to which this declaration relates is in conformity with the following standard(s) or other
normative document(s)
EN61010-1: 2010 Safety Requirements for Electrical Equipment for Measurement,
Control and Laboratory Use. General Requirements
EN61326-2-3: 2013 Electrical equipment for measurement, control and laboratory
(Class B Emissions, Use. EMC requirements. Particular requirements. Test
Industrial Immunity) configuration, operational conditions and performance criteria
for transducers with integrated or remote signal conditioning
EN50581:2012 Technical Documentation for the Assessment of Electrical and
Electronic Products with respect to the Restriction of Hazardous
Substances
and fulfils all the relevant provisions of
2014/30/EU Electromagnetic Compatibility (EMC) Directive
2011/65/EU Restriction of Certain Hazardous Substances (RoHS) Directive
Note: This declaration covers all product serial numbers from the date this Declaration was
signed onwards.
09.07.2015, Eastbourne
Larry Marini, Senior Technical Manager Date and Place
This product has been manufactured under a quality management system certified to ISO 9001:2008
Fomblin® is a registered trademark of Ausimont SpA.
LET-LOK® is a registered trademark of HAM-LET Group.
Swagelok™ is trademark of Crawford Fitting Company.
Viton™ is trademark of DuPont Dow Elastomers L.L.C.
Edwards and the Edwards logo are trademarks of Edwards Limited.
D048-50-880 Issue E
CAUTION
WARNING
1INTRODUCTION
1.1Scope
This manual provides installation, operation and maintenance instructions for the Edwards Active Ion Gauge (AIGX).
You must use the AIGX as specified in the manual. Read this manual before installing and operating the AIGX.
Important safety information is highlighted as WARNING and/or CAUTION instructions which must be followed. The
use of WARNINGS and/or CAUTIONS is defined below.
Warnings are given where failure to observe the instruction could result in injury or death to
people.
Cautions are given where failure to observe the instruction could result in damage to the equipment, associated
equipment and process.
INTRODUCTION
The units used throughout this manual conform to SI international system of measurement followed by imperial units
in parenthesis; SI (imperial).
In accordance with the recommendations of IEC1010, the following warning symbols are on the pump:
Warning - refer to accompanying
documentation.
From August 2005, Edwards will offer
European customers a recycling service.
1.2Description
The AIGX is an ion gauge head and gauge controller in a single compact unit. The gauge operates as a hot filament
ionisation gauge, in which the pressure is measured indirectly as a function of the current.
-10
The measurement range of the AIGX is 5 x 10
vacuum connections are available: DN16CF, DN40CF and NW25.
Th e A I G X r e q u i r e s a 1 4 . 5 - 30 V d.c. power supply; it has a 0 to 10 V d.c . an a l o g u e o u t p u t w h i c h is r elated to pressure.
The AIGX is compatible with Edwards Controllers. Alternatively, you can use an independent power supply and you
can read the output signal with a voltmeter or analogue to digital converter.
An 8-way electrical connector socket on the AIGX (Figure 1, item 1) is used to connect the AIGX to your Controller or
electrical supply and voltmeter. Electrical cables fitted with suitable connectors are available as accessories. A
gauge identification signal is available on the electrical connector: this signal is used by Edwards Controllers to
identify which type of active gauge is connected.
to 5 x 10
-2
torr (6.6 x 10
-10
to 6.6 x 10-2 mbar). Three types of
The AIGX is supplied with two filaments. A switch on top of the gauge selects between filament 1 and 2 (Figure 1,
item 2).
There are two levels of emission currents: 0.1 and 1 mA. The gauge will automatically select 1 mA emission current
at low pressure, and 0.1 mA at high pressure.
-2
A protection circuit is built into the gauge to prevent operating the gauge at pressures higher than 5 x 10
The AIGX can be degassed using the degas feature. During degas, the grid voltage is 500 V and the emission current
is 10 mA. The degas is inhibited at high pressure. The maximum duration of degas is three minutes, after which the
gauge returns automatically to its normal mode.
The AIGX has a set-point facility. The set-point signal is an open collector transistor output, which is on when the
pressure measured by the gauge is below a preset pressure. The pressure at which the set-point operates can be
adjusted using the set-trip switch on the top of the gauge (1, item 3).
The AIGX is fitted with a status LED (Figure 1, item 4) . This LED indicates the operating status of the gauge.
Figure 1 - General View of the AIGX (NW25 version)