Dräger Savina 300 Technical Manual

4.3 (3)

Technical Documentation IPM

Savina 300

Intensive Care Ventilator

Warning

All servicing and/or test procedures on the device require detailed knowledge of this documentation. Use of the device requires detailed knowledge and observance of the relevant Instructions for Use.

– This page has been intentionally left blank –

<![if ! IE]>

<![endif]>No.1309-0000001227_Publication

2

Savina 300

 

Table of contents

Table of contents

 

General

5

General notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . 6

Function descriptions

9

Block diagram and functional principle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . 10

Electronic assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . 12

Control panel with TFT colour display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . 18

Pneumatic assembly. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . 20

Maintenance instructions

31

Disassembling/assembling the device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . 32

Replacing the microfilter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . 38

Replacing the dust filters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . 40

Replacing the O2 sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . 42

Replacing the diaphragm of the expiratory valve . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . 47

Replacing the internal batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . 50

Replacing the external batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . 54

Replacing the consumables in the O2 gas inlet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . 65

Annex

71

<![if ! IE]>

<![endif]>No.1309_0000001227_Publication

Savina 300

3

– This page has been intentionally left blank –

<![if ! IE]>

<![endif]>No.1309-0000001227_Publication

4

Savina 300

General

This chapter contains general notes and definitions that are important for the use of this documentation.

General notes ....................................................................................................................................

6

<![if ! IE]>

<![endif]>No.1309_0000001227_Publication

Savina 300

5

General

General notes

General notes

Notes on use

Copyright and other protected rights

Read through the following notes thoroughly before applying this documentation.

Dräger reserves the right to make changes to the device and/or to this documentation without prior notice. This documentation is intended solely as an information resource for experts.

The content of this documentation, in particular its design, text, software, technical drawings, configurations, graphics, images, data and their selection and its composition and any amendments to it ("content") are protected by copyright. The content must not (in whole or in part) be modified, copied, distributed, reproduced, republished, displayed, transmitted or sold without the prior written consent of Dräger.

Definitions

WARNING!

A WARNING statement provides important information about a potentially hazardous situation which, if not avoided, could result in death or serious injury.

CAUTION

A CAUTION statement provides important information about a potentially hazardous situation which, if not avoided, may result in minor or moderate injury to the user or patient or in damage to the medical device or other property.

NOTE

A NOTE provides additional information intended to avoid inconvenience during operation and/or servicing.

Inspection

= Identification of actual condition

Maintenance

= Measures to maintain the speci-

 

fied condition

Repair

= Measures to restore specified

 

condition

Servicing

= Inspection, maintenance, repair

<![if ! IE]>

<![endif]>No.1309_0000001227_Publication

6

Savina 300

<![if ! IE]>

<![endif]>No.1309_0000001227_Publication

General

General notes

General safety precautions Read through each section thoroughly before beginning servicing. Always use the correct tools and the specified test equipment. Otherwise the device may not work correctly or may be damaged.

WARNING!

Improper servicing

If the medical product is not properly serviced, the safety of the patient and/or the operator may be put at risk.

Have the medical product checked and maintained on a regular basis by appropriately qualified experts, otherwise the proper functioning of the device may be compromised.

Have repairs to the medical product carried out only by personnel who have undergone product-specific Dräger training.

NOTE

Dräger recommends entering into a service contract with DrägerService and having all repairs likewise carried out by DrägerService.

WARNING!

Replacement parts not certified by Dräger standards

Dräger cannot guarantee or confirm the operational safety of third-party replacement parts used on the device.

Use only replacement parts certified to Dräger standards for servicing of the device, otherwise the proper functioning of the device may be compromised.

Pay attention to the „Servicing“ section of the Instructions for Use.

WARNING!

Non-conforming test values

If test values do not conform to specifications, the safety of the patient may be put at risk.

Do not put the device into operation if test values do not conform to specifications.

Contact your local service organization.

WARNING!

Impermissible modifications to the device

If impermissible modifications are made to the device, the safety of the patient may be put at risk.

Do not modify the device without Dräger's permission.

WARNING!

Risk of infection

The device may transmit pathogens following use on the patient.

Before carrying out any servicing, ensure that the device and its components have been handed over by the user cleaned and disinfected.

Service only cleaned and disinfected devices and device components.

Savina 300

7

General

General notes

WARNING!

Risk to patients

Ensure that no patient is connected to the device before starting maintenance or repair work.

NOTE

Where reference is made to legislation, regulations and standards, in respect of devices used and serviced in Germany they are based on the laws of Germany. Users and technicians in other countries must comply with their national laws and/or international standards.

<![if ! IE]>

<![endif]>No.1309_0000001227_Publication

8

Savina 300

Function descriptions

 

This chapter contains descriptions of the device's technical functions.

Block diagram and functional principle

.............................................................................................. 10

Electronic assembly ...........................................................................................................................

12

Control panel with TFT colour display................................................................................................

18

Pneumatic assembly..........................................................................................................................

20

<![if ! IE]>

<![endif]>No.1309_0000001227_Publication

Savina 300

9

Dräger Savina 300 Technical Manual

Function descriptions

Block diagram and functional principle

Block diagram and functional principle

Introduction

The following text sets out the block diagram and the functional principle of

 

the Savina 300.

Block diagram

Fig. 1 Block diagram Savina 300

<![if ! IE]>

<![endif]>No.1309_0000001227_Publication

10

Savina 300

Function descriptions

Block diagram and functional principle

Functional principle

The blower generates the necessary compressed air for ventilation. A controllable valve (non-return valve) is switched in parallel with the blower to regulate the pre-set ventilation parameter. The non-return valve opens or closes according to the pre-set ventilation parameters.To increase the oxygen concentration in the ventilation gas, an external oxygen source can be connected to the device. Sensors, the electronics and the valve block meter the oxygen concentration.

Main components

The device consists of the following main components:

Elektronic assembly

Control unit

Pneumatic assembly

<![if ! IE]>

<![endif]>No.1309_0000001227_Publication

Savina 300

11

Function descriptions

Electronic assembly

Electronic assembly

Introduction

The electronic assembly contains the following subassemblies:

 

 

 

Power supply unit

 

 

 

Central Control Board

 

 

 

Motor actuator

 

 

 

O2 Valve PCB

 

 

 

O2 Diaphragm PCB

 

 

 

Fan

 

 

 

 

 

 

 

Power supply unit

The power supply unit delivers the supply voltages for the device. The input

 

 

 

voltage range is 100 V to 240 V AC and 50 Hz to 60 Hz. The power supply

 

 

 

unit can also be operated with an external rechargeable battery (12 V or

 

 

 

24 V).

 

 

 

 

The connection to the alternating voltage is made by a power cable. The con-

 

 

 

nection for the external rechargeable battery is made by an encoded connec-

 

 

 

tor.

 

 

 

 

 

 

The power supply unit actuates the „Mains power“, „External battery or on-

 

 

 

board power supply“ and „Internal battery“ LEDs. The LEDs are mounted on

 

 

 

the membrane keypad of the control panel and indicate the respective operat-

 

 

 

ing status.

 

 

 

 

The device includes two internal rechargeable batteries (2 x 12 V) which

 

 

 

enable uninterrupted operation in the event of a mains power failure. The

 

 

 

internal rechargeable batteries supply the O2 sensors with power even when

 

 

 

the device is switched off. When the device is switched on the valid O2 val-

 

 

 

ues are present.

 

 

 

 

The power supply unit generates the following supply voltages:

 

 

 

+5 V

 

 

 

 

-15 V

 

 

 

 

+15 V

 

 

 

 

+24 V

 

 

 

 

+48 V

 

 

 

 

The output voltages are short-circuit-proof and stable at no-load.

 

 

 

The output voltages are generated according to the following priority, depen-

 

 

 

dent on the input voltages:

 

 

 

 

 

 

 

Input voltage

 

 

Priority

Action

 

 

 

 

 

 

 

 

 

 

 

 

 

AC voltage

 

 

1

Charge external and internal batteries, and maintain the

 

 

 

 

 

 

charge.

 

 

 

 

 

 

External battery/batteries

 

2

Charge internal batteries, and maintain the charge.

 

 

 

 

 

 

 

Internal batteries

 

 

3

-

 

 

 

 

 

 

 

<![if ! IE]>

<![endif]>No.1309_0000001227_Publication

12

Savina 300

Function descriptions

Electronic assembly

The fan cools the power pack.

Central Control Board

The Central Control Board is the device's central „control and monitoring unit“. It has three separate processor systems (master processor, front processor and display processor).

Three processor systems save the changeable, non-volatile data to EEPROMs.

<![if ! IE]>

<![endif]>No.1309_0000001227_Publication

Fig. 2 Block diagram of the Central Control Board

Master processor

The master processor has the following tasks:

Activation of the actuators (valves, valve block, blower)

Reading of the measured values from the sensors

Control of the ventilation

Monitoring of the front processor

The master processor data are cached in the RAM.

The EEPROM of the master processor system stores the calibration data of the sensors and set values such as volume.

Savina 300

13

Function descriptions

Electronic assembly

Front processor

The front processor has the following tasks:

Interface with the display processor

Monitoring of the master processor and display processor

Monitoring of the input logic and the data e.g. the pixel sum

The RAM caches data from the front processor.

The EEPROM with socket of the front processor stores safety parameters, set values such as volume, software options and the operating hours.

Display processor

The display processor has the following tasks:

Presentation of curves and parameters on the display, LEDs

Input and operator data (keyboard, rotary encoder, touchscreen)

Collation of data such as the pixel sum.

The EEPROM of the display processor stores user-specific display settings, such as measured value positions.

Flash EEPROMs

The device's operating system program (software) is stored in Flash

EEPROMs (rewritable memory modules).

Real-time clock

The real-time clock generates the time and date information. The real-time clock has an internal battery.

The Central Control Board incorporates the following functions:

Processing of the signals from the sensors (O2, flow, pressure, temperature)

Control of the blower and valves

Monitoring of the unit functions and the supply voltages

Actuation of the displays

Keypad interpretation

Provision of the internal and external interfaces

Set values such as the volume

Mains power failure alarm

If the mains power fails, the device switches to the internal batteries which supply the device with power. If the internal batteries are discharged and the device has no external batteries, an acoustic alarm is sounded with a GoldCap and a horn.

<![if ! IE]>

<![endif]>No.1309_0000001227_Publication

14

Savina 300

Function descriptions

Electronic assembly

Temperature hybrid

Motor actuator

The temperature hybrid reads the data of the AWT01 sensor and converts these into digital signals. An electrical isolation occurs in this process.

Internal temperature measurement

The internal temperature measurement is a safety function. In the event of overheating (if the internal temperature of the device is too high), an alarm is sounded.

Fan actuation and fans

The fan rotates quickly after the device has been switched on to remove any residual oxygen from the device. The rotation of the fan is detected and controlled in three stages (slow, medium and fast).

Three temperature sensors control the fan.

CAN

The CAN interface is a fast, serial interface. Via the CAN interface the control unit can communicate with the electronic and pneumatic assemblies. The transmission rate is 800 kbit/s.

The motor actuator controls the blower motor. The motor actuator is located in a self-contained housing. The supply voltage for the motor actuator is +48 V and is protected by a fuse (6.3AT).

The input voltage range of the motor actuator is 12 to 52.5 V. The rotation speed is set by the Central Control Board. The control voltage for the rotation speed is 0 V to +5.00 V, corresponding to a rotation speed of 0 to 12,000 rpm. The rotation speed range is 4,000 to 12,000 rpm.

The motor actuator acquires the „actual speed signal“ and forwards it to the Central Control Board. The „actual value signal“ is 6 pulses per rotation. In the event of discrepancies in the rotation speed the Central Control Board adjusts the speed according to the deviation.

<![if ! IE]>

<![endif]>No.1309_0000001227_Publication

Fig. 3 Block diagram of the motor actuator

Savina 300

15

Function descriptions

Electronic assembly

O2 Valve PCB

The O2 Valve PCB holds the pressure sensors (absolute pressure S6 and S7 and O2 supply pressure S5), the actuator for the O2 calibrating valve and the nebulizer valve and the actuator for the valve block.

The signals of the pressure sensors are amplified and routed to the Central Control Board. The supply voltage (+5 V) for the pressure sensors is generated by the O2 Valve PCB.

The valve block valves, the O2 calibrating valve and the nebulizer valve can be operated separately by an electronic switch. They are actuated by the Central Control Board.

O2 Diaphragm PCB

Fig. 4 Block diagram of the O2 Valve PCB

The O2 Diaphragm PCB amplifies the signals from the O2 sensors and measures the temperature of the O2 sensors and of the respiratory gas in the inspiration block. The temperature of the O2 sensors is required to compensate for the temperature-sensitive O2 measurements. The EEPROM on the Central Control Board stores the calibration data of the sensors. The reference voltage for the O2 sensors is generated from the voltage of the rechargeable batteries.

The supply voltages for the O2 Diaphragm PCB are +5 V and +15 V.

Fig. 5 O2 Diaphragm PCB block diagram

<![if ! IE]>

<![endif]>No.1309_0000001227_Publication

16

Savina 300

 

Function descriptions

 

Electronic assembly

 

 

Fan

The fan takes in ambient air through the cooler and cools the blower motor.

 

The air flow removes excess oxygen from the device.

 

The supply voltage for the fan is +24 V. The Central Control Board regulates

 

the speed of the fan.

Fig. 6

Fan control

 

 

 

 

Item

Designation

 

 

 

1

 

Supply voltage

 

 

 

2

 

Target speed

 

 

 

3

 

Speed signal

 

 

 

4

 

Sensor supply voltage

 

 

 

5

 

Ground

 

 

 

<![if ! IE]>

<![endif]>No.1309_0000001227_Publication

Savina 300

17

Function descriptions

Control panel with TFT colour display

Control panel with TFT colour display

Control unit

The control panel is the interface between the unit and the user. The control panel is used to enter and display the ventilation parameters.

Membrane keypad

TFT colour display

Rotary encoder

Fig. 7 Block diagram of the control panel with TFT colour display

The keypad features the control keys and the associated LEDs.

The 12.1” TFT colour display comprises the actual display and the backlight converter. The TFT colour display has a resolution of 800 x 600 pixels.

The backlight converter generates a high voltage for the display backlighting.

The rotary knob is used to set and acknowledge the ventilation parameters. The shaft encoder transmits square signals to the processor system as it rotates, and the signals are then evaluated by the Central Control Board. The voltage supply is +5 V.

<![if ! IE]>

<![endif]>No.1309_0000001227_Publication

18

Savina 300

Function descriptions

Control panel with TFT colour display

Touch-panel

Analog-resistive touch-screen

Fig. 8 Analog-resistive touch-screen

Analog systems consist of two opposing conductive indium tin oxide (ITO) layers (x and y layer) which are actuated with a constant direct voltage. Indium tin oxide is a transparent semiconducting material.

Between the two ITO layers there are a large number of small, barely visible so-called spacer dots which ensure the two layers are kept separate.

In 8-wire systems the touch-screen has eight wires routed to the controller – four for each axis.

When the touch-screen is touched at a certain point where the two ITO layers are located an electrical contact is produced. The resistance of this contact results in a different voltage at each point. The change in voltage is then used to define the x and y coordinates.

The Central Control Board controls the communication between the processor system and the touch-screen. The correct position is determined with the aid of the relevant software drivers. The analog touch-screen works very precisely, and provides a high resolution.

<![if ! IE]>

<![endif]>No.1309_0000001227_Publication

Savina 300

19

Function descriptions

Pneumatic assembly

Pneumatic assembly

Functional diagram

Fig. 9

Functional diagram

 

 

 

 

 

 

 

Item

Designation

 

 

 

 

 

 

F1

Micro-filter (AIR)

 

 

 

 

 

 

F2

Filter (O2)

 

 

 

 

 

 

F4

Filter (inspiratory flow sensor)

 

 

 

 

 

 

F5

Filter (Filter element)

 

 

 

 

 

 

Volume

Mixing chamber

 

 

 

 

 

 

SD1

Suction sound insulator

 

 

 

 

 

 

SD2

Sound insulator

 

 

 

 

 

 

DR1

Supply pressure regulator (Oxygen)

 

 

 

 

 

 

Gebläse

Blower

 

 

 

 

 

 

Kühler

Cooler

 

 

 

 

 

 

V1

Reduction valve

 

 

 

 

 

 

V3

Expiratory valve

 

 

 

 

 

 

 

 

<![if ! IE]>

<![endif]>No.1309_0000001227_Publication

20

Savina 300

<![if ! IE]>

<![endif]>No.1309_0000001227_Publication

Function descriptions

Pneumatic assembly

 

Item

 

 

Designation

 

 

 

 

 

 

 

V4

 

Switching valve (oxygen compensation)

 

 

 

 

 

 

 

V5

 

Switching valve (nebulizer)

 

 

 

 

 

 

 

V6.1

 

Calibration valve (inspiratory airway pressure sensor)

 

 

 

 

 

 

 

V6.2

 

Calibration valve (expiratory airway pressure sensor)

 

 

 

 

 

 

 

V7.1 - V7.8

 

Oxygen metering valves

 

 

 

 

 

 

 

V8

 

Pilot valve for emergency vent valve

 

 

 

 

 

 

 

V9

 

Emergency vent valve

 

 

 

 

 

 

 

S1

 

Inspiratory flow sensor

 

 

 

 

 

 

 

S2

 

Expiratory flow sensor

 

 

 

 

 

 

 

S3.1

 

O2 sensor 1 (measurement and control)

 

 

 

 

 

 

 

S3.2

 

O2 sensor 2 (Monitoring)

 

 

 

 

 

 

 

S4.1

 

Inspiratory airway pressure sensor (located in inspiratory branch)

 

 

 

 

 

 

 

S4.2

 

Expiratory airway pressure sensor (located in expiratory branch)

 

 

 

 

 

 

 

S5

 

Supply pressure sensor (oxygen)

 

 

 

 

 

 

 

S6

 

Pressure sensor 1 (absolute pressure)

 

 

 

 

 

 

 

S7

 

Pressure sensor 2 (absolute pressure)

 

 

 

 

 

 

 

D1

 

Safety pressure-limitting valve (passive, approx. 120 mbar)

 

 

 

 

 

 

 

D2

 

Emergency air valve (-3 mbar to -6 mbar)

 

 

 

 

 

 

 

D3

 

Expiratory non-return valve

 

 

 

 

 

 

 

D4

 

Inspiratory non-return valve

 

 

 

 

 

 

 

D5

 

Flush flow non-return valve

 

 

 

 

 

 

 

D6

 

Non-return valve (LPO)

 

 

 

 

 

 

 

R1

 

Flush flow metering unit (0,1 L/min at 30 mbar)

 

 

 

 

 

 

 

R2

 

O2 calibration metering unit for (0,1 L/min, integrated in valve block)

 

 

 

 

 

 

 

R3

 

Metering unit for O2 measurement (0,2 L/min at 30 mbar, Sensor 3.1)

 

 

 

 

 

 

 

R4

 

Metering unit for nebulizer (10 L/min, integrated in valve block)

 

 

 

 

 

 

 

R5.1 - R5.8

 

Metering units for the oxygen metering valves

 

 

 

 

 

 

 

R6

 

Metering unit for O2 measurement (0,2 L/min at 30 mbar, Sensor 3.2)

 

 

 

 

 

 

 

 

 

 

 

 

Main components

 

The pneumatic assembly consists of the following components:

 

 

 

 

– Plug-in unit

 

 

 

 

Valve block

 

 

 

 

Inspiratory block

 

 

 

 

Pressure measuring block

 

 

 

 

Patient system

 

 

 

 

Flow sensors

 

 

 

 

 

 

 

 

 

 

 

Savina 300

21

Function descriptions

Pneumatic assembly

Device functions

Ventilation function

When the device is switched on the power supply unit supplies the blower

 

motor with operating voltage. The blower motor draws in ambient air through

 

the microfilter F1, the volume and the sound insulators SD1 and SD3. The

 

blower motor compresses the gas intake to an overpressure up to max. 140

 

mbar at a delivery rate of up to 180 L/min. The air compressed by the blower

 

motor passes through the sound insulator SD4, the cooler, the sound insula-

 

tor SD2, the filter F4 and the inspiratory flow sensor S1 to the inspiratory non-

 

return valve D4. The controller operates the blower motor during a breath at a

 

constant speed. The bypass valve regulates the inspiratory pressure. The

 

combination of the blower motor and bypass valve V1 provides a pressure

 

source.

Fig. 10 Detail view of functional diagram; ventilation function

The sound insulators SD1 and SD2 on the inlet and outlet sides of the blower motor reduce the sound level.

The bypass valve V1 is operated such that the desired respiratory pressure is applied to the blower outlet, and thus to the patient. If the patient needs a high flow during the inspiratory phase, the gas flows in part or in its entirety from the blower motor outlet to the patient and the gas flow through the bypass valve V1 is reduced. During the expiratory phase all the blower gas flows through the bypass valve V1.

The cooler reduces the respiratory gas temperature down to a permissible range.

<![if ! IE]>

<![endif]>No.1309_0000001227_Publication

22

Savina 300

O2 mixture with O2 high pressure

Function descriptions

Pneumatic assembly

In order to be able to ventilate with an increased O2 concentration, the unit must be supplied with 2.7 to 6.0 bar O2. The oxygen is filtered by the filter F2. With the aid of the digital valve block consisting of 8 digital solenoid valves, oxygen is metered into the volume (mixing chamber). The amount of metered oxygen depends on the pre-set O2 concentration and on the inspiratory flow rate measured by the flow sensor S1. The oxygen is metered in a closed control loop. In the process, the inspiratory O2 concentration is measured by the O2 sensor S 3.1.

Fig. 11 Detail view of functional diagram; O2 mixture with O2 high pressure

<![if ! IE]>

<![endif]>No.1309_0000001227_Publication

Savina 300

23

Function descriptions

Pneumatic assembly

O2 mixture with O2 low pressure („LPO“ option)

NOTE

Connect only O2 low pressure sources without humidifier to the device!

An O2 low pressure source without humidifier feeds the oxygen into the “LPO” connection on the back of the unit. The filter (filter element) F5 protects the non-return valve D6 from coarse particles. The oxygen flows from the non-return valve D6 into the volume (mixing chamber). In the volume (mixing chamber) it is mixed with the drawn-in and filtered fresh air.

Fig. 12 Detail view of functional diagram; O2 mixture with O2 low pressure

When no O2 low pressure source is connected to the unit, the non-return valve D6 prevents gas from escaping during normal operation.

NOTE

In „LPO“ mode, the valve block in the O2 supply is not actuated.

<![if ! IE]>

<![endif]>No.1309_0000001227_Publication

24

Savina 300

Function descriptions

Pneumatic assembly

Pneumatic safety devices

The pneumatic safety valve D1 ensures that the ventilation pressure cannot rise above 120 mbar. In the event of inspiratory stenosis the pressure is limited by opening the expiratory valve. The mechanical vacuum valve D2 ensures (except in the case of inspiratory stenosis) that the patient can breathe ambient air in case of a fault.

The pneumatic emergency vent valve V9 relieves the pressure in the breathing system in a case of expiratory stenosis if the pressure cannot be relieved through the expiratory valve. To do so, the electric emergency vent valve (pilot valve) V8 actuates the emergency vent valve V9 accordingly.

Drug nebulizer

Fig. 13 Detail view of functional diagram; pneumatic safety devices

The drug nebulizer is operated with 100% oxygen. The pressure regulator DR1 ensures in the case of widely varying supply pressure (2.7 to 6.0 bar) that the pneumatic drug nebulizer receives a constant supply pressure of

2 bar.During the nebulizing phase the solenoid valve V5 operates in an “inspiration” (open position) and “expiration” (closed position) cycle. When the nebulizer function is inactive the nebulizer switching valve V5 is closed.

<![if ! IE]>

<![endif]>No.1309_0000001227_Publication

Fig. 14 Detail view of functional diagram; drug nebulier

Savina 300

25

Function descriptions

Pneumatic assembly

O2 calibration function

During operation, the switching valve (oxygen compensation) V4 is set to “measurement” – that is, the connection between the inspiratory side and the oxygen sensor is open. During oxygen sensor calibration oxygen passes to the oxygen sensor. This layout permits “online” calibration of the oxygen sensor S3.1 during ventilation. The oxygen sensor S3.2 must be calibrated manually (patient disconnected).

O2 sensor detection

Fig. 15 Detail view of functional diagram; O2 calibration function

The device has an oxygen sensor detector which is necessary for the „LPO“ option.

<![if ! IE]>

<![endif]>No.1309_0000001227_Publication

26

Savina 300

 

Function descriptions

 

Pneumatic assembly

 

 

Sensors

The pressure sensor (absolute pressure) S6 measures the atmospheric pres-

 

sure necessary for oxygen measurement and for volume application. The

 

pressure sensor (absolute pressure) S7 monitors the pressure sensor (abso-

 

lute pressure) S6.The oxygen sensor S3.1 generates the signal for the dis-

 

played „FiO2“ measured value and the signal to control the inspiratory

 

oxygen concentration. The oxygen sensor S3.2 monitors the oxygen sensor

 

S3.1

Fig. 16 Detail view of functional diagram; pressure sensors and oxygen sensor

The airway pressure sensor S4.1 measures the pressure in the inspiratory branch. The airway pressure sensor S4.2 measures the pressure in the expiratory branch. The output signals of the airway pressure sensors are needed

<![if ! IE]>

<![endif]>No.1309_0000001227_Publication

Savina 300

27

Function descriptions

Pneumatic assembly

to determine the airway pressure and for control and monitoring purposes. The airway pressure is measured on the basis of the measured value from the airway pressure sensor in the respective no-flow branch.

Fig. 17 Detail view of functional diagram; airway pressure sensors

The calibration valves V6.1 and V6.2 enable calibration of the inspiratory and expiratory airway pressure sensors. During calibration, the corresponding calibration valve interrupts the connection to the ventilation circuit and switches the airway pressure sensor to ambient pressure.

Fig. 18 Detail view of functional diagram; calibration valves

<![if ! IE]>

<![endif]>No.1309_0000001227_Publication

28

Savina 300

<![if ! IE]>

<![endif]>No.1309_0000001227_Publication

Function descriptions

Pneumatic assembly

Flow sensor S1 measures the inspiratory gas flow. The measurement variable is used to calculate the necessary oxygen flow and to actuate the oxygen metering valves V7.1 to V7.8 in order to control the breaths and monitor the device functions. The flow sensor includes a temperature measurement function to measure the inspiratory gas temperature.

Fig. 19 Detail view of functional diagram; flow sensor

The flow sensor S2 measures the gas flow through the expiratory valve. The flow sensor is a temperature-compensated hot-wire anemometer with no flow direction detector. With this signal the patient is monitored (e.g. the minute volume).

Fig. 20 Detail view of functional diagram; flow sensor

Expiration is executed with the directly actuated expiratory valve V3.

The expiratory valve has the following functions:

PEEP control during the expiratory phase

Close the breathing system during the inspiratory phase.

Savina 300

29

– This page has been intentionally left blank –

<![if ! IE]>

<![endif]>No.1309-0000001227_Publication

30

Savina 300

Loading...
+ 136 hidden pages