The HM35 digital pressure gauge is a pressure-measuring instrument with an integrated
pressure sensor for the measurement of differential, relative or absolute pressures and
vacuum. Its versatile range of functions and high precision render it suitable for a wide
range of applications. Via the infrared interface (IR) and SCPI (Standard Commands for
Programmable Instruments) commands, the HM35 can communicate with a PC. Its
operation is very simple, and supports the user in his measurement tasks.
Operating modes
• Pressure measurement / Differential pressure
• Min./Max. values
• Mean value (average)
• Pressure change rate
• Data logging
Selectable configuration possibilities
• Data logging
Æ Interval time, print/transfer, deleted memory
• Configuration
Æ Measurement units, display filter, auto. switch-off time, auto zero, lighting level, etc.
• Average period (period for determining average value)
• Date and Time (real-time clock)
• Calibration
Æ Date of last calibration date, manual recalibration
2 Safety information
•The pressure values and overload levels stated on the rating plate and quoted in
these operating instructions must not be exceeded, as otherwise the pressure sensor could be destroyed or there could be a risk of injury.
•Only use pressure hoses with a maximum loading capacity corresponding to that
necessary for the application.
•Ensure that the pneumatic hoses are securely fitted! Do not use damaged or kinked
hoses.
• Do not open up the instrument (this would void the guarantee).
• The instrument must be stored within the permissible storage temperature range.
The instrument must not be put into operation in an
explosive environment!
Wear eye protection if working with pressures > 1bar!
3
3 Operation
3.1 Switching on and off
Switching on Briefly press the On/Off key () (< 1 s)
For precise measurements, the HM35 must first be switched on for at
least 1 minute (warm-up phase).
Switching off Briefly press the On/Off key
or
automatic switch-off 3, 10 or 60 minutes after the last time key operation
(automatic switch-off does not take place during Average, Change Rate
and Data Logging measurements or in IR
Notes
• The HM35 switches on automatically when the supply voltage is connected.
• The HM35 continues to work in battery mode following an interruption of the supply
voltage
•In case of a change in temperature, the HM35 must be allowed to adapt to the new
ambient temperature for a least 30 minutes while switched off in order to attain the
best measurement accuracy.
• The day/time, battery level and accuracy will be briefly displayed at switch on:
• After switch-on, the HM35 switches to the last operational mode used, e.g.:
() (< 1 s),
and network operation).
• With the display filter activated, wait until the transient effect finishes (approx. 5 s).
4
3.2 Lighting
Switching on Press the On/Off-key
() for > 1 s
Brightness control In the Menu, select the Configuration → Lighting function and
select an adjustment of Off, Level 1, Level 2 or Level 3.
Switching off Briefly press the On/Off key
() (< 1 s)
(switch off the instrument),
or automatic switch-off after 20 s.
With mains operation, the HM35 must be switched off manually.
3.3 Pneumatic connection
Designation Pressure range
Hose 4/6 mm
NPT1/8“ internal 10 ... 90 bar
Plug in nipple „Rectus“ Type 20
M10 x 1 internal thread (for „Minimess“ connector) all
≤ 7,5 bar
≤ 30 bar
Ensure that the pneumatic hoses are connected correctly!
+P Higher pressure
S- Lower pressure (not available with the absolute and relative pressure version)
When screwing onto a coupling, it is important to hold the
coupling steady with a wrench to prevent any turning!
Never secure by holding the casing itself!
5
3.4 Functions and operating modes
Key clear hold
Functions
PRESSURE
DIFFERENCE
▲ MIN/MAX Sets Max/Min to
▼ AVERAGE 4)
CHANGE RATE 5)
DATA LOGGING
Zero: sets
measured value to
3)
zero
current measured
value
Sets measured
value to zero
3)
Freezes all
current
measurement
values
--
To Menu selection
1)
Stop/Menu
Start 2)
--
Starts measuring
Starts data
logging
Notes
1) Once a measurement procedure has been started, the menu selection is blocked.
2) During a measurement procedure (after Start has been pressed), you can switch
between functions. This permits, for example, the observation of the Min/Max
function during data logging.
3) The Clear key has no function in the absolute pressure instruments.
4) The AVERAGE function creates an arithmetic average value of all measured values
during the time period selected in the menu. After expiry of the time period, the
average value will be displayed.
5) Measurement of the leak rate (diff/gauge sensor) or tendency (abs. sensor). The
pressure change (CHANGE RATE) from the start time to the current time will be
displayed. The first display occurs 10 s after the start.
3.5 Menu selection and set-up
Navigation within the Menu selection
clear
▲▼
Brief (< 1 s) 1 level back
Long (> 1 s) Back to the function level/operating modes
Selection of Set-up/Functions
The functions shown inverted on the display will be carried out if the Function key
) is pressed.
(
The currently selected setting for values is marked with '9'. In the following table, the
default values are correspondingly marked (factory settings).
6
Key
Data logging
Interval Interval period
manual
25 M./s
10 M./s
1 s 9 2 s
5 s
10 s
30 s
1 min
2 min
5 min
10 min
30 min
1 h
3 h
6 h
12 h
24 h user Set with ▲/▼/ EDIT/OK
▲ Print Data Logging
▼ Press 'Start' Print/send via IR
Clear Memory
Press 'Clear' Deletes the data memory
Configuration
Pressure Unit 9 mbar
bar
Pa
hPa
kPa
MPa
kg/cm2
kg/m2
mmHg
cmHg
mHg
inHg
mmH2O
cmH2O
mH2O
inH2O
ftH2O
psi
lb/in2
lb/ft
torr atm
2
Notes
7 8
Key
Display Filter Filters the display values 1)
9 On Off
Auto-Off Auto. switch-off
3 min 9 10 min
60 min
Auto-Zero
On Sensor auto-zeros at switch-on
9 Off
Beep Warning beeper
9 On
Off
Lighting Off Level 1 Only Level 1 possible for Ex-models
Level 2
▲ 9 Level 3
▼ IR Interface
9On At switch-on, the automatic
Off Automatic connection is de-activated
Average period Time period for average value
10 s
30 s
1 min
2 min 9 5 min
10 min
30 min
1 h
3 h
6 h
12 h
24 h user Set with ▲/▼/EDIT/OK
Date & Time dd.mm.yyyy Set with ▲/▼/ EDIT/OKhh:mm:ss Set with ▲/▼/ EDIT/OK
Calibration
History Displays the last calibration date
Manual re-calibration Manual re-calibration of the zero
Notes
if measured value < 1% FS
connection to the PC is activated for
2 minutes
point and limit value
Note
1) With the filter function active, short-term measurement variations should be
suppressed, resulting in a steadier display. Measured values via the interface and in
the Data Logging Memory will not be filtered.