Dwyer DPMW User Manual

Bulletin E-90-DPMW
P Models
°F °C
P
SI
%
V
A KW PF
29/32 [23.1]
2-5/16 [58.7]
1
3/32
[10.2]
1-9/16 [39.7]
63/64 [
25.1]
1/8 [3]
1
/8
[3]
2
-3/8
[60.5]
17/64
[6.9]
[Ø.140]
[Ø3.6 mm]
2 PLCS
PANEL CUTOUT FOR
WINDOW MOUNTING
Series DPMW LCD Digital Panel Meters
Specifications - Installation and Operating Instructions
The Series DPMW LCD Digital Panel Meter is designed with a 3-1/2
digit, high contrast LCD display. The colored segments are available in red, amber, or green - ideal for viewing at a distance. The Series DPMW features user selectable engineering units, selectable decimal point position and adjustable zero and span. The meter accepts a 4 to 20 mA input signal for pressure, level, flow, and temperature transmitter. A 24 VDC power supply is required to illuminate the colored segments. The series DPMW can be quickly installed in a window cutout.
INSTALLATION
The Series DPMW is designed for window mounting in a 2.31˝ (58.4 mm) W x .91˝ (22.9 mm) H panel cutout.
WIRING
The unit is powered by a 4-20 mA loop and the screw terminals for wiring are located on the back of the adder board marked with + SIG -. The backlighting requires a 24 VDC power supply and should be connected to terminals identified with + B/L -.
SPECIFICATIONS Inputs: 4 to 20 mA. Input Impedance: 300Ω nominal. Accuracy: ±(0.1% FS + 2 count). Power Supply: Powered by control loop. Backlight Power Supply: 24 VDC @ 35 mA typical. Span and Zero: Adjustable. (± 1999). Display: 3-1/2 digits, 7 segments, 0.45˝ (11.4 mm) H. Decimal Points: 3-position, user selectable. Engineering Units: DPMW-XXX: °F, °C, %, PSI; DPMW-XXXP: V, A,
KW, PF.
Polarity: Automatic, “-” displayed. Operating Temperature: 32 to 122°F (0 to 50°C). Storage Temperature: -4 to 158°F (-20 to 70°C) Mounting: Window mount. Connection: Screw terminals. Conversion Rate: 3 per second. Warm-Up: 10 minutes typical. Weight: 2 oz (56.7 g). Agency Approvals: RoHS.
REAR VIEW
Note: If backlight supply is not loop supply, ground should be referenced
together.
DWYER INSTRUMENTS, INC.
P.O. BOX 373 • MICHIGAN CITY, INDIANA 46360, U.S.A. Fax: 219/872-9057 e-mail: info@dwyermail.com
Phone: 219/879-8000 www.dwyer-inst.com
OPERATION
A
J3 J2
J1
B
1
*
P Models
V A
KW
PF
1
1
3
°F °C
PSI
%
A
J3 J2
J1
B
Selecting Engineering Units
Four sets of jumper pins are located in the back of the meter, between the meter and the adder board. Move the jumper to fit over the appropriate pins which correspond to the desired engineering units. See Figure 1.
FIGURE 1
Selecting Decimal Point Position
Four decimal point positions are available on the digital panel meter. Move the jumper to correspond to the desired decimal point location. See Figure
2.
Setting Min/Max Display Value J1, J2, J3 (See Figure 3 & 4):
If:
The minimum display is ≤ 0 or Minimum display is > 0 and the (maximum display minimum display) > 5; The J1, J2 & J3 should be all set to the top jumper (see Figure 3).
FIGURE 3
If: The minimum display is > 0 and the (maximum display minimum display) < 5; The J1, J2 & J3 should be all set to the bottom jumper (see Figure 4).
FIGURE 2
FIGURE 4
MAINTENANCE/REPAIR
Upon final installation of the Series DPMW, no routine maintenance is required. The Series DPMW is not field serviceable and should be returned if repair is needed. Field repair should not be attempted and may void warranty.
WARRANTY/RETURN
Refer to “Terms and Conditions of Sale” in our catalog and on our website. Contact customer service to receive a Return Goods Authorization number before shipping the product back for repair. Be sure to include a brief description of the problem plus any additional application notes.
Phone: 219/879-8000 www.dwyer-inst.com
©Copyright 2014 Dwyer Instruments, Inc. Printed in U.S.A. 5/14 FR# R2-443464-00 Rev. 3
DWYER INSTRUMENTS, INC.
P.O. BOX 373 • MICHIGAN CITY, INDIANA 46360, U.S.A. Fax: 219/872-9057 e-mail: info@dwyermail.com
Loading...