to which this declaration relates is in conformity with the following standard(s) or other
normative document(s)
EN1012-2:1996+A1:2009 Compressors and Vacuum Pumps. Safety Requirements.Vacuum Pumps
EN61010-1:2010 Safety Requirements for Electrical Equipment for Measurement,
Control and Laboratory Use. General Requirements
EN60034-1: 2010 Rotating electrical machines. Rating and performance
C22.2 No77: 1995 Motors with inherent overheating protection
C22.2 No100: 2004 Motors and Generators
C22.2 61010-1-04: 2004 Safety requirements for electrical equipment for measurement,
Control and laboratory use – Part 1: General requirements
UL61010A: 2002 Safety requirements for electrical equipment for measurement,
Control and laboratory use – Part 1: General requirements
EN50581:2012 Technical Documentation for the Assessment of Electrical and
Electronic Products with respect to the Restriction of Hazardous
Substances
and fulfils all the relevant provisions of
2006/42/EC Machinery Directive
2014/35/EU Low Voltage Directive
2011/65/EU Restriction of Certain Hazardous Substances (RoHS) Directive
Note: This declaration covers all product serial numbers from the date this Declaration was
signed onwards.
16.05.2016, Burgess Hill
Mr Peter Meares
Senior Technical Support Manager, General Vacuum
This product has been manufactured under a quality management system certified to ISO 9001:2008
Edwards and the Edwards logo are trade marks of Edwards Limited.
A652–07–880 Issue D
CAUTION
WARNING
1Introduction
1.1Scope and definitions
This manual provides installation, operation and maintenance instructions for the Edwards E-LAB 2 Pump
(abbreviated to E-LAB 2 in the remainder of this manual). The pump must be used as specified in this manual. Read
this manual before installing and operating the pump.
Important safety information is highlighted as WARNING and CAUTION instructions; these instructions must be
obeyed. The use of WARNINGS and CAUTIONS is defined below.
Warnings are given where failure to observe the instruction could result in injury or death to
people.
Cautions are given where failure to observe the instruction could result in damage to the equipment, associated
equipment and process.
Introduction
The units used throughout this manual conform to the SI international system of units of measurement.
Refer to Figure 1. The E-LAB 2 is a complete laboratory pumping system, which has a two-stage, rotary-vane vacuum
pump. An oil mist filter (6) is fitted to the outlet of the pump, to prevent the discharge of oil mist. An inlet catchpot
(4) is fitted to the pump inlet to minimise the ingress of condensable vapours into the pump.
The E-LAB 2 has a gas ballast oil return assembly (16) which continuously returns oil trapped in the mist filter to the
pump.
The vacuum pumping mechanism is oil sealed. The level and condition of the oil in the oil box is inspected through
the sight glass (8). Diagrams on the side panels and on the oil change tag (Figure 3, item 4) give guidance on when
to change the oil. Two oil filler plugs (7) and an oil drain plug (9) are provided on the oil box.
The pump mechanism is driven directly by a single-phase electric motor through a flexible motor coupling. The motor
is totally enclosed and is cooled by the motor cooling fan which directs air along the motor fins. The pumps are cooled
by an additional fan attached to the motor coupling.
The motor has an on/off switch (12) and a thermal overload device. When the motor is too hot, the thermal overload
device switches off the pump. The thermal overload device has an automatic reset; when the motor cools down, the
device resets and the motor will restart, unless the system incorporates suitable control equipment that must be
manually reset (see Section 3.10.2).
The E-LAB 2 has a gas ballast oil return assembly, which is supplied configured to automatically deliver high gas
ballast flow rate into the pump; this prevents the condensation of vapours in the pumped gases and allows the use
the E-LAB 2 to pump high-vapour loads. If required, the gas ballast flow rate into the pump can be reduced or
switched off.
When the E-LAB 2 is switched off, an inlet valve seals the inlet and prevents the suck-back of air and oil into the
vacuum system.
The E-LAB 2 has a retractable lifting handle (15) and is mounted on a base plate on rubber feet (11).
Edwards and the Edwards logo are trade marks of Edwards Limited.
-1
-1
Figure 3 - Dimensions (mm)
1. Outlet
2. Inlet
3. Baseplate mounting holes: 4 x M8
4. Oil change tag
A652–07–880 Issue D
Technical data
2.3.2Noise and vibration data
Sound pressure: measured at ultimate vacuum 1 metre from the
end of the pump to ISO 11201, 50 Hz electrical supply
Vibration severity: measured at the inlet port to ISO 2372 (1974) Class 1C
50 dB(A)
2.4Electrical data
Note:Edwards recommends using fuses of the maximum ratings specified in Tables 1 and 2. Do not use fuses of a
higher rating.
The dual-voltage, dual-frequency motor is designed for a single-phase electrical supply and is suitable for 50 Hz or
60 Hz operation. The motor can be manually switched between nominal supply voltages of 110-120 V and 220-240 V
(refer to Section 3.10.1).
When starting a cold pump, the motor will draw a start-up current for up to several seconds, therefore use a
slow-blow fuse to prevent unnecessary fuse failure during pump start-up. Within five minutes, as the oil in the pump
warms up, the current drawn will slowly reduce to the full load current specified in Tables 1and 2.