Edwards EV1000 Setup Manual

VolumeView Set
EV1000 Clinical Platform
Setup Guide
System Setup
VolumeView sensor
VolumeView femoral arterial catheter
VolumeView thermistor manifold
PreSep oximetry catheter
TruWave pressure transducer
Oximetry cable
1
2
3
4
5
6
1
2
3
4
5
6
EV1000 Clinical Platform Startup with the VolumeView Set
Before startup, please ensure monitor, cables and patient cables are all connected.
1. Turn on monitor and databox. System self-test starts.
2. Enter new patient data OR continue with existing patient.
3. Set up and prime VolumeView sensor and TruWave CVP transducer and connect to databox.
4. Ensure the sensor and transducer are level to patient’s phlebostatic axis and zero transducers on monitor using Clinical Actions button.
5. Choose continuous parameters, and define alarms/target range.
6. Start transpulmonary thermodilution - inject bolus and review and edit boluses results.
7. Monitor patient in real time with one of the six available screens.
Thermodilution Bolus - VolumeView Set
1. Insert the VolumeView femoral arterial catheter according to the directions for use.
2. Insert a CVC (if using PreSep CVC, refer to PreSep catheter setup guide).
3. Ensure the TruWave central venous pressure transducer and VolumeView sensor are primed and then connected to the EV1000 databox.
4. Ensure the sensor and transducer are level to patient’s phlebostatic axis and zero transducers on monitor using Clinical Actions button.
5. Connect the VolumeView sensor to the VolumeView femoral arterial catheter and confirm the appropriate waveforms and pressures are displayed.
6. Connect the databox thermistor cable to the VolumeView femoral arterial thermistor and conrm appropriate temperature on the EV1000 monitor.
7. Prime the VolumeView set manifold, removing all air. Connect the TruWave central venous pressure transducer to the VolumeView manifold and then connect it to the central venous catheter and conrm the appropriate waveform and pressure are displayed.
8. Connect the in-line injectate thermistor probe to the VolumeView manifold.
9. Connect a pre-chilled syringe with (15 or 20 ml) normal saline to the VolumeView manifold valve/port.
10. Touch the Clinical Actions button and then touch the Thermodilution button to start the TPTD procedure.
11. Select the desired injectate volume on the keypad (prefer 15 or 20 ml).
12. Select Lung Resection, if appropriate, as well as the descriptor (for example RUL = right upper lobectomy).
13. Select Parameters, to choose indexed or non-indexed parameter.
14. Touch Start Set.
15. Wait is highlighted until the thermal baseline is reached.
16. When Inject is highlighted, use a rapid, smooth, continuous method to inject the cold solution with the volume amount previously selected.
17. Observe the calculation.
18. Repeat steps 14 and 15.
19. Touch review to edit the series of boluses.
20. Touch Accept.
2
Loading...
+ 2 hidden pages