≤ 0.2% I
< 1% P
Max
Max
Max
RMS
Max
Max
< 0.05% P < 0.75% P
≤ 0.85% P
Max
Max
Max
Max
Max
Max
< 0.1% U < 0.02% U < 0.05% U
≤ 0.2% U
Max
Max
Max
Max
< 1600 mVPP < 350 mV
Max. 5% U
< 0.2% I < 0.05% I < 0.15% I < 22 mA
≤ 0.2% I
< 1% P < 0.05% P < 0.75% P
≤ 0.85% P
RMS
Max
Max
Max
Max
RMS
Max
Max
Max
Max
Max
< 0.1% U < 0.02% U < 0.05% U
≤ 0.2% U < 320 mVPP
< 25 mV Max. 5% U
< 0.2% I < 0.05% I < 0.15% I < 240 mA
≤ 0.2% I
< 1% P < 0.05% P < 0.75% P
≤ 0.8% P
Max
Max
Max
Max
RMS
Max
Max
Max
Max
RMS
Max
Max
Max
Max
Max
(1 Related to the nominal values, the accuracy denes the maximum deviation between an adjusted values and the true (actual) value. Example: a 80 V model has min. 0.1% voltage accuracy, that is 80 mV. When adjusting the voltage to 5 V, the actual value is allowed to di󰀨er max. 80 mV, which means it might be between 4.92 V and 5.08 V. (2 RMS value: LF 0...300 kHz, PP value: HF 0...20MHz (3 Typical value at 100% output voltage and 100% power (4 The display error adds to the error of the related actual value on the DC output
EA Elektro-Automatik GmbH Helmholtzstr. 31-37 • 41747 Viersen Germany
Fon: +49 2162 / 3785-0 Fax: +49 2162 / 16230
www.elektroautomatik.de
ea1974@elektroautomatik.de
Page 14
PS 9000 3U Series
Model 400 V
10 kW / 15 kW
Analog interface
Set value inputs U, I, P Actual value outputs U, I Control signals DC on/o󰀨, remote control on/o󰀨 Status signals CV, OVP, OT, OCP, OPP, PF, DC on/o󰀨 Galvanic isolation to the device Max. 725 V DC Sample rate of inputs 500 Hz
Insulation Allowed oat (potential shift) on the DC output:
Negative terminal to PE Max. ±400 V DC ±725 V DC ±725 V DC ±725 V DC ±400 V DC Positive terminal to PE Max. ±600 V DC ±1000 V DC ±1000 V DC ±1000 V DC ±400 V DC
Miscellaneous
Cooling Temperature controlled fans, front inlet, rear exhaust Ambient temperature 0..50°C (32...122°F) Storage temperature -20...70°C (-4...158°F) Humidity < 80%, not condensing
Standards
Overvoltage category 2
Protection class 1 Pollution degree 2
Operational altitude <2000 m (1,242 mi)
Digital interfaces
Featured 1x USB-B, 1x Ethernet (2, 1x GPIB (optional with option 3W) Galvanic isolation from device Max. 725 V DC
Connectors
Rear side Share Bus, DC output, AC input, remote sensing, analog interface, USB-B, Ethernet
Dimensions
Enclosure (WxHxD) 19“ x 3 U x 609 mm (24”) Total (WxHxD) 483 x 133 x 714 mm (19” x 5,2” x 28,1”)
Weight
Article number
(1
(3
PS 9360-80 PS 9500-60 PS 9750-40 PS 91000-30 PS 9080-510
EN 61010-1:2010 EMC TÜV approved acc. IEC 61000-6-2:2005 and IEC 61000-6-3:2006 Class B
24 kg (52.9 lb) 24 kg (52.9 lb) 24 kg (52.9 lb) 24 kg (52.9 lb) 30 kg (66.1 lb) 06230259 06230260 06230261 06230262 06230264
(1 For technical specications of the analog interface see „3.5.4.4 Analog interface specication“ on page 58 (2 Only in the standard version (3 Article number of the standard EU version, devices with options will have a di󰀨erent number
EA Elektro-Automatik GmbH Helmholtzstr. 31-37 • 41747 Viersen Germany
Fon: +49 2162 / 3785-0 Fax: +49 2162 / 16230
www.elektroautomatik.de
ea1974@elektroautomatik.de
Page 15
PS 9000 3U Series
Model 400 V
15 kW
AC Input
Voltage (L-L) 340...460 V AC, 45 - 65 Hz
Input connection 3ph,PE Leak current < 3.5 mA Power factor > 0.99
DC Output
Max. output voltage U Max. output current I Max. output power P Overvoltage protection range 0...220 V 0...396 V 0...550 V 0...825 V 0...1650 V Overcurrent protection range 0...231 A 0...132 A 0...99 A 0...66 A 0...33 A Overpower protection range 0…16.5 kW 0…16.5 kW 0…16.5 kW 0…16.5 kW 0…16.5 kW Temperature coe󰀩cient for set
values Δ/K Output capacitance (approx.) 7560 μF 1200 μF 760 μF 310 μF 84 μF
Voltage regulation
Adjustment range 0...204 V 0...367.2 V 0...510 V 0...765 V 0...1530 V
Accuracy
(1
(at 23±5°C / 73±9°F)) < 0.1% U Line regulation at ±10% ΔU Load regulation at 0...100% load < 0.05% U Rise time 10...90% Max. 30 ms Max. 30 ms Max. 30 ms Max. 30 ms Max. 30 ms Settling time after load step < 2 ms < 2 ms < 2 ms < 2 ms < 2 ms Display: Resolution See section „1.9.6.4. Resolution of the displayed values“ Display: Accuracy
(2
Ripple
Remote sensing compensation Max. 5% U
Fall time at no load after switching
DC output o󰀨
Current regulation
Adjustment range 0...214.2 A 0...122.4 A 0...91.8 A 0...61.2 A 0...30.6 A
Accuracy
(1
(at 23±5°C / 73±9°F)) < 0.2% I Line regulation at ±10% ΔU Load regulation at 0...100% ΔU
(2
Ripple
Display: Resolution See section „1.9.6.4. Resolution of the displayed values“ Display: Accuracy
Power regulation
Adjustment range 0…15.3 kW 0…15.3 kW 0…15.3 kW 0…15.3 kW 0…15.3 kW
Accuracy
(1
(at 23±5°C / 73±9°F)) < 1% P Line regulation at ±10% ΔU Load reg. at 10-90% ΔU Display: Resolution See section „1.9.6.4. Resolution of the displayed values“ Display: Accuracy E󰀩ciency
(3
Max
Max
Max
AC
(4
AC
(4
AC
* ΔI
OUT
(4
PS 9200-210 PS 9360-120 PS 9500-90 PS 9750-60 PS 91500-30
200 V 360 V 500 V 750 V 1500 V 210 A 120 A 90 A 60 A 30 A 15 kW 15 kW 15 kW 15 kW 15 kW
Voltage / current: 100 ppm
Max
Max
Max
< 0.1% U
Max
< 0.02% U < 0.05% U
≤ 0.2% U
Max
< 550 mVPP < 65 mV
RMS
Max. 5% U
Max
< 0.02% U
≤ 0.2% U
Max
< 300 mVPP < 40 mV
RMS
Down from 100% to <60 V: less than 10 s
Max
Max
Max
Max
< 0.2% I
Max
< 0.05% I < 0.15% I < 15.6 mA
≤ 0.2% I
< 1% P
Max
Max
< 0.05% P < 0.75% P
≤ 0.85% P
OUT
OUT
Max
< 0.05% I < 0.15% I
< 66 mA
≤ 0.2% I
RMS
Max
Max
< 0.05% P < 0.75% P
≤ 0.8% P
Max
≈ 95% ≈ 94% ≈ 95% ≈ 94% ≈ 95%
Max
Max
Max
Max
Max
RMS
Max
Max
Max
< 0.1% U
Max
< 0.02% U < 0.05% U
≤ 0.2% U
Max
< 350 mVPP < 70 mV
RMS
Max. 5% U
< 0.2% I < 0.05% I < 0.15% I < 48 mA
≤ 0.2% I
< 1% P
Max
Max
Max
RMS
Max
Max
< 0.05% P < 0.75% P
≤ 0.85% P
Max
Max
Max
Max
Max
Max
< 0.1% U
Max
< 0.02% U < 0.05% U
≤ 0.2% U
Max
< 800 mVPP < 200 mV
RMS
Max. 5% U
< 0.2% I < 0.05% I < 0.15% I < 48 mA
≤ 0.2% I
< 1% P
Max
Max
Max
RMS
Max
Max
< 0.05% P < 0.75% P
≤ 0.85% P
Max
Max
Max
Max
Max
Max
< 0.1% U < 0.02% U < 0.05% U
≤ 0.2% U < 2400 mVPP
< 400 mV Max. 5% U
< 0.2% I < 0.05% I < 0.15% I < 26 mA
≤ 0.2% I
< 1% P < 0.05% P < 0.75% P
≤ 0.85% P
Max
Max
Max
Max
RMS
Max
Max
Max
Max
RMS
Max
Max
Max
Max
Max
(1 Related to the nominal values, the accuracy denes the maximum deviation between an adjusted values and the true (actual) value. Example: a 80 V model has min. 0.1% voltage accuracy, that is 80 mV. When adjusting the voltage to 5 V, the actual value is allowed to di󰀨er max. 80 mV, which means it might be between 4.92 V and 5.08 V. (2 RMS value: LF 0...300 kHz, PP value: HF 0...20MHz (3 Typical value at 100% output voltage and 100% power (4 The display error adds to the error of the related actual value on the DC output
EA Elektro-Automatik GmbH Helmholtzstr. 31-37 • 41747 Viersen Germany
Fon: +49 2162 / 3785-0 Fax: +49 2162 / 16230
www.elektroautomatik.de
ea1974@elektroautomatik.de
Page 16
PS 9000 3U Series
Model 400 V
15 kW
Analog interface
Set value inputs U, I, P Actual value outputs U, I Control signals DC on/o󰀨, remote control on/o󰀨 Status signals CV, OVP, OT, OCP, OPP, PF, DC on/o󰀨 Galvanic isolation to the device Max. 725 V DC Sample rate of inputs 500 Hz
Insulation Allowed oat (potential shift) on the DC output:
Negative terminal to PE Max. ±400 V DC ±400 V DC ±725 V DC ±725 V DC ±725 V DC Positive terminal to PE Max. ±600 V DC ±600 V DC ±1000 V DC ±1000 V DC ±1800 V DC
Miscellaneous
Cooling Temperature controlled fans, front inlet, rear exhaust Ambient temperature 0..50°C (32...122°F) Storage temperature -20...70°C (-4...158°F) Humidity < 80%, not condensing
Standards
Overvoltage category 2
Protection class 1 Pollution degree 2
Operational altitude <2000 m (1,242 mi)
Digital interfaces
Featured 1x USB-B, 1x Ethernet (2, 1x GPIB (optional with option 3W) Galvanic isolation from device Max. 725 V DC
Connectors
Rear side Share Bus, DC output, AC input, remote sensing, analog interface, USB-B, Ethernet
Dimensions
Enclosure (WxHxD) 19“ x 3 U x 609 mm (24”) Total (WxHxD) 483 x 133 x 714 mm (19” x 5,2” x 28,1”)
Weight
Article number
(1
(3
PS 9200-210 PS 9360-120 PS 9500-90 PS 9750-60 PS 91500-30
EN 61010-1:2010 EMC TÜV approved acc. IEC 61000-6-2:2005 and IEC 61000-6-3:2006 Class B
30 kg (66.1 lb) 30 kg (66.1 lb) 30 kg (66.1 lb) 30 kg (66.1 lb) 30 kg (66.1 lb) 06230265 06230266 06230267 06230268 06230269
(1 For technical specications of the analog interface see „3.5.4.4 Analog interface specication“ on page 58 (2 Only in the standard version (3 Article number of the standard EU version, devices with options will have a di󰀨erent number
EA Elektro-Automatik GmbH Helmholtzstr. 31-37 • 41747 Viersen Germany
Fon: +49 2162 / 3785-0 Fax: +49 2162 / 16230
www.elektroautomatik.de
ea1974@elektroautomatik.de
Page 17
PS 9000 3U Series
1.8.4 Specic technical data (208 V AC models)
Model 208 V
5 kW
AC Input
Voltage (L-L) 208 V, ± 10%, 45 - 65 Hz
Input connection 2ph,PE 2ph,PE 2ph,PE 2ph,PE 2ph,PE Leak current < 3.5 mA < 3.5 mA < 3.5 mA < 3.5 mA < 3.5 mA Power factor > 0.99 > 0.99 > 0.99 > 0.99 > 0.99
DC Output
Max. output voltage U Max. output current I Max. output power P Overvoltage protection range 0...88 V 0...220 V 0...396 V 0...550 V 0...825 V Overcurrent protection range 0...187 A 0...77 A 0...44 A 0...33 A 0...22 A Overpower protection range 0…5.5 kW 0…5.5 kW 0…5.5 kW 0…5.5 kW 0…5.5 kW Temperature coe󰀩cient for set
values Δ/K Output capacitance (approx.) 8500 μF 2500 μF 400 μF 250 μF 100 μF
Voltage regulation
Adjustment range 0...81.6 V 0...204 V 0...367.2 V 0...510 V 0...765 V
Accuracy
(1
(at 23±5°C / 73±9°F)) < 0.1% U Line regulation at ±10% ΔU Load regulation at 0...100% load < 0.05% U Rise time 10...90% Max. 30 ms Max. 30 ms Max. 30 ms Max. 30 ms Max. 30 ms Settling time after load step < 1.5 ms < 1.5 ms < 1.5 ms < 1.5 ms < 1.5 ms Display: Resolution See section „1.9.6.4. Resolution of the displayed values“ Display: Accuracy
(2
Ripple
Remote sensing compensation Max. 5% U
Fall time at no load after switching
DC output o󰀨
Current regulation
Adjustment range 0...173.4 A 0...71.4 A 0...40.8 A 0...30.6 A 0...20.4 A
Accuracy
(1
(at 23±5°C / 73±9°F)) < 0.2% I Line regulation at ±10% ΔU Load regulation at 0...100% ΔU
(2
Ripple
Display: Resolution See section „1.9.6.4. Resolution of the displayed values“ Display: Accuracy
Power regulation
Adjustment range 0…5.1 kW 0…5.1 kW 0…5.1 kW 0…5.1 kW 0…5.1 kW
Accuracy
(1
(at 23±5°C / 73±9°F)) < 1% P Line regulation at ±10% ΔU Load reg. at 10-90% ΔU Display: Resolution See section „1.9.6.4. Resolution of the displayed values“ Display: Accuracy E󰀩ciency
(3
Max
Max
Max
AC
(4
AC
(4
AC
* ΔI
OUT
(4
PS 9080-170 PS 9200-70 PS 9360-40 PS 9500-30 PS 9750-20
80 V 200 V 360 V 500 V 750 V 170 A 70 A 40 A 30 A 20 A 5 kW 5 kW 5 kW 5 kW 5 kW
Voltage / current: 100 ppm
Max
Max
Max
< 0.1% U
Max
< 0.02% U < 0.05% U
≤ 0.2% U
Max
< 300 mVPP < 40 mV
RMS
Max. 5% U
Max
< 0.02% U
≤ 0.2% U
Max
< 200 mVPP < 16 mV
RMS
Down from 100% to <60 V: less than 10 s
Max
Max
Max
Max
< 0.2% I
Max
< 0.05% I < 0.15% I < 22 mA
≤ 0.2% I
< 1% P
RMS
Max
Max
< 0.05% P < 0.75% P
≤ 0.8% P
Max
OUT
OUT
Max
< 0.05% I < 0.15% I < 80 mA
≤ 0.2% I
RMS
Max
Max
< 0.05% P < 0.75% P
≤ 0.8% P
Max
≈ 93% ≈ 95% ≈ 95% ≈ 95,5% ≈ 94%
Max
Max
Max
Max
Max
Max
Max
< 0.1% U
Max
< 0.02% U < 0.05% U
≤ 0.2% U
Max
< 550 mVPP < 65 mV
RMS
Max. 5% U
< 0.2% I < 0.05% I < 0.15% I < 5.2 mA
≤ 0.2% I
< 1% P
Max
Max
Max
RMS
Max
Max
< 0.05% P < 0.75% P
≤ 0.8% P
Max
Max
Max
Max
Max
Max
< 0.1% U
Max
< 0.02% U < 0.05% U
≤ 0.2% U
Max
< 350 mVPP < 70 mV
RMS
Max. 5% U
< 0.2% I < 0.05% I < 0.15% I < 16 mA
≤ 0.2% I
< 1% P
Max
Max
Max
RMS
Max
Max
< 0.05% P < 0.75% P
≤ 0.8% P
Max
Max
Max
Max
Max
Max
< 0.1% U < 0.02% U < 0.05% U
≤ 0.2% U < 800 mVPP
< 200 mV Max. 5% U
< 0.2% I < 0.05% I < 0.15% I < 16 mA
≤ 0.2% I
< 1% P < 0.05% P < 0.75% P
≤ 0.8% P
Max
Max
Max
Max
RMS
Max
Max
Max
Max
RMS
Max
Max
Max
Max
Max
(1 Related to the nominal values, the accuracy denes the maximum deviation between an adjusted values and the true (actual) value. Example: a 80 V model has min. 0.1% voltage accuracy, that is 80 mV. When adjusting the voltage to 5 V, the actual value is allowed to di󰀨er max. 80 mV, which means it might be between 4.92 V and 5.08 V. (2 RMS value: LF 0...300 kHz, PP value: HF 0...20MHz (3 Typical value at 100% output voltage and 100% power (4 The display error adds to the error of the related actual value on the DC output
EA Elektro-Automatik GmbH Helmholtzstr. 31-37 • 41747 Viersen Germany
Fon: +49 2162 / 3785-0 Fax: +49 2162 / 16230
www.elektroautomatik.de
ea1974@elektroautomatik.de
Page 18
PS 9000 3U Series
Model 208 V
5 kW
Analog interface
Set value inputs U, I, P Actual value outputs U, I Control signals DC on/o󰀨, remote control on/o󰀨 Status signals CV, OVP, OT, OCP, OPP, PF, DC on/o󰀨 Galvanic isolation to the device Max. 725 V DC Sample rate of inputs 500 Hz
Insulation Allowed oat (potential shift) on the DC output:
Negative terminal to PE Max. ±400 V DC ±400 V DC ±400 V DC ±725 V DC ±725 V DC Positive terminal to PE Max. +400 V DC +600 V DC +600 V DC +1000 V DC +1000 V DC
Miscellaneous
Cooling Temperature controlled fans, front inlet, rear exhaust Ambient temperature 0..50°C (32...122°F) Storage temperature -20...70°C (-4...158°F) Humidity < 80%, not condensing
Standards
Overvoltage category 2
Protection class 1 Pollution degree 2
Operational altitude <2000 m (1,242 mi)
Digital interfaces
Featured 1x USB-B, 1x Ethernet (2, 1x GPIB (optional with option 3W) Galvanic isolation from device Max. 725 V DC
Connectors
Rear side Share Bus, DC output, AC input, remote sensing, analog interface, USB-B, Ethernet
Dimensions
Enclosure (WxHxD) 19“ x 3U x 682 mm (26.8”) Total (WxHxD) 483 x 133 x 787 mm (19” x 5.2” x 31”)
Weight
Article number
(1
(3
PS 9080-170 PS 9200-70 PS 9360-40 PS 9500-30 PS 9750-20
EN 61010-1:2010 EMC TÜV approved acc. IEC 61000-6-2:2005 and IEC 61000-6-3:2006 Class B
17 kg (37.5 lb) 17 kg (37.5 lb) 17 kg (37.5 lb) 17 kg (37.5 lb) 17 kg (37.5 lb) 06238251 06238252 06238253 06238254 06238255
(1 For technical specications of the analog interface see „3.5.4.4 Analog interface specication“ on page 58 (2 Only in the standard version (3 Article number of the standard US version, devices with options will have a di󰀨erent number
EA Elektro-Automatik GmbH Helmholtzstr. 31-37 • 41747 Viersen Germany
Fon: +49 2162 / 3785-0 Fax: +49 2162 / 16230
www.elektroautomatik.de
ea1974@elektroautomatik.de
Page 19
PS 9000 3U Series
Model 208 V
10 kW
AC Input
Voltage (L-L) 208 V, ± 10%, 45 - 65 Hz
Input connection 3ph,PE 3ph,PE 3ph,PE 3ph,PE 3ph,PE Leak current < 3.5 mA < 3.5 mA < 3.5 mA < 3.5 mA < 3.5 mA Power factor > 0.99 > 0.99 > 0.99 > 0.99 > 0.99
DC Output
Max. output voltage U Max. output current I Max. output power P Overvoltage protection range 0...88 V 0...220 V 0...396 V 0...550 V 0...825 V Overcurrent protection range 0...374 A 0...154 A 0...88 A 0...66 A 0...44 A Overpower protection range 0…11 kW 0…11 kW 0…11 kW 0…11 kW 0…11 kW Temperature coe󰀩cient for set
values Δ/K Output capacitance (approx.) 16900 μF 5040 μF 800 μF 500 μF 210 μF
Voltage regulation
Adjustment range 0...81.6 V 0...204 V 0...367.2 V 0...510 V 0...765 V
Accuracy
(1
(at 23±5°C / 73±9°F)) < 0.1% U Line regulation at ±10% ΔU Load regulation at 0...100% load < 0.05% U Rise time 10...90% Max. 30 ms Max. 30 ms Max. 30 ms Max. 30 ms Max. 30 ms Settling time after load step < 1.5 ms < 1.5 ms < 1.5 ms < 1.5 ms < 1.5 ms Display: Resolution See section „1.9.6.4. Resolution of the displayed values“ Display: Accuracy
(2
Ripple
Remote sensing compensation Max. 5% U
Fall time at no load after switching
DC output o󰀨
Current regulation
Adjustment range 0...346.8 A 0...142.8 A 0...81.6 A 0...61.2 A 0...40.8 A
Accuracy
(1
(at 23±5°C / 73±9°F)) < 0.2% I Line regulation at ±10% ΔU Load regulation at 0...100% ΔU
(2
Ripple
Display: Resolution See section „1.9.6.4. Resolution of the displayed values“ Display: Accuracy
Power regulation
Adjustment range 0…10.2 kW 0…10.2 kW 0…10.2 kW 0…10.2 kW 0…10.2 kW
Accuracy
(1
(at 23±5°C / 73±9°F)) < 1% P Line regulation at ±10% ΔU Load reg. at 10-90% ΔU Display: Resolution See section „1.9.6.4. Resolution of the displayed values“ Display: Accuracy E󰀩ciency
(3
Max
Max
Max
AC
(4
AC
(4
AC
* ΔI
OUT
(4
PS 9080-340 PS 9200-140 PS 9360-80 PS 9500-60 PS 9750-40
80 V 200 V 360 V 500 V 750 V 340 A 140 A 80 A 60 A 40 A 10 kW 10 kW 10 kW 10 kW 10 kW
Voltage / current: 100 ppm
Max
Max
Max
< 0.1% U
Max
< 0.02% U < 0.05% U
≤ 0.2% U
Max
< 300 mVPP < 40 mV
RMS
Max. 5% U
Max
< 0.02% U
≤ 0.2% U
Max
< 320 mVPP < 25 mV
RMS
Down from 100% to <60 V: less than 10 s
Max
Max
RMS
Max
Max
< 0.2% I
Max
< 0.05% I < 0.15% I < 44 mA
≤ 0.2% I
< 1% P
RMS
Max
Max
< 0.05% P < 0.75% P
≤ 0.85% P
OUT
OUT
Max
< 0.05% I < 0.15% I < 160 mA
≤ 0.2% I
Max
Max
< 0.05% P < 0.75% P
≤ 0.8% P
Max
≈ 93% ≈ 95% ≈ 93% ≈ 95% ≈ 94%
Max
Max
Max
Max
Max
Max
Max
Max
< 0.1% U
Max
< 0.02% U < 0.05% U
≤ 0.2% U
Max
< 550 mVPP < 65 mV
RMS
Max. 5% U
< 0.2% I < 0.05% I < 0.15% I
Max
Max
Max
< 10.4 mA
≤ 0.2% I
< 1% P
Max
Max
< 0.05% P < 0.75% P
≤ 0.8% P
Max
Max
Max
Max
RMS
Max
Max
< 0.1% U
Max
< 0.02% U < 0.05% U
≤ 0.2% U
Max
< 350 mVPP < 70 mV
RMS
Max. 5% U
< 0.2% I < 0.05% I < 0.15% I < 32 mA
≤ 0.2% I
< 1% P
Max
Max
Max
RMS
Max
Max
< 0.05% P < 0.75% P
≤ 0.85% P
Max
Max
Max
Max
Max
Max
< 0.1% U < 0.02% U < 0.05% U
≤ 0.2% U < 800 mVPP
< 200 mV Max. 5% U
< 0.2% I < 0.05% I < 0.15% I < 32 mA
≤ 0.2% I
< 1% P < 0.05% P < 0.75% P
≤ 0.85% P
Max
Max
Max
Max
RMS
Max
Max
Max
Max
RMS
Max
Max
Max
Max
Max
(1 Related to the nominal values, the accuracy denes the maximum deviation between an adjusted values and the true (actual) value. Example: a 80 V model has min. 0.1% voltage accuracy, that is 80 mV. When adjusting the voltage to 5 V, the actual value is allowed to di󰀨er max. 80 mV, which means it might be between 4.92 V and 5.08 V. (2 RMS value: LF 0...300 kHz, PP value: HF 0...20MHz (3 Typical value at 100% output voltage and 100% power (4 The display error adds to the error of the related actual value on the DC output
EA Elektro-Automatik GmbH Helmholtzstr. 31-37 • 41747 Viersen Germany
Fon: +49 2162 / 3785-0 Fax: +49 2162 / 16230
www.elektroautomatik.de
ea1974@elektroautomatik.de
Page 20
PS 9000 3U Series
Model 208 V
10 kW
Analog interface
Set value inputs U, I, P Actual value outputs U, I Control signals DC on/o󰀨, remote control on/o󰀨 Status signals CV, OVP, OT, OCP, OPP, PF, DC on/o󰀨 Galvanic isolation to the device Max. 725 V DC Sample rate of inputs 500 Hz
Insulation Allowed oat (potential shift) on the DC output:
Negative terminal to PE Max. ±400 V DC ±400 V DC ±400 V DC ±725 V DC ±725 V DC Positive terminal to PE Max. ±400 V DC +600 V DC +600 V DC +1000 V DC +1000 V DC
Miscellaneous
Cooling Temperature controlled fans, front inlet, rear exhaust Ambient temperature 0..50°C (32...122°F) Storage temperature -20...70°C (-4...158°F) Humidity < 80%, not condensing
Standards
Overvoltage category 2
Protection class 1 Pollution degree 2
Operational altitude <2000 m (1,242 mi)
Digital interfaces
Featured 1x USB-B, 1x Ethernet (2, 1x GPIB (optional with option 3W) Galvanic isolation from device Max. 725 V DC
Connectors
Rear side Share Bus, DC output, AC input, remote sensing, analog interface, USB-B, Ethernet
Dimensions
Enclosure (WxHxD) 19“ x 3U x 682 mm (26.8”) Total (WxHxD) 483 x 133 x 787 mm (19” x 5.2” x 31”)
Weight
Article number
(1
(3
PS 9080-340 PS 9200-140 PS 9360-80 PS 9500-60 PS 9750-40
EN 61010-1:2010 EMC TÜV approved acc. IEC 61000-6-2:2005 and IEC 61000-6-3:2006 Class B
24 kg (52.9 lb) 24 kg (52.9 lb) 24 kg (52.9 lb) 24 kg (52.9 lb) 24 kg (52.9 lb) 06238257 06238258 06238259 06238260 06238261
(1 For technical specications of the analog interface see „3.5.4.4 Analog interface specication“ on page 58 (2 Only in the standard version (3 Article number of the standard US version, devices with options will have a di󰀨erent number
EA Elektro-Automatik GmbH Helmholtzstr. 31-37 • 41747 Viersen Germany
Fon: +49 2162 / 3785-0 Fax: +49 2162 / 16230
www.elektroautomatik.de
ea1974@elektroautomatik.de
Page 21
PS 9000 3U Series
Model 208 V
10 kW / 15 kW
AC Input
Voltage (L-L) 208 V, ± 10%, 45 - 65 Hz
Input connection 3ph,PE 3ph,PE 3ph,PE 3ph,PE 3ph,PE Leak current < 3.5 mA < 3.5 mA < 3.5 mA < 3.5 mA < 3.5 mA Power factor > 0.99 > 0.99 > 0.99 > 0.99 > 0.99
DC Output
Max. output voltage U Max. output current I Max. output power P Overvoltage protection range 0...1100 V 0...88 V 0...220 V 0...396 V 0...550 V Overcurrent protection range 0...33 A 0...561 A 0...231 A 0...132 A 0...99 A Overpower protection range 0…11 kW 0…16.5 kW 0…16.5 kW 0…16.5 kW 0…16.5 kW Temperature coe󰀩cient for set
values Δ/K Output capacitance (approx.) 127 μF 25380 μF 7560 μF 1200 μF 760 μF
Voltage regulation
Adjustment range 0...1020 V 0...81.6 V 0...204 V 0...367.2 V 0...510 V
Accuracy
(1
(at 23±5°C / 73±9°F)) < 0.1% U Line regulation at ±10% ΔU Load regulation at 0...100% load < 0.05% U Rise time 10...90% Max. 30 ms Max. 30 ms Max. 30 ms Max. 30 ms Max. 30 ms Settling time after load step < 1.5 ms < 2 ms < 2 ms < 2 ms < 2 ms Display: Resolution See section „1.9.6.4. Resolution of the displayed values“ Display: Accuracy
(2
Ripple
Remote sensing compensation Max. 5% U
Fall time at no load after switching
DC output o󰀨
Current regulation
Adjustment range 0...30.6 A 0...520.2 A 0...214.2 A 0...122.4 A 0...91.8 A
Accuracy
(1
(at 23±5°C / 73±9°F)) < 0.2% I Line regulation at ±10% ΔU Load regulation at 0...100% ΔU
(2
Ripple
Display: Resolution See section „1.9.6.4. Resolution of the displayed values“ Display: Accuracy
Power regulation
Adjustment range 0…10.2 kW 0…15.3 kW 0…15.3 kW 0…15.3 kW 0…15.3 kW
Accuracy
(1
(at 23±5°C / 73±9°F)) < 1% P Line regulation at ±10% ΔU Load reg. at 10-90% ΔU Display: Resolution See section „1.9.6.4. Resolution of the displayed values“ Display: Accuracy E󰀩ciency
(3
Max
Max
Max
AC
(4
AC
(4
AC
* ΔI
OUT
(4
PS 91000-30 PS 9080-510 PS 9200-210 PS 9360-120 PS 9500-90
1000 V 80 V 200 V 360 V 500 V 30 A 510 A 210 A 120 A 90 A 10 kW 15 kW 15 kW 15 kW 15 kW
Voltage / current: 100 ppm
Max
Max
RMS
Max
< 0.1% U
Max
< 0.02% U < 0.05% U
≤ 0.2% U
Max
< 320 mVPP < 25 mV
RMS
Max. 5% U
Max
< 0.02% U
≤ 0.2% U
Max
< 1600 mVPP < 350 mV
Down from 100% to <60 V: less than 10 s
Max
Max
Max
Max
Max
< 0.2% I
Max
< 0.05% I < 0.15% I < 240 mA
≤ 0.2% I
< 1% P
Max
Max
< 0.05% P < 0.75% P
≤ 0.8% P
Max
OUT
OUT
Max
< 0.05% I < 0.15% I < 22 mA
≤ 0.2% I
RMS
Max
Max
< 0.05% P < 0.75% P
≤ 0.85% P ≈ 95% ≈ 93% ≈ 95% ≈ 94% ≈ 95%
Max
Max
Max
Max
Max
RMS
Max
Max
< 0.1% U
Max
< 0.02% U < 0.05% U
≤ 0.2% U
Max
< 300 mVPP < 40 mV
RMS
Max. 5% U
< 0.2% I < 0.05% I < 0.15% I
< 66 mA
≤ 0.2% I
< 1% P
Max
Max
Max
RMS
Max
Max
< 0.05% P < 0.75% P
≤ 0.8% P
Max
Max
Max
Max
Max
Max
< 0.1% U
Max
< 0.02% U < 0.05% U
≤ 0.2% U
Max
< 550 mVPP < 65 mV
RMS
Max. 5% U
< 0.2% I < 0.05% I < 0.15% I
Max
Max
Max
< 15.6 mA
≤ 0.2% I
< 1% P
Max
Max
< 0.05% P < 0.75% P
≤ 0.85% P
Max
Max
Max
RMS
Max
Max
Max
< 0.1% U < 0.02% U < 0.05% U
≤ 0.2% U < 350 mVPP
< 70 mV Max. 5% U
< 0.2% I < 0.05% I < 0.15% I < 48 mA
≤ 0.2% I
< 1% P < 0.05% P < 0.75% P
≤ 0.85% P
Max
Max
Max
Max
RMS
Max
Max
Max
Max
RMS
Max
Max
Max
Max
Max
(1 Related to the nominal values, the accuracy denes the maximum deviation between an adjusted values and the true (actual) value. Example: a 80 V model has min. 0.1% voltage accuracy, that is 80 mV. When adjusting the voltage to 5 V, the actual value is allowed to di󰀨er max. 80 mV, which means it might be between 4.92 V and 5.08 V. (2 RMS value: LF 0...300 kHz, PP value: HF 0...20MHz (3 Typical value at 100% output voltage and 100% power (4 The display error adds to the error of the related actual value on the DC output
EA Elektro-Automatik GmbH Helmholtzstr. 31-37 • 41747 Viersen Germany