Dynatronics dynatron solaris 709 plus, dynatron solaris 707 plus, dynatron solaris 708 plus, dynatron solaris 705 plus, dynatron solaris 706 plus Service Manual
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 2017I
CAUTION
Federal law restricts these devices for sale by or on the order of a physician, chiropractor, physical therapist, or
dentist licensed by the law of the state in which said person practices to use or order the use of the devices.
Risk of burns and re - Do not use near conductive materials such as metal bed parts, inner spring mattresses
and the like.
DANGER - Explosion Hazard: Do not use in the presence of ammable anesthetics.
IMPORTANT: Before treating a patient with any Dynatron Solaris Plus® Device, see the “Contraindications, Warnings, and
Precautions” in this manual. Read the operating instructions for each modality carefully.
INDICATIONS FOR USE
ELECTROTHERAPY: Electrical muscle stimulation therapy (Russian, Biphasic, High Volt) for:
1. relaxation of muscle spasm;
2. prevention or retardation of disuse atrophy;
3. increasing local blood circulation;
4. muscle re-education;
5. immediate post-surgical stimulation of calf muscles to prevent venous thrombosis
6. maintaining or increasing range of motion.
Transcutaneous electrical nerve stimulation and Interferential Current erapy (Interferential, Premodulated, High Volt,
Microcurrent) for: Symptomatic relief of chronic intractable and/or management of post-traumatic or post-surgical pain.
DIRECT CURRENT THERAPY: Direct Current is indicated for relaxation of muscle spasms.
ULTRASOUND THERAPY: Ultrasound therapy is intended to generate deep heat within body tissues for the treatment
of selected medical conditions such as relief of pain, muscle spasms, and joint contractures, but not for the treatment of
malignancies.
LIGHT THERAPY: Light therapy provides topical heating for temporary increase in blood circulation, temporary relief of
minor muscle and joint aches, pain and stiness, relaxation of muscles, and treatment of muscle spasms and minor pain and
stiness associated with arthritis.
THERMOSTIM: A hand held cutaneous electrode to be used with Dynatronics Solaris devices to apply electrical stimulation
and/or apply heat and cooling to skin.
COMPLIANCE: e contents of this “Instructions For Use” manual are exactly the same in both the printed and electronic forms.
2015/05/28 - Rev. 5
Inventory 5D00160
All Rights Reserved
ISO 13485
DNV NEMKO PRESAFE AS CERTIFIED
DQS CERTIFIED
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 2017II
Introduction to the Dynatron Solaris® Plus Series ................................................................................ 1
Summary of Features by Device ...............................................................................................................................................1
Language Selection ....................................................................................................................................................................3
Before You Treat a Patient .........................................................................................................................................................3
Installation and Features ........................................................................................................................ 4
Standard Components ...............................................................................................................................................................5
Dynatron Solaris® Plus Physical Features .............................................................................................. 7
Channels and Jacks ..................................................................................................................................................................13
Current Limit ............................................................................................................................................................................15
Lead Wires ................................................................................................................................................................................17
Why Is Target Better? ..............................................................................................................................................................32
Biphasic / Russian Parameters and Defaults.........................................................................................................................39
High Volt Instructions ......................................................................................................................... 42
High Volt Electrode Setup ......................................................................................................................................................42
High Volt Probe Treatment Setup ..........................................................................................................................................43
Detailed High Volt Setup ........................................................................................................................................................44
High Volt Quick Setup ............................................................................................................................................................44
High Volt Modality Information ......................................................................................................... 48
High Volt Waveform ................................................................................................................................................................48
High Volt Settings ....................................................................................................................................................................48
High Volt Default Settings ......................................................................................................................................................49
High Volt Waveform Specications .......................................................................................................................................49
Direct Current Instructions ................................................................................................................. 58
Direct Current Quick Setup....................................................................................................................................................58
Detailed Direct Current Setup ...............................................................................................................................................59
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 2017IV
TABLE OF CONTENTS
Direct Current Modality Information ................................................................................................. 61
Direct Current Probe erapy ................................................................................................................................................61
Direct Current Waveforms .....................................................................................................................................................61
Direct Current Warnings ........................................................................................................................................................62
Direct Current Default Setting ...............................................................................................................................................62
Head Temperature Hot Display .............................................................................................................................................74
Display Watts or W/cm2 ..........................................................................................................................................................74
Ultrasound Modality Information ....................................................................................................... 78
Selecting the Appropriate Soundhead ...................................................................................................................................78
Penetration of Ultrasound Waves ..........................................................................................................................................79
Types of Delivery......................................................................................................................................................................80
Treatment Time ........................................................................................................................................................................80
Frequency of Treatment ..........................................................................................................................................................81
Potential for Burns or Periosteal Pain ...................................................................................................................................81
Soundhead Optimization Adding or Replacing Soundheads ............................................................. 83
Soundhead Temperature Too Cold ........................................................................................................................................86
No Soundhead ..........................................................................................................................................................................86
Ultrasound Regulation and Technical Information ............................................................................................................87
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 2017V
Stim rough the Soundhead ................................................................................................................................................92
Modify A Treatment ................................................................................................................................................................94
Set Up A Second Treatment ..................................................................................................................................................102
Precautions and Warnings ....................................................................................................................................................114
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 2017VI
TABLE OF CONTENTS
Section IV: Technical Information
Dynatron Solaris® Plus Descriptions and Diagrams ......................................................................... 116
Dynatron® Solaris™ Plus Description ...................................................................................................................................116
Processor Interaction with the Keyboard and Displays ....................................................................................................118
Processor Interaction with Oscillators ................................................................................................................................118
Processor Interaction with the Output Jacks ......................................................................................................................119
Processor Control of Output Waveforms ...........................................................................................................................119
General Specications ........................................................................................................................ 120
Dynatron Solaris Plus Specications ...................................................................................................................................120
Safety Features of the Dynatron Solaris ..............................................................................................................................121
Care and Cleaning Instructions ...........................................................................................................................................121
Routine Ultrasound Calibration Inspections for Solaris Plus ..........................................................................................125
Denition of Symbols and Labeling ....................................................................................................................................126
Disposal of Equipment and Accessories .............................................................................................................................127
Save New Defaults ..................................................................................................................................................................128
Safety Features of the Dynatron Solaris PLUS ...................................................................................................................129
Lead Testing ............................................................................................................................................................................130
Battery Life ..............................................................................................................................................................................135
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 2017VII
TABLE OF CONTENTS
Electromagnetic Emissions and Immunity ....................................................................................... 142
Final Quality Check (QC) Checko Sheet ......................................................................................... 146
Dynatron Solaris® Plus Limited Warranty ......................................................................................... 150
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 2017VIII
INTRODUCTION TO THE DYNATRON SOLARIS® PLUS SERIES
Introduction
to the Dynatron
Solaris® Plus Series
e Dynatron Solaris Plus devices are both powerful and versatile. All channels allow fully-independent treatment setups,
oering Interferential, Premodulated, High Volt, Biphasic, Russian, Microcurrent, and xed frequency IFC/Premod. All
units, excluding the 707, also oer Direct Current. In addition, the 708 and 709 include Dynatronics’ Ultrasound Comboplus
feature with the power to deliver up to 5 channels of Stim and Ultrasound—all at the same time. e 708 and 709 Ultrasound
units also oer 1, 2, and 3 MHz frequencies for the greatest exibility in depth of treatment. Choose 1 MHz for deep treatments,
2 MHz for moderate depth, or 3 MHz for supercial depth. All of the Solaris Plus accessories are compatible with all
Solaris Plus devices.
Summary of Features by Device
Feature709708707706705
Electrotherapy
IFCXXXXX
PremodXXXXX
Fixed Frequency IFC/PremodXXXXX
BiphasicXXXXX
RussianXXXXX
High VoltXXXXX
MicrocurrentXXXXX
Direct CurrentXXXX
Combo Electrotherapy/UltrasoundXX
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 20171
INTRODUCTION TO THE DYNATRON SOLARIS® PLUS SERIES
Feature709708707706705
Tri-Wave Light Therapy
Ultrasound
XXXXX
XX
Available Channels
Electrotherapy Channels42442
High Volt Channel11111
MultiStim Probe Channel1111
Ultrasound Channel11
Combo Channel11
Tri-Wave Light Probe Channels11211
Tri-Wave Light Pad Channels11211
ThermoStim Channels11211
e Solaris Plus Series includes the standard advantages of Dynatronics’ engineering, such as customizable treatments, electrode
conductance meters, and the popular Target touchpad to move the center of interference directly to the site of the patient’s pain.
In addition all units oer the option of battery operation, making the devices truly portable. e manufacturer’s warranty for these
devices is two years (see full warranty details at the back of this manual).
is manual provides operator information and instructions for ve Solaris Plus models: the 705, 706, 707, 708, and 709.
e section that discusses Ultrasound and Combo treatments applies only to the Dynatron 708 and 709 Solaris Plus models.
All other sections of this manual apply to all Dynatron Solaris Plus devices excluding the 707 where special instructions
may apply.
Simplied Setup
e unique design of the Solaris Plus front panel means treatment setup has never been easier. A few simple key presses are
all you need to fully set up a treatment. e User Interface intuitively groups and displays all the options for a modality setup
on the large LCD screen to ensure that treatment parameters can easily be selected and adjusted.
Each modality oers default settings which are automatically preset when the modality is selected—saving time in the treatment
setup. You can change these defaults to match your own most common treatment setups reducing setup time to a matter of seconds.
WARNING: Power-on the device before attaching electrodes to the patient.
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 20172
INTRODUCTION TO THE DYNATRON SOLARIS® PLUS SERIES
Language Selection
e default language on the Solaris Plus Family of devices is English; however, both French and Spanish are also available. To
change the default language: 1) Begin at the START UP SCREEN. 2) Press the FUNCTION KEY. 3) Use the toggle key under
the LANGUAGE WINDOW to select the desired language. 4) Press STOP to return to the START UP SCREEN.
Before You Treat a Patient
Before administering a treatment to a patient with the Solaris Plus devices, you should familiarize yourself with all the
operating instructions for the modality used, as well as the contraindications, warnings, and precautions for that modality.
You should also read the general information about each of the modalities provided in this manual. In addition to this information,
consult other published sources for additional application and safety instructions regarding use of each type of therapy.
CAUTION: Device should be at room temperature prior to treatment.
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 20173
INSTALLATION AND FEATURES
Installation and Features
Unpacking
When you receive the unit, immediately unpack it and all accessories and check for possible damage, obvious or concealed.
In case of damage, immediately notify the freight carrier and take any steps necessary to le a claim for the damage sustained.
Do not destroy or discard the shipping carton. e carton should be reused if the device must be shipped for any reason,
including calibration. e carton is specially designed to protect the unit from shipping damage. Improper packaging of the
unit during transport can result in damage and invalidate the warranty.
Complete the warranty registration form located at the back of this manual and return it to Dynatronics within 30 days
of purchase. is is essential to insure you are not billed for services that are covered by the warranty policy. Warranty
registration should include serial numbers for both the device, probe, pads, and soundheads.
Connect the AC power cord, which is provided as a hospital grade, UL listed, plug to a properly grounded 110/120V 60 Hz
AC outlet. e device will automatically switch to 220/240V 50 Hz when connected to a power source with that voltage. e
power cord must also be rmly plugged into the device itself. When the cord is properly connected, it cannot be easily pulled
out. Do not place the cord or the device in a place where the cord could be tripped over or accidentally pulled out of its socket
during a treatment.
If Tri-Wave Light erapy Probe, Pads, or a ermoStim Probe are being used in conjunction with a Solaris Plus
device, they should be plugged into the Solaris Plus console prior to powering-on the device.
Read the operating instructions in this manual before proceeding with a treatment.
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 20174
INSTALLATION AND FEATURES
Standard Components
REF e following accessories are included with the Solaris Plus units:
Qty Part No. Description: One of the following devices plus acessories as listed:
1 D715 Solaris Plus 705
1 D716 Solaris Plus 706
1 D717 Solaris Plus 707
1 D718 Solaris Plus 708
1 D719 Solaris Plus 709
1 7B0241 Power Cord (black)
1 5D00090 Operator’s Manual
1 7B0268 Protocol Reference Manual for Electrotherapy & Ultrasound (J. Stephen Guey, P.T., Ed., D.)
1 7B0284 Ultra Polys™ self-adhesive electrodes 2” x 4” (5.08cm x 10.16cm) w/ pin connector (pkg. of 4)
1 DW248 2.5” x 48” (6.35cm x 121.92cm) straps (pkg. of 2)
1 7B0191 5” x 8” (12.7cm x 20.32cm) dispersive electrode for High Volt (gray)
1 7B0201 Sponge Fabric for use with 5”x 8” (12.7cm x 20.32cm) dispersive electrodes
Dynatron 705, 706, and 707
2 7B0232 120” (304.8cm) double leads (2 red) – Solaris Plus 706 and 707 only
2 7B0233 120” (304.8cm) double leads (2 black) - Solaris Plus 706 and 707 only
1 7B0230 72” (183cm) double lead (1 red) - Solaris Plus 705 only
1 7B0231 72” (183cm) double lead (1 black) - Solaris Plus 705 only
Dynatron 708 and 709 Ultrsound
1 7B0217 DynaGel Ultrasound Gel 100 ml sample
1 7B0234 Combo lead wires –Solaris Plus
2 7B0232 120” (304.8cm) double leads (2 red) – Solaris Plus 709 only
2 7B0233 120” (304.8cm) double leads (2 black) - Solaris Plus 709 only
1 7B0230 72” (183cm) double lead (1 red) - 708 only
1 7B0231 72” (183cm) double lead (1 black) - 708 only
Soundheads
e Solaris Plus devices may be purchased with one or more applicator soundheads in the following sizes:
Part No. Size Frequencies
DSH02 2 cm2 Operates at 1, 2, and 3 MHz
DSH05 5 cm2 Operates at 1, 2, and 3 MHz
DSH10 10 cm2 Operates at 1, 2, and 3 MHz
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 20175
INSTALLATION AND FEATURES
Optional Accessories
e following optional and replacement accessories may be purchased from Dynatronics or from your Dynatronics dealer:
7B0059 3” x 5” (7.62cm x 12.7cm) carbon electrodes (red)
7B0061 3” x 5” (7.62cm x 12.7cm) carbon electrodes (gray)
7B0067 1.5” x 2.0” (3.81cm x 5.8cm) carbon electrodes (red)
7B0069 1.5” x 2.0” (3.81cm x 5.8cm) carbon electrodes (gray)
7B0260 2” x 4” (5.8cm x 10.16cm) Ultra Polys™ adhesive electrodes (w/snap or pin)
7B0261 2” x 2” (5.8cm x 5.8cm) Ultra Polys™ square adhesive electrodes (w/snap or pin)
7B0077 Bifurcated extension lead wire for High Volt use
7B0082 Pin-to-Banana adapter (black)
7B0079 Banana-to-Pin Adapter (black)
7B0001 Snap adapter
5LTRGEL Ultrasound Coupling Gel (5 liter container)
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 20176
DYNATRON SOLARIS® PLUS PHYSICAL FEATURES
Dynatron Solaris® Plus
Physical Features
Before operating the Dynatron Solaris Plus devices, acquaint yourself with the control panel by reviewing the illustrations
and descriptions on the following pages. e numbered features in the diagrams correspond to the numbered descriptions.
Before administering treatment to a patient, read the sections later in this manual that provide specic instructions for
performing treatments, discussions of each modality, denitions of the available options, along with contraindications,
warnings, and precautions for all modalities.
Note: e User Interface on Solaris Plus devices is engineered with “CapSense Touch Technology” requiring that the user
make direct contact with the keys on the faceplate using dry, bare ngers or a glove with a conductive ngertip.
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 20177
DYNATRON SOLARIS® PLUS PHYSICAL FEATURES
1. START: Press the green START key on the right side of the Treatment Display Screen to start the treatment timer
and treatment proceeds as set up. For the Solaris Light and Microcurrent Probe treatments, the START key enables
the probe(s) in preparation for the treatment. e treatment begins aer the 1/0 (ON/OFF) key on the probe handle
is pressed.
e START key can also be used to save new treatment DEFAULT settings. Aer setting up a treatment, press
and hold the START key for two seconds. At the end of two seconds, a beep will sound indicating the treatment
parameters have been saved. e next time the modality is selected, these parameters will be selected automatically.
2. STOP: Pressing the red STOP key during a treatment IMMEDIATELY stops the output and sets the treatment time
to zero for all modalities. To stop the focus treatment only, reduce the focus treatment’s time to zero or press the
FUNCTION and STOP keys simultaneously.
For Light and Microcurrent Probe treatments, press the 1/0 (ON/OFF) key located on the probe handle(s) which
immediately PAUSES/STOPS/STARTS the treatment.
3. PAUSE: e PAUSE key is designed to pause Tri-Wave Light, Microcurrent, and Ultrasound treatments. Tri-Wave
Light and Microcurrent Probe treatments may also be paused by pressing the 1/0 (ON/OFF) key located on the
probe handle(s).
4. FUNCTION: is key is used to access unique features for High Volt, Ultrasound, Combo treatments and for
entering soundhead parameters. e FUNCTION key is also used in conjunction with the STOP key to stop only
a treatment in focus. In addition, the FUNCTION KEY provides access to settings for STIM, LANGUAGE, LEAD
TESTS AND SYSTEM INFORMATION. Specic instructions for using this key are provided later in the manual as
they apply to each function or modality.
5. TREATMENT DISPLAY SCREEN: Located in the upper center of the USER INTERFACE, the TREATMENT
DISPLAY SCREEN allows the clinician to view all of the parameters of the focus treatment such as time, intensity,
frequency, duty cycle, contraction rest, ramp time, polarity, or any other setting applicable to a treatment at a glance.
In addition, the screen lists all active modalities not in focus along with their active channels and remaining treatment
times in small font under the heading RUNNING TREATMENTS. If an error occurs during treatment, an error message
will appear on the Treatment Display Screen identifying the treatment modality that triggered the error message.
6. ARROW KEYS: e UP/DOWN arrow keys are used to increase/decrease the treatment time or other parameters
that appear on the TREATMENT DISPLAY SCREEN directly next to the arrow keys being used.
7. MODALITY KEYS: e Solaris Plus Series devices have all or a combination of 11 treatment modality options: IFC,
Premod, Light Probe, Light Pad, Ultrasound, Combo, Biphasic, Russian, High Volt, Microcurrent, and Direct Current.
MODALITY KEYS appear at the bottom of the USER INTERFACE. Pressing any of the available MODALITY KEYS
will bring the selected modality into focus and the default parameters for that treatment modality will be displayed.
Treatment modality parameters may be customized once the treatment is in focus.
8. TREATMENT WINDOWS: Across the bottom of the TREATMENT DISPLAY SCREEN are ve smaller treatment
windows providing treatment options and parameters that are unique to each modality. e quick access and
visibility of these TREATMENT WINDOWS allow for quick, easy, and accurate setup. On the following pages are
illustrations of each modality’s TREATMENT WINDOWS and their associated default settings.
Note: High Volt, Sound, and Combo treatments all have a secondary set of TREATMENT WINDOWS and treatment
options that are accessed when the treatment is in focus and the FUNCTION key is pressed. e arrow between the
two boxes indicates the secondary treatment window.
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 20178
DYNATRON SOLARIS® PLUS PHYSICAL FEATURES
IFC (Interferential)Premod
BiphasicRussian
MC (Microcurrent)DC (Direct Current)
LT PROBE (Tri-Wave Light Probe) LT PAD (Tri-Wave Light Pad)
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 20179
DYNATRON SOLARIS® PLUS PHYSICAL FEATURES
ermoStim (Heat)ermoStim (Cold)
SOUND (Ultrasound)
COMBO (Combination)
SOUND (Function Key View)
COMBO (Function Key View)
HIVOLT (Hight Volt)
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 201710
HIVOLT (Function Key View)
DYNATRON SOLARIS® PLUS PHYSICAL FEATURES
9. TREATMENT WINDOW TOGGLE KEYS: TOGGLE KEYS are located below the ve TREATMENT WINDOWS.
Pressing the Toggle Key directly below a window allows one to choose an output channel, and select treatment
parameters for the treatment in focus. A treatment is in focus when the name of the treatment appears in the center
of the TREATMENT DISPLAY SCREEN and the name of the modality is highlighted green in the channels window.
10. CHANNELS WINDOW / CHANGING THE FOCUS TREATMENT
One of ve TREATMENT WINDOWS, this window lists the available channels/jacks: 1,2,3,4, HV (High Volt, Probe
(MultiStim), Sound, LT Pad or LT PD1(2) indicates Light Pad, and LT Probe or LT PB1(2) Light Probe. Some
treatment jacks do not appear in the CHANNELS window until the accessory device is inserted into the Solaris Plus
console, for example Light Probe, Light Pad, ermoStim, and Sound.
e Channel Number/Indicator lights in the CHANNELS window identify which output channels/jacks are
currently in use. e channel(s)/jack illuminated in GREEN indicates the focus treatment and the time, intensity,
and other treatment parameters for that active treatment appear on the Treatment Display Screen. A Channel/
Indicator illuminated in YELLOW identies a channel/treatment is in use and delivering current, but is not in
focus. e intensity, and treatment parameters are not displayed at this time (only one channel’s parameters may
be displayed at a time). A treatment’s parameters may only be modied when the treatment is brought into focus.
To bring a treatment into focus, press the CHANNELS TOGGLE key below the CHANNELS window to select a
channel to be brought into focus. When the channel is in focus, the name will be illuminated GREEN. If a treatment
that is active but not the focus treatment times-out, the text in the CHANNELS window will change from YELLOW
to WHITE. A channel running in the background will change from YELLOW to ORANGE when the stim output is
in “rest” mode (Russian or Biphasic).
11. TARGET PAD: e Solaris Plus TARGET feature is engineered to precisely pinpoint a patient’s pain. As the nger
is glided across the TARGET touchpad, the patient identies the point of greatest pain. When the TARGET point is
identied, the nger is lied from the Target Pad and the selected TARGET is locked in place. TARGET is available
with Interferential or any four pad stim treatment.
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 201711
DYNATRON SOLARIS® PLUS PHYSICAL FEATURES
12. CONDUCTANCE/TEMPERATURE BAR
Conductance.
e Solaris Plus devices continuously measure conductance during electrical Stim treatments for Interferential,
Premod, and Microcurrent to ensure that the treatment outcome is optimal and to minimize the possibility of
patient discomfort due to poor conductance and/or changes in current density. As conductance is measured, Solaris
Plus displays the results in graph form on the CONDUCTANCE bar located on the right side of the TREATMENT
DISPLAY SCREEN. Optimum conductance is displayed as the conductance bar ows RED - YELLOW- GREEN.
GREEN indicating the best CONDUCTANCE. If the green bar only partially lls the graph area, the conductance
is at a percentage of optimum. Lower INTENSITY may cause the bar to partially ll, but does not mean that the
treatment is not eective. Below are some helpful denitions.
Conductance and Worn Electrodes.
Conductance is how readily electrical current is passed from the electrode to the skin surface during a treatment.
Conductance aects current density. A worn electrode that does not conduct the current evenly over its entire
surface will have “hot spots” where a greater amount of current ows through a smaller area which means the
current density is higher at that point than elsewhere on the electrode. “Hot Spots” can lead to patient discomfort.
Never risk patient comfort by using worn electrodes or lead wires.
Intensity.
e intensity level is a convenient incremental measurement. However, raising the intensity increases the current
delivered to the patient but does not improve conductance.
Current Density.
Current density is the amount of current that passes through a given area of the electrode. Current density varies
depending on the size of the electrode, the conductance, and intensity setting; and has an eect on patient comfort.
With proper setup and good accessories, current is dispersed evenly over the entire surface of the electrode. e
smaller the electrode, the greater the density of the current delivered through the area. To reduce current density and
improve patient comfort, use larger electrodes, or lower the intensity setting, or both.
During a Microcurrent probe treatment, the graph is also useful in observing conductance changes since the goal of
some Microcurrent treatments is to increase conductance (reduce resistance/impedance) at a given point.
If the number of green displayed segments begin to decrease on the graph during a treatment, it is important to
determine the cause of the poor conductance. Remember with poor conductance you may inadvertently increase
current density at a small point under the electrode and cause patient discomfort. Following are some considerations
to insure proper conductance.
• Check to be sure electrodes are not worn or that self-adhesive electrodes have not lost their adhesiveness. ese
are the most common causes of poor current delivery. Both self-adhesive and carbon electrodes eventually lose
their ability to conduct current eectively. See “Electrotherapy Usage Cautions” in this manual for recommended
intensity settings and usage limits.
• Check to ensure the entire surface of the poly adhesive electrode is adhering.
• Self-adhesive electrodes do not require sterilization, however, electrodes should be clean and hydrated (see
package instructions or “Self-Adhesive Electrodes” section of this manual).
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 201712
DYNATRON SOLARIS® PLUS PHYSICAL FEATURES
• Check to be sure the snap adapters haven’t fallen o or that the lead wire has not become disconnected from the
electrodes or the device.
• Make sure carbon electrodes have a secure connection with the pin ends of the leads. Over time the carbon
electrodes may become too loose to use safely and the electrodes must be replaced.
• Check for corrosion on lead ends.
• Make sure carbon electrodes are adequately moistened and free from build-up to allow complete contact across
the surface of the electrode.
• Observe the electrode placement. Some areas of the patient’s body conduct current better than others. In areas
where resistance is high you may be unable to obtain optimum conductivity.
• Check the dryness of the patient’s skin. Dry skin does not conduct current well.
• Check to see if the electrodes do not adhere properly when a patient shis position during a treatment. Worn
electrodes could become loose and a signicant change in conductance could result.
Temperature.
e Solaris Plus devices continuously measure temperature during a Light Probe, Light Pad, Ultrasound, and Combo
Treatment. TEMPERATURE is indicated by the length of the Blue/Green indicator lights on the temperature bar.
e longer the length of the colored bar, the higher the temperature. It is not uncommon to have the temperature bar
move into the medium length ranges. If the temperature of an ultrasound treatment approaches the maximum level
of 108° Fahrenheit (42.22° Celsius), the treatment is automatically PAUSED, output power stopped, and treatment
time stops counting down. Following a cooling period, the treatment may be continued by pressing START.
Channels and Jacks
13. Front Panel Channels and High Volt Jack
Illustrated below are the output channels for delivering Interferential, Premodulated, Russian, Biphasic, and
Microcurrent treatments. ese channels are located on the front of the device. As you face the device, channels 1
and 2 are on the le, channels 3 and 4 are on the right with the dedicated High Volt jack for delivering High Volt Pad
treatments in the middle. ree channel units (708 and 705) have channels 1, 2, and High Volt.
Front Panel Channels and Jack
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 201713
DYNATRON SOLARIS® PLUS PHYSICAL FEATURES
14. Le-Side Panel Jacks
Located on the le-side of the Solaris Plus are the “Keyed and Locking” Light Probe and Light Pad jacks. Once the arrows
are aligned at the top of both the jack and the connector, they will slide together smoothly and exactly. Do not force the
connector or damage to the pins may occur. is damage is not covered by warranty. When removing the connectors,
turn the connector’s outer sleeve in the direction of the arrow and the connector will slide easily away from the jack.
Le-Side Panel Jacks
15. Right-Side Panel Jacks
Located on the right-side of the Solaris Plus are the Ultrasound, Combo, and Stim Probe Jacks for the 709 and 708.
e right-side of the 707 mirrors the le-side diagram oering additional Light Probe and Light Pad Jacks, and
ermoStim Probe setup.
Right-Side Panel Jacks
Right-Side Panel Jacks
MultiStim Probe Jack.
e universal MultiStim probe plugs into this jack for Microcurrent, High Volt or DC probe therapy. e jack is an
eight pin DIN jack that requires careful alignment of the pins to prevent damage. Align the arrow on the top of the
connector with the notch at the top of the jack and gently slide the connector into place.
Combo Jack.
e special combo lead wire for combination treatments is plugged into this jack for a combination treatment setup
providing Stim output through the Ultrasound head. e Combination Treatment (Combo) Jack is a simple banana
jack connector and requires no special alignment.
Ultrasound Jack.
e Ultrasound Jack is a “Keyed and Locking” jack. Once the arrows are aligned at the top of both the jack and the
connector, they will slide together smoothly. Do not force the connector or damage to the pins may occur. When
removing the connector, turn the connector’s outer sleeve in the direction of the arrow and the connector will slide
easily away from the jack.
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 201714
DYNATRON SOLARIS® PLUS PHYSICAL FEATURES
16. Back Panel Jacks
b
Back Panel Jacks
c
a
a. Power 1/0 (ON/OFF) Switch. Located on the back of the unit this switch is labeled “1” and “0.” Set the switch
to “1” for ON; set the switch to “0” for OFF.
b. Batter y. is jack may be used to supply power to the device using an optional battery pack. More information
about the optional battery operation is provided later in this manual.
c. SD Card Input. e SD Input provides a way for the Solaris Plus Devices to receive soware updates quickly
and easily. Complete instructions for updating the devices using an SD card are found in the “Technical”
information section of the manual.
NOTE: Patient Remote Stop. Adding the Remote Stop requires a custom order. e Patient Remote Stop Jack is
located below the Light Probe holder. e remote stop is controlled by the patient during unattended therapy, allowing
the patient to stop the treatment at any time. When the button on the remote stop cable is pressed, output for all Stim
modalities and pad treatments is stopped. During Combo treatments, both Sound and Stim outputs are stopped.
Current Limit
e Dynatron Solaris Plus devices continuously measure the actual current output during a treatment and limit the output
current to the level indicated in “Technical Information” in this manual. As the intensity of a treatment is increased the
current output is also increased.
When the maximum output current limit is reached, the device will immediately stop increasing the intensity and
automatically reduce the intensity a few increments to prevent the possibility of patient discomfort. Simultaneously, the
device will beep and one of the following CURRENT LIMIT WARNINGS will appear in the lower right-hand corner of the
Treatment Display Screen. Following is a list of CURRENT LIMIT WARNINGS that may occur.
Remember to treat at the patient’s comfort level. It is not important to reach a given intensity level. It is only
important to set the treatment at a level that is comfortable to the patient. See “Electrotherapy Usage Cautions” in
this manual for suggested intensity limits.
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 201715
DYNATRON SOLARIS® PLUS PHYSICAL FEATURES
CURRENT LIMIT ERROR MESSAGES CAUSE
“Cannot start treatment with zero intensity”Intensity not set
Error 101, Error 111, Error 120, Error 130
“Lead error: current too low!
Please check or replace your leads and pads!”
Lead issue; electrode issue
Error 100, Error 110, Error 140
“Lead error: High current delivery detected. Adjusting intensity
to a safe limit. Please check leads. Space electrodes further
apart. Ensure skin is dry between electrodes.”
Electrodes touching or too close
Hot pack may be too moist
Lead shorted
Most warnings will occur during the setup portion of a treatment. It would be rare to encounter a current limit warning during
a patient treatment as reaching the current limit would require an intensity setting that is uncomfortable and intolerable to
most patients. Below are some possible considerations for exceptions:
• e patient is unable to adequately feel the current and is unable; therefore, to report discomfort at the high intensity level.
• When using four large electrodes for a treatment, current is dispersed over a larger electrode surface area permitting
a higher intensity setting without discomfort to the patient.
• For users who need to provide intensity levels above 50 mA (not available Japan or Canada), the default may be
changed to 100mA: 1) Begin at the START UP SCREEN. 2) Press the FUNCTION KEY. 3) Use the toggle key under
the MODE WINDOW to select the 100mA option and conrm your choice when prompted. Press STOP to return
to the START UP SCREEN.
As the intensity is increased, ensure that the patient feels the current as expected. If the patient is unable to feel the current,
the current could unintentionally be raised to a level much too high and risk causing unnecessary discomfort or possibly
burn the patient. Keep the intensity very low if the patient has little or no feeling in the treatment area (see “Contraindications, Warnings, and Precautions” in this manual). If you encounter the Current Limit Warnings, it may indicate that the
patient cannot adequately feel the current. Reduce the intensity immediately.
A wide range of factors can cause the patient to lack sucient feeling in the treatment area, including, but not limited to, pain
control drugs, use of ice packs, neurological damage, etc.
Always consider these and other factors when delivering an electrotherapy treatment. Determine intensity settings based
upon your medical expertise and judgment.
Error Messages
If an error occurs during any active treatment, whether in or out of focus, the Solaris Plus will sound a beep. A white box with
a red Error message will appear in the Time area of the Treatment Display Screen, if the treatment is in focus. If the treatment
error is associated with a treatment that is not in focus, “ERR” will appear next to the active treatment listed on the le-hand
side of the screen with other treatments that are currently running but not in focus. Pressing the modality key for the treatment
indicated will bring that treatment into focus and details regarding the error will appear on the Treatment Display Screen.
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 201716
DYNATRON SOLARIS® PLUS PHYSICAL FEATURES
Ultrasound Error Messages
ULTRASOUND ERROR MESSAGESCAUSE
Lead Wires
DID YOU KNOW?
• Lead wires should be replaced at least every six months
“No soundhead connected,
cannot setup ultrasound/combo treatment.”
“Soundhead is too hot!
Output has been disabled to allow cooling.”
“Caution: soundhead is getting hot!”Soundhead is getting hot
“Thermistor on soundhead is broken!
Please get soundhead replaced.”
No soundhead attached
Soundhead is too hot
Thermistor on soundhead is broken
• Carbon electrodes should be replaced approximately every six months
• Self-adhesive electrodes should be replaced aer no more than 15 uses
• You should never use monitoring electrodes nor ordinary TENS electrodes with this device
• Some brands of electrodes are of very poor quality or are inappropriate for electrotherapy. Your patient
may experience discomfort and even skin reaction due to poor distribution of current when using these
electrodes
• Failure to replace worn lead wires and carbon electrodes or using cheap, poor quality electrodes are some
of the most common causes of patient discomfort.
Even with good care, lead wires will eventually develop breaks (open connections) simply from normal usage, and must
be replaced about every six months. Damage can occur due to jerking or pulling on the wires, excessive bending or tight
wrapping the wires, or running over the wire with a device cart. When setting up treatments, keep lead wires out of areas
where a person could trip on them. When storing, lead wires should be loosely wrapped to prevent any kinking in the lead
wire. Never use worn or damaged leads to treat a patient. Using faulty leads may result in injury to a patient.
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 201717
DYNATRON SOLARIS® PLUS PHYSICAL FEATURES
Test Leads Daily
Lead wires should be tested regularly to ensure they are functioning properly and safely. A simple test performed with
the Dynatron Solaris Plus devices makes daily lead testing convenient. Damaged or worn leads should be discarded and
replaced. Instructions for testing are provided below.
Remove Corrosion From Lead Tips
Lead tips will build up corrosion through use. e lead tips must be cleaned and kept free of corrosion in order to function
correctly. To remove corrosion from lead tips, use steel wool to gently scrape o the corrosion. Take care not to scratch the
metal plating of the tip during cleaning. If the tip’s metal surface becomes pitted or uneven, the lead must be replaced.
Testing Leads
To test leads, perform the following steps daily.
1. Power on the Solaris Plus.
2. When the device has completed INITIALIZING, press the FUNCTION key located on the right side of the USER
INTERFACE to activate the SETTINGS screen.
3. Make sure that LEAD TEST is illuminated GREEN in the MODE window.
4. Using the LEAD TEST TOGGLE KEY, select ON in the LEAD TEST window. ON will be illuminated GREEN.
5. Plug a lead into Channel 1 (no other channel is used for the lead test). Remove snap adapters, if applicable, from the leads.
WARNING
UNDER NO CIRCUMSTANCES SHOULD THE LEADS BE CONNECTED TO A PATIENT DURING THIS TEST!
6. Press START.
7. Hold the pins securely together, move the leads around, wiggle the cord, especially at the jack end of the cord. e
numbers in the CONDUCTANCE window will begin to count up. e quality of the lead is represented on a rolling
scale of 0 to 250. e higher the number the better the lead’s quality. A count of 200 or more indicates the lead is
ready to be used. If the count registers under 100, the leads are probably bad and should be replaced.
8. Aer the test, remove the lead from Channel 1. If other leads need to be tested, plug in the next lead and test in the
same way.
9. To exit the LEAD TEST function, press the STOP key.
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 201718
DYNATRON SOLARIS® PLUS PHYSICAL FEATURES
NOTE: e LEAD TEST should be used for testing patient lead wires only. is is not an accurate means of testing
carbon electrodes. Contact Dynatronics Customer Service to arrange for free testing of carbon electrodes or for
instructions for testing these electrodes.
Carbon Electrodes
Carbon electrodes provide an economical means of delivering electrotherapy to patients but should not be used with
Microcurrent. is type of electrode lasts a long time and can be used again and again. However, if they are not properly
cared for, these electrodes can fail to deliver the desired treatment and can present the possibility for injury to a patient. To
ensure greatest safety and eectiveness with your treatments, follow these rules when using carbon electrodes.
1. Carbon electrodes must be well-moistened prior to treatment setup.
Dry carbon electrodes are very poor conductors of current and should NEVER be used. ey may be moistened
with either water or an electrolyte spray. Water is adequate for short treatments, but will evaporate too quickly for
longer treatments. If water is used for longer treatments, you may need to interrupt the treatment and remoisten
the electrodes. A special sponge fabric available with some carbon electrodes may be moistened well and used as a
conductive medium (do not use ordinary sponges for this purpose). Do not use Ultrasound gel as a conductive agent
with carbon electrodes.
If you use an electrolyte spray, this liquid may be diluted with equal amounts of distilled water, if desired. is
reduces the amount of build-up on the electrodes yet usually provides adequate moistening of the electrodes.
NOTE: As you increase the intensity to higher levels during setup, if your patient feels a “biting” sensation or if the
patient feels nothing, this indicates you are not getting adequate conductivity—the electrode may be too dry or is
not moistened evenly across its entire surface. Stop the setup and correct the problem.
2. Carbon electrodes must be free from any build-up.
If electrodes have a build-up from body oils or a moistening agent such as an electrolyte spray, conductivity is greatly
impaired. If treatment is allowed to continue, intensity could be inhibited. When using carbon electrodes with any
electrotherapy device, you must make sure conductivity is not impaired due to any type of build-up on the electrodes.
3. How to Clean Carbon Electrodes.
Carbon electrodes from Dynatronics may be cleaned using a mild soap and a small brush (such as a nail brush). To
sterilize, alcohol may be used. ey may also be sterilized in an Autoclave. Daily cleaning is recommended
If seeking a commercial cleanser/disinfectant, it is recommended that a product contain only the following active
ingredients to avoid damage to the probe or pads:
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 201719
DYNATRON SOLARIS® PLUS PHYSICAL FEATURES
OctylDecyl Dimethyl Ammonium Chloride
Dioctyl Dimenthyl Ammonium Chloride
Didecyl Dimenthyl Ammonium Chloride
Alkyl (C14 50%; C12 40%; C16 10
Dimenthyl Benzyl Ammonium Chloride
Other Ingredients not published
4. Carbon electrodes eventually wear out.
Do not assume you can safely use carbon electrodes indenitely. Over time these electrodes will wear; and when worn,
the amount of current delivered through the electrode will decrease and will be inconsistent over the surface of the
electrode. As a general rule, carbon electrodes that are used regularly should be replaced at least every six months.
Do not take chances with patient safety!
Discard worn carbon electrodes!
If you think that your carbon electrodes are showing wear, you can test them with an ohm meter. Good carbon electrodes
should measure resistance between 40 and 200 ohms.
Self-Adhesive Electrodes
Dynatronics’ self-adhesive electrodes are intended for multiple but patient specic use due to the danger of cross
contamination. Improper use of the electrodes can decrease the life of the electrode and could even result in harm to your
patient. e following instructions will help you achieve maximum usage from your electrodes while ensuring patient safety
and comfort during treatment.
1. Make sure the electrode is adhering and making contact with the skin across the entire surface of the electrode.
Electrodes will lose their adhesive quality when exposed to air, dust, dry skin, etc.
To Retain Adhesiveness:
• Electrodes should be stored in a tightly sealed pouch until used.
• e patient’s skin should be thoroughly cleaned and free from oils or akiness prior to placing the electrodes.
To Restore Adhesiveness:
• Before a Treatment. Before placing the electrode on the patient, moisten the patient’s skin with a damp cloth
using plain water, then apply the electrode to the skin.
• Aer a Treatment. Apply one or two drops of water to the adhesive side of the electrode using plain water, rub
it lightly with ngertips, reapply the electrode to its plastic backing, and seal it tightly in its storage pouch. Do
not use an electrolyte spray to remoisten self-adhesive electrodes as this substance can destroy the adhesive.
Self-adhesive electrodes do not require sterilization.
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 201720
DYNATRON SOLARIS® PLUS PHYSICAL FEATURES
• With this method of re-hydration, aer a couple of hours electrodes can regain up to 90 per cent of their original
adhesive quality.
2. NEVER use a self-adhesive electrode for more than 15 treatments (maximum).
3. NEVER USE STRAPS, WEIGHTS, or other devices to attach self-adhesive electrodes to the skin. If an electrode
has lost its adhesive quality, you can use one of the methods given above to re-hydrate the adhesive, or you should
discard the electrode. Using straps and weights with self-adhesive electrodes could have an unpredictable eect on
the electrodes and could cause injury.
4. NEVER use monitoring electrodes such as ECG, or EMG, nor ordinary TENS electrodes.
5. If you see the “No Patient Current” screen message, or if you observe poor conductivity indicators, check the
electrodes and lead wires for proper connection.
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 201721
ELECTROTHERAPY INFORMATION AND USAGE CAUTIONS
Electrotherapy
Information and
Usage Cautions
e following general cautions are to be observed during Interferential, Premodulated, Russian, Biphasic, High Voltage, and
Direct Current stimulation. For Microcurrent electrotherapy, see additional cautions in the Microcurrent Section of this manual.
WARNING
• NEVER turn the power ON or OFF while the unit is connected to the patient.
• Always STOP a treatment before removing or attaching electrodes or leads. Leads and electrodes must
only be applied to the patient before a treatment is started.
• Never use worn or damaged leads or electrodes as these may result in injury to the patient.
• See the Contraindications, Warnings, and Precautions for Interferential and Premodulated treatments in
this manual before administering a treatment.
• Additional warning from the Canadian Health and Welfare Department, Health Protection branch:
WARNING: oracic applications are contraindicated. Cardiac brillation may occur if output current is
50mA RMS or greater for any output circuit. (For use in Canada and Japan, this device is limited to 50mA
output).
Electrical stimulation, by its very nature, has the ability to irritate the patient’s skin. Certain precautions should be observed
to assure maximum safety and comfort for patients. A patient’s tendency to have adverse reactions is dependent upon several
factors. ese factors are:
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 201722
ELECTROTHERAPY INFORMATION AND USAGE CAUTIONS
Current Density
is is the amount of current being delivered to the patient divided by the area through which the current is being delivered
(the surface area of the electrodes being used).
Electrode Condition
Worn or dried out electrodes cause the current to concentrate in small areas of the electrode instead being evenly distributed
over the entire surface of the electrode. is has the eect of concentrating and increasing the current density into small
areas.
Patient Susceptibility
Some patients’ skin is more sensitive to electrotherapy currents. is can cause a reaction similar to a heat rash.
Electrotherapy treatment can result in a rash, burn, or blister. e tendency to do this is dependent upon the factors listed
above and can be minimized by applying the following guidelines:
1. Use only moderate current
It is not always necessary to raise the treatment intensity to just short of the patient’s pain threshold to achieve
adequate results. Below is a chart comparing the size of the self-adhesive and carbon electrodes with their suggested
maximum intensity levels.
NOTE: e intensity settings should be considered maximum and not target intensities. ese suggested settings
apply to Interferential and Premodulated treatments. For High Voltage pulsed stimulation the intensity is displayed
in volts; therefore, these suggested settings do not apply.
For Biphasic or Russian stimulation treatments intended to eect a muscle contraction, it may sometimes be
necessary to exceed these recommended limits to achieve the desired results. However, use caution when doing so to
ensure that the patient can feel and can comfortably tolerate the electrical current. Also observe all other precautions
in this section concerning leads and electrodes to ensure the higher intensity setting is not necessary as a result of
defective accessories. In any case, do not exceed patient tolerance in setting the intensity. Consult published medical
literature for more information about treatment protocols using each of these electrotherapy modalities.
Use as large an electrode as is practical for the application.
NOTE: e current density in a 1.25" square electrode is over FOUR TIMES the current density in a 1.75" by 3.75"
electrode for the same intensity setting. Using larger electrodes allows current to be delivered over a larger area of
the body keeping the current density as low as possible and minimizing the possibility for adverse reactions. Below
are recommended intensities that correspond to electrode sizes.
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 201723
ELECTROTHERAPY INFORMATION AND USAGE CAUTIONS
Interferential / Premodulated
Electrode SizeMaximum Recommended Intensity
Carbon
Electrodes
1.75" x 3.75" (4.45cm x 9.53cm)25 - 30
Self-adhesive
Electrodes
When delivering combination Ultrasound and Stim treatments where the Stim current is delivered through the
soundhead, the following are the recommended maximum Stim intensities (refers to Premodulated, Biphasic or
Russian stimulation only):
3" round (7.62cm)25 - 30
3" x 5" (7.62cm x 12.7cm)30 - 40
1.75" square (4.45cm)10 - 15
1.25" round (3.18cm)10 - 12
2" round (5.08cm)10 - 20
3" round (7.62cm)25 - 30
Combination Treatment
Ultrasound Head SizeMaximum Recommended Intensity for Electrotherapy
2 cm2 Head 4 - 7
5 cm2 Head10 - 15
10 cm2 Head15 - 20
2. Ensure that the area on the patient’s skin where the electrode is to be placed is clean and free of all foreign matter.
Includes powders, perfumes, as well as body oils, dirt, and grime. Cleaning with an alcohol wipe should be adequate.
Allow the alcohol to fully evaporate before applying the electrodes. Iontophoresis occurs with all electrical current therapies
and can drive any of the above-surface contaminants below the epidural layer where an allergic reaction may occur.
Any electrode which is suspect should be discarded. It’s not worth the price of an electrode to risk harming a patient.
3. Make sure the electrodes being used are in good condition.
e poly adhesive electrodes should have good adhesion over the entire surface area of the electrode. e area where
the leads attach to the electrode (either through a lead or a snap) should not be damaged such that the connection to
the foil backing behind the adhesive is broken. Carbon electrodes should be deep black and should be free of cracks
in the electrode surface.
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 201724
ELECTROTHERAPY INFORMATION AND USAGE CAUTIONS
4. Some patients tend to be much more sensitive to electrotherapy treatments.
On patients with this tendency, treat with reduced intensity and/or shorter treatment times with possibly more
frequent treatments, if required. Most reactions are localized and very short-lived, so limiting the exposure should
minimize any potential for adverse reactions.
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 201725
INTERFERENTIAL / PREMODULATED INSTRUCTIONS
Interferential / Premodulated
Instructions
An Interferential treatment uses two channels and four electrodes (channel pairs 1-2 or 3-4). e device will automatically
select the rst available channel pair when you select IFC. A Premodulated treatment uses one channel and two electrodes.
e device will automatically select the rst available channel (1, 2, 3, or 4) when PREMOD is selected. If desired, multiple
treatments can be setup using available channels. Note: Channels 3 and 4 are only available on Solaris Plus 706,
707, and 709.
Interferential / Premod Quick Setup
1. Choose IFC or PREMOD.
• Plug the patient lead(s) into the output jack(s) for the channel(s) selected.
• Attach electrodes to patient at treatment site.
2. Set the treatment TIME.
3. (Optional) CUSTOMIZE FREQUENCIES now.
Select RANGE: HIGH, LOW, ALT (high/low alternating), CONSEC ( ½ high, ½ low), or FIXED (select
4,000 to 10,000 Hz).
4. Choose TARGET, SWEEP, or STATIC (for Interferential treatments only) using the TARGET TOGGLE key.
5. Increase INTENSITY (patient will feel the current) using the arrow keys located next to the INTENSITY display.
6. If TARGET is selected, use the target pad to focus therapeutic beat to desired site.
7. Press STA RT .
8. STOP. Press the FUNCTION AND STOP keys simultaneously to stop only focus treatment appearing on
the Treatment Screen. Pressing STOP alone, stops all treatments operating on the device. Using the TIME
arrow keys to bring a treatment time to zero will also stop a treatment.
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 201726
INTERFERENTIAL / PREMODULATED INSTRUCTIONS
NOTE: Prior to increasing intensity, electrodes must be placed on the patient and the lead(s) attached to the device.
Plug the lead(s) into the channel(s) the device selects for this treatment. Consult published sources for electrode
placements, treatment settings, and treatment times. Make sure electrodes make good contact with the patient’s skin
over the entire surface area of the electrode. Improper electrode contact may result in patient injury.
Detailed Interferential / Premodulated Setup
1. IFC or PREMOD MODALITY keys
Press the IFC or PREMOD MODALITY keys to choose IFC (Interferential) or PREMOD (Premodulated). When
you choose IFC, two channels are automatically selected: 1-2 or 3-4. When choosing PREMOD, the rst available
single channel is selected. Make sure the patient lead(s) is plugged into the correct jack(s) for the channel(s) selected.
e default settings for the modality are automatically selected. If you wish to use the default settings, increase the
intensity to the desired level, and press START.
Target (IFC): .......................................... On
If you wish to change or customize the treatment settings, proceed through the following steps:
2. Customize TIME
e default treatment time is displayed at 10:00 min. Use the TIME arrow keys to increase or decrease the treatment time.
3. Customize FREQUENCY (optional)
Default HIGH and LOW Frequency Settings
• HIGH range is 80 to 150 Hz.
• LOW range is 0 to 10 Hz.
• ALTERNATING range alternates every 30 seconds between HIGH and LOW, beginning with LOW.
NOTE: ALTERNATING and HIGH/LOW options may be selected aer pressing START. However, Consecutive
and Fixed options must be selected before pressing START.
• CONSECUTIVE HIGH/LOW. During the rst half of the treatment time the High frequency range is delivered.
During the second half of the treatment time the Low frequency range is delivered.
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 201727
INTERFERENTIAL / PREMODULATED INSTRUCTIONS
NOTE: Make any desired changes to the treatment time before selecting Consecutive. Treatment time changes
made aer selecting Consecutive will cause the treatment to revert to an ALTERNATING HIGH/LOW treatment.
• FIXED. A FIXED treatment automatically defaults to a STATIC setting. Press the toggle key under MODE until
FREQ is illuminated. A FIXED treatment can be set between 4,000 and 10,000 KHz. Selections progress from
4,000 KHz to 10,000 KHz in 1,000 KHz increments. Once the FIXED rate is set, return to the TIME display by
using the MODE toggle key.
NOTE: If INTENSITY has been set during the setup of a FIXED treatment option and a change is made to any
other treatment option, INTENSITY will automatically return to “0.”
e default High and Low frequency settings may be changed for a single treatment if desired, or new default settings
may be saved to apply to all future treatment setups.
• Press the toggle key under the MODE display. Select FREQ (Frequency).
• Press the toggle key under the RANGE display. Select a HIGH or LOW frequency range.
• e HIGH Frequency will appear on the le-hand side of the Treatment Display Screen. e LOW Frequency
will appear on the right-hand side of the Treatment Display Screen. Changes to the upper and lower limits are
made using the arrow keys to the side of each displayed setting. If you set both displays to the same value, the
treatment will be delivered at that single frequency rather than sweep through a frequency range.
• Aer pressing START, frequency settings will remain in eect for the duration of the treatment. If you save defaults
during this treatment, the new frequency settings you have entered become the defaults for this modality. However,
if you do not save the new settings, the unit will return to the current default settings for the next treatment.
• To SAVE these CUSTOM FREQUENCY settings, hold the START key down until a beep sounds. Once saved,
these settings will be used on all IFC or Premod treatments that follow.
• Press the toggle key under the MODE display to return to the TIME display, Aer 10 seconds with no key
presses, the Time display will automatically return.
4. Choose TARGET, SWEEP, or STATIC (for Interferential only)
Using the toggle key located under the TARGET display make your selection from the three available options.
TARGET. Pinpoint the treatment site delivering the full Interferential current where it is needed.
NOTE: e intensity must be set before using the Target Pad so the patient will be able to indicate when the
treatment site is found. Also remember, an injured area will oen be more sensitive to the current delivered.
erefore, an intensity setting that is comfortable to the patient at rst may feel uncomfortable when the treatment
site is found using the TARGET PAD. If necessary, reduce the intensity to the patient’s comfort level.
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 201728
INTERFERENTIAL / PREMODULATED INSTRUCTIONS
SWEEP. e interferential current randomly sweeps the treatment area within the electrodes allowing the general
area to be bathed with Interferential current.
STATIC. e interferential current focuses only on the point where the current between electrodes intersects as it
follows the path of least resistance.
5. INTENSITY
Press the UP/DOWN arrow keys located next to the INTENSITY display on the right-hand side of the Treatment
Screen. When the INTENSITY is increased, current to the patient begins. NOTE: e CONDUCTANCE BAR
GRAPH will be operable at this time.
Before selecting the intensity setting for an individual patient, see “Electrotherapy Information and Usage Cautions”
in this manual for recommended intensity settings. Also see the section of this manual entitled “Contraindications,
Warnings, and Precautions” for specic precautions when treating any conditions contributing to loss of sensation,
or any time the patient cannot feel the electrical stimulation.
6. Press START
When you press start, the treatment timer in the TIME display window begins counting down and the treatment
proceeds. Remember to set the intensity before pressing START.
7. SAVE DEFAULTS
If the treatment you have just set up is a frequently used, you can save the treatment parameters as new defaults by
pressing and holding the START key until a beep sounds indicating the treatment parameters have been saved. e
next time you select the modality, these parameters will be selected automatically.
8. MODIFY SETTINGS
Treatment settings can be modied while the treatment is in progress except for a CONSECUTIVE HIGH/LOW
treatment. If the TIME setting on a CONSECUTIVE HIGH/LOW treatment is altered, the treatment will be aborted
and default to an ALTERNATING HIGH/LOW treatment.
• FREQUENCY RANGE. Use the Range Toggle key to select a dierent frequency option (High, Low, Alternating
High/Low, Consecutive High/Low, or Fixed). Fixed Frequency and Consecutive High/Low cannot be selected
aer treatment is started.
• TARGET/SWEEP/STATIC. Use the Target Toggle key to select Target, Sweep, or Static (for IFC only).
• TIME. Use the Time Arrow keys to increase or decrease the treatment time.
• INTENSITY. Use the Intensity Arrow keys to increase or decrease the intensity.
• TARGET. Relocate the treatment site by touching the TARGET PAD at any time during the treatment when
TARGET has been selected.
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 201729
INTERFERENTIAL / PREMODULATED INSTRUCTIONS
9. STOP.
When the treatment time has elapsed, the current to the patient stops and a tone sounds signaling the end of a
treatment. Treatments in progress may be stopped at any time using one of the following methods.
Stop One Treatment Only. Press and hold the FUNCTION key and press STOP. is stops only the treatment in focus.
Stop All. Press the STOP key. All treatments at all channels will stop.
Stop Time. Reduce the treatment time using the Time arrow key. e output at the selected channel is stopped,
and the device then displays the parameters for the next treatment that remains in progress (if there is an active
treatment).
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 201730
INTERFERENTIAL AND PREMODULATED MODALITY INFORMATION
Interferential and Premodulated
Modality Information
Interferential (Quadpolar) Therapy
Interferential therapy uses four electrodes to deliver two currents, one current with a constant frequency of 4000 Hz and the
other current with a variable frequency of 4000 to 4150 Hz. e paths of these two currents cross resulting in a “beat” that
produces the therapeutic frequency at the treatment site.
e resulting frequency is between 1 and 150 Hz. An example of wave forms representing these currents is illustrated here.
In the Interferential mode, two output jacks (Channels 1 and 2, or 3 and 4) are utilized with four electrodes placed in a
crisscross fashion, “bracketing” the treatment site. e output from Channel 1 (or Channel 3) is the constant 4000 Hz wave,
while the output of Channel 2 (or Channel 4) is the variable 4000 to 4150 sine wave.
e “beat” phenomenon. Two waves of dierent frequencies over 4000
Hz, combine to produce a beat which is between 1 and 150 Hz.
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 201731
Stimulation produced by 4 electrodes.
INTERFERENTIAL AND PREMODULATED MODALITY INFORMATION
Premodulated (Bipolar) Therapy
Premodulated therapy utilizes one output jack and two electrodes. e current delivered is a composite wave form. In order
to produce this composite current, two Frequencies are “mixed” within the device prior to output. One frequency is 4000 Hz
while the second frequency covers a range between 4000 to 4150 Hz.
With the Dynatron Solaris Plus devices, any of the four channels may be used simultaneously to deliver up to four separate,
independent Premodulated treatments. A crisscross electrode setup pattern should not be used when setting up multiple
Premodulated treatments. Note that a Premodulated treatment usually requires a lower intensity setting than an Interferential
treatment since current is dispersed to only two electrodes rather than four (a smaller total coverage area means greater
current density at the treatment site).
Target
e Dynatronics’ TARGET (available for Interferential treatments
only) simplies placing the interferential beat directly on the
treatment site. e movement of the nger on the Target paid along
with the feedback supplied by the patient allows the user to place the
full force of the interferential “beat” directly on the treatment site
regardless of conductance variations caused by dierences in human
tissue (skin, muscle, bone, etc.) that, without Target, make placing
the Interferential “beat” a guessing game. Target eliminates the need
to move the electrodes to achieve the desired result.
In the diagram the electrodes are placed in a position to treat a
point directly in the center of the electrodes. But the center of
interference actually occurs at another point. In these illustrations
the point of interference is shown in a cloverleaf shape as
Interferential treatment aects a cloverleaf-shaped area. Using the
Target pad, the point of interference in easily moved to the desired
treatment area. Other devices increase current at one channel
while decreasing it at the other. is merely rotates the treatment.
e center of interference does not move.
Normal Interferential currents cross at a point between
electrodes. It is dicult to guess where they will cross as
shown in the le diagram. With Target, you move the point
of interference without moving electrodes (as shown in the
diagram on the right)—just press the touch pad.
Why Is Target Better?
With Target, the voltage output from both channels remains equal
at all times; so wherever the treatment is applied, a full, deep
Interferential beat occurs. Other devices attempt to achieve this
eect by increasing the current from one channel while decreasing
the current from the other channel. is method only rotates the
cloverleaf-shaped area, but the center of interference does not
move. In addition, the depth of the beat is reduced.
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 201732
Other devices increase current at one channel while
decreasing it at the other. is merely rotates the
treatment. e center of interference does not move.
INTERFERENTIAL AND PREMODULATED MODALITY INFORMATION
Target Sweep
e Sweep option literally moves the point of interference inward and outward in a somewhat spiral pattern, bathing about
80 percent of the area within the electrodes with the Interferential current. Sweep utilizes the Target feature and moves the
point of interference to cover a wider treatment area while still retaining the full Interferential beat. Sweep bathes a larger
area with the Interferential current.
Interferential Electrode Placement
When performing Interferential therapy with a two-channel or four-electrode setup, it is
important to arrange the electrodes in a crisscross manner so the current from one channel
will intersect with the current from the second channel at the point where treatment is to be
delivered. Consult published literature for electrode placements for specic sites and conditions.
Interferential Electrode Placement
Interferential / Premodulated Default Settings
e following default settings are set by the manufacturer and are selected when you select IFC or PREMOD. You may
change these defaults to your own preferred settings.
Interferential Default Settings
• High 80-150 Hz
• Target
• e rst available channel pair (1-2 or 3-4)
• Time: 10 minutes
• Frequency Ranges: High 80-150 Hz; Low 0-10 Hz
Premodulated Default Settings
• High 80-150 Hz
• Target
• e rst available channel pair (1 through 4)
• Time: 10 minutes
• Frequency Ranges: High 80-150 Hz; Low 0-10 Hz
• Interferential/Premodulated erapy: 4000 Hz sine wave frequency modulated by a 4000 to 4150 Hz variable
frequency sine wave of equal amplitude
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 201733
BIPHASIC / RUSSIAN INSTRUCTIONS
Biphasic / Russian
Instructions
In the Russian and Biphasic Stimulation modes the output of the device is a pulsed sinusoidal wave. Solaris Plus allows the
operator to choose a muscle contraction/rest cycle that is most suited for the individual patient and for the desired treatment.
Once the cycle is chosen, each muscle-stimulating burst is followed by a rest cycle. See “Russian/Biphasic Parameters” in this
manual for further discussion of pulse rate and duration, and illustrations showing the segments of the Russian Stimulation
cycle and the Biphasic Stimulation cycle.
Solaris Plus provides four treatment options in Russian and Biphasic Stimulation: Normal, Reciprocal, Co-Contraction, and
Custom. Aer deciding which treatment is to be used, attach the appropriate number of leads required to set up the treatment.
NORMAL: Use one channel with one lead wire (two electrodes). Place the electrodes so as to treat through the muscle. e
contraction/rest cycle is selected from an option list of 10/10, 10/30, 10/50, CONT (continuous/no rest cycle), and Custom.
Each time period is indicated in seconds. For example, 10/30 indicates 10 seconds of stimulation with 30 seconds of rest. e
continuous duty cycle is not recommended for electrical muscle stimulation, but may be used for settings that are intended
to eect results other than a muscle contraction.
CO-CONTRACTION: Use two channels and two lead wires (four-electrodes) for this treatment. Each pair of electrodes is
placed over a dierent muscle group. is treatment res the two muscle groups simultaneously—contraction and rest cycles
for both treatment areas occur at the same time. Two channels are required (1-2 or 3-4). Note: 3-4 are available only on the
Solaris Plus 706, 707, and 709.
RECIPROCAL: Use two channels and two lead wires (four electrodes) for this treatment. e reciprocal muscle stimulation
res two muscle groups (such as reciprocal exors/extensors) one aer the other. For example, with a duty cycle of 10/30, the
device would deliver stimulation for 10 seconds to the rst muscle, followed by 10 seconds of stimulation to the reciprocal
muscle. A 30-second rest time follows each stimulation. e timing of the two cycles will overlap (the rst muscle group is
stimulated aer a 30-second rest, even though the second muscle group is 20 seconds into its rest cycle). e Continuous
contraction/rest cycle is not available for Reciprocal treatments. Two channels are required (1-2 or 3-4). Note: 3-4 are available
only on the Solaris Plus 706, 707, and 709.
CUSTOM CONTRACTION: e Custom Contraction/Rest cycle feature allows the treatment to be customized by selecting
from a Custom Contraction ON time (1 to 60 seconds), and an OFF time (1 to 90 seconds). Treatments can be customized
by using the following steps:
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 201734
4. Using the RAMP toggle key, select RAMP time (.05, 1.0, 1.5, 2.0).
5. Set CUSTOM CONTRACTION using the arrow keys on the le-side of the Treatment Display Screen (1-60 sec.)
6. Set CUSTOM REST using the arrow keys on the right-side of the Treatment Display Screen (1-90 sec.). e REST time
can be set to a minimum of one second.
7. Press START.
Biphasic / Russian Quick Setup
1. Choose BIPHASIC or RUSSIAN.
2. Choose the TREATMENT (Normal, Co-contraction, or Reciprocal using the TREATMENT toggle key).
Plug the patient lead(s) into the output jack(s) for the channel(s) selected.
3. Choose the CONTRACTION/REST times by pressing the CONTRACTION/REST toggle key.
4. Choose the RAMP setting by using the RAMP toggle.
5. Change the treatment TIME, by pressing the UP/DOWN TIME ARROW keys, if desired.
6. Change the PULSE and DURATION. Using the MODE toggle key select RATE/DUR (rate/duration).
• Press the PULSE RATE arrow keys to change the PULSE RATE
• Press the PULSE DURATION arrow keys to change the PULSE DURATION
7. Press the MODE toggle key to return to the TIME display window.
8. Raise the INTENSITY to the desired level by pressing the INTENSITY ARROW keys on the right-side of
the Treatment Screen.
9. For co-contraction or reciprocal treatments, set the intensity for the rst channel. Press START. e device
will automatically select the second channel. Set the intensity for the second channel.
10. Press START.
11. STOP. Press the FUNCTION and STOP keys simultaneously to stop only the focus treatment appearing in
the Treatment Screen. Pressing STOP alone, stops all treatments operating on the device. Using the TIME
arrow keys to bring treatment time to zero will also stop a treatment.
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 201735
BIPHASIC / RUSSIAN INSTRUCTIONS
Detailed Biphasic / Russian Setup
If you do not understand the terms contraction, rest, ramp time, pulse duration, or pulse rate; consult the diagrams in the
section of this manual entitled “Biphasic / Russian Parameters.”
1. Press the BIPHASIC or RUSSIAN
When you select this modality, the default settings are automatically selected. If you wish to use the default settings,
you can now increase intensity to the desired level and press START.
Default Setting
Mode............................................... Normal
Cont/Rest Time ................................. 10/30
Time ..........................................10 minutes
Ramp Time .................................... 0.5 sec.
If you wish to change the treatment settings, proceed through the following steps:
2. Choose the TREATMENT.
Use the TREATMENT toggle key to select NORMAL, CO-CONT (Co-contraction), or RECIP (Reciprocal). Connect
the patient lead wire(s) to the channel(s) selected.
3. CONTRACTION/REST
Cycle Times Using the CONT/REST toggle key, choose the CONTRACTION/REST cycle times. Available options
include 10/10, 10/30, 10/50, Continuous, and Custom. e setting of 10/30, for example, means a 10-second
contraction time followed by a 30-second rest time. Note that you may not select Continuous cycle for a Reciprocal
treatment. e Continuous duty cycle is not recommended for electrical muscle stimulation, but may be used for
settings that are intended to eect results other than a muscle contraction.
4. Choose the RAMP Setting (does not aect Continuous treatments).
Press the RAMP TOGGLE key one or more times to select the desired ramp time. Available options include .5, 1.0,
1.5, and 2.0 seconds. e ramp time is applied to both the start and to the end of the contraction time. e ramp time
is in addition to the contraction time itself.
5. Change the treatment TIME (optional).
e default time is displayed. Use the TIME arrow keys to increase or decrease the treatment time.
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 201736
BIPHASIC / RUSSIAN INSTRUCTIONS
6. Change the PULSE DURATION and/or PULSE RATE (optional)
Press the MODE toggle key to select RATE/DUR (Pulse Duration). e pulse DURATION (width) and RATE may
be modied for each channel pair (1-2 and 3-4).
Press the PULSE RATE arrow keys to the le-side of the Treatment Screen to change the PULSE RATE. Press the
PULSE DURATION arrow keys to the right-side of the Treatment Screen to change the PULSE DURATION. Press
the MODE toggle key and select TIME to return the TIME display. However, if you make no key presses for 10
seconds, the display automatically returns to the TIME display. e ranges and default settings for pulse duration
(width) and pulse rate are listed later in this section.
7. Raise the INTENSITY to the desired level.
Set the intensity by pressing the INTENSITY arrow keys located on the right-side of the Treatment Display Screen.
Increasing the intensity sends current directly to the patient. Intensity can be decreased by pressing the down arrow key.
Before selecting the intensity setting for an individual patient, see section entitled “Electrotherapy Usage Cautions”
for recommended intensity settings. Also see the section of this manual entitled “Contraindications, Warnings, and
Precautions” for specic precautions when treating any conditions contributing to loss of sensation, or any time the
patient cannot feel the electrical stimulation.
e intensity, pulse rate, and pulse duration must all be considered together when setting up the treatment as all
three factors aect patient comfort. It may be necessary to adjust one or more of these parameters somewhat aer
the initial settings are selected to nd the best settings for a given treatment and patient.
8. For co-contraction or reciprocal treatments, select the SECOND CHANNEL and set the INTENSITY.
Intensity is set for each channel separately. For co-contraction or reciprocal treatments, Set the intensity for the
rst channel. Press START. e device will automatically select the second channel. Set the intensity for the second
channel. When setting intensity, only the channel with the solid GREEN light is aected. e rst channel will be
illuminated in YELLOW.
9. Press START
When you press START the treatment timer begins counting down. If the intensity is not set before pressing START,
the treatment will not begin until the intensity is set. For Reciprocal and Co-contraction treatments, the intensity
must be set for each channel separately.
10. SAVE DEFAULTS
If this treatment setup is the most common Biphasic or Russian treatment setup you use, save the treatment
parameters as your defaults. Aer setting up the treatment, press and hold the START key until a tone sounds
indicating the treatment parameters have been saved. e next time you select this modality, the saved parameters
will be selected automatically.
11. MODIFY Settings
While the treatment is in progress, the treatment settings can be modied. Carefully observe the channel indicator
lights when modifying a treatment. When a channel’s light is illuminated GREEN, the current treatment parameters
for that channel are displayed. Any changes made to the parameters will aect only the channel that is illuminated in
GREEN. Use the toggle key to display the parameters of another channel in order to modify parameters.
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 201737
BIPHASIC / RUSSIAN INSTRUCTIONS
During a Biphasic or Russian treatment you may make the following modications:
• CONTRACTION/REST cycle.
• RAMP TIME
• TREATMENT TIME
• RATE/DURATION (not available for Reciprocal treatments nor when two “Normal” treatments are running
simultaneously on a channel pair—CH 1-2 or 3-4).
• INTENSITY (separately for each channel)
12. STOP
When the treatment time has elapsed, the current to the patient stops and a tone sounds signaling the end of a
treatment. Treatments in progress may be stopped at any time using one of the following methods.
Stop One Treatment Only. Press and hold the FUNCTION key and press STOP. is stops only the treatment in focus.
Stop All. Press the STOP key. All treatments at all channels will stop.
Stop Time. Reduce the treatment time using the Time arrow key. e output at the selected channel is stopped,
and the device then displays the parameters for the next treatment that remains in progress (if there is an
active treatment).
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 201738
BIPHASIC / RUSSIAN MODALITY INFORMATION
Biphasic / Russian
Modality Information
Russian Stimulation
With Russian Stimulation mode, the output of the device is a 2500 Hz sinusoidal wave. Russian stimulation currents produce
strong muscle contractions.
e Dynatron Solaris Plus devices allows complete control over all the parameters of the Russian Stimulation treatment.
ree treatment modes include Normal for ring one muscle, Reciprocal for ring two dierent muscles at dierent times,
and Co-contraction for ring two dierent muscles simultaneously. Choose a muscle contraction/relaxation cycle from
options of 10/10 (ten seconds on and ten seconds o), 10/30, 10/50, Continuous, and Custom cycles. e Normal mode
requires use of just one output jack (Channel 1, 2, 3, or 4). e Reciprocal and Co-Contraction modes utilize a channel pair
(Channels 1-2 or 3-4).
NOTE: e continuous duty cycle is not recommended for electrical muscle stimulation, but may be used for settings that
are intended to aect other results than a muscle contraction. Pulse rate, the pulse duration, and the ramp time can all be
modied from their default settings.
Biphasic Stimulation
e Biphasic stimulation is similar to Russian stimulation in the parameters that are selected and in the available options.
It diers from Russian stimulation in the pulse duration and rate ranges (see parameters below). Additionally, the Biphasic
pulse includes just one cycle (one positive phase and one negative phase) per pulse.
Biphasic / Russian Parameters and Defaults
e default settings and the available ranges for Biphasic and Russian are as follows:
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 201739
e pulse rate and duration should not be confused with the contraction/rest times in the treatment as these are dierent
parameters; the pulse occurs only during the contraction time. e diagrams below illustrate the relationship of each of these
parameters.
e pulse duration indicates the duration (in milliseconds or microseconds) of the output cycle of the pulse, and the pulse
rate is measured in number of pulse occurrences per second. Between pulses, current is at zero.
A 50 percent duty cycle or less is the usual duty cycle for Russian stimulation (the duty cycle includes one output cycle period
and one zero-current period). A 50 percent duty cycle means that the length of the pulse duration must be equal to or less than
the zero-current period. e number of pulses per second aects the allowable range of pulse durations. A greater number of
pulses per second means a shorter pulse duration is allowed.
50 Pulses per sec.1-500
10 mSec.04 to 50 mSec
Russian Stimulation
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 201740
BIPHASIC / RUSSIAN MODALITY INFORMATION
If a given Russian stimulation treatment has a 50 percent duty cycle, this means the output cycle is continuously repeating
for half of the pulse duration (see “Rate” in the diagram above) followed by a zero-current period for the other half of the
pulse duration.
Biphasic stimulation diers from Russian stimulation in the pulse duration (width) and rate ranges, as explained above. In
addition, the Biphasic pulse includes just ONE output cycle per pulse. One pulse cycle (including one positive phase and one negative phase) occurs, followed by a zero-current period.
Biphasic Stimulation
e pulse rate and duration (width) may be modied during setup of a Russian or Biphasic Stimulation treatment or may be
modied while a treatment is in progress.
When modifying the pulse rate and duration (width) for a treatment in progress, modify the intensity as well, as all
three of these parameters will aect delivered energy and patient comfort.
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 201741
HIGH VOLT INSTRUCTIONS
High Volt
Instructions
High Volt electrical stimulation is a pulsed DC current with pulse durations in the microsecond range and pulse rates
ranging from 1 to 200 Hz, with peak amplitude of up to 500 Volts. e Solaris Plus Series devices deliver High Volt utilizing
a twin-peak monophasic waveform.
High Volt treatments may be delivered using electrodes or the optional MultiStim probe. e device provides a dedicated
channel for High Volt electrodes treatment (HV) and a separate dedicated STIM PROBE JACK for probe treatments. During
High Volt treatments, the Solaris Plus device’s other output channels (1-2-3-4) remain available for other simultaneously stim
treatments. Note, the MultiStim probe is not available on the 707 model.
High Volt Electrode Setup
is treatment setup utilizes a standard lead wire with
two electrodes; an active and a dispersive electrode.
e size of the dispersive electrode is recommended
to be double the area of the active electrode. If desired,
the active output of the lead wire may be bifurcated
by using an optional bifurcated extension (Part
no. 7B0077) to attach additional active electrodes.
However, the combined total area of the active
electrodes should be no more than half the area of the
single dispersive (passive) electrode, as illustrated.
It is best to use a lead that is marked to show polarity.
e active electrode is connected to the positive
output. e dispersive electrode is connected to the
output end that is marked “HV Dispersive” (-). If
desired, a bifurcated lead extension may be attached
to the positive (HV Active) end of the lead wire,
allowing two active electrodes to be used. In this case,
High Volt Electrode Setup
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 201742
HIGH VOLT INSTRUCTIONS
be sure the dispersive electrode is at least twice the size (in area) of the combined sizes of the active electrodes. e bifurcated
lead wire extension is an optional accessory available through Dynatronics.
During the treatment current ows in one direction
between the active and dispersive electrodes.
Changing the polarity in the treatment parameters
has the eect of reversing the direction of the current
ow between electrodes. It is important to attach the
active electrode to the HV Active (+) output of the
lead wire to ensure you are delivering the selected
polarity. e Lead wires are labeled: HV Active (+),
HV Dispersive (-) delivery.
High Volt Probe Treatment Setup
is attended form of High Volt therapy is delivered
using the Dynatronics’ optional MultiStim probe in
conjunction with a dispersive electrode. is
hand-held probe is connected to the STIM PROBE
OUTPUT JACK located on the right-side of the
Solaris Plus device. e MultiStim probe intensity is
controlled on the probe itself.
High Volt electrodes with bifurcated active lead.
When setting intensity, the probe must be in contact with the patient’s skin. Be sure that the patient can feel the
current as you increase the intensity. Do not increase intensity for probe treatments if the patient is unable to report
feeling the current.
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 201743
WARNING
HIGH VOLT INSTRUCTIONS
High Volt Quick Setup
1. Press HI VOLT.
Electrode Treatment
• Plug in lead wire to the HIGH VOLT OUTPUT JACK (HV).
• Attach electrodes to patient.
Probe Treatment
• Press CHANNEL TOGGLE key until the PROBE LED is lighted.
• Plug the optional MultiStim probe into the STIM OUTPUT JACK.
• Attach dispersive electrode.
2. Choose the POLARITY. Bipolar is not available for probe treatments. Note: Polarity must be selected
before customizing a treatment.
3. TIME. Select TIME by using the TIME ARROW keys for Electrode Pads treatments only. Probe Treatment
TIME counts up.
4. Select (or setup) a PULSE RATE RANGE (High/Low) or set a single pulse rate by pressing the MODE toggle
key and selecting RATES. Use the arrow keys next to RATE-START and RATE-STOP to set the Pulse Rate
Range.
5. Set CONTRACTION/REST times by pressing the FUNC (Function key). Use the CONT/REST toggle key
to make your selection.
6. RAMP TIME. While in FUNCTION mode use the RAMP TOGGLE key to select RAMP TIME.
7. Raise the INTENSITY to the desired level. For a probe treatment, make sure probe is in contact with
patient’s skin when raising intensity.
8. Press START. Treatment time will begin to count-down for an Electrodes treatment. For a probe treatment,
press ON/OFF on the probe handle aer pressing the device START key to activate the treatment. Time will
begin counting up.
9. S T O P. Press and hold the FUNCTION and press the STOP key to stop only the focus treatment appearing
in the Treatment Screen. Pressing STOP alone, stops all treatments operating on the device. Using the
TIME arrow keys to bring treatment time to zero will also stop a treatment.
Detailed High Volt Setup
1. Press the HI VOLT key.
e High Volt channel and the default settings for High Voltage electrode pulsed stimulation are automatically
selected. Using the dedicated High Volt (HV) channel attach leads and place electrodes on the patient now. For a
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 201744
HIGH VOLT INSTRUCTIONS
High Volt Probe, insert the optional MultiStim probe into the STIM PROBE output jack on the right-side of the
device. Attach the dispersive electrode. Press the DELIVERY toggle key until PROBE is illuminated GREEN.
Pulse Rate................................ High Range
Range: ............................... High 80-120 Hz
If you wish to use the default settings, increase the intensity to the desired level, and press START. For a probe
treatment, use the intensity arrows located on the handle of the probe. Press START on the device faceplate rst, next
press START on the probe handle to activate the treatment. To customize the settings, follow steps 2-8.
ELECTRODE PADS TREATMENTS are timed, and a treatment time in MINUTES is entered at the start of the
treatment. e timer counts DOWN for electrode treatments.
All PROBE TREATMENTS are manual, and a treatment time is not set. During the probe treatment, the timer will
start at zero and count up in seconds when current is ON, and will return to zero when current is stopped. To begin
a probe treatment, press START on the device face plate, next press the ON/OFF button on the probe. Pressing the
ON/OFF button on the probe will also stop the treatment and return the time to zero. If you press STOP on the
device, all therapies currently running will be terminated. If you press Function-Stop, only the High Volt treatment
or treatment in focus will stop.
2. Choose the POLARITY
NOTE: e Bipolar option is not available for probe treatments. When both Bipolar (negative and positive) is
selected, the device alternates between the two, delivering each polarity for approximately 30 seconds. Some
discomfort may be felt by the patient when the polarities change. If the patient nds this setup too uncomfortable,
you may consider selecting a monopolar treatment and/or reducing the intensity.
Press the Polarity toggle key to choose Polarity. Polarity options are Negative, Positive, and Bipolar.
3. Change the treatment TIME, if desired.
Electrode Pads. Use the TIME UP/DOWN ARROW keys to change the TIME for an Electrode treatment. Using
the MODE toggle key, be sure TIME is illuminated in the MODE window when entering TIME.
Probe. e TIME display shows zero until the START key on the faceplate of the device is pressed, followed by
pressing the ON/OFF button on the probe, initiating the ow of current. e timer counts up from zero (in seconds)
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 201745
HIGH VOLT INSTRUCTIONS
allowing for exible time for the delivery of current to each treatment site. When you press the button on the probe
to STOP the current, the device beeps twice and the timer resets to zero.
4. PULSE RATE RANGE.
Select a PULSE RATE RANGE (High 80-120 Hz or Low 1-10 Hz) or set a single pulse rate by pressing the MODE
toggle key and selecting RATES from 1-120 Hz. Use the arrow keys next to RATE-START and RATE-STOP to set
the Pulse Rate Range. ese ranges may be changed for a single treatment if desired, or press and hold the START
key to set new default settings to be applied to all future treatment setups. e PULSE RATE RANGE must be set
before pressing START.
5. CONTRACTION (ON) and REST (OFF) TIMES
FUNCTION KEY: To access the CONTRACTION AND REST settings, press the FUNCTION key.
Press the CONT/REST toggle key one or more times to select contraction/rest (Duty) cycle times. Available options
include 10/10, 10/30, 10/50, CONT (Continuous). and Custom. e setting of 10/30, for example, means a 10-second
contraction time followed by a 30-second rest time.
CUSTOM (Duty) CYCLE TIME SELECTIONS
• Press the FUNCTION key located on right-side of the device faceplate.
• Using the CONT/REST toggle key, select CUSTOM in the CONT/REST window.
• Using the MODE toggle key, select CUST CR in the MODE window. Selections will be illuminated GREEN.
• Custom Contraction and Custom Rest cycle times may now be set by using the Up and Down arrow keys
located next to Custom Contraction and Custom Rest windows. Available ranges for Contraction (ON) times
are 1-120 seconds, for the REST (OFF) times 1 to 300 seconds. Remember, the REST time cannot be less than the
Contraction time. Pressing and holding the START key until the beep is heard will save the current Contraction/
Rest settings as the default.
6. Choose the RAMP setting.
FUNCTION KEY: To access the RAMP settings, press the FUNCTION key.
Press the FUNCTION key on the right side of the face plate. Press the RAMP toggle key to select RAMP time.
Selections include: 0.5, 1.0, 1.5, and 2.0. Ramp time is applied to both the start and end of the contraction. Ramp
time is in addition to the contraction time itself. A ramp setting is not applied to the Continuous duty cycle except
when using a probe. With a probe treatment set to a Continuous duty cycle, a ramp time of 3 seconds is automatically
applied.
7. INTENSITY.
Using the Up and Down arrow keys to the right-side of the INTENSITY display, set the Intensity. Remember, when
you increase intensity, current to the patient begins. erefore, START should be pressed immediately aer setting
the intensity to begin the treatment timer.
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 201746
HIGH VOLT INSTRUCTIONS
NOTE: e intensity arrows on the probe handle work the same as the intensity arrows on the device faceplate.
Never use High Volt to treat any conditions which contribute to loss of sensation, or an area where the patient
cannot feel the electrical stimulation.
8. Press STA RT .
Electrodes Pad Treatment: Press START. When you press start, the treatment timer begins counting down.
Probe: Press START on the device followed by ON/OFF on the probe handle to activate a probe treatment. e timer
will begin counting up. If you fail to set the intensity on either an electrode or a probe treatment before pressing
START, the Intensity display will begin ashing, and you will be unable to start the treatment until the intensity is set.
NOTE: SAVING DEFAULTS. If the treatment you have just set up is the most common High Volt setup you use,
the treatment parameters can be saved as the defaults for your own device. Aer setting up the treatment, press
and hold the START key for two seconds. At the end of two seconds, a beep will sound indicating the treatment
parameters have been saved. e next time this modality is chosen, the parameters will be selected automatically.
9. MODIFY Settings
While the High Volt treatment is in progress, TIME, INTENSITY, AND POLARITY can be modied excluding
POLARITY for MultiStim probe. Carefully observe the channel indicator lights when modifying a treatment. When
a channel’s light is illuminated GREEN, the current treatment parameters for that channel are displayed. Any changes
made to the parameters will aect only the channel that is illuminated in GREEN. Use the Channels Toggle key to
display the parameters of another channel in order to modify parameters.
10. STOP
When the treatment time has elapsed, the current to the patient stops and a tone sounds signaling the end of a
treatment. Treatments in progress may be stopped at any time using one of the following methods.
Stop One Treatment Only. Press and hold the FUNCTION key and press STOP. is stops only the treatment in
focus.
Stop All. Press the STOP key. All treatments at all channels will stop.
Stop Time. Reduce the treatment time using the Time arrow key. e output at the selected channel is stopped,
and the device then displays the parameters for the next treatment that remains in progress (if there is an active
treatment).
Multistim Probe Stop. A probe treatment may be stopped by pressing the ON/OFF key located on the handle of
the MultiStim probe.
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 201747
HIGH VOLT MODALITY INFORMATION
High Volt
Modality Information
High Voltage pulsed stimulation is a pulsed DC current with
pulse durations in the microsecond range and pulse rates ranging
from 1 to 200 Hz, with a peak amplitude of up to 1.0 A utilizing a
twin-peak monophasic waveform.
e Dynatron Solaris Plus High Volt treatment setup uses
a dedicated channel. Each treatment utilizes the single HV
channel with one or more active electrodes and a large dispersive
electrode. Electrodes are placed on opposite sides of the
aected area so treatment is “through” the aected area. e
optional Dynatron MultiStim probe is available to accommodate a
hand-held treatment delivery.
High Volt Waveform for positive polarity
High Volt Waveform
e High Volt waveform is a twin-peak monophasic decaying waveform with either positive or negative polarity.
High Volt Settings
High and Low Sweep Pulse Rate. e user may select High (80-120 Hz) or Low (1-10 Hz) frequency ranges. During a
treatment, the device scans through the range of frequencies selected. e settings for these pulse rate frequency ranges may
be customized and saved by the user. e available custom range is 1 to 200 Hz.
Continuous Pulse Rate. e pulse rate may be set to a single continuous pulse rate instead of a range. e pulse rate is
selected from a range of 1 to 200 Hz.
Selectable Polarity. Positive or Negative monophasic current may be selected. Alternating positive and negative (Bipolar)
currents may also be selected. Note: Dual Polarity is not available with a probe treatment.
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 201748
HIGH VOLT MODALITY INFORMATION
Custom Contraction/Rest Time Cycles. e Dynatron Solaris Plus allows for the choice of muscle contraction and
relaxation time cycles (Duty Cycles) from options of 10/10 (ten seconds on and ten seconds o), 10/30, 10/50, Continuous
or Custom cycles. e Custom time cycle allows for a Contraction (ON) time from 1 to 120 seconds, and a Rest (OFF) time
from 1 to 300 seconds. Remember, the Rest time cannot be less than the Contraction time.
Selectable Ramp Speed. You can choose a ramp speed of .5 (1/2) to 2.0 seconds in half-second increments. For patient
comfort, the ramp occurs both before and aer the “Contraction” segment of the pulse.
Pulse Duration. e pulse duration is xed at 25 µs (micro-seconds).
Pulse Pair Interval. e interval between the two pulses in the wave form is xed at 100 µs.
Intensity Display in Volts. Intensity is displayed in volts (peak voltage with no load) with a range of 1 to 500 volts.
High Volt Default Settings
e following default settings are set by the manufacturer and are selected when you choose High Volt. You may change
these defaults to your own preferred settings. See “Setting Defaults” in this manual.
• High Volt Pads Treatment
• Continuous Duty Cycle
• Treatment Time Electrodes: 10 minutes
• Treatment Time Probe: 0.00
• Polarity: Negative
• Pulse Rate: High Range
Default High Range: ...........................................................................80-120 Hz
Default Low Range: ..................................................................................1-10 H
Available Range:....................................................................................1-200 Hz
Maximum Power Output: ..................Limited to less than 500 V, open load
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 201749
MICROCURRENT INSTRUCTIONS
Microcurrent
Instructions
Microcurrent treatments may be delivered using either electrodes or the optional MultiStim probe. e MultiStim probe
option is not available on the 707. For treatment with electrodes, plug one lead into the CHANNEL 1 JACK, and place the
two electrodes on the patient before setting up the treatment. For treatment with probes, plug the Dynatron MultiStim Probe
into the Stim Probe Jack located on the right-side of the device.
During the Microcurrent treatment the current is delivered to both CHANNEL 1 and to the MULTISTIM PROBE output
simultaneously. erefore, CHANNEL 1 is unavailable for any other treatment while any Microcurrent treatment is in
progress (including a single Microcurrent probe treatment).
NOTE: Microcurrent is unavailable if Direct Current is operational.
How To Use The Optional MultiStim Probe For Microcurrent Treatments
To set up a Microcurrent Probe treatment, use the end of a cotton swab (such as a Q-Tip®) inserted into the end of both the
MultiStim probe (active) and Microcurrent Ground probe. Cut the end of the swab to a short length. e cotton must touch
the metal ring at the tip of the probe. Use a conductive electrolyte spray or water to wet the cotton swab before treating. If the
cotton dries out during treatment, conductance may become erratic. Re-wet the cotton, if needed.
NOTE: When applying a Microcurrent probe treatment, it is necessary to use either a ground probe or a ground
electrode in conjunction with the active MultiStim probe. Attach the banana plug ground wire from the MultiStim
probe by plugging it into the back of the ground probe. To use a ground electrode you must use a banana-to-pin
adapter to connect the ground wire to the electrode.
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 201750
MICROCURRENT INSTRUCTIONS
Active Probe: To deliver the current to the patient through the active MultiStim probe, rst set up the probe treatment and
press Start. Hold the probe as you would hold a pencil and press the ON/OFF button on the probe to start delivery of the
current. e active probe should touch the patient’s skin at the treatment site and the ground probe should touch the patient’s
skin elsewhere near the treatment site. is completes the circuit and delivers current to the patient. Do not hold the ON/
OFF button down. Once the button has been pressed and released, the current is delivered until the button is pressed and
released again to stop the current. e unit beeps once when the button is pressed to start the current, and beeps twice when
you press the button to stop the current.
During a treatment, observe conductance by listening to the audio tone. Also, the CONDUCTANCE BAR GRAPH will
indicate the conductance between patient skin and probe tip, allowing for an optimum treatment.
While delivering current, the treatment timer counts up in seconds from zero. When the current is stopped, the timer returns
to zero. Press the ON/OFF button again to commence the next treatment cycle. Continue in this way until treatment is
completed.
Ground Probe: e optional ground probe is used to complete the circuit, allowing current to ow through patient’s tissue.
e ground probe should touch the patient’s skin at any location away from the treatment point.
Ground Electrode: As an alternative, a dispersive electrode may be used. A hand-held ground probe is not needed when
using a ground electrode. is is particularly convenient when treating in several dierent places around one point.
Unplug the ground probe from its cable, attach a banana to pin adapter to the cable, then attach an electrode to the pin. Place
the electrode on the patient at a site where it will not interfere with placement of the active probe during treatment.
Test Mode: Press and release the MultiStim Probe ON/OFF button again to stop delivery of the current. e unit beeps twice
to let you know current delivery has stopped. e probe is now in “TEST” mode allowing you to continue monitoring
conductance at any point desired. When the probe touches the patient’s skin, conductance continues to be measured and the
audio tone is produced. Press and release the button when you wish to return to the Treatment mode.
NOTE: In “TEST” mode during a Microcurrent treatment, a small amount of current is delivered to allow an
impedance reading to be taken in order to show conductance. e current delivered during “TEST” mode is either
25 micro amperes or the micro amperage setting for the treatment itself, whichever is less. For purposes of this
measurement, current is delivered in pulses of approximately 20 mSec.
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 201751
MICROCURRENT INSTRUCTIONS
Microcurrent Quick Setup
1. Press the MICRO key. Channel 1 for electrodes is automatically selected.
2. PROBE TREATMENT. Insert the optional MultiStim probe into the STIM OUTPUT jack. Use the
DELIVERY toggle key to select PROBE. Selections are illuminated GREEN.
3. Choose POLARITY (Positive, or Negative. Bipolar selection available only for electrode treatments).
4. Choose POLARITY (Positive, Negative, or Bipolar). Note: Bipolar is not available in when inProbe mode.
5. CONDUCTANCE BAR: PAD: e CONDUCTANCE BAR may be turned ON or OFF by using the COND
toggle key. PROBE: Turning the Audible Tone OFF in Probe mode will turn o the Conductance Bar.
6. FREQUENCY. Using the MODE toggle key select FREQ. Use the right-side UP/DOWN arrow keys to
change FREQUENCY if desired.
7. INTENSITY. Using the INTENSITY arrow keys or the (+) & (-) switch in the probe handle, change the
INTENSITY, if desired.
8. TIME. For electrode treatments only, change the TIME by using the TIME ARROW keys.
9. VOLUME. e Audible Tone volume can be adjusted from 0 (o) to 10 for a Probe Treatment aer selecting
VOLUME in the MODE window.
10. Press STA RT .
11. S T O P. Press and hold the FUNCTION and press the STOP key to stop only the focus treatment appearing
in the Treatment Screen. Pressing STOP alone, stops all treatments operating on the device. Using the
TIME arrow keys to bring treatment time to zero will also stop a treatment.
Detailed Microcurrent Setup
Before setting up a Microcurrent treatment with electrodes, attach a lead wire with two electrodes to CH 1. For a probe
treatment, connect the optional MULTISTIM PROBE to the STIM OUTPUT jack on the right-side of the device. When a
MULTISTIM PROBE treatment is in progress, Channel 1 is not available for any other treatment.
1. Press the MICRO (Microcurrent) Key.
Aer pressing the MICRO key, the default settings for an electrodes pads treatment are automatically selected.
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 201752
MICROCURRENT INSTRUCTIONS
Electrode Pads Default Setting:
Time.................................................20 min.
For a Probe treatment, insert the MultiStim probe into the STIM PROBE output jack located on the right side to the
device. Use the DELIVERY toggle key to select PROBE on the Treatment Display Screen. e Probe defaults will be
selected. Note: e bipolar option is not available for a Microcurrent Probe treatment.
If using the treatment default settings, press START. If you wish to use other settings, complete the following steps.
2. Choose ELECTRODE OR PROBE TREATMENT
Choose the output desired. Select CHANNEL 1 for electrodes treatment. For a probe treatment, insert the MultiStim
Probe into the STIM PROBE output jack. Press the DELIVERY Toggle key to select PROBE.
3. POLARITY
Using the POLARITY toggle key, select Positive, Negative. Bipolar selection only available for electrode (pads)
treatment.
4. TIME
For electrode treatments only, change the TIME, if desired. To change the treatment time, use the Up/Down arrow
keys located next to the TIME display. For probes treatment, the TIME display shows zero until treatment begins.
Aer pressing the START key on the faceplate of the device, press the 1/0 (On/O) button on the probe which will
initiate the ow of current. e device beeps once and the timer then counts up from zero (in seconds). When the
“1/0” key is pressed again, the current stops, the device beeps twice, and the timer resets to zero.
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 201753
MICROCURRENT INSTRUCTIONS
5. FREQUENCY
Press the MODE toggle key to select FREQ (Frequency). Using the Up/Down arrow keys located next to the
Frequency display, enter the desired frequency (0.1 - 200).
6. INTENSITY
INTENSITY is displayed in microamps (µA). Use the Up/Down arrow keys located next to the Intensity display to
set the intensity (10-990 µA).
7. CONDUCTANCE / AUDIBLE TONE
Both the Conductance Bar and the Audible Tone can be turned on and o by using the toggle key beneath their
respective display windows. Note: Turning the Audible Tone OFF in Probe mode will turn o the Conductance Bar.
When the Probe is PAUSED, sound and conductance are ON.
8. Press STA RT
Electrode Treatments. Pressing START causes the treatment timer to begin counting down and Stim is delivered
through Channel 1.
Probe Treatments. Begin by pressing START on the device faceplate, then press and release the 1/0 (On/O) button
on the probe to activate current through the probe. e device beeps once to signal that current delivery has started
(it is not necessary to hold the button down while treating). e treatment timer begins counting up in seconds from
zero to provide a convenient means of timing the delivery of current at a given point. Press and release the 1/0 button
again to stop the current through the probe. e treatment timer returns to zero and the device beeps twice to signal
that current delivery has stopped. Continue the treatment cycle in similar fashion for each treatment site. When
applying a Microcurrent Probe treatment, it is necessary to use a ground probe or ground electrode in conjunction
with the treatment.
NOTE: While the treatment is in progress, treatment parameters can be modied. Separate default parameters may
be saved for both electrodes and probe treatments.
9. STOP
When the treatment time has elapsed, the current to the patient stops and a tone sounds signaling the end of a
treatment. Treatments in progress may be stopped at any time using one of the following methods.
Stop One Treatment Only. Press and hold the FUNCTION key and press STOP. is stops only the treatment in focus.
Stop All. Press the STOP key. All treatments at all channels will stop.
Stop Time. Reduce the treatment time using the Time arrow key. e output at the selected channel is stopped, and
the device then displays the parameters for the next treatment that remains in progress (if there is an active treatment).
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 201754
MICROCURRENT MODALITY INFORMATION
Microcurrent
Modality Information
Microcurrent is low-volt pulsed microamp stimulation which has been used for symptomatic relief of chronic intractable
and post-surgical pain. With the Dynatron Solaris Plus both attended or unattended Microcurrent therapy can be applied.
Unattended therapy is delivered using electrodes. Attended therapy is delivered using the optional Dynatron hand-held
MultiStim probe.
Due to the very low frequencies generally associated with a Microcurrent treatment, the current is commonly not felt by the
patient. However, Microcurrent at high intensity settings or during a probe treatment can sometimes be felt by the patient
and may even be uncomfortable. During a probe treatment, this may be the result of current being concentrated at a very
small area.
As with all forms of electrotherapy, current density is dependent upon the current setting, the intensity setting, size, and type
of electrodes used, and conductance. Higher current density increases the possibility for patient discomfort. e current
density can be reduced by decreasing the amount of current (intensity) or increasing the area through which the current is
being delivered. e area can be increased by using larger electrodes and/or making sure that the total area of the electrode
is actually delivering current.
Microcurrent Waveforms
ree dierent output waveforms (polarities) may be
selected with the Dynatron Solaris Plus. Positive and
Negative selections are available for both probe and
electrodes. An additional bipolar option is available when
using electrodes. With Conductance Bar turned o, the
rst two are unipolar square waves having a monophasic
pulse at a 50 percent duty cycle (equal on and o times).
In the illustration, the rst unipolar waveform has negative
polarity only, and the second has positive polarity only.
e third waveform is a bipolar (negative and positive)
square wave with a 50 percent duty cycle.
Microcurrent Waveform
Negative Unipolar
Squ a r e Wave
Positive Unipolar
Squ a r e Wave
For bipolar square wave,
press the toggle key until
both negative and positive
waveforms are selected
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 201755
MICROCURRENT MODALITY INFORMATION
Microcurrent Guidelines
When delivering Microcurrent therapy, observe the following guidelines:
1. Use only moderate current. Consult published literature for recommended settings for Microcurrent treatment.
2. When using Microcurrent probes, keep in mind that all of the current is delivered through the tip of the probe
resulting in much higher current density than when using electrodes. erefore, low to moderate intensity settings
will be much more comfortable for the patient, especially when treating sensitive areas.
3. When using electrodes, ensure that the area on the patient’s skin where the electrode is to be placed is clean and free
of all foreign matter. is includes powders, perfumes, as well as body oils, dirt, and grime. Cleaning with an alcohol
wipe should be adequate. Allow the alcohol to fully evaporate before applying the electrodes. Iontophoresis occurs
with all electrical current therapies, and can drive any above-surface contaminants below the epidural layer where
an allergic reaction may occur.
4. Make sure the electrodes being used are in good condition. e poly adhesive electrodes should have good adhesion
over the entire surface area of the electrode. e area where the leads attach to the electrode (either the pin connector
or a snap) should not be damaged such that the connection to the backing behind the adhesive is broken. Carbon
electrodes should be deep black, and should be free of cracks in the electrode surface.
5. Some patients tend to be much more sensitive to Microcurrent treatments. On patients with this tendency, treat with
reduced intensity and/or shorter treatment times with possibly more frequent treatments, if required.
6. If electrodes are placed on an uneven skin surface, such as over a raised mole, the electrode may not adhere evenly
over its entire surface and current could be intensied at the raised area, causing discomfort to the patient.
Microcurrent Default Settings
e following default settings are set by the manufacturer and are selected when a Microcurrent treatment is selected. Default
Settings may be changed to your own preferred settings. See “Setting Defaults” in this manual.
Electrodes. e Microcurrent treatment setup with electrodes is the default, and the following parameters are also selected:
Time ................................................Determined by Practitioner - Counts Up
Audible Tone ................................................................................................... On
Conductance ................................................................................................... On
Available Ranges.
Frequency ........................................................................................ Between 0.1 and 200 Hz (positve or negative)
Biploar 0.0-5.0 Hz
Maximum Voltage: ....................................................................................................................... 60 V (open circuit)
Intensity: .......................... Between 10 and 990 microamps open load, constant current, increments of 10 µA.
Maximum Load Resistance ........................................................................................................................ 500 Ohms
e ranges of .1 to 200 Hz frequency and between 10 and 990 micro amperes intensity are available for setting up both
electrodes and probes treatments. For guidance in selecting the appropriate frequency and intensity for a given probes or
electrodes treatment, consult published medical literature.
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 201757
DIRECT CURRENT INSTRUCTIONS
Direct Current
Instructions
Direct Current ows in one direction. From a practical perspective, Direct Current can be dened as having a pulse duration
long enough to depolarize skeletal muscle when the nerve is not intact. On the Solaris Plus line of devices, Direct Current is
a square wave interrupted with a set interpulse duration of 500 mSec. e interrupted treatment is delivered in a continuous
train of 0.1 mSec to 500 mSec pulses set at a maximum of 20 mA. Duration, Intensity, and Polarity settings may be adjusted
separately, if desired. e Polarity may be set at Positive or Negative. Treatment is provided only through the optional
MultiStim probe. e use of various attachments and a dispersive electrode are necessary. e MultiStim probe utilizes the
Stim Probe output jack located on the right-side of the Solaris Plus device. ON/OFF controls and INTENSITY keys (+/-) for
the MultiStim probe are located on the handle. Direct Current is not available on the 707.
Direct Current Quick Setup
Choose DIRECT CURRENT.
1. Plug the MULTISTIM PROBE into the STIM PROBE output jack. Use the DELIVERY toggle key to select
PROBE. Attach the dispersive electrode.
2. Choose POLARITY. Negative or Positive. Note: Polarity defaults to Negative.
3. DURATION of each DC pulse. Press the MODE toggle key to select Duration. Set duration using the Up/
Down arrow keys next to the Duration display.
4. Place probe tip in contact with patient’s skin.
5. Press the STA RT key followed by the ON/OFF key located on the MultiStim probe.
6. RAMP UP INTENSITY by using the POWER INTENSITY KEYS (+) and (-) located on the handle of the
probe. Caution: Probe must be in contact with patient’s skin prior to starting treatment.
7. Treatment will continue until the OFF key on the MultiStim probe is pressed.
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 201758
DIRECT CURRENT INSTRUCTIONS
Detailed Direct Current Setup
1. Press the DIRECT CURRENT key.
When DIRECT CURRENT is selected, the STIM PROBE output jack automatically becomes active.
2. Plug the MultiStim probe into the STIM PROBE output jack. Use the DELIVERY toggle key to select PROBE.
Attach the dispersive electrode.
e default settings for the modality are programmed at 100 mSec, and Negative polarity. If you wish to use the
default settings, attach dispersive electrode and place probe in contact with the patient’s skin. Press START on the
device followed by the ON/OFF key on the MultiStim probe handle. ere is no need to hold down the key. Current
delivery begins immediately. Press the POWER INTENSITY (+) (-) KEYS located on the Probe handle until the
desired muscle contraction is obtained.
If you wish to change the treatment settings, complete the following steps.
3. Choose POLARITY
Press the POLARITY toggle key and selecting Positive or Negative. e selection will be illuminated GREEN. e
default setting is Negative (-). NOTE: Polarity may be changed during setup or during a treatment. Place the probe
in contact with the patient’s skin before pressing START.
4. Set PULSE DURATION
Press the MODE toggle key to select Duration. Once Duration has been selected, the DURATION DISPLAY will
be visible in the right-hand corner of the Treatment Display Screen. Set duration between 0.1 mSec and 500 mSec
per pulse by using the Up/Down arrow keys next to the DURATION DISPLAY. Pulse Duration selections available
are 0.1, 0.3, 0.5, 1.0, 10, 100, and 500, with an interpulse duration of 500 mSec. e selected pulse duration will be
delivered in a continuous train of pulses set up to a maximum of 20 mA.
5. START.
Press the START key located on the faceplate of the device followed by pressing and releasing the ON key located on
the MultiStim Probe handle. When the ON key is pressed, a single beep will be heard, followed by one beep every
second as long as the DC current continues to ow.
6. INTENSITY
Aer the ON key has been pressed, begin ramping up the intensity by pressing the POWER/INTENSITY Up arrow
key on the faceplate or by pressing the (+) (-) keys on the handle of the MultiStim probe. e INTENSITY may be
set from 1 to 20 mA.
Before selecting the intensity setting for an individual patient, see “Electrotherapy Usage Cautions” in this manual
for recommended intensity settings. Also see the section of this manual entitled “Contraindications, Warnings, and
Precautions” for specic precautions when treating any conditions contributing to loss of sensation, or anytime the
patient cannot feel the electrical stimulation.
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 201759
DIRECT CURRENT INSTRUCTIONS
7. STOP
When the desired muscle contraction has been reached, press the ON/OFF key on the MultiStim Probe handle to
end the treatment. When the ON/OFF key on the probe handle is pressed, two quick beeps will be heard indicating
that the DC treatment is complete.
If the DC treatment is in focus, pressing and holding the FUNCTION and pressing the STOP key will also STOP the
treatment.
NOTE: Treatment setup may be saved as new Default settings by pressing and holding down the START key until a
beep sounds. However, the intensity setting is not saved but must be reset with each individual patient.
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 201760
DIRECT CURRENT MODALITY INFORMATION
Direct Current
Modality Information
Direct Current is electrical current that ows in one direction. Direct Current has been clinically used for: 1) Performing
strength duration curve tests and 2) Stimulating denervated muscle following peripheral nerve injury.
Direct Current Probe Therapy
e Dynatron Solaris Plus Direct Current treatment setup uses the optional MULTISTIM PROBE with the STIM PROBE
JACK as its dedicated channel. e treatment is normally delivered with a paddle adapter accompanied by a large dispersive
electrode or ground probe acting to complete the circuit. Note: Electrodes and Probe must be in contact with the patient
prior to beginning therapy.
Direct Current Waveforms
e Direct Current wave form is a square wave
with adjustable duration times. Duration will
default to 100 mSec, but may be set from 0.1
mSec to 500 mSec. Duration selections available
are 0.1, 0.3, 0.5, 1, 10, 100, 500 with an interpulse
duration of 500 mSec. e variable Pulse Duration
is interrupted with an interpulse duration of 500
mSec. Note: Load impedance does not aect
output of the waveform.
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 201761
Direct Current with Variable Pulse Duration Times
DIRECT CURRENT MODALITY INFORMATION
Direct Current Warnings
It is important that excellent coupling be maintained any time current is owing. Increased resistance from poor coupling or
excessive current density can cause skin reactions.
• Electrodes must be attached and probe placed in contact with the patient’s skin prior to starting the treatment.
• Direct Current tends to lower surface sensory perception as the treatment proceeds. e patient must have good
sensory perception if Direct Current is to be used.
Power Output: ........................................................................................ 20 mA.
Note: Load impedance does not aect output of the DC waveform.
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 201762
DYNATRON TRI-WAVE™ OPERATING INSTRUCTIONS
™
Dynatron Tri-Wave
Operating Instructions
Tri-Wave Light erapy provides delivery of a single wavelength of light or any combination of 3-wavelengths of light (red,
infrared, and blue) that allow for 7 dierent light treatment combinations. e Tri-Wave Light Probe or 2 robust 5” x 7”
Tri-Wave Light Pads deliver the desired treatment. Dual Light Pads allow for greater coverage of a joint, treatment of 2
separate areas at the same time, and the ability to treat large areas of the body when placed side-by-side. e Solaris Plus 707
has the unique ability to utilize two Light Probes and two Light Pads simultaneously.
e Solaris Plus Tri-Wave Light erapy probe is specically designed for treating smaller areas. e probe has a treatment
area of 6 cm2 and should be applied in direct contact with the surface of the skin over the treatment area. To treat areas
larger than 6 cm2 individual treatments should be given one at a time in a grid pattern until the desired treatment area is
covered unless treating large areas where the Dynatron Tri-Wave Light Pads (5” X 7” each) would be more suitable. Please
see published literature for specic treatment parameters.
e Dynatron Tri-Wave probe and the Dynatron Tri-Wave Pads have been cleared by the FDA to provide topical heating for
temporary increase in blood circulation, temporary relief of minor muscle and joint aches, pain and stiness and relaxation
of muscles; for muscle spasms and minor pain and stiness associated with arthritis.
CAUTION
e Dynatron Tri-Wave probe and the Dynatron Tri-Wave Pads must be plugged securely into the respective output
jacks prior to turning ON the Solaris Plus base console.
WARNING
Protective eyewear is required for patient and practitioner.
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 201763
DYNATRON TRI-WAVE™ OPERATING INSTRUCTIONS
Dynatron Tri-Wave™ Light Quick Treatment Setup
1. Remove all gels and lotions from skin surface.
2. While the Dynatron Solaris Plus controlling console is turned OFF, plug the Dynatron Tri-Wave Light
probe and/or Dynatron Tri-Wave Light pads into the corresponding output jacks located on the side(s)
of the console.
3. Placing protective eyewear on both the patient and the practitioner is required.
4. Turn ON the Dynatron Solaris Plus controlling console.
5. LT PROBE and LT PADS. Press Probe or Light Pad on the Solaris Plus console. When using the 707, press
the toggle key below the channels window to select the desired Probe or Pads to activate.
6. Select WAVELENGTHS. Using the Wavelength toggle keys, select the desired wavelengths(Infrared,
Blue, Red) by toggling the wavelength ON or OFF. e selection will be illuminated in GREEN when the
wavelength is active. Note: Wavelengths cannot be changed during a treatment.
7. TIME/DOSAGE. Using the Up/Down arrow keys located next to the TIME and DOSAGE displays,
Dosage/Time may be adjusted. Dosage and Time are calculated according to the wavelengths selected. If
changes are made to Dosage, Time, will automatically be recalculated. If changes are made to Time, Dosage
will automatically be recalculated.
8. START. Press START on the Dynatron Solaris Plus control console.
Probe. Aer pressing START on the console, press the 1/0 (ON/OFF) key located on the Probe handle to
begin treatment.
Pads. e Pads treatment begins immediately when the console START key is pressed.
9. PAUSE.
Probe. Press the 1/0 (ON/OFF) key located on the Probe handle to pause a Probe treatment or press the
PAUSE key located on the console. To continue the treatment, press the 1/0 (ON/OFF) key on the Probe
handle or the PAUSE key on the console again.
Pads. With the desired set of Pads as the focus treatment, press the PAUSE key located on the Solaris Plus
console. To continue the treatment, press the PAUSE key again.
10. STOP
Probe. Press the 1/0 (ON/OFF) key on the Probe to pause or stop a probe treatment or press the FUNCTION/
STOP keys on the console simultaneously.
Pads. Press the FUNCTION/STOP keys on the console to immediately STOP a Pads treatment, reduce the
treatment time to zero or allow the treatment to automatically time out.
NOTE: Pressing only the STOP key on the Dynatron Solaris Plus console will terminate all treatments in progress.
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 201764
DYNATRON TRI-WAVE™ OPERATING INSTRUCTIONS
Detailed Treatment Setup
1. CLEANSE THE TREATMENT AREA
To insure that all gels and lotions have been removed from the surface of the skin, cleanse the treatment area
thoroughly.
CAUTION
DO NOT use gels or lotions in combination with Light erapy. Gels or lotions can clog the vents causing overheating
and damage to the internal components. Damage caused by the use of gels and lotions may void your warranty.
2. WHILE THE DYNATRON SOLARIS PLUS CONTROLLING CONSOLE IS TURNED OFF, insert the Dynatron
Tri-Wave Light Probe and/or Dynatron Tri-Wave Light Pads into their proper output jacks. e Solaris Plus Light
erapy output jacks are located on the le-side of the console. Light erapy output jacks on the 707 are located on
both sides of the console. Carefully align the connector “keys” to avoid damage to the pins when inserting the probe
and pad cable(s) into the jack(s).
3. Place PROTECTIVE EYEWEAR on both the patient and the practitioner.
POWER-ON THE CONSOLE.
Aer the Pads and/or Probe are plugged into the proper output jack(s), select the “1” (ON) position on the power
switch located on the back of the Dynatron Solaris Plus console. If a probe or pads have not been properly attached,
an error message will be displayed on the Treatment Display Screen or the probe/pads will not be recognized by
the console. It is recommended, if the probe/pads are used frequently, that the unit remain ON with the Dynatron
Tri-Wave devices plugged into the console throughout the day.
4. Select PROBE or LIGHT PAD on the Dynatron Solaris Plus console.
Probe. Pressing “PROBE” causes the LT PROBE in the Channels window of the display to be illuminated in GREEN
text indicating that the Probe is activated in preparation for treatment.
Probe. Solaris Plus 707. Two LIGHT PROBES may be used at the same time on the 707. Pressing the “PROBE” key
will automatically default to the Light Probe located on the le side of the console and LT PB1 will be highlighted in
green in the Channels Window. Press the toggle key below the Channels Window to change the focus to the Probe
located on the right side. LT PB2 will be activated and highlighted in green.
Pads. Pressing “LIGHT PAD” causes LT PAD in the Channels window of the display to be highlighted in GREEN
text indicating that the Pads are activated.
Pads. Solaris Plus 707. Two LIGHT PADS may be operated at the same time on the 707. Pressing the “LIGHT PAD”
key will automatically default to the Light Pads located on the le side of the console and LT PD1 will be highlighted
in green in the Channels Window. Press the toggle key below the Channels Window to change the focus to the Pads
located on the right side. LT PD2 will be activated and highlighted in green.
707 Note. Both Probe and Pad treatments may be operated simultaneously. Press the Toggle Key below the Channels
window to bring the desired treatment parameters into focus on the Treatment Display Screen. To set up a new
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 201765
DYNATRON TRI-WAVE™ OPERATING INSTRUCTIONS
treatment, press the TREATMENT MODALITY key for the desired treatment. When a treatment is in focus, the
selected treatment will be highlighted in GREEN text within the Channels window and the treatment name will
appear in the center of the Treatment Display Screen.
5. Select WAVELENGTHS (Total Energy Delivered)
e user has the capability of selecting a specic wavelength or blending various wavelengths into a single treatment.
Wavelengths are selected by using the WAVELENGTH TOGGLE keys to turn the Wavelengths “ON” and/or “OFF.”
e default wavelength for the Probe and Pads is Infrared (850nm) and Red (624nm).
To select a wavelength other than the Infrared (850nm) default, press the key under the light therapy option you wish
to activate: Red (624nm), Blue (464nm), or Infrared (850nm). Pressing the key under each option will cause the ON
text to become illuminated. Pressing the key again will cause the OFF text to illuminate. Selections may be activated
individually or in any combination.
Dosage (J/cm2 ) is calculated as a combination of Time and Power, and is automatically calculated once the Wavelength
selections have been made. Dosage can manually be altered by using the Up/Down arrow keys located on the
right-side of the Dosage display. When Dosage is altered, treatment time is automatically adjusted.
NOTE: Wavelength selections cannot be changed during a treatment.
Dynatron Tri-Wave™ Light Wavelength Settings
e following tables list wavelength selections or combination of selections with their respective dosage and treatment times.
TRI-WAVE LIGHT PROBE DEFAULTS
WAVELENGTHSDEFAULT DOSAGEPOWER OUTPUT mWTIME
Infrared (IR) – (850nm)6 j/cm
Blue – (464nm)6 j/cm
Red – (624nm)6 j/cm
Infrared and Red6 j/cm
Blue and Infrared6 j/cm
2
2
2
2
2
1000 mW35 sec.
500 mW1 min. 12 sec.
500 mW1 min. 12 sec.
800 mW44 sec.
800 mW48 sec.
Infrared, Blue and Red6 j/cm
Blue and Red 6 j/cm
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 201766
2
2
650 mW1 min.
525 mW1 min. 11 sec.
DYNATRON TRI-WAVE™ OPERATING INSTRUCTIONS
TRI-WAVE LIGHT PAD DEFAULTS
Infrared (IR) – (850nm)10 j/cm
Blue – (464nm)10 j/cm
Red – (624nm)10 j/cm
Infrared and Red10 j/cm
Blue and Infrared10 j/cm
Infrared, Blue and Red10 j/cm
Blue and Red 10 j/cm
2
2
2
2
2
2
2
5160 mW10 min.
2090 mW10 min.
2580 mW20 min.
4128 mW9 min. 50 sec.
4092 mW8 min. 23 sec.
3599 mW8 min. 42 sec.
2802 mW13 min. 20 sec.
6. SAVING CUSTOM SETTINGS
Newly programmed custom settings can be saved as default settings for future treatments by pressing and holding
down the START key until a tone/beep is heard. If selections are not saved, the modality settings will return to the
original default settings once the treatment is complete.
When creating Custom Settings for the Solaris Plus 707, because there are multiple identical accessories (2 Tri-Wave
Light Probes, 2 Pads, and 2 ermoStim Probes), the Custom Default Setting set for one accessory, such as the
Tri-Wave Light Probe, becomes the Default Setting for all Tri-Wave Light Probe treatments. e same is true for the
Tri-Wave Light Pads as well as the ermoStim Probes.
7. START
Probe. Aer pressing START on the console, the yellow LED on the probe handle will be illuminated indicating that
there is power to the probe. Press the 1/0 (ON/OFF) toggle key located on the Probe handle to begin treatment. e
green LED on the probe handle will be illuminated and the treatment time indicated in the Time display window will
begin counting down to zero. While the treatment is in progress and in “focus” LT PROBE (LTPB1 or LTPB2) will
be illuminated in GREEN text in the Channels window. When the countdown reaches zero a tone will sound and the
green LED on the Probe handle will turn o and the yellow LED will be illuminated. Another cycle of the same
duration and dosage may be activated by pressing the 1/0 (ON/OFF) key on the probe handle.
1/0 (On/O/Pause) Key
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 201767
DYNATRON TRI-WAVE™ OPERATING INSTRUCTIONS
CAUTION
e vents on the Probe handle should be clear and free of any obstruction during treatment. Never use gels or lotions
with a light probe treatment.
Expect to hear the probe fan go on and o during a treatment as the fan is designed to maintain a consistent internal
probe temperature. If for any reason the internal temperature of the probe exceeds set limits, the word HOT will
appear on the display and the treatment will be automatically paused until the internal probe temperature returns
to prescribed limits.
Pads. e Dynatron Tri-Wave light pad treatment will immediately begin when START is pressed on the console.
LT PAD (LTPD1 or LTPD2) will be illuminated in green text in the Channels Window when the treatment begins
and is the “focus” treatment. e treatment time will be displayed in the Time Display Window and will begin to
count down to zero. When the treatment time has counted down to “0,” a tone will sound, alerting the practitioner
that the treatment is complete. Another cycle of the same duration and dosage may be activated by pressing the
START key agai
8. PAU SE
Probe. Press the 1/0 (ON/OFF) toggle key located on the probe handle to Pause and Restart a treatment at any time.
When the treatment is active, the green LED on the handle of the probe will be illuminated. When the treatment has
been Paused or is OFF, the yellow LED on the handle will be illuminated.
Pads. With the Dynatron Tri-Wave™ Light Pad as the focus treatment, press the PAUSE key located on the Console.
To continue the treatment, press the PAUSE key again.
9. STOP
Probe. With the Probe as the “focus” treatment, pressing and holding FUNCTION followed by the STOP key will
stop the Probe treatment. Treatment may be paused by pressing the 1/0 (ON/OFF) key located on the probe handle.
Pad. With the Pad as the “focus” treatment, pressing and holding FUNCTION followed by the STOP key will stop the
Pad treatment or allow the treatment to time out.
NOTE: Pressing STOP on the Console will stop ALL TREATMENTS in progress and reset the device to the last
saved preset default values.
Dynatron Tri-Wave™ Light Treatment Notes
• Always cleanse the probe, pad, and treatment area thoroughly before beginning each treatment to avoid skin
irritation or infection.
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 201768
DYNATRON TRI-WAVE™ OPERATING INSTRUCTIONS
• DO NOT use gels or lotions in combination with Light erapy. Gels may clog the vents causing overheating and
damage to the internal components.
• DO NOT attempt to unscrew or tighten down the probe lens. is is not a threaded part. ere are no user
serviceable parts in the Tri-Wave Light probe.
• If the treatment area is larger than the size of the Probe aperture, then areas equal in size to the aperture (6 cm2)
should be treated one at a time until the entire area is covered.
• Do not “bathe” the area by moving the Probe back and forth or in some other manner over the target area.
Failure to make full contact with the treatment surface may result in outcomes that are less eective. Other
methods may reduce total energy absorption. If the area is large, then the Tri-Wave Light Pads (two 5” x 7” pads)
may be a more appropriate treatment option.
• When treating the patient, the Tri-Wave Light Probe or Pads should be placed on the skin over the treatment
area. Maintaining constant contact with the skin is essential during treatment.
• DO NOT use over open wounds unless covered with a clear protective barrier (clear plastic wrap). An open
wound may be treated in the non-contact mode without protective barrier however energy delivery will be
diminished. Undue pressure should not be exerted on the wound bed.
• Never use water or cleansing agents on the Tri-Wave Light Probe Head or Pads. Caustic cleansers (even liquid
cleansers) may cause damage. DO NOT USE ISOPROPYL ALCOHOL. A so cloth lightly damped with
water and a small amount of non-caustic so soap may be used. Wipe with a damp cloth to rinse. If seeking a
commercial cleanser/disinfectant it is recommended that a product contain only the following active ingredients
to avoid damage to the probe or pads:
OctylDecyl Dimethyl Ammonium Chloride
Dioctyl Dimenthyl Ammonium Chloride
Didecyl Dimenthyl Ammonium Chloride
Alkyl (C14 50%; C12 40%; C16 10
Dimenthyl Benzyl Ammonium Chloride
Other Ingredients not published.
WARNING
is therapy must be used cautiously where sensory nerve damage is present or in any case where there is a loss of
normal skin sensation; including areas desensitized by medication or other therapies such as any type of cryotherapy.
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 201769
DYNATRON TRI-WAVE™ LIGHT MODALITY INFORMATION
™
Dynatron Tri-Wave
Light
Modality Information
e Dynatron Tri-Wave™ light probe and Dynatron Tri-Wave™ Light Pad have been cleared by the FDA to provide topical
heating for temporary increase in blood circulation, temporary relief of minor muscle and joint aches, pain and stiness and
relaxation of muscles; for muscle spasms and minor pain and stiness associated with arthritis.
Dynatron Tri-Wave™ Light Basic Vocabulary
Watt (W) ...........................................................................Energy (1 Joule) delivered per second
Milliwatt (mW) One thousandth of a watt. Power and dose determines length of treatment
Nanometer (nm) .................................................................................... One billionth of a meter.
Joule(s) .........................................................................Watts x seconds = total energy delivered
Dose (per cm2) ............ J/cm2 – total energy delivered per square cm in a set period of time.
Dynatron Tri-Wave™ Light Probe And Light Pad Specications
Following are the General Specications for the Dynatron Tri-Wave Light Probe and Dynatron Tri-Wave Light Pad. Other
ranges, accuracy and precision values that are not provided here may be obtained from Dynatronics upon request.
e Dynatron Tri-Wave Light Probe is designed with super luminous diodes emitting three dierent wavelengths of light.
e 20 Infrared SLEDs have a wavelength of 850nm, the 8 Red SLEDs have a wavelength of 624nm and the 6 Blue SLEDs
have a wavelength of 464nm.
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 201770
DYNATRON TRI-WAVE™ LIGHT MODALITY INFORMATION
Dynatron Tri-Wave™ Light Probe Specications
Light Source .......................34 Super Luminous Diodes
Power Output ........................ Max 1000mW (+/- 20%)
Treatment Area........................................................6 cm
2
ermal reshold ............................ 104°F in 2.7 min.
Dynatron Tri-Wave™ Light Pad
e Dynatron Tri-Wave™ light pad consists of two 5” x 7” treatment pads connected with a 24” cable between them and
a 72” bifurcated cable connecting the pads to the Dynatron Solaris Plus controlling console. Dual pads allow for greater
coverage of a joint, treatment of two separate areas simultaneously, and the ability to treat large areas of the body when placed
side-by-side.
Dynatron Tri-Wave™ Light Pad Specications
Light Source .................................................................................. 176 Super Luminous Diodes (per pad)
Wavelengths ........ 112 Infrared Diodes – 850nm; 40 Red Diodes – 624nm; 24 Blue Diodes – 464nm
Power Output ....................................................................................................... Max 5160mW (+/- 20%)
Treatment area .................................................................................................................................... 352cm
2
ermal reshold ........................................................................................................... 104°F in 3.9 min.
Tri-Wave Light Pad - two 5" x 7" pads
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 201771
ULTRASOUND INSTRUCTIONS
Ultrasound
Instructions
e following Ultrasound Instructions are for Solaris Plus 708 and 709 USERS ONLY. e Dynatron Solaris Plus 705, 706, and
707 do not oer the Ultrasound feature.
Ultrasound therapy channels sound waves through muscle, nerve, bone, and connective tissue to aid in reducing pain,
muscle spasms, and joint contractures.
e physiological eect of Ultrasound therapy depends upon the frequency of the Ultrasound signal. e lower frequency
(1 MHz) penetrates deeper than a higher frequency (such as 2 MHz or 3 MHz), thus the practitioner can decide which
frequency to use according to the condition and depth to be treated.
A section in this manual entitled “Ultrasound Usage Cautions” provides some general guidelines for Ultrasound treatment
and selection of the appropriate soundhead to help ensure safe and eective treatments are delivered to your patients. Further
information about Ultrasound application may be obtained from published medical literature.
WARNING
• ALWAYS keep the applicator soundhead in constant motion.
• ALWAYS keep the soundhead properly coupled to the patient’s skin or submerged underwater when intensity
is turned on.
• Use ample conductive gel to ensure good coupling throughout the treatment. If needed, apply additional gel
during the treatment.
• See the section of this manual entitled “Contraindications, Warnings, and Precautions” for Ultrasound treatments.
• Be alert for any sign of periosteal (bone) pain.
• Be sure to read all instructions for operation before treating a patient.
• Use of controls or adjustments or performance of procedures other than those specied herein may result in
hazardous exposure to ultrasonic energy.
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 201772
ULTRASOUND INSTRUCTIONS
CAUTION
• Do not drop the soundhead on the hard surfaces.
• Do not cool the soundhead with ice water or ice packs.
• Do not allow the soundhead to overheat repeatedly.
• Do not hold the soundhead in the air while a treatment is running.
All of these conditions are likely to damage the soundhead crystal and/or to stress electronic components in the
device. Damage caused by these conditions is not covered by warranty.
Make sure a soundhead is rmly plugged into the device before turning the device on. When changing to a dierent size
soundhead, turn the machine o rst, remove the soundhead, plug in the desired soundhead, then turn the machine on
again. Please acquaint yourself with the following terms and device features prior to delivering an Ultrasound treatment.
Soundhead Warming
Soundhead Warming is an optional feature used to maintain a comfortable soundhead temperature for the patient. When
Soundhead Warming is ON, the soundhead should remain in its holder as a small amount of Ultrasound output is emitted
from the soundhead (0.1 W/cm2). e soundhead warming mode is automatically stopped during a treatment, and resumes
automatically as needed aer a treatment has ended.
If Soundhead Warming is ON and SOUND has been selected but is not the focus treatment, the word SOUND in the
CHANNELS window will be illuminated ORANGE.
Although the Soundhead Warming feature defaults to OFF, it can be turned ON at any time. To turn ON the feature, select
SOUND, followed by FUNCTION key. Using the WARMING toggle key, select ON.
Coupling
e term “coupling” refers to the ability to deliver ultrasonic waves from the soundhead to the skin surface with as little
impedance or dissipation of power as possible. Coupling (contact between the soundhead and the treatment site) may be
provided by a coupling agent such as a gel or lotion. Any material used as a coupling agent must be highly conductive of
ultrasonic waves. Air is a very poor conductor of ultrasonic waves. Holding the soundhead in the air while a treatment is
running may also damage the soundhead crystal and/or stress electronic components in the device.
If any part of the soundhead is exposed to air during the treatment, coupling is decreased. e air bubbles in a whirlpool, for
example, can decrease the eective Ultrasound therapy to the patient. Avoid allowing any air between the soundhead and
the treatment area. Water is an excellent conductor of ultrasonic waves; therefore, underwater treatments provide excellent
coupling.
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 201773
ULTRASOUND INSTRUCTIONS
During any Ultrasound treatment the soundhead should be moved continuously, covering an area approximately twice
the size of the soundhead. e full surface of the soundhead should maintain contact with the patient’s skin (except with
underwater treatments).
Head Temperature Hot Display
If coupling (the eective degree to which the Ultrasound energy is delivered from the soundhead to the patient’s body) is not
adequate during treatment, the temperature of the soundhead rises and the patient does not receive the full intended dosage.
e Dynatron Solaris Plus TEMPERATURE bar reects the amount of soundhead heating caused by poor coupling to ensure
that the patient is receiving the optimal treatment and that the soundhead crystal is protected from overheating.
When the coupling is acceptable, the length of the Blue/Green segment lights on the coupling bar will remain less than half
the length of the bar or less. If the soundhead approaches a temperature of 103 degrees Fahrenheit, the TEMPERATURE
BAR begins to increase in length and the colored bar moves past the center mark and continues to lengthen.
If the SOUNDHEAD reaches approximately 103 degrees, a caution will appear in the Treatment Display Screen: “CAUTION,
SOUNDHEAD IS GETTING HOT.” Following the caution, the treatment should be terminated and the soundhead cooled.
If the SOUNDHEAD reaches approximately 108 degrees, SOUND will be disabled and the Treatment Display Screen will
read: “SOUNDHEAD IS TOO HOT! OUTPUT HAS BEEN DISABLED TO ALLOW COOLING.”
NOTE: If the soundhead becomes too hot the SOUNDHEAD HOT warning will appear in the Treatment Screen whether
SOUND is the focus treatment or not.
e soundhead must then be cooled down before the treatment can resume. When the soundhead cools suciently, press
PAUSE or START to resume the treatment. e output power resumes, the display returns to its normal state, and the timer
resumes. e soundhead should cool quickly if placed in the soundhead holder or if held exposed to the air. Larger soundheads
take longer to cool than smaller heads. If the soundhead is not cooling as quickly as needed to resume the treatment, it can
be placed in room temperature water to quicken the cooling process. Sometimes just applying more conductive gel will
adequately cool the head.
NEVER USE ICE OR ICE PACKS TO COOL THE SOUNDHEADS as this is likely to cause thermal shock to the electronic
components of the soundhead and may necessitate a costly repair. Heads damaged by thermal shock are not covered by the
warranty.
To prevent overheating of the soundhead, maintain good coupling throughout the treatment by applying ample conductive
gel or lotion. Reducing the power when treating an area where it is dicult to obtain good coupling will also keep the
soundhead from overheating.
Display Watts or W/cm
2
Power for the Dynatron Solaris Plus may be displayed as WATTS or W/cm2. To choose the desired option, select SOUND,
then press the DISPLAY toggle key under the DISPLAY window and select WATTS or W/cm2. e default setting for power
is W/cm2; however, the display you prefer may be selected at any time before or during a treatment. Power selection may be
saved by pressing and holding down the START key until a beep sounds.
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 201774
ULTRASOUND INSTRUCTIONS
Ultrasound Quick Setup
Select SOUND.
1. FREQUENCY. Using the FREQUENCY toggle key select 1 MHz, 2 MHz, or 3 MHz.
2. DUTY CYCLE. Using the DUTY toggle key, select 10%, 20%, 50% or CONT (Continuous).
3. TIME. Change the treatment TIME, if desired using the Up/Down arrow keys.
4. INTENSITY. Using the Up/Down arrow keys, raise the INTENSITY to desired level.
5. Press STA RT .
6. S T O P. Press and hold the FUNCTION and press the STOP key to stop only the focus treatment appearing
in the Treatment Display Screen. Pressing STOP alone, stops all treatments operating on the device. Using
the TIME arrow keys to bring treatment time to zero will also stop a treatment.
Detailed Ultrasound Setup
1. Press SOUND.
Press SOUND to select an Ultrasound treatment. e Default Parameters automatically appear in the Treatment
Display Screen. Following are the Ultrasound Default Parameters:
Frequency ........................................1 MHz
Warming .............................................. OFF
If you wish to use the default settings, increase the Intensity to desired treatment level and press START. If you wish
to customize settings, follow steps outlined below.
2
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 201775
ULTRASOUND INSTRUCTIONS
2. Choose the FREQUENCY.
Press the FREQUENCY toggle key located under the FREQUENCY window to select 1, 2, or 3 MHz. Any one of the
three Frequencies may be selected with the 2 cm2, 5 cm2 or 10 cm2 soundhead.
3. Select the DUTY CYCLE.
Press the DUTY CYCLE toggle key to select one of the four available options: 10%, 20%, 50%, or Continuous duty cycles.
4. Press the FUNCTION key located on the console to display additional parameters.
5. HEAD WARMING.
Press the FUNCTION key to access the Head Warming option. Using the WARMING Toggle key the Head Warming
feature may be turned ON or OFF.
6. TIME.
e default time is set for a 5 minute treatment. Time can be changed by using the TIME Up/Down arrow keys
located to the le of the TIME display.
7. Raise the INTENSITY.
Use the INTENSITY Up/Down arrow keys to increase the power to the desired setting. For patient safety and
comfort, it is recommended that treatment begins with .1 W/cm2, then increase power to the desired level aer the
treatment begins. Valid ranges are from 0.1 to 2.0 W/cm2 (exceptions: valid ranges when using a 10 cm2 head at 3
MHz are from 0.1 to 1.0 W/cm2).
8. Press STA RT .
Press START, the treatment timer begins counting down and output is delivered to the soundhead. If you fail to
set the Intensity before pressing START, a reminder will appear in the lower-right corner of the Treatment Display
Screen: “CANNOT START TREATMENT WITH ZERO INTENSITY.”
9. SAVING DEFAULTS. If the treatment you have just set up is the most common Ultrasound setup you use, new
defaults may be saved by pressing and holding down the START key for two seconds. At the end of two seconds,
a beep will sound indicating the treatment parameters have been saved. e next time SOUND is selected, these
parameters will be selected automatically.
10. MODIFY a treatment in progress, if desired.
While the treatment is in progress, the following parameters can be modied: FREQUENCY, DUTY CYCLE, TIME,
INTENSITY, and DISPLAY of Watts or Wcm2.
11. PAUSE. Temporarily PAUSE a treatment, if necessary, while the treatment is in progress.
To temporarily PAUSE an Ultrasound treatment, press the PAUSE key. Two quick tones will sound indicating that
the treatment has been paused. e Ultrasound output from the soundhead stops and the treatment timer is paused
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 201776
ULTRASOUND INSTRUCTIONS
without ending the treatment. Press the PAUSE key again to restart the treatment. A tone will sound indicating that
the treatment is again in progress. Output resumes and the treatment timer starts from where it was paused.
NOTE: During a COMBO treatment, THE STIM OUTPUT OF THE TREATMENT IS NOT PAUSED when the
PAUSE key is pressed, although the Ultrasound output is stopped and the treatment timer is paused.
12. S T O P.
When the treatment time has elapsed, the ultrasound to the patient stops and a tone sounds signaling the end of a
treatment. Treatments in progress may be stopped at any time using one of the following methods.
Stop One Treatment Only. Press and hold the FUNCTION key and press STOP. is stops only the treatment in focus.
Stop All. Press the STOP key. All treatments at all channels will stop.
Stop Time. Reduce the treatment time using the Time arrow key. e output at the selected channel is stopped,
and the device then displays the parameters for the next treatment that remains in progress (if there is an active
treatment).
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 201777
ULTRASOUND MODALITY INFORMATION
Ultrasound
Modality Information
For Dynatron Solaris® Plus 708 and 709 users only. e Dynatron Solaris® Plus 705, 706, and 707 do not oer the Ultrasound feature.
Ultrasound, by its very nature, has the ability to irritate the patient’s skin. While the benets of Ultrasound far outweigh any
disadvantages, certain precautions should be observed to assure maximum safety and comfort for your patients.
A patient’s tendency to have adverse reactions to Ultrasound is dependent upon several factors. Some of these factors are
discussed below.
Selecting the Appropriate Soundhead
Head and Crystal Size Comparison
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 201778
ULTRASOUND MODALITY INFORMATION
e selection of the appropriate soundhead is key to the success of the treatment and is based on the size of the area to be
treated. Ultrasound treatments should be kept specic to the tissue involved in pathology. A good guideline is 2 to 4 times
the size of the soundhead. For example:
• A 2 cm2 soundhead can deliver up to 4 Watts and is appropriate for small areas (i.e. hands, ngers, feet).
• A 5 cm2 soundhead can deliver up to 10 Watts and is appropriate for medium sized areas (i.e. extremities such as
arms, legs, and cervical areas).
• A 10 cm2 soundhead can deliver up to 20 Watts and is appropriate for large areas, i.e. torso and back).
Ultrasound is a directed beam of energy. erefore, not only will the average spatial intensity be a factor in the dosage the
patient receives, but the time delivered and area covered will matter as well. For example, an area of 50 cm2 is treated for 5
minutes. en an area of 200 cm2 is treated for 5 minutes. Both receive the same intensity. e 200 cm2 area however does
not receive the same dosage (only ¼ ) because as the soundhead is moved around the area it has to cover represents 4 times
as much tissue.
e Soundhead area measurement is the ERA (eective radiating area). Each soundhead has an eective radiating area. It is
not necessarily the outside diameter of the soundhead, but the area of the crystal inside, therefore special care should be
taken in selecting the correct size soundhead for the area to be treated according to the diameter of the crystal.
NOTE: If a patient experiences pain during a treatment, the size of the soundhead maybe inappropriate for the area
being treated, the intensity maybe too high, the treatment time maybe too long, or coupling maybe poor.
Penetration of Ultrasound Waves
e correct frequency should be selected for the depth of penetration desired. e amount of penetration needed is
determined by the density of tissue and the depth of the site to be treated. Care should be taken to select a penetration level
that does not cause periosteal (bone) pain.
e frequency determines the depth of penetration of
the Ultrasonic wave.
• Select 1 MHz for deep lesions; provides a
Half-Value Distance (HVD) of about 5 cm.
• Select 2 MHz for moderate depth lesions; about
2.6 cm HVD.
• Select 3 MHz for supercial lesions; about 1.5
cm HVD.
HVD is the approximate point at which the Ultrasound
energy is reduced to half in the average human tissue.
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 201779
Multi-Frequency Ultrasound
ULTRASOUND MODALITY INFORMATION
Types of Delivery
Ultrasound can be delivered in four dierent ways. You will likely only see two of the four methods in clinical practice.
1. Direct Contact Movable. Here the soundhead is placed in direct contact with the patient. A coupling agent is used
between soundhead and the patient’s skin. e soundhead is moved slowly, but continuously. is is the method of
choice.
e rate of speed at which the applicator moves across the skin is very important in determining how much Ultrasonic
output is delivered. If the rate is too slow, the patient may feel periosteal pain (bone ache/pain). If the rate is too fast,
or if the applicator head becomes uncoupled with the skin, the amount of treatment is reduced. Uncoupling can also
cause the soundhead to overheat.
2. Immersion Method. Here the area to be treated is placed underwater. e soundhead is water tight so it can be
immersed with the area to be treated. e water becomes the coupling agent. e head is always moving around the
surface area, but not in contact (1/2 to 1 inch away).
3. Hydrogel Disk. For treating crater wounds, cover the wound with a hydrogel disk and apply the soundhead to the
disk. is allows direct wound sonation without bringing the soundhead in direct contact with the wound.
4. Stationary Soundhead. is method is dangerous. Hot spots can develop. Do not use.
Treatment Time
For Sub-Acute Conditions:
For Chronic Conditions:
For Maximal ermal Eect:
area to be treated (cm2)
= minutes of treatment
1.5 x ERA
area to be treated (cm2)
= minutes of treatment
1.0 x ERA
area to be treated (cm2)
= minutes of treatment
.8 x ERA
Treatment Intensity
Several factors come into play as one decides the level of intensity for the treatment.
1. Supercial lesions require less intensity.
2. Less intensity should be used if bone is supercial to the treatment eld.
3. Less intensity should be used when the stage of the injury makes heating questionable.
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 201780
ULTRASOUND MODALITY INFORMATION
4. Use a little lower intensity for the rst treatment to gauge response.
5. Patient feedback is key. A treatment should feel warm, but the patient should never feel heat, pain, stabbing, pricking
or dull ache.
NOTE: It is very common that intensity is always 1.5 W/cm2. is is incorrect in many cases. A more specic
intensity should be used based on patient response and stage of injury.
Frequency of Treatment
Treatment can be given daily. It is not uncommon to give Ultrasound twice daily, but this may be excessive. Some guidelines
may be helpful.
1. Daily may be the best maximum frequency.
2. Ultrasound can be eectively given every other day.
3. Ultrasound should give some positive benets by the 3rd or 4th application. If not, discontinue the treatment and
consider other options.
4. A maximum of 12 to 15 Ultrasound treatments should be given. If the result desired has not been reached by this
point, Ultrasound may not be the proper choice. EXCEPTION: Some Chronic conditions which cause adhesions.
Usage Cautions – Combination Treatments
When using a Stim device in conjunction with a Solaris Plus device to output Stim through the soundhead, observe all contraindication, warnings, precautions, and usage cautions provided by the manufacturer for all modalities involved.
Potential for Burns or Periosteal Pain
Some patients’ skin is more sensitive to Ultrasound output. is can cause a reaction similar to a heat rash. It is also possible
for a patient to suer a burn from Ultrasound therapy if the therapy is not administered properly. is can occur for the
following reasons:
• Intensity (power) too high
• Frequency too low
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 201781
ULTRASOUND MODALITY INFORMATION
• Holding the soundhead in one place on the patient’s skin
• Moving the soundhead too slowly
• Treating an area where sensory nerve damage is present with a loss of normal skin sensation
• Time (Caution: Don’t treat too long).
Bony prominences are especially susceptible, as they reect sound waves and increase intensity to the periosteum resulting
in a burning sensation. Desensitized areas can be overheated or burned without the patient realizing it, so extreme care must
be taken with these patients (e.g. diabetes, neural damage, etc.)
Burns can be avoided as long as the treatment causes no pain, tingling, excess heat or aching (for patients with normal skin
sensation). Use sucient coupling agent and make sure there are no bubbles in the gel. When treating in water, clear the
bubbles o the soundhead and o the patient’s skin.
An un-calibrated soundhead can also cause tingling, excess heat, aching, or a burning sensation.
Read Ultrasound Contraindication, Warning, & Precaution in this manual for more information.
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 201782
SOUNDHEAD OPTIMIZATION ADDING OR REPLACING SOUNDHEADS
Soundhead Optimization
Adding or Replacing
Soundheads
Entering Parameters for Solaris Plus 708 and 709
CAUTION
Soundheads CAN successfully be moved between dierent Solaris devices ONLY if the parameters for that soundhead
are entered into the device. Entering parameters ensures the optimization of soundhead output.
To ensure soundhead output is optimized, please carefully follow the instructions provided. If you have any questions about
the following instructions, contact Dynatronics’ Customer Service Department before proceeding ((800) 874-6251).
1. Head Calibration Printout.
Soundheads are shipped with a Head Calibration Printout sheet. e sheet contains calibration numbers unique
to the each specic soundhead. e soundhead serial number appears on the sheet to assist in matching the correct
soundhead with the printed parameters. e printout should be kept for future use.
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 201783
SOUNDHEAD OPTIMIZATION ADDING OR REPLACING SOUNDHEADS
Dynatronics Soundhead Sample Parameters Printout
Serial: Manufacture Date: 9/15/2012
Size (cm): 5
Cal Date: 9/15/2012
Unit #: Solaris Plus
IHT:
1 MHz2 MHz3 MHz
Frequency
Impedance
Temperature
NOTE: THE NUMBERS ABOVE ARE PROVIDED FOR ILLUSTRATION ONLY AND SHOULD NOT BE
ENTERED INTO YOUR DYNATRON DEVICE. USE ONLY THE ACTUAL NUMBERS PROVIDED WITH
YOUR OWN SOUNDHEAD.
2. Turn the Solaris Plus Device OFF. Plug in the new soundhead. Turn the Solaris Plus ON.
3. Press the SOUND key followed by the FUNCTION KEY.
NOTE: e Solaris Plus device will automatically recognize the size of the soundhead plugged into the console. To
verify that you are entering the parameters for the correct soundhead, make sure that the HEAD SIZE that appears
on the le-hand side of the Treatment Display Screen is correct. For example, if “5” is displayed, it means the device
2
senses a 5 cm
next to the HEAD SIZE display to select the correct soundhead size.
soundhead is plugged into the console. If correct soundhead size is not visible, use the arrow keys
175019502050
4611064699
Example / Do NOT USE
888
4. As a default, FREQUENCY should appear in the PARAMETER window and 1 MHz in the FREQUENCY window.
Using the Up/Down arrow keys located next to the FREQUENCY display, select the FREQUENCY value for 1 MHz
listed on the Parameter’s printout.
NOTE: Parameters must be entered for all three FREQUENCY options 1 MHz, 2 MHz, and 3 MHz shown on the
Parameters printout for Frequency, Impedance, and Temperature.
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 201784
SOUNDHEAD OPTIMIZATION ADDING OR REPLACING SOUNDHEADS
5. Using the FREQUENCY Up/Down arrow keys select 2 MHz. When it is illuminated GREEN, use the Up/Down
arrow keys next to the FREQUENCY window to select the FREQUENCY value for 2 MHz listed on the Parameter
Printout. Follow the same steps to enter the parameter value for 3 MHz.
6. Aer entering the FREQUENCY parameters, reset the FREQUENCY to 1 MHz and use the PARAMETER toggle
key to select IMPEDANCE. Follow steps 4 and 5 to enter the 1 MHz, 2 MHz, and 3 MHz parameter values for
IMPEDANCE, and TEMPERATURE.
7. STORING THE NEW PARAMETERS.
Aer entering all the parameters, press START and hold until a tone sounds indicating that the Parameters have
been stored. Press the FUNCTION key to return to the SOUND treatment screen.
e above procedure must be performed for each separate soundhead used with the Solaris Plus device. Turn the
device OFF before attaching the next soundhead, then turn the device ON again with the soundhead rmly plugged
into the console.
Ultrasound Calibration
To maintain accuracy, all soundheads must be calibrated annually to ensure proper operation. With the exception of
calibration, all service on the Dynatron Solaris Plus device should be performed by a Dynatronics service technician. If
your Dynatron Solaris Plus requires service, contact Dynatronics Customer Service at (800) 874-6251. Calibration MUST
be performed either by Dynatronics or by a qualied Ultrasound technician in your local area. Qualied technicians may
contact Dynatronics for the calibration service procedures or refer to the Solaris Plus Service Manual.
What to Calibrate. Calibrate all soundheads used with this device at 1, 2 and 3 MHz Frequencies.
Equipment Required. e process requires an ultrasound power meter capable of accurately measuring outputs up to 3
MHz. Check the manufacturer’s specications to conrm your power meter meets this qualication. Ohmic Instrument
UPM-DT1 or UPM-DT-10 is recommended for use.
Frequency Numbers. e “Frequency” numbers for each soundhead are required. e frequency numbers for each
soundhead may be obtained by calling Dynatronics Customer Service or referring to the Dynatronics Soundhead Calibration
Printout shipped with each Dynatronics’ soundhead.
Water Quality. Water used in the testing procedure must be degassed water with an oxygen content of four parts per million
(4ppm) or less.
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 201785
ULTRASOUND PROBLEM SOLVING
Ultrasound
Problem Solving
Whirlpool Treatments
If you are treating in a whirlpool, you may nd the temperature reaches a high enough temperature to read approximately
103˚F causing the overheated soundhead caution to appear in the Treatment Display Screen. is is a warning only to let
you know you are approaching the temperature limit. You may, however, continue with the treatment at this level. If your
whirlpool temperature is hot enough to cause the treatment to stop, you will need to adjust the temperature of the whirlpool.
Soundhead Temperature Too Cold
If the soundhead has been sitting in a very cold room or vehicle, it could be too cold to operate when you plug it into the
console. e keypad may not respond to key presses and you will be unable to use the device until the soundhead is suciently
warmed. You must raise the temperature of the soundhead to about 60 degrees F in order for the machine to recognize that the
soundhead is present and to proceed with setting up a treatment. You can accomplish this with any of the following methods:
1. Press the at face of the soundhead against the palm of your hand for 30 to 60 seconds to warm it slightly. is
usually provides adequate warmth to the crystal to raise the temperature to the minimum acceptable level. Once the
crystal reaches this level, you can proceed with treatment.
2. You can also place the soundhead in room temperature water to warm the crystal. However, do not place the
soundhead in very hot water when the crystal is this cold as it could damage the crystal.
No Soundhead
If the device cannot detect a soundhead during setup or delivery of an Ultrasound treatment, the error message “SOUNDHEAD
IS NOT CONNECTED, HEAD WARMING WILL BE DISABLED!” will appear in the Treatment Display Screen.
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 201786
ULTRASOUND PROBLEM SOLVING
If this error occurs, check to be sure the soundhead is rmly plugged into its connector. If you are unable to clear the message
by reconnecting the soundhead, contact Dynatronics’ customer service department at 1-800-874-6251 for assistance.
Miscellaneous
Certain conditions can cause an error in operation. When this occurs, the machine will not allow a treatment to be set up or
delivered and will display an error message. Some errors are easily resolved by the following methods.
• Press STOP to stop the treatment, and turn the machine OFF then ON again. Always wait 5-10 seconds before
restarting the device.
• Check to be sure the soundhead has not become disconnected from the machine. e soundhead should be rmly
plugged into its port. Only Dynatronics soundheads may be used with this device. If the soundhead has been
dropped, it may be damaged. If the device operates normally with one soundhead, but not with another, the problem
may be a damaged soundhead and you must contact Dynatronics Customer Service.
• Make sure the soundhead is not too hot. In this case the Soundhead alert will appear in the Treatment Display Screen.
• Check to see if conditions may have caused extreme moisture condensation in the device. is could occur when
the machine has become very cold then is brought indoors to a warm, humid environment. Condensation is a not
a serious condition. Allow the machine to sit in a dry environment until the condensation dries. e machine will
operate normally once the condensation is gone.
If you have tried all of these suggestions, the device may require service by the manufacturer. In this case, make a note of the
error message and the sequence of events that cause the error, and contact Dynatronics Customer Service at 1-800-874-6251
for further assistance. Do not send the device to Dynatronics without rst contacting the Customer Service Department.
e Dynatron Solaris Plus 708, and 709 comply with the following:
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 201787
ULTRASOUND PROBLEM SOLVING
• FDA 21CFR 1050(c)(1)(i). e error in indication of the temporal-average ultrasonic power shall not exceed ±20
percent for all emissions greater than 10 percent of the maximum emission.
• FDA 21CFR 1050(c)(1)(ii). e sum of the errors in the indications of temporal-maximum ultrasonic power and the
ratio of the temporal-maximum eective intensity to the temporal-average eective intensity shall not exceed ±20
percent for all emissions greater than 10 percent of the maximum emission.
• FDA 21CFR 1050.10(c)(2). e treatment timer must be accurate to within 0.5 minute of the preset duration of
emission for settings less than 5 minutes, to within 10 percent of the preset duration of emission for settings of from
5 minutes to 10 minutes, and to within 1 minute of the preset duration of emission for settings greater than 10
minutes.
NOTE: e Dynatron Solaris Plus 708 and 709 are accurate to within ±1% of any treatment time.
Pursuant to FDA 21CFR 1050.10(f)(1), the uncertainties in magnitude, expressed in percentage error, of the ultrasonic
frequency, eective radiating area, and the ratio of the temporal-maximum to temporal-average eective intensity, pulse
duration, and pulse repetition rate for the Dynatron Solaris Plus 708 and 709 are as follows:
(1) Ultrasonic frequency ....................................................................................................... ±15%
(2) Eective Radiating Area ................................................................................................. ±20%
(3) Ratio of the temporal-maximum to temporal-average eective intensity ............... ±20%
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 201788
ULTRASOUND BEAM PROFILES
Ultrasound
Beam Proles
(For Dynatron Solaris Plus 708, and 709 users only. e Dynatron Solaris Plus 705, 706, and 707 do not oer Ultrasound).
e following diagrams show the typical spatial distribution of the radiated eld for each size of Dynatron Solaris soundhead.
is applies to the radiation emitted into the equivalent of an innite medium of distilled, degassed water at 30˚ C and with
line voltage variations in the range of ±10 percent of the rated value.
10 cm2 Head. Near Field
1MHz
2 MHz
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 201789
3 MHz
ULTRASOUND BEAM PROFILES
5 cm2 Head. Near Field
1MHz
2 MHz
3 MHz
2 cm2 Head. Near Field
1MHz
2 MHz
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 201790
3 MHz
COMBINATION THERAPY INSTRUCTIONS
Combination
Therapy Instructions
e following combination therapy instructions are for Dynatron Solaris® Plus 708 and 709 USERS ONLY. e Dynatron
Solaris® Plus 705, 706, and 707 do not oer the Ultrasound feature.
WARNING
• DO NOT use combination therapy for underwater treatment. Placing active electrodes underwater poses a
serious hazard to the patient!
• Use VERY LOW STIM INTENSITY for COMBO treatments.
• Remember to observe all contraindications, warnings, precautions, and usage cautions for BOTH Ultrasound
and Electrical Stimulation therapy when performing combination therapy.
• Since electrical current travels between the electrode and the soundhead during a COMBO treatment, the
electrode should be placed in proximity with the treatment area. Do not place the electrode and soundhead in
positions that will cause current to pass through contraindicated areas.
• Avoid removing the soundhead from the skin surface during “Stim rough Soundhead” treatments as
this may cause a momentary interruption of Stim current which may be uncomfortable to the patient. e
soundhead should remain in full contact of the skin until current output is stopped.
• Be alert for any sign of periosteal (bone) pain.
Comboplus
™
Dynatronics’ Comboplus feature means you have almost unlimited options in setting up a combination treatment with
Solaris Plus. You can:
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 201791
COMBINATION THERAPY INSTRUCTIONS
• Combine an Ultrasound treatment with the following electrotherapy modalities provided by this device: IFC,
Premodulated, Biphasic, Russian, High Volt, or Microcurrent.
• Set up a combination treatment by using the ULTRASOUND output jack and the automatically selected default
STIM channel. When using the 708 and 709, the Microcurrent Combo will always use Channel 1, and High Volt
Combo treatments will default to the High Volt (Hi-Volt) channel.
A special “COMBO” lead wire is provided with the 708 and 709 with the standard accessories for this device to accommodate
the ComboPlus feature. is lead wire is plugged into the selected STIM jack; then the banana end of the lead is connected to
the COMBO input jack on the right-side of the device, and the pin end of the lead is connected to an electrode to be placed
on the patient. It is important to note the channel selected by the device during setup and connect the lead wire to the correct
channel before setting intensity for the treatment.
Stim Through the Soundhead
With combination therapy, the soundhead is used in place of one electrode for a Stim treatment; and electrotherapy current
is delivered through the soundhead. is means that for a normal 2-electrode Stim treatment therapy, you would place
one electrode on the patient and use the soundhead as the second electrode site to complete the setup. A patient lead wire
designed to accommodate this setup is included in the Dynatron Solaris Plus standard accessory package for devices with
Ultrasound and electrical stimulation capabilities.
During the treatment, the Stim current passes between the soundhead and the other electrode. At the same time ultrasonic
waves are introduced into patient tissue through the soundhead. Avoid touching the electrode with the soundhead during
the treatment, keep the soundhead in contact with the patient’s skin at all times, and keep the intensity low for the Stim
current.
When setting up a combination treatment, observe all contraindications, warnings, and precautions for both
therapies to be used.
REMEMBER: Use very low Stim intensities for all COMBO treatments!
In order to set up a COMBO treatment, you must be familiar with setup instructions for both the electrotherapy modality to
be used and Ultrasound as explained earlier in this manual. Also remember:
• When a modality indicator is highlighted GREEN, the treatment parameters for that output are displayed. Any
changes you make to the parameters will aect that channel only.
• When a modality indicator light is highlighted YELLOW, the channel is active, but its parameters are not being
displayed (in focus) and may not be modied at this time. To bring a treatment’s parameters into focus, you must
rst press the CHANNELS toggle key to select the treatment—the modality indicator will then become GREEN and
modications are allowed.
DYNATRON SOLARIS® PLUS SERIES | SERVICE MANUAL REV. 6 | AUGUST 29, 201792