Edwards and the Edwards logo are trademarks of Edwards Limited.
A343-10-880 Issue S
CAUTION
WARNING
1Introduction
1.1Scope and definitions
This manual provides installation, operation and maintenance instructions for the Edwards E1M18 and E2M18 rotary
vacuum pumps. The pump must be used as specified in this manual.
Read this manual before installing and operating the pump. Important safety i nformation is highli ghted as WARNING
and CAUTION instructions; these instructions must be obeyed. The use of WARNINGS and CAUTIONS is defined below.
Warnings are given where failure to observe the instruction could result in injury or death to
people. The actual symbol shown varies according to the hazard.
Cautions are given where failure to observe the instruction could result in damage to the equipment, associated
equipment and/or process.
Introduction
The units used throughout this manual conform to the SI international system of units of measurement.
The following warning labels may be present on the pump and used throughout the product documentation:
Warning - an appropriate safety instruction should be followed or a caution to a pote ntial hazard exists.
Warning - dangerous voltage. Indicates hazards arising from dangerous voltages.
Warning - hot surfaces. To indicate that the marked item can be hot and should not be touched without
taking precautions.
Warning - do not block the pump outlet.
Warning - use protective equipment. Indicates that protective equipment must be used.
This equipment is designed to meet the requirements of Group II Category 3 equipment in accordance with
Directive 94/9/EC of the European P arliament and the Council of 23rd March 1994 on the approximation of
the laws of the Member States concerning equipment and protective systems intended for use in potentially
explosive atmospheres. (The ATEX Directive)
The ATEX Category 3 applies in respect of potential ignition sources internal to the equipment. An ATEX
Category has not been assigned in respect of potential ignition sources on the outside of the equipment as
the equipment has not been designed for use where there is an external potentially explosive atmosphere.
There is no potential source of ignition within the pump during normal operation but there may be pote nt ial
sources of ignition under conditions of predictable and rare malfunction as defined in the Directive.
Accordingly, although the pump is designed to pump flammable materials and mixtures, operating
procedures should ensure that under all normal and reasonably predictable conditions, these materials and
mixtures are not within explosive limits. Category 3 is considered appropriate for the avoidance of ignition in
the case of a rare malfunction which allows flammabl e materials or mixtures to pass through the pump while
within their explosive limits.
When flammable or pyrophoric materials are present within the equipment:
Do not allow air to enter the equipment.
Ensure that the system is leak tight.
For further information, please contact Edwards: refer to the Addresses page at the end of this manual for
Edwards and the Edwards logo are trademarks of Edwards Limited.
Figure 1 - The E1M18/E2M18 pump
1. Oil filler
2. Outlet nozzle
3. Gas-ballast control
4. Gas-ballast inlet
5. Centring-ring and O-ring (supplied)
6. Inlet-port (adaptor flange)
7. Cable-gland/Amphenol connector position
8. Motor terminal box
9. On/Off switch
10.Box section skids
11.Oil drain-plug (gravity drain)
12.Oil sight-glass
13.Pump identification label
14.Oil drip tray
A343-10-880 Issue S
Introduction
Note:A pump with a single-phase motor is shown in this figure.
The motor shown in this figure is not representative of the motor used on the E1/E2M18 pumps with Item
Numbers A343-17-984 and A363-17-984. On these pumps, items 7 and 9 are transposed, with Item 7 being
an IEC60320 16-20 Amp socket.
The Edwards E1M18 and E2M18 pumps are shown in Figure 1. Refer to Figure 1 for item numbers in brackets in the
following descriptions.
The E1M18 and E2M18 pumps are direct drive, sliding vane pumps. The E1M18 is a single-stage pump and the E2M18
is a two-stage pump. The pump is oil-sealed and designed for reliable, long-term operation in both laboratory and
industrial environments. The pump is a free-standing unit. The drive is provided through a fle xible coupling by a
single-phase or three-phase (four pole) motor.
The motors are totally enclosed and are cooled by the motor-coo li ng fan wh ich directs air along the motor fins.
The single-phase motors have a thermal overload device. When the motor is too hot, the thermal overload device
switches off the pump. The thermal overload device has an automatic reset; when the motor cools down, the d evice
resets and the motor will restart. The single phase motors have an on/off switch.
An oil pressure system lubricates the pump shaft bearing surfaces and rotor sealing faces. The pump has an oil
distribution valve which prevents discharge of oil to the pump interior (suckback) after the pump stops. The pumping
chambers are air-tight, so this arrangement prevents air suckback unless the gas-ballast valve is open. For protection
in this case, refer to Section 1.4.
The level and condition of oil in the oil box reservoir can be inspected through the oil sight-glass (12). An oil fillerplug (1) is fitted to the top of the oil box. The pump has an oil drain-plug (11) to allow gravity oil drain.
The pump has an inlet-port (6), outlet nozzle (2) and gas-ballast control (3). The pump is mounted on two mild steel
box section skids (10) on rubber pads. Details of suitable vibration isolators and other recommended accessori es are
given in Section 7.4.
1.4Gas-ballast
When using the pump with high vapour throughputs, the gas-ballast facility should be used to prevent condensation
of the vapours inside the pump. The con densates will contaminate the oil, will cause performance to deteri orate and
may cause corrosion of the pump mechanism.
Air (or an inert gas) can be introduced into the pump mechanism through the gas-ballast control (Figure 1, item 3).
The gas-ballast control is a multi-turn valve:
From the closed position, the first two turns of the gas-ballast control provide an additional oil-feed to the
pump mechanism, but do not introduce gas-ballast into the pump. The additional oil-feed improves the
lubrication and sealing of the pump mechanism.
With further turns of the gas-ballast control, the gas-ballast flow into the pump increases from zero flow,
Edwards and the Edwards logo are trademarks of Edwards Limited.
A343-10-880 Issue S
2Technical data
2.1Operating and storage conditions
Note:To comply with EN 61010 and CSA standards, the pump must be installed and used indoors and within the
operating conditions specified in Table 1 below.
Table 1 - Operating and storage conditions
ParameterData
Ambient temperature range (operation)13 to 40°C
Ambient temperature range (storage)-30 to 70°C
Normal surface temperature of the pump-body
Maximum humidity (operation)90 % RH
Maximum altitude (operation)2000 m
Pollution degree2
Installation categoryII
Noise level at 1 metre57 dB(A) (at 50 Hz)
*
At ultimate vacuum, with ambient temperature of 20°C.
*
45 to 65°C
Technical data
2.2Performance
Note:Where total pressures are shown in Table 2 below, measurements were taken using an untrapped total
pressure capacitance diaphragm gauge on a header, as specified by Pneurop standards.
Table 2 - Performance data
Parameter
Maximum displacement
50 Hz electrical supply20.5 m
60 Hz electrical supply25.0 m3 h
Maximum pumping speed - Pneurop
50 Hz electrical supply17.0 m
60 Hz electrical supply20.4 m3 h
Motor rotational speed
50 Hz electrical supply1440 r min
60 Hz electrical supply1720 r min
Ultimate vacuum
without gas-ballast (partial pressure)2 x 10
without gas-ballast (total pressure)3 x 10-2 mbar
with full gas-ballast (partial pressure)6.5 x 10-1 mbar
Maximum water vapour pumping rate0.65 kg h
Maximum permitted outlet pressure
(for full pump throughout)
0.5 bar gauge
1.5 bar absolute
1.5 x 10
E1M18E2M18
-1
5
Pa
Data
0.3 kg h
0.5 bar gauge
1.5 bar absolute
1.5 x 105 Pa
-1
2.3Mechanical data
Table 3 - Mechanical data
Approximate pump mass38 kg (E1M18), 41 kg (E2M18)
DimensionsRefer to Figure 2
Degree of protection
Single-phase motorsIP44
Three-phase motorsIP54
Pump inlet portNW25 (the flange can be removed from the 1 inch BSP
threaded hole)
Pump outlet port15 mm external diameter nozzle (the nozzle can be
removed from the 3/4 inch BSP threaded hole)
2.4Lubrication data
Note:Edwards Material Safety Data Sheets for the oils specified below are available upon request.
Table 4 - Lubrication data
Hydrocarbon pumps:
Recommended oil
Maximum oil capacity
E1M181.4 litres
E2M181.05 litres
PFPE-prepared EM pumps:
Recommended o ilKrytox 1506 or Fomblin 06/6
Maximum oil capacity
E1M181.4 litres
E2M181.05 litres
*
To operate the pump when the ambient temperature is outside the range specified in Section 2.1 or to
optimise pump performance when processing condensible vapours, a different oil may be required.
Edwards and the Edwards logo are trademarks of Edwards Limited.
A343-10-880 Issue S
2.5Electrical data
Refer to Table 5 and 6. The motor start-up current is drawn for less than one second, so time-lag fuses must be used
to prevent unnecessary fuse failure when the pump starts. Fuses should be to EN60269 Section 2.2. If using the pump
at temperatures lower than 13°C, the start-up current will be drawn for longer; this may cause the motor thermal
overload device to open.
Edwards and the Edwards logo are trademarks of Edwards Limited.
3Installation
WARNING
WARNING
WARNING
WARNING
3.1Safety
Ensure that the installation technician is familiar with the safety procedures which relate to the
pump oil and the products handled by the pumping system.
If a hydrocarbon oil is used in this pump, do not use the pump to process oxygen in concentrations
greater than 25% in volume. There is a risk of fire or explosion in the oil-box of the pump.
Do not use the E1M18 or E2M18 pump to pump hazardous substances.
A343-10-880 Issue S
Installation
Obey the safety instructions listed below and take note of appropriate precautions. Failure to do
so can cause injury to people and damage to equipment.
Prevent any part of the human body from coming into contact with the vacuum.
Wear the appropriate safe ty clothing when coming into contact with contaminated components is
anticipated. Dismantle and clean contaminated components inside a fume cupboard.
Vent and purge the vacuum system before installation work is started.
Take suitable precautions to avoid the inhalation of oil mist and excessive skin contact with pump oil, as
prolonged exposure can be harmful.
Disconnect the other components in the pumping system from the electrical supply so that they cannot be
operated accidentally.
Safely route any electrical supply cables so that they cannot accidentally trip people.
Ensure that the E1M18 or E2M18 pump is suitable for the application. If in doubt as to the suitability of the
pump for the application, refer to the Edwards guidelines on vacuum pump and vacuum system safety (refer
to the Associated Publication at the end of the contents list at the front of this manual) or contact Edwards
for advice.
3.2System design
Consider the following points when designing the pumping system:
Edwards recommends the use of a foreline vacuum isolation valve to allow the pump to warm up before
pumping condensable vapours or if a vacuum needs to be maintained when the pump is not running.
Avoid high levels of heat input to the pump from the process gases, otherwise the pump may overheat and
seize, and cause the motor thermal overload device to open.
If using the pump in a high ambient temperature with a high gas throughput, the temperature of the pump
body may exceed 70°C. Edwards recommends the use of ad ditional guarding to prevent contact with hot
surfaces under these conditions.
Make sure that the exhaust pipeline canno t b ecom e blo cke d. If an ou tle t-i sola ti on val ve is fi tt ed , ma ke sur e
the pump cannot be operated with the valve closed.
Provide for a purge of inert gas when the pumping system is shut down, to dilute dangerous gases to safe
concentrations. A suitable gas-ballast control valve for introduction of purge gas into the pump is available
as an accessory (refer to Section 7.4.15).
Contact the Edwards Application team for further advice on dilution requirements if required.
3.3Unpack and inspect
Table 9 - Checklist of items
QuantityDescriptionCheck ()
1E1M18 or E2M18 rotary vacuum pump
(1)Fitting pack containing the following:
1NW25 centring-ring
1O-ring for centring-ring
1Receptacle connectors
*
Various sizes: supplied with single-phase motors except for pumps with Item Numbers A343-16-903,
A343-17-984, A363-17-984, A343-25-984 and A363-25-984.
*
1. Remove all packing materials and protective covers and check the pump. If the pump is damaged, notify th e
supplier and the carrier in writing within three days; state the Item Number of the pump together with the order
number and the supplier's invoice number. Retain all packing materials for inspection. Do not use the pump if it
is damaged.
2. Check that the package contains the items listed in Table 9. If any of these items are missing, notify the supplier
within 3 days.
If the pump is not to be used immediately, replace the protective covers. Store the pump in suitable conditions, as
described in Section 6.1.
3.4Locate the pump
Attach mechanical lifting equipment to the eye on the pump. Failure to do so can result in injury
to people or damage to the pump. Refer to Section 2.3 for the mass of the pump.
The pump can be either free-standing on its box section skids, or be fixed by bolts through the four fixing holes in
the box section skids, or be used with vibration isolators. For t he locations of the fixing holes in the box section skids,
refer to Figure 2.
Provide a firm, level platform for the pump. Locate the pump so that the oil-level sight-glass is visible and the oil
filler-plug, oil drain-plugs and gas -ballast control are accessible.
If the pump will be located inside an enclosure, make sure that there is adequate ventilation at both ends of the
pump, so that the ambient temperature around the pump does not exceed 40°C. There must be a minimum space of
25 mm between the pump and the enclosure walls.
Edwards and the Edwards logo are trademarks of Edwards Limited.
A343-10-880 Issue S
CAUTION
WARNING
3.5Fill the pump with oil
Fill the pump with oil as described below. Refer to Figure 1 for the item numbers in brackets.
1. Remove the oil filler-plug (1).
2. Pour oil into the pump until the oil-leve l reaches the MAX mark on the bezel at the top of the oil sight-glass (12).
If the oil-level goes above the MAX mark, remove the oil drain-plug (11) and drain the excess oil from the pump.
3. After a few minutes, recheck the oil-level. If the oil-level is now below the MAX mark, pour more oil into the
pump.
4. Refit the oil filler-plug. Tighten the plug firmly by hand. Do not overtighten.
3.6Electrical installation: single-phase motors
Ensure that the electrical installation of the pump conforms with the local and national safety
requirements. The pump must be connected to a suitably fused and protected electrical supply
with a suitable earth (ground) point, for recommended cord sets refer to Section 2.4.
Installation
Ensure that the motor is correctly configured for the electrical supply voltage and frequency. If not, damage to
the motor can occur.
3.6.1Introduction
Note:The pump will restart automatically when the electrical supply is restored after an interruption and when
the pump cools after it has overheated. If the pump is not required to restart automatically, use electrical
control equipment which must be reset manually.
Ensure that the electrical supply voltage corresponds with the voltage specified on the motor data pla te.
Table 5 gives the recommended fuse rating and Table 10 specifies the Figure which identifies the correct electrical
supply connection details.
The diameter of the outer sheath of the electrical supply cable must be within the range 7 to 10.5 mm. The cable
must conform in size and colour coding with the local and national electrical installation regulations. The
temperature rating of the cable must be 70°C or greater.
1. Remove the cover from the motor terminal box (Figure 1, item 8).
2. Remove the cable- gland from the inside of the terminal-box and fit the cable-gland to the cable leadthrough
hole in the side of the terminal-box. Using a tool this should be tightened to a torque of 3.75 Nm.
3. Pass the electri cal supply cable through the gland.
4. Remove the receptacle connectors from the fitting pack.
5. Fit a ring connector to the earth (ground) wire.
6. Select the correct size of receptacle connectors for the cable wires:
Use the red connectors for wire sizes 0.75 to 1.5 mm
Use the blue connectors for wire sizes 1. 5 to 2.5 mm
7. Attach the appropriate connector to the live and neutral wires.
8. Connect the electrical supply cable to the appropriate terminals as shown in Figure 3 to 5. Tighten the earth
(ground) terminal connection to a torque of 2.13 to 2.87 Nm.
9. Tighten the dome shaped nut on the cable-gland until the outer sheath of the cable is firmly gripped. Using a
tool this should be tightened to a torque of 2.5 Nm, do not overtighten.
2
.
2
.
10.Check that the cable connections are correct, then refit the cover on the motor terminal-box.
3.6.3Amphenol version single-phase motors
When selecting the mating half of the amphenol connector, please ensure that it is compatible.
The protective earth (ground) connection is required to make first and break last with respect to the other
connections.
Connect the electrical supply cable to the appropriate pins as shown in Figure 6.
3.6.4110/200-240 V 50 Hz or 115-120/200-230 V 60 Hz single-phase motors
The motor is designed for a single-phase electrical supply and is suitable for 50 Hz and 60 Hz operation. The motor
can be manually switched between nominal supply voltages of 110-120 V and 200-240 V.
Use the following procedure to check that the voltage selector switch is correctly positioned for the electrical supply
voltage:
1. Refer to Figure 7. Undo the three screws securing the terminal box cover (1) and remove the cover.
2. Check the position of the voltage selector switch (6). If the switch is in the correct position, continue at Step 3.
If the voltage selector switch (6) is in the wrong position, press the switch to select the alternative position.
3. Refit the terminal box cover and secure with the three screws.
When selecting the mating half of the connector, please ensure that it is compatible (IEC60320 C19 female).
A. Top view of motor
B. Internal view of top of motor
C. Voltage selector switch
1. Terminal box cover
2. Electrical inlet socket
3. On/Off switch
4. Position 'I' (on)
5. Position 'O' (off)
6. Voltage selector switch
7. Position 'II' (high voltage setting 200 – 240 V)
8. Position 'I' (low voltage setting 110 – 120 V)
Installation
Figure 7 - Motor voltage selection: single-phase motors, 110/200-240 V 50 Hz and 115-120/200-240 V 60 Hz
3.7Electrical installation: three-phase motors
3.7.1Connect the electrical supply to the motor
Ensure that the electrical installation of the pump conforms with the local and national safety
requirements. The pump must be connected to a suitably fused and protected electrical supply
with a suitable earth (ground) point.
Ensure that the motor is correctly configured for the electrical supply voltage and frequency. If not, damage to
the motor can occur.
Note:To prevent automatic restart of the pump-motor if the electrical supply is restored after an electrical
supply failure, connect the pump to the electrical supply through suitable control equipment which must
be reset manually after an electrical supply failure.
Edwards and the Edwards logo are trademarks of Edwards Limited.
A343-10-880 Issue S
CAUTION
Edwards recommend that you connect the electrical supply to the motor through a starter or circuit breaker which
has thermal over-current protection which can be adjusted to suit t he full load current ratings shown in Table 5. The
fuse ratings in Table 5 are provided for guidance only. The supplier of the thermal overcurrent protection device may
specify different values to ensure correct operation of the fuse and the overcurrent protection device. Ensure that
the fuse used is suitable for the starting currents given in Table 5. Refer to Figure 8 or 9 as appropriate for the correct
electrical supply connection diagram for the motor.
The diameter of the outer sheath of the electrical cable must be in the range 10-14 mm. The cable must conform in
size and colour coding with the local and national electrical installation regulations. The temperature rating of the
cable must be 70°C or greater.
1. Remove the cover from the motor terminal-box.
2. Remove the cable- gland from the inside of the terminal-box and fit the cable-gland to the cable leadthrough
hole in the side of the terminal-box. Using a tool this should be tightened to a torque of 3.75 Nm.
3. Refer to the electrical supp ly connection diagram for the pump, (Figure 8 or 9). Make sure that the metal links
are connected as shown; if they are not, move the links to the positions shown.
4. Pass the electrica l supply cable through the cable-gland and connect to the appropriate terminals, as shown in
Figure 8 and 9. Tighten the earth (ground) terminal connection to a torque of 2.13 to 2.87 Nm.
5. Tighte n the dome shaped nut on the cable gland until the outer sheath of the cable is firmly gripped. Using a
tool this should be tightened to a torque of 2.5 Nm, do not overtighten.
Installation
6. Ensure that the gasket seal for the terminal box cover is correctly positioned, then refit the cover to the
terminal-box and secure with the screws.
3.7.2Check the direction of rotation
Ensure that the motor rotates in the correct direction. If it does not, the pump and the vacuum system can
become pressurized when the pump is in operation.
1. With the pump inlet still unconnected, watch the motor cooling-fan through the motor fan-cover, switch-on the
electrical supply to the motor for a few seconds, then switch off the electrical supply again.
2. Check that the motor cooling-fan rotates in the correct direction shown by the arrow on the motor mounting
plate. If the direction of rotation is incorrect:
Switch off the electrical supply immediately.
Isolate the pump from the electrical supply.
Remove the terminal-box cover and swap wires L1 and L3: refer to Figure 8 and 9.
Refit the cover to the terminal-box.
3.8Connect the pump inlet
Take note of the following information when connecting the pump to the vacuum system. Refer to Section 7.4 for
details of the accessories mentioned below.
For optimum pumping speeds, ensure that the pipeline connected to the inlet-port is as short as possible and
has an internal diameter not less than the inlet-port diameter.
Support the vacuum pipelines to prevent loading of the coupling-joints.
If necessary, incorporate flexible bellows in the system pipeline s to redu ce the transmission of vibration and
to prevent loading of coupling-joints. If you using flexible bellows, ensure that bellows are used which have
a maximum pressure rating which is greater than the highest pressure that can be generated in the system. It
is recommended that Edwards flexible bellows are used.
Use a suitable inlet catchpot if pumping condensible vapours or if using the pump for very dusty applications.
Ensure that sealing surfaces are clean and scratch-free.
Connect the vacuum system to the inlet-port (Figure 1, item 6). To make this connection:
Connect to the NW25 flange supplied: use the NW25 centring-ring and O-ring supplied.
Remove the NW25 flange and connect to the 1 inch BSP threaded hole.
Use an NW25 to 28 mm bore tube adaptor available as an optional accessory, refer to Section 7.4.11.
Figure 8 - Electrical supply connection, three-phase motors: 200-220 V 50 Hz and 200-230 V 60 Hz
Edwards and the Edwards logo are trademarks of Edwards Limited.
A343-10-880 Issue S
WARNING
WARNING
Figure 9 - Electrical supply connection, three-phase motors: 380-415 V 50 Hz and 460 V 60 Hz
Installation
3.9Connect the pump outlet
Connect the pump outlet to a suitable treatment plant to prevent the discharge of dangerous gases
and vapours to the surrounding atmosphere.
Use a catchpot to prevent the drainage of contaminated condensate back into the pump.
The exhaust system must be configured so that the maximum pressure at the pump outlet does not exceed 0.5 bar
gauge (1.5 bar absolute, 1.5 x 10
Edwards recommend fitting an oil mist filter to the pump outlet in the following circumstances:
When using the pump with the gas-ballast control open.
When operating the pump with an inlet pressure greater than 10 mbar for extended periods.
When frequently pumping down from atmospheric pressure.
The mist filter will trap the oil exhausted from the pump: the oil can be reused if it is not contaminated.
To connect the pump to the outlet accessories or to the exhaust treatment plant:
5
Pa) at full pump throughput.
Connect 15 mm internal diameter vacuum or plastic hose to the outlet nozzle (Figure 1, item 2).
Remove the outlet nozzle and connect to the 3/4 inch BSP threaded hole.
Remove the outlet nozzle and replace it with an NW25 flange adaptor (available as an optional accessory,
refer to Section 7.4.13) and then connect to the NW25 flange.
The position of the gas-ballast inlet is shown in Figure 1 (item 4). The gas-ballast inlet has several filters (shown in
Figure 11) to trap any dust and debris if air is used as the gas-ballast supply.
If using a different gas for the gas-ballast supply or to connect a valve to the gas-ballast inlet:
1. Remove the filters (as described in Section 5.6).
2. Connect the gas supply or valve to the 1/4 inch BSP threaded hole.
3.11Leak test the system
Leak-test the system and seal any leaks found after installation of the pump to prevent leakage of substances out of
the system and leakage of air into the system.
Edwards and the Edwards logo are trademarks of Edwards Limited.
A343-10-880 Issue S
WARNING
WARNING
4Operation
Do not expose any part of the body to vacuum. Failure to obey this warning could result in injury.
Note:If the pump is PFPE-prepared, refer to Section 8.
4.1ATEX directive implications
4.1.1Introduction
This equipment is designed to meet the requirements of Group II Category 3 equipment in accordance with Directive
94/9/EC of the European Parliament and the Council of 23rd March 1994 on the approximation of the laws of the
Member States concerning equipment and protective systems intended for use in potentially explosive atmospheres.
(The ATEX Directive)
The ATEX Category 3 applies in respect of potential ignition sources internal to the equipment. An ATEX Category has
not been assigned in respect of potential ignition sources on the outside of the equipment as the equipment has not
been designed for use where there is an external potentially explosive atmosphere.
Operation
There is no potential source of ignition within the pump during normal operation but there may be potential sources
of ignition under conditions of predictable and rare malfunction as defined in the Directive. Accordingly, although
the pump is designed to pump flammable materials and mixtures, operating pr ocedures should ensure that under all
normal and reasonably predictable conditions, these materia ls and mixtures are not within explosive limits. Category
3 is considered appropriate for the avoidance of ignition in the case of a rare malfunction which allows flammable
materials or mixtures to pass through the pump whilst within their explosive limits.
4.1.2Flammable/pyrophoric materials
Obey the instructions and take note of the precautions given below to ensure that pumped gases
do not enter their flammable ranges.
When flammable or pyrophoric materials are present within the equipment:
Do not allow air to enter the equipment.
Ensure that the system is leak tight.
Use an inert gas purge (for example, a nitrogen purge) to dilute any flammable gases or vapours entering the
pump inlet, and/or use an inert gas purge to reduce the concentration of flammable gases or vapours in the
pump and in the exhaust pipeline to less than one quarter of the gases' published lower explosive limits
(LEL).
Use an inert gas purge into the pump gas ballast connection to prevent the condensation of flammable
vapours within the pump mechanism and exhaust pipeline.
If using inert gas purges to dilute dangerous gases to a safe level, ensure that the pump is shut
down if an inert gas supply fails.
Obey the instructions and take note of the precautions given below to ensure that pumped gases
do not enter their flammable ranges.
Switch on the inert gas purge to remove air from the pump and the exhaust pipeline before the process starts. Switch
off the purge flow at the end of the process only after an y remaining flammable gases or vapours have been purged
from the pump and exhaust pipeline.
If liquids that produce flammable vapours could be present in the pump foreline, then the inert gas purge to the
pump should be left on all the time this liquid is present. Flammable liquids could be present in the foreline as a
result of condensation or may be carried over from the process.
When calculating the flow rate of inert gas required for dilution, consider the maximum flow rate for the fla mmable
gases/vapours that could occur. For example, if a mass flow controller is used to supply flammable gases to the
process, assume a flow rate for flammable gases that could arise if the mass flow controller is fully open.
Continually measure the inert gas purge flow rate: if the flow rate falls below that required, stop the flow of
flammable gases or vapours into the pump.
Note:Edwards recommend obtaining and reading the Vacuum Pump and Vacuum System Safety manual
(publication number P400-40-100), available from Edwards or the supplier.
4.2Gas-ballast control
Use the gas-ballast control (Figure 1, item 3) to change the flow of gas-ballast into the low vacuum stage of the pump
and to provide an additional oil-feed to the pump mechanism.
Use the gas-ballast control closed:
To achieve ultimate vacuum.
To pump dry gases.
Turn the gas-ballast control by up to two turns anti-clockwise to introduce an additional oil-feed to the pump
mechanism. Edwards recommend doing this when pumping high throughputs of dry gases.
Turn the gas-ballast control from two to six turns anti-clockwise to increase the gas-ballast from zero flow. Use gasballast flow:
To pump high concentrations of condensable vapour.
To decontaminate the oil.
When operating the pump with the gas-ballast control open there will be an increased rate of oil loss from the pump.
Ideally, a mist filter and oil return kit should be used on clean applications.
Edwards and the Edwards logo are trademarks of Edwards Limited.
A343-10-880 Issue S
WARNING
4.3Start-up
Do not block the pump outlet or allow the outlet pressure to rise above 1.5 bar absolute. The oil
box may fracture and may cause injury to people nearby.
If the oil is contaminated, or if the pump tempera ture is below 13 °C, or if the electrical supply voltage is more than
10% below the lowest voltage specified for the pump, the pump may operate at a reduced s peed for a few minutes.
On single-phase pumps, if the pump continues to operate at reduced speed, the motor thermal overload device w ill
open and stop the pump. When the motor has cooled, the thermal overload device will reset automatically and the
pump will restart.
1. Check that the pump oil-level is between the MAX and MIN marks on the bezel of the oil-level sight-glass; if it is
not, refer to Section 5.3.
2. Turn the gas-ballast control to the required position (refer to Section 4.2).
3. Switch on the electrical supply to the pump.
4. Check that the oil-level in the sight-glass drops slightly (by 3 to 5 mm) after start-up. This shows that the pump
has primed with oil.
Operation
5. If the pump fails to prime, operate the pump with the inlet open to atmosphere for approximately 30 seconds.
Then isolate the inlet and check that the oil-level drops by 3 to 5 mm.
6. To achieve ultimate vacuum, pump condensab le vapours or decontaminate the pump oil, refer to th e procedures
in Section 4.4, 4.5 and 4.6 respectively. Otherwise, open the vacuum system isolation-valve.
4.4To achieve ultimate vacuum
If the pump does not achieve the performance specified in Section 2.2, make sure that this is not due to the system
design before you contact the supplier or Edwards for advice, In particular, the vapour pressure of all materials used
in the vacuum system (including pump oil, refer to below) must be much lower than the specified ultimate vacuum
of the pump.
Refer to Section 5.11.3 for a list of possible causes for failure to achieve the specified performance; note however
that the most common causes are:
Pressure measurement technique or gauge head is unsuitable or the gauge head is faulty.
Use of an oil other than the recommended oil and the vapour pressure of the oil is higher than the specified
ultimate vacuum of the pump.
Use the following procedure to achieve ultimate vacuum:
1. Isolate the pump from the vacuum system.
2. Turn the gas-ballast control fully anti-clockwise (fully open) and operate the pump for at least 1 hour (or
overnight) to thoroughly purge the oil of contaminants.
3. Close the gas-ballast control.
4. Open the vacuum system isolation-valve and pump down to ultimate vacuum.
Use gas-ballast (open the gas-ballast control) when there is a high proportion of condensable vapours in the process
gases:
1. Close the vacuum system isolation-valve.
2. Turn the gas-ballast control anti-clockwise to fully open and operate the pump for 30 minutes to warm the oil;
this will help to prevent vapour condensation in the pump.
3. Open the vacuum system isolation-valve and continue to operate the pump with the gas-ballast control open.
After pumping condensible vapours, decontaminate the oil (if necessary); use the procedure in Section 4.6.
4.6To decontaminate the oil
The oil in the pump should be clear. If the oil is cloudy or discoloured, it is contaminated with process vapours.
1. Look at the condition of the oil in the oil sight-glass (Figure 1, item 12). If the oil is cloudy or discoloured,
continue with the procedure at Step 2 below.
2. Close the vacuum system isolation-valve.
3. Turn the gas-ballast control fully anti-clockwise.
4. Operate the pump until the oil is clear.
4.7Unattended operation
The pump is designed for unattended operation under the normal operating conditions specified in Section 2.1.
However, Edwards recommend checking the pump at a regular interval of not more than 14 days; check the pump
more frequently if pumping high volumes of gas or if operating the pump with the gas-ballast control open.
Single-phase motors are cooled by internal fans. These motors have a thermal overload device. When the motor is
too hot, the thermal overload device switches off the pump. The thermal overload device has an automatic reset;
when the motor cools down, the device resets and the motor will restart.
When checking the pump, make sure that the pump is not going through a repetitive cycle of thermal overload
failures and automatic resets. If necessary, reduce the thermal load from the pumped gases to prevent overheating
of the pump.
4.8Shut-down
Note:If the gas-ballast control is open and the motor is switched off for any reason, the pump drive shaft may
rotate in the reverse direction, causing a system pressure rise. To prevent this, use a gas-ballast control
valve (refer to Section 7.4.15).
Edwards recommend, as described in the procedure below, decontaminating the oil before shutting down the pump;
this will prevent damage to the pump by the contaminates in the oil.
1. Refer to Section 4.6 and decontaminate the oil, as required.
2. Close the vacuum system isolation-valve (if not already closed).
3. Close gas-ballast (that is, turn the gas-ballast control clockwise).
Edwards and the Edwards logo are trademarks of Edwards Limited.
5Maintenance
CAUTION
WARNING
WARNING
5.1Safety information
Allow the pump to cool (so that it is at a safe temperature for skin contact) before starting
maintenance work. Make sure the pump is switched off in case the thermal overload device
restarts the pump.
Obey the safety instructions given below and take note of appropriate precautions. Failure to do
so can cause injury to people and damage to equipment.
Never use hydrocarbon lubricants in a PFPE-prepared pump.
A343-10-880 Issue S
Maintenance
Note:If the pump is PFPE-prepared, refer to Section 8.
A suitably trained and supervised technician must maintain the pump.
Ensure that the maintenance technician is familiar with the safety procedures which relate to the pump-oil
and the products processed by the pumping-system.
Isolate the pump and other components in the pumping system from the electrical supply so that they cannot
be operated accidentally.
After maintenance is completed, recheck the pump rotation direction if the electrical supply has been
disconnected.
Do not reuse O-rings and seals if they are damage d.
The pump and its oil will be contaminated with the process chemicals that have been pumped during
operation. Ensure that the pump is decontaminated before maintenance and that adequate precautions are
taken to protect people from the effects of dangerous substances if contamination has occurred.
Leak-test the system after maintenance work is complete if connecting or disconnecting any vacuum or
exhaust joints; seal any leaks found.
Do not touch or inhale the thermal breakdown products of fluorinated materials which may be present if the
pump has been heated to 260°C and above. These breakdown products are very dangerous. Fluorinated
materials in the pump may include oils, greases and seals. The pump may have overheated if it was misused,
if it malfunctioned or if it was in a fire. Edwards Material Safety Data sheets for fluorinated materials used in
the pump are available on request: contact the supplier or Edwards.
5.2Maintenance plan
The plan shown in Table 11 details the routine maintenance operations necessary to maintain the pump in normal
use. Instructions for each operation are given in the section shown.
More frequent maintenance may be required if the pump is used to proces s corrosive or abrasive gase s and vapours;
in these circumstances, Edwards recommend replacing the pump seals every year. If n ecessary, adjust the
maintenance plan according to experience.
When carrying out maintenance on the pump, use Edwards spares and maintenance kits; these contain all of the
components necessary to complete maintenance operations successfully. The Item Numbers of the spares and kits
are given in Section 7.3.
Examine the condition of any external accessories, filters or traps (if fitted) when carrying out maintenance on the
pump. Refer to the instructions supplied with these accessories for the necessary maintenance procedures.
Edwards and the Edwards logo are trademarks of Edwards Limited.
Table 11 - Maintenance plan
A343-10-880 Issue S
Maintenance
OperationFrequency
Check the oil levelAs required; at least monthly5.3
Replace the oilEvery 3000 hours of operation5.4
Inspect and clean the inlet filterEvery oil change5.5
Clean or replace the gas-ballast filterYearly5.6
Clean the motor fan-cover and enclosureYearly5.7
Clean and overhaul the pumpEvery 15000 hours of operation5.8
Replace the run capacitorEvery 4 years5.9
Test the motor conditionEvery 15000 hours of operation5.9
Fit new bladesEvery 30000 hours of operation5.10
Refer to
Section
5.3Check the oil-level
Notes: 1. If required, it is possible to check the oil-level while the pump is operating, however the pump must be
switched off. Isolate the pump and other compon ents in the pumping sy stem from the electrical s upply
before pouring oil into the pump.
2. Do not mix hydrocarbon lubricants with PFPE or vice versa. If the oil is mixed, drain and refill with clean
oil as described in Section 5.4.
Refer to Figure 1 for the items in brackets.
1. Check that the oil-level in the oil sight-glass (12) i s between the MAX and MIN le vel mark s on the bezel of the oi l
sight-glass.
2. If the oil-level is near to or below the MIN level mark, remove the oil filler-plug (1) and pour more oil into the
reservoir until the oil reaches the MAX level mark. If the oil-level goes above the MAX mark, remove the oil
drain-plug (11) and drain the excess oil from the pump. Refit the oil drain-plug. Refit the oil filler-plug.
3. If the oil is contaminated, drain and refill the pump with clean oil as described in Section 5.4.
5.4Replace the oil
Refer to Figure 1 for the items in brackets.
1. Operate the pump for approximately ten minutes to warm the oil, then switch off the pump. (This lowers the
viscosity of the oil and allows the oil to be drained from the pump more easily).
2. Isolate the pump from the electrical supply and disconnect it from the vacuum system.
3. Remove the oil filler-plug (1).
4. Place a suitable block under the pump-motor to tilt the pu mp an d place a suitable contai ner under the oi l drainplug (gravity drain) (11). Remove the oil drain-plug and allow the oil to drain into the container.
5. If the oil is dirty or contaminated:
Refit the oil drain-plug and pour clean oil into the pump.
Reconnect the pump to the electrical su pply and operate the pump for about 5 to 10 minutes.
Disconnect the pump from the electrical supply, remove the oil drain-plug and allow the oil to drain out of
the pump.
Repeat this step until the oil reservoir is clean.
6. Refit the oil drain- plug, remove the block and reconnect the pump to the vacuum system.
7. Fill a suitable container with clean oil and pour the oil into the filler hole until the oil-level reaches the MAX
level mark on the bezel of the oil sight-glass (12).
8. Allow a few minutes for the oil to drain into the pump. If necessary, add more oil. Refit the oil filler-plug.
5.5Inspect and clean the inlet-filter
Refer to Figure 10.
1. Unscrew the inlet adaptor (1) and remove the O-ring (2), circlip (3) and inlet-filter (4).
2. W ash the filter in a suitable cleaning solution. Allow the filter to dry.
3. Refit the inlet-filter (4), circlip (3), O-ring (2) and inlet adaptor (1).
Replace the run capacitor as described in the instructions supplied with the capacitor kit.
Test the earth (ground) continuity and the insulation resistance of the pump-motor, in accordance with local
regulations for periodic testing of electrical equipment. Edwards recommend that:
The earth (ground) continuity is less than 0.1 .
Insulation resistance is greater than 10 M.
If the motor fails these tests, the motor must be replaced.
5.10Fit new blades
Fit new blades to the pump as described in the instructions supplied with the blade kit (refer to Section 7.3).
5.11Fault finding
5.11.1Introduction
A list of fault conditions and their possible causes is provided in the following sections to assist in basic fault-fi nding.
If unable to rectify a fault when using this guide, call the supplier or nearest Edwards Service Centre for advice.
5.11.2The pump has failed to start
The electrical supply fuse has failed.
The motor is incorrectly wired.
The operating voltage does not match that of the motor.
The exhaust filter or exhaust line is blocked.
The oil temperature is below 13°C.
The oil is too viscous.
The oil is contaminated.
The pump is seized after long storage or has been left to stand after pumping contaminants.
The motor is faulty.
Edwards and the Edwards logo are trademarks of Edwards Limited.
A343-10-880 Issue S
5.11.3The pump fail to achieve specified performance (failure to reach ultimate
vacuum)
The measuring technique or gauge is unsuitable.
There is a leak in the external vacuum system.
The gas-ballast control is open.
The oil level is too low.
The pump has been filled with the wrong type of oil.
The oil is contaminated.
The pump has not primed.
The vacuum fittings are dirty.
The inlet-filter is blocked.
The pump has not warmed up.
The motor shaft rotates in the wrong direction.
Maintenance
5.11.4The pump is noisy
The motor fan-cover is damaged.
The coupling and fan assembly are poorly adjusted.
The motor coupling is loose.
The motor bearings are worn.
The oil is contaminated with solid particles.
One of the pump blades is sticking.
5.11.5The pump surface temperature is above 100°C
The ambient temperature is too high.
The cooling-air supply is insufficient or is too hot.
The cooling-air supply is blocked.
The electrical supply voltage is too high.
The exhaust filter or exhaust line is blocked.
The oil level is too low.
The pump is filled with the wrong type of oil.
The oil is contaminated.
The process gas is too hot or the throughput is too high.
5.11.6The vacuum is not maintained after the pump is switched off
The gas-ballast control is open.
O-ring(s) are damaged or missing.
The shaft seals have deteriorated.
5.11.7The pumping speed is poor
The connecting pipelines are too small in diameter.
The connecting pipelines are too long.
The inlet-filter is blocked.
5.11.8There is an external oil leak
The oil pump shaft seal is worn or damaged.
The oil box gasket has deteriorated.
There is an oil leak from gas-ballast control.
There is an oil-leak from the drain-plug.
Edwards and the Edwards logo are trademarks of Edwards Limited.
A343-10-880 Issue S
CAUTION
6Storage and disposal
6.1Storage
Observe the storage temperature limits stated in Section 2.1. Storage below these limits will permanently
damage the pump seals.
Note:If storing a new pump in conditions of high humidity, remove the pump from its cardboard packaging box;
dispose of the box (refer to Section 6.2).
Use the following procedure to store the pump:
1. Shut-down the pump as described in Section 4.8.
2. Disconnect the pump from the electrical supply.
3. Purge the vacuum system and the pump with dry nitrogen and disconnect the pump from the vacuum system.
4. Replace the oil as described in Section 5.4.
Storage and disposal
5. Place and secure protective covers over the inlet and outlet-ports.
6. Store the pump in cool, dry conditions until required for use.
When required, prepare and install the pump as described in Section 3. If the pump has been stored for more than a
year, clean and overhaul the pump before installing it as described in the instructions supplied with the clean and
overhaul kit.
6.2Disposal
Dispose of the pump, the oil and any components removed from the pump safely in accordance with all local and
national safety and environmental requirements.
Particular care must be taken with components and waste oil which have been contaminated with dangerous process
substances.
Do not incinerate fluoroelastomer seals and O-rings.
Edwards and the Edwards logo are trademarks of Edwards Limited.
A343-10-880 Issue S
7Service, spares and accessories
7.1Introduction
Edwards products, spares and accessories are available from Edwards companies in Belgium, Brazil, China, France,
Germany, Israel, Italy, Japan, Korea, Singapore, United Kingdom, U.S.A and a world-wide network of distributors.
The majority of these centres employ Service Engineers who have undergone comprehensive Edwards training
courses.
Order spare parts and accessories from the nearest Edwards company or distributor. When ordering, state for each
part required:
Model and Item Number of the equipment
Serial number
Item Number and description of part.
7.2Service
Edwards products are supported by a world-wide network of Edwards Service Centres. Each Service Centre offers a
wide range of options including: equipment decontamination; service exchange; repair; rebuild and testing to factory
specifications. Equipment which has been serviced, repaired or rebuilt is returned with a full warranty.
Service, spares and accessories
The local Service Centre can also provide Edwards engineers to support on-site maintenance, service or repair of the
equipment.
For more information about service options, contact the nearest Service Centre or other Edwards company.
7.3Spares
The maintenance kits listed in Table 12 contain all of the parts required to maintain the pump. The maintenance kits
also include instructions for the use of the kits.
Use the Clean and Overhaul kit for routine maintenance operations. Use the Blade kit together with the Clean and
Overhaul kit when renewing the blade assembly in the rotary pump.
Table 12 - Spares item numbers
SpareItem Number
Clean and Overhaul kit (Hydrocarbon filled)A363-01-131
Clean and Overhaul kit (PFPE filled)A363-01-136
E1M18 Blade kitA343-01-041
E2M18 Blade kitA363-01-020
Ultragrade 19 Oil (4 litres)H110-25-013
Fomblin Grade 06/6-500 mlH113-01-016
Capacitor kit for pump part number A343-15-903A505-91-813
Capacitor kit for pump part number A343-16-903A505-91-813
Capacitor kit for pump part number A363-15-903A505-91-813
Capacitor kit for pump part number A343-15-904A505-91-800
Capacitor kit for pump part number A363-15-904A505-91-800
Capacitor kit for pump part number A363-25-904A505-91-800
Capacitor kit for pump part number A343-15-912A505-91-811
Capacitor kit for pump part number A363-15-912A505-91-811
Capacitor kit for pump part number A363-25-912A505-91-811
Capacitor kit for pump part number A343-15-920A505-91-812
Capacitor kit for pump part number A363-15-920A505-91-812
Capacitor kit for pump part number A343-15-981A505-91-800
Capacitor kit for pump part number A363-15-981A505-91-800
Capacitor kit for pump part number A343-17-984A505-91-814
Capacitor kit for pump part number A343-25-984A505-91-814
Capacitor kit for pump part number A363-17-984A505-91-814
Capacitor kit for pump part number A363-25-984A505-91-814
7.4Accessories
7.4.1Introduction
A range of accessories is available for the E1M18 and E2M18 pumps, as shown in Figure 12.
The accessories are briefly described in the following sections. Refer to Table 13 for the Item Numbers of these
Edwards and the Edwards logo are trademarks of Edwards Limited.
A343-10-880 Issue S
7.4.3Inlet dust filter
The Inlet Dust Filter protects the pump against abrasive dust.
7.4.4Inlet desiccant trap
Use a Desiccant Trap when pumping limited quantities of water vapour at high pumping speeds to a low vapour
pressure.
7.4.5Inlet chemical trap
The Inlet Chemical Trap protects the pump against chemically active gases.
7.4.6Outlet mist filter
The Outlet Mist Filter separates and traps oil drop lets in the pump outlet to prevent oil mist discharge. The Mist Filter
is supplied with an adaptor to connect it to the pump outlet.
Edwards and the Edwards logo are trademarks of Edwards Limited.
A343-10-880 Issue S
7.4.7Solenoid operated pipeline valve
Fit the Pipeline Valve between the vacuum system and the pump inlet to provide additional system protection when
the pump is switched off.
7.4.8Foreline trap
Use a Foreline Trap on a clean pumping system to prevent back-migrati on of rotary pump oi l vapour into t he vacuum
system.
7.4.9Vibration isolators
The Vibration Isolators reduce transmission vibration and noise when the pump is floor or frame mounted and help
to reduce strain when the mounting area is uneven.
7.4.10Oil drain extension
Fit the Oil Drain Extension between the oil drain port on the pump and the oil drain-plug to make the drainage of oil
from the pump easier.
Service, spares and accessories
7.4.11Pump inlet adaptor
This is a NW25 to 28 mm bore tube adaptor, supplied with connection hardware. Attach this adaptor to the NW25
pump-inlet to connect 28 mm bore plastic or rubber tube.
7.4.12Flexible bellows
Use Flexible Bellows to connect the pump-inlet to the vacuum system. Edwards recommend the use of bellows when
using Vibration Isolators.
7.4.13Pump outlet adaptor
This is a 3/4 inch BSP to NW25 adaptor. Use this adaptor when removing the hose adaptor supplied fi tted to the pump
outlet to adapt the 3/4 inch BSP thread to NW25 pump inlet.
7.4.14Gas ballast banjo/elbow assembly
Use this 3/4 inch BSP assembly when a remote ballast supply or gas ballast valve is fitted.
7.4.15Gas ballast valve
Use this to control the gas-ballast remotely. The Gas Ballast Banjo/Elbow Assembly will also be required (refer to
Edwards and the Edwards logo are trademarks of Edwards Limited.
A343-10-880 Issue S
CAUTION
WARNING
WARNING
8PFPE-prepared EM pumps
8.1Summary
If a PFPE-prepared EM pump has been ordered, the pump will be supplied prepared for use with Edwards PFPE
mechanical pump oils, such as Fomblin YVAC 06/6 and Krytox 1506.
PFPE-prepared EM pumps are suitable for pumping high concentrations of oxygen.
Edwards recommend referring to publication P400-40-100 (Vacuum pump and Vacuum System Safety) before
installing and using a PFPE-prepared EM pump.
8.2Installation
Never use hydrocarbon lubricants in a PFPE-prepared pump.
Installation of a PFPE-prepared EM pump is as specified in Section 3, but take note of the caution above.
PFPE-prepared EM pumps
8.3Operation
PFPE-prepared EM pumps are suitable for pumping high concentrations of oxygen, but Edwards
recommend that a PFPE-prepared EM pump is not used for the pumping of hazardous materials.
Operation of a PFPE-prepared EM pump is as specified in Section 4, but take note of the warning above.
8.4Maintenance
Obey the safety instructions given below and take note of appropriate precautions. Failure to do
so can cause injury to people.
Take additional care if it is suspected that the pump (and hence the PFPE oil) has overheated.
Do not touch or inhale the thermal breakdown products of PFPE oil, which may be present, if the pump has
been heated to 260°C and above. PFPE oils are safe in normal use but can decompose into very dangerous
substances if they are heated to 260°C and above. The pump may have overheated if it was misused, if it
malfunctioned or if it was in a fire. Safety Data Sheets for PFPE oils used in the pump are available on
request: contact the supplier or Edwards.
Fomblin oil has different properties from other pump oils, therefore:
When filling the PFPE-prepared EM pump with Fomblin oil, Edwards rec omm end regular checks for oil leaks,
particularly around the shaft seals.
If an oil leak is detected, contact the supplier or Edwards for advice.