Dwyer 485B User Manual

Model 485B Digital Hygrometer
Specifications and Operating Instructions
SPECIFICATIONS Service: Humidity, temperature, wet bulb and dew point temperature
detection in clean air.
Temperature Limits:
Process: -40 to 176°F (-40 to 80°C); Ambient: 5 to 125°F (-15 to 51°C).
Range Relative Humidity: 0 to 100% (non-condensing). Accuracy Relative Humidity: ±2% FS over 10 to 90% @ 77°F (25°C). Range Temperature: -20 to 140°F (-28 to 60°C). Accuracy Temperature: ±0.9°F @ 72°F (±0.3°C @ 25°C). Display: 4.5 digit LCD. Resolution: 0.1%, 0.1 °F/°C. Probe Length: 8˝ (203 mm) insertion. Power Requirements: 9V alkaline battery included. Weight: 16 oz (454 g). Agency Approvals: CE, RoHS.
ulletin TE-485B
B
Model 485B Digital Thermo-Hygrometer is a versatile, compact, hand-
held instrument for measuring percentage of relative humidity and temperature in °F or °C. Dew point and wet bulb temperatures are derived from relative humidity and temperature measurements and displayed on the LCD. The hold key freezes the current temperature and relative humdity readings for situations where readings fluctuate. Store up to 99 readings with the non-volatile memory function – ideal for technicians needing to take multiple readings for later analysis.
Battery Installation
The unit is shipped with a separate 9 volt alkaline battery which must be installed before operation. Remove the two screws holding the bottom endcap in place and remove it. Connect the battery to the enclosed battery clip, observing correct polarity. Be careful not to trap the wires between the case or foam pad which retains the battery. This could make it difficult to install the battery or remove it later for replacement. Be sure the rubber gasket is properly seated in the gasket channel and replace the endcap. Note that the endcap will only fit one way because the holes are slightly off-center. Place the “Z” shaped wrist strap clip in one of the screw recesses and replace the screws. Do not overtighten. Attach the wrist strap to the clip.
When battery replacement becomes necessary, use only 9 volt alkaline type batteries such as a Duracell®MN1604, or Eveready®522 or equivalent. Zinc-carbon types, often labeled Heavy-Duty are not recommended because of their shorter life and an increased potential for leakage. Alkaline batteries are a better value because they last typically up to three times longer in this device.
Low Battery Indicator
A weak battery may cause improper operation or inaccurate measurements. A low battery indicator ( ) is provided on the display to show when the battery needs to be replaced. Although the unit may appear to function and read properly, the accuracy of the readings cannot be guaranteed when the ( ) indicator is empty of power segments. Replace the battery with a fresh one. Do not leave an exhausted battery in the unit due to potential battery leakage.
OPERATING INSTRUCTIONS
Duracell® is a registered trademark of The Gillette Company
®
DWYER INSTRUMENTS, INC.
Eveready
Phone: 219/879-8000 www.dwyer-inst.com
is a registered trademark of Eveready Battery Company, Inc.
P.O. BOX 373 • MICHIGAN CITY, INDIANA 46360, U.S.A. Fax: 219/872-9057 e-mail: info@dwyer-inst.com
HOLD
Memory
HOLD will appear on the home screen when the HOLD button is pressed, indicating the process reading has been frozen.
Memory Storage and Retrieval
From the home screen, push MEMORY/UP to view all previously stored values. Cycle up or down through the stored data with the MEMORY/UP or STORE/DOWN buttons. From the home screen, press STORE/DOWN to store data points to memory. For both data storage and data memory retrieval, press MENU to return to the home screen.
MENUS Maximum
Push ENTER to see the maximum recorded reading. When viewing the maximum recorded reading, hold ENTER for 4 seconds to clear.
Minimum
Press ENTER to see . Hold ENTER for 4 seconds to clear all stored memory locations.
Engineering Units
Press ENTER to see current engineering units. Press MEMORY/UP or STORE/DOWN to select another unit.
Power Management
Pu sh ENTER to see current time off setting. Select 30 or 10 values indicating minutes for power to automatically shut-off. By selecting 0, the auto shut-off is disengaged and the unit will only turn off manually by the ON/OFF/LIGHT button.
MAINTENANCE/REPAIR
Upon final installation of the Model 485B, no routine maintenance is required. The Model 485B is not field serviceable and should be returned if repair is needed. Field repair should not be attempted and may void warranty.
WARRANTY/RETURN
Refer to “Terms and Conditions of Sales” in our catalog and on our website. Contact customer service to receive a Return Goods Authorization number before shipping the product back for repair. Be sure to include a brief description of the problem plus any additional application notes.
Push ENTER to see the minimum recorded reading. When viewing the minimum recorded reading, hold ENTER for 4 seconds to clear.