Du Pont SORVALL RT6000B, SORVALL T6000B Operating Instructions Manual

OPERATING INSTRUCTIONS
SORVALL RT6000B
& T6000B
TABLETOP CENTRIFUGES
Du Pont Company SORVALL® Products Wilmington, Delaware 19898 U.S.A.
SORVALL Centrifuges
PN 07903-3 Issued July 1993
RT6000B & T6000B
This manual is a guide to the use of the
SORVALL® Centrifuges
SORVALL® RT6000B and T6000B
Tabletop Centrifuges
Data herein has been verified and validated and is believed adequate for the intended use of the centrifuge. Because failure to follow the recommendations set forth in this manual could produce personal injury or property damage, always follow the recommendations set forth herein. Du Pont does not guarantee results and assumes no obligation for the performance of rotors or other products that are not used in accordance with the instructions provided. This publication is not a license to operate under , nor a recommen­dation to infringe upon, any process patents.
Publications prior to the Issue Date of this manual may contain data in apparent conflict with that provided herein. Please consider all data in this manual to be the most current.
NOTES, CAUTIONS, and W ARNINGS within the text of this manual are used to emphasize important and critical instructions.
WARNING: A W arning informs the operator of a hazard or an unsafe practice that could
result in personal injury, affect the operator’s health, or contaminate the environment.
CAUTION: A Caution informs the operator of an unsafe practice that could result in
damage of equipment.
NOTE: A Note highlights essential information.
© 1988 by E. I. du Pont de Nemours & Co.
ii
SORVALL® Centrifuges
Important Safety Information
Certain potentially dangerous conditions are inherent to the use of all centrifuges. To ensure safe operation of this centrifuge, anyone using it should be aware of all safe practices and take all precautions described below and throughout this manual.
WARNING
When using radioactive, toxic, or pathogenic materials, be aware of all characteristics of the materials and the hazards associated with them in the event leakage occurs during centrifugation. If leakage does occur, neither the centrifuge nor the rotor can protect you from particles dispersed in the air. To protect yourself, we recommend additional precautions be taken to prevent exposure to these materials, e.g., use of controlled ventilation or isolation areas.
Always be aware of the possibility of contamination when using radioactive, toxic, or pathogenic materials. Take all necessary precautions and use appropriate decontamination procedures if exposure occurs.
The use of sealed rotors, buckets and/or sample containers will provide increased protection from contamination during routine operation. However, these items will not guarantee contamination protection from accidents resulting in damage to the rotor or buckets. Do not run hazardous materials in the centrifuge unless placed in a biohazard enclosure and operated using all appropriate safety precautions.
Safety
Never use any material capable of producing flammable or explosive vapors.
Never exceed the maximum rated speed of the installed rotor; to do so can cause rotor failure.
Always reduce (derate) rotor speed as instructed in this manual whenever:
the rotor speed/temperature combination exceeds the solubility of the gradient material and causes it to precipitate.
the compartment load exceeds the maximum allowable compartment load specified. See Chapter 3, Operation, page 3-6.
Failure to reduce rotor speed under these conditions can cause rotor failure.
CAUTION
Do not run or precool a rotor at the critical speed, as this will have a detrimental effect on centrifuge component life. See Chapter 3, page 3-3.
Do not operate the centrifuge with a rotor out of balance. To do so can cause damage to the centrifuge drive assembly.
Do not operate centrifuge without a rotor properly installed: rotor cover must be on and locked in place, and the rotor locked to the centrifuge drive. See rotor instruction manual.
The centrifuge can be damaged if connected to the wrong voltage. Check the voltage before plugging the centrifuge into a power source. Du Pont is not responsible for incorrect installation.
Always maintain the centrifuge in the recommended manner. See Chapter 4, Maintenance.
iii
Table of Contents
SORVALL® Centrifuges
Table of Contents
Safety Information Page . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . iii
Chapter 1. DESCRIPTION
Centrifuge Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
Centrifuge Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
Rotors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
Chapter 2. INSTALLATION
Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
Electrical Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
Installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Page
Chapter 3. OPERATION
Controls and Indicators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Rotor Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
Precooling a Rotor in the RT6000B Centrifuge . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
Running Hazardous Material . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
Emergency Sample Recovery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
Reducing Speed for Rotor Compartment Loads in Excess
of Design Mass . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
Chapter 4. MAINTENANCE
Inspection and Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
REV/MIN x 1000 Indicator Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
Motor Brush Inspection and Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
RT6000B CHAMBER TEMP° C Indicator Adjustment . . . . . . . . . . . . . . . . . . . . . 4-4
Parts Ordering Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
Service Decontamination Policy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
Warranty Index Decontamination Certificates
iv
APPENDIX
SORVALL® Centrifuges
Table of Contents
List of Illustrations
Figure Page
3-1. Controls and Indicators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
3-2. Location of Mechanical Override Loop . . . . . . . . . . . . . . . . . . . . . 3-6
4-1. REV/MIN x 1000 Indicator Adjustment . . . . . . . . . . . . . . . . . . . . . 4-3
4-2. Motor Brush Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
4-3. CHAMBER TEMP° C Indicator Adjustment . . . . . . . . . . . . . . . . . . 4-4
List of Tables
Table Page
3-1. Controls and Indicators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
v/vi
SORVALL® Centrifuges
A PHOTOGRAPH OF THIS INSTRUMENT WAS NOT TAKEN AND THEREFORE
WILL NOT BE AVAILABLE FOR FIRST PRINTING OF MANUAL
Figure 1-1. SORVALL® RT6000B Refrigerated
Tabletop Centrifuge (T6000B not shown)
vi
SORVALL® Centrifuges
Description
Chapter 1
DESCRIPTION
This manual provides you with the information you need to install, operate and maintain your SORVALL® RT6000B and T6000B Tabletop Centrifuges. If you encounter any problem concerning either operation or maintenance that is not covered in this manual, please contact our Marketing Technical Group for assistance. In the United States, telephone toll free (800) 551-2121. Outside the United States, contact your local distributor or agent for SORVALL® Products.
Centrifuge Description
The R T6000B combines the features of the tabletop centrifuge and those of the larger refrigerated centrifuge in a compact tabletop instrument. It is similar to other SORV ALL Tabletop Centrifuges, with an added refrigeration system that consists of an evaporator/ rotor chamber and a low-temperature condensing unit. The system is designed to maintain temperature within 1°C, while in the operating range of –5°C to +25°C. The T6000B is a non-refrigerated tabletop centrifuge.
Both the RT6000B and T6000B feature a closed-loop speed control and have a see­through chamber door that allows rotor calibration and visual inspection of a run in progress. The chamber door is counterbalanced for easy opening and safe closing. The lid latch is locked manually by turning the door release knob counterclockwise to the LOCKED position, then mechanically an interlock engages when the rotor begins spinning to prevent the chamber door from being opened during operation. The gyro-action, self­centering drive has a diamond hex-shaped drive spindle which assures proper seating of rotors. The centrifuges also have a slow-start feature that slowly accelerates the rotor from 0 to 600rpm.*
Centrifuge Specifications
Maximum Speed** . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6000 rpm
®
Operating Temperature
Range (RT6000B only) . . . . . . . . . . . . . . . . . . . . . . . –5°C to +25°C
Maximum heat output during
operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6000 Btu per hour (RT6000B)
*Speed in revolutions per minute (rpm) is related to angular velocity, ω, according to the following:
ω
= (rpm) = (rpm) (0.10472)
Where ω = rad/s. All further references in this manual to speed will be designated as rpm. ** Maximum speed is dependent on rotor and line voltage.
(
2π 60
3000 Btu per hour (T6000B)
(continued)
)
1-1
Description
SORVALL® Centrifuges
Centrifuge Specifications (continued)
Electrical Requirements:
RT6000B . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 115 Vac, 60 Hz, 15A, single phase*
220 Vac, 50 Hz, 8A, single phase*
T6000B. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 115 Vac, 60 Hz, 8A, single phase
220 Vac, 50 Hz, 4A, single phase
Dimensions:
Width (RT6000B) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81 cm (32 inches)
(T6000B) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66 cm (26 inches)
Depth . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64 cm (25 inches)
Height . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32 cm (12.5 inches)
Mass (Weight):
RT6000B . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85 kg (187 lbs)
T6000B. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78 kg (160 lbs)
*CSA and UL Certified.
Rotors
The table below lists the rotors available for the RT6000B and T6000B Centrifuges. For more information about other rotor accessories and tubes refer to the most current SORVALL® Rotors, Tubes, Bottles, and Adapters Catalog.
Rotor Specifications
Maximum Maximum Maximum Relative Operating Compartment Centrifugal
Rotor Speed (rpm) Mass (grams) Force (RCF)
A-384 Fixed-Angle Rotor 6000 27
A-500 Fixed-Angle 6000 115 Rotor
H-1000B Swinging 3200 820 Bucket Rotor
1
1
2
4890
4960
2190
A/S-400 Fixed-Angle 4380 90 Rotor
1 Includes weight of tube holder, tube or bottle, adapter, and tube contents. 2 Includes weight of adapter, tube or bottle, and tube contents.
1-2
2
3000
SORVALL® Centrifuges
Description
RT-6000B
Centrifuge Specifications
Maximum Speed** . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6000 rpm
Operating Temperature
Range (RT6000B only) . . . . . . . . . . . . . . . . . . . . . . . –5°C to +25°C
Maximum heat output during operation
Electrical Requirements:
RT6000B . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 115 Vac, 60 Hz, 15A, single phase*
T6000B. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 115 Vac, 60 Hz, 8A, single phase
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6000 Btu per hour (RT6000B)
3000 Btu per hour (T6000B)
220 Vac, 50 Hz, 8A, single phase*
220 Vac, 50 Hz, 4A, single phase
Dimensions:
Width (RT6000B) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81 cm (32 inches)
(T6000B) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66 cm (26 inches)
Depth . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64 cm (25 inches)
Height . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32 cm (12.5 inches)
Mass (Weight):
RT6000B . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85 kg (187 lbs)
T6000B. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78 kg (160 lbs)
Rotor Specifications
Maximum Maximum Relative Operating Compartment Centrifugal
Rotor Speed (rpm) Mass (grams) Force (RCF)
A-384 Fixed-Angle Rotor 6000 27
A-500 Fixed-Angle 6000 115 Rotor
1
1
Maximum
4890
4960
H-1000B Swinging 3200 820 Bucket Rotor
A/S-400 Fixed-Angle 4380 90 Rotor
2
2
2190
3000
1-3
SORVALL® Centrifuges
Installation
Chapter 2
INSTALLATION
This chapter contains instructions to prepare your SORVALL® RT6000B and T6000B for operation.
Inspection
When you receive your centrifuge, carefully inspect it for any signs of shipping damage. If you find damage, report it immediately to the transportation company and file a damage claim, then notify Du Pont.
Check the parts received with the centrifuge against the shipping list; if any parts are missing, contact Du Pont (see back cover).
WARNING
The RT6000B weighs 187 lbs. and the T6000B weighs 160 lbs. Refer to the unpacking instructions for proper care when lifting and installing the centrifuge. Failure to use proper lifting techniques can result in possible damage to the centrifuge and/or personal injury.
CAUTION
Do not lift the centrifuge by the front panel or by the lid. To do so can result in damage to these parts.
Do not place the RT6000B on its side; any position other than upright can cause compressor contamination.
Electrical Requirements
The centrifuge has specific power requirements and must be connected to the correct supply for proper performance. The nameplate on the back of the cabinet specifies one of the following:
RT6000B:
T6000B:
The current draw during acceleration in the T6000B could peak to 15 amps for 60 Hz centrifuges and 8 amps for 50 Hz centrifuges. Therefore, if the main power source is
*CSA and UL Certified
105-125 Vac, 60 Hz, single phase, 15 amps*
210-230 Vac, 50 Hz, single phase, 8 amps*
105-125 Vac, 60 Hz, single phase, 8 amps
210-230 Vac, 50 Hz, single phase, 4 amps
2-1
Installation
protected by fuses rather than circuit breakers, we recommend the use of slow-blow or time-delay fuses.
SORVALL® Centrifuges
CAUTION
If the power cord is connected to the wrong voltage, it can cause damage to the centrifuge. Check the voltage listed on the nameplate before plugging the power cord into the power source. Du Pont is not responsible for incorrect installation.
The power cord has a universal keyed plug that inserts into the receptacle at the back of the centrifuge. The other end of the power cord has a NEMA5-15P , three-prong molded cap with a ground pin and parallel blades. The plug will fit Hubbell receptacle No. 5261 for 115 volt (parallel blades) or receptacle No. 5561 for 220 volt (tandem blades). The power cord must be attached when the centrifuge is installed.
Installation
To install the centrifuge:
1. Place the centrifuge on a large, sturdy bench or work table that will support its weight
leaving space for sample preparation. Be sure to leave a minimum clearance of
5cm(2 inches) on all sides for pr oper air circulation.
2. Insert the universal keyed end of the power cord into the receptacle at the back of the
centrifuge.
3. Make sure the centrifuge is level. If necessary, place the leveling pads provided under
the corners of the centrifuge.
4. Make sure the POWER switch is set to "0" (off position), then plug the power cord into
a wall receptacle. The centrifuge is now ready for use.
2-2
SORVALL® Centrifuges
Operation
Chapter 3
OPERATION
This chapter describes the function of each operating control and indicator and provides the information necessary to operate your SORVALL
®
RT6000B and T6000B Centrifuges.
Controls and Indicators
Figure 3-1 shows the location of operating controls and indicators. Table 3-1 lists them by item number and describes their function.
Figure 3-1. Controls and Indicators
Table 3-1. Controls and Indicators
(keyed to figure 3-1)
Item Name Function
1 REV/MIN x 1000 Sets desired run speed. At full counterclockwise
dial
position the speed control is turned off and the rotor will not spin. (The settings on the dial indicate actual rotor speed.)
CAUTION
Continued operation of the centrifuge be­low 600 rpm may damage the speed control.
(continued)
3-1
Operation
Table 3-1. Controls and Indicators (continued)
(keyed to figure 3-1)
Item Name Function
2 REV/MIN x 1000 Indicates actual rotor speed from 0 to 6000 rpm
indicator
3 OPEN light* Lights at the end of the run to indicate that the door
4 FAULT light* Lights when a rotor imbalance occurs or when a
(T6000B)
4 FAULT light* Lights when a rotor imbalance occurs, when cham-
(RT6000B)
±100 rpm.
may be opened.
drive belt malfunction occurs. Centrifuge will not restart until problem has been corrected.
ber over or under temperature occurs, or when a drive belt malfunction occurs. The centrifuge will not restart until problem has been corrected.
SORVALL® Centrifuges
5 CHAMBER TEMP° C Indicates actual chamber temperature from -20°C to
indicator (RT6000B only)
6 CHAMBER TEMP° C Sets desired chamber temperature from -20°C to
dial (RT6000B only)
7 TIME dial Sets length of run time up to 30 minutes (or 35
8 Door release knob When the knob is turned to OPEN it will release the
9 BRAKE switch When set to "
10 POWER switch The power switch is an on/off toggle switch that,
+40°C.
+40°C.
minutes on 50 Hz instruments). At run time. The TIME dial acts as a start switch to begin centrifuge run.
door latches; when knob is turned to LOCKED it engages the door latches.
", the centrifuge brakes to approxi­mately 200 rpm at end of run then coasts to a stop. When set to " ", the centrifuge coasts to a stop without braking.
when set to "I" applies power to the centrifuge.
, sets indefinite
*These messages are backlit, meaning that each message lights only when the condition that it represents exists.
Rotor Considerations
a. Rotor Temperature
Any type of centrifugation creates an increase in rotor temperature. The increase in temperature is caused by many variables including ambient temperature, rotor speed, length of the run, and the type of rotor.
3-2
SORVALL® Centrifuges
Air flow through the T6000B centrifuge is designed to minimize increase in rotor temperature. Air enters through an opening in the center of the lid and exits through the exhaust duct at the top of the rotor chamber.
WARNING
Blocking the air flow entering and/or exiting the T6000B rotor chamber will cause an increase in the temperature within the rotor chamber. The temperature increase can temporarily distort the non-metal surfaces allowing particles to exit the rotor chamber resulting in personal injury and/ or centrifuge damage should tube breakage or rotor failure occur.
When running temperature-sensitive material we recommend that you do a trial run.
b. Rotor Installation, Loading and Balancing
Refer to the rotor instruction manual for loading and balancing procedures as well as information regarding the selection and use of tubes, bottles, and adapters.
Operation
WARNING
When loading the rotor, be sure not to exceed the maximum compartment mass of the rotor (see Table 1-1 or the individual rotor instruction manual). If maximum compartment mass is exceeded, maximum rotor speed must be lowered as described on page 3-6, Reducing Speed for Rotor Compartment Loads in Excess of Design Mass. Failure to do so can cause rotor failure which could result in personal injury and/or centrifuge damage.
CAUTION
Do not operate the centrifuge without a rotor installed and properly balanced. Also, when installing a rotor carefully place it on the centrifuge drive spindle. The drive can be damaged if rotors are dropped onto it.
Precooling a Rotor in the RT6000B Centrifuge
For operation at temperatures other than ambient, the rotor and rotor chamber should be precooled before the run using either of the precooling methods described.
a. Precooling a Rotor in the Centrifuge
1. Follow steps 1 through 6 of the operating instructions (pages 3-4 and 3-5).
2. Set the REV/MIN x 1000 dial to 1500 rpm.
The length of time it will take to precool a rotor will vary depending on the temperature selected, rotor weight, and rotor material.
3-3
Operation
SORVALL® Centrifuges
b. Precooling the Centrifuge Chamber Only
If the rotor has been precooled outside of the chamber (e.g., in a refrigerator or cold room), the chamber should be precooled before the run.
1. With the chamber door closed, set the CHAMBER TEMP° C dial to the temperature desired.
2. When the chamber is cooled to the proper temperature, install the precooled rotor and begin the centrifuge run.
NOTE
To expedite cooling of the centrifuge chamber, run an empty rotor at 1500 rpm.
Running Hazardous Material
Because neither centrifuge chamber of the RT6000B or T6000B are designed for biocontainment, some vapors or aerosols released from uncapped, leaking or broken tubes may leak from the chamber during operation. Once a run is completed and the chamber door is opened, the vapors or aerosols which have concentrated in the chamber will be released into the laboratory area. For this reason, when materials which are pathogenic, toxic, or otherwise hazardous in nature are to be run, the centrifuge should be placed in a biohazard safety enclosure and operated using all appropriate safety precautions. Observe the WARNING found on the Safety Information Page.
Use appropriate decontamination procedures should exposure to any hazardous material occur . See Chapter 4 for the procedure to follow if a centrifuge or rotor that has been used with a hazardous material must be returned to our service facilities for repair.
Operation
To perform the run:
1. Set the POWER switch to "I", and turn the TIME dial off. (The fan will tur n on and continue to operate whenever the POWER switch is in the on position).
NOTES
RT6000B Centrifuges ONLY:
The FAULT light may come on when the power is turned on with the door closed. The light will go off when the door is opened.
It is recommended that the chamber door be kept closed when the centrifuge is not in use to prevent condensation from forming in the rotor chamber.
3-4
SORVALL® Centrifuges
2. When the OPEN light comes on, turn the door release knob to the OPEN position, and lift the chamber door up.
3. Install the rotor on the drive shaft and close the door . Turn the door release knob to the LOCKED position (the door latches will engage).
CAUTION
Be sure the rotor is properly balanced and seated on the drive spindle. See rotor instruction manual.
4. RT6000B only: Set the CHAMBER TEMP° C dial to the run temperature desired.
5. Set the BRAKE switch to " " if braking is desired.
6. Set the TIME dial to the run time desired.
7. Set the REV/MIN x 1000 dial to the desired run speed.
Operation
NOTE
The settings on the REV/MIN x 1000 dial indicate actual rotor speed. Verify the rotor speed by checking the REV/MIN x 1000 indicator.
A zero speed switch is incorporated in the REV/MIN x 1000 dial. This switch prevents rotation of the rotor if the door is locked and the timer is on.
At the end of the run remove the rotor as described below.
WARNING
Always check that the rotor has stopped spinning before opening the chamber door. Do not open the door while the rotor is still in motion; to do so can result in possible injury.
To remove the rotor:
1. Leave the POWER switch set to "I".
2. If the TIME dial is set to
3. When the OPEN light comes on, turn the door release knob to OPEN position and lift the chamber door up.
4. Remove the rotor, close the chamber door, and set the POWER switch to "0".
, turn it off.
3-5
Operation
SORVALL® Centrifuges
Emergency Sample Recovery
If the main power shuts off because of a power failure or system malfunction while the rotor is spinning, the chamber door will not open. A mechanical override is provided to allow sample recovery in the case of an emergency.
WARNING
This procedure is included for emergency sample recovery only and should never be used for any purpose other than those explained in this section.
When the main power shuts off, the brake will not operate. Wait until the rotor stops spinning before using the mechanical override.
The mechanical override loop is located under the right, front corner of the centrifuge as shown in figure 3-2. To operate the override, insert a screwdriver or similar object into the metal loop and pull down. Turn the door release knob to the OPEN position and lift the chamber door up.
Figure 3-2. Location of Mechanical Override Loop
Leave a clearance area around the lower right corner of the centrifuge to prevent obstruction of the mechanical override.
Reducing Speed for Rotor Compartment Loads in Excess of Design Mass
There is a maximum allowable compartment mass established for each centrifuge rotor.
3-6
NOTE
SORVALL® Centrifuges
To prevent rotor failure, the total contents of any compartment, including specimen, tubes, cover, and adapters (if used), must not exceed the figur e given on page 1-2 unless rotor speed is reduced proportionately.
Strict adherence to the maximum allowable compartment mass or reduced speed is required to prevent rotor failure. Observe WARNING on the Safety Information Page
in front of this manual.
The rotor speed is reduced in proportion to the square of the ratio for the maximum allowable compartment mass to the actual compartment mass (including specimen, tubes, covers, and adapters). If the compartment mass is more than that specified for the rotor, the reduced speed can be determined by using the formula given below.
Maximum Compartment Mass
Reduced Rotor Speed = Maximum Rotor Speed x
Actual Compartment Mass
Operation
3-7
3-7/3-8
SORVALL® Centrifuges
Maintenance
Chapter 4
MAINTENANCE
This chapter describes routine maintenance procedures that you should perfor m on a regular basis. As the user, it is your responsibility to make certain that these procedures are followed when necessary. Also, to keep your centrifuge in good working condition and ensure accurate test results, we recommend that, in addition to these routine procedures, you have the speed control, timer, rotor imbalance detector and temperature control checked periodically by a Du Pont Service Representative or other qualified service personnel because occasionally, these controls may need to be recalibrated. If further service is needed, contact your local representative for SORVALL
WARNING
Because of the high voltages in this centrifuge, only personnel trained in electronics should test or repair the electrical circuits.
®
Products.
If hazardous materials have been processed in the centrifuge, take all necessary precautions when cleaning or servicing the centrifuge to avoid personal exposure to contamination.
Inspection and Cleaning
a. Inspection
Inspect the centrifuge each week for signs of wear, encrusted biological deposits, and general cleanliness. Follow the cleaning procedures described below when necessary.
b. Cleaning
Rotor Chamber
The rotor chamber should be kept clean and wiped dry. Wash the rotor chamber with a mild, nonalkaline dishwashing liquid, then rinse and dry with a soft absorbent cloth.
Use 70% ethanol to disinfect the rotor chamber or a 2% glutaraldehyde solution to sterilize it. For general radioactive decontamination, use a solution of equal parts of 70% ethanol, 10% SDS, and water. Follow this with ethanol rinses then deionized water rinses. Dry with a soft absorbent cloth. Dispose of all wash solutions in proper radioactive waste containers. Read the CAUTION on the next page.
4-1
Maintenance
Chlorides are extremely harmful to aluminum alloy rotors and can cause stress corrosion cracking. Therefore, if chlorides are used to disinfect the chamber, be sure to rinse the chamber thoroughly with water to remove all of the chloride cleanser.
Periodically defrost the rotor chamber of the RT6000B to maintain efficient cooling. The chamber can be defrosted by either leaving the door open with the power off until the frost melts, or by running a rotor for approximately five minutes at 40°C. When the chamber has been completely defrosted, wipe the chamber dry.
Drive Spindle
Before each run, wipe the spindle with a soft cloth before a rotor is installed to reduce the chance of the rotor sticking to the spindle. To keep the surface of the spindle smooth, polish occasionally with #400 Emery cloth. Always wipe the spindle clean after polishing.
SORVALL® Centrifuges
CAUTION
Cabinet
Clean the enameled cabinet panels with a household wax cleaner. Use a mild, nonalkaline detergent and water to clean the centrifuge.
Chamber Door
Clean the chamber door with a soft, wet cloth and soapy water . Avoid using abrasive materials (for example, dry papertowels) and organic solvents (for example, acetone) on the dry surface.
Condenser Fins
To maintain the efficiency of the RT6000B refrigeration system, keep the condenser fins (located in the front of the centrifuge) free of dirt and dust. Clean the fins with a brush or vacuum cleaner at one to three month intervals.
c. FAULT System Check
To check the function of the FAULT system, move the drive spindle back and forth to simulate a rotor imbalance. This should cause the FAULT indicator to light.
REV/MIN x 1000 Indicator Adjustment (figure 4-1)
With the POWER switch set to "0", use a screwdriver and adjust the zero adjustment screw until the pointer reads 0.
4-2
SORVALL® Centrifuges
Figure 4-1. REV/MIN x 1000 Indicator Adjustment
Motor Brush Inspection and Replacement
1. Unplug the centrifuge power cord.
2. Remove the four screws from the access cover and remove the cover.
Maintenance
3. Remove the plastic brush caps from the motor with a screwdriver, and pull the brushes from their holders, noting orientations.
NOTE
Before inserting the replacement brushes, remove any carbon accumula­tion from the interior of the brush holder. The brush should then move freely in the brush holder.
a. If the brushes are longer than 6 mm (1/4 inch), new brushes are not needed.
Reinstall brushes in their original brush tubes with the same orientation.
b. If the brushes are shorter than 6 mm (1/4 inch), replace worn brushes with new
brushes, Catalog No. 12732.
NOTE
If this is not the original set of brushes and the brushes are shorter than 6 mm (1/4 inch), periodically (that is, after 3 to 4 brush set changes) have the armature commutator inspected for wear and roughening (pitting) by a qualified technician when the brushes are replaced. If necessary, replace the armature.
4. Reinstall the plastic brush caps, making sure that the brush spring retainer “ears” are seated in the brush holder slots (see figure 4-2).
5. Reinstall the access cover and secure in place using the four screws removed in step2.
6. With the POWER switch set to "0", plug in the centrifuge power cord.
7. If a new armature and/or new brushes have been installed, run the centrifuge at 1000rpm with a r otor for 30 minutes to seat the brushes. Failure to do this will result
4-3
Maintenance
in immediate brush burn and commutator pitting and will severly limit the life of the new parts.
Do not stone or burnish the commutator as this will increase brush wear and may result in smear metal on the surface of the commutator.
Do not allow brushes to wear to the point where they do not make contact with the commutator. This causes severe arcing, resulting in excessive brush and commutator wear.
SORVALL® Centrifuges
CAUTION
Figure 4-2. Motor Brush Assembly
CHAMBER TEMP° C Indicator Adjustment (RT6000B only)
With the POWER switch set to "0", use a screwdriver and adjust the adjusting screw until the pointer reads -20°C. (Refer to figure 4-3.)
Figure 4-3. CHAMBER TEMP° C Indicator Adjustment
Parts Ordering Information
To order replacement parts, telephone toll free (800) 551-2121 in the United States. Outside the United States, contact your local distributor or agent for SORV ALL® Products. Be sure to provide a description of the part, the centrifuge model and serial number.
4-4
SORVALL® Centrifuges
Maintenance
Service Decontamination Policy
WARNING
Because of the characteristics of the samples likely to be processed in this centrifuge, biological or radioactive contamination may occur. Always be aware of this possibility, and take normal precautions. Use appropriate decontamination procedures should exposure occur.
If a centrifuge or rotor that has been used with radioactive or pathogenic material requires servicing by Du Pont personnel, either at the customer’s laboratory or at a Du Pont facility, comply with the following procedure to ensure the safety of all personnel:
1. Clean the centrifuge or rotor to be serviced of all encrusted material and decontami­nate it (see Maintenance Section of the centrifuge or rotor instruction manual) prior to servicing by the Du Pont representative or returning it to the Du Pont facility. There must be no radioactivity detectable by survey equipment.
The SORVALL® Rotors, Tubes, Bottles, Adapters, and Accessories Catalog contains descriptions of commonly used decontamination methods and a chart showing method compatibility with various materials. The centrifuge or rotor instruction manual contains specific guidance about cleaning and decontamination methods appropriate for the product it describes.
Clean and decontaminate your centrifuge or rotor as follows:
For tabletop centrifuges:
a. Remove rotor from the rotor chamber. b. Remove motor sealing gasket and wash with appropriate decontaminant. c. Decontaminate cover, bowl, and drive using an appropriate method.
For rotors:
a. Remove tubes, bottles, and adapters from the rotor and decontaminate rotor
using an appropriate method. If tubes or rotor caps are stuck in the rotor, or the rotor lid is stuck, notify Du Pont representative; be prepared with the name and nature of the sample so the Du Pont Chemical Hazards Officer can decide whether to authorize the rotor's return to a Du Pont facility.
2. Complete and attach Decontamination Information Certificate (SORVALL® Instru­ments Form No. IPDP-59 or E53603) to the centrifuge or rotor before servicing or return to Du Pont facility. If Certificate is not available, attach a written statement verifying decontamination (what was contaminant and what decontamination method was used).
If the centrifuge or rotor must be returned to a Du Pont facility:
1. Contact your Du Pont representative to obtain a Return Service Order Number (RSO#); be prepared with the name and serial number of the centrifuge or rotor and the repairs required.
4-5
Maintenance
2. Send item(s) with the RSO# clearly marked on the outside of packaging to the address obtained from your Du Pont representative.
SORVALL® Centrifuges
NOTE
United States federal regulations require that parts and instruments must be decontaminated before being transported.
If a centrifuge or rotor to be serviced does not have a Decontamination Information Certificate attached and, in Du Pont's opinion presents a potential radioactive or biological hazard, the Du Pont representative will not service the equipment until proper decontami­nation and certification is complete. If Du Pont receive a centrifuge or rotor at its Service facilities which, in its opinion, is a radioactive or biological hazard, the sender will be contacted for instructions as to disposition of the equipment. Disposition costs will be borne by the sender.
Decontamination Information Certificates are included with these instructions. Addi­tional certificates are available from your local Account Representative or Field Service Engineer . In the event these certificates are not available, a written statement certifying that the unit has been properly decontaminated and outlining the procedures used will be acceptable.
NOTE
The Field Service Engineer will note on the Customer Service Repair Report if decontamination was required and, if so, what the contaminant was and what procedure was used. If no decontamination was required, it will be so stated.
4-6
SORVALL® Centrifuges
Maintenance
4-7
APPENDIX
Warranty
Warranty
E. I. Du Pont de Nemours and Company makes no warranty of any kind, expressed or implied, except as stated in this warranty policy.
®
The SORVALL defects in material and workmanship for a period of one (1) year from the date of delivery. Du Pont will repair or replace and return free of charge any part which is returned to its factory within said period, transportation prepaid by user, and which is found upon inspection to have been defective in materials or workmanship. This warranty does not apply to any damage to any instrument resulting from: normal wear and tear; misuse; abuse; use of electrical currents or circuits other than those specified on the plate affixed to the instrument; or use of any rotor other than those intended for use in this instrument.
Du Pont reserves the right to change, alter, modify or improve any of its instruments without any obligation whatsoever to make corresponding changes to any instrument previously sold or shipped.
RT6000B and T6000B Tabletop Centrifuges are warranted to be free from
THE FOREGOING OBLIGATIONS ARE IN LIEU OF ALL OTHER OBLIGATIONS AND LI­ABILITIES INCLUDING NEGLIGENCE AND ALL WARRANTIES, OF MERCHANTABILITY OR OTHERWISE, EXPRESSED OR IMPLIED IN FACT OR BY LAW, AND STATE OUR ENTIRE AND EXCLUSIVE LIABILITY AND BUYER’S EXCLUSIVE REMEDY FOR ANY CLAIM OR DAMAGES IN CONNECTION WITH THE SALE OR FURNISHING OF GOODS OR PAR TS, THEIR DESIGN, SUITABILITY FOR USE, INSTALLATION OR OPERATION. DU PONT WILL IN NO EVENT BE LIABLE FOR ANY SPECIAL OR CONSEQUENTIAL DAMAGES WHATSOEVER, AND OUR LIABILITY UNDER NO CIRCUMSTANCES WILL EXCEED THE CONTRACT PRICE FOR THE GOODS FOR WHICH LIABILITY IS CLAIMED.
Index
Cabinet, Cleaning, 4-2 Centrifuge Specifications, 1-1 CHAMBER TEMP°C Indicator
Adjustment, RT6000B, 4-4
Cleaning,
Cabinet, 4-2 Chamber Door, 4-2 Condenser Fins, 4-2 Drive Spindle, 4-2
Rotor Chamber, 4-1 Compartment Loads, 3-6 Condenser Fins, Cleaning, 4-2 Controls and Indicators, 3-1–3-2 Decontamination Policy, 4-5 Dimensions, Centrifuge, 1-2 Door, Cleaning, 4-2 Door Opening, Emergency, 3-6 Electrical Requirements, 1-2, 2-1 Emergency Sample Recovery, 3-6 FAULT System Check, 4-2 Inspection, 2-1, 4-1 Installation, Centrifuge, 2-2 Installation, Rotor, 3-3 Location, Centrifuge, 2-2 Maintenance, 4-1 Mass, Centrifuge, 1-2 Mechanical Override Loop, 3-6 Motor Brush Replacement, 4-3 Ordering Information, 4-4 Operating Temperature, RT6000B, 1-1 Operation, Centrifuge, 3-4 Parts, Ordering, 4-4 Precool
Chamber, 3-4
Rotor, 3-3 Reducing Rotor Speed, 3-6 REV/MIN x 1000 Indicator
Adjustment, 4-2 Rotor Chamber , Cleaning, 4-1 Rotor Installation, 3-3 Rotor RCF Information, 1-2 Rotor Specifications, 1-2 Rotor Temperature, 3-2
Running Hazardous Material, 3-4 Safety Information, iii Sample Recovery, 3-6 Service Decontamination Policy, 4-5 Speed
Operating, 1-1
Reducing, 3-6 Temperature, Operating, RT6000B, 1-1 Temperature, Rotor, 3-2 Warranty, See Appendix
Loading...