Edwards and the Edwards logo are trademarks of Edwards Limited.
Page 7
B722-20-880 Issue D
CAUTION
WARNING
1Introduction
1.1Scope and definitions
This manual provides installation, operation and maintenance instructions for the Edwards EXT70H 24 V,
EXT70Hi 24 V, EXT255H 24 V and EXT255Hi 24 V Compound Molecular Pumps. These pumps must be used as specified
in this manual. It is not possible to purchase an EXT70Hi 24 V pump; an EXT70Hi 24 V pump is an EXT70H/ISO63 24 V
pump fitted with an interstage-port accessory: see Section 7.
The EXT Compound Molecular Pumps are designed for use with a Edwards Turbomolecular Pump Drive Module
(abbreviated to EXDC Drive Module in the remainder of this manual). Read this manual and the instruction manual
supp l i ed w i t h you r E XDC D r ive M o dule b e for e y ou at t e mpt t o install or operate the equipment. The instruction manual
supplied with your EXDC Drive Module contains details of how to set up the Module and the logic interface.
Important safety information in this manual is highlighted as WARNING and CAUTION instructions; you must obey
these instructions. The use of WARNINGS and CAUTIONS is defined below.
Warnings are given where failure to observe the instruction could result in injury or death to
people.
Introduction
Cautions are given where failure to observe the instruction could result in damage to the equipment, associated
equipment and process.
In accordance with the recommendations of EN61010, the following warning symbols may appear on the pump or its
accessories:
Warning - refer to accompanying documentation.
Warning - risk of electric shock.
Warning - hot surfaces.
Protective conducter terminal.
The units used throughout this manual conform to the SI international system of units of measurement. Also
throughout this manual, wherever flow rates are specified, the abbreviation ‘sccm’ is used to mean
‘standard cm
(1.013 x 10
min-1: this is a flow of 1 cm3 min-1 at an ambient temperature of 0 °C and a pressure of 1013 mbar
5
Pa).
Page 8
B722-20-880 Issue D
Introduction
1.2General description
The EXT70H 24 V, EXT70Hi 24 V, EXT255H 24 V and EXT255Hi 24 V pumps are compound vacuum pumps that combine
a turbomolecular pump and a drag pump on a single shaft. The turbomolecular pump is made from a multi-stage axialflow turbine which is optimised for operation in molecular flow conditions. The drag pump is designed for operation
at higher pressures than conventional turbomolecular pumps.
The angled blades of the turbomolecular pump are machined from a single piece of material. For the blades to be
effective, their speed must be close to the thermal velocity of the gas. The rotor of the EXT70H/Hi 24 V therefore
rotates at up to 90000 r min
Gas from the inlet is compressed through each of the bladed stages of the turbomolecular pump and transferred to
the drag pump. In the drag pump stages the gas is further compressed.
The rotat ing asse mbly is s upporte d at the inlet end by a frictionless magnetic bearing and by a precision ceramic ball
bearing at the outlet end. The ball bearing is lubricated from an oil reservoir and wick mechanism.
The EXT pump is supplied with an inlet-screen fitted in the bore of the inlet-flange. The EXT255Hi 24 V pump is also
supplied with an inlet-strainer which fits in the interstage-port. The inlet-screen and inlet-strainer protect the pump
against damage which would be caused by debris entering the pump. The inlet-screen also serves to protect the user
against injury from the sharp blades in the pump.
The EXT pump has a vent-port for venting the pump and vacuum system to atmospheric pressure. The pump is
supplied with a manual vent-valve fitted to the vent-port. As described in Section 3.4, the manual vent-valve can be
replaced with a TAV5 or TAV6 solenoid-operated vent-valve (available as an accessory: see Section 7).
-1
, and the rotor of the EXT255H/Hi 24 V rotates at up to 60000 r min-1.
The EXT255H 24 V and EXT255Hi 24 V pumps have a purge-port in the motor and bearing housing chamber. An inert
purge gas can be introduced through the purge port in order to protect these parts from corrosion or the bearing
lubricant from oxidisation. An optional vent-port adaptor and purge restrictor can be fitted to the purge-port to
control the flow rate of the purge gas and to filter the gas supply (see Section 7).
The EXT255Hi 24 V pump has an interstage-port between the turbomolecular and drag stages. If required, an
interstage-port accessory can be fitted to an EXT70H/ISO63 24 V pump to upgrade it to an EXT70Hi 24 V pump (see
Section 7). The interstage-port on EXT70Hi 24 V and EXT255Hi 24 V pumps can be used to pump a separate chamber
of the vacuum system or to back another turbomolecular pump.
If the EXT70Hi 24 V pump operates with an outlet pressure of less than 5 mbar (500 Pa) and an interstage inlet flow
of less than 100 sccm (1.69 mbar l s
independent of the interstage operating conditions.
If the EXT255Hi 24 V pump operates with an outlet pressure of less than 5 mbar (500 Pa) and an interstage inlet flow
of less than 70 sccm (1.17 mbar l s
independent of the interstage operating conditions.
An EXDC 24 V Drive Module is used to control the EXT 24 V pump. The EXDC 24 V Drive Module is fitted directly to the
19-way connector on the pump.
The pump may be cooled using air-cooled or water-cooled optional accessories. Refer to Section 3 for guidance on
applications and cooling requirements. Pumps with a Conflat flange are supplied with a water-cooler.
All EXT pumps have thermal sensors to monitor the motor and pump-body temperature.
-1
, 169 Pa l s-1) the vacuum performance at the turbomolecular pump-inlet is
-1
, 117 Pa l s-1), the vacuum performance at the turbomolecular pump-inlet is
1.3Vent options and vent control
To maintain the cleanliness of the vacuum system, it is recommended that, whenever the pump is switched off, the
pump (or vacuum system) is vented when the speed of the EXT pump is between full rotational speed and 50% of full
rotational speed. At and above 50% of full rotational speed, the rotor spins fast enough to suppress any backstreaming
of hydrocarbon oil from your backing pump.
However, if the pump is vented when it is at full rotational speed and the rate of pressure rise is too high, the pump
life may be reduced. It is therefore recommended that either the rate of pressure rise is limited (see Figure 1), or
that the vent-valve is only opened after the EXT pump speed has fallen to 50% of full rotational speed.
Edwards and the Edwards logo are trademarks of Edwards Limited.
Page 9
B722-20-880 Issue D
The rate of pressure rise cannot be controlled by the manual vent-valve, so if the manual vent-valve is used it must
only be opened after the EXT pump speed has fallen to 50% of full rotational speed.
If a TAV vent-valve is used, but the rate of pressure rise cannot be limited, the vent-valve must only be opened after
the speed of the EXT pump has fallen to 50% of full rotational speed.
Figure 1 - Maximum allowed rate of pressure rise during venting: pressure against time (with the pump initially
Edwards and the Edwards logo are trademarks of Edwards Limited.
Page 11
2Technical data
CAUTION
CAUTION
WARNING
WARNING
WARNING
2.1General
PerformanceSee Ta b le 1 and 2
Cooling, noise and electrical dataSee Ta bl e 1 and 2
DimensionsSee Ta bl e 2 and 3
Maximum inlet-flange temperature100 °C
Maximum magnetic field (EXT255H/Hi)3.5 mT horizontal field; 7 mT vertical field
Maximum magnetic field (EXT70H/Hi)5 mT
Installation categoryEN61010 part 1, Category 1
Pollution degreeCAN/CSA, UL, EN61010 part 1
Altitude2000 m
Humidity10 to 90% non condensing
Electrical supply24 V d.c.
Equipment typeFixed equipment, for indoor use only
2.2Pumping media
B722-20-880 Issue D
Technical data
Vent dangerous gases and gas mixtures safely. Do not expose people to these gases.
Do not use EXT pumps to pump explosive gas mixtures as the pumps are not suitable for this
purpose.
On EXT70Hi 24 V and EXT255Hi 24 V pumps, gas pumped through the interstage-port will mix with
gas pumped through the pump-inlet. Ensure that the gases will not react or combine to form
dangerous gases and substances.
Do not use an EXT to pump gases containing more than 2 0% oxygen unless the pump is gas purged. If you do, the
lubricant will polymerise and the pump will fail prematurely.
Do not use an EXT to pump mercury vapour and do not allow mercury (for example, from a McLeod gauge) to
come into contact with the pump. If you do, the pump rotor may corrode and fail.
Note:Note that concentrations of gases may be modified by the compression of the pump.
The pumps are designed to pump the following residual gases normally used in high-vacuum systems:
z Airz Carbon monoxidez Neonz Ethanez Methane
z Nitrogenz Kryptonz Argonz Propane
z Carbon dioxidez Heliumz Hydrogenz Butane
The pumps may be used to pump oxygen and water vapour, subject to the following conditions:
z OxygenThe oxygen concentration must be less than 20% by volume
z Water vapourYou must ensure that vapour does not condense inside the pump; refer to Section 3.7.3.
If you wish to pump a gas not in the list above, contact your supplier for advice. If you do not contact your supplier,
you may invalidate the warranty on the pump. The pumps are not suitable for pumping aggressive or corrosive gases.
2.2.2Pumps with gas purge (EXT255H 24 V and EXT255Hi 24 V only)
When purged with an inert gas, the pumps can be used to pump oxygen in concentrations above 20% by volume.
2.3Vent gas specification and vent control data
Although the pump may be vented to atmospheric air, high relative humidity of the air may greatly increase the
subsequent pumping time. To reduce pump-down times, vent the pump with dry, clean gases.
Vent gasDry air, nitrogen, argon or other inert gases
Maximum dew point at atmospheric pressure- 22 °C
Maximum size of particulates1 µm
Maximum concentration of oil0.1 parts per million
Time for rotational speed to reach 50%15 s
Maximum allowed rate of pressure riseSee Figure 1
Edwards and the Edwards logo are trademarks of Edwards Limited.
Page 15
B722-20-880 Issue D
2.4Purge gas specification (EXT255H 24 V and EXT255Hi 24 V only)
Purge gasDry nitrogen, argon or other inert gases
Maximum dew point at atmospheric pressure-22 °C
Maximum size of particulates1
Maximum concentration of oil0.1 parts per million
Allowable purge gas flow (when required)20 to 100 sccm (0.33 to 1.67 mbar l s
Recommended purge gas flow25 sccm (0.42 mbar l s
Maximum allowable purge gas supply pressure2 bar (gauge); 29 psi, 3 x 105 Pa
µm
167 Pa l s-1)
-1
, 42 Pa l s-1)
-1
, 33 to
2.5Cooling-water
The following cooling-water specification corresponds to a typical high-quality drinking water specification. Check
with your water supply authority if you are in doubt about the quality of your supply.
QualityMechanically clean and optically clear with no
deposits or turbidity.
pH value6.0 to 8.0
Maximum calcium carbonate concentration75 parts per million
Maximum chloride concentration100 parts per million
Minimum oxygen concentration4 parts per million
Technical data
Minimum water-cooling flow rate (at 15 °C)15 l h
Water temperatureSee Ta bl e 1 and 2
Maximum water pressure5 bar (gauge), 73.5 psig, 6 x 10
Materials exposed to cooling waterNickel plated brass
-1
5
Pa
2.6Materials exposed to gases pumped
The following materials and component types are exposed to the gases pumped: aluminium alloys, stainless steels,
PEEK (polyether- etherketone) 450G, fluoroelastomer and nitrile ‘O’-rings, hydrocarbon lubricant, felt, rare earth
magnets, silicon nitride, phenolic resin and carbon- fibre reinforced epoxy resin.
Table 1 - EXT 70H 24 V / EXT 70Hi 24 V technical data
Parameter
Mass2.8 kg2.8 kg4.7 kg2.8 kg
Inlet-flangeDN63ISO-KDN100ISO-KDN63CFDN40NW
Outlet-flangeDN16NWDN16NWDN16NWDN16NW
Vent-port1/8 inch BSP1/8 inch BSP1/8 inch BSP1/8 inch BSP
Pumping speeds are without inlet-screen or inlet-strainer. Inlet-screens and inlet-strainers reduce speed by
approximately 10% for ISO63. For ISO100 this reduction is less pronounced.
‡
Pb = backing pressure, Qi = flow through the interstage-port (EXT70Hi only), Qp = flow through pump-inlet
**
Ultimate pressure 48 hours after bakeout with 2-stage rotary vane backing-pump.
††
Ultimate pressure 48 hours after bakeout with Pb < 5 mbar (500 Pa).
‡‡
Inlet pressure has risen to 1 x 10-3 mbar (1 x 10-1 Pa).
***
Above this pressure, rotational speed drops below nominal.
†††
A larger backing-pump may be required for maximum throughput. A suitable diaphragm pump with ultimate
<
5 mbar may also be used.
Table 2 - EXT255H 24 V / EXT255Hi 24 V technical data
Edwards and the Edwards logo are trademarks of Edwards Limited.
Page 21
B722-20-880 Issue D
WARNING
3Installation
Safely route all vacuum, vent/purge gas and cooling-water pipelines, and all electrical cables and
wires, so that people cannot trip over them.
3.1Unpack and inspect
The pump is packed to prevent damage in transit. Take care when unpacking the pump to avoid excessive shocks
which could damage the bearings and reduce the life of the pump. The pump is supplied with the inlet and outlet
sealed to prevent entry of dust and vapour. Do not remove these seals until you are ready to install the pump on the
vacuum system.
Remove all packing materials and check the pump. If the pump is damaged, notify your supplier and the carrier in
writing within three days; state the Item Number of the pump together with your order number and your supplier’s
invoice number. Retain all packing materials for inspection. Do not use the pump if it is damaged.
Check that the package contains the items listed in Table 3. If any of these items is missing, notify your supplier in
writing within three days.
Installation
Table 3 - Checklist of items
QtyDescriptionCheck (9)
1EXT turbomolecular pump
1Inlet seal (trapped ‘O’ ring Co-Seal or copper compression gasket)
1Inlet-strainer (EXT255Hi only: fitted in the interstage-port)
If the pump is not to be used immediately, store the pump in suitable conditions as described in Section 6.1.
Do not discard the packing materials; retain them to repack the pump when you return it for service.
3.2Typical installation
A typical pumping system with an EXT255H 24V pump is shown in Figure 4. A typical pumping system with an EXT255Hi
24 V pump is shown in Figure 5. When necessary, purge the EXT255H/Hi 24 V pump with inert gas as described in
Section 3.5.
Typical pumping systems with an EXT70H 24 V and EXT70Hi 24 V pumps will be as shown in Figure 4 and 5, but without
the gas purge components (15, 16).
The accessories available for these EXT pumps are detailed in Section 7; the accessories are shown in Figure 7.
Install the pump in the vacuum system before you connect the EXDC Drive Module. This will ensure
that the pump cannot operate and injure people during installation.
3.3.1Inlet-screen
Do not remove the inlet-screen unless you can be sure that there is no danger that debris can fall into the pump. In
order to avoid the danger of injury from the rotor blades, do not remove the inlet-screen until you are ready to mount
the pump onto your system. If the screen is removed, the pumping speed will increase by approximately 10%. It is
not possible to remove the inlet-screen from a pump with an NW inlet-flange.
To remove the inlet-screen from a pump with an ISO or CF inlet-flange, carefully extract it from the inlet-flange using
a bent wire hook. To replace a screen which has been removed, install it as shown in Figure 6, with the Edwards logo
uppermost. Ensure that the dimples on the rim of the screen engage in the groove in the pump flange. If necessary,
gently bend the tags of the screen outwards to ensure a tight fit.
Figure 6 - Correct installation of the inlet-screen
3.3.2Mechanical fixing
Do not operate the EXT pump un til it is sec urely fixed . If th e pump seizes, the stored energy of the
rotor can cause rapid movement of the pump, which may cause damage and injury to people.
There are two ways in which the EXT pump can be securely fixed. Ideally, the EXT pump should be securely fixed by
its inlet-flange to a rigid, firmly fixed vacuum system: see Section 3.3.4. If this is not possible because of the nature
of the vacuum system, the base of the EXT pump must be securely fixed to a firm support: see Section 3.3.3.
Edwards and the Edwards logo are trademarks of Edwards Limited.
Page 25
B722-20-880 Issue D
3.3.3Base mounting
Ensure that the base of the pump is securely fixed to a firm support (refer to Figure 2 and 3 for the fixing hole
details). If the pump supports the weight of the vacuum system, the mass of the vacuum system must be no more
than 10 kg for an EXT70H/Hi pump, or no more than 20 kg for an EXT255H/Hi pump.
You must als o e n s u r e that yo ur m o u n t ing metho d m e e t s the following requirements, so that the EXT pump will remain
secure in the event of a pump seizure:
z The support mounting must be able to withstand a destructive torque of 333 Nm for an EXT70H/Hi pump or
620 Nm for an EXT255H/Hi pump.
z Fit cap-head fixing screws through the tapped fixing-holes in the base of the pump (see Figure 2 and 3): use
M4 screws for an EXT70H/Hi pump, and use M5 screws for an EXT255H/Hi pump.
z The fixing screws must comply with ISO 898-1, with a strength class of 12.9 (nominal tensile strength
1200 MPa).
z The fixing screw engagement length must be 6 mm or more.
z Tighten the fixing screws to a torque of 6 Nm (0.61 kgf m) on EXT70H/Hi pumps, and to a torque of 12 Nm
(1.22 kgf m) on EXT255H/Hi pumps.
3.3.4Inlet connection and orientation
Installation
Th e EXT pump c a n be f ixed t o the v acuum syste m by th e inle t-fla nge. T he pum p can b e mou n ted i n any a ttitu de fro m
vertical and upright through to horizontal (±2°). If the pump is mounted horizontally and you use a rotary vane pump
to back the EXT pump, the backing port must point vertically downwards (±20°) to reduce the risk of contamination
from the backing pump oil.
Make sure that the pump-inlet and all components fitted to the pump-inlet are clean and dust-free. If the pump-inlet
is not kept clean, the pump-down time may be increased.
The inlet-connection of the EXT pump is a CF flange, an ISO flange or an NW flange:
z If the pump has a CF flange, use the copper compression gasket supplied with the pump and use a full
complement of bolts to connect the inlet-flange of the pump to the vacuum system.
z If the pump has an ISO flange, use the Edwards trapped ‘O’ ring supplied with the pump and use a minimum
of four claw clamps to connect the inlet-flange of the pump to the vacuum system. Ensure that each claw
clamp is tightened to a torque of 10 Nm or more.
Alternatively, use a rotatable collar and the trapped ‘O’ ring supplied with the pump to connect the inlet-flange of
the pump to the vacuum system; use a full complement of bolts with the rotatable collar.
z If the pump has an NW flange, use the Co-Seal supplied with the pump and a suitable NW clamp to connect
the inlet-flange of the pump to the vacuum system.
Ensure that no torque or other forces are transmitted to the pump from the vacuum system or the associated
pipelines.
If necessary, fit an inlet vibration isolator between the pump-inlet and the vacuum system: refer to Section 7.4.7 for
the Item Numbers, and refer to the instruction manual supplied with the vibration isolator for installation details. If
you fit a vibration isolator, you must securely fix the base of the EXT pump as described in Section 3.3.3.
Note:The first time you pump down the system to vacuum, you must re-tighten the bolts which secure the inlet-
Use suitable vacuum tubing and connectors to connect the NW flange of the backing-port to your backing-pump. If
necessary, use flexible pipe or bellows to reduce the transmission of vibration from the backing-pump to the EXT
pump.
It is recommended that you use an Edwards EM or RV backing-pump. The minimum size of the backing-pump required
is given in Table 1 and 2. It may be necessary to use a larger backing-pump if the pump is run at a high inlet pressure
or high throughput, or if an EXT255H/Hi 24 V pump is purged with more than 25 sccm (0.42 mbar l s-1, 42 Pa l s-1) of
purge gas.
Do not use the EXT pump with a backing pressure below 1 x 10
increase the evaporation rate of the lubricating oil and so will reduce the life of the bearings.
3.3.6Interstage connection ((EXT70Hi 24 V and EXT255Hi 24 V only)
Use suitable vacuum tube and connectors to connect the interstage-port to the vacuum sy s tem or to t he outle t flange
of another turbomolecular pump (if using the EXT70Hi 24 V or EXT255Hi 24 V pump to back another turbomolecular
pump). Leave the inlet-strainer in the interstage-port of the EXT255Hi 24 V pump, unless you are sure that debris
cannot be drawn into the interstage-port.
3.4Vent-valve connection and control
-4
mbar (1 x 10-2 Pa). Lower backing pressures will
When designing your system and when you install a vent-valve, take note of the information in Section 1.3 and 2.3.
The EXT pump and vacuum system can be vented by any of the following methods:
z Use the manual vent-valve supplied. Take care not to open the manual vent valve too quickly (see below).
z Use a TAV5 or TAV6 solenoid vent-valve accessory (see Section 7) in place of the manual vent-valve.
z U s e a TAV 5 o r TAV6 s o len o i d ve n t -valve connected to a convenient flange on your vacuum system.
z Use an alternative valve connected to the vacuum system.
If using the manual vent-valve, the vent-valve must be opened only after the EXT pump speed has fallen to 50% of
full rotational speed.
If using the TAV5 vent-valve, the pump may only be vented when it is at full speed if the vacuum system has a volume
of 5 litres or more. If using the TAV6 vent-valve, the pump may only be vented when it is at full speed if the vacuum
system has a volume of 10 litres or more. I f the vo lume o f your vacuu m syst em is less than 5 litres (when using a TAV5
vent-valve), or if the volume of your vacuum system is less than 10 litres (when using a TAV6 vent-valve), you can
incorporate a suitable vent-restrictor (see Table 4) and vent the pump when it is at full speed.
If using another vent-valve, ensure that a suitable vent-restrictor is fitted to your vacuum system to limit the rate of
pressure rise: refer to Table 5. If you do not fit a suitable vent-restrictor, you must open the vent-valve only after
the speed of the EXT pump has fallen to 50% of full rotational speed.
If you connect a vent-valve to your vacuum system, select a point upstream of the EXT pump to prevent
backstreaming of oil from the backing pump. Do not connect the vent-valve to the backing pipeline. Connect the
inlet of the vent-valve to the vent gas supply (refer to Section 2.3 for the vent gas specification).
Note:If you use a vent-restrictor, you may find that the time required to vent your vacuum system is unacceptably
long. You may be able to reduce the vent time if you use a vent-valve without a vent-restrictor and wait
until the pump speed has fallen to 50% of full rotational speed before you open the vent-valve.
Edwards and the Edwards logo are trademarks of Edwards Limited.
Page 27
B722-20-880 Issue D
CAUTION
Table 4 - Vent-restrictor orifice diameter (with atmospheric pressure at the inlet of the vent-valve)
Vacuum system volume (l)Orifice diameter (mm)
< 20≤ 1.0
< 10≤ 0.7
< 5≤ 0.5
< 2≤ 0.35
3.5Purge gas connection (EXT255H 24 V and EXT255Hi 24 V only)
3.5.1Connect the purge gas
To supply a purge gas to the pump, remove the manual purge-valve/plug fitted in the purge-port, fit a vent-port
adaptor (see Section 7.4) to the purge-port and connect the purge gas supply to the vent-port adaptor. The purge
gas must comply with the specification given in Section 2.4.
The flow rate of the purge gas must be limited to the allowed range, also specified in Section 2.4. To limit the flow
rate, use a flow controller or a pressure regulator and calibrated flow restrictor. The PRX10 purge restrictor accessory
(see Section 7.4) is suitable for this purpose. Adjust the PRX10 as described in the instruction manual supplied with
the accessory.
Installation
3.5.2Recommended purge gas flow
The recommended purge gas flow for typical applications is 25 sccm (0.42 mbar l s-1, 42 Pa l s-1). This flow will protect
the pump when pumping oxygen in concentrations above 20% by volume.
3.6Electrical installation
Always make the electrical connections to the EXT pump after the pump has been installed on your vacuum system.
The EXT pump should be electrically bonded to earth (ground) using the connection provided (see Section 3.6.2).
3.6.1EXDC drive module
There are two EXDC drive modules, the EXDC80 24 V and the EXDC160 24V. The EXDC80 24 V drive should be used in
conjunction with the EXT70H/Hi 24 V pumps - an EXDC160 should not be used with this family of pumps. It is
recommended that the EXT255H/Hi 24 V pumps are driven by the EXDC160 24 V module. However an EXDC80 24 V
may be used, although this will result in significantly extended ramp times and reduced performance.
Fit the connector on the EXDC Drive Module directly to the 19-way connector on the EXT pump.
The EXDC Drive Module is designed to allow a pumping system to be configured in a variety of ways. Refer to the
EXDC Drive Module instruction manual to complete the electrical installation.
Do not disconnect or connect the EXDC drive module while the pump is rotating. If you do, you may damage the
drive module.
It is recommended that you fit a separate earth (ground) conductor to earth (ground) the EXT pump: use an
uninsulated braid or a separate insulated green/yellow conductor, and use the M5 x 10 screw and shakeproof washer
supplied (fitted to the earth (ground) hole on the pump) to secure the earth (ground) conductor to the EXT pump.
For external protective earth (ground) systems: the installation should be permanent and use the same earth (ground)
system as the EXDC Drive Module; the impedance between the pump-body and the earth (ground) connection point
must be < 0.1 Ω.
3.7Cooling
3.7.1Introduction
You must ensure that the pump is adequately cooled to prevent damage to the bearing lubricant.
We recommend that wherever possible, you cool the pump by forced-air or water-cooling. You must use watercooling in any of the following operating conditions:
z Backing pressure > 10 mbar (1000 Pa) and interstage flow > 20 sccm (0.3 mbar l s
z Backing pressure > 8 mbar (800 Pa) and interstage flow > 30 sccm (0.5 mbar l s
z Backing pressure > 5 mbar (500 Pa) and interstage flow > 80 sccm (1.3 mbar l s
-1
, 30 Pa l s-1).
-1
, 50 Pa l s-1).
-1
, 130 Pa l s-1).
z Ambient temperature > 35 °C.
z When using a bakeout band.
In all other operating conditions, forced-air cooling may be used. When using forced-air to cool the pump, ensure
that there is an adequate supply of cooling-air to the pump.
However, for light pumping duties, with ambient temperature < 30 °C and inlet-flange temperature < 70 °C, you can
use free convection to cool the EXT70H 24 V pump.
During operation, if the temperature of any surface of the pump is higher than 45 °C, the pump is too hot and cooling
must be increased.
3.7.2Forced-air cooling
An air-cooler accessory is available for the EXT pumps (refer to Section 7). Fit the air-cooler as described in the
instruction manual supplied with it. If you wish to use an alternative fan for air-cooling, ensure that the flow rate is
above 70 m
Edwards and the Edwards logo are trademarks of Edwards Limited.
Page 29
B722-20-880 Issue D
3.7.3Water-cooling
A water-cooler accessory is available for the EXT pumps (refer to Section 7.3.4). Fit the water-cooler as described in
the instruction manual supplied with it.
The cooling-water supply must comply with the specification given in Section 2.5. Pipes in the water-cooling circuit
may become blocked if the cooling-water contains too much calcium carbonate or if it contains particulates which
are too large. Corrosion of the water-cooling circuit may occur if there is too little calcium carbonate and oxygen in
the water. Good quality drinking water is usually suitable for water-cooling. If in doubt, you must check the quality
of your cooling-water supply and, if necessary, provide treatment and filtration.
Connect the cooling-water supply to the water-cooler on the pump as described below. Either of the two riffled
connectors on the water-cooler can be used for the water supply or return connections.
1. Push reinforced hose (approximately 6 mm internal diameter) over the ends of the riffled hose connectors on
the water-cooler on the pump.
2. Attach the hose with strong hose clips and make sure that they are tightened securely.
Alternatively, unscrew the riffled hose connectors and remove them from the water-cooler and make direct
connections to the 1/8 BSP female threaded fittings on the water-cooler.
You must turn off the cooling-water supply when you switch off the pump to prevent condensation of vapours inside
the pump.
Installation
To remove the pump for maintenance without breaking the cooling-water circuit, unscrew the two M4 cap-head
fixing-screws and remove the water-cooler from the pump. Make sure that there is a layer of thermal contact grease
on the water-cooler before you refit it to the pump.
Edwards and the Edwards logo are trademarks of Edwards Limited.
Page 31
B722-20-880 Issue D
WARNING
WARNING
4Operation
Do not operate the pump unless it is connected to your vacuum system. If you do, the pump rotor
can cause injury. The pump rotor rotates at very high speeds and you may not be able to see that
the pump is rotating. Always ensure the pump has stopped rotating before removing it from the
vacuum system.
Do not expose any part of your body to vacuum. If you do, you may be injured.
Note:The following sections describe the use of the EXT pump in a basic, manually-controlled pumping system
with a manual vent-valve, and with an EXDC drive module. Refer to the EXDC drive module instruction
manual for full details of installation and electrical connection.
4.1Start-up
Operation
1. Turn the manual vent-valve clockwise to close it.
2. Turn on the cooling-water supply (if water-cooling is used) or fan (if air cooling is used).
3. Start the backing-pump.
4. Start the EXT pump by supplying power to the EXDC drive module (see EXDC instruction manual).
Note:For large vacuum systems (100 litres or larger) wait until the system pressure reaches 10 mbar (1 x 103 Pa)
before supplying power to the EXDC drive module.
5. The pump will then accelerate to normal operating speed. When this has been reached, the green LED on the
EXDC will light.
Note:The first time you pump down the system to vacuum, you must re-tighten the bolts which secure the
inlet-flange: refer to Section 3.3.4.
4.2Shut-down
Note:In an emergency only, open the vent-valve quickly to decelerate the pump rotor in the shortest possible
time.
1. Switch off the backing-pump and stop the power to the EXDC drive module (see EXDC instruction manual) to
switch off the EXT pump.
2. When the EXT pump rotational speed has fallen to below 50% of full rotational speed, you may turn the manual
vent-valve anticlockwise to open it. Ensure that the rate of pressure rise does not exceed the allowed rate of
pressure rise, otherwise the pump may be damaged: refer to Section 1.3 and 2.3.
3. If water-cooling is in use, turn off the cooling-water supply.
The EXDC Drive Module monitors the temperature of the EXT pump and the electrical power consumption of the
pump. If the EXDC Drive Module detects excessive power consumption or temperature, the rotational speed of the
pump motor is reduced until the power and temperature return to normal.
Shut down the EXT pump if its speed falls below 50% of full rotational speed. If you do not, the EXT pump may
be damaged.
If the pump speed is reduced to 50% of nominal speed this means that the inlet pressure or pump temperature are
too high, or that there is a fault with the pump (for example, the bearings may have failed). If the pump speed falls
to below 50% of its full rotational speed you must therefore switch the pump off. Instructions on how to monitor pump
speed can be found in the EXDC drive module manual.
The EXDC has an electronic EXT pump speed control system. This control system prevents operation of the EXT pump
at over 100% of its normal full rotational speed (as operation above this level reduces bearing operational life). In
the unlikely event of a failure of this control circuit the pump may overspeed. If you suspect that your pump may be
running at overspeed you should switch it off and consult Edwards or your supplier.
4.4Bakeout
When baking the EXT pump to above 70 °C at the inlet-flange, the pump must be water-cooled to prevent
damage to the bearing lubricant.
If you heat your EXT pump (and your vacuum system), you will speed up the degassing process so that the pump will
reach ultimate vacuum in the shortest possible time. If you heat the pump, this will also prevent condensation of
vapours inside the pump.
The Edwards BX bakeout band may be used to heat the pump (refer to Section 7). Fit the band around the pump, just
below the inlet-flange. When baking the pump or the system, make sure that the temperature of the inlet-flange
does not exceed 100 °C.
When baking vacuum system, if the temperature of the system exceeds 200 °C, you must put a radiation shield
between the system and the EXT pump. This radiation shield will reduce the heat radiated onto the pump rotor.
Typically, a bakeout of four hours is long enough to remove water condensation from the pump. However, the
bakeout time will depend on the amount of condensation in the pump and the vacuum system, and the ultimate
pressure you want to achieve.
Edwards and the Edwards logo are trademarks of Edwards Limited.
Page 33
B722-20-880 Issue D
CAUTION
WARNING
WARNING
5Maintenance
Allow the pump-rotor to stop, then disconnect the EXDC Drive Module from the pump before you
remove the pump from your vacuum system for maintenance or fault-finding procedures.
5.1Introduction
The maintenance operations for the EXT Turbomolecular pumps are described in the following sections. The ISX inletscreen, the WCX water-cooler, the inlet-strainer and inlet-flange seals are available as spares (refer to Section 7).
Fit the ISX inlet-screen as described in Section 3.3.1. Fit the WCX water-cooler as described in Section 3.7.3.
5.2Bearing life
When supplied, the pump contains sufficient lubricant to supply the bearings for life. No routine maintenance is
therefore required between bearing replacements. The bearings are not user-serviceable. The bearings will need to
be replaced when they reach the end of their service life. This is typically more than 20,000 hours, but may be less;
this depends on the type of pumping duty on which the pump is used.
Maintenance
When the bearings need replacement, we recommend that you exchange your pump for a factory reconditioned
replacement. Alternatively, you can send your pump to an Edwards Service Centre to have the bearings replaced.
When returning EXT pumps to Edwards Service Centres please use the procedure included at the end of this manual.
However, the instruction to drain all fluids does not apply to the lubricant in the EXT pump oil-reservoirs.
5.3Rotor life
The life of the EXT pump rotor is typically 40,000 to 50,000 cycles (of acceleration to full speed, and then
deceleration to a stop). The pump rotor is not user-serviceable.
We therefore recommend that you exchange your pump for a factory reconditioned replacement every 20,000 cycles,
or 10 years of use, whichever occurs first. Alternatively, you can send your pump to an Edwards Service Centre for a
major service (which will include rotor replacement).
When you return EXT pumps to Edwards Service Centres please obey the procedure included at the end of this
manual. However, the instruction to drain all fluids does not apply to the lubricant in the EXT pump oil-reservoirs.
5.4Clean the pump
Clean the external surfaces of the EXT pump in a well-ventilated location. When you use cleaning
solutions and solvents to clean the pump, observe all precautions specified by the manufacturer.
Avoid inhalation of any particulates which may be present in the pump.
Do not attempt to clean any parts of the EXT pump other than the external surfaces. Organic solvents may
damage internal pump components. Do not use abrasive materials to clean any part of the pump.
If the inside of the EXT pump is contaminated, it may not be possible to achieve the specified ultimate vacuum, or
pump-down time may increase. In these circumstances, you should return the pump to an Edwards Service Centre,
where the pump will be dismantled and cleaned. Use the procedure given in the forms at the end of this manual to
return the pump.
You can use any organic solvent to clean the external surfaces of the EXT pump. We recommend that you use nonCFC solvents, such as isopropanol or ethanol. Use a cleaning solution which is suitable for the contaminants on the
pump surfaces.
For environmental reasons, keep wastage of cleaning solutions and solvents to a minimum.
5.5Fault finding
Refer to Table 5 for the possible causes of faults and for the recommended actions to rectify faults.
Refer to the fault finding section of the instruction manual supplied with your EXDC Pump Drive Module for further
fault finding information.
Tab l e 5 - F a u l t f i n di ng
SymptomCheckAction
The pump does not rotate.Check that the electrical supply is
on.
The pump is not rotating at full
speed.
Is the inlet pressure too high?If so reduce the pumping load or
Is the EXT pump running too hot?Increase the cooling-water flow or
Does the rotor rotate freely?If not the EXT pump-bearings are
If power is supplied and the pump
still does not rotate then there is a
fault with either the pump or
controller - consult Edwards or your
supplier.
check for a gross leak into the
system.
decrease the water temperature or
do both. You may need to change
from air-cooling to water-cooling.
(Refer to Section 2 for maximum
inlet pressure and cooling
requirements). Check that external
heat sources (such as system bakeout
heaters) are not excessive.
Edwards and the Edwards logo are trademarks of Edwards Limited.
Page 35
B722-20-880 Issue D
Table 5 - Fault finding (continued)
SymptomCheckAction
Ultimate pressure cannot be
reached.
The EXT is very noisy or there is
excessive vibration or both.
None of the above.-Consult Edwards or your supplier.
Is the pressure limited by water
vapour ?
Are any of the vacuum gauges
contaminated ?
Is the pumping speed insufficient
(due to poor conductance between
the pump and the gauge or too large
a chamber) ?
Is the interstage inlet pressure > 0.5
mbar (50 Pa) (for the EXT70Hi 24V)
or > 0.2 mbar (20 Pa) (for the
EXT255Hi 24V) ?
Is the backing pressure > 10 mbar
3
(1 x 10
or > 5 mbar (5 x 10
EXT255H/Hi 24 V ?
Is the high-vacuum area of the
system contaminated ?
Check the rest of your system for
leaks and contamination.
Remove the pump from the system
and test the ultimate pressure of the
pump alone (see Section 2 for
specification).
Is the pump rotational speed the
same as the resonant frequency of
the attached system ?
Is the vibration being transmitted
from the rotary pump ?
Is the noise irregular and getting
progressively worse ?
Is the EXT making a constant highpitched noise ?
Pa) for the EXT70H/Hi 24V,
2
Pa) for the
Bake the system and pump.
If so clean or replace them.
Increase the conductance or reduce
the volume.
If the interstage inlet pressure is too
high inlet pressure at the
turbomolecular inlet is reduced;
ensure that the interstage inlet
pressure is < 0.5 mbar (50 Pa) or
< 0.2 mbar (20 Pa) as appropriate.
If so, the backing pressure is too
high. Check for backing pipeline
leaks. If the throughput is high you
may need a larger backing-pump.
If so clean the high-vacuum system.
If found clean the contaminated
areas and repair the leaks.
If poor check the pump for
contamination and if necessary
return the pump as described in
Section 5.4. Leak-check the pump. If
the leak rate > 1 x 10
(1 x 10-5 Pa l s-1) consult Edwards or
your supplier.
If so change the natural frequency of
your system or isolate the pump
using flexible bellows.
If so fit flexible bellows or a
vibration isolator in the backing line.
If so a bearing is defective. Consult
Edwards or your supplier.
If so the rotor is out of balance.
Consult Edwards or your supplier.
Edwards and the Edwards logo are trademarks of Edwards Limited.
Page 37
B722-20-880 Issue D
WARNING
6Storage and disposal
6.1Storage
Use the following procedure to store the pump.
1. Place protective covers over the inlet, outlet, interstage (EXT70Hi 24 V and EXT255Hi 24 V only), purge and vent
ports.
2. Place the pump in its packing materials. For fastest pump-down when the pump is put back into service, seal the
pump inside a plastic bag together with a suitable desiccant.
3. Store the pump in cool, dry conditions until required for use. When required, prepare and install the pump as
described in Section 3.
4. Keep the pump upright at all times to prevent the drainage of oil from the bearing reservoir.
5. Avoid long-term storage if possible. When long-term storage is necessary, the pump should be set up and run for
at least eight hours every six months.
6.2Disposal
Storage and disposal
In the unlikely event that the rotor fails, dust may be generated from the carbon fibre reinforced
components in the pump. If the rotor has failed:
z Wear the appropriate personal protective equipment (gloves, a dust-proof mask which covers
the nose and mouth, an apron or overalls, and safety glasses or goggles) when you handle the
pump.
z Ensure that the pump inlet, outlet, and vent-, interstage- and purge-ports are all blanked off/
sealed before you dispose of the EXT pump.
Dispose of the EXT Turbomolecular Pump and any components and accessories safely in accordance with all local and
national safety and environmental requirements.
Take particular care with any components which have been contaminated with dangerous process substances.
Take appropriate action to avoid inhalation of any particulates which may be present in the pump.
Do not incinerate the pump. The pump contains phenolic and fluorosilicone materials which can decompose to very
dangerous substances when heated to high temperatures.
Edwards and the Edwards logo are trademarks of Edwards Limited.
Page 39
B722-20-880 Issue D
7Service, spares and accessories
7.1Introduction
Edwards products, spares and accessories are available from Edwards companies in Belgium, Brazil, China, France,
Germany, Israel, Italy, Japan, Korea, Singapore, United Kingdom, U.S.A and a world-wide network of distributors.
The majority of these employ service engineers who have undergone comprehensive Edwards training courses.
Order spare parts and accessories from your nearest Edwards company or distributor. When you order, please state
for each part required:
z Model and Item Number of your equipment
z Serial number (if any)
z Item Number and description of the part
7.2Service
Edwards products are supported by a worldwide network of Edwards Service Centres. Each Service Centre offers a
wide range of options including: equipment decontamination; service exchange; repair; rebuild and testing to factory
specifications. Equipment which has been serviced, repaired or rebuilt is returned with a full warranty.
Service, spares and accessories
Your local Service Centre can also provide Edwards engineers to support on-site maintenance, service or repair of
your equipment.
For more information about service options, contact your nearest Service Centre or other Edwards company.
7.3Spares
7.3.1ISX inlet-screen
An inlet-screen is fitted to your pump as supplied to prevent damage from the entry of debris into the pump. The
Item Numbers of replacement inlet-screens are given below. Select the inlet-screen according to the pump inletflange size. You cannot replace the inlet-screen on a pump with an NW inlet-flange.
A water-cooler can be fitted to the EXT pump. Please refer to Section 3 to check the suitability of water cooling for
a particular application. Pumps with a Conflat flange are supplied with a water-cooler.
Water-coolerItem Number
WCX500B736-01-036
7.3.4Inlet-flange seals
EXT pumps are supplied with an inlet seal. The Item Numbers of replacement seals are given below.
Inlet-flangeInlet-sealItem Number
DN63ISO-KISO63 trapped ‘O’ ring (fluoroelastomer)B271-58-17F
DN40NWDN40NW Co-Seal (fluoroelastomer)B271-58-453
DN100ISO-KISO100 trapped ‘O’ ring, fluoroelastomerB271-58-171
DN63CF63CF copper compression gasket (pack of 5)C081-00-003
DN100CF100CF copper compression gasket (pack of 5)C082-00-003
7.4Accessories
7.4.1Installation
The accessories available for use with the EXT turbomolecular pumps are described in the following Sections. Figure 7
shows how the accessories are fitted to an EXT pump.
7.4.2EXDC drive modules
There are two EXDC drive modules, the EXDC80 24 V and the EXDC160 24 V. The EXDC80 24V drive should be used in
conjunction with the EXT70H/Hi 24 V pumps - an EXDC160 should not be used with this family of pumps. It is
recommended that the EXT255H/Hi 24 V pumps are driven by the EXDC160 24 V module. However an EXDC80 24 V
may be used, although this will result in significantly extended ramp times and reduced performance.
Drive ModuleItem Number
EXDC80 24 VD396-45-000
EXDC160 24 VD396-46-000
7.4.3BX bakeout band
A BX bakeout band accelerates the degassing of the pump to enable it to achieve lower pressures. It may also be used
to protect the pump from condensation of contaminants. The bakeout bands are available in 110-120 V or 220-240 V
versions.
Edwards and the Edwards logo are trademarks of Edwards Limited.
Page 41
B722-20-880 Issue D
7.4.4FL20K foreline trap
The foreline trap minimises oil vapour backstreaming from the backing pump and is recommended where the highest
system cleanliness is required.
Foreline trapItem Number
FL20KA133-05-000
7.4.5TAV vent-valve and vent-port adaptor
Two solenoid-operated vent-valves are available for system venting. The valves are 24 V d.c., normally-open. The
solenoid-valve is fitted in place of the manual-valve, or alternatively can be fitted with an adaptor (supplied with
the valve) and be used with any suitable NW10 flanged port on your vacuum system.
The vent-port adaptor allows the vent-port (or the purge-port, o n EXT255H pumps) to be us e d w i t h a n y suitable NW10
fitting: see Figure 7, item 7.
An ACX air-cooler can be fitted to the EXT pump. However, please refer to Section 3 to check the suitability of aircooling in a particular application.
In applications where the small amount of vibration generated by the turbomolecular pump is a problem, a vibration
isolator can be fitted. The isolator consists of two special flanges separated by a flexible bellows and a rubber, antivibration, outer collar. The isolator required depends on the pump inlet-flange size.
The PRX10 is a modified DN10NW centring-ring which filters the purge gas and restricts its flow rate to the
recommended flow of 25 sccm. The restrictor is suitable for all EXT pumps fitted with a purge-port.
Note that you must fit a vent-port adaptor (see Section 7.4.5) to the purge-port on an EXT255H 24V or EXT255Hi 24V
pump, in order to connect the purge restrictor to the pump.
Use a VRX fixed orifice vent-restrictor to restrict the flow of vent gas into the EXT pump. You can fit a VRX ventrestrictor to the inlet of a TAV5 or TAV6 vent-valve or to a PRX10 purge-restrictor. Refer to Section 3.4 for
information on the selection of the correct VRX vent-restrictor.
On the EXT70H/ISO63 24 V pump, the interstage-port in the pump-envelope is blanked off. You can remove the
blanking cap and fit the interstage-port accessory to upgrade the pump to an EXT70Hi pump, so that you can use the
interstage-port of the pump as described in Section 3.3.4.
The accessory kit consists of an ‘O’ ring and a flange; the flange has an inlet strainer to prevent the entry of debris
into the pump. To fit the accessory:
z Use a suitable pin-spanner to remove the blanking plug from the interstage-port. The holes in the blanking
plug are 3.2 mm in diameter, and the centres of the holes are 14 mm apart.
z Fit the ‘O’ ring supplied, then use a 27 mm diameter ‘C’ spanner to screw the flange into the interstage-
port; the holes in the flange accessory are 3.2 mm in diameter.
z Tighten the flange sufficiently, to prevent the flange being loosened by vibration or stress transmitted from
pipelines connected to the flange.
AccessoryItem Number
EXT70H/ISO63 interstage-port accessory kitB580-69-001
Pin-spannerB279-29-029
27 mm ‘C’ spannerB279-29-028