Edwards SIGA-PHS Quick Start Manual

LIFE SAFETY & INCIDENT MANAGEMENT
Intelligent 3D Multisensor Detector
SIGA-PHS
EDWARDS Catalog u Intelligent Initiating
Overview
The Signature Series Model SIGA-PHS Intelligent 3D Multisensor Detector gathers analog information from each of its two fire sens­ing elements (photoelectric and heat) and converts it into digital signals. The detector’s on-board microprocessor measures and analyzes these signals separately with respect to a third element
- Time. We call this technology 3D. It compares the information to historical readings, time patterns, and several known fire charac­teristics to make an alarm decision. Digital filters remove signal patterns that are not typical of fires. Unwanted alarms are virtually eliminated.
Standard Features
• Integrates photoelectric smoke and 135 º F (57 º C) fixed-tem­perature heat sensing technologies
• Non-volatile memory
• Automatic device mapping
• Electronic addressing
• Environmental compensation
• Integral microprocessor - intelligent detector
• Wide 0.67% to 3.77%/ft. sensitivity range
• Twenty pre-alarm sensitivity values, set in 5% increments
• Identification of dirty or defective detectors
• Automatic day/night sensitivity adjustment
• Twin RED/GREEN status LEDs
• Standard, relay, fault isolator, and audible mounting bases
• Designed and manufactured to ISO 9001 standards
Note: Some features described here may not be supported by all control systems. Check your control panel’s Installation and Operation Guide for details.
MEA
7272-1657:
0126
S853
Signature Series Overview
The microprocessor in each detector provides four additional benefits - Self-diagnostics and History Log, Automatic Device Mapping, and Fast, Stable Communication.
Self-diagnostics and History Log - Each Signature Series
detector constantly runs self-checks to provide important mainte­nance information. The results of the self-check are automatically updated and permanently stored in the detector’s non-volatile memory. This information is accessible for review any time at the control panel, PC, or by using the SIGA-PRO Signature Program/ Service Tool.
Automatic Device Mapping - The loop controller learns where
each device’s serial number address is installed relative to other devices on the circuit. This mapping feature provides supervision of each device’s installed location to prevent a detector from being reinstalled (after cleaning etc.) in a different location from where it was originally. The history log for the detector remains relevant and intact regardless of its new location.
The Signature Series Data Entry Program also uses the mapping feature. With interactive menus and graphic support, the wired cir­cuits between each device can be examined. Layout or “as-built” drawing information showing wire branches (T-taps), device types and their address are stored on disk for printing hard copy. This takes the mystery out of the installation. The preparation of as-built drawings is fast and efficient.
Page 1 of 4 DATA SHEET E85001-0247
Not to be used for installation purposes. Issue 1
Fast Stable Communication - On-board intelligence means less
information needs to be sent between the detector and the loop controller. Other than regular supervisory polling response, the de­tector only needs to communicate with the loop controller when it has something new to report. This provides very fast control panel response time and allows a lower baud rate (speed) to be used for communication on the circuit.
Environmental Compensation - Detection sensitivity for the
SIGA-PHS is virtually independent of its installed environment and its physical condition. Environmental compensation means each sensing element adapts to long-term changes caused by dirt, humidity, temperature, aging etc. It even compensates for small amounts of normal ambient smoke.
Sensitivity Range - The SIGA-PHS has a sensitivity range or
window of 0.67 - 3.70% obscuration per foot. The installer selects the detector’s ALARM sensitivity level. Five settings ranging from the least sensitive at 3.5% obscuration/ft to the most sensitive at
1.0% obscuration/ft are available.
Automatic Day/Night Sensitivity Selection - Signature Series
detectors may be programmed for different sensitivities during day and night periods. This allows the detector to be more sensitive during unoccupied periods when lower ambient background con­ditions are expected.
Stability - The 3D detector’s sensitivity remains stable in wind
velocities up to 500 ft/min (2.53 m/sec). Ambient temperature has very little affect on the detector. The detector may be installed in rooms with ambient temperatures up to 100oF (38oC).
Electronic Addressing - The loop controller electronically ad-
dresses each detector, saving valuable time during system commis­sioning. Setting complicated switches or dials is not required. Each detector has its own unique serial number stored in its on-board memory. The loop controller identifies each device on the circuit and assigns a “soft” address to that device’s serial number. If desired, detectors can be addressed using the SIGA-PRO Signature Pro­gram/Service Tool.
Testing & Maintenance
Each detector automatically identifies when it is dirty or defective and causes a “dirty detector” message. The detector’s sensitiv­ity measurement can also be transmitted to the loop controller. A sensitivity report can be printed to satisfy NFPA sensitivity meas­urements which must be conducted at the end of the first year and every two years thereafter.
The user-friendly maintenance program shows the current state of each detector and other pertinent messages. Single detectors may be turned off temporarily, from the control panel. Availability of maintenance features is dependent on the fire alarm system used. Scheduled maintenance (Regular or Selected) for proper detec­tor operation should be planned to meet the requirements of the Authority Having Jurisdiction (AHJ). Refer to current NFPA 72 and ULC CAN/ULC 536 standards.
Installation
Signature Series detectors mount to North American 1-gang boxes, 3-1/2 inch or 4 inch octagon boxes, and to 4 inch square electrical boxes 1-1/2 inches (38 mm) deep. They mount to Euro­pean BESA and 1-gang boxes with 60.3 mm fixing centers.
Compatibility
The SIGA-PHS detectors are compatible only with the Signature Loop Controller.
Application
Although photoelectric detectors have a wide range of fire sens­ing capabilities they are best suited for detecting slow, smoldering fires. The table below shows six standard test fires used to rate the sensitivity of smoke and heat detectors. The table indicates
SIGA-HRS and
Test Fire SIGA-PS Photo
Open Wood unsuitable optimum very suitable optimum Wood Pyrolysis optimum unsuitable optimum optimum Smouldering Cotton optimum unsuitable optimum optimum Poly Urethane Foam very suitable suitable very suitable optimum n-Heptane very suitable very suitable optimum optimum Liquid Fire without
Smoke
unsuitable optimum very suitable very suitable
SIGA-HFS Rate-of­Rise/ Fixed Temp.
that no single sensing element is suited for all test fires.
EDWARDS recommends that this detector be installed according to latest recognized edition of national and local fire alarm codes.
SIGA-PHS Photo Heat 3D
SIGA-IPHS Ion/Photo/Heat 4D
Page 2 of 4 DATA SHEET E85001-0247
Not to be used for installation purposes. Issue 1
Loading...
+ 2 hidden pages