to which this declaration relates is in conformity with the following standard(s) or other
normative document(s)
EN61010-1: 2010 Safety Requirements for Electrical Equipment for Measurement,
Control and Laboratory Use. General Requirements
EN61326-2-3: 2013 Electrical equipment for measurement, control and laboratory
(Class B Emissions, Use. EMC requirements. Particular requirements. Test
Industrial Immunity) configuration, operational conditions and performance criteria for
transducers with integrated or remote signal conditioning
EN50581:2012 Technical Documentation for the Assessment of Electrical and
Electronic Products with respect to the Restriction of Hazardous
Substances
and fulfils all the relevant provisions of
2014/30/EU Electromagnetic Compatibility (EMC) Directive
2011/65/EU Restriction of Certain Hazardous Substances (RoHS) Directive
Note: This declaration covers all product serial numbers from the date this Declaration was
signed onwards.
09.07.2015, Eastbourne
Larry Marini, Senio
Technical Manager Date and Place
This product has been manufactured under a quality management system certified to ISO 9001:2008
Edwards and the Edwards logo are trademarks of Edwards Limited.
D048-60-880 Issue E
CAUTION
WARNING
1Introduction
1.1Scope and definitions
This manual provides installation, operation and maintenance instructions for the Edwards Active Ion Gauge (AIGX).
You must use the AIGX as specified in the manual.
Read this manual before you install and operate the AIGX. Important safety information is highlighted as WARNING
and CAUTION instructions; you must obey these instruction s. The us e of WARNINGS a nd CA UTIONS is def ined below .
Warnings are given where failure to observe the instruction could result in injury or death to
people.
Cautions are given where failure to observe the instruction could result in damage to the equipment, associated
equipment and process
Introduction
The following IEC warning labels appear on the pump:
Warning - refer to accompanying documentation.
Warning - Edwards offer European customers a recycling service.
The units used throughout this manual conform to the SI international system of units of measurement.
1.2Description
The AIGX is an ion gauge head and gauge controller i n a single compact unit. The gauge operates as a hot filament
ionisation gauge, in which the pressure is measured indirectly as a function of the current.
The measurement range of the AIGX is 5 x 10
connections are available: DN16CF, DN40CF and NW25.
The AIGX requires a 14.5 - 30 V d.c. power supply; it has a 0 to 10 V d.c. analogue output which is related to pressure.
A 9-way electrical connector socket on the AIGX (Figure 1, item 1) is used to connect the AIGX to your electrical
supply and voltmeter.
-10
to 5 x 10-2 Torr (6.6 x 10
-10
to 6.6 x 10-2 mbar). Three types of vacuum
The AIGX is supplied with two filaments. A switch on top of the gauge selects between filament 1 and 2 (Figure 1,
item 2).
There are two levels of emission currents: 0.1 and 1 mA. The 0.1 mA emission current should be used at pressures
higher than 10
A protection circuit is built into the gauge to prevent operating the gauge at pressures higher than 5 x 10
Torr and the 1mA emission current should be used at pressures lower than 10-5 Torr.
-2
Torr.
D048-60-880 Issue E
3
1
2
4
ACTIVE ION GAUGE
AIGX-D-1.33CF
P NO. D04861000
EDWARDS LIMITED
Manor Royal, Crawley, West Sussex
RH10 2LW, ENGLAND
Introduction
The AIGX can be degassed using the degas feature. During degas, the grid voltage is 500 V and the emission current
is 10 mA. The degas is inhibited at high pressure. The maximum duration of degas is three minutes, after which the
gauge returns automati cally to its normal mode.
The AIGX is fitted with a status LED (Figure 1, item 3) . This LED indicates the operating status of the gauge.