DTS is a world leader in ultra-small, low power, high shock rated, high sample rate data
acquisition and sensing systems.
DTS is based in Seal Beach, California, USA, just south of Los Angeles. DTS has offices in 6
times zones for fast, expert technical support.
1.1. DTS Technical Support
For the fastest technical support, please contact your local DTS technical support
engineer or e-mail support@dtsweb.com
1.2. DTS Web Site
For the most up to date specifications, user’s manuals and other information, please
go to www.dtsweb.com
1.3. DTS Offices
.
.
United States
Eastern Standard Time (EST) (GMT-5)
Please call DTS North American Technical Center
+1 248 427-0045
7:00 AM - 5:00 PM, Monday through Friday
After hours, please e-mail support@dtsweb.com
Pacific Standard Time (PST) (GMT-8)
Please call DTS Corporate Headquarters
+1 562 493-0158
7:00 AM - 5:00 PM, Monday through Friday
After hours, please e-mail support@dtsweb.com
Japan
(GMT+9)
Please contact DTS Japan Technical Center
Rihito Shoji
9:00 AM - 5:00 PM (Japan), Monday through Friday
After hours, please e-mail support.japan@dtsweb.com
(GMT+1)
Please contact Dave Martin, European Regional Manager
+49 17 11 286 033 (German and English language support)
8:00 AM - 5:00 PM (Germany), Monday through Friday
After hours, please e-mail support@dtsweb.com
Asia/Pacific
(GMT+11)
Please call Steve Mitchell, Asia Pacific Regional Manager
+61 438 507 449
8:00 AM - 5:00 PM (Australia), Monday through Friday
After hours, please e-mail support@dtsweb.com
China
(GMT+8)
Please call Xi Tianlu, Asia Pacific Regional Manager
+86 21-6386-7559
8:00 AM - 5:00 PM (China), Monday through Friday
After hours, please e-mail support@dtsweb.com
India
(GMT+5.30)
Please call Dave Martin, Regional Manager
+49 17 11 286 033
8:00 AM - 5:00 PM (Europe), Monday through Friday
After hours, please e-mail support@dtsweb.com
For a quick overview of SLICE, please see the SLICE Buyer’s Guide in Appendix A.
SLICE is an ultra small, low power, high shock rated data acquisition system. SLICE is a
standalone system with microprocessor, memory, sensor excitation and signal conditioning
with options for built-in battery and internal sensors. Systems from 3 to hundreds of
channels can be built-up in 3 channel increments.
2.1. SLICE MICRO and SLICE NANO
SLICE comes in two sizes:
! SLICE MICRO (42 x 42 mm)
! SLICE NANO (26 x 31 mm)
They have the exact same function and circuit boards inside. SLICE MICRO has builtin connectors; SLICE NANO has wires with connectors.
31 mm
26 mm
SLICE NANO SLICE MICRO
42 mm
42 mm
2.2. SLICE Modular Concept
Modular system – Plug multiple SLICEs onto Base SLICE to make a Stack
! Each SLICE “Stack” can accommodate 8 Sensor Input SLICES. Each Bridge
SLICE has 3 analog input channels. You may want multiple “Stacks” if more
channels are needed or placement in different locations makes sense for your
application.
! Each SLICE “Stack” consists of 1 Base SLICE and up to 8 additional Sensor
Up to 8 Bridge SLICEs can be stacked on top of the Base SLICE. Each Bridge
SLICE has these components:
! 3 channels of analog input
! Sensor excitation
! 16 bit, 100 kHz ADC, one per channel
! Software adjustable gain, anti-alias filters, offset, and shunt check
! TEDS sensor ID
A Bridge SLICE MICRO is shown below.
Connector to
next Bridge
SLICE
Sensor input
channels
2.3.3. IEPE SLICE
Available in fourth quarter 2010 in SLICE MICRO size only.
Features:
! 3 input channels
! Power: 26 VDC
! Current source: 2-4 mA
! Supply will have isolated +5 V
! 8 isolated inputs
! Input modes supported:
o CCNO - Contact closure normally open
o CCNC - Contact closure normally closed
o VIHL - Voltage Input High-Low
o VILH - Voltage Input Low-High; 3 input channels
2.3.5. ACCEL SLICE
The ACCEL SLICE has Bridge SLICE electronics with the addition of a built-in 3axis accelerometer. The following specifications apply:
! MSI Model 3038 accelerometers are used (see www.meas-spec.com
)
! Options from 50 to 2000 g
! DC response
Accel and ARS
SLICEs can be
positioned in any
order in a SLICE
MICRO stack
ARS SLICE
ACCEL SLICE
2.3.6. ARS SLICE
The ARS SLICE has Bridge SLICE electronics with the addition of a built-in 3-axis
angular rate sensor. The following specifications apply:
The Battery SLICE is connected to the bottom of the Base SLICE. It is only
available in the SLICE NANO version. The Battery SLICE is only a back-up
battery in case main power is lost. Specifications:
! 2 cell LiPo design, with charging directly from Base SLICE
! ~10 second run time
! Only 3.5 mm thick
2.3.8. Stack Extender
The Stack Extender is only available for the SLICE NANO package. The Stack
Extender allows the user to create a flatter, longer package.
Stack Extender
! Example: 12 channel system
! 1 Base SLICE
! 4 Bridge SLICEs
2.3.9. End-of-Chain (EOC) Terminal
See Appendix C for detailed specifications.
The EOC Terminal provides the easiest method to attach a battery, trigger signal
and status lamp to the SLICE system. It is ruggedized for high shock use.
Maximum channels:
1 chain x 2 stacks x 8 Bridge SLICEs x 3 chan/Bridge = 48 channels
SLICE systems use lightweight, rugged plastic connectors with reliable, gold
plated contacts. These are a MIL-STD-type pin and socket configuration where
the socket is exposed, instead of the pin, which is mechanically more robust.
For some bench-top units, LEMO-style as well as industry standard USB,
Ethernet and SubD connectors are used.
A typical SLICE rugged plastic connector is shown below.
Finger actuated
retention latch
Exposed sockets
Recessed pins
2.4. Batteries
DTS offers some commercial-off-the-shelf batteries for operation of SLICE systems.
Batteries must be disconnected from the SLICE system before connecting to a charger.
2.4.1. 9.6 V Rechargeable NiMH Batteries
! Allows up to 40 min runtime with a 6 channel SLICE System.
! Package of 4 batteries ensures you’ll always have back-ups ready to use.
2.4.2. 11.1 V Rechargeable Lithium-Polymer Batteries
! Available in 3 capacities: 2200, 4400, and 6600 mAh
2.5. SLICE Software
See Section 4.0 for detailed software information.
The SLICEWare software application allows for easy:
! Test set-up
! Sensor database management
! Real-time sensor check-out
! Test execution
! Data download and viewing
! Data export
A SLICE API (Application Programmers Interface) and LABView driver (fourth quarter
2010) are also available.
Please contact technical support (support@dtsweb.com
) for the latest update to your
software version.
3. Mounting and Connecting SLICE Hardware
This section gives details on how to connect your SLICE hardware. Choose the connection
method you have from the options below for the quickest information.
3.1. General Connection Guidelines
Great care should always be taken when connecting any power, switch, sensor or any
other device to the SLICE system.
! DO NOT exceed the rated voltage input range for the device. Whenever
possible use the power supply or battery pack supplied with your SLICE system.
! DO NOT connect directly to vehicle power or other noisy power sources.
! ALWAYS disconnect the battery from the SLICE system before connecting to a
battery charger.
! ALWAYS use SLICE NANO with a heat sink as the SLICE NANO case is very thin
aluminum with very little heat sinking ability. Never use SLICE NANO mounted
to a thermally non-conductive surface like wood or plastic.
! Refer to proper grounding procedures described in Appendix H.
! Check that all cables show no signs of physical damage.
! Be sure all sensors have their cable shields ungrounded at the sensor end and
grounded at the SLICE input connector.
3.2. Guidelines for High Shock and Vibration Testing
SLICE MICRO and SLICE NANO components can generally be used in test environments with maximum acceleration levels as high as 500 g. In some cases, SLICE
NANO can be used up to 5,000 g if proper care is used.
Proper mounting of the SLICE system, cables, and accessories is critical to successful
testing.
! DO NOT mount SLICE components in an area where they may be directly
impacted by an object.
! Use damping material whenever possible to help protect the SLICE system
from excessive shock or vibration, but remember that SLICE NANO requires a
heat sink.
! Be sure that connectors and wiring are properly secured.
3.3. SLICE Connectors and Cables
3.3.1. SLICE Connectors
SLICE is an ultra small data acquisition system. One challenge with a small
system is electrical connections. Although connectors such as Bendix, Amphenol,
and LEMO are common for instrumentation, all of these are much too large to be
practical for SLICE.
The SLICE system uses circular plastic connectors manufactured by Omnetics
Corporation (www.omnetics.com
contacts and are used in many military, aerospace and other high shock applications. Connectors are available direct from Omnetics or can be purchased from
DTS.
DTS provides a number of different SLICE cable options depending on the
connection needs. SLICE stack to stack connection cables are shown below.
SLICE MICRO Base units can be connected together via a daisy-chain cable.
SLICE NANO Base units can be connected to each other directly or via a daisychain extension cable.
See Appendix C for detailed information on the SLICE End-of-Chain Terminal.
3.6. Using the SLICE System Interface
The SLICE System Interface is similar to the EOC Terminal except it is used between
the first/only SLICE Stack and the PC. See the example diagram below.
! Although this shows a SLICE MICRO, connections with a SLICE NANO are
similar.
! The SLICE System Interface is a shock rated item.
! The AUX Terminal can be used for easy connection to a battery, start record,
trigger, and status signal.
! Voltage input through the SLICE System Interface can be provided via the
included power supply a battery, or any voltage source between 9 and 20 VDC.
Warning: Do not exceed the 20 VDC input voltage range as damage may result.
! The SLICE System Interface has a recessed ON switch. Alternately the ON
signal can be connected to the GND on the EOC Terminal.
! If you connect an LED between the STS and GND terminals on the EOC
Terminal you will get a Status light when the system is armed.
3.7. Using the SLICE USB Interface
The SLICE USB Interface is similar to the SLICE System Interface. The main
difference is that the SLICE USB Interface is designed for bench-top, non-rugged use.
See the example diagram below.
! Although this shows a SLICE MICRO, connections with a SLICE NANO are
similar.
! The SLICE USB Interface is NOT a shock rated item.
! Voltage input can be with the included power supply a battery, or any voltage
source between 9 and 15 VDC. Warning: Do not exceed the 15 VDC input
voltage range as damage may result.
! The AUX connector is a standard D-sub HD15. This can be used to hardwire a
Start Record or Event switch or monitor the Status line.
3.8. Using the SLICE Ethernet Interface
The SLICE Ethernet Interface is similar to the SLICE USB Interface. The main
difference is that the SLICE Ethernet Interface has an Ethernet connection to the PC
instead of a USB connection. This allows for a longer communications cable between
the PC and the SLICE system. The SLICE Ethernet Interface can also connect 2 SLICE
chains for large system configurations. See the example diagram below.
Notes:
! Although this shows a SLICE MICRO, connections with a SLICE NANO are
! The SLICE Ethernet Interface is NOT a shock rated item.
! Voltage input can be with the included power supply a battery, or any voltage
source between 9 and 15 VDC. Warning: Do not exceed the 15 VDC input
voltage range as damage may result.
! The AUX connector is a standard D-sub HD15. This can be used to hardwire a
Start Record or Event switch or monitor the Status line.
3.9. Using the SLICE Distributor
The SLICE Distributor allows for the connection of up to 4 SLICE chains and converts
the communications signals from USB to Ethernet. This allows for a longer communications cable between the PC and the SLICE system. The most common application
for the SLICE Distributor is for an embedded system with a high channel count as
shown for the in-dummy (manikin) configuration below.
This section covers basic information regarding SLICE compatible sensors and sensor ID.
More detailed information regarding sensor connections can be found in Appendix I.
The diagram below shows a common 4-wire bridge sensor connection to a Bridge SLICE
input channel.
4.1. Sensor ID
Sensor ID is also referred to as electronic ID (EID). The function of EID is for the
SLICE hardware to automatically read and determine what sensor is attached to each
sensor input channel.
SLICE uses EID chips from Maxim IC, model DS2401 (see http://www.maxim-
ic.com/quick_view2.cfm/qv_pk/2903). To make soldering of the EID more easy, DTS
provides EID microcards, which have a chip scale packaged DS2401 soldered to a
circuit board with wires attached (see below).
The Bridge SLICE supplies 5 VDC excitation up to 20 mA and supports many types of
accelerometers, load cells, pressure sensors and other sensor types. The following
general sensor types are supported:
! Full (4-wire) or half bridge (2- or 3-wire) resistive and piezo-resistive types
! Voltage input: Input range 0.1 to 4.9 V; larger range with voltage expander
circuit
! Conditioned sensors with 5 V excitation and 2.5 V centered signal output
The IEPE SLICE MICRO (fourth quarter 2010) will support common piezo-electric
sensor types.
If you have questions regarding what sensors are supported please contact
support@dtsweb.com
and provide the sensor manufacturer and model number if
available.
5. Software
This section covers software installation and use.
See Appendix J for additional software reference information regarding files formats and
firmware updates.
5.1. Basic Requirements
®
SLICEWare is a Windows
! Windows XP, Windows Vista, or Windows 7 32-bit. 64-bit support is coming
soon
! 1 GHz or faster processor
! 2 GByte RAM minimum. More RAM is important for longer/higher sample rate
data acquisition.
! 100 MByte disk space for Software plus storage for test data
! 1024 x 768 minimum screen resolution
This section discusses the basics of data collection with SLICE.
5.2.1. Standalone Operation
SLICE is a standalone data logger. This means that once it is armed, the PC can
be disconnected if desired. After receiving a Start Record or Trigger signal, the
SLICE autonomously collects data, storing it to flash memory with no user
interaction. After the test, the user can reconnect the PC to download the data.
There is also a real-time mode in the SLICEWare software application that
allows the user to check channel inputs on an oscillograph-looking screen.
5.2.2. Data Collection Modes
SLICE supports two data collection modes, Circular Buffer and Recorder.
5.2.2.1. Circular Buffer Mode
In circular buffer mode, the user can program SLICE to record pre-trigger
data. For example, the test set-up can specify to record x seconds pretrigger and x seconds post trigger. Time Zero (T=0) is marked when the
trigger signal is received.
5.2.2.2. Recorder Mode
5.3. SLICEWare
This section discusses the basics of data collection with SLICE using the SLICEWare
application.
The SLICEWare software application allows for easy:
! Test set-up
! Sensor database management
! Real-time sensor check-out
! Test execution
! Data download and viewing
! Data export
Recorder mode starts when a Start Record signal is received and continues
for the time specified in the test set-up. If a trigger signal is received
sometime after the Start Record, this marks the T=0 point.
for each of the screens: set-up wizard, driver installation,
SLICE User’s Manual support@dtsweb.com
You must allow the driver to be installed. You may see these screens:
For Windows
®
Vista, click to “Install this driver software anyway.”
For Windows
®
XP, click “Continue Anyway.”
Note: Windows
®
will ask you to reinstall the hardware driver each time you
connect the SLICE stack to a different USB port.
To start SLICEWare, either double-click the
icon or go to:
5.3.2. Menu Descriptions
Click Prepare tab …
This tab identifies the relationship between available sensors and the
attached SLICE units. Sensor information can be added, removed, viewed,
edited or assigned to channels on connected data acquisition hardware.
! Add: Will add a blank Sensor Details field to all for creating a new sensor
entry.
! Delete: Deletes the sensor that is highlighted in the sensor list.
! Undo Edit: Reverts all edits made in the sensor details fields.
! Read ID: Reads the EID from the channel selected in the Connected
Sensor list. The returned ID value will be populated in the Sensor ID field
for the sensor that is highlighted in the sensor list.
! Measure Bridge: Measures a sensors bridge resistance from the channel
selected in the Connected Sensor list. The returned bridge resistance will
be populated in the Bridge resistance (ohms) field for the sensor that is
highlighted in the sensor list.
! Save: Saves edits made in the Sensor Detail area.
Manual Sensor Assignment
! For sensors connectors without an EID.
! You cannot un-assign or overwrite an auto-assigned channel.
! Assign: After highlighting a sensor in the Sensor List and highlighting an
un-assigned channel in the Connected Sensors area, use this to assign the
sensor.
! Un-assign: Remove the highlighted channel in the Connected Sensors area.
! Apply: Commits the sensor set-up information to SLICE.
! Note: A sensor that is manually applied should not have a value in the
Sensor ID field and the SLICE should not have an EID installed on the
connector. If the Sensor ID field is populated or an EID exists on the
channel, the sensor will need to be re-applied after switching away from
and then back to the Prepare tab.
! This field is automatically updated whenever a new sensitivity is applied to
the sensor attributes. You cannot enter directly into this field.
Options
! Invert: When checked, the data will be inverted.
! Shunt Check and Bridge Resistance: When Shunt Check is checked, the
sensor will have the bridge resistance measured during diagnostics and
compared to the value entered in Bridge resistance.
! SW Filter: Choose the frequency of a software filter to be applied to the
data when viewing. This only affects the viewed data as all data stored will
be as collected with the hardware anti-alias filter.
! Zero Method (post download software zeroing):
o Use Pre-Cal Zero: The Zero Measured Output (ZMO) of the sensor
during calibration will be used to set the EU zero of the downloaded
data.
o Average Over Time: Used in conjunction with ZeroStart and
ZeroEnd, the average EU value during the Start and End window
will be used to Zero the collected data. The Zero Start/End window
must be set to data that will be collected. If using a negative time,
then the Acquire tab must include this window.
o None: The actual recorded input will not be adjusted or
compensated for zero level. This setting can be used to show the
actual mV offset. An example may be to record a logic level signal
and see the actual on/off state.
o Remove Offset (hardware): When checked, this will remove the
ZMO during diagnostics. This will “electrically” zero the input.
o ZeroStart/End: See Zero Method!Average Over Time.
o Offset Check: Used in conjunction with Limit Low/High during
diagnostics. When checked, the ZMO is measured and compared
the Low/High limits as a pass/fail criteria during diagnostics.
Step-by-Step Procedure to Add a New Sensor
1. On the PREPARE tab,
click the “Add” button in
the “Current Sensor”
button group
This tab ensures that the connected hardware is operating normally.
Hardware diagnostics include checks for battery level, excitation voltage,
noise and expected offset.
This tab shows data input to the attached SLICE hardware in real-time. It is
a useful tool for establishing confidence in the current hardware configuration prior to data collection.
This tab configures and controls the data acquisition process. The user
enters the test name, description, sample rate, acquisition mode (circular
buffer or recorder mode), pre- and post-trigger times and then prepares the
system for data acquisition with the arm command. After the test is
completed, the user can use the download command to view the data.
This section covers what to expect when powering up a SLICE system and running a test.
The LEDs on the Base SLICE indicate the status of the system.
6.1. Status (STS) LED
Action Result
Power up
Communicating with PC
Recording Data (Recorder Mode) -or- Armed (Circular Buffer)
Armed in Recorder Mode
Unit received Event
Idle
The status LED is red, green or blue. At system power up, the LED cycles from red to
green to blue followed immediately by the power LED boot-up sequence.
The status LED indicates communication and arm status.
! When the unit is not armed, the status LED will blink green when handling a
command from the PC.
! For Recorder Mode
o When the unit is first armed, the LED will go solid blue to indicate that it
is waiting for the START RECORD signal but not taking data.
o When it receives the START RECORD signal, the LED will turn green to
indicate that it is actively recording data.
o The LED will turn off when data collection has completed.
o If an EVENT signal is received while the unit is recording data, the LED
will turn red and then turn off when data collection has completed.
! For Circular Buffer Mode
o When the unit is armed, the LED will go solid green to indicate that it is
collecting data and waiting for the EVENT signal.
o When an EVENT signal is received the LED will turn red and then turn off
Lid ~9 grams 2 (0.079)
Accel ~33 grams 9 (0.354)
ARS ~33 grams 9 (0.354)
Bridge ~25 grams 7 (0.276)
Base ~28 grams 8 (0.314)
Assembly Screw
Length (FH)
Mounting Screw Length (min)
(BH or SHC)
Total Stack Height
mm (inch)
M3* M4** 6-32***/8-32**
17 (0.67) 16 mm 22 mm 7/8"
24 (0.95) 20 mm 30 mm 1-1/4"
26 (1.02) 25 mm 35 mm 1-1/4"
31 (1.22) 30 mm 35 mm 1-1/2"
33 (1.30) 30 mm 40 mm 1-1/2"
35 (1.38) 35 mm 40 mm 1-5/8"
38 (1.50) 35 mm 45 mm 1-3/4"
40 (1.57) 40 mm 45 mm 1-3/4"
42 (1.65) 40 mm 50 mm 2"
45 (1.77) 45 mm 50 mm 2"
47 (1.85) 45 mm 55 mm 2-1/4"
49 (1.93) 45 mm 55 mm 2-1/4"
52 (2.05) 50 mm 60 mm 2-1/4"
54 (2.13) 50 mm 60 mm 2-1/2"
56 (2.21) 55 mm 70 mm 2-1/2"
59 (2.32) 55 mm 70 mm 2-1/2"
61 (2.40) 60 mm 70 mm 3"
63 (2.48) 60 mm 70 mm 3"
66 (2.60) 65 mm 70 mm 3"
68 (2.68) 65 mm 80 mm 3"
70 (2.76) 70 mm 80 mm 3"
73 (2.87) 70 mm 80 mm 3"
75 (2.95) 75 mm 80 mm 3-1/2”
77 (3.03) 75 mm 90 mm 3-1/2”
80 (3.15) 80 mm 90 mm 3-1/2”
82 (3.23) 80 mm 90 mm 3-1/2”
84 (3.31) 80 mm 90 mm 3-1/2”
Specifications may be revised without notice. Torque specs: * 5.2 in-lb (0.59 Nm); ** 19.8 in-lb (2.24 Nm); *** 9.6 in-lb (1.1 Nm)
1 /ON 1 /START, CC to ground
2 /START 2 +Status out
3 /EVENT 3 /EVENT, CC to ground
4 STATUS 6 Ground
5 12.6 VDC out 7 -Status out
6 12.6 VDC out 8 Ground
7 Ground
8 Ground
9 USB power
10 USB_DP
11 USB_DM 15V IN
12 Ground (ECG.2B.304.CLL)
1
This is a standard USB (“B”) interface. A commercial, off-the-shelf
USB cable is acceptable.
Electromagnetic Interference (EMI), Radio Frequency Interference (RFI) and
Electrostatic Discharge (ESD) can seriouslydegrade the performance of
electronic equipmentif not addressed. DTSSLICEsystemscontain protection
f
or
EMI/RFI/ESD, h
owever, many dynamic testing environments
(pyrotechnics, blast) are particularly noisy and require the utmost attention
to grounding and shielding practices. The following recomm endations are intended tomaximize protection and keep systems functioning properly in
the harshest environments
.
Ground all DAS equipment,powersupplies andsensor mounting fixtures
wheneverpossible. This is an extremelyimportant s tep toward ensuringthe
best performance from yourSLICE system.
•Always connect a cable from a good Earth groundto the test article, test fixture or
instrumentedvehicle. Notonly doesthis help divertpotentially disruptive electrical
energy,itis alsogoodsafetypractice. Forremote testing applications, a metalgroundrod driven 3 ftintothe soilcan be an effective Earthingdevice.
•Ground
all SLICE enclosures to the test article o
rv
ehicle
•Installgroundcablesbetween allSLICEStacks andthe test article orvehicle.
will cause large ground-loop currents that can increase noise or cause damage.
• P C Grounding?
g
•Thisismore importantthan you mightthink.
•If the Laptop usedtocommunicate with SLICEispowered from a source that has a
significantly differentgroundpotential than the SLICE system,communication with the
SLICE can be impaired. In severe cases damage tothe laptoporSLICE can occur .
•Eitherrun the la
ptop
on battery power or use a voltmeter to make sure the AC outlet
ppyp
g
roundis not at a significantlydifferentpotentialthan the groundconnectedto the test
article orvehicle.
• Carefullyconsider routing andcable design for any high currentsignalsto air bags,
cameras, lights, etc.
•
Route these cables away from sensor wiring
.
•Cross sensor wiring at90" angles if the cables must cross.
•Flat braided ground cable has lower impedance than typical round
wires and hence makes a bettergroundconnection.
•Never assume that co nnections are good until you checkthemwith
an ohmmeter. Shouldbe <1# forshortrunsor <5# forlongruns.
•Groundcablesinside testdummiesshouldbe braidedtype with a 12-
gage equivalentsize.
•Groundcables from a test article orvehicle tothe SLICEStackshould
be braidedstraptype with a 15-gage equivalentsize.
•The cable from a test article orvehicle tothe Earth connection
should be large enough tocreate an a low impedance connection
given the distance between Earth connection and test vehicle. 8 to 12gage equivalent iscomm o n.
ggq
•If braided cable is not available, any ground wire is better than none!
DAC usedto null
offset.Mustzero
to < 2% A/ D f/ s.
Electronic ID
Interface
+ID
Sta
ndard 4-wire Bridge Connecti
on
-SIG
+SIG
Excitation
5-volts 20 mA
+EX
-EX
+ID
-ID
Shield
3
$EX
+EX
3K
3K
Half-Bridge Switch
1
2
5
+SIG
$SIG
+ex
+sig
-sig
-
SensorBridge
Amplifie
r
6
7
ex
$ID
Shield
2.5V
4
+ID
! InternaltoSLICEBridge
Electronic ID
Interface
SensorSide "
Strai
nGage 3-wire Connecti
on
-SIG
+SIG
Excitation
5-volts 20 mA
+EX
-EX
+ID
-ID
Shield
350# 0. 1%
3
$EX
+EX
3K
3K
Half-Bridge Switch
350# Gag
e
1
2
5
+SIG
$SIG
Amplifie
r
6
7
$ID
2.5V
4
+ID
! InternaltoSLICEBridge
Electronic ID
Interface
SensorSide "
Strai
nGage 2-wire Connecti
on
-SIG
+SIG
+EX
-EX
+ID
-ID
Shield
Excitation
5-volts 20 mA
3
$EX
+EX
3K
3K
Half-Bridge Switch
350#
0.1%
350# Gag
e
1
2
5
+SIG
$SIG
Amplifie
r
6
7
$ID
2.5V
4
+ID
! InternaltoSLICEBridge
Electronic ID
Interface
SensorSide "
Mea
suring Large Voltages - Example
(5V) + EX
1
LemoPlug
+ SIG
3
- SIG
Amplifie
r
_
+
10K
10K
10K
R
External
Voltage
4
2
10K
R
SLICEBridge
(0V) - EX
Resistance
R
ApproxMAX
ExternalVoltage
Vmax
* Sensitivity
mV/V
*
95.3K
49.850
+ / -40V
150K
32.258
+ / -60V
45.3K
99.404
+ / -20V
Sensitivitycalculation
…
.
5
(5 + R)
( )
X1000
Contents
1 Overview•
2 XML Structure
2.1 <Test>
2.1.1
<Modules>
2.1.1.1
<TriggerSampleNumbers>
⋅
2.1.1.2
<Channels>
2.1.1.2.1
<AnalogInputChanel>
•
⋅
◊
♦
•
Overview
The .DTS file is an XML based file that contains information about the overall test and the individual
channels. Some of the information may be redundant with information stored in the binary channel header.
The attributes and relationships of each XML node are described below
XML Structure
<Test>
The Test tag is the outer most tag. It's attributes describe details common to the entire test. It contains the
following attributes.
NameData typeDescription
IdString
The name of the test, typically the same as the .DTS
file name
DescriptionStringThe description of the test provided by the user.
InlineSerializedDataBoolean
GuidWindows UUID stringA unique identifier assigned to each event
<Modules>
Within the Test node will be a list of modules contained within a <Modules></Modules> tag. A module
corresponds to a data acquisition system - for example an entire stack. Each module will have it's own
<Module> tag with the following attributes.
Name
Data
type
Description
AaFilterRateHzInteger
Appendix J: Software Information
XML Structure
The name cut off frequency of the hardware anti-alias filter
used during the test
NumberIntegerA sequential number assigned to each module within the test
SerialNumberStringThe factory assigned serial number of the base
NumberOfSamplesInteger
The number of samples stored in each channel file. This will be
fewer than the number of samples originally requested by the
user if the data has been sub sampled or if only a portion of the
data was downloaded.
UnsubsampledNumberOfSamples IntegerThe total number of samples collected during data acquisition
PostTriggerSecondsDouble
The number of seconds of recorded data that the user requested
after t=0.
PreTriggerSecondsDouble
The number of seconds of recorded data that the user requested
before t=0.
RecordingModeString
Either the value RecorderMode or CircularBuffer. Other values
will be added in the future.
SampleRateHzIntegerThe rate at which sampling occurred during data collection.
StartRecordSampleNumberInteger
The sample number at which the start signal was first detected.
The value will always be 0 when
RecordingMode=CircularBuffer
NumberOfChannelsIntegerThe number of user configured channels within the module
InlineSerializedDataBoolean
<TriggerSampleNumbers>
This is a list (possibly 0 length) of trigger sample numbers. In the circular buffer case there will be one trigger
sample number. In recorder mode the trigger is optional. In the case of multiple event mode, there may be
more than one trigger sample number.
<Channels>
The Channels tag contains a list of channel elements. It should have the same number of entries as
NumberOfChannels in the Module tag. The type of the child elements will depend on the type of signal
conditioning SLICE used.
<AnalogInputChanel>
The AnalogInputChanel tag corresponds to a Bridge SLICE channel.
Note:
There is a typo in the tag name. Channel is misspelled. It is left for backwards compatibility.
Many of the attributes indicate how the channel was configured during the test. The AnalogInputChanel
element has the following properties:
NameData typeDescription
ChannelTypeStringThis identifies the representation of the data contained in
the .BIN file. Currently this value is always expected to be
SLICEWare_DTS_File_Format
<Modules>
DTS.Serialization.Test+Module+AnalogInputChannel
NumberInteger
The channel number within the signal conditioning unit. In
a Bridge SLICE channels are numbered 0 - 2
StartDateCurrently unused
BridgeStringEither FullBridge or HalfBridge
BridgeResistanceOhmsInteger
The specified bridge resistance used during the shunt
check.
ChannelDescriptionStringStringThe user provided description for the channel
DescriptionString
The user provided description for the sensor. Currently the
same as ChannelDescriptionString
SensitivityDouble
The sensitivity of the sensor in either mv/V/EU or mv/EU
depending on ProportionalToExcitation
SoftwareFilterString
The requested filtering to apply to this channel. Stored data
is unfiltered, and this value must be used to apply proper
filtering. Typical values are "1650hz" for CFC1000.
ProportionalToExcitationBoolean
Indicates if the output of this sensor is proportional to
excitation. Used in conjunction with Sensitivity.
IsInvertedBoolean
(Optional)
Indicates if the data should be inverted before
presenting to the user. If missing, this attribute is
considered 'false'.
IsSubsampledBoolean
(Optional) Indicates if the data stored on disk is at a lower
sample rate than the original data collection. If missing,
this attribute is considered 'false'.
ExcitationVoltageString
Indicates the nominal excitation voltage. Valid values are
currently Volt5 and Volt2.5.
EuString
The user provided Engineering Units (EU). For example
"mm", "g", or "msec2"
SerialNumberStringThe serial number of the sensor used with this channel
ShuntEnabledBoolean
Indicates if the user requested the channel be shunted
during diagnostics.
RemoveOffsetBoolean
Indicates if the user requested hardware offset
compensation be used during diagnostics.
ZeroMethodString
Identifies the type of software offset compensation that
should be used. If the value is "UsePreCalZero" then the
Pre Calibration zero value stored in the channel file should
be used. If the value is "AverageOverTime" then an
average value computed from the channel data should be
used.
ZeroAverageWindowBeginDouble
If ZeroMethod=AverageOverTime this is the beginning of
the window to be used for computing the average.
ZeroAverageWindowEndDouble
If ZeroMethod=AverageOverTime this is the end of the
window to be used for computing the average.
InitialEuDouble
A value provided by the user that should be subtracted
from all scaled data in addition to the selected ZeroMethod
UnsubsampledSampleRateHzInteger
The sampling rate used during data collection. Valid only if
IsSubsampled=true
SLICEWare_DTS_File_Format
<AnalogInputChanel>
MeasuredShuntDeflectionMvDouble
(Optional)
If a shunt test was performed, the actual
deflection of the shunt
TargetShuntDeflectionMvDouble
(Optional) If a shunt test was performed, the expected
shunt deflection
MeasuredExcitationVoltageDouble
(Optional) The measured excitation voltage, if available.
Used by SliceWARE for scaling proportional-to-excitation
sensor data if "factory" excitation voltage is not available.
FactoryExcitationVoltageDouble
(Optional) The factory excitation voltage, if available.
Used by SliceWARE for scaling proportional-to-excitation
sensor data.
SLICEWare_DTS_File_Format
<AnalogInputChanel>
SLICE Base Firmware
Update Procedure
To update the SLICE MICRO™ or SLICE NANO™ Base firmware, you need:
1. Hardware: SLICE USB cable or SLICE SSI Cable Kit.
2. Software: ZIP file (e.g., eSensing Firmware Updater 1.06.zip) extracted on your
PC. The ZIP file contains the SLICE Firmware Updater program
(eSENSING_FirmwareUpdater.exe) and required support files.
3. Firmware: Firmware version (e.g., SLICEBaseFirmware_BASE_00G6.sfw) you
want to install.
4. Procedure:
1. With PC on, connect the SLICE Base to the PC via USB. Power-up the SLICE
Base.
2. Start the SLICE Firmware Updater (eSENSING_FirmwareUpdater.exe). The
screen will show the serial number of the connected Base: