Purchasing products from Electro-Chemical Devices, Inc. provides you with the finest liquid analytical
instrumentation available. If this is your first purchase from ECD, please read this manual before
installing and commissioning your new equipment.
If there are any questions concerning this equipment, please contact your local ECD representative, or
the factory directly at:
PREFACE ....................................................................................................................................................... iii
TABLE OF CONTENTS .................................................................................................................................... iv
WARRANTY ................................................................................................................................................. vii
IMPORTANT SERVICE INFORMATION ......................................................................................................... vii
1.0 GENERAL DESCRIPTION ........................................................................................................................... 1
1.0.3 Temperature Compensation ........................................................................................................ 1
1.0.4 Cleaning and Maintenance .......................................................................................................... 1
1.1 FEATURES ............................................................................................................................................ 2
1.1.1 Features and Benefits .................................................................................................................. 2
1.3.1 Parts List: ...................................................................................................................................... 4
4.0 START UP ............................................................................................................................................... 18
4.1 SET-UP AND CONFIGURATION .......................................................................................................... 18
4.1.1 Set Up Menus, Standard ............................................................................................................ 18
4.1.2 Set Up Menu, MASTER/SLAVE ................................................................................................... 19
4.1.3 Configure and Trim Menu, Standard ......................................................................................... 20
4.1.4 Configure and Trim Menu, MASTER/SLAVE ............................................................................... 21
8.3 SPARE PARTS ..................................................................................................................................... 33
Page vi HYDRA NH
+
-N
4
WARRANTY
Electro-Chemical Devices, Inc. (ECD) warrants all products it manufactures to be free from defect in materials and factory
workmanship, and agrees to repair or replace any product that fails to perform, as specified, within one (1) year after date of
shipment. This warranty shall not apply to any product that has been:
1. Subjected to misuse, negligence or accident;
2. Connected, installed, adjusted or otherwise used not in accordance with the instructions furnished by ECD;
3. Repaired, modified or altered by persons not authorized by ECD, resulting in injury to the performance, stability or
reliability of the product.
This warranty is in lieu of any other warranty, expressed or implied. ECD reserves the right to make changes in the design or
construction of its products at any time, without prior notification, and without incurring any obligation to make any changes in
previously delivered products.
Seller’s sole liabilities and the buyer’s sole remedies under this agreement shall be limited to a refund in the purchase price, or
at ECD’s discretion, to the repair or replacement of any product that proves, upon ECD’s examination, to be defective, when
returned to the factory, transportation prepaid by the buyer, within one (1) year of the product’s original shipment date. Seller
shall not be liable for damages consequential or incidental to defects in any product, for failure of delivery in whole or in part,
for injuries resulting from its use, or for any other cause.
This warranty and the writing attached constitute the full understanding of seller and the buyer, and no terms, conditions,
understanding, or agreement purporting to modify or vary the terms hereof shall be binding unless hereafter made in writing
and signed by an authorized official of Electro-Chemical Devices, Inc.
This warranty does not cover pH, ORP or Specific Ion measurement, reference or combination electrodes or electrode
cartridges that have been commissioned in service.
IMPORTANT SERVICE INFORMATION
Use only factory authorized components for repair. Tampering or unauthorized substitution of components may
adversely affect the operation of this product and may void the warranty.
If service or repair is required, please obtain the serial number(s) or sales order number of the product(s) in
question and contact ECD’s Service Department at:
+1-800-729-1333 (USA/Canada) or +1-949-336-6060
or email Service@ecdi.com
A Return Material Authorization (RMA) number must be obtained from the service department before returning
any material to ECD. All material returned to ECD shall be shipped prepaid to the factory.
HYDRA NH
+
-N Page vii
4
UNPACKING THE INSTRUMENT
Your Electro-Chemical Devices instrument has been carefully packaged to protect it from damage during
shipment and dry storage. Upon receipt please follow the procedure outlined below.
1. Before unpacking, inspect the condition of the shipping container to verify proper handling by
the carrier. If damage is noted, save the shipping container as proof of mishandling for the
carrier.
2. Check the contents of the shipping container with the items and quantities shown on the
packing list. Immediately report any discrepancies to ECD.
3. Save the original packing material until you are satisfied with the contents. In the event the
product(s) must be returned to ECD, the packing material will allow you to properly ship it to
ECD.
4. Familiarize yourself with the instrument before installation, and follow proper installation and
wiring procedures.
Page viii HYDRA NH
+
-N
4
1.0 GENERAL DESCRIPTION
1.0.0 Description
The HYDRA Ammonium Analyzer measures the concentration of dissolved
ammonium as nitrogen (NH
determine the NH
+
-N concentration, an Ammonium Ion Electrode, a Potassium Ion
4
Electrode and a pH electrode. It is designed for use in all kinds of water. Typical
applications include monitoring environmental waters, lakes, streams and wells as
well as wastewater treatment in aeration basins and effluent.
1.0.1 Potassium Ion Compensation
The Ammonium Ion Electrode provides the primary measurement. Any potassium
ion in the sample, due to its similar size and charge to the ammonium ion, causes a
positive interference in the measurement. A Potassium Ion Electrode measures the
amount of potassium ion present in the sample and HYDRA C22 Analyzer subtracts
the appropriate amount of signal from the Ammonium Measurement.
+
-N) in water. The sensor uses three electrodes to
4
1.0.2 pH Compensation
The Ammonium Ion Electrode only measures the ammonium ion (NH
+
) not
4
ammonia (NH3). Ammonium ion and ammonia coexist in a pH dependent ratio in
solution. At pH 9.2 the ratio is 1:1, at pH 7 nearly all of the ammonia is in the
ammonium state and at pH 11 it is nearly all ammonia which is invisible to the
sensor. The more acidic values favor the NH
+
and the more basic values favor
4
ammonia gas, NH3. The pH Electrode measures the pH and the HYDRA C22 Analyzer
calculates the total NH
+
-N concentration based on the pH vs. concentration profile.
4
1.0.3 Temperature Compensation
Temperature is also measured and used to compensate each of the three electrode measurements.
While the pH electrode response is well defined with respect to temperature the ion electrodes tend to
be less well behaved. For the best results
calibrate the sensors near the process
temperature.
1.0.4 Cleaning and Maintenance
The HYDRA Ammonium Analyzer is configured to
NH3/NH
% concentration vs pH
100
80
+
4
periodically actuate a cleaning cycle using the
integral spray cleaner in the sensor. The period
and duration of the cleaning cycle is user
configurable. The 4-20 mA output is held at
either the last value or a preset value during the
cleaning. This minimizes the formation biofilms
60
%
40
20
0
681012
pH
% NH4+
% NH3
HYDRA NH
+
-N Page 1
4
or other coatings on the electrodes keeping maintenance to a minimum.
The rugged HYDRA sensor has a 1 ¼” NPT rear facing thread for attaching an extension/immersion tube
for easy installation from catwalks or handrails. The HYDRA sensor is submersible with an IP68 degree of
ingress protection. A removable electrode guard facilitates easy electrode replacement when necessary.
The HYDRA sensor features internal signal conditioning that allows the sensor to be mounted up to 200
meters from the analyzer.
1.1 FEATURES
1.1.1 Features and Benefits
• Separate, economical, easily replaceable Ammonium, Potassium and pH electrodes
• Fast and Accurate Ammonium Measurement NH4+ or NH4+-N
• pH and temperature compensated NH4+-N measurement
• Automatic compensation for K+ interference
• Rugged PVC design with removable electrode guard for easy maintenance
• Integral Spray Head Cleaner
Cleans sensor in situ with pressurized air
• Internal Signal Conditioning allows up to 200 meters between Sensor and Analyzer
1.2 Specifications
1.2.1 Hydra Sensor
• Three Electrode System with spray cleaner
• Ammonium ISE (NH4+- N) is the primary measurement
• Potassium ISE and pH glass electrodes are used for compensation of the NH4+ signal.
• The Sensor is waterproof with an ingress rating of IP 68.
Measurement Range
• NH4+- N: 0.1 to 1000 ppm
• K+: 0.1 to 10,000 ppm
• pH: 4-12 pH
Operating Temperature
• 0° C to 40° C (32° F to 104° F)
Min/Max Flow Rate by the sensor
• Minimum 0.1 m/s
• Maximum 3.0 m/s
Wetted Materials
• PVC, PES, PVDF, PTFE, Viton, Glass, 316 SS
Page 2 HYDRA NH
+
-N
4
Accuracy
• ± 3% of reading, dependent on Calibra0on
Response Time
• T90 approximately 1 minute
Electrode Life
• ISEs: 4- 6 months, typical
• pH electrode: 6-12 months, typical
1.2.2 C22 Analyzer
Measurements
• Ammonium: 0.1 ppb to 1000 ppm as NH4+- N
• Potassium: 0.1 ppb to 1000 ppm
• pH: 0 to 14 pH
• Temperature: 0° C to 100° C (32° F to 212° F)
Compensation
• pH 4 - 12 pH
• Potassium: 0.1 to 1000 ppm
• 0° - 50°C
Display
• 2.5” X 1.75” backlit LCD
• 4 lines of Text
• Graphical display, NH4+-N vs. Time
Enclosure
• NEMA 4X, LxWxD: 5.7” x 5.7” x 7”
HYDRA NH
+
-N Page 3
4
Relay
Standard Configuration
Master
Configuration
Slave Configuration
1 Spray Cleaner
(NO)
PV 1 Alarm
(NO)
NH4-N Alarm
(NO)
2 NH4-N Alarm
(NO)
Digital Input to Slave
(NO)
pH Alarm
(NO)
3
Spray Cleaner
(NO)
4
Diverter Valve
(NO)
Outputs
• (2) 4-20 mA maximum load 800 ohms @ 24 VDC
• mA #1: 0.1 to 50 mg/l NH4+- N (User Configurable)
• mA #2: 2 - 12 pH (User Configurable)
• Optionally up to (4) 4-20 mA outputs*
Input Power
• 110/220 VAC @ 50/60 Hz
Alarm Relay Ratings
• SPDT 230 VAC/5A or 30 VDC/5A resistive max. (NO) Normally Open
1.3 Hydra Dimensional Drawing
Figure 1.3
1.3.1 Parts List:
Page 4 HYDRA NH
o HYDRA Sensor with Electrodes, Calibration Cap,
o 30’ of ¼” air tubing, Electrode Removal Tool
+
-N
4
2.0 INSTALLATION
2.1 UNPACKING
Carefully remove the HYDRA sensor from its shipping container. Inspect the sensor for damage. Verify
the electrodes, Ammonium, Potassium and pH and the Spray Cleaner Nozzle are installed in the sensor
housing. The electrodes should be hand tightened into place so that the sealing o-rings are not visible.
The electrodes are supplied with protective caps that must be removed before start up. Do not remove
the caps until ready to use.
### NEVER SUPPORT THE HYDRA SENSOR BY THE CABLE, IRREPAIRABLE DAMAGE WILL OCCUR ###
2.2 ASSEMBLY
The HYDRA Sensor is shipped completely assembled, before use it must be connected to the analyzer
and an air supply. Connect the spray cleaner feed tube to a compressed air supply controlled by a
solenoid valve (not supplied). Attach an immersion/support tube (not supplied). Finally connect the
sensor wires and 110 VAC power to the analyzer as shown on the wiring diagram. Connect the Cleaner
Relay to the solenoid and connect the 4-20 mA Output(s) and Alarm Relay to the Control System, (PLC or
DCS). The HYDRA is then ready to use.
2.2.1 Spray Cleaner
The Spray Cleaner uses compressed air to create turbulence around the electrodes which removes dirt
and films from the measurement surfaces. Depending on the process being measured the cleaner
should be actuated for 15-30 seconds every 0.25 -2 hour period. The spray cleaner connection is a ¼”
compression fitting and requires between 25-75 psi air pressure. The air supply is controlled with a user
supplied solenoid valve through Relay 1 in the C-22 analyzer. The solenoid valve should be a 110 AC
powered, <10 watt device. Wire the solenoid to Relay 1 as described in Section 2.4.4 below.
If no air supply is available use the ECD Model AC10 Air Blast Spray Cleaner compressor. The AC10 uses
redundant intake air filters and redundant fuses on both the relays and the compressor. A highly reliable
high current contactor assures years of trouble free service. The 115 VAC 3.0 Amp or 220 VAC 1.3 Amp
Air Compressor is housed in a rugged, corrosion resistant, hot compression molded, fiberglass
reinforced, polyester enclosure with a stainless steel piano hinge to secure the cover to the base. The
AC10 can be wall mounted or use the optional 2” handrail mounting system that uses stainless steel
support rails with two 2” galvanized pipe clamps.
### NEVER SUPPORT THE HYDRA SENSOR BY THE CABLE, IRREPAIRABLE DAMAGE WILL OCCUR ###
2.2.2 Immersion /Support Tube
Feed the cable and compressed air line through the immersion/support tube (not supplied). Connect a
support tube to the 1 ¼” MNPT thread at the rear of the sensor. The Hydra sensor weighs approximately
5.3 lbs. The support tube must be able to support the weight of the HYDRA sensor in the user’s
installation, whether vertical or angled. The recommended material for the immersion tube is 1.5”
Schedule 80 PVC pipe with a reducer fitting to 1 ¼” FNPT.
HYDRA NH
+
-N Page 5
4
2.3 MOUNTING
2.3.1 C-22 Analyzer
Mount the C-22 in a location where there is easy access to the analyzer and sensors. Install the system
in an area where vibrations, electromagnetic and radio frequency interference are minimized or absent.
Do not mount in direct sunlight or areas of extreme heat (temperature > 120°F). The NEMA 4X C-22 is
suitable for outdoor use but it is best to mount it with a protective cover or sunshield.
There are three basic installation methods for the C-22, Wall Mount (UM), Panel Mount (PM) and Pipe
Mount (HM). The Pipe Mounting hardware includes 2” “U” bolts for attaching to standard 2” handrails.
(See the installation drawings in Section 9.3 at the rear of the manual)
2.3.2 Hydra Sensor
Install the sensor where the measured sample is representative of the entire process. Although the
sensor will function in a quiescent sample, flow improves the measurement. The recommended
minimum flow is 0.1 m/sec.
Securely mount the HYDRA sensor with the measuring end at least 6”away from the tank wall and
bottom. Ensure that the sensor is immersed at least 6” at all times.
Use care when servicing the sensor to ensure that the sensor does not hit the tank wall or bottom which
could break the sensing electrodes.
2.4 WIRING
Electrical wiring should only be conducted by qualified personnel. See the wiring color code for the
Hydra Sensor and the C-22 wiring diagram in Section 7.3 below.
2.4.1 Wiring, power
Attach power cable as shown in the diagram in Section 7.3 or the inside of the C22 cover. Feed the cable
through the gland fitting on the right hand side of the C-22. Tighten the cable gland to provide a good
seal to the cable. The instrument can be powered up at this point with no harm to the analyzer but it is
best to wait until the sensors are installed.
2.4.2 Wiring, Sensor
Attach the sensor wires as described on the diagram inside the C22 cover. Feed the sensor cable
through the gland fitting on the left hand side of the C-22. Do not use the same gland fitting for the AC
power or Alarm/Relays. The green terminal strip connectors are detachable from the circuit boards.
Remove the connector by pulling straight back from the circuit board.
2.4.3 Wiring, 4-20 mA Outputs
The 4-20 mA output is an unpowered output, 24V can be supplied from an external source, the PLC or
DCS receiving the signal or from the internal PS2 power supply.
If the internal PS2 Power Supply is used, first connect a jumper wire between Terminal #1 (RTN) on the
PS2 Power Supply Board and Terminal #2 of the 4-20 mA 1 output connector, the two slot connector on
Page 6 HYDRA NH
+
4
-N
the top of the Channel 2 input card. (See diagram in Section 7.3) Then connect a shielded 22 gauge
twisted pair communication cable to Terminal #2 (+24V) on the PS2 power supply board and Terminal
#1 of the 4-20 mA 1 connector.
If a second mA output is supplied connect it in the same manner, jumper Terminal #1 of the PS2 board
to Terminal #2 of the 4-20 mA 2 output terminal strip then connect the two wire communication cable
to PS2 Terminal #2 and mA 2 Terminal #1.
If the 4-20 mA loop is powered from an external source then simply connect a shielded 22 gauge twisted
pair communication wire to Terminals #1 and #2 of the 4-20 mA terminal on the Channel 2 input card.
The standard Hydra C-22 is configured with the NH4+-N on mA 1 and pH on mA 2.
2.4.4 Wiring, Contact Relay Outputs
The standard configuration has two SPDT 230V 5 A relays that can be wired either normally open (NO)
or normally closed (NC). The default configuration is set to use the relays as normally open. Relay 1
actuates the solenoid for the air blast spray cleaner and Relay 2 is an NH
+
-N alarm relay, either a High or
4
Low Alarm.
Relay 1
Connect the Line terminal of the 110VAC power to the C (common) terminal of Relay 1. Connect the
Cleaner solenoid wires to the neutral terminal of the 110 VAC power and the NO (normally open)
terminal of Relay 1. (See Section 7.3)
Relay 2
Wire Relay 2 as a NO relay. Relay 2 is an alarm relay and should be connected to the Control System or
an external alarm.
Do not run the cable through the same cable gland fitting as the electrodes or the 4-20 mA signals.
The MASTER configuration has (4) SPDT 230V 5A relays that are wired as normally open (NO) relays.
Relay 1
Wire Relay 1 as a NO relay. Relay 1 is an alarm relay and should be connected to the Control System or
an external alarm.
Do not run the cable through the same cable gland fitting as the electrodes or the 4-20 mA signals.
Relays 2 (NO), 3 (NO), 4 (NO)
Connect to AC10 per wiring diagram in AC10 Air Blast Spray Cleaner
The SLAVE configuration has (2) SPDT 230V 5 A relays that are wired as normally open (NO) relays.
Relays 1, 2
Wire Relays 1 and 2 as a NO relays. Relays 1 and 2 are alarm relays and should be connected to the
Control System or an external alarm.
HYDRA NH
+
-N Page 7
4
Do not run the cable through the same cable gland fitting as the electrodes or the 4-20 mA signals.
Page 8 HYDRA NH
+
-N
4
3.0 OPERATION
This section will provide a basic overview of the C22 Analyzer/Controller. It covers the basic Menu
structure, the function of the MENU SELECT keys and the CALIBRATE keys.
The C-22 HYDRA analyzer has three measurement channels:
Channel 1, Ammonium as Nitrogen, NH
Channel 2, Potassium, K+ and
Channel 3, pH
+
4
-N,
3.1 KEYS
The blinking cursor indicates the active point where menus can be selected or numerical values
adjusted. There are two sets of keys on the C22 analyzer, the MENU SELECT keys and the CALIBRATE
keys.
The MENU SELECT keys are used to move the cursor vertically changing the displayed menu. These keys
are also used to Save/Accept the calibration data and exit the calibration menu.
The CALIBRATE keys are used to enter menus, change numerical values and move the cursor.
***To enter a calibration menu or parameter adjustment line simultaneously press both of the
Horizontal CALIBRATE keys, ◄CALIBRATE►***
Pressing either of the Horizontal CALIBRATE keys separately will move the cursor horizontally to the
point under the digit to be adjusted. The Vertical CALIBRATE keys are used to adjust numeric values.
Pressing the upper key will increase the value and pressing the lower key will decrease the value. When
the cursor is on a menu line, not in the menu, pressing the down CALIBRATE key will return the cursor to
the HOME Screen from any menu.
HYDRA NH
+
-N Page 9
4
Screen Displayed
Button Pressed
Home Screen
MENU SELECT
Graphical display
MENU SELECT
Parameter Selection Screen (
Ch 1 Buffer
) MENU SELECT
Parameter Selection Screen
(Ch 1 Set
-Up) MENU SELECT
Parameter
Selection Screen (
Ch 1 Status
) MENU SELECT
Configure/ Trim Menu
MENU SELECT
..... To return to the Home screen
Screen Displayed
Button Pressed
Home Screen
MENU SELECT
pH Manual Mode Screen
MENU SELECT
Parameter Selection Screen (
Ch 2 Status
) MENU SELECT
Parameter Selection Screen (
Ch 2 Set
-Up) MENU SELECT
Parameter Selection Screen (
Ch 2 Buffer
) MENU SELECT
To channel 3
MENU SELECT
….. To return to the Home screen
Screen Displayed
Button Pressed
Parameter Selection Screen (
Ch 2 Buffer
) MENU SELECT
Parameter Selection Screen (
Ch 3
Status
) MENU SELECT
Parameter Selection Screen (
Ch 3
Set-Up)
MENU SELECT
Parameter Selection Screen (
Ch 3
Buffer
) MENU SELECT
Contrast
(adjustment screen)
MENU SELECT
….. To return to the Home screen
3.2 MENUS
There are three sets of menus on the Model C-22. The Channel 1 menus (PV1) the Ammonium Ion
channel (NH
the pH channel. Pressing the MENU SELECT UP key will first access the Potassium channel and then the
pH channel, pressing the MENU SELECT DOWN key will access the Ammonium channel. See the guide
below for the general menu structure.
3.2.1 Ammonium Ion Channel 1 Menu Structure
3.2.2 Potassium Ion Channel 2 Menu Structure
+
-N), the Channel 2 menus (PV2) the Potassium Ion channel and the Channel 3 menus (PV3)
4
▼
▼
▼
▼
▼
▲
▲
▲
▲
▲
▲
▼
3.2.3 pH Channel 3 Menu Structure
▲
▲
▲
▲
▼
3.2.4 Graphical Display Screen
This Graphical Display is only available in the Channel 1 menu. It is located one screen down from the
Home Screen. It displays a user assigned output value, for example, 4-20 mA1 vs. time. The Graphical
Page 10 HYDRA NH
+
-N
4
Screen
Displayed
Button Pressed
Ch1 Buffer
CALIBRATE
(enter
the calibration menu by pressing
both
◄►
simultaneously
)
1 NH4-N
4.0
0 ppm
Cal
.0 mV
CALIBRATE
(enter line
, zero pt. Cal 1)
Sensor should be in the Cal 1 solution.
1
NH4-N
4.00
ppm
Cal
393.5 mV
To set Calibration Sol. Value use the CALIBRATE
or ►
to
move cursor. Use the
to adjust the integer value.
1
NH4-N
5.00 ppm
Cal
393.5 mV
MENU SELECT
(accept Calibration)
1 NH4-N
5.00 ppm
Cal
393.5 mV
MENU SELECT
(move to Cal line)
1
NH4-N
5.00
ppm
C
al
393.5 mV
MENU SELECT
(move to Cal 2, Slope)
Remove sensor, rinse and place in Cal 2 solution
2 NH4-N
00.0 ppm
Cal
29.6 mV/dec
CALIBR
ATE ►
(enter screen)
2
NH4-N
00.0
ppm
Cal
29.6 mV/dec
To set Calibration Sol. Value use the CALIBRATE
or ►
to
move cursor. Use the
to adjust the integer value.
2
NH4-N
5
0.0 ppm
Cal
29.6 mV/dec
MENU SELECT
(accept Calibration)
2 NH4-N
5
0.0 ppm
Cal
29.2 mV/dec
CALIBRATE
(return to Home Screen)
Display variables; which Output to display and which Time interval is used are configured in the Channel
1 Set-Up Menu. See Section 3.2.6, below, for more information on the Set-Up Menu.
3.2.5 Buffer Menu
The Buffer Menus are Calibration Menus. To access the Buffer Menu the cursor must be flashing in front
of the “Ch X Buffer” line. Simultaneously press both of the Horizontal CALIBRATE keys to enter the
Calibration screens. The calibration is structured for a two point calibration.
The first calibration “1 NH4-N” is the “zero point” calibration. This sets a base millivolt value to a specific
ion concentration. The concentration is adjustable to suit the measurement range, see procedure
below. When adjusting a value first move the cursor to the largest integer to be adjusted, perform the
adjustment and then move to the next lower value. When a calibration is performed the Cal line will
display the millivolt value associated with calibration standard solution. The Cal value can be reset to the
Factory default setting by “entering” the Cal line.
The second calibration screen is the “slope, mV/decade” calibration. This calibration should use a
standard solution that is at least 10X the concentration of the solution used in the first calibration. Scroll
down to the “2 NH4-N” line and enter the value of the standard solution and the Cal line will display the
new slope, mV/decade. Follow the same procedure for calibration of Channels 2 & 3.
◄
◄
◄
►
►
◄
▲or▼
▼
▼
▼
◄
▲or▼
▼
▼
3.2.6 Set-Up Menu
The Setup menu allows the various parameters to be configured for the user’s requirements. Each
Channel has a Set-Up Menu that includes the parameters associated with that specific Channel. The
HYDRA NH
+
-N Page 11
4
Screen Displayed
Button Pressed
Ch1 Setup
◄
CALIBRATE
►
(enter screen)
P
lot 4-
20 1
MENU SELECT
Plot
4-20 1
MENU SELECT
4mA1
.0 ppt
OUTPUT SET
-UP
CALIBRATE
►
(enter 4mA line)
4mA1
.0 ppt
To adjust the value
use the CALIBRATE
or ►
to move
4mA1 00.1 ppb
MENU SELECT
(accept the Value)
4mA1
00.1
ppb
MENU SELECT
4mA1
00.1
ppb
CALIBRATE
►
(enter 20mA line)
4mA1
00.1
ppb
To adjust the value
use the CALIBRATE
or ►
to move
4mA1
00.1
ppb
MENU SELECT
(accept the Value)
4mA1
00.1
ppb
MENU SELECT
2 On
> .0 ppt
RELAY SET
-UP
CALIBRATE
►
(enter Relay Set Point)
2 On > .
0
ppt
To adjust the value
use the CALIBRATE
or ►
to move
2
On >
10.00
ppm
MENU SELECT
(save
the Value)
2 On >
1
00.0
ppm
MENU SELECT
2
On >
10.00
ppm
CALIBRATE
►
(enter Relay Hysteresis Point)
2 On
> 10.00
ppm
To adjust the value
use the CALIBRATE
or ►
to move
2 On
< 10.00
ppm
MENU SELECT
(save
the Value)
** Whether the relay is a
2
On
< 10.00
ppm
(Low Alarm
MENU SELECT
(If the NH
+
-
N drops below 10
.
0 ppm the
Graphical Display, 4-20 mA Outputs, Alarm Relays and Timers are all configured in the Setup Menu. Ion
selective electrodes have unique Iso-Potential points, the mV value for the concentration where changes
in temperature have no effect. This value is supplied with each specific ion electrode and must be
entered in the “isopot” entry line below. This provides the analyzer with the base point for temperature
compensation.
▼
Sample .1 m
(the “4-20 1” indicates the displayed value, variable)
▼
Sample .1 m
(the Sample time is 0.1 minutes, variable)
◄
20mA1 999.9 ppt
◄
20mA1 999.9 ppt
cursor. Use the ▲or▼ to adjust the integer value.
▼
20mA1 999.9 ppt
▼
20mA1 999.9 ppt
◄
20mA1 999.9 ppt
◄
20mA1 999.9 ppt
cursor. Use the ▲or▼ to adjust the integer value.
▼
20mA1 100 ppm
▼
20mA1 100 ppm
(4-20 is set to 00.1 ppb to 1000 ppm)
◄
2 Off < .0 ppt
◄
2 Off < .0 ppt
cursor. Use the ▲or▼ to adjust the integer value.
▼
2 Off < .0 ppt
▼
2 Off < .0 ppt
◄
2 Off < .0 ppt
◄
2 Off < .0 ppt
cursor. Use the ▲or▼ to adjust the integer value.
▼
2 Off > 11.00 ppm
2 Off > 11.00 ppm example)
2 On > 10.00 ppm (High Alarm
2 Off < 9.500 ppm example)
High or Low Alarm is determined by the Off value **
▼
4
contact closes until it is higher than 11 ppm. If the Off value
is set to 9.5 ppm then the contact will switch to a High
Alarm, closing at 10.0 ppm and re-opening below 9.5 ppm.)
Page 12 HYDRA NH
+
-N
4
T
imer 1: periodic
CALIBRATE
►
P
er 0d00 :00 :00
Cleaning Timer
CALIBRATE
►
Per 0d0
0 :15 :00
To adjust the value
use the CALIBRATE
or ►
to move
Per 0d0
0 :15
:00
CALIBRATE
►
(enter the
Off line, the duration of the
Timer 2: oneshot
↑
CALIBRATE
►
N
ow 0 00 :00 :00
Manual Mode Timer
MENU SELECT
, MENU SELECT
CALIBRATE
►
Ks
.120
MENU SELECT
NH4
electrode selectivity ratio for K
+
Ks
.120
MENU SELECT
K+ compensation should be
Off for
T
C .333 %/
°C
MENU SELECT
T
C .333 %/
°C
CALIBRATE
►
(Enter Iso
-
Potential Entry line)
T
C .333 %/
°C
To adjust the value
use the CALIBRATE
or ►
to move
T
C .333 %/
°C
MENU SELECT
(accept the Value)
T
C .333 %/
°C
MENU SELECT
Noise Filter 5
CALIBRATE
(return to Home Screen)
◄
←Clock
◄
Now 0 00 :00 :00 (Per sets the time
On 0 00 :00 :00 between cleaning
Off 0 00 :00 :00 cycles)
(Set the Period to 15 minutes then every ¼ hour the Now
Clock will reset to 0:00:00 and trigger the On Timer which is
set at 0:00:00, starting the cleaning cycle.)
◄
Now 0 00 :00 :00
On 0 00 :00 :00
Off 0 00 :00 :00
cursor. Use the ▲or▼ to adjust the integer value, then
MENU SELECT ▼ (save the Value)
MENU SELECT ▼, MENU SELECT ▼, MENU SELECT ▼
◄
Now 0 00 :00 :00 (Set the Off time
On 0 00 :00 :00 at 30 seconds)
Off 0 00 :00 :30
cleaning cycle) To adjust the value use the CALIBRATE ◄ or
► to move cursor. Use the ▲or▼ to adjust the integer
value, then MENU SELECT ▼ (save the Value)
◄
←timer 1←clock
On 0 00 :00 :00 (Hold Function)
Off 0 00 :01 :30 (Set at 1 min 30 sec)
4-20 mA hold time after cleaning cycle, (CT + 1 minute)
▼
▼, ◄
To adjust the value use the CALIBRATE ◄ or ► to move
cursor. Use the ▲or▼ to adjust the integer value.
**Iso-Potential value supplied with each electrode**
Dissoc 1.000 Off
◄
Isopot 20.0 mV
cursor. Use the ▲or▼ to adjust the integer value.
Equilib 9.24 pH
Dissoc 1.000 Off
▼
Isopot 187.0 mV
Equilib 9.24 pH
Dissoc 1.000 Off
MENU SELECT ▼ Equilibrium value, No Input Needed
(Equilib= pH value equal to the pKa, the acid dissociation
constant, where [NH4 = NH3 + H+])
▼
Isopot 187.0mV
Equilib 9.24 pH
Dissoc 1.000 On
Dissoc = the pH correction factor for the [NH4 ↔ NH3 + H+]
dissociation , (the Measured NH4)/Dissoc = Total NH4-N
(should be on at all times)
▼
HYDRA NH
+
-N Page 13
4
Screen Displayed
Button Pressed
C
h1 Status
◄
CALIBRATE
►
(enter screen)
Fri 09:11:28
MENU SELECT
Input 364.2 mV
CALIBRATE
(return to Home Screen)
Screen Displayed
Button Pressed
Co
nfigure/Trim
◄
CALIBRATE
►
(enter screen)
Passwords
MENU SELECT
(allows menus to be
password protected)
Default Display
MENU SELECT
(defines Home Screen)
4-20 assign
MENU SELECT
(Assigns parameter/function to mA output)
Manual Mode
MENU SELECT
(Allows manual control of mA outputs)
Relay assign/Test
MENU SELECT
(Assigns triggering
parameter to a relay)
PID Assign
MENU SELECT
(Assigns PID to a PV)
PWM Assign
MENU SELECT
(Assigns Pulse Width Modulation to a PV)
Clock and timers
MENU SELECT
(Sets Clock function, assigns Timers)
Logic Gates
MENU SELECT
(assigns AND/OR gates)
Ion Species
MENU SELECT
(Sets Ion Measured)
General
MENU SELECT
( Not Used)
°C/°F & Temp Cal
MENU SELECT
( allows temperature trim/cal, choose °C/°F)
4-20 Trim/Test
MENU SELECT
(allows mA trim)
Millivolt Trim
CALIBRATE
(return to Home Screen)
3.2.7 Status Menu
The Status menu displays the actual uncompensated millivolt value from the sensor, very handy
information for troubleshooting. The day and time, hours: minutes: seconds, are also displayed.
▼
▼
3.2.8 Configuration and Trim Menu
The Configure/Trim menu provides access to the menus used to assign parameters, define functions and
trim input and output signals. This group of menus should rarely be needed as the instrument was
configured at the factory before testing and shipment. The basic structure is as follows. See the “System
22 Configuration Manual” for details about these menus, available at www.ecdi.com.
▼
▼
▼
▼
▼
▼
▼
▼
▼
▼
▼
▼
▼
▼
3.3 OUTPUTS
The Standard C-22 HYDRA Analyzer has two 4-20 mA outputs. The default configuration has “mA 1”
configured for NH4-N, 4-20 mA = 00.1 ppb to 50 ppm and “mA 2” is configured for pH, 4-20 mA = 2-12
pH. To reconfigure the outputs see Section 3.2.6, Set-Up Menu, above.
Page 14 HYDRA NH
+
-N
4
The Outputs, mA 1 and mA 2, are held at the “Last Value” by Timer 2 during the Air Blast Cleaning Cycle.
The Outputs return to live readings one minute after the cleaning cycle ends. Timer 2 is a “one-shot”
timer triggered by Timer 1. When Timer 1 triggers, it starts Timer 2 which is set to run for a period of
time equal to the cleaning time plus one minute.
3.3.1 Manual Mode (4-20 mA Hold Function)
The % Output is displayed on the Home Screen for the NH4-N output, mA 1, and one screen above the
Home Screen for the pH % Output, mA 2.
The outputs can be set for manual control by simultaneously pressing both horizontal CALIBRATE keys
when the cursor is in front of the % Output line. An “M” will be displayed in front of the % Output value.
The Output will remain frozen at the last value until the Manual Mode is turned off. The Manual Mode
hold function is commonly used to freeze the outputs during maintenance and calibration cycles. While
in Manual Mode the Output can be adjusted between 0 and 99.9% using the CALIBRATE keys, very
useful for troubleshooting loop communication problems.
To exit Manual Mode press the left CALIBRATE arrow, ◄, three times and the “M” will disappear, the
Output will return to a live reading.
3.4 CONTACTS-ALARM RELAYS
3.4.1 Standard Configuration
Relay 1 is configured to control the Air Blast Spray Cleaner. It is triggered by Timer 1. Timer 1 is
configured as a periodic timer. The default setting for the period is 15 minutes with the duration set at
30 seconds. The period and duration of the cleaning cycle is user configurable and dependent on the
application. For water going to the clarifier an adequate cleaning cycle may be, every two hours for 15
seconds, where incoming water to an aeration basin may require a 30 second cleaning cycle every 15
minutes.
The Now clock is an arbitrary timer. The Period determines the amount of time that must pass before
the Now clock resets to 00 hours, 00 minutes and 00 seconds, 00: 00: 00. The On time is the start time
and it is set at 00: 00: 00, so every time the clock resets Relay 1 is triggered. The Off time sets the
duration of Timer 1 and is set at 00: 00: 30, thirty seconds. (See Section 3.2.6 above)
To Change how often the Spray Cleaner actuates, change the Per line to the desired period. The display
is set for days, hours: minutes: seconds, 0d00: 00: 00.
To Change the duration of the Cleaning cycle change the Off line. If the duration is changed from the
default value of 30 seconds then the Manual Mode Timer should be adjusted to allow the sensor to
stabilize after the cleaning cycle. Adjust Timer 2; Off line; to the new Clean Time (CT) + 1 minute.
Relay 2 is configured as a high alarm triggering at 10.0 ppm NH
+
-N. The value can be changed in the Set-
4
Up Menu as shown in Section 3.2.6. Changing the > sign to a < sign in the On line changes the set point
from a high to a low alarm. This can also be accomplished by setting the Off value higher than the On
HYDRA NH
+
-N Page 15
4
value which forces an automatic sign change. The Off value is the hysteresis point that the PV has to
correct to before the relay disengages.
3.4.2 One AC10 with Two HYDRAs
When two HYDRA Analyzers are connected to one AC10 Air Blast Spray Cleaning System then one C22
Analyzer becomes the MASTER and the other C22 Analyzer is a SLAVE.
3.4.2.1 Master Configuration
Relay 1 is configured as a high alarm triggering at 10.0 ppm NH
+
-N. The value can be changed in the Set-
4
Up Menu as shown in Section 3.2.6. Changing the > sign to a < sign in the On line changes the set point
from a high to a low alarm. This can also be accomplished by setting the Off value higher than the On
value which forces an automatic sign change. The Off value is the hysteresis point that the PV has to
correct to before the relay disengages.
Relay 2 is configured to trigger the Digital Input (Aux In) of the SLAVE analyzer. This initiates the Manual
Mode function of the SLAVE analyzer, placing the mA outputs in a Last Value Hold (Bumpless).
Relay 3 is configured to control the Air Blast Spray Cleaner. It is triggered by Timer 1. Timer 1 is
configured as a periodic timer. The default setting (Set Up Menu) for the period is 15 minutes with the
duration set at 30 seconds. This triggers a cleaning cycle every 15 minutes alternating between the
MASTER and the SLAVE units, so each HYDRA sensor is cleaned every 30 minutes. (see Relay 4
below)The period and duration of the cleaning cycle is user configurable and dependent on the
application. For water going to the clarifier an adequate cleaning cycle may be, every two hours for 15
seconds, where incoming water to an aeration basin may require a 30 second cleaning cycle every 15
minutes.
The Now clock is an arbitrary timer. The Period determines the amount of time that must pass before
the Now clock resets to 00 hours, 00 minutes and 00 seconds, 00: 00: 00. The On time is the start time
and it is set at 00: 00: 00, so every time the clock resets Relay 1 is triggered. The Off time sets the
duration of Timer 1 and is set at 00: 00: 30, thirty seconds. (See Section 3.2.6 above)
To Change how often the Spray Cleaner actuates, change the Per line to ½ the desired period. The
display is set for days, hours: minutes: seconds, 0d00: 00: 00.
To Change the duration of the Cleaning cycle change the Off line. If the duration is changed from the
default value of 30 seconds then the Manual Mode Timer should be adjusted to allow the sensor to
stabilize after the cleaning cycle. Adjust Timer 2; Off line; to the new Clean Time (CT) + 1 minute.
Relay 4 is configured to control the diverter valve which alternates the Air Blast between the two
sensors. It is triggered by Timer 2. Timer 2 is configured as a periodic timer. The default setting (Set Up
Menu) for the period is 30 minutes with the duration set at 30 seconds. The Period of Timer 2 must be twice the Period of Timer 1 and have the same duration.
Page 16 HYDRA NH
+
-N
4
% NH4+
% NH3
3.5 ION SELECTIVE FEATURES
The Hydra C22 analyzer has several features unique to Ion Selective Electrode measurements,
Isopotential, Interfering Ion Correction and pH Correction. The measurement range of the sensors is
from some ppb level to some ppm level, XXX ppb – XXX ppm, not the traditional 0 – XXX common to
most measurements, there is no zero value. This is most evident when configuring the 4-20 mA outputs.
3.5.1 The Isopotential point
Isopot in the menu is the mV value of the specific sensor where changes in the temperature of the
solution do not change the output of the sensor. This is the base point for temperature compensation
correction. Most pH electrodes have an isopotential near pH 7, but each type of Ion Selective electrode
has a unique isopotential value. Each NH4+ and K+ electrode is supplied with an Isopot that must be
entered in the respective Set-Up Menu.
3.5.2 Potassium Ion Interference Correction
Potassium ions and Ammonium ions have a similar ionic size and the same electrical charge. If
Potassium ions are present in the measured solution then they will cause a positive Interference on the
Ammonium Ion Measurement. The corrections for the Interfering Ions are in the Ks and K+ comp screen
in the Set-Up menu. The Ks, potassium ion selectivity, is the correction factor, ppm NH
+
/ppm K+;
4
Example: (50 ppm K+) x (0.10 ppm NH
With the Ks set at 0.10 (10 K+/ NH
+
) then 50 ppm potassium ion would cause the ammonium
4
+
/ ppm K+) = 5 ppm NH
4
+
interference
4
measurement to read 5 ppm high. When the K+ comp line is On the C22 Analyzer uses the potassium
ion concentration from the Potassium Ion Electrode and the Ks factor to calculate the amount of
interference on the Ammonium Ion Electrode and subtracts it from the measured ammonium value
before displaying the NH4-N value.
3.5.3 pH Correction
The Ammonium Ion Electrode only measures the ammonium ion (NH
+
) not ammonia (NH3). Ammonium
4
ion and ammonia coexist in a pH dependent ratio in
solution. At pH 9.24 the ratio is 1:1, at pH 7 nearly all of the
ammonia is in the ammonium state and at pH 11 it is nearly
all ammonia which is invisible to the sensor. The more acidic
values favor the NH
+
and the more basic values favor
4
ammonia gas, NH3. The pH Electrode measures the pH and
the HYDRA C22 Analyzer calculates the total NH
+
-N
4
concentration based on the pH vs. concentration profile.
The Dissociation should be On at all times.
% concentration vs pH
100
80
60
%
40
NH3/NH
+
4
HYDRA NH
20
0
681012
pH
+
-N Page 17
4
Ch 1 Setup Menu
Settings
Settings
Plot 4
-
20 1
4mA1
0.1 ppb
2 On > 10.00 ppm
Timer 1 : periodic
Per 0d0
0 :15
:00
Timer 2 :
oneshot
↑ Now 0 00 :00 :00
Ks 0.10
TC .333 %/°C
Noise filter 5
Ch 2
Setup Menu
Settings
Settings
TC .333 %/°C
Noise filter 5
4.0 START UP
The HYDRA Analyzer and Sensor was assembled and tested at the factory before shipment. The sensor
was then disconnected from the analyzer, cleaned and packaged for shipment. The HYDRA C22 analyzer
is configured as shown below and should be ready to use aside from any changes the user may wish to
make to the Outputs, Cleaning Timers and Alarm Set points.
Complete all of the steps in Section 2 INSTALLATION (above) before proceeding further. These steps
include:
1. Install the Analyzer in a suitable environment.
2. Determine an installation method and representative sample point for the sensor.
3. Connect the Air Blast Spray Cleaner tube to a solenoid controlled air supply, 1-4 bar (15-60 psig).
4. Wire the Power, Outputs, Contact Relays and the HYDRA Sensor to the HYDRA C22 Analyzer.
Familiarize yourself with the MENU Structure of the HYDRA C22 Analyzer, Section 3 (above).
4.1 SET-UP AND CONFIGURATION
The HYDRA C22 analyzer is configured as shown below. Changes to the default configuration can be
easily made as described in Section 3 above.
4.1.1 Set Up Menus, Standard
Sample .1 m
20mA1 50.00 ppm
2 Off < 9.500 ppm
K+ comp On
Isopot XXX.X mV
Equilib 9.24 pH
Dissoc 1.000 On
Now 0 00 :00 :00
On 0 00 :00 :00
Off 0 00 :00 :30
On 0 00 :00 :00
Off 0 00 :01 :30
Page 18 HYDRA NH
Isopot XXX.X mV
+
-N
4
Ch 3 Setup Menu
Settings
Settings
4mA 2.00 pH
Noise filter 5
Ch 1
Setup Menu
Settings
Settings
Plot 4
-
20 1
4mA1 0.1 ppb
1 On > 10.00 ppm
Timer 1 : periodic
Per 0d00 :15 :00
Timer 2 : periodic
Per 0d00 :30 :00
Timer 3 : oneshot
↑ Now 0 00 :00 :00
Ks 0.10
TC .333 %/°C
Noise filter 5
Ch 2 Setup Menu
Settings
Settings
TC .333 %/°C
Noise filter 5
Ch 3 Setup Menu
Settings
Settings
4mA 2.00 pH
Noise filter 5
Ch 1 Setup Menu
Settings
Settings
Plot 4
-
20 1
4mA1 0.1 ppb
20mA 12.00 pH
4.1.2 Set Up Menu, MASTER/SLAVE
Master
Sample .1 m
20mA1 50.00 ppm
1 Off < 9.500 ppm
K+ comp On
Isopot XXX.X mV
Equilib 9.24 pH
Dissoc 1.000 On
Now 0 00 :00 :00
On 0 00 :00 :00
Off 0 00 :00 :30
Now 0 00 :00 :00
On 0 00 :00 :00
Off 0 00 :00 :30
On 0 00 :00 :00
Off 0 00 :01 :30
Slave
HYDRA NH
Isopot XXX.X mV
20mA 12.00 pH
Sample .1 m
20mA1 50.00 ppm
+
-N Page 19
4
1 On > 10.00
ppm
Timer 1 : oneshot
↑ Now 0 00 :00 :00
Ks 0.10
TC .333 %/°C
Noise filter 5
Ch 2 Setup Menu
Settings
Settings
TC .333 %/°C
Noise filter 5
Ch 3 Setup Menu
Settings
Settings
4mA 2.00 pH
1 On > 7.50 pH
Noise filter 5
Menu
Settings
Passwords
No Pass Words (not configured)
Default display
0 (not configured)
4-20
assign
4-
20 1 : Ch1 PV
Manual Mode
Ch1 man mode on: 4
-
20 1
Ch2 man mode on:
4-20 2
Re
lay assign/test
Relay
1 timer 1
PID Assign
PID A unused
Clock and timers
Clock: Software
Ch 1 Setup Menu,
cont’d
1 Off < 9.500 ppm
On 0 00 :00 :00
Off 0 00 :01 :30
K+ comp On
Isopot XXX.X mV
Equilib 9.24 pH
Dissoc 1.000 On
Isopot XXX.X mV
20mA 12.00 pH
1 Off < 7.40 pH
4.1.3 Configure and Trim Menu, Standard
The Configure and Trim is configured as shown be and does not require any adjustments.
4-20 2 : Ch3 PV
Initial ch1 manual setting: Bumpless
Ch1 auto entry: timer 2
Initial ch2 manual setting: Bumpless
Ch2 auto entry: timer 2
Relay 2 Ch1 PV
(Relays 3-8 set to missing)
PID B unused
Hide Clock N
Hide Rate Y
Hide Flash N
Man trig OK Y
Page 20 HYDRA NH
+
-N
4
Timer 1: periodic
Timer2:
oneshot
↑
Logic gates
All unused
Ion Species
Ch1 ion NH4
-N
General
No Menu
°C/°F & temp cal
1 Trim °C .0
4-20 Trim/Test
(not configured)
Millivolt trim
(not configured)
Menu
Settings
Passwords
No Pass Words (not configured)
Default display
0 (not configured)
4-20 assign
4-
20 1 : Ch1 PV
Manual Mode
Ch1 man mode on:
4-20 1
Ch2 man mode on: 4
-
20 2
Relay assign/test
Relay 1 Ch1 PV
PID
Assign
PID A
& B
unused
Clock and timers
Clock: Software
Timer 1: periodic
Timer 2: periodic
Timer
3: oneshot
↑
Logic gates
AND Gate A
Retriggerable N
Resettable N
Trig 2 timer 1
Ch2 ion K+
Unit 200.0 °C
4.1.4 Configure and Trim Menu, MASTER/SLAVE
MASTER
4-20 2 : Ch3 PV
Initial ch1 manual setting: Bumpless
Ch1 auto entry: timer 3
Initial ch2 manual setting: Bumpless
Ch2 auto entry: timer 3
Relay 2 timer 2
Relay 3 timer 1
Relay 4 timer 2
Hide Clock N
Hide Rate Y
Hide Flash N
Man trig OK Y
Retriggerable N
Resettable N
Trig 3 AND A
HYDRA NH
+
-N Page 21
4
In Timer 1
In Timer/2
OR Gates A, B, C Not Used
Ion Species
Ch1 ion NH4
-N
General
No Menu
°C/°F & temp cal
1 Trim °C .0
4-20 Trim/Test
(not
configured)
Millivolt trim
(not configured)
SLAVE
Menu
Settings
Passwords
No Pass Words (not configured)
Default display
0 (not configured)
4-20 assign
4-
20 1 : Ch1 PV
Manual Mode
Ch1 man mode on: 4
-
20 1
Ch2
man mode on: 4
-
20 2
Relay assign/test
Relay 1
Ch 1 PV
PID Assign
PID A unused
Clock and timers
Clock: Software
Timer
1: oneshot
↑
(Timers 3
-
8 unused)
Logic gates
All unused
Ion
Species
Ch1 ion NH4
-N
General
No Menu
°C/°F & temp cal
1 Trim °C .0
4-20 Trim/Test
(not configured)
Millivolt trim
(not configured)
AND Gate B, C Not Used
Ch2 ion K+
Unit 200.0 °C
4-20 2 : Ch3 PV
Initial ch1 manual setting: Bumpless
Ch1 auto entry: timer 1
Initial ch2 manual setting: Bumpless
Ch2 auto entry: timer 1
Relay 2 Ch 2 PV
(Relays 3-8 set to missing)
PID B unused
Hide Clock N
Hide Rate Y
Hide Flash N
Man trig OK Y
Retriggerable N
Resettable N
Trig 1 Ext/1
Ch2 ion K+
Page 22 HYDRA NH
Unit 200.0 °C
+
-N
4
4.2 CALIBRATION
The HYDRA Ammonium Analyzer was calibrated at the factory using the Two Point Calibration described
below. The sensor was assembled and placed beaker of equilibration solution for 8-12 hours before
starting the calibration. The Potassium Ion and Ammonium Ion electrodes sensing membranes require
several hours to properly hydrate/ equilibrate to the measured solution. The calibration sets the zero
potential and slope for each of the sensors into the analyzers memory.
Since the factory calibration characterizes the complete measurement system the initial user calibration
should only require a standardization of the Ammonium Ion Electrode once the HYDRA has equilibrated
to the process solution.
The Factory Calibration and all subsequent calibrations should be performed in the following sequence:
1. Temperature Calibration
2. pH Electrode calibration in pH 4.01 buffer and pH 7.00 buffer (2 point)
3. Potassium Electrode calibration in 1 ppm and 100 ppm solutions (2 point)
4. Ammonium Electrode calibration in 1 ppm and 100 ppm solutions (2 point)
Calibration checks should be performed weekly by comparison to a grab sample or immersing the
HYDRA sensor into a standard solution with known pH, K+ and NH4-N values. Performing a Single Point
Calibration will eliminate any offset found in the calibration check. Continuous drifting of the potassium
or ammonium ion reading may indicate the electrode needs to be replaced. (See Section 6 Trouble
Shooting below)
Two Point Calibration is required whenever an electrode is replaced, pH, Potassium or Ammonium. Two
Point Calibrations are also recommended every two months to verify the response of the electrodes is
greater than 50 mV/decade.
Place the HYDRA sensor in a beaker of water so that the bottom 3 inches (7.6 cm) of the body is
immersed. Allow the sensor to equilibrate for 10 minutes. Verify the displayed temperature agrees with
the thermometer. If not, then adjust the temperature in the Configure/Trim > °C/°F & temp cal > 1 Trim
°C menu to agree with the thermometer. This calibration was performed in the factory calibration but it
is advisable to perform a temperature check every 2-3 months.
4.2.3 Single point Calibration
Single Point Calibrations are used to standardize the sensor’s reading to a known value typically to
eliminate errors caused by drift or large changes in the makeup of the measured solution. A Single Point
Calibration adjusts the base potential of the sensor by shifting the response curve to a higher or lower
level but it does not affect the slope (mV/ppm) of the curve.
Single point calibration is available for each of the measured parameters in the appropriate Buffer
Menu, Channel 1 Buffer for the NH
+
-N, Channel 2 Buffer for the K+ and Channel 3 Buffer for pH. (See the
4
Single Point Calibration Screen table below)
Immerse the front end of the sensor into a container of the calibration solution, enter the value of the
calibration solution or the value derived from a grab sample into the “1 (ion) xxx ppm” line of the
Buffer menu as described in Section 3.2.5 above. Accept the reading when stable by pressing either of
the MENU SELECT keys. The lower line in the menu, the Cal line will show the new mV value associated
with the entry.
***Important Note for Ammonium Calibrations*** When using the 1 ppm and 100 ppm NH
+
-N
4
calibration solutions to calibrate the HYDRA C22 analyzer the potassium ion compensation, the “K+
comp” line in the Channel 1 Setup menu should be turned on, it is on by default. These calibration
solutions contain no potassium ions but the ammonium ions interfere with the potassium ion electrode
and the potassium ion electrode registers a K+ reading, this value must be accounted for in the NH
+
-N
4
calibration.
If the Single Point Calibration is the first point in a Two Point Calibration it is customary to use the
solution with the lowest value for the single point calibration. There are no restrictions on the value
entered in the “1 (ion) xxx ppm” line of the Buffer menu; it can be higher or lower than the value in line
2 of the menu. For the pH calibration the zero point is set with pH 7.0 calibration buffer and the slope is
set with the pH 4.01 calibration buffer.
Single Point Calibration Screens
Cal 320.0 mV
Cal 165 mV
or Grab Sample value
or Grab Sample value
Page 24 HYDRA NH
Cal 3.1 mV
or Grab Sample value
+
-N
4
4.2.4 Two point Calibration
The second point of a Two Point Calibration sets the slope of the sensor, the mV per decade. The slope is
calculated by comparing the millivolt values and ppm values in the “1NH4-N 1.00 ppm” line to the
values in the “2NH4-N 100 ppm” line of the Buffer Menu. The concentration of the solution used for
the second point should be at least 10 times higher than the value used in the first point of the
calibration. The recommended calibration standards for both the ammonium ion and potassium ion
calibrations are 1 ppm and 100 ppm. Perform a two point calibration whenever an electrode is replaced.
After completing the single point calibration, see section 4.2.3 above, rinse the sensor with distilled
water and gently dab it dry with a paper towel or soft tissue. Carefully cleaning the sensor prevents
errors in the calibration due to carryover from the first solution. Place the sensor in the second solution,
either the pH 4.01 buffer or one of the 100 ppm solutions and wait for the reading to stabilize before
accepting the value. See section 3.2.5 above for further instruction.
4.2.5 Potassium Ion Compensation
Potassium ions have a positive interference on the Ammonium Ion Electrode, i.e. the ammonium
reading is higher than the actual value. The Potassium Ion Compensation, K+ comp, adjusts the
measured potassium ion concentration using the interference ratio value, the Ks value, and subtracts
the resulting amount from the Ammonium Ion Signal. The Ks value in the NH4-N Set Up screen allows
the interference ratio to be adjusted as needed. The default setting is Ks = .100 which is 10 potassium
ions produce the same signal as 1 ammonium ion, Ks = 0.04 would change the correction to 25:1. The
0.100 correction factor is good for most solutions with low levels of potassium, less than 20 ppm.
Each Potassium Ion Electrode should have similar characteristics but adjustments are necessary when
the potassium electrode is replaced or when high levels of potassium are present in the solution. If the
ammonium ion concentration is less than expected when compared to a grab sample measurement
then adjust the Ks value to a smaller value. Adjusting from 0.100 > 0.082 for example would increase the
potassium to ammonium ion ratio from 10:1 to 12:1 thereby slightly increasing the displayed
ammonium ion concentration by subtracting less in the compensation.
HYDRA NH
+
-N Page 25
4
5.0 MAINTENANCE
5.1 MAINTENANCE SCHEDULE
The HYDRA sensor requires little maintenance since most of the required cleaning is accomplished by
the Air Blast Spray Cleaner. Determining the proper cleaning cycle and duration for the application will
keep the front end of the sensor clean for extended periods but weekly inspection of the sensing end is
recommended.
Weekly calibration checks versus a grab sample are recommended to minimize any drift in the sensor.
Once the stability of the sensor has been established in the process, the time between calibration
checks can be adjusted. A calibration check must occur at least once per month.
The HYDRA should be checked monthly in calibration solutions for proper span of both the ammonium
and potassium electrodes. The electrodes start life with a span in the 55 mV/decade range and drop off
into the 40 mV/decade range as they age. Readings below 40 mV/decade indicate the electrode should
be replaced. Depending on the characteristics of the water being measured the Ion Selective electrodes
should last 3-6 months.
5.2 CLEANING THE SENSOR
The HYDRA sensor can be rinsed with water and wiped with a soft brush or cloth to remove most
coatings. DO NOT clean the potassium or ammonium ion electrodes with a stiff brush or vigorous wiping
as the sensing membrane is easily torn. DO NOT use strong detergent solutions or Spray Glass cleaners
to clean the sensor, the potassium and ammonium electrodes will be irreparably harmed. The sensor
can be soaked in dilute HCl, 2-3%, for 15 minutes to help remove stubborn coatings. Use a weak
detergent solution, a couple of drops detergent per liter of water, to remove any oily coatings. The pH
electrode and HYDRA housing can be cleaned in a more rigorous manner since the plastic body and glass
membrane are more durable than the Ion Selective Electrodes.
5.3 REPLACING THE ELECTRODES
The three electrodes, Ammonium, Potassium and pH are easily replaceable by simply removing the
electrode guard, screwing out the old sensor and screwing in the new one using the supplied installation
tool. Make sure no water gets inside the sensor when removing the electrodes. Water will cause the
internal electronics to fail. If the electronics are compromised the HYDRA sensor must be replaced.
The potassium and ammonium electrodes are PVC membrane combination electrodes (indicator
electrode and reference electrode in one body) and the pH electrode is a glass membrane combination
electrode.
Apply a thin film of o-ring lubricant to the o-rings on the electrode before installing it into the sensor
housing. Use the supplied electrode installation tool to remove or install electrodes into the sensor. The
use of pliers or other tools is not recommended as they may crack the electrode housing.
Page 26 HYDRA NH
+
-N
4
Problem
Possible Cause
Remedy
NH
+
-
N Reading Low
Dirty Coated electrode
Clean sensor:
Old Expired
NH4 ISE Perform two point calibration,
Inaccurate
Potassium
Verify
Potassium
ISE reading
NH
+
-N Reading High
Ammonium
Sensor has drifted
Perform a single point
Low or No
Potassium
ion
Verify
K
comp
is ON
Interfering Ions Present in
Influx of interfering ions causing
No Response to Changing
Bad Electrode
Visually inspect membrane for
Bad Sensor
Switch out old electrode with a
6.0 TROUBLESHOOTING
4
Soak in 2% HCl for 5 minutes to
remove algae or
Rinse with weak detergent 10-15
seconds solution to remove oils
Soak in tap water for 30 minutes
after either cleaning method.
Increase frequency of the air
blast cleaner, See Section 3.4
the slope > 40 mV/decade if not
replace NH4 Sensor
4
Ammonium
compensation
Compensation
Sample
with single point calibration
Adjust K coef (see section 4.2.5)
calibration.
Verify Potassium ISE reading
with single point calibration
Adjust K coef (see section 4.2.5)
error, wait for ions to clear the
process, Sodium and Calcium
ions would be the most
common, >500 ppm levels
cuts, if torn replace electrode
Remove the electrode examine
for breakage, torn o-rings,
corrosion or moisture, if any of
these issues are present then,
Switch out old electrode with a
new electrode, check response,
if good recalibrate the HYDRA
HYDRA NH
+
-N Page 27
4
new electrode, check response,
if still bad examine electrode
tube for signs of moisture.
If present remove sensor from
installation and allow the sensor
to dry out. Retest with a new
electrode, if it still fails return it
to the factory for refurbishment.
7.0 ENGINEERING DOCUMENTATION
7.1 SPECIFICATIONS
Sensor
Three Electrode system with spray cleaner, Ammonium ISE (NH
Potassium ISE and pH glass electrodes are used to compensate the NH
proof with an ingress rating of IP 68.
Measurement Range
+
NH
-N: 0.1 to 1000 ppm
4
Operating Temperature
0° C to 40° C (32° F to 104° F)
Min/Max Flow Rate
Minimum 0.1 m/s
Maximum 3.0 m/s
Wetted Materials
PVC, PES, PVDF, PTFE, Viton, Glass, 316 SS
Accuracy
± 3% of reading, dependent on Calibration
Response Time
Ammonium: 0.01 to 1000 ppm as NH
Potassium: 0.01 to 1000 ppm
pH: 0 to 14 pH
Temperature: 0° C to 100° C (32° F to 212° F)
Compensation
pH 4 – 10 pH
Potassium: 0.1 to 1000 ppm
Display
2.5” X 1.75” backlit LCD, 4 lines of Text & Graphical
Enclosure
NEMA 4X, LxWxD: 5.7” x 5.7” x 7
Outputs
(2) 4-20 mA maximum load 800 ohms @ 24 VDC
Configured: 0.1 to 20 mg/l NH
+
4
Input Power
+
- N
4
- N; 0 – 14 pH; optionally up to (4) 4-20 mA outputs
+
- N) is the primary measurement. The
4
+
signal. The Sensor is water-
4
Page 28 HYDRA NH
+
-N
4
110/220 VAC @ 50/60 Hz
Alarm Relay Ratings
(2) SPDT 230 VAC/5A or 30 VDC/5A resis0ve max.
Relay (1) Spray Cleaner, Relay (2) Alarm
Optionally up to (8) relays
7.2 OUTLINE & DIMENSIONAL DRAWING
7.2.1 Hydra Sensor Dimensions
7.2.2 C-22 Dimensions
Panel Mount Dimensions and Cut-out Pattern
HYDRA NH
+
-N Page 29
4
Wall Mounting Dimensions, Screw Hole Pattern
7.2.3 Sun Shield Mounting
Page 30 HYDRA NH
+
-N
4
7.3 WIRING DIAGRAM
7.3.1 Sensor Color Coded Diagram
7.3.2 Instrument – Terminal Layout/Connections
HYDRA NH
+
-N Page 31
4
Page 32 HYDRA NH
+
-N
4
Part No.
Model and Product Descrip
tion
1290030
-1
HYDRA NH4
-N Sensor,
1290030
-2
HYDRA NH4
-
N Sensor, (No Potassium Sensor)
16KA2221.KA00
HYDRA C22 NH4
-
N Analyzer,
16KA2221.JA00
HYDRA C22 NH4
-
N Analyzer,
16KA2221.KAL0
HYDRA C22 NH4
-
N Analyzer,
(SLAVE for (1)AC10 x (2)
HYDRAs)
16KA2421.KAK0
HYDRA C22 NH4
-
N Analyzer,
(MASTER for (1)AC10 x (2) HYDRAs)
Part No.
Accessories
Descrip
tion
2010445
-1
Ammonium Calibra
ti
on solu
ti
on, NH4
-N, 1 ppm
(500 ml)
2010446
-1
Ammonium Calibra
ti
on solu
ti
on, NH4
-N, 100 ppm
(500 ml)
2010443
-1
Potassium Calibra
ti
on solu
ti
on, 1 ppm
(500 ml)
2010444
-1
Potassium Calibra
ti
on solu
ti
on, 100 ppm
(500 ml)
2010100
pH 4 Buffer Calibra
ti
on solu
tion
(500 ml)
2010101
pH 7 Buffer Calibra
ti
on solu
tion
(500 ml)
1000260
-1
Pole Mount Instrument Sunshield
1000260
-2
Rail Mount Instrument Sunshield
Part No.
Spare Parts Descrip
tion
2005084.VIT
Ammonium Electrode Cartridge (recommended spare)
2005035.VIT
Potassium Electrode Cartridge (recommended spare)
2005145.VIT
pH Electrode Cartridge (recommended spare)
3300854
-1
Replacement Spray Nozzle
3501050
-1
PVC Front Sensor Guard
8.0 ORDERING INFORMATION
The HYDRA Ammonium Analyzer is a complete measurement system including the C22
analyzer/controller (p/n 16AK2221.1K00) and the multiple parameter HYDRA Sensor (p/n 1290030-1).
The Ammonium Ion measurement is fully compensated for the effects of changes in the pH and
potassium ions in the measured solution. The sensor features an Air Blast Cleaner that minimizes
maintenance in the dirtiest applications.
8.1 PART NUMBERS/MODEL NUMBERS
Complete, NH4, K+, pH, Temp, Spray Cleaner head and 30 ft. cable
Complete, NH4, pH, Temp, Spray Cleaner head and 30 ft. cable
pH & K+ compensated, (2) 4-20 mA output, 0.1 – 50 ppm NH4-N and (2) relays
(No K+ compensation), (2) 4-20 mA output, 0.1 – 50 ppm NH4-N and (2) relays
pH & K+ compensated, (2) 4-20 mA output, 0.1 – 50 ppm NH4-N and (2) relays
pH & K+ compensated, (2) 4-20 mA output, 0.1 – 50 ppm NH4-N and (4) relays
8.2 ACCESSORIES
1000254-XX
1000254-99
2000255-1 2” Handrail Swivel Mounting Bracket
Immersion Assembly, 1 ½” Slip x 45°, 1 ¼” NPT sensor connection and 1 ½” Slip x
Cable feed through Elbow End Cap, XX= Length in ft.
Immersion Assembly, 1 ½” Slip by 45° 1 ¼” NPT sensor connection and 1 ½” Slip x
Cable feed through Elbow End Cap, Fittings Only , User supplied Pipe (1 ½” Pipe)
8.3 SPARE PARTS
HYDRA NH
+
-N Page 33
4
3500007
Electrode Replacement Tool
3500008
Calibration Cup,
(remove guard and it threads onto the front, ~250 ml)
Page 34 HYDRA NH
+
-N
4
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.