Model DW-USB-LITE Mini Temperature Data Logger
Specifications - Installation and Operating Instructions
The Model DW-LITE Data Logger is a low cost temperature data logger
that measures and stores up to 4080 temperature readings over a mea­surement range of 14 to 122°F. The stand-alone device has a fixed 30-min sample rate and a battery life of 1 month with continuous use. High/low alarm settings are user programmable through easy to install windows based software. The DW-LITE features a visual LED alarm indicator. Stored data can be viewed by directly plugging the data logger into a PC’s USB port without any cables. The provided software allows downloaded data to be graphed, printed and exported to other applications.
LED FLASHING MODES
Features a red/green LED.
(Single Flash) Logging, no alarm.
(Single Flash) Logging, low alarm.
(Double Flash) Logging, high alarm.
BATTERY REPLACEMENT Note: Product is shipped with battery installed. Remove white tab from
device to begin logging.
SPECIFICATIONS Temperature Range: 14 to 122°F (-10 to 50°C). Temperature Accuracy: ±4.5°F (±2.5°C). Temperature Resolution: 2°F (1°C). Memory Size: 4080 readings. Sampling Mode: Continuous. Sampling Rate: 30 min (fixed). Computer Requirements: Windows®98, Windows®2000, Windows®XP,
and Windows Vista™ (software included).
Power Requirements: 3V coin cell battery (included). Alarms: Programmable high/low. Interface: USB. Weight: 1 oz (30.0 g).
SOFTWARE INSTALLATION Note: This software must be installed prior to the USB Driver. Please fol-
low the instructions below then refer to the USB Driver Installation based on the Windows platform you are using.
1. Visit our website www.dwyer-inst.com/DW-USBSoftware and click on the download link. If the installation folder doesn’t open directly, go to your computer’s “Downloads” folder and double-click on the “EasyLogUSBV610_Dwyer” folder.
2. Double-click on “setup” then follow the on screen instructions to complete the Configuration Software Installation. An EasyLog USB icon will now be placed on your desktop.
Windows® and Windows Vista®are registered trademarks of Microsoft Corporation.
DWYER INSTRUMENTS, INC.
Phone: 219/879-8000 www.dwyer-inst.com
P.O. BOX 373 • MICHIGAN CITY, INDIANA 46360, U.S.A. Fax: 219/872-9057 e-mail: info@dwyer-inst.com
USB DRIVER INSTALLATION
Windows®XP/Windows Vista™: Note: You must install the Configuration Software as instructed in the pre­vious section before going through these steps.
1. Insert the data logger directly into a USB port on the computer. Installation screens should automatically appear.
2. Follow the on screen instructions to complete the Configuration Software Installation. If a compatibility error window appears, click “Continue Anyway”. Note: This will not cause any adverse changes/effects on your computer.
3. You can now begin using your DW-USB-LITE data logger by double clicking on the EasyLog USB icon on your computers desktop.
Windows®98/Windows®2000: Note: You must install the Configuration Software as instructed in the pre­vious section before going through these steps.
1. Insert the data logger directly into a USB port on the computer. Installation screens should automatically appear.
2. Follow the on screen instructions to complete the USB Driver Installation.
3. You can now begin using your DW-USB-LITE data logger by double clicking on the EasyLog USB icon on your computer’s desktop.
MAINTENANCE
Upon final installation of the DW-USB-LITE Data Logger, no routine main­tenance is required. Leaving the logger plugged into your computer for an extended period of time will considerably lower its battery life. A periodic check of system calibration is recommended. The Model DW-USB-LITE is not field serviceable and should be returned if repair is needed (field repair should not be attempted and may void warranty). Be sure to include a brief description of the problem plus any relevant application notes. Contact customer service to receive a return goods authorization number before shipping.