Dräger X-am 5100, MQG 0020 Technical Manual

Dräger X-am® 5100
(MQG 0020)
Technical Manual
Contents
1 For Your Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.1 Follow the Instructions for Use . . . . . . . . . . . . . . . . 4
1.2 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.3 Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.4 Safe coupling with electrical devices . . . . . . . . . . . . 4
1.5 Use in areas subject to explosion hazards . . . . . . . 4
1.6 Safety symbols used in this Technical Manual . . . . 4
2 Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2.1 Product overview . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2.2 Intended Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2.3 Intended operating area and operating conditions . 5
2.4 Tests and Approvals . . . . . . . . . . . . . . . . . . . . . . . . 6
3 Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
3.1 Preparations for use . . . . . . . . . . . . . . . . . . . . . . . . 7
3.2 Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
3.3 Performing the bump test . . . . . . . . . . . . . . . . . . . 11
3.4 During use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
3.5 Identifying Alarms . . . . . . . . . . . . . . . . . . . . . . . . . 12
4 Menu functions . . . . . . . . . . . . . . . . . . . . . . . . . . 13
4.1 Calling the Info Mode . . . . . . . . . . . . . . . . . . . . . . 13
4.2 Calling the Info-Off Mode . . . . . . . . . . . . . . . . . . . 13
4.3 Quick Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
4.4 Adjustment menu . . . . . . . . . . . . . . . . . . . . . . . . . 14
5 Calibrating the instrument . . . . . . . . . . . . . . . . . 15
5.1 Adjustment interval: . . . . . . . . . . . . . . . . . . . . . . . . 15
5.2 Performing a fresh air adjustment . . . . . . . . . . . . . 15
5.3 Performing a sensitivity adjustment . . . . . . . . . . . 16
6 Replacing the Sensors . . . . . . . . . . . . . . . . . . . . 18
7 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . 19
7.1 Warning messages . . . . . . . . . . . . . . . . . . . . . . . . 19
7.2 Fault messages . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
8 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
8.1 Maintenance intervals . . . . . . . . . . . . . . . . . . . . . . 21
8.2 Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
9 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
9.1 Disposal of electrical and electronic appliances . . 22
9.2 Electrochemical sensors . . . . . . . . . . . . . . . . . . . . 22
10 Technical Data . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
11 Ordering list . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
12 Declaration of Conformity . . . . . . . . . . . . . . . . . 25
Contents
Dräger X-am 5100 3
For Your Safety

1 For Your Safety

1.1 Follow the Instructions for Use

Any use of the instrument requires full understanding and strict observation of the Instructions for Use supplied with the instrument. The instrument is only to be used for the purposes specified here.

1.2 Maintenance

The maintenance intervals and measures specified in this Technical Manual as well as the specifications in the Instructions for Use/data sheets sensors used must be carefully observed. Repairs to the instrument may only be carried out by trained service personnel.

1.3 Accessories

Only the accessories specified in the ordering list may be used.

1.4 Safe coupling with electrical devices

Devices which are not mentioned in the Instructions for Use or in this Technical Manual can only be coupled electronically after consultation with the manufacturers or an expert.

1.5 Use in areas subject to explosion hazards

Devices or components for use in explosion-hazard areas, which have been tested and approved according to national, European or international Explosion Protection Regulations, may only be used under the conditions explicitly specified in the approval and with consideration of the relevant legal regulations. Devices and components must not be altered. The use of faulty or incomplete parts is forbidden. The applicable regulations must be observed at all times when carrying out repairs on these devices or components.
1
for the DrägerSensor
® 2

1.6 Safety symbols used in this Technical Manual

This Technical Manual contains a number of warnings for risks and hazards which might occur when using the instrument. These warnings contain signal words to alert you to the degree of hazard you may encounter. These signal words and corresponding hazards are as follows:
WARNING
!
Death or serious physical injury may occur as a result of a potential hazard situation if appropriate precautionary measures are not taken.
CAUTION
!
Indicates a potentially hazardous situation which, if not avoided, could result in injury or damage to property or to the environment. Can also be used to warn against any wanton actions.
NOTICE
i
i
Additional information on the use of the product.
1 Instructions for Use/data sheets for the utilized sensors and the PC software
CC-Vision for Dräger X-am 5100 can be downloaded on the product page of the X-am 5100 at the following Internet address: www.draeger.com. See also the enclosed Instructions for Use and data sheets for the sensors used.
2 DrägerSensor
4 Dräger X-am 5100
®
is a registered trademark of Dräger.
Description

2 Description

2.1 Product overview

1
2
X-am 5100
3
45
0
1
1
6
0.0
HCl ppm
7
8
9
10
11

2.2 Intended Use

Portable single gas detection instrument for the continuous monitoring of the concentration of HCl, HF, H in the ambient air within the working area and in explosion­hazard areas.
or hydrazine
2O2

2.3 Intended operating area and operating conditions

2.3.1 Hazardous areas classified by zones

This instrument is intended for use in hazardous areas classified as Zone 0, Zone 1 or Zone 2, or in mines in which there is a danger of firedamp, in a temperature range of -20 °C to +50 °C, and in areas where gases of explosion groups IIA, IIB or IIC and temperature class T3 or T4 (depending on batteries and rechargeable battery) may be present. In mines, the instrument may only be used in areas with a low risk of mechanical influence.
12 13 14
00133279.eps
1 Alarm LED 8 Fastening clip 2 Gas entry 9 Type plate 3 Display 10 Power pack 4 key 11 Charging contacts
OK
5 key 12 Measured gas display 6 Buzzer 13 Measured gas display 7 IR interface 14 Special symbols
Special symbols:
Fault message 1-button sensitivity
Warning message Standard sensitivity
Display peak value Password required
Show TWA Battery 100 % full
Show STEL Battery 2/3 full
Bump test mode Battery 1/3 full
Fresh air adjustment Battery empty
adjustment
adjustment
Dräger X-am 5100 5
Description

2.4 Tests and Approvals

2.4.1 Marking

TC RU C-DE.ГБ06.В.00095 PO Ex ia l X 0 Ex ia IlC T4/T3 X
Dräger Safety Type: MQG 0020
23560 Lübeck, Germany
`
C
I M1/II 1G Ex ia I/IIC T4/T3 Ma/Ga Um=4.6V Im=1.3A BVS 10 ATEX E 080X IECEx BVS 10.0053X
Intrinsically safe Ex ia, CSA 11 1800517 CSA: Class I, Div. 1, Groups A,B,C,D TC T4/T3
-20°C Ta +50/+40°C: see Battery Pack! For TC T4/T3: see Battery Pack! Warning: Read manual for safety precautions. Do not change or charge batteries in haz loc.
Serial No.

2.4.2 Permissible power packs

Power pack 83 22 237, approved as Type ABT 0100
Temperature class T4
-20 °C Ta +50 °C use with alkaline batteries Duracell Procell MN1500
0158
Class I, Zone 0, A/Ex ia IIC T4/T3 /Ga
1
on a separate label

2.4.3 Safety instructions

WARNING
!
Read the safety measures in the Instructions for Use.
Do not replace or charge batteries in potentially explosive areas. Danger of explosion!
To reduce the danger of explosion, do not mix new batteries with old batteries and do not mix batteries made by different manufacturers.
Always disconnect the instrument from the power pack before carrying out any maintenance operations.
Substitution of components may impair intrinsic safety.
High off-scale readings may indicate an explosive concentration.
CAUTION
!
Not tested in oxygen enriched atmospheres (>21 % O2).
Only use power packs ABT 0100 (Order No, 83 22 237), HBT 0000 (Order No. 83 18 704) or HBT 0100 (Order No. 83 22 244). See marking on power pack for approved batteries and related temperature class.
Temperature class T3
-20 °C Ta +40 °C use with NiMH batteries GP 180AAHC (1800 mAh)
use with alkaline batteries Varta Powerone 4006 Varta Powerone 4106 Panasonic Powerline LR6
Power pack 83 18 704; approved as HBT 0000
Temperature class T4
-20 °C Ta +50 °C
Power pack 83 22 244; approved as HBT 0100
Temperature class T4
-20 °C Ta +50 °C
1 The year of manufacture is coded by the third capital letter of the serial number:
Y = 2007, Z = 2008, A = 2009, B = 2010, C = 2011, D = 2012, E = 2013, etc. Example: Serial No. ARCH-0054: the third letter is C, so the year of manufacture is 2011.
6 Dräger X-am 5100

3 Use

Use

3.1 Preparations for use

Before using the instrument for the first time, the
enclosed batteries or a charged T4 NiMH power pack (Order No. 83 18 704 / 83 22 244) must be inserted, see "Replacing the batteries / rechargeable batteries" on page 8.
The X-am-5100 is ready for operation.

3.1.1 Charging the rechargeable batteries

WARNING
!
Danger of explosion! Do not charge underground or in explosion-hazard areas!
The chargers are not designed in accordance with the regulations for firedamp and explosion protection.
Charge power packs type HBT 0000 or HBT 0100 with the appropriate Dräger charger. Charge single NiMH cells for battery holder ABT 0100 in accordance with the manufacturer's specifications. Ambient temperature during the charging procedure: 0 to +40 °C.
NOTICE
i
i
Even if the instrument is not in use, Dräger recommends that you store it in the charger (chargermodule X-am 1/2/5000, Order No. 83 18 639).
To maintain the lifetime of the batteries, charging is only
performed within a temperature range of 5 to 35 °C. Outside this temperature range, the charging is automatically interrupted and resumes automatically after the temperature is within the range again.
The charging time is typically 4 hours.A new NiMH power pack reaches its full capacity after three
complete charging/discharging cycles.
Never store the instrument for extended periods without
being connected to a power source (maximum of 2 months) because the internal buffer battery will drain.
Charging with the multiple charging station
A maximum of 20 instruments can be charged at the same
time on the power pack (Order No. 83 18 805) of the multiple charging station.
When attaching the charging modules, disconnect the
power pack from the mains supply!
CAUTION
!
Always connect or disconnect the charging modules individually and not in groups in order to prevent the charging station from becoming damaged. During transportation, the power pack and the charging modules should also always be handled individually and without inserted instruments.
2
1
34 5 5
Position the instrument on an even and level surface.
X-am 5100
0.0 0.0
HCl ppm
0
2
1
HCl ppm
X-am 5100
0
1
00733280.eps
1. Turn the slots of the interlock into a horizontal position by
using a screwdriver or coin.
2. Insert the projecting tongue (2) of the charging module
(doubles as power feed) until it engages.
3. Close the lock (1) with a quarter turn (slot is positioned
vertically).
4. Attach additional charging modules in the same way.
5. Connect the power pack to the mains.
The green "Mains" LED (1) lights.
6. Insert the switched-off instrument into the charging module.
LED indicator (5) on the charger:
Charge
Fault
Full
If a fault occurs:
Remove the instrument from the charging module and
insert it again.
If the fault still occurs, have the charging module repaired.
It takes approx. 4 hours to fully charge an empty
rechargeable battery.
CAUTION
!
A short circuit of the charging contacts in the charging modules, e.g., by metallic objects that have fallen in, will not result in damage to the charging station. It should, however, be avoided due to possible heating hazards and incorrect displays on the charging module.
In the event of a short circuit or if the power pack is
overloaded:
The red "Overload" LED (3) lights, and an audible alarm
sounds.
After the fault has been corrected, the alarm is switched
off automatically and the charging process is restarted.
In the event of a power failure, the instruments already
charged will be protected from discharging.
Dräger X-am 5100 7
Use
Charging with charger module and plug-in power pack or vehicle charging adapter
+
2
+
83 16 994 (100 ... 240 V) 83 15 635 (100 ... 240 V)
HCl ppm
X-am 5100
0.0
0
45 30 057
00933280.eps
When the power pack is used (Order No. 83 16 994) up to
5 instruments, or with power pack (Order No. 83 15 635), up to 2 instruments can be charged at the same time.
The power pack contained in the rechargeable battery and
charging set (Order No. 83 18 785) is suitable for charging one instrument.
When the vehicle charging adapter is used (Order
No. 45 30 057) it is recommended that each charging module is supplied separately.
The charging process is carried out analogue to charging
with the multiple charging station.

3.1.2 Replacing the batteries / rechargeable batteries

WARNING
!
Danger of explosion! Do not replace the batteries / rechargeable batteries in areas where there is a danger of explosion.
Do not throw used batteries into fire or try to open them by force. Dispose of the batteries in accordance with national regulations.
Batteries / rechargeable batteries are part of the Ex approval. Only the following types may be used:
Alkaline batteries – T4 – (not rechargeable)
Duracell Procell MN1500
Alkaline batteries – T3 – (not rechargeable)
Varta Powerone 4006Varta Powerone 4106Panasonic Powerline LR6
NiMH rechargeable batteries – T3 – (rechargeable)
GP 180AAHC (1800)
max. 40 °C ambient temperature.
3
1
00633280.eps
1. Switch off the instrument if necessary (see "Switching off the instrument" on page 8).
2. Loosen the screw (2.0 mm hexagon socket) on the power pack and remove the power pack.
3. Replace the alkaline batteries with new ones or the rechargeable NiMHy batteries with charged ones – ensure correct polarity.
4. Completely replace the T4 power pack (with sealed rechargeable batteries, Order No. 83 18 704 / 83 22 244).
5. Insert the power pack into the instrument and tighten the screw, the instrument switches on automatically.
After replacing the power pack T4, it is recommended that
a complete charging is carried out.
After the batteries have been replaced:
The settings and data are stored when the battery is
replaced. The sensors warm up again.

3.1.3 Switching on the instrument

1. Press and hold the key for approx. 3 seconds until the countdown » 3 . 2 . 1 « shown in the display has elapsed.
All the display segments, including the visual, audible
and vibration alarms, are activated for a short time.
The software version is displayed.The instrument performs a self test.The next sensor to be calibrated is shown, together
with the days remaining until the next calibration, e.g. » HCl ppm CAL 20 «.
The time until the bump test interval elapses is
displayed in days, e.g., » bt 123 «.
All alarm thresholds A1 and A2 as well as » « (TWA)
and » « (STEL)1, are displayed consecutively.
During the sensor warm-up period, the respective
display of the measured value flashes and the special symbol » « (for warning) is displayed. No alarms are issued during the sensor warm-up period.
2. Press the key to cancel the display of the activation
OK
sequence.

3.1.4 Switching off the instrument

1. Press and hold the key and key at the same time until the countdown » 3 . 2 . 1 « shown in the display has elapsed.
Before the instrument is switched off, the visual, audible
and vibration alarms are activated for a short time.
OK
OK
1
1 Only when activated in the instrument configuration. Delivery status: not
activated.
8 Dräger X-am 5100

3.1.5 Before entering the workplace

WARNING
!
Before safety-related measurements, check the calibration and recalibrate if necessary. A bump test must be carried out in accordance with the national regulations.
1. Switch on the instrument. The current measured values are shown in the display.
Observe any warning » « or fault messages » «.
The instrument can be operated normally. If the warning message does not disappear automatically during operation, the instrument must be serviced after the end of use.
The instrument is not ready to measure and requires maintenance.
2. Check that the gas inlet opening on the instrument is not covered.
Use
Dräger X-am 5100 9
Use

3.2 Configuration

3.2.1 Standard gas configuration

Alarm A1
1
Alarm A2
1
DrägerSensor Measuring range
XS EC HF/HCI [ppm]
2
XS EC H2O2 [ppm]
XS EC N2H4 [ppm]
1 Different settings can be selected to meet customer requirements on delivery. The current setting can be checked and changed with the Dräger CC Vision
software. A version of the CC-Vision software that can be used for Dräger X-am 5100 is available for download from the product page for the X-am 5100 at the following web address: www.draeger.com
2 Valid for HCl.
0 to 30 5 Yes No 10 No Yes
0 to 20 1 Yes No 2 No Yes
0 to 3 0.1 Yes No 0.2 No Ye s

3.2.2 Standard instrument configuration

1
threshold
1

3.2.3 Configuring the instrument

can be acknowledged
self-latching
threshold
can be acknowledged
To customise its configuration, the instrument must be
Dräger X-am 5100
Bump test mode Quick bump test
connected to a PC with the USB-DIRA adapter (Order No. 83 17 409). Dräger CC-Vision PC software is used to perform the configuration.
Fresh air adjustment On Life signs (visual only) On Disabling permitted / blocked for A2 Averaging time 15 minutes for STEL
8 hours for TWA
0
IR
X-am 5100
USB 1.1
self-latching
1 Different settings can be selected to meet customer requirements on
delivery. The current setting can be checked and changed with the Dräger CC Vision software.
00433280.eps
Changing the standard configuration
The installed Dräger CC Vision PC software is used for
configuration.
Observe the documentation and online help of the software.A version of the CC-Vision software that can be used for
Dräger X-am 5100 is available for download from the product page for the X-am 5100 at the following web address: www.draeger.com.
Reading the database and displaying it Graphically
The installed PC software Dräger GasVision is used for
reading and displaying the database.
Observe the documentation and online help of the software.Changing the configuration: See “Troubleshooting” on
page 19.
WARNING
!
After a basic initialization has been carried out with the Dräger CC Vision PC software, individual alarm settings may have been changed.
10 Dräger X-am 5100
Use

3.3 Performing the bump test

3.3.1 Manual testing without documentation of results in
1. Prepare a test gas source (e.g., cylinder, permeation oven);
2. Fit the calibration adapter
3. Connect the test gas
4. Vent the test gas into
5. Switch on the instrument.
6. Open the valve on the test gas source to let test gas flow
7. Recommendation: Wait until the instrument displays the
8. Close the valve on the test gas source.
If the concentration has now fallen under the A1 alarm
If the displays are outside of the above-mentioned ranges:
3.3.2
The "Quick bump test" or the "Extended bump test" is
Setting on delivery: Quick bump test.
the instrument memory
NOTICE
i
i
Sensor-specific features affecting the bump test are described in the respective sensor data sheets.
the volume flow must be 0.5 L/min and the gas concentration must be higher than the alarm threshold concentration to be tested.
(68 06 291) to the sensor cap.
source to the calibration adapter.
a fume cupboard or into
0
x-am 5100
the open air (with a hose connected to the second connector of the calibration adapter).
00533280.eps
WARNING
!
Never inhale the test gas. Danger to health! Observe the hazard warnings in the relevant Safety Data Sheets.
over the sensor.
test gas concentration with sufficient tolerance. However, wait at least until alarm threshold A1 or A2 has been exceeded.
If the alarm thresholds are exceeded, the instrument
displays the gas concentration in alternation with » A1 « or » A2 « depending on the test gas concentration.
threshold:
Acknowledge the alarm.
Calibrate the instrument, see "Calibrating the instrument"
on page 15.

Bump test with documentation of results in the instrument memory.

selected using the Dräger CC Vision PC software. The "Quick bump test" checks whether the gas concentration has exceeded the Alarm 1 threshold (with oxygen, the check is whether the concentration has fallen below the Alarm 1 threshold). The "Extended bump test" checks whether the gas concentration has exceeded the Alarm 1 threshold (with oxygen, the check is whether the concentration has fallen below the Alarm 1 threshold) and whether the gas concentration has reached the preset bump test concentration.
1. Prepare a test gas source (e.g., cylinder, permeation oven); the volume flow must be 0.5 L/min and the gas concentration must be higher than the alarm threshold concentration to be tested.
2. Fit the calibration adapter (68 06 291) to the sensor cap.
3. Connect the test gas source to the calibration adapter.
4. Vent the test gas into a fume cupboard or into
0
x-am 5100
the open air (with a hose connected to the second connector of the calibration adapter).
00533280.eps
WARNING
!
Never inhale the test gas. Danger to health! Observe the hazard warnings in the relevant Safety Data Sheets.
5. Switch on the instrument.
6. Call up the Quick Menu and select the bump test, see "Quick Menu" on page 13.
The current gas
concentration values and the special symbol » « (for bump test) flash.
7. Press the key to start the bump test.
OK
HCl ppm
8. Open the valve on the test gas source to let test gas flow over the sensor.
If gas concentration
exceeds the alarm thresholds A 1 the
10433280.eps
corresponding alarm will occur.
If a gas alarm (Quick
bump test) is triggered or the preset bump test concentration (Extended bump test) is reached within the specified time:
The display containing the
OK
HCl ppm
current gas concentration changes with the display » OK «.
The bump test that was
carried out is documented with the result and date in
10533280.eps
the instrument memory.
9. Close the valve on the test gas source.
If the concentration has now fallen below the A1 alarm
threshold, the instrument returns to the measuring mode.
If the set bump test concentration is not reached after a
sensor-specific time interval, an instrument error is generated.
Dräger X-am 5100 11
Use

3.4 During use

CAUTION
!
To ensure correct measurement operation, the gas inlet opening on the instrument must not be covered or dirty.
The sensor cap must not be twisted.
During operation, the measured values for the measured
gas are displayed.
If a measuring range is exceeded or a negative drift occurs,
the following displays are shown instead of the measured value display:
» «
» «
In the event of an alarm, the corresponding displays,
including the visual, audible and vibration alarms, are activated, see "Identifying Alarms" on page 12.
i
i
(Too high concentration) or (Negative drift).
NOTICE
Special states in which there is no measuring operation (quick menu, calibration menu, warm-up of sensors, password input) are indicated by a visual signal (slow flashing of the alarm LED
).
In the case of A2, a double tone is audible and the alarm
LED flashes twice.
After leaving the area, if the concentration is less than the alarm threshold A2:
Press the key. The alarm messages are switched off.
OK

3.5.3 STEL / TWA exposure alarm

WARNING
!
Leave the area immediately. After this alarm, the deployment of personnel is subject to the relevant national regulations.
The alarm is indicated by an intermittent alarm message:
Display » A2 « and » « (STEL) or » « (TWA) and measured value alternating:
The STEL and TWA alarm cannot be acknowledged
or cancelled.
Switch off the instrument. The values for the exposure
evaluation are deleted after the instrument is switched on again.

3.5.4 Battery pre-alarm

The alarm is indicated by an intermittent alarm message:
Flashing special symbol » « on the right side of the display:

3.5 Identifying Alarms

An alarm is displayed visually, audibly and through vibration in a specific pattern.

3.5.1 Concentration pre-alarm A1

The alarm is indicated by an intermittent alarm message:
Display » A1 « and measured value alternating.
The pre-alarm A1 is not self-latching and stops when the concentration has dropped below the alarm threshold A1.
In the case of A1 a single tone is audible and the alarm
LED flashes.
Acknowledging the pre-alarm:
Press the key. Only the audible alarm and the vibration
alarm are switched off.

3.5.2 Concentration main alarm A2

!
The alarm is indicated by an intermittent alarm message:
Display » A2 and measured value alternating.
OK
WARNING
Danger to life! Leave the area immediately. A main alarm is self-latching and cannot be acknowledged or cancelled.
Acknowledging the pre-alarm:
Press the key. Only the audible alarm and the vibration
OK
alarm are switched off.
The battery lasts for at least another 20 minutes after the
first battery pre-alarm.

3.5.5 Battery main alarm

The alarm is indicated by an intermittent alarm message:
Flashing special symbol » « on the right side of the display:
The battery main alarm cannot be acknowledged or cancelled:
The instrument is automatically switched off again after
10 seconds.
Before the instrument is switched off, the visual, audible
and vibration alarms are activated for a short time.

3.5.6 Instrument alarm

The alarm is indicated by an intermittent alarm message:
Special symbol » « displayed on the right side of the display:
The instrument is not ready for operation.For remedies, see "Troubleshooting" on page 19 to page 20.Commission maintenance personnel or DrägerService to
rectify the error.
12 Dräger X-am 5100
Menu functions

4 Menu functions

4.1 Calling the Info Mode

In measuring mode, press the key for approx. 3 seconds.
If any warning or fault messages are present, the
corresponding information or error codes are displayed (see "Troubleshooting" on page 19 to page 20).
Press the key successively for the next display.
The peak values and the exposure values TWA STEL
OK
1
are displayed.
Warning messages are displayed. Numerical codes of warning messages: see "Warning messages" on page 19.
OK
key
Fault messages are displayed. Numerical codes of fault messages: see "Fault messages" on page 20.
OK
key
The peak values = the maximum measured values are displayed.
OK
key
The average values of the exposures based on a shift of, e.g., 8 hours (TWA) are displayed
OK
key
The short-term values (STEL) = average values of the concentrations over the average value duration are displayed
OK
key
The instrument is in measuring mode again
If no key is pressed for 10 seconds, the instrument reverts
automatically to measuring mode.
OK
1
and

4.3 Quick Menu

4.3.1 Quick menu functions

Bump test, see "Performing the bump test" on page 11.
Fresh air adjustment, see "Performing a fresh air adjustment" on page 15.
Display and deletion of the peak values, see "Quick menu "Displaying and deleting peak values"" on page 13.

4.3.2 Calling the Quick Menu

The fresh air adjustment and the bump test are activated in the quick menu on delivery. The function for displaying and deleting peak values can additionally be activated with the Dräger CC Vision PC software.
1. In measuring mode, press the key three times.
If functions in the quick menu have been activated using
the Dräger CC-Vision PC software, you can select these functions using the key. If no functions have been activated in the quick menu, the instrument remains in measuring mode.
2. You can select the activated functions of the quick menu by pressing the key.
Press the key to call the selected function.Press the key to cancel the active function and to
switch to measuring mode.
If no key is pressed for 60 seconds, the instrument
reverts automatically to measuring mode.

4.3.3 Quick menu "Displaying and deleting peak values"

After the function has been
selected, the current peak values are displayed; the special peak values symbol appears in the display at the same time.
OK
12
HCl ppm

4.2 Calling the Info-Off Mode

When the instrument is in a deactivated state, press the
key.
The name of the gas, measuring unit and measuring
range limit value are displayed.
Pressing the key again exits the Info Off mode (or via
OK
timeout).
1. The peak values can be deleted by pressing the
OK
key for 5 seconds.
The next display appears.
2. Press the key to end
OK
the function.
HCl ppm
1 Only when activated in the instrument configuration. Delivery status:
not activated.
Dräger X-am 5100 13
10133280.eps
0
10233280.eps
Menu functions

4.4 Adjustment menu

4.4.1 Functions of the adjustment menu

Fresh air adjustment, see "Performing a fresh air adjustment" on page 15
1-button sensitivity adjustment
Standard sensitivity adjustment

4.4.2 Calling the adjustment menu

The adjustment menu can only be accessed by entering
a password. Password on delivery: » 001 «
The default password on delivery can be changed using
the PC software Dräger CCVision.
1. In measuring mode, press the key for at least 5 seconds.
The function for entering the password is selected.The special symbol » « (for the "Enter password"
function) is displayed.
The display shows » 000 «, with the first digit flashing.
2. Use the key to set the flashing digit.
3. Press the key, the second digit starts flashing.
4. Use the key to set the flashing digit.
5. Press the key, the third digit starts flashing.
6. Use the key to set the flashing digit.
7. Press the key to confirm the password once it has been set completely.
OK
000
OK
OK
10333280.eps
8. The adjustment menu functions can now be selected by pressing the key.
Press the key to call the selected function.Press the key to cancel the active function.If no key is pressed for 10 minutes, the instrument
OK
reverts automatically to measuring mode.
14 Dräger X-am 5100
Calibrating the instrument

5 Calibrating the instrument

WARNING
!
Always adjust the zero point first, before the sensitivity. Otherwise the adjustment will be incorrect!
Adjustment may not be possible due to instrument and
channel errors.
Allow the sensor to warm up before the adjustment.Warm-up time: see Instructions for Use/data sheet for the
installed DrägerSensor.

5.1 Adjustment interval:

Observe the relevant specifications in the Instructions for
Use/data sheet for the DrägerSensor installed.
For critical applications according to EN 60079-29-2
EN 45544-4
Improvement of zero point accuracy – perform a fresh air
2
and national regulations.
adjustment, see "Performing a fresh air adjustment" on page 15.

5.2 Performing a fresh air adjustment

A fresh air adjustment can be performed to improve the zero point accuracy.
Calibrate the instrument to fresh air, free of measured
gases or other interfering gases.
The zero point of the sensor is set to 0 during the fresh
air calibration.
Sensors that are faulty or not warmed-up will prevent the
adjustment.
In the case of sensors which are in the warm-up phase,
the message » 159 « is displayed with the special symbol » « (for warning message).
In the case of a sensor or instrument error, the
message » 109 « is displayed with the special symbol » « (for a fault message).
The message is cleared after 5 seconds and the
function is available again in the menu.
1. Switch on the instrument.
2. Depending on instrument configuration:
Call up the Quick Menu and select the fresh air
adjustment function » «, see "Quick Menu" on page 13.
or
Call up the Adjustment menu and select the fresh air
adjustment function » «, see "Adjustment menu" on page 14.
1
or
The measured values
flash.
When the measured
values have stabilized:
3. Press the key to
OK
carry out the fresh air adjustment.
The display containing the
current gas concentration changes with the display » OK «.
4. Press the key to exit
OK
the calibration or wait for approx. 5 seconds.
If a fault has occurred during the fresh air adjustment:
The fault message » «
appears and » « is displayed for the respective sensor instead of the measured value.
In this case, repeat the
fresh air adjustment.
If necessary, replace the
sensor, see "Replacing the Sensors" on page 18.
HCl ppm
0.0
OK
HCl ppm
HCl ppm
--
10733280.eps
10833280.eps
10933280.eps
1 EN 60079-29-2 – Guidelines for selection, installation, use and maintenance
of instruments for the detection and measurement of flammable gases and oxygen.
2 EN 45544-4 – Electrical instruments for the direct detection and direct
concentration measurement of toxic gases and vapors – Part 4: Guidelines for selection, installation, use and maintenance.
Dräger X-am 5100 15
Calibrating the instrument

5.3 Performing a sensitivity adjustment

The sensitivity adjustment can optionally be performed by the standard or by the 1-button method.

5.3.1 Standard sensitivity adjustment

CAUTION
!
Never inhale the test gas. Danger to health! Observe the hazard warnings of the relevant Safety Data Sheets.
NOTICE
i
i
To minimise adsorption effects, keep the length of the hose as short as possible (maximum hose length: 1 m). Dräger recommends using PTFE hoses.
During the sensitivity calibration, the sensitivity of the sensor is set to the value of the test gas.
1. Fit the calibration adapter (68 06 291) to the sensor cap.
2. Connect the test gas source to the calibration adapter.
3. Vent the test gas into a fume cupboard or into
0
the open air (with a hose connected to the second connector of the calibration adapter).
4. Switch on the instrument.
5. Call up the adjustment menu, enter the password and select the standard sensitivuty adjustment » «, see Chapter 4.4.2 on page 14.
6. Press the key to start the sensitivity calibration.
The name of the gas
OK
HCl ppm
flashes.
7. Confirm with the key.
The calibration gas
OK
concentration is displayed.
8. Press the key to
OK
confirm the calibration gas concentration or use the key to change the calibration gas concentration and complete the process by pressing the key.
The set calibration gas concentration flashes.
9. Press the key to confirm the set value.
OK
10. Open the valve on the test gas source to let test gas flow over the sensor.
The currently displayed measured values start to flash.The displayed measured values change to the values
according to the gas supplied.
x-am 5100
00533280.eps
0.0
11033280.eps
OK
When the calibration is complete:
The display of the
current gas concentration alternates with the » OK « display.
12. Press the key or wait for 5 seconds to quit the calibration.
The instrument
OK
OK
HCl ppm
changes to the measuring mode.
13. Close the valve on the test gas source.
11133280.eps
If a fault has occurred during the sensitivity calibration:
The fault message » «
appears and » « is displayed for the respective sensor instead of the measured value.
In this case, repeat the
sensitivity adjustment.
If necessary, change the
HCl ppm
--
sensor, see "Changing sensors" on page 18.
11233280.eps
When the measured value is stabilized:
11. Press the key to carry out the calibration.
OK
16 Dräger X-am 5100
Calibrating the instrument

5.3.2 1-button sensitivity adjustment

CAUTION
!
Never inhale the test gas. Danger to health! Observe the hazard warnings of the relevant Safety Data Sheets.
NOTICE
i
i
To minimise adsorption effects, keep the length of the hose as short as possible (maximum hose length: 1 m). Dräger recommends using PTFE hoses.
During the 1-button sensitivity adjustment, the sensitivity of the sensor is set to the value of the test gas.
Using the Dräger CC Vision PC software, the preset
concentration values of the test gas cylinder used must be changed in the instrument to the target values of the mixed gas used.
1. Fit the calibration adapter (68 06 291) to the sensor cap.
2. Connect the test gas source to the calibration adapter.
3. Vent the test gas into a fume cupboard or into
0
x-am 5100
the open air (with a hose connected to the second connector of the calibration adapter).
00533280.eps
4. Switch on the instrument.
5. Call the calibration menu, enter the password and select the 1-button sensitivity adjustment function » «, see Chapter 4.4.2 on page 14.
6. Press the key to start the sensitivity calibration.
7. Open the valve on the test
OK
HCl ppm
0.0
gas source to let test gas flow over the sensor.
The currently displayed
measured values start to flash.
The flashing stops after
a static measured value has been reached.
The adjustment is
now carried out
11333280.eps
automatically.
The displayed measured values change to the values
according to the gas supplied.
The automatic stability monitoring can be terminated by
pressing the key. A calibration then takes place
OK
immediately.
When the calibration is complete and the displayed measured values have stabilized:
The display of the current
gas concentration alternates with the » OK « display.
8. Press the key or wait for 5 seconds to quit the calibration.
The instrument
OK
OK
HCl ppm
changes to the measuring mode.
9. Close the valve on the test gas source.
If a fault has occurred during the 1-button sensitivity adjustment:
The fault message » «
appears and » « is displayed for the respective sensor instead of the measured value.
In this case, repeat the
sensitivity adjustment.
If necessary, change the
HCl ppm
--
sensor, see "Changing sensors" on page 18.
11433280.eps
11533280.eps
Dräger X-am 5100 17
Replacing the Sensors

6 Replacing the Sensors

1. Connect the instrument to the PC with the USB-DIRA adapter (Order No. 83 17 409).
0
X-am 5100
IR
USB 1.1
00433280.eps
2. Deactivate the slot using the CC-Vision PC software.
3. 4 screws on the lower shell should now be unfastened.
4. Carefully release the upper shell upwards from the lower shell. Do not tilt the upper shell when doing this.
00133280.eps
NOTICE
i
i
When the upper shell is being released, it is possible that the DrägerSensor will be pulled out of its mount at the same time and remain in the upper shell.
5. Follow the remaining instructions in the CC Vision PC software.
18 Dräger X-am 5100
Troubleshooting

7 Troubleshooting

Fault Cause Remedy
Not possible to switch on the instrument
Not possible to switch off the instrument
Display » – – « Measuring range calibrated incorrectly Recalibrate the measuring range,
To display the numerical codes of the warning and fault messages in the info mode, see "Calling the Info Mode" on page 13.

7.1 Warning messages

Discharge the power pack Charge the power pack, see "Charging the
Discharge the alkaline batteries Insert new alkaline batteries, see "Replacing
The instrument is not set to measuring mode
The instrument is configured to "Disable prohibited"
Electronics or sensors defective Must be repaired by DrägerService.
rechargeable batteries" on page 7.
the batteries / rechargeable batteries" on page 8.
Select measuring mode.
Configure the instrument to "Disable allowed" with Dräger CC Vision.
see "Calibrating the instrument" on page 15.
Special symbol » « and displayed numerical code:
152
153
154
155
159
351
352
353
354
355
356
Cause Remedy
Customer's service life counter about to elapse
Database 90 % full Read the database soon and then clear
Database full Read the database and clear memory.
Interval for the bump test has elapsed Carry out the bump test, see "Performing the
Adjustment not possible. The menu function cannot be carried out because of a message which is preventing the function (e.g., sensors in warm-up phase).
DrägerSensor XS EC1 in the warm-up phase
DrägerSensor XS EC1 in the warm-up phase
EC1 concentration has drifted into the negative range
The temperature is too high Operate the instrument within the allowed
The temperature is too low Operate the instrument within the allowed
The calibration interval for DrägerSensor XS EC1 has elapsed
Reset the service life counter using Dräger CC Vision.
the memory.
bump test" on page 11.
Identify the message code via the info menu and switch off, if necessary.
Wait until warm-up time is complete.
Wait until warm-up time is complete.
Perform a fresh air calibration, see "Performing a fresh air adjustment" on page 15.
temperature range.
temperature range.
Perform a sensitivity adjustment for DrägerSensor XS EC1, see "Calibrating the instrument" on page 15.
Dräger X-am 5100 19
Troubleshooting

7.2 Fault messages

Special symbol » « and displayed numerical code:
102 The customer's service life counter
103 The instrument is defective The instrument must be repaired by
104 Check sum error program code The instrument must be repaired by
105 The bump test interval has elapsed Carry out the bump test, see "Calibrating the
106 The calibration interval has elapsed
107 Bump test error Perform bump test, see "Performing the
108 The instrument is defective The instrument must be repaired by
109 The menu function cannot be carried out
301 No valid zero point adjustment of the
302 No valid sensitivity adjustment of the
303 The measured value of DrägerSensor XS
304 Dräger Sensor XS EC1 not inserted or
305 Error during bump test of Dräger Sensor
326 Error during accelerated warm-up of
Cause Remedy
has elapsed
(at least 1 calibration interval has elapsed)
because of an error.
Dräger Sensor XS EC1
Dräger Sensor XS EC1
EC 1 is in the negative range
defective
XS EC1
Dräger Sensor XS EC1
Reset the service life counter using Dräger CC Vision.
DrägerService.
DrägerService.
instrument" on page 15.
Perform a sensitivity adjustment, see "Calibrating the instrument" on page 15.
bump test" on page 11, or snsitivity adjustment, see "Calibrating the instrument" on page 15.
DrägerService.
Identify the error code via the info menu and switch off, if necessary.
Perform a fresh air calibration, see "Performing a fresh air adjustment" on page 15.
Perform sensitivity adjustment or fresh air adjustment, see "Calibrating the instrument" on page 15.
Perform a fresh air calibration, see "Performing a fresh air adjustment" on page 15.
Check DrägerSensor XS EC1, see "Replacing the Sensors" on page 18.
Repeat bump test. If necessary, adjust or replace the Dräger Sensor XS EC1, see "Replacing the Sensors" on page 18.
Remove the power pack and insert it again or replace the sensor. Sensor must not be exposed to gas during the first 5 minutes.
20 Dräger X-am 5100

8 Maintenance

8.1 Maintenance intervals

The instrument should be inspected and maintained by suitably qualified persons annually (consult: EN 45544-4 – Electrical apparatus used for the direct detection and direct concentration measurement of toxic gases and vapours - Part 4: Guide for selection, installation, use and maintenance and national regulations). Calibration intervals: see Instructions for Use for the respective DrägerSensors.
Depending on instrument configuration:
Replace the alkaline batteries, see "Replacing the
batteries / rechargeable batteries" on page 8 or charge the battery see "Charging the rechargeable batteries" on page 7 – after each use, at the latest after the battery alarm has been triggered or after 2 weeks.
Calibrate the instrument – see "Calibrating the instrument"
on page 15.
In regular intervals, according to the sensors used and
the operating conditions. For sensor-specific calibration data, refer to the Instructions for Use/data sheets for the Dräger sensors used
Before you carry out safety-related relevant
measurements, the zero point and sensitivity of the instruments should be tested in accordance with national regulations.
Inspection by suitably qualified persons – every year.
The inspection intervals must be established in each
individual case and shortened if necessary, depending on safety related considerations, engineering conditions and the technical requirements of the equipment.
Dräger recommend that a service agreement be
concluded with Dräger Service and that repairs also be carried out by them.
Replace the sensors, see "Replacing the Sensors" on
page 18 – if necessary, when it is no longer possible to adjust them.
1
.
Maintenance

8.2 Cleaning

CAUTION
!
Abrasive cleaning implements (brushes etc.), cleaning agents and cleaning solvents can destroy the dust and water filters.
The instrument does not need any special care.If the instrument is very dirty, clean it with a cloth.
1 Instructions for Use/data sheets for the utilized sensors and the PC software
CC-Vision for Dräger X-am 5100 can be downloaded on the product page of the X-am 5100 at the following Internet address: www.draeger.com. See also the enclosed Instructions for Use and data sheets for the sensors used.
Dräger X-am 5100 21
Disposal

9 Disposal

9.1 Disposal of electrical and electronic
appliances
This product must not be disposed of as municipal waste. This is indicated by the adjacent icon. You can return this product to Dräger free of charge. For information please contact the national sales organisations and Dräger.
Batteries and rechargeable batteries must not be disposed of as municipal waste. This is indicated by the adjacent icon. Collect and dispose of batteries and rechargeable batteries at battery collection centres, in accordance with applicable regulations.

9.2 Electrochemical sensors

WARNING
!
Acid burn risk! Do not throw them into fires or use force to open them.
As for batteries, only dispose of as special waste in accordance with local waste disposal regulations. Further information can be obtained from the relevant local authority and from appropriate waste disposal companies.
22 Dräger X-am 5100

10 Technical Data

Ambient conditions:
During operation and storage -20 to +50 °C
Instrument data
Protection class IP 54 for instruments with sensors
Alarm volume Typically 90 dB (A) in 30 cm distance
Operation time (with 24 hours' use per day, 1 minute alarm per day)
Alkaline battery (Order No.
83 20 240 / 83 22 239)
NiMH rechargeable battery (Order No. 83 18 704)
NiMH HC rechargeable battery (Order No. 83 22 244)
Dimensions approx. 130 mm x 48 mm x 61 mm (H x W x D)
Weight approx. 220 g to 250 g
CE markings Electromagnetic compatibility (Directive 2004/108/EC)
Approvals: See “Description” on page 5.
Sensor data: See data sheets for the sensors used.
(-20 to +40 °C for NiMH Type 180AAHC single cells and T3 alkaline batteries) 700 to 1300 hPa 10 to 90 % (short-term up to 95 %) relative humidity
Typically 180 hours
Typically 150 hours
Typically 180 hours
Explosion protection (Directive 94/9/EC)
Technical Data
Dräger X-am 5100 23
Ordering list

11 Ordering list

Name and Description
Order No.
Dräger X-am 5100 83 22 750
Power supply units:
NiMH power pack T4 83 18 704
NiMH power pack HC T4 83 22 244
Alkaline power pack T3/T4 (without alkaline batteries)
Alkaline batteries T4 (2 ea.)
1
83 22 237
83 20 240
for alkaline power pack
Alkaline batteries T3 (2 ea.)
83 22 239
for alkaline power pack
Rechargeable battery and charging set
83 18 785 (includes NiMH T4 power pack, charging module for Dräger X-am 1/2/5000 and plug-in power pack)
Chargers:
Charging module for Dräger X-am 1/2/5000 83 18 639
Power pack with connecting cord (worldwide)
83 15 805 for a maximum of 20 Dräger X-am 1/2/5000 charging modules)
Plug-in power pack (worldwide)
83 16 994 for a maximum of 5 Dräger X-am 1/2/5000 charging modules)
Plug-in power pack (worldwide)
83 15 635 for a maximum of 2 Dräger X-am 1/2/5000 charging modules)
Vehicle connecting line 12 V/24 V
45 30 057 for Dräger X-am 1/2/5000 charging module
Vehicle installation set
83 18 779 for 1 Dräger X-am 1/2/5000 charging module
Additional strap for vehicle mounting bracket 83 18 788
1 The T3/T4 alkaline power pack (Order No. 83 18 703) is not a subject of the
BVS 08 ATEX G 002 X and PFG 08 G 001 certificates.
Name and Description
Order No.
Accessories for measured value acquisition and configuration:
Dräger GasVision 83 14 034
Dräger CC Vision 64 08 515
PC communications set for 1 Dräger
83 18 761 X-am 1/2/5000 with USB port and Dräger CC Vision
PC communications set for 2 Dräger
83 18 762 X-am 1/2/5000 with USB port, Dräger CC Vision and bar code reader
USB DIRA with USB cable (USB infrared
83 17 409 adaptor for communication Dräger X-am 1/2/5000 – PC)
Calibration/adjustment accessories:
Test gas cylinder 10 ppm HCl 68 12 107
Test gas cylinder 10 ppm SO
2
68 10 645
On demand controller 83 16 556
Standard controller 68 10 397
Calibration adapter 68 06 291
Sensors:
DrägerSensor XS EC HF/HCl 68 09 140
DrägerSensor XS EC H2O
DrägerSensor XS EC N2H
2
4
68 09 170
68 09 190
24 Dräger X-am 5100

12 Declaration of Conformity

Declaration of Conformity
Dräger X-am 5100 25
Dräger Safety AG & Co. KGaA
Revalstraße 1 23560 Lübeck, Germany Tel +49 451 882 0 Fax +49 451 882 20 80 www.draeger.com
90 33 280 - TH 4638.230 © Dräger Safety AG & Co. KGaA Edition 04 - April 2014 (Edition 01 - July 2011) Subject to alteration
Loading...