PS 9000 3U Series
1. General
1.1 About this document
1.1.1 Retention and use
This document is to be kept in the vicinity of the equipment for future reference and explanation of the operation of
the device. This document is to be delivered and kept with the equipment in case of change of location and/or user.
1.1.2 Copyright
Reprinting, copying, also partially, usage for other purposes as foreseen of this manual are forbidden and breach
may lead to legal process.
1.1.3 Validity
This manual is valid for the following equipment with colour TFT display, including derived variants.
Warning and safety notices as well as general notices in this document are shown in a box with a symbol as follows:
Symbol for a life threatening danger
Symbol for general safety notices (instructions and damage protection bans) or important information for operation
Symbol for general notices
1.2 Warranty
EA Elektro-Automatik guarantees the functional competence of the applied technology and the stated performance
parameters. The warranty period begins with the delivery of free from defects equipment.
Terms of guarantee are included in the general terms and conditions (TOS) of EA Elektro-Automatik.
1.3 Limitation of liability
All statements and instructions in this manual are based on current norms and regulations, up-to-date technology
and our long term knowledge and experience. The manufacturer accepts no liability for losses due to:
• Usage for purposes other than designed
• Use by untrained personnel
• Rebuilding by the customer
• Technical changes
• Use of not authorized spare parts
The actual delivered device(s) may dier from the explanations and diagrams given here due to latest technical
changes or due to customized models with the inclusion of additionally ordered options.
EA Elektro-Automatik GmbH
Helmholtzstr. 31-37 • 41747 Viersen
Germany
Fon: +49 2162 / 3785-0
Fax: +49 2162 / 16230
www.elektroautomatik.de
ea1974@elektroautomatik.de
Page 5
PS 9000 3U Series
1.4 Disposal of equipment
A piece of equipment which is intended for disposal must, according to European laws and regulations (ElektroG,
WEEE) be returned to the manufacturer for scrapping, unless the person operating the piece of equipment or another, delegated person is conducting the disposal. Our equipment falls under these regulations and is accordingly
marked with the following symbol:
1.5 Product key
Decoding of the product description on the label, using an example:
PS 9 080- 510 3U zzz
Field for identication of installed options and/or special models
S01...S0x = Special models
HS = High Speed option installed
3W = Option 3W installed (GPIB port instead of Ethernet port)
US208V = 208 V model for US market
Construction (not stated everywhere):
3U = 19" frame with 3U
Maximum current of the device in Ampere
Maximum voltage of the device in Volt
Series : 9 = Series 9000
Type identication:
PS = Power Supply, usually programmable
1.6 Intended usage
The equipment is intended to be used, if a power supply or battery charger, only as a variable voltage and current
source, or, if an electronic load, only as a variable current sink.
Typical application for a power supply is DC supply to any relevant user, for a battery charger the charging of various battery types and for electronic loads the replacement of an ohmic resistor by an adjustable DC current sink
in order to load relevant voltage and current sources of any type.
• Claims of any sort due to damage caused by non-intended usage will not be accepted.
• All damage caused by non-intended usage is solely the responsibility of the operator.
EA Elektro-Automatik GmbH
Helmholtzstr. 31-37 • 41747 Viersen
Germany
Fon: +49 2162 / 3785-0
Fax: +49 2162 / 16230
www.elektroautomatik.de
ea1974@elektroautomatik.de
Page 6
1.7 Safety
1.7.1 Safety notices
Mortal danger - Hazardous voltage
• Electrical equipment operation means that some parts can be under dangerous voltage.
Therefore all parts under voltage must be covered! This basically applies to all models,
though 40 V models according to SELV can not generate hazardous DC voltage!
• All work on connections must be carried out under zero voltage (output not connected to
load) and may only be performed by qualied and informed persons. Improper actions
can cause fatal injury as well as serious material damage!
• Never touch cables or connectors directly after unplugging from mains supply as the
danger of electric shock remains!
• Never touch the contacts on DC output terminal directly after switching o the DC
output, because there still can dangerous voltage present, sinking more or less slowly
depending on the load! There also can be dangerous potential between negative DC
output to PE or positive DC output to PE due to charged X capacitors.
• Always follow 5 safety rules when working with electric devices:
• Disconnect completely
• Secure against reconnection
• Verify that the system is dead
• Carry out earthing and short-circuiting
• Provide protection from adjacent live parts
PS 9000 3U Series
• The equipment must only be used as intended
• The equipment is only approved for use within the connection limits stated on the product label.
• Do not insert any object, particularly metallic, through the ventilator slots
• Avoid any use of liquids near the equipment. Protect the device from wet, damp and conden-
sation.
• For power supplies and battery chargers: do not connect users, particularly low resistance, to
devices under power; sparking may occur which can cause burns as well as damage to the
equipment and to the user.
• For electronic loads: do not connect power sources to equipment under power, sparking may
occur which can cause burns as well as damage to the equipment and to the source.
• ESD regulations must be applied when plugging interface cards or modules into the relative slot
• Interface cards or modules may only be attached or removed after the device is switched o.
It isn’t necessary to open the device.
• Do not connect external power sources with reversed polarity to DC input or outputs! The
equipment will be damaged.
• For power supply devices: avoid where possible connecting external power sources to the
DC output, and never those that can generate a higher voltage than the nominal voltage of
the device.
• For electronic loads: do not connect a power source to the DC input which can generate a
voltage more than 120% of the nominal input voltage of the load. The equipment isn’t protected
against over voltage and may be irreparably damaged.
• Never insert a network cable which is connected to Ethernet or its components into the master-slave socket on the back side of the device!
• Always congure the various protecting features against overvoltage overpower etc. for sensitive loads to what the currently used application requires
EA Elektro-Automatik GmbH
Helmholtzstr. 31-37 • 41747 Viersen
Germany
Fon: +49 2162 / 3785-0
Fax: +49 2162 / 16230
www.elektroautomatik.de
ea1974@elektroautomatik.de
Page 7
PS 9000 3U Series
1.7.2 Responsibility of the user
The equipment is in industrial operation. Therefore the operators are governed by the legal safety regulations.
Alongside the warning and safety notices in this manual the relevant safety, accident prevention and environmental
regulations must also be applied. In particular the users of the equipment:
• must be informed of the relevant job safety requirements
• must work to the dened responsibilities for operation, maintenance and cleaning of the equipment
• before starting work must have read and understood the operating manual
• must use the designated and recommended safety equipment.
1.7.3 Responsibility of the operator
Operator is any natural or legal person who uses the equipment or delegates the usage to a third party, and is
responsible during its usage for the safety of the user, other personnel or third parties.
The equipment is in industrial operation. Therefore the operators are governed by the legal safety regulations.
Alongside the warning and safety notices in this manual the relevant safety, accident prevention and environmental
regulations must also be applied. In particular the operator has to
• be acquainted with the relevant job safety requirements
• identify other possible dangers arising from the specic usage conditions at the work station via a risk assessment
• introduce the necessary steps in the operating procedures for the local conditions
• regularly control that the operating procedures are current
• update the operating procedures where necessary to reect changes in regulation, standards or operating con-
ditions.
• dene clearly and unambiguously the responsibilities for operation, maintenance and cleaning of the equipment.
• ensure that all employees who use the equipment have read and understood the manual. Furthermore the users
are to be regularly schooled in working with the equipment and the possible dangers.
• provide all personnel who work with the equipment with the designated and recommended safety equipment
Furthermore, the operator is responsible for ensuring that the device is at all times technically t for use.
1.7.4 User requirements
Any activity with equipment of this type may only be performed by persons who are able to work correctly and
reliably and satisfy the requirements of the job.
• Persons whose reaction capability is negatively inuenced by e.g. drugs, alcohol or medication may not operate
the equipment.
• Age or job related regulations valid at the operating site must always be applied.
Danger for unqualied users
Improper operation can cause person or object damage. Only persons who have the necessary
training, knowledge and experience may use the equipment.
Delegated persons are those who have been properly and demonstrably instructed in their tasks and the attendant dangers.
Qualied persons are those who are able through training, knowledge and experience as well as knowledge of
the specic details to carry out all the required tasks, identify dangers and avoid personal and other risks.
EA Elektro-Automatik GmbH
Helmholtzstr. 31-37 • 41747 Viersen
Germany
Fon: +49 2162 / 3785-0
Fax: +49 2162 / 16230
www.elektroautomatik.de
ea1974@elektroautomatik.de
Page 8
PS 9000 3U Series
1.7.5 Alarm signals
The equipment oers various possibilities for signalling alarm conditions, however, not for danger situations. The
signals may be optical (on the display as text) acoustic (piezo buzzer) or electronic (pin/status output of an analog
interface). All alarms will cause the device to switch o the DC output.
The meaning of the signals is as follows:
Signal OT
(OverTemperature)
• Overheating of the device
• DC output will be switched o temporarily
• Non-critical
Signal OVP
(OverVoltage)
• Overvoltage shutdown of the DC output due to high voltage entering the device or generated by the device itself due to a defect
• Critical! The device and/or the load could be damaged
Signal OCP
(OverCurrent)
Signal OPP
(OverPower)
Signal PF
(Power Fail)
• Shutdown of the DC output due to excess of the preset limit
• Non-critical, protects the load from excessive current consumption
• Shutdown of the DC output due to excess of the preset limit
• Non-critical, protects the load from excessive power consumption
• DC output shutdown due to AC undervoltage or defect of the AC input circuit
• Critical on overvoltage! AC mains input circuit could take damage
Display: Color TFT display, 480pt x 128pt
Controls: 2 rotary knobs with button function, 5 pushbuttons
The nominal values for the device determine the maximum adjustable ranges.
EA Elektro-Automatik GmbH
Helmholtzstr. 31-37 • 41747 Viersen
Germany
Fon: +49 2162 / 3785-0
Fax: +49 2162 / 16230
www.elektroautomatik.de
ea1974@elektroautomatik.de
Page 9
PS 9000 3U Series
1.8.3 Specic technical data (400 V AC models)
Model 400 V
3.3 kW / 5 kW
AC Input
Voltage (L-L)340...460 V AC, 45 - 65 Hz
Input connection2ph,PE
Leak current< 3.5 mA
Power factor> 0.99
DC Output
Max. output voltage U
Max. output current I
Max. output power P
Overvoltage protection range0...44 V0...88 V0...220 V0...396 V0...550 V
Overcurrent protection range0...187 A0...187 A0...77 A0...44 A0...33 A
Overpower protection range0…3.63 kW0…5.5 kW0…5.5 kW0…5.5 kW0…5.5 kW
Temperature coecient for set
Adjustment range0...40.8 V0...81.6 V0...204 V0...367.2 V0...510 V
Accuracy
(1
(at 23±5°C / 73±9°F))< 0.1% U
Line regulation at ±10% ΔU
Load regulation at 0...100% load< 0.05% U
Rise time 10...90%Max. 30 msMax. 30 msMax. 30 msMax. 30 msMax. 30 ms
Settling time after load step< 1.5 ms< 1.5 ms< 1.5 ms< 1.5 ms< 1.5 ms
Display: ResolutionSee section „1.9.6.4. Resolution of the displayed values“
Display: Accuracy
(2
Ripple
Remote sensing compensationMax. 5% U
Fall time at no load after switching
DC output o
Current regulation
Adjustment range0...173.4 A0...173.4 A0...71.4 A0...40.8 A0...30.6 A
Accuracy
(1
(at 23±5°C / 73±9°F))< 0.2% I
Line regulation at ±10% ΔU
Load regulation at 0...100% ΔU
(2
Ripple
Display: ResolutionSee section „1.9.6.4. Resolution of the displayed values“
Display: Accuracy
Power regulation
Adjustment range 0…3.36 kW0…5.1 kW0…5.1 kW0…5.1 kW0…5.1 kW
Accuracy
(1
(at 23±5°C / 73±9°F))< 1% P
Line regulation at ±10% ΔU
Load reg. at 10-90% ΔU
Display: ResolutionSee section „1.9.6.4. Resolution of the displayed values“
Display: Accuracy
Eciency
(1 Related to the nominal values, the accuracy denes the maximum deviation between an adjusted values and the true (actual) value.
Example: a 80 V model has min. 0.1% voltage accuracy, that is 80 mV. When adjusting the voltage to 5 V, the actual value is allowed to dier max. 80 mV, which
means it might be between 4.92 V and 5.08 V.
(2 RMS value: LF 0...300 kHz, PP value: HF 0...20MHz
(3 Typical value at 100% output voltage and 100% power
(4 The display error adds to the error of the related actual value on the DC output
EA Elektro-Automatik GmbH
Helmholtzstr. 31-37 • 41747 Viersen
Germany
Fon: +49 2162 / 3785-0
Fax: +49 2162 / 16230
www.elektroautomatik.de
ea1974@elektroautomatik.de
Page 10
PS 9000 3U Series
Model 400 V
3.3 kW / 5 kW
Analog interface
Set value inputsU, I, P
Actual value outputsU, I
Control signalsDC on/o, remote control on/o
Status signalsCV, OVP, OT, OCP, OPP, PF, DC on/o
Galvanic isolation to the deviceMax. 725 V DC
Sample rate of inputs500 Hz
InsulationAllowed oat (potential shift) on the DC output:
Negative terminal to PEMax. ±400 V DC±400 V DC±400 V DC±400 V DC±725 V DC
Positive terminal to PEMax. ±400 V DC±400 V DC±600 V DC±600 V DC±1000 V DC
Miscellaneous
CoolingTemperature controlled fans, front inlet, rear exhaust
Ambient temperature0..50°C (32...122°F)
Storage temperature-20...70°C (-4...158°F)
Humidity< 80%, not condensing
Standards
Overvoltage category2
Protection class1
Pollution degree2
Operational altitude<2000 m (1,242 mi)
Digital interfaces
Featured1x USB-B, 1x Ethernet (2, 1x GPIB (optional with option 3W)
Galvanic isolation from deviceMax. 725 V DC
Connectors
Rear sideShare Bus, DC output, AC input, remote sensing, analog interface, USB-B, Ethernet
Dimensions
Enclosure (WxHxD)19“ x 3 U x 609 mm (24”)
Total (WxHxD)483 x 133 x 714 mm (19” x 5,2” x 28,1”)
EN 61010-1:2010
EMC TÜV approved acc. IEC 61000-6-2:2005 and IEC 61000-6-3:2006 Class B
≈17 kg (37.5 lb) ≈17 kg (37.5 lb) ≈17 kg (37.5 lb) ≈17 kg (37.5 lb) ≈17 kg (37.5 lb)
0623025006230251062302520623025306230254
(1 For technical specications of the analog interface see „3.5.4.4 Analog interface specication“ on page 58
(2 Only in the standard version
(3 Article number of the standard EU version, devices with options will have a dierent number
EA Elektro-Automatik GmbH
Helmholtzstr. 31-37 • 41747 Viersen
Germany
Max. output voltage U
Max. output current I
Max. output power P
Overvoltage protection range0...825 V0...44 V0...44 V0...88 V0...220 V
Overcurrent protection range0...22 A0...374 A0...561 A0...374 A0...154 A
Overpower protection range0…5.5 kW0…7.26 kW0…11 kW0…11 kW0…11 kW
Temperature coecient for set
Adjustment range0...765 V0...40.8 V0...40.8 V0...81.6 V0...204 V
Accuracy
(1
(at 23±5°C / 73±9°F))< 0.1% U
Line regulation at ±10% ΔU
Load regulation at 0...100% load< 0.05% U
Rise time 10...90%Max. 30 msMax. 30 msMax. 30 msMax. 30 msMax. 30 ms
Settling time after load step< 1.5 ms< 1.5 ms< 1.5 ms< 1.5 ms< 1.5 ms
Display: ResolutionSee section „1.9.6.4. Resolution of the displayed values“
Display: Accuracy
(2
Ripple
Remote sensing compensationMax. 5% U
Fall time at no load after switching
DC output o
Current regulation
Adjustment range0...20.4 A0...346.8 A0...520.2 A0...346.8 A0...142.8 A
Accuracy
(1
(at 23±5°C / 73±9°F))< 0.2% I
Line regulation at ±10% ΔU
Load regulation at 0...100% ΔU
(2
Ripple
Display: ResolutionSee section „1.9.6.4. Resolution of the displayed values“
Display: Accuracy
Power regulation
Adjustment range 0…5.1 kW0…6.72 kW0…10.2 kW0…10.2 kW0…10.2 kW
Accuracy
(1
(at 23±5°C / 73±9°F))< 1% P
Line regulation at ±10% ΔU
Load reg. at 10-90% ΔU
Display: ResolutionSee section „1.9.6.4. Resolution of the displayed values“
Display: Accuracy
Eciency
750 V40 V40 V80 V200 V
20 A340 A510 A340 A140 A
5 kW6.6 kW10 kW10 kW10 kW
Voltage / current: 100 ppm
Max
Max
RMS
Max
< 0.1% U
Max
< 0.02% U
< 0.05% U
≤ 0.2% U
Max
< 320 mVPP
< 25 mV
RMS
Max. 5% U
Max
< 0.02% U
≤ 0.2% U
Max
< 800 mVPP
< 200 mV
Down from
100% to <60 V:
--
less than 10 s
Max
Max
Max
Max
< 0.2% I
Max
< 0.05% I
< 0.15% I
< 160 mA
≤ 0.2% I
< 1% P
Max
Max
< 0.05% P
< 0.75% P
≤ 0.7% P
Max
OUT
OUT
Max
< 0.05% I
< 0.15% I
< 16 mA
≤ 0.2% I
RMS
Max
Max
< 0.05% P
< 0.75% P
≤ 0.8% P
Max
≈ 94%≈ 93%≈ 93%≈ 93%≈ 95%
Max
Max
Max
Max
Max
RMS
Max
Max
< 0.1% U
Max
< 0.02% U
< 0.05% U
≤ 0.2% U
Max
< 320 mVPP
< 25 mV
RMS
Max. 5% U
< 0.2% I
< 0.05% I
< 0.15% I
< 120 mA
≤ 0.2% I
< 1% P
Max
Max
Max
RMS
Max
Max
< 0.05% P
< 0.75% P
≤ 0.7% P
Max
< 0.1% U
< 0.02% U
Max
< 0.05% U
Max
≤ 0.2% U
< 320 mVPP
< 25 mV
Max. 5% U
Max
Down from
100% to <60 V:
less than 10 s
< 0.2% I
< 0.05% I
< 0.15% I
< 160 mA
≤ 0.2% I
< 1% P
< 0.05% P
Max
< 0.75% P
Max
≤ 0.8% P
Max
RMS
Max
Max
Max
Max
Max
Max
Max
Max
Max
Max
RMS
Max
Max
< 0.1% U
< 0.02% U
< 0.05% U
≤ 0.2% U
< 300 mVPP
< 40 mV
Max. 5% U
Down from
100% to <60 V:
less than 10 s
< 0.2% I
< 0.05% I
< 0.15% I
< 44 mA
≤ 0.2% I
< 1% P
< 0.05% P
< 0.75% P
≤ 0.85% P
Max
Max
Max
Max
RMS
Max
Max
Max
Max
RMS
Max
Max
Max
Max
Max
(1 Related to the nominal values, the accuracy denes the maximum deviation between an adjusted values and the true (actual) value.
Example: a 80 V model has min. 0.1% voltage accuracy, that is 80 mV. When adjusting the voltage to 5 V, the actual value is allowed to dier max. 80 mV, which
means it might be between 4.92 V and 5.08 V.
(2 RMS value: LF 0...300 kHz, PP value: HF 0...20MHz
(3 Typical value at 100% output voltage and 100% power
(4 The display error adds to the error of the related actual value on the DC output
EA Elektro-Automatik GmbH
Helmholtzstr. 31-37 • 41747 Viersen
Germany
Fon: +49 2162 / 3785-0
Fax: +49 2162 / 16230
www.elektroautomatik.de
ea1974@elektroautomatik.de
Page 12
PS 9000 3U Series
Model 400 V
5 kW / 6.6 kW / 10 kW
Analog interface
Set value inputsU, I, P
Actual value outputsU, I
Control signalsDC on/o, remote control on/o
Status signalsCV, OVP, OT, OCP, OPP, PF, DC on/o
Galvanic isolation to the deviceMax. 725 V DC
Sample rate of inputs500 Hz
InsulationAllowed oat (potential shift) on the DC output:
Negative terminal to PEMax. ±725 V DC±400 V DC±400 V DC±400 V DC±400 V DC
Positive terminal to PEMax. ±1000 V DC±400 V DC±400 V DC±400 V DC±600 V DC
Miscellaneous
CoolingTemperature controlled fans, front inlet, rear exhaust
Ambient temperature0..50°C (32...122°F)
Storage temperature-20...70°C (-4...158°F)
Humidity< 80%, not condensing
Standards
Overvoltage category2
Protection class1
Pollution degree2
Operational altitude<2000 m (1,242 mi)
Digital interfaces
Featured1x USB-B, 1x Ethernet (2, 1x GPIB (optional with option 3W)
Galvanic isolation from deviceMax. 725 V DC
Connectors
Rear sideShare Bus, DC output, AC input, remote sensing, analog interface, USB-B, Ethernet
Dimensions
Enclosure (WxHxD)19“ x 3 U x 609 mm (24”)
Total (WxHxD)483 x 133 x 714 mm (19” x 5,2” x 28,1”)
EN 61010-1:2010
EMC TÜV approved acc. IEC 61000-6-2:2005 and IEC 61000-6-3:2006 Class B
≈17 kg (37.5 lb) ≈24 kg (52.9 lb) ≈30 kg (66.1 lb) ≈24 kg (52.9 lb) ≈24 kg (52.9 lb)
0623025506230256062302630623025706230258
(1 For technical specications of the analog interface see „3.5.4.4 Analog interface specication“ on page 58
(2 Only in the standard version
(3 Article number of the standard EU version, devices with options will have a dierent number
EA Elektro-Automatik GmbH
Helmholtzstr. 31-37 • 41747 Viersen
Germany
Fon: +49 2162 / 3785-0
Fax: +49 2162 / 16230
www.elektroautomatik.de
ea1974@elektroautomatik.de
Page 13
PS 9000 3U Series
Model 400 V
10 kW / 15 kW
AC Input
Voltage (L-L)340...460 V AC, 45 - 65 Hz
Input connection3ph,PE
Leak current< 3.5 mA
Power factor> 0.99
DC Output
Max. output voltage U
Max. output current I
Max. output power P
Overvoltage protection range0...396 V0...550 V0...825 V0...1100 V0...88 V
Overcurrent protection range0...88 A0...66 A0...44 A0...33 A0...561 A
Overpower protection range0…11 kW0…11 kW0…11 kW0…11 kW0…16.5 kW
Temperature coecient for set
Adjustment range0...367.2 V0...510 V0...765 V0...1020 V0...81.6 V
Accuracy
(1
(at 23±5°C / 73±9°F))< 0.1% U
Line regulation at ±10% ΔU
Load regulation at 0...100% load< 0.05% U
Rise time 10...90%Max. 30 msMax. 30 msMax. 30 msMax. 30 msMax. 30 ms
Settling time after load step< 1.5 ms< 1.5 ms< 1.5 ms< 1.5 ms< 2 ms
Display: ResolutionSee section „1.9.6.4. Resolution of the displayed values“
Display: Accuracy
(2
Ripple
Remote sensing compensationMax. 5% U
Fall time at no load after switching
DC output o
Current regulation
Adjustment range0...81.6 A0...61.2 A0...40.8 A0...30.6 A0...520.2 A
Accuracy
(1
(at 23±5°C / 73±9°F))< 0.2% I
Line regulation at ±10% ΔU
Load regulation at 0...100% ΔU
(2
Ripple
Display: ResolutionSee section „1.9.6.4. Resolution of the displayed values“
Display: Accuracy
Power regulation
Adjustment range 0…10.2 kW0…10.2 kW0…10.2 kW0…10.2 kW0…15.3 kW
Accuracy
(1
(at 23±5°C / 73±9°F))< 1% P
Line regulation at ±10% ΔU
Load reg. at 10-90% ΔU
Display: ResolutionSee section „1.9.6.4. Resolution of the displayed values“
Display: Accuracy
Eciency
360 V500 V750 V1000 V80 V
80 A60 A40 A30 A510 A
10 kW10 kW10 kW10 kW15 kW
Voltage / current: 100 ppm
Max
Max
Max
< 0.1% U
Max
< 0.02% U
< 0.05% U
≤ 0.2% U
Max
< 350 mVPP
< 70 mV
RMS
Max. 5% U
Max
< 0.02% U
≤ 0.2% U
Max
< 550 mVPP
< 65 mV
RMS
Down from 100% to <60 V: less than 10 s
Max
Max
RMS
Max
Max
< 0.2% I
Max
< 0.05% I
< 0.15% I
< 32 mA
≤ 0.2% I
< 1% P
RMS
Max
Max
< 0.05% P
< 0.75% P
≤ 0.85% P
OUT
OUT
Max
< 0.05% I
< 0.15% I
< 10.4 mA
≤ 0.2% I
Max
Max
< 0.05% P
< 0.75% P
≤ 0.8% P
Max
≈ 93%≈ 95%≈ 94%≈ 95%≈ 93%
Max
Max
Max
Max
Max
Max
Max
Max
< 0.1% U
Max
< 0.02% U
< 0.05% U
≤ 0.2% U
Max
< 800 mVPP
< 200 mV
RMS
Max. 5% U
< 0.2% I
< 0.05% I
< 0.15% I
< 32 mA
≤ 0.2% I
< 1% P
Max
Max
Max
RMS
Max
Max
< 0.05% P
< 0.75% P
≤ 0.85% P
Max
Max
Max
Max
Max
Max
< 0.1% U
< 0.02% U
< 0.05% U
≤ 0.2% U
Max
Max
Max
Max
< 1600 mVPP
< 350 mV
Max. 5% U
< 0.2% I
< 0.05% I
< 0.15% I
< 22 mA
≤ 0.2% I
< 1% P
< 0.05% P
< 0.75% P
≤ 0.85% P
RMS
Max
Max
Max
Max
RMS
Max
Max
Max
Max
Max
< 0.1% U
< 0.02% U
< 0.05% U
≤ 0.2% U
< 320 mVPP
< 25 mV
Max. 5% U
< 0.2% I
< 0.05% I
< 0.15% I
< 240 mA
≤ 0.2% I
< 1% P
< 0.05% P
< 0.75% P
≤ 0.8% P
Max
Max
Max
Max
RMS
Max
Max
Max
Max
RMS
Max
Max
Max
Max
Max
(1 Related to the nominal values, the accuracy denes the maximum deviation between an adjusted values and the true (actual) value.
Example: a 80 V model has min. 0.1% voltage accuracy, that is 80 mV. When adjusting the voltage to 5 V, the actual value is allowed to dier max. 80 mV, which
means it might be between 4.92 V and 5.08 V.
(2 RMS value: LF 0...300 kHz, PP value: HF 0...20MHz
(3 Typical value at 100% output voltage and 100% power
(4 The display error adds to the error of the related actual value on the DC output
EA Elektro-Automatik GmbH
Helmholtzstr. 31-37 • 41747 Viersen
Germany
Fon: +49 2162 / 3785-0
Fax: +49 2162 / 16230
www.elektroautomatik.de
ea1974@elektroautomatik.de
Page 14
PS 9000 3U Series
Model 400 V
10 kW / 15 kW
Analog interface
Set value inputsU, I, P
Actual value outputsU, I
Control signalsDC on/o, remote control on/o
Status signalsCV, OVP, OT, OCP, OPP, PF, DC on/o
Galvanic isolation to the deviceMax. 725 V DC
Sample rate of inputs500 Hz
InsulationAllowed oat (potential shift) on the DC output:
Negative terminal to PEMax. ±400 V DC±725 V DC±725 V DC±725 V DC±400 V DC
Positive terminal to PEMax. ±600 V DC±1000 V DC±1000 V DC±1000 V DC±400 V DC
Miscellaneous
CoolingTemperature controlled fans, front inlet, rear exhaust
Ambient temperature0..50°C (32...122°F)
Storage temperature-20...70°C (-4...158°F)
Humidity< 80%, not condensing
Standards
Overvoltage category2
Protection class1
Pollution degree2
Operational altitude<2000 m (1,242 mi)
Digital interfaces
Featured1x USB-B, 1x Ethernet (2, 1x GPIB (optional with option 3W)
Galvanic isolation from deviceMax. 725 V DC
Connectors
Rear sideShare Bus, DC output, AC input, remote sensing, analog interface, USB-B, Ethernet
Dimensions
Enclosure (WxHxD)19“ x 3 U x 609 mm (24”)
Total (WxHxD)483 x 133 x 714 mm (19” x 5,2” x 28,1”)
EN 61010-1:2010
EMC TÜV approved acc. IEC 61000-6-2:2005 and IEC 61000-6-3:2006 Class B
≈24 kg (52.9 lb) ≈24 kg (52.9 lb) ≈24 kg (52.9 lb) ≈24 kg (52.9 lb) ≈30 kg (66.1 lb)
0623025906230260062302610623026206230264
(1 For technical specications of the analog interface see „3.5.4.4 Analog interface specication“ on page 58
(2 Only in the standard version
(3 Article number of the standard EU version, devices with options will have a dierent number
EA Elektro-Automatik GmbH
Helmholtzstr. 31-37 • 41747 Viersen
Germany
Fon: +49 2162 / 3785-0
Fax: +49 2162 / 16230
www.elektroautomatik.de
ea1974@elektroautomatik.de
Page 15
PS 9000 3U Series
Model 400 V
15 kW
AC Input
Voltage (L-L)340...460 V AC, 45 - 65 Hz
Input connection3ph,PE
Leak current< 3.5 mA
Power factor> 0.99
DC Output
Max. output voltage U
Max. output current I
Max. output power P
Overvoltage protection range0...220 V0...396 V0...550 V0...825 V0...1650 V
Overcurrent protection range0...231 A0...132 A0...99 A0...66 A0...33 A
Overpower protection range0…16.5 kW0…16.5 kW0…16.5 kW0…16.5 kW0…16.5 kW
Temperature coecient for set
Adjustment range0...204 V0...367.2 V0...510 V0...765 V0...1530 V
Accuracy
(1
(at 23±5°C / 73±9°F))< 0.1% U
Line regulation at ±10% ΔU
Load regulation at 0...100% load< 0.05% U
Rise time 10...90%Max. 30 msMax. 30 msMax. 30 msMax. 30 msMax. 30 ms
Settling time after load step< 2 ms< 2 ms< 2 ms< 2 ms< 2 ms
Display: ResolutionSee section „1.9.6.4. Resolution of the displayed values“
Display: Accuracy
(2
Ripple
Remote sensing compensationMax. 5% U
Fall time at no load after switching
DC output o
Current regulation
Adjustment range0...214.2 A0...122.4 A0...91.8 A0...61.2 A0...30.6 A
Accuracy
(1
(at 23±5°C / 73±9°F))< 0.2% I
Line regulation at ±10% ΔU
Load regulation at 0...100% ΔU
(2
Ripple
Display: ResolutionSee section „1.9.6.4. Resolution of the displayed values“
Display: Accuracy
Power regulation
Adjustment range 0…15.3 kW0…15.3 kW0…15.3 kW0…15.3 kW0…15.3 kW
Accuracy
(1
(at 23±5°C / 73±9°F))< 1% P
Line regulation at ±10% ΔU
Load reg. at 10-90% ΔU
Display: ResolutionSee section „1.9.6.4. Resolution of the displayed values“
Display: Accuracy
Eciency
200 V360 V500 V750 V1500 V
210 A120 A90 A60 A30 A
15 kW15 kW15 kW15 kW15 kW
Voltage / current: 100 ppm
Max
Max
Max
< 0.1% U
Max
< 0.02% U
< 0.05% U
≤ 0.2% U
Max
< 550 mVPP
< 65 mV
RMS
Max. 5% U
Max
< 0.02% U
≤ 0.2% U
Max
< 300 mVPP
< 40 mV
RMS
Down from 100% to <60 V: less than 10 s
Max
Max
Max
Max
< 0.2% I
Max
< 0.05% I
< 0.15% I
< 15.6 mA
≤ 0.2% I
< 1% P
Max
Max
< 0.05% P
< 0.75% P
≤ 0.85% P
OUT
OUT
Max
< 0.05% I
< 0.15% I
< 66 mA
≤ 0.2% I
RMS
Max
Max
< 0.05% P
< 0.75% P
≤ 0.8% P
Max
≈ 95%≈ 94%≈ 95%≈ 94%≈ 95%
Max
Max
Max
Max
Max
RMS
Max
Max
Max
< 0.1% U
Max
< 0.02% U
< 0.05% U
≤ 0.2% U
Max
< 350 mVPP
< 70 mV
RMS
Max. 5% U
< 0.2% I
< 0.05% I
< 0.15% I
< 48 mA
≤ 0.2% I
< 1% P
Max
Max
Max
RMS
Max
Max
< 0.05% P
< 0.75% P
≤ 0.85% P
Max
Max
Max
Max
Max
Max
< 0.1% U
Max
< 0.02% U
< 0.05% U
≤ 0.2% U
Max
< 800 mVPP
< 200 mV
RMS
Max. 5% U
< 0.2% I
< 0.05% I
< 0.15% I
< 48 mA
≤ 0.2% I
< 1% P
Max
Max
Max
RMS
Max
Max
< 0.05% P
< 0.75% P
≤ 0.85% P
Max
Max
Max
Max
Max
Max
< 0.1% U
< 0.02% U
< 0.05% U
≤ 0.2% U
< 2400 mVPP
< 400 mV
Max. 5% U
< 0.2% I
< 0.05% I
< 0.15% I
< 26 mA
≤ 0.2% I
< 1% P
< 0.05% P
< 0.75% P
≤ 0.85% P
Max
Max
Max
Max
RMS
Max
Max
Max
Max
RMS
Max
Max
Max
Max
Max
(1 Related to the nominal values, the accuracy denes the maximum deviation between an adjusted values and the true (actual) value.
Example: a 80 V model has min. 0.1% voltage accuracy, that is 80 mV. When adjusting the voltage to 5 V, the actual value is allowed to dier max. 80 mV, which
means it might be between 4.92 V and 5.08 V.
(2 RMS value: LF 0...300 kHz, PP value: HF 0...20MHz
(3 Typical value at 100% output voltage and 100% power
(4 The display error adds to the error of the related actual value on the DC output
EA Elektro-Automatik GmbH
Helmholtzstr. 31-37 • 41747 Viersen
Germany
Fon: +49 2162 / 3785-0
Fax: +49 2162 / 16230
www.elektroautomatik.de
ea1974@elektroautomatik.de
Page 16
PS 9000 3U Series
Model 400 V
15 kW
Analog interface
Set value inputsU, I, P
Actual value outputsU, I
Control signalsDC on/o, remote control on/o
Status signalsCV, OVP, OT, OCP, OPP, PF, DC on/o
Galvanic isolation to the deviceMax. 725 V DC
Sample rate of inputs500 Hz
InsulationAllowed oat (potential shift) on the DC output:
Negative terminal to PEMax. ±400 V DC±400 V DC±725 V DC±725 V DC±725 V DC
Positive terminal to PEMax. ±600 V DC±600 V DC±1000 V DC±1000 V DC±1800 V DC
Miscellaneous
CoolingTemperature controlled fans, front inlet, rear exhaust
Ambient temperature0..50°C (32...122°F)
Storage temperature-20...70°C (-4...158°F)
Humidity< 80%, not condensing
Standards
Overvoltage category2
Protection class1
Pollution degree2
Operational altitude<2000 m (1,242 mi)
Digital interfaces
Featured1x USB-B, 1x Ethernet (2, 1x GPIB (optional with option 3W)
Galvanic isolation from deviceMax. 725 V DC
Connectors
Rear sideShare Bus, DC output, AC input, remote sensing, analog interface, USB-B, Ethernet
Dimensions
Enclosure (WxHxD)19“ x 3 U x 609 mm (24”)
Total (WxHxD)483 x 133 x 714 mm (19” x 5,2” x 28,1”)
EN 61010-1:2010
EMC TÜV approved acc. IEC 61000-6-2:2005 and IEC 61000-6-3:2006 Class B
≈30 kg (66.1 lb) ≈30 kg (66.1 lb) ≈30 kg (66.1 lb) ≈30 kg (66.1 lb) ≈30 kg (66.1 lb)
0623026506230266062302670623026806230269
(1 For technical specications of the analog interface see „3.5.4.4 Analog interface specication“ on page 58
(2 Only in the standard version
(3 Article number of the standard EU version, devices with options will have a dierent number
EA Elektro-Automatik GmbH
Helmholtzstr. 31-37 • 41747 Viersen
Germany
Max. output voltage U
Max. output current I
Max. output power P
Overvoltage protection range0...88 V0...220 V0...396 V0...550 V0...825 V
Overcurrent protection range0...187 A0...77 A0...44 A0...33 A0...22 A
Overpower protection range0…5.5 kW0…5.5 kW0…5.5 kW0…5.5 kW0…5.5 kW
Temperature coecient for set
Adjustment range0...81.6 V0...204 V0...367.2 V0...510 V0...765 V
Accuracy
(1
(at 23±5°C / 73±9°F))< 0.1% U
Line regulation at ±10% ΔU
Load regulation at 0...100% load< 0.05% U
Rise time 10...90%Max. 30 msMax. 30 msMax. 30 msMax. 30 msMax. 30 ms
Settling time after load step< 1.5 ms< 1.5 ms< 1.5 ms< 1.5 ms< 1.5 ms
Display: ResolutionSee section „1.9.6.4. Resolution of the displayed values“
Display: Accuracy
(2
Ripple
Remote sensing compensationMax. 5% U
Fall time at no load after switching
DC output o
Current regulation
Adjustment range0...173.4 A0...71.4 A0...40.8 A0...30.6 A0...20.4 A
Accuracy
(1
(at 23±5°C / 73±9°F))< 0.2% I
Line regulation at ±10% ΔU
Load regulation at 0...100% ΔU
(2
Ripple
Display: ResolutionSee section „1.9.6.4. Resolution of the displayed values“
Display: Accuracy
Power regulation
Adjustment range 0…5.1 kW0…5.1 kW0…5.1 kW0…5.1 kW0…5.1 kW
Accuracy
(1
(at 23±5°C / 73±9°F))< 1% P
Line regulation at ±10% ΔU
Load reg. at 10-90% ΔU
Display: ResolutionSee section „1.9.6.4. Resolution of the displayed values“
Display: Accuracy
Eciency
80 V200 V360 V500 V750 V
170 A70 A40 A30 A20 A
5 kW5 kW5 kW5 kW5 kW
Voltage / current: 100 ppm
Max
Max
Max
< 0.1% U
Max
< 0.02% U
< 0.05% U
≤ 0.2% U
Max
< 300 mVPP
< 40 mV
RMS
Max. 5% U
Max
< 0.02% U
≤ 0.2% U
Max
< 200 mVPP
< 16 mV
RMS
Down from 100% to <60 V: less than 10 s
Max
Max
Max
Max
< 0.2% I
Max
< 0.05% I
< 0.15% I
< 22 mA
≤ 0.2% I
< 1% P
RMS
Max
Max
< 0.05% P
< 0.75% P
≤ 0.8% P
Max
OUT
OUT
Max
< 0.05% I
< 0.15% I
< 80 mA
≤ 0.2% I
RMS
Max
Max
< 0.05% P
< 0.75% P
≤ 0.8% P
Max
≈ 93%≈ 95%≈ 95%≈ 95,5%≈ 94%
Max
Max
Max
Max
Max
Max
Max
< 0.1% U
Max
< 0.02% U
< 0.05% U
≤ 0.2% U
Max
< 550 mVPP
< 65 mV
RMS
Max. 5% U
< 0.2% I
< 0.05% I
< 0.15% I
< 5.2 mA
≤ 0.2% I
< 1% P
Max
Max
Max
RMS
Max
Max
< 0.05% P
< 0.75% P
≤ 0.8% P
Max
Max
Max
Max
Max
Max
< 0.1% U
Max
< 0.02% U
< 0.05% U
≤ 0.2% U
Max
< 350 mVPP
< 70 mV
RMS
Max. 5% U
< 0.2% I
< 0.05% I
< 0.15% I
< 16 mA
≤ 0.2% I
< 1% P
Max
Max
Max
RMS
Max
Max
< 0.05% P
< 0.75% P
≤ 0.8% P
Max
Max
Max
Max
Max
Max
< 0.1% U
< 0.02% U
< 0.05% U
≤ 0.2% U
< 800 mVPP
< 200 mV
Max. 5% U
< 0.2% I
< 0.05% I
< 0.15% I
< 16 mA
≤ 0.2% I
< 1% P
< 0.05% P
< 0.75% P
≤ 0.8% P
Max
Max
Max
Max
RMS
Max
Max
Max
Max
RMS
Max
Max
Max
Max
Max
(1 Related to the nominal values, the accuracy denes the maximum deviation between an adjusted values and the true (actual) value.
Example: a 80 V model has min. 0.1% voltage accuracy, that is 80 mV. When adjusting the voltage to 5 V, the actual value is allowed to dier max. 80 mV, which
means it might be between 4.92 V and 5.08 V.
(2 RMS value: LF 0...300 kHz, PP value: HF 0...20MHz
(3 Typical value at 100% output voltage and 100% power
(4 The display error adds to the error of the related actual value on the DC output
EA Elektro-Automatik GmbH
Helmholtzstr. 31-37 • 41747 Viersen
Germany
Fon: +49 2162 / 3785-0
Fax: +49 2162 / 16230
www.elektroautomatik.de
ea1974@elektroautomatik.de
Page 18
PS 9000 3U Series
Model 208 V
5 kW
Analog interface
Set value inputsU, I, P
Actual value outputsU, I
Control signalsDC on/o, remote control on/o
Status signalsCV, OVP, OT, OCP, OPP, PF, DC on/o
Galvanic isolation to the deviceMax. 725 V DC
Sample rate of inputs500 Hz
InsulationAllowed oat (potential shift) on the DC output:
Negative terminal to PEMax. ±400 V DC±400 V DC±400 V DC±725 V DC±725 V DC
Positive terminal to PEMax. +400 V DC+600 V DC+600 V DC+1000 V DC+1000 V DC
Miscellaneous
CoolingTemperature controlled fans, front inlet, rear exhaust
Ambient temperature0..50°C (32...122°F)
Storage temperature-20...70°C (-4...158°F)
Humidity< 80%, not condensing
Standards
Overvoltage category2
Protection class1
Pollution degree2
Operational altitude<2000 m (1,242 mi)
Digital interfaces
Featured1x USB-B, 1x Ethernet (2, 1x GPIB (optional with option 3W)
Galvanic isolation from deviceMax. 725 V DC
Connectors
Rear sideShare Bus, DC output, AC input, remote sensing, analog interface, USB-B, Ethernet
Dimensions
Enclosure (WxHxD)19“ x 3U x 682 mm (26.8”)
Total (WxHxD)483 x 133 x 787 mm (19” x 5.2” x 31”)
EN 61010-1:2010
EMC TÜV approved acc. IEC 61000-6-2:2005 and IEC 61000-6-3:2006 Class B
≈17 kg (37.5 lb) ≈17 kg (37.5 lb) ≈17 kg (37.5 lb) ≈17 kg (37.5 lb) ≈17 kg (37.5 lb)
0623825106238252062382530623825406238255
(1 For technical specications of the analog interface see „3.5.4.4 Analog interface specication“ on page 58
(2 Only in the standard version
(3 Article number of the standard US version, devices with options will have a dierent number
EA Elektro-Automatik GmbH
Helmholtzstr. 31-37 • 41747 Viersen
Germany
Max. output voltage U
Max. output current I
Max. output power P
Overvoltage protection range0...88 V0...220 V0...396 V0...550 V0...825 V
Overcurrent protection range0...374 A0...154 A0...88 A0...66 A0...44 A
Overpower protection range0…11 kW0…11 kW0…11 kW0…11 kW0…11 kW
Temperature coecient for set
Adjustment range0...81.6 V0...204 V0...367.2 V0...510 V0...765 V
Accuracy
(1
(at 23±5°C / 73±9°F))< 0.1% U
Line regulation at ±10% ΔU
Load regulation at 0...100% load< 0.05% U
Rise time 10...90%Max. 30 msMax. 30 msMax. 30 msMax. 30 msMax. 30 ms
Settling time after load step< 1.5 ms< 1.5 ms< 1.5 ms< 1.5 ms< 1.5 ms
Display: ResolutionSee section „1.9.6.4. Resolution of the displayed values“
Display: Accuracy
(2
Ripple
Remote sensing compensationMax. 5% U
Fall time at no load after switching
DC output o
Current regulation
Adjustment range0...346.8 A0...142.8 A0...81.6 A0...61.2 A0...40.8 A
Accuracy
(1
(at 23±5°C / 73±9°F))< 0.2% I
Line regulation at ±10% ΔU
Load regulation at 0...100% ΔU
(2
Ripple
Display: ResolutionSee section „1.9.6.4. Resolution of the displayed values“
Display: Accuracy
Power regulation
Adjustment range 0…10.2 kW0…10.2 kW0…10.2 kW0…10.2 kW0…10.2 kW
Accuracy
(1
(at 23±5°C / 73±9°F))< 1% P
Line regulation at ±10% ΔU
Load reg. at 10-90% ΔU
Display: ResolutionSee section „1.9.6.4. Resolution of the displayed values“
Display: Accuracy
Eciency
80 V200 V360 V500 V750 V
340 A140 A80 A60 A40 A
10 kW10 kW10 kW10 kW10 kW
Voltage / current: 100 ppm
Max
Max
Max
< 0.1% U
Max
< 0.02% U
< 0.05% U
≤ 0.2% U
Max
< 300 mVPP
< 40 mV
RMS
Max. 5% U
Max
< 0.02% U
≤ 0.2% U
Max
< 320 mVPP
< 25 mV
RMS
Down from 100% to <60 V: less than 10 s
Max
Max
RMS
Max
Max
< 0.2% I
Max
< 0.05% I
< 0.15% I
< 44 mA
≤ 0.2% I
< 1% P
RMS
Max
Max
< 0.05% P
< 0.75% P
≤ 0.85% P
OUT
OUT
Max
< 0.05% I
< 0.15% I
< 160 mA
≤ 0.2% I
Max
Max
< 0.05% P
< 0.75% P
≤ 0.8% P
Max
≈ 93%≈ 95%≈ 93%≈ 95%≈ 94%
Max
Max
Max
Max
Max
Max
Max
Max
< 0.1% U
Max
< 0.02% U
< 0.05% U
≤ 0.2% U
Max
< 550 mVPP
< 65 mV
RMS
Max. 5% U
< 0.2% I
< 0.05% I
< 0.15% I
Max
Max
Max
< 10.4 mA
≤ 0.2% I
< 1% P
Max
Max
< 0.05% P
< 0.75% P
≤ 0.8% P
Max
Max
Max
Max
RMS
Max
Max
< 0.1% U
Max
< 0.02% U
< 0.05% U
≤ 0.2% U
Max
< 350 mVPP
< 70 mV
RMS
Max. 5% U
< 0.2% I
< 0.05% I
< 0.15% I
< 32 mA
≤ 0.2% I
< 1% P
Max
Max
Max
RMS
Max
Max
< 0.05% P
< 0.75% P
≤ 0.85% P
Max
Max
Max
Max
Max
Max
< 0.1% U
< 0.02% U
< 0.05% U
≤ 0.2% U
< 800 mVPP
< 200 mV
Max. 5% U
< 0.2% I
< 0.05% I
< 0.15% I
< 32 mA
≤ 0.2% I
< 1% P
< 0.05% P
< 0.75% P
≤ 0.85% P
Max
Max
Max
Max
RMS
Max
Max
Max
Max
RMS
Max
Max
Max
Max
Max
(1 Related to the nominal values, the accuracy denes the maximum deviation between an adjusted values and the true (actual) value.
Example: a 80 V model has min. 0.1% voltage accuracy, that is 80 mV. When adjusting the voltage to 5 V, the actual value is allowed to dier max. 80 mV, which
means it might be between 4.92 V and 5.08 V.
(2 RMS value: LF 0...300 kHz, PP value: HF 0...20MHz
(3 Typical value at 100% output voltage and 100% power
(4 The display error adds to the error of the related actual value on the DC output
EA Elektro-Automatik GmbH
Helmholtzstr. 31-37 • 41747 Viersen
Germany
Fon: +49 2162 / 3785-0
Fax: +49 2162 / 16230
www.elektroautomatik.de
ea1974@elektroautomatik.de
Page 20
PS 9000 3U Series
Model 208 V
10 kW
Analog interface
Set value inputsU, I, P
Actual value outputsU, I
Control signalsDC on/o, remote control on/o
Status signalsCV, OVP, OT, OCP, OPP, PF, DC on/o
Galvanic isolation to the deviceMax. 725 V DC
Sample rate of inputs500 Hz
InsulationAllowed oat (potential shift) on the DC output:
Negative terminal to PEMax. ±400 V DC±400 V DC±400 V DC±725 V DC±725 V DC
Positive terminal to PEMax. ±400 V DC+600 V DC+600 V DC+1000 V DC+1000 V DC
Miscellaneous
CoolingTemperature controlled fans, front inlet, rear exhaust
Ambient temperature0..50°C (32...122°F)
Storage temperature-20...70°C (-4...158°F)
Humidity< 80%, not condensing
Standards
Overvoltage category2
Protection class1
Pollution degree2
Operational altitude<2000 m (1,242 mi)
Digital interfaces
Featured1x USB-B, 1x Ethernet (2, 1x GPIB (optional with option 3W)
Galvanic isolation from deviceMax. 725 V DC
Connectors
Rear sideShare Bus, DC output, AC input, remote sensing, analog interface, USB-B, Ethernet
Dimensions
Enclosure (WxHxD)19“ x 3U x 682 mm (26.8”)
Total (WxHxD)483 x 133 x 787 mm (19” x 5.2” x 31”)
EN 61010-1:2010
EMC TÜV approved acc. IEC 61000-6-2:2005 and IEC 61000-6-3:2006 Class B
≈24 kg (52.9 lb) ≈24 kg (52.9 lb) ≈24 kg (52.9 lb) ≈24 kg (52.9 lb) ≈24 kg (52.9 lb)
0623825706238258062382590623826006238261
(1 For technical specications of the analog interface see „3.5.4.4 Analog interface specication“ on page 58
(2 Only in the standard version
(3 Article number of the standard US version, devices with options will have a dierent number
EA Elektro-Automatik GmbH
Helmholtzstr. 31-37 • 41747 Viersen
Germany
Max. output voltage U
Max. output current I
Max. output power P
Overvoltage protection range0...1100 V0...88 V0...220 V0...396 V0...550 V
Overcurrent protection range0...33 A0...561 A0...231 A0...132 A0...99 A
Overpower protection range0…11 kW0…16.5 kW0…16.5 kW0…16.5 kW0…16.5 kW
Temperature coecient for set
Adjustment range0...1020 V0...81.6 V0...204 V0...367.2 V0...510 V
Accuracy
(1
(at 23±5°C / 73±9°F))< 0.1% U
Line regulation at ±10% ΔU
Load regulation at 0...100% load< 0.05% U
Rise time 10...90%Max. 30 msMax. 30 msMax. 30 msMax. 30 msMax. 30 ms
Settling time after load step< 1.5 ms< 2 ms< 2 ms< 2 ms< 2 ms
Display: ResolutionSee section „1.9.6.4. Resolution of the displayed values“
Display: Accuracy
(2
Ripple
Remote sensing compensationMax. 5% U
Fall time at no load after switching
DC output o
Current regulation
Adjustment range0...30.6 A0...520.2 A0...214.2 A0...122.4 A0...91.8 A
Accuracy
(1
(at 23±5°C / 73±9°F))< 0.2% I
Line regulation at ±10% ΔU
Load regulation at 0...100% ΔU
(2
Ripple
Display: ResolutionSee section „1.9.6.4. Resolution of the displayed values“
Display: Accuracy
Power regulation
Adjustment range 0…10.2 kW0…15.3 kW0…15.3 kW0…15.3 kW0…15.3 kW
Accuracy
(1
(at 23±5°C / 73±9°F))< 1% P
Line regulation at ±10% ΔU
Load reg. at 10-90% ΔU
Display: ResolutionSee section „1.9.6.4. Resolution of the displayed values“
Display: Accuracy
Eciency
1000 V80 V200 V360 V500 V
30 A510 A210 A120 A90 A
10 kW15 kW15 kW15 kW15 kW
Voltage / current: 100 ppm
Max
Max
RMS
Max
< 0.1% U
Max
< 0.02% U
< 0.05% U
≤ 0.2% U
Max
< 320 mVPP
< 25 mV
RMS
Max. 5% U
Max
< 0.02% U
≤ 0.2% U
Max
< 1600 mVPP
< 350 mV
Down from 100% to <60 V: less than 10 s
Max
Max
Max
Max
Max
< 0.2% I
Max
< 0.05% I
< 0.15% I
< 240 mA
≤ 0.2% I
< 1% P
Max
Max
< 0.05% P
< 0.75% P
≤ 0.8% P
Max
OUT
OUT
Max
< 0.05% I
< 0.15% I
< 22 mA
≤ 0.2% I
RMS
Max
Max
< 0.05% P
< 0.75% P
≤ 0.85% P
≈ 95%≈ 93%≈ 95%≈ 94%≈ 95%
Max
Max
Max
Max
Max
RMS
Max
Max
< 0.1% U
Max
< 0.02% U
< 0.05% U
≤ 0.2% U
Max
< 300 mVPP
< 40 mV
RMS
Max. 5% U
< 0.2% I
< 0.05% I
< 0.15% I
< 66 mA
≤ 0.2% I
< 1% P
Max
Max
Max
RMS
Max
Max
< 0.05% P
< 0.75% P
≤ 0.8% P
Max
Max
Max
Max
Max
Max
< 0.1% U
Max
< 0.02% U
< 0.05% U
≤ 0.2% U
Max
< 550 mVPP
< 65 mV
RMS
Max. 5% U
< 0.2% I
< 0.05% I
< 0.15% I
Max
Max
Max
< 15.6 mA
≤ 0.2% I
< 1% P
Max
Max
< 0.05% P
< 0.75% P
≤ 0.85% P
Max
Max
Max
RMS
Max
Max
Max
< 0.1% U
< 0.02% U
< 0.05% U
≤ 0.2% U
< 350 mVPP
< 70 mV
Max. 5% U
< 0.2% I
< 0.05% I
< 0.15% I
< 48 mA
≤ 0.2% I
< 1% P
< 0.05% P
< 0.75% P
≤ 0.85% P
Max
Max
Max
Max
RMS
Max
Max
Max
Max
RMS
Max
Max
Max
Max
Max
(1 Related to the nominal values, the accuracy denes the maximum deviation between an adjusted values and the true (actual) value.
Example: a 80 V model has min. 0.1% voltage accuracy, that is 80 mV. When adjusting the voltage to 5 V, the actual value is allowed to dier max. 80 mV, which
means it might be between 4.92 V and 5.08 V.
(2 RMS value: LF 0...300 kHz, PP value: HF 0...20MHz
(3 Typical value at 100% output voltage and 100% power
(4 The display error adds to the error of the related actual value on the DC output
EA Elektro-Automatik GmbH
Helmholtzstr. 31-37 • 41747 Viersen
Germany
Fon: +49 2162 / 3785-0
Fax: +49 2162 / 16230
www.elektroautomatik.de
ea1974@elektroautomatik.de
Page 22
Loading...
+ 50 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.