Eizo RadiForce GS520 Brochure & Specs

Page 1
Page 2
Selecting the Optimum Digital Mammography Monitor
High-Resolution
High resolutions are required to display the correct “information volume” of a digital mammography image.
High-Definition & High-Density
For the detection of mammary gland disorders which appear as “distortion,” the monitor needs this required performance in order to display the subtle structures.
Multi-Grayscale
For the detection of small tumors which appear as delicate “density” differences, the monitor needs the correct display of extremely subtle grayscale shadings.
DICOM Part 14 Compliance with Quality Control Software
For unified image display between multiple monitors, the monitor is required to adjust the characteristics to the grayscale standard DICOM Part 14 through calibration.
Simple Quality Control Procedures
Quality characteristics gradually change over time and therefore in order to maintain a consistent display, the monitor needs to be quality-controllable.
Image Sharpness
For detection of subtle masses and calcifications, the monitor needs to display the outlines of the images with a high degree of sharpness.
Brightness Uniformity
For correct luminance display of delicate “density” differences, the monitor needs uniformity in brightness across the entire screen.
1
2
1
Requirement
2
Requirement
3
Requirement
4
Requirement
5
Requirement
6
Requirement
7
Requirement
Page 3
12-Bit Simultaneous Grayscale Display
Many current monitors use anti-glare panels, which have a waffled surface to diffuse the reflection of surrounding light. However, this also diffuses the light from the monitor’s backlight, which affects the outline of an image, causing a viewing strain on the observer. With a distinctive glare panel, there is no diffused reflection by the waffled surface, and the outline of the image is displayed more clearly. Therefore, the subtle images such as calcification and masses will be easier to observe reducing the strain on the observer.
Sharper Resolution for Clearer Image Outline
MTF (Modulation Transfer Function) is a parameter to define the display sharpness and resolution of a monitor
panel. The glare panel has better MTF characteristics than the anti-glare panel meaning the glare panel can be expected to display the outline of a detailed image more sharply.
8bit 256 Tones Image 12b it4,096 Tones Image
12-bit graphics board and 12-bit viewer software needed for 12-bit display.
1.2
1
0.8
0.6
0.4
0.2
0
MTF
Spatial Frequency (Line pairs/mm)
Anti-Glare Panel
Glare Panel
0.0 0.5 1.0 1.5 2.0 2.5 3.0 3.5
High-Definition and High-Density with Minute Pixels
Enlarged Image
Enlarged Image
Digital
Mammography
2048
2560
2048
1536
3.14 million pixels
5.24 million pixels
0.249mm 0.207mm 0.165mm
Pixel Size Comparison
Brightness Stabilization
At startup or upon wakeup, the EIZO patented drift correction function quickly stabilizes the brightness level. In addition, a sensor measures the backlight brightness and compensates for brightness fluctuations caused by the ambient temperature and the passage of time.
Compatibility with Various Graphics Boards
Supports various input signals for compatibility with a wide range of graphics boards so the user can select the board according to budget and performance requirements. This includes single link and dual link signals, as well as packed­pixel signal for users requiring 12-bit or 10-bit simultaneous display.
Backlight Saver
With ScreenManager Pro for Medical utility software installed, the Backlight Saver function allows for the monitor’s backlight to turn off when the screen saver is activated and the monitor’s backlight to turn on when the computer comes out of the screen saver mode. This function helps to reduce power consumption when the monitor is used for a prolonged period of time.
High-Resolution for Delicate Grayscale Shadings
When selecting a monitor, it’s important to consider the “information volume” of the medical images to be displayed.
Provided by Anjo Kosei Hospital
Please note that the reflection of surrounding light may increase when used in a bright environment. Available with either a glare panel (GS520-CLG, GS520-BLG) or anti-glare panel (GS520-CL, GS520-BL).
This brightness stabilization function is EIZO patented technology (Japan patent numbers 3171808 and 3193315, US patent number 6188380).
0.165mm pixel size is the smallest pixel size among current EIZO medical monitors providing high-definition and high­density
display without shadings
appearing granular.
CT
512
512
1
024
1280
1600
1200
2048
1536
Mammography
1914
2294
Chest CR
1760
2150
2
560
2048
DSA
1000
1000
Displaying only 72%
DUE for Brightness Uniformity
Input Level 80% = 204/255
CCD Camera
Brightness Map
Input Level 80% = 204/255
CCD Camera
Brightness Map
100%
70%
lower
100%
70%
lower
EIZO's confidence in its product quality extends to brightness stability which is also covered during the usage time specified in the warranty.
Brightness Stability Within Usage Time Guaranteed
EIZO and its authorized distributors offer a five-year limited warranty for RadiForce GS520.
Full 5-Year Warranty
Calibration Mode Selection
Selectable with the front panel buttons, the CAL Switch function allows for various calibration modes of different modalities such as digital mammography, CR, and CT images. Furthermore, with ScreenManager Pro for Medical installed, auto mode settings can be made with the Auto CAL Switch function. 
Mammo Mode
Brightness : 500 cd/m
2
Grayscale : DICOM Part 14
Brightness : 450 cd/m
2
Grayscale : DICOM Part 14
Brightness : 350 cd/m
2
Grayscale : γ=2.2
CR Mode CT/MRI Mode
Without DUE With DUE
1
Requirement
3
Requirement
5
Requirement
4
Requirement
2
Requirement
To sustain image consistency, brightness uniformity is required for medical imaging monitors. However, unified
luminance of the backlight is difficult to attain due to the characteristics of LCD monitors. The Digital Uniformity Equalizer (DUE) function provides optimum backlight luminance uniformity.
3
4
Distinctive Glare Panel
Featuring high-resolution, high-definition and multi-grayscale, the RadiForce GS520 is
Featu ring high -reso lutio n, hi gh-de finit ion and m ulti- grays cale, the RadiF orce GS520 is
designed specifically for digital mammography with a choice of distinctive glare panel to
designed spec ificall y for d igital mammo graphy with a choice of dis tinctiv e glare panel to
meet diversified environment usage.
meet di versified envir onment usage.
MONOCHROME LCD MONITOR
Medical Images (Example)
Monitor Resolution
Glare Panel
Incoming Light
Reflections
Anti-Glare Panel
Incoming Light
Reflections
Light from Backlight Light from Backlight
Change modes with one touch.
Backlight
Sensor
Backside of Panel
Hole for Light Transmission
Other Manufacturer's Monitor with Drift Correction
EIZO RadiForce Monitors with Drift Correction
Seconds
Drift Correction Comparison
Backlights switch off
automatically
Cuts power consumption by 11%
RegularUse
18 hours / day
2 hours
16 hours / day
Along with EIZO's new frame rate control technology (patent pending), 4,096 (12-bit) grayscale tones can be displayed simultaneously from an abundant palette of
13,771 tones for high-definition digital mammography.
Main Board
Brightness (cd/m2)
“Information volume” of a digital mammography image should exceed 5 million pixels. When a lower resolution
monitor displays this “information volume,” the monitor stretches the information forcing the mosaic to appear as shadings. With a 2048 x 2560 resolution or 5.24 million pixels, the stretching effect is minimized and the mosaic becomes suitable for rendering subtle masses and calcifications, within the mammography image.
Same Brightness !
Page 4
Quality Control Sensor
Digital Mammography Monitor
Quality Control Tool
Four options are available:
Displaying the Optimum Digital Mammography Image
EIZO offers all the necessary components for correct digital mammography image display.
Bit
Compatible Operating System
Frame Buffer Memory
Bus Interface
Maximum Disply Resolutions
Output Terminals
Dot Clock
Scanning Frequency (H, V)
Maximum Power Consumption
12-bit /10-bit / 8-bit Windows XP / 2000 Professional 128 MB PCI 32-bit (33 MHz / 66 MHz) 2048 x 2560 DVI-I x 2 Up to 148 MHz
103.9 kHz 50 Hz
16.8 watts
High-Performance Graphics Board
Standard Configuration
10-bit / 8-bit Windows XP / 2000 Professional 256 MB PCI 32-bit (33 MHz) / 64-bit (66 MHz) 2048 x 2560 DVI-I x 2 Up to 148 MHz
103.9 kHz 50 Hz 40 watts
8-bit Windows XP 1 GB PCI Express x 16 2048 x 2560 DVI-D x 2
214.25 MHz 131 kHz 50 Hz 75 watts
8-bit Windows XP 256 MB PCI Express x 16 / x 1 2048 x 2560 DVI-D x 2 162 MHz 75 kHz 60 Hz 22 watts
Accessories
Monitor Cleaning Kit
Keep your screen free from dust and fingerprints with this screen cleaner kit. Includes pump spray and cloth. (Bundled with GS520-CLG and GS520-BLG.)
Panel Protector
Protect against scratches and dust with high light transmission panel.
Dual Height Adjustable Stand
Mount two panels in either portrait or landscape orientation.
5
6
PAS 1054
“Requirements and Testing of Digital Mammography Equipment” established by the Normenausschusses Radiologie (DIN NAR) in cooperation with the Deutschen Röntgengesellschaft (DRG) and others. In line with the QS-RL (Qualitätssicherungs-Richtlinie) quality assurance ordinance, it is mandatory in Germany to refer to this standard when conducting monitor quality assurance.
EIZO's easy-to-use RadiCS quality control software performs precise calibration conforming to DICOM Part 14 and enables quality control complying with ACR Practice Guideline and other QC standard for digital mammography monitors.
Once attached to the top of the monitor the Clip-On Swing Sensor G1 can remain in place
and is ready for use. This saves the time and trouble of attaching and removing it whenperforming quality control tasks.
Quality Control of Digital Mammography Monitor
As with the use of film mammography, image quality testing of the monitor at installation and regularly during use should be carried out. This ensures that the monitor maintains a consistent display of quality digital mammography.
●An option of
An opt ion of
clear
cle ar
base and
bas e and
blue
blue
base backlights are available.
base backlights are available.
●Paired monitors which provides the same
Paired monitors which provides the same
color and grayscale for two monitors
co lor a nd gr ays cal e for two m oni tors
bundled as a pair in one box, also available.
bundled as a pair in one box, also available.
ACR“Practice Guideline for Determinants of Image
 Quality in Digital Mammography”
A guideline established by the American College of Radiology (ACR) in cooperation with the American Association of Physicists in Medicine (AAPM), Radiological Society of North America (RSNA), and Society for Imaging Informatics in Medicine (SIIM).
EUREF “European Guidelines for Quality Assurance
in Breast Cancer Screening and Diagnosis”
A standard issued by the European Commission in cooperation with the European Reference Organization for Quality Assured Breast Screening and Diagnostic Services (EUREF), and European Breast Cancer Network (EBCN), European Society of Mastology (EUSOMA).
Page 5
54 cm 21.3" Monochrome LCD Monitor
Dimensions
[
Unit: mm
]
EIZO NANAO CORPORATION supports Pink Ribbon activities.
Model Variations
Cabinet Color
Active Display Size (H x V)
Viewable Image Size
Native Resolution
Pixel Pitch
Grayscale Tones
Viewing Angles (H, V)
Brightness
Recommended Brightness for Calibration
Contrast Ratio
Response Time
Scanning Frequency (H, V)
Dot Clock
Input Signals
Input Terminal
USB Ports
USB Standard
Power Requirements
Power Consumption
Power Save Mode
Sensor
Power Management
OSD Languages
Net Weight
Hole Spacing
Certifications and Standards
Supplied Accessories
GS520-CLG: Clear Base with Glare Panel GS520-BLG: Blue Base with Glare Panel GS520-CL: Clear Base GS520-BL: Blue Base
Black
TFT Monochrome LCD Panel (IPS)
337.9 x 422.4 mm
Diagonal: 540 mm
2048 x 2560
0.165 x 0.165 mm
4,096 from a palette of 13,771 tones
170°, 170°
700 cd/m
2
(typical)
500 cd/m
2
800:1 (typical)
100 ms (typical)
31- 135 kHz, 19-51 Hz Frame synchronous mode: 49 - 51 Hz
300 MHz
DVI Standard 1.0
DVI-D 24 pin
1 upstream, 2 downstream
Standard Rev. 2.0
AC 100 -120 V, 200 -240 V: 50 / 60 Hz
90 watts
Less than 1.2 watts
Backlight Sensor
DVI DMPM
English, French, German, Italian, Japanese, Simplified Chinese, Spanish, Swedish, Traditional Chinese
With Stand: 9.6 kg Without Stand: 6.6 kg
VESA standard (100 x 100 mm)
CE (Medical Device Directive), TÜV/GM (EN60601-1), cTÜVus (UL60601-1, CSA C22.2 No. 601-1), CB (IEC60601-1), VCCI-B, FCC-B, Canadian ICES-003-B, c-Tick, FDA 510(k) for Mammography and General Radiography, RoHS
AC power cord, user's manual, signal cable (DVI-D ~ DVI-D), USB cable, EIZO LCD Utility Disk (RadiCS LE, ScreenManager Pro for Medical software), ScreenCleaner (GS520-CLG, GS520-BLG only), warranty card
90°
40°
512.5
340
388
424
480
88
61.5
Height Adjustment 82