EDAN SE-601A, SE-601B, SE-601C User Manual

I
About this Manual
P/N: 01.54.455420-10
Release Date: November 2011
© Copyright EDAN INSTRUMENTS, INC. 2011. All rights reserved.
Statement
EDAN owns the copyrights of this manual. Without prior written consent of EDAN, any materials contained in this manual shall not be photocopied, reproduced or translated into other languages.
Materials protected by the copyright law, including but not limited to confidential information such as technical information and patent information are contained in this manual, the user shall not disclose such information to any irrelevant third party.
The user shall understand that nothing in this manual grants him, expressly or implicitly, any right or license to use any of the intellectual properties of EDAN.
EDAN holds the rights to modify, update, and ultimately explain this manual.
Product Information
Product Name: Electrocardiograph Model: SE-601A, SE-601B, SE-601C
Responsibility of the Manufacturer
EDAN only considers itself responsible for any effect on safety, reliability and performance of the equipment if:
Assembly operations, extensions, re-adjustments, modifications or repairs are carried out by persons authorized by EDAN, and
The electrical installation of the relevant room complies with national standards, and
The instrument is used in accordance with the instructions for use.
Upon request, EDAN may provide, with compensation, necessary circuit diagrams, and other information to help qualified technician to maintain and repair some parts, which EDAN may define as user serviceable.
II
Terms Used in this Manual
This guide is designed to give key concepts on safety precautions.
WARNING
A WARNING label advises against certain actions or situations that could result in personal
injury or death.
CAUTION
A CAUTION label advises against actions or situations that could damage equipment, produce
inaccurate data, or invalidate a procedure.
NOTE A NOTE provides useful information regarding a function or a procedure.
III
Table of Contents
Chapter 1 Safety Guidance ...........................................................................................................1
1.1 Intended Use ........................................................................................................................... 1
1.2 Warnings and Cautions ........................................................................................................... 1
1.2.1 Safety Warnings ...............................................................................................................1
1.2.2 Li-ion Battery Care Warnings ..........................................................................................4
1.2.3 General Cautions.............................................................................................................. 5
1.3 List of Symbols....................................................................................................................... 6
Chapter 2 Introduction..................................................................................................................9
2.1 Top Panel .............................................................................................................................. 10
2.2 Keyboard and Keys .............................................................................................................. 11
2.3 Rear Panel............................................................................................................................. 12
2.4 Right Panel ...........................................................................................................................13
2.5 Bottom Panel ........................................................................................................................16
2.6 Function Features .................................................................................................................17
Chapter 3 About SE-601 Application.........................................................................................18
3.1 Selecting Menu Functions .................................................................................................... 18
3.2 Entering Data........................................................................................................................ 19
3.3 Selecting an Option from a List............................................................................................ 20
3.4 About the main Screen.......................................................................................................... 20
3.5 About the System Setup Screen............................................................................................ 24
3.6 About the File Manage Screen..............................................................................................26
Chapter 4 Operation Preparations.............................................................................................29
4.1 Power and Earthing ..............................................................................................................29
4.2 Loading/Replacing Recorder Paper...................................................................................... 30
4.3 Preparing the Patient............................................................................................................. 33
4.3.1 Instructing the Patient ....................................................................................................33
4.3.2 Preparing the Skin.......................................................................................................... 33
4.4 Connecting the Patient Cable to the Electrocardiograph and Electrodes ............................. 33
4.5 Attaching Electrodes to the Patient.......................................................................................34
4.5.1 Reusable Electrodes ....................................................................................................... 35
4.5.2 Disposable Electrodes ....................................................................................................37
4.6 Inspection before Power-On................................................................................................. 38
Chapter 5 Switching On the Electrocardiograph......................................................................40
Chapter 6 Entering Patient Information...................................................................................41
6.1 Entering Patient ID ............................................................................................................... 41
6.2 Entering Other Information .................................................................................................. 42
Chapter 7 Printing ECG Reports ...............................................................................................43
7.1 Auto Mode ............................................................................................................................ 43
7.2 Manual Mode........................................................................................................................ 45
7.3 Rhythm Mode ....................................................................................................................... 47
7.4 R-R Mode ............................................................................................................................. 49
7.5 Transmitting ECG Data to the PC ........................................................................................50
7.5.1 Transmitting ECG Data through the Serial Port ............................................................ 50
7.5.2 Transmitting ECG Data through the Net Port ................................................................ 51
IV
7.6 Copy Printing........................................................................................................................ 51
7.7 ECG Reports......................................................................................................................... 52
7.7.1 ECG Reports in the Auto Mode..................................................................................... 52
7.7.2 ECG Reports in the Rhythm Mode ................................................................................ 54
7.7.3 ECG Reports in the Manual Mode................................................................................. 55
7.7.4 ECG Reports in the R-R Mode ......................................................................................56
7.7.5 ECG Reports Printed by the USB Printer ...................................................................... 58
Chapter 8 Managing Files...........................................................................................................60
8.1 Storage Upgrade Function .................................................................................................... 62
8.2 Transmitting Files to the PC ................................................................................................. 62
8.2.1 Transmitting Files through the Serial Port ..................................................................... 63
8.2.2 Transmitting Files through the Net Port......................................................................... 64
8.3 Copying Files between SE-601 and the U Disk ................................................................... 66
8.4 Editing Patient Information .................................................................................................. 68
8.5 Deleting Files........................................................................................................................ 69
8.6 Previewing a File (Only for SE-601B/C) ............................................................................. 70
8.7 File Printing .......................................................................................................................... 72
Chapter 9 System Setup...............................................................................................................73
9.1 Work Mode Setup ................................................................................................................. 73
9.1.1 Specifying Work Mode ..................................................................................................74
9.1.2 Specifying Manual Style................................................................................................ 75
9.1.3 Specifying Rhythm Style ...............................................................................................75
9.1.4 Specifying Auto Style .................................................................................................... 75
9.1.5 Specifying Sampling Mode............................................................................................ 76
9.1.6 Specifying Recording Sequence ....................................................................................77
9.1.7 Inputting Sampling Time................................................................................................77
9.2 Filter Setup ...........................................................................................................................78
9.2.1 Setting AC Filter ............................................................................................................ 78
9.2.2 Setting EMG Filter......................................................................................................... 78
9.2.3 Setting DFT Filter .......................................................................................................... 79
9.2.4 Setting Lowpass Filter.................................................................................................... 79
9.3 Record Setup......................................................................................................................... 79
9.3.1 Specifying Recording Device ........................................................................................80
9.3.2 Setting Paper Marker......................................................................................................81
9.3.3 Selecting Patient Information......................................................................................... 83
9.3.4 Setting Speed.................................................................................................................. 83
9.3.5 Setting Gain.................................................................................................................... 83
9.3.6 Selecting Template ......................................................................................................... 84
9.3.7 Selecting Measure .......................................................................................................... 84
9.3.8 Selecting Minnesota Code.............................................................................................. 84
9.3.9 Selecting Analysis.......................................................................................................... 84
9.3.10 Selecting Position Marker............................................................................................ 85
9.4 Lead Setup ............................................................................................................................ 85
9.4.1 Setting Rhythm Lead1/2/3 ............................................................................................. 86
9.4.2 Setting Lead Sequence ................................................................................................... 86
9.5 Transmission Setup............................................................................................................... 87
9.6 Display & Sound Setup ........................................................................................................88
9.6.1 Setting Brightness (Only for SE-601B) ......................................................................... 89
9.6.2 Selecting Display Colors (Only for SE-601C)............................................................... 89
V
9.6.3 Selecting Antialising (Only for SE-601B/C)..................................................................89
9.6.4 Setting Key Volume .......................................................................................................89
9.6.5 Setting Hint Volume ....................................................................................................... 90
9.6.6 Setting QRS Volume ......................................................................................................90
9.6.7 Setting Notification Volume........................................................................................... 90
9.7 Patient Question.................................................................................................................... 90
9.7.1 Specifying ID Mode....................................................................................................... 91
9.7.2 Selecting ID Hint............................................................................................................ 91
9.7.3 Selecting Gender/Height/Weight/BP/Race/Medication/Ward NO/Doctor/Technician.. 92
9.7.4 Specifying H/W Unit...................................................................................................... 92
9.7.5 Specifying BP Unit.........................................................................................................92
9.7.6 Selecting Next Patient.................................................................................................... 93
9.7.7 Setting Prompt................................................................................................................ 93
9.7.8 Inputting Extra Question................................................................................................ 94
9.8 Date & Time Setup ...............................................................................................................94
9.8.1 Setting Current Date/Current Time ................................................................................ 95
9.8.2 Setting Date Mode.......................................................................................................... 95
9.8.3 Setting Time Mode......................................................................................................... 95
9.8.4 Setting Period Interval and Period Duration .................................................................. 95
9.8.5 Setting Power-Off Time ................................................................................................. 96
9.8.6 Setting LCD Off Time.................................................................................................... 96
9.9 More Setup ...........................................................................................................................96
9.9.1 Choosing a Language..................................................................................................... 97
9.9.2 Setting Pacemaker Detection Sensitivity ....................................................................... 97
9.9.3 Setting Save Option........................................................................................................97
9.9.4 Entering Institution......................................................................................................... 98
9.9.5 Restoring Default Settings .............................................................................................98
9.9.6 Setting Extern Input/Extern Output ...............................................................................99
Chapter 10 Switching Off the Electrocardiograph.................................................................100
Chapter 11 Hint Information....................................................................................................101
Chapter 12 Troubleshooting......................................................................................................102
Chapter 13 Cleaning, Care and Maintenance.........................................................................106
13.1 Cleaning............................................................................................................................ 106
13.1.1 Cleaning the Main Unit and the Patient Cable........................................................... 106
13.1.2 Cleaning the Reusable Electrodes.............................................................................. 106
13.1.3 Cleaning the Print Head ............................................................................................. 106
13.2 Disinfection ......................................................................................................................107
13.3 Care and Maintenance ...................................................................................................... 107
13.3.1 Recharge and Replacement of Battery....................................................................... 107
13.3.2 Recorder Paper........................................................................................................... 109
13.3.3 Visual inspection ........................................................................................................109
13.3.4 Maintenance of Main Unit, Patient Cable and Electrodes ......................................... 109
Chapter 14 Accessories ..............................................................................................................112
Chapter 15 Warranty & Service............................................................................................... 114
15.1 Warranty............................................................................................................................ 114
15.2 Contact information.......................................................................................................... 114
Appendix 1 Technical Specifications ........................................................................................ 115
VI
A1.1 Safety Specifications ....................................................................................................... 115
A1.2 Environment Specifications............................................................................................. 116
A1.3 Physical Specifications .................................................................................................... 116
A1.4 Power Supply Specifications ........................................................................................... 116
A1.5 Performance Specifications ............................................................................................. 117
Appendix 2 EMC Information..................................................................................................119
Appendix 3 Abbreviation...........................................................................................................124
SE-601 Series Electrocardiograph User Manual Safety Guidance
- 1 -
Chapter 1 Safety Guidance
This chapter provides important safety information related to the use of SE-601 series electrocardiograph (hereinafter called SE-601).
1.1 Intended Use
The intended use of SE-601 is to acquire ECG signals from adult and pediatric patients through
body surface ECG electrodes. The electrocardiograph is only intended to be used in hospitals or
healthcare facilities by doctors and trained healthcare professionals. The cardiogram recorded by
the electrocardiograph can help users to analyze and diagnose heart disease. However, the ECG
with measurements and interpretive statements is offered to clinicians on an advisory basis only.
WARNING
1. This equipment is not designed for internal use or direct cardiac application.
2. This equipment is not intended for home use.
3. This equipment is not intended for treatment or monitoring.
4. This equipment is intended for use on adult and pediatric patients only.
5. The results given by the equipment should be examined based on the overall clinical condition of the patient, and they can not substitute for regular checking.
1.2 Warnings and Cautions
In order to use the electrocardiograph safely and effectively, and avoid possible dangers caused by improper operation, please read through the user manual and be sure to be familiar with all functions of the equipment and proper operation procedures before use.
Please pay more attention to the following warning and caution information.
1.2.1 Safety Warnings
WARNING
1. The electrocardiograph is intended to be used by qualified physicians or personnel
professionally trained. They should be familiar with the contents of this user manual before operation.
2. Only qualified service engineers can install this equipment, and only service engineers
authorized by the manufacturer can open the shell.
SE-601 Series Electrocardiograph User Manual Safety Guidance
- 2 -
WARNING
3. EXPLOSION HAZARD - Do not use the electrocardiograph in the presence of
flammable anesthetic mixtures with oxygen or other flammable agents.
4. SHOCK HAZARD - The power receptacle must be a hospital grade grounded outlet.
Never try to adapt the three-prong plug to fit a two-slot outlet.
5. Make sure that the power is turned off and the power cord is disconnected from the AC
socket before connecting or disconnecting equipment. Otherwise, electrical shock or other injuries may happen to the patient or operator.
6. If the integrity of the external protective conductor is in doubt, the equipment should be
powered by an internal rechargeable battery.
7. Do not use this equipment in the presence of high static electricity or high voltage
equipment which may generate sparks.
8. Only the patient cable and other accessories supplied by the manufacturer can be
used. Or else, the performance and electric shock protection can not be guaranteed.
9. Make sure that all electrodes are connected to the patient correctly before operation.
10. Ensure that the conductive parts of electrodes and associated connectors, including
neutral electrodes, do not come in contact with earth or any other conducting objects.
11. If reusable electrodes with electrode gel are used during defibrillation,
electrocardiograph recovery will take more than 10 seconds. The manufacturer recommends the use of disposable electrodes at all times.
12. Electrodes of dissimilar metals should not be used; otherwise it may cause a high
polarization voltage.
13. The disposable electrodes can only be used for one time.
14. The electrocardiograph has been safety tested with the recommended accessories,
peripherals, and leads, and no hazard is found when the electrocardiograph is operated with cardiac pacemakers or other stimulators.
15. Do not touch the patient, bed, table or the equipment while using the ECG together
with a defibrillator.
16. Do not touch accessible parts of non-medical electrical equipment and the patient
simultaneously.
17. The use of equipment that applies high frequency voltages to the patient (including
electrosurgical equipment and some respiration transducers) is not supported and may produce undesired results. Disconnect the patient cable from the electrocardiograph, or detach the leads from the patient prior to performing any procedure that uses high frequency surgical equipment.
SE-601 Series Electrocardiograph User Manual Safety Guidance
- 3 -
WARNING
18. If the wireless AP technology is used, the minimum distance between the radiator and
the human body should be not less than 20cm. Use the supplied antenna only. There should be no shield in or around the room where the wireless AP is used.
19. Fix attention on the examination to avoid missing important ECG waves.
20. SHOCK HAZARD - Don’t connect non-medical electrical equipment, which has been
supplied as a part of the system, directly to the wall outlet when the non-medical equipment is intended to be supplied by a multiple portable socket-outlet with an isolation transformer.
21. SHOCK HAZARD - Don’t connect electrical equipment, which has not been supplied
as a part of the system, to the multiple portable socket-outlets supplying the system.
22. Do not connect any equipment or accessories that are not approved by the
manufacturer or that are not IEC/EN 60601-1-1 approved to the electrocardiograph. The operation or use of non-approved equipment or accessories with the electrocardiograph is not tested or supported, and electrocardiograph operation and safety are not guaranteed.
23. Any non-medical equipment (such as the external printer) is not allowed to be used
within the patient vicinity (1.5m/6ft.).
24. Do not exceed the maximum permitted load when using multiple portable
socket-outlets to supply the system.
25. Multiple portable socket-outlets shall not be placed on the floor.
26. Do not use the additional multiple portable socket-outlet or extension cord in
the medical electrical system, unless it’s specified as part of the system by manufacturer. And the multiple portable socket-outlets provided with the system shall only be used for supplying power to equipment which is intended to form part of the system.
27. Accessory equipment connected to the analog and digital interfaces must be certified
according to the respective IEC/EN standards (e.g. IEC/EN 60950 for data processing equipment and IEC/EN 60601-1 for medical equipment). Furthermore all configurations shall comply with the valid version of the standard IEC/EN 60601-1-1. Therefore anybody, who connects additional equipment to the signal input or output connector to configure a medical system, must make sure that it complies with the requirements of the valid version of the system standard IEC/EN 60601-1-1. If in doubt, consult our technical service department or your local distributor.
SE-601 Series Electrocardiograph User Manual Safety Guidance
- 4 -
WARNING
28. Parts and accessories used must meet the requirements of the applicable IEC/EN 601
series safety standards, and/or the system configuration must meet the requirements of the IEC/EN 60601-1-1 medical electrical systems standard.
29. Connecting any accessory (such as external printer) or other device (such as the
computer) to this electrocardiograph makes a medical system. In that case, additional safety measures should be taken during installation of the system, and the system shall provide:
a) Within the patient environment, a level of safety comparable to that provided by
medical electrical equipment complying with IEC/EN 60601-1, and
b) Outside the patient environment, the level of safety appropriate for non-medical
electrical equipment complying with other IEC or ISO safety standards
30. If multiple instruments are connected to a patient, the sum of the leakage currents
may exceed the limits given in the IEC/EN 60601-1 and may pose a safety hazard. Consult your service personnel.
31. The potential equalization bar can be connected to that of other equipment when
necessary. Make sure that all these devices are connected to the potential equalization terminal.
1.2.2 Li-ion Battery Care Warnings
WARNING
1. Improper operation may cause the internal li-ion battery (hereinafter called battery) to
be hot, ignited or exploded, and it may lead to the decrease of the battery capacity. It is necessary to read the user manual carefully and pay more attention to warning messages.
2. Only qualified service engineers authorized by the manufacturer can open the battery
compartment and replace the battery, and batteries of the same model and specification should be used.
3. DANGER OF EXPLOSION -- Do not reverse the anode and the cathode when
installing the battery.
4. Do not heat or splash the battery or throw it into fire or water.
5. Do not destroy the battery; do not pierce battery with a sharp object such as a needle;
do not hit with a hammer, step on or throw or drop to cause strong shock; do not disassemble or modify the battery.
SE-601 Series Electrocardiograph User Manual Safety Guidance
- 5 -
WARNING
6. When leakage or foul smell is found, stop using the battery immediately. If your skin or
cloth comes into contact with the leakage liquid, cleanse it with clean water at once. If the leakage liquid splashes into your eyes, do not wipe them. Irrigate them with clean water first and go to see a doctor immediately.
7. Properly dispose of or recycle the depleted battery according to local regulations .
8. Only when the device is off can the battery be installed or removed.
9. Remove the battery from the electrocardiograph when the electrocardiograph is not
used for a long time.
10. If the battery is stored alone and not used for a long time, we recommend that the
battery should be charged at least once every 6 months to prevent overdischarge.
1.2.3 General Cautions
CAUTION
1. Visually examine the package prior to unpacking. If any signs of mishandling or
damage are detected, contact the carrier to claim for damage.
2. Avoid liquid splash and excessive temperature. The temperature must be kept
between 5 ºC and 40 ºC during operation, and it should be kept between -20 ºC and 55 ºC during transportation and storage.
3. Do not use the equipment in a dusty environment with bad ventilation or in the
presence of corrosive.
4. Make sure that there is no intense electromagnetic interference source around the
equipment, such as radio transmitters or mobile phones etc. Attention: large medical electrical equipment such as electrosurgical equipment, radiological equipment and magnetic resonance imaging equipment etc. is likely to bring electromagnetic interference.
5. Ruptured fuse must only be replaced with that of the same type and rating as the
original.
6. The device and accessories are to be disposed of according to local regulations after
their useful lives. Alternatively, they can be returned to the dealer or the manufacturer for recycling or proper disposal. Batteries are hazardous waste. Do NOT dispose of them together with house-hold garbage. At the end of their lives hand the batteries over to the applicable collection points for the recycling of waste batteries. For more detailed information about recycling of this product or battery, please contact your local Civic Office, or the shop where you purchased the product.
7. Federal (U.S.) law restricts this device to sale by or on the order of a physician.
SE-601 Series Electrocardiograph User Manual Safety Guidance
- 6 -
1.3 List of Symbols
External output
External input
Equipment or part of CF type with defibrillator proof
Caution
Consult Instructions for Use
Potential equalization
Mains Supply
Battery indicator
Battery recharging indicator
Delete key
Enter key
Esc key
Space key/Feed paper key
Shift key
SE-601 Series Electrocardiograph User Manual Safety Guidance
- 7 -
Fn key
Power On/Off key
MODE key
RESET key
1mV/COPY key
START/STOP key
Tab key
Recycle
Part Number
Serial Number
Manufacturer
Date of Manufacture
The symbol indicates that the device complies with the European Council Directive 93/42/EEC concerning medical devices.
SE-601 Series Electrocardiograph User Manual Safety Guidance
- 8 -
Authorized Representative in the European Community
With respect to electrical shock, fire and mechanical hazards only in accordance with UL 60601-1and CAN/CSA C22.2 No. 601.1
It indicates that the device should be sent to the special agencies according to local regulations for separate collection after its useful life
Rx only (U.S.)
Federal (U.S.) Law restricts this device to sale by or on the order of a physician
SE-601 Series Electrocardiograph User Manual Introduction
- 9 -
Chapter 2 Introduction
The 6-channel electrocardiograph gathers ECG signals of 12 leads simultaneously. It displays the
operation menu, ECG parameters as well as electrocardiograms.
The 6-channel ECG waves can be viewed on the LCD screen and printed out by using a
high-quality thermal recorder.
The manual, auto, rhythm or R-R mode can be chosen freely.
SE-601 can be powered by the mains supply or a battery.
With a high resolution thermal recorder, a 32-bit processor and a large-capacity memorizer, the
6-channel electrocardiograph has advanced performance and high reliability. The compact size
makes it suitable for clinic and hospital uses.
The 6-channel electrocardiograph has three models: SE-601A, SE-601B and SE-601C.
SE-601A adopts 192×64 dot single color LCD screen; SE-601B adopts 320×240 dot single color
LCD screen; SE-601C adopts 640×480 dot multicolor TFT LCD screen.
Configuration: main unit, power cord, patient cable, chest electrodes, limb electrodes,
disposable electrodes, alligator clips, thermal recorder paper, fuses and battery.
NOTE: The pictures and windows in this manual are for reference only.
SE-601 Series Electrocardiograph User Manual Introduction
- 10 -
2.1 Top Panel
Figure 2-1 SE-601A
Figure 2-2 SE-601B / SE-601C
Indicator
Symbol Name Explanation
A
Mains supply indicator
When the device is powered by the mains supply, this indicator is lit.
B
Battery indicator
When the device is powered by the battery, this indicator is lit.
C
Battery recharging indicator
When the battery is being recharged, this indicator is lit.
LCD Screen
Recorder
Keyboard
Indicators From left to right on the top panel: A, B, C.
LCD Screen
Recorder
Keyboard
Indicators From left to right on the top panel: A, B, C.
SE-601 Series Electrocardiograph User Manual Introduction
- 11 -
2.2 Keyboard and Keys
Figure 2-3 SE-601A / SE-601B / SE-601C Keyboard
Name Explanation
A Function Key
Selecting menu functions on the screen (Pressing F1, F2, F3 or F4
can move the cursor)
B Indicator
1 indicates that the device is powered by the mains supply
2 indicates that the device is powered by the battery
3 indicates the battery charging status
C
Power on / off
D Esc Canceling operation
E Tab
Pressing Tab can move the cursor forward, and pressing Shift + Tab can move the cursor backward.
In the MANU mode, press Tab to switch the lead group.
F
Space: adding a space between typed characters
Feed Paper: before printing, if Paper Marker is set to Style1 or Style2, you can press the Space key to advance the recorder paper to the next black marker; if Paper Marker is set to No, you can press the Space key to advance the paper for about 2cm. Press the Space key again to stop advancing the paper.
G MODE
Press this key to select a working mode among AUTO, MANUAL and RHYTHM.
NOTE: Only in the Work Mode Setup window can the R-R
mode be selected.
H RESET Draw the baseline to zero quickly in the case of baseline drift
I 1mV/COPY
In the AUTO mode, pressing the 1mV/Copy key can print the ECG
report which was printed out last time.
A
N
M
G
H
I
L
J
E
D
F
C
K
B (1, 2, 3)
SE-601 Series Electrocardiograph User Manual Introduction
- 12 -
In the MANU mode, pressing the 1mV/Copy key can insert a 1mV
calibration mark in the printing course.
J START/STOP Start/Stop printing reports
K Fn Inputting special characters. Press Fn + a to type è. L Shift Inputting a capital letter. Press Shift + p to type a capital P.
M Enter Confirming operation
N Delete Deleting characters
2.3 Rear Panel
Figure 2-4 SE-601A / SE-601B / SE-601C Rear Panel
Name Explanation
A
Potential Equalization Conductor
Potential equalization conductor provides a connection between the unit and the potential equalization bus bar of the electrical installation.
B Handle Part for people to hold
C
Mains Supply Socket
AC SOURCE: alternating current supply socket
A
C
B
SE-601 Series Electrocardiograph User Manual Introduction
- 13 -
2.4 Right Panel
Figure 2-5 SE-601A / SE-601B / SE-601C Right Panel
Name Explanation
A Patient Cable Socket Connecting to the patient cable
B Serial Port Connecting to a PC
C USB Socket 1 (Optional) Standard USB socket, connecting to a PC
D
USB Socket 2 (Optional)
Standard USB socket, connecting to a U disk or a USB printer recommended by the manufacturer
E Net port Standard net port, connecting to a PC
F External Input / Output Socket Connecting to the external signal device
1) Patient Cable Socket
: Applied part of type CF with defibrillator proof
: Caution
Definitions of corresponding pins:
Pin Signal Pin Signal Pin Signal
1
C2 / V2
6
SH
11
F / LL
2
C3 / V3
7
NC
12
C1 / V1
or NC
D
A
F
C
B
E
SE-601 Series Electrocardiograph User Manual Introduction
- 14 -
3
C4 / V4
8
NC
13
C1 / V1
4
C5 / V5
9
R / RA
14
RF (N) /RL
or NC
5
C6 / V6
10
L / LA
15
RF (N) / RL
NOTE: The left side of “/” is European standard; and the right side is American standard.
2) Serial Port
WARNING
The isolated intensity of serial port is 1500V AC and the maximum voltage applied should not exceed +15V DC.
Definitions of corresponding pins:
Pin Signal Pin Signal Pin Signal
1
NC
4
NC
7
NC
2
RxD (input)
5
GND
8
NC
3
TxD (output)
6
NC
9
NC
3) External Input / Output Socket
Definitions of corresponding pins:
Pin Signal Pin Signal
1
GND
4
GND
2
GND
5
ECG Signal (input)
3
GND
6
ECG Signal (output)
SE-601 Series Electrocardiograph User Manual Introduction
- 15 -
4) USB Socket 1 / USB Socket 2 (Optional)
WARNING
Only the USB equipment recommended by the manufacturer can be connected to the USB interface.
Definitions of corresponding pins:
Pin Signal Pin Signal
1
VBUS
3
D+
2
D-
4
GND
WARNING
1. Accessory equipment connected to the analog and digital interfaces must be certified
according to the respective IEC/EN standards (e.g. IEC/EN 60950 for data processing equipment and IEC/EN 60601-1 for medical equipment). Furthermore all configuration shall comply with the valid version of the standard IEC/EN 60601-1-1. Therefore anybody, who connects additional equipment to the signal input or output connector to configure a medical system, must make sure that it complies with the requirements of the valid version of the system standard IEC/EN 60601-1-1. If in doubt, consult our technical service department or your local distributor.
2. If multiple instruments are connected to a patient, the sum of the leakage currents may
exceed the limits given in the IEC/EN 60601-1 and may pose a safety hazard. Consult your service personnel.
SE-601 Series Electrocardiograph User Manual Introduction
- 16 -
2.5 Bottom Panel
Figure 2-6 SE-601 A / SE-601B / SE-601C Bottom Panel
Name Explanation
A Speaker Hole Path for sound from speaker (Only for SE-601B/C)
B Battery Compartment Compartment for battery
C Heat Emission Hole Path for internal heat emission
D Label Position for product information label
E Fuse The fuse specification is T1AL250VP Ø5×20.
1) Battery Compartment
The rated voltage and the rated capacity of the battery are as follows:
Rated Voltage: 14.8V; Rated Capacity: 2200mAh.
WARNING
1. Improper operation may cause the battery to be hot, ignited or exploded, and it may
lead to the decrease of the battery capacity. Therefore, it is necessary to read the user manual carefully and pay more attention to warning messages.
2. When leakage or foul smell is found, stop using the battery immediately. If your skin or
cloth comes into contact with the leakage liquid, cleanse it with clean water at once. If the leakage liquid splashes into your eyes, do not wipe them. Irrigate them with clean water first and go to see a doctor immediately.
A
B
C
D
C
E
SE-601 Series Electrocardiograph User Manual Introduction
- 17 -
WARNING
3. Only qualified service engineers authorized by the manufacturer can open the battery
compartment and replace the battery, and batteries of the same model and specification must be used.
4. Only when the device is off can the battery be installed or removed.
NOTE: If the battery has not been used for two months or more, you should recharge it
before using it again.
2) Fuse
There is a fuse installed on the bottom of the main unit. The specification is T1AL250VP Ø5×20.
WARNING
Ruptured fuse must only be replaced with that of the same type and rating as the original.
2.6 Function Features
Low weight and compact size Multi-language support ECG signals of 12 leads are gathered and amplified simultaneously, 6-channel waves
are displayed and recorded simultaneously
High resolution thermal recorder, recording frequency response ≤150Hz Flexible printing formats Full alphanumeric keyboard The auto, manual, rhythm, and R-R modes are optional Convenient operation of system setup and file management Measurement function and interpretation function are optional Hint information of lead off, lack of paper and low battery capacity etc. Battery with large capacity Automatic baseline adjustment for optimal printing Supporting real-time sampling, pre-sampling, period sampling and arrhythmia
triggering sampling
LCD backlight and the device can be turned off automatically according to the set time The patient information to be printed is optional ECG data can be transmitted to the PC software through the serial cable, net cable, or
wireless AP (optional).
SE-601 Series Electrocardiograph User Manual About SE-601 Application
- 18 -
Chapter 3 About SE-601 Application
The following sections provide an overview of the main operations and functions in the SE-601
application.
3.1 Selecting Menu Functions
To select Patient, press the function key F1 below Patient. To select , press the function key F5 below .
SE-601 Series Electrocardiograph User Manual About SE-601 Application
- 19 -
3.2 Entering Data
1. When the main screen1 is displayed, press the function key F1 below Patient to open the
Patient Information window.
2. Press Tab or Shift + Tab to move the cursor to the Name textbox.
3. Press on the keyboard to erase the typed information.
4. Press the letter or number keys on the keyboard to input patient name. Press Fn and a letter
key to input the special character in the top right corner of the key. Press Shift and a number key to input the special character in the top right corner of the key. Press Shift and a letter key to input a capital letter. For example, press Fn + a to input è, press Shift + 3 to input #, and press Shift + a to input a capital A.
5. Press Enter to confirm, or press Tab or Shift + Tab to move the cursor to the OK button, and
then press Enter to confirm.
6. Press Esc to cancel the operation, or press Tab or Shift + Tab to move the cursor to the
Cancel button, and then press Enter to cancel the operation.
SE-601 Series Electrocardiograph User Manual About SE-601 Application
- 20 -
3.3 Selecting an Option from a List
For SE-601B/C
1. Press Tab or Shift + Tab to move the cursor to the Work Mode item.
2. Press F1 or F2 to select MANU, AUTO, RHYT or R-R.
3. Press Enter to confirm, or press Tab or Shift + Tab to move the cursor to the OK button, and
then press Enter to confirm.
4. Press Esc to cancel the operation, or press Tab or Shift + Tab to move the cursor to the
Cancel button, and then press Enter to cancel the operation.
For SE-601A
1. Press F1, F2, Tab or Shift + Tab to move the cursor to the Work Mode item.
2. Press F3 or F4 to select MANU, AUTO, RHYT or R-R.
3. After setup, press Enter to confirm, or press Esc to display a hint to prompt you to save these
modifications.
3.4 About the main Screen
After you turn on the device, the main screen1 pops up. Press the function key F5 below
to open the main screen2.
SE-601 Series Electrocardiograph User Manual About SE-601 Application
- 21 -
Figure 3-1 SE-601B / SE-601C Main Screen1
Figure 3-2 SE-601B / SE-601C Main Screen2
Figure 3-3 SE-601A Main Screen1
L
K
J
I
H
C
D
M
C
G
B
A
E
F
P
O
N
E
G
D
F B
M
L
K
H
J
I
SE-601 Series Electrocardiograph User Manual About SE-601 Application
- 22 -
Figure 3-4 SE-601A Main Screen2
Name Explanation
A Name
Patient Name: within 12 characters (for SE-601B/C)
within 10 characters (for SE-601A)
B ID
When ID Mode is set to Auto or Time, the patient ID is within
10 characters.
When ID Mode is set to Manu, the patient ID is within 18
characters.
C Hint Information
Including Lead Off, No Paper, Paper Error, Battery Weak,
Modu Error, Demo, Sampling, Analyzing, Recording, Learning, Transmitting, Transmit Fail, Detecting, Memory Full, Overload, U Disk, USB Printer, Testing
D Work Mode MANU, AUTO, RHYT or R-R
E Heart Rate Actual Heart Rate
F Current Time Refer to Section 9.8, “Date & Time Setup”
G Battery Symbol Identify the current capacity of the battery
H
Press the function key F5 below to open the main
screen2
I 100Hz
EMG Filter: 25Hz, 35Hz or 45Hz
Lowpass Filter: 75Hz, 100Hz or 150Hz
J 25mm/s
Paper Speed: In the MANU mode, you can set Speed to 5mm/s, 6.25mm/s, 10mm/s, 12.5mm/s, 25mm/s or 50mm/s. In the AUTO and RHYT modes, only 25mm/s and 50mm/s are available. In the R-R mode, only 25mm/s is available.
K 10mm/mV
Gain: 10 mm/mV, 20 mm/mV, 10/5 mm/mV, AGC, 2.5 mm/mV or 5 mm/mV
P
O
N
SE-601 Series Electrocardiograph User Manual About SE-601 Application
- 23 -
L Patient
Press the function key F1 below Patient to display the Patient Information window
M ECG waveform Display ECG waveform
N
Press the function key F5 below to return to the main
screen1
O File
Press the function key F2 below File to open the File Manage screen. For details, please refer to Chapter 8, “Managing Files”.
P Setup
Press the function key F1 below Setup to display the System Setup screen. For details, please refer to Chapter 9, “System
Setup”.
To Open the main Screen3
If Next Patient is set to On, in the AUTO mode, press the START/STOP key to print ECG
reports, and the system will automatically open the main screen3 after a complete ECG report is printed.
Figure 3-5 SE-601B / SE-601C Main Screen3
Figure 3-6 SE-601A Main Screen3
On the main screen3, pressing the function key F1 below Same can return to the main screen1, and all the patient information will keep the same; pressing the function key F2 below Next can return to the main screen1, all the patient information will be cleared, and patient ID will be
refreshed.
Loading...
+ 102 hidden pages